Protocol: Genetisure Pre-Screen Kit For Single Cell Analysis
Protocol: Genetisure Pre-Screen Kit For Single Cell Analysis
Protocol: Genetisure Pre-Screen Kit For Single Cell Analysis
Protocol
Revision C0
Notices
Manual Part Number Warranty Safety Notices
G4410-90003 The material contained in this document
Revision C0, August 2020 is provided “as is,” and is subject to being
changed, without notice, in future
CAUTION
Copyright editions. Further, to the maximum extent
permitted by applicable law, Agilent A CAUTION notice denotes a
© Agilent Technologies, Inc. 2018, 2020 disclaims all warranties, either express or hazard. It calls attention to an
No part of this manual may be implied, with regard to this manual and operating procedure, practice, or
reproduced in any form or by any means any information contained herein, the like that, if not correctly
(including electronic storage and retrieval including but not limited to the implied performed or adhered to, could
or translation into a foreign language) warranties of merchantability and fitness
result in damage to the product or
without prior agreement and written for a particular purpose. Agilent shall not
consent from Agilent Technologies, Inc. be liable for errors or for incidental or loss of important data. Do not
as governed by United States and consequential damages in connection proceed beyond a CAUTION
international copyright laws. with the furnishing, use, or performance notice until the indicated
Agilent Technologies, Inc.
of this document or of any information conditions are fully understood
contained herein. Should Agilent and the and met.
5301 Stevens Creek Blvd.
user have a separate written agreement
Santa Clara, CA 95051 USA
with warranty terms covering the
material in this document that conflict WARN ING
Technical Support with these terms, the warranty terms in
the separate agreement shall control. A WARNING notice denotes a
For US and Canada
hazard. It calls attention to an
Call 800-227-9770 (option 3, 4, 2) Technology Licenses
Or send an email to operating procedure, practice, or
genomics@agilent.com The hardware and/or software described the like that, if not correctly
in this document are furnished under a performed or adhered to, could
For all other regions license and may be used or copied only in
accordance with the terms of such
result in personal injury or death.
Agilent’s world-wide Sales and Support
license. Do not proceed beyond a
Center contact details can be obtained at
www.agilent.com/en/contact-us/page. WARNING notice until the
Restricted Rights Legend indicated conditions are fully
U.S. Government Restricted Rights.
understood and met.
Software and technical data rights
granted to the federal government
include only those rights customarily
provided to end user customers. Agilent
provides this customary commercial
license in Software and technical data
pursuant to FAR 12.211 (Technical Data)
and 12.212 (Computer Software) and, for
the Department of Defense, DFARS
252.227-7015 (Technical Data -
Commercial Items) and DFARS
227.7202-3 (Rights in Commercial
Computer Software or Computer
Software Documentation).
In This Guide…
This guide describes the Agilent recommended operational procedures to analyze
DNA copy number variations (CNVs) in single cell samples using the GenetiSure
Pre-Screen Kit. This protocol is specifically developed and optimized to amplify
and enzymatically label DNA from single cell samples, and then hybridize that DNA
to 8×60K and 4×180K CGH microarrays to obtain same day results.
2 Sample Amplification
This chapter describes the protocol to amplify DNA from single cell samples
alongside samples of reference DNA (male and female). For each slide, one
male and one female reference sample is amplified along with the single cells.
The reference DNA samples are provided with the GenetiSure Pre-Screen Kits
(P/N G9500A and G9501A) and the GenetiSure Pre-Screen Labeling Kit (P/N
G9502A).
3 Sample Labeling
This chapter describes the steps to differentially label the amplified DNA
samples with fluorescent-labeled nucleotides using reagents from the SureTag
DNA Labeling Kit that is included in the GenetiSure Pre-Screen Kit.
4 Microarray Processing
This chapter describes the steps for microarray processing, which consists of
hybridization, washing, and scanning.
5 Troubleshooting
This chapter contains potential causes for an array failure.
6 Reference
This chapter contains reference information that pertains to this protocol.
2 Sample Amplification 17
Step 1. Preparation of Samples 18
Step 2. Whole Genome Amplification 18
Step 3. Quantitation of Amplified DNA using Qubit Fluorometer 21
3 Sample Labeling 23
Step 1. Fluorescent Labeling of DNA 24
Step 2. Purification of labeled DNA 26
4 Microarray Processing 27
Hybridization 28
Step 1. Prepare the 10× Blocking Agent 28
Step 2. Prepare labeled DNA for hybridization 29
Step 3. Prepare the hybridization assembly 31
Step 4. Hybridize 38
Step 5. Prewarm Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2
(during hybridization) 39
Microarray Wash 40
Step 1. Prepare the equipment 40
Step 2. Prewarm Stabilization and Drying Solution (Wash Procedure B
Only) 42
Step 3. Wash microarrays 43
Step 4. Put slides in a slide holder 47
Microarray Scanning and Analysis 49
Step 1. Scan the microarray slides 49
5 Troubleshooting 55
If the whole genome amplification fails 56
If the labeling efficiencies for Cy3 and Cy5 are dissimilar 57
If you have post-labeling signal loss 58
If you have high BGNoise values 59
If you have poor reproducibility 60
6 Reference 61
Reagent Kit Components 62
“Secure Fit” Slide Box Opening Instructions 64
Microarray Handling Tips 66
Agilent Microarray Layout and Orientation 67
Array/Sample tracking on microarray slides 68
Make sure that you read and understand the information in this chapter and have
the necessary equipment and reagents listed before you start an experiment.
