AXIOM Aristos Image System

AX

Installation

Software
Software Installation Instructions, VA20 or Higher

© Siemens AG 2004
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.

English
Print No.: AXB4-320.816.03.02.02 Doc. Gen. Date: 07.04
Replaces: AXB4-320.816.03.01.02
0-2 Revision
Chapter Page Revision
all all 01

Document revision level


The document corresponds to the version/revision level effective at the time of system delivery. Revi-
sions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.

AXIOM Aristos Image System AXB4-320.816.03 Page 2 of 4 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Contents 0-3

Page
1 _______Requirements____________________________________________________ 1

General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

2 _______Windows NT Installation ___________________________________________ 1

General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Installing the NT OS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Installing OS Patch Disk for Primergy 170 (Black housing) . . . . . . . . . . . . . . . . 4
Installing OS Patch Disk for Celsius 460 (green housing) . . . . . . . . . . . . . . . . . 5
Installing Virus Patch for Celsius 460 (green housing) and Celsius M410 (silver housing)
only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Final Check (for Celsius 460 and Celsius M410 only) . . . . . . . . . . . . . . . . . 6
For Primergy 170 (black housing) only . . . . . . . . . . . . . . . . . . . . . . . . 6

3 _______Image System Aristos Appl. SW Installation __________________________ 1

General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Computer Name Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Calibrating the Detector(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

4 _______Installing the Patch Software _______________________________________ 1

General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Installing OS Patch Disk for Celsius 460 (green housing) . . . . . . . . . . . . . . . . . 1
Installing Image System Patch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
System Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

5 _______Backup and Restore ______________________________________________ 1

Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Backup of SW Settings and IQAP Database . . . . . . . . . . . . . . . . . . . . . 1
Note teh Service Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Note the Host Name and IP Addresses on Paper . . . . . . . . . . . . . . . . . . . 2
Note the Local AEs on Paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Note the Customer and Site Settings on Paper . . . . . . . . . . . . . . . . . . . . 2
Note the RIS Options on Paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Note the Window Image Processing Parameters on Paper . . . . . . . . . . . . . . 2
Note the Customer Settings on Paper . . . . . . . . . . . . . . . . . . . . . . . . . 2
Organ Program Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Restore . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Restoring the SW settings and IQAP Database . . . . . . . . . . . . . . . . . . . . 5
Generating Backup Packages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Siemens AG AXB4-320.816.03 Page 3 of 4 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
0-4 Contents

Page
Restoring the Organ Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Reconfigure the Host Name and IP Addresses. . . . . . . . . . . . . . . . . . . . . 6
Reconfigure the Local AEs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Reconfigure the Customer and Site Settings . . . . . . . . . . . . . . . . . . . . . . 6
Reconfigure the RIS Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Reconfigure the Window Image Processing Parameters . . . . . . . . . . . . . . . . 7
Reconfigure the Customer Settings . . . . . . . . . . . . . . . . . . . . . . . . . . 7

6 ______ Forms for a Manual Configuration Backup ____________________________ 1

7 ______ Changes to Previous Version _______________________________________ 1

AXIOM Aristos Image System AXB4-320.816.03 Page 4 of 4 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Engineering
Requirements 1 1-1

General Remarks 1

This procedure describes the steps necessary to upgrade or reinstall software on Image
System AXIOM Aristos.
For system breakdowns, it is reasonable to perform the complete installation (Win NT +
Application).
Unless otherwise specified, it is sufficient to reinstall only the application SW for updates
of the Image System AXIOM Aristos.

Required Tools 1

Standard service tools.

Safety Information 1

The “Safety Information“ contained in the system manual and in the particular unit instruc-
tions must be observed.

Protection of Electrostatically-sensitive Components


Integrated circuits and circuit boards on which MOS components are installed call for par-
ticularly careful handling because of their highly sensitive electrical structures and
because of their extremely high input resistance. The ESD regulations that apply for elec-
trostatically-sensitive components must be observed.

