Untitled
Untitled
Untitled
Manufactured for:
Sakura Finetek USA, Inc., Torrance, CA 90501 U.S.A.
Sakura Finetek Japan Co., Ltd., Tokyo, 135-0007, Japan
Sakura Finetek Europe B.V., 2804 AV Alphen aan den Rijn, NL
Made in U.S.A.
1320109-01 Rev C
TABLE OF CONTENTS
Section Page
1. INTRODUCTION 1.1
Safety Instructions ........................................................................................................................................ 1.1
Safety Precautions ....................................................................................................................................... 1.1
General Description ...................................................................................................................................... 1.2
Physical Characteristics................................................................................................................................ 1.3
User Interface ............................................................................................................................................... 1.5
Safety Standards .......................................................................................................................................... 1.7
2. INSTALLATION 2.1
General Information ...................................................................................................................................... 2.1
Environmental Factors .................................................................................................................................. 2.1
Unpacking..................................................................................................................................................... 2.1
Positioning the Instrument ............................................................................................................................ 2.3
5. ACCESSORIES 5.1
Standard Accessories ................................................................................................................................... 5.1
Optional Accessories .................................................................................................................................... 5.3
7. TROUBLESHOOTING 7.1
General Information ...................................................................................................................................... 7.1
Revised 3/30/2010 i
TABLE OF CONTENTS
ii Revised 3/30/2010
Section 1
INTRODUCTION
Safety Instructions Safety Precautions
The Tissue-Tek® Prisma® Automated Slide Stainer CAUTIONS and NOTES are provided throughout this
operating manual includes important instructions and manual to indicate levels of potential hazards or helpful
information related to the operating safety and information as defined below.
maintenance of the instrument. It is important to read the
CAUTION: Indicates a potential hazard in which failure
operating manual carefully prior to the operation of the
to follow instructions may result in damage to the
instrument for the first time and to keep the manual with
Prisma and/or other property.
the instrument at all times.
This instrument has been built and tested with the NOTE: Indicates a reminder or other helpful information.
following safety regulations on electrical measuring
control, regulating and laboratory devices: Safety Features
nd
IEC 61010-1 2 Ed.
nd
The Tissue-Tek Prisma Slide Stainer is equipped with
CAN/CSA C22.2 No. 61010-1 2 Ed. several safety features, which help to keep the operator
nd
UL 61010-1 2 Ed. and the instrument safe from harm.
In order to ensure safe operation, the operator must • The robotic arm will stop if the cover is opened.
always comply with the instructions and warnings • The robotic arm does not make access to the stations
contained in the operating manual. in the front row when the door is opened.
• Heater circuit is blocked off when the drying unit
CAUTION: The use of iodine on this instrument may
reaches an abnormal temperature (85ºC).
cause oxidation on some parts. It is recommended
• Heater circuit is blocked when the heating stations
that if iodine must be used in some staining methods,
reach an abnormal temperature (90ºC).
that it be removed immediately upon completion of
• Heater circuit is blocked off when the fan of the drying
the stain. Failure to remove the iodine may cause
unit abnormally stops.
oxidation.
• Heater circuit of the drying unit is blocked off when
CAUTION: Power supply plug is used as mains power the fan of the fume control system abnormally stops.
disconnecting device. Locate unit where power • Electrical circuit is blocked off when excessive current
supply is easily accessible. flows into the instrument.
NOTE: The protective devices installed on both the
instrument and accessories may neither be removed nor
modified. Only authorized and qualified persons may
access and repair the internal components of the
instrument.
The safety devices installed in this instrument by the
manufacturer only establish the basis for accident
prevention. Primary responsibility for accident-free
operation lies, above all, with the institution that owns the
instrument and, in addition, the designated personnel that
operate, service, or repair the instrument.
