Luna g3 Cpap Apap User Manual
Luna g3 Cpap Apap User Manual
Luna g3 Cpap Apap User Manual
Table of Contents
1. Introduction ··························································································· 1
2. Symbols ································································································· 1
2.1 Control Buttons ············································································· 1
2.2 Device Symbols ············································································ 1
3. Warning, Caution and Important Tip ························································ 3
4. Intended Use ························································································· 3
5. Contraindications ··················································································· 4
6. Specifications ························································································ 5
7. Available Therapies ················································································ 8
8. Glossary································································································· 8
9. Model ································································································· 10
10. Package Contents ············································································· 10
11. System Features ·················································································· 11
12. First Time Setup ··················································································· 13
12.1 Placing the Device···································································· 13
12.2 Installing the Air Filter and Filter Cap / PM2.5 Filter ························ 13
12.3 Connecting to Power ································································ 14
12.4 Assembling the Tubing / Heated Tubing and Mask ······················· 15
12.5 Using Oxygen with the Device···················································· 17
12.6 Inserting the SD Card (Only for the device that equipped with SD
card) ······························································································ 18
12.7 Starting Treatment····································································· 18
13. Routine Use ························································································ 19
13.1 Connecting the Tubing······························································ 19
13.2 Adjusting the Tubing ·································································· 19
13.3 Turning on the Airflow ································································ 19
13.4 Heating the Water ···································································· 19
13.5 Using the Ramp Feature ···························································· 19
13.6 Accessing the iCode ································································· 20
13.7 Turning the Device Off ······························································· 20
14. Heated Humidifier··············································································· 20
14.1 Filling the Water Chamber ························································· 20
14.1.1 Removing the Water Chamber ········································ 20
14.1.2 Filling Water ··································································· 21
14.1.3 Returning the Water Chamber········································· 22
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
1. Introduction
Thank you for your purchase of the Luna® G3 APAP / Luna® G3 CPAP. This User
Manual will introduce you to your device. Please read it carefully. If you experience
any difficulties or problems during use, please contact your homecare provider or
physician.
2. Symbols
2.1 Control Buttons
Home Button
Knob
Operating Instructions
AC Power
DC Power
Hot Surface
Manufacturer
1 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Date of manufacture
Use-by Date
Disassembly is prohibited
Lot number
Non-Ionizing Radiation
SD Card
WEEE Marking
Air Inlet
Air Outlet
2 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
CAUTION!
Indicate the possibility of damage to the device.
IMPORTANT TIP!
Place emphasis on an operating characteristic.
Warnings, Cautions, and Important Tips appear throughout this manual as they
apply.
4. Intended Use
The Luna® G3 APAP / Luna® G3 CPAP is a CPAP (Continuous Positive Airway Pressure)
device designed for the treatment of adult Obstructive Sleep Apnea (OSA). The
integrated humidifier is indicated for the humidification and warming of air from the
flow generator device. These devices are intended for single-patient use by
prescription in the home or hospital/institutional environment on adult patients. It is
to be used on patients > 66 lbs/30 kg for whom CPAP therapy has been prescribed.
WARNINGS!
• This device is intended for adult use only.
• This device is not intended for life support.
• The instructions in this manual are not intended to supersede established medical
protocols.
• To ensure that you receive the safe, effective therapy prescribed for you, use only
3B MEDICAL accessories.
• Do not bring the device or accessories into a Magnetic Resonance (MR)
environment as it may cause unacceptable risk to the patient or damage to the
device or MR medical devices. The device and accessories have not been
evaluated for safety in an MR environment.
• Do not use the device or accessories in an environment with electromagnetic
equipment such as CT scanners, Diathermy, RFID and electromagnetic security
systems (metal detectors) as it may cause unacceptable risk to the patient or
damage to the device. Some electromagnetic sources may not be apparent, if you
notice any unexplained changes in the performance of this device, if it is making
unusual or harsh sounds, disconnect the power cord and discontinue use. Contact
your home care provider.
CAUTIONS!
• U. S. federal law restricts this device to sale by or on the order of a physician.
3 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
IMPORTANT TIP!
• Read and understand the entire user manual before operating this system. If you
have any questions concerning the use of this system, contact your home care
provider or health care professional.
