Fluke 77 Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 20

®

Model 77 Series IV
Digital Multimeter

Calibration Information
Introduction
XWWarning
To avoid electric shock or injury, do not perform the performance tests or
calibration adjustment procedures unless qualified to do so.
The information provided in this document is for the use of qualified personnel
only.
The Model 77 Series IV Calibration Information provides the information necessary to adjust and verify
the performance of the Fluke Model 77 Series IV Digital Multimeter (hereafter known as the Meter).
The following information is included in this document:
• Safety Information (page 2)
• International Electrical Symbols (page 3)
• Test Lead Alert (page 3)
• Specifications (page 4)
• Testing and Replacing the Fuses (page 7)
• Replacing the Battery (page 8)
• Cleaning (page 9)
• Replacing the LCD (page 9)
• Performance Tests (page 10)
• Calibration Adjustment (page 14)
• Replaceable Parts and Accessories (page 17)
• Complete Warranty (page 19)

See the Model 77 Series IV Users Manual for complete operating instructions.

Service Information
To contact Fluke, call one of the following telephone numbers:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's Web site at www.fluke.com.
To register your product, visit register.fluke.com

September 2006
© 2006 Fluke Corporation. All rights reserved.
1
Model 77 Series IV
Calibration Information

Safety Information
Warning and Caution Statements
A XW Warning identifies hazardous conditions and actions that could cause bodily harm or death.
A W Caution identifies conditions and actions that could damage the Meter, the equipment under test, or
cause permanent loss of data.

XWWarnings and Precautions


To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter
or to the equipment under test, adhere to the following practices:
• Use the Meter only as specified in this manual or the protection provided by the Meter might
be impaired.
• Do not use the Meter or test leads if they appear damaged, or if the Meter is not operating
properly. If in doubt, have the Meter serviced.
• Always use the proper terminals, switch position, and range for measurement.
• Verify the Meter's operation by measuring a known voltage.
• Do not apply more than the rated voltage, as marked on the Meter, between the terminals or
between any terminal and earth ground.
• Use caution with voltages above 30 V ac rms, 42 V ac peak, or 60 V dc. These voltages pose
a shock hazard.
• Disconnect circuit power and discharge all high-voltage capacitors before testing
resistance, continuity, diodes, or capacitance.
• Do not use the Meter around explosive gas or vapor.
• When using the test leads, keep your fingers behind the finger guards.
• Remove test leads from the Meter before opening the Meter case or battery door.

2
Digital Multimeter
International Electrical Symbols

International Electrical Symbols


The following international symbols appear in this document, and on the Meter.

Symbols Description Symbols Description

B AC (Alternating Current) I Fuse

F DC (Direct Current) P Conforms to European Union directives

F
B
DC/AC $ Conforms to relevant Canadian Standards
Association directives

J Earth ground T Double-insulated

W Important Information; see manual. X Hazardous Voltage

Battery. (Low battery when shown on Underwriters Laboratories, Inc.


b display)
!
Meter in accordance with IEC 61010-1. 54CJ

Inspected and licensed by TÜV


;
s (Technischer Überwachungs Verein) N10140 Conforms to relevant Australian standards
Product Services.

# Verband Deutscher Electroniker. German electronics association.

Test Lead Alert


XWWarning
Attempting to make a measurement with a test lead in an incorrect terminal
might blow a fuse, damage the Meter, and cause serious personal injury.
As a reminder to check that the test leads are in the correct terminals, LEAd is momentarily displayed when
the Rotary Switch is moved to or from the mA or A position.

