GE Healthcare

Innova™ IGS 530


PRODUCT DATA SHEET
Table of Contents
I Image Chain. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

II Dose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

III Applications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

IV Gantry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

V Patient Table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

VI User interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

VII In room display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

VIII Image management, connectivity and workflow. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

IX Room Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

2 | GE Healthcare
I Image Chain

X-Ray Generator
The IGS system uses a 100 kW high-frequency Jedi three-phase power unit that provides grid pulsed fluoroscopy capability.

Maximum power available 100 kW

Maximum kVp available 125 kVp

Maximum continuous input power 3200W for the tube unit

Pulse frequency 0.5 to 50 fps

Radiographic/recording mode 50 to 125 kVp

Radiographic mA ratings 1 - 1000 mA

Max. continuous power in fluoro mode 3200 W

Fluoroscopy mode 60 - 120 kVp

Fluoroscopy mA ratings 1 – 130 mA (up to 140 mA with PCI ASSIST†)

Fluoroscopic timer Yes

X-Ray Tube
The IGS system is equipped with the high-performance, highly reliable Performix™ 160A metal X-ray tube, which meets requirements for
all vascular applications.

Anode diameter 160 mm brazed graphite

Anode rotation 7800 rpm/ 130 HZ

Anode Target angle 11,25°

Anode heat storage capacity 2.7MJ (3.7 MHU)

Anode steady state heat dissipation 6.72 kW

Cathode Bi-filament design

Coincident focal spot sizes 0.3, 0.6 and 1.0

• 3200 W (continuous)
Fluoroscopic power
• 4500 W (peak capability for maximum of 10 minutes)

Maximum casing heat storage 5.14 MJ (6.9 MHU)

Continuous casing heat dissipation 3200 W

Maximum anode cooling rate 544 KHU/min (6.72 kW)

Total filtration (IEC 60601-1-3) 1.0 mm Al

Leakage radiation (IEC 60601-1-3) <50mR/h measured at 3.2kW (125kv, 25.4mA)

Product Data Sheet | 3


Tube Cooling Unit COOLIX 4100

Cooling type Closed-loop remote water chiller

Maximum Cooling capacity 4100 W

Flow rate 12 l/min

Coolant volume 23 liters in chiller + 17 liters in pipes

Dry weight 120kg+/- 5

Collimation

Number of collimation blades 2 pairs

0.1, 0.2 and 0.3 mm of copper

3 integrated contour filter blades


Spectral filtration

motorized, tapered filter blade that can be rotated 360° as well as translated
in and out using a simple joystick control at tableside

Detector1

Detector manufacturer GE

Size of the detector 31 cm x 31 cm

Material Amorphous silicon photodiode array on a continuous-substrate

Pixel size 200 x 200 µm

Image matrix 1536 x 1536

Bit acquisition 14 bit

Nyquist Frequency 2.5 LP/mm

Anti-scatter Grid: The system is configured with an anti-scatter grid to enhance image quality during routine imaging. Removal of the
grid can improve the X-ray dose efficiency for infants (e.g. less than one-year-old) for field of view (FOV) smaller than 20 cm (7.9 in).

Grid ratio 13:1

Focal distance of the grid 100 cm (39in)

Grid Line frequency 70 LP/cm

Mode @ Dose/Fr

DQE values at average fluoro and record dose operating points

Record, e.g. DSA 175 nGy (20 uR) 84%

1
In clinical use, the results of dose reduction techniques will vary depending on the clinical task, patient size, anatomical location and clinical practice.
Physicians assisted by a physicist as necessary have to determine the appropriate settings for each specific clinical task

4 | GE Healthcare
Fluoro 8.8 nGy (1 uR) 81%

Additional DQE values at minimum fluoro dose operating point

Fluoro 2.2 nGy (0.25 uR) 73%

Note: DQE values given are typical at f = 0 cycles/mm with RQA5 conditions as defined by IEC62220-1-3 standards

Image acquisition
Fluoroscopy modes Non-subtracted, subtracted, roadmap, Blended Roadmap†

Fluoroscopy frame rate 30 fps, 15 fps, 7.5 fps and 3.75* fps

Fluorostore 450 fluororoscopy images (up to 900)

