221-79041C AOC20i 20s IM EN
221-79041C AOC20i 20s IM EN
221-79041C AOC20i 20s IM EN
Dec. 2020
AUTOINJECTOR
AOC™-20i/AOC™-20i Plus
AUTOSAMPLER
AOC™-20s
Instruction Manual
Read this manual thoroughly before you use the product.
Keep this manual for future reference.
This page is intentionally left blank.
Introduction
IMPORTANT
•If the user or usage location changes, ensure that this Instruction Manual is always kept together with the product.
•If this manual or a product warning label is lost or damaged, immediately contact your Shimadzu representative to
request a replacement.
•To ensure safe operation, read all Safety Instructions before using the product.
•To ensure safe operation, contact your Shimadzu representative if product installation, adjustment, re-installation
(after the product is moved), or repair is required.
Notice •Information in this manual is subject to change without notice and does not represent
a commitment on the part of the vendor.
•Any errors or omissions which may have occurred in this manual despite the utmost
care taken in its production will be corrected as soon as possible, although not
necessarily immediately after detection.
•All rights are reserved, including those to reproduce this manual or parts thereof in
any form without permission from Shimadzu Corporation.
•AOC, GCsolution, Chromatopac, and GCMSsolution are trademarks of Shimadzu
Corporation.
Windows is a registered trademark of Microsoft Corporation in the United States and/
or other countries.
Third-party trademarks and trade names may be used in this publication to refer to
either the entities or their products/services, whether or not they are used with
trademark symbol "TM" or "®".
•In this booklet, Autoinjector AOC-20i/AOC-20i Plus is indicated as "AOC-20i" or
"Autoinjector".
AOC-20i/AOC-20i Plus/AOC-20s i
Indications Used in This Manual
Indicates a potentially hazardous situation which, if not avoided, may result in minor to moderate injury or
Caution
equipment damage.
Emphasizes additional information that is provided to ensure the proper use of this product.
Note
Safety Instructions
The AOC-20i Autoinjector and AOC-20s Autosampler automatically inject liquid samples into a gas chromatograph.
To ensure safe operation of the instrument, read these Safety Instructions carefully before use. Observe all of the
WARNINGS and CAUTIONS described in this section. They are extremely important for safety.
1. Do not use for any purpose other than the above-mentioned types of analyses.
2. Follow the procedures described in the Instruction Manual, otherwise the instrument may not be used safely.
3. Observe all warnings and precautions.
4. Do not disassemble or modify the unit without the express approval of an authorized Shimadzu representative.
5. For internal repair of the product, contact your Shimadzu representative.
Product Applications
! WARNING
• Do NOT use the product for purposes other than those specified.
This product is automatically inject liquid samples into a gas chromatograph.
Using it for an unspecified purpose may result in an accident.
This product (including accessories) is manufactured for measuring and industrial purposes, and not suitable for
household usage.
Installation
To ensure safe operation, contact your Shimadzu representative for product installation, adjustment, or re-
installation (after the product is moved) is required.
ii AOC-20i/AOC-20i Plus/AOC-20s
Operation
! WARNING
• USE CARE WHEN HANDLING FLAMMABLE SOLVENTS OR SAMPLES
To prevent flammable solvents or samples (such as carbon disulfide) from evaporating and filling the installation
site, securely tighten the cap that holds the septum in place on all vials (including rinse solvent and waste liquid
vials).
@ NOTE
Installation Precautions
Be sure to use the optional sample cooling fan when working with solvents with particularly strong flammability, such
as carbon disulfide.
For other flammable solvents or samples, it is also recommended to use the sample cooling fan.
! WARNING
• Contact your Shimadzu Service Representative for removing the cover of the power supply unit.
Otherwise, electric shock could occur.
For the autoinjector/autosampler featuring a built-in power supply unit in the gas chromatograph, bring the GC
to an emergency shutdown by the following procedure.
1. Turn OFF the gas chromatograph power switch.
2. Shut off the power supply to the gas chromatograph.
•If the power cable is fixed with screws to the switchboard, turn OFF the switch on the switchboard.
•If the power cable is connected with a plug, unplug the power cable.
Power switch
Power cable
Figure 1 Power switch and power cable (External power supply unit)
iv AOC-20i/AOC-20i Plus/AOC-20s
Warning Labels
In order to ensure safety, warning labels are attached in places requiring caution.
If a warning label is lost or damaged, obtain a new label through your Shimadzu representative and attach it in the
correct position. See "Warning Labels on the Equipment" in this Instruction Manual for details on the positions where
the labels are attached.
Autoinjector
Figure 2
HIGH
VOLTAGE
Figure 3
AOC-20i/AOC-20i Plus/AOC-20s v
Warranty
1. Period: Please contact your Shimadzu representative for information about the period of this
warranty.
2. Description: If a product/part failure occurs for reasons attributable to Shimadzu during the warranty
period, Shimadzu will repair or replace the product/part free of charge. However, in the
case of products which are usually available on the market only for a short time, such as
personal computers and their peripherals/parts, Shimadzu may not be able to provide
identical replacement products.
3. Limitation of Liability: 1) In no event will Shimadzu be liable for any lost revenue, profit or data, or for special,
indirect, consequential, incidental or punitive damages, however caused regardless of the
theory of liability, arising out of or related to the use of or inability to use the product, even
if Shimadzu has been advised of the possibility of such damage.
2) In no event will Shimadzu's liability to you, whether in contract, tort (including negligence),
or otherwise, exceed the amount you paid for the product.
4. Exceptions: Failures caused by the following are excluded from the warranty, even if they occur
during the warranty period.
1) Improper product handling
2) Repairs or modifications performed by parties other than Shimadzu or Shimadzu
designated companies
3) Product use in combination with hardware or software other than that designated by
Shimadzu
4) Computer viruses leading to device failures and damage to data and software, including
the product's basic software
5) Power failures, including power outages and sudden voltage drops, leading to device
failures and damage to data and software, including the product's basic software
6) Turning OFF the product without following the proper shutdown procedure leading to
device failures and damage to data and software, including the product's basic software
7) Reasons unrelated to the product itself
8) Product use in harsh environments, such as those subject to high temperatures or
humidity levels, corrosive gases, or strong vibrations
9) Fires, earthquakes, or any other act of nature, contamination by radioactive or hazardous
substances, or any other force majeure event, including wars, riots, and crimes
10) Product movement or transportation after installation
11)Consumables and equivalent items
Recording media such as CD-ROMs are considered consumable items.
* If there is a document such as a warranty provided with the product, or there is a separate contract agreed upon that includes warranty
conditions, the provisions of those documents shall apply.
vi AOC-20i/AOC-20i Plus/AOC-20s
After-Sales Service and Availability of Replacement Parts
After-Sales Service If any problem occurs with this product, perform an inspection and take
appropriate corrective action as described in "6 Troubleshooting" P.151 of this
manual.
If the problem persists, or the symptoms are not covered in the
Troubleshooting section, contact your Shimadzu representative.
Replacement Parts Availability Replacement parts for this product will be available for a period of seven (7)
years after the product is discontinued. Thereafter, such parts may cease to be
available.
Note, however, that the availability of parts not manufactured by Shimadzu
shall be determined by the relevant manufacturers.
In order to maintain the instrument's performance and obtain accurate measurement data, daily inspection and
periodic inspection/calibration are necessary.
•For daily maintenance, inspection, and replacement parts, see "5 Maintenance" of this Instruction Manual.
•Periodic inspection/calibration should be requested to your Shimadzu representative.
•Replacement cycles described for periodic replacement parts are rough estimate.
Replacement may be required earlier than the described replacement cycles depending on usage environment and
frequency.
Disposal Precautions
This product must be disposed of in compliance with the respective national or local government regulations.
Equipment marked with this symbol indicates that it was sold on or after 13th
August 2005, which means it should not be disposed of with general
household waste. Note that our equipment is for industrial/professional use
only.
With your co-operation we are aiming to reduce contamination from waste electronic and electrical equipment and
preserve natural resources through re-use and recycling.
Do not hesitate to ask Shimadzu service representative, if you require further information.
This product complies with European standard EN61326-1, class A for electromagnetic interference (Emissions) and
basic electromagnetic environment requirement (Table1) for electromagnetic susceptibility (Immunity).
AOC-20i/AOC-20i Plus/AOC-20s ix
This page is intentionally left blank.
x AOC-20i/AOC-20i Plus/AOC-20s
Contents
Introduction.............................................................................................................................. i
Warning Labels....................................................................................................................... v
Autoinjector ....................................................................................................................................... v
Power Supply Unit............................................................................................................................. v
Warranty ................................................................................................................................ vi
1 Overview
1.1 Introduction ..........................................................................................................1
1.4 Unpacking............................................................................................................5
1.5 Accessories List...................................................................................................6
1.5.1 Autoinjector Standard Accessories .................................................................... 6
1.5.2 AOC Power Supply Unit ..................................................................................... 6
1.5.3 Autoinjector Installation Parts............................................................................. 8
1.5.4 Autosampler Standard Accessories ................................................................. 11
1.5.5 Autosampler Installation Parts (for GC-2014, GC-14A/B) ................................ 11
AOC-20i/AOC-20i Plus/AOC-20s xi
Contents
2 Autoinjector/Autosampler Installation
2.1 Installing the Autoinjector to the Gas Chromatograph....................................... 21
2.1.1 Installation to the Gas Chromatograph ............................................................ 21
2.1.2 Mounting Spacers on GC-2014/GC-14A/B ...................................................... 23
2.3 Setting Vial Racks to the Rack Guide, Loading Rack into the Autoinjector....... 38
2.3.1 Setting Vial Racks to the Short Rack Guide..................................................... 38
2.3.2 Setting Vial Racks to the Long Rack Guide (Optional) .................................... 39
2.3.3 Loading Rack into the Autoinjector .................................................................. 41
2.3.4 Washing with 3 Different Solvents ................................................................... 45
3 Operation
3.1 Basic Operation .................................................................................................79
3.1.1 Loading Vials in the Rack................................................................................. 79
3.1.2 Turning the Power ON/OFF ............................................................................. 80
3.1.3 AOC-20i Keypad .............................................................................................. 81
3.1.4 Setting Parameters .......................................................................................... 83
3.1.5 Basic Injection Procedure ................................................................................ 85
3.1.6 Operation Precautions ..................................................................................... 88
4 External Control
4.1 Preparation ...................................................................................................... 134
5 Maintenance
5.1 Cleaning the Vials............................................................................................ 145
6 Troubleshooting
6.1 Error Codes .....................................................................................................151
6.1.1 Autoinjector Error Codes ................................................................................ 151
6.1.2 Autosampler Error Codes............................................................................... 153
6.1.3 Sub-injector Error Codes................................................................................ 154
6.1.4 Miscellaneous Error Messages ...................................................................... 154
AOC-20i/AOC-20i Plus/AOC-20s xv
Contents
1.1 Introduction 1
Shimadzu Autoinjector AOC-20i/Autosampler AOC-20s automatically injects liquid samples to gas
chromatographs (GC). It can be mounted to Shimadzu gas chromatographs such as GC-14A/B, GC-15A,
GC-16A, GC-17A Ver.1/2/3, GC-1700, GC-18A, and GC-2010/2010Plus/2014/2025. 1
AOC-20i/AOC-20s for GC-2010/2010Plus/2014/2025 is a built-in power type, where the AOC power
supply unit is incorporated in the GC main unit.
The existing "External power type" can also be used with GC-2010/2010Plus/2014/2025, if the AOC-20i
1
installation kit is attached. See "5.5 Optional Parts" P.149.
1
1
1
1
1
1
1
1
1
1
1
1
1
AOC-20i/AOC-20i Plus/AOC-20s 1
1 Overview
1.2 Components
Autoinjector AOC-20i/Autosampler AOC-20s consists of three units: 1 Autoinjector, 2 Power supply unit,
and 3 Autosampler.
(The configuration without the autosampler is also available.)
1
1
3 3
Built-in
2 2
Figure 1-1
2 AOC-20i/AOC-20i Plus/AOC-20s
1.3 System Configuration and Operation
Figure 1-2
Complete the setting to make the autoinjector/autosampler available via the PC software.
In principle, set the parameters for the autoinjector/autosampler via the PC software.
Operate the gas chromatograph keypad only for the function settings not accessible via the
GCsolution software.
Operate the autoinjector keypad only for the function settings not available even via the gas
chromatograph.
Figure 1-3
In principle, set the parameters for the autoinjector/autosampler via the gas chromatograph.
Operate the autoinjector keypad for the function settings not accessible via the gas chromatograph.
Configure the AOC on the analytical flow line using the [SET] key [Line Config] on the gas
chromatograph. (Refer to the instruction manual of the gas chromatograph.)
AOC-20i/AOC-20i Plus/AOC-20s 3
1 Overview
Figure 1-4
In principle, set the parameters for the autoinjector/autosampler via the computer or chromatopac.
Operate the autoinjector keypad for the functions not accessible via the operating software.
Figure 1-5
@ NOTE
When GC-2010/2010Plus/2014/2025 or CBM-101/102 is not linked with AOC-20i/AOC-20s, perform the
operation procedure for (4).
4 AOC-20i/AOC-20i Plus/AOC-20s
1.4 Unpacking
1.4 Unpacking
Carefully inspect the condition of the shipping container before unpacking your AOC-20i/AOC-20s. If you
find any severe damage or other problems with the container, contact your Shimadzu representative
immediately.
1
After unpacking the container, verify that all items have been included by referring to the Accessories List
given in "1.5 Accessories List" P.6. If there are any missing parts or damaged items, contact your
Shimadzu representative immediately.
AOC-20i/AOC-20i Plus/AOC-20s 5
1 Overview
6 AOC-20i/AOC-20i Plus/AOC-20s
1.5 Accessories List
4 Stand S221-44992-92 2
4 Stand S221-44992-92 2
AOC-20i/AOC-20i Plus/AOC-20s 7
1 Overview
6 Spacer S037-38786-29 1
8 Washer, M4 S023-66140-01 2
9 Nut, M4 S023-04140 1
0 Wrench, 7 × 8 S086-03047-04 1
6 Washer, M4 S023-66240-01 2
7 Nut, M4 S023-04140 1
8 Wrench, 7 × 8 S086-03047-04 1
a Washer, M3 S023-77030 2
b Clamp S072-60330-01 2
8 AOC-20i/AOC-20i Plus/AOC-20s
1.5 Accessories List
4
Spacer, rear right
S221-73794
1
1
1
5 Spacer, adjuster S221-45215 1
6 Spacer S037-38786-29 1
8 Washer, M4 S023-66140-01 2
9 Nut, M4 S023-04140 1
0 Wrench, 7 × 8 S086-03047-04 1
0 Screw, M3 × 6 S020-46134 4
a Washer, M3 S023-66130-01 4
c Washer, M4 S023-66140-01 2
d Nut, M4 S023-04140 1
e Wrench, 7 × 8 S086-03047-04 1
AOC-20i/AOC-20i Plus/AOC-20s 9
1 Overview
c Screw, M3 × 6 S020-46534 2
d Screw, M4 × 12 S020-46549 4
f Nut, M4 S023-04140 1
g Wrench, 7 × 8 S086-03047-04 1
10 AOC-20i/AOC-20i Plus/AOC-20s
1.5 Accessories List
3 4 ml vial rack
S221-34266-93
2
1
1
4
e Screw, M4 × 8 S020-46547 1
f Screw, M4 × 12 S020-46549 2
g Screw, M4 × 16 S020-46551 3
h Screw, M4 × 25 S020-46553 4
i Screw, M4 × 30 S020-46554 4
j Screw, M3 × 8 S221-46535 6
AOC-20i/AOC-20i Plus/AOC-20s 11
1 Overview
1.6 Features
1.6.1 Autoinjector Features
(1) Variable injection speed
(3) Three different solvents can be used for washing the syringe.
(4) You can set AOC parameters via the gas chromatograph main unit (when using GC-2010/2010Plus/
2014/2025).
12 AOC-20i/AOC-20i Plus/AOC-20s
1.7 Specifications
1.7 Specifications
1.7.1 Autoinjector Specifications
Injection Method: Liquid sample injection via microsyringe
0.5 μl, 5 μl, 10 μl, 50 μl, and 250 μl (optional except 10 μl)
1
! CAUTION
• Be sure to use only a designated syringe. Failure to do so may damage the microsyringe,
vials, or other parts, or cause instrument malfunctions.
AOC-20i/AOC-20i Plus/AOC-20s 13
1 Overview
Number of
1.5 ml Vials 4 ml Vials
Rack
Auto Sampler installation bracket (Standard accessory) 6 150 96
When Using the Heating 5 125 80
Optional Cooling/
Heating Unit Cooling 3 75 48
@ NOTE
When using the Mounting bracket for GC-2014 (for AOC-20), the vial racks indicated by arrows cannot be
mounted.
Dimensions, Weight: 320 (dia.) × 135 (H) mm, 2.4 kg
14 AOC-20i/AOC-20i Plus/AOC-20s
1.8 Part Names
9 4
7
6
Figure 1-7
5 Needle guide
AOC-20i/AOC-20i Plus/AOC-20s 15
1 Overview
1.8.2 Autosampler
Figure 1-8
1 Arm 4 Tray
3 Vial rack
16 AOC-20i/AOC-20i Plus/AOC-20s
1.8 Part Names
! WARNING
1
• HIGH VOLTAGE
Risk of electric shock. Only authorized Shimadzu service personnel can remove the cover.