3
Model 77 Series IV
Calibration Information

Specifications
Accuracy is specified for 1 yr after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy
specifications take the form of ±([ % of Reading ] + [ Counts ]).
Maximum voltage between any
terminal and earth ground ................................... 1000 V
Surge protection ................................................... 8 kV peak per IEC 61010
W Fuse for mA inputs .......................................... 440 mA, 1000 V FAST Fuse
W Fuse for A inputs ............................................. 11 A, 1000 V FAST Fuse
Display
Digital.............................................................. 6000 counts, updates 4/sec
Bar Graph ....................................................... 33 segments; Updates 32/sec
Frequency....................................................... 10,000 counts
Capacitance.................................................... 1,000 counts
Altitude
Operating ............................................................ 2,000 meters
Storage ............................................................... 12,000 meters
Temperature
Operating ........................................................ -10 °C to +50 °C
Storage ........................................................... -40 °C to +60 °C
Temperature coefficient ....................................... 0.1 X (specified accuracy / °C (<18 °C or >28 °C)
Electromagnetic Compatibility
(EN 61326-1:1997) ................................................. In an RF field of 3 V/M, accuracy = specified accuracy except in temperature: specified
accuracy ± 5 °C (9 °F)
Relative Humidity (Maximum Non-condensing). 90 % to 35 °C
75 % to 40 °C;
45 % to 50 °C
Battery Life ............................................................ 400 hrs typical (Alkaline)
Size (H x W x L) ..................................................... 4.3 cm x 9 cm x 18.5 cm
Weight.................................................................... 420 g
Safety Compliance................................................ ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010 to 1000 V Measurement Category
III, 600 V Measurement Category IV
Certifications......................................................... CSA, TÜV (EN61010), UL, P, ; (N10140),VDE

4
Digital Multimeter
Specifications

Function Range Resolution Accuracy ± ( [ % of Reading ] + [ Counts ] )

AC Volts (Average 6.000 V 0.001 V


responding) 60.00 V 0.01 V 2.0 % + 2
600.0 V 0.1 V (45 Hz to 1 kHz)
1000 V 1V
DC mV 600.0 mV 0.1 mV 0.3 % + 1
DC Volts 6.000 V 0.001 V
60.00 V 0.01 V
0.3 % + 1
600.0 V 0.1 V
1000 V 1V
Continuity 600 Ω 1Ω Meter beeps at <25 Ω, beeper turns off at >250 Ω;
detects opens or shorts of 250 μs or longer.
Ohms 600.0 Ω 0.1 Ω 0.5 % + 2
6.000 kΩ 0.001 kΩ 0.5 % + 1
60.00 kΩ 0.01 kΩ 0.5 % + 1
600.0 kΩ 0.1 kΩ 0.5 % + 1
6.000 MΩ 0.001 MΩ 0.5 % + 1
50.00 MΩ 0.01 MΩ 2.0 % + 1
Diode test 2.400 V 0.001 V 1%+2
Capacitance 1000 nF 1 nF 1.2 % + 2
10.00 μF 0.01 μF 1.2 % + 2
100.0 μF 0.1 μF 1.2 % + 2
9999 μF 1 μF
[1]
10 % typical
AC Amps (Average 60.00 mA 0.01 mA
[2] [3]
responding) 400.0 mA 0.1 mA 2.5 % + 2
6.000 A 0.001 A (45 Hz to 1 kHz)
[4]
10.00 A 0.01 A
Notes:
[1] In the 9999 μF range for measurements to 1000 μF, the measurement accuracy is 1.2 % + 2.
[2] Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.
[3] 400.0 mA accuracy specified up to 600 mA overload.
[4] >10 A unspecified.

Function Range Resolution Accuracy ± ( [ % of Reading ] + [ Counts ] )


[3]
DC Amps 60.00 mA 0.01 mA
[4]
400.0 mA 0.1 mA
1.5 % + 2
6.000 A 0.001 A
[5]
10.00 A 0.01 A
[1], [2]
Hz 99.99 Hz 0.01 Hz
(ac voltage input ) 999.9 Hz 0.1 Hz 0.1 % + 1
9.999 kHz 0.001 kHz
99.99 kHz 0.01 kHz
MIN MAX AVG For dc functions, accuracy is the specified accuracy of the measurement function ±12 counts for changes longer than
350 ms in duration.
For ac functions, accuracy is the specified accuracy of the measurement function ±40 counts for changes longer than
350 ms in duration.
Notes:
[1] Frequency is specified from 2 Hz to 99.99 kHz.
[2] Below 2 Hz, the display shows zero Hz.
[3] Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.
[4] 400.0 mA accuracy specified up to 600 mA overload.
[5] >10 A unspecified.

