Artis Zee Stand Adjustment
Artis Zee Stand Adjustment
Artis Zee Stand Adjustment
Adjustment
Stand
Floor and Biplane Stands
2 Document Version / Disclaimer / Copyright
Document Version
Siemens Healthcare GmbH reserves the right to change its products and services at any
time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens Healthineers office to order a current version or refer to
our website http://www.siemens-healthineers.com.
Disclaimer
Siemens Healthcare GmbH provides this documentation “as is” without the assumption of
any liability under any theory of law.
The content described herein requires superior understanding of our equipment and may
only be performed by qualified personnel who are specially trained for such installation
and/or service.
Copyright
“© Siemens Healthineers, 2007” refers to the copyright of a Siemens Healthineers entity
such as:
Siemens Healthcare GmbH - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany
1 General Information 6
1.1 Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.1.1 Required Aids and Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.2 Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.1 General Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.2 Product-specific Safety Information . . . . . . . . . . . . . . . . . . . . . . . 8
1.2.3 Product-specific Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.2.4 Conventions and Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.2.5 Tolerance Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.2.6 Screenshots & Graphics in this Document . . . . . . . . . . . . . . . . . . 10
2 Definition of Movements 11
3.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.2 C-arm rotational potentiometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.3 C-arm orbital potentiometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.4 Receptor lift potentiometer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.5 Swivel arm potentiometer (base rotation) . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.6 Multispace F potentiometer (manual rotation). . . . . . . . . . . . . . . . . . . . . . . 26
3.7 Card collimator, rotatable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.8 AMC collimator, rotatable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.9 Detector Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
5.1.1 Resetting or Deleting Adjustments . . . . . . . . . . . . . . . . . . . . . . . . 54
5.2 Linearity Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
5.2.1 CarmRot. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
5.2.2 CarmOrb . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
5.2.3 ReceptorA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
5.2.4 BaseRotA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.2.5 Collimator RotA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.2.6 CameraRotA / Detector Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . 61
5.2.7 Multispace F (Manual Rotation) . . . . . . . . . . . . . . . . . . . . . . . . . . 63
5.2.8 Final Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
7.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
7.2 End Position Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
7.2.1 C-ArmRotA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
7.2.2 CarmOrbA:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
7.2.3 ReceptorA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
9 Geometry Adjustment 88
9.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
9.2 AXIOM Artis - Single-plane System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
9.2.1 Saving the adjustment values . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
9.2.2 Final Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
9.3 AXIOM Artis - Biplane System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
9.3.1 Saving the adjustment values . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
9.3.2 Final Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
10 Mechanical Settings 93
11 Procedures 105
14 Index 112
1.1 Requirements
Service may be performed only by CSEs who have participated in a product course.
All specified tools, test equipment, and aids, with the exception of the "standard installa‐
tion tool kit” and the general purpose test meter, are listed and specified in the Service
Parts Catalog.
*1 regarding the ground wire test meter
n Ground wire test meter (the test meter must meet the specifications of IEC 61557 / EN
61557 / VDE 0413 / IEC 62353),
e.g.: SECUTEST♦SIII general purpose test meter or PROFITEST♦ 0100S-II *.
1.2 Notes
After repair work is completed and before the system is turned over to the customer, the
ground wire resistance of the equipment ( Ground wire test in accordance with IEC
62353 / Page 107) must be tested.
CAUTION
When performing all activities on the system, the safety information con‐
tained in the Operator Manuals must be observed.
hm_serv_Operator_manuals
WARNING
Incorrect service can lead to damage for personnel and system.
Lack of training can result in failure.
Service is only permitted by trained personnel.
hm_serv_Training
WARNING
Unwanted disconnection of electrical cables.
Interruption of interventional procedure.
Check fixation of connectors and cables after service procedures.
hm_serv_FixCablesAfterService
WARNING
Improperly established protective ground conductor connections during installa‐
tion, maintenance, or service.
Personal injury due to electric shock.
Protective conductor tests shall be performed in accordance with the instruc‐
tions in the technical documentation prior to initially switching on the device/
system, as well as after finishing all types of service / installation activities
and prior to handing over the system to the customer.
hm_serv_ProtectiveEarthMeasurement
After finishing the repair but before turning the system over to the customer, complete
the measurement of the protective earth resistance (ground connection) for every com‐
ponent where the grounding wire was removed. This measurement must be performed
in accordance with the IEC 62353 requirements. Document the measurement in the
service report for future reference.
WARNING
Even after switching off the system (shutdown), power is still present in the con‐
trol cabinet and generator.
Disregarding safety precautions can result in death or serious bodily injury.
Observe the warning labels in the electronic cabinets.
Switch off the power supply to the X-ray system with the on/off switch instal‐
led on-site.