1 1
1
2 2
2
3 3
3
4 4 4
5 5 5
6 6 6
Figure 1-9 For GC-2010/2010Plus Figure 1-10 For GC-2014 Figure 1-11 For GC-2025
AOC-20i/AOC-20i Plus/AOC-20s 17
1 Overview
! WARNING
• Be sure to use only a fuse of the specified type and rating.
Be sure to unplug the power supply cable before replacing a fuse.
Figure 1-12
9 8 7 6 5 4 3 2
Figure 1-13
5 RS-232C terminal
18 AOC-20i/AOC-20i Plus/AOC-20s
1.9 Operation Environment
AOC-20i/AOC-20i Plus/AOC-20s 19
1 Overview
20 AOC-20i/AOC-20i Plus/AOC-20s
2 2 Autoinjector/
Autosampler Installation
2
1 Place the autoinjector directly upon the four installation pins located on top of the gas
chromatograph. Press the autoinjector down until it is set onto the septum nut and
2
completely secured.
Then, plug in the AOC cable to connect the AOC to the GC, if it is unplugged.
! CAUTION
2
• Be sure to connect/remove the AOC cable when the AOC power is OFF.
Connecting/removing the cable while the AOC power is ON will result in instrument malfunction 2
or failure.
1 Place directly upon the installation pins until the autoinjector is completely secured. 2
2 AOC cable: Align the parts indicated by arrows to plug/unplug the cable.
3 Place the AOC cable on the hook (accessory) to avoid interrupting the rack movement.
2
1
2
2
2
2
3
2
2
Figure 2-1
2
2
AOC-20i/AOC-20i Plus/AOC-20s 21
2 Autoinjector/Autosampler Installation
@ NOTE
If the autoinjector is not completely secured, the error message "-11" appears on the display when
the power is turned ON. In this case, check whether the autoinjector is properly mounted, and press
it thoroughly until it is securely fixed.
@ NOTE
Even though the AOC cable is connected, you can install/remove the autoinjector. Remove the AOC
cable when moving the autoinjector to places where the AOC cable cannot reach (e.g. when
carrying the autoinjector away).
@ NOTE
If the autoinjector is used for a GC-2014 or GC-14A/B, Shimadzu personnel remove at installation
the plate spring, screws, and spacers (shown in Figure 2-2) used for mounting to GC-2010 or other
gas chromatographs. These parts are necessary when using the autoinjector with GC-2010, etc.
Please store them in a safe place.
The plate spring, screws, and The hole for the septum nut
spacers here have been
removed. (They are required
for a GC-2010, GC-17A or
other chromatographs.) Bottom view of autoinjector
Figure 2-2
22 AOC-20i/AOC-20i Plus/AOC-20s
2.1 Installing the Autoinjector to the Gas Chromatograph
Spacer, Rear
S221-70185
1 2 3 4
Figure 2-3
When installing the autoinjector on GC-2014 and GC-14A/B, you can mount up to two sets of the spacers
simultaneously. The positions can be selected from the following combinations.
(1 and 2), (2 and 3), (3 and 4), and (1 and 4)
AOC-20i/AOC-20i Plus/AOC-20s 23
2 Autoinjector/Autosampler Installation
To install the autoinjector to the injection port with no spacer mounted, move the spacers by the following
procedure:
For GC-2014
M3 Spring washer
M4 Spring
washer
Figure 2-4
The front left spacer is an assembly of Spacer, front left bottom (S221-44892), Nut (S023-04140), and
Spacer, adjuster (S221-70184).
2 Loosen the knurled screw fixing the INJ/DET cover, and remove the cover.
3 Using the wrench (accessory), remove two of the front spacers mounted to the INJ/DET
mounting plate. Remount them to the designated positions by referring to Figure 2-3.
4 Using the wrench, remove two of the rear spacers mounted to the INJ/DET mounting
plate. Remount them to the designated positions by referring to Figure 2-3.
5 Reinstall the INJ/DET cover to the original position, and tighten the knurled screw to fix
the cover.
@ NOTE
• Be careful not to lose the spring washers, which have been set with spacers.
• Pay extra attention not to confuse the right and left front spacers.
• If the autoinjector is unstable when it is installed, open the oven door and adjust the length of the front left
spacer.
Loosening the nut in the middle of the front left spacer allows you to adjust its height of the spacer top.
After completing the adjustment, tighten the nut to fix the spacer top securely.
24 AOC-20i/AOC-20i Plus/AOC-20s
2.1 Installing the Autoinjector to the Gas Chromatograph
For GC-14A/B
Perform procedures similar to that for GC-2014 to mount spacers designated in the Figure 2-5 below.
M4 Spring washer
M4 Spring washer
2
Spacer, Front Left (Height
Adjuster) (S221-44726)
Adjust the spacer height when
the autoinjector is unstable after
installation. Fix with the M4 nut.
(Use the accessory wrench.)
M4 Nut
Figure 2-5
AOC-20i/AOC-20i Plus/AOC-20s 25
2 Autoinjector/Autosampler Installation
2 Hook your finger at the half-circle depression on the autoinjector top, and pull the door
open.
@ NOTE
Check that "OP" appears on the display when the door is open. If it does not appear, contact your
Shimadzu representative.
[STOP] key
Figure 2-6
26 AOC-20i/AOC-20i Plus/AOC-20s
2.2 Mounting and Removing the Syringe
4 Inspect the new syringe. Check if the needle is fixed to the syringe barrel in a straight
manner.
You can check the straightness of the needle by rolling the syringe on a flat surface (Avoid the flange
interference). If the needle is set at an angle, the distance to the needle tip from the surface keeps
changing during the rotation.
If you mount the microsyringe with an angled needle to the autoinjector, the needle may be bent during
the injection operation.
When the needle is set at an angle, straighten the needle by slightly bending from the needle base (i.e.,
where the needle is fixed to the syringe barrel) with your fingers.
2
5 Attach the barrel holder (used for syringe removal: P/N S221-44780) onto the syringe
barrel.
The barrel holder should be positioned within the range shown in Figure 2-7.
1 Flange
2 4 - 6 mm
Barrel holder Flange
3
The syringe is positioned correctly
when the graduations are visible.
Figure 2-7
AOC-20i/AOC-20i Plus/AOC-20s 27
2 Autoinjector/Autosampler Installation
6 Lift the syringe drive unit to its highest position. Pull up the plunger drive unit at least
16 mm by holding the knurled screw.
If it is difficult to pull up the plunger drive unit, rotate the plunger motor pulley by hand.
Knurled screw
Lift up at
least 16 mm.
Figure 2-8
@ NOTE
Until doing procedure 3 (pressing the [STOP] key while the door is open), you cannot lift up the
plunger drive unit since the plunger motor power is still ON. Be sure to complete procedure 3 before
proceeding.
7 Open the syringe clip while the syringe drive unit is located at its highest position.
8 Press the syringe plunger down to its lowest position. Tilt the syringe top toward you
and insert the needle tip to the needle guide in an oblique manner.
At this point, the scale marks on the syringe barrel should be directed toward you.
28 AOC-20i/AOC-20i Plus/AOC-20s
2.2 Mounting and Removing the Syringe
9 Set the syringe barrel all the way until the clips of the syringe drive unit hold the barrel
securely. Make sure that the plunger flange is fit on Clip 3.
If the barrel is not properly set here, the needle may be bent during operation.
Verify that the barrel holder is located in the position specified in Figure 2-9, and that the needle tip is
inserted into the needle guide.
9-1, 9-2
Clip 3
9-3
Clip 3
Clip 2
Clip 2
8-1 10
7
Clip 1
8-3 Clip 1
Needle guide
9-5
8-2 9-4
Left side view Front view
Figure 2-9
@ NOTE
With the barrel holder and syringe flange placed in the direction specified in Figure 2-9, install the
syringe to the syringe drive unit until it is completely secured. If the placement direction is not correct
or the syringe is not mounted securely, the needle may be bent during operation.
AOC-20i/AOC-20i Plus/AOC-20s 29
2 Autoinjector/Autosampler Installation
@ NOTE
Be sure to close the syringe clip before operation. If the syringe drive unit operates with the syringe
clip opened, the clip may fall off or be broken.
An extra syringe clip is supplied with the autoinjector as a standard accessory. To install the new
syringe clip, follow the instructions in Figure 2-10.
Syringe clip
(P/N S221-48989-91)
Figure 2-10
@ NOTE
Before mounting the syringe, be sure to press down the syringe plunger to its lowest position. If the
plunger is not completely pressed down, it may be bent by being caught on the plunger drive unit.
30 AOC-20i/AOC-20i Plus/AOC-20s
2.2 Mounting and Removing the Syringe
11 Verify that the syringe is securely set to the syringe drive unit, and the syringe plunger
pressed down to its lowest position. Attach the plunger holder to the plunger button.
Be sure to insert the holder all the way until it contacts the plunger button.
Plunger holder
Plunger button 2
Figure 2-11
12 Loosen the knurled screw on the plunger drive unit to create a gap of 7 mm or longer
between the screw and the unit.
@ NOTE
Loosen the knurled screw on the plunger drive unit sufficiently. If the top of the plunger holder
contacts the screw, you cannot fasten the screw properly and the plunger may be damaged.
Hole
Knurled screw
Figure 2-12
AOC-20i/AOC-20i Plus/AOC-20s 31
2 Autoinjector/Autosampler Installation
13 Pull down the plunger drive unit, which was lifted up 16 mm in procedure 6, until its
bottom hole is set on the plunger holder top.
To pull down, place your fingers on the upper part of the plunger drive unit or the knurled screw. If it is
hard to pull down, rotate the pulley of the plunger motor by hand.
Hole
1
Knurled screw
Top of plunger holder 2
Figure 2-13
Moves downward.
Small clearance
should be found.
Figure 2-14
! CAUTION
• DO NOT place your hand near the drive unit while the plunger drive unit is operating.
The plunger drive unit will then stop, leaving a small gap between it and the syringe drive unit.
32 AOC-20i/AOC-20i Plus/AOC-20s
2.2 Mounting and Removing the Syringe
15 Lightly press down the plunger holder so that the syringe plunger tip contacts the
syringe barrel bottom.
Check to be sure that the plunger tip is located at the barrel zero point.
Knurled screw
Figure 2-15
16 Tighten the knurled screw on the plunger drive unit to fix the plunger holder.
@ NOTE
The plunger tip should already be in contact with the barrel bottom and located at the zero point
upon completing procedure 8. Lightly press down the plunger holder hook with your fingers, and
tighten the knurled screw securely.
DO NOT press down the plunger holder too strongly as you may bend the plunger.
Figure 2-16
AOC-20i/AOC-20i Plus/AOC-20s 33
2 Autoinjector/Autosampler Installation
@ NOTE
After fixing the plunger, make sure that the plunger tip is in contact with the syringe barrel bottom
(i.e., the plunger tip is located at the zero point of the barrel). When the plunger is not in contact with
the barrel bottom, replace the plunger holder and follow the syringe mounting procedure from the
beginning.
Verify that there is a small gap between the plunger drive unit and the syringe drive unit as shown in
Figure 2-16, illustrating the plunger drive unit in its home position.
! CAUTION
• Always turn ON the instrument before mounting the microsyringe. If the power is OFF,
the plunger drive unit cannot move to its home position, which may cause improper
sample aspiration.
@ NOTE
Once the plunger holder is used, the knurled screw creates a small dent in it. Therefore, it is recommended
to replace the plunger holder when replacing the microsyringe.
@ NOTE
The plunger motion becomes constricted after repeated air aspiration/discharge, or when the syringe is left
unused for a long time. Before mounting the microsyringe to the autoinjector, check to be sure that the
plunger moves smoothly. Additionally, always maintain the plunger in good condition by rinsing it with
solvents, etc.
If the plunger motion is too constricted, the autoinjector aborts the operation, displaying the error "-03". To
avoid this, always maintain the plunger in good condition.
The motion of 50 μl syringes and 250 μl syringes (both optional) especially become constricted in shorter
period of time than the standard 10 μl syringes. After using these syringes, be sure to perform maintenance
on them immediately.
@ NOTE
Remember that syringes are consumables. The plunger and barrel become loose after long periods of use,
which may cause sample leaks. The sample leaks will deteriorate analysis reproducibility. Periodically check
that there is no leak when replacing the gas chromatograph injection port septum, etc.
34 AOC-20i/AOC-20i Plus/AOC-20s
2.2 Mounting and Removing the Syringe
! CAUTION
• OCI Syringe (for only when using GC-17A (OCI-17))
Use the OCI syringe (P/N S221-37282-02: optional) when using the OCI as an injection port.
The needle of the OCI syringe thins towards its tip, allowing the introduction of samples directly to the
capillary column. To set the OCI syringe to the autoinjector, use the same procedure as that for other
syringes, except set the syringe injection position 3 mm higher than usual. If you use the autoinjector
without the parameter change, you may break your syringe barrel.
When using a standard syringe again, be sure to reset to the original value, 0 mm. If you keep using
the autoinjector with the injection position 3 mm higher, it may deteriorate analysis reproducibility.
2
1 Press the [FUNCTION] key.
(The FUNCTION number set in the last operation is displayed.)
2 Access F27 using the arrow keys, and press the [ENTER] key.
3 Change the displayed value to "003", and press the [ENTER] key.
@ NOTE
This parameter can be set within the range from 0 up to 22 mm (upward direction).
1 If the autoinjector power is ON, press the [RESET] key to lift up the syringe drive unit
and to return the plunger to its home position.
Figure 2-17
AOC-20i/AOC-20i Plus/AOC-20s 35
2 Autoinjector/Autosampler Installation
If the autoinjector power is OFF, open the door and verify that the syringe drive unit is located at its
highest position. If not, lift the syringe drive unit up to its highest position with your fingers.
Figure 2-18
3 Loosen the knurled screw fixing the plunger, and lift up the plunger drive unit 16 mm or
more.
If it is difficult to pull up the plunger drive unit, rotate the plunger motor pulley by hand. When the AOC
power is OFF, even though the plunger drive unit is not located in its home (lowest) position, you can
remove the syringe by lifting the plunger drive unit up 16 mm or more from the current unit position.
Top of plunger
holder
Figure 2-19
4 Remove the plunger holder from the plunger. Then, open the syringe clip.
36 AOC-20i/AOC-20i Plus/AOC-20s
2.2 Mounting and Removing the Syringe
5 Verify that the plunger is fully pressed down to its lowest position. Hold the barrel
holder and tilt the barrel down toward you to remove the syringe from the syringe drive
unit.
Barrel holder
Figure 2-20
AOC-20i/AOC-20i Plus/AOC-20s 37
2 Autoinjector/Autosampler Installation
Figure 2-21
The standard accessories are: 1 short rack guide (S221-45608-91), 2 4 ml vial rack (S221-32949-01)
and 3 1.5 ml vial rack (S221-45609-92) without sockets. The short rack guide accommodates 6 vials.
To load the vial racks, spread the rack guide slightly apart and insert both 3 1.5 ml vial rack and 2 4 ml
vial rack, so that the side pegs fit into the corresponding holes in each side of 1 rack (as shown in Figure
2-22).
Figure 2-22
38 AOC-20i/AOC-20i Plus/AOC-20s
2.3 Setting Vial Racks to the Rack Guide, Loading Rack into the Autoinjector
Figure 2-23
Figure 2-24
To assemble the long vial racks, first connect two 1.5 ml vial racks, 3 and 3' end-to-end, by inserting the
pegs of 3 vial rack into the sockets of 3' vial rack.
Sockets
3 1.5 ml vial rack (without sockets)
Pins
Figure 2-25
AOC-20i/AOC-20i Plus/AOC-20s 39
2 Autoinjector/Autosampler Installation
The assembled vial racks are inserted into the long rack guide along with 2 4 ml vial rack (for solvent/
waste liquid vials), as shown in Figure 2-26. This is the configuration at shipment.
3' 3 2
Figure 2-26
As for the short rack guide, the pegs of 2, 3, and 3', fit into the sockets of 1 long rack guide. Spread the
rack guide apart slightly, and insert the vial racks.
3'
Figure 2-27
40 AOC-20i/AOC-20i Plus/AOC-20s
2.3 Setting Vial Racks to the Rack Guide, Loading Rack into the Autoinjector
@ NOTE
When using PC software such as GCsolution, be sure to perform this setting via the PC software. Otherwise,
the set values will be erased. For details, refer to the system configuration section of the software instruction
manual.
For GC-2010/2010Plus/2014/2025 2
1 Press the [OPTION] key on the gas chromatograph to display the [AOC Parameters] window.
2 Press [OtherPara (PF3)] of the PF menus to display the [AOC Other Parameters] window.
3 Change the rack guide setting according to the type of rack guide used.
Figure 2-28
@ NOTE
When using GC-2010/2010Plus/2014/2025, be sure to change the parameters via the gas chromatograph.
DO NOT change the parameters by operating the autoinjector keypad, because when rebooting the
instrument all parameters on the autoinjector are updated to the AOC parameters set on the gas
chromatograph.
For details on the gas chromatograph operation, refer to the instruction manual of the gas chromatograph.
