Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Associated Research 7850 Datasheet

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

HypotULTRA

ALL-NEW AVAILABLE INTERFACES


® SCAN TO VIEW
QUICK START
VIDEO

THE MOST FLEXIBLE AND FEATURE-RICH USB


USB RS-232
RS-232 Ethernet
Ethernet GPIB
GPIB

AUTOMATED DIELECTRIC
ANALYZER AVAILABLE SAFETY & PRODUCTIVITY
FEATURES
EN 61010-1
EN 61010-31

Warranty
Smart
Smart GFI
GFI®®
Remote Safety
Remote Safety Data Transfer
Data Transfer
Automatic Interlock
Interlock Easily import/
operator shock Easily disable export test
protection HV output files and data
via USB

The best dielectric analyzer available just got better. We’ve combined
superior testing power and ease of use, with an innovative sleek design
that showcases all of our productivity and safety enhancing features.
Our touch screen capability allows you to interact with your instrument
as intuitively as you would with a smart phone. This simplifies setting up Barcode
Barcode
Capability
Multiple
Multiple
Languages
BatchTEST
Batch ®
Test™
Reduce test
Capability Languages
your system and test parameters. Get even more out of your instrument Direct barcode
connection
Multi-Language
user interface
time with
simultaneous
with direct barcode connection, this all-new feature increases efficiency DUT testing

and production throughput. HypotULTRA® will improve the productivity


and safety of your production line in every single way.

ProVOLT™
ProVOLT™ Internal
Internal Modular
Modular
Multi-dwell Scanner
Scanner Scanner
Scanner
cycles at Available with Compatible
different optional HV with SC6540
voltages for scanning matrix scanning
ACW/DCW/IR matrix

FailCHEK™
FailCHEK ®
Prompt
Prompt & Hold
& Touch Screen
Touch Screen
Confirms Provides
Holdalerts Interact
failure & instructions with your
detection between tests instrument like
a smartphone

Autoware®®3
Autoware 3 Advanced
Advanced Ramp-HI®
Ramp-HI ®

Advanced User Security


User Security Reduce ramp
Automation Customize ID time during
Find the Right Model that Fits Your Testing Needs Control
Software
& password
protection
DC Hipot

AC Hipot DC Hipot Ground Insulation 500 VA


Continuity Resistance Charge-LO®
Charge-LO ®
Accredited
Accredited Negative
Negative
Confirms Cal
Cal DC Hipot
DC Hipot
proper DUT Accredited Reverse
7800* • • • • • connection calibration polarity
options DC Hipot
7820 • • available (optional)

7850 • • • •
*Meets 200 mA short circuit requirements

6 SAFETY IS OUR ONLY FOCUS® arisafety.com


HypotULTRA®

INPUT SPECIFICATIONS INSULATION RESISTANCE (7850 & 7800 ONLY)


Voltage 100 - 120 VAC / 200 - 240 VAC±10% Auto Range Output Voltage, Range: 10-1,000 VDC
Frequency 50/60 Hz ± 5%" DC Resolution: 1 VDC
Fuse 7820 and 7850: 6.3A / 250 VAC Slow-Blow, Accuracy: ± (2% of reading + 2 counts)
7800: 15 A / 250 VAC Fast- Blow Range: 1001-6000 VDC
Resolution: 1 VDC
AC WITHSTAND TEST MODE
Accuracy: ± (2% of setting + 5 V)

Charging Current Maximum > 10 mA peak


Output Voltage Range: 0-5,000 VAC
HI & LO-Limit Range: 0.10 M – 99.99 MΩ (HI-Limit: 0 = OFF)
Resolution: 1 VAC
1.00 - 99.99 when voltage > 1,000 V
Accuracy: ± (2% of setting + 5 V)
Resolution: 0.01 M
Output Frequency 50/60 Hz ± 0.1% , User Selection
Accuracy: ±(2% if setting + 2 counts)
Output Waveform Sine Wave , Crest Factor = 1.3 - 1.5
Range: 100.0 M – 999.9 M
Output Regulation ±(1% of output + 5 V)
Resolution: 0.1 M
HI and LO-Limit Total Range: 0.000 – 9.999 mA
Accuracy: 1,000-9,999 ±(5% if setting + 2 counts)
Resolution: 0.001 mA
Range: 1,000 M – 50,000 M
Range: 10.00 – 30.00 mA
Resolution: 1M
(10-99.99 mA, Model 7800)
Accuracy: 10,000-50,000 M ±(15% if setting
Resolution: 0.01 mA
+ 2 counts)
Accuracy: 7820 & 7850 ± (2% of setting
+ 2 counts), 7800: ± (2% of setting
Ramp Up Timer Range: 0.1 – 999.9 sec.
+ 6 counts)
Ramp Down Timer Range: 0.0, 1.0 – 999.9 sec.
Real Range: 0.000 – 9.999 mA
Dwell Timer Range: 0, 0.5 – 999.9 sec. or 0
Resolution: 0.001 mA
Delay Timer Range: 0, 0.5 – 999.9 sec. or 0
Range: 10.00 – 30.00 mA
Charge-LO 0.000-3.500 μA or Auto Set
(10-99.99 mA, Models 7800)
Resolution: 0.01 mA
Accuracy: ± (3% of setting + 50 μA) CONTINUITY TEST
Ramp Up Timer Range: 0.1 – 999.9 sec. Output Current, DC 1A for 0.000 - 1.000 Ω
Ramp Down Timer Range: 0.0 – 999.9 sec. 0.1A for 1.01-10.00 Ω
Dwell Timer Range: 0, 0.2 – 999.9 sec. (0=Continuous) 0.01 A for 10.01 - 100 Ω
Ground Continuity Current: DC 0.1 A ± 0.01 A, fixed 0.001 A for 101-1,000 Ω
Current Max. ground resistance: 1.0 Ω ± 0.1 Ω 0.0001 A for 1001-10,000 Ω
Arc Detection Range: 1 - 9 ranges (9 is most sensitive) 1 A is Max

