Operating Instructions
Deltapilot S FMB70
Hydrostatic level measurement
Deltapilot S FMB70 with 4...20 mA HART
Serial number
2. www.endress.com/deviceviewer Endress+Hauser
Operations App
3.
A0023555
Make sure the document is stored in a safe place such that it is always available when
working on or with the device.
To avoid danger to individuals or the facility, read the "Basic safety instructions" section
carefully, as well as all other safety instructions in the document that are specific to
working procedures.
The manufacturer reserves the right to modify technical data without prior notice. Your
Endress+Hauser Sales Center will supply you with current information and updates to these
Instructions.
2 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART
Table of contents
1 Document information . . . . . . . . . . . . . . 4 7.3 Selecting language and measuring mode . . . . . . 39
7.4 Position adjustment . . . . . . . . . . . . . . . . . . . . . . . . 40
1.1 Document function . . . . . . . . . . . . . . . . . . . . . . . . . . 4 7.5 Level measurement . . . . . . . . . . . . . . . . . . . . . . . . 42
1.2 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 7.6 Pressure measurement . . . . . . . . . . . . . . . . . . . . . 46
1.3 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . 5
1.4 Terms and abbreviations . . . . . . . . . . . . . . . . . . . . . 6
1.5 Turn down calculation . . . . . . . . . . . . . . . . . . . . . . . 7 8 Maintenance . . . . . . . . . . . . . . . . . . . . . 47
8.1 Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
2 Basic safety instructions . . . . . . . . . . . . . 8
2.1 Requirements concerning the staff . . . . . . . . . . . . . 8 9 Trouble-shooting . . . . . . . . . . . . . . . . . 48
2.2 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 9.1 Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
2.3 Workplace safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 9.2 Response of outputs to errors . . . . . . . . . . . . . . . . 56
2.4 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 9.3 Confirming messages . . . . . . . . . . . . . . . . . . . . . . . 58
2.5 Hazardous area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 9.4 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
2.6 Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 9.5 Repair of Ex-certified devices . . . . . . . . . . . . . . . . 59
2.7 Functional Safety SIL3 (optional) . . . . . . . . . . . . . . 9 9.6 Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
9.7 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
3 Identification . . . . . . . . . . . . . . . . . . . . . 10 9.8 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
9.9 Software history . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
3.1 Product identification . . . . . . . . . . . . . . . . . . . . . . 10
3.2 Device designation . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . 12 10 Technical data . . . . . . . . . . . . . . . . . . . . 60
3.4 CE mark, declaration of conformity . . . . . . . . . . 12
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
4 Installation . . . . . . . . . . . . . . . . . . . . . . . 13
4.1 Incoming acceptance and storage . . . . . . . . . . . . 13
4.2 Installation conditions . . . . . . . . . . . . . . . . . . . . . 13
4.3 General installation instructions . . . . . . . . . . . . . 13
4.4 Installation instructions . . . . . . . . . . . . . . . . . . . . 14
4.5 Post-installation check . . . . . . . . . . . . . . . . . . . . . 19
5 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.1 Connecting the device . . . . . . . . . . . . . . . . . . . . . 20
5.2 Connecting the measuring unit . . . . . . . . . . . . . . 22
5.3 Potential matching . . . . . . . . . . . . . . . . . . . . . . . . 24
5.4 Overvoltage protection (optional) . . . . . . . . . . . . 24
5.5 Post-connection check . . . . . . . . . . . . . . . . . . . . . 24
6 Operation. . . . . . . . . . . . . . . . . . . . . . . . . 25
6.1 On-site display (optional) . . . . . . . . . . . . . . . . . . 25
6.2 Operating elements . . . . . . . . . . . . . . . . . . . . . . . 26
6.3 On-site operation –
on-site display not connected . . . . . . . . . . . . . . . 28
6.4 On-site operation –
on-site display connected . . . . . . . . . . . . . . . . . . . 31
6.5 HistoROM®/M-DAT (optional) . . . . . . . . . . . . . . 33
6.6 Operation via SFX100 . . . . . . . . . . . . . . . . . . . . . . 36
6.7 Endress+Hauser operating program . . . . . . . . . . 36
6.8 Locking/unlocking operation . . . . . . . . . . . . . . . 36
6.9 Factory setting (reset) . . . . . . . . . . . . . . . . . . . . . 37
7 Commissioning . . . . . . . . . . . . . . . . . . . . 39
7.1 Configuring messages . . . . . . . . . . . . . . . . . . . . . 39
7.2 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Endress+Hauser 3
Document information Deltapilot S FMB70 with 4...20 mA HART
1 Document information
Symbol Meaning
DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
A0011189-DE seriousor fatal injury.
WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
A0011190-DE seriousor fatal injury.
CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
A0011191-DE
minoror medium injury.
NOTICE!
NOTICE This symbol contains information on procedures and other facts which do not result in
A0011192-DE personalinjury.
Symbol Meaning
Allen key
A0011221
Hexagon wrench
A0011222
4 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Document information
Symbol Meaning
Permitted
Indicates procedures, processes or actions that are permitted.
A0011182
Forbidden
Indicates procedures, processes or actions that are forbidden.
A0011184
Tip
Indicates additional information.
A0011193
Reference to documentation
A0028658
A
Reference to page
A0028659
Reference to graphic
A0028660
Series of steps
1. , 2. , 3. …
A0031595
Visual inspection
A0028673
Symbol Meaning
Series of steps
1. , 2. , 3. …
A0031595
A, B, C, D, ... Views
Symbol Meaning
Safety instructions
Observe the safety instructions contained in the associated Operating Instructions.
A0019159
Endress+Hauser 5
Document information Deltapilot S FMB70 with 4...20 mA HART
p
0
LRL LRV URV URL MWP OPL
A0029505
6 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Document information
1=2 3
LRL LRV URV URL
A0029545
Fig. 1:
1 Calibrated/Adjusted measuring span
2 Zero-based span
3 Upper range limit
Example
• Sensor: 10 bar (150 psi) • Calibrated/Adjusted measuring span: 0...5 bar
• Upper range limit (URL) = 10 bar (150 psi) (0...75 psi)
• Lower range value (LRV) = 0 bar
• Upper range value (URV) = 5 bar (75 psi)
Turn down (TD):
URL
TD =
|URV - LRV|
Endress+Hauser 7
Basic safety instructions Deltapilot S FMB70 with 4...20 mA HART
8 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Basic safety instructions
Endress+Hauser 9
Identification Deltapilot S FMB70 with 4...20 mA HART
3 Identification
3.2.1 Nameplate
• The MWP (maximum working pressure) is specified on the nameplate. This value refers
to a reference temperature of +20 °C (68°F) and may be applied to the device for an
unlimited time. Observe temperature dependency of the MWP. The pressure values
permitted at higher temperatures can be found in the standards EN 1092-1: 2001 Tab. 18
(With regard to their stability-temperature property, the materials 1.4435 and 1.4404 are
grouped together under 13EO in EN 1092-1 Tab. 18. The chemical composition of the two
materials can be identical.), ASME B 16.5a – 1998 Tab. 2-2.2 F316, ASME B 16.5a –
1998 Tab. 2.3.8 N10276, JIS B 2220.
• The test pressure corresponds to the over pressure limit (OPL) of the device = MWP x 1.5.