Procedural Notes
• Do not use a pipette to transfer or mix single cell samples until the whole
genome amplification is complete as cells may adhere to the pipette tip.
• Follow the procedure described in this document to amplify DNA from single
cells, to increase the likelihood of a successful experiment.
• To prevent contamination of reagents by nucleases, always wear powder-free
laboratory gloves, and use dedicated solutions and pipettes with
nuclease-free aerosol-resistant tips.
• Maintain a clean work area.
• Do not mix stock solutions and reactions containing DNA or enzymes on a
vortex mixer. Instead, mix the solutions and reactions by gently tapping the
tube with your finger.
• Avoid repeated freeze-thaw cycles of solutions containing DNA or enzymes.
• When preparing frozen reagent stock solutions for use:
1 Thaw the aliquot as quickly as possible without heating above room
temperature.
2 Mix briefly on a vortex mixer, and then spin in a microcentrifuge for 5 to 10
seconds to drive the contents off the walls and lid.
3 Store on ice or in a cold block until use.
• In general, follow Biosafety Level 1 (BL1) safety rules.
If you are using this product for embryo screening please make sure you adhere
N OTE to your country specific laws and regulations for human assisted reproductive
technologies. Your country might have banned sex selection for non-medical
purposes, as well as the commercial use of gametes, zygotes, and embryos.
Agilent shall have no liability for any direct, indirect, consequential, or incidental
damages arising out of the use, the results of use, or the inability to use this
product.
Safety Notes
The Agilent part numbers, sub-kit components, and storage conditions for the
GenetiSure Pre-Screen Kits are listed in Table 1.
P/N G9500A REPLI-g Single Cell Kit*, Sufficient reagents for whole Store at –20°C
GenetiSure Pre-Screen Kit, P/N 5191-4065 genome amplification of 42
8×60K, 48 Reactions experimental samples and 6
reference samples
Hybridization Chamber Gasket Three gasket slides, 8 gaskets Store at room temperature
Slide Kit, P/N G2534-60018 per slide
P/N G9501A REPLI-g Single Cell Kit*, Sufficient reagents for whole Store at –20°C
GenetiSure Pre-Screen Kit, P/N 5191-4065 genome amplification of 36
4×180K, 48 Reactions experimental samples and 12
reference samples
Hybridization Chamber Gasket Six gasket slides, 4 gaskets per Store at room temperature
Slide Kit, P/N G2534-60018 slide
P/N G9502A REPLI-g Single Cell Kit*, Sufficient reagents for whole Store at –20°C
GenetiSure Pre-Screen P/N 5191-4065 genome amplification of 36
Labeling Kit, 48 Reactions experimental samples and 12
reference samples
* The full list of components provided in the REPLI-g Single Cell Kit is available in Table 23 on page 62.
** The full list of components provided in the SureTag DNA Labeling Kit, –20°C Components, is available in Table 24 on page 62.
Table 2, Table 3, and Table 4 list the materials that are required for the whole
genome amplification, labeling, and CGH microarray hybridization protocol, but
are not provided in a GenetiSure Pre-Screen Kit.
Table 2 Required reagents
1× Phosphate Buffered Saline (pH 7.4) Thermo Fisher Scientific p/n 10010-023 or
equivalent
Qubit dsDNA BR Assay Kit, for use with the Qubit Thermo Fisher Scientific p/n Q32850
fluorometer, 100 assays (Optional. Used for
quantitation of amplified DNA on Qubit Fluorometer.)
96-well PCR plates, centrifuge for 96-well plates, and Plates: Agilent p/n 401334, or equivalent
thermal plate sealer (Optional. Can substitute with Centrifuge: Eppendorf p/n 5810, or equivalent
200-L PCR tubes.) Plate sealer: Agilent p/n G5402A/G, or equivalent
Magnetic stir plate with heating element Corning p/n 6795-420 or equivalent
Microcentrifuge with rotor for 200-L tubes and 1.5-mL Eppendorf p/n 5430 or equivalent
tubes (and 500-L tubes if single cells were collected in
these tubes)
Qubit Fluorometer (Optional. Used for quantitation of Thermo Fisher Scientific p/n Q32857
amplified DNA on Qubit Fluorometer.)
Thin wall, clear 0.5 mL PCR tubes (Optional. Used for Thermo Fisher Scientific p/n Q32856 or
quantitation of amplified DNA on Qubit Fluorometer.) VWR p/n 10011-830
P10, P20, P200 and P1000 pipettes Pipetman P10, P20, P200, P1000 or equivalent
Magnetic stir bar, 7.9 × 38.1 mm (×2 or ×4)† VWR p/n 58948-150 or equivalent
250 mL capacity slide-staining dish, with slide rack Wheaton p/n 900200 or
(×3 or ×5)† Thermo Shandon p/n 121
Ice bucket
Clean forceps
Powder-free gloves
Timer
Vortex mixer
* Optional. Recommended when processing arrays with a G2565CA scanner in environments in which
ozone levels are 5 ppb or higher.