5
Model 77 Series IV
Calibration Information

Overload Input Impedance Common Mode Rejection Ratio


Function [1] Normal Mode Rejection
Protection (Nominal) (1 kΩ Unbalanced)
Volts AC 1000 V >10 MΩ <100 pF >60 dB @ dc, 50 Hz or 60 Hz
Volts DC 1000 V >10 MΩ <100 pF >120 dB @ dc, 50 Hz or 60 Hz >60 dB @ 50 Hz or 60 Hz
2
mV 1000 V >10 MΩ <100 pF >120 dB @ dc, 50 Hz or 60 Hz >60 dB @ 50 Hz or 60 Hz
Open Circuit Test Full Scale Voltage To:
Short Circuit Current
Voltage 6.0 MΩ 50 MΩ
2
Ohms/Capacitance 1000 V < 8.0 V dc <660 mV dc <4.6 V dc <1.1 mA
2
Continuity/Diode test 1000 V <8.0 V dc 2.4 V dc <1.1 mA
[1] 10 7 V-Hz maximum.
[2] For circuits <0.3 A short circuit. 660 V for high energy circuits.

Function Overload Protection Overload


mA Fused, 440 mA, 1000 V FAST Fuse 600 mA overload for 2 minutes maximum, 10
minutes rest.
A Fused, 11 A, 1000 V FAST Fuse 20 A overload for 30 seconds maximum, 10
minutes rest.

6
Digital Multimeter
Testing the Fuses

Testing the Fuses


XWWarning
To avoid electrical shock or personal injury:
• Remove the test leads and any input signals before replacing the battery or
fuses.
• To avoid electrical shock, arc blast, or damage to the Meter, install only
fuses with the amperage, interrupt, voltage, and speed ratings specified in
Table 4.
To test the fuses (refer to Figure 1):
1. Set the Rotary Switch to e.
2. Plug a test lead into the VeG terminal and touch the probe to the 400 mA to test the 440 mA Fuse or
10 A terminal to test the 11 A Fuse.
• If the display shows a resistance value in the range shown in Figure 1, the fuse is good.
• If the display reads OL, replace the fuse and test again.
• If the display shows any other value, have the Meter serviced. See “Service Information” earlier in
this manual.

440 mA 11 A
<12 Ω OK <0.5 OK
HOLD MIN MAX RANGE
OK OK

aik12f.eps
Figure 1. Testing the Current Fuses

7
Model 77 Series IV
Calibration Information

Replacing the Fuses


XWWarning
To avoid electrical shock, arc blast, or damage to the Meter, install only fuses
with the amperage, interrupt, voltage, and speed ratings specified in
Table 4.
To replace the Fuses (refer to Figure 2):
1. Turn the Rotary Switch to OFF.
2. Disconnect the test leads and/or any connectors from the terminals.
3. Remove the four screws from the Case Bottom and separate the Case Top from the Case Bottom.
4. Remove the fuses F1 and F2 by gently prying one end loose, then slide the fuse out of its bracket, and
replace with exact replacement only.
5. Rejoin the Case Bottom, Case Top, Battery compartment door, and reinstall the four screws.

Replacing the Battery


To replace the Battery (refer to Figure 2):
1. Turn the Rotary Switch to OFF and remove the test leads from the terminals.
2. Remove the two screws from the Battery Compartment Door, and remove it from the Case Bottom.
3. Remove the Battery (B1) from the battery connector.
4. Replace the Battery with a new 9 V battery (NEDA A1604, 6F22, or 006P).
5. Reattach the Battery Compartment Door to the Case Bottom, and reinstall the two screws.