Required Parts

Description Quantity

Windows NT OS CD, VA10A (for PC Primergy 170 only = black cabinet) 1


or or
Windows NT OS CD, VA20C or higher (for Celsius 460 only = green cabinet) 1
or or
Windows NT OS CD, VA21A or higher (for Celsius M410 only = grey cabinet) 1
O.S. VA21 CD-R-SCSI-Boot-Disc VA00A or higher (for Celsius M410 with 1
SCSI CD-Writer only) also in:
http://cs.med.siemens.de >For Service > AX >AXIOM Aristos > Software

RADIS Application CD, VA20 or higher 1

Customer backup organ programs (Disk) 1

Backup Image of FD System SW settings (Disk) 1

Necessary Patch (Disk) e.g. VA20L 1

Patch VA10E (for PC Primergy 170 only)


also in: 1
http://cs.med.siemens.de >For Service > AX >AXIOM Aristos > Software

Siemens AG AXB4-320.816.03 Page 1 of 2 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
1-2 Requirements
This page intentionally left blank.

AXIOM Aristos Image System AXB4-320.816.03 Page 2 of 2 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Windows NT Installation 2 2-1

General Remarks 2

The purpose of this procedure is to give instructions to load the NT OS into the Image
System AXIOM Aristos after the system drive is replaced or Re-install over an existing
Windows NT. It is recommended to load the NT OS before loading the Image System
AXIOM Aristos Application. It requires that the proper set up and configuration of the hard-
ware and associated BIOS setting has been configured correctly.
The NT OS CD must be run from a bootable CD ROM device.

Backup 2

NOTE Before reinstalling the software, a backup of the configuration


and an organ program backup is imperative.
To do this, see chapter -5- and start a backup process.
Make a note of all customer-specific configurations such as key-
board layout, regional settings,...Please use the blank screen
shots in chapter -6- as a backup medium. After finishing the SW
upgrade, these pages can also be used as a backup for later ser-
vice.

Installing the NT OS 2

General Remarks 2

NOTE All information in the drives will be destroyed. Archive all the
images or put them on film before carrying out this installation.

Follow the instructions in the chapter -5- to back up the AXIOM Aristos Image System
Configuration and ORGAN programs, if this is necessary.

Siemens AG AXB4-320.816.03 Page 1 of 8 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
2-2 Windows NT Installation

NOTE If problems during bootup occur or error messages appear that could
lead to a stop of the boot routine following a new installation, it may be
necessary to COMPLETELY format the hard disks.

Formatting the hard disks (necessary only if the prob-


lems described occur)
• boot up the system
• login into the Adaptec controller by pressing CRTL - A, when the
message appears on the screen
• select SCSI utilities
• select a disk drive
• select format disk
--hint: formatting takes approximately 40min per disk
• do it with the second disk also

To install the NT Workstation operating system on to the Image System AXIOM Aristos,
perform the following steps:
• Switch on power to the Image System AXIOM Aristos.

For Celsius M410 grey cabinet with SCSI Plexwriter 40/12/40S only:
• Insert the O.S. VA21 CD-R-SCSI-Boot-Diskette 4787631 G2192
End of Celsius M410 grey cabinet with SCSI Plexwriter 40/12/40S only

• While the system is running initial power-on diagnostics, insert the NT OS CD into the
bootable CD-ROM device.
• Wait for the system to boot.

NOTE This operation repartitions and formats the primary hard disk
drive; all information on the drive will be destroyed.

• A prompt will be displayed indicating that any information located on the primary hard
disk device will be destroyed.
• Respond to the prompt by pressing the F1 key on keyboard.

NOTE Upon completion of formatting, the computer will reboot.

• After the reboot, a prompt will be displayed indicating that a secondary hard disk
device has been detected.

AXIOM Aristos Image System AXB4-320.816.03 Page 2 of 8 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Windows NT Installation 2-3

SIEMENS MEDICAL SYSTEM AX - NT OS VA.....


A second hard drive has been detected.

Please select from the following choices

1. Continue Installation
(Second hard drive will not be formatted/partitioned)
2. Partition second hard drive and continue installation
(Caution: Remove all patient information and images from the system)

Please enter the corresponding number:

• Select "2"
The setup window is shown as below.

Windows NT Workstation Setup

Please wait while setup copies files to your hard disk.

Setup is copying file%

• When the file copy operation is completed, the computer will reboot.
• When prompted, remove the NT OS CD from the CD ROM drive.