Left side
Right side
1.2.4.1 Abbreviations
Abbreviation Description
ISK Allen screw
SW Wrench size
MCU MCM4 motor control unit
SCM Stand Control Module
TCM Table Control Module
DMG Deadman's Grip
CAN Controller Area Network
CW Clockwise
CCW Counterclockwise
Limit values for over 3 mm over 6 mm over 30 over 120 over 400 over 1000 over 2000
the nominal to 6 mm to 30 mm mm to 120 mm to 400 mm to mm to mm to
range mm mm 1000 mm 2000 mm 4000 mm
Admissible tol‐ ± 0,5 mm ± 1 mm ± 1,5 mm ± 2,5 mm ± 4 mm ± 6 mm ± 8 mm
erance
These tolerances apply to all dimensions given in these instructions, unless another toler‐
ance is expressly indicated after the value.
The following tables provide an overview of the designations for the potentiometer, en‐
coder, reference tag, and switches for the various axes:
3.1 General
Fig. 5: Stand - Component Adjustment
Adjustment of the potentiometers in the different axes is always made according to the
following principle:
n Start the service interface. Select Tune up > Stands > Potentiometer Adjustment,
Floor Stand using the Select button and press Start.
n The axis to be adjusted is moved to a defined calibration position (use a spirit level or
tape measure).
n Decouple the potentiometer from the axis or loosen the set screw on the potentiome‐
ter shaft.
n The actual value for the active potentiometer is displayed on the DDIS in the “Stand
message line” (examination room). All potentiometer values are displayed in the serv‐
ice interface (control room).
Remedy:
Press the <Set> button: the first axis is selected in the "Potentiometer Adjustment, Floor
Stand” window (marked with "*"). Keep pressing the <Set> button until the correct axis
is selected.
Look through the event log for AX_SCU_ERR_1100, _1001,_1105. This error is generated
if the potentiometer was not correctly adjusted.
When moving the axis, make sure that there is not a drop below or a rise above the A/D
values, because this will destroy the potentiometer.
Max. range of the potentiometer: 0...4096
Do not drop below or exceed the range: 50...4046.
This value is only a reference value and values significantly below this can appear for
some unit movements. If the AX_SCU_ERR_1100 error occurs (unit movement no longer
possible, the A/D value of the potentiometer is outside of the permissible range and can
also be destroyed in the mechanical end areas. Here, a readjustment of the particular po‐
tentiometer is required.
The axes available depend on the construction. This includes table tilt (tilt, cradle) and
collimator/detector rotation on the floor stand.
Once the potentiometer adjustment is completed, perform the linearity adjustment and
the end position adjustment.
Axis MCU 1 Floor Stand MCU 2 Ceiling Stand MCU 3 Patient Table
Axis 1 C-Arm rotation C-Arm rotation Table Lift
Axis 2 C-Arm orbital C-Arm orbital Table Tilt
Axis 3 Receptor Lift Receptor Lift Table Longitudinal
Axis 4 Gantry Rotation (pivot) Gantry Longitudinal Table Cradle
Axis 5 Detector rotation n.a. n.a.
Axis 6 Collimator Rotation n.a. n.a.
n Remove the cover from the floor stand column ( Covers - Floor Stand / Page 105).
n To start the service interface, select Tune up > Stands > Potentiometer Adjustment,
Floor stand using the Select button and press Start .
n Move the axis to the 0 position and use a level to align it.
n Decouple the potentiometer from the axis and adjust it until the reference value in the
table is reached.
n The actual value for the active potentiometer is displayed on the DDIS in the “Stand
message line” (examination room). All potentiometer values are displayed in the serv‐
ice interface (control room monitor).
n Reattach the potentiometer. Compare the reference value to the actual value; be sure
to adhere to specified tolerances.
n Perform the linearity adjustment.
n Perform adjustment of the end position.
n Move the axis over the entire range of movement and check the event log for
AX_SCU_ERR_1101. This error is generated when the potentiometer has not been ad‐
justed correctly.
When moving the axes, make sure that the A/D values are neither fallen below or excee‐
ded, because this will destroy the potentiometer.
Max. range of the potentiometer: 0...4096
Do not drop below or exceed the range: 50...4046.
This value is only a reference value and values significantly below this can appear for
some unit movements. If the AX_SCU_ERR_1100 error occurs (unit movement no longer
possible, the A/D value of the potentiometer is outside of the permissible range and can
also be destroyed in the mechanical end areas. In this case, the potentiometer must be
readjusted.
1 2
n Remove the left cover on the floor stand C-Arm (LAO side).
n Move the C-arm into the 0 position (align the bore on the C-arm ( 1/Fig. 7 Page 19)
with the reference tag ( 2/Fig. 7 Page 19)
The 0 position for the C-arm must be set very precisely in order to avoid actuating the
limit switches.
n To start the service interface, select Tune up > Stands > Potentiometer Adjustment,
Floor stand using the Select button and press Start .
n Remove the M5 Allen screw( 2/Fig. 8 Page 19); loosen the set screw on the potenti‐
ometer shaft.
n Adjust the potentiometer to the reference value in the table.
n The actual value for the active potentiometer is displayed on the DDIS in the “Stand
message line” (examination room). All potentiometer values are displayed in the serv‐
ice interface (control room).
n Tighten the set screw on the potentiometer and reattach and tighten the M5 Allen
screw. Compare the reference value to the actual value; be sure to adhere to specified
tolerances.
n Perform the linearity adjustment.
n Perform adjustment of the end position.
n Move the axis over the entire range of movement and check the event log for
AX_SCU_ERR_1101. This error is generated when the potentiometer has not been ad‐
justed correctly.
