FILING INSTRUCTIONS

File this documentation in binder: Image Detection


Subsystem 19 manual

{ IDS19
{ XTV19 Camera head
{ TVB
{ II shield 38(31) cm
{ II generator 38(31) cm
{ Basic Lens 38(31) cm
{ ID measuring field Y
{ IDC
{ PC host
{ II-tube
{ ………………………………
{ ………………………………
{ ………………………………

(mark the presence of this manual on the contents of the card of that binder with ⊗)

IDS19 (03.1) 0.1


Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED


SERVICE MANUAL UNIT

Image Detection Subsystem 19

For serial numbers, see List Of Pages And Drawings

&

Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.

Reproduction in whole or part of this manual is prohibited without the written consent of the copyright owner.
Use of the information contained herein, in any form and/or by any means whatsoever, is strictly reserved for Philips and its
licensees.
Use of this manual by unauthorised persons is strictly prohibited.

Printed in The Netherlands (03.0) 1


Unit Manual Image Detection Subsystem 19 LOPAD

SERVICE MANUAL: UNIT


Image Detection Subsystem 19
TYPE NO:

Manual Code Number: 4522 981 22823 Date: 2004-07-01

LIST OF PAGES and DRAWINGS

0.1 (03.1)

1 (03.0)

2.1 (04.0)

3 (03.1)
--- (03.1)
33 (03.1)

P (03.1)

Z0 (03.1)
Z1-1 (03.1)
Z6-1 (03.0)

4522 981 22812


4522 981 23161
4522 981 23171
4522 981 23632
4522 981 22831

IDS (04.0) 2.1


Copyright (c) 2003-2004 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED


Unit Manual IMAGE DETECTION SUBSYSTEM 19

Image Detection Subsystem 19

Contents

1. INTRODUCTION AND TECHNICAL DATA ...........................................................................................5


1.1. Purpose of manual ...............................................................................................................................5
1.2. Items supplied ......................................................................................................................................5
1.3. Compatibility.........................................................................................................................................6
1.4. Technical data ......................................................................................................................................6
1.5. Safety information ................................................................................................................................6
1.6. Compliance information .......................................................................................................................6
1.7. Identification and labeling.....................................................................................................................7
1.8. Abbreviations and definitions ...............................................................................................................8
1.9. Manual history ......................................................................................................................................8

2. INSTALLATION.......................................................................................................................................9

3. FAULT FINDING ...................................................................................................................................10


3.1. Fault Isolation procedure....................................................................................................................10
3.1.1. Power Concept ...........................................................................................................................11
3.1.2. LEDs and ERROR codes............................................................................................................12
3.1.3. Hardware configuration...............................................................................................................12
3.1.4. Symptom description ..................................................................................................................13
3.1.5. No image.....................................................................................................................................14
3.1.6. Distorted image...........................................................................................................................17
3.1.7. Dose Control Loop ......................................................................................................................20
3.1.8. No communication with SYSCO .................................................................................................23
3.2. Error messages ..................................................................................................................................24
3.3. Logging...............................................................................................................................................24
3.4. Indicators............................................................................................................................................24
3.5. Test facilities.......................................................................................................................................25
3.5.1. POST ..........................................................................................................................................25
3.5.2. BIST ............................................................................................................................................26
3.5.3. Functional test.............................................................................................................................26
3.5.4. Performance test.........................................................................................................................27

4. REPLACEMENTS .................................................................................................................................28
4.1. What to do table .................................................................................................................................28
4.2. Replacement procedures ...................................................................................................................28

IDS (03.1) 3
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

5. ADJUSTMENTS ....................................................................................................................................31

6. PROGRAMMING...................................................................................................................................32
6.1. Hardware............................................................................................................................................32
6.2. Software .............................................................................................................................................32

7. ACCEPTANCE ......................................................................................................................................33
7.1. Acceptance test 1...............................................................................................................................33

P. PARTS LIST ........................................................................................................................................... P

Z. DRAWINGS ............................................................................................................................................ Z

4 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

1. INTRODUCTION AND TECHNICAL DATA

1.1. PURPOSE OF MANUAL


This manual covers the fault finding for Image Detection Subsystem (IDS) 19 and the controllers. This
manual will lead the FSE to the defect / malfunctioning FRU. Replacement instructions will be described in
the unit manuals.

