LOTO PC Oscilloscope Software User Manual Version22.11.23
1 Welcome
Thanks for choosing LOTO USB oscilloscope solutions. LOTO Instruments is committed to
the development of high-performance software for virtual instrumentation products. It has been
committed to the research and development of virtual instruments for many years, improving the
cost, the functional architecture of traditional instruments and providing cost-effective products.
Welcome to contact us (173393190@qq.com) for any suggestion, customization development
or business cooperation.
2 Software Version
The hardware specifications of different models can be found in their respective data
sheets. This manual is based on the software version in the list below.
OSCA02D
OSCA02L
OSCA02M
OSCA02S
OSCA02X
OSCA02F
OSCA02H
OSC2002
OSC2002E
OSC2002L
OSC2002D
OSC2002 OSC2002S version 22.11.17 2022/11/17
OSC2002X
OSC2002F
OSC2002H
OSC2002M
OSC980
OSC980 version 22.11.17 2022/11/17
OSC984
SIG851 version 1.0.1 2020/12/10
SIG852 version 1.0.2 2022/01/09
OSCH02
OSCH02L
OSCH02D
OSCH02 OSCH02S version 22.11.17 2022/11/17
OSCH02X
OSCH02F
OSCH02H
To make sure LOTO users could have the best experiences and better products, we keep
updating oscilloscope software-correcting bugs, adding practical features, and improving
software operations.
The latest version of software &driver &documents will be updated to the link:
http://www.loto-ins.com/en/
https://drive.google.com/open?id=1Py0I-_Uud0JqKU4wiCyk8YOeHiVqj_y-
No matter when you download, you will download the latest version. There is no fixed
schedule of updates, we usually release a new version in every one or two months.
-7-
LOTO PC Oscilloscope Software User Manual Version22.11.23
You can also contact us via Facebook public group: LOTO USB oscilloscopes
4 Safety Warning
△1 Make sure that the input of the device does not exceed the voltage range.
△2 Make sure that the black clip of the probe is well grounded and equipotential
to the PC power supply during measurement.
△3 Do not connect the probe to the mains voltage socket (110V / 220V).
△1 note: the oscilloscope device has an input range of ±5v and it should never be exposed to
higher voltages although it has an over-voltage protection circuit which can protect the device
from voltages even ten times the maximum allowed.
To avoid the risk of damage, it is required to select an input probe with the attenuation
factor according the nature of the voltage to be measured to ensure that the device is used
within the voltage range.
The device is provided with a 1X/10X probe: when the 1X position is selected, the voltage
input range is ± 5v. When the 10X position is selected, the voltage input range is ±50v. If an
optional 100X probe is used, the voltage input range is ±500v. Optional isolated differential
module can measure ±800v.
△2 note: the virtual oscilloscope device takes power directly from the USB port of the PC, so
its ground is connected to the PC ground as also the black clip of the probe. If the PC is
powered by the power supply from the mains voltage socket, then the PC should be connected
-8-
LOTO PC Oscilloscope Software User Manual Version22.11.23
to the ground through a 3-hole power socket. In case the device is connected to a PAD or laptop
and these are powered just by a battery, then the ground will be suspended. But in most cases,
the device ground, PC ground and probe black clips are all connected to the ground. If the black
clip is used to connect a non-earth potential in the circuit under test, it is equivalent to short-
circuiting the circuit under test with the black clip directly to the ground.
Warning
The software application requires Microsoft .NET framework 2.0. This component is an
integral part of Windows systems since Win 7, so there is no need to install it on these
systems, however on Windows XP system it may be required to install this component to
use the software. The software installation process will determine if this component is
required and will download and install it if necessary.
We designed this virtual oscilloscope to be as simple as possible. Even connecting it for the
first-time can be quick by following the steps below.
(1) : Connect the virtual oscilloscope to the PC via the supplied USB cable. You should use
only the supplied USB cable, or use a cable of a better quality. Poor quality cables may cause
communication problems or may become unusable. If you are using the device on a desktop
PC, be sure to connect it to the USB port on the back of the chassis. The front ports may also
be unstable due to power supply issues. If you use a laptop, it is easy to connect to the USB
ports on the sides.
(2) : Install the driver. When the virtual oscilloscope is connected to the PC for the first
time, in Win 8.1 and Win10/11 systems the device driver will be automatically recognized and
installed. In Win XP and Win 7 32-bit operating systems the device will be recognized, and the
system will require the driver installation. The steps for installing the driver are different
according the operating system. See Sections 8.1 and 8.2 for more information on the specific
installation process.