† The number varies depending on if wash procedure A or B is selected.
• Refer to the Agilent Scanner manual and Agilent CytoGenomics manuals for
minimum memory requirements and other specifications for the PC used to
the run these software programs. Go to www.agilent.com to download the
manuals.
• You can download the design files needed for data extraction and analysis in
CytoGenomics (version 3.0 or later) from the Agilent SureDesign website. Go
to www.agilent.com/genomics/suredesign.
Figure 1. Overview of the GenetiSure Pre-Screen workflow for single cell analysis
This chapter describes the protocol to amplify DNA from single cell samples
alongside samples of reference DNA (male and female). For each slide, one male
and one female reference sample is amplified along with the single cells. The
reference DNA samples are provided with the GenetiSure Pre-Screen Kits (P/N
G9500A and G9501A) and the GenetiSure Pre-Screen Labeling Kit (P/N G9502A).
The single cell samples can be any of the following:
• Single cell collected from a day-3 human embryo biopsy
• Cells (3–10 cells total) collected from a day-5 human embryo biopsy
• Single cell sample (1–2 cells) of tissue-culture cells
• Small tissue-culture sample consisting of 5–15 cells
• Multi-cell tissue-culture sample consisting of 15–50 cells or more
This section describes the preparation of the male and female reference DNA
samples and the single cell samples.
1 Prepare the reference DNA samples. Perform these steps for both the Human
Reference DNA, Female, P/N 5190-3797 and Human Reference DNA, Male,
P/N 5190-3796 that are provided in the GenetiSure Pre-Screen Kit.
a In a fresh 200-L PCR tube, combine 2 L of Human Reference DNA with
104.6 L of PBS. Mix briefly on a vortex mixer.
b Transfer 4 L of the diluted Human Reference DNA into a fresh PCR tube.
Keep on ice until required.
2 Prepare samples of single cells.
• If the single cell is stored in a 500 L-tube, use a pipette to remove the
mineral oil and leave the cell in the tube.
• If the single cell is stored in a 4-well dish, petri dish, or other similar culture
vessel, use a denudation pipette under the microscope to transfer the cell
from the dish into a PCR tube containing 3 L of PBS (for a total volume of
approximately 4 L).
This section describes the Agilent recommended procedure to amplify DNA from
single cells and male and female reference DNA using the REPLI-g Single Cell Kit.
1 Prepare Buffer DLB.
a Add 500 L of nuclease-free distilled water to the provided tube of Buffer
DLB to resuspend the lyophilized pellet. Mix thoroughly then spin briefly in
a centrifuge.
b Prepare aliquots of the reconstituted Buffer DLB. Store the aliquots at
–20°C for up to 6 months. Buffer DLB is pH-labile and should not be stored
longer than 6 months.
2 Prepare Buffer D2 (denaturation buffer).
a Mix the components listed in Table 5.
3 Add Buffer D2 to the samples (single cell samples and reference samples).
a Briefly spin all sample tubes in a microcentrifuge.
b Add 3 L of Buffer D2 to each sample tube for a total volume of 7 L.
c Mix carefully by flicking the tubes. Do not mix on a vortex mixer or by
pipetting up and down.
d Briefly spin all sample tubes in a microcentrifuge then transfer to ice.
4 Set the thermal cycler (with heated lid) to run the program listed in Table 6.
Table 6 Incubation protocol to lyse the cells
5 Transfer the sample tubes to the thermal cycler, and then start the program.
If your samples are in 500-L tubes, and your thermal cycler does not
N OTE accommodate tubes of this size, use a 65°C water bath for step 1 then transfer
the samples to an ice bath.
6 At the end of the thermal cycler program, remove the tubes from the thermal
cycler. Briefly spin them in a microcentrifuge, then transfer to ice.
7 Without allowing the pipette tip to contact the sample, add 3 L of Stop
Solution to each sample for a total volume of 10 L. Mix carefully by flicking
the tubes. Briefly spin them in a microcentrifuge, then transfer to ice.
8 Prepare the Amplification Master Mix by mixing the components in Table 7
on ice. Mix the components in the order listed in the table, and briefly vortex
the master mix before adding the Amplification DNA Polymerase. Once you
add the Amplification DNA Polymerase, mix the Amplification Master Mix by
flicking the tube or pipetting up and down. Then, immediately proceed to
step 9. The components are included in the REPLI-g Single Cell Kit that is
included in the GenetiSure Pre-Screen Kit.
Thaw the Amplification DNA Polymerase on ice. For all other components, thaw
N OTE at room temperature, vortex to mix, then briefly spin in a microcentrifuge.
The Amplification Reaction Buffer may form a precipitate after thawing. Mixing it
on a vortex mixer for 10 seconds dissolves the precipitate.
11 Transfer the sample tubes to the thermal cycler, and then start the program.
12 At the end of the thermal cycler program, remove the tubes from the thermal
cycler. Briefly spin them in a microcentrifuge, then transfer to ice.
1 Set up clear, thin-walled 0.5-mL PCR tubes for all samples (the two standards
and the amplified DNA samples that you are processing).
