Epa Qaqc PCR
Epa Qaqc PCR
Epa Qaqc PCR
October 2004
Office of Water (4607)
EPA 815-B-04-001
www.epa.gov/safewater
October 2004
The authors are grateful to the following people for reviewing and contributing to the document.
The Technical Support Center, Standards and Risk Management Division, of the U.S. EPA
Office of Ground Water and Drinking Water, Cincinnati, OH, and the National Exposure
Research Laboratory of the U.S. EPA Office of Research and Development, Cincinnati, OH,
have prepared this guidance manual. Support for preparation of the manual was provided by the
CSC Biology Studies Group under contract number GS-10F-0135K. The manual has been
subjected to the Agency’s peer and administrative review and it has been approved for
publication as an EPA document.
This manual is not a regulation; EPA offers it as guidance for laboratories developing polymerase
chain reaction (PCR) based-analyses on contaminants in environmental samples and for decision
makers who need to judge the quality of PCR data.
The mention of trade names or commercial products in this manual does not constitute
endorsement or recommendation for use.
Keya Sen
U.S. EPA Office of Ground Water and Drinking Water
Technical Support Center
26 West Martin Luther King Drive
Cincinnati, OH 45268-1320
sen.keya@epa.gov
(513) 569-7026
(513) 569-7191 (facsimile)
G. Shay Fout
U.S. EPA Office of Research and Development
National Exposure Research Laboratory
26 West Martin Luther King Drive
Cincinnati, OH 45268-1320
fout.shay@epa.gov
(513) 569-7387
(513) 569-7117 (facsimile)
TABLE OF CONTENTS
Section 1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Section 7. Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
From the isolation of specific genes to the sequencing of entire genomes, the polymerase chain reaction
(PCR) has become one of the most widely used technologies for conducting biological research.
Advances have led to the development of specific and sensitive high-throughput PCR methods for the
detection of a variety of microorganisms, and these methods are increasingly being applied to analysis of
environmental samples.
The successful application of PCR requires the proper use of techniques and interpretation of results.
Many PCR methods offer a level of sensitivity equal to, or greater than, more traditional environmental
microbiological methods. Due to the ability to amplify small amounts of nucleic acid, PCR can be used to
detect organisms that are difficult to culture in vitro or that cannot be cultured. However, the advantages
of these techniques can be offset by the demanding assay protocols and the need to follow quality
assurance/quality control (QA/QC) procedures carefully. These QA/QC procedures are necessary
because the ability of PCR to produce many copies of target DNA creates the possibility of
contamination by previously amplified products, which can lead to false-positive results. In addition,
environmental samples may inhibit the PCR, which can lead to false-negative results. As efforts are made
to standardize PCR protocols for analyses of environmental samples, it is essential to establish
standardized QA/QC procedures.
1.1 Purpose
This guidance manual has been developed to serve as a resource and a reference for the following:
• Laboratory QA/QC practices for environmental sample analyses involving PCR techniques
• Laboratory QA/QC practices for new U.S. Environmental Protection Agency (EPA) PCR method
protocols
• Laboratory QA/QC practices for EPA grants and assistance agreements involving PCR analyses
• A basis for researchers, managers, and quality assurance officers to evaluate the quality of PCR data
of research projects and technical papers
1.2 Scope
This manual is intended to serve as general guidance for the development of laboratory and method-
specific QA/QC procedures for PCR analysis of environmental samples. However, because PCR analysis
includes a broad range of nucleic acid targets and procedures, all QA/QC procedures necessary for a
particular protocol may not be addressed.
This document does not address federal, state, and local regulations governing waste management,
hazardous materials, and radioactive material; it is the laboratory's responsibility to comply with relevant
regulations. Furthermore, this guidance does not address related safety issues; it is the laboratory’s
responsibility to establish appropriate safety and health practices.
1 October 2004
Section 1: Introduction
• Section 2. Laboratory Quality Assurance. This section provides guidance on general laboratory
QA/QC applicable to PCR analysis of environmental samples. It provides recommendations for
personnel, facility design, workflow, equipment, disposables, and cleaning.
• Section 3. Reagents, Kits, Primer Sets, and Enzymes. This section provides QA/QC guidance for
reagents, kits, primer sets, and enzymes used in PCR analysis of environmental samples, including
the information for these materials that should be recorded in laboratory logbooks and storage
conditions.
• Section 4. Method Development and Assessment. This section provides guidance on the
development of a PCR method and its component parts, including factors to consider when selecting
appropriate PCR techniques and confirmation of amplicons.
• Section 5. Quality Control Samples for Methods Using PCR. This section provides a summary of
the controls necessary for PCR analysis of environmental samples. Recommended corrective actions
for control failures and proficiency testing are also covered.
• Section 6. Data Recording, Record Keeping, and Data Evaluation. This section provides
guidance on data recording, record keeping, and data interpretation, including variability, equivocal
results, and limitations.
2 October 2004
SECTION 2. LABORATORY QUALITY ASSURANCE
Section 2 provides guidance on the recommended qualifications and training for personnel who perform
PCR analyses and on the use of protective wear by laboratory personnel to prevent sample
contamination. Facility design and workflow are considered, as they are essential for preventing both
sample contamination and erroneous results. In addition, calibration, maintenance, and cleaning
procedures for laboratory equipment and the standards for disposable labware are discussed.
2.1 Personnel
Personnel working in the laboratory performing PCR analysis should meet background and training
specifications outlined below and should follow the guidance provided concerning protective outerwear.
Although the amount of time required for training will vary depending on the analyst and the technique,
each analyst should demonstrate that they can successfully perform the method through analyses of
positive and negative control samples before being allowed to analyze environmental samples without
supervision. An initial demonstration of capability should include at least four replicate analyses of
seeded reagent water, two replicates of seeded environmental water, and a method blank (see Section
5.2.2), as well as proficiency testing (see Section 5.6). These recommendations for the initial
demonstration of capability are based on current requirements for EPA’s conventional microbiological
methods. Each analyst also is expected to be knowledgeable in laboratory safety and QA/QC procedures.
The laboratory should maintain a training record for each analyst that documents the following:
3 October 2004
Section 2: Overall Laboratory Quality Assurance for Environmental PCR Analyses
2.1.2 Outerwear
Dedicated laboratory coats and powder-free gloves should be available in each laboratory room (see
description of laboratory rooms in Section 2.2.1). Laboratory coats should be removed and gloves
discarded before leaving each room. Changing laboratory coats and gloves reduces the possibility of
contamination with template (the nucleic acid from which the PCR is performed) or amplified nucleic
acid. Gloves should be changed after working with seeded or environmental samples, after handling
template or amplified nucleic acids, and after contact of the outside of the gloves with skin. The latter
prevents introduction of enzymes prevalent on the skin, such as DNases and RNases, that degrade nucleic
acids. In the absence of a laminar flow hood or biological safety cabinet (see Section 2.4.7), protective
dust masks or surgical masks may be worn to further reduce the risk of contamination of airborne nucleic
acids from the analyst.
Laboratory coats should be cleaned regularly to reduce the possibility of contamination of the designated
workspace and the PCR reaction. Laboratory coats should be separated from non-laboratory clothing
(i.e., laboratory coats should not to be taken home and washed with other clothes), and cleaned only with
other laboratory coats that were in the same work area (e.g., laboratory coats from the sample preparation
room should not be cleaned with laboratory coats from the amplification and product room). Many
regions have specialized companies that are qualified to clean laboratory coats. These companies often
clean coats in lots to prevent the mixing of coats from a designated laboratory area with other coats or
materials from another area. The frequency of cleaning is dependant on the amount of PCR work the
laboratory is performing, and busy laboratories may want to clean coats weekly or daily. Laboratory
coats used to prepare positive controls should be dedicated to this task and cleaned more frequently. To
eliminate the need for cleaning laboratory coats, disposable (single-use) laboratory coats also may be
used.
4 October 2004
Section 2: Overall Laboratory Quality Assurance for Environmental PCR Analyses
A unidirectional workflow will reduce the opportunity for contamination to occur (see Section 2.2.2 and
Figure 2-1). No materials, supplies, or equipment from the sample preparation room should be taken into
the reagent preparation room. Nothing from the amplification and product detection room should be
taken into the sample preparation room or the reagent preparation room. To reduce the threat of sample
contamination from sample processing activities, separate biological safety cabinets (see Section 2.4.7)
should be dedicated for positive control and test sample processing. Vacant areas in the rooms may be
used for non-PCR activities, such as analysis of chemical samples, but the personnel associated with
these activities should maintain the unidirectional workflow, and the equipment should not be moved
between the rooms used for any PCR sample processing and analysis steps.
* The sample preparation room should be divided into two physically separated areas (hoods) or rooms, one for
sample and negative control preparation and the other for positive control preparation.
5 October 2004
Section 2: Overall Laboratory Quality Assurance for Environmental PCR Analyses
The reagent preparation room should have dedicated, adjustable pipettes with plugged, aerosol-barrier or
positive-displacement pipette tips, laboratory coats, and disposable gloves. Fresh gloves and laboratory
coats should be worn at all times to control contamination. Personnel should perform tasks in this room
before working in the sample processing or amplification/detection rooms and should not move from
these rooms back to the reagent preparation room.
Important: Materials from other rooms (including amplified, template, and target nucleic acid or
positive controls, supplies, or equipment) should not be brought into the reagent preparation room.
The sample preparation room should have dedicated adjustable pipettes with plugged, aerosol-barrier tips
or positive-displacement tips. Fresh gloves and laboratory coats should be worn at all times to control
contamination from this room to any other location. This room should be kept under negative pressure to
prevent contamination outside of the room.
Ideally, two biological safety cabinets (see Section 2.4.7) should be used within the room—one for
sample and negative control preparation, and the other for positive control preparation—to protect the
samples from cross-contamination and to protect workers from exposure to pathogens. Separate pipettes
and laboratory coats also should be designated for work in each hood.
Important: Nothing from this room should be taken to the reagent preparation room.
In addition to separating positive control preparation from sample and negative control preparation,
laboratories may want to include a designated enclosed area or room for sample receiving and storage.
This room would contain refrigerators for sample storage and should be isolated from any of the other
areas. Personnel should not move from this room to the reagent preparation room.
6 October 2004
Section 2: Overall Laboratory Quality Assurance for Environmental PCR Analyses
and removed before leaving the room to control amplicon contamination of other locations. All
equipment used for amplification and product detection should be dedicated to this room, including
adjustable pipettes with plugged, aerosol-barrier or positive displacement pipette tips. This room should
be kept under negative pressure. Although PCR amplification and post-PCR analyses may be performed
in the same room, laboratories may want to separate these activities in different areas or different rooms
to reduce the risk of contamination from amplified products.
Important: Nothing from this room should be taken into either the sample preparation room or the
reagent preparation room.