@ NOTE
When the [AOC Parameters] window does not appear even if you press the [OPTION] key many times,
press the [SET] key on the gas chromatograph and select [LineConfg] to check if the AOC is specified for the
current analytical flow line.
AOC-20i/AOC-20i Plus/AOC-20s 41
2 Autoinjector/Autosampler Installation
Follow the procedure below to install the rack guide and racks into the autoinjector.
1 The autoinjector power can be ON or OFF. Verify that the syringe drive unit is at its
highest position.
If it is not all the way up, it will strike the rack when it is inserted. Turn OFF the power and be sure to lift up
the syringe drive unit to its highest position.
Figure 2-29
42 AOC-20i/AOC-20i Plus/AOC-20s
2.3 Setting Vial Racks to the Rack Guide, Loading Rack into the Autoinjector
2 Insert the rack guide into the autoinjector track so that the rack gears catch on the
motor pinion.
Once the gears catch, insert the rack guide a further 20 mm. The rack guide can be inserted from the left
or right if no autosampler is installed. If an autosampler is in use, insert the rack guide from the right.
Figure 2-30
3 When the autoinjector power is turned ON, the rack moves to the home position (all the
way to the left) as shown in Figure 2-31.
If the power is already ON, press the [RESET] key to move the rack to the home position.
Figure 2-31
AOC-20i/AOC-20i Plus/AOC-20s 43
2 Autoinjector/Autosampler Installation
@ NOTE
When the rack is in the home position, the motor is released, and the rack can be removed.
@ NOTE
During analysis, when the GC is in the READY state, the motor is engaged. The rack can still be removed
and inserted without pressing the [RESET] key; however, this will be more difficult because the motor has
not been released. Remove the rack, if necessary, then re-insert as in step (2). Press the [START] key once
to begin the analysis.
If the [RESET] key is pressed before removing or after inserting the rack, analysis will not be performed
continuously. Once the [RESET] key is pressed, pressing the [START] key begins a new analysis, not a
continued analysis.
44 AOC-20i/AOC-20i Plus/AOC-20s
2.3 Setting Vial Racks to the Rack Guide, Loading Rack into the Autoinjector
C Waste
(Positions for sample vials) liquid vial
B Waste
liquid vial
A Waste
liquid vial
Solvent Vial Solvent Vial
Solvent Vial
2
1
Figure 2-32
For GC-2010/2010Plus/2014/2025
1 Press the [OPTION] key on the gas chromatograph to display the [AOC Parameters]
window.
2 Press [OtherPara (PF3)] of the PF menus to display the [AOC Other Parameters]
window.
Figure 2-33
AOC-20i/AOC-20i Plus/AOC-20s 45
2 Autoinjector/Autosampler Installation
2 Access F28 using the arrow keys, and press the [ENTER] key.
3 Change the displayed value to "001", and press the [ENTER] key.
The default value is "000" (Use one solvent).
@ NOTE
Compared to the solvent flush mode using the autosampler, this function has the following differences.
• You cannot add standard samples.
• You cannot use the solvent for solvent flush method in a designated vial.
• The wash solvent is used as the solvent for solvent flush method. (The 3rd wash solvent when three
solvents are used)
@ NOTE
When FUNCTION 08 is set to the initial value ("000"), three solvents can be used (see "3.2.8 FUNCTION
Key" P.92, Solvent select F08).
46 AOC-20i/AOC-20i Plus/AOC-20s
2.4 Vial Types and Vial Mounting Procedure on Rack Guide
@ NOTE
Use the specified vials only. Use the vials for their intended purpose. 4 ml vial can be used for samples. But
optional parts are required. See "2.7.3 Using 4 ml Vials (Option)" P.70.
Be sure to use the septa and caps listed in Table 2-2 according to the vial size.
Table 2-2
CAP TYPES
SEPTUM TYPES
Concerning the part numbers for above items, see "1.5 Accessories List" P.6.
! CAUTION
• The vial caps for use with the autoinjector must be white. If the vials have no caps or have
caps of another color, the autoinjector may not recognize the vials. If a part other than that
specified is used, the sensor will be unable to detect it, resulting in possible damage. Use
only the specified caps.
AOC-20i/AOC-20i Plus/AOC-20s 47
2 Autoinjector/Autosampler Installation
! CAUTION
• A loose cap could damage the unit.
@ NOTE
Pay attention to the septum direction. The PTFE side must be directed downwards.
Cap
Septum
Vial
Figure 2-34
Replace septa frequently. The effectiveness of the septum seal deteriorates over time, so avoid using the
same septum repeatedly. Also, once the septum has been punctured, certain solvents can deteriorate the
septum and release contaminants into the sample.
The minimum amount of sample needed in a vial is determined by the length of the syringe needle, which
is 43 mm. At least approximately 3.5 mm of solvent is required for a 1.5 ml vial, as shown in Figure 2-35.
For a 4 ml vial, a minimum of approx. 7.5 mm is required.
approx. 7.5 mm
approx. 3.5 mm
Figure 2-35
48 AOC-20i/AOC-20i Plus/AOC-20s
2.4 Vial Types and Vial Mounting Procedure on Rack Guide
Changing the Depth of the Syringe Needle when Inserted into Sample Vials
Reducing the depth of the syringe needle when inserted into sample vials
• For GC-2010/2010Plus/2014/2025
Press the [SET] key on the gas chromatograph and select [LineConfg] to check if the AOC is specified
for the current analytical flow line.
1 Press the [OPTION] key on the gas chromatograph to display the [AOC Parameters] window.
2
3
Press [OtherPara (PF3)] of the PF menus to display the [AOC Other Parameters] window.
Set [Syringe Height (↑)] in millimeter (setting range: 0 - 20 mm). 2
4 Set [Syringe Height (↓)] to [0].
Figure 2-36
AOC-20i/AOC-20i Plus/AOC-20s 49
2 Autoinjector/Autosampler Installation
Increasing the depth of the syringe needle when inserted into sample vials
To increase the depth of the syringe needle when inserted into vials with small sample volume, perform the
following procedure according to the vial types.
For GC-2010/2010Plus/2014/2025
1 Press the [OPTION] key on the gas chromatograph to display the [AOC Parameters] window.
2 Press [OtherPara (PF3)] of the PF menus to display the [AOC Other Parameters] window.
3 Set [Kinds of Vials] to [1.5 ml] (default).
4 Set [Syringe Height (↓)] in millimeter (setting range: 0 - 20 mm, default value: 0).
5 Set [Syringe Height (↑)] to [0].
Figure 2-37
50 AOC-20i/AOC-20i Plus/AOC-20s
2.4 Vial Types and Vial Mounting Procedure on Rack Guide
For GC-2010/2010Plus/2014/2025
1
2
Press the [OPTION] key on the gas chromatograph to display the [AOC Parameters] window.
Press [OtherPara (PF3)] of the PF menus to display the [AOC Other Parameters] window.
2
3 Set [Kinds of Vials] to [4 ml].
4 Set [Syringe Height (↓)] in millimeter (setting range: 0 - 10 mm).
5 Set [Syringe Height (↑)] to [0].
Figure 2-38
AOC-20i/AOC-20i Plus/AOC-20s 51
2 Autoinjector/Autosampler Installation
To change 4 ml vials to 1.5 ml vials, complete the parameter setting described below. If this is not
done, the needle will hit the 1.5 vial bottom, breaking both the needle and vials.
For GC-2010/2010Plus/2014/2025
1 Press the [OPTION] key on the gas chromatograph to display the [AOC Parameters] window.
2 Press [OtherPara (PF3)] of the PF menus to display the [AOC Other Parameters] window.
3 Set [Kinds of Vials] to [1.5 ml].
4 Set [Syringe Height (↓)] to [0].
5 Set [Syringe Height (↑)] to [0].
Figure 2-39
For details on the gas chromatograph operation, refer to the instruction manual of the gas
chromatograph.
52 AOC-20i/AOC-20i Plus/AOC-20s
2.4 Vial Types and Vial Mounting Procedure on Rack Guide
The number of injections is limited by the capacity of the 4 ml rinse solvent and waste liquid vials. Assume
that a 10 μl syringe is used, with aspiration of 8 μl per wash.
The rinse solvent vial volume limit refers to the maximum number of sample injections possible before a full
solvent vial (4 ml) is emptied by aspirations.
For example, a maximum of 125 injections can be made when specifying 3 solvent washes (as indicated
by dashed line (Graph 2-1)). Assuming 125 injections, 3 × 125, or 375 solvent washes will be performed.
This will use all of the available solvent; even if more solvent washes are attempted, no solvent will be
aspirated. (The actual solvent vial volume limit will be less than the amount indicated here due to loss by
evaporation according to the solvent type.) 2
The other factor that must be considered is the waste liquid vial volume limit. This refers to the number of
sample injections possible before the waste liquid vial overflows. Since the autoinjector performs sample
washes and solvent washes, this limit value indicates the sum of both wash times.
There are several possible combinations of solvent and sample washes which yield a total of 3 washes, as
shown in Table 2-3.
Table 2-3
Frequency
Sample Wash 0 1 2 3
Solvent Wash 3 2 1 0
Use the graph to help determine whether the limiting factor is the solvent vial volume or the waste liquid
vial capacity. Use the smallest number of injections indicated by the graph.
AOC-20i/AOC-20i Plus/AOC-20s 53
2 Autoinjector/Autosampler Installation
Possible # of injections
Was
te liq
uid vial c
apac
ity limit
Solv
ent v
ial vo
lu me li
m it
! Warning
• USE CARE WHEN HANDLING FLAMMABLE SOLVENTS OR SAMPLES
To prevent flammable solvents or samples (such as carbon disulfide) from evaporating and filling the
installation site, securely tighten the cap that holds the septum in place on all vials (including rinse
solvent and waste liquid vials).
@ NOTE
Installation Precautions
Be sure to use the optional sample cooling fan when working with solvents with particularly strong
flammability, such as carbon disulfide.
For other flammable solvents or samples, it is also recommended to use the sample cooling fan.
54 AOC-20i/AOC-20i Plus/AOC-20s
2.4 Vial Types and Vial Mounting Procedure on Rack Guide
Figure 2-40
See Figure 2-23 for the assembly of 2 and 3. Figure 2-41 shows the vial arrangement in the assembled
rack.
Figure 2-41
1.5 ml vials can be placed in any position of 3. Injection starts from the vial closest to the injection port,
and proceeds in numeric order, as shown in Figure 2-42.
@ NOTE
Injection can proceed in an arbitrary order when performing an analysis schedule on GC-2010/2010Plus/
2014/2025 or batch analysis on GCsolution, CLASS-GC10, and Chromatopac GC net.
2
Injection port
Figure 2-42
AOC-20i/AOC-20i Plus/AOC-20s 55
2 Autoinjector/Autosampler Installation
Figure 2-43
Place the samples in any position. The injection starts from the position closest to the injection port (1 → 2
→ 3 in the example above).
@ NOTE
The vial mounting procedure and injection order is the same for the long rack, except that there are more
1.5 ml vial positions (12).
@ NOTE
Ensure that a waste liquid vial is in place. If not, a "-14" error message appears.
56 AOC-20i/AOC-20i Plus/AOC-20s
2.5 Replacing INJ Port Septum
! WARNING
• DO NOT perform septum replacement before the INJ cools down to 50 °C or lower.
Attempting replacement work when the INJ is still hot will cause burn injuries.
! CAUTION 2
• To prevent the screw part from being burnt, DO NOT tighten the screw and nut when
the INJ is still hot.
The injection port septa are durable for only approximately 100 injections. Exceeding this number of
injections may cause gas leaks and deteriorate analysis reproducibility. Be sure to replace the septum
before it exceeds its service life.
@ NOTE
By setting the analysis counter function on GC-2010/2010Plus/2014/2025, GC-17A Ver.2,3, 1700, and 18A,
a message informing you of the septum replacement time is displayed on the GC screen. For details, refer to
the instruction manual of the gas chromatograph.
Remove the autoinjector and replace the injection port septum by the procedures described below. They
are basically the same regardless of the gas chromatograph type, except for procedure 1.
For GC-2010/2010Plus/2014/2025
If the gas chromatograph power is ON, switch OFF the autoinjector as follows:
1 Press the [OPTION] key on the gas chromatograph to display the [AOC Parameters] window.
AOC-20i/AOC-20i Plus/AOC-20s 57
2 Autoinjector/Autosampler Installation
Figure 2-44
2 Open the autoinjector door, and verify that the syringe drive unit is located at its highest
position.
If not, lift up the syringe drive unit as shown in Figure 2-45.
Figure 2-45
! CAUTION
• If the autoinjector is removed when the syringe drive unit is not at its highest position,
the syringe needle may be broken or damaged.
58 AOC-20i/AOC-20i Plus/AOC-20s
2.5 Replacing INJ Port Septum
3 Lift the autoinjector straight up and set it aside during septum replacement. (e.g. on the
back of the INJ/DET cover in the case of GC-2010/2010Plus or GC-17A)
Connection terminal
Figure 2-46
By unplugging the cable from the terminal, you can move the autoinjector to a place where the cable
cannot reach.
@ NOTE
Be sure to use the needle guide as well as septum nut designated for the autoinjector.
(They are standard accessories for GC-2010/2010Plus/2014/2025.)
When using GC-14A/B, 15A, and 16A, pay extra attention to the shape of the septum nut. Differently
shaped septum nut must be used according to the injection port types.
Note that the OCI-14 needle guide is designated only for OCI-14, unlike those for the other injection
ports. This needle guide is longer than others. Figure 2-47 below shows the shapes and dimensions
of various autoinjector needle guides and septum nut.
AOC-20i/AOC-20i Plus/AOC-20s 59
2 Autoinjector/Autosampler Installation
φ4.9
φ24.8
φ24.8
5.0
φ24.8
12
Figure 2-47
@ NOTE
In the case of GC-14A/B, 15A, or 16A:
When there are other INJ ports next to the one currently used (e.g. DUAL packed INJ), they also
must have the septum nut and needle guide designated for the autoinjector.
If standard items (i.e., not those designated for the autoinjector) are mounted to the other INJ ports,
the septum nut may obstruct the proper installation of the autoinjector due to its height.
You can install the autoinjector by removing the septum nut and needle guide, if those INJ ports are
not used.
Note that the standard septum nut for GC-2014 does not obstruct the autoinjector installation.
INJ unit to be used
Autoinjector
Replace these septum nuts and needle guides
with those for the autoinjector.
Figure 2-48
60 AOC-20i/AOC-20i Plus/AOC-20s
2.5 Replacing INJ Port Septum
5 Following the installation procedure described in "2.1 Installing the Autoinjector to the
Gas Chromatograph" P.21, reinstall the autoinjector to the original position.
6 If you turned OFF the AOC power via GC-2010/2010Plus/2014/2025 in procedure 1, turn
ON the autoinjector as follows.
1 Press the [OPTION] key on the gas chromatograph to display the [AOC Parameters] window.
2 Set [AOC POWER] to [On], and press the [ENTER] key.
Figure 2-49
For details on the gas chromatograph operation, refer to the instruction manual of the gas
chromatograph.
AOC-20i/AOC-20i Plus/AOC-20s 61
2 Autoinjector/Autosampler Installation
Figure 2-50
62 AOC-20i/AOC-20i Plus/AOC-20s
2.6 Available Options
@ NOTE
When installing the autoinjector to the INJ located at the front of the FPD, use the FPD main unit cover
attached to the standard cooling fan.
2
Cover attached to the FPD cooling fan
Figure 2-51
If the autoinjector is installed to an INJ other than that located at the front of the FPD, use the standard FPD
cover (FPD main unit accessory). This prevents accidental contact with the FPD surface, which can cause
burns.
AOC-20i/AOC-20i Plus/AOC-20s 63
2 Autoinjector/Autosampler Installation
1 The autosampler sample tray is divided into sections for positioning the vial racks. Seat
the vial racks between the sections, in numerical order, as shown in Figure 2-52.
An empty rack does not need to be placed in the autosampler for it to function.
Figure 2-52
2 The arm of the autosampler can be rotated manually when the power is ON (unless the
autosampler is operating). If necessary, the arm can be moved out of the way when
placing a rack on the autosampler.
Figure 2-53
Figure 2-54
64 AOC-20i/AOC-20i Plus/AOC-20s
2.7 Installing the Autosampler (AOC-20s) [Optional]
! CAUTION
• Although the arm rotates manually, do not move it up and down or horizontally unless
you need to do so. Do not attempt to rotate the arm beyond the movement range.
When carrying the autosampler, do not hold it by its arm. Otherwise, it may be
damaged.
To identify vial positions, the labels shown below are available. See Figure 2-55 and Figure 2-56 for the vial
position numbers.
@ NOTE
2
When vial racks are not placed continuously, sample vial numbers are assigned sequentially skipping the
places with no vial rack. For details, see "3.4.1 Preparation", "(3) Placing vials in the autosampler tray" P.118.
AOC-20i/AOC-20i Plus/AOC-20s 65
2 Autoinjector/Autosampler Installation
Figure 2-57
66 AOC-20i/AOC-20i Plus/AOC-20s
2.7 Installing the Autosampler (AOC-20s) [Optional]
1 1.5 ml vial rack for 2 4 ml vial rack for autosampler 3 Solvent/waste liquid vial
autosampler × 1 (S221-45182) (Option) rack (4 ml) × 2
(S221-45181) (This holder is included in the 4 ml (S221-32949-01)
sample vial rack set (S221-41583-91).)