For Celsius M410 grey cabinet with SCSI Plexwriter 40/12/40S only:
• When prompted, remove the O.S. VA21 CD-R-SCSI-Boot-Diskette from the Disk drive.
End of Celsius M410 grey cabinet with SCSI Plexwriter 40/12/40S only.

• Close the CD ROM drive and hit any key to continue.


Several boot routines take place.
• A prompt will be displayed to enter a "Computer Name" (= Host Name).

Siemens AG AXB4-320.816.03 Page 3 of 8 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
2-4 Windows NT Installation
• Respond to the prompt by entering a computer name.
(With a new installation, it is absolutely necessary to use the original name according to
the save.)

NOTE Be sure the computer name is correct before selecting "Next".


You cannot return to this screen to change the name. If there are
no special requirements from the customer site, always use the
original name.

• Select NEXT. Select NEXT again to continue with the installation.

NOTE During the following procedure a message box with a DHCP error
may appear several times. Skip the message box by pressing the
’NO’ button.
In addition an error message may pop up. Confirm the error mes-
sage by pressing the ’OK’ button. Subsequently a message win-
dow may appear asking for the path for a missing file. Confirm
this window with "IGNORE".

• Wait for the installation procedure to complete.

NOTE The entire installation operation will require approximately 15


minutes to complete and should require no further user interven-
tion. During the installation the computer will reboot several
times and restart.

• Wait until the system returns with the prompt Ctrl+Alt +Delete to log on.
• Log on to the system by pressing Ctrl+Alt +Delete.
• Enter Customer for the user.
• Enter "customer" for the password.
• After logging on to the system, the "Date/Time properties" dialog box will be displayed.
- Select the proper time zone and click on Apply.
- Set the date and time, click on Apply and click on OK.
• This completes Windows NT installation.
• Follow the instructions in the next chapter to install the Image System AXIOM Aristos
Application software.

Installing OS Patch Disk for Primergy 170 (Black housing) 2

• After complete installation of OS VA10 insert the floppy with the OS VA10E patch.
(Disk is also available as SI AX104/02/I available at http://cs.med.siemens.de >For Ser-
vice > AX >AXIOM Aristos > Software.)

AXIOM Aristos Image System AXB4-320.816.03 Page 4 of 8 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Windows NT Installation 2-5
• Install Radis OS VA10 Patch2 by opening the Windows Explorer and double-clicking on
the file "ParamUpdate.bat".

NOTE Since application SW VA20, it is not possible to install OS VA10E


or a higher patch via Local Service > Param Update. It has to be
done just after completion of OS installation.

Installing OS Patch Disk for Celsius 460 (green housing) 2

• After complete installation of OS VA20C insert the floppy with the O.S. VA 20 D Patch 2
7728319
(Disk is also available as SI AX104/02/I at http://cs.med.siemens.de >For Service > AX
>AXIOM Aristos > Software)
• Install Radis OS VA20 D Patch2 by opening the Windows Explorer and double-clicking
on the file "ParamUpdate.bat".

Installing Virus Patch for Celsius 460 (green housing) and Celsius
M410 (silver housing) only 2

• Insert "SW-CD Radis Application VA20J" 7555456 into Radis PC.


• Open the Explorer and go to Drive (R)
> Virus Patches for NT >MS03-039-KB824146 >NT4 Work
• Double click on "WindowsNT Workstation KB824146-x86-ENU.EXE".
• Confirm the window by clicking on <OK>, the system will then perform a restart.
• Remove CD.

Siemens AG AXB4-320.816.03 Page 5 of 8 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
2-6 Windows NT Installation
Final Check (for Celsius 460 and Celsius M410 only) 2

• Go to >Start > Settings > Control Panel and open "Add/Remove Programs"

• Check the existence of Hotfix

NOTE You must reinstall this patch if system software was reloaded.

• Close all windows

For Primergy 170 (black housing) only 2

NOTE The Security Patch does not work with Primergy 170. To protect the
system against viruses it is necessary to switch off the DCOM service
and close the appropriate port.

• Click on 'Start' on the Start bar at the bottom of the screen.


• Select 'Run' from the menu.
• Enter 'Dcomcnfg.exe' .
• Confirm with "OK".

