XRD SOP Vs 2013
XRD SOP Vs 2013
XRD SOP Vs 2013
X-Ray Diffractometer
For the instructions below, hardware switches on the machine are underlined. Buttons
or icons within the computer program which you point at and click (or double click) are
in bold. The titles of dialog boxes or command lines which require you to type
something in are in “quotes.” Files or commands that you choose from a list of
possibilities (possibly by using a slide bar) are in italics.
1. Press Control Power On button on the front of the Scintag. It will be lit when it is
on and “interlock open” will be lit.
2. Press X-ray On button to let x-ray tube warms up.
3. Open the door and insert your sample into the holder assembly.
4. Verify the slits sizes. This is done by pulling the metal covers that hide the slits
toward you. The two slits on the tube assembly should normally be 2mm (closest
to the tube) and 4mm (furthest from the tube). The two slits on the detector
assembly should normally be 0.5mm (furthest from the detector) and 0.3mm
(closest to the detector). If you desire to change slits check with Hogan Martin or
Scott Chumbley.
Tube slits Detector slits
5. Close the door and press the Interlock Reset button. This button must be pressed
each time the door is opened and closed.
BEFORE turning on the x-ray tube, you must first push and hold the XRAY OFF button in
order to see the voltage and current that the tube will go to when turned on. The LCD
displays on the console should read 2 mA and 10kV (both figures OK +/- 10%) while the
button is being held. If the console reads the 2mA/10kV, go ahead and push the XRAY
ON button.
[Note: if ever you turn on the x-ray tube and it starts to energize past about 2mA/10kV,
immediately push the XRAY OFF button and follow the procedure in the appendix]
II. Computer Set-up
1. If the computer is not up and running you may have to press Control+Alt+Del to
make the computer active. The small window appears. Hit the Login.
2. Select “Use an ISU Net ID” option to go further. Then hit to continue to sign in.
There will be two events listed in the column on the left of the window for the
“Eventlist.evt” file. These are Generator and Scan. When you selected the
“Evenlist.evt” window the program will have defaulted to the first event on the list. So
the right side of the page shown above is actually the “Generator” set-up page. (Note:
The following instructions refer to choosing events from the list. The buttons marked
Generator and Scan are only used if you want to add new events. The student.evt
template described already includes the Generator and Scan set-up.)
4.1 Under “General”, there is a command line entitled “Raw File Name”
which will look similar to this:
This command line shows you all the directory are in. It is recommended that you
save your data in the “C:” drive. After you have completed your experiment,
remember to transfer your data to your network folder! You can now name your
data file. It is recommended that you use the Browse... button since this will save
time and prevent errors. This will open a window similar to the one below:
Example of the window that opens when you use the Browse... button
Please save your data to C drive under the XRAY DATA folder. The path should be
exactly this: C: \ XRAY DATA\ Your name. If you don’t have your own folder, please
create a new folder for you. Type a name for your file and it will replace the
highlight text. The file will automatically be placed in the directory and given a
“.raw” suffix.
1. Check to see that the green X-ray on button is lit. If it is not lit, push this button
and wait 5 minutes for the tube to warm up before proceeding. [note: if ever you
turn on the x-ray tube and it starts to energize past about 2mA/10kV, immediately
push the XRAY OFF button and follow the procedure in the appendix]”Review the
status of the diffractometer hardware, i.e. the cover is closed, the interlock reset
has been pushed (it shouldn’t be lit.)
2. To execute the scan you have specified press the Go from Top button.
3. “Save changes to Events?” hit No.
4. You may get the message “Do you want to initialize the hardware?” hit Yes. The
window below appears:
Hardware initialization window Position of Tube and Detector angle readouts
5.1 Go to the Scintag unit and check the angular position of the detector and
tube. You may have to open the door to see the angular readouts
clearly. The “OM” reading corresponds to the tube and “TH”
corresponds to the detector. The blue arrow in the above picture
corresponds to the angle reading for OM/tube whereas the red arrow
indicates the angle reading for OM/tube. To read the angular value of
the detector, you should look at the reading display under the detector
at 6 o’clock position. If the hardware readouts match the software value
in the program, hit OK.
5.2 If the hardware and software values do NOT match, hit Calibrate. This
will cause the detector and tube to move to the nearest whole angle.
5.3 Go to the Scintag unit and note the actual position of the detector and
tube. Type these values in to the “Set Axis Position” box for both
detector and tube. Hit Set after entering each true value.
5.4 If you had to open the door on the Scintag unit, remember to press the
Interlock Reset button before running the scan.
5.5 Verify that X-ray On is lit and hit OK. This will initiate the scan. The
detector and tube will move to the appropriate positions, the shutters
will open, and the scan will begin.
5.6 A small window, as shown below, will appear noting the progress. Click
on the Realtime Display button to watch the progress of the scan.
5.7 When the scan is done you will get a message “Event list execution complete.”
Hit OK.
This section includes instructions related to viewing your data, carrying out a
background fit and correction, identifying the peak locations, and carrying out a search
of the JCPDS files to determine the phase(s) present in your sample.
1. If you have been watching the Realtime Display as described above then your
scan will be visible upon completion. If you want to open a saved data file choose
Open from the “File” menu and select your saved data file.
2. An opened scan may appear similar to the example shown below.
x
Example of a typical scan
You can zoom in on any particular area of the scan by using the right mouse
button.
3. The menu at the top of the scan allows you to carry out an analysis of the scan.
This menu is shown in close-up below. These are discussed in turn.
3.1 This button brings up the Background window shown below. Using the
functions in this window you can fit a Background Curve to the raw data
and remove it, apply a data Smoothing routine, and apply Corrections to
the data for various artifacts e.g. Κα2 radiation etc.
Normally the Box Car Fit and Perform Κα2 Strip buttons will be checked. Press
Calculate to perform the selected functions. A box will appear asking you to save
files “C: Name.” Click Yes.
3.2 Peak Find. Pressing this button brings up the page shown below.
Peak Find page
Be sure that Peak Finder using Digital Filtering is selected and hit Calculate. A window
will come up saying “Found x Peaks.” Click OK.
A scan after the background has been stripped and the peak ID routine has been run
4. Hit OK. The text file will automatically be saved under the folder you created.
APPENDIX:
If the power settings on the tube are not at 2mA/10kV you must go through the
software's hardware/manual/generator control window to reset a few things. It will
insist that you calibrate the goniometer. When you’ve finished the calibration, the
following dialog box will appear:
First be sure that the button in the upper right hand corner says "set local". If that
button says "set remote" then the dials on the front of the diffractometer are in
control, but we want the software to control the voltage and current. So if it says "set
remote" then click that button and it will change to saying "set local" which means it is
in "set remote" mode (yes, this is a confusing arrangement). "REMOTE" means
computer, and "local" means diffractometer dials. After putting things into the
"REMOTE" mode, move the sliders to 10kV and 2mA and click the lower left button
which is "set new power".
Re-check the power supply to the tube by pushing the XRAY OFF button as above.
When you are leaving this dialog box it will ask you:
Answer “Yes”.