Figure 2-58
First, remove the current vial racks (for 6 vials) supplied with the autoinjector, by pulling them up. Then,
insert the new vial racks in the arrangement shown in Figure 2-59. Spread the rack guide apart slightly to
insert the vial racks. See Figure 2-60 and Figure 2-61 for details on vial placement.
Figure 2-59
AOC-20i/AOC-20i Plus/AOC-20s 67
2 Autoinjector/Autosampler Installation
1 1.5 ml vial rack for autosampler 3 Solvent/waste liquid vial rack (4 ml)
(S221-45181) (S221-32949-01)
S = Solvent
W = Waste liquid
For sample (1.5 ml) For solvent flush injection or internal standard (4 ml)
2 4 ml vial rack for autosampler (S221-45182) (option) 3 Solvent/waste liquid vial rack (4 ml)
(This holder is included in the 4 ml sample vial rack set (S221-32949-01)
(S221-41583-91).)
S = Solvent
W = Waste liquid
For sample (4 ml) For solvent flush injection or internal standard (4 ml)
As indicated in Figure 2-60 and Figure 2-61, solvent A (S) is discharged into the waste liquid A vial (W),
and likewise for B and C. When washing with 3 solvents, each of the waste liquid is discharged into the
waste liquid vials marked A to C. When washing with only one solvent, the waste liquid of sample wash is
discharged into the waste liquid vial corresponding to the used wash solvent.
Both sizes of vial racks have pegs on the bottom. When inserting the racks, the pegs should appear at the
end of the rack guide, as shown in Figure 2-62.
Figure 2-62
68 AOC-20i/AOC-20i Plus/AOC-20s
2.7 Installing the Autosampler (AOC-20s) [Optional]
@ NOTE
Use the vial racks supplied with AOC-20s autosampler by mounting to AOC-20i rack guide. There are two
types of AOC rack guides of gray or black material. There is no problem in using vial racks with colors
different from that of a rack guide.
@ NOTE
To use the autosampler, complete the following settings.
• For GC-2010/2010Plus/2014/2025
1 Press the [OPTION] key on the gas chromatograph to display the [AOC Parameters] window.
2 Press the toggle key to display [Sampler] on the PF menu. 2
3 Press the [Sampler] (PF1) to display the [Sampler] window.
4 Set [Use of the Sampler] to [Use].
Figure 2-63
For details on the gas chromatograph operation, refer to the instruction manual of the gas
chromatograph.
A solvent or standard needs to be placed in the second vial position shown in Figure 2-60 and Figure 2-61
only in solvent flush injection mode.
The 4 ml vial for solvent flush injections is placed in the vial rack. The autosampler allows use of a different
vial for solvent flush and solvent rinse to minimize cross contamination problems.
AOC-20i/AOC-20i Plus/AOC-20s 69
2 Autoinjector/Autosampler Installation
When using 4 ml vials, both the vial racks for both the autosampler and autoinjector must be changed. For
the installation procedure, see "2.7.1 Installing the Vial Racks" P.64 and "2.7.2 Setting up Autoinjector Vial
Racks" P.67.
(1) Make sure that the caps of the vials being used are securely fastened.
If a cap is loose or is at an angle, the autosampler may not be able to grip the vial properly, resulting
in an error.
(2) If the autoinjector has been removed, in order to replace the septum or glass insert, make sure that
the autoinjector is reattached in the correct position.
If the autoinjector is at an angle, the autosampler may not be able to grip the vial properly, resulting
in an error.
(3) Set the position of the autosampler arm correctly (in [Rack position] for GC-2010/2010Plus/2014/
2025, and in FUNCTION 94 for other GC). If the arm position is not adjusted correctly to the cap of
the vial in the autoinjector rack, perform the following procedure to teach the autosampler the
correct position (teaching function).
1 Turn the autoinjector on while pressing both the [Function] and [Monitor] keys. (For GC-
2010/2010Plus/2014/2025, set "AOC POWER" to "On" in the [AOC Parameters] window
on the GC, and press the [ENTER] key.)
2 After the initialization process has completed, place a vial (1.5 ml or 4 ml vial) on the left
side of the autoinjector turret.
3 Set the rack guide position for the autoinjector by the following procedure:
70 AOC-20i/AOC-20i Plus/AOC-20s
2.7 Installing the Autosampler (AOC-20s) [Optional]
For GC-2010/2010Plus/2014/2025
1 Press the [OPTION] key on the gas chromatograph to display the [AOC Parameters] window.
2 Press [OtherPara (PF3)] of the PF menus to display the [AOC Other Parameters] window.
3 Specify the rack guide position.
[0] : GC-2014/2025 (All INJ positions)
[1] : GC-2010/2010Plus Front
[2] : GC-2010/2010Plus Rear = Standard position
Per factory default, the autosampler has been programmed so that the arm moves to the rack guide
on the autoinjector mounted to the positions above (the far left INJ for GC-2014/2025). You can also
specify the desired arm destinations by using "teaching" procedures. 2
Figure 2-64
! CAUTION
• When setting the rack guide position (FUNCTION 94), be sure to set the value for the
correct rack position beforehand, and perform the teaching procedure.
AOC-20i/AOC-20i Plus/AOC-20s 71
2 Autoinjector/Autosampler Installation
5 Place a vial on the turret. Gently move the autosampler arm so that it is positioned level
with the vial.
To position the arm in the horizontal plane, first extend the arm slightly beyond the amount required and
then align it to the correct position. (The teaching will not be performed properly when the arm is fully
extended.)
To position the arm in the vertical plane, use the [SAMPLE WASH], [SOLVENT WASH], and [NUMBER
OF INJECTIONS] keys to move the arm to the position shown in Figure 2-65. (Never move the arm up or
down while holding it near the end.)
@ NOTE
There is a slight delay between when the keys are pressed and when the arm moves.
[SAMPLE WASH] key: up
[SOLVENT WASH] key: stop
[NUMBER OF INJECTIONS] key: down
Arm
2 - 4 mm
Vial
Turret
Figure 2-65
6 After the position is aligned, adjust the function value to "001", then press the [Enter]
key.
The arm will grip the cap of the vial once, then return to the home position.
72 AOC-20i/AOC-20i Plus/AOC-20s
2.7 Installing the Autosampler (AOC-20s) [Optional]
Vial destination ∗ W S W S W S
2
GC front side
for INJ4
Figure 2-66
To use this option with GC-2014, press the gas chromatograph [OPTION] key to display the [AOC
Parameters] window, and complete the following settings for each sub screen.
<Sampler sub screen>
Use of the Sampler: [Use]
<AOC Other Parameters sub screen>
Rack: [Long]
Rack Position: [0]
To use this option with GC-14A/B (GC-15A/16A), complete the following settings via the autoinjector.
FUNCTION 40........ "001" (Use sampler)
FUNCTION 93........ "001" (Long rack guide)
FUNCTION 94........ "000" (Rack guide position: 0)
@ NOTE
Be sure to perform the teaching process.
@ NOTE
DO NOT use the short rack guide when the autosampler is set for the long rack guide.
If you do, the vial racks will fall from the autoinjector gear, causing an error.
@ NOTE
When using in the solvent flush injection mode
Regardless of the INJ position, the solvent vial for solvent flush method should be set at the position marked
with (∗) in Figure 2-66. Therefore, when using this option in the solvent flush injection mode, it is required to
prepare another sampler vial rack for the solvent flush vial.
• 1.5 ml vial rack for autosampler P/N S221-45181
• 4 ml vial rack for autosampler P/N S221-45182
AOC-20i/AOC-20i Plus/AOC-20s 73
2 Autoinjector/Autosampler Installation
! CAUTION
• When using the solvent flush injection mode at INJ1
If the vial racks for autosampler are mounted as shown in Figure 2-3, the autosampler gripper
interferes with the solvent flush vial, making the vial transportation impossible.
To use the solvent flush injection mode at INJ1, perform the teaching procedure described below.
1 Mount the vial rack for autosampler to the "vial destination for INJ2" described in Figure 2-66.
2 In the [AOC Other Parameters] window, set [Rack Position] to [1]. Then, perform the teaching
procedure.
74 AOC-20i/AOC-20i Plus/AOC-20s
2.8 Controlling via Workstation
Operate the autoinjector keypad for the parameters not available via GCsolution or CLASS-GC10. For
details on the setting procedure, refer to the instruction manual for each software.
AOC-20i/AOC-20i Plus/AOC-20s 75
2 Autoinjector/Autosampler Installation
(1) The valid ranges for conventional parameters supported by CLASS-5000 are the same as those for
AOC-17.
(2) The CLASS-5000 software does not support new parameters introduced with AOC-20, such as
Aspiration to needle tip, Needle lower position, etc. However, these functions can be controlled
using the AOC-20i keys.
When connected to the GCMS-QP5000, a 2-m fiber optic cable (P/N S070-92025-52) is required, similar to
that for CLASS-GC10 software. A conventional AOC-17 Link Kit is not required.
76 AOC-20i/AOC-20i Plus/AOC-20s
2.8 Controlling via Workstation
AOC-20i/AOC-20i Plus/AOC-20s 77
2 Autoinjector/Autosampler Installation
78 AOC-20i/AOC-20i Plus/AOC-20s
3 3 Operation
Figure 3-1 3
Vial positions are designated by numbers 1 - 6 (1 - 12 for the optional long rack), as shown in the figure
above. A single vial can be placed in any of these positions; it will be automatically detected and injected. If
there are two or more vials in the rack, injections occur from the lowest numbered position to the highest.
3
For example, if vials occupy all 6 positions, injections will occur, in order, from position 1 to position 6.
3
3
3
3
3
3
3
3
AOC-20i/AOC-20i Plus/AOC-20s 79
3 Operation
For GC-2010/2010Plus/2014/2025
@ NOTE
When using GC-2010/2010Plus/2014/2025 (AOC built-in power type), when the autoinjector does not
operate even if turning ON the gas chromatograph:
• Verify that the AOC cable is connected properly.
• Press the [OPTION] key to display the [AOC Parameters] window, and verify that [AOC POWER] is set to
[On].
@ NOTE
When the [AOC Parameters] window is not displayed even if pressing the [OPTION] key many times,
perform the following procedure:
1 Press the [SET] key on the gas chromatograph to display the [Analysis Settings] window.
2 Select [LineConfg] (PF2 key) to display the [Line Configuration] window.
3 Specify the AOC for the analytical line.
@ NOTE
If the message "AOC link ok." is not displayed, check if the RS-232C cable is properly connected to the AOC
power supply unit terminal on the GC back panel.
@ NOTE
When using gas chromatographs other than GC-2010/2010Plus/2014/2025 (AOC built-in power type), turn
OFF the autoinjector power supply unit first, then turn OFF the gas chromatograph. If the GC is turned OFF
while the autoinjector is waiting for analysis, the autoinjector may start an injection. If it is necessary to turn
OFF the gas chromatograph before the autoinjector, verify beforehand that the autoinjector is in the home
position.
80 AOC-20i/AOC-20i Plus/AOC-20s
3.1 Basic Operation
Display
Arrow keys
Parameter keys
3
MONITOR key
FUNCTION key
ENTER key
RESET key
RUN light
Figure 3-2
! CAUTION
• GC-2010/2010Plus/2014/2025 [AOC Parameters] and AOC-20i Keypad
For GC-2010/2010Plus/2014/2025, operate the gas chromatograph when changing the AOC
parameters accessible from the gas chromatograph. When you have set those parameters using the
autoinjector keypad, perform the following procedure to change the AOC parameters set on the GC
according to your change:
AOC-20i/AOC-20i Plus/AOC-20s 81
3 Operation
• Display
All set values, sample numbers and error codes (numerical values displayed at an error occurrence)
are displayed here. All parameter settings are either 000 or positive numbers. Error codes are negative
numbers or numbers that begin with "E". When an invalid value is entered, "- - -" is displayed.
@ NOTE
When using GC-2010/2010Plus/2014/2025, error messages also appear on the gas chromatograph display.
• Arrow keys
These three keys change the values displayed. Each key corresponds to the right, center and left digit,
respectively. Values changed with these keys do not become valid until the [ENTER] key is pressed.
• Parameter keys
These display and set the various parameters. When one of these keys is pressed, the LED to its left
illuminates.
@ NOTE
To set the parameters corresponding to these keys on GC-2010/2010Plus/2014/2025, be sure to operate via
the gas chromatograph.
• MONITOR key
During analysis, the [MONITOR] key is automatically activated, so that the current sample number is
displayed. To display a parameter's current value, press the [MONITOR] key, then press the parameter
key (within one second).
• FUNCTION key
This key is used to set parameters (except for those parameters which have a parameter key, such as
SAMPLE WASH), and for customizing or specifying optional accessories.
When the [FUNCTION] key is pressed, a function number beginning with "F" is displayed (this is the
most recently accessed function number). To view a different function, use the arrow keys to change
the function number. Then press [ENTER] to display the current function value. To change the current
value, use the arrow keys, then press [ENTER] to save the setting.
@ NOTE
When using GC-2010/2010Plus/2014/2025, use these keys only for setting the parameters not accessible
via the gas chromatograph.
• ENTER key
This saves the displayed numerical value as the current setting. When the [ENTER] key is pressed, the
LED to its left illuminates momentarily.
• RESET key
This key returns the autoinjector to the home position. All the parameter values remain in battery-
powered reserve memory even if the [RESET] key is pressed.
@ NOTE
When using GC-2010/2010Plus/2014/2025, the same function can be accessed through [Reset] (PF menu)
in the gas chromatograph [AOC Parameters] window.
• Run light
This light is either on or blinking, according to the autoinjector status.
Light on .................indicates that the autoinjector is performing a task.
Light blinking .........indicates that the autoinjector is waiting for completion of the gas chromatograph
analysis (waiting for the GC READY status).
82 AOC-20i/AOC-20i Plus/AOC-20s
3.1 Basic Operation
• START key
This initiates the injection process.
@ NOTE
When using GC-2010/2010Plus/2014/2025, the same function can be accessed through [Start] (PF menu) in
the gas chromatograph [AOC Parameters] window.
• STOP key
Pressing the [STOP] key causes the autoinjector to stop after completing the current operation. The
process resumes where it left off when the [START] key is pressed.
@ NOTE
When using GC-2010/2010Plus/2014/2025, the same function can be accessed through [Stop] (PF menu) in
the gas chromatograph [AOC Parameters] window.
3
3.1.4 Setting Parameters
The parameter default values have been preset to the autoinjector at the factory. Additionally, those
autoinjector parameter default values have also been preset to the gas chromatograph GC-2010/
2010Plus/2014/2025 at the factory. You can change these parameters as your analytical purposes dictate.
@ NOTE
When changing the parameters accessible from the gas chromatograph GC-2010/2010Plus/2014/2025, be
sure to operate from the GC.
The parameters not accessible from GC-2010/2010Plus/2014/2025 are indicated with "∗" in "Corresponding
AOC Parameters/Other Parameters on GC" in the table shown in "3.2.8 FUNCTION Key" P.92.
To set autoinjector parameters when using GC-2010/2010Plus/2014/2025, turn ON the gas chromatograph
to link it to the autoinjector. The parameter values in the file loaded on the GC are automatically transmitted
and set to the autoinjector. You can also change parameters in the GC [AOC Parameters] window
(including sub screens). After changing the parameters, press the [ENTER] key on the GC. The changed
parameters will be transmitted to the autoinjector.
Note that even though you can change the parameters by pressing the [ENTER] key on the autoinjector,
the AOC parameters set on the gas chromatograph remain unchanged.
This means that the next time the GC is linked to the autoinjector (e.g. when the GC is restarted, etc.), the
autoinjector parameters (accessible via the GC) are overwritten by all the AOC parameters transmitted
from the GC.
For GC-2010/2010Plus/2014/2025
AOC-20i/AOC-20i Plus/AOC-20s 83
3 Operation
3 Enter [3] from the gas chromatograph keypad and press the [ENTER] key.
@ NOTE
DO NOT change the parameter by pressing the [SAMPLE WASH] key on the autoinjector.
Figure 3-3
@ NOTE
When setting parameters, the [ENTER] key must be pressed for the new parameter to take effect.
Otherwise, the old parameter value will be used. If an invalid value is entered, the autoinjector does
not process it. "Input parameter out of range" is displayed, and the previously set value remains in
effect.
84 AOC-20i/AOC-20i Plus/AOC-20s
3.1 Basic Operation
@ NOTE
When setting parameters, the [ENTER] key must be pressed for the new parameter to take effect.
Otherwise, the old parameter value will be used. If an invalid value is entered, the autoinjector does not
process it. "- - -" is displayed, and the previously set value remains in effect.
@ NOTE
Pressing any of the parameter keys causes its current value to be displayed. However, the SAMPLE
NUMBER parameter displays automatically without pressing the key. (See "3.1.6 Operation Precautions",
"Vial Detection" P.88.) Therefore, if an injection sequence is started while setting a parameter, the SAMPLE
NUMBER may be displayed. In this case, reset the parameter.