(3): Open the virtual oscilloscope software, click the Start Device button in the lower right
corner, then you will see that channel A has been turned on by default, with the zero-voltage
baseline in the middle of the screen. Channel B can be enabled in its setting area. When the
zero-voltage baseline is in the vertical center, this indicates that the virtual oscilloscope is
operating normally.
-----
- 10 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
As shown in the figure above, after turning on all the oscilloscope channel switches, there
are noise lines, which means that the virtual oscilloscope can work normally.
If the driver is not installed successfully, or the oscilloscope device is not connected to the
computer, there will be a yellow warning bar above the LOTO oscilloscope software:
- 11 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
(4): Use the standard probe calibration signal to display your first waveform on the screen.
Some kind of device has an extended metal contact next to the two BNC input connectors that
output a square wave with amplitude of approximately 3.3V and a frequency of 1 KHz. The
other kind of device has this square wave output in pin 6 of a DE-15 interface.
This signal allows a quick check whether the product is working normally when there is no
other AC signal at hand. Set one probe with x1 selector; connect it to the oscilloscope channel A
- 12 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
port from the BNC connector side and to the metal contact on the other side. Adjusts the time
knob in the software to 1ms position. The square wave will appear on the screen.
Measuring the built-in square wave is an important way to check the quality of the
oscilloscope. If you are not sure where the problem is in your actual application, please
disconnect the oscilloscope from the target system as soon as possible, and use the
oscilloscope to test the built-in square wave.
Wave to troubleshoot whether it is the problem of the oscilloscope. If it is normal for the
oscilloscope to measure its own square wave, then the oscilloscope can basically conclude that
there is no problem.
- 13 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
7 Drivers Installation
This chapter describes the detailed steps of installing drivers for this product under various
types of Windows operating systems, which can cover the guidance most users need to install
drivers. However, in some users' operating systems, there will be exceptional situations of driver
conflict, which will lead to the failure of driver installation or the failure of normal opening of the
device when the installation is successful. Users can refer to the detailed solutions for driver
installation under exceptional circumstances listed in Appendix 1 of this manual.
- 14 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
2: Click on Browse, specify the search path as "(software installation package path)\driver". For
example "Driver_XP_Win7\driver", then click next:
- 15 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
3: Once the installation is complete, you can see the installed driver in the Device Manager:
- 16 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Warning:
Windows 7 – (32bit and 64 bit) cannot install the driver automatically. It is required to install
the driver manually.
Windows 7 will automatically search for the driver through Windows Update when the
device is connected, but this automatic installation will fail since the driver is not present in the
system:
Manual installation:
- 17 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
1. Right click on Computer, then select Manage. Enter in the Device Manager, search for the
device with name OSCxxx with the yellow exclamation mark in Other devices, and then
right click on it and select Update the Driver Software:
3. In the next dialog box specify the driver location as "(Installation package path)\driver" via
the Browse... button. For example ":\ Driver_XP_Win7\driver":
- 18 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
- 19 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Manual installation
1. Right click on Computer, then select Manage. Enter in the Device Manager, search for the
device with name OSCxxx with the yellow exclamation mark in Other devices, and then
right click on it and select Update the Driver Software.
2. Use the Browse... button to specify the driver location as "(Installation Package Path)\driver".
For example ":\ Driver_Win7_64\driver".
3. Other steps are the same as for the above windows 7 32-bit system.
The system will automatically install the drivers for the device, and the user does not
need to proceed with any manual installation. Some systems cannot recognize the oscilloscope
device normally, please follow the steps in Appendix 1 to install the driver.
- 20 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
If the above steps still do not allow to install the driver properly, in Appendix I you can find a
detailed solution.
This section describes the basic concepts and usage of the virtual oscilloscope software.
When displaying the voltage-time relationship, the abscissa represents the time value
(from left to right for the direction of time growth), and the ordinate represents the voltage value
(from bottom to top for lower voltages to higher voltages).
Unit Description
(Time)
s Seconds
ms Milliseconds, that is, one thousandth of a second (10-3 s)
us Microsecond, one thousandth of a millisecond (10-6 s)
ns Nanoseconds, one thousandth of a microsecond (10-9 s)
v Voltage in Volts
mv Millivolt, that is, one thousandth of a Volt (10-3 V)
Waveform plots: In the oscilloscope software the drawing area of the waveform is evenly
divided with grids which serve for quickly measurement on the signal.
The oscilloscope software interface provides two types of knobs for setting the grid units,
one for the time and one for the voltage.
- 21 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
All the input channels of this oscilloscope are synchronized in time, so they share the
same time axis and the same time settings from the corresponding knob.