• The Pressure Equipment Directive (2014/68/EU) uses the abbreviation "PS".
The abbreviation "PS" corresponds to the MWP (maximum working pressure) of the
measuring device.
1
2 Order code: Ser. no.: 5
3 Ext. order code:
4 6
A0016056
Fig. 2: Nameplate
1 Device name
2 Order code (for re-orders)
3 Extended order code (complete)
4 Technical data
5 Serial number (for identification)
6 Address of manufacturer
10 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Identification
Devices for use in hazardous areas are fitted with an additional nameplate.
A0021222
1
2
3 Order code:
4 Ext. ord. cd.:
5 Ser. no.:
7
6
A0021552
Fig. 4: Nameplate
1 Device name
2 Address of manufacturer
3 Order code (for re-orders)
4 Extended order code (complete)
5 Serial number (for identification)
6 Technical data
7 Approval-specific information and document number for safety instructions or drawing number
Endress+Hauser 11
Identification Deltapilot S FMB70 with 4...20 mA HART
12 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Installation
4 Installation
4.1.2 Transport
! WARNING
Incorrect transportation
Housing and diaphragm may become damaged, and there is a risk of injury!
‣ Transport the measuring device to the measuring point in its original packaging or by the
process connection (with secure transport protection for the diaphragm).
‣ Follow the safety instructions and transport conditions for devices weighing more than
18 kg (39.6 lbs).
4.1.3 Storage
The device must be stored in a dry, clean area and protected against damage from impact
(EN 837-2).
Storage temperature range:
• –40 to +90°C (–40 to +194°F)
• On-site display: –40 to +85°C (–40 to +185°F)
• Separate housing: –40 to +60°C (–40 to +140°F)
4.2.1 Dimensions
For dimensions, please refer to the Technical Information for Deltapilot S TI00416P,
"Mechanical construction" section.
Endress+Hauser 13
Installation Deltapilot S FMB70 with 4...20 mA HART
SLANIMRET DLEIF
• Keep the pressure compensation and GORE-TEX® filter (1) free from contaminations.
• Do not clean or touch process isolating diaphragm with hard or pointed objects.
• The device must be installed as follows in order to comply with the cleanability
requirements of the ASME-BPE (Part SD Cleanability):
SLANIMRET DLEIF
IF
T DLE
IMRE
SLAN
INALS
TERM
FIELD
SLANIMRET DLEIF
14 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Installation
Level measurement
P01-PMP75xxx-11-xx-xx-xx-000
Endress+Hauser 15
Installation Deltapilot S FMB70 with 4...20 mA HART
1 2
A0017743
Fig. 6:
1 Process isolating diaphragm
2 Seal
ø42...60 (1.65...2.36)
70 (2.76)
52 (2.05)
86 (3.39)
ø6 (0.24)
A0028493
16 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Installation
6
8
5
3
2
1 r ³ 120 mm
P01-FMB70xxx-11-xx-xx-xx-003
Endress+Hauser 17
Installation Deltapilot S FMB70 with 4...20 mA HART
2 3
A0019996
1. T14 and T15 housing: Loosen setscrew with a 2 mm (0.08 in) Allen key.
T17 housing: Loosen setscrew with a 3 mm (0.12 in) Allen key.
2. Rotate housing (max. up to 380 °).
3. Retighten setscrew with 1 Nm (0,74 lbf ft.
NOTICE
Devices with EPDM cover seal - transmitter leakiness!
Mineral-based, animal-based or vegetable-based lubricants cause the EPDM cover seal to
swell and the transmitter to become leaky.
‣ The thread is coated at the factory and therefore does not require any lubrication.
NOTICE
The housing cover can no longer be closed.
Damaged thread!
‣ When closing the housing cover, please ensure that the thread of the cover and housing
are free from dirt, e.g. sand.If you feel any resistance when closing the cover, check the
thread on both again to ensure that they are free from dirt.
P01-FMB70xxx-17-xx-xx-xx-001
The covers for the terminal and electronics compartment are hooked into the casing and
closed with a screw. These screws should be finger-tightened (2 Nm (1.48 lbf ft)) to the stop
to ensure that the covers sit tightly.
18 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Installation
Endress+Hauser 19
Wiring Deltapilot S FMB70 with 4...20 mA HART
5 Wiring
➀ ➅ 10.5 V DC
➆ 11.5 V DC
4…20 mA
Test
Test
➂ ➄
➇
4... 20mA Test
➃
P01-xMx7xxxx-04-xx-xx-xx-001
20 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Wiring
A B – +
7
+ – 8 6
1
2 5
3 4
Han7D
+ –
A0019990
Fig. 10:
A Electrical connection for devices with Harting plug Han7D
B View of the plug connector at the device
1 Signal +
2 Not assigned
4 3
3 Signal –
1 4 Earth
2
A0011175
rd +
+
bk –
–
gnye
PE
4...20 mA
P01-PMx4xxxx-04-xx-xx-xx-010
Endress+Hauser 21
Wiring Deltapilot S FMB70 with 4...20 mA HART
Electronic version Jumper for 4 to 20 mA test signal Jumper for 4 to 20 mA test signal
in "Test" position (Delivery status) in "Non-Test" position
5.2.2 Terminals
• Supply voltage and internal ground terminal: 0.5 to 2.5 mm2 (20 to 14 AWG)
• External ground terminal: 0.5 to 4 mm2 (20 to 12 AWG)
22 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Wiring
5.2.4 Load
➀ RLmax ➁ RLmax
[Ω] [Ω]
Test 1500 Test
1456
1282 1239
847 804
➃ ➃
413
➂ 369
➂
10.5 20 30 40 45 U 11.5 20 30 40 45 U
[V] [V]
U – 10.5 V U – 11.5 V
RLmax ≤ RLmax ≤
23 mA 23 mA
P01-xMD7xxxx-05-xx-xx-xx-005
Fig. 12: Load diagram, observe the position of the jumper and the explosion protection. ( ä 22, Section 5.2.1 "Taking 4 to 20
mA test signal" part.)