5. Contraindications
Studies have shown that the following pre-existing conditions may contraindicate
the use of positive airway pressure therapy for some patients:
IMPORTANT TIPS!
• An irregular sleep schedule, alcohol consumption, obesity, sleeping pills, or
sedatives may aggravate your symptoms.
• Please use a mask which meets ISO 17510: 2015.
CAUTION!
• Contact your health care professional if symptoms of sleep apnea recur. Contact
your health care professional if you have any questions concerning your therapy.
4 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
6. Specifications
Device Size
Dimensions (L x W x H): 265 mm × 145 mm × 114 mm
Weight: 1.7 kg
Water capacity: To maximum fill line 360 mL
Heated Humidifier
Humidifier Settings: Off, Auto, 1 to 5 (95˚F to 154.4˚F / 35°C to 68°C)
Humidifier Output: No less than 15 mg H2O/L
Environmental Conditions: Maximum airflow, 35°C, 15% relative humidity
Maximum Operating Pressure: 40 hPa
Pressure Drop with Humidifier: < 0.4 hPa at 60 LPM flow
Maximum Delivered Gas Temperature: ≤ 43°C
Mode of Operation
Continuous
Work Mode
LG3600: CPAP,AutoCPAP
LG3500: CPAP
SD Card
The SD card can record patient data and fault information
AC Power Consumption
100 - 240 V ~ , 50 / 60 Hz, 2 A Max
Pressure Range
4 to 20 hPa (in 0.5 hPa increments) ≤ 30 hPa under single fault conditions.
5 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
10 ±1 ±1 ±1
20 ±1 ±1 ±1
Ramp
The ramp time ranges from 0 to 60 minutes.
Maximum Flow
When the working pressure is set to the values listed in the table, the average
flow rate at the patient end should be greater than 80% of the corresponding
flow value in the table.
Air Tubing
Air tubing Length Inner diameter
Tubing 6 ft. (1.83 m) 19 mm
Heated Tubing 6 ft. (1.83 m) 19 mm
Air filter
Type Material Average arrestance
Standard filter Polyurethane > 20% for 10 micron
Polypropylene and Poly
PM2.5 filter > 90% for 2.5 micron
(ethylene terephthalate)
6 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Cellular Module
Shock, severe vibration, and moisture should be
Transportation Requirements
avoided in transportation
Effective Radiated Power LTE LTE Cat M1/ NB1: ≤ +23 dBm (2100 mW), Class 3
FCC ID XPY2AGQN4NNN
1 The LTE bands supported by Cellular Module are defined in above, while the
following Table 1 describes the Transmitting and Receiving frequencies.
Table 1 Transmitting and Receiving frequencies
Frequency
Uplink 703 748 MHz Module transmit
range FDD
Band 28
(700 MHz) Downlink 758 803 MHz Module receive
Frequency
Uplink 777 787 MHz Module transmit
range FDD
Band 13
(700 MHz) Downlink 746 756 MHz Module receive
7 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
WARNING!
• All other wireless technology emitters must be kept at least 30 cm (12 inches) from
the Cellular Module.
CAUTION!
• Considering the requirements of network security, the cpu on this equipment only
supports the standard of our product software and does not support the operation
of other foreign software.
7. Available Therapies
The device delivers the following therapies:
AutoCPAP – Delivers CPAP therapy and provides an air pressure no less than the
prescribed one based on the patient’s needs.
8. Glossary
Apnea
A condition marked by the cessation of spontaneous breathing.
AutoCPAP
Adjust CPAP pressure automatically to improve patient comfort based on
monitoring of apnea and snoring events.
Auto Off
When this feature is enabled, the device automatically discontinues therapy
whenever the mask is removed.
8 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Auto On
With this feature, the device automatically initiates therapy when you breathe into
the mask. This feature is always enabled.
SmartC
With this feature, the device adjusts Treat P according to the patient's respiratory
event during a certain time.
SmartA
With this feature, the device adjusts Min APAP, Smart P and Max APAP according to
the patient's respiratory event during a certain time.
CPAP
Continuous Positive Airway Pressure.
iCode
A feature that is intended to give access to compliance and therapy management
information. The “iCode” consists of six separate codes displayed in the Patient
Menu, each code is a sequence of numbers. The “iCode QR” and “iCode QR+”
display two-dimensional codes.
