11.2. Output, Voltage and Current Diagrams
11.2. Output, Voltage and Current Diagrams
11.2. Output, Voltage and Current Diagrams
200
200
150 150
100
100
50 50
0 0
10 100 1000 10000 10 100 1000 10000
ARC350MCTUR_SE_080124_1 R [Ω] ARC350MCTUR_SE_080124_2 R[ ]
U [Vp]
P [W]
1000
300
Blend 8, 9
900
Blend 6, 7
250 800
Blend 3, 4 5
700
Blend 2
200
600 Blend 0, 1
500
150
400
100 300
200
50
100
0
0 0 50 100 150 200 250 300
0 0 100 150 200 250 300
ARC350MCTUR_SE_080125_4 displayed value
ARC350MCTUR_SE_080125_3 displayed value
Setting Program
MicroCUT 7, 10,
11 (up to 30 W)
MacroCUT 6, 8, 9
P [W] P [W]
35 80
Blend 9 Blend 9
Blend 8 Blend 8
70
30
Blend 7 Blend 7
Blend 6 60 Blend 6
25
Blend 5 Blend 5
Blend 4 50
Blend 4
20
Blend 3 Blend 3
40
Blend 2 Blend 2
15
Blend 1 Blend 1
30
Blend 0 Blend 0
10
20
5
10
0 0
10 100 1000 10000 10 100 1000 10000
ARC350MPC_SE_080124_1 ARC350MPC_SE_080124_2
P [W] U [Vp]
80 700
Blend 9
Pmax Macro, Local Anestesia,
HNO Blend 8
70 600 Blend 7
Blend 6
60
500
Blend 5
50 Blend 4
400 Blend 3
40 Blend 2
Pmax Micro Programme 300 Blend 1
30 Pmax MicroCUT
Blend 0
200
20
100
10
0 0
0 10 20 30 40 50 60 70 80 0 10 20 30 40 50 60 70
ARC350MPC_SE_080124_3 displayed value ARC350MPC_SE_080124_4 displayed value
Setting Program
DryCUT 12
P [W]
P [W]
250
DryCut 200W
Blend 0-9 200
DryCut 100W
Blend 0-9 180
200
160
150 140
120
100 100
80
50
60
40
0
10 100 1000 10000 20
ARC350DC_SE_080124_1
0
0 20 40 60 80 100 120 140 160 180 200
Measurement of resistances
ARC350DC_SE_080125_2 displayed value
1600
1400
1200
1000
800
600
400
200
0
0 20 40 60 80 100 120 140 160 180 200
ARC350DC_SE_080124_3 displayed value
Setting Program
GastroCUT Pol 21
P [W]
600
500
Blend
400 1000
300 Cut Phase
1000
Coag Phase Blend 9
Coag Phase Blend 5-8
1000
200
Coag Phase Blend 4
Coag Phase Blend 3
1000
Coag Phase Blend 2
100
Coag Phase Blend 1 1000
Coag Phase Blend 0
0 1000
10000
1000
100
10
R [Ohm]
ARC350GCPol_SE_100111_1
1000
Measurement of resistances 1000
• Diagram of power output P [W] as a 1000
function of the load resistance R [Ω] for 1000
the setting “GastroCUT Pol” “Blend 0-9” • Diagram of HF output voltage U [Vp] as
a function of the “GastroCUT Pol”
“Blend 0-9” (idle mode)
P [W]
300
250
200
150
Coag Phase
Cut Phase = 520 W
100
50
0
0 1 2 3 4 5 6 7 8 9
ARC350GCPol_SE_100111_2 Blend
Setting Program
GastroCUT Pap 22
P [W] P [W]
70
600
60
500
50
400
40
Coag Phase Blend 9
Coag Phase Blend 8
300
Coag Phase Blend 7
30 Coag Phase Blend 6
Coag Phase Blend 5
Coag Phase Blend 4
200
20 Coag Phase Blend 2,3
Coag Phase Blend 0,1
10 100
0
0
10
10000
1000
100
10000
1000
100
10
ARC350GCPap_SE_10011_1 R [Ohm]
ARC350GCPap_SE_100111_3 R [Ohm]
60
Blend
50
1000
40
1000
30
Coag Phase
Cut Phase = 520 W
1000
20
1000
10
1000
0
0 1 2 3 4 5 6 7 8 9
1000
ARC350GCPap_SE_100111_2 Blend
1000
• Diagram of power output P [W] as a 1000
function of the “Blend” setting for 1000
“GastroCUT Pap” 1000
• Diagram of HF output voltage U [Vp] as
a function of the “GastroCUT Pap”
“Blend 0-9” (idle mode)
Setting Program
CardiacCUT 14, 15
P [W] P [W]
100 90
CardiacCut
90W
90 CardiacCut
80
45W
80
70
70
60 60
50 50
40
40
30
30
20
10 20
0 10
10 100 1000 10000
ARC350CCUT_SE_080124_1
0
0 10 20 30 40 50 60 70 80 90
ARC350CCUT_SE_080124_2 displayed value
Measurement of resistances
• Diagram of power output P [W] as a • Diagram of power output P [W] as a
function of the load resistance R [Ω] for function of the “CardiacCUT” setting.