2.2.2 Workflow
A unidirectional workflow should be used to reduce the potential for contamination. Color-coding of
equipment, reagents, laboratory coats, and supplies can help maintain the unidirectional workflow by
designating colors specific to each laboratory room. Laboratory manuals and notebooks also should not
be moved from room to room. Laboratories may want to consider using electronic data reporting to avoid
this potential source of contamination.
On a given day, analysts should not return to the reagent or sample preparation rooms after working in
the amplification and product room. Similarly, analysts should not return to the reagent preparation room
after working in the sample preparation room. If an analyst must unexpectedly return to an upstream
room in the workflow process, the analyst should first shower and change clothes. If a breach in this
workflow process does occur, the affected laboratory rooms and equipment should be cleaned according
the guidance provided in Sections 2.4 and 2.6, and the results should be monitored as defined in Sections
5.2 and 5.4.
2.4 Equipment
The equipment used to perform a PCR method should function properly to generate reliable data. To
verify that equipment is functioning properly, the laboratory should have a schedule for maintaining
equipment. The schedule should include the setup, calibration, repair, record keeping, and normal
operation of all equipment used in sample analysis, as stated in the SOP for each individual instrument or
7 October 2004
Section 2: Overall Laboratory Quality Assurance for Environmental PCR Analyses
method. The results of all tests should be documented in an equipment logbook and/or electronic
database. The logbook or database should be checked monthly by QC personnel or the laboratory
supervisor, and any problems and corrective actions noted. Equipment should be dedicated to a specific
laboratory room, and the instrument manuals from the manufacturer should be available. Individual
laboratories should decide which tests to perform to assess the functionality of the instruments and the
frequency at which to test them. Recommended QA for specific types of instruments used for PCR
analyses are provided in Sections 2.4.1 through 2.4.11.
Block temperature should be tested with an external probe that has been calibrated against a temperature
standard. For testing, the probe is placed in several of the wells in the periphery and center of the
instrument. All temperatures should be within the manufacturers’ specifications. The amplification
program used in each run should be printed to further verify the conditions of the PCR. This also allows
the analyst to determine if there were any instrument malfunctions during the reaction. If the
thermocycler does not have software with this capability, then the cycling program used should be
documented in the laboratory’s SOP with results of monthly calibration checks.
Real-time PCR instruments are equipped to perform fluorescence excitation and detection to monitor
amplification throughout the PCR cycles. For this reason, the design is often different from the standard
thermocycler, and calibration may be specific to the instrument design and should be carried out
according to manufacturers’ specifications. For example, several real-time platforms do not have heating
blocks but rotor designs, with samples spinning in an air chamber that is heated and cooled. Temperature
can be calibrated for this type of instrument using a calibration rotor provided by the manufacturer.
Laser performance, alignment, and safety devices should also be checked and optical systems calibrated.
This may involve analyzing fluorescent test trays to confirm consistent detection from each well. For
rotor designs, each tube is passed by the same detector, providing greater sample to sample consistency.
Spectral calibration files may also need to be recalibrated to confirm that the excitation signals from the
range of fluorophores used can be separated, and gain settings adjusted. Analysts can perform part of this
calibration, but service technicians trained by the equipment supplier will be required to carry out a full
maintenance and calibration service. A full service should be preformed at least annually.
2.4.2 Centrifuges
Separate centrifuges, including microfuges, should be used for pre- and post-PCR procedures.
Centrifuges should have speed capabilities that meet the criteria required at different stages in the overall
method. The manufacturers’ instructions for calibration should be followed. The centrifuge should be
balanced before use to increase bearing life and minimize vibrations that can unsettle concentrates.
8 October 2004
Section 2: Overall Laboratory Quality Assurance for Environmental PCR Analyses
Care should be taken with the hybridization bottles. These bottles should be inspected periodically for
chips and cracks and for deterioration of the O-ring. Bottles that are chipped or cracked should be
discarded and replaced. Deteriorated or damaged O-rings should be replaced immediately. Bottles and O-
rings that contact radioactive material should be thoroughly cleaned of any radiation before they are
discarded. These bottles also should be cleaned after each use in accordance with manufacturers’
specifications and in accordance with radiation removal protocols (if applicable).
2.4.6 Sequencers
Sequencers should be calibrated according to the manufacturers’ recommendations, and all calibrations
should be documented. For sequencers using capillaries, it may be necessary to perform both spatial and
spectral calibrations using standards provided by the manufacturer. Calibrations should be performed at
the frequency specified by the manufacturer and according to the manufacturers’ instructions. If the
sequencer uses sample blocks, the blocks should be cleaned regularly and checked for fluorescence
contamination. Instrument performance should be checked regularly using standards provided by the
manufacturer. Preventative maintenance should be performed annually by a qualified technician.
9 October 2004
Section 2: Overall Laboratory Quality Assurance for Environmental PCR Analyses
Class II or III biological safety Intake and exhaust All aspects of PCR
Yes
cabinet filtered preparation
Although several types of chambers are available, many may not provide satisfactory contaminant
removal for the preparation of reagents and samples for PCR. Laminar-flow hoods (also listed by
suppliers as “enclosures” or “chambers”) provide air that is cleaner than the laboratory air and reduce the
chance of airborne contamination. These hoods provide protection from outside contamination to
material inside the hood, but do not protect the worker or the environment unless they have a filtered
exhaust. Biological safety cabinets (BSCs) have a classification system that ranges from Class I to Class
III (33, 47). Class I cabinets have inward air flow and HEPA-filtered exhaust that provides personal and
environmental protection, but no product protection. Class II and III BSCs filter both air intake and
exhaust, and prevent contaminants from entering and leaving the hood (reducing the likelihood of sample
and work area contamination). Laboratories purchasing Class III BSCs should be aware that they require
a larger air intake than Class II cabinets. This can produce problems with air balance if the hood is vented
to an outside environment.
Hoods without an air-flow system or those with filtered air intakes may be used for reagent preparation,
if they are equipped with UV lights. For all other purposes, laboratories should use Class II or III BSCs
with UV lights. These types of hoods and cabinets provide efficient laminar air flow to keep
contaminants from entering the work area or the laboratory. Other chambers that lack these qualities
increase the likelihood of sample contamination. An alternative to the recommended BSCs is a Fed 209
Class 100 clean room with an appropriate ISO-class HEPA-filtered air intake and UV light.
Before use, hoods should be decontaminated using UV light for at least 8 hours (14, 34) and cleaned with
bleach or other effective nucleic acid inactivating agent. After a spill or accident (see Section 2.6 for
guidance on cleaning), hoods also should be cleaned with the nucleic acid inactivating agent and then
decontaminated further using UV light. The airflow and HEPA filtration in all hoods should be
monitored and certified per manufacturers’ recommendations on at least annually.
2.4.9 Pipettes
Due to the numerous small volume transfers involved in PCR methods, automatic pipettes, fixed-volume,
adjustable, positive-displacement, and/or micropipettes should be used, and should be calibrated
10 October 2004
Section 2: Overall Laboratory Quality Assurance for Environmental PCR Analyses
quarterly by the manufacturer or a technician using the tips commonly used in the laboratory.
Documentation on the precision of the recalibrated micropipette should be retained. Calibration records
should be kept on file in the laboratory’s QC logbook. Each pipette should be sterilized according to
manufacturers’ recommendation on a regular basis or whenever contamination is suspected. Calibration
of the pipette should be performed after sterilization.
If a micropipette calibration problem is suspected, the laboratory should tare an empty weigh boat using
an analytical balance and pipette the following volumes of reagent water into the weigh boat using the
pipette in question: 100% of the maximum dispensing capacity of the micropipette, 50% of the capacity,
and 10% of the capacity (do not go beyond the recommended range of the pipette). Ten replicates should
be performed at each weight. The weight of the water (assume that 1.00 mL of reagent water weighs 1.00
g) should be recorded and the relative standard deviation (RSD) calculated for each. The balance used
must have sufficient accuracy in the range that will be used for calibration (e.g., for 1-50 µl volumes the
balance should be able to accurately weigh 0.0001g). When measuring ultra-low volumes, a moisture
source or other device should be used to prevent evaporation.
If the weight of the reagent water is within the manufacturers’ recommended limits using the tips
commonly used in the laboratory, then the pipette remains acceptable for use. If the weight of the reagent
water is outside the acceptable limits, the instruction manual for the pipette should be consulted to
determine appropriate adjustments/actions that can be taken and the calibration checked again. If
problems with the pipette persist, the laboratory should send the pipette to the manufacturer for
recalibration. Spectrophotometric calibration of pipettes also may be performed (1). Kits for this type of
calibration are available from commercial vendors.
For equipment used in PCR analysis, the following temperature ranges should be applied:
• Incubators, water baths, and heating blocks: ± 0.5 °C of the temperature required by the protocol
• Refrigerators: 1°C to 5°C
• Standard laboratory freezers: -20°C ± 5°C
• Ultra-low freezers: -70°C ± 10°C
Temperatures of equipment should be monitored and recorded at least once a day for each workday in
use. Alternatively, temperatures may be monitored continuously using a computer-based alarm system.
Important: “Frost-free” freezers should not be used. Temperature fluctuations lead to reagent
degradation.
11 October 2004
Section 2: Overall Laboratory Quality Assurance for Environmental PCR Analyses
Separate refrigerators and freezers for samples, reagents, and final amplification products should be
maintained in the appropriately designated laboratory room. Amplified product should always be kept in
a separate freezer from reagents (such as master mix), samples, and sample concentrates.
2.5 Disposables
Disposable materials used in PCR analysis include pipette tips, sample tubes, PCR tubes, and gloves. To
reduce the contamination and degradation of the target nucleic acids, disposable materials should meet
the standards discussed in Sections 2.5.1 to 2.5.3.
2.5.3 Gloves
Disposable gloves should be available in each section of laboratories used for PCR analysis. Gloves
should be changed before leaving and entering each section of the laboratory and each time
contaminating DNA is potentially encountered. In addition to reducing potential contamination from
samples, wearing gloves may protect the technician from potential chemical exposure and prevent sample
contamination due to human DNases and RNase.
12 October 2004
Section 2: Overall Laboratory Quality Assurance for Environmental PCR Analyses
Thermocyclers and centrifuges should be cleaned with the diluted bleach solution (see above) whenever
contamination is suspected. Pipettes should be cleaned according to manufacturers’ instructions. Racks
and trays should be soaked in the 0.6% NaOCl solution and thoroughly rinsed with water after each use.
Gel-trays, gel combs, and glassware used for blot hybridization should be rinsed with water or a mild
detergent after each use.
13 October 2004
SECTION 3. REAGENTS, KITS, PRIMER SETS, AND ENZYMES
This section addresses the general QA procedures that should be followed when working with reagents,
kits, primer sets, and enzymes used in PCR analysis. Guidance on the record keeping for reagents, kits,
primer sets, and enzymes is provided in Section 6.
