Manual Rosemount 6888c in Situ Combustion Oxygen Analyzer en 1260898

Download as pdf or txt
Download as pdf or txt
You are on page 1of 86

Manual

00809-0100-4891, Rev AB
January 2023

Rosemount™ 6888C In Situ Oxygen


Analyzer
For Hazardous Areas
Essential instructions
Read this page before proceeding!
Emerson™ designs, manufactures, and tests its products to meet many national and international standards. Because
these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure
they continue to operate within their normal specifications. The following instructions must be adhered to and
integrated into your safety program when installing, using, and maintaining Emerson products. Failure to follow the
proper instructions may cause any one of the following situations to occur: loss of life, personal injury, property
damage, damage to this instrument, and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Emerson representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
• Install equipment as specified in the installation instructions of the appropriate instruction manual and per
applicable local and national codes. Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the
product.
• Use only factory documented components for repair. Tampering and unauthorized substitution and parts can
affect product performance and cause unsafe operation of your process.

NOTICE
The Field Communicator must be upgraded to System Software 2.0 with graphic license for operation with the
Rosemount™ 6888C O2 Transmitter. The AMS software must be upgraded to AMS 8.0 or above. Contact Emerson's
Global Service Center (GSC) at +1-800-833-8314 to upgrade the Field Communicator software to System Software 2.0
with graphic license.

Symbols

Earth (ground) terminal

Protective conductor terminal

Risk of electrical shock

Hot surface

Refer to Manual.

The following definitions apply to Warnings, Cautions, and Notices found throughout this publication.

WARNING
Highlights an operation or maintenance procedure, practice, condition, statement, etc., which if not strictly observed,
could result in injury, death, or long-term health hazards of personnel.

CAUTION
Highlights an operation or maintenance procedure, practice, condition, statement, etc., which if not strictly observed,
could result in damage to or destruction of equipment or loss of effectiveness.

2
NOTICE
Highlights an essential operating procedure, condition, or statement.

WARNING
EXPLOSION
Do not open when an explosive atmosphere may be present.

WARNING
ELECTRIC SHOCK
Do not open while energized.

WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’
equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict
physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the
facility.

CAUTION
EQUIPMENT DAMAGE
For the standard housing probe and direct replacement probe, only use supply cables & certified cable glands rated >
105 °C (221 °F).

CAUTION
EQUIPMENT DAMAGE
For the autocal housing, only use supply cables and certified cable glands rated > 85 °C (185 °F).

Special conditions for safe use (ATEX) and conditions of acceptability (CSA)
1. The unit is to be connected to the supply mains by qualified personnel in accordance with local and national
codes.

2. Mounting flange temperature shall not exceed 190 °C during combustion process.

3. The 6888C O2 Analyzers are used with the 6888 Xi Advanced Electronics (associated equipment not part of this
certification) which must be installed in a safe area.

4. Calibration air lines and reference air lines shall not contain pure oxygen or combustible gas other than inert/
oxygen gas mixture of which oxygen represents no more than that normally present in air.

5. The pressure within the enclosure and gas lines shall not be higher than 1.1 times the atmospheric pressure
during the normal operations of the equipment.

6. Fasteners property class must be A2-70 Stainless Steel

7. Flameproof joints are not intended to be repaired.

3
Safety instructions
SAFETY INSTRUCTIONS FOR THE WIRING AND INSTALLATION OF THIS APPARATUS
The following safety instructions apply specifically to all EU member states. They should be strictly adhered to in order
to ensure compliance with the Low Voltage Directive. Non-EU states should also comply with the following unless
superseded by local or national standards.
1. Adequate earth connections should be made to all earthing points, internal and external, where provided.

2. After installation or troubleshooting, all safety covers and safety grounds must be replaced. The integrity of all
earth terminals must be maintained at all times.

3. Main supply cords should comply with the requirements of IEC227 or IEC245.

4. All wiring shall be suitable for use in an ambient temperature of greater than 75 °C (167 °F).

5. All cable glands used should be of such internal dimensions as to provide adequate cable anchorage.

6. To ensure safe operation of this equipment, connection to the mains supply should only be made through a
circuit breaker which will disconnect all circuits carrying conductors during a fault situation. The circuit breaker
may also include a mechanically operated isolating switch. If not, then another means of disconnecting the
equipment from the supply must be provided and clearly marked as such. Circuit breakers or switches must
comply with a recognized standard, such as IEC947. All wiring must conform with any local standards.

7. All graphical symbols used in this product are from one or more of the following standards: EN61010, IEC417,
and ISO3864.

8. Where equipment or labels are marked "Do Not Open While Energized" or similar, there is a danger of ignition
in areas where an explosive atmosphere is present. This equipment should only be opened when the power
is removed and adequate time as specified on the label or in the instruction manual has been allowed for the
equipment to cool down - and then only by trained service personnel.

The following table illustrates symbols used on equipment or covers and explains the meaning of each.

Description Symbol

Where equipment or covers are marked with the symbol to the right,
hazardous voltages are likely to be present beneath. These covers should
only be removed when power is removed from the equipment - and then
only by trained service personnel.

Where equipment or covers are marked with the symbol to the right, there
is a danger of hot surfaces beneath. These covers should only be removed
by trained service personnel when power is removed from the equipment.
Certain surfaces may remain hot to the touch.

Where equipment or covers are marked with the symbol to the right, refer to
the Manual for instructions.

4
Manual Contents
00809-0100-4891 January 2023

Contents
Chapter 1 General information.................................................................................................. 7
1.1 Overview....................................................................................................................................... 7
1.2 System configurations................................................................................................................ 7

Chapter 2 Typical system package.......................................................................................... 11


2.1 Rosemount 6888C Hazardous Area Probe product matrix..................................................12

Chapter 3 Specifications............................................................................................................15
3.1 Measurement specifications.................................................................................................... 15
3.2 Environmental specifications...................................................................................................16
3.3 Optional General Purpose 6888 Xi Electronics...................................................................... 16
3.4 Installation specifications - probe........................................................................................... 17
3.5 Installation specifications - Rosemount 6888 Xi with Rosemount 6888C Transmitter
probe............................................................................................................................................ 18
3.6 Installation specifications for traditional architechture Rosemount 6888 Xi for use
with DR or other probe.............................................................................................................. 19

Chapter 4 Install.........................................................................................................................21
4.1 Mechanical installation............................................................................................................. 21
4.2 Mounting the Rosemount 6888 Xi General Purpose Advanced Electronics...................... 21
4.3 Electrical installation................................................................................................................. 25
4.4 Connecting the Hazardous Area Transmitter Probe to the Rosemount General
Purpose 6888 Xi Electronics...................................................................................................... 26
4.5 Connecting the analyzer probe to the single-channel safe area Xi and flame safety
interlock....................................................................................................................................... 28
®
4.6 Connect the analyzer probe with integral autocalibration to HART communications... 33

4.7 Connect the analyzer probe with integral autocalibration to FOUNDATION Fieldbus
communications......................................................................................................................... 35
4.8 Connecting the traditional architecture to the direct replacement hazardous area
replacement probe (no electronics inside)..............................................................................41
4.9 Connect traditional architecture system to the direct replacement probe....................... 43
4.10 Pneumatic installation............................................................................................................ 48

Chapter 5 Power up................................................................................................................... 51


5.1 Powering up the standalone Rosemount 6888 Transmitter probe (without
Rosemount 6888 Xi)....................................................................................................................51
5.2 Power up the Rosemount 6888 Analyzer with single/dual channel or single
channel and flame safety interlock Rosemount 6888Xi.........................................................52
5.3 Power up the Rosemount 6888C direct replacement probe (no electronics inside)
with traditional architecture Rosemount 6888Xi....................................................................52

Manual 5
Contents Manual
January 2023 00809-0100-4891

5.4 Run Rosemount 6888Xi Quick Start Wizard........................................................................... 53


5.5 Re-initiating Rosemount 6888 Xi wizard.................................................................................54

Chapter 6 Calibration................................................................................................................ 55
6.1 Manual/semi-automatic calibration........................................................................................55
6.2 Fully automatic calibration.......................................................................................................56

Chapter 7 HART® menu trees................................................................................................... 59


Chapter 8 Maintenance and service........................................................................................ 65
8.1 Overview.....................................................................................................................................65
8.2 Maintenance intervals.............................................................................................................. 65
8.3 Repair..........................................................................................................................................66

Chapter 9 Replacement parts...................................................................................................81


Chapter 10 Service support.........................................................................................................83
Appendix A Product certifications..............................................................................................85

6 Rosemount 6888C
Manual General information
00809-0100-4891 January 2023

1 General information
1.1 Overview
The 6888 is Rosemount's latest combustion flue gas oxygen analyzer. This
product is intended for measuring the flue gases resulting from any combustion
process.
This product uses an in situ sensor, i.e., the sensor is placed at the end of the
probe, and the probe extends directly into the flue gas duct or stack at a given
length. The sensor is like a thermocouple, generating its own millivolt signal
based on the difference between a reference gas (ambient or instrument air -
always 20.95% O2) and the flue gases being measured. This manual covers the
Rosemount 6888C, which is the hazardous area probe version.
This probe can be configured as a blind stand-alone transmitter where HART® or
FOUNDATION Fieldbus™ communications can be used as a method of accessing
the electronics for setup, operation, and diagnostics or with one of several remote
electronics options, as noted below in System configurations. Wiring diagrams for
each of these configurations follow in Electrical installation.
There are several different arrangements of probes, electronics, and features that
are explained below and in the wiring diagrams.

1.2 System configurations


1.2.1 Hazardous Area Transmitter probe only
The Rosemount 6888 probe has the electronics in the blue housing that controls
the heater temperature and also amplifies the raw O2 millivolt signal to a linear
4-20 mA. The 4-20 mA signal lines can be run directly to the control room and also
power the transmitter electronics. As with most other Rosemount transmitters
measuring pressure, temperature, and flow, setup is conducted through HART
communications via a Field Communicator or Asset Management Solutions (AMS).