2 Make a Qubit working solution. For each standard and amplified DNA sample
to be quantified, mix the components in Table 9.
Table 9 Qubit Working Solution
Component Volume
3 Add 190 L of Qubit working solution to the two 0.5-mL tubes labeled for the
standards.
4 Add 199 L of Qubit working solution to the 0.5-mL tubes labeled for your
amplified DNA samples.
5 Add 10 L of Qubit dsDNA BR standard #1 to the tube labeled for standard #1,
and add 10 L of Qubit dsDNA BR standard #2 to the tube labeled for
standard #2.
6 Add 1 L of amplified DNA sample to the remaining 0.5-mL tubes.
7 Mix the contents of all tubes on a vortex mixer for 2–3 seconds, taking care
not to create bubbles.
8 Incubate the tubes at room temperature for 2 minutes.
9 Calibrate the Qubit.
a On the home screen of the Qubit, use the up or down arrow to select
dsDNA Broad Range Assay as assay type, and then press GO. The
standard screen is automatically displayed.
b Select Run new calibration, and then press GO.
c Insert the tube with the first standard into the Qubit Fluorometer, close the
lid and press GO. After the reading is done, remove the standard.
d Insert the tube with the second standard into the Qubit Fluorometer, close
the lid, and press GO. After the reading is done, remove the standard.
The calibration is complete after the second standard has been read.
10 Measure the concentrations of the amplified DNA samples.
a Insert a sample and press GO.
b When the measurement is complete (approximately 5 seconds later),
make a note of the reading. The result is displayed on the screen. The
number displayed is the concentration of the nucleic acid in the assay
tube.
c Remove the sample from the instrument, insert the next sample, and press
GO.
d Repeat sample readings until all samples have been read.
e Calculate the concentration of your original sample.
The Qubit Fluorometer gives a value for the Qubit dsDNA BR assay in g/mL.
This value corresponds to the concentration after your samples were diluted
into the assay tube. To calculate the concentration of your sample, use the
equation below.
Sample concentration = QF value × 200
where
QF value = the value given by the Qubit Fluorometer
The expected concentration for all samples of amplified DNA is 300 ng/L.
If all of your samples, including the reference samples, have a low concentration
N OTE (i.e., lower than 300 ng/L) then the whole genome amplification failed and
Agilent does not recommend proceeding to sample labeling. If some, but not all,
of the samples have a low concentration, then Agilent recommends repeating the
whole genome amplification of those samples, if possible.
Maintain the samples at 4°C for short-term storage (up to 3 days), or store at
–20°C for long-term storage (up to 1 year) until ready for labeling. When ready,
proceed to Chapter 3, “Sample Labeling.”
This chapter describes the steps to differentially label the amplified DNA samples
with fluorescent-labeled nucleotides using reagents from the SureTag DNA
Labeling Kit that is included in the GenetiSure Pre-Screen Kit.
The procedure uses random primers and the exo(-) Klenow fragment to
differentially label amplified DNA samples with fluorescent-labeled nucleotides.
For your single cell samples, you differentially label the amplified DNA with
cyanine-3 (Cy3) and cyanine-5 (Cy5) dyes, and then combine the samples in
Cy3-Cy5 pairs. You also differentially label the amplified DNA from your male and
female reference samples with Cy3 and Cy5 dyes and combine as a Cy3-Cy5
pair.
Cyanine 3-dUTP and cyanine 5-dUTP are light sensitive and are subject to
N OTE degradation by multiple freeze thaw cycles. Minimize light exposure throughout
the labeling procedure.
Each array on the microarray slide needs to be hybridized with a Cy3-Cy5 pair.
N OTE
If you have an odd number of single cell samples, you need to set up an
additional tube for a second male reference sample and label it with the
appropriate dye in order to pair it with the odd single cell sample. Pairing of Cy3-
and Cy5-labeled single cells samples is described in step 12.
Similarly, if you have an insufficient number of samples to fill all of the arrays on a
4×180K or 8×60K slide, set up additional tubes of references samples and label
them with Cy3 and Cy5 in order to create enough Cy3-Cy5 pairs to fill the empty
arrays.
5 Transfer the PCR tubes to the thermal cycler, and then start the program.
6 At the end of the thermal cycler program, remove the tubes from the thermal
cycler. Briefly spin them in a microcentrifuge, then transfer to ice.
7 Mix the components in Table 11 on ice in the order indicated to prepare one
cyanine-3 (Cy3) and one cyanine-5 (Cy5) Labeling Master Mix. The
components are included in the SureTag DNA Labeling Kit that is included in
the GenetiSure Pre-Screen Kit.
Table 11 Labeling Master Mix
8 Add 9.5 μL of Labeling Master Mix to each sample tube that contains the
amplified DNA to make a total volume of 25 μL. Mix well by gently pipetting up
and down. Briefly spin them in a microcentrifuge, then transfer to ice.
• For the male reference sample and the female reference sample, use the
Cy3 Labeling Master Mix for one of the two samples and use the Cy5
Labeling Master Mix for the other sample.
• For the single cell samples, use the Cy3 Labeling Master Mix for half of the
samples and use the Cy5 Labeling Master Mix for the other half of the
samples. If you have an odd number of single cell samples, use the second
male reference sample to create an even number.
