Inteliq Cardiac Advance Control Levels 1, 2, 3 and 4

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

InteliQ Cardiac Advance Control

Levels 1, 2, 3 and 4
Level 1 97311T
12016484 Level 1 6 x 3 mL
Level 2 97312T
12016485 Level 2 6 x 3 mL 2025-11-30 97310T Level 3 97313T
12016486 Level 3 6 x 3 mL
Level 4 97314T
12016487 Level 4 6 x 3 mL

https://www.myeinserts.com/97310T

INTENDED USE LIMITATIONS


InteliQ Cardiac Advance Control is intended for use as an assayed quality control serum to 1. This product should not be used past the expiration date.
monitor the precision of laboratory testing procedures for the analytes listed in this 2. This product is not intended for use as a standard.
package insert. 3. If there is evidence of microbial contamination or excessive turbidity in the product, discard
SUMMARY AND PRINCIPLE the tube.
The use of quality control materials is indicated as an objective assessment of the precision of 4. For Laboratory/Professional Use Only
methods and techniques in use and is an integral part of good laboratory practices. Multiple 5. Values may gradually decrease over the product shelf life for the following analyte:
levels of control are available to allow monitoring of the test system’s reliability. CK-MB Isoenzyme.
For customers in Germany: Quality control materials are required for assessment of laboratory WARNING
performance as described in the “Guideline for Quality Assurance of Medical Laboratory
Examinations following the German Medical Association” (Rili-BÄK regulation). Biological source material. Treat as potentially infectious.
REAGENT The human source material used to manufacture this product was tested as required by FDA
This product is prepared from human source material with added constituents of human and accepted methods. Tests results were non-reactive or negative for evidence of infection due to
animal origin, stabilizers and preservatives. This product is provided in liquid form Human Immunodeficiency Virus (HIV), Hepatitis B Virus (HBV) and Hepatitis C Virus (HCV). This
for convenience. product may also contain other human source materials for which there are no approved tests. In
STORAGE AND STABILITY accordance with good laboratory practice, all human source material should be considered
This product will be stable until the expiration date when stored unopened at -20 to -70°C. potentially infectious and handled with the same precautions used with patient specimens.
Refrigerator Storage (All Instruments, Off-Board, 2 to 8°C):
This product will be stable as follows:
- All analytes: 10 days
Except: Warning
- Troponin I and Homocysteine: 5 days Contains 5-Chloro-2-methyl-3(2H)-isothiazolone, mixture with 2-methyl-3(2H)-isothiazolone
Instrument Storage (Siemens Atellica, On-Board, 2 to 8°C): Hazard (H) and Precautionary (P) Statements
This product will be stable as follows: H317 May cause an allergic skin reaction.
- All analytes: 10 days H412 Harmful to aquatic life with long lasting effects.
- Galectin-3 and Troponin T: No claim is made for expected values or the stability. P333+P313 If skin irritation or rash occurs: Get medical advice/attention.
This product is shipped under frozen conditions. P273 Avoid release to the environment.
PROCEDURE P501 Dispose of contents/container in accordance with local/ regional/ national/
This product should be treated the same as patient specimens and run in accordance with the international regulations.
instructions accompanying the instrument, kit, or reagent being used. P280 Wear protective gloves/protective clothing/eye protection/face protection.
This product can only be used following either instrument storage or refrigerator storage P302+P352 IF ON SKIN: Wash with plenty of soap and water.
procedure, but not a combination of both. Safety Data Sheet (SDS) available for professional users on www.bio-rad.com.
To Thaw the Product: SPECIFIC PERFORMANCE CHARACTERISTICS
If the product has been stored frozen, allow it to stand at room temperature (18 to 25°C) until it This product is a stabilized liquid product manufactured under rigid quality control standards. To
is completely thawed. Once thawed, gently invert the tube several times to ensure homogeneity. obtain consistent assay values, the control requires proper storage and handling as described.
Refrigerator Storage (All Instruments, Off-Board, 2 to 8°C): ASSIGNMENT OF VALUES
Before use, gently invert the tube several times to ensure homogeneity. Remove the cap and The mean values and corresponding ±3SD ranges in the Assignment of Values Data Charts
place the product in the instrument for testing. Once testing is completed, promptly replace the (available separately) were derived from replicate analyses and are specific for this lot of
cap and return the product to refrigerator storage. For optimal analyte stability, minimize the product. Data from Unity™ Interlaboratory Program are included in the determination of some
time on-board the instrument and exposure to room temperature. ranges. The tests listed were performed by the manufacturer and/or independent laboratories
Instrument Storage (Siemens Atellica, On-Board, 2 to 8°C): using manufacturer supported reagents and a representative sampling of this lot of product. It is
Remove the cap and place the product in the instrument for testing. Store the product opened for recommended that each laboratory establish its own acceptable ranges and use those provided
the duration of the on-board stability claim. only as guides. Laboratory established ranges may vary from those listed during the life of this
Safety Precautions control. [Customers in Germany have to follow the requirements as described in the Rili-BÄK
• Caution should be used when handling this product to prevent splashing. Wear appropriate regulation.] Variations over time and between laboratories may be caused by differences in
eye/face protection when using this product to protect from splashes. laboratory technique, instrumentation and reagents, or by manufacturer test
• If, during the use of this device or as a result of its use, a serious incident occurs, please method modifications.
report it to Bio-Rad Laboratories and to your national health authority, as required.
• Dispose of any discarded materials in accordance with the requirements of your local Unity™ Interlaboratory Program is a proprietary Bio-Rad software program holding more than 2
waste management authorities. In the event of damage to packaging, contact the local billion QC data points from thousands of laboratories.
Bio-Rad Laboratories Sales Office or Bio-Rad Laboratories Technical Services.
INSTRUCTIONS FOR OBTAINING THE DATA CHARTS
To obtain the Data Charts use the checkboxes on the Customize My eInserts page to select your
chart items. You can View & Print your selections or Save Selections. Follow the directions on
this website to receive email notifications of insert updates. Alternate methods for receiving data
charts are available by contacting your local Bio-Rad Laboratories Office.
BARCODE LABELS
For some instruments, supplemental barcode labels may be available at https://blg.qcnet.com.
Ensure the product tube is clean and dry prior to applying the label.

