Fuji FP232 Service Manual
Fuji FP232 Service Manual
Fuji FP232 Service Manual
MINILAB
FILM PROCESSOR
FP232B
UL & CE Types
First Edition
PP3-C829UE
INTRODUCTION
This Servicing Manual outlines the maintenance and servicing procedures for the FP232B Film Processor.
This manual is a professional publication provided for qualified service personnel or persons fully trained in
equipment service procedures. All other personnel and operators are restricted from servicing the FP232B.
When maintenance service is needed, be sure to contact qualified service personnel.
The manual is divided into 13 chapters, with chapters 1-6 providing generally applicable information and
chapters 7-13 giving a detailed description of each part and each function. In order for specific details to be
accessed as quickly as possible, the headlines on the right edge of the contents page match the headlines of
the pages of each chapter. The contents of a particular chapter are listed on the first page of that chapter.
! CAUTION
The light source section and dryer will be hot to the touch. Wait 15 minutes after turning off the power before commencing
with servicing procedures.
• To keep the parts free from fingerprints, wear gloves when handling the optical system parts.
• When mounting machine parts, take care not to sandwich any of the wires.
• Each connector is provided with a symbol indicating what it is to be connected to. Make sure the connectors are connected
to connectors, harness ends or parts bearing the corresponding symbols.
• The plastic connectors have a locking catch on the plug (male) end. To disconnect one of these connectors, loosen the
catch first; to connect it, make sure the catch engages (locks).
1. All rights are reserved by the Fuji Photo Film Co., Ltd. (FUJIFILM).
3. MAINTENANCE SCHEDULE
4. ERROR MESSAGES
5. CONDITION SETTINGS
7. LOADER
8. PROCESSOR
11. DRYER
1. SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1
CONTENTS
2
6.4 Input Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
6.4.1 Input Check Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
6.5 Output Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
6.5.1 Output Check Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
6.6 Setting Data Saving, Reading and Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
6.6.1 Setting Data Saving . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
6.6.2 Stored Data Reading and Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
6.7 Error Log Clear . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
6.7.1 Error Log Clear . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
6.8 Replenishment Control Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
6.8.1 Replenishment Control Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
6.9 IX240 Replenishment Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
6.9.1 IX240 Replenishment Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
6.10 Water Replenishment Log/Water Replenishment Rate Measurement . . . . . . . . . . . . . . . . . . . 6-21
6.10.1 Water Replenishment Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21
6.10.2 Water Replenishment Rate Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21
6.11 Water Replenishment Condition Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
6.11.1 DZP Switch Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
6.11.2 Evaporation Rates (During Stand-by) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
6.11.3 Evaporation Rates (During Processing) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
6.11.4 Evaporation Rates (During Power OFF) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
6.11.5 Water Replenishment Rate Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-24
6.11.6 Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-24
6.11.7 Room Temperature Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-25
6.11.8 Circuit Board Adjustments for Initial Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-25
6.12 Demonstration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-26
6.12.1 Demonstration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-26
6.13 Default Values/Hidden Running Total . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
6.13.1 Data Resetting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
6.13.2 Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
6.13.3 Hidden Running Total . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-29
7. LOADER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-0
3
CONTENTS
7.1.8 E402, E403 “# Perforation Sensor Abnormal” E404, E405 “# Film Sensor Abnormal”
E411 “Left 120 Sensor Abnormal” E412, E413 “# IX240 Sensor Abnormal” . . . . . . 7-27
7.1.9 E406, E407 “#. Perforation or Film Sensor Abnormal”
E409 “Left 120 or Film Sensor Abnormal”
E414, E415 “# IX240 or Film Sensor Abnormal” . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-29
7.2 Loader Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-31
7.2.1 Input Check of Loader Cover Detector (S6) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-31
7.2.2 Output Check of Film Press Roller (SOL4/SOL5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-31
7.2.3 Film Press Roller Solenoid (SOL4/SOL5)
and Safety Thermostat (TH5) Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-32
7.3 Magazine Support Plate, Loader Bottom Cover and Auto Lock . . . . . . . . . . . . . . . . . . . . . . . . 7-34
7.3.1 Output Check of Auto Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-34
7.3.2 Magazine Support Plate Cover Removal/Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-34
7.3.3 Loader Cover Detector (S6) Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-35
7.3.4 Loader Bottom Cover Removal/Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-36
7.3.5 Auto Lock Release Switch Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-37
7.3.6 Auto Lock Release Switch (S11) Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-38
7.3.7 Auto Lock Solenoid (SOL1) Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-38
7.4 Cutter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-40
7.4.1 Input Check of Press Roller and Cutter Operating Position Detectors . . . . . . . . . . . . . . 7-40
7.4.2 Output Check of Cutter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-40
7.4.3 Cutter Blade Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-41
7.4.4 Cutter Operating Position in Power Failure Detector Inspection . . . . . . . . . . . . . . . . . . . 7-42
7.4.5 Press Roller Operating Position Detector (S2/S3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-42
7.4.6 Cutter Operating Position Detector (S4/S5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-43
7.4.7 Power OFF Cutter Operating Position Detector (S8/S9) . . . . . . . . . . . . . . . . . . . . . . . . . 7-43
7.4.8 Cutter Solenoid (SOL2/SOL3) Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-44
7.5 Feeder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-46
7.5.1 Input Check of Film Sensors (FMS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-46
7.5.2 Film Sensor (FMS) Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-46
7.5.3 Loader Cover Interlock Detector (S10) Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-48
7.5.4 Feeding Roller Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-49
8. PROCESSOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
4
8.3 Processing Racks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12
8.3.1 Processing Rack Belt/Roller Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12
8.4 Processing Section Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14
8.4.1 Input Check of Processing Section Cover Detector (S7) . . . . . . . . . . . . . . . . . . . . . . . . 8-14
8.4.2 Processing Section Cover Detector Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14
8.4.3 Processor Ventilation Fan Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-15
5
CONTENTS
6
13. CONTROL PANEL AND ELECTRICAL EQUIPMENT SECTION . . . . . . . . . . . . . . . . . 13-1
7
1. SPECIFICATIONS
1-1
1.1 Main Specifications
Item Specifications
Processing Capacity 22 rolls/hour (135 film 24 exposures, or IX240 film 25 exposures, 2 lanes)
CN-16Q
Temperature
Process Step Code Time
(°C/°F)
Color Developer N1 38 ± 0.2/100 ± 0.3 3 min. 19 sec.
Bleach N2 35 – 41/95 – 105 1 min. 10 sec.
Bleach-Fix N3 35 – 41/95 – 105 3 min. 12 sec.
Processing Stages
Super Rinse 1 NS-1 32 – 38/90 – 100 1 min. 31 sec.
Super Rinse 2 NS-2 32 – 38/90 – 100 1 min. 46 sec.
Stabilizer N4 32 – 38/90 – 100 42 sec.
Dry DRY 40 – 65/104 – 149 3 min. 00 sec.
Total 14 min. 40 sec.
Waste solution Tanks Built-in type, with solution level sensors, 5-liter capacity x 2
1-2
Item Specifications
Character liquid crystal display Display items : Pre-checks, post-checks, film types, error alarms, number of
Display rolls processed,
processing solution temperatures, film loading display, etc.
FP232B UL FP232B CE
Voltage Single-Phase AC Single-Phase AC 220/230/240 V
Power Requirements
200/210/220/230/240 V 60 Hz 50 Hz
Line Frequency
Dimensions
1088 mm x 563 mm x 1170 mm (42.83 x 22.17 x 46.06 in)
(L x W x H)
Weight Unit weight approx. 130 kg (287 lbs)/Running weight approx. 200 kg (441 lbs)
1-3
1.2 Design
1.2 Design Specifications
Specifications
Item Specifications
Temperature setting Standard Setting range Temperature alarm Heater alarm Sensor alarm
range and alarm N1 38.0 °C 30 to 40 °C Setting ± 1 °C 1 °C/10 min. 51.1 °C
monitoring range N2 37 °C 30 to 40 °C Setting ± 7 °C 1 °C/10 min. 51.1 °C
N3 37 °C 30 to 40 °C Setting ± 7 °C 1 °C/10 min. 51.1 °C
NS-1 35 °C 30 to 40 °C Setting ± 7 °C 1 °C/10 min. 51.1 °C
NS-2 35 °C 30 to 40 °C Setting ± 7 °C 1 °C/10 min. 51.1 °C
N4 35 °C 30 to 40 °C Setting ± 7 °C 1 °C/10 min. 51.1 °C
DRY (135) 50 °C 40 to 70 °C Setting ± 20 °C 5 °C/ 5 min. 102.3 °C
DRY (120) 60 °C and 35 to 75 °C
Capacity N1 360 W
N2
Common 360 W
N3
NS-1
NS-2 Common 360 W
N4
Processing solution Type Self-recovery type mounted on processing solution heater aluminum body surface
heater safety thermostat
Open
64° ± 3 °C
temperature
Close
42° ± 3 °C
temperature
1-4
Item Specifications
Water replenishment Evaporation rates Evaporation rates Evaporation rates Water replenishment
condition (standard during (standard during (standard power OFF) rate correction
stand-by) processing)
Output N1-R
N2-R
N3-R 50 [Hz] : 50 to 60 ml/30 sec.
NS-R 60 [Hz] : 60 to 72 ml/30 sec.
N4-R
N1W
N2W
System 3
Setting + 1.0 °C Setting + 0.5 °C
(NS-1, NS-2, N4)
Dryer heater Capacity 1000 W
Voltage 240 V ( )
Fan Amount
50 Hz 2.3 cubic meters/min (81 cubic foot/min)
of air
60 Hz 2.6 cubic meters/min (92 cubic foot/min)
expelled
Voltage AC100 V
Open
temperature 85 °C ± 5.6 °C (185 °F ± 10 °F)
1-5
1.2 Design Specifications
Item Specifications
50/60 Hz line 50 Hz 60 Hz
frequency Drive motor gear head “300k” mark “360k” mark
selection circulation pump N1 orifices 5 mm I.D. 4.5 mm I.D.
1-6
2. OUTLINE AND OPERATION SEQUENCES
2-0
2.7 Operation Sequences of Drive System . . . . . . . . . . . . . . . . . . . . . . . . 2-22
2.7.1 Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
2.7.2 Drive Motor Shutdown Timing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
2-1
2.1 Operation Sequences of Loader
2.1.1 Outline
· Auto locking of loader cover to shield the inside of the loader from light to prevent film fogging.
· Detection of type and length of film (information used to replenish processing solutions and dry film).
Film trailing end cutting (135 size only) Film length identification
2-2
2.1.2 Film Supply/Auto Lock
START
Load film leader and film into loader. Load only film leader into loader.
Film trailing end passes by film sensor. 60 seconds later (Film sensor OFF)
45 seconds later
END
2-3
2.1 Operation Sequences of Loader
This sequence serves to detect films and their sizes in the left and right lanes of the film leader, and this information is relayed to the
CPU. The CPU then displays the information and proceeds with the processing solution replenishment and drying, both corresponding
to the size of the film.
Film Sensor
Left Lane Right Lane
START
Film sensor is ON
Film starts to pass by film sensor.
for 1 or more seconds.
1 second later
Film sensor ON
END
· Drying temperature is set to “120 drying temperature” 9 minutes after 120 film ON.
· Drying temperature is set back to “normal drying temperature” 7 minutes and 30 seconds after 120 film OFF.
· The mechanical counter of the electrical equipment section is increased by 1 when the identification of the film type has been
completed.
2-4
2.1.4 Film Trailing End Cutting (135 Size)
In this operation sequence, the trailing end of the 135 mm film is detected and cut at a constant position.
START
YES
E302
Alarm
E303
END
2-5
2.1 Operation Sequences of Loader
In the event of a power failure during film processing, the cutter is operated once if the film leaders and films are transported from the
loader to the processor using the manual handle :
START
END
2-6
2.1.6 Film Sensor OFF
This operation sequence serves to control the steps from after the film has been transported from the loader until the next film can be
installed.
STRAT
Film sensor
Film trailing end passes by film sensor.
OFF
1 second later
45 seconds later
END
2-7
2.2 Outline
2.2 Operation of ofProcessing
Sequences Solution
Automatic Loader (FP362B AL) Circulation System
The processing tanks are formed by molding, and nozzles for discharging the processing solutions and for circulation are provided at the
very bottom of each tank. These nozzles are connected to the drain valves and processing solution circulation pumps.
N2
N1
NS/N4 Heater
Subtank
N1 Heater
N2/N3 Heater
N1 Cooling Fan
Waste Solution
N1 Radiator Level Detector
Circulation Pump
2-8
Concentration Correction Addition
N1-W N2-W
Chain Drain
Color
Developer Bleach Bleach-Fix Super Rinse Super Rinse Stabilizer
N1 N2 N3 NS-1 NS-2 N4
W1 W2
2-9
2.2 Outline of Processing Solution Circulation System
The temperature of the CN-16Q processing solutions can be set to between 30 ˚C and 40 ˚C for each processing tank, but the
temperature which will chemically guarantee its performance are indicated in the table below.
· Circulation filter
· Level detector
· Temperature detector
· Water filling nozzle (N1, N2 only)
· Replenisher nozzle (N1, N2, N3, NS-2, N4 only)
The processing solution which has been filtered by the circulation filter in the temperature control tank is returned to the bottom of the
processing tank via the radiator (N1 only) and the processing solution heater. The N1 line radiator and NS and N4 line heater are
provided with a cooling fan to cool the processing solution and enable the temperature to be controlled even when the room temperature
is high.
Circulation Filter
Cooling Fan
(for NS and N4)
Processing Rack
Subtank
Cooling Fan
Solution Flow
(for N1)
Radiator
Processing Tank
(for N1)
Recirculation Pump
F595
2-10
2.2.4 Processing Rack Roller Arrangement
Hard white
Flat tube
Cloth
Hard black
Squeegee
F298
2-11
2.3 Operation Sequences of Processing Solution Circulation System
2.3.1 Outline
There are four operation sequences for temperature control : warm-up, normal operation, heat save operation and cool-down operation.
The diagram below shows the relationship between these sequences.
Up mode
Warm-up
operation
Preheat operation
N1 solution temperature
N1 temperature
setting
N1 heat save
operation temperature
setting (difference
from setting
temperature) or
temperature of N2/N3
solutions is 35 ˚C, NS-1/
NS-2/N4 solutions is
2-12
2.3.2 Warm-up Operation
Objective The warm-up operation is designed to enable the temperature of the solutions in the six processing tanks to reach the
temperature settings after the power has been switched on or the processing solution heater started up.
Start condition Temperature control starts under the following conditions from the no temperature control mode.
Power ON operation
Processing solution heater startup operation (timer-activated start-up/instruction given by operator)
Solution The solution temperature is monitored at 1-second intervals.
temperature
monitoring
interval
Mode Processing
Condition solution heater Circulation pump Cooling fan
Objective Regular operation is designed to keep the temperature of each solution at its temperature setting after the warm-up
operation.
Start condition The warm-up operation has been completed.
The condition for terminating the heat save operation has been satisfied during the heat save operation
(see subsection 2.3.4).
Solution 1 second
temperature
monitoring
interval
Condition Mode Heater Circulation pump Cooling fan
The solution temperature is lower than the ON ON OFF
temperature setting
2-13
2.3 Operation Sequences of Processing Solution Circulation System
Objective The heat save operation is designed to keep the solution temperature at the heat save temperature in order to
achieve the following objectives.