the setting “CardiacCUT” = 90 W/45 W Measured load resistance = 500 Ω
U [Vp]
3000
2500
2000
1500
1000
500
0
0 10 20 30 40 50 60 70 80 90
ARC350CCUT_SE_080124_3 displayed value
Setting Program
SimCOAG 13
P [W]
P [W]
140
SimCOAG 120
P=120W
SimCOAG
120 P=60W
100
100
80
80
60
60
40
40
20
20
0
0 0 20 40 60 80 100 120
10 100 1000 10000 ARC350SCOAG_SE_080125_2 displayed value
ARC350SCOAG_SE_080125_1
Measurement of resistances
• Diagram of power output P [W] as a • Diagram of power output P [W] as a
function of the load resistance R [Ω] for function of the “SimCOAG” setting.
the setting “SimCOAG” = 120 W/60 W Measured load resistance = 500 Ω
U [Vp]
3500
3000
2500
2000
1500
1000
500
0
0 20 40 60 80 100 120
ARC350SCOAG_SE_080125_3 displayed value
Setting Program
Moderate COAG Standard 0-6, 9, 12, 14-16, 17,
11 (31 W and above),
23 (only socket 1)
24-99 (socket 1+2)
P[W]
P [W]
140
P=120W 120
P=60W
120
100
100
80
80
60 60
40
40
20
20
0
10 100 1000 10000
ARC350MBCOAG_SE_080124_1 0
0 20 40 60 80 100 120
ARC350MBCOAG_SE_080124_2 displayed value
Measurement of resistances
• Diagram of power output P [W] as a • Diagram of power output P [W] as a
function of the load resistance R [Ω] for function of the “Moderate COAG
the setting “Moderate COAG Standard” Standard” setting. Measured load
= 120 W/60 W resistance = 75 Ω
U [Vp]
200
180
160
140
120
100
80
60
40
20
0
0 20 40 60 80 100 120
ARC350MBCOAG_SE_080124_3 displayed value
Setting Program
Moderate COAG Dessication 1 0-6, 9, 12, 14-16, 17, 24-99 (socket 1+2),
23 (only socket 1)
P [W] P [W]
140 120
Desiccate 1
120W
Desiccate 1
120
60W
100
100
80
80
60 60
40
40
20
20
0
10 100 1000 10000
ARC350MCD1_SE_080124_1
0
0 20 40 60 80 100 120
ARC350MCD1_SE_080124_2 displayed value
Setting Program
Moderate COAG Dessication 2 0-6, 9, 12, 14-16, 17, 24-99 (socket 1+2),
23 (only socket 1)
P [W] P [W]
140
Desiccate 2 120
120W
120 Desiccate 2
60W 100
100
80
80
60 60
40
40
20
20
0
10 100 1000 10000
ARC350MCD2_SE_080124_1
0
0 20 40 60 80 100 120
ARC350MCD2_SE_080124_2 displayed value
Measurement of resistances
• Diagram of power output P [W] as a • Diagram of power output P [W] as a
function of the load resistance R [Ω] for function of the “Moderate COAG
the setting “Moderate COAG Dessication 2” setting. Measured load
Dessication 2” = 120 W/60 W resistance = 200 Ω
U [Vp]
400
300
200
100
0
0 20 40 60 80 100 120
ARC350MCD2_SE_080125_3 displayed value
Setting Program
Moderate COAG Dessication 3 0-6, 9, 12, 14-16, 17, 24-99 (socket 1+2),
23 (only socket 1)
P [W] P [W]
140
Dessicate 3 120
120W
120 Dessicate 3
60W
100
100
80
80
60 60
40
40
20
20
0
10 100 1000 10000
ARC350MCD3_SE_080124_2 R[ ]
0
0 20 40 60 80 100 120
ARC350MCD3_SE_080124_2 displayed value
Measurement of resistances
• Diagram of power output P [W] as a • Diagram of power output P [W] as a
function of the load resistance R [Ω] for function of the “Moderate COAG
the setting “Moderate COAG Dessication 3” setting. Measured load
Dessication 3” = 120 W/60 W resistance = 500 Ω
U [Vp]
600
500
400
300
200
100
0
0 20 40 60 80 100 120