AXIOM Aristos Image System AXB4-320.816.03 Page 6 of 8 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Windows NT Installation 2-7
• Wait until the following window appears:

• Choose the Default Properties tab.


• Clear the "Enable Distributed COM on this Computer" check box.
• Click "OK" to apply the changes and exit.
• Close all windows.

Siemens AG AXB4-320.816.03 Page 7 of 8 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
2-8 Windows NT Installation
This page intentionally left blank.

AXIOM Aristos Image System AXB4-320.816.03 Page 8 of 8 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Image System Aristos Appl. SW Installation 3 3-1
The installation of the Image System AXIOM Aristos Application Software requires the NT
OS to be installed first. Perform a complete installation of the PC operating system prior to
installing the Image System AXIOM Aristos software to remove a previous version of
Image System AXIOM Aristos.

General Remarks 3

Computer Name Verification 3

Before installing the Image System AXIOM Aristos application software ensure that the
computer name is unique if more than one PC is to be used on a network. Duplicate
names on a single Network "Workgroup" or "Domain" will prevent the application software
from operating properly.

NOTE The computer name is used to create the data base. Therefore, if
the NT OS CD was used to set up more than one computer on the
same network, the computer names, as well as the configured IP
addresses, must be changed.
If there are no special requirements from the customer site, the
Image System AXIOM Aristos PC Ident number should be used as
computer name, e.g. Radis01234.
For reinstallation, if at all possible, the same name as was used
for the original installation should be used.

To check the computer name perform the following steps:


• Click the 'Start' button on the task bar located at the bottom of the screen.
• Select 'Settings'.
• Select 'Control Panel'.
• From the control panel select 'Network'.
• The Network settings dialog will be displayed. Verify that the Computer name is unique
for each of the PC connected to the 'Workgroup' or 'Domain'.

NOTE The name of the PC can be changed only if it is not being added
to or is already part of an 'NT Network Domain'. If the PC is a
member of an 'NT Network Domain' a systems administrator for
the domain will be required to add the PC to the domain.

• Select the 'Protocols' tab card.


• Double-click on the 'TCP/IP Protocol' entry.
• Check with the system administrator to determine whether the network you are
connected to supports DHCP services. If so, select the 'Obtain IP address from DHCP
server' radio button.
• If the network does not support DHCP services, select the 'Specify an IP Address' radio
button, and enter unique information.
• Select "Apply" and "OK".
• Close the "Network" an "Control Panel" windows.

Siemens AG AXB4-320.816.03 Page 1 of 4 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
3-2 Image System Aristos Appl. SW Installation
• Restart the system (Start Menu > Shut Down...).

To change the computer name, perform the following steps:


• On the 'Identification' tab card, select the 'Change' button.
• The Identification Changes dialog will be displayed.
• In the "Computer name" dialog box, enter the new name for the Image System AXIOM
Aristos.
• Click on "OK".
• Click on "OK" in the Confirm Network configuration dialog.
• A message will be displayed stating that the changes will not take effect until the system
is restarted.
• Acknowledge the message but do not restart the system yet.
• Select the 'Protocols' tab card.
• Double-click on the 'TCP/IP Protocol' entry.
• Check with the system administrator to determine whether the network you are
connected to supports DHCP services. If so, select the 'Obtain IP address from DHCP
server' radio button.
• If the network does not support DHCP services, select the 'Specify an IP Address' radio
button, and enter unique information.
• Select "Apply" and "OK".
• Close the "Network" an "Control Panel" windows.
• Restart the system (Start Menu > Shut Down...).

Installation 3

• Login to the system with the user name 'Customer' and the password 'customer'. This
is the user installed by the OS CD.
• Insert the Image System AXIOM Aristos Application CD into the drive.
• Click on 'Start' in the Start bar at the bottom of the screen.
• From the menu select 'Run'.
• Enter 'R:\Setup.exe' (R being the letter of the CD-ROM drive) or "Browse.." to
R:\Setup.exe and double-click.
• Press 'OK'.
A screen window is displayed as shown below.

Please wait. Copying files.

AXIOM Aristos Image System AXB4-320.816.03 Page 2 of 4 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Image System Aristos Appl. SW Installation 3-3

• The installation process will take approximately 5 minutes.