Reduction of the processing solution heater operating time to conserve electric power
Shutdown of the operation of all circulation pumps to prevent deterioration of the processing solutions
Objective The preheat operation is designed to keep each solution temperature above a prescribed level when it might
otherwise fall below freezing point, such as at night or at other times when the power is OFF.
The operating time can be set once or twice in units of two hours.
Start condition The operating time assigned by “Pre-Heat Timer Set” has arrived.
Solution 1 second
temperature
monitoring
interval
Mode Processing
Condition Circulation pump Cooling fan
solution heater
The solution temperature is less than 25 ˚C ON ON OFF
The solution temperature is more than 25 ˚C OFF ON OFF
End condition The power switch is turned to STANDBY when the preheat end time (2 hours) is reached.
Operation transfers to the warm-up operation when the power switch is turned ON during preheating.
Operation transfers to the cool-down operation when any solution temperature enters an abnormal temperature
range.
2-14
2.3.6 Cool-down Operation
Objective The cool-down operation is designed to shut down the temperature control by the processing solution heater OFF
operation.
Start condition A warm-up, normal or heat save operation is in progress.
The processing solution heater shut down operation has been started.
Solution 1 second
temperature
monitoring
interval
Mode Processing solution heater Circulation pump Cooling fan
OFF OFF OFF
End condition The power switch is turned to STANDBY.
The processing solution heater OFF operation* is released.
The processing solution heater forced OFF operation is released.
* Solution heater OFF operation occurs: · When the solution level lowers.
· When solution temperature abnormality, solution temperature detector abnormality,
or solution heater abnormality occurs.
2-15
2.3 Operation Sequences of Processing Solution Circulation System
Level detector ON ?
(Waste solution tank is full) OFF
ON
Alarm
* The level is checked continuously until waste solution is processed (The alarm cannot be cleared until waste solution is processed.)
2-16
2.4 Outline of Processing Solution Replenishment System
Replenisher
Nozzles
Solution
Temperature
Detector
Processing Solution
Level Detector
Circulation Filter
N4-R
Replenisher NS-R
Tanks N3-R
N2-R
N1-R
Replenisher
Level Detector
Drain Valves
F368
2-17
2.5 Operation Sequences of Processing Solution Replenishment System
· The pump discharge volume shall be the amount of solution discharged when the pump is operated continuously for 30 seconds
under the control of the timer.
The figure below shows the relationship between the amount of film processed and the replenishment timing.
N1-R: Every
1.1 rolls
N2-R: Every
2.5 rolls
N3-R: Every
1.7 rolls
NS-R: Every
1.7 rolls
N4-R: Every
2.5 rolls
1 2 3 4 5 6 7 8 9 10 11 12 13 14
NOTE: · The figure above shows the replenishment timing when the pump discharge volume is 50 ml.
· The actual replenishment timing is corrected by changing the difference between the reference value and the actual discharge
volume measured by “Output check” into the number of film.
2-18
2.5.4 Replenisher Levels in Replenisher Tanks
The replenisher level detector detects a drop in replenisher level in one of the replenisher tanks drops, and alerts the operator with an
alarm.
The amount of replenisher remaining in each of the replenisher tanks when the alarm sounds is listed in the table below.
Normal mode:
· Energize the replenisher level detector (float switch) for 1 second once every 5 minutes, and check the replenisher level.
· If the replenisher level is low, the level detector will be set ON and the operator will be alerted by an alarm.
(The alarm can be cleared by pressing the [YES] key.)
The alarm is repeated every 5 minutes until replenisher is added.
(The level is checked continuously until replenisher is added after the 2nd alarm.)
· Energize the replenisher level detector (float switch) for 1 second continuously, and check the replenisher level.
· If the replenisher level is low, the level detector will be set ON and the operator will be alerted by an alarm.
Then check the replenisher level continuously. (The replenisher alarm cannot be cleared until replenisher is added)
2.5.5 Constitution
NS-1 NS-2 N4
N2 N3
N1
FS8 FS9 FS10 FS11 FS12
NS-R
N1-R
N1W
N4-R
N2-R
N2W
Replenisher filter
N3-R
Replenisher pump Replenisher level detector
FS
2-19
2.6 Operation Sequences of Dryer
2.6.1 Outline
Temperature control of the dryer is characterized by the following modes : no control, warm-up, normal temperature control and
cool-down. These modes stand in the relationship illustrated below.
Dryer temperature
Low
Dryer temperature
(Monitored second by second)
Objective The warm-up operation is designed to enable the temperature of the dryer to reach the temperature setting after
the drive has started.
Start condition Temperature control starts under the following conditions from the no temperature control mode.
· Power ON
· Processing section and loader covers Closed
· Drive start ON
Temperature The dryer temperature is monitored at 1-second intervals.
monitoring interval
Mode Condition Dryer heater Dryer fan
Dryer temperature is lower than drying ON ON
temperature setting
Dryer temperature has reached drying OFF ON
temperature setting
End condition The warm-up operation is terminated when the dryer temperature has reached its setting.
2-20
2.6.3 Normal Temperature Control Mode
Objective Normal temperature control is designed to keep the temperature inside the dryer at its setting after the warm-up
operation.
Start condition This mode is established as soon as the warm-up operation is completed.
Temperature The temperature inside the dryer is monitored at 1-second intervals.
monitoring interval
Mode Condition Dryer heater Dryer fan
Temperature inside the dryer is lower than ON ON
the drying temperature setting
Temperature inside the dryer is higher than OFF ON
the drying temperature setting
End condition Under the following conditions, the normal temperature control mode is terminated and the cool-down operation
starts.
· Drive stops (the dryer heater is set OFF as soon as drive stops and the dryer fan is turned OFF after 60 seconds).
· Power has been turned OFF or the heater OFF operation has started.
The dryer heater OFF operation starts under the following conditions.
· Drive stops after the auto-shut off period.
· The emergency stop switch is pressed.
· Drying temperature is abnormal, dryer heater is abnormal or abnormally overheated, or drying temperature
sensor is abnormal.
Objective The cool-down operation is designed to shut down the temperature control by the heater OFF operation.
Start condition · Drive has stopped in the normal temperature control mode.
· The power OFF or heater OFF operation has been started in the normal temperature control mode or during the
warm-up operation.
Temperature The temperature inside the dryer is monitored at 1-second intervals.
monitoring interval
Mode Condition Dryer heater Dryer fan
For 60 seconds after operation has moved to OFF ON
cool-down
60 seconds after operation has moved to OFF OFF
cool-down
End condition This mode is terminated when the dryer fan is turned OFF.
2-21
2.7 Operation Sequences of Drive System
2.7.1 Functions
· The power generated by the drive motor is reduced by the gearbox and transmitted to the loader and processing and drying racks
through the drive chain and sprocket to transport the film leaders and films at a constant rate.
· When a power failure occurs, the film being processed can be saved by rotating the drive shaft using the manual crank handle at a
rate of approximately 13 seconds per turn. (During this operation if power has come back on, “FORCED DRIVE” function may be
selected and resume powered drive.)
· When the operator has installed the film leader and films into the loader after the “Ready For Processing” message has appeared,
and has closed the loader cover and pressed the [START] key.
· When the operator has selected “YES” for “Test Drive” in the pre-check mode and has pressed the [YES] key, then set the leader,
closed the loading section cover and pressed the [START] key.
· When the operator has on the “Process Control Strip” screen, installed the film leader and films into the loader, closed the loader
cover and pressed the [START] key.
· When the next film leader/film is not installed within 17 min after the drive motor has been turned ON.
· When the next film leader/film is not installed within 17 min after the film leader/film has passed through the loader.
· When the operator has pressed the emergency stop switch.
· When the operator has opened the processor section cover.
2-22
2.8 Electrical Equipment
Common Grounding
Line Side Side Wire
F580
Circuit Breaker
2-23
2.8 Electrical Equipment
continued
SSR
F577
Voltage Used
2-24
2.8.2 CTL32 Circuit Board
CN6 CN7 (CN12) CN10 CN4 (CN9)
RF
GND3 RP
RS
(RB)
HMD LF
BZ N3 (DIP SW) LP CN8
TMP LS
N1 N2 OFF ON
LB
CN3
GND4
DRY N4
NS1 NS2
VR1
GND2
LED
CN5
CN13
5V
JP1
ON
GND1
CN2 CN11
F346
2-25
2.8 Electrical Equipment
continued
Not Used 5 - -
Not Used 6 - -
Not Used 7 - -
Demo Mode 8 NA A
2-26
2.8.3 PWA32/PWB32 Circuit Board
CN26 CN20 CN33 CN34 CN29
GND +12V
(+5) +5V GND
CN32 (+12)
COM
CN25
C2 C1
+LC
+RC CN27
CN24 SW
(CN36)
CP17 CP18
CN35
CN21 CP15
CN31
CP19
CP16
(U1) CN30
(1L) (2L)
CP12 CP11 CN38
(U2)
CN28 (V)
(2N) CP13
CN22 (1N) CP14
CN23 CN39
F348
CP18 Circuit protector Cutter solenoid (SOL2 and 3), power OFF cutter operating position detectors
(S8 and S9)
CP19 Circuit protector Cooling fans (F1 and F3), exhaust fan (F4), dryer fan (F10)
2-27
2.8 Electrical Equipment
(CN3)
RV51
FU1
CN1
F347
2-28
3. MAINTENANCE SCHEDULE
3-1
3.1 Routine Maintenance Schedule
In order to keep the FP232B operating at peak performance, conduct regular maintenance and inspections as indicated in the table
below.
(NOTE) : Replace these solutions when they have not been used for 1 week or more or when, as a result of processing the
control strips, they are considered to be unfit for development processing.
3-2
3.2 Regular Maintenance Schedule
To keep the FP232B in peak working order, perform the inspection, cleaning or replacement procedures listed below at the interval
specified.
Every 1
Every year Every 3 years Reference section
month
Cutter blades (NOTE) Replace 7.4.3
Loader cover detector Inspect 7.2.1
Press roller operating position detector Inspect 7.4.5
Cutter operating position detector Inspect 7.4.6
Power OFF cutter operating position detector Inspect 7.4.7
Auto lock release switch Inspect 7.3.5
Film sensors Inspect 6.4.1
Auto lock solenoid Inspect 7.3.7
Film press roller solenoid Inspect 7.2.3
Cutter solenoid Inspect 7.4.8
Processing rack Clean Instruction Manual 11.6
Conveyor belt Inspect Replace –
Rollers Inspect –
Cooling fan Clean –
Dryer rack Clean –
Drive motor Inspect 12.2.1
Circulation pump Inspect 9.3.1
Replenisher pump Inspect 5.5.1
Temperature control heater Inspect –
Dryer heater Inspect –
Processing section cover detector Inspect 8.4.2
Processor temperature detector Inspect –
Dryer temperature detector Inspect 11.3.3
Processing solution level detector Inspect 9.2.1
Replenisher solution level detectors Inspect 10.2.1
Waste solution level detectors Inspect 9.4.1
Dryer fan Clean 11.3.6
Flow meter Inspect –
Circuit board, power supply & fuses Inspect –
Room temparature/humidity detector Inspect 13.3.14
(NOTE) : As a general guideline, the cutter blade must be replaced once a year. If large quantities of film are being processed,
the cutter will become dull in less than a year.
3-3
4. ERROR MESSAGES
4-1
4.1 General Description
4.1.1 Outline
The FP232B comes with self-diagnosis functions to deal with errors that may conceivably occur. When an error occurs, an alarm sounds
and an error message appears on the display.
When this happens, first check the “Error Messages” on the following pages to get the reference sections relevant to the error number.
Then proceed with the troubleshooting, and inspect and repair the machine.
! WARNING
· When engaged in inspection or repair work inside the machine during troubleshooting, make absolutely sure that the power switch is
first turned to STANDBY, and the built-in circuit breaker and power distribution switch are turned OFF.
The machine may sustain damage and operating personnel may suffer injury if the machine is activated while this work is being done
because the power was not switched off.
· The capacitors (C1 and C2) for cutting the end of the 135 film in the event of a power failure must be discharged before working on the
electrical equipment section (see subsection 13.3.2).
5313
· Connectors (male and female) which are connected to each other have numbers marked on them or on their harness ends.
Connectors with the same numbers must always be connected together.
· Plastic connectors are provided with locking catches at the plug (male) end. These catches must first be loosened before
disconnecting them. When re-connecting them, tighter them until they are locked properly.
Connectors
Catch
0305
4-2
4.2 Error Messages E000 to E201
The waste solution level detector (float switch) is energized once every 5
minutes to check the level except during water addition. If it is ON, the error
message is displayed. When the detector is continuously ON, the error
message can not be cleared until the solution level detector is OFF.
E000 Drain W1 Waste Tank. If this message is displayed before water addition, water can not be added until 9.1.1
the solution level detctor is OFF.
If this message is displayed before water adding to the N1 or N4 under the
water adding mode, water can not be added to the N1 or N4 until the solution
level detector is OFF.
The waste solution level detector (float switch) is energized once every 5
minutes to check the level except during water addition. If it is ON, the error
message is displayed. When the detector is continuously ON, the error
E001 Drain W2 Waste Tank. message can not be cleared until the solution level detctor is OFF. 9.1.1
If this message is displayed before water addition or under the water additing
mode to the NS or N2, water can not be added to the NS or N2 until the
solution level detctor is OFF.
E002 N1 Replenisher Tank Empty. The replenisher level detector (float switch) is energized once every 5 minutes 10.1
to check the level. If it is ON, the error message is displayed. When the
detector is continuously ON, the error message can not be cleared until the
E003 N2 Replenisher Tank Empty. 10.1
solution level detctor is OFF.
E051 N1 Working Tank Level Low. Not Driving: When the solution level is low continuously 1 second, this error 9.1.2
message is displayed. This error message can not be cleared
until the tank is refilled.
E052 N2 Working Tank Level Low. 9.1.2
Operation returns to "M005 Solution Warming Up" after the
message is cleared.
E053 N3 Working Tank Level Low. During Driving: When the solution level is low continuously 5 seconds, this 9.1.2
error message is displayed. This error message can not be
cleared during driving even if the tank is refilled. After the drive
E055 NS1 Working Tank Level Low. 9.1.2
is stopped, the error message can be cleared if the tank is
refilled.
E056 NS2 Working Tank Level Low. Operation returnes to "M005 Solution Warming Up" after the 9.1.2
message is cleared.
E057 N4 Working Tank Level Low. 9.1.2
If the processing section cover is opened when the [START] key is pressed, this
E200 Close Processing Section Cover. error message is displayed. 8.1.1
This error message can be cleared with the [YES] key.
If the loader cover is opened during the film sensor is ON, this error message is
E201 Close Film Load Section Cover. displayed. 7.1.1
The message is cleared when the loader cover is closed.
4-3
4.2 Error Messages
Reference
Error Messages Description
Section
E202 Processing!! Turn ON Power When the power switch is turned to STANDBY during driving, this error
Switch. message is displayed. 13.1.1
It is cleared when the power switch is turned ON.
When the drive stop switch is pressed during processing, this error message is
displayed after drive is stopped.