1 Jumper for the 4 to 20 mA test signal inserted in "Non-Test" position
2 Jumper for the 4 to 20 mA test signal inserted in "Test" position
3 Supply voltage 10.5 (11.5) to 30 V DC for Ex ia, 1/2 D, 1 GD, 1/2 GD, FM IS and CSA IS, IEC Ex ia, NEPSI Ex ia
4 Supply voltage10.5 (11.5) to 45 V DC for devices for non-hazardous areas, 1/3 D, Ex nA, FM DIP, FM NI
RLmax Maximum load resistance
U Supply voltage
When operating via a handheld terminal or via PC with an operating program, a minimum
communication resistance of 250 must exist within the loop.
Endress+Hauser 23
Wiring Deltapilot S FMB70 with 4...20 mA HART
For the following Endress+Hauser instruments you need the "ToF Adapter FXA291" as an
additional accessory:
• Cerabar S PMC71, PMP7x
• Deltabar S PMD7x, FMD7x
• Deltapilot S FMB70
24 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Operation
6 Operation
Feature 20 "Output; operation" in the order code provides you with information on the
operating options available to you.
Header line
Main line
Unit
Information
line
– + E Symbol
Bargraph
Editing modes
Operating keys
Selection
options
Value that
can be edited
Endress+Hauser 25
Operation Deltapilot S FMB70 with 4...20 mA HART
The following table illustrates the symbols that can appear on the on-site display. Four
symbols can occur at one time.
Symbol Meaning
Alarm symbol
– Symbol flashing: warning, device continues measuring.
– Symbol permanently lit: error, device does not continue measuring.
Note: The alarm symbol may overlie the tendency symbol.
Lock symbol
The operation of the device is locked. Unlock device, ä 36.
Communication symbol
Data transfer via communication
Note: The alarm symbol may overlie the communication symbol.
➃➄
t ➅
on
➀
➀ ➁
off
➂
1 2
PC
P01-PMx7xxxx-19-xx-xx-xx-009 P01-xxxxxxxx-19-xx-xx-xx-104
Fig. 13: Operating keys, external Fig. 14: Operating keys, internal
1 Operating keys on the exterior of the device under the 1 Operating keys
protective flap 2 Slot for optional display
3 Slot for optional HistoROM®/M-DAT
4 DIP-switch for locking/unlocking measured-value-
relevant parameters
5 DIP-switch for damping on/off
6 Green LED to indicate value being accepted
26 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Operation
Position adjustment
Reset all parameters. The reset via operating keys corresponds to the software reset
and and code 7864.
Copy the configuration data from the optional HistoROM®/M-DAT module to the
and device.
Copy the configuration data from the device to the optional HistoROM®/M-DAT
and module.
F – Confirm entry
– Jump to the next item
Endress+Hauser 27
Operation Deltapilot S FMB70 with 4...20 mA HART
28 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Operation
Carry out position adjustment.1) Setting lower pressure value. Setting upper pressure value.
Pressure is present at device. Desired pressure for lower Desired pressure for upper
pressure value (EMPTY pressure value (FULL PRESSURE1)
PRESSURE2)) is present at device. is present at device.
Does the LED on the electronic Does the LED on the electronic Does the LED on the electronic
insert light up briefly? insert light up briefly? insert light up briefly?
Applied Applied The pressure The pressure The pressure The pressure
pressure for pressure for present was present was not present was present was not
position position saved as the saved as the saved as the saved as the
adjustment has adjustment has lower pressure lower pressure upper pressure upper pressure
been accepted. not been value (EMPTY value. Observe value (FULL value. Observe
accepted. PRESSURE2) the input limits. PRESSURE2) the input limits.
Observe the and assigned to and assigned to
input limits. the lower level the upper level
value (EMPTY value (FULL
CALIB.2). CALIB.2).
Endress+Hauser 29
Operation Deltapilot S FMB70 with 4...20 mA HART
Carry out position adjustment.1) Setting lower range value. Setting upper range value.
Pressure is present at device. Desired pressure for lower range Desired pressure for upper range
value is present at device. value is present at device.
Does the LED on the electronic Does the LED on the electronic Does the LED on the electronic
insert light up briefly? insert light up briefly? insert light up briefly?
Applied Applied Applied Applied Applied Applied
pressure for pressure for pressure for pressure for pressure for pressure for
position position lower range lower range upper range upper range
adjustment has adjustment has value has been value has not value has been value has not
been accepted. not been accepted. been accepted. accepted. been accepted.
accepted. Observe the Observe the
Observe the input limits. input limits.
input limits.
30 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Operation
P01-xxxxxxxx-19-xx-xx-xx-017
P01-xxxxxxxx-19-xx-xx-xx-033
P01-xxxxxxxx-19-xx-xx-xx-034
Endress+Hauser 31
Operation Deltapilot S FMB70 with 4...20 mA HART
P01-xxxxxxxx-19-xx-xx-xx-023
P01-xxxxxxxx-19-xx-xx-xx-027
P01-xxxxxxxx-19-xx-xx-xx-028
P01-xxxxxxxx-19-xx-xx-xx-029
Use to save the new value and exit the editing mode.
See next graphic.
P01-xxxxxxxx-19-xx-xx-xx-031
P01-xxxxxxxx-19-xx-xx-xx-032
32 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Operation
P01-xxxxxxxx-19-xx-xx-xx-035
P01-xxxxxxxx-19-xx-xx-xx-036
P01-xxxxxxxx-19-xx-xx-xx-037
P01-xxxxxxxx-19-xx-xx-xx-035
Endress+Hauser 33
Operation Deltapilot S FMB70 with 4...20 mA HART
➁
on
t
off
➀
1 2
PC
HART
R
250002271-–
FIELD COMMUNICATION PROTOCOL
HW-Version: 1
SW-Version: 2
P01-xxxxxxxx-19-xx-xx-xx-099
34 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Operation
Endress+Hauser 35
Operation Deltapilot S FMB70 with 4...20 mA HART
If operation is locked by means of the DIP-switch, you can only unlock operation again by
means of the DIP-switch. If operation is locked by means of the on-site display or remote
operation e.g. FieldCare, you can only unlock operation again by means of the on-site display
or remote operation.