When moving the axes, make sure that the A/D values are neither fallen below or excee‐
ded, because this will destroy the potentiometer.
Max. range of the potentiometer: 0...4096
Do not drop below or exceed the range: 50...4046.
This value is only a reference value and values significantly below this can appear for
some unit movements. If the AX_SCU_ERR_1100 error occurs (unit movement no longer
possible, the A/D value of the potentiometer is outside of the permissible range and can
also be destroyed in the mechanical end areas. In this case, the potentiometer must be
readjusted.
n Remove the M4 Allen screw( 1/Fig. 9 Page 21); loosen the set screw on the potenti‐
ometer shaft.
When moving the axes, make sure that the A/D values are neither fallen below or excee‐
ded, because this will destroy the potentiometer.
Max. range of the potentiometer: 0...4096
Do not drop below or exceed the range: 50...4046.
This value is only a reference value and values significantly below this can appear for
some unit movements. If the AX_SCU_ERR_1100 error occurs (unit movement no longer
possible, the A/D value of the potentiometer is outside of the permissible range and can
also be destroyed in the mechanical end areas. In this case, the potentiometer must be
readjusted.
40 cm receptor axle
Name Position (mm) Voltage AD Value
Potentiometer Potentiome‐ 1175 3.67 ± 0.05 V 1505 ± 20 Bit
ter 3
Reference Tag Xref3 1025+/-10
n To start the service interface, select Tune up > Stands > Potentiometer Adjustment,
Floor stand using the Select button and press Start .
n Detach the potentiometer (spring balancer) and adjust it until the reference value in
the table is reached.
n The actual value for the active potentiometer is displayed on the DDIS in the “Stand
message line” (examination room). All potentiometer values are displayed in the serv‐
ice interface (control room).
n Reattach the potentiometer. Compare the reference value to the actual value; be sure
to adhere to specified tolerances.
n Perform the linearity adjustment.
n Move the axis over the entire range of movement and check the event log for
AX_SCU_ERR_1101. This error is generated when the potentiometer has not been ad‐
justed correctly.
When moving the axes, make sure that the A/D values are neither fallen below or excee‐
ded, because this will destroy the potentiometer.
Max. range of the potentiometer: 0...4096
Do not drop below or exceed the range: 50...4046.
This value is only a reference value and values significantly below this can appear for
some unit movements. If the AX_SCU_ERR_1100 error occurs (unit movement no longer
possible, the A/D value of the potentiometer is outside of the permissible range and can
also be destroyed in the mechanical end areas. In this case, the potentiometer must be
readjusted.
3
1
1
2 2
n Remove the bottom cover on the swivel arm of the floor stand.
n Move the pivot arm into the working position to the mechanical end stop. Make sure
that the pivot arm is all the way at stop.
Fig. 14: Manual Rotation Potentiometer Adjustment
n Move the patient table to the foot end and into the center. The table may not be ro‐
tated.
n To start the service interface, select Tune up > Stands > Potentiometer Adjustment,
Floor stand using the Select button and press Start .
n The actual value for the active potentiometer is displayed on the DDIS in the “Stand
message line” (examination room). All potentiometer values are displayed in the serv‐
ice interface (control room).
n Loosen the set screw on the potentiometer shaft ( 1/Fig. 13 Page 26).
n Adjust the potentiometer using the adjustment disc ( 3/Fig. 13 Page 26) attached to
the shaft to the reference value in the table.
n Reattach the potentiometer. Compare the reference value to the actual value; be sure
to adhere to specified tolerances.
n Perform the linearity adjustment.
n Move the axis over the entire range of movement and check the event log for
AX_SCU_ERR_1101. This error is generated when the potentiometer has not been ad‐
justed correctly.
When moving the axes, make sure that the A/D values are neither fallen below or excee‐
ded, because this will destroy the potentiometer.
Max. range of the potentiometer: 0...4096
Do not drop below or exceed the range: 50...4046.
This value is only a reference value and values significantly below this can appear for
some unit movements. If the AX_SCU_ERR_1100 error occurs (unit movement no longer
possible, the A/D value of the potentiometer is outside of the permissible range and can
also be destroyed in the mechanical end areas. In this case, the potentiometer must be
readjusted.
3 5
1
4
Card Collimator
Name Position Voltage AD Value
Potentiometer R181 0 5.0 ± 0.05 V 2048, ± 20 Bit
N Collimator
Name Position Voltage AD Value
Potentiometer R181 0 5.0 ± 0.05 V 2048 ± 20 Bit
2. Select <Test tools> <Stand& Table Tests> <Position Check Floor Stand> in the Service
menu and click <Start>.