9 Set the thermal cycler (with heated lid) to run the program listed in Table 12.
Table 12 DNA labeling using a thermal cycler
10 Transfer the sample tubes to the thermal cycler, and then start the program.
11 At the end of the thermal cycler program, remove the tubes from the thermal
cycler. Briefly spin them in a microcentrifuge, then transfer to ice.
12 Combine the differentially labeled male and female reference samples to
create a Cy3-Cy5 pair with a final volume of 50 L. Similarly, for the single cell
samples, combine two differentially labeled samples to create Cy3-Cy5 pairs
with a final volume of 50 L. If you have an odd number of single cell samples,
use the second male reference sample to create a Cy3-Cy5 pair.
Paired samples of labeled DNA can be stored for up to a month at -20°C in the
N OTE dark.
This chapter describes the steps for microarray processing, which consists of
hybridization, washing, and scanning.
Hybridization
The 10× Blocking Agent can be prepared in advance and stored at -20°C.
N OTE
* Included in the Agilent Oligo aCGH/ChIP-on-chip Hybridization Kit, p/n 5188-5220 (25 slides) or
5188-5380 (100 slides)
* Included in the Agilent Oligo aCGH/ChIP-on-chip Hybridization Kit, p/n 5188-5220 (25 slides) or
5188-5380 (100 slides)
2 Add the appropriate volume of the Hybridization Master Mix to each PCR tube
or plate well that contains 15 L of paired DNA to make the total volume listed
in Table 15.
4-pack 95 μL 110 μL
8-pack 35 μL 50 μL
5 Transfer the PCR tubes to the thermal cycler, and then start the program.
6 At the end of the thermal cycler program, remove the tubes from the thermal
cycler. Briefly spin them in a microcentrifuge and proceed to the hybridization
steps.
1 With tweezers, carefully lift up the corner of the clear plastic covering and
slowly pull back the protective film.
2 With clean, powder-free gloved fingers, remove the gasket slide from its
package. Handle the slide only on its edges.
2 Locate the four chamber base guideposts and rectangular barcode guide in
the chamber base.
3 Position the gasket slide between the 4 chamber base guide posts (see
Figure 4) with the barcode label resting over the base’s rectangular barcode
guide.
The “drag and dispense” method helps to distribute the sample evenly across
the surface of the well and avoids spillover of sample over the gasket edge.
Start with the pipette tip near the top edge of the well. Do not directly touch the
gasket or the glass with the pipette tip. Then, dispense the mixture while you
move your pipette tip to the opposite end of the well so that the sample is
distributed across the well space. Avoid creating large air bubbles as you
dispense the mixture as they could lead to spillover.
This image is for demonstration purposes only. Always put the gasket slide in
the chamber base before you dispense the hybridization sample mixture.
Figure 6. Drag and dispense method – Start dispensing when the pipette tip is near the top of the well.
Finish dispensing when the pipette tip is near the bottom of the well.
Figure 7. Chamber base with gasket and microarray slide applied, guide posts denoted with arrows
Do not drop the microarray slide onto the gasket slide as this increases the
CAU TI O N chances of sample mixing between gasket wells.
Once placed, do not attempt to move the chamber and sandwiched slides as
this can cause leakage of the hybridization solution.
Correct Wrong
Figure 8. Chamber cover in correct (left) and incorrect (right) orientations
2 From the rounded corner of the chamber base, slip the clamp onto the
chamber base and cover until it stops firmly in place, resting at the center of
the two pieces.
Keep the chamber assembly flat on the lab bench to avoid spilling the
hybridization solution.
Wrong
Correct
Figure 12. The slide on the left shows a stray, stationary bubble (denoted with arrow), which must be
removed before hybridization. The slide on the right shows only large mixing bubbles, which move freely
around the chamber when rotated. Bubbles are acceptable, as long as they move freely when you rotate the
chamber.
Step 4. Hybridize
1 Load each assembled chamber into the oven rotator rack, starting from the
center of the rack (position 3 or 4 when counting from left to right) Refer to the
figure below for correct and incorrect orientations.
Figure 13. Assembled chambers in correct (left) and incorrect (middle and right) orientations
If you are not loading all the available positions on the hybridization rotator
CAU TI O N
rack, be sure to balance the loaded hybridization chambers on the rack similar
to a centrifuge to prevent unnecessary strain on the oven motor.
You must calibrate the hybridization oven regularly for accuracy of the
CAU TI O N
collected data. Refer to Agilent G2545A Hybridization Calibration Procedure
(publication G2545-90002) for more information.
Microarray Wash
> 5 ppb < 10 ppb “Wash Procedure A (without Stabilization and Drying Yes
Solution)” on page 43
> 10 ppb “Wash Procedure B (with Stabilization and Drying Solution)” on Yes
page 45
Always use clean equipment when doing the hybridization and wash steps to
avoid wash artifacts on your slides and images.
Use only dishes that are designated and dedicated for use in GenetiSure
Pre-Screen experiments.
1 Add the slide rack and stir bar to the slide-staining dish.
2 Transfer the slide-staining dish with the slide rack and stir bar to a magnetic
stir plate.