Figure 1 Scope of IDS 19 manual

II TUBE &
HTGEN

II shield
PC HOST
Basic Lens
TVB
IDC

ID
CHA

Supplied unit manuals:


− IDS 19 unit manual
− II generator 38(31) cm unit manual
− II shield 38(31) cm unit manual
− Basic lens 38(31) cm unit manual
− Image Distributor Measuring Field Carrousel unit manual
− XTV19 Camera head unit manual
− XTV19 TVB unit manual
− IDC unit manual
− PC Host unit manual

1.2. ITEMS SUPPLIED


The IDS 19 consists of the following units:
− Camera head (XTV19)
− Image Distributor Measuring Field Carrousel
− Basic Lens 38(31)cm
− II shield 38(31)cm (II tube 38(31) cm)
− II generator for 38(31) cm tube
− PC Host (industrial computer)
− TVB board (PCB)
− IDC board (PCB)

IDS (03.1) 5
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

1.3. COMPATIBILITY
The described IDS 19 is used in:

URF systems:
Multi Diagnost Eleva
Omni Diagnost Eleva
Easy Diagnost Eleva

Options:
No options applicable.

1.4. TECHNICAL DATA


Room temperature : 10 °C to 35 °C.

Environmental data for transport:


Operation ambient temperature : +10 to +40 degrees Celsius.
Storage ambient temperature : -25 to +70 degrees Celsius.
Relative humidity : 20% to 90 %
Air Pressure : 70 – 110 kPa

1.5. SAFETY INFORMATION

WARNING

This equipment generates dangerous high voltages and should be handled with caution.

________________

WARNING

For safe handing of the II generator refer to the II generator unit manual.
________________

1.6. COMPLIANCE INFORMATION

Philips products comply with relevant international and national standards and laws. Information on
compliance will be supplied on request by your local PMS representative, or by:

Philips Medical Systems


PO Box 10.000
5680 DA Best
The Netherlands

Facsimile: +31 40 276 2205

Philips products comply with relevant international and national law and standards on EMC (Electro-
magnetic compatibility) for this type of equipment when used as intended.

6 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

Such laws and standards define both the permissible electromagnetic emission levels from equipment and
its required immunity to electromagnetic interference from external sources.

General:
• IEC 60601-1 "Medical electrical equipment. Part 1: General requirements for safety" (second edition
1988, including amendment no. 1 (1991), amendment no. 2 (1995))
• IEC 60601-1-2 "Medical electrical equipment. Part 1: General requirements for safety. 2. Collateral
Standard: Electromagnetic compatibility - Requirements and tests" (second edition, 2001)
• IEC 60601-1-3 "Medical electrical equipment. Part 1: General requirements for safety. 3. Collateral
Standard: General requirements for radiation protection in diagnostic X-ray equipment" (first edition,
1994-07)
• IEC 60601-1-4 "Medical electrical equipment. Part 1: General requirements for safety. 4. Collateral
Standard: Programmable electrical medical systems" (first edition, 1996-05)
• UL 2601-1 "Medical electrical equipment, Part 1: General requirements for safety" (second edition,
October 1997)
• CAN/CSA -C22.2 No. 601.1-M90 "Medical Electrical Equipment. Part 1: General requirements for safety"
(November 1990, including Supplement C22.2 No. 601.1S1-94 and Amendment no. 2 (1998))
• 21CFR, Subchapter J

1.7. IDENTIFICATION AND LABELING

Figure 2 Overview IDS 19 (hardware)

Label

For other labeling per unit refer to relevant unit manual.

IDS (03.1) 7
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

1.8. ABBREVIATIONS AND DEFINITIONS

BIST Built In Self Test


CAN Central Area Network
DI Digital Imaging
FRU Field Replaceable Unit
FSE Field Service Engineer
Htgen High Tension generator
ID Image Distributor
IDC II and Dose Control board
II Image Intensifier
LED Light Emitting Diode
PCB Printed Circuit Board
PMS Philips Medical Systems
POST Power On Self Test
SAN Synchronization Area Network
SRM System Reference Manual
SYSCO System Controller
TVB TV Control Board
URF Universal Radiography and Fluoroscopy

1.9. MANUAL HISTORY

Date Version Name Reason of changes


12-01-2003 1.0 Image Detection Subsystem 19
17-07-2003 2.0 Image Detection Subsystem 19 Information added for IDC Power Supply

8 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

2. INSTALLATION

The IDS19 based on XTV19 camera is factory installed and adjusted and for this reason no installation
instructions are necessary.