Fig. 20: Position check
3. Use the potentiometer to set an AD value of 2048 under the point “Camera Rotation”
( 1/Fig. 20 Page 33).
To prevent unintentional rotation of the detector during the adjustment, the two power
cables on the motor ( 5/Fig. 21 Page 34) should be disconnected for this adjustment!
Fig. 21:
1. Remove the cover from the ceiling carriage ( Covers - Ceiling Stand (2nd
Plane) / Page 106).
2. To start the service interface, select Tune up > Stand > Potentiometer Adjustment,
Second Plane using the Select button and press Start.
3. Move the axis to 0˚.
4. Decouple the potentiometer from the axis and adjust it until the reference value in the
table is reached.
5. The actual value for the active potentiometer is displayed on the DDIS in the “Stand
message line” (examination room). All potentiometer values are displayed in the serv‐
ice interface (control room).
6. Mesh the potentiometer again. Compare the reference value to the actual value;
when doing this, observe the tolerance data.
7. Perform the linearity adjustment.
8. Move the axis over the entire movement range and look through the event log for
AX_SCU_ERR_1101. This error is generated when the potentiometer has not been ad‐
justed correctly.
When moving the axis, make sure that there is not a drop below or a rise above the A/D
values, because this will destroy the potentiometer.
Max. range of the potentiometer: 0...4096
Do not drop below or exceed the range: 50...4046
1. Due to the mechanical tolerances (installation of the ceiling rails, mech. tolerances of the overhead carriage and c-arm)
these values can differ. The particular listed value is used only as a reference
2 1
The 90° position for the C-Arm must be set very exactly to avoid moving into the limit
switches.
3. To start the service interface, select Tune up > Stands > Potentiometer Adjustment,
Second Plane using the Select button and press Start.
4. Remove the M5 Allen screw; loosen the set screw and then adjust the potentiometer
shaft until the reference value listed in the table is reached.
5. The actual value of the active potentiometer is displayed on the DDis in the “Stand
message line” (examination room). All potentiometer values are displayed on the serv‐
ice interface (control room).
6. Tighten the set screw on the potentiometer, and retighten the M5 Allen screw. Com‐
pare the reference value to the actual value; when doing this, observe the tolerance.
7. Perform the linearity adjustment.
8. Perform an end position adjustment.
9. Move the axis over the entire movement range and look though the event log for
AX_SCU_ERR_1101. This error is generated when the potentiometer was not correctly
adjusted.
When moving the axes, make sure that there is not a drop below or a rise above the A/D
values, because this will destroy the potentiometer.
Max. range of the potentiometer: 0...4096
Range which may not be dropped below or exceeded: 50...4046
1. Because of mechanical tolerances (installation of the ceiling rails, mech. tolerances of the overhead carriage and C-Arm),
these values can be different. The particular values should be seen only as references
4.3 Receptor B
Fig. 24: Receptor B
4. Remove the M4 Allen screw; loosen the grub screw on the potentiometer axis and
then adjust the potentiometer axis until the reference value in the table is reached.
5. The actual value for the active potentiometer is displayed on the DDIS in the “Stand
message line” (examination room). All potentiometer values are displayed in the serv‐
ice interface (control room).
6. Tighten the grub screw at the potentiometer and the M4 allen screw. Compare the
reference value to the actual value; when doing this, observe the tolerance data.
7. Perform the linearity adjustment.
8. Perform adjustment of the end position.
9. Move the axis over the entire movement range and look through the event log for
AX_SCU_ERR_1101. This error is generated when the potentiometer has not been ad‐
justed correctly.
When moving the axis, make sure that there is not a drop below or a rise above the A/D
values, because this will destroy the potentiometer.
Max. range of the potentiometer: 0...4096
Do not drop below or exceed the range: 50...4046
1. Due to the mechanical tolerances (installation of the ceiling rails, mech. tolerances of the overhead carriage and c-arm)
these values can differ. The particular listed value is used only as a reference
pot8a
pot8b
3. To start the service interface, select Tune up > Stands > Potentiometer Adjustment,
Second Plane using the Select button and press Start.
4. Disengage the potentiometer and then adjust the potentiometer until the reference
value listed in the table is reached.
5. The actual value of the active potentiometer is displayed on the DDis in the “Stand
message line” (examination room). All potentiometer values are displayed on the serv‐
ice interface (control room).
6. Mesh the potentiometer. Compare the reference value to the actual value; when do‐
ing this, observe the tolerance.
7. Perform the linearity adjustment.
8. Move the axis over the entire movement range and look though the event log for
AX_SCU_ERR_1101. This error is generated when the potentiometer was not correctly
adjusted.
When moving the axes, make sure that there is not a drop below or a rise above the A/D
values, because this will destroy the potentiometer.