ARC350MCD3_SE_080125_3 displayed value
Setting Program
Moderate COAG Dessication 4 0-6, 9, 12, 14-16, 17, 24-99 (socket 1+2),
23 (only socket 1)
P [W]
P [W]
140
P = 120 W 120
120 P = 60 W
100
100
80 80
60
60
40
40
20
0 20
10 100 1000 10000
ARC350MCD4_SE_091217_1 R [Ω]
0
0 20 40 60 80 100 120
1600
1400
1200
1000
800
600
400
200
0
0 20 40 60 80 100 120
ARC350MCD4_SE_091217_3 displayed value
Setting Program
Micro Moderate COAG 7, 8, 10,
11 (up to 30 W)
P [W]
P [W]
35
Micro Moderate
30
COAG 30W
30 Micro Moderate
COAG 15W
25
25
20 20
15
15
10
10
5
0 5
10 100 1000 10000
ARC350MMCOAG_SE_080124_1
0
0 5 10 15 20 25 30
180
160
140
120
100
80
60
40
20
0
0 5 10 15 20 25 30
ARC350MMCOAG_SE_080125_3 displayed value
Setting Program
Forced COAG Non-cutting, Cutting, Mixed 0-6, 9, 12, 16, 20,
11 (31 W and above),
23 (only socket 1),
24-99 (socket 1+2)
P [W]
P [W]
140 Forced Coag 120
P=120W
Forced Coag
120 P=60W
100
100
80
80
60
60
40 40
20
20
0
10 100 1000 10000
0
ARC350FCOAG_SE_080124_1 0 20 40 60 80 100 120
ARC350FCOAG_SE_080124_2 displayed value
Measurement of resistances
• Diagram of power output P [W] as a • Diagram of power output P [W] as a
function of the load resistance R [Ω] for function of the "Forced COAG Mixed,
the setting “Forced COAG Mixed, Non- Non-cutting, Cutting" setting. Measured
cutting, Cutting” = 120 W/60 W load resistance = 500 Ω
U [Vp]
U [Vp] 2500
5000
4500
2000
4000
3500
1500
3000
2500
1000
2000
1500
500
1000
500
0
0 0 20 40 60 80 100 120
0 20 40 60 80 100 120
ARC350FCC_SE_080125 displayed value
ARC350FCNC_SE_080125 displayed value
U [Vp]
3500
3000
2500
2000
1500
1000
500
0
0 20 40 60 80 100 120
ARC350FCOAG_SE_080313_3 displayed value
Setting Program
Forced COAG Micro Mixed, Micro Non-cutting, 7, 10,
Micro Cutting 11 (up to 30 W)
P [W]
35 P [W]
Micro Forced
Coag P=30W 30
30 Micro Forced
Coag P=15W
25
25
20 20
15
15
10
10
5
0 5
3500
2500
3000
2000 2500
2000
1500
1500
1000
1000
500 500
0
0 0 5 10 15 20 25 30
0 5 10 15 20 25 30 ARC350MFCOAG_SE_080124_5 displayed value
ARC350MFCOAG_SE_080124_3 displayed value
U [Vp]
1200
1000
800
600
400
200
0
0 5 10 15 20 25 30
ARC350MFCOAG_SE_080124_4 displayed value
Setting Program
GastroCUT ForcedCOAG 21, 22 (only Socket 2)
P [W]
60 P [W]
P=50W
50
P=25W
50 45
40
40
35
30 30
25
20
20
15
10
10
0
5
10 100 1000 10000
ARC350GCFC_SE_080125_1
0
0 5 10 15 20 25 30 35 40 45 50
ARC350GCFC_SE_080125_3 displayed value
Measurement of resistances
• Diagram of power output P [W] as a • Diagram of power output P [W] as a
function of the load resistance R [Ω] for function of the “GastroCUT
the setting “GastroCUT ForcedCOAG” ForcedCOAG” setting. Measured load
= 50 W/25 W resistance = 500 Ω
U [Vp]
4000
3500
3000
2500
2000
1500
1000
500
0
0 5 10 15 20 25 30 35 40 45 50
ARC350GCFC_SE_080125_2 displayed value
Setting Program
Micro Spray COAG 7, 10,
11 (up to 30 W)
P [W]
P [W]
35
Micro Spray 30
Coag P=30W
30 Micro Spray
Coag P=15W
25
25
20 20
15
15
10
10
5
5
0
10 100 1000 10000
ARC350MSCOAG_SE_080125_1
0
0 5 10 15 20 25 30
Measurement of resistances
ARC350MSCOAG_SE_080125_2 displayed value
3000
2500
2000