E203 Processing!! Press START Key The message is cleared and drive is re-started when the [START] key is 13.1.2
to Start. pressed.
The "E210" appears after re-starting.
When the processing section cover is opened during driving, this error
E204 Processing!! Close Processing message is displayed after drive is stopped. 8.1.2
Cover. The message is cleared and drive is re-started when the processing section
cover is closed.
E206 Control Strip Not Cut, Check If the film sensor is not turned OFF within 60 seconds after the cutter has been
Cutter Unit. operated with the DIP switch on the CTL32 circuit board is set to cut the control 7.1.2
strip, this error message is displayed.
When the power switch is turned to STANDBY with the processing section
E207 Open Processing Section cover is closed, this error message is displayed. 8.1.3
Cover. The message is automatically cleared when the power switch is turned ON.
The power is turned OFF when the processing section cover is opened.
When re-starting after forced drive stopping, or after the E203 is cleared, this
E210 Wait Until Display Shows error message is displayed. 12.1.1
[Ready For Proc.] In the specified time after the film sensor is OFF, this message is cleared
automatically.
E300 Right Lane 135 Film Not Cut If the cutter does not operate after the film sensor is OFF when loading the 135
Check. film on the right lane, this error message is displayed instead of "Ready for 7.1.3
processing".
E301 Left Lane 135 Film Not Cut If the cutter does not operate after the film sensor is OFF when loading the 135
Check. film on the left lane, this error message is displayed instead of "Ready for 7.1.3
processing".
When a film is in the loader section and the right cutter operating position
E302 R, Film Cutting Fail or detector is not OFF in spite of three times of right lane cutter operation, this
7.1.4
Mag. Guide Not Retrn. error message is displayed.
(Cutter operation=300 ms ON and 200 ms OFF)
When a film is in the loader section and the left cutter operating position
E303 L, Film Cutting Fail or detector is not OFF in spite of three times of left lane cutter operation, this
7.1.4
Mag. Guide Not Retrn. error message is displayed.
(Cutter operation=300 ms ON and 200 ms OFF)
E304 Right Lane Film Length When the right lane film sensor ON time exceeds the 400 seconds, this error
7.1.5
Abnormal. message is displayed.
E305 Left Lane Film Length When the left lane film sensor ON time exceeds the 400 seconds, this error
7.1.5
Abnormal. message is deisplayed.
4-4
E202 to E305/E306to E409
E306 Right 110 Film Cut, Check If the right lane cutter operates when loading the 110 film on the right lane, this
7.1.6
Magazine. error message is displayed instead of "Ready for Processing".
E307 Left 110 Film Cut, Check If the left lane cutter operates when loading the 110 film on the left lane, this
7.1.6
Magazine. error message is displayed instead of "Ready for Processing".
E308 Right 126 Film Cut, Check If the right lane cutter operates when loading the 126 film on the right lane, this
7.1.6
Magazine. error message is displayed instead of "Ready for Processing".
E309 Left 126 Film Cut, Check If the left lane cutter operates when loading the 126 film on the left lane, this
7.1.6
Magazine. error message is displayed instead of "Ready for Processing".
E311 Left 120 Film Cut, Check If the left lane cutter operates when loading the 120 film on the left lane, this
7.1.6
Magazine. error message is displayed instead of "Ready for Processing".
E320 Right IX240 Film Cut, If the right lane cutter operates when loading the IX240 film on the right lane,
7.1.6
Check Intermed. Ctrdg this error message is displayed instead of "Ready for Processing".
E321 Left IX240 Film Cut, If the left lane cutter operates when loading the IX240 film on the left lane, this
7.1.6
Check Intermed. Ctrdg error message is displayed instead of "Ready for Processing".
E400 Right Magazine Guide Did Not If the right press roller operation position right detector is ON for more than 5
Return. seconds during driving, this error messsage is displayed after turning OFF the 7.1.7
press roller.
E401 Left Magazine Guide Did Not If the left press roller operation position right detector is ON for more than 5
Return. seconds during driving, this error message is displayed after turning OFF the 7.1.7
press roller.
E402 Right Perforation Sensor If the right perforation sensor is ON when the [START] key is pressed during
7.1.8
Abnormal. the drive is OFF, this error message is displayed.
E403 Left Perforation Sensor If the left perforation sensor is ON when the [START] key is pressed during the
7.1.8
Abnormal. drive is OFF, this error message is displayed.
If the right film sensor is ON when the [START] key is pressed during the drive
E404 Right Film Sensor Abnormal. 7.1.8
is OFF, this error message is displayed.
If the left film sensor is ON when the [START] key is pressed during the drive is
E405 Left Film Sensor Abnormal. 7.1.8
OFF, this error message is displayed.
E406 R. Perforation or Film Sensor If the right perforation sensor is ON when the right film sensor is OFF during
7.1.9
Abnormal. the drive is OFF, this error message is displayed.
E407 L. Perforation or Film Sensor If the left perforation sensor is ON when the left film sensor is OFF during the
7.1.9
Abnormal. drive is OFF, this error message is displayed.
E409 Left 120 or Film Sensor If the left 120 sensor is ON when the left film sensor is OFF during driving, this
7.1.9
Abnormal. error message is displayed.
4-5
4.2 Error Messages
If the left 120 sensor is ON when the [START] key is pressed during the drive is
E411 Left 120 Sensor Abnormal. 7.1.8
OFF, this error message is displayed.
If the right IX240 sensor is ON when the [START] key is pressed during the
E412 Right IX240 Sensor Abnormal. 7.1.8
drive is OFF, this error message is displayed.
If the left IX240 sensor is ON when the [START] key is pressed during the drive
E413 Left IX240 Sensor Abnormal. 7.1.8
is OFF, this error message is displayed.
E414 Right IX240 or Film Sensor If the right IX240 sensor is ON when the right film sensor is OFF during driving,
7.1.9
Abnormal. this error message is displayed.
E415 Left IX240 or Film Sensor If the left IX240 sensor is ON when the left film sensor is OFF during driving,
7.1.9
Abnormal. this error message is displayed.
E500 N1 Temperature Sensor If solution temperature sensor reading exceeds 51.1 °C, this error message is
9.1.3
Abnormal. displayed. If the drive is OFF, the 200V power supply for solution heater is
turned OFF. This error message can not be cleared until the power switch is
E501 N2 Temperature Sensor turned OFF.
9.1.3
Abnormal.
If dryer temperature sensor reading exceeds 102.3 °C, this error message is
E507 DRYER Temperature Sensor displayed. If the drive is OFF, the 200V power supply for solution heater is 11.1.1
Abnormal. turned OFF.
This error message can not be cleared until the power switch is turned OFF.
E508 Thermometer (Room) If room temperature detector reading exceeds 51.1°C, this error message will
Abnormal. be displayed one at the next water replenishment.
E509 Hygrometer (Room) If room humidity detector reading is 100% RH, this error message will be
Abnormal. displayed one at the next water replenishment.
If the N1 solution temperature does not increase more than 1°C within ten
minutes during the N1 warming up, this error message is displayed. When the
E600 Heater 1 (N1) Abnormal. solution level is low (during E051 is displayed) or after warming up has been 9.1.4
completed, this check is not done. This error message can not be cleared until
the power switch is turned OFF.
4-6
E411 to E600/E601 to E706
If the N2, N3 solution temperature does not increase more than 1°C within ten
minutes during the N2 or N3 warming up, this error message is displayed.
If the N2 or N3 solution temperature does not increase more than 0.2°C
E601 Heater 2 (N2/N3) Abnormal. within ten minutes after either the N2 or N3 has been warmed up. 9.1.4
When the solution level is low (during E052 or E053 is displayed) or after
warming up has been completed, this check is not done. This error message
can not be cleared until the power switch is turned OFF.
If the NS1, NS2, or N4 solution temperature does not increase more than 1°C
within ten minutes during the NS1, NS2, or N4 warming up, this error message
is displayed.
E604 Heater 3 (NS1/NS2/N4) If the NS1, NS2, or N4 solution temperature does not increase more than 0.2°C
Abnormal. within ten minutes after either the NS1, NS2, or N4 has been warmed up. 9.1.4
When the solution level is low (during E055, E056, or E057 is displayed) or
after warming up has been completed, this check is not done. This error
message can not be cleared until the power switch is turned OFF.
If the dryer temperature does not increase more than 5°C within five minutes
after the drive is ON, this error message is displayed. When the solution level is 11.1.2
E607 DRYER Heater Abnormal. low (during E051 is displayed) or after warming up has been completed, this
check is not done. This error message can not be cleared until the power
switch is turned OFF.
E701 N2 Temperature Abnormal. If the solution temperature exceeds the setting temperature ±7°C for 30 9.1.6
seconds after warming up has been completed, this error message is
displayed.
E702 N3 Temperature Abnormal. 9.1.6
Then the power supply for solution heaters is turned OFF during the drive is
OFF and the temperature exceeds 45°C.
E704 NS1 Temperature Abnormal. This error message can not be cleared until the power switch is turned OFF. 9.1.6
4-7
4.2 Error Messages E707 to E913
If the dryer temperature does not increase more than 45°C within 9 minutes
E708 DRYER Temperature
after the drive starts, this error message is displayed. 11.1.4
Increasing.
When it reaches to 45°C, the error message is automatically cleared.
E902 Water Replenishment When communication error occurs during communicating to the host computer,
10.1.2
Not Working. this error message is displayed.
The setting data are checked every 6 minutes during the drive is OFF. If they
exceed the upper or lower limit, this error message is displayed. Then the
check is discontinued but the error flag can not be cleared. The error message
can be cleared by pressing the [YES] key, however, it is displayed again when
the drive is attempted to start.
When the program is not running properly, reset the CPU and run the program
again. When the program error has occurres, this error message is displayed,
then the data initialization is performed. The data is reloaded from the
E913 System Malfunction
EEPROM.
Initialized Data.
If the drive is ON, the message is cleared by pressing the [YES] key and drive
is automatically re-started. The “Forced Drive is ON” message appears after
re-starting.
4-8
5. CONDITION SETTINGS
5-1
5.1 Menu
5.1 Menu Table Table
MODE
MODE + 4
Present Time
M400 Machine Condition Settings M410 Present Time Set ?
YES YES
Setting 1 ? Y/N Y/N
5-2
M411 Present Time (###)
MODE
YYYY/MM/DD HH:MM ?
YES
YES
M431
YES
5-3
5.1 Menu Table
MODE + 5
Processing Temperature
M500 Machine Condition Settings M510 Temperature Set ?
YES YES
Setting 2 ? Y/N Y/N
Replenishment Rate
Settings
M520 Replenishment
SUB 2 YES
Rate Set ? Y/N
Replenishment Pump
Output Check
M530 Pressent Pump
SUB 3 YES
Output Data ? Y/N
Replenishment Pump
Output Measurement/
Settings
M540 P.Output Check ? Y/N
SUB 4 YES
Press START key to Start
MODE + 6
Emergency Processing
M600 Machine Condition Mode M610 Emergency
YES YES
Setting 3 ? Y/N Process Mode ? Y/N
5-4
N1:##.# N2/N3:## DRY110/126/135/240:##
YES MODE
NS1/NS2/N4:## (˚ C ) ? DRY120:## (˚C)?
M511 M512
YES
YES
M521 M522
YES
M531 M532
YES
M541 M542
M007 Ready for Process.
YES
N 1 : # # . #˚ C
M621
5-5
5.2 Timers
NOTE: If the present time is not set correctly, the calender 8) Using the [YES] key, move the cursor to the
timer and preheat timers will not activate the system minute position.
at the correct time, and the error messages will not
show the correct date or time. 9) Using the numeral keys, change the minute
display to the present minute.
1. Press the [MODE] key.
10) Press the [YES] key.
2. Press the [4] key while holding the [MODE] key down. 6. Verify the present time and press the [YES] key to
return to “M420 Next ON Time Setting ?” screen or
press the [MODE] key to return to the mode setting
M400 Machine Condition screen.
Setting 1 ? Y / N
5. Verify the set date and time, perform the settings with
the following procedures.
5-6
5.2.2 Weekly Timer Settings
The scheduled work days and the timer ON times can be <Example>
preset in the weekly timer one week in advance.
Once the weekly timer is turned ON, the processing solution The screen below shows an example of settings where the
circulation pumps and heaters are activated to raise the timer is set to come ON at 8:30 AM.
temperatures of the solutions by the scheduled work
commencement time and then to control the temperatures
within the set ranges. If the power switch is not manually M421 Next On Time
turned ON within 3 hours after the timer is turned ON, the
0 8 : 30 ?
system turns the main power supply off.
1. Press the [MODE] key. 1. Press the [YES] key after settings the timer ON time
on the “M421 Next ON Time” screen.
MOOB M o d e 0-9?
M430 Next On Day Setting ?
(1:ON 0:OFF) Y/ N
2. Press the [4] key while holding the [MODE] key down.
2. Press the [YES] key.
3) Enter the timer ON minute using the numeral 4. Press the [YES] key to move to “M440 Pre-Heat Timer
keys. Set ?” screen or press the [MODE] key to return to the
mode setting screen.
4) Press the [YES] key.
5-7
5.2 Timers
The preheat timers are used to prevent crystallization of the When the preheat timers are not going to be used:
processing solutions not being used.
This may be caused by falling temperatures in the winter 1) Since the cursor is positioned on the hour
months. position, press the [YES] key twice and enter “0”
Depending on the climatic conditions, the timers may by set (“OFF”).
to come ON either once or twice. By setting the preheat
timer(s), the heaters will by activated every day at the preset 2) Press the [YES] key.
times and stop operating two hours later.
When ceasing to use of the preheat timers:
NOTE: · Ensure that the preheat timer ON setting does not
coincide with the weekly timer ON setting. If it does, Since “1” (“ON”) is set, enter “0” and press the [YES] key.
the first set timer will by given priority and all the
power will be shut down at the OFF time.
For example, if the weekly timer setting is 5:00 and M441 Pre-Heat Timer 1
the preheat timer ON setting is 3:30, then at 5:30 (2
1:ON 0:OFF HH:MM ?
hours after the preheat timer ON time) all the power
will be shut down.
· Generally, the preheat timers are used when the • The Pre-Heat Timer 2 screen now appears.
room temperature is expected to drop below 10 °C
(50ÜV) at night when the system is not operated When the preheat timers are going to be used:
and when temperature control will not be exercised
for 8 or more hours. 1) Since the cursor is positioned on the hour
position, enter the hour using the numeral keys.
2 hours
ON OFF NOTE: The hours must be entered on a 24-hour
basis.
3. Press the [SUB] and [4] keys. 5. Set the timer on the Pre-Heat Timer 2 screen with the
same procedures as the timer 1.
5-8
<Example>
5-9
5.3 Processing Temperatures
N1:38. 0 N2/N3:37
NS1/NS2/N4:35 (˚ C ) ?
M511
DRY110/126/135/240:50
DRY120:60 (˚ C ) ?
M512
5-10
5.3.2 Save Mode Settings
All the heaters and circulation pumps will automatically stop 6. Verify the temperature range setting and press the
operating when “Ready for Processing” is displayed (drive [YES] key to move to “Replenishment Rate Set ?”
motor is not operating) once the save mode has been screen or press the [MODE] key to return to the mode
established. When the temperature of the N1 solution drops setting screen.
below the set temperature range or the temperature of N2/N3
solution drop below 35°C or NS-1/NS-2/N4 solution drops
below 33°C, the heating operation commences in order to
return the solution concerned to the operating temperature.