Figure 1-A
Physical Characteristics The special stain reservoirs can be combined with the
standard or expanded configurations. The special stain
configuration (Figure 1-D) includes eight solution
Solution Reservoir Area reservoirs in a special removable reservoir tray and two
heated solution reservoirs (only available on product #
The solution reservoirs are arranged in a single-level 6131). The special stain reservoirs accommodate a
configuration and are positioned in reservoir trays. These solution volume of 160 to 180 ml.
trays can hold multiple solution reservoirs to allow for easy
placement into and removal from the instrument. The trays
are designed to ensure proper orientation when reservoirs
are placed into and, when the tray is set into the
instrument.
There are three different solution reservoirs: standard,
expanded, and special stain.
The standard configuration (Figure 1-B) contains a total
of 30 stations with six stations in the front row and eight Figure 1-D
stations each in three other rows. The standard reservoir
Also included in the solution reservoirs are the
accommodates 680-820 ml of solution.
start/end stations, wash stations, and dryer stations.
Start/End Stations - the start/end stations are located
in the front row of the reagent reservoirs. These reservoirs
have a handle for easy removal and can be programmed
as 1 to 3 start stations and up to 5 end stations. The
reservoir volume is 830 ml.
Wash Stations - the wash stations are located in the
right rear of the instrument. A maximum of four and a
Figure 1-B minimum of zero can be used at one time, depending on
The expanded configuration (Figure 1-C) contains a the setup. The reservoir volume is 850 ml.
total of 54 stations and is arranged with 6 stations in the Drying Stations - There are a total of two drying
front row and eight stations in the other six rows. The stations programmable from 30º C (86 º F) to 65º C
expanded reservoir can accommodate 255 to 285 ml of (149ºF)
solution.
Figure 1-C
Figure 1-G
Figure 1-E
Fume Control
The activated carbon filters (Figure 1-F) are part of an
effective fume control system. As air circulates throughout
the inside of the instrument, fumes generated by various
solutions are captured and extracted by the carbon filter.
The fume control housing can be accessed at the back,
inside of the instrument. It is recommended that the
carbon filters be replaced after two weeks with routine
operation of the stainer.
Figure 1-F
Figure 1-J
Figure 1-I
Figure 1-K
Figure 1-L
Figure 1-M
Specifications Dimensions
Models Centimeters: 125 (W) x 71.3 (D) x 63 (H) cm
Inches: 49.3 (W) x 28 (D) x 24.8 (H) in.
Model Name/Description
Weight: 150 kg (330 lbs.)
Tissue-Tek® Prisma® Automated Slide Stainer (115 VAC)
6130
(USA)
Operating Conditions
Tissue-Tek® Prisma® Automated Slide Stainer with Special
6131
Stains (115 VAC) (USA) Operating Temperature: 10º C to 40º C (50º F to 104º F)
Tissue-Tek® Prisma® Automated Slide Stainer (230 VAC) Relative Humidity: 30-85% (non-condensing)
6132
(Europe)
Atmospheric Pressure: 80 to 106 kPa
Tissue-Tek® Prisma® Automated Slide Stainer with Special
6133
Stains (230 VAC) (Europe)
Storage Conditions
Tissue-Tek® Prisma® Automated Slide Stainer (100 VAC)
6120 Storage Temperature: -20º C to 65º C (-4ºF to 149º F)
(Asia)
Tissue-Tek® Prisma®
Automated Slide Stainer with Special Relative Humidity: 20-90% (non-condensing)
6121
Stains (100 VAC) (Asia)
Atmospheric Pressure: 50 to 106 kPa
Safety Standards
Complies with IEC 61010-1 2nd ed., CAN/CSA C22.2 No.
61010-1, UL 61010-1
CE registered in compliance with the requirements of
EMC Directive, LV Directive in Europe, In Vitro Diagnostic
Device Directive.