The sampling rate from the Analog to Digital converter in the oscilloscope hardware vary
automatically according the selected time scale so to make a better usage of the device internal
memory, without losing resolution in the time analysis.
- 22 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
- 23 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
After the software interface is opened, the OSCxxx device status will be monitored in real
time. When the device is connected on the USB port of the PC, the software interface will
display a message on blue background.
If the device gets disconnected physically or logically, the software interface will display a
message on yellow background.
If the software always displays a yellow bar after plugging in the device, it means that the
connection is abnormal or the oscilloscope driver is not installed.
- 24 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
After the driver has been successfully installed, if the oscilloscope is connected when the
software is opened, the software automatically turns on the device and starts the real-time
acquisition. If the device is connected before opening the software, it is required to start the
device manually by clicking on the Start Device button in the lower right corner of the software
or by pressing the Enter key (shortcut key (Enter) - start/stop) on the keyboard.
After the device is started the Start Device button becomes the Stop Device button which
allows to stop the acquisition at any time. The acquisition can be stopped also by pressing the
Space Bar on the keyboard or when the USB cable is unplugged during the operations.
The start/stop button can only control the device real-time acquisition, but it is not used for
play/stop operations on past recorded data. In this case there are specific control buttons. For
details, please refer to Chapter 9.2.8.
Pressing the boot device button usually encounters two error messages: 0 and 1.
- 25 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
If the error code 0 is reported, the possible reasons are that the USB cable is
damaged/device is damaged, the driver is not installed, the device is not connected, or the
software and the oscilloscope model do not correspond.
If Error code2 is reported, the possible reason is that the USB power supply is insufficient,
or the driver of the oscilloscope device conflicts with other drivers, or the oscilloscope is set as a
slave device when it is cascaded.
Note
When the device is not used it is recommended to close the software and then to unplug
the device's USB cable.
The waveforms are drawn on the plot area using different colors to distinguish them. The
default background color of the plot area is dark blue. The waveform for channel A (chA) and
channel B (chB) are drawn respectively in cyan and yellow colors: .
- 26 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The channel’s waveforms are drawn according their zero-voltage baselines. These lines
represent 0 Volts; positive voltages are drawn above while negative voltages are drawn below
the baselines. The zero-voltage baseline position of each channel is indicated by a horizontal
line and a triangle on the side which can be dragged with the mouse to move the baseline. A
number inside the triangle indicates the corresponding channel number. By default, the zero-
voltage baseline is set in the middle of the screen for all channels. Channel A (chA) waveform
and the channel B (chB) waveform uses respectively blue and red colors for the zero-voltage
baselines, as: , . When the mouse is placed on the triangle of the zero-voltage baseline
and the left button is clicked, you can drag up and down the line to change its position.
The zero-voltage baseline for each channel is identified by a triangle on the right side of
the waveform display area. Channel A baseline is in blue while channel B baseline is in red. You
can use the mouse to slide up and down these triangles to adjust the levels for the zero-voltage
baselines. The grid values on the left axis will change accordingly. By default, the zero-voltage
baselines for channel A and channel B are in the middle of the display area.
The user can self-calibrate the zero voltage level using this function. Zero level calibration
is required since different environment conditions (as temperature) or aging of the hardware
components. When calibrating the zero position, please be careful not to input any signal, just
let the channel input hang.
The above chapters have explained the accuracy of the equipment caused by factors such
as ambient temperature and component aging. Our products also support Calibration correction
function for zero position and voltage amplitude. Customers can calibrate the oscilloscope with
their own standard source.
- 27 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Click will pop up the calibration window. Before calibrating, there are a few precautions:
1. Before calibrating the zero position, you need to turn on both channels and adjust the
time base gear to the ms level, such as 0.5ms,1ms etc. The current zero position calibration
does not support permanent storage after the user self-calibration. If you need to calibrate zero,
you need to perform calibration when opening the device software. Note that when calibrating
the zero position, do not connect any signal to the probe of the oscilloscope and keep it away
from any interference.
2. When performing voltage calibration, first adjust the voltage knob of the channel to be
calibrated to the voltage range that needs to be calibrated, and then Click the calibration
button to call up the calibration window, select the channel voltage slider to calibrate, after
calibration, click the save button.
For example: We feel that the 1V voltage scale measurement value of channel chA is not
accurate enough. The calibration steps are as follows:
a. Set the voltage wheel gear of channel chA to 1V;
b. Click the button and the calibration window will pop up;
- 28 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
c. Connect the probe of channel chA to a signal source that we consider to be standard,
and set the signal source to output a standard sine Wave, the amplitude is for example ±2V.
false. Assume. The display of the oscilloscope is shown in the figure below: From the figure, we
can see the waveform and peak Both the trough and the trough exceed the voltage grid line of
±2V.
d. Adjust the voltage calibration slider of chA to adjust the waveform to the ±2V position,
and click the Save button to save.