Page 1 English • 2023-11 • 30027-00W-00


InteliQ Cardiac Advance Control, Levels 1, 2, 3 and 4
ANALYTES
B-type Natriuretic Peptide (BNP) Digitoxin Myoglobin Troponin I
CK-MB Isoenzyme Galectin-3 N-terminal pro-Brain Natriuretic Peptide Troponin T
(NT-proBNP)
C-Reactive Protein (CRP) Homocysteine

UNITED STATES, Bio-Rad Laboratories Inc.


9500 Jeronimo Road, Irvine, CA 92618

France, Bio-Rad
3 boulevard Raymond Poincare, 92430 Marnes-la-Coquette
Phone:(33) 1-4795-6000 / Fax:(33) 1-4741-9133

SWITZERLAND, DiaMed GmbH


Pra Rond 23, 1785 Cressier FR

Bio-Rad
Laboratories, Inc. SWITZERLAND, Bio-Rad Laboratories AG
Pra Rond 23, 1785 Cressier FR

Australia, Bio-Rad Laboratories Pty. Ltd., u1A, 62 Ferndell Street, South Granville, New South Wales 2142 • Phone +61 (2) 9914 2800 • Fax +61 (2) 9914 2888
Clinical Austria, Bio-Rad Laboratories Ges.m.b.H., Hummelgasse 88/3-6, A-1130 Vienna • Phone +43 (0) 1 877 89 01 9 • Fax +43 (0) 1 876 56 29
Belgium, Bio-Rad Laboratories N.V., Winninglaan 3, BE-9140 Temse • Phone +32 (0) 3 710 53 00 • Fax +32 (0) 3 710 53 01
Diagnostics Group Brazil, Bio-Rad Laboratórios Brasil Ltda, Avenida Doutor Chucri Zaidan, 1.240 cj 1902 e 1904 Morumbi Corporate Santo Amaro, São Paulo, SP CEP 04711-130 • Phone +55 11 3065 7550
Canada, Bio-Rad Laboratories Ltd., 2403 Guénette Street, Montréal, Québec H4R 2E9 • Phone +1 514 334 4372 • Fax +1 514 334 0872
China, Bio-Rad Laboratories (Shanghai) Co., Ltd. Room 601, Allian Plaza, No. 168 Jingzhou Road, Yangpu District, Shanghai 200082 • Phone +86 21 6169 8500 • Fax +86 21 6169 8599
9500 Jeronimo Road Czech Republic, Bio-Rad spol. s r.o., Pikrtova 1737/1a, 14000 Prague 4 • Phone +420 241 431 660
Irvine, California 92618 Denmark, Bio-Rad Laboratories, Symbion Science Park, Fruebjergvej 3, DK-2100 Copenhagen East • Phone +45 44 52 10 00 • Fax +45 44 52 10 01
(800) 854-6737 Finland, Bio-Rad Finland Oy, Kutomotie 16 FI-00380, Helsinki • Phone +358 9 804 22 00
France, Bio-Rad, 3 boulevard Raymond Poincaré, 92430 Marnes-la-Coquette • Phone +33 (0) 1 47 95 60 00 • Fax +33 (0) 1 47 41 91 33
FAX (949) 598-1550 Germany, Bio-Rad Laboratories GmbH, Kapellenstraße 12, D-85622 Feldkirchen, Munich • Phone +49 (0) 89 31884 393 • Fax +49 (0) 89 31884 136
bio-rad.com/qualitycontrol Greece, Bio-Rad Laboratories M E.P.E, 2-4 Mesogion Ave. (Athens Tower) 11527 Ampelokipi, Athens • Phone +30 210 7774396 • Fax +30 210 7774376
Hong Kong, Bio-Rad Pacific Ltd., Unit 1101, 11/F DCH Commercial Centre, 25 Westlands Road, Quarry Bay • Phone +85 2 2789 3300 • Fax +85 2 2789 1257
Hungary, Bio-Rad Hungary Ltd., Futó utca 47-53, 1082, Budapest • Phone +36 1 459 6190 • Fax +36 1 459 6101
Technical Service: India, Bio-Rad Laboratories India Pvt. Ltd., EMAAR Digital Greens, 9th Floor, Tower A- Sector 61 Gurugram-122 102, Haryana 122 015 • Phone +91 124 4029300 • Fax +91 124 2398115