LPM
Liters Per Minute.
OSA
Obstructive Sleep Apnea.
Patient Menu
The display mode in which you can change patient-adjustable device settings,
such as the starting pressure for the Ramp feature.
Ramp
A feature that may increase patient comfort when therapy is started. It can reduce
pressure and then gradually increase the pressure to the prescription setting so the
patient can fall asleep more comfortably.
Reslex
A therapy feature that is enabled by your home care provider to provide pressure
relief during exhalation.
Standby State
The state of the device when power is applied but the airflow is turned off.
min
Means the time unit “minute”.
h
Means the time unit “hour”.
yy mm dd / mm dd yy / dd mm yy
Denotes date.
9 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
9. Model
Product Description Maximum
Work Work
Model Product
Optional Accessory Mode Pressure
Contents (hPa)
Tubing(optional),
Device CPAP,
LG3600 Cellular Module
(3.5-inch LCD) AutoCPAP
(optional),
20
Heated Tubing
Device
LG3500 (optional) CPAP
(3.5-inch LCD)
PM2.5 Filter (optional)
All parts and accessories are not made with natural rubber latex.
The service life of the device is five years if the use, maintenance, cleaning and
disinfection are in strict accordance with the User Manual. The shelf life of the
device is ten years.
The service life of the Tubing and the Heated Tubing is six months. The shelf life of the
Tubing and the Heated Tubing is 3 years.
WARNINGS!
• The device should only be used with the mask and accessories manufactured or
recommended by 3B MEDICAL or with those recommended by your prescribing
physician. The use of inappropriate masks and accessories may affect the
performance of the device and impair the effectiveness of therapy.
• Do not pile up the long tubing at the head of the bed, as it may wrap around the
10 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
IMPORTANT TIPS!
• If any of the above parts are missing, contact your home care provider.
• Contact your home care provider for additional information on the available
accessories of this device. When using optional accessories, always follow the
instructions enclosed with the accessories.
Display Screen
Home Button
Knob
Fig. 10-1
Name Function
Start / Stop Button Start / Stop delivering air
Display menus for operation, messages, monitoring
Display Screen
data, etc.
Home Button Return to the previous menu or main interface
Knob Adjust device settings
11 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Air Outlet
DC Inlet
Fig. 10-2
Name Function
Air Outlet Deliver pressurized air; connect to the tubing
Heated Tubing Port Connected to the plug of the heated tubing
DC Inlet An inlet for the DC power supply
Cover and secure the air filter, which is used to filter dust
Filter Cap (Air Inlet)
and pollen in the air entering the device
Transfer Box
SD Card Slot
Fig. 10-3
Name Function
Transfer Box For the connection of the device to the water chamber
SD Card Slot Insert the SD card into this slot
CAUTION!
• The pictures in this manual are only for reference, if they are different from the
material object, the latter shall prevail.
12 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
WARNINGS!
• If the device has been dropped or mishandled, if the enclosure is broken, or if
water has entered the enclosure, disconnect the power cord and discontinue use.
Contact your home care provider immediately.
• If the room temperature is warmer than 95°F (35°C), the airflow produced by the
device may exceed 109.4°F (43°C). The room temperature must be kept below 95°F
(35°C) while the patient uses the device.
CAUTIONS!
• Always ensure that the device is placed in an area where the screen and
indicators are clearly visible.
• If the device has been exposed to either very hot or very cold temperatures, allow
it to adjust to room temperature (approximately 2 hours) before beginning setup.
• Make sure the device is away from any heating or cooling equipment (e.g.,
forced air vents, radiators, air conditioners).
• The device is not suitable for use in high humidity environments. Make sure that no
water enters the device.
• Make sure that bedding, curtains, or other items are not blocking the filter or vents
of the device.
• Keep pets, pests or children away from the device and avoid small objects being
inhaled or swallowed.
• To avoid explosion, this device must not be used in the presence of flammable
gases (e.g. anesthetics).
• Tobacco smoke may cause tar build-up within the device, leading to the
malfunctioning of the device.
• Air must flow freely around the device for it to work properly.