- 29 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
e. At this time, there will be one more folder (calibrate) in the root directory of the
oscilloscope software, and one more file in the folder (chA1v.set).
f. For other voltage range calibration steps, follow the steps described in a ~ e. After the
calibration is completed, you need to close the oscilloscope software,restart the software, then
the user calibration parameters of the voltage range will take effect.
If you want to restore the factory default calibration parameters of all voltage ranges, click
Reset of default button, and restart the oscilloscope software.
The display control functions can be accessed using the waveform display toolbar located
on the right side of the waveform display area. When the mouse is placed over one of the
buttons of this toolbar, a tooltip will show the name of the function corresponding to the button.
When button is clicked, the waveform display area is restored to the default state,
which is the original state according the knobs settings and without any zoom.
When button is clicked, the mouse cursor will become a cross and will allow to select
an area for zooming. Press on the left button of the mouse to start the zoom area selection until
the area of interest is no completely selected. Releasing the left button of the mouse, the
selected area will be enlarged to the size of the full plot area.
- 30 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
This operation can be repeated also in already zoomed areas. When this function is used
a thumbnail appears in the lower right corner of the drawing area to allow to find the area
currently selected and its position respect the full waveform area. This function will enlarge the
horizontal and vertical coordinates simultaneously. If you want to zoom only according the
horizontal axis and do not change the zoom level of the voltage axis, you can use the X Axis
Zoom button, which will be introduced later.
The Default Display button mentioned above allows to end the zoom-in state and to
return to the original view state.
When button is clicked, the mouse cursor will become a cross, allowing to select an
area. Pressing the left button of the mouse the selection starts until the mouse button is not
released. After the area has been selected the voltage and time span of the selected area will
be displayed. Just one area can be selected with this tool so new selections overwrite the
previous ones.
In white it is reported the time span. In the same color as each channel waveform is
reported the voltage span selected of the corresponding channel. This measurement is
convenient and mainly used for calculating time and voltage intervals.
- 31 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Clicking on button, two green horizontal lines and vertical lines will appear in the
display area. Dragging with the mouse the green triangle on the side of these marking lines, it is
possible to adjust the position of the markings for data measurement.
- 32 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
These two measurement methods are mutually exclusive so that only one kind can be
chosen at time. The Marking Measurement is more suitable for waveforms. The Cursor
Measurement is more suitable for punctual measurement on certain points of the waveform.
When paused, a lock icon will appear on the vertical marking. Click the lock icon to lock the
vertical marking. After locking the vertical marking, you can use the paused buffer slider to move
left and right to measure the data in the buffer.
The button zoom the display area along the X axis for a better observation of the
period and frequency of the signal. When this button is clicked, a little window with the full
waveform will appear in the lower right corner.
- 33 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
In the display area, select the area to be enlarged by dragging the mouse (the zooming
operation is limited only to the X axis). The selected area will cover the entire display interface.
In the little windows at the lower right corner, two white lines identify the current visualized area
of the waveform.
The button is enabled only when the waveform has been zoomed in. By clicking on it
the visualized waveform can be continuously reduced by clicking on the visualized waveform
until it is drawn fully on the screen.
The button saves the data collected by the oscilloscope in the user's computer in plain
text format (.txt). The destination path can be chosen during the process. The detailed function
of this button is described in detail in 8.2.32.
- 34 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The button takes a screenshot of the current screen and save it as a picture for later
analysis and viewing. You can select the destination path, file name and the file format among
the followings: .jpg, .bmp, and .gif ( shortcut key (Ctrl + P) - screen-shots).
The button shows the Fourier Transform of the original measured data coming from
the enabled channels, to allow the frequency analysis of the signals.
Analyze the frequency domain of the signal and support dB display.
Support mouse frame selection or input some frequency bands for display.
In advanced settings, it supports specified fundamental and harmonic analysis, harmonic
comparison analysis, and waveform distortion THD detection.
Supports the setting of fundamental wave and harmonic rules, and alarms for waveforms
that do not meet user-defined rules. The alarm supports software alarms and IO port alarms.
Suitable for application scenarios such as audio analysis.
Enter the frequency range in the text box and click the set button to specify the frequency
range and select the frequency band to display the observation frequency.
- 35 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
As shown in the figure: Some models such as: OSC980 have the measuring line function
enabled by default.
Click the Save Export button to export the FFT data (frequency and voltage value)
as a file, supporting CSV and TXT formats.