1.2.2 Standard housing Hazardous Area Transmitter


probe with General Purpose Rosemount 6888 Xi
Electronics
The Rosemount 6888 Xi Electronics serve as a local operator interface unit,
with a back-lit display and keypad. It is capable of two channels, serving two
Rosemount 6888 probes. The Rosemount 6888 Xi also carries these optional
advanced features:
• Fully automatic calibration. Requires Rosemount 6888 Xi O2 Cal Auto
calibration system

Manual 7
General information Manual
January 2023 00809-0100-4891

• Loss of flame contact for powering down the heater in the event of a flame-out
condition in a furnace.
• Heaterless operation at process temperatures above 550 °C (1022 °F). This
feature will also permit operation above the heater setpoint of 736 °C (1357
°F). Sensing cell life will be shortened by operation above 800 °C (1472 °F),
however.
• The plugged diffuser diagnostic operates by measuring the return-to-process
rate after calibration gas has been stopped. This feature also includes auto gas
switching when the reading settles out versus waiting for configured gas flow
time to expire.
• Stoichiometer - If a furnace goes into a reducing condition (0% O2), this feature
will determine how far.
• Programmable reference - Permits more accurate readings at near-ambient O2
levels (20.95% O2).
• A cal check capability. New calibration values are not automatically stored after
a calibration. An accept/reject calibration feature can be enabled or disabled so
that the technician or operator can decide to accept or reject a potentially large
change in calibration values.
• Tolerance check that will alarm if the wrong test gases are being used or if a
bottle runs out in the middle of a calibration. Care must be taken to ensure gas
1 and gas 2 calibration gases are properly configured if the tolerance check
feature is enabled.

1.2.3 Hazardous Area Rosemount 6888C probe with


General Purpose Rosemount Xi Electronics with
flame safety interlock
A flame safety interlock by Emerson™ is available for heater power disconnect
whenever there is a loss of the process flame or a heater runaway condition
(heater over-temperature) in the O2 probe. This input is internally powered by the
Rosemount 6888 Xi and is actuated via a dry contact input from the user's flame
scanner. A closed contact indicates a flame is present. An open contact indicates a
loss of flame. This feature is also available with the Integral autocal housing.

1.2.4 Hazardous Area Rosemount 6888C probe with


Integral Autocal General Purpose Rosemount 6888
Xi Electronics
This probe contains gas-switching solenoids so that the Rosemount 6888 Xi
Electronics can control the introduction of calibration gases. Calibrations can be
intiated via a calibration recommended diagnostic, time since last calibration,
manually via external dry contact, HART communications, or from the Rosemount
6888 Xi local operator interface keypad. The integral autocal feature can only be
implemented when the probe is used with a Rosemount 6888 Xi.

8 Rosemount 6888C
Manual General information
00809-0100-4891 January 2023

1.2.5 Hazardous Area Rosemount 6888C probe with


Integral Autocal and Foundation Fieldbus (FF)
communications
This probe contains gas-switching solenoids that can control the introduction of
calibration gases for calibration. Calibrations can be initiated automatically via
a calibration recommended diagnostic, time since last calibration, or manually
via optional Xi keypad, FF communications via the Field Communicator, or AMS
console. Unlike the HART transmitter electronics, the FF version can execute
automatic calibrations either with or without the optional Rosemount 6888 Xi
Electronics. Likewise, advanced features can be implemented either with or
without the optional Xi.

1.2.6 Hazardous Area Rosemount 6888C direct


replacement (DR) probe with General Purpose
Traditional Architecture Rosemount 6888 Xi
Electronics
Here there are no electronics inside the probe head, so the raw sensor signals for
the heater thermocouple and zirconium oxide O2 sensor are sent to a remote
Rosemount 6888 Xi Electronics. The Rosemount 6888 Traditional Architecture
Electronics will also directly apply power to the probe heater in order to maintain
the correct sensor temperature. This arrangement calls for a 7-conductor cable
to carry this power and the sensor signals. Maximum length for this cable is
200 feet. This probe will also operate on previous Westinghouse/Rosemount
electronics (World Class and Oxymitter), as well as many competitive electronics.

WARNING
HAZARDOUS AREAS
Cables are supplied with Exd Rated Glands; the cable is not ExD rated, and
installation in a hazardous area is the responsibility of the installer/customer.
Cables must be installed in accordance with national and local electrical codes.

1.2.7 Wireless capability


Both the analyzer electronics in the head of the probe and the Rosemount 6888Xi
Electronics communicate over HART communications and can implement wireless
communications via Emerson Wireless 775 THUM™ Adapter.

Manual 9
General information Manual
January 2023 00809-0100-4891

10 Rosemount 6888C
Manual Typical system package
00809-0100-4891 January 2023

2 Typical system package


Figure 2-1: Typical System Package

A. Optional Rosemount SPS 4001B or Rosemount IMPS Autocalibration Sequencer


(Requires use of Rosemount 6888 Xi Advanced Electronics option)
B. Quick Start Guide
C. Optional mounting or adapter plate
D. Optional traditional architecture cable
E. HART® Field Communicator package (optional)
F. Optional reference & calibration gas accessories
G. Rosemount 6888C probe
H. Optional Rosemount 6888 XI Advanced Electronics

Manual 11
Typical system package Manual
January 2023 00809-0100-4891

2.1 Rosemount 6888C Hazardous Area Probe


product matrix
Table 2-1: Rosemount 6888C Product Matrix
Model Description
Rosemount O2 Transmitter approved for hazardous locations
6888C
Measurement
1OXY Oxygen, standard sensing cell
2OXY Oxygen, high sulfur resistant sensing cell
Probe length(1)
1A 18-in. probe, 3-in. ANSI Class 150 flange
1D 18-in. probe, DIN flange
2A 3-ft probe, 3-in. ANSI Class 150 flange
2D 3-ft probe, DIN flange
3A 6-ft probe, 3-in. ANSI Class 150 flange
3D 6-ft probe, DIN flange
Diffuser
1 Snubber diffuser, 400 °C (572 °F) with process flame arrestors
2 Ceramic diffuser, 825 °C (1517 °F) with process flame arrestors
3 Alloy diffuser 705 °C (1301 °F) with both ambient and process flame arrestors
Housing and electronics
1HT Standard housing, transmitter electronics, HART communications
2HT Integral autocal, transmitter electronics, HART communications
4FF Integral autocal, transmitter electronics, FOUNDATION Fieldbus™ communications
5DR Standard housing, direct replacement, requires remote electronics
6DRY Standard housing, direct replacement, for use with YEW Electronics
Certifications
A ATEX/IECEx
C CSA
Mounting Plate
00 None
04 New installation - square weld plate with ANSI 2 in.(50.8 mm)- 150# studs and flange (2.5 in.
(63.5 mm) process hole required)
05 New installation - square weld plate with DIN studs and flange (2.5 in. (63.5 mm) process
hole required)
08 Adapter to existing ANSI 3 in. (76.2 mm), 150# flange

12 Rosemount 6888C
Manual Typical system package
00809-0100-4891 January 2023

Table 2-1: Rosemount 6888C Product Matrix (continued)


09 Adapter to existing ANSI 4 in. (101.6 mm), 150# flange
10 Adapter to existing ANSI 6 in. (152.4 mm), 150# flange
11 Adapter to existing ANSI 3 in. (76.2 mm), 300# flange
12 Adapter to existing ANSI 4 in. (101.6 mm), 300# flange
99 Special adapter - provide existing flange dimensions, including thru-hole diameter
Manual calibration accessories
00 None
01 Calibration and reference gas flowmeters and reference regulator/filter diffuser
02 Calibration/reference panel
Stoichiometer function(2)
0 No
1 Yes
Programmable reference function(2)
0 No
1 Yes
Extended temperature reference function(2)
0 No
1 Yes
Diffuser warning function(2)
0 No
1 Yes

(1) Flanges are flat-faced and for mounting only. Flanges are not pressure-rated.
(2) FOUNDATION Fieldbus only (for HART versions, order this feature with Rosemount 6888 Xi Electronics)

Manual 13
Typical system package Manual
January 2023 00809-0100-4891

14 Rosemount 6888C
Manual Specifications
00809-0100-4891 January 2023

3 Specifications
3.1 Measurement specifications
3.1.1 Net O2 range
Variable 0 - 10% to 0-50% (Xi electronics of 0-50% range)

3.1.2 Accuracy in oxidizing condition


±0.75% of reading at 0.05% O2, whichever is greater. ±1% of full scale for 0-10%
ranges and above.

3.1.3 Lowest detectable limit


0.02% O2

3.1.4 Process temperature effect


Less than 0.05% O2 from 100 to 700 °C (212 to 1292 °F)

3.1.5 System speed of response to calibration gas


Initial response in less than 3 seconds; T90 in less than 8 seconds. Response to
process gas changes varies depending on process gas velocity and particulate
loading of the diffuser.

3.1.6 Calibration validity


Presentation of calibration gases matches the bottle value to within 0.02% O2

3.1.7 Accuracy in reducing conditions (requires


stoichiometer feature)
±10% of reading or 0.1% O2, whichever is different size

3.1.8 System response in reducing conditions (requires


stoichiometer feature)
Going from oxidizing to reducing - T90 in 120 seconds
Going from reducing to oxidizing - T90 in 30 seconds

Manual 15
Specifications Manual
January 2023 00809-0100-4891

3.2 Environmental specifications


3.2.1 Transmitter probe
Process-wetted materials are 316L stainless steel.

3.2.2 Transmitter probe process temperature limits


0 to 705 °C (32 to 1300 °F)
550 to 825 °C (1022 to 1517 °F) with Xi heaterless operation feature(1)
Optional bypass and jacket accessories permit operation to 1050 °C (1922 °F).

3.2.3 Probe electronics ambient temperature limits


-40 to 70 °C (-40 to 158 °F)

3.2.4 Temperature limit as measured inside probe


electronics
-40 to 85 °C (-40 to 185 °F)

3.2.5 DR probe, no electronics inside, ambient


temperature limits
-40 to 90 °C (-40 to 194 °F)

3.3 Optional General Purpose 6888 Xi


Electronics
3.3.1 Material
NEMA® 4X, polycarbonate material

3.3.2 Rosemount 6888 Xi ambient temperature limits


-20 to -50 °C (-4 to 122 °F)

(1) Reduced cell life can be expected if operated continuously at temperatures above 705 °C (1300 °F).

16 Rosemount 6888C
Manual Specifications
00809-0100-4891 January 2023

3.3.3 Rosemount 6888 Xi temperature limits as


measured inside the electronics housing
-20 to 70 °C (-4 to 158 °F)

3.4 Installation specifications - probe


3.4.1 Probe mounting flange
Vertical or horizontal-3 in. 150 6 in. (152.4 mm) bolt circle

CAUTION
Flanges are flat-faced and for mounting only. Flanges are not pressure-rated. A
2.5 in. (63.55 mm) diameter hole in the process is required.

Spool piece P/N 3D39761G02 is available to offset probe electronics housing from
hot ductwork.
Many adapter flanges are available to mate to existing flanges.