12.2 Installing the Air Filter and Filter Cap / PM2.5 Filter
(1) Attach the air filter to the filter cap, as shown in Fig. 12-1.
Fig. 12-1
13 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
(2) Install the filter cap containing the air filter to the device, as shown in Fig. 12-2.
Fig. 12-2
(3) Change the air filter and filter cap to the PM2.5 filter, as shown in Fig. 12-3.
PM2.5 Filter
Fig. 12-3
WARNINGS!
• Do not block the gas INTAKE PORT, thereby interfering with the therapy.
• Nebulisation or humidification can increase the resistance of breathing system
filters and the operator must monitor the breathing system filter frequently for
increased resistance and blockage to ensure the delivery of the therapeutic
pressure.
CAUTIONS!
• The air filter or the PM2.5 filter must be in place when the device is operating.
• Device must be unplugged when installing the air filter and filter cap or PM2.5
filter.
14 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
DC Inlet
Power Adapter
Power Cord
Fig. 12-4
Note: The length of the power cord and power adapter is 1.5 m and 1.8 m
respectively without the function of preventing electromagnetic interference.
WARNINGS!
• The device is powered on for use when the power cord and power adapter is
connected. The Knob turns the blower On / Off.
• Use of the device at an AC voltage beyond the stated range (see Section 5 “AC
Power Consumption”) may damage the device or cause device failure.
• Connect to appropriate power for proper operation of the device.
• Inspect the power cord often for any signs of damage. Replace a damaged cord
immediately.
IMPORTANT TIPS!
• After interruption and restoration of the power supply, the device will restore its
pre-interruption working status automatically.
• To remove AC power, disconnect the power cord from the power outlet.
Air Outlet
Tubing
Fig. 12-5
15 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
(2) Connect the heated tubing joint to the air outlet of the device, and then insert
the power plug into the heated tubing port on the back of the device, as shown in
Fig. 12-6.
Air Outlet Heated Tubing
Power Plug
Heated Tubing
Fig. 12-6
CAUTION!
• As the ambient temperature decreases, the humidifier can be reduced to
improve or avoid condensation in the pipeline. If the ambient temperature is too
low, in order to avoid condensation, it is recommended to use the heated tubing.
If the heated tubing is connected correctly, the icon will become a number in
the Main Interface on the screen of the device, as shown in Fig. 12-7.
Fig. 12-7
Turn the Knob to turn on or turn off the heated tubing and to adjust the heat
level according to instructions of the Patient Menu of the device.
There are five heat levels available, and the number of heat level will appear in the
Main Interface on the screen of the device. The number 3 next to the icon
indicating the heat is adjusted to Level 3, as shown in Fig. 12-8.
16 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Fig. 12-8
(3) Connect the other end of the tubing to the mask according to the user manual
for the mask. Wear the mask.
WARNINGS!
• If you are using a mask with a built-in exhalation port, connect the mask’s
connector to the tubing.
• If you are using a mask with a separate exhalation port, connect the tubing to the
exhalation port. Position the exhalation port so that the vented air is blowing away
from your face. Connect the mask’s connector to the exhalation port.
• If you are using a full-face mask (a mask covering both your mouth and nose), the
mask must be equipped with a safety (entrainment) valve.
• In order to minimize the risk of CO2 rebreathing, the patient should observe the
following instructions:
- Use only tubing and mask provided by 3B MEDICAL.
- Do not wear the mask for more than a few minutes while the device is not
operating.
- Use only masks with vent holes. Do not block or try to seal the vent holes in the
exhalation port.
WARNINGS!
• Connect the oxygen tube to the oxygen inlet of the mask.
• The oxygen supply must comply with the local regulations for medical oxygen.
• Turn on the device before turning on the oxygen. Turn off the oxygen before
turning off the device. Explanation of Warning: When the device is turned off, but
the oxygen flow still exists, oxygen may accumulate within the device's enclosure
and pose a fire hazard. Turning off the oxygen before turning off the device will
prevent oxygen accumulation in the device and reduce the risk of fire. This warning
applies to CPAP devices.
• Oxygen supports combustion. Keep the device and the oxygen container away
17 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
from heat, open flames, any oily substances, or other sources of ignition. DO NOT
smoke in the area near Luna® G3 APAP / Luna® G3 CPAP or the oxygen container.