B1
F1
F2

aik13f.eps
Figure 2. Battery and Fuse Replacement

8
Digital Multimeter
Cleaning

Cleaning
XWWarning
To avoid electrical shock:
• Remove test leads and any input signals before cleaning.
• Do not reinstall the pca until it is completely dry.
• Dirt or moisture in the terminals can affect readings.

W Caution
To avoid damaging the Meter, do not use aromatic hydrocarbons or chlorinated
solvents for cleaning. These solutions will react with the plastics used in the
instruments.
Do not use detergents of any kind for cleaning the pca.
Clean the instrument case with a damp cloth and mild detergent.
The pca may be washed with isopropyl alcohol or hot deionized water and a soft brush. Remove excess
cleaning material with clean dry air at low pressure, then dry the pca at 50 °C.

Replacing the LCD


W Caution
To prevent contamination, do not handle the conductive edges of the LCD or
the LCD Elastomeric Connectors.
If the edges are contaminated, clean them with alcohol. Allow the alcohol to dry
before reassembling.
To remove and replace the LCD, perform the following procedure (refer to Figure 3):
1. Turn the rotary switch to OFF and remove the test leads from the Meter terminals.
2. Remove the four screws from the case bottom, and separate the case top from the case bottom
The circuit board remains attached to the case bottom.
3. Remove the battery compartment and disconnect the battery from the battery connector.
4. Remove the pca screw located under the 440 mA fuse, and lift the pca out of the case bottom. The
beeper is loose in the case. Exercise caution not to damage it.
5. Loosen the four screws in the area of the LCD on the back of the pca. This will facilitate reinstalling
the LCD.
6. Insert a small, flat-head screwdriver under the LCD mask edge and gently pry the LCD mask from the
snaps.
W Caution
Take care to not break the LCD with the screwdriver.
7. Lift out the LCD.
8. Make sure that all connector contact points are clean. Refer to “Cleaning” for more information.
9. Install a new LCD, taking care that it is correctly oriented.
10. Reattach the LCD Mask to the LCD assembly by snapping the LCD mask into place.
11. Tighten the 4 screws on the back of the pca.
12. Lay the pca back into the case bottom, and screw it down.
13. Reinsert the 440 mA Fuse.
14. Reinstall the battery.
15. Rejoin the case bottom, case top, battery compartment door, and reinstall the four case screws.

9
Model 77 Series IV
Calibration Information

LCD Mask

aik15f.eps
Figure 3. Removing the LCD Mask

Performance Tests
The following performance tests verify the complete operation of the Meter and check the accuracy of
each Meter function against its specifications. The recommended calibration interval is 12 months. If the
Meter fails any part of the test, calibration adjustment and/or repair is indicated.
In the performance tests, the Meter is referred to as the unit under test (UUT).

10
Digital Multimeter
Performance Tests

Required Equipment
A Fluke 5500A Multi-Product Calibrator (or equivalent) is required for the performance test procedures in
this document.
If an equivalent calibrator is used, it must meet the accuracy specifications shown in Table 1.

Table 1. Calibrator Specifications

Recommended Equipment Measurement Function Minimum required accuracy

5500A Multi-Product Calibrator DC Volts 30 mV to 1000 V ±0.075 %


(or equivalent)

