Installation
Software
Software Installation Instructions, VA20 or Higher
© Siemens AG 2004
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
English
Print No.: AXB4-320.816.03.02.02 Doc. Gen. Date: 07.04
Replaces: AXB4-320.816.03.01.02
0-2 Revision
Chapter Page Revision
all all 01
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
Page
1 _______Requirements____________________________________________________ 1
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Installing the NT OS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Installing OS Patch Disk for Primergy 170 (Black housing) . . . . . . . . . . . . . . . . 4
Installing OS Patch Disk for Celsius 460 (green housing) . . . . . . . . . . . . . . . . . 5
Installing Virus Patch for Celsius 460 (green housing) and Celsius M410 (silver housing)
only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Final Check (for Celsius 460 and Celsius M410 only) . . . . . . . . . . . . . . . . . 6
For Primergy 170 (black housing) only . . . . . . . . . . . . . . . . . . . . . . . . 6
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Computer Name Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Calibrating the Detector(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Installing OS Patch Disk for Celsius 460 (green housing) . . . . . . . . . . . . . . . . . 1
Installing Image System Patch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
System Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Backup of SW Settings and IQAP Database . . . . . . . . . . . . . . . . . . . . . 1
Note teh Service Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Note the Host Name and IP Addresses on Paper . . . . . . . . . . . . . . . . . . . 2
Note the Local AEs on Paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Note the Customer and Site Settings on Paper . . . . . . . . . . . . . . . . . . . . 2
Note the RIS Options on Paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Note the Window Image Processing Parameters on Paper . . . . . . . . . . . . . . 2
Note the Customer Settings on Paper . . . . . . . . . . . . . . . . . . . . . . . . . 2
Organ Program Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Restore . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Restoring the SW settings and IQAP Database . . . . . . . . . . . . . . . . . . . . 5
Generating Backup Packages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Page
Restoring the Organ Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Reconfigure the Host Name and IP Addresses. . . . . . . . . . . . . . . . . . . . . 6
Reconfigure the Local AEs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Reconfigure the Customer and Site Settings . . . . . . . . . . . . . . . . . . . . . . 6
Reconfigure the RIS Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Reconfigure the Window Image Processing Parameters . . . . . . . . . . . . . . . . 7
Reconfigure the Customer Settings . . . . . . . . . . . . . . . . . . . . . . . . . . 7
General Remarks 1
This procedure describes the steps necessary to upgrade or reinstall software on Image
System AXIOM Aristos.
For system breakdowns, it is reasonable to perform the complete installation (Win NT +
Application).
Unless otherwise specified, it is sufficient to reinstall only the application SW for updates
of the Image System AXIOM Aristos.
Required Tools 1
Safety Information 1
The “Safety Information“ contained in the system manual and in the particular unit instruc-
tions must be observed.
Required Parts
Description Quantity
General Remarks 2
The purpose of this procedure is to give instructions to load the NT OS into the Image
System AXIOM Aristos after the system drive is replaced or Re-install over an existing
Windows NT. It is recommended to load the NT OS before loading the Image System
AXIOM Aristos Application. It requires that the proper set up and configuration of the hard-
ware and associated BIOS setting has been configured correctly.
The NT OS CD must be run from a bootable CD ROM device.
Backup 2
Installing the NT OS 2
General Remarks 2
NOTE All information in the drives will be destroyed. Archive all the
images or put them on film before carrying out this installation.
Follow the instructions in the chapter -5- to back up the AXIOM Aristos Image System
Configuration and ORGAN programs, if this is necessary.
NOTE If problems during bootup occur or error messages appear that could
lead to a stop of the boot routine following a new installation, it may be
necessary to COMPLETELY format the hard disks.
To install the NT Workstation operating system on to the Image System AXIOM Aristos,
perform the following steps:
• Switch on power to the Image System AXIOM Aristos.
For Celsius M410 grey cabinet with SCSI Plexwriter 40/12/40S only:
• Insert the O.S. VA21 CD-R-SCSI-Boot-Diskette 4787631 G2192
End of Celsius M410 grey cabinet with SCSI Plexwriter 40/12/40S only
• While the system is running initial power-on diagnostics, insert the NT OS CD into the
bootable CD-ROM device.
• Wait for the system to boot.
NOTE This operation repartitions and formats the primary hard disk
drive; all information on the drive will be destroyed.
• A prompt will be displayed indicating that any information located on the primary hard
disk device will be destroyed.
• Respond to the prompt by pressing the F1 key on keyboard.
• After the reboot, a prompt will be displayed indicating that a secondary hard disk
device has been detected.
1. Continue Installation
(Second hard drive will not be formatted/partitioned)
2. Partition second hard drive and continue installation
(Caution: Remove all patient information and images from the system)
• Select "2"
The setup window is shown as below.
• When the file copy operation is completed, the computer will reboot.
• When prompted, remove the NT OS CD from the CD ROM drive.
For Celsius M410 grey cabinet with SCSI Plexwriter 40/12/40S only:
• When prompted, remove the O.S. VA21 CD-R-SCSI-Boot-Diskette from the Disk drive.