Sub/no Sub simultaneous display† Yes

• DSA (digital subtracted angiography) at 0.5 - 7.5 fps


Angio Acquisition Package
• Multi-segment DSA with flexible frame rate and duration and single shot
capabilities

Dynamic Acquisition Package 30 fps and 15 fps

Innova Chase acquisition 5 fps

Digital output 1024 x 1024

Field-of-view adjustment from tableside with


30 cm, 20 cm, 16 cm, and 12 cm
four magnification selections (records mode)

Image flip capability Horizontal and vertical image flip capability for all acquisition

Shutter Automated electronic shutter matched to collimated portion of image

Integrated X-ray dose tracking and in-room display of air kerma and dose area product

A configurable beep is activated when using the fluoro mode

*: non-subtracted Fluoro mode only

Image Processing and Review


Immediate auto-review of acquisition

Next and prior sequences or images

Slow and fast review of sequences, forward and reverse

Pause, adjust brightness and contrast during review

Image review with or without edge enhancement filters

Mask select, pixel shift

Store/recall reference images

Product Data Sheet | 5


II Dose

Dose Reduction Dose Awareness


Dose Personalization Integrated dose monitoring
Several image-quality and dose strategies are available and can The user can monitor air kerma rate, integrated air kerma over
be customized for the various clinical protocols in both fluoro and the exam, and the total dose area product received by the patient
record acquisitions . during a procedure. The threshold of cumulated dose displayed
on a gauge icon is customizable to warn operator when such
InnovaSense† threshold has been reached. The threshold is customizable
InnovaSense is an advanced patient contouring technology depending on the protocol.
that uses an intelligent algorithm during gantry motion to
select the optimal position for the image receptor relative Dose reporting
to the patient . y reducing the distance from receptor to The system provides DICOM** compatible Radiation Dose
patient, the system optimizes imaging geometry and helps Structured Report allowing the export of the dose and related
reduce radiation exposure . The user also can position both the acquisition parameters.
gantry and detector with one integrated operation . Capacitive
sensor technology and optimized collision avoidance software Dose Map†
enable a speed of pivot and C-arm, of up to 20° Dose Map is a feature used to calculate, display and record an
per second . estimated local cumulated dose during procedures done on
the GE X-Ray angiographic system. It is designed to provide to the
user a visualization of the distribution of the local cumulated
Dose Efficiency dose throughout the exam as well as the current projection
Dynamic exposure optimization – AutoEx of the beam.
A neural network technology allows advanced exposure
management algorithms to dynamically control X-ray technique Estimated local
Yes
and beam filtration . he user can select system settings, from cumulated dose
maximizing image quality to minimizing skin dose .
Pediatric patient Yes
Based on those settings, the system automatically optimizes
in real time the contrast-to-noise ratio within the image by
selecting the optimal technique factors and spectral filter Real-time calculation Yes
thickness . pecific filtration of 0 .1, 0 .2, and 0 .3 mm Cu is
incorporated in the collimator . Configurable alert Yes

6 | GE Healthcare
.

StentViz†
InnovaChase™
The StentViz option enhances visibility of the stent structure .
InnovaChase™ is a dynamic, unsubtracted acquisition at a fixed It is particularly useful in verifying placement and deployment of
frame rate of 5 fps with manual and remote panning of the table . stents during coronary interventions where moving arteries could
It is optimized for visualization of a run off . make visibility challenging . StentViz† processing is fully
automated and can be launched at the press of a button on the
Innova Central touch screen . The result is automatically displayed
on the reference monitor and shows two zoomed and enhanced
images of the stent: One with the guidewire in view and a second
one where the guidewire is subtracted out in the area between
the two balloon markers to allow better visualization of the stent
struts or borders .

PCI ASSIST†
PCI ASSIST† is an ASSIST package containing several applications
including High Contrast Fluoro that increases the mA peak up to
36%, and decreases the pulse width by 38% . While the dose is
equivalent, it is delivered in an efficient way that helps signifi-
cantly reduce the blurring in the image due to organ motion .