3
3.1.5 Basic Injection Procedure
The parameters required for sample injection are already set as default values in the autoinjector. (See
"3.2.2 Parameter Keys" P.89 and "3.2.8 FUNCTION Key" P.92.) Therefore you can easily perform sample
injections by (1) setting vials on the rack guide, (2) turning ON the power, and (3) pressing the [START] key
(alternatively for GC-2010/2010Plus/2014/2025, [Start] (PF menu) in [AOC Parameters] on GC.)
Create the desired injection conditions by changing the following 4 parameters, etc.: SAMPLE WASH
(Default: 2 times), SOLVENT WASH (Once), NUMBER OF INJECTIONS (Once), and SAMPLE SIZE
(1.0 μl). (See "3.2.2 Parameter Keys" P.89.)
The changed parameters are saved to the autoinjector even after the power is turned OFF.
Table 3-1
GC-2010/2010Plus/
Key Name 2014/2025 Description Setting Range Default Value
Item Name
SAMPLE WASH Sample Wash Sample washing times of the 0 - 99 times 2 times
syringe before injection
SOLVENT WASH Solvent Wash Solvent washing times of the 0 - 99 times Once
syringe after injection
NUMBER OF Number of Injection Injection times for the same 1 - 99 times Once
INJECTIONS samples
SAMPLE SIZE Sample Size (μl) Sample injection volume 0.1 - 8.0 μl 1.0 μl
If the [START] key is pressed on the autoinjector (For GC-2010/2010Plus/2014/2025, [Start] (PF menu) in
[AOC Parameters]), the injection is performed by the following procedure:
(1) Start
If the GC is READY when the [START] key is pressed, the autoinjector immediately starts injection
sequences. If not, the autoinjector starts the sequences after the GC becomes READY. However, an
injection cannot be started if an error code (-01 to -04: see "6.1 Error Codes" P.151) is displayed. In
such a case, identify and remove the cause of the error, press the [RESET] key to return the AOC to
its home position (the indication returns to "000".), and press the [START] key again.
If the GC is not READY when the [START] key is pressed, the autoinjector stands by until the GC
becomes READY, blinking the RUN light. If the [START] key is pressed again during the stand-by,
the injection process begins, regardless of the GC status.
AOC-20i/AOC-20i Plus/AOC-20s 85
3 Operation
@ NOTE
When pumping and washing the syringe with sample or solvent, the plunger rises to aspirate the
designated amount (8 μl using 10 μl syringe in the default setting), then holds that position for a
certain time. The plunger is then pushed down to expel the contents into the waste vial. When
aspirating high viscosity samples, increase the plunger hold time, since it takes some time for the
sample to enter the syringe after the plunger is raised.
For GC-2010/2010Plus/2014/2025
Set the waiting time during sample wash with the parameter of [Viscosity(s)]. For example, if the set
value is [0.2], the wait time is 0.2 seconds.
(3) Pumping
If bubbles are included in aspirated samples, the reproducibility of the sample aspiration quantity is
deteriorated. The autoinjector performs "pumpings" to pump out the bubbles from the samples.
The default value is 5 times. The number of pumpings can be changed.
For GC-2010/2010Plus/2014/2025: Pumping
For other gas chromatographs: FUNCTION 02
After the pumping is complete, the autoinjector aspirates samples with the specified volume. For
example, if the value is set to [1.0] ("01.0") when using 10 μl syringe, the autoinjector aspirates 1 μl
of the sample.
@ NOTE
The set value of [Viscosity(s)] or VISCOSITY also applies for the wait time during pumping. For
example, if the value is set to [0.2] ("00.2"), the wait time is 0.2 seconds. After aspirating the sample
for the actual injection, however, the autoinjector always waits for 4 seconds, even if the value is set
to 4 seconds or smaller. If the value is set to 4 seconds or longer, the autoinjector waits as long as
the set value.
For GC-2010/2010Plus/2014/2025
Select the plunger speed at injections from [Slow], [Middle], and [Fast] in [Inj. Speed(Plunger)] in the
[AOC Parameters] window. Select the syringe up/down speed from [Slow] and [Fast] in [Inj.
Speed(Syringe)] in the [AOC Other Parameters] window.
86 AOC-20i/AOC-20i Plus/AOC-20s
3.1 Basic Operation
For GC-2010/2010Plus/2014/2025
The wash times can be set in [Solvent Wash] in the [AOC Parameters] window. For example, if the
value is set to [1], one solvent wash is performed.
@ NOTE
3
The wait time for solvent wash is fixed to 0.2 seconds regardless of the [Viscosity(s)] and
[VISCOSITY] settings.
AOC-20i/AOC-20i Plus/AOC-20s 87
3 Operation
Parameter Settings
When an invalid parameter value is entered, the previous set value remains unchanged. (When using GC-
2010/2010Plus/2014/2025, "Input parameter out of range" appears in the GC display, and when the value
is entered with the autoinjector keypad, "- - -" is displayed.)
If a parameter is changed after an injection sequence has begun, the previous value is applied to the
current sequence. For example, if a SAMPLE WASH value is changed from 2 to 3 during a sample wash,
only two sample washes will be performed. The new setting will take effect upon the next injection, when
sample wash will be performed three times.
Vial Detection
The autoinjector uses a photo-sensor to automatically detect vials in the rack. If there are no sample vials,
the injection process is not performed; if there are no solvent vials, the solvent wash procedure is not
performed.
A vial can be placed in any position of the rack to start injection. When vials are placed in several positions,
injections are made in ascending order from the lowest position number.
@ NOTE
If the autoinjector detects no waste vial, an error occurs and the autoinjector stops operating as the contents
of the syringe cannot be expelled. Always place a waste liquid vial in the rack.
@ NOTE
To allow error-free vial detection, use only white vials and caps. Avoid allowing strong light to shine directly
on the vial detection sensor, as this can cause detection errors.
88 AOC-20i/AOC-20i Plus/AOC-20s
3.2 Display and Keypad Functions
! CAUTION
• For GC-2010/2010Plus/2014/2025, operate the following keys only for setting the
parameters not accessible via the gas chromatograph.
The parameters not accessible via GC-2010/2010Plus/2014/2025 are indicated with "∗" in
"Corresponding AOC Parameters/Other Parameters on GC" in the table shown in "3.2.8 FUNCTION
Key" P.92.
AOC-20i/AOC-20i Plus/AOC-20s 89
3 Operation
@ NOTE
Be sure to set these parameters on the gas chromatograph. Otherwise, the autoinjector parameters between
the autoinjector and GC become inconsistent.
Table 3-2
90 AOC-20i/AOC-20i Plus/AOC-20s
3.2 Display and Keypad Functions
@ NOTE
That if the [RESET] key is pressed during a sequence of analyses, all information on the number of
completed analyses and injections is lost.
3
• For GC-2010/2010Plus/2014/2025, [Reset] (PF menu) in the [AOC Parameters] window on GC has the
same function.
AOC-20i/AOC-20i Plus/AOC-20s 91
3 Operation
• The function setting will display, and the FUNCTION LED stays on, until a key other than an arrow key or
[ENTER] is pressed.
• For GC-2010/2010Plus/2014/2025, use the autoinjector keypad only when setting the FUNCTIONs not
accessible via the GC. (i.e., FUNCTIONs indicated with "∗" in the following table, "Corresponding AOC
Parameters/Other Parameters on GC")
92 AOC-20i/AOC-20i Plus/AOC-20s
3.2 Display and Keypad Functions
(1) FUNCTION 01 - 08
The FUNCTION 01 - 08 are the parameters for injection sequence.
Table 3-4
Corresponding AOC
Setting Default
Other Parameter on FUNCTION Name Description
Range Value
GC
Inj. Speed (Syringe) F06 INJ SPEED- Injection speed of the 0: Low 1
SYRINGE syringe 1: High
Inj Mode F07 INJECTION MODE Injection mode 0 - 4 (Note 3) 0
Note 1) FUNCTION 01 is available only in Normal injection mode (when FUNCTION 07 is set to [0]).
Note 2) Wait times vary, depending on the injector operation. The wait time for both sample wash and pumping is
the FUNCTION 03 set time. When aspirating a sample for injection, the plunger holds either the set time,
or 4 sec, whichever is longer. The wait time for solvent washing is fixed at 0.2 sec.
Note 3) 0: sample + (air) --- Normal injection mode
1: Solvent + air + sample + (air) --- standard solvent flush mode (See Figure 3-14.)
2: Solvent + sample + (air)
3: Solvent + air + standard (internal standard) + air + sample + (air)
4: Solvent + standard (internal standard) + sample + (air)
Settings 3 and 4 are available only when an autosampler is connected.
The option of drawing air last is set with FUNCTION 20.
Note 4) 0: 3 vials of solvent are used.
1: Only solvent A is used.
2: Only solvent B is used. (See Figure 3-18.)
3: Only solvent C is used.
FUNCTION 08 is available only when the autosampler is connected and when FUNCTION 28 is set to [1].
AOC-20i/AOC-20i Plus/AOC-20s 93
3 Operation
(2) FUNCTION 10 - 13
These functions designate priority samples.
For details, see "3.3.2 Priority Analysis" P.111 when not using the autosampler, or "3.4.5 Priority
Analysis" P.123 when using the autosampler.
@ NOTE
For GC-2010/2010Plus/2014/2025, the [Priority] PF menu is displayed in the gas chromatograph [AOC
Parameters] window, only when the autoinjector is operating.
Table 3-5
94 AOC-20i/AOC-20i Plus/AOC-20s
3.2 Display and Keypad Functions
(3) FUNCTION 20 - 35
These functions customize the instrument.
Table 3-6
AOC-20i/AOC-20i Plus/AOC-20s 95
3 Operation
Note 1) Specify whether to aspirate 1 μl of air after sample has been drawn. This helps to avoid discrimination for
sample components with a wide boiling point range.
! CAUTION
• When using a high initial column temperature or a column with high polarity (such as
polyethylene glycol (PEG) phase), injecting air may shorten the column service life.
Note 2) The parameter settings for solvent wash before sample injection in SOLVENT FLUSH mode (when
FUNCTION 07 is set to [1] - [4].)
0: Number of times set by the SOLVENT WASH key (the same number before and after injection)
1: Number of times set by the SAMPLE WASH parameter
2: Number of times specified in F01 (PRE SOLVENT WASH). (This is invalid for GC-2010/2010Plus/2014/
2025.)
If the solvent flush mode is set for GC-2010/2010Plus/2014/2025, the GC [AOC Parameters] window does
not display [Pre Solvent Wash], and the solvent wash is performed the same number of times before and
after injections. If the F21 is set to [1], however, [Pre Solvent Wash] is displayed in the solvent flush mode,
which allows specifying different sample wash times before and after injections.
Note 3) The aspiration speed for sample wash and sample aspiration is variable. The aspiration speed for pumping
and solvent wash is fixed at HIGH speed.
Note 4) [5%] can be selected when Injection mode (FUNCTION 07) is set to [3] or [4]. Otherwise, a value of [10%]
is applied regardless of the setting.
(4) FUNCTION 40 - 42
These functions specify whether or not to use optional units.
Table 3-7
(5) FUNCTION 50 - 51
These functions set the dual injector parameters. For details, refer to the Dual Injector User's
Manual (P/N 221-40483).
Table 3-8
Corresponding AOC
Corresponding AOC Default
Other Parameter on FUNCTION Description Setting Range
Parameter on GC Value
GC
Alloc F50 Specifies the 0-8 0
sample allocation in
DUAL mode. (Note 1)
Use the Same Param F51 Specifies how to set 0: Set on SUB 0
the parameters for injector.
the SUB injector in 1: Use MAIN
DUAL mode. parameters.
Note 1) This FUNCTION uses the same sample vial for both the MAIN and SUB injectors in DUAL mode.
96 AOC-20i/AOC-20i Plus/AOC-20s
3.2 Display and Keypad Functions
(6) FUNCTION 60 - 63
These functions specify events such as RAM initialization.
Table 3-9
∗ F63
parameters.
Specifies which injector receives 0: Receive on 0
3
transmitted commands. MAIN
1: Receive on
SUB (via MAIN
injector)
After normal start-up, setting F60 to "001" initializes the values for FUNCTION 01 - 13. To initialize
all the values, turn on the autoinjector while pressing the [SAMPLE WASH] key and set F60 to
"001".
@ NOTE
For GC-2010/2010Plus/2014/2025, when the gas chromatograph and autoinjector are linked after
the F60 initialization, the autoinjector parameters are updated to those set on the GC.
AOC-20i/AOC-20i Plus/AOC-20s 97
3 Operation
(7) FUNCTION 80 - 86
These functions set the transmission parameters.
@ NOTE
Except FUNCTION 84, all changed parameters become effective after restarting the autoinjector.
Table 3-10
Note 1) The AOC-20 can be used in AOC-17 mode. AOC-17 compatible mode is selected by setting FUNCTION
84 to "001" (the default value). Then set FUNCTION 83 to "000" (transmission protocol: "simple
procedure"). When AOC-17 compatible mode is selected, valid parameter ranges for the AOC-20 become
the same as those of the AOC-17. This is useful when upgrading from an AOC-17 autoinjector, because
injection conditions created for the AOC-17 from software that supports only the AOC-17, can still be used.
0: Normal mode ..................................... numerical values consist of three digits.
1: AOC-17 compatible mode ................. numerical values consist of two digits.
Note 2) CH2 is fixed at LEVEL 2, EVEN, STOP 1. Only the baud rate can be changed.
98 AOC-20i/AOC-20i Plus/AOC-20s
3.2 Display and Keypad Functions
(8) FUNCTION 90 - 97
These functions specify features not categorized above.
Table 3-11
3
parameter.)
Rack F93 Specifies the rack type. 0: Short 0
1: Long
Rack Position F94 Specifies the rack position when using the 0-2 1
autosampler. (Note 2)
∗ F95 Specifies the injector in DUAL mode 0: MAIN 0
automatically. (DO NOT change this 1: SUB
parameter.)
∗ F96 (DO NOT change this parameter.) 0
∗ F97 (DO NOT change this parameter.) (Note 3) 0, 1 0
Note 1) Stops the second and subsequent injections, when the GC READY signal is "OPEN" even though the GC
is not READY (e.g. when the READY signal cable is cut).
Note 2) When using the autosampler, specifies which injection port on the GC the autoinjector is installed. (See
Figure 3-4.)
GC-14A/B, GC-15A, GC-16A, GC-2014, and GC-2025 ...........................Position "0"
GC-17Aver.1/2/3, GC-1700, GC-18A, GC-2010, and GC-2010Plus ........Position "1" or "2"
Note 3) Set to [1] only when using the DUAL injector without the autosampler, which disables the autosampler
regardless of the "Use of the Sampler" (F40) setting.
GC-2014 GC-2010
(GC-14A/B, GC-15A, GC-16A, GC-2025) (GC-17A ver.1/2/3, GC-1700, GC-18A, GC-2010Plus)
2
0
1
Figure 3-4
AOC-20i/AOC-20i Plus/AOC-20s 99
3 Operation
@ NOTE
The figure above shows the INJ port positions under the factory default setting.
Change the positions using the teaching function.
(See "2.7.4 Precautions for Using the Autosampler" P.70.)
The GC-2010/2010Plus with single injection port should be installed so that 2 INJ can be used
when "Rack Position" (FUNCTION 94) is set to [1] ("001").
AOC Parameters
Sample Wash SAMPLE WASH (When the injection mode F07 is "0")
Number of Injection NUMBER OF INJECTIONS
Sample Size SAMPLE SIZE
Pre Solvent Wash F01 (When the injection mode F07 is "0")
SAMPLE WASH (When the injection mode F07 is other than "0", and F21 is
"1")
Solvent Wash SOLVENT WASH 3
Pumping F02
Viscosity F03
Dwell Time F04
Inj. Speed (Plunger) F05
Use the Same Param F51
(Only for MAIN injector in Dual injection mode)
AOC POWER None (Use the power switch on the external power supply unit.)
PF Menus
Start START
Stop STOP
Priority F10
OtherPara None (window switching key)
Sampler None (window switching key)
Reset RESET
Upload None (Used for data transfer from AOC to GC.)
Print None (Used for outputting AOC parameters on Chromatopac.)
Next None (window switching key)
PF Menus
Set [ENTER] key after setting F10
PF Menus
Inj Mode F07
Injection Mode
0 Sample + (Air)
1 Solvent + Air + Sample + (Air)
2 Solvent + Sample + (Air)
3 Solvent + Air + Standard (Internal standard) + Air + Sample + (Air)
4 Solvent + Standard (Internal standard) + Sample + (Air)
PF Menus
F21 F21
Autosampler
Use of the Sampler F40
Use of the Sub AOC F41
Bar Code Reader F42
PF Menus
Alloc F50
Key
MONITOR
FUNCTION
F11 INJECTION SAMPLE
F12 FINAL SAMPLE
F13 INJECTION SAMPLE2
F30 Checking vial return position
F31 Analysis time (time before the next injection)
F32 Analysis start time 3
F35 Aspiration volume of the solvent in solvent flush mode
F60 RAM initialization
F61 ROM version display
F62 Setting the SUB injector parameter by the MAIN injector key operation
F63 Transmit commands to the SUB injector via the MAIN injector
F80 CH1 baud rate
F81 CH1 parity during simple procedure transmission
F82 CH1 stop bit (1/2) during simple procedure transmission
F83 CH1 protocol
F84 Simple procedure transmission format
F85 CH2 baud rate
F90 Auto-stop function
F91 READY polarity
F92 Injector: Single/Dual
F95 In Dual mode: MAIN/SUB
@ NOTE
Specifying batch schedules allows the same operations as FUNCTION 11 to 13. To use other functions,
operate the autoinjector keypad.