Capacity:
Standard Reservoirs: 680-820 ml
Expanded Reservoirs: 255-285 ml
Wash Reservoirs: 850 ml (volume up to the
overflow level)
Start/End Reservoirs: 690-830 ml
Special Stain Reservoir: 160-180 ml
Heater Solution reservoir*: 240-250 ml
Solution Names
Up to 100 solution names (20 characters per name)
can be stored in memory.
Solution Configurations
Up to 50 solution configurations can be stored in
memory.
INSTALLATION
General Information
This section provides detailed installation and setup
instructions for the Tissue-Tek® Prisma® Automated Slide
Stainer. The installation steps must be followed correctly
to ensure proper operation and service. Read this
operating manual carefully before attempting to operate
the instrument. Follow all instructions carefully.
The Tissue-Tek Prisma is a precision instrument and
must be handled accordingly. Rough handling or dropping
the instrument will disturb or damage internal components.
Always handle the instrument with care.
Select a place where sufficient clearance can be
provided around the instrument. Ample working space is
necessary in front and on top of the instrument. Figure 2-A
2. Remove the lid of the box (Figure 2-B).
Environmental Factors
As with all sensitive electronic instruments, prolonged
exposure to excessive humidity and temperature should
be avoided. Temperature and humidity should be held
relatively constant. The ambient temperature for operating
the instrument is 10º C to 40º C (50º F to 104º F). The
ambient operating humidity range is 30-85% relative
humidity.
Unpacking
Figure 2-C
7. Remove all tape and packing foam from the outside
of the instrument.
8. Open the main cover on the instrument and locate the
accessory box positioned inside.
9. Cut the tie-wraps holding the accessory box. Remove
the tie-wraps carefully from around the support post.
10. Remove the accessory box from inside the instrument
(Figure 2-D).
Figure 2-E
14. Remove the packing and adhesive tape from the
display.
15. Remove the power cord from the plastic bag.
NOTE: Be certain that all tape and foam packing material
is removed from the instrument prior to operation.
Figure 2-D
Unpacking Check List Optional Accessories (Link System)-For use with the
When opening the accessory box, confirm that all Tissue-Tek Film Automated Coverslipper. Tissue-Tek Link
accessories have been included with the instrument. System (1)-6134 consists of:
r Operating Manual (1) r Installation Bar (1)
r Warranty Card (1)-located inside the operating r Bolts, SW, W (5 Each)
manual (USA only) r Communication Data Cable (1)
r Slide Baskets (10) r Compact Flash (program for connection) (1)
r Slide Basket Adapter (10) If any of these items are missing, contact the Technical
r 820 ml Solution Reservoir (22) Support Department at Sakura Finetek USA. In countries
r 830 ml Load/Unload Reservoir (6) outside the USA, contact the nearest Sakura distributor or
representative.
r 850 ml Wash Reservoir (4)
r 3-position Reservoir Lid (3)
r 4-position Reservoir Lid (3) Positioning the Instrument
r 3-position Reservoir Tray (3) Locate the instrument in a well-ventilated area,
r 4-position Reservoir Tray (3) avoiding exposure to corrosive vapors and extreme
r Activated Carbon Filters (2/cs) variations in temperature or humidity. The area should be
clean, dust free, and have a firm, level surface capable of
r Compact Flash Card (1) holding at least 150 kg (330 lbs.) of weight. Be sure it is
r Control Panel Protection Sheet (installed) (1) near a power source that meets the electrical
r Drain Hose (1) requirements specified on the rating label located on the
r Drain Hose Clamp (1) right side of the instrument. The power receptacle must be
grounded and should be a dedicated line. Avoid proximity
r Drain Elbow (1)
to direct sunlight, open windows, ovens, hot plates, open
r Water Supply Hose Set (1) burners, or radiators.
r Station Labels (1)
r Wash Station Plug (4) (4 installed) Leveling the Instrument
r Single Reservoir Lid (6)
Once the instrument is placed on a sturdy, level
surface, it is important that it is leveled.