36 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Operation
1) Parameters which refer to how the display appears, e.g. LANGUAGE and DISPLAY CONTRAST can still be
altered.
Damping [t]
Damping [t]
➀
➁ ➂
on on
1 2
PC
1 2 1 2
off off
HART
R
250002271-–
FIELD COMMUNICATION PROTOCOL
HW-Version: 1
SW-Version: 2
– + E
P01-xxxxxxxx-19-xx-xx-xx-133
Description
Endress+Hauser 37
Operation Deltapilot S FMB70 with 4...20 mA HART
38 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Commissioning
7 Commissioning
The device is configured for the level measuring mode as standard. The measuring range and
the unit in which the measured value is transmitted correspond to the specifications on the
nameplate.
! WARNING
Exceeding the maximum allowable working pressure!
Risk of injury due to bursting of parts! Warning messages are generated if pressure is too
high.
‣ If a pressure greater than the maximum permitted pressure is present at the device, the
messages "E115 Sensor overpressure" and "E727 Sensor pressure error - overrange" are
output in succession! Use the device only within the sensor range limits
NOTICE
Shortfall of the allowable working pressure!
Output of messages if pressure is too low.
‣ If a pressure smaller than the minimum permitted pressure is present at the device, the
messages "E120 Sensor low pressure" and "E727 Sensor pressure error - overrange" are
output in succession! Use the device only within the sensor range limits
Endress+Hauser 39
Commissioning Deltapilot S FMB70 with 4...20 mA HART
POS. ZERO ADJUST (685) Position adjustment – the pressure difference between zero (set point) and the
Entry measured pressure need not be known.
Example:
– MEASURED VALUE = 2.2 mbar (0.032 psi)
– Correct the MEASURED VALUE via the POS. ZERO ADJUST parameter with the
"Confirm" option. This means that you are assigning the value 0.0 to the pressure
present.
– MEASURED VALUE (after pos. zero adjust) = 0.0 mbar
– The current value is also corrected.
The CALIB. OFFSET parameter displays the resulting pressure difference (offset) by
which the MEASURED VALUE was corrected.
Factory setting:
0.0
POS. INPUT VALUE (563) Position adjustment – the pressure difference between zero (set point) and the
Entry measured pressure need not be known. To correct the pressure difference, you
need a reference measurement value (e. g. from a reference device).
Example:
– MEASURED VALUE = 0.5 mbar (0.0073 psi)
– For the POS. INPUT VALUE parameter, specify the desired set point for the
MEASURED VALUE, e.g. 2.0 mbar (0.029 psi).
(MEASURED VALUEnew = POS. INPUT VALUE)
– MEASURED VALUE (after entry for POS. INPUT VALUE) = 2.0 mbar (0.029 psi)
– The CALIB. OFFSET parameter displays the resulting pressure difference (offset)
by which the MEASURED VALUE was corrected.
CALIB. OFFSET = MEASURED VALUEold – POS. INPUT VALUE,
here: CALIB. OFFSET = 0.5 mbar (0.0073 psi) – 2.0 mbar (0.029 psi) = – 1.5
mbar (0.022 psi))
– The current value is also corrected.
Factory setting:
0.0
40 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Commissioning
CALIB. OFFSET (319) Position adjustment – the pressure difference between zero (set point) and the
Entry measured pressure is known.
Example:
– MEASURED VALUE = 2.2 mbar (0.032 psi)
– Via the CALIB. OFFSET parameter, enter the value by which the MEASURED
VALUE should be corrected. To correct the MEASURED VALUE to 0.0 mbar, you
must enter the value 2.2 here.
(MEASURED VALUE new = MEASURED VALUEold – CALIB. OFFSET)
– MEASURED VALUE (after entry for calib. offset) = 0.0 mbar
– The current value is also corrected.
Factory setting:
0.0
Endress+Hauser 41
Commissioning Deltapilot S FMB70 with 4...20 mA HART
42 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Commissioning
The measured variable is LEVEL SELECTION: Via OUTPUT UNIT – Calibration with – Incorrect entries are The measured value
in direct proportion to Level Easy Pressure parameter: %, level, reference pressure – wet possible display and the LEVEL
the measured pressure. volume or mass calibration, see Operating – SIL mode possible BEFORE LIN parameter
Calibration is performed units. Instructions BA00274P. – Customised units are show the measured
by entering two – Calibration without not possible value.
pressure-level value reference pressure – dry
pairs. calibration, see Operating
Instructions BA00274P.
The measured variable is LEVEL SELECTION: Via OUTPUT UNIT – Calibration with – Incorrect entries are The measured value
in direct proportion to Level Easy Height parameter: %, level, reference pressure – wet possible display and the LEVEL
the measured pressure. volume or mass calibration, see Operating – SIL mode not possible BEFORE LIN parameter
Calibration is performed units. Instructions BA00274P. – Customised units are show the measured
by entering the density – Calibration without not possible value.
and two height-level reference pressure – dry
value pairs. calibration, see Operating
Instructions BA00274P.
The measured variable is LEVEL SELECTION: Via LIN. – Calibration with – Incorrect entries are The measured value
in direct proportion to Level standard/ MEASURAND reference pressure – wet rejected by the device display and the LEVEL
the measured pressure. LEVEL MODE: parameter: calibration, see Operating – SIL mode not possible BEFORE LIN parameter
Linear – % (level) Instructions BA00274P. – Customised level, show the measured
– Level – Calibration without volume and mass value.
– Volume reference pressure – dry units are possible
– Mass calibration, see Operating
Instructions BA00274P.
The measured variable is LEVEL SELECTION: Via LINd – Calibration with – Incorrect entries are The measured value
not in direct proportion Level standard/ MEASURAND reference pressure: rejected by the device display and the
to the measured LEVEL MODE: parameter: semiautomatic entry of – SIL mode not possible TANK CONTENT
pressure as, for example, Pressure linearized – Pressure + % linearisation table, see – Customised level, parameter show the
with containers with a – Pressure + Operating Instructions volume and mass measured value.
conical outlet. A volume BA00274P. units are possible
linearisation table must – Pressure + mass – Calibration without
be entered for the reference pressure:
calibration. manual entry of
linearisation table, see
Operating Instructions
BA00274P.
– Two measured LEVEL SELECTION: Via COMB. – Calibration with – Incorrect entries are The measured value
variables are required Level standard/ MEASURAND reference pressure: wet rejected by the device display and the
or LEVEL MODE: parameter: calibration and – SIL mode not possible TANK CONTENT
– The container shape is Height linearized – Height + volume semiautomatic entry of – Customised level, parameter show the
given by value pairs, – Height + mass linearisation table, see volume and mass 2nd measured value
such as height and – Height + % Operating Instructions units are possible (volume, mass or %).
volume. – %-Height + BA00274P.