3.3.1.1 Preparation
Before performing sequential analyses, it is first necessary to configure the instrument, GC, and
peripherals as shown in Figure 3-5 to Figure 3-8. For more information on the connection procedure, refer
to the appropriate operation manuals.
AOC-20i
GC-2010
GCsolution
Figure 3-5 Connection Example (Using GCsolution): Built-in Power Type (Back view)
AOC-20i
GC-2010
Chromatopac
Figure 3-6 Connection Example (NOT Using GCsolution): Built-in Power Type (Back view)
AOC-20i
AOC cable
GC-17A
GCsolution
External power
supply unit
Figure 3-7 Connection Example (Using GCsolution): External Power Type (Back view)
AOC-20i
AOC cable
GC-17A
Figure 3-8 Connection Example (NOT Using GCsolution): External Power Type (Back view)
If the autoinjector is connected as shown in Figure 3-5 to Figure 3-8, two-way communication is made
possible. A START signal is sent to the gas chromatograph when the sample is injected, starting the gas
chromatograph and data processor. When the analysis is complete and the gas chromatograph becomes
READY, the GC sends a READY signal to the autoinjector, which then begins its next injection.
If the READY cable connecting the gas chromatograph and the autoinjector is not connected, the injector
will not perform the subsequent injections.
Additionally, the autoinjector will not perform the subsequent injections if the gas chromatograph was not
programmed correctly. A GC temperature program or time program is required for a READY signal to be
sent.
1 The gas chromatograph goes into the READY status, when the column oven temperature decreases 3
to the initial temperature after the temperature program is complete.
2 The READY signal is output to the autoinjector.
3 The autoinjector starts the next injection operation.
For GC-2010, 2010Plus, 2014, 2025, 17A, 1700 and 18A, specify an equilibration time (EQUIB TIME). After
completing the temperature program, the GC pauses to generate a READY signal for the specified time after
the column oven temperature decreases to the initial temperature. This allows the autoinjector to start
injections when the column temperature is stable at its initial temperature, which improves the repetition
reproducibility. (Default value on GC: 3 min.)
T2
T1
0 READY TIME
Note that the equilibration time cannot be set for GC-14A/B, 15A, and 16A. To improve the reproducibility
as mentioned above, change the temperature program as shown in Figure 3-10 to that in Figure 3-11. In
this case, set the second stage temperature rate to a value larger than "0". This allows the GC to become
READY when the column oven temperature is stable at its initial temperature, at which point the
autoinjector starts the next injection.
TEMP
T2
T1
0 READY TIME
T2
T1
0 READY TIME
• Isothermal Analysis
When using GC-17A, 1700, 18A, 2010, 2010Plus, 2014, and 2025, an isothermal program can be set by
specifying initial temperature and time in the column temperature program, and setting the first stage
temperature rate to "0". ([END] is displayed for GC-2010/2010Plus/2014/2025.) By doing so, the GC does
not generate a READY signal at the set initial time since the temperature program is running, and the
autoinjector does not start the next injection.
Note that for GC-14A/B, 15A, and 16A, the isothermal program cannot be set only by specifying the initial
temperature and time. Be sure to set the first stage temperature rate to a value other than [0], and set the
first stage end temperature to the same value as the initial temperature. Additionally, the time values must be
set to satisfy the next equation: [Initial time + End time] = [Analysis time].
When using GC-14A/B, 15A, and 16A, a time program with a STOP time must be used to specify the
analysis run time. The GC becomes READY after the specified time for the time program elapses, and
generates a READY signal to trigger the next injection.
For example, to set 5 minutes of analysis time for the GC-14A or -14B, perform the following key operations
on the GC keypad:
Table 3-12
1 CITP Display continues to show the previous parameter (in this case, column
SHIFT.D oven temperature). S.DOWN light on.
6
STOP
3
0
When using GC-17A, 1700, 18A, 2010, 2010Plus, 2014, 2025, the analysis run time can also be specified
using the time program. For details, refer to the instruction manual of the gas chromatograph.
@ NOTE
When both a temperature and time program are loaded, the gas chromatograph reverts to the READY state
when the longer of the two programs ends. For example, if STOP time settings for the temperature program
and time program are five minutes and ten minutes, respectively, the READY signal is not generated until
the end of the time program (ten minutes), even though the oven temperature has already returned to the
initial temperature.
@ NOTE
When the gas chromatograph is READY, the autoinjector starts injection. The data processor should also be
READY when injection starts. For this reason, the gas chromatograph STOP time setting should be longer
than the STOP time setting for the data processor. (For Chromatopac, add the time to finish printing results
to the STOP.TIME setting.)
3.3.1.2 Operation
@ NOTE
When using GCsolution, CLASS-GC10, or Chromatopac GC network, operate the computer or chromatopac
to start the autoinjector.
@ NOTE
During GC analysis, the autoinjector RUN light blinks. The autoinjector pauses until the GC sends its
READY signal. If the [STOP] key is pressed, the injection process is interrupted, and the RUN light
goes out. To restart the process, press [START].
3 When all of the samples have been analyzed, the autoinjector returns to the home
position.
@ NOTE
When the NUMBER OF INJECTIONS parameter is changed during a sequence, the new setting will
take effect starting with the next sample.
For example, if the setting is changed from 2 to 5 while sample No. 1 is being analyzed, sample
No. 1 will be analyzed twice. The new setting will take effect on the analysis of the next sample, and
sample No. 2 will be analyzed 5 times.
For GC-2010/2010Plus/2014/2025
Priority Analysis
When the autoinjector is operating, [Priority] (PF menu) is displayed in the gas chromatograph [AOC
Parameters] window.
Press this menu to display the window for specifying priority samples.
Enter the sample No. to preferentially inject, then [Set] (PF menu) to start the priority analysis.
After the priority analysis, the autoinjector resumes analyses from the next sample to the priority samples.
For example, if the analysis of sample No. 2 is interrupted for the priority analysis of sample No. 4, sample
3
No. 5 will be analyzed after completing the analysis of sample No. 4.
For details on the procedure, refer to the section "AOC priority analysis" in the instruction manual of the
gas chromatograph.
@ NOTE
If a parameter for priority analysis is changed before its execution, the new parameter is applied to the
priority analysis. If "0" is set, the priority analysis will be canceled.
@ NOTE
If a priority analysis is set during a repeated analysis of the same sample (i.e., "Number of Injection" is set to
2 or more), the priority analysis will be executed after the current injection.
@ NOTE
DO NOT set a priority sample while the autoinjector is actively functioning.
If you do, a new injection sequence for the specified priority sample is started during the current sequence,
which may cause a problem in that the syringe is not washed by the priority sample. To avoid this, be sure to
set a priority analysis after the autoinjector completes the current injection sequence (i.e., during standby for
the next injection).
1 FUNCTION Example) Displays the last setting. (In this example, FUNCTION 03.)
F03
2 Arrow key F10 Press the arrow keys until "F10" displays.
3 ENTER 000 The current FUNCTION 10 set value is displayed. (A value of 0 indicates
no PRIORITY SAMPLE.)
4 Arrow key 002 Pressing the right arrow key twice displays "002".
• In this example, the priority sample is set up before the [START] key is pressed. Sample No. 2 will
be the first sample injected, even if a sample vial is in position 1. Once executed, the PRIORITY
SAMPLE setting returns to "000".
• If a priority analysis is executed during a repeated analysis, the autoinjector resumes analyses
from the next sample to the priority samples.
@ NOTE
Once a PRIORITY SAMPLE setting has been made, it can be overridden before it is executed by
making another setting. For example, once sample No. 3 has been set as a PRIORITY SAMPLE, the
setting can be canceled by then specifying 0 as the PRIORITY SAMPLE.
@ NOTE
When a PRIORITY SAMPLE is scheduled during repeated analysis of the same sample (NUMBER
OF INJECTIONS set to 2 or more), the PRIORITY SAMPLE will be executed at the next injection.
@ NOTE
Do not set a priority sample when the autoinjector is actively functioning. Otherwise, all the rinses
may not be completed before the syringe moves to the priority sample. Make the priority setting
during an analysis or when the injector is not in use.
@ NOTE
When using GC-2010/2010Plus/2014/2025, batch schedules can be substituted for the FUNCTION 10-12
(features available when the autoinjector is on hold). For details, refer to the section "Batch Schedule" in the
instruction manual of the gas chromatograph.
Figure 3-12
In the solvent flush method, as shown in Figure 3-13, the autoinjector draws solvent, air, and sample, in
that order, into the syringe. After injection, solvent, not sample, is left in the syringe needle. Since the
plunger completely expels the sample, all of the sample vaporizes at the same time, reducing
discrimination.
The solvent flush method is recommended when analyzing compounds with a wide boiling point range.
Solvent
Sample
Air
Figure 3-13
The following Figures show the sample aspiration patterns in 3 injection modes with FUNCTION 07.
Sample
@ NOTE
It is recommended to select 10% (i.e., 1 μl of solvent) since the needle volume of the standard 10 μl syringe
is 0.6 μl.
The wash solvent is used for solvent flush, normally. However, when FUNCTION 28 is set to "001" (three
solvent positions are used), or for GC-2010/2010Plus/2014/2025, when [Using 3 Solvent Vials] is set to [Yes]
in the GC [AOC Other Parameters] window, the solvent at position C (the least contaminated) will be used
for solvent flush. (See Figure 3-17.) In this mode, the maximum number of samples that can be analyzed is
1 for the short rack, and 7 for the long rack. (Position 1 in Figure 3-17 is only for 4 ml vials. When 4 ml vials
are used as sample vials, the maximum number of sample vials is 2 for the short rack and 8 for the long
rack). However, 4 ml vial rack for autosampler (optional, P/N S221-45182) is separately required.
Figure 3-17
When using the solvent flush injection mode (FUNCTION 07 is set to "001" - "004", or [Inj Mode] is set to
[1] - [4] for GC-2010/2010Plus/2014/2025), the parameter upper limit and set items change as follows:
To inject volumes larger than the syringe capacity, use FUNCTION 33 (For GC-2010/2010Plus/2014/2025,
[Multi-Inj.] in [AOC Other Parameters] on GC) to increase the number of syringe injections after which the
GC receives the START signal from the autoinjector. For example, when using the large volume 250 μl
syringe, although 200 μl (80% of full scale) is the maximum volume for a single injection, setting
FUNCTION 33 to "005" (For GC-GC-2010/2010Plus/2014/2025, [Multi-Inj.] to [5]) starts the GC analysis
after five injections, so that a total of 1000 μl is injected (200 μl × 5 = 1000 μl).
@ NOTE
Possible injection volumes for CLASS-GC10 range from 0.1 μl to 8 μl. When using a large volume syringe,
injection volumes need to be converted as follows:
10 μl syringe: 0.1 μl - 8 μl (1 - 80% of 10 μl)
↓
50 μl syringe: 0.5 μl - 40 μl (1 - 80% of 50 μl)
250 μl syringe: 2.5 μl - 200 μl (1 - 80% of 250 μl)
@ NOTE
When using the standard syringe (10 μl), be sure to set FUNCTION 34 to "000" ([Kinds of the Syringe] to
[10 μl] for GC-2010/2010Plus/2014/2025). Otherwise, poor reproducibility may result.
3.4.1 Preparation
(1) Setting sample gripping position
For GC-2010/2010Plus/2014/2025
Using [Rack Position] in the [AOC Other Parameters] window, specify the position of the GC
injection port where the autoinjector is installed. 3
GC-2010/2010Plus Rear (=Standard position): 2 (The autosampler can be adjusted for use under
the setting of "1".)
GC-2010/2010Plus Front: 1
GC-2014/2025: 0 (Specify [1] when using the optional mounting bracket (for AOC-20) to use the
injection port located at far left. See "2.1.2 Mounting Spacers on GC-2014/GC-14A/B" P.23.)
Then set [Use of the Sampler] to [Use] in the [Sampler] window.
When using only the autoinjector (without the autosampler), set [Use of the Sampler] to [Not Use].
For GC-14A/B
When using the autosampler with a GC-14A/B, set FUNCTION 94 to "000".
For both the GC-14 and -17 series, set FUNCTION 40 to "001". This places the autoinjector in
autosampler mode. (This setting is saved when the power is turned OFF.)
The autoinjector can be used without the autosampler; set FUNCTION 40 to "000".
@ NOTE
If the position of the autosampler arm in relation to the vial is not correct at gripping position, teach the
autosampler the correct position (teaching function). (See "2.7.5 Adjusting Sample Gripping Position
(Teaching)" P.70.)
Waste liquid
vial B
Solvent vial
Figure 3-18
It is not necessary to place empty racks in the sample tray. Place only the racks with samples. The
racks do not have to be in sequential order. For example, when 1.5 ml vials are placed as shown in
Figure 3-19, the vial rack at position 1 holds vial numbers 1 to 25. At position 4, the rack holds vial
numbers 26 to 50, and at position 5, numbers 51 to 75. If a rack is loaded for different size vials than
the size set by FUNCTION 29 ([Kinds of Vials] for GC-2010/2010Plus/2014/2025), the rack is
ignored.
Figure 3-19
Set vials in the sample rack in sequential order from position 1. The first vial used for analysis is the
vial at position 1. See Figure 3-20 for a diagram of the position of the sample vials.
Figure 3-20
@ NOTE
1 After pressing the autoinjector [START] key, the autosampler detects the position of the sample
rack on the tray. In this case, vial numbers are assigned to only vials in that sample rack.
To avoid malfunction, DO NOT remove the rack after sample vial numbers are assigned.
2 If you want to add racks after analysis has been started, add them behind the rack containing
the vial currently being analyzed. Numbers will be reassigned to the vials. Racks added in front
of the current rack will be ignored.
As shown in the following examples, if a position in the rack is left empty, analysis on that rack ends,
and the autosampler starts analysis on the next vial rack. If no vial is placed in the starting position
of that rack, operation stops.
For example, when 1.5 ml vials are positioned as shown in Figure 3-21 (the black circles indicate
vials), samples 1 through 18 are analyzed. Next, the autosampler searches for, but does not detect,
a vial in position 19. It then searches for a vial in position 26 (the first vial of the next rack). If the
autosampler does not detect a vial there, it stops operation.
Figure 3-21
When the vials are positioned as shown in Figure 3-22, after analyzing samples in positions 1 to 18,
the autosampler does not detect a vial in position 19. It checks in position 26, finds a vial there, and
analyzes the samples in positions 26 to 31. The autosampler finds no vial in position 32, detects no
vial in position 51, then stops.
Figure 3-22
When the vials are placed as shown in Figure 3-23, after analyzing samples in positions 1 to 18, the
autosampler does not detect a vial in position 19. It skips the vials in positions 20 to 25, and
searches for a vial at position 26. Detecting no vial in this position, the autosampler stops operation,
skipping the samples in positions 51 to 53. Only the samples in positions 1 to 18 are analyzed.
Figure 3-23
As shown in the examples above, if a position in the rack is left empty, analysis on that rack ends,
and the autosampler starts analysis on the next vial rack. If no vial is placed in the starting position
of that rack, operation stops.
3.4.2 Power ON
The autosampler shares the power supply with the autoinjector.
Turn ON the switch on the power supply unit.
For GC-2010/2010Plus/2014/2025, you can toggle the power ON/OFF in [AOC POWER] in the gas
chromatograph [AOC Parameters] window. When the power is turned ON, the power light starts blinking
green, the initial operation is performed (illuminates orange), the autosampler arm moves to its home
position, and the autosampler goes into stand-by.
Arm
Gripper
Vial rack
Tray
Power light
Figure 3-24
@ NOTE
When the light is orange, the autosampler is operating, so do not touch the arm. Applying excessive force
may break the motor.
@ NOTE
Perform all key operations when the autosampler is standing by after initial operations are completed.
(Except if an error occurs.)
• Sample Vial Size FUNCTION 29/ AOC Other Parameters Kinds of Vials
Specify the sample vial size from 1.5 ml and 4 ml. This setting cannot be changed once analysis has begun.
The change made after START will be ignored. To change the size, press the [RESET] key and set the
parameter.
You can START and RESET analyses via the gas chromatograph display.
3
• Checking Vial Return Position FUNCTION 30
It is specified whether to check the rack position (to ensure that it is empty) before replacing a vial. "000"
indicates do not check, and "001" indicates check. Set this parameter with the autosampler keypad, even
when using GC-2010/2010Plus/2014/2025.
(1) While the gas chromatograph is analyzing (i.e., the autoinjector is standing by), press [Priority]
(PF2) in the AOC parameter to display the priority analysis window.
(2) Specify the desired priority samples and press [Set] (PF2).
(3) After the analysis currently executed is complete, the autoinjector starts injecting the specified
priority samples.
(4) After the priority analysis is complete, the autoinjector starts injections from the sample next to the
one completed before interruption.
(5) The set values for priority analyses are reset to [0] after the completion of the analyses.
W S W S W S
S = Solvent
W = Waste liquid
Solvent for solvent flush (4 ml vial)
For sample
Figure 3-25
Figure 3-26 to Figure 3-29 shows each injection mode available through FUNCTION 07 ([Inj Mode] for GC-
2010/2010Plus/2014/2025) when an autosampler is used.