• Sources of oxygen should be located more than 1 m from the device.
• When using oxygen with this system, a Pressure Valve must be placed in-line with
the patient circuit between the device and the oxygen source. The pressure valve
helps prevent the backflow of oxygen from the patient circuit into the device when
the unit is off. Failure to use the pressure valve could result in a fire hazard.
• Do not connect the device to an unregulated or high pressure oxygen source. The
pressure of oxygen source does not exceed the work pressure of the device.
SD Card Slot
SD Card
Fig. 12-9
CAUTIONS!
• If the SD card is not inserted, there will not be a symbol appear in the Main
Interface on the screen of the device.
• To avoid data loss or any damage to the SD card, the SD card can only be
removed after the device stops delivering air.
WARNINGS!
• Be sure to follow your physician’s instructions on adjusting the settings! To order any
accessories not included with this device, contact your equipment supplier.
• DO NOT connect any ancillary equipment to this device unless recommended by
3B MEDICAL or your physician. If you suffer from chest discomfort, shortness of breath,
18 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
stomach bloating, or severe headache when using the device, contact your
physician or qualified medical personnel immediately.
CAUTION!
• Before each use, examine the tubing for any damage or debris. If necessary,
clean the tubing to remove the debris. Replace any damaged tubing. Make sure
that the mask does not leak.
CAUTION!
• Observe the water level of the water chamber before using the humidifier. Make
sure there is sufficient water in the water chamber, and avoid heating the device
with an empty water chamber.
CAUTIONS!
• You can use the ramp feature as often as you wish during sleep.
• The ramp feature is not prescribed for all users.
19 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Fig. 13-1
Fig. 13-2
CAUTION!
• Do not position the device where it is difficult to disconnect the device.
20 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Fig. 14-1
WARNING!
• Turn the device off and allow approximately 15 minutes for the heater plate and
water to cool.
Fig. 14-2
Fig. 14-3
(2) Open the cap, and fill the water chamber with approximately 360 ml of water,
as shown in Fig. 14-4. Make sure that the water does not exceed the maximum
water level line.
21 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Fig. 14-4
WARNING!
• Change water before every use and do not surpass the maximum water level line.
CAUTIONS!
• Empty the water chamber when the heated humidifier is not in use.
• Distilled water is recommended.
IMPORTANT TIP!
• It is not necessary that remove the water chamber from the device. The users can
open the cap of the water chamber directly to fill the water.
Fig. 14-5
Fig. 14-6
22 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
WARNING!
• For safety purposes, the device must be placed on a flat surface at a level lower
than the patient's head on a bed, so that the condensation flows back to the water
chamber rather than remain in the tubing causing rainout.
CAUTIONS!
• Avoid moving or tilting the device when the water chamber has water in it.
• Take precautions to protect furniture from water damage.
Fig. 14-7
CAUTION!
• Empty and air-dry the water chamber when the device is not in use.
There are five humidity levels available, and the number of humidity level will
appear in the Main Interface on the screen of the device. The number 2 next to the
icon indicating the humidity is adjusted to Level 2, as shown in Fig. 14-8. The
temperature of the water in the water chamber maintains a constant set level.
Fig. 14-8
23 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
WARNING!
• Do not touch the heater plate of the device when it is working, otherwise you may
get burned. Turn off the heat when the heated humidifier is not in use.
CAUTIONS!
• Generally speaking, the humidity inside the mask is low when the water
temperature is low.
• The greater the difference between the temperature inside the air tubing and
room temperature is, the more easily condensation occurs inside the tubing.
• If there are only a few condensed water droplets inside the tubing in the morning
after therapy, it means that the humidity level is appropriate; if there is lots of
condensed water droplets inside the tubing and / or mask, it means that the
humidity level is too high and should be set lower; Nasal dryness means that the
humidity level is too low and should be set higher.
(1) Insert the Cellular Module into the device, and turn on the device. The device
screen displays the Main Screen shown in Fig. 15-1.
Fig. 15-1
(2) The Cellular Module starts searching for signals in a few seconds. Once a signal is
found, the module will automatically connect to it, and a signal icon will appear in
the status bar at the top of the device screen.
24 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Note:
(1) When the signal is weak, data transmission may become slow and even stop.