See the Installation manual on system level, to install the II/TV cabinet.

IDS (03.1) 9
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

3. FAULT FINDING

3.1. FAULT ISOLATION PROCEDURE


Two major parts can be divided where errors can occur.

-Dose control loop


-XTV video path

Figure 3 IDS 19 functionality’s XGEN cabinet

xray tube generator CAN

collimator
room geometry

Ixxat
PCI

CAN
interface
DOSE
dose link

LOOP Host

CAN
I-cabinet

Processor
board
HTGEN

II
Tube
HT generator cable IDC
ID cable
mf.carr
BD2
BD1

SAN
ID

local SAN

RIO
Ibis-PC rack

XTV19 XTV19
camera cable
CHA TVB

VIDEO
serial link

Video Link

PATH
Digital
Imaging

10 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

3.1.1. POWER CONCEPT


The picture below shows the power concept of the IDS19 sub system. You can use this picture to check if all
power supplies are correct. The table tells what voltages are applicable for the components.

Image XTV19
II-Segment distributor camera

legend:
power interface
IDS.XTV19cable
IDS.IIcable IDS.IDcable protective earth

PC-Host rack + backpanel


PCI
< power>
Host

(structure)
power CAN interface
XTV19 PC-Host
supply: IDC board
HSL TVB processor board
PCI, (IXXAT)
<power>
HSL

IDC PS

IDS.Power

PC Host power supply IDC power


Physical component supply**
+3.3V +12V -12V +5V +15V -15V
II-segment * *

Image Distributor * *
XTV19 * * * *

IDC-board [MF. carrousel] * *


IDC-board [other circuits] * * * * *
CAN i/f board *

** IMPORTANT:
When one, or both, Power Supplies for the IDC are replaced they must be adjusted to the correct output
voltage of 15Vdc, as at delivery they are set to 12Vdc.
Adjustment must be done by means of the pot-meter on the unit.

IDS (03.1) 11
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

3.1.2. LEDS AND ERROR CODES


First check if there are any red LED's on either the TVB board or the IDC board. See chapter 3.4 for
explanation of the LED colors.
Look in the Field Service Framework for error codes. Refer to system documentation for explanation of the
error codes.

3.1.3. HARDWARE CONFIGURATION


The IDS19 subsystem will not function properly if either TVB or IDC is missing in the PC Host or the
interfaces and devices are not connected. The location of the TVB and IDC in the PC host is fixed. The slots
in the PC host are not numbered. Counting from left to right the IDC has to be inserted in slot 15. The TVB
has to be inserted in slot 14. This is illustrated in figure 4.

Figure 4 Hardware configuration

TVB IDC
14 15

PC host

Slot number 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Left side PC host Right side PC host

The POST of the TVB and the IDC are independent of the status of the PC Host. But without Windows NT
operating or the PC Host POST the field service framework will not work.

NOTE

In the picture the +/- 15V on the front plate is not shown. To have a working system this power supply has to
be present (BHFLD-X2).
When it's necessary to disconnect the cable between the IDC and TVB the PC host has to be kept powered
on.

________________

12 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

3.1.4. SYMPTOM DESCRIPTION


To locate the problem following fault trees must be followed. Starting point is that all connections are
checked on their presence and all calibrations are carried out.

Start

error messages on
Yes Act according error message
FSF or LED?

No

Is there communication No Go to communication tree


between SYSCO and
IDS?

Yes

Do you have an image No Go to no image tree


on monitor?

Yes

Is the problem dose


Yes Go to dose control loop tree
related?

Yes

Do you have an image


Yes Go to distorted image tree
but is it distorted?

No

Contact Helpdesk

IDS (03.1) 13
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

3.1.5. NO IMAGE

The fault finding tree below helps to locate the cause if no image is visible at the exam (and reference)
monitor.

Preconditions:

PC host and CPU are working and communication between SYSCO and IDS 19 is possible. If this is not the
case then go to no communications fault finding tree.
Also the Digital Imaging, the monitors and the link between the Digital Imaging and the monitors should be
working. If this is not the case refer to the Digital Imaging unit manual.

14 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

No Image

No Image

make continuous
fluoro

Look at kV/mA
values on desk

kV between 60 No kV or mA
kV >110 kV <40
and 80 values

Problem is in dose Problem is in Problem is in dose


Problem is in X-ray
control loop. Go to videopath. Go to video control loop. Go to
generator or tube
dose control loop tree path tree dose control loop tree

IDS (03.1) 15
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

No Image but kV and mA settings are ok.


