) Envirotech Calibration Laboratory

Quantity Measured Range/Frequency *Calibration Measurement Remarks


/ Instrument Capability (±)

FLUID FLOW CALIBRATION


I. FLOW MEASURING DEVICES

1. Air Flow Rate$ 35 LPM to 100 LPM 1.02 % Using PD Meter by


Comparison Method

2. Air Flow Rate$ 0.6 m3/min to 1.5 m3/min 0.78 % Using PD Meter by
Comparison Method,
USEPA IO2.1

3. Air Flow Rate# 0.6 m3/min to 1.5 m3/min 1.10 % Using Top Loading
Calibrator by
Comparison Method,
USPEA IO2.1

0.25 lpm to 5 lpm 2.36 % Using Diaphragm Gas


5 lpm to 35 lpm 35 1.31 % Meter by Comparison
lpm to 65 lpm 0.65 % Method

3 lpm to 20 lpm 1.00 % Using Low Flow


Calibrator by
Comparison Method,
40CFR Part 50
Appendix L

10 ml/min to 200 ml/min 0.9 ml/min Using Piston Flow


Calibrator by
0.2 lpm to 3 lpm 0.032 lpm Comparison Method,
ASTM D5337-0
3 lpm to 30 lpm 0.12 lpm

4. Velocity$ 3 m/s to 35 m/s 1.12 % Reading Using ‘L’ Type Pitot


Tube in Wind Tunnel
by Comparison method

5. Volume$ 0.18 m3 to 0.5 m3 at Flow 1.95 % Using PD Meter by


Rate of Comparison Meter
34 lpm to 100 lpm
2.) Enpro Envirotech & Engineers Pvt Ltd

Fluid Flow Parameter

 Ultrasonic Flow Meter (5 -450 m3/h)


 LFE Gas Flow Calibrator/1 to 75 LPM)
 LFE Gas Flow Calibrator/1 to 75 LPM)
 Hot Wire Anemometer (0.2 to 35 m/s)
 Electronic Gas Volume Meter (0.1 to 2 LPM)
 Pitot Tube (L Type) (Up to 30 m/s)
3.) NPK calibration laboratory

Although some parameters are yet to be covered under NABL accreditation, NPKLAB fulfils
all requirements of traceability and technical competency for its unique facility for Flow
calibration:

Parameter & Instruments / Equipment Range of


calibration
Liquid Flow measuring instruments 0.05 to 25 ml/min
Gas flow measuring instruments using dry air, Nitrogen, and 0.5 to 2000 ml/min
Hydrogen as medium gas
Anemometer / Air velocity meter 0.2 to 40 m/s
Wet gas flow meter 1000 ml

Measurand or Reference Measurement range


Material/Type of * Calibration
S.No Discipline / Group instrument or material to Calibration or and additional and
Measurement parameters where Measurement
be calibrated or Method or procedure applicable(Range and
measured / Quantity Capability(CMC
Frequency) )(±)
Measured /Instrument
Using pitot tube with
FLUID FLOW- Air velocity meters, sensors differential manometer and
126 FLOW & anemometers wind tunnel by comparison 4.5 m/s to 30 m/s 3.01% of reading
MEASURING method
DEVICES

Air velocity meters, sensors & Using Reference


127 FLUID FLOW- anemometers. anemometer and wind tunnel 0.3 m/s to 4 m/s 7.5%
FLOW by comparison method
MEASURING
DEVICES

Calibration by gravimetric
method using constant flow
Liquid flowmeter / flow pump and semi micro 0.1 ml/min to 5 ml/min 0.0058 ml/min to
128 FLUID FLOW- measuring sensors 0.0038 ml/min
weighing balance (0.1mg) in
FLOW
auto mode.
MEASURING
DEVICES
4.) SBEM PVT LTD, SBEM FLOW LABORATORY
Measurand or Reference Measurement range and
Material/Type of additional parameters * Calibration and
Calibration or
S.No Discipline / Group instrument or material to Measurement Method where applicable Measurement Capability
be calibrated or measured (Range and Frequency) (CMC) (±)
or procedure
/ Quantity Measured
/Instrument
LOCATION 1-UNIT NO 9, 691/A/2, BIBWEWADI INDUSTRIAL ESTATE, PUNE-SATARA ROAD, PUNE, MAHARASHTRA, INDIA
Permanent Facility