3 Fill the slide-staining dish with 100% acetonitrile.
4 Turn on the magnetic stir plate and adjust the speed to 350 rpm (medium
speed).
5 Wash for 5 minutes at room temperature.
6 Slowly remove the slide rack from the slide-staining dish.
The liquid tension created by removing the slide rack slowly helps to limit the
amount of acetonitrile that adheres to the slides.
7 Repeat step 1 through step 5 with fresh acetonitrile. Discard the used
acetonitrile as is appropriate for your site.
8 Air dry all of the equipment in the vented fume hood, then proceed to “Milli-Q
ultrapure water wash (all equipment)”, below.
Some detergents may leave fluorescent residue on the dishes. Do not use any
CAU TI O N
detergent in the washing of the staining dishes, slide racks, or stir bars. If
detergent is used, all traces must be removed by copiously rinsing with Milli-Q
ultrapure water.
1 Put a clean magnetic stir bar into the Stabilization and Drying Solution bottle
and recap.
2 Partially fill a plastic bucket with hot water at approximately 40°C to 45°C (for
example from a hot water tap).
3 Put the Stabilization and Drying Solution bottle into the hot water in the plastic
bucket.
4 Put the plastic bucket on a magnetic stirrer (not a hot-plate) and stir.
5 The hot water cools to room temperature. If the precipitate has not all
dissolved replenish the cold water with hot water.
6 Repeat step 5 until the solution is clear.
Do not filter the Stabilization and Drying Solution, or the concentration of the
CAU TI O N
ozone scavenger may vary.
8 When all slides in the group are put into the slide rack in slide-staining dish #2,
stir at 350 rpm for 5 minutes. Adjust the setting to get good but not vigorous
mixing.
9 Wash the slides in Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 2:
a Prior to transferring the slide rack, activate the magnetic stirrer in
slide-staining dish #3, which contains Agilent Oligo aCGH/ChIP-on-Chip
Wash Buffer 2 at 37°C. Adjust the setting to get thorough mixing.
b Transfer slide rack to slide-staining dish #3. If necessary, further adjust the
setting on the magnetic stirrer to get thorough mixing without disturbing
the microarray slides.
c Wash microarray slide for at least 1 minute and no more than 2 minutes.
10 Slowly remove the slide rack trying to minimize droplets on the slides. It
should take 5 to 10 seconds to remove the slide rack.
11 Discard used Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1 and Agilent
Oligo aCGH/ChIP-on-Chip Wash Buffer 2.
12 Repeat step 1 through step 11 for the next group of five slides using fresh
Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1 and Agilent Oligo
aCGH/ChIP-on-Chip Wash Buffer 2 warmed to 37°C.
13 Scan slides immediately to minimize the impact of environmental oxidants on
signal intensities. If necessary, store slides in orange slide boxes in a N2 purge
box, in the dark.
The Stabilization and Drying Solution must be set-up in a fume hood. Put the
WA R N I N G
Wash Buffer 1 and Wash Buffer 2 set-up areas close to, or preferably in, the
same fume hood. Use gloves and eye/face protection in every step of the
washing procedure.
Table 19 lists the wash conditions for the Wash Procedure B with Stabilization
and Drying Solution.
Table 19 Wash conditions
The acetonitrile and the Stabilization and Drying Solution may be reused for
N OTE washing of up to two batches of five slides (that is, total 10 microarray slides) in
one experiment. Pour the Stabilization and Drying Solution to a different marked
bottle, and protect from light with other flammables. After each use, rinse the
slide rack and the slide-staining dish that were in contact with the Stabilization
and Drying Solution with acetonitrile followed by a rinse in Milli-Q ultrapure water.
8 Repeat step 1 through step 7 for the next group of five slides using fresh
Agilent Oligo aCGH/ChIP-on-Chip Wash Buffer 1 and Agilent Oligo
aCGH/ChIP-on-Chip Wash Buffer 2 prewarmed to 37°C.
9 Dispose of acetonitrile and Stabilization and Drying Solution as flammable
solvents.
• In environments in which the ozone level is below 5 ppb, put the slides with
Agilent barcode facing up in a slide holder.
Scan resolution 3 μm
6 Verify that the Scanner status in the main window is Scanner Ready.
7 Click Scan Slot m-n on the Scan Control main window, where the letter m
represents the Start slot where the first slide is located and the letter n
represents the End slot where the last slide is located.
After scanning is completed, load the microarray TIF images into the Agilent
CytoGenomics software (version 3.0 or higher) for feature extraction and
analysis using one of the preloaded analysis methods designed for single cell
samples. These analysis methods are described in Table 21. Details on the
analysis method parameters are accessible from the Configure Settings screen
of the CytoGenomics software.
For users of the Innopsys InnoScan 710 Microarray Scanner If you are using
CytoGenomics version 3.0 or 4.0, then you must first open the scanner-generated
TIFF image in the Feature Extraction for CytoGenomics software module (see the
Feature Extraction for CytoGenomics User Guide for instructions on launching the
software module and opening TIFF files from within the software). As the
Feature Extraction for CytoGenomics software opens the TIFF image, it converts
the image to the Agilent format and saves the converted image to the original file
folder (with “converted” appended to the file name). Use this converted image as
the sample file in your analysis workflow in CytoGenomics 3.0 or 4.0. If you have
already upgraded to CytoGenomics 5.0, then this separate conversion step for
InnoScan files is not necessary.