(2) The Cellular Module will keep searching for GPRS signals until one is found.
If the signal is strong, the signal icon appears in the Main Screen, as shown in Fig.
15-2 (the signal icons of different strength appear in a similar way).
Fig. 15-2
The device screen will not show the signal icon, if the Cellular Module is connected
to the device improperly or if the Module is not working properly.
WARNING!
• To ensure successful data transmission through the Cellular Module, computers,
televisions, radios or similar devices should not be placed near the Cellular Module.
25 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Fig. 16-1
Note: The above interface only applicable to the device does not activate the
SmartC or SmartA. If the SmartC or SmartA is enabled, and the symbol will
appear in the status bar at the top of the screen, as shown in Fig. 16-2.
Fig. 16-2
The first icon on the upper part of the screen indicates the Preheat Function Icon,
the second icon indicates the Accessorie, and the third icon indicates Mask
Setup Icon, the fourth icon indicates the Report Interface Icon, the fifth icon
indicates the Initial Setup Icon. As you turn the Knob , the cursor switches among
the five icons, and the interface displayed on the screen changes accordingly.
Note: As the humidity levels is off, the Preheat Function Icon will become gray, as
shown in Fig. 16-2.
Fig. 16-3
26 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Note: The Heated Tubing option can only be adjusted when the device is
connected to the heated tubing, as shown in Fig. 16-4.
Fig. 16-4
Fig. 16-5
Fig. 16-6
27 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Fig. 16-7
Fig. 16-8
28 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
29 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
00:00
Time — Set time by adjusting this option.
23:59
12-hour / Turn the Knob to choose between the two
Time Format
24-hour time formats.
30 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
17. Alert
Alert Message Description
An audible alert will sound in 6 s if the device is accidentally
disconnected from power when it is delivering air.
Note:
Power Failure!!!
(1) The alert will not sound if power failure occurs when the
device is in standby state.
(2) No alert message on the screen during a power failure.
An audible alert will sound if no airflow comes out of the
Device Fault!!!
machine; the screen will display “Device Fault!!!”.
When the airflow is on, an audible alert will sound if the air
Leak!!
leak rate is excessive; the screen will display “Leak!!”.
If the voltage supplied by power adaptor is lower than 22V,
Low Input
an audible alert will sound and the screen will display “Low
Voltage!!
Input Voltage!!”.
When humidifier is applied, an audible alert will sound when
Humidifier
the humidifier fails to work; the screen will display “Humidifier
Failure!!
Failure!!”.
When the Filter reminder feature is enabled, an audible alert
Please Change will sound if the preset replacement time reaches but without
Filter! replacing the air filter; the screen will display “Please Change
Filter!”.
When the tubing reminder feature is enabled, an audible
Please Change alert will sound if the preset replacement time reaches but
Tube! without replacing the tubing; the screen will display “Please
Change Tube!”.
When the Mask reminder feature is enabled, an audible alert
Please Change will sound if the preset replacement time reaches but without
Mask! replacing the mask; the screen will display “Please Change
Mask!”.
The screen will display “SD Card Full!” if the SD card has
SD Card Full!
reached its maximum capacity.
The screen will display “Reinsert SD card!” if the SD card fails
Reinsert SD card!
to work.
31 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
18. Cleaning
WARNINGS!
• Regular cleaning of the device and its accessories is very important for the
prevention of respiratory infections.
• To avoid electric shock, always unplug the device before cleaning.
• Use mild soap that is nontoxic to humans.
• Follow the manufacturer's instructions on cleaning the mask and tubing and on
determining the frequency of cleaning.
• Before cleaning, check whether the device has been disconnected from the
power supply, whether the power cord has been unplugged, and whether the
water chamber of the device has cooled down. Make sure the plate has cooled
down to room temperature, to avoid the risk of burns.
• Do not open or modify the device. There are no user serviceable parts inside.
Repairs and servicing should only be performed by an authorized service agent.
CAUTIONS!
• Overheating of the materials could lead to early fatigue of these materials.
• Do not use solutions containing chlorinated lime, chlorine, or aromatic to clean
the device and its accessories. Liquid soap containing moisturizing agents or
antimicrobials should not be used either. These solutions may harden cleaned
materials or reduce their lifespan.