Max. range of the potentiometer: 0...4096
Range which may not be dropped below or exceeded: 50...4046
1. Because of mechanical tolerances (installation of the ceiling rails, mech. tolerances of the overhead carriage and C-Arm),
these values can be different. The particular values should be seen only as references
pot8a
pot8b
3. To start the service interface, select Tune up > Stands > Potentiometer Adjustment,
Second Plane using the Select button and press Start.
4. Disengage the potentiometer and then adjust the potentiometer until the reference
value in the table is reached.
5. The actual value of the active potentiometer is displayed on the DDis in the “Stand
message line” (examination room). All potentiometer values are displayed on the serv‐
ice interface (control room).
6. Mesh the potentiometer. Compare the reference value to the actual value; when do‐
ing this, observe the tolerance.
7. Perform the linearity adjustment.
8. Move the axis over the entire movement range and look though the event log for
AX_SCU_ERR_1101. This error is generated when the potentiometer was not correctly
adjusted.
When moving the axes, make sure that there is not a drop below or a rise above the A/D
values, because this will destroy the potentiometer.
Max. range of the potentiometer: 0...4096
Range which may not be dropped below or exceeded: 50...4046
1. Due to the mechanical tolerances (installation of the ceiling rails, mech. tolerances of the overhead carriage and c-arm)
these values can differ. The value specified is for reference only
3 5
1
4
Card Collimator
Name Position Voltage AD Value
Potentiometer R181 0° 5.0 ± 0.05 V 2048 ± 20 Bit
N Collimator
Name Position Voltage AD Value
Potentiometer R181 0° 5.0 ± 0.05 V 2048 ± 20 Bit
With the ceiling stand, detector rotation is available only in combination with the 30 x
40 detector!
2. To start the service interface, select Tune up > Stands > Potentiometer Adjustment,
Floor stand using the Select button and press Start .
Fig. 33: Position check
To prevent unintentional rotation of the detector during the adjustment, the two power
cables on the motor ( 5/Fig. 35 Page 51) should be disconnected for this adjustment!
4. Tension the toothed belt for the potentiometer using the tensioning device
( 1/Fig. 34 Page 51).
The toothed belt must be tensioned so that no return play is generated when the direc‐
tion of rotation is changed!
Camera / Detector
Name Position Voltage AD Value
Potentiometer n.a. 0° 5.0 ± 0.02 V 2048 ± 10 Bit
Tune Up
Select UP
5.1 General
Prerequisite for the linearity adjustment is that the potentiometers have been correctly
adjusted.
A tape measure and a spirit level are needed as aids.
In the service screen, select Tune Up > Stand > Movement Linearity Range Adjustment
usingSelect and then press Start. The results of the last adjustment will be displayed in
the Result box. The individual axes can be selected by using the Set button.
Fig. 37: Movement Linearity Range Adjustment
Tune up
Movement Linearity Range Adjustment,
Second Plane
Stand
Adjustment Duration : ....
GROUP scu_axle_32767
GROUP SecondPlane_Carm_RotB_GRP
axle_ad_step_resolution : "7.572130"
axle_ad_pot_offset : "-15507.719727"
axle_ad_step_resolution_2 : "7.572130"
axle_ad_pot_offset_2 : "-15507.719727"
axle_ref_tag_position : 0
GROUP SecondPlane_Carm_OrbB_GRP
axle_ad_step_resolution : "3.159231"
axle_ad_pot_offset : "-12477.826172"
GROUP SecondPlane_ReceptorB_GRP
axle_ad_step_resolution : "-2.775208"
axle_ad_pot_offset : "16571.138672"
axle_ref_tag_position : 10867
GROUP SecondPlane_Carm_RotB_GRP
axle_ad_step_resolution : "-7.601351"
axle_ad_pot_offset : "5731.824219"
axle_ad_step_resolution_2 : "-7.516413"
axle_ad_pot_offset_2 : "5993.609863"
axle_ref_tag_position : 0
The axis is displayed on the data display and the direction or position in which the axis is
to be moved. Depending on the axis, movement must be into 1 to 3 positions.
Once the axis has been aligned using the spirit level or the tape measure, this position is
temporarily saved by pressing the Prog_1, Prog_2, Prog_3 buttons on the stand control
module (SCM).
The result of the adjustment is displayed on the data display. The axis adjustment is saved
by pressing the Store button and the next axis is displayed.
If the values are rejected, the message “Invalid values press System Reset” appears (see
“Reset” below). The adjustment can then be repeated.
If errors occur during the adjustment process, full or individual adjustments can be reset.
The adjustment must be saved before any readjustment can take place:
Use <Next> to move to the end of the adjustment.
n
5.2.1 CarmRot
Fig. 39: Linearity adjustment C-arm rotation
The result of the adjustment will be displayed. If a reference tag is not displayed, the ad‐
justment was not successful. In this case, check the reference tag!