Single Cell Recommended Agilent’s recommended analysis method for analysis of single cell
Analysis Method samples. This analysis method is appropriate for most single cell
analyses.
Single Cell Small Aberration This analysis method is appropriate for analysis of single cell sample in
Analysis Method which you want to focus on a few particular loci of interest because the
Aberration Filter used in this analysis method has a less stringent
threshold for aberration size than that used in the Single Cell
Recommended Analysis Method. However, because of the reduced
stringency, the risk for false positives is higher with this analysis method
compared to the Single Cell Recommended Analysis Method.
Single Cell Long Low This analysis method is appropriate for analysis of mosaic samples
Aberration Analysis Method consisting of just a few cells. The Aberration Filter used in the analysis
method is capable of finding large aberrations with a compressed log2
ratio.
Figure 16. Example slide and resulting comparisons for two single cells samples on the slide (Sample1
and Sample2)
* On the Agilent SureScan microarray scanner (model G4900DA or G5761AA), the threshold is 15. On
the Agilent microarray C scanner model G2565CA, the threshold is 20.
or G5761AA), the threshold is 15. On
† On the Agilent SureScan microarray scanner (model G4900DA
the Agilent microarray C scanner model G2565CA, the threshold is 20.
If you have low post-amplification yield, as determined using the Qubit dsDNA BR
kit, the whole genome amplification may have been inefficient.
• Do not mix solutions containing single cells by pipetting up and down, as this
may cause the cells to adhere to the pipette tip. Instead, mix the samples by
flicking the tubes.
• When preparing cells for lysis (after adding the denaturation buffer), avoid
vigorous mixing, as this may shear the DNA.
• After cell lysis, proceed immediately to the next step of the protocol (the
addition of the Stop Solution, followed by amplification).
• Make sure that the cell lysis and amplification incubations are performed at
the correct temperatures. Use a thermal cycler with a heated lid that is set to
at least 70°C.
• To avoid degradation of the DNA within the single cells, do not store the cells
for extended periods of time, and make sure that the cells are always stored at
the appropriate temperature.
• Do not use reconstituted Buffer DLB that has been stored at –20°C for longer
than 6 months. After 6 months, prepare new aliquots of the reconstituted
Buffer DLB.
After you pair the Cy3- and Cy5-labeled samples, the paired sample should be
purple in color. A paired sample that is too pink indicates inefficient Cy5 labeling.
A paired sample that is too blue indicates inefficient Cy3 labeling. Inefficient
labeling can result from sub-optimal whole genome amplification and labeling
conditions such as too many freeze thaws of the buffers or Cyanine dUTP,
enzyme degradation due to being left warm for too long, wrong temperatures or
times, volume mistakes, or too much exposure of the dyes to light or air.
• Inefficient labeling can be caused by sub-optimal whole genome
amplification. See the troubleshooting suggestions in “If the whole genome
amplification fails” on page 56.
• Keep amplification and labeling enzymes on ice while setting up reactions,
and return them to –20°C as quickly as possible. Make sure to store Cyanine
dUTP at –20°C.
• Double check incubation times and temperatures (use a calibrated
thermometer), and use a thermal cycler with heated lid.
• Evaporation can be a problem when you process samples at high
temperatures. Make sure that sample tubes are well closed or use a plate heat
sealer to avoid evaporation.
• Make sure that the pipettes are not out of calibration.
• Make sure that the reagents and master mixes are well mixed. Tap the tube
with your finger or use a pipette to move the entire volume up and down. Then
spin in a microcentrifuge for 5–10 seconds to drive the contents off the walls
and lid. Do not mix the stock solutions and reactions that contain amplified
DNA or enzymes on a vortex mixer.
Signal loss can be due to wash or hybridization conditions that are too stringent,
or degradation of the cyanine-5 signal.
Cyanine-5 signal degradation can be caused by ozone or NOx compounds
coming from pollution and/or compressors and centrifuges. Cyanine-5 signal
degradation can result in less red signal around the edges of the features, a
visible gradient of red intensity (especially on the left side of the slide and on
slides scanned later in a batch), and poor red reproducibility of the cyanine
5-labeled single cell samples and green reproducibility of the cyanine 5-labeled
reference.
• Check that the oven temperature is 67°C. If needed, recalibrate the
hybridization oven. Follow the steps in Agilent G2545A Hybridization
Calibration Procedure (p/n G2545-90002).
• Check that the temperature of Wash 2 is 37°C.
• Check that Wash 2 was not accidentally used instead of Wash 1.
• Wash and scan slides in an ozone controlled environment (<5 ppb), such as
an ozone tent.
• Use small batches that can be washed and scanned in about 40 minutes to
minimize exposure to air.
• For Agilent Scanner C, use the Agilent Ozone-Barrier Slide Cover (p/n
G2505-60550). The SureScan scanner has built-in ozone protection.
• Use the Stabilization and Drying Solution as described in “Wash Procedure B
(with Stabilization and Drying Solution)” on page 45.