End of Celsius M410 grey cabinet with SCSI Plexwriter 40/12/40S only.
NOTE During the following procedure a message box with a DHCP error
may appear several times. Skip the message box by pressing the
’NO’ button.
In addition an error message may pop up. Confirm the error mes-
sage by pressing the ’OK’ button. Subsequently a message win-
dow may appear asking for the path for a missing file. Confirm
this window with "IGNORE".
• Wait until the system returns with the prompt Ctrl+Alt +Delete to log on.
• Log on to the system by pressing Ctrl+Alt +Delete.
• Enter Customer for the user.
• Enter "customer" for the password.
• After logging on to the system, the "Date/Time properties" dialog box will be displayed.
- Select the proper time zone and click on Apply.
- Set the date and time, click on Apply and click on OK.
• This completes Windows NT installation.
• Follow the instructions in the next chapter to install the Image System AXIOM Aristos
Application software.
• After complete installation of OS VA10 insert the floppy with the OS VA10E patch.
(Disk is also available as SI AX104/02/I available at http://cs.med.siemens.de >For Ser-
vice > AX >AXIOM Aristos > Software.)
• After complete installation of OS VA20C insert the floppy with the O.S. VA 20 D Patch 2
7728319
(Disk is also available as SI AX104/02/I at http://cs.med.siemens.de >For Service > AX
>AXIOM Aristos > Software)
• Install Radis OS VA20 D Patch2 by opening the Windows Explorer and double-clicking
on the file "ParamUpdate.bat".
Installing Virus Patch for Celsius 460 (green housing) and Celsius
M410 (silver housing) only 2
• Go to >Start > Settings > Control Panel and open "Add/Remove Programs"
NOTE You must reinstall this patch if system software was reloaded.
NOTE The Security Patch does not work with Primergy 170. To protect the
system against viruses it is necessary to switch off the DCOM service
and close the appropriate port.
General Remarks 3
Before installing the Image System AXIOM Aristos application software ensure that the
computer name is unique if more than one PC is to be used on a network. Duplicate
names on a single Network "Workgroup" or "Domain" will prevent the application software
from operating properly.
NOTE The computer name is used to create the data base. Therefore, if
the NT OS CD was used to set up more than one computer on the
same network, the computer names, as well as the configured IP
addresses, must be changed.
If there are no special requirements from the customer site, the
Image System AXIOM Aristos PC Ident number should be used as
computer name, e.g. Radis01234.
For reinstallation, if at all possible, the same name as was used
for the original installation should be used.
NOTE The name of the PC can be changed only if it is not being added
to or is already part of an 'NT Network Domain'. If the PC is a
member of an 'NT Network Domain' a systems administrator for
the domain will be required to add the PC to the domain.
Installation 3
• Login to the system with the user name 'Customer' and the password 'customer'. This
is the user installed by the OS CD.
• Insert the Image System AXIOM Aristos Application CD into the drive.
• Click on 'Start' in the Start bar at the bottom of the screen.
• From the menu select 'Run'.
• Enter 'R:\Setup.exe' (R being the letter of the CD-ROM drive) or "Browse.." to
R:\Setup.exe and double-click.
• Press 'OK'.
A screen window is displayed as shown below.
An 'Enter Information' dialog will be displayed. The default User name "Customer" is
displayed in the dialog box.
• If the Image System AXIOM Aristos does not connect to NT networking, click on
"NEXT".
• Click on "Next" for the default password, and click on "NEXT" for domain.
• If the Image System AXIOM Aristos is connected to the NT network domain:
- Enter new user name, password and domain into the "Enter Information" dialog box.
• If no domain is used in the network, leave the "Domain" box blank.
• When prompted, select "Yes, I want to restart my computer now" and select 'Finish'.
• The system will reboot.
• Remove the CD from the drive.
NOTE When the system comes back up, there will be a long delay as all
of the databases and Syngo back ends come up. Allow all of the
databases and back ends to be initially created before proceed-
ing.
NOTE The first time Image System AXIOM Aristos is run, import of the
service patient images into the database causes a delay of
approximately 3 minutes. Wait for the prompt in the status bar
that states "5 Second start up delay" before proceeding.
After the successful system boot up, the message "NOT READY- NO PATIENT
SELECTED " is displayed at the bottom of the screen.
NOTE Only if required, install the Image System AXIOM Aristos patch
according to chapter -4- of these instructions. This finishes the
Image System AXIOM Aristos software download.
General Remarks 4
For successful function of the Image System AXIOM Aristos, install the existing Patch SW
following the new installation.
NOTE Select the IGNORE button when the timeout message is displayed.
• When the update is complete, a message will appear indicating that reboot is required;
close the message box by selecting "OK".
• A second message box will appear indicating the update was successful; close the
message box by selecting "OK".
• Close the local service window by selecting the "X" in the upper right hand corner of the
window.
• Remove the "Image System AXIOM Aristos Patch" floppy from drive A:
• Restart the application by selecting "Patient" and "Exit" in the main menu and restart
the Image System AXIOM Aristos.
