099 Vitamin B12-IFU-V3.06-en-EU
099 Vitamin B12-IFU-V3.06-en-EU
099 Vitamin B12-IFU-V3.06-en-EU
INTENDED USE
The kit has been designed for the quantitative determination of
Vitamin B12 in human serum.
The method can be used for samples over the range of
Shenzhen New Industries Lotus Medical Equipment 12.5-2000.0 pg/ml.
Biomedical Engineering Co., Ltd Limited The test has to be performed on MAGLUMI Fully-auto
No.16, Jinhui Road, 26B Cameron Court, chemiluminescence immunoassay (CLIA) analyzer (Including
Pingshan New District, Cork Street, Dublin 8, Maglumi 600, Maglumi 800, Maglumi 1000, Maglumi 1000 Plus,
Shenzhen, 518122, P.R.China. Ireland Maglumi 2000, Maglumi 2000 Plus, Maglumi 4000).
Tel: +86-755-21536601 Tel: +353-1-6571034 Catalog Number Specification
Fax:+86-755-28292740 E-mail: peter@lotusme.org 130213002M 100 tests
130613002M 50 tests
KIT COMPONENTS
Material Supplies
Component 100 tests 50 tests
Magnetic Microbeads: coated with
sheep anti-FITC polyclonal antibody, 2.5 ml 2.0 ml
containing BSA, 0.09% NaN3.
(recommended).
Accessories Required But Not Provided
After each servicing of MAGLUMI Fully-auto
MAGLUMI Reaction Module REF: 630003
chemiluminescence immunoassay (CLIA) analyzer.
MAGLUMI Starter 1+2 REF: 130299004M
If controls are beyond the expected range.
MAGLUMI Wash Concentrate REF: 130299005M
Whenever room temperature changes exceed 5 °C
MAGLUMI Light Check REF: 130299006M (recommended).
Please order accessories from Shenzhen New Industries
Biomedical Engineering Co., Ltd (SNIBE) or our representative. SPECIMEN COLLECTION AND PREPARATION
Sample material: serum
Collect samples using standard procedures. Store at 2-8°C: 24
Preparation of the Reagent Integral hours, for longer storage periods: freeze to below - 20°C
Avoid repeated freezing and thawing more than one time, stored
Before the sealing is removed, gentle and careful horizontal
samples should be thoroughly mixed prior to use (Vortex mixer).
shaking of the Reagent Integral is essential (avoid foam formation!)
Please ask local representative of SNIBE for more details if you
Remove the sealing and turn the small wheel of the magnetic
have any doubt.
microbeads compartment to and fro, until the color of the
Specimen Conditions
suspension has changed into brown. Place the Integral into the
• The sample which has been placed at the room temperature
reagent area and let it stand there for 30 min. During this time, the
more than 8 hours cannot be used again.
magnetic microbeads are automatically agitated and completely
• The sample serum with high concentration of protein
resuspended.
(>90g/L) cannot be used to do the tests. As the high dose of
Do not interchange integral components from different
protein will form the gel and block the needle
reagents or lots!
•Do not use specimens with the following conditions:
Preparation of Displacing Reagent
(a) heat-inactivated specimens;
1. Use 300 μl distilled water to dissolve the lyophilized DTT
(b) Cadaver specimens or body fluids other than human
completely in DTT reagent vial, then aliquot into 3 empty tubes by
serum;
100μl each (Eppendof tube, 0.5 ml type is recommended). The
(c) Obvious microbial contamination.
dissolved DTT tube should be sealed and stored at -20°C and can
Use caution when handling patient specimens to prevent
be stable for 2 month. Take it out to room temperature before
cross contamination. Use of disposable pipettes or pipette tips
preparing Displacing Reagent.
is recommended.
2. Pipette 5ml NaOH into the empty displacing reagent bottle, then
Inspect all samples for bubbles. Remove bubbles with an
pipette 100μl DTT solution from DTT tube and add in the displacing
applicator stick prior to analysis. Use a new applicator stick for
reagent bottle. Using the NaOH solution in the bottle to rinse the wall,
each sample to prevent cross contamination.
make sure DTT solution is dissolved with NaOH solution completely.
Serum specimens should be free of fibrin, red blood cells or
Then horizontally shake the bottle in round cycles for mixing,
other particulate matter.
avoided forming bubbles.
Ensure that complete clot formation in serum specimens has
3. Place Displacing Reagent bottle into the 4th position of the kit
taken place prior to centrifugation. Some specimens,
integral. This bottle can be used for about 40 tests. (It is
especially those from patients receiving anticoagulant or
recommended customer collect enough samples and run the tests
thrombolytic therapy, may exhibit increased clotting time. If the
together).
specimen is centrifuged before a complete clotting, the
Note:
099 Vitamin B12-IFU-V3.06-en-EU 2/4
presence of fibrin may cause erroneous results. the microbeads requires mixing to re-suspend microbeads
that have settled during shipment.
Preparation for Analysis For microbeads mixing instructions, refer to the KIT
Patient specimens with a cloudy or turbid appearance must be COMPONENTS, Preparation of the Reagent Integral section
centrifuged prior to testing. Following centrifugation, avoid the of this package insert.
lipid layer (if present) when pipetting the specimen into a To avoid contamination, wear clean gloves when operating a
consistency in the results. Multiple freeze-thaw cycles of refer to the SNIBE service information.
specimens should be avoided.