3.1 Reagents
The reagents used in PCR amplification can be purchased or prepared in-house. Care should be taken to
ensure the reagents are maintained contamination-free. All reagents should be clearly labeled with name,
expiration date, and relevant safety information. Reagents from different lot numbers should not be
interchanged without prior functional validation (see Section 3.1.1 through 3.4).
Molecular-grade water or its equivalent from commercial sources should be used for all assays.
Laboratories also may use water purification systems that produce high-quality pyrogen and
DNase/RNase-free water. If a laboratory uses a water purification system, it should be installed in a
nucleic acid-free area (such as the reagent preparation area) to reduce the possibility of contamination.
Diethylpyrocarbonate (DEPC) treatment can be used to eliminate RNase from water used in RNA
analysis. Addition of DEPC results in the covalent modification of nucleases (such as RNase), causing
them to lose their function. Reagent water is treated with a solution of 0.1% DEPC for several hours and
then autoclaved to degrade the DEPC completely. Proper autoclaving is necessary, because trace amounts
of DEPC in a solution will lead to the modification of the purine residues in RNA by
carboxymethylation. This leads to downstream effects in RNA experimentation (e.g., removing the
ability of reverse transcriptase to bind RNA and synthesize DNA from an RNA template) (3).
14 October 2004
Section 3: Reagents, Kits, Primer Sets, and Enzymes
the kits should be evaluated before use for the analysis of field samples by running an appropriate
positive control (See section 5.1).
Purity may be assessed by HPLC or by separation on an acrylamide gel of appropriate concentration (24).
Every lot of new oligos should be checked for contamination by being used in a PCR negative control
(Section 5.2.1). No positive results should be found. Primers and probes should be added to the PCR
master mix in the reagent preparation area.
Functional validation also should be performed on new lots of primers and probes by comparing their
performance against older sets of known quality (31). The new oligos should be rejected or the
concentration adjusted, if performance is significantly different from the proven lot. When real-time PCR
is used, the validation of new sets can be done by comparing the PCR efficiencies of old and new sets.
The PCR efficiency (E) is calculated from the slope of a standard curve (S) using the formula, E=10-1/S -1
(36). The samples used for generating standard curves should be carefully prepared dilutions of a stock
solution containing a known number of purified PCR template nucleic acids or total nucleic acids
extracted from the target organism. A PCR efficiency of 1 (or 100%) is obtained when each target
sequence present in the PCR reaction doubles during each round of amplification. A high efficiency is
needed to obtain accurate and reproducible results. A low efficiency indicates that primers and/or probes
are performing poorly. No single efficiency (as a percentage) has been set as a limit for data to be
acceptable, although 80% or greater is an accepted norm.
3.3.2 Storage
Most oligos and DNA templates should be stored at -20°C or -70°C in either TE buffer (10 mM Tris-HCl
and 0.1mM EDTA, pH 8.0) or molecular grade water. PCR products may also be stored at -20°C or -
70°C. RNA templates should be aliquoted and stored at -70°C. TE buffer generally is the preferable
storage buffer for oligos and DNA templates, because it may prevent DNA degradation, however,
molecular-grade water may be more suitable for certain purposes, such as taking spectrophotometric
readings. The pure, concentrated oligos should be stored in the original tube from the manufacturer and
labeled with the primer name and concentration. To minimize the chance of contamination and
degradation, these concentrated stocks should not be used on a regular basis. Diluted working stocks
should be made for each oligo, and these working stocks should be used for all experiments. Before use,
oligos should be thawed and mixed completely. Laboratories should either set expiration dates of one
year for primers and probes, or check their sensitivity when they are a year old and then on a regular
basis to see if any degradation has occurred. The sensitivity of primers can be checked by running a PCR
15 October 2004
Section 3: Reagents, Kits, Primer Sets, and Enzymes
positive control using the old and new primers as described in Section 5.1.1. The sensitivity of probes
can be checked by comparing old and new probes using a blotting technique as described in Section
4.4.2. Alternatively, the purity of the old primers can be checked using HPLC or polyacrylimide gel
electrophoresis (PAGE). If degradation has occurred, new primers and probes should be prepared.
3.4 Enzymes
Because the enzyme is the true machinery of all PCR procedures, enzymes should be purchased from a
commercial source to ensure purity. Laboratories should select vendors that provide QA information with
the enzymes.
3.4.1 Quality
The enzyme manufacturer should provide documentation on the quality control that has been performed
to ensure enzyme purity. Information about the minimum activity characteristics and the conditions under
which those characteristics are tested should be provided. This documentation should be stored in the
enzyme logbook in addition to all other pertinent information.
After receipt from commercial sources, each new lot of enzyme should be compared with old lots using
known controls and environmental samples (31). The new enzymes should be rejected or the
concentration adjusted, if performance is significantly below that of the proven lot.
Important: The analyst should have only one container open at a time when working with enzymes.
Pipette tips should be discarded after each dispense to prevent cross-contamination.
3.4.2 Storage
The manufacturers’ instructions on enzyme storage and use should be followed carefully. Enzymes
typically are stored at -20°C, and should never be left at room temperature in the laboratory. Insulated
bench-top coolers or ice can be used to keep the enzyme cold in the laboratory, when used on the bench
top.
Important: “Frost-free” freezers should not be used. Temperature fluctuations lead to reagent
degradation.
16 October 2004
SECTION 4. METHOD DEVELOPMENT AND ASSESSMENT
When developing a new PCR method, laboratories should design and select the individual components of
the method, including nucleic acid isolation and PCR analysis, to optimize the method’s ability to
recover and detect the target analyte. The laboratory also should consider how samples will be collected
and processed for the method, and assess the performance of the individual method components and the
entire analytical process before using the method for environmental sample analyses.
The following guidance on method development and assessment is provided in this section:
• The major components that should be considered when developing a PCR method for environmental
monitoring (Section 4.1 through 4.4)
• The determination of method sensitivity and precision (Section 4.5)
• The factors that should be considered for validating a method after it has been developed and
assessed internally (Section 4.6)
The sample processing component of an environmental PCR method is performed to isolate the organism
or nucleic acid of interest from the environmental matrix, while minimizing co-purification of potential
contaminants and components of the matrix that may inhibit PCR. The isolation procedure also should
stabilize the target nucleic acid from nucleases and reduce the sample to a volume small enough to be
analyzed by PCR. Positive controls that demonstrate that the target organism or nucleic acid has been
successfully isolated and that the PCR inhibitors have been removed are discussed in Section 5.
The efficiency of nucleic acid isolation varies with sample type and extraction procedure. Ideally, an
isolation technique should meet the following goals:
17 October 2004
Section 4: Method Development and Assessment
18 October 2004
Section 4: Method Development and Assessment
Conventional DNA Amplification and • Easiest of the PCR • Normally produces only
detection of DNA types to perform qualitative results
sequences • Low cost of equipment • Requirement for post-
and supplies amplification analyses
increases time and
labor as well as risk of
cross-contamination
and human error
• Products should be
confirmed by probe
hybridization or
sequencing
19 October 2004
Section 4: Method Development and Assessment
One repetition or thermal cycle of these three steps theoretically doubles the amount of DNA present in
the reaction. The number of repetitions needed for a PCR application is determined by the amount of
DNA present at the start of the reaction and the number of amplicon copies desired for post-PCR
applications. Typically 25 to 40 cycles are performed. (10)
A schematic representation of the events that occur at different temperatures is shown in Figure 4-1.
Conventional PCR normally is not a quantitative assay. However, it is possible to generate quantitative
results by using most probable number (MPN) or competitive PCR approaches. The MPN approach is
performed by serially diluting a sample to its detection limit (see Section 4.5). Five or more replicates are
run from each dilution, and the MPN value is determined using an MPN calculator (a free calculator is
available at www.epa.gov/microbes). With the competitive PCR approach, an internal standard is made by
cloning a modified version of the amplicon. The modified version usually contains a small deletion or
insertion, which produces a different length product, and often contains an alternative hybridization probe
recognition sequence that allows the product to be differentiated from the normal amplicon by probe
hybridization. After preparation, the modified clone is purified, quantified by spectrophotometry, and
serially diluted. Replicate samples containing the target nucleic acids to be quantified are mixed with
different concentrations of the clone, and the mixes are amplified by PCR. The products are then run on a
gel and visualized by staining. The concentration of the target nucleic acid from the sample is assumed to
be equal to the number of copies of the clone in the reaction that produces equal amounts of the clone and
the target. In this method, it is important to ensure that the clone and target sequences are amplified at the
same rate. This can be facilitated by minimizing the sequence and size differences between the target
amplicon and the clone.
Competitive RT-PCR can also be performed, but in this case it is necessary to first generate and quantitate
an RNA transcript of the clone. Accurate quantification of the transcript can be difficult because of RNA
degradation and because of the presence of the clone used to generate the transcript even after DNase
treatment. Transcripts of variable lengths can also be generated, making it more difficult to correlate
spectrophotometric reading with the true copy number. The results of competitive RT-PCR must therefore
be interpreted with caution.
20 October 2004
Section 4: Method Development and Assessment
*Annealing and elongation can be combined into one step at about 60°C for two-step PCR processes.
Real-time PCR also differs from conventional PCR in that the target selection for real-time PCR is more
restricted due to requirements of smaller target fragment and the need to select probes with a higher
melting temperature than the primers to ensure that the probe is fully hybridized during primer extension.
In addition, the annealing and elongation temperatures are usually combined in a two-step PCR process
that is performed at an intermediate temperature (e.g., 60°C) for one to two minutes.
21 October 2004
Section 4: Method Development and Assessment
There are several different fluorescent detection chemistries used for real-time PCR, including the
following:
• Dual-labeled fluorogenic oligonucleotide probes are most frequently used. These probes (e.g.,
TaqMan® probes) have a reporter fluorescent dye at the 5' end and a quencher dye at the 3' end. The
probes are added to the PCR master mix along with the PCR primers. During the PCR, if the target
sequence is present, the probe anneals downstream from a primer site and is cleaved by the 5' nuclease
activity of Taq DNA polymerase during polymerization (PCR product formation). This cleavage
releases the reporter dye from the probe and away from the quencher dye, resulting in fluorescence
that is detected by the instrument. Both the polymerization and 5' nuclease activities of Taq
polymerase are used in this process.
• Fluorescent resonance energy transfer (FRET) probes involve the hybridization of two probes to
adjacent sequences within the amplified product. The upstream probe has a fluorescent dye at the 3’
end and the adjacent probe has a fluorescent dye at the 5’ end. Correct hybridization of these probes
brings the two dyes into close proximity. The laser excites the first fluorescent dye, which emits light
at a different wavelength. This light then excites the second fluorescent dye by FRET between the
adjacent probes. The real-time PCR machine detects the wavelength of light emitted by the second
fluorescent dye.