Optional Accessories –Special Stain Set (If Product #
Using a wrench, adjust the feet by turning the nut on
6131 was ordered)
each foot. Turning the nut clockwise will raise the
r 10-slide Basket (10)-6137 instrument, counter-clockwise will lower the instrument.
r 10-slide Basket Adapter (10)-6138 Place a level on the inside, middle of the instrument
r 10-slide Basket Load/Unload Adapter (3)-6139 and adjust the feet until the instrument is level.
r 160 ml Special Stain Solution Reservoir (8)-6140 NOTE: Being certain that the instrument is level will
r Heated Solution Reservoir (2)-6141 ensure proper water drainage and overall instrument
r Special Stain Reservoir Lid (8)-6144 operation.
r Special Stain Reservoir Tray (1)-6156
Figure 2-H
2. Insert the elbow fitting.
3. Place the drain hose assembly onto the elbow fitting
Figure 2-F and tighten the connecting hose clamp to ensure that
4. Lower the filter cover. it is properly seated.
5. Remove the plastic wrapping from both carbon filters. 4. Attach the drain hose to the drain hose assembly. Be
sure that the clamped end of the hose attaches to the
6. Place the carbon filters side by side in a horizontal elbow. Tighten the clamp by using a Phillips
position in the appropriate location (Figure 2-G).
screwdriver. The other end of the drain hose should
be routed to a floor or sink drain that is capable of
draining at a rate of 15 liters (4 gallons) per minute.
CAUTION: Make sure that the entire length of the
drain hose is lower than the drain port. It should be
straight and not twisted or looped. Do not extend the
drain hose by coupling it with another hose. The end
of the drain hose should be at least 4 inches (100 mm)
or higher from the water surface in the floor or sink
drain.
Figure 2-G
Installing the Water Supply Hose 3. Grasp the reservoir by the handle and insert it into
one of six positions located in the front row of the
1. Connect the pre-assembled water supply hose to the
solution reservoirs (Figure 2-J).
water supply port located at the bottom, right of the
instrument (Figure 2-I).
Figure 2-J
4. Close the front door.
Dryer Stations
Figure 2-I
CAUTION: Make sure that the instrument is switched
2. Connect the other end of the hose to a faucet or off and the power cord is unplugged before
water source. beginning.
NOTE: Use plumbers tape to be certain that the water 1. Remove the 2-piece drying station assembly from the
supply hose is securely connected to the instrument. accessory box.
CAUTION: When installing the water supply hose, 2. Place the large piece inside the drying station (Figure
make sure it is securely connected to the instrument 2-K) using the diagram provided near the drying
and to the faucet. Failure to do so may allow the water station as a guide.
pressure to loosen the connection and cause an
overflow.
CAUTION: It is recommended that the water supply to
the instrument be turned off when the laboratory is
left unattended, i.e. overnight or weekends.
Figure 2-L
4. Repeat the procedure in the second drying station.
Figure 2-N
Wash Stations
2. Replace the wash nozzle with a wash station plug
The wash stations are positioned in the right rear of the and secure it in place using the wash cap.
solution reservoirs. A maximum of four and a minimum of
3. Place a solution reservoir at that position.
zero wash stations may be used at one time. The factory
default setting is set at four wash stations. The solution To transform a solution station to a wash station:
configuration may need to be set up before a decision is
made to make changes to remove the factory set wash 1. Turn the wash cap counter-clockwise and remove the
stations. wash cap, and gasket.
1. Position the wash stations with the opening in the rear 2. Replace the wash station nozzle and secure all items
of the reservoir toward the back (Figure 2-M). This will in place.
allow for the water to overflow from the reservoir into 3. Place a wash station at this position.
the drain. NOTE: Verify that the station configuration reflects the
proper wash positions.
Figure 2-M
Figure 2-O
Solution Reservoirs
The Prisma comes with six solution reservoir trays. The
trays can be easily removed to fill and position the solution
reservoirs.