3.4.2 Probe lengths and approximate shipping weights


18 in. (457 mm) package: 16 lb (7.3 kg)
3 ft (0.91 m) package: 21 lb (9.5 kg)
6 ft (1.83 m) package: 27 lb (12.2 kg)
9 ft (2.74 m) package: 33 lb (15.0 kg)
12 ft (3.66 m) package: 39 lb (17.7 kg)

3.4.3 Reference air (optional)


Maximum 2 scfh (1 L/min), clean, dry, instrument-quality air (20.95% O2) regulated
to 5 psi (34 kPa)

3.4.4 Calibration
Semi-automatic or automatic

3.4.5 Cal gases


0.4% O2 and 8% O2 balance N2 recommended. Instrument air may be used as a
high cal gas but is not recommended. 100% nitrogen cannot be used as the low
cal gas. No explosive gases or gases with % O2 greater than that found in ambient
air (i.e., 20.95% O2) can be used.

Manual 17
Specifications Manual
January 2023 00809-0100-4891

3.4.6 Calibration gas flow


5 scfh (2.3 L/min) at 15 psi, maximum

3.4.7 Heater electrical power


120/240 V ±10%, 50/60 Hz, 1/2 in. - 14 NPT conduit ports

3.4.8 Traditional architecture cable


200 ft (61 m) maximum length

3.4.9 Power consumption of transmitter probe/integral


autocal probe
776 W maximum during warm-up

3.5 Installation specifications - Rosemount


6888 Xi with Rosemount 6888C Transmitter
probe
3.5.1 Electrical power of optional Rosemount 6888 Xi
Electronics
120/240 V ±10%, 50/60 Hz

3.5.2 Power consumption of Rosemount 6888 Xi


10 W maximum

3.5.3 Rosemount 6888 Xi alarm relays


2 provided- 2 amps, 20 Vdc

3.5.4 Rosemount 6888 Xi optional loss of flame contact


Removes heater power

3.5.5 Electrical noise

3.5.6 Traditional architecture cable


200 ft (61 m) maximum length

18 Rosemount 6888C
Manual Specifications
00809-0100-4891 January 2023

3.5.7 Transmitter electrical 4-20 mA power


12 - 42 Vdc (looped-powered from the control room or from the Rosemount 6888
Xi box)

3.6 Installation specifications for traditional


architechture Rosemount 6888 Xi for use
with DR or other probe
3.6.1 Power consumption of Rosemount 6888 Xi
120/240 V ±10%, 50/60 Hz, 260/1020 VA max for 120 V probes
120 V ±10%, 50/60 Hz, 450 VA max for 44 V probes

3.6.2 Alarm relay outputs


2 provided - 2 Amperes, 30 Vdc, Form-C

3.6.3 Optional loss of flame input


Internally powered input to remove heater power actuated via dry contact output
form proof of flame device.
Emerson has satisfied all obligations coming from the European legislation to
harmonize the product requirements in Europe.

Manual 19
Specifications Manual
January 2023 00809-0100-4891

20 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

4 Install
WARNING

Failure to follow safety instructions could result in serious injury or death.


Before installing this equipment, read the Essential instructions at the front of this
Reference Manual.

WARNING

ELECTRIC SHOCK
Failure to install covers and ground leads could result in serious injury or death.
Install all protective equipment covers and safety ground leads after installation.

4.1 Mechanical installation


Most combustion processes run only slightly negative or positive in pressure, so
that the probe flange is for mechanical mounting only. The probe is not rated for
high pressures. If this is a new installation, Emerson can supply a weld plate for
welding to the flue gas duct.

WARNING

Electric shock
Failure to install covers and ground leads could result in serious injury or death.
Install all protective covers and safety ground leads after installation.

4.2 Mounting the Rosemount 6888 Xi General


Purpose Advanced Electronics
Follow the steps below to mount the Rosemount 6888 Xi.

Prerequisites
The Rosemount 6888 Xi Advanced Electronics is available in a panel mounting or
wall/pipe mounting configuration. Refer to Figure 4-1 or Figure 4-2 for the panel,
wall, or pipe mounting details. The Rosemount 6888 Xi is rated for use in general
purpose locations only.
Ensure all components are available to install the Rosemount 6888 Xi.

Procedure
1. Select a mounting location near or removed from the O2 Probe.

Manual 21
Install Manual
January 2023 00809-0100-4891

Consider the temperature limitations of the Rosemount 6888 Xi (see


Rosemount 6888 Xi ambient temperature limits) when selecting the
mounting location.
2. Mount the Roseount 6888 Xi at a height convenient for viewing and
operating the interface.
Approximately 5 ft (1.5 m) is recommended.
3. The keypad window on the Rosemount 6888 Xi may have interior and
exterior protective membranes. Remove the protective membranes prior
to use of the Rosemount 6888 Xi enclosure.

CAUTION
EQUIPMENT DAMAGE
Failure to remove the protective membranes may cause the display to
appear distorted. The membrane may be difficult or impossible to remove
after extended use at elevated temperatures.

22 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

Figure 4-1: Wall/surface mount and pipe mount

A. 4X cover screw
B. Front panel
C. 6X 1/2-in. NPT conduit openings
D. Mounting bracket
E. U-bolts
F. 2-in. pipe supplied by customer
Dimensions are in inches with millimeters in parentheses.

Manual 23
Install Manual
January 2023 00809-0100-4891

Figure 4-2: Panel mount

A. Maximum panel thickness 0.375 (9.52)


B. Panel mount gasket
C. 4X mounting brackets and screws provided
D. 6X 1/2 in. NPT conduit openings

NOTICE
Dimensions are in inches with millimeters in parentheses.

24 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

NOTICE
The front panel is hinged at the bottom. The panel swings down for easy access to
wiring locations.

4.3 Electrical installation


All wiring must conform to local and national codes. Multiple wiring diagrams are
shown in this section. Always refer to the diagrams that apply to your analyzer
configuration and disregard all other wiring diagrams.

WARNING
ELECTRIC SHOCK
Disconnect and lock out power before connecting the power supply.
Install all protective covers and safety ground leads after installation. Failure to
install covers and ground leads could result in serious injury or death.
To meet the safety requirements of IEC 61010-1 (EC requirement) and ensure safe
operation of this equipment, connect the main electrical power supply through a
circuit breaker (min 10 A) which will disconnect all current-carrying conductors
during a fault situation. This circuit breaker should also include a mechanically
operated isolating switch. If not, keep another external means of disconnecting
the supply from the equipment located close by. Circuit breakers or switches must
comply with a recognized standard such as IEC 947.

WARNING
Before installing this equipment, read the Essential instructions at the front of this
instruction manual. Failure to follow safety instructions could result in serious
injury or death.

CAUTION
EQUIIPMENT DAMAGE
If external loop power is used, the power supply must be a safety extra low
voltage (SELV) type.

NOTICE
To maintain proper earth grounding, ensure a positive connection exists between
the analyzer housing and earth. The connecting ground wire must be 14 AWG
minimum.

NOTICE
Line voltage, signal, and relay wiring should be rated for at least 105 °C (221 °F).

Manual 25
Install Manual
January 2023 00809-0100-4891

NOTICE
If metal conduit is used with the Rosemount 6888 Xi, the conduit should be
reliably bonded to protective earth (PE). The grounding plate inside the
Rosemount 6888 Xi is not bonded to PE and does not provide adequate
grounding.

4.4 Connecting the Hazardous Area


Transmitter Probe to the Rosemount
General Purpose 6888 Xi Electronics
The Rosemount 6888 Xi Electronics serve as an operator interface unit with a
back-lit display and keypad. It is capable of two channels, serving two Rosemount
6888 probes.

Procedure
1. Remove cover screws from the front cover of the Rosemount 6888 Xi. Swing
down the front cover of the interface box.
2. Pull out the I/O board on the right side of the card rack inside the
Rosemount 6888Xi.
If your system is configured to operate two transmitter probes, there are
two I/O interface boards.
3. See Figure 4-3. Connect the 4-20 mA signal wires at J4 of the I/O board.
Attach the supplied ferrite clamp over the 4-20 mA OUT wires that extend
past the shield.

NOTICE
Installation of the ferrite clamp over the 4-20 mA OUT wires is required for
compliance with the European EMC directive.

4. Terminate the shield of the 4-20 mA signal wires at the designated ground
terminal of the Rosemount 6888 Xi. Do not allow bare shield wires to
contact the circuit boards. Insulate the shield wires prior to termination.
5. Connect the signal wires from the SPS or IMPS (if used) to the applicable
terminals of J3. Refer to the SPS or IMPS instruction manual for wiring
details.
6. Reinstall the I/O board at the card rack of the Rosemount 6888 Xi.
7. If your system is configured for two channel operation, repeat steps 2
through 6 to connect the other probe's signal wires.
8. Remove the probe's connector from the power supply board located on the
left-hand side of the card rack inside the Rosemount 6888 Xi.
9. Connect the line, or L1, wire to the L1 terminal and the neutral, or L2, wire
to the N terminal.
10. Reinstall the power supply connector in the power supply board.

26 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

Figure 4-3: Single/dual channel wiring diagram

Manual 27
Install Manual
January 2023 00809-0100-4891

NOTICE
1. Except for JP5, JP7, and JP8 on IO board, jumper and switch settings are factory set and are shown for
reference only.
2. IO board 4-20 mA/HART® loop power settings
• JP5
— Pins 1-2: internal loop power Rosemount 6888 Xi to Rosemount 6888 Transmitter
— Pins 2-3: external power Rosemount 6888 Xi to Rosemount 6888 Transmitter (requires 2,500
resistor across J4 PR+ to PR-)
• JP7/JP8
— Pins 1-2: internal power from Rosemount 6888 Xi to DCS
— Pins 2-3: external power Rosemount 6888 Xi to DCS

4.5 Connecting the analyzer probe to the


single-channel safe area Xi and flame
safety interlock
A flame safety interlock by Emerson is available for the heater power disconnect
whenever there is a loss of the process flame or a heater runaway condition
(heater over-temperature) in the O2 probe. This input is internally powered by the
Rosemount 6888 Xi and is actuated via a dry contact output from the user's flame
scanner. A closed contact indicates a flame is present. An open contract indicates
a loss of flame.