FLUID FLOW- Using 3000 kg weighing


FLOW Liquid Mass /Flow system as per ISO 4185 By
1 MEASURING Measuring Devices Weighing Method 1000 kg to 3000 kg 0.05%
DEVICES

FLUID FLOW- Using 3000 kg weighing


FLOW Liquid Mass Flow Rate / system as per ISO 4185 By
2 MEASURING Flow Measuring Devices Weighing Method 1.5 t/hr to 220 t/hr 0.17%
DEVICES

FLUID FLOW- Using 3000 kg weighing


FLOW Liquid Volume /Flow system as per ISO 4185 By
3 MEASURING Measuring Devices Weighing Method 1000L to 3000L 0.05%
DEVICES

FLUID FLOW- Using 3000kg Weighing


FLOW Liquid Volume Flow Rate System as per ISO 4185 By 1.5 m3/hr to 220 m3/hr
4 MEASURING (Water) / Flow measuring Weighing Method 0.5%
DEVICES devices

Measurand or Reference Measurement range and


Material/Type of additional parameters * Calibration and
Calibration or
S.No Discipline / Group instrument or material to Measurement Method where applicable Measurement Capability
be calibrated or measured (Range and Frequency) (CMC) (±)
or procedure
/ Quantity Measured
/Instrument
LOCATION 2 - GAT NO. 326, INDUSTRIAL SHED NO E, SHRIRAMNAGAR, GAUD DARA ROAD, KHEDSHIVAPUR, PUNE,
MAHARASHTRA, INDIA
Permanent Facility

Using Electromagnetic
FLUID FLOW- Flowmeter by Comparison
FLOW Liquid Volume Flow Rate Method (Comparative 500 m3/hr to 3500 m3/hr
1 MEASURING (Water) Method) 0.2%
DEVICES
Using Electromagnetic
FLUID FLOW- Flowmeter by Comparison
FLOW Liquid Volume Method (Comparative
2 MEASURING Measured Quantity Method) 7 m3 to 35 m3 0.2%
DEVICES (Water)

5.) Calibration Laboratory – Aarohi embedded systems Pvt. Ltd.

Measurand or Reference Measurement range


Material/Type of * Calibration
S.No Discipline / Group instrument or material to Calibration or and additional and
Measurement Method parameters where Measurement
be calibrated or or procedure applicable (Range
measured / Quantity Capability
and Frequency)
Measured /Instrument (CMC) (±)

Fluid flow -Quantity by


Mass (EMF, Rota meter, Using 10000 kg Weighing
1 FLUID FLOW- USFM, Turbine, Oval Gear, System by Gravimetric 1500 kg to 10000 kg 0.086% rdg
FLOW Vortex, Orifice Plate, Method as per ISO: 4185
MEASURING Venturi, Critical flow
DEVICES venturi, Flow Nozzles, Pitot
tube, Wedge type, Annubar,
Coriolis FM,Thermal, Posi

Fluid flow- Quantity by


Mass (EMF, Rota meter,
USFM, Turbine, Oval Gear,
FLUID FLOW- Using 1500 kg Weighing
2 Vortex, Orifice Plate, 300 kg to 1500 kg 0.092% rdg
FLOW System by Gravimetric
Venturi, Critical flow
MEASURING Method as per ISO 4185
venturi, Flow Nozzles, Pitot
DEVICES
tube, Wedge type, Annubar,
Coriolis FM,Thermal, Posi

Using Electro Magnetic


3 FLUID FLOW- Digital Water Flow Meter Flow Meter By Comparison 700 m3/hr to 2500 m3/hr
FLOW Method 0.2% Rdg
MEASURING
DEVICES