• Molecular beacon probes use a variation of this same process, wherein reporter and quencher dyes
are held together by a hairpin structure in the probes but become sufficiently separated by
linearization of the probe after annealing with the template to allow the reporter dye fluorescence to
be detected.
• SYBR® Green I, a fluorescent dye, is another frequently used real-time detection chemistry. When
this dye intercalates into double-stranded DNA, including PCR products, its fluorescence increases
greatly, and this increase can be detected by the real-time PCR thermocyclers. Because this detection
chemistry is not target sequence specific, it is more versatile than probe-based detection, but is
susceptible to false positives due to the formation of nonspecific PCR products or primer-dimers.
Melting curve analyses (Section 4.4.5) are often used as an additional product confirmation for
procedures using SYBR ® Green.
While multiplex PCR provides a potential time savings by allowing simultaneous detection of multiple
targets, significant optimization is required to obtain all of the products with equal efficiency and
sensitivity (see Section 4.5). Extra precaution must be taken to design primers that do not have adverse
primer interactions (see Section 4.3.3). The best concentration ratios of different primer sets need to be
determined carefully. Even with extra precaution and optimization (16, 26), the amplification of non-
specific targets and reduced sensitivity for some primer sets can be problematic (24, 28). Optimization of
PCR is discussed in Section 4.3.4.
22 October 2004
Section 4: Method Development and Assessment
(an oligos consisting of only thymidine residues), random hexamers (primers made of six random
nucleotides), or a specific primer. Oligos dT will hybridize to the poly A tail of messenger and certain
viral RNAs and prime DNA from the 3'-end of the RNA molecule. This approach may not be adequate for
amplification of RNA segments near the 5'-end of the molecule. Random hexamers work with any RNA,
but require an extra initial incubation at 25°C. Specific primers can be either the PCR primer that
hybridizes to the RNA at the 3' side of the amplification region or a primer that hybridizes further
downstream from the PCR primers. RNase inhibitor should be added to RT reactions to prevent the
degradation of the RNA target sequence by RNase present in the sample or introduced as contamination.
The reverse transcription and the PCR amplification can be performed in a one- or two-step process. In
general, the two-step process is more sensitive, while the single-step reactions are less likely to be
contaminated, because the tube is not opened after reverse transcription. The determination of which
process should be used depends on the level of sensitivity required and the likelihood of contamination.
There are many types of reverse transcriptases available for RT-PCR. The characteristics of the enzymes
make some better suited for a one- or two-step reactions and other downstream applications. Some
enzyme characteristics that impact the type of reverse transcriptase used for RT-PCR include: the presence
or absence of RNase H activity that degrades RNA in an RNA:cDNA hybrid, processivity of the enzyme,
divalent ion requirements, specificity and sensitivity, ability to incorporate dUTP for UNG carryover
contamination (see Section 5.5.1), and optimum temperature for function. (7)
Nested PCR is a conventional PCR with a second round of amplification using a different set of primers.
This second set of primers is specific to a sequence found within the DNA of the initial conventional PCR
amplicon. The use of a second amplification step with the "nested" primer set results in a reduced
background from products amplified during the initial PCR due to the nested primers’ additional
specificity to the region. The amount of amplicon produced is increased as a result of the second round of
amplification and due to a reduction in any inhibitor concentrations. The presence of the expected second
PCR amplicon is usually viewed as confirmation of the presence of the target organism in a sample.
Used correctly, the multiple rounds of nested PCR should increase both the sensitivity and specificity of
the PCR. However, there is an increased chance of carryover or cross-contamination when taking product
from the first round of PCR and putting it in the tubes for the second round of PCR. Additional steps and
precautions may need to be taken to reduce the chance of sample contamination and false-positives,
including the following:
• Never opening more than one tube at a time
• Using a separate thermocycler for the first and second amplifications
• Adding additional negative controls beyond those recommended in Section 5.2.1. The additional
controls should be interspersed with tubes containing samples
• Including first round negative controls in the second round of amplification to check for false-
positives
• Designating a fourth room or separate area for sample preparation after the first amplification
23 October 2004
Section 4: Method Development and Assessment
When selecting an enzyme type for a method or study, the analyst should evaluate the different strengths
and weaknesses of the DNA polymerases available to determine which individual polymerase, or
combination of polymerases, will work with their template nucleic acid. Before switching manufacturer,
enzyme type, or lot number, the new enzyme’s effectiveness should be tested against the enzyme
originally used to validate the PCR using both QC and environmental samples and the results documented.
A variety of computer programs are available to aid in the creation of the best possible primers and
probes. These programs also can help determine the optimum annealing temperature for newly created
oligos and check for the formation of intra- and intermolecular dimers and hairpin loops. Laboratories
should consider repeating the design process with more than one computer program, because these
programs represent a simulated environment that may not include all the variables that affect oligo design.
For laboratories that are performing real-time PCR, the software provided with the real-time PCR
instrument may be used for primer design. New primers and probes should always be tested
experimentally for sensitivity as described in Section 4.5 and specificity before use in any method (10,
22).
The specificity of a chosen sequence should be evaluated using BLAST (Basic Local Alignment Search
Tool) (32) or its equivalent. Versions of BLAST are available on the WEB at a number of sites, including
www.ncbi.nlm.nih.gov and www.embl-heidelberg.de. BLAST compares the designed oligo sequences to
known nucleic acid databases such as GenBank and EMBL. The search determines the potential of
hybridization of the chosen oligo with sequences from other organisms. The results of this search should
24 October 2004
Section 4: Method Development and Assessment
be used to define any relevant, closely matched sequences for specificity testing. Specificity testing should
be conducted as described for the PCR positive control (Section 5.1.1) by substituting the heterologous
sample in place of the target organism or nucleic acid. Primers and/or probes should be redesigned if any
heterologous sample is positive.
It is not always possible to design perfect primers and probes. Primers and probes should be selected
based on the best theoretical considerations, and negative features should be documented, and results
interpreted accordingly.
25 October 2004
Section 4: Method Development and Assessment
In addition to required reagents, enhancers that act on different aspects of the reaction may be added to the
PCR master mix. These include dimethyl sulfoxide (DMSO), betaine, tetramethylammonium chloride
(TMAC), and bovine serum albumin (BSA) (25, 27). Other proprietary enhancers and additives are
available from various manufactures. Some of the enhancers act by increasing thermostability of the
polymerase and some by giving the polymerase greater processivity through GC rich regions. The effects
of many others are not completely understood, and although they may be beneficial in some reactions,
their effect on all reactions cannot be predicted. Each additive should be empirically tested with each
combination of template and primers.
A master mix containing optimized reagents should be prepared in the reagent preparation room. The
master mix contains all of the reagents necessary for the reaction to occur except the template, thus
minimizing the potential for reaction-to-reaction variability associated with the pipetting of these reagents.
During master mix preparation, pipette tips should be changed after handling each reagent. The master
mix should be prepared on ice, unless the manufacturers’ instructions specifically state that the reagents
can be mixed at room temperature. Due to the potential interference of the glycerol solution in which
some enzymes are stored, the volume of enzyme used in the reaction should not exceed 10% of the total
reaction volume (9). The master mix should be aliquoted into reaction tubes in the reagent preparation
room. If desired, laboratories can prepare a large quantity of master mix and store aliquots at -20°C after
demonstrating that the freeze/thawing of aliquots does not reduce sensitivity.
Although sequencing provides the most reliable confirmation of a PCR result, the use of two or more
other determinative techniques increases the confidence in the result. The inherent use of internal
hybridization probes in real-time PCR procedures provides a degree of amplicon confirmation that is
similar to the use of dot blot hybridizations in conventional PCR (Section 4.4.2.2.). Further confirmation
of the products of this method by gel electrophoresis and sequencing may still be required. Some real-time
PCR thermocyclers are equipped with capabilities to determine melting point curves of the products that
can also be used for their confirmation.
Note: Due to the small size of some amplicons, especially in real-time PCR detections, confirmation by
sequencing may not be possible.
26 October 2004
Section 4: Method Development and Assessment
Table 4-2. Advantages and Disadvantages of Techniques for Detection and Confirmation of
Amplicons
Technique Objective Advantages Disadvantages
Southern blot Detection of a specific • Combines product • Requires increased time for
sequence of nucleic acid size and probe- transfer of PCR product from
by size on a solid based confirmation gel to membrane and for
support hybridization
• Requires a preliminary
electrophoresis step
• Confirmation by other
techniques, e.g. sequencing,
may still be needed
Dot blot Detection of a specific • Faster than Southern • Confirmation based only on
sequence of nucleic acid blot probe
on a solid support • Confirmation by other
techniques, e.g. sequencing,
may still be needed
Restriction Determinatio n of the size • Relatively easy and • Requires the presence of a
mapping of nucleic acid fragments well-documented restriction site in the amplicon
after enzyme digestion • Requires additional enzymes
and buffers
• Requires additional
electrophoresis after digestion
with the enzymes
• Slow
• Confirmation by other
techniques may still be needed
Melting curve Determination of the • Simple and quick; • Resolution of multiple products
analysis melting temperature of analysis carried out may be poor, e.g., two different
an amplicon at the end of real- amplicons can have the same
time PCR melting temperatures
• Confirmation by sequencing or
other techniques may still be
needed
27 October 2004
Section 4: Method Development and Assessment
The most common medium used for separation of DNA is agarose, a linear polymer that forms a solid gel
of uniform density. An alternative medium is acrylamide, which can form a polyacrylamide gel as a result
of covalent cross-linking of the acrylamide molecules. This type of medium provides higher resolution
than agarose. To visualize the DNA, gels are stained with fluorescent dyes, such as ethidium bromide or
SYBR® Green I, after electrophoresis. Ethidium bromide, but not SYBR Green, also can be added to the
agarose before electrophoresis. Various running dyes can be added along with the DNA in a loading
buffer to monitor DNA migration through the gel.
Many different buffers and concentrations of matrices are used to make and run gels. Different
combinations can provide more or less resistance to the DNA being separated to achieve the desired
resolution and separation.
Note: Electrophoresis alone usually is not sufficient to confirm PCR products, because unrelated
amplicons of similar size to the target of interest may be detected.
28 October 2004
Section 4: Method Development and Assessment
Commercial sequencing facilities that have the necessary equipment and expertise may be used for DNA
sequencing. These facilities should be provided with the appropriate forms of amplified DNA for their
sequencing platform and technology. If confirmation of an amplicon by sequence analysis is anticipated,
the requirements of the sequencing facility should be considered in the development of the PCR
application.
