Elma Metrel+MI3200 Manual UK
Elma Metrel+MI3200 Manual UK
Elma Metrel+MI3200 Manual UK
MI 3200
User Manual
Version 4.2; Code No. 20 751 097
Distributor:
Producer:
METREL d.d.
Ljubljanska 77
SI-1354 Horjul
Mark on your equipment certifies that this equipment meets the requirements of the EU
(European Union) concerning safety and interference causing equipment regulations
© 2019 Metrel
No part of this publication may be reproduced or utilized in any form or by any means
without permission in writing from METREL.
2
MI 3200 TeraOhm 10 kV Table of contents
Table of contents
3
MI 3200 TeraOhm 10 kV General introduction
1. General introduction
1.1. Features
The TeraOhm 10 kV Tester is a portable battery / mains powered test instrument
intended for the testing of Insulation Resistance by using high test voltages of up to
10kV.
The instrument is designed and produced with the extensive knowledge and experience
acquired through many years of of working in this sector.
A dot matrix LCD offers easy-to-read results and all associated parameters. The
operation is straightforward and clear to enable the user to operate the instrument
without the need for special training (except reading and understanding this Users
Manual).
Test results can be stored on the instrument. The new professional PC SW enables
straightforward transfer of test results and other parameters in both directions between
the test instrument and PC.
4
MI 3200 TeraOhm 10 kV Instrument Description
2. Instrument Description
2.1. Instrument Casing
The instrument is housed in a plastic box that maintains the protection class defined in
the general specifications.
5
MI 3200 TeraOhm 10 kV Instrument Description
Legend:
2.3. Accessories
The accessories consist of standard and optional accessories. Optional accessories can
be delivered upon request. See attached list for standard configuration and options or
contact your distributor or see the METREL home page: http://www.metrel.si.
6
MI 3200 TeraOhm 10 kV Instrument Description
2.4.1. High voltage shielded test lead with High voltage tip
Application notes:
This test lead is designed for hand held testing of insulation.
Insulation ratings:
- High voltage tip (red): 10kV d.c (double insulation);
- High voltage banana connector (red): 5 kV d.c (double
insulation), 10 kV d.c. (basic insulation);
- Guard banana connector (green): 600V CAT IV (double
insulation);
- Cable (yellow): 12kV (shielded).
2.4.2. High voltage shielded test leads with High voltage alligator clips.
Application notes:
These test leads is designed for diagnostic testing of
insulation.
Insulation ratings:
- High voltage banana connector (red, black): 5kV d.c
(double insulation), 10 kV d.c. (basic insulation);
- Alligator clibs (red, black): 5kV d.c (double insulation),
10 kV d.c. (basic insulation);
- Guard banana connector (green): 600V CAT IV (double
insulation);
- Cable (yellow): 12kV (shielded).
Insulation ratings:
- Guard test lead with banana connectors (green): 600V CAT IV (double insulation);
- Alligator (green): 600V CAT IV (double insulation).
7
MI 3200 TeraOhm 10 kV Warnings
The symbol on the test equipment means »Read the Instruction manual
with special care for safe operation«. The symbol requires an action!
If the test equipment is used in a manner not specified in this Instruction manual,
the protection provided by the equipment could be impaired!
Read this Instruction manual carefully, otherwise the use of the test equipment
may be dangerous for the operator, the test equipment itself or for the tested
object!
Do not use the test equipment or any of the accessories if any damage is noticed!
Consider all generally known precautions in order to avoid risk of electric shock
while dealing with hazardous voltages!
Do not connect the test equipment to a mains voltage different from the one
defined on the label adjacent to the mains connector, otherwise it may be
damaged.
Service intervention or adjustment is only allowed to be carried out by competent
authorized personnel!
Hazardous voltages exist inside the test equipment. Disconnect all test leads,
remove the power supply cable and switch off the test equipment before opening
the battery compartment.
All normal safety precautions must be taken in order to avoid risk of electric shock
while working on electrical installations!
Do not use the equipment in a wet environment, around explosive gas, vapour or
dust.
Only adequately trained and competent persons may operate the equipment.
8
MI 3200 TeraOhm 10 kV Warnings
Always connect accessories to the test equipment and to the test object before
starting high voltage measurement. Do not touch test leads or crocodile clips
during measurement.
Only hand-held part of high voltage test lead with tip is allowed to be touched
(held) during measurement.