The 1st measured volume – Calibration without The LEVEL BEFORE LIN
– %-Height + mass reference pressure: dry parameter displays the
variable %-height or
– %-Height + % calibration and manual 1st measured value (%-
height must be in direct
proportion to the entry of linearisation height or height).
table, see Operating
measured pressure. The
Instructions BA00274P.
2nd measured variable
volume, mass or % must
not be in direct
proportion to the
measured pressure. A
linearisation table must
be entered for the 2nd
measured variable. The
2nd measured variable is
assigned to the 1st
measured variable by
means of this table.
Endress+Hauser 43
Commissioning Deltapilot S FMB70 with 4...20 mA HART
GROUP SELECTION
Select MEASURING MODE.
MEASURING MODE
Select "Level" option.
LEVEL SELECTION
Select level mode. For an overview see ä 43.
GROUP SELECTION
Select QUICK SETUP menu.
DAMPING TIME
Enter damping time (time constant ). The damping affects the speed at
which all subsequent elements, such as the on-site display, measured value
and current output react to a change in the pressure.
44 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Commissioning
For on-site operation, see also ä 27, Section 6.2.3 "Function of the operating elements –
on-site display connected" and ä 31, Section 6.4 ff "Operating elements".
Endress+Hauser 45
Commissioning Deltapilot S FMB70 with 4...20 mA HART
GROUP SELECTION
Select MEASURING MODE.
MEASURING MODE
Select "Pressure" option.
GROUP SELECTION
Select QUICK SETUP menu.
SET URV
Set the measuring range (enter 20 mA value).
Specify a pressure value for the upper current value (20 mA value). A
reference pressure does not have to be present at the device.
DAMPING TIME
Enter damping time (time constant ). The damping affects the speed at
which all subsequent elements, such as the on-site display, measured value
and current output react to a change in the pressure.
For on-site operation, ä 27, Section 6.2.3 "Function of the operating elements – on-site
display connected" and ä 31, Section 6.4 ff "Operating elements".
46 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Maintenance
8 Maintenance
Keep the pressure compensation and GORE-TEX® filter (1) free from contaminations.
P01-FMB70xxx-17-xx-xx-xx-003
Endress+Hauser 47
Trouble-shooting Deltapilot S FMB70 with 4...20 mA HART
9 Trouble-shooting
9.1 Messages
The following table lists all the possible messages that can occur.
The device differentiates between the error types "Alarm", "Warning" and "Error". You may
specify whether the instrument should react as if for an "Alarm" or "Warning" for "Error"
messages.
See "Error type/NA 64" column and Section 9.2 "Response of outputs to errors".
In addition, the "Error type/NA 64" column classifies the messages in accordance with
NAMUR Recommendation NA 64:
• Break down: indicated with "B"
• Maintenance need: indicated with "C" (check request)
• Function check: indicated with "I" (in service)
Error message display on the on-site display:
• The measured value display shows the message with the highest priority. See "Priority"
column.
• The ALARM STATUS parameter shows all the messages present in descending order of
priority. You can scroll through all the messages present with the key or key.
Message display via the digital communication:
• The ALARM STATUS parameter shows the message with the highest priority. See
"Priority" column.
• If the device detects a defect in the on-site display during initialization, special error
messages are generated. For the error messages, ä 55, Section 9.1.1 "On-site
display error messages".
• For support and further information, please contact Endress+Hauser Service.
• See also Section 9.4 ff.
101 (A101) Alarm Failure (F) B>Sensor electronic – Electromagnetic effects are – Wait a few minutes. 17
B EEPROM error greater than specifications in – Restart the device. Perform
the technical data. (See reset (Code 62).
Section 10.) This message – Block off electromagnetic
normally only appears briefly. effects or eliminate source of
disturbance.
102 Warning Maintenance C>Checksum error in – Main electronics defect. Correct – Replace main electronics. 53
(W102) C request (M) EEPROM: peakhold measurement can continue as
segment long as you do not need the
peak hold indicator function.
106 Warning Funktion C>Downloading - please – Downloading. – Wait for download to complete. 52
(W106) C check (C) wait
110 (A110) Alarm Failure (F) B>Checksum error in – The supply voltage is – Reestablish supply voltage. 6
B EEPROM: configuration disconnected when writing. Perform reset (Code 7864) if
segment necessary. Carry out calibration
again.
48 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Trouble-shooting
113 (A113) Alarm Failure (F) B>ROM failure in – Main electronics defect. – Replace main electronics. 1
B transmitter electronic
115 (E115) Error Out of B>Sensor overpressure – Overpressure present. – Reduce pressure until message 29
B specification disappears.
factory (S)
– Sensor defect. – Replace sensor.
setting:
Warning C
116 Warning Maintenance C>Download error, repeat – The file is defect. – Use another file. 36
(W116) C request (M) download
– During the download, the data – Check cable connection PC –
are not correctly transmitted to transmitter.
the processor, e.g. because of – Block off electromagnetic
open cable connections, spikes effects or eliminate sources of
(ripple) on the supply voltage or disturbance.
electromagnetic effects. – Perform reset (Code 7864) and
carry out calibration again.
– Repeat download.
120 (E120) Error Out of B>Sensor low pressure – Pressure too low. – Increase pressure until message 30
B specification disappears.
factory (S)
setting: – Sensor defect. – Replace sensor.
Warning C
121 (A121) Alarm Failure (F) B>Checksum error in – Main electronics defect. – Replace main electronics. 5
B factory segment of
EEPROM
122 (A122) Alarm Failure (F) B>Sensor not connected – Cable connection sensor –main – Check cable connection and 13
B electronics disconnected. repair if necessary.
– Electromagnetic effects are – Block off electromagnetic
greater than specifications in effects or eliminate source of
the technical data. disturbance.
(See Section 10.)
130 (A130) Alarm Failure (F) B>EEPROM is defect. – Main electronics defect. – Replace main electronics. 10
B
131 (A131) Alarm Failure (F) B>Checksum error in – Main electronics defect. – Replace main electronics. 9
B EEPROM: min/max
segment
132 (A132) Alarm Failure (F) B>Checksum error in – Main electronics defect. – Replace main electronics. 7
B totalizer EEPROM
133 (A133) Alarm Failure (F) B>Checksum error in – An error occurred when writing. – Perform reset (Code 7864) and 8
B History EEPROM carry out calibration again.