After clicking the button , you can use the left mouse button to select an area in the FFT
waveform display area, and perform the mouse selection to observe and display. Click
to restore the FFT default display. Note: When using the dB mode to display FFT, you need
to restore the FFT default display. The dB mode currently does not support the mouse to
select the frequency band display.
- 36 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Check the dB checkbox, FFT will display frequency domain waveform in dB. Note: The dB
mode display does not support the specified frequency domain. If the FFT frequency
domain was specified in the previous operation, you need to restore the FFT default display.
Click the Set button to call out the FFT advanced window:
- 37 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
1. Select Auto mode, the oscilloscope will automatically identify the fundamental wave of
the waveform, and display up to the 6th harmonic. Select Manual mode, the user enters the
frequency as the fundamental wave in the text box of the 1st wave, and the software calculates
and displays up to 6 harmonics based on the fundamental wave input by the user.
2. Value tolerance slide bar, set the software to find a wide range for each harmonic.
Setting too narrow or too wide will affect the precise positioning of the harmonic peak search.
3. The Total harmonic distortion option is used to measure the THD of waveform distortion,
mainly for sine waves.
4. With the Harmonic component (p-p V) function, the user can specify the dB value
corresponding to the peak-to-peak value (p-p) of the waveform, and set the peak-to-peak
comparison rule. When the rule is not met, it will alarm. The alarm supports software alarm and
IO port alarm.
5. Harmonic component (K Hz), the user can specify the comparison rules between
different harmonics, and alarm when the rules are not met. The alarm supports software alarm
and IO port alarm.
6. The Comparison function displays the final result of the comparison between Harmonic
component and Harmonic component rules. As long as one does not meet the rules, the final
result is displayed in red, and all qualified are displayed in green.
7. Load the button to save the historical setting parameters. The user can save the set FFT
harmonic comparison rules into a file, use the Load Setup button to browse the file to load the
historical rules, and use the Save Setup button to save the settings into a file.
- 38 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
8. As shown in the figure below, the FFT function also supports the phase-frequency curve
display, which can visually display the phase-frequency curve of the waveform participating in
the FFT.
The button performs mathematical operations between the signals from channel A
and channel B like addition, subtraction, multiplication, or XY plotting. Once clicked on this
button, you can select the desired operation:
- 39 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
“A+B” means that the channel A waveform is added to the channel B waveform.
Selecting this operation, the display area will show both channel A and channel B as stacked
waveforms and the waveform resulting from the operation in purple color.
“A-B” is used to realize the subtraction between channel A and channel B signals. In similar
way as in the addition both channels’ waveforms are displayed together with the result of the
subtraction, but this time in green color.
The “AxB” operation is used to multiply channel A and the channel B signals. The result is
shown in the orange color.
- 40 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The "·A·B" button is used to hide the signal of the AB channel. After hiding, only the results
of the A and B mathematical operations are displayed, as shown in the following figure:
XY Plot or “Lissajous curves” plot. This display mode plots one channel against another on
the screen. One channel is plotted on X-axis, the other channel is plotted on Y-axis, together
with the information of the measured frequency of each signal. From the resulting curve it is
possible to calculate the frequency and phase relation of the signals.
- 41 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Through the "Lissajous figure", the relationship between the frequency and phase of the
two measured signals can be displayed intuitively.
- 42 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Through the "Lissajous graph", the UI characteristic curve of the diode can also be
measured. We connect the scanning voltage of the diode to channel A, and connect the current
flowing through the diode to channel B using LOTO's mA current module, and then we can
directly The UI characteristic curve of the measured diode is shown in the figure below:
The picture above shows the UI curve of a normal diode. If the diode is damaged as an
open circuit, the curve will be a horizontal line, and if it is short-circuited, the curve will be a
vertical line.
- 43 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Through the "Lissajous figure", the UI characteristic curve of the triode can also be
measured.
The same as the UI curve of the diode, we connect the scanning voltage UCE of the triode
to channel A, and use the current IC of the corresponding triode flowing through the C pole to
use LOTO.
When the mA current module is connected to channel B, the UI characteristic curve of the
triode can be directly measured under a certain fixed base current IB.
Different from diodes, triodes need to measure multiple sets of such UI curves under
different base current IB and display them on the graph at the same time, so we have a right-
click function, which is to obtain a UI curve after changing IB every time, and then select Keep
this curve fixed on the interface, as shown in the figure below:
- 44 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The button switches the background color from dark to white and the other way
around. The dark background is good for long-term observations to alleviate the eye fatigue.
The white background is convenient for taking screen shots for making reports or for further
image processing.