1. Grasp a reservoir tray by the handles located on each Figure 2-Q
side and remove the tray by lifting it straight out
(Figure 2-P).
Figure 2-P
Figure 2-R
Figure 2-S
3. Press flash card in fully.
2. Place insert into start stations (Figure 2-T).
4. Power instrument on with the power switch.
Figure 2-T
NOTE: The instrument will automatically detect that only
one 20-slide holder is being placed into the start station.
Setting up the Link Connection (Optional) NOTE: If three-arm basked adapter is not removed, the
basket holder may become misaligned resulting in a
A data cable is part of the optional link connection that robotic arm malfunction.
connects the Prisma to the Tissue-Tek Film Coverslipper.
1. Connect one end of the data cable to the link inlet
located on the lower left of the instrument, and the
other end to the rear of the Film Coverslipper.
NOTE: Optional link station must be installed by an
authorized Sakura Representative.
Figure 2-W
NOTE: The insert must be placed into the start stations if
a 10-slide basket is used. The instrument will
automatically sense that a 10-slide basket is being placed
into the start station.
NOTE: It is important to remove the special stain
reservoir tray if running other configurations that do not
include those reservoirs. If left in place, the other
configurations will not show up.
CUSTOMIZATION OF SETTINGS
General Information
The Tissue-Tek® Prisma® Automated Slide Stainer
allows for the customization of a number of system
options and the configuration of certain operating
parameters to suit the specific requirements of a given
installation. The Prisma Menu provides the means of
viewing and modifying current system settings.
This section outlines the procedures to perform initial
setup of Tissue-Tek Prisma system settings. Factory
default settings for each parameter are listed where
applicable.
Perform the following setup procedures prior to initial
use of the system. In addition, perform these procedures
as needed to reset system settings, add or remove system
users, etc. 3. Touch the Menu button to display the main menu.
To change User ID
1. From the Administrator/User list on the left of the
screen, select the user or administrator for whom you
want to create a user id then touch the User’s ID
button.
• The Create User’s ID screen displays.
Stain Process – sets permissions for Staining Edit – sets permissions for Edit Program,
Mode Selection, Abort Stain Process, Reset Program Check, Edit Solution Configurations and
Solution Usage Information, and Solutions Solutions Name and Delete Data. The default
Configuration Selection. The default setting for setting for each function is Allowed. Touch each
each function is Allowed. Touch each function as function as necessary to change permissions
necessary to change permissions from Allowed from allowed to Not Allowed. Touch Save to save
to Not Allowed. Touch Save to save settings settings and return to the Password Setup screen
and return to the Password Setup screen or or touch exit to return without saving changes.
touch exit to return without saving changes.
Utility – sets permissions for System Setup,
Export Data, Language Selection, Water Flow.
The default setting for each function is Allowed.
Touch each function as necessary to change
permissions from allowed to Not Allowed. Touch
Save to save settings and return to the Password
Setup screen or touch exit to return without
saving changes.
4. Use the touch keyboard to enter an abbreviated name 2. Touch the Add Configuration button.
for the new solution up to eight characters long. This • The Edit Solution Configuration Screen displays.
name is use for on-screen display.
• Touch Save to save the abbreviated solution
name and return to the Edit Menu or
• Touch Cancel to return without saving the new
name.
11. Under Set Limit Value, touch Modify. Note: The programmed end stations will automatically be
• A numeric keypad displays. positioned in the end station located on the far right.
5. Touch Save.
• The Create Solution Configuration Name screen
displays.
6. If necessary, use the touch keypad to modify the
solution configuration name, up to 20 characters long
and then touch Save.
• Touch Cancel to exit without saving.
• The Create Abbreviated Solution Name Screen
Displays.
7. If necessary, use the touch keyboard to enter an
abbreviated solution name (to display on the station
button), up to 8 characters long.