Product Data Sheet | 7


InnovaBreeze™ peripheral angiography† Blended Roadmap†
InnovaBreeze† lets the user follow the contrast using variable Blended Roadmap† is a vascular road mapping application that
panning speed control in the control room while looking at superimposes a previously acquired vascular image over live
subtracted images in real time . InnovaBreeze† includes Advantage fluoroscopy. This advanced application helps clinicians view the
Paste progression of guidewires and devices through the vessels.
Clinicians can select any DSA or bolus image as a reference
roadmap image. By using this image multiple times, Blended
Roadmap has the potential to minimize contrast media injections
during road mapping.
Blended Roadmap† provides additional features to enhance
road mapping procedures, including:
• Adjustment of the subtraction level
• Adjustment of vessel transparency
• Automatic resizing of the roadmap image to adapt to the
fluoroscopic field-of-view
• Pixel shift of the vessel image to compensate for motion
• Blended Roadmap† is available on systems with either the
Omega V or InnovaIQ and OR tables. The Innova Central touch
screen is recommended for table side control.

Advantage Paste*
Provides the ability to reconstruct and visualize the entire length
of the subtracted bolus chasing acquisition on a single image .

InnovaSpin™†
The offset C-arm permits fast-spin rotational angiography
over a maximum range of 200° at variable speeds from 20°
to 40°/sec (maximum speed of 30°/s from LAO to RAO) with flexible
cranio/caudal oblique angulations . The enhanced nnovaSpin™
trajectories are not constrained to a single transverse plane and
can be used at oblique angulations
within physical constraints . A total of seven trajectories can
be preset (limited to five on a tilting table) . The entire workflow for
the test run and spin acquisition can take place at tableside . The
spin acquisitions can be performed in the cardiac record mode for
coronaries, or peripheral image quality can be optimized in the
InnovaChase™ mode .

8 | GE Healthcare
.

Innova 3D Innova CT

Spin duration 5 sec. 5, 10, 20sec.

Frame rate 30 fps 30 fps

Reconstruction
5 to 20 seconds 5 to 20 seconds
time

Reconstructed 512x512x512 512x512x512


3D model
resolution 256x256x256 256x256x256

Product Data Sheet | 9


IV Gantry
Positioner specifications

L-arm rotation on vertical axis ±100° (±95° motorized limit set)

C-arm angulation 50° cranial and 45° caudal

Combination of movements of the C-arm and L-arm Permits -/+55° cranial and caudal angulations

C-arm angulations speed 0 to 15° (Up to 20°/sec with InnovaSense†)

Offset C-arm -117°/+105° RAO/LAO rotations

107 cm (42 in) with L-arm at 0° provides femoral coverage on


Offset C-arm throat depth
most patients without rotating the L-arm

Fully motorized SID 8.9 cm/s (3.5 in/s)

SID range 89 cm to 119 cm (35 in to 47 in)

Tube Focal Spot distance to Isocenter (SOD) 72 cm (28 in)

Isocenter to floor distance 107 cm (42 in)

• Anatomical mode to hold the view while moving the L-arm to


an optimum patient access position.
Positioning modes
• Mechanical angulation allows movement of any one of the
three axes independently for maximum positioning flexibility
and vessel profiling

Support Floor-mounted

Weight ~ 705 kg

10 | GE Healthcare
V Patient Table
Omega V

Maximum total load 304 kg (670 lbs.)

Maximum patient weight 204 Kg (450 lbs.)

100kg (220 lbs.) while complying with the following -maximum weight
requirements:

Maximum accessories weight


• 40 kg (88lbs.) on each of the two fixed side rails

• 20 kg (44 lbs.) of accessories on the table foot-end rail (option)

CPR 50 kg (110 lbs.) additional maximum load

Tabletop absorption Less than 0 .8 mm Al Equivalence, 100 KVp

Tabletop Material Radio translucent carbon fiber tabletop

Tabletop length 333 cm (131”)

Tabletop width 46 cm (18”) in patient trunk area

Horizontal Float Movement 8-way

Longitudinal travel Up to 170 cm (67”)

Longitudinal Speed 15cm/s

Imaging coverage with table panning Up to 195 cm (76’’)

Transverse Travel ± 14 cm (± 5.5”)

Vertical travel above floor From 78 cm (30.7”) to 108 cm (42.7”)

2 cm/s (0.8”/s) at 50Hz.