29 October 2004
Section 4: Method Development and Assessment
The detection limit of quantitative methods is defined in this guidance as the minimum concentration of a
substance that can be measured with a given level of confidence that the analyte concentration is greater
than zero (the confidence level should typically be 95% to 99%, but the analytical limitations of some
methods may require the use of lower levels). For presence/absence methods, the detection limit as
defined in this guidance is the minimum concentration of analyte that produces a positive response with a
given level of confidence. The detection limit can be expressed either as the minimum number of
organisms or of the target sequence copy number in a given volume. There are many uncertainties that can
affect the detection limit. Some of them are:
• The type of target nucleic acid being detected (e.g., DNA, mRNA, tRNA, rRNA, etc.)
• The secondary structure and the GC content of the nucleic acid target molecule
• The matrix from which the organism is isolated
• The detection of microbes that are inactivated by physical and chemical disinfectants
There are two different detection limits that are of interest when analyzing environmental samples using
PCR: the detection limit of the PCR procedure and the detection limit of the entire method. For bacteria,
protozoa, and fungi these detection limits often are measured in terms of the minimum detectable counts
or colony forming units (CFU). For viruses, the detection limit should be measured in terms of the
minimum number of detectable physical viral particles, if possible.
The two types of detection limits are determined by analysis of replicate, seeded samples containing
increasingly lower levels of the target organism or target nucleic acid. One way of accomplishing this is to
prepare serial dilutions of an enumerated stock suspension of the target organism. Cell concentrations in
the stock suspension can be determined by direct counting on filters or in a hemacytometer under a
microscope; by automated direct counting methods such as Coulter counting or flow cytometry; or, for
organisms that can be cultured, by counting of colony forming units (CFU) following plating on a suitable
growth medium. Physical particle counts for some viruses can be determined spectrophotometrically (16).
If no method is available to determine physical particles, enumeration can be by infectious units (e.g.,
plaque forming units [PFU]). However, laboratories should recognize that a number of animal viruses
have a high physical to infectious particle ratio, and for these viruses, PCR detects more virus particles in
a sample than an infectious assay can detect (16). The number of replicates that should be analyzed at
each dilution level will depend on the variability of the procedure, the accuracy and variability of the
spike enumerations, the desired level of confidence in the detection limit, and the application of the
method. Depending on the application of the method, it may be desirable to determine the detection limit
for reagent water and a variety of sample matrices to determine the effect of matrix inhibition on the
detection limit.
30 October 2004
Section 4: Method Development and Assessment
It should be noted that manually enumerated spikes containing low numbers of organisms may be
inaccurate, and the results obtained from such spikes should be interpreted with care. Errors associated
with serial dilutions should be minimized by using appropriate volumes that can be accurately measured
and by thoroughly homogenizing stocks immediately before removing an aliquot for the next dilution. If
possible, multiple replicates of each dilution should be enumerated independently of the PCR analysis to
assess the accuracy of the dilutions.
The detection limit can also be expressed in terms of copy number of target sequences (41). This requires
that the nucleic acid sequence of interest be amplified and cloned in a suitable vector and that the
concentration of these molecules be determined, as described below. This is especially helpful when the
organism cannot be cultured. The lowest copy number of the nucleic acid fragment that can be detected by
the PCR method can be calculated from the lowest concentration of DNA detected in g/mL. Ideally, the
detection limit should be in the single copy range.
The following formula can be used to calculate the copy number (CN):
Where:
M = minimum concentration of nucleic acid detected (g/mL)
N = Avogadro’s number (6.022 × 1023 molecules/mole)
L = length of nucleic acid in kilobase pairs (total length of plasmid + insert)
D = conversion factor from 1 Kb of nucleic acid to daltons
ds DNA = 6.6x105 g/mole/Kb
Using spectrophotometry, the approximate concentration of the nucleic acid in grams is calculated by
measuring the optical density (OD) per 1 cm light path, according to the following relationship:
1 OD of ds DNA at 260 nm = 50 µg/mL (5×10-5 g/mL)
Nucleic acid concentrations can be estimated by measuring fluorescence after staining with a fluorescent
dye such as Hoechst 33258. Nucleic acid concentrations also can be calculated by measuring UV-induced
fluorescence emitted by intercalating dyes such as ethidium bromide. Both the fluorimetry and the UV-
transillumination methods require the preparation and measurement of standards (40). A summary of the
three approaches is presented in Table 4-3.
31 October 2004
Section 4: Method Development and Assessment
Detection limits may also be expressed in terms of PCR or RT-PCR units (16), especially for analysis of
organisms (e.g., certain viruses) that cannot be enumerated. PCR units are determined by performing PCR
on serial dilutions of the organism or the nucleic acid from the organism. A unit is defined in terms of the
lowest spike level or highest dilution (lowest concentration) that gives a positive result. These limits
should be estimated from analysis of at least five replicates at each dilution and expressed using most
probable number (MPN) statistics. It should be noted that this method for expressing detection limits can
only be used for relative comparisons of samples using the same primer set, rather than the absolute
values.
When reporting detection limits for the entire method or the PCR procedure, the procedure for calculating
the detection limit should be clearly described, including the units, whether the detection limit is for whole
organisms or copy number, and the associated confidence limits. The units used to report the detection
limit should be based on the procedure used to enumerate the spikes (e.g., if bacterial plate counts are used
to determine CFU, then the method detection limit should be reported in terms of CFU).
32 October 2004
Section 4: Method Development and Assessment
The method detection limit can be determined by processing reagent water, standardized matrix, or field
samples that have been seeded with different concentrations of the target organism prior to any sample
processing. It is not uncommon for the method detection limit to be higher than the PCR detection limit,
depending on the efficiency of the concentration and purification steps, inhibitors present in the field
sample, and other sample processing variables.
4.5.2 Precision
Good analytical methods will have a high precision. Precision is a measure of how closely values from
replicate measurements of a sample agree with each other.1 For quantitative PCR analyses on
environmental samples, method precision can be expressed as the relative standard deviation (RSD) or the
relative percent difference (RPD). The RSD is the standard deviation from three or more replicate samples
divided by the mean and multiplied by 100%. The RPD is the absolute difference of duplicate samples
divided by the mean and multiplied by 100%. The RSD or RPD decreases with increasing precision and
increases with increased variability.
Variability (i.e., low precision) may result from differences in analyst performance, reagents or
equipment, or matrix characteristics. The precision of a method should be tested at the levels at which the
target organism will be found in the environment. If environmental levels are unknown, the precision
should be tested at levels 10 to 100 times greater than the detection limit. The number, range, mean value,
and RSD or RPD of the replicates analyzed to determine precision should be documented, and the nature
of any imprecision characterized.
4.5.3 Recovery
In addition to high precision, good analytical methods will have low bias. Bias is a measure of
disagreement between the concentration of an analyte as measured by a method and the true concentration
in the environmental sample that was tested. A PCR method’s recovery of the target sequence of interest
is an important factor for characterizing method bias at levels higher than the detection limit. Bias can
occur at any component of a PCR method. If a method is designed to detect a group of related organisms,
members of the group may have different recoveries and thus different biases through the sampling
protocol, nucleic acid extraction, and the PCR steps. Bias at the PCR level can occur from the favoring of
the amplification of certain target sequences due to properties of the target, the flanking sequences, or the
overall genome (35). PCR bias may be seen for mixed-template PCRs, including reactions using internal
controls, and may make interpretation of results difficult, because the final results do not reflect the
original makeup of the target templates. Bias usually results when the amplification efficiencies of the
target sequences are not the same. Target sequences may amplify at different efficiencies because the
sequences are not equally accessible to primer hybridization after denaturation, primer-template hybrids
do not form with equal efficiency for all templates, and/or the polymerase acts on the templates with
different efficiencies (42). One factor that may effect primer-template hybridization is the GC content of
the template (35, 42). PCR bias may be reduced by running a low number of PCR cycles and by using
high template concentrations, when possible (35).
1
From a statistical perspective, precision is expressed as the reciprocal of the coefficient of
variation, which would increase as precision increased and decrease as variability increased. However,
because this guidance document is designed for use by environmental laboratories, which commonly use
RSD or RPD, method precision should be determined using this approach. The use of RSD or RPD also
is consistent with the approach used in quantitative EPA microbiology methods; further information on
the use of RSD and RPD for environmental measurements is available in the draft EPA Microbiological
Alternate Test Procedure [ATP] Protocol for Drinking Water, Ambient Water, and Wastewater
Monitoring Methods (EPA- 821-B-03-004) (48).
33 October 2004
Section 4: Method Development and Assessment
Recovery of the PCR analysis and the entire method can be assessed by the analysis of seeded samples.
For quantitative PCR analyses, recovery is determined as the total amount of the analyte found in a seeded
sample minus the background (i.e., amount detected in the unseeded sample) divided by the amount of the
analyte seeded into the sample. To determine recovery for presence/absence methods, serial dilutions of
the seeded sample should be performed and five replicates at each dilution analyzed. The amount of
analyte in the seeded and unseeded samples should then be calculated using MPN analysis as described in
Section 4.3.1.1. The recovery would then be calculated similar to the quantitative analysis by subtracting
the MPN-estimated background concentration from the MPN-estimated concentration of the seeded
sample and then dividing by the amount seeded (section 5.1).
Results that are obtained using methods that have low recoveries should be qualified as potentially biased.
Laboratories should also be aware that recovery of target organisms that are endogenous to a matrix (e.g.,
enteric pathogens associated with fecal material in river water) may not be the same as that measured by
seeded samples (e.g., where the organisms are not associated with other material). Unless it can be shown
that the recoveries of endogenous and seeded organisms are similar, sample results should be recorded
as biased.
The results of these studies should be carefully considered before the method is used for environmental
sample monitoring. If the validation study data indicate that the method is not consistent enough for
routine use in a single laboratory or cannot be performed by other qualified laboratories, the method
should not be used until it is standardized. If the validation study data indicate that the method does not
perform acceptably in many of the matrices in which it would be used for routine monitoring, additional
development work is needed to address matrix interferences. If the results of the single-laboratory or
interlaboratory validation study demonstrate that the method can be consistently performed, and is
characterized by acceptable accuracy, sensitivity, and specificity, then it may be used for monitoring.
34 October 2004
SECTION 5. QUALITY CONTROL SAMPLES FOR METHODS USING
PCR
Laboratories using PCR should analyze positive and negative QC samples on a routine basis to
demonstrate adequate performance of PCR based methods. General guidance on both positive and
negative control types and the frequency in which they should be performed is provided in Sections 5.1
and 5.2. The actual number of controls will depend on the experimental design and expected variability. A
summary of these controls is provided in both Tables 5-1 and 5-2 and Figure 5-1. A summary of QC
samples for detection and confirmation procedures is provided in Section 5.3. Corrective actions for QC
sample failures and false positive/false negative prevention are discussed in Section 5.4 and 5.5,
respectively. Guidance for participating in proficiency testing studies is suggested in Section 5.6.
The method protocol should be documented in the laboratory’s SOPs, and concise versions of the SOP for
a particular part of the method should be posted at the bench where the procedures will be performed.