Fluid flow-Quantity By
FLUID FLOW- Mass (EMF, Rota meter, Using 400 kg Weighing
4 FLOW MEASURING USFM, Turbine, Oval Gear, System by Gravimetric 100 kg to 400 kg 0.08% rdg
DEVICES Vortex, Orifice Plate, Method as per ISO 4185
Venturi, Critical flow
venturi, Flow Nozzles, Pitot
tube, Wedge type, Annubar,
Coriolis FM,Thermal, Posi.
Liquid (Water Medium)
FLUID FLOW- Mass Flow Rate (EMF, Rota Using 400 kg Weighing
FLOW MEASURING meter, USFM, Turbine, Oval System by Gravimetric
5 DEVICES Gear, Vortex, Orifice Plate, Method as per ISO 4185 0.5 t/h to 25 t/h 0.11% rdg
Venturi, Critical flow
venturi, Flow Nozzles,
Wedge type, Annubar,
Coriolis FM, Thermal, Posi.
disp

Liquid (Water Medium)


FLUID FLOW- Mass Flow Rate (EMF, Rota Using 1500 kg Weighing
FLOW MEASURING meter, USFM, Turbine, Oval System by Gravimetric
6 DEVICES Gear, Vortex, Orifice Plate, Method as per ISO 4185 5 t/h to 125 t/h 0.095% rdg
Venturi, Critical flow
venturi, Flow Nozzles,
Wedge type, Annubar,
Coriolis FM, Thermal, Posi.
disp

Liquid (Water Medium)


FLUID FLOW- Mass Flow Rate (EMF, Rota Using 10000 kg Weighing
FLOW MEASURING meter, USFM, Turbine, Oval System by Gravimetric
7 DEVICES Gear, Vortex, Orifice Plate, Method as per ISO 4185 50 t/h to 1000 t/h 0.11% rdg
Venturi, Critical flow
venturi, Flow Nozzles,
Wedge type, Annubar,
Coriolis FM, Thermal, Posi.
disp

Liquid Volume Flow Rate


FLUID FLOW- (EMF, Rota meter, USFM, Using 400 kg Weighing
8 FLOW MEASURING Turbine, Oval Gear, Vortex, System by Gravimetric 0.5 m3/h to 25 m3/h 0.11% rdg
DEVICES Orifice Plate, Venturi, Method as per ISO 4185
Critical flow venturi, Flow
Nozzles, Wedge type,
Annubar, Coriolis
FM,Thermal, Posi. dis
Liquid Volume Flow Rate
FLUID FLOW- (EMF, Rota meter, USFM, Using 10000 kg Weighing
FLOW MEASURING Turbine, Oval Gear, Vortex, System by Gravimetric 50 m3/h to 1000 m3/h
9 DEVICES Orifice Plate, Venturi, Method as per ISO 4185 0.12% rdg
Critical flow venturi, Flow
Nozzles, Pitot tube, Wedge
type, Annubar, Coriolis
FM,Thermal, Posi. dis

Liquid(Water) Volume Flow


FLUID FLOW- Rate (EMF, Rota meter, Using 1500 kg Weighing
FLOW MEASURING USFM, Turbine, Oval Gear, System by Gravimetric 5 m3/hr to 125 m3/hr
10 DEVICES Vortex, Orifice Plate, Method as per ISO 4185 0.093% rdg
Venturi, Critical flow
venturi, Flow Nozzles, Pitot
tube, Wedge type, Annubar,
Coriolis FM,Thermal, Posi.
dis

Quantity by Volume (EMF,


FLUID FLOW- Rota meter, USFM, Turbine, Using 400 kg Weighing
11 FLOW MEASURING Oval Gear, Vortex, Orifice System by Gravimetric 100 L to 400 L 0.11% rdg
DEVICES Plate, Venturi, Critical flow Method as per ISO 4185
venturi, Flow Nozzles,
Wedge type, Annubar,
Coriolis FM,Thermal, Posi.
Disp.)