• Do not clean or dry the device and its accessories when the temperature is higher
than 80°C (176°F). High temperatures could reduce product life.
• Do not immerse the device in any fluids.
Fig. 18-1
(2) Cleaning the Water Chamber: You may also clean the water chamber with a
soft cloth which does not scratch the water chamber (dip the soft cloth in liquid
soap if necessary), rinse it thoroughly, and then wipe it dry with a soft cloth.
32 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
WARNINGS!
• Emptying and cleaning the water chamber daily will help prevent mold and
bacteria growth.
• Allow the water in the chamber to cool down to room temperature before
removing it from the device.
CAUTIONS!
• Clean the water chamber only after the water in it cools. Make sure that no water
enters the device.
• After cleaning, rinse the water chamber throughly in clean water to make sure
that no soap residue is left; then wipe it dry with a lint-free cloth, so as to prevent
calcareous accumulations.
• Inspect the water chamber for any leak or damage. Replace the water chamber
if any damage is present.
• It is recommended to clean the water chamber and change the water daily.
Transfer Box
Fig. 18-2
(2) Cleaning the Transfer Box: Rinse the transfer box throughly in clean water. You
may also clean the transfer box with a soft cloth which does not scratch it (dip the
soft cloth in liquid soap if necessary), rinse it thoroughly, and then wipe it dry with a
soft cloth.
Transfer Box
Fig. 18-3
CAUTION!
• It is recommended to clean the transfer box once a week.
33 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
CAUTIONS!
• The device can only be used after the enclosure is dry, so that no moisture enters
the device.
• It is recommended to clean the enclosure once a week.
Fig. 18-4
Fig. 18-5
(3) Hold both the cuff of the air tubing and the swivel of the mask, then gently pull
apart as shown in Fig. 18-6.
Fig. 18-6
34 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
(4) Clean the components with a soft bristled brush for one minute while soaking in
detergent solution (see the table below). Pay particular attention to all crevices and
cavities.
Heated
Detergent Water temperature Tubing
Tubing
AlconoxTM Warm water
√ √
(diluted at 1%) (approx 113 to 140°F or 45 to 60°C)
(5) Run the detergent solution through the air tubing repeatedly until no
contamination is visible.
(6) Thoroughly rinse each component according to the detergent manufacturer's
instructions.
(7) Thoroughly rinse the tubing in drinking quality water (five liters per assembly) by
immersing it completely for a minimum of one minute in duration.
(8) Repeat the rinse procedure two additional times using fresh water for a total of
three rinses.
(9) Air dry out of direct sunlight and/or heat.
(10) Inspecting
Perform a visual inspection of the components. If any visible deterioration is
apparent (holes, tears or cracks etc.), the components should be discarded and
replaced. Slight discoloration may occur and is acceptable.
WARNINGS!
• Please wash by hand.
•The tubing should be cleaned daily.
• If the heated tubing is damaged (such as broken hole, tear, exposed heated wire,
etc.) or poor function, please do not repair and use it by yourself and replace it
immediately.
• Failure to clean in accordance with the Manual may result in reduced
performance of the heated tubing or reduced product life.
• After cleaning and prior to reuse, the Breathing Tubing should be inspected for
holes, creases and tears.
Fig. 18-7
35 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
(2) Install the filter cap containing the air filter to the device, as shown in Fig. 18-8.
Fig. 18-8
(3) Disassemble the PM2.5 Filter from the device, as shown in Fig. 18-9, then change
a new one.
PM2.5 Filter
Fig. 18-9
CAUTIONS!
• To avoid material damage, do not place the spare air filter / PM2.5 Filter in direct
sunlight, humid environments, or temperatures below the freezing point. The air filter
/ PM2.5 Filter should be replaced every 6 months (It may be replaced more
frequently based on actual sanitary conditions).
• Operating the device with a dirty air filter may stop it from working properly and
may cause damage to the device.
36 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
(4) Security Stations: For convenience at security stations, there is a note on the
bottom of the device stating that it is medical equipment. It may be helpful to bring
this manual along with you to help security personnel understand the device.
CAUTIONS!
• Empty the water chamber before packing the device for your trip; in order to
prevent any remaining water from entering the device.