Figure 3-26 When FUNCTION 07 is "001" ([Inj Mode] is [1] for GC-2010/2010Plus/2014/2025)
Figure 3-27 When FUNCTION 07 is "002" ([Inj Mode] is [2] for GC-2010/2010Plus/2014/2025)
Figure 3-28 When FUNCTION 07 is "003" ([Inj Mode] is [3] for GC-2010/2010Plus/2014/2025)
Figure 3-29 When FUNCTION 07 is "004" ([Inj Mode] is [4] for GC-2010/2010Plus/2014/2025)
Set FUNCTION 07 ([Inj Mode] for GC-2010/2010Plus/2014/2025) to "003" or "004" when performing
internal standard analysis and solvent flush injection mode. This automatically aspirates internal standard
and injects it along with the sample. The internal standard is placed in position 1 of the rack. The solvent for
solvent flush is then taken from the solvent in position C of the rack (Figure 3-30).
W S W S W S
Sample
Standard sample (4 ml)
S = Solvent
W = Waste liquid 3
Figure 3-30
When injecting internal standard along with the sample, not only 1 μl but 0.5 μl can be selected for solvent
aspiration volume for solvent flush. To do so, set FUNCTION 35 using the autoinjector keypad.
0: 10% of syringe volume
1: 5% of syringe volume
(2) Set the program time of the gas chromatograph longer than the data processor analysis time. (For
Chromatopac, add the printout time to STOP.TIME.)
The data processor has to be READY before the gas chromatograph in order to correctly follow the
sequence of [Sample injection → Gas chromatograph starts → Data processor starts]. Allow time for
any reports to print before starting the next injection, or the data of the next injection may not be
acquired.
(5) Inspect and maintain the syringe (See "7.1 Precautions on Use of the Microsyringe" P.155 and "7.2
Maintenance of the Microsyringe" P.156.)
When the plunger becomes stiff, the LED to the left of the [SAMPLE SIZE] key blinks as a warning
during analysis (when not in solvent flush mode). When the plunger becomes too stiff to move
normally, error "-03" displays and the autoinjector stops operation.
1
FUNCTION NUMBER
__
3
FUNCTION NUMBER
CUSTOMIZE 18 __
FUNCTION ESCAPE
3.6 GCsolution
3.6.1 System Configuration
Select a type of the autosampler to be used under [Available Modules] in the [System Configuration]
window. Add it to the analytical line as a [Modules Used for Analysis].
Figure 3-31
For details on the dual injection system, refer to the instruction manual of AOC-20 Dual Injection System
(P/N 221-40483).
Complete the required setting in the properties of the AOC unit icon under Analytical line.
Figure 3-32
Figure 3-33
Figure 3-34
@ NOTE
For details on the parameter settings, refer to the instruction manual of GCsolution.
For the setting procedures for GCMSsolution, refer to the instruction manual of GCMSsolution.
3.7 CLASS-GC10
3.7.1 System Configuration
Select the following items from the system configuration autosampler drop-down list.
AOC-20i with short rack: AOC-20i (short)
AOC-20i with long rack: AOC-20i (long)
AOC-20i + AOC-20S: AOC-20s
Dual injection system: AOC-20d
∗ For a dual injection system, refer to the instruction manual of AOC-20 Dual Injection System (P/N 221-40483).
Specify the injection volume and the number of injections per sample, when registering samples or creating
the sample schedule.
@ NOTE
For setting up large volume sample injections, see "3.3.3 Injection Modes", "Large Volume Injection Mode
(Low-speed Injection Method)" P.115. The range of injections per sample is 0 - 10.
@ NOTE
The default values of [Injection Speed] and [Inj. Port Dwell Time] are [Medium] and [0.3 sec], respectively.
The recommended AOC-20 values are [High speed] and [0 sec].
3.8 CLASS-VP
3.8.1 Autoinjector/Autosampler Configuration
To configure CLASS-VP for the AOC-20 autoinjector and/or autosampler, choose Configuration from the
CLASS-VP Main Menu. Then select the Injection button. The autoinjector primary configuration screen is
displayed; choose Shimadzu AOC from the drop-down list. Click on Settings.
Figure 3-37
Choose the appropriate Comm. Port, Model, Vial Size, and Syringe Type from the drop-down list
selections. To test the connection, click on the Test button. Restart the program for the new configuration to
take effect. Please see the CLASS-VP manual for more information.
Figure 3-38
The parameter functions correspond to the autoinjector parameters described in "3.2 Display and Keypad
Functions" P.89. Set them up as you would on the autoinjector. Then, save the pretreatment file by
choosing [Save] from the [File menu]. The file is saved with an.aoc extension, which is then specified when
setting up the batch.
@ NOTE
When the autoinjector is controlled by CLASS-VP, FUNCTION 31 must be set to 655 min.
4
This transmission method is used to connect the autoinjector to a C-R7A Chromatopac. This method is also
used for communication between the gas chromatograph and the C-R7A Chromatopac.
Either a fiber optic or RS-232C interface can be used for the external control connection. However, both
cannot be used at the same time. 4
The pin configuration for the RS-232C connector is shown below. It is the same configuration as that of
IBM PC/AT compatible computers. Use a commercially available null modem cable for an RS-232C
connection to external devices. 4
Table 4-1 RS-232C pin configuration
4
Pin No. Function
1
2
NC (Not used)
RxD (Received Data)
4
4
3 TxD (Transmitted Data)
4 DTR (Always ON)
5 Ground
6 NC (Not used) 4
7 RTS (Always ON)
8
9
NC (Not used)
NC (Not used)
4
4
4
4
4
4
AOC-20i/AOC-20i Plus/AOC-20s 133
4 External Control
4.1 Preparation
Make the following settings to prepare for the autoinjector transmission. Restart the unit for the settings to
take effect.
Level 2 Transmission
F80 ....................Baud rate. Leave the default value of 2.
F81 ....................Setting not necessary
F82 ....................Setting not necessary
F83 ....................Leave the default value of 1.
F84 ....................Setting not necessary
F86 ....................Leave it as 0 (default value).
MODE setting of the Chromatopac (communication port parameter) .......... Change to 12917
2 If the setting was successfully made, the autoinjector returns "AK CR ". If the setting
was not successfully made, then "NK CR " is returned.
@ NOTE
For the list of transmission commands, see "4.2.3 Command List" P.140.
When setting autoinjector parameters, use the format described in 1. Always designate parameters with 3
digits (for example, to set 3, use 003). However, use only 2 digits when FUNCTION 84 is set to "001".
If the setting was successfully completed, the autoinjector returns "AK CR ". If the setting was not
performed (mistaken transmission code, invalid parameter, etc.), "NK CR " is returned.
For an example, a BASIC program for setting the PUMPING frequency to 6 from a Chromatopac C-R7A or
C-R4A is shown below. (Port No. 3 is used.)
OPEN TRS 3 Port open
C$ = "KPUMP 006" + CHR$(0DH) Command for PUMPING 6 times
PRINT #3,C$; Transmission
INPUT #3,A$ Input response from the autoinjector
CLOSE TRS 3 Port close
@ NOTE
Be sure to add a semi-colon (;) at the end of the sentence when sending a command.
@ NOTE
Use "M" at the beginning of a command, instead of an "S", to monitor operations (receive actual
values). See the command list in "4.2.3 Command List" P.140.
@ NOTE
The reset command is ignored when F84 is set to 1.
(3) If a transmission error occurs, and the condition cannot be returned to normal, send CHR$(05H)
(ENQ command) to the autoinjector. CHR$(05H) resets the autoinjector transmission line, and "AK
CR " is returned.
(4) When " " (space) in the transmission command is replaced with "$", the autoinjector returns the
command along with the value. This makes it easy to distinguish to which command the response
corresponds.
For example:
Command to be transmitted Autoinjector
@ NOTE
Follow the procedures below to prevent transmission errors.
(5) When a command is transmitted to the autoinjector, make sure the response is returned from the
autoinjector before sending the next command.
(6) As in the program example shown in "4.2.1.1 Parameter Transmission (Setting)" P.135, be sure to
add a semi-colon (;) at the end of the sentence when sending a command. When a program is
executed without including the semi-colon, no carriage returns or line feeds occur.
OPEN TRS 3
Physical port number
For the physical port number, refer to the GC network instruction manual.
REPT#3=3
WTPP#3=30
Logical port number
For the logical port number, refer to the GC network instruction manual.
In Simple procedure transmission, the number must be transmitted in 3 digits, so that [1] becomes [001]
(or, when F84 is set to 1, in two digits). However, in Level 2 transmission, the number of digits is not taken
into consideration.
PRINT WMOD#3
LPRINT ISNO#3M
Logical port number
Receiving the set value, as shown in Example 1, is a G mode operation.
Monitoring the status, as shown Example 2, is an M mode operation. Therefore, M is added to the
command.
STRT#3
STOP#3
RSET#3
Logical port number
CLOSE TRS 3
Physical port number
Command Default
Mode Contents Valid Range
Name Value
STRT K START
STOP K STOP
RSET K RESET
WRPT K, G, S SAMPLE WASH 000 - 099 002
WMOD K, G, S SOLVENT WASH 000 - 099 001
REPT K, G, S NUMBER OF INJECTIONS 001 - 099 001
IVOL K, G, S SAMPLE SIZE 001 - 080 010
WPRS K,G.S F01 PRE SOLVENT WASH 000 - 099 000
PUMP K, G, S F02 PUMPING 000 - 099 005
WTPP K, G, S F03 VISCOSITY 000 - 999 002
WAIT K, G, S F04 DWELL TIME 000 - 999 000
ISPD K, G, S F05 INJ SPEED (PLUNGER) 000 - 002 002
SSPD K, G, S F06 INJ SPEED (SYRINGE) 000 - 001 001
SAND K, G, S F07 INJECTION MODE 000 - 004 000
SOLV K, G, S F08 SOLVENT SELECT 000 - 003 000
SINT K, G, S F10 PRIORITY SAMPLE 000 - 150 000
SSNO K, G, S F11 INJECTION SAMPLE 000 - 150 000
FSAM K, G, S F12 FINAL SAMPLE 000 - 150 000
SNO2 K, G, S F13 INJECTION SAMPLE2 000 - 150 000
AAIR K, G, S F20 WITH/WITHOUT AIR ASPIRATION 000 - 001 000
WKEY K, G, S F21 KEY FOR SOLVENT WASH BEFORE 000 - 001 000
INJECTION
USPD K, G, S F22 PLUNGER ASPIRATION SPEED 000 - 002 002
DSPD K, G, S F23 SPEED TO LOWER PLUNGER 000 - 002 002
UVOL K, G, S F24 SUCTION VOLUME AT SAMPLE WASH AND 000 - 001 000
PUMPING
HIGH K, G, S F25 SYRINGE HEIGHT WHEN ASPIRATING (UP) 000 - 020 000
LOWS K, G, S F26 SYRING HEIGHT WHEN ASPIRATING (DOWN) 000 - 010 000
INJH K, G, S F27 SYRING HEIGHT ATINJECTION (UP) 000 - 022 000
SLMD K, G, S F28 3 SOLVENT VIALS IN USE 000 - 001 000
VIAL K, G, S F29 VIAL SIZE 000 - 001 000
CKTR K, G, S F30 WITH/WITHOUT TRAY CHECK 000 - 001 000
Command Default
Mode Contents Valid Range
Name Value
TANL K, G, S F31 ANALYSIS TIME 000 - 655 000
TSTR K, G, S F32 ANALYSIS START TIME 000 - 999 000
STRI K, G, S F33 NUMBER OF INJECTIONS PER ONE 001 - 099 001
ANALYSIS
LSYR K, G, S F34 SYRINGE SIZE 000 - 002 000
SVOL K, G, S F35 SOLVENT ASPIRATION VOLUME IN SOLVENT 000 - 001 000
FLUSH MODE 3, 4
SAMU K, G, S, M F40 USE/NOT USE AUTOSAMPLER 000 - 001 000
SUBU K, G, S, M F41 USE/NOT USE SUB AOC 000 - 001 000
BARC K, G, S, M F42 USE/NOT USE BAR-CODE READER 000 - 001 000
SPMD K, G, S F50 SAMPLE DISTRIBUTION OF DUAL AOC 000 - 008 000
PAR1 K, G, S F51 OPERATING SUB-INJECTOR BY
PARAMETERS OF MAIN-INJECTOR
000 - 001 000
4
SET2 K, G, S F63 SEND PARAMETER COMMAND TO SUB AOC 000 - 001 000
GLPM K, G, S VALIDATION MODE SETTING 000 - 001 000
CH1B K, G, S F80 CH1 BAUD RATE 000 - 004 002
CH1P K, G, S F81 CH1 PARITY FOR SIMPLE PROCEDURE 000 - 001 000
TRANSMISSION
CH1S K, G, S F82 CH1 STOP BIT FOR SIMPLE PROCEDURE 000 - 001 001
TRANSMISSION
CH1L K, G, S F83 CH1 SIMPLE PROCEDURE/LEVEL 2 000 - 001 001
TRANSMISSION
CH1M K, G, S F84 TRANSMISSION BY SIMPLE PROCEDURE 000 - 001 001
NORMAL/AOC-17 COMPATIBLE
CH2B K, G, S F85 CH2 BAUD RATE 000 - 004 000
ATSP K, G, S F90 AUTOMATIC STOP FUNCTION OFF/ON 000 - 001 001
ARSG K, G, S F91 READY SIGNAL POLARITY OPEN/CLOSE 000 - 001 000
SORD K, G, S F92 AOC SINGLE/DUAL 000 - 001 000
TLET K, G, S F93 RACK SHORT/LONG 000 - 001 000
TSEL K, G, S F94 RACK POSITION WHEN AUTOSAMPLER IS 000 - 002 001
USED
DUAL K, G, S F95 MAIN-AOC/SUB-AOC 000 - 001 000
TLT2 K, G, S F96 RACK FOR AOC-17/FOR AOC-20i 000 - 001 001
RDYF M MONITOR WHETHER ANALYSIS IS BEING 000 - 001
PERFORMED
NINJ M MONITOR INJECTION NUMBER 000 - 099
NSPS M MONITOR SAMPLE WASH NUMBER 000 - 099
NSLW M MONITOR SOLVENT WASH NUMBER 000 - 099
NPMP M MONITOR PUM FREQUENCY 000 - 099
NTIM M MONITOR ANALYSIS (START) TIME 000 - 999
ISNO M MONITOR SAMPLE No. 000 - 150
ISN2 M MONITOR SAMPLE No. OF SUB-AOC 000 - 150
STRF M MONITOR START STATUS 000 - 001
100 PO=3
110 A$="MISNO" +CHR$(0DH)
120 B$="MNINJ" +CHR$(0DH)
130 S$="KSTRT" +CHR$(0DH)
140 OPEN TRS PO : CT=0
150 PRINT #PO,S$; : GOSUB 340
160 IF ((AS$="NK" OR AS$="") AND CT<4) THEN CT=CT+1 : GOTO 150
170 IF CT>=4 THEN LPRINT "ERROR" : CLOSE TRS PO : GOTO 420
180 CLOSE TRS PO
190 WAIT START
200 OPEN TRS PO : CT=0
210 PRINT #PO,A$; : GOSUB 340
220 IF ((AS$="NK" OR AS$="") AND CT<4) THEN CT=CT+1 : GOTO 210
230 IF CT>=4 THEN LPRINT "ERROR" : CLOSE TRS PO : GOTO 420
240 SNS=AS$ : CT=0
250 PRINT #PO,B$; :GOSUB 340
260 IF ((AS$="NK" OR AS$"") AND CT<4) THEN CT=CT+1 : GOTO 250
270 IF CT>=4 THEN LPRINT "ERROR" : CLOSE TRS PO : GOTO 420
280 NI$=AS$
290 CLOSE TRS PO
300 IF SN$="1*" THEN SN$="100"
305 LPRINT : LPRINT
310 LPRINT "SAMPLE No. : ";SN$;" REPEAT : ";NI$"
315 LPRINT
320 WAIT STOP
330 GOTO 190
340 REM *INPUT SUBROUTINE
350 T=TIME : WT=30 : AS$="" : AB$=""
360 IF T+WT<TIME THEN GOTO 410
370 IF (TRSF%(PO) AND 01H)<>0H THEN AB$=INPUT$(1.PO)
380 IF AB$=CHR$(0DH) THEN GOTO 410
390 IF AB$=<>"" THEN AS$ [LEN(AS$)+1]=AB$
400 AB$="" : GOTO 360
410 RETURN
420 END
Program explanation
Operation procedure
Program explanation
@ NOTE
Clean the cover of the unit with a soft cloth dampened with water or a solution of water and a mild detergent. 5
5.1 Cleaning the Vials 5
Sample and solvent vials can be re-used if they are cleaned after each use. Follow the procedure below to
ensure that vials are as clean as possible. 5
1 Remove and discard the caps and septa.
5
5
2 Place the glass part of the vials in a flask made of material that can tolerate a vacuum,
and add the cleaning solution (descaler, organic solvent, water etc.).
5
3 Reduce the pressure in the flask, with an aspirator etc., to extract the air from the vials.
5
4 Gradually allow the pressure to return to normal. The vials will fill with cleaning
solution.