No image kV mA
is oke

Make testimage
via sysco
(generated on
TVB)

Do you have
Do video link test
Do Local SAN test Yes an image on No
the monitor?

Measure where Problem is in DI or


ctrl_i is lost and Test oke? yes
No Test oke? Monitor
replace according
to result
No
Yes

Do internal video
Go to IDC or TVB Do global SAN path test
unit manual test

Measure where Go to TVB unit


ctrl_i is lost and Test oke? No manual
No Test Oke?
replace according
to result
Yes
Yes

Go to IDC unit Do internal DI


manual or RIO DACQ test
Check ctrl_i signal
manual
from RIO (See
RIO manual)

Test Oke? No Problem is in


DACQ board in DI
Go to RIO manual No ctrl_i oke?

Yes
Yes

Go to XTV 19
camera head unit Problem is in cable or
manual interface circuits from
TVB to DACQ

16 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

3.1.6. DISTORTED IMAGE

The fault finding tree below helps to locate the cause if a distorted image is visible at the exam (and
reference) monitor. A distorted image can typically be identified as an image with bad contrast and
brightness and an image with artifacts (e.g. defect pixels).

Preconditions: PC host and CPU are working and communication between SYSCO and IDS 19 is possible.
If this is not the case than go to no communications fault finding tree.
Also the Digital Imaging, the monitors and the link between the Digital Imaging and the monitors should be
working. If this is not the case refer to the Digital Imaging unit manual.

IDS (03.1) 17
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

Distorted Image

No Image

make continuous
fluoro

Look at kV/mA
values on desk

kV >110 kV between 60 kV <40


and 80

Problem is in dose Problem is in Problem is in dose


control loop. Go to videopath. Go to video control loop. Go to
dose control loop tree path tree dose control loop tree

18 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

Distorted image but kV and mA are ok


distorted image,
kV mA is oke

Check HT generator manual for


Is the image
Yes control voltages and HT Voltages
out of focus?
and measure them

No

Do videolink test

Check interface
Test Oke? No
TVB-DI

Yes
Make testimage
via sysco
(generated on
TVB)

Do analogue Did the


Make testimage
camera head Yes complaint go No
on DI
pattern test away?

Did the
Go to XTV19 Did the complaint go No Problem is in DI or
camera head unit No complaint go away? Monitor
manual away?

Yes
Yes

Do internal DI
Do life image test DACQ test

Go to XTV19 camera Problem is in


Test oke? No DACQ board in DI
head or Basic lens No Test Oke?
unit manual

Yes
Yes

Problem is in
Problem is in II or cable from TVB to
ID DACQ

IDS (03.1) 19
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

3.1.7. DOSE CONTROL LOOP

The fault finding tree below helps to locate the cause if there is a problem in the dose control loop.
With dose control loop also doserate control loop is meant.

Preconditions: PC host and CPU are working and communication between SYSCO and IDS 19 is possible.
If this is not the case than go to no communications fault finding tree.

20 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

Dose control problem

Dose control
problem

Is result IDC Go to IDC unit


No
POST oke? manual

Yes

Measure where
ctrl_i is lost and Is global SAN
No
replace according test oke?
to result

Yes

Got to IDC unit


manual or RIO Check ctrl_i signal
manual from RIO (See
RIO manual)

Go to RIO manual No ctrl_i oke?

Yes

Problem is in X-
Is doselink test
No ray generator,
oke?
tube or collimator

Yes

Is Measuring
Go to ID measuring field
field carousel No
carrousel unit manual
test oke?

Yes

Check dose(rate)voltages
on X-ray generator
according to table below

IDS (03.1) 21
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

kV Value Voltage at measuring point Fault location


> 110kV << 0V Dose sensor in ID
>> 0V X-ray generator
< 40kV << 0V X-ray generator
>> 0V Dose sensor in ID

If the voltage at the measuring point of the X-ray generator is 0V and the dose(rate) is a factor 2 of what it
should be then the fault is in the interface of ID to IDC.

22 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

3.1.8. NO COMMUNICATION WITH SYSCO

When on the Field Service Framework the message is shown that there is no communication with IDS follow
next Fault tree.

Precondition:
All CAN cabling and terminators have to be connected.