DC Current 3 mA to 9 A
±0.375 %

AC Volts 50 mV to 1000 V ±0.5 % @ 45 Hz to 1 kHz

AC Current 0.5 mA to 9 A
±0.625 % @ 45 Hz to 1 kHz

Resistance 25 Ω to 5 MΩ
±0.125 %

10 MΩ to 40 MΩ
±0.5 %

Capacitance 900 nF
±0.3 %

Frequency 5 V, 50 kHz
±0.025 %

Preparing for the Performance Test


XWWarning
To avoid possible electric shock or personal injury:
• Do not perform the following procedures unless qualified to do so. Some
procedures involve the use of high voltages.
• Before handling the test connections, and in between tests, make sure the
calibrator is in standby (STBY) mode.
• Do not perform the performance test procedures unless the Meter is fully
assembled
To prepare for the performance test:
1. Make sure that you have the required equipment (refer to Table 1).
2. Warm up the calibrator as required by its specifications.
3. Allow the temperature of the UUT to stabilize at room temperature ( 23 °C ± 5 °C [73 °F ± 9 °F] ).
4. Check the fuses and Battery, and replace them if necessary. Refer to Testing the Fuses, Replacing the
Fuses, and Replacing the Battery.

11
Model 77 Series IV
Calibration Information

Performance Tests
Note
When calibrating or measuring ac current, avoid the potential for errors from coupled noise by either:
1. Shorting the VeG to the COM input
or
2. Having no connection to the VeG input.

1. Set the Meter’s Rotary Switch to the position called for in Table 2, and for all measurements other than
current (amps), connect the calibrator to the VeG and COM input terminals.
• If testing the milliamps function, connect the calibrator to the Meter 400 mA and COM input
terminals.
• If testing the amps function, connect the calibrator to the Meter 10A and COM input terminals.
2. Referring to Table 2 or Table 3, apply the indicated calibrator output voltages to the UUT for each test.
3. Verify that the UUT displays a reading that is within the limits shown in the Meter Response column.

Table 2. Performance Tests

Meter Response
Steps Test
(Switch Position) 5500 Output Lower Limit Upper Limit

1 e 500 Ω 497.3 Ω 502.7 Ω


Ohms[2]
2 50 kΩ 49.74 kΩ 50.26 kΩ

3 5 MΩ 4.974 MΩ 5.026 MΩ

4 10 MΩ 9.79 MΩ 10.21 MΩ
5 40 MΩ 39.19 MΩ 40.81 MΩ

6 R 25 Ω Beeper On
Continuity
7 250 Ω Beeper Off
8 K 50 mV 45 Hz 0.47 V ac 0.53 V ac
AC Volts
9 5 V 45 Hz 4.898 V ac 5.102 V ac
10 5 V 1 kHz 4.898 V ac 5.102 V ac
11 50 V 45 Hz 48.98 V ac 51.02 V ac
12 50 V 1 kHz 48.98 V ac 51.02 V ac
13 500 V 45 Hz 489.8 V ac 510.2 V ac
14 500 V 1 kHz 489.8 V ac 510.2 V ac
15 1000 V 45 Hz 978 V ac 1022 V ac
16 1000 V 1 kHz 978 V ac 1022 V ac
17 K Hz 5 V 50 kHz 49.94 kHz 50.06 kHz
AC Volts Frequency[1]

12
Digital Multimeter
Performance Tests

Table 2. Performance Tests (cont)

Meter Response
Steps Test
(Switch Position) 5500 Output Lower Limit Upper Limit

18 5V 4.984 V dc 5.016 V dc
19 300 V 299.0 V dc 301.0 V dc
L
20 DC Volts 1000 V 996 V dc 1004 V dc
21 -1000 V -1004 V dc -996 V dc

22 30 mV 29.8 mV dc 30.2 mV dc

23 E -300 mV -301.0 mV dc -299.0 mV dc

24 DC Millivolts 600 mV 598.1 mV dc 601.9 mV dc

25 E Apply 900 nF 887 nF 913 nF


Capacitance[1]

26 G 0.5 mA 45 Hz 0.47 mA ac 0.53 mA ac


Diode Test[1]
27 50 mA 1 kHz 48.73 mA ac 51.27 mA ac

28
F 400 mA 1 kHz 389.8 mA ac 410.2 mA ac
AC Milliamps

29 4.0 A 45 Hz 3.898 A ac 4.102 A ac

30 ? 9.0 A 1 kHz 8.75 A ac 9.25 A ac


AC Amps

31 3 mA, 0 Hz 2.93 mA dc 3.07 mA dc

32 mI 50 mA 49.23 mA dc 50.77 mA dc
DC Milliamps[1]
33 -400 mA −406.2 mA dc −393.8 mA dc

34 A 4.0 A 3.938 A dc 4.062 A dc


DC Amps1
35 -9.0 A −9.16 A dc −8.84 A dc
1. Press the YELLOW button to access this function.
2. Does not include test lead resistance.