Do not touch any conductive parts of equipment under test during the test, risk of
electric shock!
Make sure that the tested object is disconnected (mains voltage disconnected)
and
de-energized before connecting the test leads and starting the measurement
(except Voltage measurement)!
In case of a capacitive test object, automatic discharge of the object may not be
done immediately after finishing the measurement.
Do not connect test terminals to an external voltage higher than 600 V DC or AC
(CAT IV environment) to prevent any damage to the test equipment!
In rare cases (internal fault) the test equipment can behave in an uncontrolled
manner (LCD blinking, freezing, not responding to keys, etc.). In this case
consider the test equipment and the test object as hazardous live, and perform all
safety measures to turn off (reset) the test equipment and to discharge the test
object manually!
Note:
For manual discharge Metrel recommends the use of A 1513 Discharge link.
The internal discharge resistors of A 1513 ensure a damped discharge up to 10
uF at 10 kV.
Disconnect all test leads, main supply cable and switch the power off before
opening the Battery cover!
9
MI 3200 TeraOhm 10 kV Performing measurements
4. Performing measurements
4. 1. Switching on the instrument
Autocalibration
The instrument is switched ON by pressing the ON/OFF key. After turning on, the
instrument will execute the autocalibration (Fig. 3).
Note:
If batteries are defective or missing and the instrument is powered from mains supply,
the instrument will not turn ON
Auto-calibration prevents the reduction in accuracy when measuring very low currents. It
compensates the effects caused by ageing, temperature and humidity changes etc.
A new auto-calibration is recommended when the temperature changes by more than
5C.
Note:
If the instrument detects an incorrect state during the autocalibration, the following
warning message will be displayed:
ERROR!
-TEST LEADS CONNECTED:
DISCONNECT AND SWITCH ON THE INSTRUMENT AGAIN
Possible reasons for out of range conditions are excessive humidity, excessively high
temperature, etc. In this case it is possible to perform measurements by pressing the
START/STOP button again but results could be out of technical specification.
10
MI 3200 TeraOhm 10 kV Performing measurements
Note: If batteries are defective or missing, the charger will not work. In button left corner
of LCD screen only plug character will be appeared.
If the instrument is connected to the mains supply when the instrument is turn ON, the
instrument will automatically switch from the battery supply to the main supply. In button
left corner of the LCD screen, the plug character will appear. If instrument is not in
measuring mode*, the internal charger will begin to charge the batteries. In button left
corner of LCD screen battery indicator will start to flash, indicating that the batteries are
charging.
Off function
The instrument can be switched OFF only by pressing the ON/OFF key. The auto-off
function is not available to allow long-term measurements to be performed..
4.2.1 Configuration
The configuration function enables the selection and adjustment of parameters (Table
1a.) of the measurements. In addition memorized results can be cleared in this menu.
(Fig. 5a).
In the lower section of the display the power supply status is shown.
The following procedure must be carried out when adjusting some of the configuration
parameters:
11
MI 3200 TeraOhm 10 kV Performing measurements
4.2.2 Setup
The setup function enables the selection and adjustment of general parameters (Table
1b.) of the instrument (Fig. 5b).
In the lower section of the display the power supply status is shown.
The following procedure must be carried out when adjusting some of the configuration
parameters:
12
MI 3200 TeraOhm 10 kV Performing measurements
13
MI 3200 TeraOhm 10 kV Measurements
5. Measurements
5.1. Generally Information about High DC voltage
testing
The purpose of insulation tests
Insulating materials are important parts of almost every electrical product. The material’s
properties depend not only on its compound characteristics but also on temperature,
pollution, moisture, ageing, electrical and mechanical stress, etc. Safety and operational
reliability require the regular maintenance and testing of the insulation material to ensure
it is kept in good operational condition. High voltage tests are used to test insulating
materials.
DC vs. AC testing voltage
Testing with a DC voltage is widely accepted as being useful as testing with AC and / or
pulsed voltages. DC voltages can be used for breakdown tests especially where high
capacitive leakage currents interfere with measurements using AC or pulsed voltages.
DC is mostly used for insulation resistance measurement tests. In this type of test, the
voltage is defined by the appropriate product application group. This voltage is lower
than the voltage used in the withstanding voltage test so the tests can be applied more
frequently without stressing the test material.