Posting bench SOPs provides technicians with easy access to method procedures, and helps promote
consistency between analysts. SOPs for a particular method also should document QA procedures, such as
the types and frequency of QC controls, and corrective actions for positive and negative control failures.
35 October 2004
Section 5: Quality Control Samples for Methods Using PCR
PCR positive controls are prepared by the addition of an exogenous control to the master mix. Exogenous
controls can be any of the following:
• A purified total nucleic acid extract from the organism containing the sequence of interest.
• The whole organism, which may be used when the nucleic acid target of interest can be released from
the seeded organism by heating before or during PCR.
• A specific nucleic acid fragment containing the entire sequence to be amplified, including primer
binding sites (e.g., a low concentration of a previously amplified, and/or a cloned DNA fragment that
has been sequenced for confirmation).
• A cloned DNA fragment containing a modified form of the target sequence (see Section 4.3.1.1).
Modified forms produce PCR products that can be distinguished from the target sequence by size, by
restriction mapping, and/or by an alternative probe recognition sequence. Modified forms, which also
are called “internal controls,” are prepared by in vitro generation of insertions, deletions or other
sequence changes.
Note: Modified forms of the target that differ significantly from the target sequence in length or GC
content will have different amplification efficiencies from target sequences, and thus are not
appropriate for use as positive controls.
• A heterologous sequence that has been previously shown to be amplified with an efficiency that is
comparable to the target sequence.
• RNA transcribed from a cloned DNA fragment containing the target sequence or a modified target
sequence (for use in RT-PCR applications).
For quantitative PCR methods, the PCR positive control is evaluated by determining the total amount of
the target nucleic acid or organism in the control divided by the amount added to the reaction as a spike.
For qualitative PCR methods, the PCR positive control typically is evaluated in terms of ‘detect’ or ‘non-
detect’ (although a detection signal that is weaker than normal may still indicate a problem).
Endogenous controls are target sequences that are expected to always be present in the sample (e.g.,
ribosomal DNA or RNA). These controls should be used only if it is demonstrated that there are
consistent occurrence and recovery of the endogenous control templates in different samples, and if the
control and target assays show comparable susceptibility to inhibition in different matrices. They are
commonly used for analyses of clinical samples, but are unlikely to be applicable for the analyses of most
environmental samples.
36 October 2004
Section 5: Quality Control Samples for Methods Using PCR
Exogenous controls can be used to analyze for PCR inhibition in several ways:
• In separate aliquots of the same nucleic acid extract from the same sample
• In extracts from separate, replicate samples processed in parallel
• In the same aliquot of the same sample using modified internal controls
The absence of detectable PCR product from this control signals PCR inhibition. For quantitative PCR
methods, different degrees of inhibition can be assessed directly by comparing the results from the control
with the results from the PCR positive control (section 5.1.1). However, this requires that both controls be
amended with the same amount of the positive control template. For qualitative PCR methods, the PCR
inhibition positive control is evaluated in terms of a detect or non-detect (although a detection signal that
is weaker than that of the PCR positive control may still indicate a problem).
Assessing different degrees of inhibition can be difficult using spikes of the template itself as the positive
control since this template may be present in varying concentrations in the test sample. Use of a modified
form of the target template as the positive control is generally recognized to be the preferred approach,
because it is likely that the nearly identical sequences of the target and control template will confer equal
susceptibility to PCR inhibition. However, this approach requires the custom design and preparation of a
surrogate template for each primer set. Analyses for an added heterologous control template are generally
easier to accomplish, but it is particularly important to demonstrate that the susceptibility of these assays
to PCR inhibition is comparable to that of the target sequence in all samples. Primer sets for heterologous
exogenous controls also should be checked for compatibility with the target sequence primers, when used
in same tube reactions. Care should be taken when designing same-tube reactions, because the different
assays will be competing for the same reagents (e.g., dNTPs, enzymes, and, in the case of modified target
templates, primers), and target detection sensitivity may be affected adversely.
Because environmental matrices, such as river water, are constantly changing, inhibition positive controls
should be performed for every sample and every target. If no matrix effects on the targets of interest are
detected, then the frequency with which this control is performed can be reduced to a periodic check to
assess potential changes in the matrix over time.
37 October 2004
Section 5: Quality Control Samples for Methods Using PCR
A second approach to the matrix spike is to seed the environmental sample with a different organism than
the target organism. This approach eliminates the need to process and analyze duplicate samples, since the
target and surrogate organism nucleic acids can be independently analyzed in the same sample. It should
only be used if the surrogate has similar properties to the target organism and if it is not present in the
matrix tested. The same amounts of surrogate organism should also be added with the target organism to
the method positive control (Section 5.1.3) for comparison of results with the matrix spike samples.
However, this approach should not be used unless it is shown that the environmental inhibitors of PCR
affect detection of surrogate and target organisms similarly. In addition, it should be shown that the
recoveries of the surrogate and target organisms throughout the entire method are similar. The use of a
different organism than the target for the control may also be limited if the front-end sample processing
techniques are specific for the target (e.g., immunomagnetic separation that uses magnetic beads coated
with antibodies specific to the target), or if there is competition between the target and the surrogate
organism.
The processing and analysis of additional samples required for matrix spikes can be minimized by
analyzing this control for only the first sample analyzed from a particular matrix. As described above for
inhibition controls, however, some environmental matrices are constantly changing. As a result, MS
samples should be performed regularly on a particular matrix to assess if any potential changes in the
matrix have affected method performance, or until it can be documented that the variability of the
recoveries within a given environmental matrix is acceptable.
Verify that no contaminating nucleic acid At least 10% of the number of field
PCR negative control
has been introduced into the master mix samples analyzed per PCR batch
38 October 2004
Section 5: Quality Control Samples for Methods Using PCR
Figure 5-1. Sample Flow and Controls of PCR Methods on En vironmental Samples
39 October 2004
Section 5: Quality Control Samples for Methods Using PCR
A DNA ladder (a mixture of DNA fragments of known sizes), should also be run on each gel to provide a
standardized gauge of the size of DNA fragments seen in the test samples and controls. The size of the
expected product should be within the size range covered by the standard. Extrapolation beyond the range
of the standard should not be performed.
A sample of the amplicon, prepared and processed in the same manner as the positive control, but with no
addition of the restriction enzyme, should be run to verify that digestion of the nucleic acids resulted from
the added restriction enzyme, and not contaminating enzymes.
Probe-based quantitative PCR and melting curve analysis are both part of real-time PCR, and the same
positive and negative controls should be used for these confirmation techniques as are used for any type of
PCR (see Sections 5.1 and 5.2).
In the event of unacceptable results from sequence analyses (e.g., the fluorescent signal is weak
throughout the sequencing ladders of all samples), a sequencing reaction control using a template and
primer provided by the manufacturer should be run according to the manufacturers’ instruction. The
source of the problem (e.g., poor PCR amplification, expired fluorescent dyes) should be determined and
corrected, and any affected samples reanalyzed.
40 October 2004
Section 5: Quality Control Samples for Methods Using PCR
If PCR negative controls or method blanks produce specific amplification products, all samples associated
with the failed controls should be considered invalid, and all positive samples should be listed as
potentially false-positive samples. The source of contamination should be identified and eliminated. For
recommendations on cleaning, see Sections 2.4 and 2.6. Once determined, the source of the contamination
should be documented, and the samples in the batch should be recollected and reanalyzed, if possible. If
the source of the contamination cannot be identified, additional types of negative controls should be added
at various steps in the method to determine where the contamination is being introduced. Three such
controls are the following.
• Equipment blank. This control is performed by passing sterile reagent grade water or buffer through
the equipment and processing the water as if it were a PCR negative control. If this control is found to
be positive, all analysis should cease until the source of the problem is identified. Equipment blanks
should then be run more frequently until it is shown that the problem has been corrected.
• Wipe test. This control is performed by wiping an area, pipette, or other equipment with a sterile,
nucleic acid-free, gauze or q-tip, then suspending the wipe in molecular-grade water and processing it
as a PCR negative control. To determine whether nucleic acids detected are part of the normal
background or the source of sample contamination, wipe tests should be performed routinely to
establish a baseline for comparison.
• Room QC controls. This control is prepared by adding sterile water in place of template to a master
mix in the reagent preparation room. These tubes are placed in working areas in the reagent
preparation and sample preparation rooms and opened for 15 minutes. The tubes are then closed and
run as a PCR negative control. These controls may be run on a monthly basis. If any of these controls
are positive, all work should cease until the source of the contamination can be identified and
corrected.
41 October 2004
Section 5: Quality Control Samples for Methods Using PCR
The most widely used approach, enzymatic inactivation, involves the introduction of dUTP (the
nucleotide deoxyuradine) in lieu of TTP (the nucleotide thymine), along with the enzyme uracil
N–glycosylase (UNG, also known as uracil DNA glycosylase, UDG), into the PCR master mix. As a result
of this substitution, all amplified product will contain deoxyuradine instead of thymidine. The
deoxyuradine incorporated into the DNA becomes a target for site-specific digestion by UNG. Prior to
amplification, an incubation step is performed to activate UNG and enzymatically degrade carryover
amplified DNA from previous amplifications. The UNG is then inactivated by heating to 94°C, and the
actual PCR cycles are begun. This procedure cannot be used with nested PCR, because it would destroy
the product of the first amplification. Similarly, it normally cannot be used for the reverse transcription
phase of a RT-PCR assay because it would destroy the cDNA produced by reverse transcriptase.
However, recently there have been reports of the use of a heat labile UNG that may overcome this
problem (43).
42 October 2004
Section 5: Quality Control Samples for Methods Using PCR
There are several procedures that are reported to reduce inhibition, including the incorporation of 200 to
400 ng/µL of BSA (bovine serum albumin) to the PCR (27). However, there are no procedures that work
with all applications, which is why the rigorous use of each positive control discussed in Section 5.2.1 is
necessary. Good laboratory practices, as discussed in Sections 2 and 3, should be followed to prevent
contamination with DNase.
PT samples should be composed of a matrix similar to that normally analyzed by the method and should
be seeded with whole target organisms. Currently there are few PT programs for molecular methods. If PT
programs or samples are unavailable for a particular assay, laboratories should implement other
procedures to test for internal consistency in their results, such as testing duplicate or split seeded
samples. If possible, the identification of one of the duplicates should be blind so that the technician
performing the analyses does not know that the two samples are duplicates. Another alternative to
approved PT programs is sending a duplicate or split sample to an outside laboratory for comparison with
the laboratory’s own results.
43 October 2004
SECTION 6. DATA RECORDING, RECORD KEEPING, AND DATA
EVALUATION
The generation of all environmental data should be completely and consistently documented, so the results
can be assessed by an independent party. Guidance for data recording and record keeping are provided in
Section 6.1. A discussion of issues related to the evaluation of PCR data from the analysis of
environmental samples is provided in Section 6.2.