Procedure
1. Refer to Figure 4-4. Connect the signal wires from the burner management
system flame status output to the flame status input terminals of J2.
You must supply the flame status sensing device. Refer to the applicable
OEM documents for signal wiring details.
2. Remove the J1 and J2 connectors from the AC relay board.
3. Connect the AC line input to the J1 connector.
4. Connect the AC power to the Rosemount 6888C probe to the J2 connector.
5. Reinstall connector J1 and J2 to the AC relay board.
Rosemount 6888C product matrix
Compare the configuration matrix below to the model number on the probe tag
to confirm the features present in this specific probe.
Table 4-1: Housing and Electronics
1HT Standard housing, transmitter electronics, HART communications
2HT Integral autocal, transmitter electronics, HART communications

28 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

Table 4-1: Housing and Electronics (continued)


4FF Integral autocal, transmitter electronics, Fieldbus communications
5DR Standard housing, direct replacement, no electronics
6DRY Standard housing, direct replacement, YEW electronics

Rosemount 6888 Xi product matrix - safe area only


Compare the configuration matrix below to the model number on the probe tag
to confirm the features present in this specific probe.
Table 4-2: Remote Type
1OXY Single channel O2
2OXY Single channel O2 with flame safety interlock for heater
3OXY Dual channel O2
4OXY Single channel O2 traditional architecture for 120 V probes

Manual 29
Install Manual
January 2023 00809-0100-4891

Figure 4-4: Single channel with flame safety wiring diagram

30 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

NOTICE
A. See Rosemount 6888 Xi Instruction Manual for additional installation and operating instructions.
B. All wiring marked with an asterisk (*) is factory wiring inside the Rosemount 6888 Xi.
C. Except for JP5, JP7, and JP8 on the input/output (IO) board, jumper and switch settings are factory set and are
shown for reference only.
D. IO board: 4-20 mA/HART loop power settings
• JP5
— Pins 1-2: internal power Rosemount 6888Xi to Rosemount 6888 analyzer
— Pins 2-5: external power Rosemount 6888Xi to Rosemount 6888 analyzer (requires 2,500 resistor across
J4, PR+ to PR-)
• JP7/JP8
— Pins 1-2: internal power to DCS
— Pins 2-3: external power Rosemount 6888Xi to DCS

Manual 31
Install Manual
January 2023 00809-0100-4891

Figure 4-5: Rosemount 6888 Xi front and bottom views

A. Power supply board


B. Channel #2 IO board
C. Shield ground
D. Channel #1 IO board
E. AC input to P/S
F. Plug
G. Channel #2 alarm relay, SPS/IMPS
H. Channel #2 4-20 mA/HART output
I. Channel #1 Alarm relay, SPS/IMPS
J. Channel #1 4-20 mA HART output

32 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

4.6 Connect the analyzer probe with integral


autocalibration to HART® communications
This probe contains gas-switching solenoids, Field Communicator, or Asset
Management Solutions (AMS) software, so that the Rosemount 6888Xi Electronics
can control the introduction of calibration gases.
You can initiate calibrations in the following ways
• Automatically with a calibration recommended diagnostic
• Automatically with time since last calibration
• Manually with external dry contact
• Manually with HART communications
• With the Rosemount 6888Xi local operator interface (LOI). You can only use
integral autocalibration with a Rosemount 6888Xi.

Procedure
1. Remove the two covers from the analyzer.
2. Refer to Figure 4-6. Connect the line (L1 wire) to the L1 terminal, the neutral
(L2 wire) to the L2/N terminal, and the ground wire to the ground lug.
The analyzer accepts line voltage at 120/240 Vac ±10 percent, 50/60 Hz. No
setup is required.
3. Connect the 4-20 mA signal wires from the Rosemount 6888Xi to the
connections in the side chamber of the analyzer.
Do not connect the signal wires to the terminals in the main chamber where
the AC input wires are connected. Use a shielded twisted wire pair. Do not
allow bare shield wires to contact the circuit boards. Insulate the shield
wires prior to termination. The 24 VDC loop power is sourced from the
Rosemount 6888Xi.
4. Terminate the shield at both the probe and the Rosemount 6888Xi.
Note
The 4-20 mA signal represents the O2 value and also powers the
probe-mounted electronics. Superimposed on the 4-20 mA signal is
HART information accessible through a Field Communicator or Asset
Management Solutions (AMS) software.

5. Reinstall both covers on analyzer.


6. Follow the remaining electrical installation instructions for the Rosemount
6888Xi included with your system configuration.

Manual 33
Install Manual
January 2023 00809-0100-4891

Figure 4-6: Integral autocalibration and HART communications

A. Ferrite clamp
B. Signal
C. Test points
D. #8 pan htd scr (internal ground)
E. Power
F. Test point group

34 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

Note
A. Except for JP5, JP7, and JP8 on the input/output (IO) board, jumper and switch settings are factory set and are
shown for reference only.
B. IO board: 4-20 mA/HART loop power settings
• JP5
— Pins 1-2: internal power Rosemount 6888Xi to Rosemount 6888 analyzer
— Pins 2-5: external power Rosemount 6888Xi to Rosemount 6888 analyzer (requires 2,500 resistor across
J4, PR+ to PR-)
• JP7/JP8
— Pins 1-2: internal power to DCS
— Pins 2-3: external power Rosemount 6888Xi to DCS

Note
IO board Channel 2 is a duplicate of Channel 1.

4.7 Connect the analyzer probe with integral


autocalibration to FOUNDATION™ Fieldbus
communications
This probe contains gas-switching solenoids enabling the Rosemount 6888 Xi
Electronics, Field Communicator, or Asset Management Solutions (AMS) software
to control the introduction of calibration gases.
You can initiate calibrations in the following ways
• Automatically with a calibration recommended diagnostic
• Automatically with time since last calibration
• Manually with external dry contact
• With the Rosemount 6888 Xi local operator interface (LOI). You can only use
integral autocalibration with a Rosemount 6888 Xi.

Procedure
1. Remove the two covers from the analyzer.
2. Connect the line (L1 wire) to the L1 terminal, the neutral (L2) wire to the
L2/N terminal, and the ground wire to the ground lug.
The analyzer accepts line voltage at 120/240 VAC ±10 percent, 50/60 Hz. No
setup is required.
3. Connect the FOUNDATION Fieldbus wires from the analyzer side housing to
the FF segment.

Manual 35
Install Manual
January 2023 00809-0100-4891

WARNING

The Rosemount 6888 probe is not rated as intrinsically safe (IS) and will
render any IS or FISCO segment it is wired to as non-IS.
Use a shielded twisted wire pair.
Do not allow bare shield wires to contact the circuit boards.

4. Terminate the shield at both the probe and the Rosemount 6888 Xi
Advanced Electronics.
Note
The FOUNDATION Fieldbus signal represents the O2 value and also powers the
probe-mounted electronics.

5. Reinstall both covers on analyzer.


6. Follow the remaining electrical installation instructions for the Rosemount
6888 Xi included with your system configuration.
Rosemount 6888C ordering information
Compare the ordering information below to the model number on the probe tag
to confirm the features present in this specific probe.
Table 4-3: Housing and electronics ordering information
1HT Standard housing, analyzer electronics, HART communications
2HT Integral autocal, analyzer electronics, HART communications
4FF Integral autocal, analyzer electronics, Fieldbus communications
5DR Standard housing, direct replacement, no electronics
6DRY Standard housing, direct replacement, YEW electronics

36 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

Figure 4-7: Integral autocalibration and FOUNDATION Fieldbus communication


without optional Rosemount 6888Xi

A. Signal
B. Not used
C. #8 pan htr scr (internal ground)
D. Power
E. Probe test point group

NOTICE
A. All wiring marked with an asterisk (*) is factory wiring inside the Rosemount
6888 Xi.
B. Except for JP5, JP7, and JP8 on the input/output (IO) board, jumper and switch
settings are factory set and are shown for reference only.
C. IO board: 4-20 mA/HART loop power settings
• JP5
— Pins 1-2: internal power Rosemount 6888Xi to Rosemount 6888 analyzer
— Pins 2-5: external power Rosemount 6888Xi to Rosemount 6888 analyzer
(requires 2,500 resistor across J4, PR+ to PR-)
• JP7/JP8
— Pins 1-2: internal power to DCS
— Pins 2-3: external power Rosemount 6888Xi to DCS

Rosemount 6888Xi ordering information: safe area only


Compare the ordering information below to the model number on the probe tag
to confirm the features present in this specific probe.

Manual 37
Install Manual
January 2023 00809-0100-4891

Table 4-4: Remote type ordering information


1OXY Single channel O2
2OXY Single channel O2 with flame safety interlock for heater
3OXY Dual channel O2
4OXY Single channel O2 traditional architecture for 120 V probes

38 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

Figure 4-8: Integral autocalibration and FOUNDATION Fieldbus communication


with optional Rosemount 6888Xi

A. Ribbon cable to display board J2 Sensor 1


B. Signal
C. HART® connection (Used as a communication bus from probe analyzer
electronics to optional Rosemount 6888Xi. Not accessible to Field Communicator
or AMS)
D. #8 pan htr scr (internal ground)
E. Power
F. Probe test point group
IO board switch/jumpers
Jumper settings

Manual 39
Install Manual
January 2023 00809-0100-4891

JP1 Pins 2-3


JP2 Pins 2-3
JP5 Pins 1-2: internal power
Pins 2-3: external power
JP7 Pins 1-2: internal power
Pins 2-3: external power
JP8 Pins 1-2: internal power
Pins 2-3: external power

SW4 switch settings


• Position 1: Off
• Position 2: Off
• Position 3: Off
• Position 4: Off

NOTICE
A. Except for JP5, JP7, and JP8 on the input/output (IO) board, jumper and switch
settings are factory set and are shown for reference only.
B. IO board 4-20 mA/HART loop power settings
JP5
• Pins 1-2: internal power Rosemount 6888Xi to Rosemount 6888 Analyzer.
• Pins 2-3: external power Rosemount 6888Xi to Rosemount 6888 Analyzer
(requires 2,500 resistor across J4, PR+ to PR-).
JP7/JP8
• Pins 1-2: internal power Rosemount 6888Xi to DCS.
• Pins 2-3: external power Rosemount 6888Xi to DCS.

40 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

Figure 4-9: Rosemount 6888 Xi front and bottom view

A. Power supply board


B. AC relay board
C. Shield ground
D. IO board
E. AC input to power supply
F. Plug
G. AC input to relay board
H. AC output to analyzer
I. Alarm relay, Rosemount SPS

4.8 Connecting the traditional architecture to


the direct replacement hazardous area
replacement probe (no electronics inside)
Here there are no electronics inside the probe head, so the raw sensor signals
for the heater thermocouple and zirconium O2 sensor are sent to a remote
Rosemount 6888 Xi Electronics. The Rosemount 6888 Xi electronics will also
directly apply power to the probe heater in order to maintain the correct sensor

Manual 41
Install Manual
January 2023 00809-0100-4891

temperature. This arrangement calls for a 7-conductor cable to carry this power
and the sensor signals. Maximum length for this cable is 200 feet.