Itest
surface
+
material
Riss1
Cpi Itest
Riso Ciso IPI
Guard Rpi
Riss2 ICiso
IRiso IRiss
-
Fig. 6 Fig. 7
14
MI 3200 TeraOhm 10 kV Measurements
Riss1 and Riss2 - the surface resistivity (position of optional guard connection)
Riso – the actual insulation resistance of material
Ciso – capacitance of material
Cpi, Rpi - represents polarization effects.
The Fig. 7 shows typical currents for that circuit.
Itest = overall test current (Itest= IPI+ IRISO+ IRISS)
IPI = polarization absorption current
IRISO = actual insulation current
IRISS = surface leakage current
15
MI 3200 TeraOhm 10 kV Measurements
After applying a high voltage to an insulator the electric dipoles distributed in the
insulator align themselves with the applied electrical field. This phenomenon is called
polarization. As the molecules polarize, a polarization (absorption) current lowers the
overall insulation resistance of the material.
The absorption current (IPI) typically collapses after a few minutes. If the overall
resistance of the material does not increase, this means that other currents (e.g. surface
leakages) dominate the overall insulation resistance.
- PI is defined as the ratio of the measured resistances in two time slots. The most
typical ratio is 10 min value to 1 min value but this is not a rule.
- The test is typically performed at the same voltage as the insulation resistance test.
- If the one-minute insulation resistance is greater than 5000 M, then this
measurement may not be valid (new modern types of insulation).
- Oiled paper used in transformers or motors is a typical insulation material that
requires this test.
In general, insulators that are in good condition will show a “high” polarization index
while insulators that are damaged will not. Note that this rule is not always valid. Refer to
Metrel’s handbook Insulation Testing Techniques for more information.
Example for minimum acceptable values for motor insulation (IEEE 43):
Class A =1.5, Class B = 2.0, Class F =2.0, Class H =2.0.
DIELECTRIC DISCHARGE
An additional effect of polarization is the recovered charge (from Cpi) after the regular
discharging of a completed test. This can also be a supplementary measurement for
evaluation of the quality of insulating material. This effect is generally found in insulating
systems with large capacitance Ciso.
The polarisation effect (described in “Polarisation Index”) causes a capacitance to form
(Cpi). Ideally this charge would dissipate immediately a voltage was removed from the
material. In practice, this is not the case.
In conjunction with the polarisation index (PI), Dialectic Discharge (DD) is another way to
check the quality and suitability of a insulation material. A material that discharges
quickly would provide a low value while a material that takes a long time to discharge will
provide a higher value (described in the table below, for more information see section
5.6).
16
MI 3200 TeraOhm 10 kV Measurements
When testing outside the reference ambient conditions, the quality of the insulation
resistance measurements can be affected by humidity. Humidity adds leakage paths
onto the surface of the complete measuring system, (i.e. the insulator under test, the test
leads, the measuring instrument etc). The influence of humidity reduces accuracy
especially when testing very high resistances (i.e. Tera ohms). The worst conditions
arise in environments containing high condensation, which can also reduce safety. In the
case of high humidity, it is recommended to ventilate the test areas before and during
the measurements. In the case of condensed humidity the measuring system must dry
and it can take several hours or even few days to recover.
17
MI 3200 TeraOhm 10 kV Measurements
IL
IM IM
IL
IA IA IL
+OUT -OUT +OUT -OUT GUARD
Ut A Ut A
where:
Ut ........ Test voltage
IL ......... Leakage current (resulted by surface dirt and moisture)
IM ......... Material current (resulted by material conditions)
IA ......... A-meter current
Result without using GUARD terminal: RINS = Ut / IA = Ut / (IM + IL) …incorrect result.
Note:
The guard terminal is protected by an internal impedance (400 K).
The instrument has two guard terminals to allow easy connection of shielded
measuring leads.
18
MI 3200 TeraOhm 10 kV Measurements
Note:
Any of the selected filter options increases the settling time with Fil1 to 60 s, Fil2 to 70 s,
and Fil3 to 120 s.
- It is necessary to pay close attention to the selection of time intervals when using the
filters.
- The recommended minimum measuring times when using filters are the settling
times of the selected filter option.
Example:
A noise current of 1 mA / 50 Hz adds approximately 15 % distribution to the measured
result when measuring 1 G.
By selecting FIL1 option the distribution will reduce to less than 2 %.