- 45 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The button hide or show again the grid displayed in the waveform area.
The “ T “ button determines which axis will be control by the mouse wheel when the cursor
- 46 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
is located in the waveform area. When this button shows “T”, we scroll the mouse wheel will
give the time axis a fine adjustment to zoom in or out the waveform horizontally.
By clicking “T”, this button will show “chA”, and another clicking will show a “chB”, which
means the mouse wheel scrolling will lead to a fine adjustment to zoom chA or chB vertically.
Also, we can do this with three controls in the right lower corner as bellow. However, we
may not see these controls when running on a PC with a small screen of screen resolution
below 1280*1024.
- 47 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Click the "F" button in the menu to change the waveform display area to full screen
display .Click the button again to restore the default.
Click the "P" button in the menu, and a 720° phase measurement line will appear in
the waveform display area. Click the button again to restore the default.
This function currently only appears in some models of software, which is convenient
for the measurement needs of industries such as automobile maintenance.
- 48 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
- 49 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Let's take the OSC984 host computer software as an example. The blue waveform
of channel A in the figure below is the crankshaft waveform of Baojun Automobile. With
such a dense waveform, we cannot see whether the speed is stable with the naked eye.
When we open the speed curve, as shown in the figure As shown by the purple
waveform in , we can see that the speed drops intermittently.
We collected another crankshaft waveform and compared it, we can clearly see
that the speed of the following waveform is very stable, and there is no obvious speed
drop, as shown in the figure below:
- 50 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
For this example automotive waveform diagnostic application, we can analyze the
problem represented by the waveform:
- 51 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
In the "Advanced" tab page, we can set the frequency of the coordinate axis of the
speed-frequency curve or which speed to use, and also set the magnification of this
curve, which is convenient for intuitive viewing of fluctuations:
When the acquisition is paused, a waveform thumbnail as well as its slider of the whole
memory buffer appears in the lower right corner of the waveform area. This slider shows the
ratio and the position of the waveform data currently displayed on the screen, related to the size
of the entire data buffer. The entire length of the slider represents the entire data buffer. The
position and length of the light slider represents the data block displayed in the current drawing
area relative to the entire data buffer. The length of the light slider can be used also to
understand the ratio between the display data respect the total acquired data. The waveform
thumbnail shows the waveform data displayed and the other waveform data in the buffer.
With the mouse you can drag the slider left and right in the entire data buffer, to show new
areas of interest. Another way to move within the memory buffer is to use the left and right arrow
keys of the keyboard. (Shortcut key (Ctrl + <-) / (Ctrl + ->) - Waveform move between left and
right).
- 52 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
When the PC buffer function is checked, the software will queue up each frame of data
collected by the oscilloscope on a first-in-first-out basis for frame buffer. When you find that
there is a waveform of interest passing by on the screen, click the (Pause) button of the
software, and you can choose to review a certain frame of waveform. The amount of one frame
of data is the total data size of the time base gear buffer selected by the current user. Different
time base gears have different buffer sizes, so how long a specific frame can be stored depends
on the time base gear selected by the user.
The maximum number of PC buffered frames supports 500 frames, which can be set in the
advanced page:
- 53 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
After the oscilloscope software is paused, the PC buffer function also supports: exporting
buffered data frames to a file, and importing historical buffered frame data. In order to cache the
frames of the historical PC of interest, perform local file persistent storage and review.
- 54 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
When saving you can set the destination path; the file suffix used for the saved file is .set. After
saving, you can load a saved setup using the menu File → Load Setup.
Following are the settings saved with this operation:
No. Setting
1 Time knob selection
You can select a system printer and print the acquired signal data.
You can choose the paper size and print format according your needs.
Enter the print preview window and allows you to see what the waveform will look like on the
screen before printing a hard copy.
- 55 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Save user interface custom settings, with memory capability. For example: FFT function is
turned on by default when the software is turned on.
- 56 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
All the channels of the virtual oscilloscope share the same time settings. The change of the
time knob position affects each channel at the same time. The value selected by the time knob
represents the time span represented by each horizontal division in the waveform drawing area.
In the same way the time span of each horizontal division is the time value indicated by the
position of the time knob. Therefore, the time axis coordinates changes according the different
positions of the time knob.
When the time axis is changed by the time knob, the device will select the most
appropriate sampling rate, according the current time span and the memory depth on the device.
- 57 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The time knob can be selected within a range from 5/50ns to 60s / division (depending on
the model). Since there are 10 divisions on the horizontal axis, the time range of the waveform
displayed on the screen vary from 50/500ns to 600s.