1500
1000
500
0
0 5 10 15 20 25 30
ARC350MSCOAG_SE_080125_3 displayed value
Setting Program
Spray COAG 0-3, 6, 9, 12-20,
4, 5 (only socket 2)
11 (31 W and above),
23 (only socket 1),
24-99 (socket 1+2)
Argon Flex Spray COAG I 21-23 (only socket 2)
Argon OPEN Spray COAG II 4, 5 (only socket 1)
P [W]
140 P [W]
Argon / Spray
Coag P=120W 120
Argon / Spray
120 Coag P=60W
100
100
80
80
60 60
40
40
20
20
0
10 100 1000 10000
ARC350ASCOAG_SE_080125_1 0
0 20 40 60 80 100 120
ARC350ASCOAG_SE_080125_2 displayed value
Measurement of resistances
• Diagram of power output P [W] as a • Diagram of power output P [W] as a
function of the load resistance R [Ω] for function of the setting “Spray COAG,
the setting “Spray COAG, Argon-Flex Argon-Flex Spray COAG I, Argon-
Spray COAG I, Argon-OPEN Spray OPEN Spray COAG II”. Measured load
COAG II und Argon Ligation” = resistance = 500 Ω
120 W/60 W
U [Vp]
6000 U [Vp]
6000
5500
5500
5000
5000
4500
4500
4000
4000
3500
3500
3000
3000
2500
2500
2000
2000
1500
1500
1000
1000
500
500
0
0 20 40 60 80 100 120
0
ARC350ASCOAG_SE_080125_3 displayed value 0 20 40 60 80 100 120
ARC350ASCOAG_SE_080125_5 displayed value
Setting Program
Spray COAG 0-3, 6, 9, 12-20,
4, 5 (only socket 2)
11 (31 W and above),
23 (only socket 1),
24-99 (socket 1+2)
Argon Flex Spray COAG I 21-23 (only socket 2)
Argon OPEN Spray COAG II 4, 5 (only socket 1)
U [Vp]
4500
4000
3500
3000
2500
2000
1500
1000
500
0
0 20 40 60 80 100 120
ARC350ASCOAG_SE_080125_6 displayed value
Setting Program
Local Anesthesia COAG 8
P [W]
50 P [W]
Local
Anesthesia 40
45 P=40W
40 35
Local
Anesthesia
35 P=20W 30
30
25
25
20
20
15 15
10
10
5
5
0
10 100 1000 10000
ARC350SCFC_SE_080124_1 0
0 5 10 15 20 25 30 35 40
ARC350SCFC_SE_080124_2 displayed value
Measurement of resistances
• Diagram of power output P [W] as a • Diagram of power output P [W] as a
function of the load resistance R [Ω] for function of the “Local Anesthesia
the setting "Local Anesthesia COAG" = COAG” setting. Measured load
40 W/20 W resistance = 500 Ω
U [Vp]
2500
2000
1500
1000
500
0
0 10 20 30 40
ARC350SCFC_SE_080124_3 displayed value
Setting Program
Local Anesthesia Spray COAG 8
P [W]
30
25
20
15
10
0
0 5 10 15 20 25 30
2500
2000
1500
1000
500
0
0 5 10 15 20 25 30
ARC350SCSC_SE_080125_3 displayed value
Setting Program
CardiacCOAG 14, 15
P [W]
100 P [W]
CardiacCoag
90W 90
90 CardiacCoag
45W
80 80
70 70
60
60
50
50
40
30 40
20 30
10
20
0
10 100 1000 10000 10
ARC350CCOAG_SE_080124_1
0
0 10 20 30 40 50 60 70 80 90
Measurement of resistances ARC350CCOAG_SE_080124_2 displayed value
U [Vp]
3000
2500
2000
1500
1000
500
0
0 10 20 30 40 50 60 70 80 90
ARC350CCOAG_SE_080124_3 displayed value
Setting Program
Bipolar CUT 0, 1, 5, 6, 9, 12-15, 17,
2 (only socket 4 for ligation),
16 (only socket 4),
11 (31 W and above),
21-99 (socket 3+4)
Bipolar Micro CUT 7, 8, 10,
11 (up to 30 W)
P [W]
P [W]
160 35
Blend 4-9
P=30W Blend 2-9
140 Blend 2, 3
P=30W Blend 2, 3
30
Blend 0, 1
P=15W Blend 0, 1
120
25
100
20
80
15
60
10
40
20 5
0 0
10 100 1000 10000 10 100 1000 10000
ARC350MBC_SE_080125_1 R [Ω] ARC350MBC_SE_080125_2 R [Ω]