In general using FIL2 and FIL3 will further improve the noise reduction.
Measurement procedure:
- Connect the test leads to the instrument and to the measured source.
19
MI 3200 TeraOhm 10 kV Measurements
- Press the START key to start the measurement, continuous measurement starts to
run.
- Press the START key again to stop the measurement.
- The result (see the right figure in the Fig. 9) can optionally be saved by pressing the
MEM key twice, see chapter 6.1 Store, Recall and Clear Operation.
Warning!
Refer to Warnings chapter for safety precautions!
Fig. 10. Insulation Resistance function display states - Graph R(t) disabled
The Fig. 11 shows states when Graph R(t) is enabled. When Graph R(t) is enabled you
can simple switching initial state and state with results after the completion of the
measurement, between numerical and graphical mode with pressing or keys.
graphical mode
numerical mode
Note:
- It is not possible to switching mode of presentation when measurement running!!!
20
MI 3200 TeraOhm 10 kV Measurements
Fig. 11. Insulation Resistance function display states - Graph R(t) enabled
Measurement procedure:
- Connect the test leads to the instrument and to the test object.
- Select INSULATION RESISTANCE function in MAIN MENU.
- Press the START/STOP key and release it, continuous measurement will begin.
- Wait until the test result has stabilized then press the START/STOP key again to
stop the measurement or, if enabled, wait for the set timer runs out.
- Wait for the object under test to discharge.
- The result can optionally be saved pressing the MEM key twice, see chapter 6.1
Store, Recall and Clear Operation.
Notes:
- If the timer is disabled then OFF is displayed instead of the timer value.
- During a measurement, the timer information displays the time needed to the
complete the measurement (tr) while after the completion the test duration (tm) is
displayed.
- A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a potentially dangerous test voltage.
- Value of capacitance is measured during the final discharge of the test object.
21
MI 3200 TeraOhm 10 kV Measurements
Timer and Time1 are independent timers. Maximum time for each of them is
99 min 60 s.
22
MI 3200 TeraOhm 10 kV Measurements
Notes:
- If the Timer is OFF is not possible to Enable the Graph R(t).
- The time duration of Graph R(t) is equal to the value of Timer.
- The Timer value could be very long (up to 100 minutes), so the Special automatic
decimation algorithm is use to write the Graph R(t) to the LCD.
- The cursors of the Graph R(t) could be activated with key
- The cursors of the Graph R(t) could be moved with and keys.
Warning!
Refer to Warnings chapter for safety precautions!
23
MI 3200 TeraOhm 10 kV Measurements
The Fig. 16 shows states when Graph R(t) is enabled. When Graph R(t) is enabled you
can simple switching initial state and state with results after the completion of the
measurement, between numerical and graphical mode with pressing or keys.
graphical mode
numerical mode
Note:
- It is not possible to switching mode of presentation when measurement running!!!
24
MI 3200 TeraOhm 10 kV Measurements
Diagnostic test is a long duration test for evaluating the quality of the insulation material
under test. The results of this test enable the decision to be made on the preventive
replacement of the insulation material. DIELECTRIC ABSORPTION RATIO (DAR)
DAR is ratio of Insulation Resistance values measured after 15s and after 1 minute. The
DC test voltage is present during the whole period of the test (also an Insulation
Resistance measurement is continually running). At the end, the DAR ratio is displayed:
DAR R 1min
iso
R 15s
iso
Note: When determining Riso (15s) pay attention to the capacitance of the test object. It
has to be charged-up in the first time section (15s). Approximate maximum capacitance
using:
t s.10 3
C F
max
U V
where:
t ........... period of first time unit (e.g. 15s)
U ......... test voltage.
Analysing the change in the measured insulation resistance over time and calculating
the DAR and PI are very useful maintenance tests of an insulating material.
25
MI 3200 TeraOhm 10 kV Measurements
PI R 10 min
iso
R 1min
iso
Note: When determining Riso (1min) pay close attention to the capacitance of the object
under test. It has to be charged-up in the first time section (1 min). Approximate
maximum capacitance using:
t s.10 3
C F
max
U V
where:
t ........... period of first time unit (e.g. 1min)
U ......... test voltage.
Analysing the change in the measured insulation resistance over time and calculating
the DAR and PI are very useful maintenance tests of an insulating material.
where:
Idis 1min ...... discharging current measured 1 min after regular discharge
U ................. test voltage
C ................. capacitance of test object.