Vertical Speed
2.5 cm/s (1”/s) at 60 Hz

Available on the base section of the tabletop to mount tableside controls


Accessories rails
and IV pole

Table weight ~ 590 kg

2
Typical values

Product Data Sheet | 11


InnovaIQ Table and OR Table

Maximum total load 320 kg (705 lbs.)

Maximum patient weight 204 kg (450 lbs.)

116kg (255 lbs.) while complying with the following -maximum weight
requirements:

• 40 kg (88 lbs.) on each of the two fixed side rails


Maximum accessories weight
• 20 kg (44 lbs.) on each of the two removable side rails

• 16 kg (35 lbs.) for other miscellaneous components/accessories

50 kg (110 lbs.) additional maximum load


CPR
Position for CPR accessible using a dedicated auto-positioner memory
position

Imaging coverage with table panning Up to 195 cm (76 in)

Tabletop absorption Less than 0.85 mm Al Equivalence, 100 KVp

Tabletop material Radio translucent carbon fiber tabletop

Tabletop length 333 cm (131in)

Tabletop width 46 cm (18 in) in the patient trunk area

Longitudinal travel Up to 170 cm (67’’)

Longitudinal speed Up to 25 cm/s

• +/-14 cm (5.5’’) in manual mode


Transverse travel
• +/-13 cm (5.1’’) with motorized panning

• Without tilt: From 78 cm (30.7”) to 108 cm (42.7”)


Vertical travel
• With tilt: From 80 cm (31.5’’) to 137 cm (53.9’’)

Vertical speed Up to 2.5 cm/s (1’’/s)

Tilting angles 20° Trendelenburg (he ad down), 12° reverse Trendelenburg (head up)2

Tilting speed Up to 2°/sec

• Isocenter tracking mode to help maintain the region of interest centered


during the tilt
Tilting modes
• Incidence-keeping mode: The gantry moves in synchronization with the
table tilt. In this mode, clinicians can also pan the table to follow the
contrast media

Compatibility The tilt feature is compatible with InnovaBreeze peripheral angiographic


imaging

2
Typical values

12 | GE Healthcare
InnovaIQ Table and OR Table

5 cm (2 in) mattress
Mattress is a sandwich of 2 Visco Elastic foams of 1” thickness each, top layer
of which with density of 85Kg/m3 (provides support and envelops quickly and
Patient mattress
efficiently adapting to patient body shape) is laminated to a lower layer with
density of 110Kg/m3 (capable of redistributing loading forces), with
X-Ray absorption of < 1 .2mm Al Equivalence @ 3 .6HVL, 100K

Table weight ~ 580 kg

The OR Table is IPX 4 certified

Memorizes the table and gantry position simultaneously or separately. Table


Auto-positioning features auto-positioning includes longitudinal, lateral, vertical and tilt movements of
the tabletop.

Auto-positioning control From tableside or from the control room†

Includes connectors for injectors, two potential equalization outlets, gantry


Interconnect panel
and tableside status control

Available on the base section of the tabletop to mount tableside controls


Accessories rails
and IV pole

Table accessories for InnovaIQ and OR tables


Dedicated shoulder and foot rests provide optimal patient comfort when
Tilt accessories†
using maximum tilt

Malleable screen holder that helps separate the patient’s head from the
Anesthesia screen holder†
surgical site during procedures

Standard size table rails4 that can be adapted on the trunk section of the
Removable rails3
table to accommodate tableside user interface or comfort accessories

Arm boards with cushions† Arm boards with thick pads that can be mounted on the removable rails4

Patient comfort accessories† • Table width extenders

• Patient restraint straps

• Head support widener

Sockets† • Universal clamp

• Easy Lock socket

TSUI adaptor rail† Allows attachment of TSUI on removable rails

To be adapted on the head section of the table; can handle up to 15 kg


Rail extenders†
(33 lbs.) of accessories on each of the two side rails.