No communication
with Sysco

Is keyboard
Connect keyboard Did PC host
No and mouse No
and mouse start up?
connected?

Yes Yes

Go to PC host unit Did CAN board


manual Replace CAN
in IDS pass No
board
self test?

A monitor needs to be connected to the Host PC for this test.

Yes

Did node Check


No
check pass? connections

Yes

Contact helpdesk

IDS (03.1) 23
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

3.2. ERROR MESSAGES


The messages that are visible for the Field Service Engineer are displayed on the Field Service Framework
user interface. Refer to system documentation.

3.3. LOGGING
The logging that is visible for the Field Service Engineer are displayed on the Field Service Framework user
interface. Refer to system documentation.

3.4. INDICATORS
The following PCB’s contain LEDS
− TVB
− IDC

Meaning LED’s TVB:


The front strip contains 6 LEDS, grouped in 3 rows of 2 LEDS each. The LEDS have the following
color scheme:

LED Left column Right column


Color Scheme
Top-row : RED RED
Center-row : GREEN GREEN
Bottom-row : YELLOW YELLOW

Only the RED and GREEN LED’s in the left-column are used (top-bottom orientation as
in drawing Z-1 TVB unit manual)

Information coding of these LED’s is as follows :

RED GREEN Function / message : Action:


off Off No power
off On Self-Test OK
on Off Check interfaces and
Self-Test Failed
power supplies
on On Self-Test Running

Meaning LED’s IDC:


The front strip contains 6 LEDS, grouped in 3 rows of 2 LEDS each. The LEDS have the following
color scheme:

LED Left column Right column


Color Scheme
Top-row : RED RED
Center-row : GREEN GREEN
Bottom-row : YELLOW YELLOW

The left three LED’s are not used for service. These LED’s will flash during startup.

The right three LED’s have the following meaning:

24 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

LED Right column Meaning


Color Scheme
Top-row : RED ON Error during
startup and
POST failed
Center-row : GREEN ON Processor
busy
Bottom-row : YELLOW ON/OFF Heartbeat of
IDC.

When red LED is on check the following


- +/- 15V of IDC power supply
- San interface
- HT generator interface
- ID interface

The yellow LED will blinking a specified way shown in the following table:

ON / OFF % Meaning
50 / 50 Start up
After about 1 minute: irregular Application started with correct
firmware
After about 1 minute: 50 / 50 Wrong Firmware

3.5. TEST FACILITIES


Test facilities are available via the Field Service Framework. To activate test on IDS 19 level select the
relevant test on the SYSCO via the Field Service Framework.

3.5.1. POST
The following units have a POST:
− Camera head (XTV19)
The POST tests the camera head, the TVB and the cable between the camera head and the TVB.

− PC Host (industrial computer)


Interfaces and CPU are tested.

− TVB board (PCB)


When started the POST of the TVB tests the following: the TVB (PCB), the PCI bus of the TVB to the
PC host.

− IDC board (PCB)


When started the POST of the IDC tests the following: the IDC (PCB), the PCI bus of the IDC to the PC
host, the cable between IDC and Image distributor, the connection between the IDC and the RIO and
the connection between the IDC and the II generator (High Tension generator).

− CAN
The CAN POST will test the CAN (PCB), the connection between the CAN board and the PC host (PCI
bus) and the connection between the CAN board and the RIO.

IDS (03.1) 25
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

3.5.2. BIST
The following units have a BIST:
− PC Host (industrial computer)
− IDC board (PCB)

3.5.3. FUNCTIONAL TEST


The functional tests can be started by the Field Service Framework at the SYSCO. The following functional
tests are available:
− XTV video path test
− Local SAN test
− Global SAN test
− Test Images
− Measuring Field Carrousel test
− XTV video link test
− Doselink test
− Life image
− Average video level test
− Digital Image Correction board pattern test

XTV video path test:


This test covers the XTV19 Camera Head, the camera cable and the TVB.

Local SAN test:


This test will test the local SAN connection between the IDC and the TVB.

Global SAN test:


The global SAN test will check the connection between the IDC and the RIO.
Test Images:
The following test images can be selected by the Field service Engineer:
Test image 1: analog test pattern generated by camera head and send to Digital Imaging
Test image 2: digital test pattern generated by camera head send to Digital Imaging
Test image 3: Digital test pattern generated by TVB and send to Digital Imaging
Test image 4: A defect pixel map send to Digital imaging

Measuring Field Carrousel test:


In this test the measuring field carrousel is moving mechanically to all the available measuring field positions.