Measurement procedure:
- Select DIAGNOSTIC TEST function in MAIN MENU.
- Connect the test leads to the instrument and to the measured object.
- Press the START/STOP key to start the measurement.
- Wait until set timer runs out, the result is displayed.
- Wait until the object under test has discharged
26
MI 3200 TeraOhm 10 kV Measurements
- The result can optionally be saved by pressing the MEM key twice, see the chapter
6.1. Store, Recall and Clear Operation.
Notes:
- A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a potentially dangerous test voltage.
- The value of the capacitance is measured during the final discharge of the test
object.
- If enabled, the instrument measures Dielectric Discharge (DD) when the
capacitance is in the range of 5 nF to 50 F.
27
MI 3200 TeraOhm 10 kV Measurements
Time1, Time2 and Time3 are timers with the same start point. The value of each
presents the duration from the start of the measurement. The maximum time is 100 min.
The following Fig.18 shows the timer relationships.
Start Stop
R15s R01min R03min
DAR PI
(if enabled DD)
Time1 Time2
Time2 Time3
28
MI 3200 TeraOhm 10 kV Measurements
Notes:
- The time duration of Graph R(t) is equal to the value of Timer 3.
- The Timer value could be very long (up to 100 minutes), so the Special automatic
decimation algorithm is use to write the Graph R(t) to the LCD.
- The cursors of the Graph R(t) could be activated with key
- The cursors of the Graph R(t) could be moved with and keys.
Warning!
Refer to Warnings chapter for safety precautions!
29
MI 3200 TeraOhm 10 kV Measurements
Fig. 21. Step Voltage function display states - Graph R(t) enabled
The Fig. 22 shows states when Graph R(t) is enabled. When Graph R(t) is enabled you
can simple switching initial state and state with results after the completion of the
measurement, between numerical and graphical mode with pressing or keys.
graphical mode
numerical mode
Note:
- It is not possible to switching mode of presentation when measurement running!!!
Fig. 22. Step Voltage function display states - Graph R(t) enabled
30
MI 3200 TeraOhm 10 kV Measurements
In this test, the insulation is measured in five equal time periods with test voltages from
one fifth of the final test voltage up to full voltage (see the Fig. 23). This function
illustrates the relationship of a materials Insulation resistance and its applied voltage.
Measurement procedure:
- Connect the test leads to the instrument and to the test object.
- Press the START/STOP key to start the measurement.
- Wait until set timer runs out, (the result will be displayed).
- Wait for the object under test to discharge.
- The results can be saved by pressing the MEM key twice, see the chapter 6.1. Store,
Recall and Clear Operation.
U
U
0.8U
0.6U
0.4U
0.2U
0 T 2T 3T 4T 5T t
Fig. 23. Step-up test voltage
31
MI 3200 TeraOhm 10 kV Measurements
Notes:
- Timer information is displayed from the start of the measurement until the completion
of each step measurement.
- Timer information shows the complete measurement period after the completion of
the measurement.
- A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a potentially dangerous test voltage.
- The value of capacitance is measured during the final discharge of test object.
-
Setting up parameters for the Step Voltage test:
- Press the SELECT key, (the Set-up menu (Fig. 24)
will appear on the display);
- Select the parameter (line) to be set using the and
keys;
- Adjust the parameter using the and keys.
Complete the parameter adjustments by pressing either
the ESC or the START/STOP key (to run the
measurement directly). The settings displayed last will
be saved Fig. 24. Set-up menu in Step
Voltage Test
32
MI 3200 TeraOhm 10 kV Measurements
Notes:
- The time duration of Graph R(t) is equal to the value of Step Time Multiplied by 5.
- The Timer value could be very long (up to 150 minutes), so the Special automatic
decimation algorithm is use to write the Graph R(t) to the LCD.
- The cursors of the Graph R(t) could be activated with key
- The cursors of the Graph R(t) could be moved with and keys.
Warning!
Refer to Warnings chapter for safety precautions!
Both functions require breakdown current detection. The test voltage increases step by
step from the Start up to the Stop value over a predefined time and it is kept at the Stop
value for a predefined test time, see the figure below.
Both functions require breakdown current detection. In the function, the test voltage is
increased from the starting voltage to the stop voltage over a predefined time (set by the
parameters). The Stop voltage is then maintained for a predefined test time, (see the
Fig. 27).