• The installation will start and a 'Welcome' dialog will be displayed.
• Select "Next" to continue.
• A "Check setup Information" dialog will be displayed confirming the installation type
and target directory and the Syngo installation type and target directory.
• Selecting "Next" will begin the installation.

Please wait!! Installing Syngo

An 'Enter Information' dialog will be displayed. The default User name "Customer" is
displayed in the dialog box.
• If the Image System AXIOM Aristos does not connect to NT networking, click on
"NEXT".
• Click on "Next" for the default password, and click on "NEXT" for domain.
• If the Image System AXIOM Aristos is connected to the NT network domain:
- Enter new user name, password and domain into the "Enter Information" dialog box.
• If no domain is used in the network, leave the "Domain" box blank.
• When prompted, select "Yes, I want to restart my computer now" and select 'Finish'.
• The system will reboot.
• Remove the CD from the drive.

NOTE When the system comes back up, there will be a long delay as all
of the databases and Syngo back ends come up. Allow all of the
databases and back ends to be initially created before proceed-
ing.

NOTE The first time Image System AXIOM Aristos is run, import of the
service patient images into the database causes a delay of
approximately 3 minutes. Wait for the prompt in the status bar
that states "5 Second start up delay" before proceeding.

After the successful system boot up, the message "NOT READY- NO PATIENT
SELECTED " is displayed at the bottom of the screen.

Siemens AG AXB4-320.816.03 Page 3 of 4 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
3-4 Image System Aristos Appl. SW Installation

NOTE Only if required, install the Image System AXIOM Aristos patch
according to chapter -4- of these instructions. This finishes the
Image System AXIOM Aristos software download.

Restore and Generating Packages


NOTE After completing installation, the restores and reconfigurations
described in Chapter -5- under "Restore" must be performed and
the packages generated.

Installing Patch Software


NOTE If required, download the software patches as described in Chap-
ter -4-.

Calibrating the Detector(s) 3

> Options > Detector Calibration...


• Perform calibration of the detector or detectors per the instructions on the screen.

AXIOM Aristos Image System AXB4-320.816.03 Page 4 of 4 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Installing the Patch Software 4 4-1

General Remarks 4

For successful function of the Image System AXIOM Aristos, install the existing Patch SW
following the new installation.

NOTE Downloading the Patch SW is necessary only if it is part of the corre-


sponding Image System AXIOM Aristos SW Status. See the Compatibil-
ity List.
If there is more than one Image System AXIOM Aristos Patch, always
download only patch that belongs with the last basic version to be
installed. It contains all previous patches.

For Primergy 170 only.

NOTE Since application SW VA20, it is not possible to install OS VA10E


or higher patch via Local Service > Param Update. It has to be
done just after completion of OS installation according to the
description in chapter 2.

Installing OS Patch Disk for Celsius 460 (green housing) 4

• Insert the floppy with the O.S. VA 20 D Patch 2 7728319


(Disk is also available as SI AX104/02/I at http://cs.med.siemens.de >For Service > AX
>AXIOM Aristos > Software)
• In the main menu, select "Options" and "Local service".
• Enter a valid password for service in the Service Window and click on "OK".
• Select the "RADIS Tools" in the Service Home menu window.
• The RADIS Tools menu window will be displayed, select "Parameter Update".
• Insert the Image System AXIOM Aristos patch floppy into the drive and select "Start
Update" in the Parameter Update menu.

NOTE Select the IGNORE button when the timeout message is displayed.

• When the update is complete, a message will appear indicating that reboot is required;
close the message box by selecting "OK".
• A second message box will appear indicating the update was successful; close the
message box by selecting "OK".
• Close the local service window by selecting the "X" in the upper right hand corner of the
window.
• Remove the "Image System AXIOM Aristos Patch" floppy from drive A:
• Restart the application by selecting "Patient" and "Exit" in the main menu and restart
the Image System AXIOM Aristos.