MOOB M o d e 0-9?
2. Press the [5] key while holding the [MODE] key down.
4. Using the numeral key, enter “1” (“ON”) and press the
[YES] key.
<Example>
5-11
5.4 Replenishment Rates
Standard Replenishment Rates 9. After verifying the replenisher rate settings, press the
[YES] key to move to “M530 Present Pump Output
Replenisher Standard replenishment Data ?” screen or press the [MODE] key to return to
the mode setting screen.
N1-R 45.0 ml
N2-R 20.0 ml
N3-R 30.0 ml
NS-R 30.0 ml
N4-R 20.0 ml
MOOB M o d e 0-9?
2. Press the [5] key while holding the [MODE] key down.
M520 Replenishment
Rate Set ? Y / N
N1R:45.0 N2R:20.0
N3R:30.0 (ml) ?
M521
6. Using the [YES] key, move the cursor and set the rates
for the “N2-R” and “N3-R” replenisher in sequence.
NSR:30.0 N4R:20.0
(ml) ?
M522
8. Using the [YES] key, move the cursor and set the rates
for the “NSR” and “N4R” replenisher in sequence.
5-12
5.5 Replenisher Pump Output
NOTE: When installing the machine, the replenishment 8. Open the sub-tank section cover, remove the N1-R
condition is set using the water filling. At this time, set replenisher discharge nozzle from the sub-tank and
the default values instead of measuring the output. insert it in the 100 ml measuring cylinder (accessory).
At this point, hold the nozzle at the same height as its
1. Press the [MODE] key. installation position.
M540 P.Output Check ? Y / N 11. Repeat steps 8 to 10 three times and obtain the
Press START key to Start average value.
5-13
5.5 Replenisher Pump Output
continued
<Example>
13. Using the [YES] key, move the cursor from one
position to the next, and measure and set the outputs
for replenishers “N2R”, “N3R”, “NSR” and “N4R” by
following steps 8 to 12.
N1W:50 N2W:50
(ml) ?
M542
15. Measure and set the outputs for N1W and N2W by
following steps 8 to 12.
16. Verify the settings and press the [YES] key to move to
the “M007 Ready for Process-” screen, or press the
[MODE] key to return to the mode setting screen.
5-14
5.6 Emergency Processing Mode
5.6.1 When the Processor Not 5.6.2 When the Processor Working
Working
Process films as they are, using the side with an active
An error message is displayed as follows (e.g. E411) in case sensor.
of sensor's abnormality.
E 4 1 1 L e f t 1 2 0 S e n s o r
A b n o r m a l Y
MOOB M o d e 0-9?
M610 Emergency
Process Mode ? Y / N
5-15
5.7 Error Log
MOOB M o d e 0-9?
2. Press the [6] key while holding the [MODE] key down.
<Example>
The above error log screen shows that the latest error (with
“#” mark) is E002 “N1 Replenisher Tank Empty” occurred at
17:21 on Sep 27.
5. Press the [YES] key to display the next errors and the
[BACK SHIFT] key to display the previous errors in
succession.
5-16
5.8 Room Temperature and Humidity Check
MOOB M o d e 0-9?
2. Press the [6] key while holding the [MODE] key down.
5-17
6. MAINTENANCE SERVICE MODE
6-0
6.12 Demonstration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-26
6.12.1 Demonstration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-26
6-1
6.1 Menu
6.1 Menu Table Table
MODE
MODE + YES
MODE + 7
Processing Temperature
Calibration
YES YES
Maintenance Mode 1
SUB 1
Temperature Correction
Value Check
SUB 2 YES
YES
Input Check Mode
SUB 3
BACK
SHIFT
YES YES
BACK BACK
SHIFT SHIFT
YES YES
BACK BACK
M738
SHIFT SHIFT
YES
MODE
BACK
SHIFT
M73C
1 2 3
6-2
YES MODE
M711 M712
YES
YES MODE
YES
YES YES
BACK BACK
M732
SHIFT SHIFT
YES YES
BACK BACK
M736 M737
SHIFT SHIFT
YES YES
BACK BACK
M73A
SHIFT SHIFT
MODE
M000
4
6-3
6.1 Menu Table
1 2 3
YES
YES
Output Check Mode
SUB 4
BACK
SHIFT
YES YES
BACK BACK
SHIFT SHIFT
YES
MODE
BACK
SHIFT
YES
M000
Setting Data Saving
SUB 5 YES
YES YES
BACK BACK
SHIFT SHIFT
YES YES
BACK BACK
SHIFT SHIFT
5 6
6-4
4
YES YES
BACK BACK
SHIFT SHIFT
YES YES
BACK BACK
SHIFT SHIFT
MODE
1:YES
1 YES
YES MODE
0:NO
YES YES
BACK BACK
SHIFT SHIFT
YES YES
BACK BACK
SHIFT SHIFT
1:YES
YES 1 YES
BACK YES
SHIFT 0:NO
6-5
6.1 Menu Table
5 6
MODE
MODE + 8
Maintenance Mode 2
SUB 1
YES MODE
M824
YES
MODE + 9
Water Replenishment
Log
YES YES
Maintenance Mode 3
SUB 1
Water Replenishment
Rate Measurement
SUB 2 YES
8 9
6-6
u
MODE
1 YES
0:NO 1:YES
YES
YES YES
M822
MODE
M831
YES
MODE
M911
START MODE
YES
MODE
M931
YES
MODE
M941
YES
MODE
M951
YES
10
6-7
6.1 Menu Table
8 9
Water Replenishment
Rate Correction
SUB 6 YES
Environmental Conditions
Settings
SUB 7 YES
Demo. Mode ( )
* SUB 9 YES
YES YES
YES YES
YES YES
YES YES
YES YES
MODE
NOTE (*): Demo Mode is available when DIP switch 8 (DS8) is ON.
6-8
10
MODE
YES
MODE
YES
MODE
M973
YES
YES YES
YES YES
YES YES
YES YES
YES YES
YES MODE
YES
YES MODE
YES MODE
6-9
6.2 Processing
6.2 Processing Temperature Calibration
Temperature Calibration
NOTE: Accurate calibration of the processing temperature is 8. Using the [YES] key, move the cursor from one
performed one hour after the “Ready for Processing” position to another, and measure and calibrate the
message appears. temperatures of the “N2”, “N3”, and “NS1” solutions by
following steps 6 and 7 above.
1. Press the [MODE] key.
Thermometer
3. Press the [7] key while pressing the [MODE] key. F489
NS2:35.0 N4:35.0 (˚ C )
M710 Temperature
DRY:35.0 ?
Calibration ? Y / N
M712
5. Press the [YES] key. 10. Following steps 6 and 7, measure and calibrate the
temperatures of the “NS2”, “N4” solutions.
N1:38.0 N2:37.0 (˚ C ) 11. Move the cursor to “DRY” by pressing the [YES] key
twice.
N3:37.0 NS1:35.0 ?
M711 NOTE: Definition of dry temperature : On the FP232,
the dry temperature is the temperature
6. Open the processor cover, immerse a thermometer to recorded by the dry temperature detector
a depth of approximately 50 mm (2 in) behind the which minimizes the difference between the
processing rack (toward the dryer) in the N1 machines. This criterion is chosen because it
processing tank, and measure and record the solution is difficult to measure the dry temperature
temperature. accurately.
NOTE: The thermometer reading will fluctuate as the 12. Remove the heater cover (see subsection 11.3.2).
heater is turned on and off, but the average
value of the readings is considered to be the
actually measured temperature.
6-10
13. Remove the screw and the dry temperature detector. 17. If the actually measured temperature and the
temperature displayed on the screen differ, rewrite the
NOTE: Do not disconnect the harness connector. temperature displayed by entering the measured
temperature value using the numeral keys.
Dry Temperature Detector Screw
18. Verify the setting and press the [YES] key to move to
“M720 Temperature Correct?” screen or the [MODE]
key to return to the mode setting screen.
F491
Thermometer F490
NS2:35.0 N4:35.0 (˚ C )
DRY:35.0 ?
M712
6-11
6.3 Temperature
6.3 Temperature Correction Correction
The temperature correction screen is on the next page of the 6. Press the [YES] key.
maintenance mode selection screen.
When the processing temperature calibration (see subsection (Initial setting)
6.2.1) is performed, the temperature correction factors are
changed automatically.
If the circuit board replacement or data reset is required, M722 NS2: 0.0 N 4 : 0 . 0˚ C
record the temperature correction factors and enter the
Dry: 0.0 Y
corrected values (temperature correction factor + displayed
temperature) on the M711 and M712 screens (see subsection
6.2.1) after the replacement or data reset. This eliminates the (Example after calibration)
processing temperature calibration procedures.
1. Press the [MODE] key. 7. Press the [YES] key to move to “M730 Input Check?”
screen or the [MODE] key to return to the mode setting
screen.
MOOB Mode 0-9?
3. Press the [7] key, while holding the [MODE] key down.
M720 Temperature
Correction ? Y / N
(Initial setting)
6-12
6.4 Input Checks
Sensor and detector function can be checked in the input NOTE: The dip switch numbers are allocated as
check mode. shown in the following.
Display the required Input Check screen by carrying out the
following steps :
DS ON OFF
Not used 1 - -
1. Press the [MODE] key.
Auto water
2 NA A
replenishment
MOOB Mode 0-9? Control strip auto cut 3 A NA
Room temp./
4 A NA
humid.detect
Not used 5 - -
2. Press the [YES] key while holding the [MODE] key Not used 6 - -
down. Not used 7 - -
Demo mode 8 A NA
M010 Maitenance Mode A : Applied
NA : Not applied
1-3? (Mode 7-9?)
6. Check the settings for “DS1” to “DS3” and press the
3. Press the [7] keys while pressing the [MODE] key. [YES] key.
Y / N DS-7:0 DS-8:0 Y
M732
4. Press the [SUB] and [3] keys.
7. Check the settings for “DS4” to “DS8” and press the
[YES] key.
M730 lnput Check ?
Y / N M733 Film Senaor Check
RSE1:0 RSE2:0 RSE4:0 Y
NOTE: The “Input Check Mode” can not be selected
during driving the processor.
NOTE: The displayed symbols signify the following.
5. Press the [YES] key. RSE1: Right film sensor
RSE2: Right perforation sensor
RSE4: Right IX240 sensor
M731 DlP Switch Check
8. Check the settings for “RSE1”, “RSE2” and “RSE4”
DS-1:0 DS-2:0 DS-3:0 Y
and press the [YES] key.
6-13
6.4 Input Checks
continued
9. Check the settings for “LSE1” to “LSE4” and press the 14. Press the [YES] key.
[YES] key.
NOTE: The previous screen is returned when the NOTE: The displayed value “0” signifies “normal” and
[BACK SHIFT] key is pressed. the “1” signifies “abnormal”.
10. Press the [YES] key. 15. Press the [YES] key.
11. Press the [YES] key. NOTE: The displayed value “1” signifies “normal” and the “0”
signifies “abnormal”.
S4:0 S5:0 S6:1 17. Upon completion of the input check, press the [MODE]
key to return to the mode setting screen. Or press the
(RCUT) (LCUT) (L D R C ) Y
[YES] key to move to the “M740 Output Check”
M737 screen.
S7:1
(PRSC) Y
M738
6-14
6.5 Output Checks
The function of the lamps, motors, solenoids, etc. can be 6. The previous screen is returned when the
checked in the output check mode. [BACK SHIFT] key is pressed, and every time the
Display the required Output Check screen by carrying out the [YES] key is pressed, the output check screens shown
following steps : below are displayed in succession.
6-15
6.6 Setting
6.6 Setting Data
Data Saving, Saving,
Reading and Loading Reading and Loading
3. Press the [7] key while holding the [MODE] key down.
M700 Maitenance Mode 1 ?
Y / N
4. Press the [SUB] and [6] keys.
1:Tes 0:No 0?
6. The previous screen is returned when the [BACK
SHIFT] key is pressed, and every time the [YES] key is
6. When saving data, enter “1” (“Yes”) using the numeral pressed, the setting data reading screens shown
keys and press the [YES] key. below are displayed in succession.
6-16
M764 Temp. Correction Y
N1:+0.7 N2:-1.3 N3:-1.4
6-17
6.7 Error
6.7 Error Log Clear
Log Clear
6-18
6.8 Replenishment Control Data
1. Press the [MODE] key. 9. Press the [YES] key to move to “M830 IX240
Replenishment Factor?” screen or the [MODE] key to
return to the mode setting screen.
MOOB Mode 0-9?
M820 Replenishment
Control Data ? Y / N
M821 Replenishment
Counter Y
M823 Replenishment
Count Date Y
N1:0274 N2:0616 N 3 :0 4 1 1
NS:0411 N4:0616 Y
M824
6-19
6.9 IX240 Replenishment Factor
3. Press the [8] key while holding the [MODE] key down.
6-20
6.10 Water Replenishment Log/Water Replenishment Rate Measurement
Day P-val. TS TD TO Y
M921 N1W:0053 N2W:0023
# 30 / 0 8 . 3 / 0 0 5 7 / 0 0 4 5 / 1 3 3 8
(ml) NSR:0082 N4R:0044 Y
M911
NOTE: The displayed symbols signify the following. 6. Select the pump to be measured by moving the cursor
P-val : Steam pressure calculated from room using the [YES] key and press the [START] keys.
temperature/humidity map (mmHg)
TS : STANDBY (rack drive stop) time (min.)
TD : Drive time (min.) M922 N1W Pump is Running
TO : Stop (power off) time (min.)
027Sec
NOTE: “#” mark is displayed on the latest log
in formation. • The water replenish rate is measured and the
W.Pump Output data screen now appears.
6. Press the [YES] key.
• Every time the [YES] key is pressed, the next log M921 N1W:0053 N2W:0023
information is displayed in succession.
(ml) NSW:0082 N4W:0044 Y
NOTE: Every time the [BACK SHIFT] key is pressed,
the previous log information is displayed in
succession. 7. After moving the cursor to the “Y” position, press the
[YES] key to move to “M930 Evaporation (Stand-by)?”
7. Press the [MODE] key to return to the mode setting screen or press the [MODE] key to return to the mode
screen. setting screen or press the [MODE] key to return to the
mode setting screen.
6-21
6.11 Water
6.11 Water Replenishment
Replenishment Condition Settings Condition Settings
3. Press the [9] key while holding the [MODE] key down.
M930 Evaporation ?
(Stand-by) Y / N
7. Use the [YES] key to move the cursor and enter the
“N2”, “NS” and “N4” evaporation rates in succession.
6-22
6.11.3 Evaporation Rates 6.11.4 Evaporation Rates
(During Processing) (During Power OFF)
Processing Standard evaporation rates during Processing Standard evaporation rates during
solution processing solution Power OFF
N1 11.5 ml/hour N1 3.7 ml/hour
N2 11.1 ml/hour N2 2.4 ml/hour
NS 22.0 ml/hour NS 6.4 ml/hour
N4 16.5 ml/hour N4 3.1 ml/hour
1. Display the Evaporation (Processing) rate setting 1. Display the Evaporation (Power OFF) rate setting
screen (see subsection 6.11.2). screen (see subsection 6.11.3).