8. Touch Save to save the name.
• The Edit Configuration 早Solution Configuration
12. Using the touch keypad, enter a number value to set Name旨 screen displays
the limit for either runs or days • Each square represents a station.
13. Touch Save. 9. Select the station to assign the configuration to.
• A message dialog box displays.
Note: Start/End, drying and wash stations are blocked.
14. Touch Save and Exit to save the new solution
• The Edit Solution Configuration> Selected
configuration.
Station screen Displays.
To edit a solution configuration: 10. Under Solution Name, use the Solution List to select
1. From the Edit Menu, touch the Edit Solution and highlight a solution to attribute to the selected
Configuration button. station.
• The Edit Solution Configuration screen displays. 11. Select a Management Method days or runs
2. From the Solution Configuration List, select to 12. Under Set Limit Value, touch Modify.
highlight the solution configuration to be modified and • A numeric keypad displays.
then touch the Edit button. 13. Using the touch keypad, enter a number value to set
3. The Edit Solution Configuration Screen displays. the limit for either runs or days
4. Set the following parameters for the solution 14. Touch Save.
configuration: • A message box displays.
Solution Configuration Type Standard or 15. Touch Save and Exit to save the new solution
Expansion configuration.
Number of Start Stations 1 to 3 Stations
Number of End Stations 1 to 3 Stations
Program End (PE) Stations In Use or Not in Use
Number of Wash Stations 0 to 4 Stations
Special Solution Stations In Use or Not in Use
Heating Stations In Use or Not in Use
Drying Stations In Use or Not in Use
Note: If the wrong step was selected use the up and To delete a step
down arrows to position the cursor to the desired field.
• Touch the screen to select and highlight the step
3. Touch Save. to be deleted and then touch Delete Step.
Setting Mix Function Confirming Program Changes
1. Touch to highlight the Mix field in the step to be When all changes have been made to the selected
modified. program:
• The following screen displays: 1. Touch Save.
• A Confirmation dialog box displays.
OPERATING INSTRUCTIONS
Initial Setup Preparation
Ensure that the initial setup procedures have been CAUTION: The use of iodine on this instrument may
performed in accordance with the requirements of Section cause oxidation on some parts. It is recommended
3, Customization of Settings, prior to operating the Tissue- that if iodine must be used in some staining methods,
Tek® Prisma®. that it be removed immediately upon completion of
the stain. Failure to remove the iodine may cause
oxidation.
Understanding the Control Panel
Display Loading of slides
The control panel provides the controls and indicators
Slides must be correctly inserted in the slide baskets
necessary to initiate a program, monitor the status of
prior to loading them on the instrument. The following
processes currently running, and configure system
precautions must be followed:
settings.
• The specimen side of the slide must face forward
(towards the words “upside” on the basket).
Process Monitor Screen • The label end of the slide must be outside (top)
The Process Monitor Screen (Figure 4-A) is displayed of the basket.
when the system is ready for operation. It shows the • Each slide must be in parallel grooves
current solution configuration and allows for log on and to
• Each basket holds 20 slides; the empty grooves
monitor any processes currently running.
in a partially filled basket may be anywhere in the
basket.
Before starting a program, verify the following:
• The water supply and drain hose are securely
connected (see Section 2, Installation) and the
water run up time is set according to the
procedure in Section 3.
• All solution reservoirs are installed and filled with
the appropriate solutions.
• Dryer stations, wash stations, and start/end
stations are properly installed (see Section 2)
• Carbon filters have been installed (see section 2)
Figure 4-A
• Start Confirmation - Confirms the start action. Reset Solution Usage Information
There are two options: This function allows users to reset the solution stations
In Use that have reached their usage limit. (See section 3 for
Not In Use setting reagent management features)
• Process Start Notice - Reconfirmation that the 1. From the Stain Process Menu, touch the Reset
process has started. There are two options: Solutions Usage Information button.
In Use • The Reset Solution Usage Information screen
Not in Use displays.