Siemens AG AXB4-320.816.03 Page 1 of 2 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
4-2 Installing the Patch Software

Installing Image System Patch 4

• Insert the disk with the latest Application patch e.g. "Image System FD Applic. VA20L
Patch".
• Enter Options > Local Service > Service Key.
• > Radis Tools > Parameter Update.
• Press the key: "Start Update".
• Confirm the message with "OK".
• Remove the diskette.
• Exit the service program with "Home" and close the wi.ndow.
• Reboot the workstation: Patient > Exit > Restart
• Upon reboot, the welcome screen should indicate the numeric version of Image System
AXIOM Aristos Patch version under the "Close the GAP" logo.

System Verification 4

• Perform calibration of all detectors in the system and reboot Image System AXIOM
Aristos.
• The Image System AXIOM Aristos should boot up properly.

AXIOM Aristos Image System AXB4-320.816.03 Page 2 of 2 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Backup and Restore 5 5-1

Backup 5

The Image System AXIOM Aristos Configuration and Organ programs can be backed up
or restored by using a floppy. The backup is necessary before updating or reloading the
software into the system. The Image System AXIOM Aristos configuration consists of 2
backup packages:

- Software Settings (Syngo, Radis and user configuration)


- IQAP_Database

Backup of SW Settings and IQAP Database 5

> Options > Local Service


• Enter the Service password level #7 into the Password edit boxes, check the "Default
password" check box, then click on "OK".
• Select "Backup and Restore".
• The backup and restore window is displayed.
In the "Command" drop down list box, select "Backup"
• In the "Drives" list box, select A: Floppy Disk
• Insert the Image System AXIOM Aristos Config floppy labeled "Backup Image System
FD SW Settings" into drive A.
• In the Packages list box, select "Software Settings".
• Click on "GO" in the lower right corner of the window.
The backup process will start. After a few minutes the end backup date and time is dis-
played. "Ready" is displayed in the left corner of the window.
• In the Packages list box, select "IQAP_Database".
• Click on "GO" in the lower right corner of the window.
The backup process will start. After a few minutes the end backup date and time is dis-
played. "Ready" is displayed in the left corner of the window.
• Select "Home"
• Close the window by clicking on "X"

In addition, some configurations have to be recorded on paper because they are


not transferred with the installation:

NOTE Please use the blank screen shots in chapter -6- as a backup
medium. After finishing the SW upgrade, these pages can also be
used as a backup for later service.

Note teh Service Key 5

> Options > Local Service


• When starting the service software, in the Authentication page, note the visible 14
characters of the service key (the last 6 characters must be known anyway).
• Use the blank table in chapter -6-.

Siemens AG AXB4-320.816.03 Page 1 of 8 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
5-2 Backup and Restore

Note the Host Name and IP Addresses on Paper 5

( > Options > Local Service )> Configuration


• Select "TCP/IP LAN" under local host.
• Use the blank table in chapter -6-.

Note the Local AEs on Paper 5

( > Options > Local Service > Configuration )


• Select in Local AEs under DICOM.
• Use the blank table in chapter -6-.

Note the Customer and Site Settings on Paper 5

(> Options > Local Service > Configuration)


• Select "customer and site" under local host.
• Use the blank screen shots in chapter -6-.
• Select "Home".

Note the RIS Options on Paper 5

( > Options > Local Service ) > Radis Tools >RIS Options
• Use the blank screen shots in chapter -6-.
• Select "Radis Tools Menu".

Note the Window Image Processing Parameters on Paper 5

( > Options > Local Service > Radis Tools > )


• Select Image Proc Params.
• Use the blank screen shots in chapter -6-.
• Select "Home".
• Close the window by clicking on "X"

Note the Customer Settings on Paper 5

> Option > Configuration>

AXIOM Aristos Image System AXB4-320.816.03 Page 2 of 8 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Backup and Restore 5-3
• Select Pre-registered Patients.

• Use the blank screen shots in chapter -6-.


• Select "Cancel".

• Select Examined patients.

• Use the blank screen shots in chapter -6-.


• Select "Cancel".

• Select Processing Configuration.

• Use the blank screen shots in chapter -6-.


• Select "Cancel".

• Select "Regional Settings".