Quantity by Volume (EMF,


FLUID FLOW- Rota meter, USFM, Turbine, Using 1500 kg Weighing
12 FLOW MEASURING Oval Gear, Vortex, Orifice System by Gravimetric 300 L to 1500 L 0.092% rdg
DEVICES Plate, Venturi, Critical flow Method as per ISO 4185
venturi, Flow Nozzles, Pitot
tube, Wedge type, Annubar,
Coriolis FM,Thermal, Posi.
Disp.
Quantity by Volume (EMF,
FLUID FLOW- Rota meter, USFM, Turbine, Using 10000 kg Weighing
13 FLOW MEASURING Oval Gear, Vortex, Orifice System by Gravimetric 1500 L to 10000 L 0.12% rdg
DEVICES Plate, Venturi, Critical flow Method as per ISO 4185
venturi, Flow Nozzles, Pitot
tube, Wedge type, Annubar,
Coriolis FM,Thermal, Posi.
Disp.)

Measurand or Reference Measurement range


Material/Type of * Calibration
S.No Discipline / Group instrument or material to Calibration or and additional and
Measurement Method parameters where Measurement
be calibrated or or procedure applicable (Range
measured / Quantity Capability
and Frequency) (CMC) (±)
Measured /Instrument
Site Facility
FLUID FLOW- Digital and Analog Water Portable Ultrasonic Flow
FLOW Flow Meter (Range DN meter & Clamp on sensors 211 m3/hr to 2500 m3/hr
3 1.0% Rdg
MEASURING 100 - DN 4000) by comparison method
DEVICES

FLUID FLOW- Digital and Analog Portable Ultrasonic Flow


FLOW Water Flow Meter meter & Clamp on sensors 30 m3/hr to 1280 m3/hr
4 1.0% Rdg
MEASURING (Range DN 50 - DN by comparison method
DEVICES 300)

FLUID FLOW- Digital or Analog Water Portable Ultrasonic Flow


FLOW Flow Meter (Range: meter & Clamp on sensors
5 1 m3/hr to 93 m3/hr 1.0% Rdg
MEASURING DN15 - DN 65) by comparison method
DEVICES
6.) WORLDONE INSTRUMENTS, #42, M.E.C LAYOUT, 8TH CROSS,
DUGGALAMMA
TEMPLE ROAD, PEENYA INDUSTRIAL AREA, BENGALURU, KARNATAKA, INDIA
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated or Measurement
Method or procedure where applicable(Range
measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

LOCATION 2 - #47, GANAPATHI NAGAR, RAJGOPAL NAGAR MAIN ROAD, PEENYA INDUSTRIAL
AREA, BENGALURU, KARNATAKA, INDIA
Permanent Facility

FLUID FLOW- Using Flow Control unit


FLOW Volume flow rate (Media: and Top Loading Orifice 257 lpm to 1700 lpm
1 Air) Top Loading calibrator 1.72%
MEASURING calibrator by Comparison
DEVICES method

Using Wind tunnel


FLUID FLOW- control unit,
2 FLOW Anemometer, Pitot Tube 0.23 m/s to 5 m/s 0.5m/s
Anemometer and Pitot
MEASURING & Velocity measuring
Tube -L Type by
DEVICES Devices
Comparison method

Using Wind tunnel


FLUID FLOW- control unit,
3 FLOW Anemometer, Pitot Tube 5 m/s to 19 m/s 0.67m/s
Anemometer and Pitot
MEASURING & Velocity measuring
Tube -L Type by
DEVICES Devices
Comparison method

FLUID FLOW- Using Flow Control unit


FLOW Volume flow (Media: Air) and Gas Flow meter by
4 Dry Gas Meter- (Digital 0.001 m³ to 20 m³ 1.2%
MEASURING Comparison method
DEVICES /Mechanical)
Volume flow (Media: Water) Using Flow control Unit
FLUID FLOW- Analog and Digital Flow and Electro Magnetic Flow
5 FLOW meters, Flow measuring Meter by Comparison 0.5 m³ to 12 m³ 0.80%
MEASURING Devices method
DEVICES