Procedure
1. Remove cover from probe.
2. Feed all DR probe wiring through the conduit port of probe.
3. Refer to Figure 4-10. Connect DR probe heater power leads to DR probe
connector.
4. Connect O2 signal and thermocouple wires to DR probe connector.
Model 6888C product matrix
Compare the configuration matrix below to the model number on the probe tag
to confirm the features present in this specific probe.
Table 4-5: Housing and Electronics
1HT Standard housing, transmitter electronics, HART communications
2HT Integral autocal, transmitter electronics, HART communications
4FF Integral autocal, transmitter electronics, Fieldbus communications
5DR Standard housing, direct replacement, no electronics
6DRY Standard housing, direct replacement, YEW electronics

Model 6888Xi product matrix - safe area only


Compare the configuration matrix below to the model number on the probe tag
to confirm the features present in this specific probe.
Table 4-6: Remote Type
1OXY Single channel O2
2OXY Single channel O2 with flame safety interlock for heater
3OXY Dual channel O2
4OXY Single channel O2 traditional architecture for 120 V probes

42 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

Figure 4-10: Traditional architecture with direct replacement hazardous area replacement
probe (no electronics inside)

NOTICE
A. See Rosemount 6888 Xi Instruction Manual for additional installation and operating instructions.
B. All wiring marked with an asterisk (*) is factory wiring inside the Rosemount 6888 Xi.
C. Except for JP5, JP7, and JP8 on the input/output (IO) board, jumper and switch settings are factory set and are
shown for reference only.

4.9 Connect traditional architecture system to


the direct replacement probe
Use a traditional architecture configuration to provide for remote location of the
analyzer electronics. All electronics are housed inside the Rosemount 6888Xi. A
multi-conductor power/signal cable connects the probe to the Rosemount 6888Xi.

Manual 43
Install Manual
January 2023 00809-0100-4891

Use the following procedure to connect the traditional architecture probe to the
Rosemount 6888Xi.

NOTICE
The traditional architecture cable is provided at the specified length and is ready
for installation. The cable glands must be properly terminated to maintain EMC/
electromagnetic interference (EMI) noise protection.

Procedure
1. Run the seven-conductor cable between the traditional architecture probe
and the installation site for the Rosemount 6888Xi.
Use new cable conduit or trough as needed.
2. Install the cable and lead wires to the probe per manufacturer's
instructions.
3. Install the cable at the probe housing and at the Rosemount 6888Xi
enclosure.
a) Unscrew locking nut from gland assembly and slide locking nut back
along cable.
b) Pull the gland body away from the plastic insert.

NOTICE
Use care not to damage the cable shield braid.

c) Insert the cable wires into the proper entry port in either the probe
housing or the Rosemount 6888Xi enclosure.
d) At the probe housing, apply PTFE tape or similar sealing compound
to the tapered pipe threads. Thread the gland body into the probe
housing until properly seated.
e) At the Rosemount 6888Xi enclosure, insert the gland body into the
left front cable port from the inside of the enclosure. Use the rubber
O-ring provided to seal the cable port.
f) Ensure the cable shield braid is evenly formed over the gray insert.
When properly formed, the braid should be evenly spaced around
the circumference of the insert and not extend beyond the narrow
diameter portion.

g) Carefully press the gray insert into the gland body.


The grooves on the insert should align with similar grooves inside the
gland body. Press the insert in until it bottoms out in the gland body.

h) Slide the locking nut up and thread it onto the gland body. Tighten
the locking nut so the rubber grommet inside the plastic insert
compresses against the cable wall to provide an environmental seal.

44 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

4. At the Rosemount 6888Xi, connect the cable leads to the connectors on the
analyzer input/output (IO) board.

Figure 4-11: Traditional architecture general purpose cable gland assembly

WARNING
ELECTRIC SHOCK
Disconnect and lock out power before working on any electrical components.
There is voltage up to 240 Vac.

Manual 45
Install Manual
January 2023 00809-0100-4891

Figure 4-12: Traditional architecture with direct replacement (DR) probe (no
electronics inside)

Note
A. See the Rosemount 6888 Xi Reference Manual for additional installation and
operating instructions.
B. All wiring marked with an asterisk (*) is factory wiring inside the Rosemount
6888 Xi.
C. Except for JP7 and JP8 on IO board, jumper and switch settings are factory set
and are shown for reference only.

46 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

Figure 4-13: Traditional architecture with direct replacement probe (no


electronics inside)

A. Power supply board


B. DR board
C. Shield ground
D. IO board
E. Plug
F. Probe cable
G. AC input
H. Alarm relay, Rosemount SPS
I. 4-20 mA/HART® output

Manual 47
Install Manual
January 2023 00809-0100-4891

4.10 Pneumatic installation


4.10.1 Reference air package
After the Rosemount 6888C is installed, connect the reference air set to the
analyzer unit.
Refer to the schematic diagram in Figure 4-14 for a locally assembled reference
air supply.
Instrument air (reference air): 5 psig (0.34 barg) minimum, 8 psig (0.55 barg)
maximum at 2.0 scfh (1.01 L/min) maximum; less than 40 parts per million
total hydrocarbons. Set regulator outlet pressure at 5 psig (0.34 barg). Use the
reference air set or the optional Rosemount SPS 4001B to supply reference air.
Reference air is recommended, or the reference air fittings can be left open to
atmosphere. Rosemount SPS 4001B autocalibration boxes contain reference air
sets.

Figure 4-14: Plant schematic diagram, standard housing

A. Vent
B. Calibration gas: ¼-in. tube
C. To analyzer
D. Reference air flow meter
E. Instrument air supply: 10 psig (0.69 barg) to 80 psig (5.52 barg) pressure
F. ¼-in.-18 NP female inlet connection
G. ¼-in. or 6 mm O.D. tubing (supplied by customer)
H. Reference gas: ¼-in. tube

48 Rosemount 6888C
Manual Install
00809-0100-4891 January 2023

Figure 4-15: Plant air schematic diagram, accessory housing

A. Vent
B. Calibration gas 1: ¼-in. tube
C. Calibration gas 2: ¼-in. tube
D. To analyzer
E. Reference air flow meter
F. Instrument air supply: 10 psig (0.69 barg) to 80 psig (5.52 barg) pressure
G. ¼-in.-18 NPT female inlet connection
H. ¼-in. or 6 mm O.D. tubing (supplied by customer)
I. Reference gas: ¼-in. tube

Manual 49
Install Manual
January 2023 00809-0100-4891

Figure 4-16: Manual calibration panel

Note
Dimensions are in inches with millimeters in parentheses.

50 Rosemount 6888C
Manual Power up
00809-0100-4891 January 2023

5 Power up
5.1 Powering up the standalone Rosemount
6888 Transmitter probe (without
Rosemount 6888 Xi)
Complete the following steps to power up the Rosemount 6888C Transmitter
probe.

Procedure
1. Apply AC line power to the transmitter.
2. Apply 24 Vdc loop power to the transmitter.
3. Using either the DCS control or a field communicator, verify
communications to the transmitter.
The analyzer probe takes approximately 45 minutes to warm up to the 736 °C
(1357 °F) heater setpoint. The 4-20 mA signal remains at a default value of 3.5 mA,
and the O2 reading remains at 0% through this warm-up period. After warm up,
the probe begins reading oxygen, and the 4-20 mA output is based on the default
range of 0-10% O2.
If there is an error condition at startup, an alarm message is displayed on the
Rosemount 6888Xi.

Manual 51
Power up Manual
January 2023 00809-0100-4891

5.2 Power up the Rosemount 6888 Analyzer


with single/dual channel or single channel
and flame safety interlock Rosemount
6888Xi
Procedure
1. Apply AC line power to the transmitter.
2. Apply AC line power to the Rosemount 6888Xi. Run the Quick Start Wizard
as described in Run Rosemount 6888Xi Quick Start Wizard. At the Auto Cal
Device screen, select the calibration method based on the Rosemount 6888
Analyzer as follows:
a. Standard probe housing configuration - Select None, SPS, or IMPS as
appropriate. Do not select Integral or calibration will not be possible.
b. Integral autocal probe housing - Select Integral only. If Integral is not
selected, calibration will not be possible.

3. Verify communications between the analyzer and the Rosemount 6888Xi.


The Rosemount 6888Xi display is preconfigured to display O2 and cell
temperature for single channel configurations and both O2 readings for
dual channel configurations.

The analyzer probe takes approximately 45 minutes to warm up to the 736 °C


(1357 °F) heater setpoint. The 4-20 mA signal remains at a default value of 3.5 mA,
and the O2 reading remains at 0% through this warm-up period. After warm up,
the probe begins reading oxygen, and the 4-20 mA output is based on the default
range of 0-10% O2.
If there is an error condition at startup, an alarm message is displayed on the
Rosemount 6888Xi.

5.3 Power up the Rosemount 6888C direct


replacement probe (no electronics inside)
with traditional architecture Rosemount
6888Xi
Procedure
1. Apply AC line power to the Rosemount 6888Xi.
2. Run the Quick Start Wizard as described in Run Rosemount 6888Xi Quick
Start Wizard.
3. At the Auto Cal Device screen, select None, SPS, or IMPS as appropriate.
Do not select Integral or calibration will not be possible.

52 Rosemount 6888C
Manual Power up
00809-0100-4891 January 2023

The analyzer probe takes approximately 45 minutes to warm up to the 1,357 °F


(736 °C) heater set point. The 4-20 mA signal remains at a default value of 3.5
mA, and the O2 reading remains at 0 percent through this warm-up period. After
warm-up, the probe begins reading oxygen, and the 4-20 mA output is based on
the default range of 0 to 10 percent O2.
If there is an error condition at start-up, the Rosemount 6888Xi displays an alarm
message.

5.4 Run Rosemount 6888Xi Quick Start Wizard


When you first power up the Rosemount 6888Xi, a short wizard program guides
you through the basic setup. Once configured, the Rosemount 6888Xi retains the
setup, and the wizard will not repeat.

Procedure
1. Apply power to the Rosemount 6888Xi.
Once the device powers on, the Quick Start Wizard screen appears. With
a dual channel Rosemount 6888Xi, the wizard runs for both channels in
succession.
2. Press Enter to continue.
3. On the Sensor Type screen, use the Up and Down keys to select O2.
Do not select CO, as this option is reserved for future use.
4. Press Enter to continue.
5. On the Device Type screen use the Up and Down keys to select HART® or FF
(FOUNDATION™ Fieldbus), whichever applies.
6. At the Auto Cal Device screen, use the Up and Down keys to select the
calibration method to be used. The methods are defined as follows:
• None: Manual calibration with the standard probe housing
configuration.
• SPS: Automatic calibration with the standard probe housing
configuration using the Rosemount SPS 4001B.
• IMPS: Automatic calibration with the standard probe housing
configuration using the Rosemount IMPS.
• Integral: Automatic calibration with the integral autocal probe housing
configuration.