13
Model 77 Series IV
Calibration Information

Calibration Adjustment
Perform the calibration adjustment procedures if the Meter fails the performance tests.
The Meter buttons behave as follows when the calibration mode is enabled:
H Press and hold this button to test the present function. This measurement is uncalibrated and
may be inaccurate. This is normal.
M Press and hold this button to display the required input.
B Press this YELLOW button to store the calibration value and advance to the next step. This
button is also used to exit calibration mode after the calibration adjustment sequence is
complete.

Calibration Adjustment Procedure


Use the following steps to adjust the Meter’s calibration:
Switch the Meter to E DC.
1. Turn the Meter over and find the Calibration Seal located near the top of the Meter (refer to Figure 4).
2. With a small probe, break the Calibration Seal and press the Calibration Button for 1 second. The
Meter will beep and change to the calibration mode. The display reads C-01, designating the first
calibration step. The Meter remains in calibration mode until the Rotary Function Switch is turned off.
3. Proceed through the calibration steps by entering the input value listed in the table for each step.

Note
When calibrating or measuring ac current, avoid the potential for errors from coupled noise by either:
1. Shorting the VeG to the COM input
or
2. Having NO connection to the VeG input.

Note
After pressing the yellow button, wait until the step number advances before changing the calibrator
source or turning the Rotary Switch.
4. After each input value is applied, press the yellow button to accept the value and proceed to the next
step (C-02 and so forth).
5. When the last step in a function is reached, turn the Rotary Function Knob to the next required
function. The Meter will not allow a step to be completed if the Rotary Function Knob is turned to the
wrong function.

Note
If the calibration adjustment procedure is not completed correctly, the Meter will not operate
correctly. When calibration adjustment is not performed correctly, the Meter displays the messages
CAL and Errand the Meter must be recalibrated.
The Meter is damaged and requires service if:
• CAL and Err messages continue to appear after a proper recalibration.
• EEPr and Err messages are alternating on the display.
• EEPr message appears on the display.

14
Digital Multimeter
Calibration Adjustment

Calibration Seal

Calibration Button

aik14f.eps
Figure 4. Calibration Access

15
Model 77 Series IV
Calibration Information

Table 3. Calibration Steps

Function
(Switch Position) Calibration Step Input Value
E C-01 600.0 mV dc
(DC Millivolts) 120.0 mV dc
C-02
L C-03 6.000 V dc
(DC Volts)
C-04 60.00 V dc
C-05 600.0 V dc

C C-06 600.0 mV, 60 Hz


(AC Volts) C-07 600.0 V, 60 Hz

e C-08 600.0 Ω
(Ohms)
C-09 6.000 kΩ
C-10 60.00 kΩ
C-11 600.0 kΩ
C-12 6.000 MΩ
O C-13 5.000 V dc
(Diode Test)

mA C-14 400.0 mA dc
(Milliamps)
C-15 400.0 mA ac, 60 Hz
A C-16 6.000 A dc
(Amps)
C-17 6.000 A ac, 60 Hz
* If the Meter is not connected correctly, or if the rotary switch is in the wrong position, the Meter will beep 2 times to alert the user.

16
Digital Multimeter
Replaceable Parts and Accessories

Replaceable Parts and Accessories


Replaceable parts and accessories are shown in Table 4 and Figure 5.

XW Warning
To avoid electric shock, injury, or damage to the meter, use exact replacement
parts only.