Israel, Bio-Rad Laboratories Ltd., 14 Homa Street, New Industrial Area, Rishon Le Zion 75150 • Phone +972 3 963 6025 • Fax +972 3 951 4129
(800) 854-6737
Italy, Bio-Rad Laboratories S.r.l., Via Cellini 18/A, 20090 Segrate, Milan • Phone +39 024 94 86 600 • Fax +39 02 21609399
Japan, Bio-Rad Laboratories K.K., Tennoz Central Tower 20F, 2-2-24 Higashi-Shinagawa, Shinagawa-ku, Tokyo 140-0002 • Phone +81 3 6361 7070 • Fax +81 3 5463 8481
Korea, Bio-Rad Korea Ltd., 10th Floor, Hyunjuk Building, 832-41, Yeoksam-dong, Gangnam-gu, Seoul 135-080 • Phone +82 080 007 7373 • Fax +82 (2) 3472 7003
Mexico, Bio-Rad, S.A., Avenida Eugenia 197, Piso 10-A, Colonia Narvarte, Delegación Benito Juárez C.P. 03020 Mexico, D.F. • Phone +52 (55) 5488 7670 • Fax +52 (55) 1107 7246
The Netherlands, Bio-Rad Laboratories B.V., Postbus 222, 3900 AE Veenendaal • Phone +31 (0) 318 540 666 • Fax +31 (0) 318 542 216
New Zealand, Bio-Rad New Zealand, 189 Bush Road Rosedale, Auckland • Phone +64 (9) 415 2280 • Fax +64 (9) 415 2284
Norway, Bio-Rad Laboratories, Nydalsveien 33, 0484 Oslo • Phone +47 23 38 41 30 • Fax +47 23 38 41 39
Poland, Bio-Rad Polska Sp. z o.o., Przyokopowa 33, Level 4, Building A, 01-208, Warsaw • Phone +48 22 331 99 99 • Fax +48 22 331 99 88
Portugal, Bio-Rad Laboratories, Lda., Edificio Prime, Ave. Quinta Grande, 53 - Fracção 3B Alfragide 26114-521, Amadora • Phone +351 21 47 27 700 • Fax +351 21 47 27 777
Russia, Bio-Rad Laboratorii, 5 Nizhny Susalny Lane, Property 5A, Moscow • Phone +7 (495) 721-14-04 • Fax +7 (495) 721-14-12
Singapore, Bio-Rad Laboratories (Singapore) Pte. Ltd., 3A International Business Park Road, #11-10/16, ICON @ IBP Tower B, Singapore 609935 • Phone +65 6415 3170 • Fax +65 6415 3189
South Africa, Bio-Rad Laboratories (Pty) Ltd., 34 Bolton Ave, Rosebank, Johannesburg • Phone +27 11 442 8508 • Fax +27 11 442 8525
Spain, Bio-Rad Laboratories, S.A., C/ Caléndula, 95, Edificio M. Miniparc II, El Soto de la Moraleja, 28109 Alcobendas • Phone +34 91 490 6580 • Fax +34 91 590 5211
Sweden, Bio-Rad Laboratories A.B., Solna Strandväg 3, SE-171 54 Solna, P.O. Box 1097, SE-172 22 Sundbyberg • Phone +46 844 98053 • Fax +46 8 55 51 27 80
Switzerland, Bio-Rad Laboratories AG, Pra Rond 23, CH-1785 Cressier • Phone +41 (0) 61 717 9555 • Fax +41 (0) 61 717 9550
Taiwan, Bio-Rad Laboratories Taiwan Ltd., 14th F B, No. 126, Sec. 4, Nan-King East Road, Taipei 10546 Taiwan, R.O.C. • Phone +886 (2) 2578-7189 • Fax +886 (2) 2578-6890
Thailand, Bio-Rad Laboratories Ltd., 1st & 2nd Floor, Lumpini I Bldg., 239/2 Rajdamri Road, Lumpini, Pathumwan, Bangkok 10330 • Phone (662) 651 8311 • Fax (662) 651 8312
United Kingdom, Bio-Rad Laboratories Ltd., The Junction Station Road, Watford, Hertfordshire, WD17 1ET • Phone +44 (0) 1923 471301 • Fax +44 (0) 1923 471340 10/22

© 2023 Bio-Rad Laboratories Inc. Printed in the USA


InteliQ is a trademark of Bio-Rad Laboratories in certain jurisdictions. Page 2 English • 2023-11 • 30027-00W-00

You might also like