7. Press Enter to continue.


Note
If you select SPS, IMPS, or Integral, you must still configure automatic
calibration as On. Verify other parameters, such as test gas values and
gas times, as well. Refer to the Rosemount 6888Xi Reference Manual for
calibration setup details.

8. When prompted by Setup Correct?, use the Up and Down keys to select
Yes.

Manual 53
Power up Manual
January 2023 00809-0100-4891

If you select No, the wizard restarts.


9. Press Enter to continue.
The Rosemount 6888Xi displays several screens while saving the configuration,
resets itself, and then returns to the main screen.

5.5 Re-initiating Rosemount 6888 Xi wizard


To re-initiate the wizard setup, reset the I/O board to default factory conditions.

Procedure
1. Apply power to the Rosemount 6888 Xi.
2. When the main screen appears, press Menu several times until the System
menu appears.
3. Use the Up and Down keys to select Configure IOB.
4. Press Enter to continue.
5. When the Configure IOB screen appears, use the Up and Down keys to
select I/O Board 1.
With a dual channel Rosemount 6888 Xi, you can reset and reconfigure
either or both I/O boards.
6. Press Enter to continue.
7. When the I/O board 1 menu appears, use the Up and Down keys to select
Reset I/O Board.
8. Press Enter to continue.
9. When the Reset menu appears, use the Up and Down keys to select Factory
Defaults.
10. Press Enter to continue.
11. When prompted, use the Up and Down keys to select Yes.
12. Press Enter to continue.
The Rosemount 6888 Xi displays several screens while saving the configuration,
resets itself, and then displays the Wizard screen.

54 Rosemount 6888C
Manual Calibration
00809-0100-4891 January 2023

6 Calibration
You can calibrate the Rosemount 6888C in the installed condition without
removing the instrument from the process duct and also while the combustion
process is online.
A stainless steel tube runs the length of the probe and delivers the calibration
gases into the cell area. Factory calibration is usually satisfactory for initial
startup and operation, but most accurate measurement is gained by executing
a calibration under normal operating conditions.
Emerson recommends using 0.4% O2 and 8% O2 as calibration gases, with the
balance of nitrogen in the gas bottles, but you can use other values as long as the
electronics are configured identically. Emerson does not recommend instrument
air or pure nitrogen as calibration gas values.
Use a two-stage pressure regulator to establish a pressure of 15 psi from the
bottles and set the flowmeter to 5 scfh flow rateor less.

WARNING

Explosion
Do not use explosive gases or gases with an O2 percentage greater than that
found in ambient air (i.e., 20.95 percent O2).

6.1 Manual/semi-automatic calibration


A technician can calibrate the Rosemount 6888C probe with standard housing
by following prompts via the display of the Rosemount 6888Xi Electronics or via
HART® communications to a Field Communicator or Asset Management Solutions
(AMS) console.
Manually switch the gases based upon these prompts. We recommend using
0.4 percent O2 and 8 percent O2, balance nitrogen as calibration gases. Always
use a two-stage pressure regulator set to 20 psig (1.38 barg). Set the calibration
gas flow meter for a maximum of 5 scfh with the cal gas fitting removed from
the probe. Do not use explosive gases or gases with greater percentage of O2
than that found in ambient air, such as 20.95 percent O2. A diffuser/filter that is
plugged over time may cause the flow meter to deliver less flow to the sensing
cell, but never readjust the flow rate until a new diffuser is installed. Readjusting
the flow meter back up to the 5 scfh level could pressurize the cell during
calibration and cause the O2 reading to shift downwards.
The electronics determine if the calibration was successful and calculate new
calibration values. New calibration values are not automatically loaded into the
electronics after a successful calibration, however. You have the choice to accept
or reject the new values.
A significant calibration change may cause a bump in the O2 readings at the
DCS console, causing operator concern. Record the calibration data on a log
(cell slope, constant, and impedance, as well as the speed of response data).

Manual 55
Calibration Manual
January 2023 00809-0100-4891

If the electronics is used, it stores calibration data for the past ten successful
calibrations.

CAUTION

Reading errors
Make sure the calibration gas cap is replaced tightly after calibration is complete.
Many combustion processes operate at a slight negative pressure (draft pressure)
and can draw ambient air down the cal gas lines and into the sensing cell, causing
a falsely elevated O2 reading. The same phenomenon is possible if the calibration
gas hoses are permitted to become degraded or loose.

6.2 Fully automatic calibration


For fully automatic calibration, the Rosemount 6888Xi Electronics must manage
the actuation of solenoids to introduce gases into the probe.

6.2.1 Probe with standard electronics housing


In addition to the Rosemount™ 6888 Xi, this arrangement requires a separate
single probe sequencer (SPS), which is a solenoid box for switching calibration
gases or a larger intelligent multiprobe sequencer (IMPS) which can handle the
autocal for up to four probes in one box.
The automatic calibrations can be initiated in several ways:
• Calibration recommended diagnostic that is periodically checking cell
impedance.
• Push button on the Rosemount 6888 Xi Electronics.
• HART® communications from a Field Communicator or AMS.
• External contact closure.
• Time since the last successful calibration.
Note
To manually initiate an automatic calibration, select 9999 for CAL INTERVAL in the
setup menu.

If using the O2 measurement for automatic control, always place the O2 control
loop into manual prior to calibrating. Always inform the operator prior to
calibrating. The Rosemount 6888 Xi Electronics provides an in cal contact closure
for this purpose as well as an initiate cal contact.
The Rosemount 6888 Xi Electronics sequences the calibration gases in turn into
the sensing cell. A 300 second flow time is the factory default for both gases and
also for the purge cycle, which lets the probe signal come back to the normal
flue gas readings. The 4-20 mA signal representing O2 can be held during the
calibration cycle or permitted to vary with the bottled gases, in which case a
record of the calibration can be trended at the DCS.
Calibration setup is found under the detailed setup menu.

56 Rosemount 6888C
Manual Calibration
00809-0100-4891 January 2023

6.2.2 Probe with integral autocal housing


This probe contains the autocal solenoids within the blue electronics housing,
eliminating the need and cost for an SPS or IMPS solenoid enclosure. Both
calibration gases are permanently piped into two ports on the probe. It's
important to confirm that there are no piping leaks or the calibration bottles will
leak down permanently.
Note
The calibration sequence from the Rosemount™ 6888 Xi Electronics is identical
to that for the manual/semiautomatic calibration, but note that with the integral
autocal version of this probe it is not possible to conduct a manual calibration.
The factory offers a probe rebuild capability if solenoid or other failures occur.

CAUTION

Leaks
Calibration gas bottles are piped and under pressure at all times, so be sure to
leak-check all fittings, tubing, and connections. Always use dual-stage pressure
regulators.

6.2.3 Other features associated with calibration


1. Cal Check capability - You can look at new calibration values prior to having
them load in and become active. If this feature is selected, a calibration
changed alarm comes up after the calibration is complete. Push Diagnostics
button on the Rosemount Xi to require an Accept Cal step, so that the
technician and operator can decide if they are ready to accept a potentially
large change in calibration values.
2. Plugged diffuser diagnostic - this feature measures the return-to-process
rate after calibration gases are removed and will alarm when this time
exceeds 75% of the purge time configured. A purge time too short alarm
is an indication that the diagnostic could not work because of a short purge
time and is another indication that the diffuser is plugged. Increase purge
time in this case.
Another feature that comes with the plugged diffuser diagnostic is auto
gas switching. This feature switches cal gases and also ends the purge
sequence when the readings settle out vs. waiting for a configured flow
time. This saves time and gas.
3. Tolerance check - alarms if the cell mV signal is significantly different from
the expected signal of the bottles noted in the setup. This indicates if the
wrong bottles are being used or if a bottle runs out in the middle of a
calibration.
Calibration setup is under the Detailed Setup menu on the Rosemount Xi menu.
See Figure 6-1.

Manual 57
58
Calibration
January 2023

Figure 6-1: Calibration Defaults

Defaults

Probe #1 Detailed setup Calibration Analog output Track cal gases

Tolerance check Off

Calibration recommended diagnostic


Off

Calibration acknowledge
Off
Cal gas #1 0.4% O2

Cal gas #2 8% O2

Cal gas flow time


300 seconds

Purge time
300 seconds
Diffuser Plugged diffuser Rate

Auto cal advance On

Auto calibration Enable autocal No – Std. housing


Yes – Autocal housing

Start on cal recommended No

Cal interval 9999 = manually


672 hours (1 week)
initiated autocal

Next cal time


Variable

Rosemount 6888C
00809-0100-4891
Manual
Manual HART® menu trees
00809-0100-4891 January 2023

7 HART® menu trees


Figure 7-1: Rosemount 6888 Xi (page 1)

Manual 59
HART® menu trees Manual
January 2023 00809-0100-4891

Figure 7-2: Rosemount 6888 Xi (continued)

60 Rosemount 6888C
Manual HART® menu trees
00809-0100-4891 January 2023

Figure 7-3: Rosemount 6888 Xi (continued)

Manual 61
HART® menu trees Manual
January 2023 00809-0100-4891

Figure 7-4: Field Communicator

62 Rosemount 6888C
Manual HART® menu trees
00809-0100-4891 January 2023

Manual 63
HART® menu trees Manual
January 2023 00809-0100-4891

64 Rosemount 6888C
Manual Maintenance and service
00809-0100-4891 January 2023

8 Maintenance and service


8.1 Overview
This section identifies the calibration methods available and provides the
procedures to maintain and service the Rosemount 6888C O2 transmitter.

WARNING
ELECTRIC SHOCK
Install all protective equipment covers and safety ground leads after equipment
repair or service. Failure to install covers and ground leads could result in serious
injury or death.

8.2 Maintenance intervals


The required maintenance interval depends on the ambient and process
conditions the analyzer is exposed to.
The zirconium oxide sensing cell is non-depleting and has no specific shelf life
or a defined life in flue gas operation. The sensitivity of a sensing cell that is
mounted inside a boiler that is burning natural gas may shift slightly over several
years. Acidic compounds are the main aggressors to the sensing cell, typically
SO2, resulting from sulfur contained in coal and heavy oil fuels, and also HCl from
the combustion of plastics in municipal incinerators and in industrial thermal
oxidizers. Sensing cells may experience significant degradation and signal shift in
this type of service, particularly if the operating levels of O2 are very low (below
1% O2).
A calibration check is recommended on a quarterly basis (every three months)
by flowing bottled gas to the probe. (Verify that the operations personnel are
notified when doing this and also make sure that the O2 control loop is placed
in manual mode). If the probe readings vary significantly from the bottle values,
then conduct a formal calibration.
The optional Rosemount 6888Xi offers a calibration recommended diagnostic
that indicates when the probe needs to be calibrated.
Combustion processes with a high level of ash or other particulate content will
cause the diffusion element on the end of the probe to plug. A significantly
plugged diffuser causes a slower speed of response to changing O2 levels in the
process. This can usually be seen on the recorded trends in the control room. In
addition, a significantly plugged diffuser can introduce error during a calibration
and negatively affect accuracy.
When performing a calibration check or actual calibration, the calibration flow
meter may read lower if the diffuser is significantly plugged. Never increase the
flow rate, however, as this can cause a shifted calibration. Adjust the calibration
flow rate only when a new diffuser is installed. Always record the response time

Manual 65
Maintenance and service Manual
January 2023 00809-0100-4891

back to the process after the calibration gases are removed. Diffuser plugging can
be tracked through the calibration record.
Visually inspect the probe during plant outages, paying particular attention
to condensed components. Condensation can be reduced or eliminated by
insulating the probe installation, including the probe mount, flange, and blue
housing.