Table 4. Replaceable Parts

Item Description Part No Qty


-- (Not shown) TL75 Test Lead Set* 855705 1
-- (Not Shown)
77 Series IV Users Manual (English only) 2695884 1
77 Series IV Users Manual CD (English, and all translated versions) 2695825 1
1 Case Top 2695766 1
2 Window 648714 1
3 Shock Absorber 428441 1
4 Keypad 1560052 1
5 LCD Mask 2695775 1
6 LCD 1560856 1
7 LCD Light Pipe 1564806 1

8 WFuse, 11 A, (fast acting), 1000 V ac/dc, minimum interrupt rating 17 kA 803293 1


9 WFuse, 440 mA (fast acting), 1000 V ac/dc, minimum interrupt rating 10 kA 943121 1
10 PCA Screw 1626602 1

11 AC Shield 648755 1

12 AC Shield and Internal Cover screws 448456 5


13 Case Bottom (includes bottom shield and calibration button) 2095692 1
14 Battery, 9 V (Alkaline, 9 V, 0-200 mA) 614487 1
15 Battery Compartment Door 1564799 1
16 Case Screws 832246 4
17 Tilt Stand 648961 1
18 LCD Elastomeric Connector 650264 2
19 Internal Cover 1564786 1
20 Battery Connector 1988201 1
21 Beeper 2041050 1
22 Swtich Detent Spring 822643 1
23 Switch Knob 648706 1
24 Calibration Button 1564889 1
* Fluke accessories are available from your authorized Fluke distributor.

17
Model 77 Series IV
Calibration Information

17
16

15

13
14

12 24

21 11

18

19
20

10

9 7

8
6

22 4
2

3
1

23

aik16f.eps
Figure 5. Replaceable Parts and Accessories

18
Digital Multimeter
Replaceable Parts and Accessories

Lifetime Limited Warranty


Each Fluke 20, 70, 80, 170, and 180 Series DMM will be free from defects in material and workmanship for its lifetime. As
used herein, “lifetime” is defined as seven years after Fluke discontinues manufacturing the product, but the warranty
period shall be at least ten years from the date of purchase. This warranty does not cover fuses, disposable batteries,
damage from neglect, misuse, contamination, alteration, accident or abnormal conditions of operation or handling, including
failures caused by use outside of the product’s specifications, or normal wear and tear of mechanical components. This
warranty covers the original purchaser only and is not transferable.
For ten years from the date of purchase, this warranty also covers the LCD. Thereafter, for the lifetime of the DMM, Fluke
will replace the LCD for a fee based on then current component acquisition costs.
To establish original ownership and prove date of purchase, please complete and return the registration card
accompanying the product, or register your product on http://www.fluke.com. Fluke will, at its option, repair at no charge,
replace or refund the purchase price of a defective product purchased through a Fluke authorized sales outlet and at the
applicable international price. Fluke reserves the right to charge for importation costs of repair/replacement parts if the
product purchased in one country is sent for repair elsewhere.
If the product is defective, contact your nearest Fluke authorized service center to obtain return authorization information,
then send the product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB
Destination). Fluke assumes no risk for damage in transit. Fluke will pay return transportation for product repaired or
replaced in-warranty. Before making any non-warranty repair, Fluke will estimate cost and obtain authorization, then
invoice you for repair and return transportation.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY
CAUSE OR THEORY. AUTHORIZED RESELLERS ARE NOT AUTHORIZED TO EXTEND ANY DIFFERENT
WARRANTY ON FLUKE’S BEHALF. Since some states do not allow the exclusion or limitation of an implied warranty or
of incidental or consequential damages, this limitation of liability may not apply to you. If any provision of this warranty is
held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the
validity or enforceability of any other provision.
Fluke Corporation Fluke Europe B.V.
P.O. Box 9090 P.O. Box 1186
Everett, WA 98206-9090 5602 BD Eindhoven
U.S.A. The Netherlands

2/02

19