4. The CarmRot axis is now adjusted and must be saved using the <Store> button on the
SCM control console.( Fig. 62 Page 76)
5. Skip to the next axis by pressing the “Next” button on the SCM ( Fig. 62 Page 76).
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently by clicking on the Save buttons in
the service interface.
5.2.2 CarmOrb
Fig. 40: Linearity adjustment C-arm orbital
The result of the adjustment will be displayed. If a reference tag is not displayed, the ad‐
justment was not successful. In this case, check the reference tag!
4. The CarmOrb axis is now adjusted and must be saved using the <Store> button on the
SCM control console.( Fig. 62 Page 76)
5. Skip to the next axis by pressing the “Next” button on the SCM ( Fig. 62 Page 76).
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently by clicking on the Save buttons in
the service interface.
5.2.3 ReceptorA
Fig. 41: Linearity Adjustment, Receptor A
The result of the adjustment will be displayed. If a reference tag is not displayed, the ad‐
justment was not successful. In this case, check the reference tag!
4. The ReceptorA axis is now adjusted and must be saved using the <Store> button on
the SCM control console.( Fig. 62 Page 76)
5. Skip to the next axis by pressing the “Next” button on the SCM ( Fig. 62 Page 76).
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently by clicking on the Save buttons in
the service interface.
5.2.4 BaseRotA
Fig. 42: Linearity Adjustment, Base Rotation
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently by clicking on the Save buttons in
the service interface.
The position of the collimator/detector must be set for this adjustment so that rotation of
+/- 90 degrees is possible.
X - Under fluoroscopy, align the vertical plates in the collimator so they are parallel to
the soldering wire mark ( 1/Fig. 43 Page 60).
- Save the position at the SCM control console by pressing the <Prog 1> button.
Fig. 43: Head-end position (FD system)
The collimator rotation direction is relative to the collimator itself and not to the symbols
on the CCM.
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently by clicking on the Save buttons in
the service interface.
The position of the collimator/detector must be set for this adjustment so that rotation of
+/- 90 degrees is possible.
Prior to adjusting the camera/detector rotation, you must adjust the collimator.
X
2. CameraRotA: 2 Pos, move to cw 90 ap, press Prog_2
Rotate the camera/detector clockwise approx. 90 degrees and save with the <Prog 2>
button.
Fig. 46: Linearity Adjustment, Collimator Rotation
5. Skip to the next axis by pressing the “Next” button on the SCM ( Fig. 38 Page 54).
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently by clicking on the Save buttons in
the service interface.
30 ˚
30 ˚
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently by clicking on the Save buttons in
the service interface.
If the table lift and table tilt linearity adjustments are performed without a subsequent
isocenter adjustment, it can result in incorrect distance measurements and incorrect re‐
sults in the Quantification option.
6.1.1 CarmRotB
Fig. 48: Linearity Adjustment L-Arm Rotation
90˚
Grafik mit
Head Foot
Markierungen
side side
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently only when the Save button is
clicked on in the service interface.
6.1.2 CarmOrbB
Fig. 49: Linearity adjustment L-arm orbital
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently only when the Save button is
clicked on in the service interface.
6.1.3 GantryLongB
6.1.3.1 Preparation
1. Place a steel ball or the tip of an Allen wrench with the ball end at the head end so it is
centered on the tabletop.
2. Mark the center of both image intensifiers with a washer on each.
6.1.3.2 Adjustment
B Ceiling Stand
towards foot end
2 1500 mm 1
4. Skip to the next axis by pressing the “Next” button on the SCM.
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently only when the Save button is
clicked on in the service interface.
6.1.4 ReceptorB
Fig. 53: Linearity adjustment receptor plane B
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently only when the Save button is
clicked on in the service interface.
6.1.5 CollimatorRotB:
This adjustment is required only in combination with the large 30x40 detector in
plane B in biplane systems!
1. Move the collimator and the detector into the following starting position (this position
can be reached easily by using the Portrait/Landscape buttons on the detector
( 1/Fig. 55 Page 70)):
- The long side of the collimator must be parallel to the C-Arm ( Fig. 54 Page 70).
- Move the detector into the landscape position ( Fig. 54 Page 70).
The collimator and the detector can be rotated 90˚ in both directions only in
this position!
Fig. 54: Detector/collimator linearity adjustment Fig. 55: Detector control panel
<Image System> <electronic Shutter> and open the shutter for Zoom 0 (the shutter
opens wider with a lower value; make a note of the original value).
X Under fluoroscopy, align the vertical plates in the collimator so they are parallel to the
soldering wire mark ( Fig. 56 Page 71). Save the position at the SCM control con‐
sole by pressing the <Prog 1> button.
Fig. 56: Collimator/camera rotation
The collimator rotation direction is relative to the collimator itself and not to the symbols
on the CCM.
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently by clicking on the Save buttons in
the service interface.
6.1.6 CameraRotB
This adjustment is required only in combination with the large 30x40 detector in
plane B in biplane systems.
