Esu BASICS 5
Esu BASICS 5
Esu BASICS 5
The cytochemical changes do in fact occur. However, these changes are reversible, and
the cells return to their normal function when the temperature returns to normal values.
Above 45°C,-irreversible changes take place that inhibit normal cell functions and lead
to cell death. First, between 45°C and 60°C, the proteins in the cell lose their
1
Medical Instrumentation I كلية االسراء الجامعة
Lec asst. Noor Talib
Lec. 11 قسم هندسة تقنيات االجهزة الطبية
quaternary oat, configuration and solidify into a glutinous substance, this process,
termed coagulation, are accompanied by tissue blanching. Further increasing the
temperature up to 100°C leads to tissue drying this process is called desiccation. If the
temperature is increased beyond 100°C, the solid contents of the tissue reduce to
carbon, a process referred to as carbonization. Tissue damage depends not only on
temperature, however, but also on the length of exposure to heat.
In the monopolar mode, the active electrode either touches the tissue directly or is
held a few millimeters above the tissue. When the electrode is held above the tissue, the
electric current bridges the air gap by creating an electric discharge arc. A visible arc
forms when the electric field strength exceeds 1 kV/mm in the gap and disappears when
the field strength drops below a certain threshold level.
Monopolar Electrosurgery
2
Medical Instrumentation I كلية االسراء الجامعة
Lec asst. Noor Talib
Lec. 11 قسم هندسة تقنيات االجهزة الطبية
Bipolar Electrosurgery
In bipolar electrosurgery, two electrodes (generally the tips of a pair forceps or scissors)
serve as the equivalent of the active and dispersive leads in the monopolar mode.
Bipolar electrosurgery does not require a patient plate. Electrosurgical current in the
patient is restricted-to a small volume of tissue in the immediate region of application of
the forceps. This affords greater control over the area to be coagulated. Damage to
sensitive tissues in close proximity to the instrument can be avoided. There is less
chance of current capacitive or directly arcing to surrounding structures such as the
bowel. Patient burns are virtually eliminated.
3
Medical Instrumentation I كلية االسراء الجامعة
Lec asst. Noor Talib
Lec. 11 قسم هندسة تقنيات االجهزة الطبية
Dispersive electrodes
Over the years electrosurgery has advanced, so too have the types and styles of
dispersive electrodes. Early on in electrosurgery the only choice was a solid pad (at first
a stainless steel plate) that was placed on the patient to disperse the heat of the RF
energy. If the solid plate was not applied correctly or began to move off the patient
during the case, the ESU would continue to deliver energy to the tissue, causing a
potentially dangerous situation.
Current Density
Current density is one of the most important concepts in electrosurgery. Simply stated,
current density is the amount of current concentration at a given point. In electrosurgery
all of the RF current is force flow through the tiny area where the active electrode
makes contact with the skin. At this point, the current flow is concentrated intensely.
4
Medical Instrumentation I كلية االسراء الجامعة
Lec asst. Noor Talib
Lec. 11 قسم هندسة تقنيات االجهزة الطبية
The heat at the site is great enough to achieve cutting and coagulation. Current leaves
the body via a dispersive electrode (grounding pad).
The pad has a large surface area thus the current density is quite low. As long as this
large, so-called "dispersive" electrode makes good contact with the skin it should offer
a passage of least resistance for safe exit of the RF current from the patient. The large
surface area generates little heat. A generator supplies RF to the active electrode.
Current passes through the patient, exiting by way of the return electrode. It returns to
the generator to complete the circuit. Without complete circular path, from generator to
patient, back to the generator, the current should not flow.
5
Medical Instrumentation I كلية االسراء الجامعة
Lec asst. Noor Talib
Lec. 11 قسم هندسة تقنيات االجهزة الطبية
ESU waveform:
6
Medical Instrumentation I كلية االسراء الجامعة
Lec asst. Noor Talib
Lec. 11 قسم هندسة تقنيات االجهزة الطبية
The first electrosurgical instrument used spark gap technique and figure below shows
the principle of work of spark gap instrument. Consist of high voltage transformer and
electric spark gap and C 1/L1 circuit. The transformer will increase the voltage from
220 v to 3000-4000v which is able to ionize the air in the gap between the two points of
tungsten. When the gap-start to spark during production of electric arc by alternating
pattern, it will produce currents which radio frequencies which start to oscillate in the
circuit (Cl/L1). This circuit is coupled "connect" with output circuit (Ll/L2) by
induction. The output energy which gone to the patient can be control its intensity
through taken different level of energy from L2 through switch S2 which is connect
with the active electrode. Depending on the type and design of the instrument the power
is between 25 watt and few hundred watts. The coils RF1 and RF2 are used for
protection and prevent the effect or return back of radio frequencies to the input power.
In some instrument the capacitor is used parallel with the secondary coil of the
transformer T1.
7
Medical Instrumentation I كلية االسراء الجامعة
Lec asst. Noor Talib
Lec. 11 قسم هندسة تقنيات االجهزة الطبية
These generators are used in cauterization and coagulation to stop the bleeding from
blood vessels. Most of these instruments exist in two mode cut and coagulate. The
rough foot selector switch. The wave for cutting is pure sinusoidal wave while the
coagulation wave is damped oscillation wave according to the type and technique of the
instrument.
The electronics circuits are used now in electrosurgical instruments so its small in size
and weight, the circuit in (figure below) is one of the types of amplifier of the power of
radio frequency.
The circuit is push-pull/parallel circuit. There are two raw or sides of 3 transistors
from Q1 -Q3 "the first", and from Q4_Q6 "the second". Each three transistors are
connected parallel, and the two raw are connected by push-pull circuit. The
transformers used in the circuit in a form of "toroid-shaped" to determine the magnetic
field in the circuit.
8
Medical Instrumentation I كلية االسراء الجامعة
Lec asst. Noor Talib
Lec. 11 قسم هندسة تقنيات االجهزة الطبية
The radio frequency signal can be produced by Oscillator as shown in figure below.
When the transistor (QI) in forward deviation state, the circuit will be oscillate with
radio frequency according the value of circuit's element. The transistor (Q2) works as
control switch to control (Q1). When the (Q2) works, it caused to (Q1) works also
"forward deviation". In the cutting mode the (Q2) continue working and the output
signal is continuous sinusoid / wave (figure below), but during coagulation mode
rectangular wave is given to the base of transistor (Q2) which is going to cut it to
produce chopped sinusoid wave (figure below). There are another circuits used another
technique according to the manufacture company.