All samples (patient specimens or controls) should be tested TEST PROCEDURE
within 3 hours of being placed on board the MAGLUMI To ensure proper test performance, strictly adhere to the operating
System. Refer to the SNIBE service for a more detailed instructions of MAGLUMI Fully-auto chemiluminescence
discussion of onboard sample storage constraints. immunoassay (CLIA) analyzer. Each test parameter is identified
Storage via a RFID tag on the Reagent Integral. For further information
If testing will be delayed for more than 8 hours, remove serum
please refer to the operating instructions of MAGLUMI Fully-auto
from the serum separator, red blood cells or clot. Specimens chemiluminescence immunoassay (CLIA) analyzer .
removed from the separator gel, cells or clot may be stored up 100 μl Sample , Calibrator
to 12 hours at 2-8°C. +100 μl Displacing reagent
Specimens can be stored up to 30 days frozen at -20° C or 2 min Incubation
colder. +110 μl ABEI label
Shipping
+110 μl FITC label
Before shipping specimens, it is recommended that +20 μl Magnetic microbeads
specimens be removed from the serum separator, red blood
15 min Incubation
cells or clot. When shipped, specimens must be packaged
400 μl Wash cycle
and labeled in compliance with applicable state, federal and
3 s Measurement
international regulations covering the transport of clinical
specimens and infectious substances. Specimens must be
shipped frozen (dry ice). DILUTION
Sample dilution by analyzer is not available in this reagent kit
WARNING AND PRECAUTIONS FOR USERS Samples with concentrations above the measuring range can be
diluted manually. After manual dilution, multiply the result by the
dilution factor.
For use in IN-VITRO diagnostic procedures only. Please choose applicable diluents or ask SNIBE for advice before
Package insert instructions must be carefully followed. manual dilution must be processed.
Reliability of assay results cannot be guaranteed if there are
any deviations from the instructions in this package insert. QUALITY CONTROL
Observe quality control guidelines for medical laboratories
Safety Precautions Use suitable controls for in-house quality control. Controls
CAUTION: This product requires the handling of human should be run at least once every 24 hours (a run cannot
specimens. exceed 24 hours), once per reagent kit and after every
All samples, biological reagents and materials used in the
calibration. The control intervals should be adapted to each
assay must be considered potentially able to transmit laboratory’s individual requirements. Values obtained should
infectious agents. They should therefore be disposed of in fall within the defined ranges. Each laboratory should
accordance with the prevailing regulations and guidelines of establish guidelines for corrective measures to be taken if
the agencies holding jurisdiction over the laboratory, and the values fall outside the range.
regulations of each country. Disposable materials must be
incinerated; liquid waste must be decontaminated with sodium
hypochlorite at a final concentration of 5% for at least half an
LIMITATIONS OF THE PROCEDURE
hour. Any materials to be reused must be autoclaved using an 1) Limitations
overkill approach. A minimum of one hour at 121°C is usually Assay results should be utilized in conjunction with other clinical and
considered adequate, though the users must check the laboratory data to assist the clinical decision-making for individual
effectiveness of their decontamination cycle by initially patients.
validating it and routinely using biological indicators. A skillful operation and strict adherence to the instructions are
It is recommended that all human sourced materials be
necessary to obtain reliable results.
considered potentially infectious and handled in accordance Procedural directions must be followed exactly and careful operation
with the 29 CFR. 1910.1030 Occupational exposure to must be used to obtain valid results. Any modification of the
bloodborne pathogens. Biosafety Level 2 or other appropriate procedure is likely to alter the results.
biosafety practices should be used for materials that contain Bacterial contamination or repeated freeze-thaw cycles may affect
or are suspected of containing infectious agents. the test results.
This product contains Sodium Azide; this material and its
2) Interfering Substances
container must be disposed of in a safe way. The assay is unaffected by bilirubin<20 mg/dl, haemoglobin<150
Safety data sheets are available on request.
mg/dl or triglycerides< 3000 mg/dl.
3) HAMA
Handling Precautions Patient samples containing human anti-mouse antibodies (HAMA)
Do not use reagent kits beyond the expiration date.
may give falsely elevated or decreased values. Although
Do not mix reagents from different reagent kits.
HAMA-neutralizing agents are added, extremely high HAMA
Prior to loading the Reagent Kit on the system for the first time,
serum concentrations may occasionally influence results.
3) Specificity
The specificity of the VB12 assay system was assessed by
measuring the apparent response of the assay to various
potentially cross reactive analytes.
Compound Concentration Cross reactivity
FA 100 ng/ml 0.2%
4) Recovery
Consider Calibrator High of known concentration as a sample,
dilute it by 1:2 ratio with diluents, and measure the diluted
concentration for 10 times. Then calculate the expected
concentration and recovery of measured concentration. The
recovery should be within 90% -110%.
5) Linearity
Use VB12 calibrator to prepare the six-point standard curve,
measuring all points’ RLU except point A, and then do
four-parameter linear fitting in log-log coordinate, the absolute
linear correlation coefficient(r) should be bigger than 0.9800.
Calibrator Concentration Absolute linear
Point pg/ml correlation coefficient (r)
A 0
B 150 r=0.9910
C 300
D 600