Ut Ut
Ustop
Ustep
Ustart Ustart
Ub
0 t 0 t
Tstep Tend
Fig. 27. Test voltage presentation without breakdown (left part) and with breakdown
(right part)
Ut ........ Test voltage
Ustop... End test voltage
33
MI 3200 TeraOhm 10 kV Measurements
Selecting this function displays the following states. The Fig. 28 shows the initial screen
and a screen with results after the completion of a measurement.
Measuring procedure:
- Connect the test leads to the instrument and to the measured object.
- Press the START/STOP key to start the measurement.
- Wait until the set timers run out or until breakdown occurs, (the result will be
displayed).
- . Wait for the object under test is discharge.
- The result can be saved by pressing the MEM key twice, see the chapter 6.1 Store,
Recall and Clear Operation.
Note:
- Breakdown is detected when the measured current reaches or exceeds the set
current level Itrig.
Notes:
- The timer shows the time needed to complete each step during the measurement
and it shows the total measurement period after the completion of the measurement.
- A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a potentially dangerous test voltage.
34
MI 3200 TeraOhm 10 kV Measurements
If Tstep is set to 00min 00s, then the ramp voltage increases by approximately 25 V
every 2s.
Warning!
Refer to Warnings chapter for safety precautions!
35
MI 3200 TeraOhm 10 kV Operation with Results
The nnnn means ser. number of stored Fig. 30. Storage operation menu
result.
-
There are the following possibilities are selectable using the or arrow keys:
- To store result: Highlight SAVE and confirm by pressing the MEM key. If graph R(t)
is enabled in the measurement, it will be automatically saved with the measurement.
- To recall stored result: Highlight RCL and confirm by pressing the MEM key. The last
stored result will be displayed. The menu is replaced with:
“0006” and “0007” represents the serial number of the stored results. G letter means
the graph R(t), if added. Results can be scrolled through by using the and keys.
To see the Graph R(t) press the SELECT key, to go back to the numerical
measurement result press the ESC key.
The recall function can be exited by pressing the ESC or Start key.
- To clear the last stored result: highlight CLR and confirm by pressing the MEM key.
In addition to the main result also, the subresults and parameters of the selected
function are also recorded. The following is a list of all the data stored for each function.
36
MI 3200 TeraOhm 10 kV Operation with Results
37
MI 3200 TeraOhm 10 kV Operation with Results
38
MI 3200 TeraOhm 10 kV Maintenance
7. Maintenance
7.1. Inspection
To maintain the operator’s safety and to ensure the reliability of the instrument it is
advisable to inspect the instrument on a regular basis. Check that the instrument and its
accessories are not damaged. If any defect is found please consult your service center,
distributor or manufacturer.
Connect the instrument to the mains power supply for 20 hours to fully charge batteries.
(Typical charging current is 600 mA).
When you charge the batteries for the first time, it normally takes about 3 charge and
discharge cycles for the batteries to regain full capacity.
Note:
The operator does not need to disconnect the instrument from mains supply after the
full recharging period. The instrument can be connected permanently.
Fully charged rechargeable batteries can supply the instrument for approx. 4 hours.
(Continues testing at 10kV)
If the batteries have been stored for a long time, it normally takes about 3 charge and
discharge cycles for the batteries to regain full capacity.
39
MI 3200 TeraOhm 10 kV Maintenance
Battery cells are stored in the bottom section of the instrument casing under the battery
cover (see Fig. 31). In case the batteries become defective please note the following:
Turn the power off and disconnect any measurement accessories or mains
supply cable connected to the instrument before opening the battery cover to
avoid electric shock.
Remove the battery cover.
All six cells have to be replaced and they have to be of the same type.
To easy remove batteries from battery holder first remove the middle batteries
in each columns and then the others (see the Fig. 31).
Insert the batteries correctly (see Fig. 31), otherwise the test instrument will
not operate and battery may be discharged!
To easy insert batteries to battery holder first insert the upper and the lower
batteries in each columns and at end the middle batteries.
The battery cover has to be placed and fixed back.
The Instrument will only work when rechargeable batteries are inside the
instrument
Nominal power supply voltage is 7.2 V DC. Use six NiMH with D cells (dimensions:
diameter = 33 mm, height = 58 mm). See the Fig. 31 for correct polarity of the batteries.