8.3 Repair
Each of the following procedures details how to remove and replace a specific
component of the Rosemount 6888C.

WARNING
BURNS
Remove the transmitter from the stack for all service activities. Allow the unit to
cool and take it to a clean work area. Failure to comply may cause severe burns.

WARNING
ELECTRIC SHOCK
Disconnect and lock out power before working on any electrical components.
There is voltage up to 240 Vac.

8.3.1 Remove and replace probe


Remove probe
Complete the following steps to remove the probe from the stack.

Procedure
1. Turn off power to the system.
2. Shut off the calibration gases at the cylinder and the instrument air.
3. Disconnect the calibration gas and instrument air lines from the
transmitter.
4. Remove housing cover.
5. Remove all signal and power wiring to the probe.
6. Remove insulation to access the mounting probe.
7. Unbolt the transmitter from the stack and take it to a clean work area.
8. Allow the unit to cool to a comfortable working temperature.

66 Rosemount 6888C
Manual Maintenance and service
00809-0100-4891 January 2023

Replace probe
Complete the following steps to replace the probe in the stack.

Procedure
1. Bolt the transmitter to the stack and install the insulation.
2. Connect all signal and power leads at the probe.
Refer to Install for detailed wiring instructions.
3. Connect the calibration gas and instrument air lines to the probe.
4. Install the housing cover.
5. Turn on instrument air.
6. Restore power to the system.
Refer to Power up.

NOTICE
Recalibration is required whenever the electronic cards or sensing cell are
replaced.

8.3.2 Replace transmitter board


The transmitter board is not repairable and must be replaced if any component
fails.
See Figure 8-1 for item number references. The transmitter board is available as a
standalone kit or with the plastic cover and mounting plate.

WARNING
ELECTRIC SHOCK
Disconnect and lock out power before working on any electrical components.

CAUTION
MEASUREMENT ERRORS
Failure to resynchronize the calibration parameters with a Rosemount 6888Xi
after replacing the transmitter board or recalibrating the instrument may cause
an inaccurate O2 measurement. When the transmitter board is replaced and the
calibration parameters are not synchronized with the Rosemount 6888Xi, the
parameters in the Rosemount 6888Xi will be used as default to calculate the O2
measurement.

Manual 67
Maintenance and service Manual
January 2023 00809-0100-4891

Figure 8-1: Transmitter board mounting

A. Cover
B. Plastic cover
C. Transmitter board
D. Mounting plate
E. Screws
F. Thermal pad
G. Screws

Procedure
1. Remove cover (A) from the transmitter housing.
2. Disconnect the power and signals from the transmitter board (C).
3. Remove the three screws (E) securing the plastic cover (B) and transmitter
board (C) to mounting plate (D).
4. The plastic cover snaps onto the transmitter board with three tabs. Pull the
cover and board away from the mounting plate.
The thermal pad (F) will most likely remain attached to the mounting plate.
5. Disconnect the two wire harness connectors from the back of the
transmitter board and remove the plastic cover and transmitter board.
Replacement transmitter boards come in two configurations:
a. Transmitter board with thermal pad and mounting screws
b. Transmitter board, thermal pad, plastic cover, mounting plate, and
mounting screws

68 Rosemount 6888C
Manual Maintenance and service
00809-0100-4891 January 2023

6. Do one of the following:


• If only the transmitter board is being replaced, carefully unsnap the
plastic cover from the old transmitter board and then snap it onto the
new board. Make sure to place the cover so that the screw terminals for
the power and signal wiring fit into the openings.
• If the transmitter board, plastic cover, and mounting plate are being
replaced, the plastic cover should already be snapped onto the
transmitter board.

7. Remove the old thermal pad from the mounting plate.


8. Attach the new thermal pad to the mounting plate 1 in. up from the flat
edge of the pad.

CAUTION

Do not reassemble without the thermal pad in place.

If the transmitter board, plastic cover, and mounting plate are being
replaced, the thermal pad should already be in place on the mounting plate.
9. If the mounting pad is being replaced, remove the four screws (H) securing
the plate to the housing. Install the new mounting plate using four screws.

NOTICE
Replacement screws are provided if any become lost or damaged.

10. Reconnect the two wire harness connectors to the back of the transmitter
board.

NOTICE
The connectors are different and individually keyed to prevent
misconnection.

11. Carefully push the wire harnesses through the hole in the mounting plate.
12. Align the transmitter board with plastic cover over the three standoffs on
the mounting plate.
13. Install the three screws to secure the board to the mounting plate.
Replacement screws are provided if any become lost or damaged.
14. Reconnect the power and signal wires to the transmitter board.
15. Reinstall the cover to the transmitter housing.
16. Reapply power to the transmitter. Prior to using the transmitter for
measurement purposes, perform one of the following:
• Transmitter without Rosemount 6888Xi: Recalibrate the transmitter. See
Power up analyzer without Rosemount 6888Xi.
• Transmitter with Rosemount 6888Xi: When initialization is complete, the
transmitter begins to heat up, but an alarm condition is shown on

Manual 69
Maintenance and service Manual
January 2023 00809-0100-4891

the Rosemount 6888Xi. Synchronize the calibration parameters via the


Rosemount 6888Xi as follows:
a. Press DIAG once to view the alarms. A Probe Changed alarm
appears.
b. Press EXIT once to enter the Diagnostic menu.
c. Use the Up and Down keys to select 3-Acknowledge Alarms. Press
ENTER.
d. Use the Up and Down keys to select 3-Ack Probe Changed. Press
ENTER.
e. Use the Up and Down keys to select 1-Send To Probe. Press ENTER.
f. When transfer is complete, press EXIT twice to return to the main
menu.

8.3.3 Replace DR terminal board


The DR terminal board is not repairable and must be replaced if any component
fails. See Figure 8-2 for letter references. The DR board is available as a
standalone kit or with the plastic cover and mounting plate.

WARNING
ELECTRIC SHOCK
Disconnect and lock out power before working on any electrical components.

70 Rosemount 6888C
Manual Maintenance and service
00809-0100-4891 January 2023

Figure 8-2: DR terminal board mounting

A. Cover
B. Plastic cover
C. DR terminal board
D. Mounting plate
E. Screws
F. Screws

Procedure
1. Remove cover (A) from the DR probe housing.
2. Disconnect the power and signal wires from the DR terminal board (C).
3. Remove the three screws (E) securing the plastic cover (B) and DR terminal
board (C) to mounting plate (D).
4. Disconnect the two wire harness connectors from the back of the
transmitter board.
5. Remove the plastic cover and transmitter board.
Replacement DR terminal boards come in two configurations:
a. DR terminal board with thermal pad and mounting screws
b. DR terminal board, thermal pad, plastic cover, mounting plate, and
mounting screws

6. If the mounting pad is being replaced, remove the four screws (F) securing
the plate to the housing.
7. Install the new mounting plate using four screws.
Replacement screws are provided if any become lost or damaged.

Manual 71
Maintenance and service Manual
January 2023 00809-0100-4891

8. Reconnect the two wire harness connectors to the back of the DR terminal
board.
The connectors are different and individually keyed to prevent
misconnection.
9. Carefully push the wire harnesses through the hole in the mounting plate
and align the DR terminal board with plastic cover over the three standoffs
on the mounting plate.
10. Install the three screws to secure the board to the mounting plate.
Replacement screws are provided if any become lost or damaged.
11. Reconnect the power and signal wires to the transmitter board.
12. Reinstall the cover to the DR probe housing.
13. Reapply power to the DR probe.

8.3.4 Heater strut replacement


If heater strut replacement is needed, order a replacement heater strut kit (Table
9-1). Refer to Figure 8-3 for item letter references.

WARNING
BURNS
Use heat resistant gloves and clothing when removing probe. Do not attempt to
work on the probe until it has cooled to room temperature. The probe can be as
hot as 427 °C (800 °F). This can cause severe burns.

72 Rosemount 6888C
Manual Maintenance and service
00809-0100-4891 January 2023

Figure 8-3: Housing/Strut Exploded View

A. Cover
B. Plastic cover
C. Transmitter board
D. Mounting plate
E. Housing screws
F. Thermal pad
G. Screws
H. Strut clamp
I. O-ring
J. Screws
K. Viton tube
L. Tube clamp

Procedure
1. Follow the instructions in Remove and replace probe to remove the
transmitter from the stack or duct.
2. Follow the instructions in Replace transmitter board or Replace DR terminal
board to remove the electronics from the housing.
3. Remove four screws (E). Remove housing from probe using care not to
damage the strut wire harnesses or O-ring (I).
4. Remove the strut clamp (H) from inside the finned housing on the probe.
5. Remove tube clamps (L) and Viton tubes (K) from the CAL and REF gas ports
and the CAL and REF gas lines. Leave the clamps in place on the hoses for
later reassembly.
6. Grasp the wire loop and carefully slide the heater strut assembly out of the
probe tube.

Manual 73
Maintenance and service Manual
January 2023 00809-0100-4891

7. When replacing the strut, orient the probe so that the small calibration gas
tube is at the 6 o'clock position of the probe tube.
8. Align the slot on the heater plate with the calibration gas line in the probe
tube.
9. Slide the strut into the probe tube.
The strut turns to align the hole on the back plate of the strut with the
calibration gas line. When the hole and calibration gas line are aligned
correctly, the strut slides in the rest of the way.
10. As the strut insertion nears completion, install the guide rod into the
calibration gas tube to assist in guiding the calibration gas tube through
the hole in the end of the strut.
11. Replace the CAL and REF gas Viton tubes (K).
12. Insert the strut clamp (H) back into the finned housing on the probe.
The opening should be down to allow for clearance for the gas tubes.
13. Carefully install the O-ring (I) back into the groove between the two
housings.
A small amount of silicone grease may be used to hold the O-ring in place.
Do not use petroleum based grease.
14. Insert the strut wire harnesses into the housing.
15. Turn the conduit ports of the housing to the CAL and REF gas ports side of
the probe and position the housing on the probe.
16. Install and tighten the four screws (J).
17. Follow the instructions in Replace transmitter board to install the electronics
into the housing.
18. Follow the instructions in Remove and replace probe to install the
transmitter into the stack or duct.