- 58 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
As shown in the figure above, our time wheel is set to the 1ms gear, as shown in the A
position. Then the abscissa axis collected in our screen has a span of 1ms and is shown in
position C. There are 10 horizontal grids in the waveform area, and the width of each grid is
1ms, such as position B. The entire screen width is 10ms.
We set the time gear through the time wheel, and this function can only be used when the
oscilloscope is in the acquisition state. If the oscilloscope is stopped, the time wheel is not
allowed to operate, because the operation of the time wheel requires a real hardware setup
action.
But when the oscilloscope is stopped, we can use the mouse or the slider to fine-tune the
time slot. As shown in position A in the figure below, we stopped the oscilloscope device, and
then we can pull the slider at position C to fine-tune the time gear. An equivalent and more
convenient way of this operation is to scroll the mouse wheel. Both methods will make small
changes to the current 1ms time slot, and the changed gear value will be displayed in position B.
As shown below:
- 59 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
- 60 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
When the selected time base is large, such as greater than 200ms/per division, the drawing
area will display the dividing line (green) for new and old data waveform refresh. If you do not
need the new and old data dividing line, you can cancel the selection. dividing line), which is
shown in the position A in the figure below.
In the figure above, the green vertical line is the old and new waveform data separation line.
Its left side, as indicated by the N position, is the newly collected data waveform, and its right
side, which is indicated by the O position, is the last time waveform. waveform data.
But when the time base gear is selected to switch between the large time gear and the
small time gear, the oscilloscope needs time to complete the hardware switching, and Loading..
will appear in the software to remind the user to wait patiently. This prompt will disappear
automatically after the switching is successful.
In LOTO's oscilloscope series products, some products support combined acquisition. The
principle of combined acquisition is to merge the ADC resources of A and B channels into A
channel, and the B channel will be disconnected and cannot be used. The original dual-channel
ADC will use interleaved sampling technology to double the original sampling rate on chA. For
example: OSCA02E series, OSC2002, OSCH02 series all support this function.
When operating the oscilloscope software, when the time base gear is adjusted to the us
level, button will appear on some products, indicating that the current time base gear
supports the synthetic acquisition function.
Click the button to turn on the synthetic collection function, and click the button
again to release the synthetic collection function.
- 61 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
For some high-frequency signals, such as 50Mhz, 80MHz signals, etc., synthetic
acquisition is not used. During normal acquisition, waveform signals will also be acquired due to
the limitation of channel bandwidth. The actual situation is that the sampling rate is not enough
to cause the acquisition to be aliased, and finally a complete waveform is presented, which is
actually wrong.
After the synthetic acquisition function is enabled, the sampling rate and bandwidth of
channel A are doubled. Channel B will lose the ability to collect.
Each channels of this virtual oscilloscope have its own voltage knobs. The value selected
by the voltage knob of one channel determines the voltage span of each vertical division in the
grid, referred to that channel waveform. The voltage grades differ depending on the model.
Changing the voltage scale with the voltage knob, the ordinate of the waveform in the
display area will change accordingly.
If the amplitude of the signal has exceeded the upper or lower boundaries of the drawing
area, it means that the selected voltage scale is too small, or the amplitude of the signal is
greater than the maximum value supported from the device.
In this case, a red exclamation mark warning sign will appear on the voltage knob of the
corresponding channel.
- 62 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
At this time, we need to increase the voltage gear selected by the voltage wheel. If the
maximum voltage gear has been selected, there is still an over-range warning, indicating that
the signal has exceeded the range, and the attenuation factor of the probe needs to be
increased up.
For the four-channel OSC984 model, in the corresponding host computer software, in
addition to the above-mentioned voltage gear knobs of the A and B channels, the voltage gears
of the other two channels are set at the positions shown in the figure below:
Using different probe divider affects the software's attenuation multiplier of the input signal,
in particularly the Y-axis of the drawing area, as well as the measured value.
- 63 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
As shown in the figure, the probe multiple set by the software needs to be consistent with
the attenuation multiple of the probe actually connected to the corresponding channel. If it is not
the multiple relationship in the software setting list, we need to use the custom probe function of
the LOTO oscilloscope, please refer to Chapter 13.
For the four-channel OSC984 model, in the corresponding host computer software, the
attenuation multiple settings of the C-channel and D-channel probes are next to their
corresponding gear settings. As shown below:
- 64 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Note
The probe selection switch and the software probe selection are mechanical and human
operations, and they should be kept on the same values. It is inevitable to forget to keep
their settings consistent at some time. This may result in errors on waveforms or measured
values.
Channel A and channel B can be enabled and disabled with the following switch buttons.
When the channel is enabled the button is respectively in cyan or yellow color, while when the
channel is disabled the button become gray.