Ensure batteries are used and disposed of in accordance with Manufacturers guidelines
and in accordance with Local and National Authority guidelines!
40
MI 3200 TeraOhm 10 kV Maintenance
7.3. Cleaning
Use a soft cloth, slightly moistened with soapy water or spirit to clean the surface of the
instrument and leave the instrument to dry totally before using it.
Notes!
Do not use liquids based on petrol or hydrocarbons!
Do not spill cleaning liquid over the instrument!
7.4. Calibration
It is essential that all measurement instruments be regularly calibrated. For occasional
daily use, we recommend an annual calibration to be carried out. When the instrument is
used continuously every day, we recommend calibrating the instrument every six
months.
7.5. Service
For repairing under or out of warranty period contact your distributor for further
information.
41
MI 3200 TeraOhm 10 kV Specifications
8. Specifications
8.1. Measurement specifications
Note: All data regarding accuracy is given for nominal (reference) environment condition.
Insulation resistance
Nominal test voltage: Any within 500 and 10000 V
Current capability of test generator: >1 mA
Short-circuit test current: 5 mA 10 %
Automatic discharge of tested object: yes
Measuring range Riso: 0.12 M up to 10 T
Display range Riso Resolution
5 999 k 1 k
1.00 9.99 M 10 k
10.0 99.9 M 100 k
100 999 M 1 M
1.00 9.99 G 10 M
10.0 99.9 G 100 M
100 999 G 1 G
1.00 10.00 T 10 G
Notes:
Full-scale value of resistance (RFS) depends on nominal test voltage (UN) and is
defined according to the following equation:
R FS
1*109 Ω/V U N V
DC test voltage:
Voltage value: Any value within 500 V and 10 kV, steps by 25 V.
Accuracy: -0 / +10 % + 20 V.
Output power: 10 W max.
Display range Test voltage (V) Resolution Accuracy
0 9999 V 1V (3 % of reading + 3 V)
10 kV 0.1 kV (3 % of reading)
42
MI 3200 TeraOhm 10 kV Specifications
Current:
Display range I (mA) Resolution Accuracy
1.00 5.50 mA 10 A
100 999 A 1 A
10.0 99.9 A 100 nA
1.00 9.99 A 10 nA (5 % of reading + 0.05 nA)
100 999 nA 1 nA
10.0 99.9 nA 0.1 nA
0.00 9.99 nA 0.01 nA
12
10
8
[ kV ]
0
0.1 1 10 100
[ M ]
Utest=10kV Utest=5kV
Polarization index PI
Display range PI Resolution Accuracy
43
MI 3200 TeraOhm 10 kV Specifications
Step voltage
DC test voltage:
Voltage value: Any value within 2000 V (400 V, 800 V, 1200 V, 1600 V, 2000 V)
and 10 kV (2000 V, 4000 V, 6000 V, 8000 V, 10 kV), steps by
125 V.
Accuracy: -0 / +10 % + 20 V.
Display range Test voltage (V) Resolution Accuracy
0 9999 V 1V (3 % of reading + 3 V)
10 kV 0.1 kV (3 % of reading)
Withstanding voltage DC
DC test voltage:
Voltage value: Any value within 500V and 10kV.
Accuracy: -0 / +10 % + 20 V.
Leakage current
Display range Itrigg (mA) Resolution Accuracy
0.000 0.009 1 A (3 % of reading + 3 digits)
0.01 5.50 10 A (3 % of reading)
Voltage
Voltage AC or DC
Display range External Voltage (V) Resolution Accuracy
0 600 1V (3 % of reading + 4 V)
Note:
- for frequency between 0 and 45 Hz displayed <45 Hz
- for frequency over 65 Hz displayed >65 Hz
Input resistance: 3 M 10 %
44
MI 3200 TeraOhm 10 kV Specifications
Capacitance
Measuring range C: 50 F*
Display range C Resolution Accuracy
0.0 99.9 nF 0.1 nF
100 999 nF 1 nF (5 % of reading + 2 digits)
1.00 50.00 F 10 nF
*Full-scale value of capacitance is defined according to the following equation:
CFS = 10 F * Utest[kV]
45
MI 3200 TeraOhm 10 kV Specifications
46
MI 3200 TeraOhm 10 kV Specifications
CLOCK
Built-in Real time clock ................................... Displayed permanently and stored as a
parameter in combination with the result.
47