3. Using the numeral keys, enter the “N1” evaporation 3. Using the numeral keys, enter the “N1” evaporation
rate during processing. rate during power OFF.
NOTE: When the system has just been installed, the NOTE: When the system has just been installed, the
standard evaporation rates during stand-by standard evaporation rates during power OFF
listed above are set as the default values. listed above are set as the default values.
4. Use the [YES] key to move the cursor and enter the 4. Use the [YES] key to move the cursor and enter the
“N2”, “NS” and “N4” evaporation rates in succession. “N2”, “NS” and “N4” evaporation rates in succession.
5. Verify the rate settings and press the [YES] key. 5. Verify the rate settings and press the [YES] key.
• “M950 Evaporation (power OFF) ?” screen is • “M960 water Replenishment Rate Correct ?” screen
displayed. is displayed.
6. Press the [SUB] key to return to the sub mode setting 6. Press the [SUB] key to return to the sub mode setting
screen or [MODE] key to return to the mode setting screen or [MODE] key to return to the mode setting
screen. screen.
6-23
6.11 Water Replenishment Condition Settings
Rate Correct. ? Y / N
3. Using the numeral keys, select the environment
condition.
2. Press the [YES] key.
NOTE: When the system has just been installed, “1
(Auto)” is set as the default condition.
M961 N1:030 N2:030 (ml)
! CAUTION
NS:060 N4:030 ?
Always select “1 (Auto)” unless a special room
condition applies.
3. Using the numeral keys, enter the “N1” water
replenishment correction rate. 4. Press the [YES] key to move to “M972 Room
Temperature Calibration?” screen or press the [MODE]
NOTE: When the system has just been installed, the key to return to the mode setting screen.
standard water replenishment rate corrections
listed above are set as the default values.
4. Use the [YES] key to move the cursor and set the
“N2”, “NS” and “N4” correction rates in succession.
6-24
6.11.7 Room Temperature Calibration 6.11.8 Circuit Board Adjustments for
Initial Installation
NOTE: When performing “Room Temperature Calibration”
right after “Environmental Conditions” (see NOTE: When the system is shipped from the manufacturing
subsection 6.11.6), step 1 can be skipped. plant, the DIP switch-2 (DS-2) is set to ON so that the
water replenishment system will not operate. When
1. Display the Room Temperature Calibration screen the system has just been installed, ensure that DS-2
(see subsection 6.11.6). is set to OFF by following the steps described below
after the water replenishment conditions have been
set.
M972 Room Temperature
1. Set the power switch to STANDBY, and the built-in
Calibration ? Y / N
circuit breaker and main power to OFF.
2. Press the [YES] key. 2. Remove the electrical equipment cover and discharge
capacitors C1 and C2 (see subsection 13.3.2).
Calib.:26.3 Corre.:-0.6 3. Set the DIP switch-2 (DS-2) on the CTL32 circuit board
from ON to OFF.
Disp.:26.9 (˚ C ) ?
M973 DIP Switch
6-25
6.12 Demonstration
6.12 Demonstration Mode Mode
1. Turn the power switch to STANDBY, and the built-in 9. Press the [SUB] and [9] keys.
circuit breaker and the main power supply OFF.
2. Remove the electric equipment cover (see subsection M990 Demo. Mode ?
13.3.1).
Y / N
3. Turn the Dip switch (DS-8) on CTL32 circuit board to
ON. 10. Press [YES] keys.
DIP Switch
E000 Drain W1 Waste Tank
Y
8. Press the [9] key while pressing the [MODE] key down.
E005 NS Replenisher
M900 Maintenance Mode 3 ?
Tank Empty Y
Y / N
6-26
E006 N4 Replenisher E301 Left Lane 135 Film
Tank Empty Y Not Cut Check Y
6-27
6.13 Default
6.13 Default Values/Hidden
Values/Hidden Running Total Running Total
Incorrect data in the CPU and RAM is cleared and the default Data Default
values in the ROM are set by conducting the following steps: Temperature setting N1 38.0 °C
37 °C
! CAUTIO
· These steps must be taken when replacing the CTL32
N2/N3
NS-1/NS-2/N4
DRY 110/126/
35 °C
6-28
6.13.3 Hidden Running Total
Data Default The hidden counter, different from the running total, can be
Evaporation (standby) N1 6.7 ml/h displayed by performing the following steps:
N2 6.4 ml/h
1. Press the [MODE] key.
NS 11.9 ml/h
N4 7.6 ml/h
Evaporation N1 11.5 ml/h MOOB Mode 0-9?
(processing) N2 11.1 ml/h
NS 22.0 ml/h
N4 16.5 ml/h
Evaporation N1 3.7 ml/h 2. Press the [2] key.
(power OFF) N2 2.4 ml/h
NS 6.4 ml/h
M200 Post-operational
N4 3.1 ml/h
Water replenishment N1 30 ml/h Check ? Y / N
rate correction N2 30 ml/h
NS 60 ml/h
3. Press the [9] key while holding the [SUB] key down.
N4 30 ml/h
Cumulative standby duration 0 min.
Cumulative processing duration 0 min. M290 Running Total ?
Cumulative power OFF duration 0 min.
Y / N
IX240 replenishment N1 58 %
factor N2 69 %
NS 69 % 4. Press the [YES] key.
N4 100 %
6-29
7. LOADER
7-0
7.4 Cutter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-40
7.4.1 Input Check of Press Roller and Cutter Operating Position Detectors . . . 7-40
7.4.2 Output Check of Cutter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-40
7.4.3 Cutter Blade Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-41
7.4.4 Cutter Operating Position in Power Failure Detector Inspection . . . . . . . . 7-42
7.4.5 Press Roller Operating Position Detector (S2/S3) . . . . . . . . . . . . . . . . . . . 7-42
7.4.6 Cutter Operating Position Detector (S4/S5) . . . . . . . . . . . . . . . . . . . . . . . . 7-43
7.4.7 Power OFF Cutter Operating Position Detector (S8/S9) . . . . . . . . . . . . . . 7-43
7.4.8 Cutter Solenoid (SOL2/SOL3) Replacement . . . . . . . . . . . . . . . . . . . . . . . 7-44
7-1
Parts Location
7.1 Troubleshooting
Magazine Guide
Feeding Roller
Auto Lock Release Switch (S11)
F563
7-2
7.1 Troubleshooting
Trouble: The loader cover was opened while film was being inserted.
Yes
Was the auto lock release switch pressed
when the loader cover was opened?
END
No
CN7 BrBr
10 1
Repair or replace the harness.
5315
OK
To Next Page
B : Black O : Orange
S6: Loader cover detector Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
: Action : Check
Gr : Gray Y : Yellow
7-3
7.1 Troubleshooting
continued
7-4
7.1.2 E206 “Control Strip Not Cut, Check Cutter Unit”
Trouble: During control strip processing the film sensor fails to go OFF within specified seconds after the cutter is operated.
No q
OK
To Next Page
: Action : Check
7-5
7.1 Troubleshooting
continued
O (Rh) Y (Lh)
CN27
W W
Repair or replace the harness.
0341
OK
No END
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
: Action : Check
Gr : Gray Y : Yellow
7-6
Open the processing section cover. Set
the power switch STANDBY/ON. Proceed
Yes
with “input check film sensor check”
selected in the maintenance mode. Is the
film sensor set OFF/ON?
(see subsection 7.5.1)
END
No
5306
OK
No END
7-7
7.1 Troubleshooting
Trouble: · The alarm will appear after the film sensor has been OFF. The cutter fails to operate even when the left-lane or right-lane
film being processed is 135 mm film.
· The cutter fails to operate when a film detection error occurs (Although the film being processed is IX240, 110 or 126, the
film sensor detects it as a 135 mm film).
· A 135 mm film is being processed using the 126 film magazine.
Yes
Is the 126 film magazine used?
No END
No
Rewind the film into the film cartridge and
re-process.
Remove the dark bag and open the No
loader cover. Does the magazine guide
move smoothly?
Yes
Remove the foreign matter or other obstruction
so that the guide moves smoothly.
E300 or E301 error may occur when the 135 damaged film rewound into the film cartridge for 120 is processed.
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
: Action : Check
Gr : Gray Y : Yellow
7-8
Is S4 or S5 mode to go OFF/ON with Yes
“input check” selected in the maintenance
mode? (see subsection 7.5.1)
No q
S4 or S5
5318
Yes
R (Rh)
O (Lh)
CN7
Repair or replace the harness.
1
10
R O
0349
OK
To Next Page
7-9
7.1 Troubleshooting
continued
0340
OK
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
: Action : Check
Gr : Gray Y : Yellow
7-10
Remove the electrical equipment cover
and discharge capacitors C1 and C2
(see subsection 13.3.2).
Disconnect the CN27 connector from the
PWA32/PWB32 board. Check the
continuity between the cutter solenoid
connector and the CN27 connector.
NG
O (Rh) Y (Lh)
W W CN27
Repair or replace the harness.
0341
OK
No END
7-11
7.1 Troubleshooting
continued
Yes
No
Interlock Detector
5319
Yes
7-12
Disconnect the CN29 connector from the
PWA32/PWB32 board and check the
continuity between the CN29 connector
and the connector of the interlock
detector.
NG
CN29
R R
R
R
OK
No END
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
: Action : Check
Gr : Gray Y : Yellow
7-13
7.1 Troubleshooting
7.1.4 E302, E303 “#. Film Cutting Fail or Mag. Guide Not Return.”
Trouble: The cutter operating position sensor fails to go OFF even though the left-lane or right-lane cutter has operated 3 times.
No
Open the loader cover. Does the
magazine guide move smoothly?
Yes
Remove foreign matter or other obstruction so
that the guide moves smoothly.
No
Is the cutter blade installed properly?
Yes
Install the cutter blade properly.
Yes END
7-14
Open the processing section cover. Set
the power switch STANDBY/ON. Is S4 or Yes
S5 made to go OFF/ON with “input check”
selected in the maintenance mode?
(see subsection 7.4.1)
Proceed to page 7-16 1
No
S4 or S5
5318
Yes
To Next Page
7-15
7.1 Troubleshooting
continued
R (Rh)
O (Lh)
CN7
1 Repair or replace the harness.
R O 10
0349
OK
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
: Action : Check
Gr : Gray Y : Yellow
7-16
Proceed with “output check” for the other No
cutter. Is the sound of the cutter
operating heard? (see subsection 7.4.2)
Proceed to 7-19 2
Yes
0340
OK
W W CN27
Repair or replace the harness.
0341
OK
To Next Page
7-17
7.1 Troubleshooting
continued
No END
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
: Action : Check
Gr : Gray Y : Yellow
7-18
From page 7-17 2
Yes
No
OK
To Next Page
7-19
7.1 Troubleshooting
continued
R
R
Repair or replace the harness.
5320
OK
No END
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
Gr : Gray Y : Yellow
7-20
7.1.5 E304, E305 “# Lane Film Length Abnormal”
Trouble: The length of the left-lane or right-lane film being processed is 1,787 mm (70.4 in).
No
To Next Page
: Action : Check
7-21
7.1 Troubleshooting
continued
5306
OK
No END
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
: Action : Check G : Green W : White
Gr : Gray Y : Yellow
7-22
7.1.6 E306, E307, E308, E309, E311 “#. #Film Cut, Check Magazine”
E320, E321 “# IX240 Film Cut Check Intermed. Ctrdg”
Trouble: The cutter operates when the left-lane or right-lane film being processed is not 135 film.
No
Is the film not 135 film?
Yes
Yes
No
END
To Next Page
: Action : Check
7-23
7.1 Troubleshooting
continued
5306
OK
No END
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
Gr : Gray Y : Yellow
7-24
7.1.7 E400, E401 “# Magazine Guide Did Not Return”
Trouble: The left or right press roller operating position detector remains ON for 5 seconds or more.
No
Open the loader cover. Does the
magazine guide move smoothly?
To Next Page
: Action : Check
7-25
7.1 Troubleshooting
continued
0367
OK
OK
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
Gr : Gray Y : Yellow
7-26
7.1.8 E402, E403 “# Perforation Sensor Abnormal”
E404, E405 “# Film Sensor Abnormal”
E411 “Left 120 Sensor Abnormal”
E412, E413 “# IX240 Sensor Abnormal”
Trouble: The sensor indicated in the message has been set ON when the [START] key is pressed to start the drive.
No
Clean the film sensor with the blower brush.
END
To Next Page
: Action : Check
7-27
7.1 Troubleshooting
continued
5306
OK
END
No
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
Gr : Gray Y : Yellow
7-28
7.1.9 E406, E407 “#. Perforation or Film Sensor Abnormal”
E409 “Left 120 or Film Sensor Abnormal”
E414, E415 “# IX240 or Film Sensor Abnormal”
Trouble: · The perforation sensor is ON although the left-lane or right-lane film sensor is OFF during operation.
· The 120 sensor is ON although the left film sensor is OFF during operation.
· The IX240 sensor is ON although the left-lane or right-lane film sensor is OFF during operation.
No
To Next Page
: Action : Check
7-29
7.1 Troubleshooting
continued
B Br
Br
WGr Bu Y R R
17 1
CN8
B B B Br
5306
OK
No END
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
Gr : Gray Y : Yellow
7-30
7.2 Loader Cover
7.2.1 Input Check of Loader Cover 7.2.2 Output Check of Film Press
Detector (S6) Roller (SOL4/SOL5)
1. Display the Input Check “S6” screen 1. Display the Press Roller output check screen (see
(see subsection 6.4.1). subsection 6.5.1).
M737
2. Open the loader cover.
2. Try opening and closing the loader cover. If the right or left press roller output system is
If detector S6 is operating properly, its display will operating properly, the right or left press roller will
change from 0 (OFF; open) to 1 (ON; close). operate for 1 second when the [1] or [2] key is pressed
and will not operate when the [0] key is pressed.
Loader Cover
Film Press Rollers
7-31
7.2 Loader Cover
7.2.3 Film Press Roller Solenoid (SOL4/SOL5) and Safety Thermostat (TH5)
Replacement
Removal 7. Disconnect the connector of the film press roller
solenoid.
1. Turn the power switch to STANDBY, and the built-in
circuit breaker and main power supply OFF.
3. Remove the cover stay arm E-ring and rest the loader
cover on the processor side.
F428
E-ring F426
Screws F429
Screws (2) Inter Lock Detector Cover Safety Thermostat Screws (2)
F427 F431
7-32
10. Remove the E-ring and take out the press roller, spring
and plain washers.
Installation
Bear in mind the cautionary notes given below and install the
parts in the reverse order of their removal.
7-33
7.3 Magazine
7.3 Magazine Support Support Plate,Cover
Plate, Loader Bottom Loader
and AutoBottom
Lock Cover and Auto Lock
7.3.1 Output Check of Auto Lock 7.3.2 Magazine Support Plate Cover
Removal/Installation
1. Display the Auto Lock output check screen
(see subsection 6.5.1). Removal
1. Turn the power switch to STANDBY, and the built-in
circuit breaker and main power supply OFF.
2. If the auto lock output system is operating properly, the 3. Remove the right and left cutter covers by removing
auto lock solenoid will operate for 1 second when the the two screws.
