OmniScan X3 EN 202307 Web
OmniScan X3 EN 202307 Web
OmniScan X3 EN 202307 Web
2
Confirm Your
Coverage in
Advance
The Acoustic Influence Map (AIM)
tool provides you with an instant
visual model of the sensitivity
based on your mode, settings, and
simulated reflector.
3
Improved Phased Array
Innovations for
Efficiency
Single time-of-flight
diffraction (TOFD) menu for
accelerated workflow
4
5
Reliable and Easy to Use
If you are an existing OmniScan™ user, the transition from the MX2 is fast. If you are new to phased array ultrasonic
testing or TFM, the OmniScan X3 flaw detector is easy to learn.
(1) A single scan using 4 TFM groups optimized for a thickness of 50 mm.
(2) Scan up to a single 7.6 m × 1.7 m (25 ft × 5.6 ft) storage tank plate with a 1 mm × 1 mm resolution using a HydroFORM™ scanner.
(3) A single pass of a wind tower weld using 4 sector scans, 2 linear scans, and 2 TOFD groups.
6
The Workhorse of Your Inspection Fleet
The OmniScan™ X3 flaw detector offers tools to help you complete your work efficiently. Its range of applications include
welds, pipelines, pipes, corrosion resistant alloys, corrosion mapping, HTHA inspection, detection of stepwise cracking,
composite inspection, flaw imaging, and more.
A Better OmniScan
❯ IP65 certified rain and dust proof ❯ Onboard GPS to record the
❯ User-replaceable cooling fan can location of your data
be changed without opening up the ❯ Connect wirelessly to the Evident
instrument or voiding the calibration Connect cloud platform to download
the latest software as soon as
it’s available
7
Specifications
Size (W × H × D) 335 mm × 221 mm × 151 mm (13.2 in. × 8.7 in. × 5.9 in.) Acoustic Specifications
Weight 5.7 kg (12.6 lb) (with 1 battery) PA Channel UT Channels
Onboard Storage 64 GB internal SSD storage, extendable as needed with an external USB Pulser Voltage 40 V, 80 V, and 115 V 85 V, 155 V, and 295 V
drive; 25 GB maximum file size
Pulse Width Adjustable from 30 ns Adjustable from 30 ns
Storage Devices SDHC™ and SDXC™ cards or most standard USB storage devices to 500 ns; resolution to 1,000 ns; resolution
of 2.5 ns of 2.5 ns
GPS Yes (unless specified otherwise for some regions)
Fall Time < 10 ns < 10 ns
Alarms 3
Pulse Shape Negative square Negative square pulse
Wireless Connection Yes (wireless LAN dongle included in the package)
pulse
Connectors 1 PA connector, 2x UT channels (2 P/R connectors each)
Output 28 Ω in pulse-echo < 30 Ω
Number of Groups 8 groups (16:128PR and 32:128PR); Impedance 24 Ω in pitch-catch
16:64PR offers either 2 groups (PA, UT, or TFM) or 2PA + 1 UT
Receiver Gain Range 0 dB to 80 dB 0 dB to 120 dB maximum
Certifications ISO maximum input input signal; 30 Vp-p
18563‑1:2015 signal; 800 mVp-p (full-screen height)
ISO 22232-1:2020 (full-screen height)
Display Input Impedance 57 Ω ± 10% at 9 MHz 50 Ω in pulse-echo mode
Type TFT LCD with resistive touch screen in pulse-echo 50 Ω in pulse-receive
100 Ω ± 10% at 9 MHz mode
Size 269 mm (10.6 in.) in pitch-catch
Resolution 1280 × 768 pixels System 0.5 MHz to 18 MHz 0.25 MHz to 28 MHz
Inputs and Outputs Bandwidth
Ports 2 USB ports (one hidden behind the battery), 1 USB 3.0, HDMI video Beam Formation Scan Type Single, linear,
output, SDHC memory card, and Ethernet communication port sectorial, compound,
and TFM
Encoder 2-axis encoder line (quadrature or clock/direction), 3rd encoder ready
Maximum OMNIX3-PA16128PR
Digital Input 6 digital inputs, TTL (enabling acquisition ON/OFF) Aperture and OMNIX3-
Digital Output 5 digital outputs, TTL PA16:64PR =
16 elements
Power Output Line 5 V nominal, 1 A (short-circuit protected), and 12 V output at 1 A OMNIX3-PA32128PR =