Accessories cart† For accessories storage

3
Available as an option with the InnovaIQ Table, included with the OR Table
4
Available in US (0.4 x 1.1 in) and European (10 x 25 mm) standard sizes

Product Data Sheet | 13


VI User interface
In the examination room

Table and gantry controls

SmartBox or SmartHandle (A second SmartBox or SmartHandle can be added at tableside)

The Tableside Status Control (TSSC) provides simple access to key acquisition and review parameters throughout the exam.
A second TSSC can be added at tableside or in the control room. TSSC controls: Acquisition preference settings, Fluoro level, Fields-
of-view, Subtracted/non-subtracted fluoro, Fluoro landscape, Room light on/off, Fluoro timer reset, Fluorostore and Auto Positioner.

The Innova Central touch screen provides safe and simple access to key features throughout the exam. It lets the user control the
system functions as well as integrated equipment. Central Touch Screen controls: Image acquisition, image review, dose settings,
Mac-Lab™ hemodynamic recording systems†, CardioLab™ EP recording systems†, AW advanced applications†, Large Display Monitor†
layout management, Favorite tab where functionalities can be grouped based on user preferences.

Smart Nav is an innovative solution to control some IGS system functionalities from tableside and from the control room. It allows
fast function access in displaying menu controls on the reference monitor upon user request. With Smart Nav, the user can keep
his/her attention on the screen monitors where clinical images are also displayed. Smart Nav is controlled from the Central Touch
Screen, local keypad or remote keypad, providing intuitive and context-based navigation. Smart Nav controls: In-room Browser,
Review and processing tools, Review sequence/photo, Stopwatch (play, pause, stop, reset), Dose Map†, StentViz†, StentVesselViz†

In the control room

A remote stand is provided to use the second tableside user interface† in the control room

The IGS system includes one B&W 48 cm (19 in) LCD monitor for display of live images in the control room.
Additional reference and subtracted roadmap repeater monitors† are available

Dedicated keypad for convenient control of commonly used review functions provides an image shuttle knob
to control playback and one-touch access to image review functions

Pause, adjust brightness and contrast during review

Flat graphic display with easy point-and-click mouse control supports patient management and
advanced processing and analysis features

Keyboard enables patient data entry

14 | GE Healthcare
VII In room display
IGS system can be installed with LCD monitors as first display or GE LDM monitor.

48 cm (19in) live and reference imaging monitors


RX 150

Diagonal 48cm (19in)

Active Display 376 x 301 mm

Display matrix 1280 x 1024

800 cd/m² (typical)


Brightness
400 cd/m² (calibrated)

Viewing angles 176°, 176° (typical)

LUT 10 Bit

Frequency 60 – 75 Hz

Contrast ratio 1000:1 (typical)

Greyscale Tones 1024

Power consumption 65 W

AW color monitor†
MX 193

Diagonal 48cm (19in)

Active Display 376 x 301 mm

Display matrix 1280 x 1024

Brightness 330 cd/m² (typical)

Viewing angles 178°, 178° (typical)

Frequency 31-64 Hz

Contrast ratio 900:1 (typical)

Power consumption < 31W

Product Data Sheet | 15


GE Large Display Monitor†
The GE Large Display Monitor† (LDM) is an in-room primary monitor designed to streamline procedure workflow. It includes a video server
solution and is fully integrated with the Central Touch Screen at tableside.

GE Large Display Monitor specifications

Diagonal 148 cm (58in)

Active Display 127 x 72 cm

8 megapixels
Display matrix
3840 x 2160-pixel array

560 cd/m² (typical)


Brightness
400 cd/m² (minimum native white level)

176°, 176° (typical)


Viewing angles
160°, 160° (minimum at CR>=20)

LUT 11 Bit

Frequency 59.7 – 60.3 Hz

Contrast ratio 3500:1 (minimum), 5000:1 (typical)

• 10 video inputs for Live, Reference, AW and optional subtracted Fluoro monochrome
signals as well as for a wide variety of other video signals usually used in an
Video inputs
interventional environment - including 3 free open inputs compatible with VGA
and DVI video formats.

• up to 200

Layouts • Organized into user or application groups

• Digital zoom (up to 200%)

User interface Layouts are selectable from the Central Touch Screen

48 cm (19 in) live and reference monitors attached at the back of the LDM or on another
Back-up monitors
suspension†

Additional HD output† For HD video compatible solutions (second 8MP monitor, 2MP HD monitor, recorder…)

An optional kit to interface a third-party suspension according to GE Healthcare specifications allows users to install another third-party
suspension instead of the standard GE Healthcare monitor suspension offering.