3. Touch the System Setup Tab and make changes to
the following settings as necessary.
• Heating Station Always Heated - When the unit
is powered on, the heating stations are always
on. There are two options:
Yes
No
NOTE: This option is only available on product # 6131-
Tissue-Tek Prisma with Special Stains.
• Solution Name Display - This allows the user to
view the solution names on the Stain Process
Monitor Screen. The options are:
Yes
No
• Link with Coverslipper - If the Prisma® is
• To select a specific station touch the station
connected to the Film® Coverslipper there is a
number.
choice to have the instrument transfer the slides
automatically or the instrument can be used • To select the stations that have reached their
independently. Options are: usage limit, touch Overused Stations.
Yes • To select all stations, touch All Stations.
No 2. Once stations have been selected, touch Reset to
Note: This option is only available if the Prisma is reset values to zero.
connected with the Tissue-Tek® Film® Coverslipper. • A confirmation dialog box appears, answer Yes
4. Touch Exit to return to the Stain Process Menu. to reset the solution usage information.
3. Touch Exit to return to the Stain Process Menu.
Detailed Monitor
While a staining run is in progress, the operator can
check the time left for a particular run and get an
estimated time for completion. The Detailed Monitor
Screen will display the run number for each program
currently running, the status of that run, the step and
station currently in, the time left in that station, and the
predicted end time for that run.
1. From the Stain Process Monitor screen, touch the
Detailed Monitor button.
3. Touch the Exit button to return to the Stain Process
Monitor Screen.
ACCESSORIES
Standard Accessories
Basic Unit
Product Product
Code Description Code Description
20-Slide Basket
6135 Load/Unload
Adapter Special Stain
6140 Solution Reservoir,
160 ml*
Standard Solution
6147 Reservoir,
680 ml
6137 10-Slide Basket*
3-Position
6152
Reservoir Tray
3-Position Reservoir
6153
Lid
4-Position Reservoir
6154
Tray
4-Position Reservoir
6155
Lid
Special Stain
6156
Reservoir Tray*
Optional Accessories
Product
Code Description
Small Solution
6145 Reservoir,
260 ml
Small Solution
6146
Reservoir Lid
Duct Connection
6165 Adapter,
38mm dia.
Duct Connection
6166 Adapter,
75mm dia
Daily Maintenance
Exterior/Interior Surfaces
Figure 6-A
Daily clean the plastic and painted areas by wiping with
a clean cloth. Clean the interior of the solution area by 3. Lower the filter cover.
wiping with a clean cloth. 4. Remove the plastic wrapping from both carbon filters.
5. Remove the activated carbon filters from the
Bi-Monthly Maintenance instrument and dispose of them properly.
6. Place the carbon filters side by side in a horizontal
Replacement of the Carbon Filter position in the appropriate location (Figure 6-B).
It is recommended that the carbon filters be changed
after two weeks or sooner if fume monitoring reveals
elevated reagent vapors.
Keep the filters in place even if there is an internal duct
system attached.
To replace the carbon filters:
CAUTION: Make sure that the instrument is switched
off and the power cord is unplugged before
beginning.
1. Open the main cover of the instrument and locate the
Figure 6-B
filter cover located on the rear panel inside of the
instrument. 7. Push firmly against the filter cover to close and latch
2. Open the filter covers by pressing down on the button the hook.
located at the top center of the filter cover (Figure 8. Lower the main cover.
6-A).
TROUBLESHOOTING
General Information If addition assistance is required concerning an
instrument problem, or if the problem cannot be isolated or
The following Troubleshooting Chart contains general is beyond the scope of this manual, please contact the
problems that could occur during operation of the Tissue- Sakura Finetek USA, Technical Support Department. If
Tek® Prisma® Automated Slide Stainer. Probably causes outside the USA, call the nearest Sakura representative.
and recommended remedies are included for each
problem listed, to aid in diagnosis and correction.