External DC Supply 32 elements
DC-IN Voltage 15 VDC to 18 VDC (min. 50 W) Number of Focal Up to 1024
Laws
Connector Circular, 2.5 mm pin diameter, center-positive
Delay Range 0 µs to 10 µs in 2.5 ns
Battery Transmission increments
Type Lithium-ion battery Delay Range 0 µs to 6.4 µs in
Capacity 93 Wh Reception 2.5 ns increments
Number of Batteries 2 TFM/FMC
Life 5 hours using 2 batteries (hot-swap capable) Supported Modes Pulse echo: L-L, TT, and TT-TT
Self-Tandem: TT-T, LL-L, LT-T, TL-T, TT-L, TTT-TT, and TL-L
PA/UT Configuration
Parallel Multimode 4 simultaneous TFM groups (wave sets)
Frequency Effective Up to 100 MHz TFM
Digitizing
Frequency Live Envelope Yes
Process
Max PRF 20 kHz
Maximum Aperture 64-element extended aperture (32-128PR only)
Display Refresh Rate A-scan: 60 Hz; S-scan: 20 Hz to 30 Hz 32-element extended aperture for 16:64PR and 16:128PR
Envelope (Echo Yes: Volume-corrected S-scan (30 Hz) Image Resolution Up to 1024 × 1024 (1 MM points) (for each TFM wave set)
Dynamic Mode)
Operating Environment
A-Scan Height Up to 800%
Ingress Protection IP65 certified (completely protected against dust and water jets from
Data Specifications Rating all directions (6.3 mm nozzle))
Processing Maximum Up to 16,384 Shockproof Rating Drop tested according to MIL-STD-810G
Number of
A-Scan Data Intended Use Indoor and outdoor use
Points Operating -10°C to 45° C (14 °F to 113°F)
Real-Time PA: 2, 4, 8, 16 Temperature
Averaging UT: 2, 4, 8, 16, 32, 64 Storage -20 °C to 60 °C (-4 °F to 140 °F) (with battery inside)
Rectification RF, full wave, half wave+, half wave- Temperature -20 °C to 70 °C (-4 °F to 158 °F) (with no battery inside)
Filtering PA channel: 3 low-pass, 6 band-pass, and 4 high-
pass filters Three Available Options
UT channel: 8 low-pass, 6 band-pass, and 4 high- The OmniScan X3 flaw detector is available in 16:64PR*, 16:128PR, and 32:128PR models.
pass filters (3 low-pass filters when configured It is easy to upgrade to the 32:128PR model if you decide you need more pulsers.
in TOFD) * Limited to 2 groups (PA, UT, or TFM) or 2PA + 1 UT.
Video Filtering Smoothing (adjusted to the probe frequency
Standard Inclusions (32:128PR)
range)
OmniScan X3 phased array instrument, including FMC/TFM functionality and 2 UT channels,
Programmable TCG Number of Points 32: One TCG (time-corrected gain) curve per focal and regionally configured power cord with printed instructions. Includes the latest version of
law OmniScan MXU software, a rigid transport case, calibration certificate, 93 Wh lithium-ion battery,
spare screen protector, DC charger with power cord, USB key with OmniScan software and
Range PA (standard): 40 dB per step of 0.1 dB
user manuals, wireless LAN dongle, SDHC™ card, empty USB key for file transfer purposes, and
PA (extended): 65 dB per step of 0.1 dB
OmniPC analysis software. GPS functionality restricted in some regions. Contact your Evident
UT: 100 dB per step of 0.1 dB
representative for more details.
Maximum Slope PA (standard): 40 dB/10 ns
PA (extended): 0.1 dB/10 ns
E0440132EN
UT: 40 dB/10 ns
Evident Scientific, Inc. Evident Canada Inc. EVIDENT CORPORATION is certified to ISO 9001, ISO 14001, and OHSAS 18001.
48 Woerd Avenue 3415 Rue Pierre-Ardouin, *All specifications are subject to change without notice.
Waltham, MA 02453, USA Québec, QC G1P 0B3, Canada All brands are trademarks or registered trademarks of their respective owners and third party entities.
Evident, the Evident logo, OmniScan, HydroFORM, Dual Linear Array, and Dual Matrix Array are trademarks of
(1) 781-419-3900 +1-418-872-1155 Evident Corporation and its subsidiairies.
Copyright © 2023 by Evident.