8.3.5 Replace oxygen sensing cell


If cell replacement is needed, order the cell replacement kit (Table 9-1).
The cell replacement kit (Figure 8-4) contains a cell and flange assembly,
corrugated seal, setscrews, socket head cap screws, and anti-seize compound.
The items are carefully packaged to preserve precise surface finishes. Do not
remove items from the packaging until they are ready to be used. Spanner
wrenches and hex wrenches needed for this procedure are part of an available
special tools kit (Table 9-1).

WARNING
BURNS
Use heat-resistant gloves and clothing when removing the probe. Do not attempt
to work on these components until they have cooled to room temperature. Probe
components can be as hot as 300 °C (572 °F). This can cause severe burns.
Disconnect and lock out power before working on any electrical components.
There is voltage up to 240 Vac.

74 Rosemount 6888C
Manual Maintenance and service
00809-0100-4891 January 2023

CAUTION
EQUIPMENT DAMAGE
Do not remove the cell unless you are certain it needs to be replaced. Removal
may damage the cell and platinum pad. Follow the complete troubleshooting
procedure to make sure the cell needs to be replaced before removing it.

Figure 8-4: Cell Replacement Kit

A. ANSI gasket
B. Wire and pad assembly
C. Anti-seize compound
D. 22 gauge wire
E. Closed end connector
F. Set screws
G. Teflon tubing
H. Socket head cap screws
I. Cell and flange assembly
J. Corrugated seal
K. Probe (not included in kit)
L. Calibration gas passage

Procedure
1. Follow the instructions in Remove and replace probe to remove the
transmitter from the stack or duct.
2. If the probe uses the standard diffusion element, use a wrench to remove
the diffuser assembly.
3. If equipped with the optional ceramic diffusion assembly, remove and
discard the setscrews and remove the vee deflector.

Manual 75
Maintenance and service Manual
January 2023 00809-0100-4891

4. Use spanner wrenches from the probe disassembly kit (Table 9-1) to turn
the hub free from the retainer.
5. Inspect the diffusion element. If it is damaged, replace the element.
6. Loosen the four socket head cap screws from the cell and flange assembly.
7. Remove the assembly and the corrugated seal.

NOTICE
The cell flange has a notch that may be used to gently pry the flange away
from the probe. The contact pad inside of the probe sometimes fuses to the
oxygen sensing cell. If the cell is fused to the contact pad, push the cell
assembly back into the probe (against spring pressure) and quickly twist the
cell assembly.

The cell and contact pad should separate. If the contact pad stays fused
to the cell, a new contact/thermocouple assembly must be installed.
Disconnect the cell and the thermocouple wires at the crimp connections
and withdraw the cell with the wires still attached.
8. If the contact assembly is damaged, replace the strut or contact pad.
9. Follow the instructions in Heater strut replacement, steps 2 through 4, to
remove the electronics housing,
10. Follow the instructions for replacing the contact pad provided in the cell
replacement kit.
11. Remove and discard the corrugated seal.
12. Clean the mating faces of the probe tube and retainer.
13. Remove burrs and raised surfaces with a block of wood and crocus cloth.
14. Clean the threads on the retainer and hub.
15. Rub a small amount of anti-seize compound on both sides of the new
corrugated seal.
16. Assemble the cell and flange assembly, corrugated seal, and probe tube.
17. Make sure the calibration tube lines up with the calibration gas passage in
each component.
18. Apply a small amount of anti-seize compound to the screw threads and use
the screws to secure assembly.
19. Torque to 60 in.-lb (6.8 N m).
20. Follow the instructions in Heater strut replacement, steps 9 through 15, to
reassemble the electronics housing.
21. Apply anti-seize compound to the threads of the cell assembly, hub, and
setscrews.
22. Reinstall the hub on the cell assembly.
23. Using pin spanner wrenches, torque to 10 ft-lb (14 N m).
24. If applicable, reinstall the vee deflector, orienting apex toward gas flow.
25. Secure with setscrews and anti-seize compound.
26. Torque to 25 in-lb (2.8 N m).

76 Rosemount 6888C
Manual Maintenance and service
00809-0100-4891 January 2023

27. On systems equipped with an abrasive shield, install the dust gaskets with
joints 180 ° apart.
28. Follow the instructions in Remove and replace probe to install the
transmitter to the stack or duct.
29. Turn on power and monitor thermocouple output.
It should stabilize at 29.3 ±0.2 mV.
30. Set reference air flow at 1 l/min (2 scfh).
31. After the transmitter stabilizes, calibrate the probe.
32. If new components have been installed, repeat calibration after 24 hours of
operation.

8.3.6 Diffusion element replacement


The diffusion element protects the cell from particles in process gases. Normally,
it does not need to be replaced, because the vee deflector protects it from
particulate erosion.
In severe environments, the diffusion element may be broken or subject to
excessive erosion. Examine the element whenever removing the probe for any
purpose. Replace if damaged.
Damage to the diffusion element may become apparent during calibration.
Compare probe response with previous response. A broken diffusion element
will cause an inadequate response to calibration gas. Hex wrenches needed
to remove setscrews and socket head screws in the following procedure are
available as part of a probe disassembly kit, Table 9-1.

NOTICE
This section refers to the ceramic diffusion element only.

Manual 77
Maintenance and service Manual
January 2023 00809-0100-4891

Figure 8-5: Ceramic diffuser element replacement

A. Retainer
B. Spanner wrench
C. Optional ceramic diffusion element
D. Set screw
E. Vee deflector
F. Cement fillet
G. Cement port
H. Hub

Procedure
1. Follow the instructions in Remove and replace probe to remove the probe
from the stack or duct.
2. Loosen set screws (D) using the hex wrench from the probe disassembly kit
and remove the vee deflector (E). Inspect sets crews. If damaged, replace
with stainless sets crews coated with anti-seize compound.
3. Follow the instructions in Remove and replace probe to install the probe
into the stack or duct.

78 Rosemount 6888C
Manual Maintenance and service
00809-0100-4891 January 2023

8.3.7 Replace blind cover


This section covers the replacement of the main housing cover and the
application of the correct field wiring label.

NOTICE
Field wiring labels are supplied for each product configuration. Ensure the correct
label is applied to the inside of the cover that matches your configuration.

Figure 8-6: Field wiring labels

A. Label for full transmitter


B. Label for DR probe
C. Label for DR/YEW probe

Procedure

Manual 79
Maintenance and service Manual
January 2023 00809-0100-4891

1. Remove existing cover from the transmitter housing.


2. Select the new field wiring label that matches the existing label on the
inside of the old cover.
A graphic of each label and its application is shown in Figure 8-6.
3. Peel off the backing.
4. Attach the new label to the inside of the new cover.
5. Install the new cover onto the housing.

80 Rosemount 6888C
Manual Replacement parts
00809-0100-4891 January 2023

9 Replacement parts
Table 9-1: Replacement Part Kits for O2 Transmitter
Part Number Description
6A00448G01 Kit, transmitter board
6A00448G02 Kit, transmitter board with mounting plate and cover
6A00449G01 Kit, DR terminal board
6A00449G02 Kit, DR terminal board with mounting plate and cover
6A00449G10 Kit, DR terminal board, YEW
6A00449G11 Kit, DR terminal board, YEW with mounting plate and cover
6A00452G01 Kit, blind cover, standard, and autocal housing
6A00450G01 Kit, Cell gaskets (pack of five)
6A00451G01 Kit, cover O-rings, standard housing (pack of five)
6A00453G01 Kit, snubber diffuser
6A00453G02 Kit, snubber diffuser with dust seal
6A00453G03 Kit, ceramic diffuser with V-deflector
6A00453G04 Kit, ceramic diffuser with dust seal with V-deflector
6A00453G05 Kit, Hastelloy diffuser with V-deflector
6A00453G06 Kit, Hastelloy diffuser with dust seal with V-deflector
6A00454G01 Kit, strut, 18 inch
6A00454G02 Kit, strut, 3 foot
6A00454G03 Kit, strut, 6 foot
6A00455G01 Kit, cell replacement, 18 inch, standard cell
6A00455G02 Kit, cell replacement, 3 foot, standard cell
6A00455G03 Kit, cell replacement, 6 foot, standard cell
6A00455G11 Kit, cell replacement, 18 inch, acid resistant cell
6A00455G12 Kit, cell replacement, 3 foot, acid resistant cell
6A00455G13 Kit, cell replacement, 6 foot, acid resistant cell
6A00456G01 Flange gasket kit, ANSI, general purpose, (pack of five)
6A00456G03 Flange gasket kit, DIN, general purpose (pack of five)
6A00457G01 Spare transmitter board cover
6A00457G02 Spare DR term board cover
6A00475G01 Tool kit - 6888
1A99119G01 Calibration gas bottles - 0.4% and 8% O2, balance nitrogen - 550 liters
each(1)

Manual 81
Replacement parts Manual
January 2023 00809-0100-4891

Table 9-1: Replacement Part Kits for O2 Transmitter (continued)


1A99119G02 Two flow regulators (for calibration gas bottles)
1A99119G03 Bottle rack

(1) Calibration gas bottles cannot be shipped via air freight.

82 Rosemount 6888C
Manual Service support
00809-0100-4891 January 2023

10 Service support
To expedite the return process outside of the United States, contact the nearest
Rosemount™ representative.
Within the United States, call the Emerson™ Instrument and Valves Response
Center using the 1-800-654-RSMT (7768) toll-free number. This center, available
24 hours a day, will assist you with any needed information or materials.
The center will ask for product model and serial numbers and will provide a
Return of Material Authorization (RMA) number. The center will also ask for the
process material to which the product was last exposed.

Manual 83
Service support Manual
January 2023 00809-0100-4891

84 Rosemount 6888C
Manual Product certifications
00809-0100-4891 January 2023

A Product certifications
For product certifications and the Declaration of Conformity, see the Rosemount
6888C Quick Start Guide.
The most recent revision of the Declaration of Conformity can be found at
Emerson.com/Rosemount.

Manual 85
00809-0100-4891
Rev. AB
2023

For more information: Emerson.com


©2023 Emerson. All rights reserved.

Emerson Terms and Conditions of Sale are available


upon request. The Emerson logo is a trademark and
service mark of Emerson Electric Co. Rosemount is a
mark of one of the Emerson family of companies. All
other marks are the property of their respective owners.