Temperature Low (Drying Temperature has not reached the Temperature control at drying station B is Stop processes using drying stations.
Station B) set temperature at Drying Station stopped. All other functions are running Contact Sakura Technical Support
B in the allotted time. normally. Department.
Thermal Fuse (Drying Internal temperature of Drying Temperature control at drying station B is Stop processes using drying stations.
Station B) Station B has reached 85º C. The stopped. All other functions are running Contact Sakura Technical Support
over-temperature prevention normally. Department.
function is running.
Fan Stop (Drying Station Drying Station B fan has stopped Temperature control at drying station B is Stop processes using drying stations.
B) functioning. stopped. All other functions are running Contact Sakura Technical Support
normally. Department.
Sensor Opened (Heating Heating Station A sensor is Temperature control at Heating Station A is Stop processes using Heating
Station A) disconnected. stopped. All other functions are running Stations. Contact Sakura Technical
normally. Support Department.
Sensor Shorted (Heating Heating Station A sensor has Temperature control at Heating Station A is Stop processes using Heating
Station A) short-circuited. stopped. All other functions are running Stations. Contact Sakura Technical
normally. Support Department.
Temperature Low (Heating Temperature has not reached the Temperature control at Heating Station A is Stop processes using Heating
Station A) set temperature at Heating Station stopped. All other functions are running Stations. Contact Sakura Technical
A in the allotted time. normally. Support Department.
Thermal Fuse (Heating Internal temperature of Heating Temperature control at Heating Station A is Stop processes using Heating
Station A) Station A has reached 85º C. The stopped. All other functions are running Stations. Contact Sakura Technical
over-temperature prevention normally. Support Department.
function is running.
Sensor Opened (Heating Heating Station B sensor is Temperature control at Heating Station B is Stop processes using Heating
Station B) disconnected. stopped. All other functions are running Stations. Contact Sakura Technical
normally. Support Department.
Sensor Shorted (Heating Heating Station B sensor has Temperature control at Heating Station B is Stop processes using Heating
Station A) short-circuited. stopped. All other functions are running Stations. Contact Sakura Technical
normally. Support Department.
Temperature Low (Heating Temperature has not reached the Temperature control at Heating Station B is Stop processes using Heating
Station A) set temperature at Heating Station stopped. All other functions are running Stations. Contact Sakura Technical
B in the allotted time. normally. Support Department.
Thermal Fuse (Heating Internal temperature of Heating Temperature control at Heating Station B is Stop processes using Heating
Station A) Station B has reached 85º C. The stopped. All other functions are running Stations. Contact Sakura Technical
over-temperature prevention normally. Support Department.
function is running.
No Communication With Settings show that the Film and Communication between the Prisma and Check to see if the Film is ready for
Film Coverslipper Prisma should be connected but Film cannot be established. use or communication cable between
communication cannot be reached. the Prisma and Film is correctly
connected.
Film Operation Stop Hood was open during operation Slides will be transferred to and End Station Close hood and perform recovery
and could not transfer the Film until the Film loading station is available. process; reestablish communication
loading station to the Prisma. with Film.
File System Error (A Drive) Memory card in the A Drive is Data cannot be transferred into memory Use another readable/writeable
unrecognizable. card in the A Drive. All functions including memory card.
staining processes are usable.
File Not Found (A Drive) Failed to detect file in A Drive. Data cannot be transferred into memory Insert memory card in the A Drive. If
card in the A Drive. All functions including error still occurs, use another
staining processes are usable. readable/writeable memory card.
No Space in Memory Card Shortage of memory card capacity Data cannot be transferred into memory Replace the memory card in A Drive.
Cover (Hood) is Open. Cover (Hood) was opened during Emergency stop condition. Close the cover and perform recovery
Robotic arm has stopped. the staining process and the operation.
robotic arm has stopped.
Door is Open. Staining Door was opened when the robotic Emergency stop condition. Close the door and perform recovery