Resistance Display Max & Min


DC WITHSTAND VOLTAGE (7850 & 7800 ONLY) Max-Lmt Range: 0.000 – 1.000 Ω
Resolution: 0.001 Ω
Output Voltage Range: 0 -6000 VDC
Accuracy: ± (1 % of setting + 3 counts)
DC Output Ripple <4 % (6KV/10mA at Resistive Load) Range: 1.01 – 10.00 Ω
HI and LO-Limit
Range: 0.0000-0.9999 μA Resolution: 0.01 Ω
Resolution: 0.0001 μA Accuracy: ± (1 % of setting + 3 counts)
Accuracy: ± (2% of setting + 10 counts) Range: 10.1 – 100.0 Ω
Low Range is ON. Resolution: 0.1 Ω
Range: 1.000 - 9.999 μA Accuracy: ± (1 % of setting + 3 counts)
Resolution: 0.001 μA Range: 101 – 1,000 Ω
Accuracy: ± (2% of setting + 10 counts) Resolution: 1Ω
Low Range is ON. Accuracy: ± (1 % of setting + 3 counts)
Range: 10.00 - 99.99 μA Range: 1001 – 10,000 Ω
Resolution: 0.01 μA Resolution: 1Ω
Accuracy: ± (2% of setting + 10 counts) Accuracy: ± (1 % of setting + 10 counts)
Low Range is ON. Dwell Timer Range: 0, 0.4 – 999.9 sec. (0=Continuous)
Range: 100.0 - 999.9 μA Resistance Offset Range: 0.000-10.00 Ω
Resolution: 0.1 μA
Accuracy: ± (2% of setting + 2 counts) GENERAL SPECIFICATIONS
Range: 1000 - 10000 μA
Resolution: 1 μA Memory 2,000 steps
Accuracy: ± (2% of setting + 2 counts) 200 steps per test file max
Interface Standard: USB/RS232,
Ramp Up Timer Range: 0.4 - 999.9 sec. Optional: GPIB (IEEE-488.2),
0.5-999.9 sec., Low Range is ON RS232/Ethernet or USB Printer.
Ramp Down Timer Range: 0.0, 1.0 - 999.9 sec. (0=OFF) SmartGFI® 0, 0.4-5.0-5.0mA (0=OFF)
Dwell Timer Range: 0, 0.4 - 999.9 sec.,(0=Continuous)
Dimensions Bench or rack mount (2U height) w/ feet
0,1.0-999.9 sec., Low Range is ON (W x H x D)
RAMP-HI Selectable Range: 0-10 mA, Selectable 16.92 x 3.50 x 15.75in, (430 x 88.1 x 400) mm
Charge-LO Range: 0.0 - 350.0 μA DC or Auto Set,
Weight 7820 = 34 lb (15.4 kg), 7850 = 35 lb (15.9 kg),
Discharge Time < 50 ms for no load 7800 = 45 lb (20.4 kg)
< 100 ms for capacitive load
Maximum 1μF < 1kV 0.08 μF < 4 kV
Capacitive 0.75 μF < 2 kV 0.04 μF < 5 kV Why We Use Counts
Load DC Mode 0.5 μF < 3 kV 0.015 μF < 6 kV Associated Research publishes some specifications using “counts” which allows us
Arc Detection Range: 1 - 9 ranges (9 is most sensitive) to provide a better indication of the tester’s capabilities across measurement ranges.
A count refers to the lowest resolution of the display for a given measurement range.
For example, if the resolution for voltage is 1V then 2 counts=2V.

Specifications subject to change without notice.

(USA) TOLL FREE: +1.800.858.TEST (8378) Phone: +1.847.367.4077 7