• Using the device at an incorrect elevation setting could result in airflow pressures
higher than the prescribed setting. Always verify the elevation setting when
traveling or relocating.
• If the device is used when the atmospheric pressure is out of the stated range (See
Section 6), the accuracy of the leakage alert will be affected.
21. Reordering
Contact your home care provider to order accessories or replacement filters.
The device does not require routine servicing.
WARNINGS!
• If you notice any unexplained changes in the performance of the device, if it is
making unusual or harsh sounds, if it has been dropped or mishandled, if the
enclosure is broken, or if water has entered the enclosure, discontinue use. Contact
your home care provider.
• If the device malfunctions, contact your home care provider immediately. Never
attempt to open the enclosure of the device. Repairs and adjustments must be
performed by 3B MEDICAL -authorized service personnel only. Unauthorized service
could cause injury, invalidate the warranty, or result in costly damage.
• If necessary, contact your local authorized dealer or 3B Medical, Inc., for
technical support and documents.
37 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
23. Disposal
Electrical product components contain chemical substance which may pollute
environment, when the device reaches the end of its service life, dispose of the
device and packaging in accordance with local laws and regulations.
24. Troubleshooting
The table below lists common problems you may have with the device and possible
solutions to those problems. If none of the corrective actions solve the problem,
contact your home care provider.
38 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
39 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
40 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Acceptable latency
As the user information is not viewed by the doctor in real time, sometimes it can be
delayed for 24 or more hours.
41 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
42 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
43 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Note 1: At 80 MHz and 800 MHz, the higher frequency range applied.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular /
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength
in the location in which the device is used exceeds the applicable RF
compliance level above, the device should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the device.
b Over the frequency range 150 kHz to 80 MHz, the field strengths should be less
than 10 V/m.
44 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
45 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
Note: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
46 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
WARNINGS!
• The device should not be used in the vicinity of other electronic equipment such
as diathermy, electrocautery and radio frequency identification (RFID), security
systems (such as electromagnetic anti-theft systems and metal detectors), cell
phone, transceiver or radio control products. If you have to do so, the device should
be observed to verify normal operation.
• Use of the device adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are
operating normally.
• Use of accessories, transducers and cables other than those specified or provided
by the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in
improper operation.
• Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the Luna® G3 APAP / Luna® G3 CPAP, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could
result.
• The device may be interfered with by other equipment, even if that other
equipment complies with CISPR EMISSION requirements.
• The device may be interfered by the electromagnetic field of some known or
unknown radio frequency transmitters in the environment during use. If interference
occurs, please stay away from the interfered electromagnetic environment, or find
and turn off the electromagnetic field interference source before continuing to use
it.
• When the device is exposed to soldering, electrosurgery, defibrillation, X-ray (γ
ray), infrared radiation, transient electromagnetic field, including nuclear magnetic
resonance (MRI) and radio interference environment, the product may be
damaged.
• During operation of the device, due to electrostatic interference, the following
phenomena may occur: (1) Temporary loss of function or degradation of
performance, such as abnormal screen display, etc. The device will recover to
normal after being restarted; (2) Automatic restart of the device. These phenomena
will not affect the normal use of the device, and will not cause permanent
performance degradation or function loss of the device.
47 / 48
Luna® G3 APAP / Luna® G3 CPAP User Manual V1.0
3B MEDICAL, Inc. DISCLAIMS ALL LIABILITY FOR ECONOMIC LOSS, LOSS OF PROFITS,
OVERHEAD OR CONSEQUENTIAL DAMAGES WHICH MAY BE CLAIMED TO ARISE
FROM ANY SALE OR USE OF THIS PRODUCT. SOME STATES DO NOT ALLOW THE
EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGES, SO THE
ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU.
To exercise the rights under this warranty, contact the local authorized dealers or:
3B Medical, Inc.
203 Avenue A NW, Suite 300,
Winter Haven, FL 33881
T: (863) 226-6285
F: (863) 226-6284
Contract Manufacturer:
BMC (Tianjin) Medical Co., Ltd.
2/F North Area and 3/F, Building No.4, No.1 Xinxing Road, Wuqing District, (301700)
Tianjin, P.R.China
Tel: +86-22-82939881
48 / 48