IP344-Determination of Light Hydrocarbons in Stabilized Crude Oils-Gas Chromatography Method
IP344-Determination of Light Hydrocarbons in Stabilized Crude Oils-Gas Chromatography Method
IP344-Determination of Light Hydrocarbons in Stabilized Crude Oils-Gas Chromatography Method
for
IP 344:1988 (R2010) Determination Of Light Hydrocarbons In Stabilized Crude Oils - Gas
Chromatography Method
Licensee: Azecolab
Date: 01-Jun-2018
COPYRIGHT NOTICE
This material is electronically reproduced by SAI Global under license from the Energy Institute who retain full copyright in the document.
No part of the printed publication, nor any part of this electronic file, may be reproduced or transmitted in any form, including transmittal by e-
mail, by file transfer protocol (FTP), or by being made part of a network-accessible system, without the prior written permission of the Publisher
or SAI Global.
SAI Global makes no guarantees or warranties as to the correctness of the document or as to the results arising from the purchase and use of
the document and is not responsible for problems in the delivery of the document. Any difficulties or queries should be addressed to SAI Global
below.
SAI Global, Forest Road Office Centre, 210 Route 4 East, Paramus, NJ 07652, USA
In Europe Contact:-
In Asia/Pacific Contact:-
SAI Global - Information Services, Level 37, 680 George Street, Sydney NSW 2000, Australia
Phone: 131 242 (within Australia) | +61 2 8206 6010 (outside Australia)
Fax: 1300 65 49 49 (within Australia) | +61 2 8206 6020 (outside Australia)
E-mail: sales@saiglobal.com
Web: http://infostore.saiglobal.com/store/
To read the full licence agreement, simply click within the red box above and scroll through with your cursor
3.4 2,3-dimethylbutane, 95% pure. 5.1.1.1 Bend the column (no bend being of less
than 50 mm in diameter) so as to be suitable for
3.5 Cyclohexane, 95% pure. fitting into the chromatograph.
3.6 Heptane 5.1.1.2 Insert a small glass wool plug into one end
of the tube and feed the Porapak Q into the other
3.7 Carrier Gas, nitrogen or helium, free from end by means of a funnel. An even distribution of
water and hydrocarbons. the packing may be achieved by gently vibrating the
344.1
LIGHT HYDROCARBONS IN CRUDE OILS BY GAS CHROMATOGRAPHY, IP 344
column whilst appplying suction. Approximately 6.4 Shake the vial vigorously for at least 1 min
2,5 g of Porapak Q are required to pack the column. to completely mix the crude and internal standard.
5.1.1.3 Pack and prepare the 100 mm pre-column NOTE - Crude samples should be stored in a refrigerator
in the same manner as the main column. There will before an analysis. The preparation of internal standard
be no need to bend the pre-column. blend may be done at ambient temperature, but care
should be taken to minimize loss of light ends.
5.1.2 Column conditioning
7 Procedure
Connect the pre-column and main column together
and fit into the gas chromatograph. Do not connect 7.1 Inject 1 to 5:l as necessary, of the crude oil
the outlet end to the detector. Condition the column plus internal standard on to the column. Start the
until the base line drift resulting from column bleed recorder and integrator at the same time as
is less than 1% of full-scale deflection at a column injection.
temperature of 150°C and at the working
sensitivity. NOTE - The sample size should be as large as possible,
consistent with the linear range of the system, in order to
Licensed to Azecolab on 01-Jun-2018. 1 user personal license only. Copying, copy/pasting, storage & distribution or use on network prohibited.
5.2 Chromatograph
6 Sample preparation
344.2
LIGHT HYDROCARBONS IN CRUDE OILS BY GAS CHROMATOGRAPHY, IP 344
1 1,114
2 1,047 Ethane 0,01-0,06 0,155X 0,390X
The precision values (as defined in IP 367) have a) a reference to this standard;
been obtained by statistical examination of
inter-laboratory results and were first published in b) the type and complete identification of the
1988. product tested;
The difference between two test results, obtained d) any deviation, by agreement or otherwise,
by the same operator with the same apparatus from the procedure specified;
under constant operating conditions on identical test
material would in the long run, in the normal and e) the date of the test.
344.3
LIGHT HYDROCARBONS IN CRUDE OILS BY GAS CHROMATOGRAPHY, IP 344
Annex A
(informative)
Figure A.1: Arrangement for back-flush of Figure A.2: Arrangement for back-flush of
column and pre-column pre-column only
Get permission to copy from or network this publication www.saiglobal.com/licensing
344.4