602 Warning Funktion C>Linearisation curve not – The linearisation table is not – Add to linearisation table or 57
(W602) C check (C) monoton monotonic increasing or perform linearisation again.
decreasing.
Endress+Hauser 49
Trouble-shooting Deltapilot S FMB70 with 4...20 mA HART
604 Warning Funktion C>Linearisation table not – The linearisation table consists – Add to linearisation table. If 58
(W604) C check (C) valid. Less than 2 points or of less than 2 points. necessary, perform linearisation
points too close again.
613 Warning Funktion I>Simulation is active – Simulation is switched on, i.e. – Switch off simulation. 60
(W613) I check (C) the device is not measuring at
present.
620 (E620) Error Out of C>Current output out of The current is outside the – Check pressure applied, 49
C specification range permitted range 3.8 to 20.5 mA. reconfigure measuring range if
Factory (S) – The pressure applied is outside necessary (See also
setting: Operating Instructions
the set measuring range (but
Warning C BA00274P)
within the sensor range).
– Perform reset (Code 7864) and
carry out calibration again.
700 Warning Maintenance C>Last configuration not – An error occurred when writing – Perform reset (Code 7864) and 54
(W700) C request (M) stored or reading configuration data or carry out calibration again.
the power supply was
disconnected.
702 Warning Maintenance C>HistoROM data not – Data were not written correctly – Repeat upload. 55
(W702) C request (M) consistent. to the HistoROM, e.g. if the – Perform reset (Code 7864) and
HistoROM was detached during carry out calibration again.
the writing process.
705 (A705) Alarm Failure (F) B>Measurement error – Fault in the main electronics. – Briefly disconnect device from 21
B the power supply.
50 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Trouble-shooting
706 Warning Maintenance C>Configuration in – Configuration (parameters) in – Copy data from the device to 59
(W706) C request (M) HistoROM and device not the HistoROM and in the device the HistoROM. (See also
identical is not identical. ä 34, Section 6.5.1 "Copying
configuration data".)
– Copy data from the HistoROM
to the device. (See also ä 34,
Section 6.5.1 "Copying
configuration data".) The
message remains if the
HistoROM and the device have
different software versions. The
message goes out if you copy
the data from the device to the
HistoROM.
– Device reset codes such as 7864
do not have any effect on the
HistoROM. That means that if
you do a reset, the
configurations in the HistoROM
and in the device may not be
the same.
707 (A707) Alarm Funktion B>X-VAL. of lin. table out – At least one X-VALUE in the – Carry out calibration again. ( 38
B check (C) of edit limits. linearisation table is either See also Operating Instructions
below the value for HYDR. BA00274P)
PRESS MIN. or MIN. LEVEL or
above the value for HYDR.
PRESS. MAX. or LEVEL MAX.
710 Warning Funktion B>Set span too small. Not – Values for calibration (e.g. – Adjust calibration to suit 51
(W710) C check (C) allowed. lower range value and upper sensor.
range value) are too close (See also Operating
together. Instructions BA00274P,
parameter description
MINIMUM SPAN)
711 (A711) Alarm Funktion B>LRV or URV out of edit – Lower range value and/or upper – Reconfigure lower range value 37
B check (C) limits range value undershoot or and/or upper range value to suit
overshoot the sensor range the sensor. Pay attention to
limits. position factor.
713 (A713) Alarm Funktion B>100% POINT level out of – The sensor was replaced. – Carry out calibration again. 39
B check (C) edit limits
Endress+Hauser 51
Trouble-shooting Deltapilot S FMB70 with 4...20 mA HART
715 (E715) Error Out of C>Sensor over temperature – The temperature measured in – Reduce process temperature/ 32
C specification the sensor is greater than the ambient temperature.
Factory (S) upper nominal temperature of
setting: the sensor. (See also
Warning C Operating Instructions
BA00274P, parameter
description Tmax SENSOR)
– Unsuitable download carried – Check configuration and
out. perform download again.
716 (E716) Error Failure (F) B>Process isolating – Sensor defect. – Replace sensor. 24
B diaphragm broken – Reduce pressure.
Factory
setting:
Alarm B
717 (E717) Error Out of C>Transmitter over – The temperature measured in – Reduce ambient temperature. 34
C specification temperature the electronics is greater than
Factory (S) the upper nominal temperature
setting: of the electronics (+88 °C (+190
Warning C °F)).
– Unsuitable download carried – Check configuration and
out. perform download again.
718 (E718) Error Out of C>Transmitter under – The temperature measured in – Increase ambient temperature. 35
C specification temperature the electronics is smaller than Insulate device if necessary.
Factory (S) the lower nominal temperature
setting: of the electronics (–43 °C (-45
Warning C °F)).
– Unsuitable download carried – Check configuration and
out. perform download again.
719 (A719) Alarm Funktion B>Y-VAL of lin. table out – At least on Y-VALUE in the – Carry out calibration again. ( 40
B check (C) of edit limits linearisation table is below the See also Operating Instructions
MIN. TANK CONTANT or above BA00274P)
the MAX. TANK CONTENT.
720 (E720) Error Out of C>Sensor under – The temperature measured in – Increase process temperature/ 33
C specification temperature the sensor is smaller than the ambient temperature.
Factory (S) lower nominal temperature of
setting: the sensor. ( See also
Warning C Operating Instructions
BA00274P, parameter
description Tmin SENSOR)
721 (A721) Alarm Funktion B>ZERO POSITION level – LEVEL MIN or LEVEL MAX has – Perform reset (Code 2710) and 41
B check (C) out of edit limits been changed. carry out calibration again.
722 (A722) Alarm Funktion B>EMPTY CALIB. or FULL – LEVEL MIN or LEVEL MAX has – Perform reset (Code 2710) and 42
B check (C) CALIB. out of edit limits been changed. carry out calibration again.
723 (A723) Alarm Funktion B>MAX. FLOW out of edit – FLOW-MEAS. TYPE has been – Carry out calibration again. 43
B check (C) limits changed.
52 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Trouble-shooting
725 (A725) Alarm Failure (F) B>Sensor connection error, – Electromagnetic effects are – Block off electromagnetic 25
B cycle disturbance greater than specifications in effects or eliminate source of
the technical data. (See disturbance.
Section 10.)
– Setscrew loose. – Retighten setscrew with 1 Nm
(0,74 lbf ft) (see Chap. 4.4.5).