[1] key is pressed and will be turned OFF when the
[0] key is pressed. Left Cutter Cover Right Cutter Cover
The solenoid operation can be checked by movement
of the auto lock lever.
7-34
7.3.3 Loader Cover Detector (S6)
Replacement
6. Disconnect the harness connector of the loader cover
detector and remove the magazine support plate Removal
cover.
1. Remove the magazine support plate cover
Loader Cover Detector Connector (see subsection 7.3.2).
Installation
Install the parts in the reverse order of their removal. Magazine Support Plate Cover F613
NOTE: Take care not to catch the harness on the other parts.
Installation
7-35
7.3 Magazine Support Plate, Loader Bottom Cover and Auto Lock
F442
Installation
Install the parts in the reverse order of their removal.
E-ring F426
F457
7-36
7.3.5 Auto Lock Release Switch Inspection
1. Actuate the auto lock solenoid by conducting the auto 5) If the switch is operating normally, remove the
lock output check (see subsection 7.3.1). electrical equipment cover and discharge
capacitors C1 and C2 (see subsection 13.3.2).
2. Press the auto lock release switch. If the auto lock Inspect the continuity between the CN27
solenoid is released, the auto lock release switch is connector of the PWA32/PWB32 circuit board and
operating normally. the switch connector.
If there is no continuity, repair or replace the
• If the solenoid is not released, inspect the auto lock harness.
release switch by following the steps described
below.
White
4) While setting the auto lock release switch from
ON to OFF, check the continuity between the
switch terminals.
If continuity results with the switch OFF and does 5394
7-37
7.3 Magazine Support Plate, Loader Bottom Cover and Auto Lock
7.3.6 Auto Lock Release Switch (S11) 7.3.7 Auto Lock Solenoid (SOL1)
Replacement Replacement
Removal Removal
1. Remove the loader bottom cover 1. Remove the loader bottom cover (see subsection
(see subsection 7.3.4). 7.3.4).
2. Remove the screw, and remove the auto lock switch. 2. Remove the solenoid bracket by removing the four
screws securing the latch.
3. Disconnect the blue and white wires from the switch. Solenoid Bracket Screws (4)
Wires Screw
Latch F445
Auto Lock Switch F443
Install the parts in the reverse order of their removal. 6. Remove the lock nut, nut, screw and the hook.
Lock Hook Lock Nut Nut
Screw F447
7-38
Installation
7-39
7.4 Cutter
7.4 Cutter
7.4.1 Input Check of Press Roller and 7.4.2 Output Check of Cutter
Cutter Operating Position
Detectors 1. Display the Cutter output check screen (see
subsection 6.5.1).
1. Display the Input Check “S2/S3” or “S4/S5” screen
(see subsection 6.4.1).
M737
2. Open the loader cover and try moving the left or right
magazine guide by hand.
If the detectors are operating properly, they will change
from 0 (OFF) to 1 (ON) when pressed.
Magazine Guides
F576
7-40
7.4.3 Cutter Blade Replacement
Removal Installation
1. Turn the power switch to STANDBY, and the built-in Installation is essentially in the reverse order of removal,
circuit breaker and main power supply OFF. nothing the following.
2. Remove the electrical equipment cover and discharge NOTE: Align the holder fulcrum boss with the hole in the
capacitors C1 and C2 (see subsection 13.3.2). cutter base.
Holder (2)
Holder
7-41
7.4 Cutter
4. Open the loader cover and while pressing down on the Detector (S3) Detector (S2) Screws (2)
magazine guide with the fingers, check the continuity
between the wire terminals of the CN29 connector.
Orange Installation
Orange
Install the detector in the reverse order of removal.
Yellow
CN29
Yellow
5395
7-42
7.4.6 Cutter Operating Position 7.4.7 Power OFF Cutter Operating
Detector (S4/S5) Position Detector (S8/S9)
Removal Removal
1. Remove: 1. Remove:
· Magazine support plate cover (see subsection 7.3.2). · Magazine support plate cover (see subsection 7.3.2).
· Loader bottom cover (see subsection 7.3.4). · Loader bottom cover (see subsection 7.3.4).
2. Disconnect the wires from the detector. 2. Remove the detector bracket by removing the two
screws.
3. Remove the detector by removing the two screws.
F450
Wires Screws (2) Wires
F449
Installation
7-43
7.4 Cutter
· Magazine support plate cover (see subsection 7.3.2). Cutter Solenoid Bracket Flat Head Screws (2)
· Loader bottom cover (see subsection 7.3.4).
· Cutter blades (see subsection 7.5.3).
CL CR
F450
F454
Clamp
F455
7-44
9. Remove the lock nut, washer, collars, three plastic
washers, bolt and arm from the solenoid.
Installation
7-45
7.5 Feeder
7.5 Feeder
7.5.1 Input Check of Film Sensors 7.5.2 Film Sensor (FMS) Replacement
(FMS)
Removal
1. Display the Input Check <Film Sensor> screen
1. Turn the power switch to STANDBY, and the built-in
(see subsection 6.4.1).
circuit breaker and main power supply OFF.
2. Remove:
M734
Cover Stay Arm
Sheet of Paper
F567
Screws (2)
F458
7-46
6. Remove the loader left cover by removing the two 9. Remove the film sensor cover by removing the two
screws. screws.
7. Remove the cutter guide and spacer by removing the 10. Loosen the two screws securing the leader guide.
three screws.
11. Disconnect the connector (812-107) and remove the
Cutter Guide film sensor by removing the two screws.
Screws (2)
F436
Installation
7-47
7.5 Feeder
Wires F434
Installation
E-ring F426
Screws (2)
F458
7-48
7.5.4 Feeding Roller Replacement
Removal Installation
1. Remove the film sensor bracket (see subsection Install the feeding roller in the reverse order of removal.
7.5.2).
NOTE: Apply thread locking agent to the hex. socket head
2. Remove the two springs from the chain tensioner. set screw.
F339
Feeding Roller
F440
7-49
8. PROCESSOR
8-1
Parts Location
8.1 Troubleshooting
Crossover Racks
No.1 No.2 No.3 No.4 No.5
N4
N1 N3
F564
8-2
8.1 Troubleshooting
Trouble: An attempt was made to start processing even though the processor cover is still open.
No
Processing
section cover O Yes
Substitute a known good PWA32/PWB32
detector 1 board (see subsection 13.3.4).
Has the trouble now been remedied?
4
O No
S7 CN33
5312
END
8-3
8.1 Troubleshooting
continued
Yes
O O
OO
1
4
CN33
5314
: Action : Check
8-4
8.1.2 E204 “Urgently Close Processing Cover”
Yes
Was the processing section cover opened
during operation?
Yes
Is the dryer exit roller turning?
No
If there is any film currently being processed
wait until it comes out and drive motor stops.
To Next Page
F571
8-5
8.1 Troubleshooting
continued
O
S7
No
5312
No
Discharge capacitors C1 and C2
(see subsection 13.3.2).
Substitute a known good CTL32 board
Remove the processing section cover (see subsection 13.3.3).
detector (see subsection 8.4.2).
Check the continuity between the detector
terminals.
Is the detector set OFF/ON? END
No
S7
Yes
O O
OO
1
4 CN33
5314
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
: Action : Check G : Green W : White
Gr : Gray Y : Yellow
8-6
8.1.3 E207 “Open Processing Section Cover”
Trouble: The processing section cover was closed when the power switch was set STANDBY upon completion of the work or in any
other circumstance.
Yes
Is the processor cover closed?
No
Open the processing section cover.
The power is automatically switched off.
END
No
Processing Section O
Cover Detector Substitute a known good PWA32/PWB32 Yes
board (see subsection 13.3.4). Has the
4 1 trouble now been remedied?
O
S7 CN33 NO
5312
END
8-7
8.1 Troubleshooting
continued
S7
Yes
O O
OO
1
4
CN33
5314
: Action : Check
8-8
8.2 Crossover Rack
Roller
Roller F279
8-9
8.2 Crossover Rack
continued
3. Drive Gear Shaft Replacement 2) Remove the roller from the shaft.
1) Remove the gear holder by removing the two Shaft Bearings (2)
screws.
Screws (2)
Gear Holder
Assembly
F280
F282
8-10
8.2.3 Disassembly/Assembly of Crossover Rack No.5
Crossover Rack No.5 2) Remove the guide and rollers by removing the E-
ring and shaft.
F284
C-ring
Assembly
<N4 Rack>
Side Plate
8-12
4. Remove the opposite side plate of the sprocket side, 6. Disassemble the roller.
squeegee rollers, bearings, and front and rear guides.
Roller Tubes Roller Flanges Roller
Rear Guide Front Guide
Belt
Upper
8-13
8.4 Processing
8.4 Processing Section Cover Section Cover
F571
Detector Screws (2)
Wires F304
Installation
8-14
8.4.3 Processor Ventilation Fan Replacement
Removal Installation
1. Turn the power switch to STANDBY, and the built-in Install the parts in the reverse order of removal, noting the
circuit breaker and the main power supply OFF. following.
2. Open the processing section cover. NOTE: Install the fan with “ AIR FLOW” mark facing to the
processing section cover so that air can flow from the
3. Remove the four screws, and remove the fan cover. processor section to the outside.
Fan Cover
Cover
8-15
9. PROCESSING SOLUTION CIRCULATION SYSTEM
9-1
Parts Location
9.1 Troubleshooting
N2
N1
NS/N4 Heater
Subtank
N1 Heater
N2/N3 Heater
N1 Cooling Fan
Waste Solution
N1 Radiator Level Detector
Circulation Pump
F562
9-2
9.1 Troubleshooting
Trouble: The level detector of the waste solution tank, which is checked once every 5 minutes, is now ON.
NOTE: After the error message is displayed twice continuously, the error message can not be cleared unless the waste solution is
discharged.
Yes
END
No
To Next Page
: Action : Check
9-3
9.1 Troubleshooting
continued
Is W1 or W2 found to be problem-free
Yes
with “Level Detector” selected for the
maintenance mode input chedk
(see subsection 9.4.1)?
No
END
F124
Yes
W (W2)
Bu (W1) CN4
Repair or replace the harness.
W (W2)
Bu (W1)
5300
OK
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
: Action : Check G : Green W : White
Gr : Gray Y : Yellow
9-4
9.1.2 E051, E052, E053, E055, E056, E057 “W# Working Tank Level Low”
Trouble: One of the solution level detectors in the processing tanks is set OFF continuously for 1 second while the processor is not
driving.
One of the solution level detectors in the processing tanks is set OFF continuously for 5 seconds while the processor is driving.
NOTE: Clear the error message after stopping the processor drive. It can not be cleared while driving even if the solution level is
corrected.
No
No
Has any processing solution collected in
the machine pan?
Yes
END
Inspect for leaks in the processing tank
hoses and joints and repair. Drain off the
solution which has collected in the machine No
pan and recycle it in the waste recovery
tank. Pour more solution into the tank.
Press the [YES] key. Does the error recur ?
Yes
To Next Page
NOTE : Remove the float and move it up and down to inspect float for trapping or improper operation due to trapped air.
9-5
9.1 Troubleshooting
continued
Yes
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
: Action : Check G : Green W : White
Gr : Gray Y : Yellow
9-6
9.1.3 E500, E501, E502, E504, E505, E506 “N# Temperature Sensor Abnormal”
Trouble: · The readout temperature for N1 is 38.0 ± 0.2˚C ± even though the setting temperature range is 51.1˚C (124.0°F)
· The readout temperature for N2 to N4 is 51.1˚C even though the setting temperature range is 37 ± 3˚C (for N2 and N3) or
35 ± 3˚C (for NS1, NS2 and N4).
Yes
Set properly.
5323
OK
To Next Page
9-7
9.1 Troubleshooting
continued
16 15
5324
OK
: Action : Check
9-8
9.1.4 E600, E601, E604 “Heater (N#) Abnormal”
Trouble: The temperature rise of the heater during warm-up (until the temperature setting is reached) is 1 ˚C (1.8 ˚F) or less over a
period of 10 minutes.
Yes
Is the E608 displayed?
No
Does the processing solution recirculate?
Yes
Check the recirculation pump system.
Yes
Set properly.
To Next Page
9-9
9.1 Troubleshooting
continued
W W
5326
OK
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
: Action : Check G : Green W : White
Gr : Gray Y : Yellow
9-10
Remove the electrical section cover and
discharge capacitors C1 and C2
(see subsection 13.3.2).
Disconnect the CN21 connector from the
PWA32/PWB32 board and check the
continuity between the CN21 connector
and the heater connector.
CN21
4 8
Repair or replace the harness.
1 5
5327
OK
No END
9-11
9.1 Troubleshooting
continued
From page 9-10
qDefect in temperature detector system
5323
OK
11 and 12 NS-2
13 and 14 N4
15 and 16 DRY
CN3 Repair or replace the harness.
(N1) 2 1 (N1)
16 15
5324 END
OK
: Action : Check
9-12
9.1.5 E608 “Working Heater Overheated”
Trouble: Some of the safety thermo-switches for the heaters 1, 2 or 3 are activated.
To Next Page
* Safety thermo-switch open temperature : 64° ± 3 °C (148° ± 5.4°F); close temperature : 42° ± 3 °C (108° ± 5.4°F)
9-13
9.1 Troubleshooting
continued
5331
CONTINUITY EXISTS
5332
OK
: Action : Check
9-14
Substitute a known good PWA32/PWB32 Yes
board (see subsection 13.3.4).
Has the trouble now been remedied?
No END
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
Gr : Gray Y : Yellow
9-15
9.1 Troubleshooting
continued
No
Clean the cooling fan filters.
Is this error message displayed again?
Yes END
No
Connect the connector securely.
1 4
5333
OK
: Action : Check
9-16
Try replacing the cooling fan
Yes
(see subsection 9.3.4 or 9.3.5).
Try again the output check conducted two
steps before. Is the cooling fan operated?
No END
9-17
9.1 Troubleshooting
continued
5332
OK
END
5331
Yes
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
: Action : Check G : Green W : White
Gr : Gray Y : Yellow
* Safety thermo-switch open temperature : 64° ± 3 °C (148° ± 5.4°F); close temperature : 42° ± 3 °C (108° ± 5.4°F)
9-18
Remove the electrical section cover and
discharge capacitors C1 and C2
(see subsection 13.3.2).
Disconnect the CN21 connector from the
PWA32/PWB32 board and check the
continuity between the CN21 connector
and the heater connector.
CN21
4 8
Repair or replace the harness.
1 5
5327
OK
No END
9-19
9.1 Troubleshooting
9.1.6 E700, E701, E702, E704, E705, E706 “N# Temperature Abnormal”
Trouble: After the temperature of the solution (N2 to N4) has reached its setting, it has either risen or dropped more than 7 ˚C (12.6 ˚F)
above or below the setting during operation. The N1 temperature exceeds the setting 1 ˚C ( 1.8 ˚F) during operation.
Yes
Set properly.
Lower
Is the temperature of the solution higher
or lower than the setting?
: Action : Check
9-20
No
Clean the cooling fan filter.
Does the error recur?
Yes END
No
Connect the connector securely.