6.1.1 Equipment
All equipment calibration and maintenance should be documented in appropriate logbooks. A schedule of
equipment calibration and maintenance should be established with the procurement of each new piece of
equipment, and a copy of the schedule should be posted in the laboratory near the equipment. Each of
these activities should be documented in the laboratory’s SOPs and followed by laboratory staff involved
in the use of the equipment.
44 October 2004
Section 6: Data Recording, Record Keeping, and Data Evaluation
• Storage buffer
• Number of units (enzyme)
• Associated buffer (enzyme)
For all amplicon confirmation procedures, data should be presented in a clear and concise way to ensure
that results are interpreted properly. All laboratory SOP procedures for the method should be followed,
including appropriate composition of the gel, buffers, voltage, and length in time of run. Any differences
to the SOP should be noted. Specific results for each amplicon confirmation should be recorded as
indicated in Sections 6.1.3.1 to 6.1.3.5.
Before being reported to the end user, all data should be reviewed by an analyst familiar with the method
to ensure that all method requirements were met, including use of appropriate testing conditions (e.g.,
reagents, thermocycler conditions), analysis of appropriate controls, and analysis of samples within
method holding times. Any information that impacts the quality or validity of the data should be noted on
the data report.
6.1.3.1 Electrophoresis
A photograph or digital image should be taken of the gel under UV light, and reactions associated with
each lane identified.
6.1.3.2 Blots
The following data should be recorded for blot procedures:
• Southern blot. A photograph or digital image of the gel should be taken and labeled prior to the
transfer of the DNA to the membrane. A graduated scale should be included in the picture to compare
it to the membrane. After the DNA is transferred to the membrane and probed, the x-ray film is taken
and compared to the gel picture to verify that the hybridized band is in the same position as the
amplicon band on the gel. The x-ray film should be a part of the record. Many laboratories have
replaced the x-ray film with other equipment such as the phosphorimager. In such cases, the
computer-generated image of the hybridization signal should be compared to the gel picture, and be
included as part of the record.
• Dot blot. The exposed x-ray film or digital image should clearly show the positive and negative
hybridization.
45 October 2004
Section 6: Data Recording, Record Keeping, and Data Evaluation
6.1.3.3 Sequences
All raw fluorescent chromatographs of the sequences should be archived and the associated sequences
recorded. The sequencer and the manufacturer of the sequencing reaction should be recorded in addition
to the date, time, and analyst. A comparison should be conducted to ensure that the determined nucleotide
bases match between the sequenced strands for both the forward and reverse sequences, and a combined
edited sequence should be reported.
46 October 2004
SECTION 7. GLOSSARY
Accuracy—The degree of agreement between an observed value and an accepted reference value.
Accuracy includes random error (precision) and systematic error (bias or recovery) that are caused by
sampling and analysis.
Amplification—The process by which extra copies of a gene or a DNA sequence are formed.
Annealing—In PCR, the step at which primers (oligonucleotides) bind to complementary segments on
template/sample DNA (typically done at 47°C - 60°C).
Base pair—The pair of nitrogenous bases, consisting of a purine linked by hydrogen bonds to a
pyrimidine, that connects the complementary strands of DNA or of hybrid molecules joining DNA and
RNA. The base pairs are adenine-thymine and guanine-cytosine in DNA, and adenine-uracil and guanine-
cytosine in RNA.
Bias—A systematic or a persistent distortion of a measurement process that deprives the result of
representativeness; i.e., the expected sample measurement is different than the sample’s true value
expressed in terms of recovery. A data quality indicator.
Blot—Transfer of nucleic acids to a membrane. The bound nucleic acid is often then detected by probe
hybridization.
Cloning—The use of an organism to produce more copies of a specific fragment of nucleic acid.
Normally done using vectors.
Confirmatory test—A test used to validate the results obtained (e.g., the presence or absence of a PCR
amplicon).
Complementary—A strand of DNA or RNA that is capable of forming specific pairing of the purine and
pyrimidine base sequences with those of a second strand.
Quality control sample—A standard sample included in an assay used to determine the validity of the
test based on a predetermined outcome or range of outcomes.
Denaturation—In PCR, breaking hydrogen bonds between base pairs in double-stranded nucleic acid
molecules to produce single-stranded nucleic acid molecules (typically done at 94°C - 97°C).
47 October 2004
Section 7: Glossary
DNA—Deoxyribonucleic acid. A nucleic acid that carries the genetic information of an organism. DNA is
capable of self-replication, and is used as a template for the synthesis of RNA. DNA consists of two long
chains of nucleotides twisted into a double helix and joined by hydrogen bonds between the
complementary bases adenine and thymine or cytosine and guanine. The sequence of nucleotides
determines individual hereditary characteristics.
Endogenous—A region of DNA or RNA that always is found naturally in a test sample and which differs
from the target sequence of a PCR method.
Functional validation—Testing that demonstrates that a method or procedure works the way it was
intended to work.
Gene—A hereditary unit consisting of a sequence of DNA that occupies a specific location on a
chromosome and determines a particular characteristic in an organism. Genes undergo mutation when
their DNA sequence changes.
Hairpin loops—A secondary structure of a nucleic acid that can be formed due to inverted repeats in the
sequence.
Hybridization—The binding of one nucleic acid to another by the formation of hydrogen bonds between
the bases on the two molecules. The sequences must be completely or nearly complementary to each other
in order for this type of bonding to occur. This process is usually used in conjunction with a probe to
determine the presence of a specific sequence of DNA.
Hybridize—To form a double-stranded nucleic acid by the pairing of complementary regions of two
strands of DNA that were not originally paired.
48 October 2004
Section 7: Glossary
Inhibition—The reduction in efficiency in a PCR caused by elements in the environmental sample that
interfere with the normal reaction.
Inhibition positive control—A positive control sample used to verify that the constituents of the matrix
carried over from the isolation of the organism do not inhibit either the isolation of the nucleic acid or the
PCR.
Limit of detection—The minimum concentration of a substance that can be measured (or gives a positive
response) with a given level of confidence that the analyte concentration is greater than zero.
Master mix—Solution containing all reagents (except for the test or sample nucleic acid) that is required
to perform PCR.
Matrix spike—A QC sample prepared by adding a known amount of target analyte to a specified amount
of sample matrix for which an independent estimate of target analyte concentration is available. A matrix
spike is used, for example, as a positive control determining the effect of the matrix on the overall
method's recovery and to verify it does not have an inhibitory effect on the PCR.
Method blank—An aliquot of reagent water that is treated exactly as a sample, including exposure to all
glassware, equipment, solvents, and procedures that are used with samples. The method blank is used as a
negative control to verify that no contamination is present in the laboratory environment, the reagents, or
the apparatus.
Method positive control—A positive control used to verify that the entire method is performing properly
and is capable of amplifying the target nucleic acid from the organism of interest.
Nested PCR—PCR in which two pairs of PCR primers are used for a single locus. The first pair amplifies
the locus. The second pair of primers (nested primers) bind within the first PCR product and produce a
second PCR product that will be shorter than the first one. The specificity of the assay is increased
because the inner primers amplify only if the first PCR reaction yielded a specific product.
Nucleic acid—Any of a group of complex compounds found in all organisms, composed of purines,
pyrimidines, carbohydrates, and phosphoric acid. Nucleic acids in the form of DNA and RNA control
cellular function and heredity.
Oligo—See oligonucleotide.
Oligonucleotide—A short segment of single stranded nucleic acid or a short sequence of nucleotides
(e.g., PCR primers and real-time probes).
PCR—Polymerase Chain Reaction. A technique for amplifying DNA sequences in vitro by separating the
DNA into two strands and incubating it with nucleotides, oligonucleotide primers, and DNA polymerase.
It can amplify a specific sequence of DNA by more than a billion times (e.g., 40 cycles = 240 = one trillion
amplicons).
49 October 2004
Section 7: Glossary
PCR Batch—A group of samples set-up and amplified at the same time under the same conditions, using
the same PCR master mix, and in the same thermocycler.
PCR negative control—A negative control used to verify that no contaminating nucleic acid has been
introduced into the master mix.
PCR positive control—A positive control used to verify that the PCR master mix and reaction were
designed correctly in order to produce amplification of the target nucleic acid.
Precision—The degree to which a set of observations or measurements of the same property, usually
obtained under similar conditions, conform to themselves; a data quality indicator. Precision is usually
expressed as standard deviation, variance, or range, in either absolute or relative terms.
Primer—A segment of DNA or RNA that is complementary to a given DNA sequence and that is needed
to initiate replication by DNA polymerase.
Primer binding sites—Sequences in the DNA that are complementary to the primer used in the PCR.
Primer dimer—Two primer sequences that form a small region of double stranded DNA due to
complementary sequences being present on both primers. This formation inhibits the primers use in the
polymerase chain reaction.
Probe—A substance, such as DNA, that is fluorescently, radioactively, or otherwise labeled and used to
detect or identify another substance in a sample.
Processivity—The ability of DNA polymerase enzyme to amplify DNA, often referring to the
amplification of longer sections of DNA.
Proficiency testing (PT) samples—Samples qualitatively and quantitatively unknown to the analyst that
can be obtained commercially or by agreement from another laboratory.
Proofreading—Capability of a polymerase to be able to check for errors in the nucleotide sequence that it
synthesizes as well as being able to repair any errors that occur.
Reagent—A substance used in a chemical reaction to detect, measure, examine, or produce other
substances.
Real-time PCR—Detection of specific nucleic acid amplification products as they accumulate during
PCR in real-time using various fluorescence-based chemistries.
Recovery—The total amount of the analyte found in the sample, corrected for background, divided by the
amount of the analyte added into the sample.
Relative percent difference (RPD)—An estimate of the variability of two numbers expressed in relative
terms. Calculated as the absolute value of the difference of the two numbers, divided by their mean:
50 October 2004
Section 7: Glossary
Equal to the relative standard deviation of the two numbers multiplied by the square root of 2.
Relative standard deviation (RSD)—The standard deviation expressed as a percentage of the mean
(100F/X); i.e., the coefficient of variation.
Reverse transcriptase—A polymerase that catalyzes the formation of complementary DNA (cDNA)
from an RNA template.
RNA—Ribonucleic acid. A polymeric constituent of all living cells and many viruses, consisting of a
long, usually single-stranded chain of alternating phosphate and ribose units with the bases adenine,
guanine, cytosine, and uracil bonded to the ribose. The structure and base sequence of RNA are
determinants of protein synthesis and the transmission of genetic information.
Sensitivity—The sensitivity of a test can be described as the proportion of all positive results detected
that were truly positive. All positives are the sum of (detected) true positives (TP) and (undetected) false
negatives (FN). Sensitivity is therefore: TP / (TP + FN) × 100%
Sample batch—A set of test samples set-up and processed together through all steps of a PCR method.