High BGNoise can cause lower signal-to-noise values (see Table 22 on page 53
for thresholds) and higher DLRSD values. If the BGNoise is high, examine the
microarray image for visible non-uniformities. High BGNoise is often introduced
during hybridization steps or washes.
• Make sure that the oven is calibrated. Follow the steps in Agilent G2545A
Hybridization Calibration Procedure (p/n G2545-90002).
Sample hybridization at incorrect temperatures affects the stringency of the
hybridization.
• Make sure that wash dishes, racks and stir bars are clean. Do not use tap
water or detergents to clean wash equipment. If needed, rinse wash
equipment with 2-Propanol (for equipment that was not exposed to
Stabilization and Drying Solution) or acetonitrile (for equipment that was
exposed to Stabilization and Drying Solution) followed by rinses with MilliQ
water. See “Milli-Q ultrapure water wash (all equipment)” on page 41.
• If high background is observed, perform an additional acetonitrile wash of the
slides and then rescan:
1 In the fume hood, fill a slide-staining dish approximately three-fourths full
with acetonitrile.
2 Add a magnetic stir bar and put this dish on a magnetic stir plate.
3 Put the slides in a slide rack and transfer the slide rack to the slide-staining
dish containing acetonitrile, and stir at 350 rpm for
1 minute.
4 Slowly remove the slide rack and scan the slides immediately.
The contents of the Agilent reagent kits used in this protocol are listed here.
Table 23 REPLI-g Single Cell Kit, p/n 5191-4065
Component Quantity
DNA Polymerase 2 × 55 L
PBS 2 × 1.5 mL
DTT, 1M 1 mL
H20 2 × 1.5 mL
Component Quantity
Cyanine 5-dUTP 78 L
Cyanine 3-dUTP 78 L
BSA* 15 L
* This component is not used in the GenetiSure Pre-Screen Kit for Single Cell Analysis protocol.
Table 25 Oligo aCGH/ChIP-on-chip Hybridization Kit, p/n 5188-5220 (25 slides) or 5188-5380
(100 slides)
Component Quantity
Component Quantity
Agilent now ships all microarray slides in a newly designed “secure fit” slide box.
The instructions below describe how to remove the slide box from the shipping
pouch, how to open the slide box, and how to properly remove a microarray slide.
1 Use scissors to cut below the seal and remove box from its foil pouch.
After breaking foil on microarray pouch, store microarray slides in the slide
box at room temperature (in the dark) under a vacuum desiccator or nitrogen
purge box.
Figure 17. Opening foil pouch (left) and removing the slide box (right)
2 Place the slide box on a flat surface. While stabilizing the box from the top
with one hand, use a sharp edge to cut the sealing tape on both sides of the
box before opening.
3 With one hand, firmly hold the base of the box on the sides with the
indentations (or dimples) for added grip.
Figure 19. Gripping the base at the indentations (top) and close-up of the indentations (bottom)
4 Using your free hand, grasp the lid and gently lift it away from the base as if it
is hinged. Set the lid aside.
Depending on your preference, you can reverse the hand positions so that the
left hand holds the base while the right hand grasps the lid.
Figure 20. Grasping the lid (left) and lifting the lid from the base (right)
Each microarray is printed on the side of the glass slide containing the
“Agilent”-labeled barcode. This side is called the “active” side. The numeric
barcode is on the inactive side of the slide.
You must familiarize yourself with the assembly and disassembly instructions
CAU TI O N
for use with the Agilent Microarray Hybridization Chamber (G2534A) and
gasket slides. Practice slide kits are available.
Agilent oligo microarrays formats and the resulting “microarray design files” are
based on how the Agilent microarray scanner images 1-inch × 3-inch glass slides.
Agilent designed its microarray scanner to scan through the glass slide (back
side scanning). The glass slide is securely placed in an Agilent microarray slide
holder with the “Agilent” labeled barcode facing the opening of the slide holder
(on SureScan Microarray Scanner) or facing the inside of the slide holder
(Scanner C). In this orientation, the “active side” containing the microarrays is
protected from potential damage by fingerprints and other elements. Once
securely placed, the numeric barcode, non-active side of the slide, is visible from
the outside of the slide holder.
Figure 21 depicts how the Agilent microarray scanner reads the microarrays and
how this relates to the “microarray design files” that Agilent generates during the
manufacturing process of its in situ-synthesized oligonucleotide microarrays.
Use the forms below to make notes to track your samples on microarray slides.
Position the gasket slide in the SureHyb chamber base with the label to the left
and load the samples: top row, left to right, then lower row, left to right. The array
suffix assignments from Feature Extraction will then occur in the order shown.
Arrays
Arrays
p
Sample: p
Sample: p
Sample]ËËËËËËËËËËËËËËËËËËËËË.ample:
B
A
R
C
O Sample: Sample: Sample]ËËËËËËËËËËËËËËËËËËËËË.ample:
D
E
A
Array 2
2_1
1 Array
A 2_2
2 2 Array
A 2_3
2 3 Array
A 2_4
2 4
www.agilent.com
Agilent Technologies, Inc. 2018, 2020
*G4410-90003*
G4410-90003 Revision C0