1 4
5333
OK
To Next Page
9-21
9.1 Troubleshooting
continued
No END
: Action : Check
9-22
From page 9-20, 9-21
qDiagnosis of the heater system
5332
OK
END
5331
Yes
To Next Page
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
Gr : Gray Y : Yellow
* Safety thermo-switch open temperature : 64° ± 3 °C (148° ± 5.4°F); close temperature : 42° ± 3 °C (108° ± 5.4°F)
9-23
9.1 Troubleshooting
continued
CN21
4 8
Repair or replace the harness.
1 5
5327
OK
No END
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
: Action : Check G : Green W : White
Gr : Gray Y : Yellow
9-24
From page 9-20
wDiagnosis of the temperature detector system
5323
OK
16 15
5324
OK
9-25
9.1 Troubleshooting
No
: Action : Check
9-26
9.2 Subtank Section
1. Display the Input Check screen for “Working Tank 4. Try moving the float of the processing solution level
Level Defector Check” (see subsection 6.4.1). detector up and down.
If the signal system of the processing solution level
detector is operating properly, the display will change
from 0 (OFF; Low) to 1 (ON; High).
Float
F575
Screws (3)
F405
Screws (2)
9-27
9.2 Subtank Section
2. Open the replenisher supply port cover. 2. Remove the detector protection cover
(see subsection 9.2.2).
3. Remove the two screws, and remove the detector
protection cover. 3. Disconnect the connector of the processing solution
level detector.
Detector Protection Cover
4. Remove the two screws, and remove the processing
solution level detector.
Connector
Screws (3)
F405
Installation
Level Detector Screws (2) F406
Install the parts in the reverse order of removal, noting the
following.
Installation
NOTE: Ensure that the harness is not pinched by the cover.
Install the parts in the reverse order of removal, noting the
following.
9-28
9.2.4 Solution Temperature Detector
Replacement
Removal
1. Turn the power switch to STANDBY, and the built-in
circuit breaker and the main power supply OFF.
Connector
Solution Temperature
Detector Screw
F407
Installation
9-29
9.3 Cooling
9.3 Cooling Fan and
Fan and Circulation Circulation Pump Section
Pump Section
2. If the circulating pump output system is operating 2. If the cooling fan 1 or 3 output system is operating
properly, the sound of the pump operating will be properly, the sound of the cooling fan operating will be
heard when the [1] key is pressed and it will stop when heard when the [1] or [3] key is pressed and the fan
the [0] key is pressed. will be switched off when the [0] key is pressed.
9-30
9.3.3 Circulation Pump Section Cover 9.3.4 N1 Cooling Fan (F1)
Removal/Installation Replacement
Removal Removal
1. Open the processor cover. 1. Turn the power switch to STANDBY, and the built-in
circuit breaker and the main power supply OFF.
2. Remove the three screws, raise the cover slightly and
remove it. 2. Remove the circulation pump section cover
(see subsection 9.3.3).
Screws (3)
3. Disconnect the connector (N1F) from the cooling fan.
Installation Install the parts in the reverse order of removal, noting the
following.
Install the parts in the reverse order of removal, noting the
following. NOTE: · Install the cooling fan with “ AIR FLOW” mark
toward the inside.
NOTE: There are two engagement hooks inside the · Pass the right screw through the cord tie as shown.
circulation pump cover. Make sure that the hooks are
hooked in place.
9-31
9.3 Cooling Fan and Circulation Pump Section
Installation Installation
Install the parts in the reverse order of removal, noting the Install the radiator in the reverse order of removal.
following.
In Side
“ AIR FLOW” Mark F391
9-32
9.3.7 Safety Thermo-Switch 9.3.8 Processing Solution Heater (H1,
(TH1 to TH3) Replacement H2, H3) Replacement
Removal Removal
1. Turn the power switch to STANDBY, and the built-in 1. Turn the power switch to STANDBY, and the built-in
circuit breaker and the main power supply OFF. circuit breaker and the main power supply OFF.
2. Remove the circulation pump section cover 2. Remove the circulation pump section cover (see
(see subsection 9.3.3). subsection 9.3.3).
3. For TH3: Remove the cooling fan (see subsection 3. Drain all processing solutions from the processing
9.3.5). tanks.
4. For TH3: Remove the two screws securing the 4. For H3: Remove the cooling fan (see subsection
processing solution heater to release the heater, to 9.3.5).
enable access to the thermo-switch screws.
5. Remove the harness duct protection cover by
5. Remove the connector cover and disconnect the wires removing the three screws.
from the safety thermo-switch.
Screws (3)
6. Remove the two screws, and remove the safety
thermo-switch from the processing solution heater.
Install the parts in the reverse order of removal. 8. Disconnect the grounding wire from the heater by
removing the screw.
9-33
9.3 Cooling Fan and Circulation Pump Section
Screws (2) Safety Screws (2) 2. Drain all processing solutions from the processing
Thermo-Switch tanks.
3. Remove:
Installation
Circulation Pump Connector F400
9-34
7. Remove the lower protection cover by removing the 12. Remove the two screws and the pump with the
two screws. bracket.
Screws
8. Remove the seven screws securing the drain valves to 13. Discharge the solution remaining inside the circulation
the pump. pump.
Screws (7)
11. Loosen the hose clamps, and disconnect the inlet and
outlet hoses from the pump.
9-35
9.3 Cooling Fan and Circulation Pump Section
continued
N1 Orifices
F403
0681
9-36
9.4 Waste Solution Tank Section
9.4.1 Input Check of Waste Solution 9.4.2 Waste Solution Level Detector
Level Detectors Replacement
1. Display the Input Check <Level Detector> screen Removal
(see subsection 6.4.1).
1. Turn the power switch to STANDBY, and the built-in
circuit breaker and the main power supply OFF.
Waste Solution Tank Section Cover F386 Connector (W1: Red, W2: Yellow) F387
3. Try moving the float of the waste solution tank level 4. Remove the tank cap.
detector up and down.
If the signal system of the respective detector is 5. Cut the detector wire at the connector, rotate the
operating properly, the W1 or W2 display will change detector, and remove it from the cap.
from 0 (OFF; Low) to 1 (ON; High).
Cap
Tank Level Detector
Detector F388
Float F573
9-37
9.4 Waste Solution Tank Section
continued
Installation
Cap
9-38
10. PROCESSING SOLUTION REPLENISHMENT SYSTEM
10-1
Parts Location
10.1 Troubleshooting
Replenisher
Nozzles
Solution
Temperature
Detector
Processing Solution
Level Detector
Circulation Filter
N4-R
Replenisher NS-R
Tanks N3-R
N2-R
N1-R
Replenisher
Level Detector
Drain Valves
F368
10-2
10.1 Troubleshooting
10.1.1 E002, E003, E004, E005, E006 “N# Replenisher Tank Empty”
Trouble: The level detector of the replenisher tank, which is checked once every 5 minutes, is now OFF.
NOTE: After the error message is displayed twice continuously, the error message can not be cleared unless the replenisher is added in
the tank.
Yes
Is the solution in the replenisher tank
indicated by the message at a low level?
No
Free or disengage the replenisher level
detector float.
To Next Page
: Action : Check
10-3
10.1 Troubleshooting
continued
No
0308
Yes
: Action : Check
10-4
Remove the electrical equipment
cover,and discharge capacitors C1and C2
(see subsection 13.3.2).
Disconnect the CN4 connector from the
CTL32 board and inspect the continuity
between the CN4 connector and the
replenisher level detector connector.
Replenisher level
CN4 Terminal
detector
1 and 2 (Red) N1 NG
3 and 4 (Brown) N2
5 and 6 (Yellow) N3
7 and 8 (Black) NS
9 and 10 (Orange) N4
Repair or replace the harness.
CN4
5301
OK
10-5
10.1 Troubleshooting
Trouble: This error message appears only once when the power switch is set ON if DIP switch “2” on the CTL32 board is in the ON
position.
END
: Action : Check
10-6
10.2 Replenisher Tanks
Screws (3)
M73A
F572
Replenisher Tank Section Cover
F367
Installation
10-7
10.2 Replenisher Tanks
Connector F370
Drain Valve
F369
4. After all the replenisher has been drained off, close the
drain valve.
Screws (4)
8. Remove the replenisher tank by lifting it slightly and
drawing it toward you to disengage its locating pin from
the bracket.
Bracket F373
10-8
9. Clamp the outlet hose of the replenisher tank securely Installation
using a pinch cock to prevent replenisher leakage.
Install the parts in the reverse order of removal, noting the
NOTE: Drain the replenisher from the hose when the following:
hose is disconnected in the following step 10 if
a pinch cock is not available. NOTE: · Connect the hoses securely so that they extend as
far as the base of the connection openings.
10. Loosen the hose clamp and remove the drain value. · Inspect the hose clamps and replace them if the
docking portion is damaged.
11. Disconnect the outlet hose and drain hose from the
replenisher tank by cutting the cord ties. Hose Clamp Docking Portions
0681
Hose Clamps Outlet Hose F376 · Check that the hose clamps are not twisted when
they are installed, hold each end of the clamp with a
pair of pliers, and tighten it.
10-9
10.2 Replenisher Tanks
Removal Installation
1. Turn the power switch STANDBY, and the built-in Install the parts in the reverse order of removal, noting the
circuit breaker and the main power supply OFF. following:
Detector
3. Pass the wire of a new detector through the
replenisher tank.
Connector
5167
Torque : 10 kgf-cm
10-10
10.2.5 Replenisher Filter Case 10.2.6 Replenisher Pump
Replacement Replacement
Removal Removal
1. Remove the waste solution tank section cover (see 1. Turn the power switch to STANDBY, and the built-in
subsection 9.4.1). circuit breaker and the main power supply OFF.
3. Clamp the outlet and inlet hoses of the filter case using · waste solution tank section cover (see subsection
pinch cocks. 9.4.1).
· W2 waste solution tank.
4. Disconnect the hoses from the filter case. · circulation pump section cover (see subsection
9.3.3).
Filter Case
3. Remove the replenisher pump section cover by
removing the two screws.
Screws (2)
Installation
Replenisher Pump Section Cover F379
Install the parts in the reverse order of removal.
10-11
10.2 Replenisher Tanks
continued
6. Disconnect the connector of the pump wire harness. 9. Clamp the inlet and outlet hoses of the pump using
pinch cocks to prevent replenisher leakage, and
disconnect the hoses from the pump.
Pinch Cocks
F381
F384
Installation
Color Pump
7. Remove the screw to disconnect the grounding wire
for the pump. N1-R, NS-R, N1W,
White
N2W
8. Remove the pump by removing the screw. Yellow N2-R, N3-R, N4-R
10-12
10.2.7 Replenisher Pump Valve Replacement
Removal Installation
1. Turn the power switch to STANDBY, and the built-in Install the parts in the reverse order of removal, noting the
circuit breaker and the main power supply OFF. following:
2. Remove the replenisher pump section cover. · Note the identification color of the pumps since the two
different valves are used as shown in the table below.
Screws (2)
Color Pump
N1-R, NS-R, N1W,
White
N2W
Yellow N2-R, N3-R, N4-R
Pinch Cocks
F384
Valves 5737
10-13
11. DRYER
11-1
Parts Location
11.1 Troubleshooting
Dryer Rack
Dryer Temperature
Detector
Dryer Safety
Thermo-switch (TH4)
Dryer Duct
Dryer Heater
Dryer Fan
F565
11-2
11.1 Troubleshooting
Trouble: The readout temperature for the dryer exceeds the specified limit.
5323
OK
To Next Page
: Action : Check
11-3
11.1 Troubleshooting
continued
OK
: Action : Check
11-4
11.1.2 E607 “DRYER Heater Abnormal”
Trouble: The temperature rise of the dryer heater is 5 ˚C (9 ˚F) or less 5 minutes after the machine has started operation.
To Next Page
11-5
11.1 Troubleshooting
continued
F615
NO CONTINUITY
No
Yes
The safety thermostat shuts off the heater at 95 ˚C (203 ˚F). Reset it by pushing in the push switch.
: Action : Check
11-6
From page 11-6 1
CN23
R Br R Br
Repair or replace the harness.
5328
OK
OK
Repair or replace the harness.
To Next Page
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
Gr : Gray Y : Yellow
11-7
11.1 Troubleshooting
continued
No END
No END
OK
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
: Action : Check G : Green W : White
Gr : Gray Y : Yellow
11-8
Substitute a known good PWA32/PWB32 Yes
board (see subsection 13.3.4).
Has the trouble now been remedied?
END
No
No
No
To Next Page
11-9
11.1 Troubleshooting
continued
WW 5330
OK
END
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
: Action : Check G : Green W : White
Gr : Gray Y : Yellow
11-10
11.1.3 E707 “DRYER Temperature Abnormal”
Trouble: After the dryer temperature has reached the setting, it has either risen or dropped more than 20 ˚C above or below the setting
during operation. Alternatively, it has dropped below 40 ˚C or risen above 75 ˚C.
To Next Page
11-11
11.1 Troubleshooting
continued
F615
NO CONTINUITY
No
F616
Yes
The safety thermostat shuts off the heater at 85˚C (185 ˚F). Reset it by pushing in the push switch.
: Action : Check
11-12
From page 11-12 2
NG
CN23
R Br R Br
5328
OK
NG
Check the continuity between the CN22
and the SSR4.
OK
Repair or replace the harness.
No END
No END
To Next Page
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
Gr : Gray Y : Yellow
11-13
11.1 Troubleshooting
continued
: Action : Check
11-14
Turn the power switch to STANDBY, the
built-in circuit breaker and the power
distribution switch OFF.
Remove the electrical equipment cover NG
and discharge capacitors C1 and C2
(see subsection 13.3.2).
Disconnect the CN25 connector from the
PWA32/PWB32 board and check the
continuity between the CN25 connector
and the dryer fan connector.
Repair or replace the harness.
OK
No END
No
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
Gr : Gray Y : Yellow
11-15
11.1 Troubleshooting
continued
No
CN3
2 1
WW 5330
OK
END
: Action : Check
11-16
11.1.4 E708 “DRYER Temperature Increasing”
Trouble: The dryer temperature fails to rise above 45 ˚C (113 ˚F) within 9 minutes after the machine started operating.
(This error message is automatically canceled when the temperature exceeds 45 ˚C/113 ˚F.)
Yes
Set properly.
Yes
Is the air above the dryer warm?
No
Is the room temperature unusually low?
(Under 10 ˚C/50 ˚F)
Yes
Yes
Press the [START] key and display the
temperature. Does the temperature rise?
No
* (NOTE) Yes
Is the dryer fan filter installed properly?
No
To Next Page
* NOTE : At this point of time, E607 or E707 will appear. Perform the troubleshooting for each error.
11-17
11.1 Troubleshooting
continued
F615
NG
: Action : Check
11-18
Press the safety thermostat switch. Is a
clicking sound heard?
No
Yes
The safety thermostat shuts off the heater at 95 ˚C (203˚F). Reset it by pushing in the push switch.
11-19
11.1 Troubleshooting
continued
CN23
R Br R Br
Repair or replace the harness.
OK
END
NG
Check the continuity between the CN22
and the SSR4.
OK
Repair or replace the harness.
Yes
Replace the SSR4 (see subsection
13.3.12).