Volume flow (Media: Water) Using Flow control Unit


FLUID FLOW- Analog and Digital Flow and Electro Magnetic Flow
6 FLOW meters, Flow measuring Meter by Comparison 12 m³ to 100 m³ 0.70%
MEASURING Devices method
DEVICES

Volume flow rate (Media :


Air) Analog & Digital Flow Using Flow Control unit
meter, Rota meter, Flow 1 ml/minute to 1000 1.49%
7 FLUID FLOW- and Mass Flow Meter by ml/minute
measuring Devices- Comparison method
FLOW
(Air/Gas)
MEASURING
DEVICES
Volume flow rate (Media :
8 Air) of Digital and Analog Using Flow Control unit
Flow Meter, Rotameter & 257 lpm to 1700 lpm 1.6%
FLUID FLOW- and Top Loading Orifice
Flow measuring devices Calibrator by Comparison
FLOW
MEASURING method
DEVICES
Volume flow rate (Media : Using Flow Control unit
FLUID FLOW- Air) of Respirable Dust and Top Loading Orifice 0.3 m³/minute to 1.7
9 FLOW Sampler and PM10 Sampler calibrator by Comparison m³/minute 1.66%
MEASURING method
DEVICES

Volume flow rate (Media :


Water), Analog and Digital Using Flow control Unit
Flow meters, and Electro Magnetic Flow 200 lpm to 2500 lpm
10 0.70%
,Ultrasonic Flow Meter, rota Meter by Comparison
FLUID FLOW- meters and Flow measuring method
FLOW Devices
MEASURING
DEVICES
Volume flow rate (Media : Using Flow control Unit, 1"
Water), Analog and Digital (25 mm) Electro Magnetic
Flow meters, Ultrasonic Flow Meter, Ultrasonic
11 Flow Meter, rota meters and Flow Meter by Comparison 5 lpm to 200 lpm 0.7%

FLUID FLOW- Flow measuring Devices method


FLOW
MEASURING
DEVICES
Volume flow rate (Media :
Water), Rotameter, Analog Using Weighing Balance &
12 and Digital Flow meters and 5 ml/minute to 5000 0.40%
FLUID FLOW- Digital Time Totaliser by ml/minute
Flow measuring Devices Gravimetric method
FLOW
MEASURING
DEVICES
Volume Flow Rate: (Media: Using Flow Control unit
Air) Analog & Digital Flow and Gas Flow meter by
Meter, Rotameter, Dry Gas Comparison method as per
13 Meter, PM2.5 Sampler and ASTM D3195, ASME 5 lpm to 300 lpm 1.95lpm

FLUID FLOW- Flow measuring Devices MFC-


21.2-2010/ ASTM D5337-
FLOW 04
MEASURING
DEVICES
Volume Flow Rate:(Media: Using Flow Control unit
Air) Analog & Digital Flow and Gas Flow meter by
Meter, Rotameter, Dry Gas Comparison method as per
14 Meter, PM2.5 Sampler and ASTM D3195, ASME 0.1 lpm to 5 lpm 0.65%

FLUID FLOW- Flow measuring Devices MFC-


FLOW 21.2-2010/ ASTM D5337-
MEASURING 04
DEVICES

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated or Measurement
Method or procedure where applicable (Range
measured / Quantity Capability CMC)
and Frequency)
Measured /Instrument (±)

LOCATION 2 - #47, GANAPATHI NAGAR, RAJGOPAL NAGAR MAIN ROAD, PEENYA INDUSTRIAL
AREA, BENGALURU, KARNATAKA, INDIA
Site Facility

FLUID FLOW- Using Flow Control unit


FLOW Volume flow rate (Media: and Top Loading Orifice 257 lpm to 1700 lpm
1 Air) Top Loading calibrator 1.72%
MEASURING calibrator by Comparison
DEVICES method