5
5 Place the flask in an ultrasonic cleaning device.
5
6 Discard the cleaning solution, and wash with water.
If acid solution was used, rinse with water at least three times using the procedure described above. 5
7 If water remains, dry the vials with acetone, alcohol or another highly volatile solvent.
5
8 Use new septa on clean vials.
5
5
5
5
5
AOC-20i/AOC-20i Plus/AOC-20s 145
5 Maintenance
! CAUTION
• Since a syringe is manufactured precisely, always use only the plunger supplied with the
syringe. Plungers are not interchangeable.
@ NOTE
The table above only lists the main syringes with 23 gauge needles.
@ NOTE
The design life of this instrument is seven years. After being installed for seven years, the instrument's
maintenance parts will need to be inspected.
! WARNING
For continued fire protection, replace only with fuses of the same type and rating.
-02
Install the rack and press the [RESET] key.
Syringe not in home position or not operating properly
6
• Syringe not completely in home position.
↓
Press the [RESET] key. 6
-03 Plunger not in home position or not operating properly
6
• Plunger not completely in home position.
↓
Press the [RESET] key.
• Plunger does not move smoothly. (See Note 1.)
↓
Turn the power off, remove the syringe and perform maintenance on or replace the plunger. 6
-04 Autoinjector is not READY to start.
• The [START] key was pressed after the autoinjector experienced an error.
↓
Press the [RESET] key to clear the error and then, press the [START] key.
6
6
-05 Autoinjector memory error (RAM) and values initialized. (See Note 2.)
↓
Re-enter the parameters.
6
-06 Autoinjector memory error (ROM) (See Note 3.)
-08 CH2 error (transmission for sub-injector or the autosampler.)
↓
6
Press the [RESET] key.
(For GC-2010/2010Plus/2014/2025, check if the analytical line setting is correctly configured.)
-09 No PRIORITY sample (set by transmission - see Note 4) or no sample in the position specified
by the INJECTION SAMPLE.
↓
Press the [RESET] key to restart after the sample has been placed in the rack.
6
-10 Memory (RAM) does not operate correctly. (See Note 3.)
-11 Autoinjector not correctly installed
↓
6
Re-install the autoinjector.
6
AOC-20i/AOC-20i Plus/AOC-20s 151
6 Troubleshooting
Note 1) If the plunger movement becomes slightly stiff, no error is generated, but a warning is displayed (the LED to the left
of the [SAMPLE SIZE] key blinks while waiting for analysis) and operation continues.
Note 2) The error code "-05" is displayed when a RAM error occurs, and also when parameters are initialized. If this error
occurs every time power is supplied, the RAM back-up battery may be dead. For battery replacement, contact your
Shimadzu representative.
Note 3) When errors "-06" or "-10" occur, the autoinjector does not operate. When error "-12" occurs, the autoinjector
cannot be controlled externally. When error "-13" occurs, the sub-injector or autosampler cannot be used. If these
errors occur, contact your Shimadzu representative.
Note 4) When a PRIORITY SAMPLE is set up by the keypad, no error is generated when there is no sample in the
position; the next sample number is injected. An error is generated only when a PRIORITY SAMPLE has been set
up by transmission from an external device, and there is no vial in that position. When an INJECTION SAMPLE is
set up, whether by keypad or transmission, an error is generated if no vial is found in the position specified.
Note 1) This error is generated only when the priority sample setting was entered by transmission (through an external
device) before start.
When it has been set up after start, an error is generated whether the priority sample was set up by the keypad or
by transmission.
Analysis Reproducibility
Please observe the following precautions when using the AOC-20i to ensure reliable analysis results.
@ NOTE 7
The microsyringe (P/N S221-34618) used in the autoinjector is specifically designed for the AOC. Be sure to
7
use the specified microsyringe.
7
ensure that an excessive force will not be applied at the end of the needle, which would reduce the life of the
needle.
• Do not operate the plunger of the microsyringe when solvent or sample is not supplied sufficiently. This may
cause the barrel glass and the plunger of the autoinjector to seize up against each other and be damaged.
• Before starting to use the autoinjector, check that fragments of the septum, etc. are not clogged at the end of
the syringe needle. (In the first operation of a sequence using the autoinjector, check that the microsyringe is
7
smoothly filled with solvent or sample.)
7
! CAUTION
• The microsyringe is manufactured with extremely high precision. Always use the syringe
main unit and the plunger in the original combination. It is not permitted to replace only the
7
plunger and use the original syringe and a new plunger together.
For attachment and removal of the microsyringe, see "2.2 Mounting and Removing the Syringe" P.26.
7
7
After attaching the microsyringe and fixing the plunger, check that the plunger is touching the lower position
of the syringe barrel.
7
7
7
7
AOC-20i/AOC-20i Plus/AOC-20s 155
7 To Achieve Good Analysis Reproducibility
Figure 7-1
@ NOTE
The silica wool offered as standard accessory with the GC-2010/2010Plus/2014/2025 (SPL) is different from
conventional silica wool, and is inert. This silica wool has thick fibers and low density. Accordingly, even if the
weight is the same, its volume is greater than that of conventional silica wool.
@ NOTE
For standard GC-2014/2025 split analysis, silica wool is packed at 20 mm from the insert end. In this case,
however, the needle tip is inserted approximately 5 mm into the silica wool. This is for obtaining stable
reproducibility regardless of how samples are discharged from the needle tip. If you wish to further improve
reproducibility, place the silica wool at 25 mm from the end, as in the case of GC-2010/2010Plus.
GC-2010/2010Plus/2014/2025
Splitless and GC-14B: 25 mm
GC-17A and GC-1700: 21 mm
GC-2014/2025 Split: 20 mm
2 mg∗1 5 - 6 mm
10 - 15 mm 10 mg
Figure 7-2
@ NOTE
For splitless analysis in which the flow velocity inside the insert is slow, and WBI direct injection in
which the inlet pressure is low, it may be better to pack the silica wool at a lower position. This is due
to INJ port pressure fluctuations during vaporization and the distance from the silica wool to the
insert top end. The optimal amount of silica wool and its position also vary depending on the type of
solvent, the temperature in the injection port, and the volume of the sample.
(2) In order to improve the reproducibility, it is advisable to replace the silicon septum used in solvent
vials and waste liquid vials with PTFE sheeting. By doing so, the chances of fragments of the
septum becoming clogged in the syringe are reduced.
For the same reason, it is also advisable to reduce the number of times for sample washing to 0 or 1
when analyzing multiple times using the same sample.
When using the GC-2010/2010Plus/2014/2025 (SPL), be sure to also observe the following
precautions.
(4) When attaching the column, be sure to use the small nut (column nut for injection port) on the INJ
side.
A B
AC power inlet .....................................................18 Bar code reader ................................................. 96
Accessories ...........................................................6 Barrel .................................................................. 27
autoinjector .....................................................6 Barrel holder ............................................... 27, 147
autosampler ..................................................11
Air suction ...........................................................95
Alloc ....................................................................96 C
Analysis (precautions) .......................................126
Cap (vial) .............................................. 47, 48, 147
Analysis start time ...............................................95
Case ................................................................... 15
Analysis time .......................................................95
CBM-101/102 ................................................. 4, 75
AOC cable ...........................................................21
CLASS-5000 ...................................................... 76
AOC other parameters ......................................102
CLASS-GC10 ..................................... 75, 110, 130
AOC parameters ...............................................101
CLASS-VP ....................................................... 132
AOC-20i installation kit ..............................149, 150
Cleaning ........................................................... 145
AOC-20i installation parts .........................149, 150
after injection ................................................ 87
Arm (autosampler) ..................................16, 64, 72 before injection ............................................. 86
Arrow keys ....................................................82, 89 Components ......................................................... 2
Aspirating air (syringe) ................................95, 116 Configuration
Aspiration amount system ........................................................... 3
pumping ........................................................95 Configure the instrument and peripherals ........ 105
sample wash .................................................95
Consumable parts ............................................ 147
Aspiration volume
Cooling/heating .......................................... 14, 149
solvent flush mode ........................................95
Autoinjector
accessories .....................................................6 D
error codes ..................................................151
installation parts ..............................................8
DEWELL TIME ................................................... 93
installing ........................................................21
Discrimination ..................................... 96, 113, 116
names ...........................................................15
part No. ...........................................................2 Display ................................................... 15, 82, 89
specifications ................................................13 Door ................................................................... 15
washing with 3 different solvents ....................45 Dwell time ........................................................... 93
Autosampler
4 ml vial rack .................................................70
accessories ...................................................11 E
arm ...................................................16, 64, 72
autosampler tray ..........................................118 ENTER key .................................................. 82, 91
error codes ..................................................153 Equilibration time .............................................. 107
installation parts (for GC-2014, GC-14A/B) .....11 Error codes ....................................................... 151
installing ........................................................64 autoinjector ................................................ 151
name ............................................................16 autosampler ............................................... 153
part No. ...........................................................2 sub-injector ................................................ 154
specifications ................................................14
Error messages ................................................ 154
Autosampler control terminal ........................17, 18
External control ................................................ 133
Autosampler tray ...............................................118 command ................................................... 140
Aux control terminal ............................................18
F I
F21 ..................................................................... 95 Initializes the RAM ............................................. 97
Fiber optic cable ............................................... 149 INJ port septum .................................................. 57
Fiber optic link terminal ................................ 17, 18 INJ SPEED-PLUNGER ...................................... 93
Final priority sample No. .................................... 94 INJ SPEED-SYRINGE ....................................... 93
FINAL SAMPLE ......................................... 94, 112 Inj. speed (plunger) ............................................ 93
Flange ................................................................ 27 Injection
FPD fan ...................................................... 62, 149 mode ............................................ 93, 104, 113
FUNCTION key ............................................ 82, 92 sample ......................................................... 86
speed ........................................................... 93
Fuse ........................................................... 18, 148
INJECTION MODE ............................................ 93
Fuse holder ........................................................ 18
Injection port septum .......................................... 59
Injection procedure ............................................. 85
G INJECTION SAMPLE ................................. 94, 112
INJECTION SAMPLE2 ....................................... 94
GC network ................................................ 77, 110 Injection volume ................................................. 85
GC14/2014 Injector 1 control terminal ............................. 17, 18
mounting bracket .......................................... 73
Injector 2 control terminal ............................. 17, 18
GC-14A/B
Installation parts
mounting bracket .......................................... 11
autoinjector .................................................... 8
spacers ........................................................ 25
autosampler ................................................. 11
GC-2010 GC-14A/B, 15A, and 16A .............................. 10
AOC other parameters ................................ 102 GC-17A, 18A, and 1700 .................................. 9
AOC parameters ........................................ 101 GC-2010 ........................................................ 8
GC-2010Plus GC-2010Plus .................................................. 8
AOC other parameters ................................ 102 GC-2014 ........................................................ 8
AOC parameters ........................................ 101 GC-2014, GC-14A/B ..................................... 11
GC-2014 GC-2025 ........................................................ 9
AOC other parameters ................................ 102 Installing
AOC parameters ........................................ 101 autoinjector .................................................. 21
mounting bracket .......................................... 11 autosampler ................................................. 64
spacers ........................................................ 23 Internal standard ........................................ 68, 125
GC-2025 Isothermal analysis .......................................... 108
AOC other parameters ................................ 102
AOC parameters ........................................ 101
GCMS-QP5000 .................................................. 76 K
GCMSsolution .................................................... 75
GCsolution ......................................... 75, 110, 128 Keyboard unit ..................................................... 15
Gripper (autosampler) .......................... 16, 72, 122 Keypad ............................................................... 81
Gripping position ................................................ 70 Kinds of vials ...................................................... 95
Ground terminal ................................................. 18
H
Heating ............................................................... 14
Home position .............................................. 80, 87
Humidity ............................................................. 19
L O
Large volume injection mode ............................115 OCI syringe ................................................ 35, 149
Light Operation environment ....................................... 19
power ..............................................16, 18, 122 OPTION ............................................................. 80
run ................................................................82 Option ........................................................... 62, 70
Line .....................................................................80 autosampler ................................................. 70
Loading Optional ................................................ 64, 73, 149
rack ...............................................................41 autoinjector ................................................ 149
Long rack ..............................................39, 62, 149 autosampler ......................................... 64, 149
Low-speed injection method .............................115 cooling/heating unit ....................................... 14
C-R7A ........................................................ 149
GC-2010 .................................................... 150
M GC-2010Plus .............................................. 150
GC-2014 .............................................. 73, 149
Maintenance ..............................................145, 156 long rack guide ............................................. 39
end of the syringe needle .............................157
plunger ........................................................156
syringe ........................................................156 P
vials ............................................................145
Maintenance parts .............................................148 Parameter
precautions .................................................. 88
Mode (injection) ...................................................93
Parameter keys ............................................ 82, 89
MONITOR key ...............................................82, 90
Parameters ......................................................... 83
Mounting
list .............................................................. 101
syringe ..........................................................26
setting .......................................................... 83
Mounting bracket
Plunger
GC14/2014 ...................................................73
maintenance ............................................... 156
Multi-Inj. .......................................................95, 115 speed ..................................................... 93, 95
Plunger button .................................................... 31
N Plunger drive unit ......................................... 15, 28
Plunger holder ............................................ 31, 147
Name Plunger suction speed ........................................ 95
autoinjector ...................................................15 Power indicator .................................................. 18
autosampler ..................................................16 Power light ....................................................... 122
power supply .................................................17
Power ON ......................................................... 122
Needle guide
Power supply
autoinjector .............................................15, 29
[AOC Parameters] window ............................ 80
injection port ..................................................60
AOC power supply .......................................... 6
Number (sample) ................................................90 built-in power ................................................ 17
Number of injection .......................................85, 90 external power .......................................... 7, 18
NUMBER OF INJECTIONS ............85, 89, 90, 110 GC-2010 ........................................................ 6
Number of injections ...........................................95 GC-2010Plus .................................................. 6
Number of priority injections ................................94 GC-2014 ........................................................ 7
GC-2025 ........................................................ 7
name ............................................................ 17
part No. ...................................................... 2, 6
Power switch ...................................................... 18
Power/status indicator ........................................ 16
PRE SOLVENT WASH ...................................... 93
Pre solvent wash ................................................ 93
Precautions S
analysis ...................................................... 126
other .......................................................... 161 Sample
septum ....................................................... 160 injection ........................................................ 86
silica wool .................................................. 158 number ........................................... 79, 90, 119
syringe ....................................................... 155
Sample aspiration
Pretreatment setup speed ........................................................... 95
GC network ................................................ 132 waiting time .................................................. 93
Priority ................................................................ 94 Sample cooling fan ..................................... 62, 149
Priority analysis SAMPLE SIZE ........................................ 85, 89, 90
autoinjector ................................................ 111
Sample size .................................................. 85, 90
autosampler ............................................... 123
SAMPLE WASH ..................................... 85, 89, 90
PRIORITY SAMPLE ........................... 94, 111, 123
Sample wash ................................................ 85, 90
Priority sample No. ............................................. 94
aspiration amount ......................................... 95
PTV .................................................................. 115 speed ........................................................... 95
PUMPING .......................................................... 93 Septum ....................................................... 48, 147
Pumping ....................................................... 86, 93 INJ port ........................................................ 57
aspiration amount ......................................... 95 injection port ................................................. 59
speed ........................................................... 95 precautions ................................................ 160
vial ............................................................... 47
Septum nut ................................................... 21, 59
R
Sequential analyses ......................................... 104
Rack ............................................................. 38, 99 SET .................................................................... 80
loading ......................................................... 41 Set .................................................................... 147
long rack guide ............................................. 39 Settings
vial placement ........................................ 55, 79 parameters ................................................... 83
Rack gear ........................................................... 15 precautions .................................................. 88
Rack guide ......................................................... 15 Short rack guide ................................................. 38
Rack position ........................................ 70, 99, 117 Silica wool ........................................................ 158
READY signal polarity ........................................ 99 Solvent flush ....................................... 74, 113, 124
Removing aspiration volume ......................................... 95
autoinjector .................................................. 57 solvent ......................................................... 68
syringe ......................................................... 35 SOLVENT SELECT ........................................... 93
Reset .................................................................. 88 Solvent select ..................................................... 93
RESET key ................................................... 82, 91 Solvent vial (volume limit) ................................... 53
ROM version ...................................................... 97 SOLVENT WASH ................................... 85, 89, 90
RS-232C terminal ......................................... 17, 18 Solvent wash ................................................ 85, 90
Run light ............................................................. 82 Spacers
GC-14A/B ..................................................... 25
GC-2014 ...................................................... 23
Specifications ..................................................... 13
autoinjector .................................................. 13
autosampler ................................................. 14
Speed
injection ........................................................ 93
plunger ......................................................... 93
pumping ....................................................... 95
sample aspiration ......................................... 95
sample wash ................................................ 95
syringe ......................................................... 93
Speed of plunger ................................................ 95
START key ................................................... 83, 91
T
Temperature ........................................................19
Temperature programmed analysis ..................107
The position when returning ................................95
Transmission interface ......................................149
Tray .............................................................16, 118
Turning the power ON/OFF .................................80
U
Upload .................................................................81
Use of the sampler ..............................................96
Use of the sub AOC ............................................96
Use the same param ...........................................96
Using 3 solvent vials .....................................45, 95