Prior to adjusting the detector rotation, adjustment of the collimator must be performed.
1. Move the collimator and the detector into the following starting position (this position
can be reached easily by using the Portrait/Landscape buttons on the detector
( 1/Fig. 59 Page 73)):
- The long side of the collimator must be parallel to the C-Arm ( Fig. 58 Page 73).
- Move the detector into the landscape position ( Fig. 58 Page 73).
The collimator and the detector can be rotated 90˚ in both directions only in
this position!
Fig. 58: Detector/collimator linearity adjustment Fig. 59: Detector control panel
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently by clicking on the Save buttons in
the service interface.
Tune Up
Select UP
7.1 General
Due to the mechanical tolerances and to different loads exerted by a 40 cm I.I., the end
limit switches can respond in the rotation, orbital and receptor axes. With adjustment of
the end position, the position of the software end limit switches is redefined and thus
moving to the limit switches is avoided. Aids are not required for this adjustment.
In the service screen, select Tune Up > Stand > End Position Adjustment usingSelect
and then press Start. The results of the last adjustment will be displayed in the Result
box. The individual axes can be selected by using the Set button.
The axis is displayed on the data display and the direction or position which the axis is to
be moved. When the end position is reached (software end limit), the axis must be
moved further by pressing the Override keys (simultaneously press the Prog_1 and
Prog_2 keys) until the hardware limit switches are reached.
The result of the last adjustment is displayed on the data display. The axis is saved by
pressing the <Store> key and there is a skip to the next axis.
After completing the adjustment, the <Save> key in the service interface must be press‐
ed.
The adjustment must be saved in the service interface by pressing <Save>, otherwise the
entire adjustment will be lost.
Next
Reset
Store
Prog 2
7.2.1 C-ArmRotA
Fig. 63: End position adjustment - Carm RotA
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently only when the Save button is
clicked on in the service interface. A time frame is no longer specified.
7.2.2 CarmOrbA:
Fig. 64: End position adjustment - CarmOrbA
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently only when the Save button is
clicked on in the service interface. A time frame is no longer specified.
7.2.3 ReceptorA
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently only when the Save button is
clicked on in the service interface. A time frame is no longer specified.
8.1.1.1 CarmOrbB
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently by clicking on the Save buttons in
the service interface.
8.1.2.1 CarmOrbB
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently by clicking on the Save buttons in
the service interface.
8.1.3 ReceptorB
Fig. 69: End position adjustment - receptor B
1. If no other axis needs to be adjusted, continue to press the <Next> button on the SCM
until the message "PMS finished" is displayed on the DDIS.
2. The adjustment parameters are saved permanently by clicking on the Save buttons in
the service interface.
8.2.1 Adjustment
Fig. 70: Adjusting the position sensor
20 +5/-0mm 2 20 +5/-0mm
1
2
A sensor plate with 2 bifurcated photo-sensitive cells is installed on the side of the over‐
head carriage. The top photo-sensitive cell is used as the Ref Tag; the bottom photo-sensi‐
tive cell ( 2/Fig. 70 Page 86) together with the two position sensors
( 1/Fig. 70 Page 86) is used as the head-end and foot-end UL limit switches for longitudi‐
nal drive.
n Move the ceiling stand into the park position:
The ceiling stand must stop 20 mm in front of the position sensor.
n If there is a difference, loosen the Allen screw on the particular position sensor, move
the position sensor accordingly and lightly retighten the Allen screw.
n Move the ceiling stand into the left position:
The ceiling stand must stop 20 mm in front of the position sensor.
n If there is a difference, loosen the Allen screw on the second position sensor, move
the position sensor accordingly and lightly retighten the Allen screw.
n Highlight the "End Limit Switch Check, Second Plane" and select it with Select, then
press Start.
Fig. 71: End Limit Switch Check, Second Plane
n Move the ceiling stand out of the left position in the direction of the position sensor.
To do this, press the <PROG1> and <PROG2> buttons simultaneously on SCM1 and
use the joystick on SCM2 to move the ceiling stand to the limit switches.
The status for "Gantry longitudinal B, Low end" in the service screen changes
from --- to OK. At the same time, the counter increases by 1 each time the "lim‐
it switch" is actuated.
n Move the ceiling stand into the park position.
n Move the ceiling stand out of the park position in the direction of the position sensor.
To do this, again press the <PROG1> and <PROG2> buttons simultaneously on SCM1
and use the joystick on SCM2 to move the ceiling stand to the limit switches.
The status for "Gantry longitudinal B, Low end" in the service screen changes
from --- to OK. At the same time, the counter increases by 1 each time the limit
switch is actuated.
n Move the ceiling stand back out of the "limit switch" into the park position.
n Select <Home> and acknowledge the message The Process will be exited”
with <OK>.
9.1 General
In the service screen, select Tune Up > Stand > Geometry Adjustment usingSelect and
then press Start. The results of the last adjustment will be displayed in the Result box. The
individual axes can be selected by using the Set button.