• Note the settings for language, date format and keyboard


• Use the blank table in chapter -6-.
• Close the window by clicking on "X"
• Close "Configuration Panel" window by clicking on "X"

Siemens AG AXB4-320.816.03 Page 3 of 8 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
5-4 Backup and Restore
Organ Program Backup 5

NOTE To ensure the functionality of the organ program backup, a special pro-
cedure must be observed:
Open the OGP Editor
> Options > Edit Regions/Studies
and exit it again with "Cancel" (not with Save!!!).
(When this is done, the changes in the filming layout links are accepted
into the OGP database).

> Options > Local Service > Radis Tools > Organ Programs
• Insert the "Backup Customer Organ Programs" floppy.
• In the "RADIS Tools Menu", select "Organ Programs".
• Click on "Backup".
The backup process will start. After a while, the message "Successfully backed up
organ programs...." is displayed.
• Remove the floppy from the drive.
• Select "Home".
• Close the window by clicking on "X".

AXIOM Aristos Image System AXB4-320.816.03 Page 4 of 8 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Backup and Restore 5-5

Restore 5

Restoring the SW settings and IQAP Database 5

Restore of the Configuration and Organ programs is carried out after finishing the soft-
ware update or reloading the software.
> Options > Local Service
• Enter the Service password level #7 into the Password edit boxes, check the "Default
password" check box, then click on "OK".
• Select "Backup and Restore".
• The backup and restore window is displayed. In the "Command" dropdown list box,
select "Restore".
• Insert the backup floppy into the drive.
• In the "Drives" list box, select A:Floppy Disk.
• In the Archive list box, select "SoftwareSettings".
• Select all groups with the left mouse button while holding down the Shift key.
• Click on "GO" in the lower right corner of the window.
The message "BackupRestore Running" is displayed in the left corner of the screen.
The restore process will start. After a few minutes the status shows "Success" and the
start restore and end restore dates and times. "Ready" is displayed in the left corner of
the screen.
• In the Archive list box, select "IQAP_Database".
• Click on "GO" in the lower right corner of the window.
The message "BackupRestore Running" is displayed in the left corner of the screen.
The restore process will start. After a few minutes the status shows "Success" and the
start restore and end restore dates and times. "Ready" is displayed in the left corner of
the screen.
• Select "Home"

Generating Backup Packages 5

To generate the "IQAP","Burn Defective Images" and "Winnie" archive packages, the
Archive Package Creator, is used. With it, the packages that were lost during the update
are restored.
• Insert
"AXIOM Aristos MX/VX System SW CD" 7555639
or
"AXIOM Aristos TX System SW CD" 7555621
or
"AXIOM Aristos FX System SW CD " 7119741
into the Radis PC
• Go to Options >Configuration > Printers > Drive(R) > Radis_Backup_Packages
• Double-click on Archive-package.bat

Siemens AG AXB4-320.816.03 Page 5 of 8 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
5-6 Backup and Restore
A DOS window in which the copy routine is displayed will open briefly.
• Close the windows by clicking on "X".
• Check the existence of the packages:
Options > local Service > Configuration
under Service > backup/restore; all packages must visible.
• Select "Home".
• Close the window by clicking on "X".

Restoring the Organ Program 5

( > Options > Local Service > ) Radis Tools >Organ Programs
• Insert the "Backup Customer Organ Programs" floppy.
• Click on "Scan Floppy".
After a while, the message "Ready to Replace or Merge" is displayed.
• Click on "Replace".
The organ programs will be restored and after a while the message "Replacement suc-
cessfully completed" is displayed.
• Remove the "Backup Customer Organ Programs" floppy.
• Select "Home".

Some configurations had to be recorded on paper because they were not trans-
ferred with the backup.

NOTE Please use the screen shots in chapter -6- .

Reconfigure the Host Name and IP Addresses 5

( > Options > Local Service > Configuration )


• Select "TCP/IP LAN" under local host.
• Use the data recorded on paper in chapter -6-.
• Select "Save" and confirm.

Reconfigure the Local AEs 5

( > Options > Local Service > Configuration )


• Select "Local AEs" under DICOM.
• Use the data recorded on paper in chapter -6-.
• Select "Save" and confirm.

Reconfigure the Customer and Site Settings 5

(> Options > Local Service > Configuration)

AXIOM Aristos Image System AXB4-320.816.03 Page 6 of 8 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Backup and Restore 5-7
• Select "customer and site" under local host.
• Use the data recorded on paper in chapter -6-.
• Select "Save" and confirm.
• Select "Home".