No END
No END
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
G : Green W : White
Gr : Gray Y : Yellow
11-20
From page 11-19
2 Defect in dryer fan system
OK
No END
No
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
: Action : Check G : Green W : White
Gr : Gray Y : Yellow
11-21
11.1 Troubleshooting
continued
No
CN3
2 1
No
END
B : Black O : Orange
Br : Brown P : Purple
Bu : Blue R : Red
: Action : Check G : Green W : White
Gr : Gray Y : Yellow
11-22
11.2 Dryer Rack
Screws (2)
Exit Guide
F308
Nuts (4)
Inlet Side Cover Outlet Side Cover Side Plate Screws (9)
F312
11-23
11.2 Dryer Rack
continued
Belt
NOTE: Note the shape of the inlet and outlet side covers.
Inlet Side Cover Outlet Side Cover
Rollers F314
F316
11-24
11.3 Dryer Heater and Dryer Fan
11.3.1 Output Check of Dryer Fan 11.3.2 Dryer Heater Section Cover
Removal/Installation
1. Display the Dryer Fan output check screen
(see subsection 6.5.1). Removal
1. Remove the negative film receiver.
2. If the dryer fan output system is operating properly, the Screws (2)
sound of the fan operating will be heard when the [1]
key is pressed and it will stop when the [0] key is
pressed.
Installation
11-25
11.3 Dryer Heater and Dryer Fan
2. Remove the dryer heater section cover 2. Remove the dryer heater section cover (see
(see subsection 11.3.2). subsection 11.3.2).
3. Disconnect the dryer temperature detector connector. 3. Disconnect the two wires from the safety thermo-
switch.
4. Remove the screw and remove the dryer temperature
detector. 4. Remove the safety therm-switch by removing the two
screws.
Dryer Temperature Grounding Wire
Detector Screws (2) Safety Thermo-switch
Installation
Installation
Install the parts in the reverse order of removal, noting the
following. Install the safety thermo-switch in the reverse order of
removal.
IMPORTANT:
The detector element is bared at the tip of the dryer
temperature detector. Be careful not to touch the tip of the
sensor to the surrounding parts while installing or removing
the detector.
Detector Element
5115
11-26
11.3.5 Dryer Heater (H4) Replacement 11.3.6 Dryer Fan (F10) Replacement
Removal Removal
1. Turn the power switch to STANDBY, and the built-in 1. Turn the power switch to STANDBY, and the built-in
circuit breaker and the main power supply OFF. circuit breaker and the main power supply OFF.
4. Remove the screw and disconnect the grounding wire. 3. Remove the harness duct protection cover by
removing the three screws.
5. Remove the four screws and remove the dryer heater.
Install the parts in the reverse order of removal, noting the Connector (DFAN)
following.
! CAUTION
The machine used at 200V or 210V must install the optional
dryer heater (Part No.117G03701) instead of the standard
type.
11-27
11.3 Dryer Heater and Dryer Fan
continued
5. Remove the two screws securing the fan duct. 9. Remove the bracket from the fan by removing the four
screws.
Screws (4)
11-28
12. DRIVE SYSTEM
12-1
Parts Location
Feeding Roller
Chain
Tensioner
Idler Sprocket
Drive Motor
Drive Chain
F566
12-2
12.1 Troubleshooting
Trouble: This appears when the [START] key is used to restart operation after it has been stopped by the emergency stop switch during
film processing.
Wait until the “Ready for processing”
display appears.
The alarm and error messages are
automatically cleared 45 seconds after
the film has passed through if film is
present in the loader and 45 seconds
after the [START] key is pressed if no
film is present.
12-3
12.2 Drive Train
12.2.1 Output Check of Drive Motor 12.2.2 Drive Motor Bracket Removal/
Installation
1. Display the Drive Motor output check screen
(see subsection 6.5.1). Removal
1. Turn the power switch to STANDBY, and the built-in
circuit breaker and the main power supply OFF.
2. Ensure that the processing section cover is closed. 3. Remove the two chain tensioner springs.
F339
Cord Tie
12-4
12.2.3 Drive Motor (M1) Replacement
6. Remove the four screws, and remove the drive motor Removal
bracket.
1. Remove the drive motor bracket
Drive Motor Bracket (see subsection 12.2.2).
Installation
Installation
Install the motor in the reverse order of removal, noting the
following.
Drive Motor
12-5
12.2 Drive Train
1. Remove the drive motor bracket 2. Remove the circulation pump section cover
(see subsection 12.2.2). (see subsection 9.3.3).
2. Remove the drive gear/sprocket and bearings by 3. Using the manual crank handle, turn the drive shaft
removing the E-ring. until the master link of the drive chain can be seen.
Bearings (2) 4. Remove the chain clip using a pair of pliers, then
remove the master links.
12-6
12.3 Film Leader Collector
Spacer
Bracket F319
12-7
12.3 Film Leader Collector
continued
Hard Roller Bracket Install the rollers in the reverse order of removal, noting the
following.
· Align the hole in the rubber roller shaft with the hex
socket head screw.
12-8
13. CONTROL PANEL AND ELECTRICAL
EQUIPMENT SECTION
13-1
Parts Location
13.1 Troubleshooting
PU Circuit Board
CTL32 Circuit Board
Counter (CNT)
Circuit Protector
(CP1 to CP3)
F349
13-2
13.1 Troubleshooting
Yes
Was the power switch turned STANDBY
during operation?
Power Switch
0333
OK
OK
Repair or replace the harness.
: Action : Check
13-3
13.1 Troubleshooting
Trouble: The drive stop switch was pressed during operation (film processing).
Yes
Was the drive stop switch pressed
during operation?
Yes
Press the [START] key.
Is the dryer exit roller rotating?
No
If there is any film currently being processed,
wait until it comes out and drive motor stops.
No
: Action : Check
13-4
Turn the power switch to STANDBY, the
built-in circuit breaker and the power
distribution switch OFF.
Remove the electrical equipment cover OK
and discharge capacitors C1 and C2
(see subsection 13.3.2).
Disconnect the CN6 connector from the
CTL32 board and set the drive stop
switch ON/OFF and check the continuity. Substitute a known good CTL32 board
(see subsection 13.3.3).
NG
OK
Replace the drive stop switch
(see subsection 13.2.5).
13-5
13.1 Troubleshooting
Trouble: Communication error occurs during communication between the machine and host computer.
Communication does not complete within 2.5 sec.
Communication is performed during operation.
Yes
Is communication performed during
operation?
No
Wait until all the films get out and the operation
stops.
Yes END
No END
Yes
Replace the computer.
Has the trouble now been remedied?
No END
: Action : Check
13-6
13.2 Control Panel
13.2.1 Output Check of Film Loading 13.2.2 Input Check of Drive Stop
Lamp (Green Lamp) Switch (S1)
1. Display the Ready Lamp output check screen 1. Display the Input Check “S1” screen
(see subsection 6.5.1). (see subsection 6.4.1).
M741 M736
2. If the green or red lamp output system is operating 2. Turn the drive stop switch on and off. If switch S1 is
properly, the green or red lamp will come on when the operating properly, it will change 0 (OFF) to 1 (ON)
[1] key is pressed and go off when the [0] key is when pressed.
pressed.
M736
13-7
13.2 Control Panel
4. Open the replenisher tank cover. NOTE: Take care not to damage the power switch
terminals.
5. Remove the control panel by removing the four
screws. 5. Press the power switch tabs, and remove the switch.
F408
Power Switch Yellow 2 F411
Installation Installation
Install the control panel in the reverse order of removal. Installation is essentially in the reverse order of removal,
noting the following:
13-8
13.2.5 Drive Stop Switch Replacement 13.2.6 Film Loading Lamp (Green
Lamp) Replacement
Removal
Removal
1. Remove the control panel (see subsection 13.2.3).
1. Remove the control panel (see subsection 13.2.3).
2. Disconnect the terminal box from the drive stop switch
by pressing both tabs of the terminal box. 2. Disconnect the lamp connector and cut the lam wires
at the connector.
Terminal Box
3. Remove the lamp by removing the nut.
Pins
Installation
Connector F410
Install the parts in the reverse order of removal.
13-9
13.2 Control Panel
Removal 5. Remove the display and spacer from the circuit board
by removing the four screws and nuts.
1. Remove the control panel (see subsection 13.2.3).
Display Screws (4)
2. Remove the two harness duct covers and disconnect
the CN5 connector.
Installation
CN5 Connector F415 Install the display in the reverse order of removal. If
necessary, adjust display contrast (see subsection 13.2.8).
3. Remove the display by removing the two screws.
Display
Screws (2)
F413
Display
Flat Cable
F414
13-10
13.2.8 Display Contrast Adjustment 13.2.9 Keyboard Replacement
VR1
Installation
13-11
13.3 Electrical
13.3 Electrical Equipment Section
Equipment Section
1. Turn the power switch to STANDBY, and the built-in 2. Remove the electrical equipment cover
circuit breaker and the main power supply OFF. (see subsection 13.3.1).
2. Remove the two screws securing the electrical 3. Push the red button on the PWA32/PWB32 circuit
equipment cover. board for approximately 5 seconds to discharge
capacitors C1 and C2.
3. Lift the cover to remove it.
Screws (2)
Installation
13-12
13.3.3 CTL32 Circuit Board
Replacement
4. Measure the capacitors voltages as described below to
check if they are fully discharged. If voltage is 7V or Removal
more after measurement according to the procedures
below, repeat step 3. 1. If possible, record the setting data, such as the
replenisher amount, etc.
1) Left cutter capacitor (C2) voltage measurement
2. Turn the power switch to STANDBY, and the built-in
NOTE: Set the digital voltmeter to a range in circuit breaker and the main power supply OFF.
which a voltage exceeding DC100V can
be measured. 3. Remove the electrical equipment cover
(see subsection 13.3.1).
To measure the voltage, connect the (+) terminal
of the voltmeter to the +LC terminal on the PWT ! WARNING
circuit board and connect the (–) terminal to COM The main power supply MUST be switched OFF.
terminal. Even if the power switch is at STANDBY, electricity will
still be supplied in the power supply area, so that it can
C2 Capacitor cause electric shocks and short-circuiting.
F618
C1 Capacitor
F357
Installation
F619
13-13
13.3 Electrical Equipment Section
continued
3. Check the DIP switch settings as in the table below. 8. Measure the output voltages using a digital voltmeter,
and check the findings against the values in the table
below.
DS OFF ON
Not used 1 - - NOTE: Measure each voltage on the checking
Auto water
2 A NA terminal of the CPT circuit board.
replenishment
Control strip auto cut 3 NA A
Voltage Tolerance Checked between terminals
Termometer/Hygrometer 4 NA A
+5V +5V±0.2V +5V and GND1
Not used 5 - -
Not used 6 - - Power Supply Circuit Board
Not used 7 - -
Demo mode 8 NA A
A: Applied
NA: Not applied
DIP Switch
F620
E902
6. Set the main power supply to ON, then set the built-in
circuit breaker to ON, keeping from an electric shock.
13-14
Power Supply (PU) Circuit Board 14. Perform each input and output check
(see subsection 6.4.1 to 6.5.1).
! CAUTION
Ensure that all parts are safe before performing the
output check. The motor or solenoid is actually
operated when the output check is done.
16. Turn the DIP switch (DS-2) to OFF to set the normal
water replenishment condition (see subsection 6.11.8).
12. Set the following conditions in the order listed. 19. Set the processing temperature calibration again and
recalibrate if necessary (see subsection 6.2.1).
· Processing temperatures (see subsection 5.3.1).
· Replenisher rates
(see subsection 5.4.1). ...................... (NOTE1)
· Replenisher pump output
(see subsection 5.5.1). ...................... (NOTE1)
· Timers (see subsection 5.2.1).
· Error logging
(see subsection 5.7.1). ...................... (NOTE2)
· Processing temperature calibrations
(see subsection 6.2.1). ...................... (NOTE3)
13-15
13.3 Electrical Equipment Section
Removal Installation
1. Turn the power switch to STANDBY, and the built-in 1. Install the new PWA32/PWB32 circuit board and
circuit breaker and the main power supply OFF. connect the connectors and wires.
F359
13-16
13.3.5 Power Supply (PU) Circuit 13.3.6 Leakage Breaker Test
Board Replacement
1. Remove the electrical equipment cover
Removal (see subsection 13.3.1).
1. Turn the power switch to STANDBY, and the built-in 2. Turn ON the main power supply and built-in circuit
circuit breaker and the main power supply OFF. breaker.
! WARNING 3. Check that the breaker turns OFF when the red button
The main power supply MUST be switched OFF. is pressed. The leakage breaker is faulty if it does not
Even if the power switch is at STANDBY, electricity will turn OFF.
still be supplied in the power supply area, so that it can
cause electric shocks and short-circuiting. Breaker Switch
Installation
13-17
13.3 Electrical Equipment Section
5. Remove the relay by removing the two screws. 5. Remove the leakage breaker by removing the two
screws.
6 4
Leakage Breaker N31 N41
Installation
Installation
Install the relay in the reverse order of removal, noting the
following. Install the breaker in the reverse order of removal, noting the
following.
NOTE: Connect the wires to the correct terminals.
NOTE: Connect the wires (N31, N32, N41 and N42) properly
to the terminals on the breaker.
13-18
13.3.9 Circuit Protector (CP1-CP3) 13.3.10 Noise Filter (NF1)
Replacement Replacement
Removal Removal
1. Turn the power switch to STANDBY, and the built-in 1. Turn the power switch to STANDBY, and the built-in
circuit breaker and the main power supply OFF. circuit breaker and the main power supply OFF.
2. Remove the electrical equipment cover 2. Remove the electrical equipment cover
(see subsection 13.3.1). (see subsection 13.3.1).
4. Make sure no button of the circuit protectors project. If 4. Disconnect the wires from the noise filter terminals by
the button projects, it must be pressed. removing the five screws.
5. Disconnect the wires from the circuit protector and 5. Remove the noise filter by removing the two screws.
check for continuity between the terminals. If there is
no continuity, the circuit protector is opened and must Wires Nuts (5)
be replaced.
Installation
Installation
13-19
13.3 Electrical Equipment Section
N11 N21
Installation
Circuit Breaker N12 N22 F352 Install the SSR in the reverse order of removal.
Installation
Install the breaker in the reverse order of removal, noting the
following.
NOTE: Connect the wires (N11, N12, N21 and N22) properly
to the terminals on the breaker.
13-20
13.3.13 Counter (CNT) Replacement 13.3.14 Room Temperature/Humidity
(optional) Detector (THT) Replacement
Removal Removal
1. Turn the power switch to STANDBY, and the built-in 1. Turn the power switch to STANDBY, and the built-in
circuit breaker and the main power supply OFF. circuit breaker and the main power supply OFF.
2. Remove the electrical equipment cover 2. Remove the electrical equipment cover
(see subsection 13.3.1). (see subsection 13.3.1).
4. Remove the harness duct cover. 4. Remove the detector cover by removing the two
screws.
5. Disconnect the CN32 connector from the PWA32/
PWB32 circuit board. Screws (2)
Detector Cover
F365
Installation
13-21
13.3 Electrical Equipment Section
50Hz 60Hz
Orifice Selection 5 mm ID 4.5 mm ID
N1 Orifices
F403
50Hz 60Hz
Identification with label of 300K with label of 360K
F344
13-22