726 (E726) Error Out of C>Sensor temperature – Electromagnetic effects are – Block off electromagnetic 31
C specification error - overrange greater than specifications in effects or eliminate source of
Factory (S) the technical data. disturbance.
setting: (See Section 10.)
Warning C
– Process temperature is outside – Check temperature present,
permitted range. reduce or increase if necessary.
727 (E727) Error Out of C>Sensor pressure error - – Electromagnetic effects are – Block off electromagnetic 28
C specification overrange greater than specifications in effects or eliminate source of
Factory (S) the technical data. disturbance.
setting: (See Section 10.)
Warning C
– Pressure is outside permitted – Check pressure present, reduce
range. or increase if necessary.
– Sensor defect. – If the pressure is within the
permitted range, replace sensor.
728 (A728) Alarm Failure (F) B>RAM error – Fault in the main electronics. – Briefly disconnect device from 2
B the power supply.
729 (A729) Alarm Failure (F) B>RAM error – Fault in the main electronics. – Briefly disconnect device from 3
B the power supply.
730 (E730) Error Out of C>LRV user limits – Pressure measured value has – Check system/pressure 46
C specification exceeded undershot the value specified measured value.
Factory (S) for the Pmin ALARM WINDOW – Change value for Pmin ALARM
setting: parameter. WINDOW if necessary. (See
Warning C also Operating Instructions
BA00274P, parameter
description Pmin ALARM
WINDOW)
Endress+Hauser 53
Trouble-shooting Deltapilot S FMB70 with 4...20 mA HART
731 (E731) Error Out of C>URV user limits – Pressure measured value has – Check system/pressure 45
C specification exceeded overshot the value specified for measured value.
Factory (S) the Pmax ALARM WINDOW – Change value for Pmax ALARM
setting: parameter. WINDOW if necessary. (See
Warning C also Operating Instructions
BA00274P, parameter
description Pmax ALARM
WINDOW)
732 (E732) Error C>LRV Temp. User limits – Temperature measured value – Check system/temperature 48
C exceeded has undershot the value measured value.
Factory specified for the Tmin ALARM – Change value for Tmin ALARM
setting: WINDOW parameter. WINDOW if necessary. (See
Warning C also Operating Instructions
BA00274P, parameter
description Tmin ALARM
WINDOW)
733 (E733) Error Out of C>URV Temp. User limits – Temperature measured value – Check system/temperature 47
C specification exceeded has overshot the value specified measured value.
Factory (S) for the Tmax ALARM WINDOW – Change value for Tmax ALARM
setting: parameter. WINDOW if necessary. (See
Warning C also Operating Instructions
BA00274P, parameter
description Tmax ALARM
WINDOW)
736 (A736) Alarm Failure (F) B>RAM error – Fault in the main electronics. – Briefly disconnect device from 4
B the power supply.
– Main electronics defect. – Replace main electronics.
737 (A737) Alarm Failure (F) B>Measurement error – Fault in the main electronics. – Briefly disconnect device from 20
B the power supply.
738 (A738) Alarm Failure (F) B>Measurement error – Fault in the main electronics. – Briefly disconnect device from 19
B the power supply.
739 (A739) Alarm Failure (F) B>Measurement error – Fault in the main electronics. – Briefly disconnect device from 23
B the power supply.
– Main electronics defect. – Replace main electronics.
740 (E740) Error Maintenance C>Calculation overflow, – Level measuring mode: the – Check configuration and carry 27
C request (M) bad configuration, measured pressure has out calibration again if
Factory hardware defect undershot the value for HYDR. necessary.
setting: PRESS. MIN. or overshot the – Select a device with a suitable
Warning C value for HYDR. PRESS MAX. measuring range.
741 (A741) Alarm Funktion B>TANK HEIGHT out of – LEVEL MIN or LEVEL MAX has – Perform reset (Code 2710) and 44
B check (C) edit limits been changed. carry out calibration again.
54 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Trouble-shooting
742 (A742) Alarm Failure (F) B>Sensor connection error – Electromagnetic effects are – Wait a few minutes. 18
B (upload) greater than specifications in – Perform reset (Code 7864) and
the technical data. (See carry out calibration again.
Section 10.) This message
normally only appears briefly.
743 (E743) Alarm Failure (F) B>Electronic PCB error – This message normally only – Wait a few minutes. 14
B during initialisation appears briefly. – Restart the device. Perform
reset (Code 62).
744 (A744) Alarm Failure (F) B>Main electronic PCB – Electromagnetic effects are – Restart the device. Perform 11
B error greater than specifications in reset (Code 62).
the technical data. – Block off electromagnetic
(See Section 10.) effects or eliminate source of
disturbance.
745 Warning Maintenance C>Sensor data unknown – Sensor does not suit the device – Replace sensor with a suitable 56
(W745) C request (M) (electronic sensor nameplate). sensor.
Device continues measuring.
746 Warning Funktion C>Sensor connection error – Electromagnetic effects are – Wait a few minutes. 26
(W746) C check (C) - initialising greater than specifications in – Restart the device. Perform
the technical data. (See reset (Code 7864).
Section 10.) This message – Block off electromagnetic
normally only appears briefly. effects or eliminate source of
disturbance.
748 (A748) Alarm Failure (F) B>Memory failure in signal – Electromagnetic effects are – Block off electromagnetic 15
B processor greater than specifications in effects or eliminate source of
the technical data. disturbance.
(See Section 10.)
Message Measure
Initialization, VU Electr. Defect A110 Exchange on-site display.
Endress+Hauser 55
Trouble-shooting Deltapilot S FMB70 with 4...20 mA HART
Current output Assumes the value specified via the Device continues measuring. For this error, you can enter whether the
OUTPUT FAIL MODE1), ALT. CURR. device should react as in the event of an
OUTPUT1 and SET MAX. ALARM1 alarm or as in the event of a warning.
parameter. See also the following See corresponding "Alarm" or "Warning"
section "Configuring current output for column. (See also Operating
an alarm". Instructions BA00274P, parameter
description SELECT ALARM TYPE)
Bargraph The bargraph adopts the value defined The bargraph adopts the value which See this table, "Alarm" or "Warning"
(on-site display) by the OUTPUT FAIL MODE 1 corresponds to the current value. column, depending on selection.
parameter.
On-site display – The measured value and message are – The measured value and message are – The measured value and message are
displayed alternately displayed alternately displayed alternately
– Measured value display: -symbol is – Measured value display: -symbol – Measured value display: see
permanently displayed. flashes. corresponding "Alarm" or "Warning"
column
56 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Trouble-shooting
I
max.