• Insert the disk with the latest Application patch e.g. "Image System FD Applic. VA20L
Patch".
• Enter Options > Local Service > Service Key.
• > Radis Tools > Parameter Update.
• Press the key: "Start Update".
• Confirm the message with "OK".
• Remove the diskette.
• Exit the service program with "Home" and close the wi.ndow.
• Reboot the workstation: Patient > Exit > Restart
• Upon reboot, the welcome screen should indicate the numeric version of Image System
AXIOM Aristos Patch version under the "Close the GAP" logo.
System Verification 4
• Perform calibration of all detectors in the system and reboot Image System AXIOM
Aristos.
• The Image System AXIOM Aristos should boot up properly.
Backup 5
The Image System AXIOM Aristos Configuration and Organ programs can be backed up
or restored by using a floppy. The backup is necessary before updating or reloading the
software into the system. The Image System AXIOM Aristos configuration consists of 2
backup packages:
NOTE Please use the blank screen shots in chapter -6- as a backup
medium. After finishing the SW upgrade, these pages can also be
used as a backup for later service.
( > Options > Local Service ) > Radis Tools >RIS Options
• Use the blank screen shots in chapter -6-.
• Select "Radis Tools Menu".
NOTE To ensure the functionality of the organ program backup, a special pro-
cedure must be observed:
Open the OGP Editor
> Options > Edit Regions/Studies
and exit it again with "Cancel" (not with Save!!!).
(When this is done, the changes in the filming layout links are accepted
into the OGP database).
> Options > Local Service > Radis Tools > Organ Programs
• Insert the "Backup Customer Organ Programs" floppy.
• In the "RADIS Tools Menu", select "Organ Programs".
• Click on "Backup".
The backup process will start. After a while, the message "Successfully backed up
organ programs...." is displayed.
• Remove the floppy from the drive.
• Select "Home".
• Close the window by clicking on "X".
Restore 5
Restore of the Configuration and Organ programs is carried out after finishing the soft-
ware update or reloading the software.
> Options > Local Service
• Enter the Service password level #7 into the Password edit boxes, check the "Default
password" check box, then click on "OK".
• Select "Backup and Restore".
• The backup and restore window is displayed. In the "Command" dropdown list box,
select "Restore".
• Insert the backup floppy into the drive.
• In the "Drives" list box, select A:Floppy Disk.
• In the Archive list box, select "SoftwareSettings".
• Select all groups with the left mouse button while holding down the Shift key.
• Click on "GO" in the lower right corner of the window.
The message "BackupRestore Running" is displayed in the left corner of the screen.
The restore process will start. After a few minutes the status shows "Success" and the
start restore and end restore dates and times. "Ready" is displayed in the left corner of
the screen.
• In the Archive list box, select "IQAP_Database".
• Click on "GO" in the lower right corner of the window.
The message "BackupRestore Running" is displayed in the left corner of the screen.
The restore process will start. After a few minutes the status shows "Success" and the
start restore and end restore dates and times. "Ready" is displayed in the left corner of
the screen.
• Select "Home"
To generate the "IQAP","Burn Defective Images" and "Winnie" archive packages, the
Archive Package Creator, is used. With it, the packages that were lost during the update
are restored.
• Insert
"AXIOM Aristos MX/VX System SW CD" 7555639
or
"AXIOM Aristos TX System SW CD" 7555621
or
"AXIOM Aristos FX System SW CD " 7119741
into the Radis PC
• Go to Options >Configuration > Printers > Drive(R) > Radis_Backup_Packages
• Double-click on Archive-package.bat
( > Options > Local Service > ) Radis Tools >Organ Programs
• Insert the "Backup Customer Organ Programs" floppy.
• Click on "Scan Floppy".
After a while, the message "Ready to Replace or Merge" is displayed.
• Click on "Replace".
The organ programs will be restored and after a while the message "Replacement suc-
cessfully completed" is displayed.
• Remove the "Backup Customer Organ Programs" floppy.
• Select "Home".
Some configurations had to be recorded on paper because they were not trans-
ferred with the backup.
( > Options > Local Service ) > Radis Tools >RIS Options
• Use the data recorded on paper in chapter -6-.
• Confirm changes.
• Select "Radis Tools Menu".
NOTE Please use the blank screen shots as a backup medium. After
completing the SW Installation, these pages can also be used as a
backup or later service.
Service key
Service key = 20 characters -------------------- 14 characters ---------- -- last 6 characters --
Host name
IP Address
Subnet mask
gateways
New beginning with Version VA20D - RIS Accession Number and Request ID Options
If one or more studies (manual) are added during an RIS transferred study, there are two
alternatives:
"Do not Copy data": the Accession Number /Request Procedure ID are not automatically
added; they can be added in the Study Manager.
or
"Copy data from the first study of session": Accession Number / Requested Procedure ID
of the 1st study. The study is automatically added.
Order
Tab. 1
Predefined List
Tab. 2
Language
Time format