FLUID FLOW- Using Flow Control unit


2 FLOW Volume flow (Media: Air) and Gas Flow meter by 0.001 m³ to 20 m³ 1.2%
MEASURING Dry Gas Meter- (Digital Comparison method
DEVICES /Mechanical)

Volume flow rate (Media:


FLUID FLOW- Air) Analog & Digital Flow Using Flow Control unit
3 1 ml/minute to 1000 1.49%
FLOW meter, Rota meter, Flow and Mass Flow Meter by
ml/minute
MEASURING measuring Devices- (Air/Gas) Comparison method
DEVICES

Volume flow rate (Media:


FLUID FLOW- Air) of Digital and Analog Using Flow Control unit
4 257 lpm to 1700 lpm 1.6%
FLOW Flow Meter, Rotameter & and Top Loading Orifice
MEASURING Calibrator by Comparison
DEVICES Flow measuring devices method
FLUID FLOW- Using Flow Control unit
FLOW Volume flow rate (Media: and Top Loading Orifice 0.3 m³/minute to 1.7
5 MEASURING Air) of Respirable Dust calibrator by Comparison m³/minute 1.66%
DEVICES Sampler and PM10 Sampler method

Volume flow rate (Media:


Water), Analog and Digital
6 FLUID FLOW- Flow meters, Using Ultrasonic flow meter 5 lpm to 13500 lpm 1.66%
FLOW with multiple clamp on
MEASURING ,Ultrasonic Flow Meter, rota sensors by Comparison
DEVICES meters and Flow measuring method
Devices
Using Flow Control unit
Volume Flow Rate: (Media: and Gas Flow meter by
Air) Analog & Digital Flow Comparison method as per
7 FLUID FLOW- Meter, Rotameter, Dry Gas 5 lpm to 300 lpm 1.95lpm
ASTM D3195, ASME
FLOW Meter, PM2.5 Sampler and MFC-
MEASURING Flow measuring Devices
DEVICES 21.2-2010/ ASTM D5337-
04
Using Flow Control unit
Volume Flow Rate:(Media: and Gas Flow meter by
Air) Analog & Digital Flow Comparison method as per
8 FLUID FLOW- Meter, Rotameter, Dry Gas 0.1 lpm to 5 lpm 0.65%
ASTM D3195, ASME
FLOW Meter, PM2.5 Sampler and MFC-
MEASURING Flow measuring Devices
DEVICES 21.2-2010/ ASTM D5337-
04

7.) ABB INDIA LTD LIQUID FLOW CALIBRATION LAB

Measurand or Reference Measurement range


Material/Type of * Calibration
S.No Discipline / Group instrument or material to Calibration or and additional and
Measurement Method parameters where Measurement
be calibrated or or procedure applicable (Range
measured / Quantity Capability
and Frequency) (CMC) (±)
Measured /Instrument
Permanent Facility
FLUID FLOW- Using Electromagnetic
FLOW Flow Meter by 3
1 Water Flow 0 to 6 m /hour 0.15 %
MEASURING Comparison Method
DEVICES

FLUID FLOW- Using Electromagnetic


FLOW Flow Meter by 3
2 Water Flow 200 m /hour to 600 0.15 %
MEASURING Comparison Method 3
DEVICES m /hour

FLUID FLOW- Using Electromagnetic


FLOW Flow Meter by 3
3 Water Flow 30 m /hour to 58 0.15 %
MEASURING Comparison Method 3
DEVICES m /hour
FLUID FLOW- Using Electromagnetic
FLOW Flow Meter by 3
4 Water Flow 58 m /hour to 200 0.15 %
MEASURING Comparison Method 3
DEVICES m /hour

FLUID FLOW- Using Electromagnetic


FLOW Flow Meter by 3
5 Water Flow 6 m /hour to 30 0.15 %
MEASURING Comparison Method 3
DEVICES m /hour

FLUID FLOW- Using Electromagnetic


FLOW Flow Meter by 3
6 Water Flow 600 m /hour to 0.15 %
MEASURING Comparison Method 3
DEVICES 1600 m /hour