Reconfigure the RIS Options 5

( > Options > Local Service ) > Radis Tools >RIS Options
• Use the data recorded on paper in chapter -6-.
• Confirm changes.
• Select "Radis Tools Menu".

Reconfigure the Window Image Processing Parameters 5

( > Options > Local Service > Radis Tools > )


• Select Image Proc Params.
• Use the data recorded on paper in chapter -6-.
If the sensitivity is 1.05 then it is a default value. Please adjust the FD signal value. See
the Service Instructions, Replacement of Parts.
• Select "Save" and confirm.
• Select "Home".
• Close the window by clicking on "X"

Reconfigure the Customer Settings 5

> Option > Configuration>


• Select Pre-registered Patients.

• Use the data recorded on paper in chapter -6-.


• Select "OK".

• Select Examined patients.

Siemens AG AXB4-320.816.03 Page 7 of 8 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
5-8 Backup and Restore
• Use the data recorded on paper in chapter -6-.
• Select "OK".

• Select Processing Configuration

• Use the data recorded on paper in chapter -6-.


• Select "Done".

• Select "Regional Settings".

• Reconfigure the settings for language, date format and keyboard.


• Use the data recorded on paper in chapter -6-.
• Select "OK".
• Close the window by clicking on "X".
• Close "Configuration Panel" window by clicking on "X".

AXIOM Aristos Image System AXB4-320.816.03 Page 8 of 8 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Forms for a Manual Configuration Backup 6 6-1

NOTE Please use the blank screen shots as a backup medium. After
completing the SW Installation, these pages can also be used as a
backup or later service.

Service key
Service key = 20 characters -------------------- 14 characters ---------- -- last 6 characters --

Host name
IP Address
Subnet mask
gateways

Local AEs HIS/RIS HRI_


Study/Transfer AN_
Print PR_

Siemens AG AXB4-320.816.03 Page 1 of 8 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
6-2 Forms for a Manual Configuration Backup

AXIOM Aristos Image System AXB4-320.816.03 Page 2 of 8 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Forms for a Manual Configuration Backup 6-3

New beginning with Version VA20D - RIS Accession Number and Request ID Options
If one or more studies (manual) are added during an RIS transferred study, there are two
alternatives:
"Do not Copy data": the Accession Number /Request Procedure ID are not automatically
added; they can be added in the Study Manager.
or
"Copy data from the first study of session": Accession Number / Requested Procedure ID
of the 1st study. The study is automatically added.

Siemens AG AXB4-320.816.03 Page 3 of 8 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
6-4 Forms for a Manual Configuration Backup

AXIOM Aristos Image System AXB4-320.816.03 Page 4 of 8 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Forms for a Manual Configuration Backup 6-5

Order

Tab. 1

Siemens AG AXB4-320.816.03 Page 5 of 8 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
6-6 Forms for a Manual Configuration Backup

AXIOM Aristos Image System AXB4-320.816.03 Page 6 of 8 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Forms for a Manual Configuration Backup 6-7

Predefined List

Tab. 2

Language

Keyboard (Input Locales)

Date format style: separator:

Time format

Siemens AG AXB4-320.816.03 Page 7 of 8 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
6-8 Forms for a Manual Configuration Backup
This page intentionally left blank.

AXIOM Aristos Image System AXB4-320.816.03 Page 8 of 8 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions
Changes to Previous Version 7 7-1
Chapter 1 page 1-1 at table, add boot floppy
Chapter 2 page 2-2 add desription of boot floppy
Chapter 2 page 2-3 add desription of boot floppy
Chapter 2 page 2-5ff. add patch installation desription
Chapter 4 page 4-1 revision of whole desription
Chapter 5 page 5-5ff change generating the packup packages

Siemens AG AXB4-320.816.03 Page 1 of 2 AXIOM Aristos Image System


Medical Solutions Rev. 02 07.04 CS PS 21 Installation
7-2 Changes to Previous Version
This page intentionally left blank.

AXIOM Aristos Image System AXB4-320.816.03 Page 2 of 2 Siemens AG


Installation Rev. 02 07.04 CS PS 21 Medical Solutions