The axis is displayed on the data display and the direction or position in which the axis is
to be moved. After reaching the end position (software limit), the axis must be moved
further by pressing the override buttons (simultaneously press the Prog_1 and Prog_2
buttons) until the hardware limit switches are reached.
The result of the adjustment is displayed on the data display. The axis adjustment is saved
by pressing the <Store> button and the next axis is displayed.
After completing adjustment, the Save button in the service interface must be pressed.
The adjustment must be saved in the service interface with <Save>, otherwise the entire
adjustment is rejected.
2. The geometry adjustment is completed. Save the values to the clipboard with the
Store button on the control console.
3. If necessary (for example if covers or protective wire connections were removed) per‐
form a protective conductor test.
2. The geometry adjustment is completed. Save the values to the clipboard with the
Store button on the control console.
2
5 1
1 3 4 6
If the play cannot removed using the one eccentric, then the second eccentric needs ad‐
justing.
1. Loosen the hex nut ( 1/Fig. 77 Page 93) and the set screw ( 2/Fig. 77 Page 93).
2. Move the eccentric ( 3/Fig. 77 Page 93) in the direction of the arrow and then check
the play:
There should be no more play between the gear teeth of the two latches and
the pinion gear. Both the gear teeth must lay cleanly against the contact surfa‐
ces ( 1/Fig. 78 Page 93) of the pinion gear. This is true when the stand cannot
be rotated by hand, i.e. there is no discernible play.
The unlock button needs to release easily.
If the play cannot be eliminated with the one eccentric, then move the second
eccentric ( 6/Fig. 77 Page 93) in the direction of the arrow. For this purpose,
loosen the hex nut ( 4/Fig. 77 Page 93) and the set screw
( 5/Fig. 77 Page 93) beforehand.
3. Secure the eccentrics with the two set screws and reaffix the hex nuts.
4. Check the function again.
1 3
3
/-0.
7+
1. By rotating the bracket ( 1/Fig. 79 Page 94) move the photo-sensitive cell
( 2/Fig. 79 Page 94) parallel to the ratchet lever ( 1/Fig. 79 Page 94) . Between the
the gray areas of the ratchet lever ( 1/Fig. 79 Page 94) and the front of the bifurca‐
ted photo-sensitive cells, set the distance to 7 ± 0,3 mm .
Fig. 83: Adjustment of the slip clutch Fig. 84: Detector rotation test
2. Place a loop on one corner of the detector (e.g. a cable tie) ( 7/Fig. 83 Page 96).
3. Use a spring balance ( Required Aids and Tools / Page 6) to measure the force that is
required to rotate the detector; to do this, hook the spring balance onto the loop and
measure the force at an angle of 45˚ ( Fig. 84 Page 96).
CARD \ 20x20 Detector:
If a force of 63 N, +/- 15 N is required, no adjustment is needed!
If the force is less than 48 N, an adjustment is needed!
Angio \ 30x40 Detector:
If a force of 34 N, +/- 8.5 N is required, no adjustment is needed!
If the force is less than 25.5 N, an adjustment is needed!
4. Tighten each of the adjustment screws for the slip clutch ( 1/Fig. 85 Page 97) 1/4
turn across from each other.
Fig. 85: Adjustment of detector rotation
10.4 Tensioning the Toothed Belt for Detector Rotation and the
Potentiometer
10.4.1 Prerequisite
The belt for detector rotation and the potentiometer must be loose.
n To release the detector rotation motor belt, loosen the 2 screws ( Fig. 87 Page 99)
and ( Fig. 88 Page 99), followed by the eccentric locking screw.
Fig. 90: Adjusting the tension with the eccentric Fig. 91: Locking the eccentric with the screw
6. If the tension is still a bit too high or too low, loosen all three screws and repeat steps
1 - 5 until the tension is correct.
Fig. 93: Tensioning the potentiometer
Fig. 94: Pressing the potentiometer belt together for fine adjustment of the tension
In accordance with IEC 62353, the measured values must be compared with those from
the last maintenance or repair session, and then documented.
If a significant difference is found or if the measured value exceeds 200 mOhm, it may
be necessary to initiate corrective measures.
Usually, the problem is that the ground wires are not connected, or incorrectly connec‐
ted.
n Power to the system must be switched off, and all cover panels must be in place.
n Measurement Ground wire test instrument, e.g. SECUTEST SIII ,
device: PROFITEST 0100S-II, Metriso G 500, Metraohm
413, Metraline RLO Check, MTECH*
11.3.1 Notes
11.3.1.2 Ground Wire Measurement on the Patient Table (not with MP)
3
1
n Because of the way the patient table is designed, the following parts cannot come in‐
to contact with voltage-conducting parts and are therefore excluded from the meas‐
urement ( Fig. 98 Page 108):
- Rail cover panel (1)
- Front lift cover panel (2)
siemens-healthineers.com