Specificity—The specificity of a test can be described as the proportion of all negatives it detects that
truly were negative. All negatives are the sum of (detected) true negatives (TN) and false positives (FP).
Specificity is therefore: TN / (TN + FP) × 100%
Target nucleic acid—The specific sequence of nucleic acid that the PCR is designed to amplify.
Template—A molecule of a nucleic acid, such as DNA, that serves as a pattern for the synthesis of
replicate copies of DNA amplicons.
Thermocycler—Device which can be programed to raise and lower the temperature of reaction mixes and
is used to perform PCR.
51 October 2004
Section 7: Glossary
Ultraviolet (UV)—Of or relating to the range of invisible radiation wavelengths from about four
nanometers, on the border of the x-ray region, to about 380 nanometers, just beyond the violet in the
visible spectrum.
Uracil N–glycosylase (UNG; a.k.a. uracil-DNA-glycosylase, UDG)—An enzyme that hydrolyses DNA
strands with deoxyuradine incorporated in its sequence. It can be used to eliminate carryover
contamination from previous PCR.
Vector—A DNA molecule that is inserted into a host cell that allows replication of itself and any inserted
sequence as the organism lives and reproduces.
52 October 2004
SECTION 8. REFERENCES AND RECOMMENDED READING
1. AACTG. 1999. General Pipetting Operational Considerations/Tips. The Adult AIDS Clinical Trials
Group. [Online.] http://aactg.s-3.com/pub/download/imm/pipettecalibdocmay99.doc. Accessed 31
December 2003.
2. American Public Health Association, American Water Works Association, and Water
Environment Federation. 1995. Standard Methods for Water and Wastewater, 20th Edition.
Sections: 9020, 9221, 9222.
4. Barker, K. 1998. At the Bench a Laboratory Navigator. Cold Spring Harbor Laboratory Press,
Plainview, NY.
5. Basic Laboratory Incorporated. 2000. Quality Assurance Plan, revision 4.0. Basic Laboratory Inc.
[Online.] http://www.basiclabinc.com/qapp/qapp2000.pdf. Accessed 30 May 2003.
6. Becker, S., P. Böger, R. Oehlmann, and A. Ernstet. 2000. PCR bias in ecological analysis: A case
study for quantitative taq nuclease assays in analyses of microbial communities. Appl. Environ.
Microb. 66(11):4945-4953.
9. Choi J. S., J. S. Kim, C. O. Joe, S. Kim, K. S. Ha, and Y. M. Park. 1999. Improved cycle
sequencing of GC-rich DNA template. Exp Mol Med. 31(1):20-24.
10. Coyne, V. E., M.D. James, S. J. Reid, and E.P. Rybicki. 2003. Molecular Biology Techniques
Manual, 3rd ed, University of Cape Town. [Online.]
http://www.mcb.uct.ac.za/manual/MolBiolManual.htm. Accessed 30 December 2003.
11. Dahllöf, I. 2002. Molecular community analysis of microbial diversity. Curr. Opin.Biotech. 13:213-
217.
12. DeLeon, R., P. Hacker, and P. A. Rochelle. 2002. Development of stringent verification procedures
for molecular detection of enteric viruses in water. American Water Works Association - Water
Quality Technology Conference.
14. Fairfax, M. R., M. A. Metcalf, and R. W. Cone. 1991. Slow inactivation of dry PCR templates by
UV light. PCR Meth. Appl. 1:142-143.
53 October 2004
Section 8: References and Recommended Reading
15. Federal Bureau of Investigation. 2000. Quality Assurance Standards for Forensic DNA Testing
Laboratories. Federal Bureau of Investigation. [Online.]
http://www.fbi.gov/hq/lab/fsc/backissu/july2000/codis2a.htm. Accessed 27 February 2004.
16. Fout, G. S., B. C. Martinson, M. W. Moyer, and D. R. Dahling. 2003. A multiplex reverse
transcription-PCR method for detection of human enteric viruses in groundwater. Appl. Environ.
Microb. 69:3158-3164.
17. Francy D. S., R. N. Bushon, C. Kephart, A. Gifford, E. Granger, K. Mauch, and D. M. Stoeckel.
2004. Quality Assurance/ Quality Control Manual. Ohio District Microbiology Laboratory. [Online.]
http://www-oh.er.usgs.gov/micro/qcmanual/manual.html. Accessed 29 February 2004.
18. GeneScan. 2003. Quality control/Quality Assurance. GeneScan USA Inc. [Online.]
http://www.gmotesting.com/qa.html. Accessed 30 May 2003.
19. Glick, B. R. and J. J. Pasternak. 1998. Molecular Biotechnology Principles and Applications of
Recombinant DNA, 2nd ed, ASM Press, Washington, D.C.
20. Greer, C. W., L. G. Whyte, J. R. Lawrence, L. Masson, and L. Brousseau. 2001. Genomics
technologies for environmental science. Environ. Sci. Technol. 35:360A-366A.
21. Heid, C. A., J. Stevens, K. J. Livak, and P. M. Williams. 1996. Real time quantitative PCR.
Genome Res. 6(10):986-994.
22. Heuvel, J. V. 2004. Quantitative Reverse Transcription Polymerase Chain Reaction (RT-PCR) and
Other PCR Procedures. Penn State University. [Online.]
http://www.cas.psu.edu/docs/CASDEPT/VET/jackvh/jvhpcr.html. Accessed 27 February 2004.
23. Hochmeister, M. 1995. DNA technology in forensic Applications. Mol. Aspects Med. 16:315-473.
24. Innis, M. A., D. H. Gelfand, J. J. Sninsky, T. J. White. 1990. PCR Protocols: A Guide to Methods
and Applications. Academic Press, San Diego, CA.
26. Kousuke, I. and F. Manabu. 2001. Optimization of annealing temperature to reduce bias caused by
primer mismatch in multitemplate PCR. Appl. Environ. Microb. 67(8):3753-3755.
27. Kreader, C. A. 1996. Relief of amplification inhibition in PCR with bovine serum albumin or T4
gene protein. Appl. Environ. Microb. 62(3):1102-1106.
28. Markoulatos, P., N. Siafakas, and M. Moncany. 2002. Multiplex polymerase chain reaction: A
practical approach. J. Clin. Lab. Anal. 16:47–51.
29. Millar, D. S., S. J. Withey, M. L. V. Tizard, J. G. Ford, and J. Herman-Taylar. 1995. Solid-phase
hybridization capture of low-abundance target DNA sequences: Application to the polymerase chain
reaction detection of Mycobacterium paratuberculosis and Mycobacterium avium subsp. silvaticum.
Anal. Biochem. 226:325–330.
54 October 2004
Section 8: References and Recommended Reading
30. Muir, P., A. Ras, P. E. Klapper, G. M. Cleator, K. Korn, C. Aepinus, A. Fomsgaard, P. Palmer,
A. Samuelsson, A. Tenorio, B. Weissbrich, A. M. van Loon. 1999. Multicenter quality assessment
of PCR methods for detection of enteroviruses. J. Clin. Microbiol. 37(5):1409-1414.
31. National Committee for Clinical Laboratory Standards (NCCLS). 1995. Molecular Diagnostic
Methods for Infectious Diseases; Approved Guideline. NCCLS MM3-A.
33. NuAire. 2004. Selecting a Biological Safety Cabinet. NuAire, Inc. [Online.]
http://www.nuaire.com/bscs/selectbsc.html. Access 27 February 2004.
34. Ou, C., J. L. Moore, and G. Schochetman. 1991. Use of UV irradiation to reduce false positivity in
polymerase chain reaction. BioTechniques. 10(4):442, 444, 446.
35. Polz, M. F. and C. M. Cavanaugh. 1998. Bias in template-to-product ratios in multitemplate PCR.
Appl. Environ.Microb. 64(10):3724-3730.
36. Pfaffl M. W. 2001. A new mathematical model for relative quantification in real-time RT-PCR.
Nucleic Acids Res. 29(9):e45.
38. Ririe, K. M., R. P. Rasmussen, and C. T. Wittwer. 1997. Product differentiation by analysis of
DNA melting curves during the polymerase chain reaction. Anal. Biochem. 245:154–60.
39. Rochelle, P. A. 2001. Environmental Molecular Microbiology: Protocols and Applications. Horizon
Scientific Press, Norfolk, England.
40. Sambrook, J. and D. W. Russel. 2001. Molecular Cloning: A Laboratory Manual. Cold Spring
Harbor Laboratory Press, lainview, NY.
41. Sen, K. 2000. Rapid identification of Yersinia enterocolitica in blood by the 5' nuclease PCR assay. J.
Clin. Microbiol. 38:1953-1958.
42. Suzuki, M. T. and S. J. Giovannoni. 1996. Bias caused by template annealing in the amplification of
mixtures of 16s rRNA genes by PCR. Appl. Environ. Microb. 62(2):625-630.
43. Taggart E. W., K. C. Carroll, C. L. Byington, G. A. Crist, D. R. Hillyard. 2002. Use of heat labile
UNG in an RT-PCR assay for enterovirus detection. J Virol Methods. 105(1):57-65.
44. Theron, J. and T. E. Cloete. 2000. Molecular techniques for determining microbial diversity and
community structure in natural environments. Crit. Rev. Microbiol. 26:37-57.
45. Weaver, R. F. 1999. Molecular Biology, 1st ed. WCB/McGraw Hill, Boston, MA.
46. Wilhelm, J. and A. Pingoud. 2003. Real-time polymerase chain reaction. Chembiochem.
4(11):1120-1128.
55 October 2004
Section 8: References and Recommended Reading
47. University of Victoria. 2003. Laminar Flow Hoods & Biological Safety Cabinets. University of
Victoria. [Online.] http://ohs.uvic.ca/biosafety/biosafetycabinets.html. Accessed 1 March 2004.
48. USEPA. 2003. EPA Microbiological Alternate Test Procedure (ATP) Protocol for Drinking Water,
Ambient Water, and Wastewater Monitoring Methods. EPA- 821-B-03-004. Office of Water,
Washington, DC.
49. USEPA. 2003. Workshop to Develop a Protocol for Reliable Genetic Methods for the Detection of
Viruses, for Use in EPA’s Water Programs. Office of Water, Cincinnati, OH.
50. USEPA. 1997. Manual for the Certification of Laboratories Analyzing Drinking Water. EPA-815-R-
97-001. Office of Water, Cincinnati, OH.
51. Zimmermann, K. and J. W. Mannhalter. 1998. Comparable sensitivity and specificity of nested
PCR and single-stage PCR using a thermally activated DNA polymerase. BioTechniques. 24:222-224.
56 October 2004
Office of Water (4607)
EPA 815-B-04-001
www.epa.gov/safewater
October 2004