16 | GE Healthcare
VIII Image management, connectivity and workflow

Record images stored in 8 bits, maximum 450 images per sequence. Storage capacity: 136,000 record images

DSA images with 12 bits data stored in 16 bits, maximum 450 images per sequence. Storage capacity: 68,000 DSA images

DICOM** image output on 100Mbit Ethernet with Autosend and background transfer for fast transmission
with minimal user interaction.

Patient Worklist capability provides a single point of entry of patient data, increasing staff productivity and
eliminating clerical errors: patient information can easily be imported into the digital system from information systems
that support DICOM** Worklist Service Class Provider.

Multi-destination Push enables images to be sent to multiple remote DICOM** destinations sequentially
(one after another). Multi-destination helps to support a clinical scenario of handling post processing and archival activities
in multiple destinations independently of each other (workstation, PACS). Multi-destination provides a seamless integration
of the system into clinical workflow.

MPPS: Modality Performed Procedure Step allows the IGS to share the main exam parameters with the hospital information system.

For the Innova 3D option, users can direct-push the 3D acquisition directly to the pre-configured AW,
even if the images of the exam are pushed to a PACS or another archiving system.

For further information about DICOM** conformance statement:


http://www3.gehealthcare.com/en/products/interoperability/dicom/x-ray_and_mammography_dicom_conformance_statements

Product Data Sheet | 17


IX Room Requirements

Load distribution for the gantry 2660 kg/m²

Load distribution for the Omega Table 2410 kg/m²

Load distribution for the IQ/OR Table 2000 kg/m²

Minimum ceiling height 2.74 m (9 ft)

Minimum Room dimensions with Omega IV table


5.47 m (17 ft 11 in) length & 4.4 m (14 ft 5 in) width
(without accessories)

Minimum Room dimensions with Omega V table


6.04 m (19 ft 10 in) length & 4.4 m (14 ft 5 in) width
(without accessories)

Minimum Room dimensions with InnovaIQ or OR table


6.14 m (20 ft 2 in) length & 4.4 m (14 ft 5 in) width
(without accessories)

Humidity 30 to 70 %

Range of temperature 15 to 32 °C in exam room

Atmospheric pressure 79.4 to 106 kPa

System Power
Nominal 380 to 480 Volts AC, three phase 50 or 60 Hz without neutral.

Maximum momentary power demand: 150 kVA

Emergency power supply


In case of power outage, a 20kVA Uninterruptible Power Supply (UPS)† provides vital minutes of emergency table control and fluoroscopy
to position the table in an optimal position and continue critical aspects of a case in progress, should power be interrupted.

18 | GE Healthcare
Typical room layout

Customer service
Advanced remote connectivity allows GE to monitor systems and intervene if necessary.

Serviceability
The Digital System Manager simplifies troubleshooting and minimizes downtime with built-in equipment error logging and power-up
diagnostics in real time. Resident software monitors the entire system, including peripheral hardware. The IGS system features 24-hour
InSite™ remote service diagnostics and repair. InSite™ service is available to systems covered by the original warranty or by a GE service
contract (broadband required).

Extended service†
An optional full-service contract ensures uptime even after the original warranty expires and provides advanced remote diagnostics
through a broadband or phone connection.

Product Data Sheet | 19


Product may not be available in all countries and regions.
Full product technical specification is available upon request.
Contact a GE Healthcare Representative for more information.
Please visit www.gehealthcare.com/promotional-locations.
Data subject to change.
© 2017 General Electric Company. DOC1629995
GE, the GE Monogram, imagination at work are trademarks of General
Electric Company.
† Option
* Requires AW.
**DICOM is a registered trademark of National Electrical Manufacturers
Association, All third party trademarks are the property of their respective
owners.
Reproduction in any form is forbidden without prior written permission
from GE. Nothing in this material should be used to diagnose or treat any
disease or condition. Readers must consult a healthcare professional.

GEA33028 3/2017

You might also like