XTV video link test:


The video link test will test the connection between the TVB and the Digital Imaging

Doselink test:
The Dose link test will test the connection between the IDC and the X-ray generator.

Life image:
With the life image the path from the camera head, TVB and video link to the Digital Imaging is tested.

Average video level test:


This test is not implemented for XTV19.

26 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

Running Bar Test:


This test is only used for development.
To check dose control:

Digital Image Correction board pattern test:


This test outputs on the video link a test pattern which is generated by the TVB

3.5.4. PERFORMANCE TEST


There are no performance tests available.

IDS (03.1) 27
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

4. REPLACEMENTS

4.1. WHAT TO DO TABLE


With the help of the Faultfinding section you have located which item has to be replaced.
The standard replacement procedure is:
• Verify hardware programming (jumpers/cables etc.).
• Replace the defective item.
• Perform adjustments/software programming.
• Hand-over to the customer.

For WHAT TO DO TABLES refer to the applicable unit manual.

4.2. REPLACEMENT PROCEDURES

The replacement procedures of the units of the IDS 19 are described in the unit manuals which can be found
in this binder.
− Replacement II generator => II generator 38(31) cm unit manual
− Replacement II tube => II shield 38(31)unit manual
− Replacement II shield => II shield 38(31)unit manual
− Replacement Basic Lens assembly => Basic lens assembly 38(31) cm unit manual
− Replacement Basic Lens optics => Basic lens assembly 38(31) cm unit manual
− Replacement Measuring field Carrousel => Image Distributor Measuring Field Carrousel
31/38 cm Unit manual
− Replacement Photo-amplifier board HG => Image Distributor Measuring Field Carrousel
31/38 cm Unit manual
− Replacement Photo Diode assembly => Image Distributor Measuring Field Carrousel
31/38 cm Unit manual
− Replacement Lamp holder assembly => Image Distributor Measuring Field Carrousel
31/38 cm Unit manual
− Replacement XTV19 camera head => XTV19 Camera head unit manual
− Replacement XTV19 TVB => XTV19 TVB unit manual
− Replacement IDC => IDC unit manual
− Replacement PC Host +CPU => PC host unit manual
− Replacement Power supply/Fan Tray => PC host unit manual
− Replacement CPU => PC host unit manual
− Replacement Floppy/CD-Rewriter assembly => PC host unit manual
− Replacement Hard disk assembly => PC host unit manual
− Replacement Video Card => PC host unit manual
− Replacement Ethernet => PC host unit manual

28 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

Figure 5 Rear view II segment

IDS (03.1) 29
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

Figure 6 Side view II segment

30 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

5. ADJUSTMENTS

The following adjustments have to be performed after a replacement:


-II Focussing
-Optical Focus
-Centring Camera (refer also to XTV 19 camera head unit manual)
-Horizon Camera (refer also to XTV 19 camera head unit manual)
-II diameter
-Centring Measuring Field
-Dose/Doserate

Refer to Field Service Framework and system documentation for instructions.

IDS (03.1) 31
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

6. PROGRAMMING

6.1. HARDWARE
The IDS 19 is not hardware programmable.

6.2. SOFTWARE

For installation and configuration of the Software see following documents:

Installation: Release Bulletin.


Configuration: Field Service Framework documentation
SRM

32 (03.1) IDS
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

7. ACCEPTANCE

7.1. ACCEPTANCE TEST 1


Perform all adjustments mentioned in chapter 5 after replacing a unit of the IDS 19. If the II generator is
replaced check the voltages as described in the II generator unit manual.

IDS (03.1) 33
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

P. PARTS LIST

For parts refer to the following:

On TECHNET or PHILIPS Intranet go to the Electronic Spare-parts Catalog or refer to the Spare Parts
Finder on CD-ROM.

IDS (03.1) P
Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.


Unit Manual IMAGE DETECTION SUBSYSTEM 19

Z DRAWINGS

Z Figure 7: Overview IDS 19


Z1-1 Simplified Diagram Overview IDS19
Z6-1 Exploded view IDS19

Figure 7 Overview IDS 19

IDS (03.1) Z.0


Copyright (c) 2003 Koninklijke Philips Electronics N.V.

ALL RIGHTS RESERVED.