120
400 Blend 0, 1
Pmax Micro Bipolar Cut
100
300
80
60 200
40
Pmax Micro Bipolar Cut 100
20
0
0 0 20 40 60 80 100 120 140
0 20 40 60 80 100 120 140 160 ARC350MBC_SE_080125_4 displayed value
ARC350MBC_SE_080125_3 displayed value
Setting Program
Bipolar Standard COAG 0, 1, 5, 6, 9, 11 (51 W and above), 12-17, 21-99
(socket 3+4), 2, 3, 4, 16, 18-20 (only socket 4)
P[W]
P [W]
140
P=120W 120
P=60W
120
100
100
80
80
60 60
40
40
20
20
0
10 100 1000 10000
ARC350MBCOAG_SE_080124_1 0
0 20 40 60 80 100 120
ARC350MBCOAG_SE_080124_2 displayed value
Measurement of resistances
• Diagram of power output P [W] as a • Diagram of power output P [W] as a
function of the load resistance R [Ω] for function of the “Bipolar Standard
the setting “Bipolar Standard COAG” = COAG” setting. Measured load
120 W/60 W resistance = 75 Ω
U [Vp]
200
180
160
140
120
100
80
60
40
20
0
0 20 40 60 80 100 120
ARC350MBCOAG_SE_080124_3 displayed value
Setting Program
Bipolar Soft COAG 0, 1, 5, 6, 9, 12-17, 21-99 (socket 3+4),
2, 3, 4, 16, 18-20 (only socket 4)
P [W]
P [W]
140
Bipolar Soft 120
COAG 120W
120 Bipolar Soft
COAG 60W
100
100
80
80
60
60
40
40
20
20
0
10 100 1000 10000
ARC350BSCOAG_SE_080125_1
0
0 20 40 60 80 100 120
U [Vp]
200
180
160
140
120
100
80
60
40
20
0
0 20 40 60 80 100 120
ARC350BSCOAG_SE_080125_2 displayed value
Setting Program
Micro Bipolar COAG 7, 8, 10
P [W] P [W]
60 50
Bipolar Micro
COAG 50W
Bipolar Micro
50 COAG 25W
40
40
30
30
20 20
10
10
0
10 100 1000 10000
ARC350BMCOAG_SE_080125_1
0
0 10 20 30 40 50
ARC350BMCOAG_SE_080125_3 displayed value
Measurement of resistances
• Diagram of power output P [W] as a • Diagram of power output P [W] as a
function of the load resistance R [Ω] for function of the “Micro Bipolar COAG”
the setting “Micro Bipolar COAG” = setting. Measured load resistance =
50 W/25 W 75 Ω
U [Vp]
200
180
160
140
120
100
80
60
40
20
0
0 10 20 30 40 50
ARC350BMCOAG_SE_080125_2 displayed value
Setting Program
Micro bipolar Neuro COAG 11 (up to 50 W)
Measurement of resistances
• Diagram of power output P [W] as a • Diagram of power output P [W] as a
function of the load resistance R [Ω] for function of the “Micro bipolar Neuro
the setting “Micro bipolar Neuro COAG” COAG” setting. Measured load
= 50 W/25 W resistance = 125 Ω
U [Vp]
200
180
160
140
120
100
80
60
40
20
0
0 10 20 30 40 50
ARC350BNCOAG_SE_080125_2 displayed value
Setting Program
Bipolar LIGATION 2, 3, 4, 16 (only socket 3)
P [W]
P [W]
300 300
250 250
200
200
150
150
100
100
50
50
0
0
10
100
1000
1 2 3 4
R [Ohm] Effekte
ARC350BipLIG_SE_100111_1 ARC350BipLIG_SE_100111_2
Setting Program
Bipolar TUR CUT 18-20 (only Socket 3)
P [W]
450 Effekt 9
Effekt 8
400
Effekt 7
350 Effekt 6
300 Effekt 5
Effekt 4
250 Effekt 3
200 Effekt 2
Effekt 1
150 Effekt 0
100
50
0
10 100 1000 10000
ARC350BTUR_SE_080125_1
Measurement of resistances
Setting Program
Bipolar TUR COAG 18-20 (only Socket 3)
P [W]
250
Effekt 7-9
Effekt 6
Effekt 5
200
Effekt 4
Effekt 3
150 Effekt 2
Effekt 1
Effekt 0
100
50
0
10 100 1000 10000
ARC350BTURC_SE_080125_1
Measurement of resistances