➀ 21...23 mA
min.
➂ 3.6 mA
t
P01-xxxxxxxx-05
Factory setting:
• OUTPUT FAIL MODE: Max. Alarm (110%)
• SET MAX. ALARM: 22 mA
Use the ALT. CURR. OUTPUT parameter to set the current output value for the error
messages E 120 "Sensor low pressure" and E 115 "Sensor overpressure". You have the
following options:
– Normal/NE43: the current output assumes the value set via the OUTPUT FAIL MODE and
SET MAX. ALARM parameters.
– Special
– Lower sensor limit undershot (E 120 "Sensor low pressure"): 3.6 mA
– Upper sensor limit overshot (E 115 "Sensor overpressure") overshot: current output
assumes the value set via the SET MAX ALARM parameter.
Attention : when using the case "special", the behavior is limited to an over/underpressure
in a range LRL -10% up to LRL -30% and URL +10% up to URL +30%.
Factory setting:
• ALT. CURR. OUTPUT: Normal/NE43
Endress+Hauser 57
Trouble-shooting Deltapilot S FMB70 with 4...20 mA HART
Settings 1) Measures
– ALARM DISPL. TIME = 0 s – Rectify cause of the message (see also Section 9.1).
– ACK. ALARM MODE = off
– ALARM DISPL. TIME > 0 s – Rectify cause of the message (see also Section 9.1).
– ACK. ALARM MODE = off – Wait for the alarm display time to elapse.
– ALARM DISPL. TIME = 0 s – Rectify cause of the message (see also Section 9.1).
– ACK. ALARM MODE = on – Confirm message using ACK. ALARM parameter.
– ALARM DISPL. TIME > 0 s – Rectify cause of the message (see also Section 9.1).
– ACK. ALARM MODE = on – Confirm message using ACK. ALARM parameter.
– Wait for the alarm display time to elapse. If a message appears and the
alarm display time elapses before the message has been acknowledged,
the message will be cleared once it has been acknowledged.
1) Menu path for ALARM DISPL. TIME and ACK. ALARM MODE: (GROUP SELECTION) OPERATING MENU
DIAGNOSTICS MESSAGES
If the on-site display displays a message, you can suppress it with the -key.
If there are several messages, the on-site display shows the message which has the highest
priority (see also Section 9.1). Once you have suppressed this message using the -key, the
message with the next highest priority is displayed. You can use the -key to suppress each
message, one after the other.
The ALARM STATUS parameter continues to display all the messages present.
58 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART Trouble-shooting
9.4 Repair
The Endress+Hauser repairs concept provides for measuring devices to have a modular
design and also the customer may carry out repairs (see Chap. 9.6 "Spare Parts" on ä 59).
• For certified devices, please consult Chapter "Repair of Ex-certified devices".
• For more information on service and spare parts contact the Endress+Hauser Service.
See www.endress.com/worldwide.
9.7 Return
The measuring device must be returned if repairs or a factory calibration are required, or if
the wrong measuring device has been ordered or delivered. According to legal regulations,
Endress+Hauser, as a ISO-certified company, is required to follow certain procedures when
handling returned products that are in contact with process fluids.
To ensure swift, safe and professional device returns, please read the return procedures and
conditions on the Endress+Hauser website at www.services.endress.com/return-material.
9.8 Disposal
When disposing, separate and recycle the device components based on the materials.
Endress+Hauser 59
Technical data Deltapilot S FMB70 with 4...20 mA HART
BA332P/00/EN/12.07 BA274P/00/EN/07.07
71043309 71061022
BA332P/00/EN/05.08 BA274P/00/EN/05.08
71071716 71071855
BA332P/00/EN/08.08 BA274P/00/EN/05.08
71077293 71071855
BA332P/00/EN/06.09 BA274P/00/EN/06.09
71095397 71095452
BA332P/00/EN/05.10 BA274P/00/EN/05.10
71114100 71118244
BA00332P/00/EN/13.11 BA00274P/00/EN/13.11
71139745 71139795
BA00332P/00/EN/14.12 BA00274P/00/EN/13.11
71161907 71139795
BA00332P/00/EN/16.14 BA00274P/00/EN/15.14
71254265 71254474
06.2014 02.20.zz HART7 protocole revision has been implemented. BA00332P/00/EN/17.14 BA00274P/00/EN/16.14
71260298 71260321
BA00332P/00/EN/18.14 BA00274P/00/EN/17.14
71270395 71270402
10 Technical data
For technical data, please refer to the Technical Information TI00416P for Deltapilot S.
60 Endress+Hauser
Deltapilot S FMB70 with 4...20 mA HART
Index
Numerics Position adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4 to 20 mA test signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Potential matching . . . . . . . . . . . . . . . . . . . . . . . . . . . 23–24
Pressure measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
A Pressure measurement, on-site operation . . . . . . . . . . . . 30
Alarm messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 Pressure measurement, Quick Setup menu . . . . . . . . . . . 46
Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
C
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Q
Commubox FXA195 anschließen . . . . . . . . . . . . . . . . . . 23 Quick Setup menu level . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Connecting Commubox FXA291 . . . . . . . . . . . . . . . . . . . 24 Quick Setup menu pressure . . . . . . . . . . . . . . . . . . . . . . . . 46
Connecting ToF Adapter FXA291 . . . . . . . . . . . . . . . . . . 24
R
D Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Repair of Ex-certified devices. . . . . . . . . . . . . . . . . . . . . . . 59
Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Returning devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
E Rotating the housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Error messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 S
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
F Screening . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Factory setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Separate housing, assemble and mount. . . . . . . . . . . . . . 17
FieldCare. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Service Interface FXA291. . . . . . . . . . . . . . . . . . . . . . . . . . 24
Sicherheitshinweise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
H SIL3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Hazardous area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Software history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
HistoROM/M-DAT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
I Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . 14 T
Trouble-shooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
L
Language, selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 U
Level measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . 42–43 Unlocking operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Level measurement, on-site operation . . . . . . . . . . . . . . 28
Level measurement, Quick Setup menu . . . . . . . . . . . . . 44 W
Load. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Wall mounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Locking operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Workplace safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
M
Measuring mode, selection . . . . . . . . . . . . . . . . . . . . . . . 39
N
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
O
On-site display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Operating elements, position . . . . . . . . . . . . . . . . . . . . . . 26
Operating keys, function, with on-site display . . . . . . . 27
Operating keys, function, without on-site display. . . . . 27
Operating keys, position . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Operating menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . . . . 24