The signal coupling refers to the way the input signal is connected to the oscilloscope.
When DC coupling (DC) is selected, the signal enters directly into the oscilloscope. When AC
coupling (AC) is selected, the oscilloscope internally isolates the DC component of the
measured signal by connecting a 0.1u capacitor in series to the input so that only the AC
component of the signal is acquired by the oscilloscope. You can select the Input Coupling (DC
or AC) for each channel. The default coupling mode is DC coupling.
- 65 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
For the four-channel OSC984 model, in the corresponding host computer software, these
channel settings are next to their corresponding gear settings. As shown below:
- 66 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The INVERT switch, the waveform of the corresponding channel is symmetrical about the X
axis (upside down), as shown in the figure, the same output signal, before and after turning on
the invert switch: Function disabled: The waveform is on the positive semi-axis above the
baseline.
After the function is turned on: The waveform is flipped over the negative half axis below the
baseline.
- 67 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
This function is deleted in the latest version of the software, and it can be perfectly
replaced by using the PC cache function. For the PC cache function, please refer to Section
8.2.5.
During the acquisition the signal statistics are computed automatically in real-time and
showed in the measurement area below the waveform display area. These values are displayed
only for the active channels.
Measurement Description
Max The highest voltage value of the current channel waveform on the
screen, in Volts
Min The lowest voltage value of the current channel waveform on the
screen, in Volts
P-P Peak to Peak Voltage (Max – Min), in Volts
Frequency The average frequency of the current channel waveform on the
screen, in Kilohertz (KHz).
Average The average voltage of the current channel waveform on the screen,
in Volts.
Period The signal period, in microseconds (us).
+/-Width Respectively the Width of a positive pulse at 50% crossings and the
Width of the negative pulse at 50% crossings, in microseconds (us).
Duty cycle Positive pulse width as percentage of period.
Rise Time The time it takes for the waveform to rise from the 5% position of the
peak value to the 95%, in microseconds (us).
Vrms The root mean square value of the current channel waveform, in
Volts.
Note
If the waveform of the current channel is displayed on the screen for less than one
- 68 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
period or for more than 50 periods, the Period measurement will be displayed as 0
because the measure cannot be performed, but this don’t mean that the signal period is
actually 0.
The automatic measurements described in the previous chapter will be shown below the
software, and sometimes we need to focus on a few measurements, and hope to be more
intuitive and convenient to observe them at any time, we can use this magnifier function. In the
Advanced tab on the right, we can choose which measurement to magnify. This function does
not conflict with the automatic measurement described in the previous chapter, they can coexist.
In the Advanced tab on the right side, we can choose to start the deep measurement
- 69 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
function, as shown by the green arrows in the figure. When the deep measurement function is
turned on, the data waveform displayed will automatically add numbered marker points near the
zero-crossing point.
The number of marker points depends on the waveform displayed on the current screen.
The numbered marker points are arranged in the table on the right .The time value of the
marker points displayed in red as a column, as shown by the long green arrow in the figure.
The next column displays the time interval between two adjacent numbered marker points,
as shown by the blue arrow. The time interval of each row is obtained by subtracting the time of
the current marker points from the previous one.
8.2.13 Annotations
The first item can be found in the right-click menu of waveform display area, and floating
data annotation can be added. Data annotation can be dragged to any position in the waveform
display area with the mouse, which will display the channel voltage value and the time value of
the annotation point. Data annotations can be added and dragged continuously to places of
interest, adding up to 10.
- 70 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
In the second row of the right-click menu, there are options to clear all data annotations, as
shown in the following figure:
8.2.14 Trigger
LOTO oscilloscopes are divided into two trigger types, software trigger and hardware
trigger. The OSC482 series is an entry-level model. For cost considerations, this system uses a
software trigger mechanism. Other high-end models are hardware trigger mechanisms, such as
(OSC802, OSCA02, OSC2002, OSCH02, OSC980). The principle of software triggering is to
- 71 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
collect a piece of data, and use software methods to find the position that meets the trigger
condition in the data, which requires CPU computing time, so it is not 100% guaranteed to find
the trigger position, because the trigger signal may not be in the current section data. The
principle of hardware triggering is that after the trigger conditions are set, these settings are
monitored by hardware logic, and the hardware logic circuit searches for the position in the
signal that meets the conditions, and then collects data after finding it, so for a single burst
signal, the hardware trigger can guarantee 100% triggering. But for signals that appear
repeatedly, the difference between software trigger and hardware trigger is not very different.
You can set the trigger in the trigger setting window and enable the trigger by selecting the
checkbox.
In addition, hardware triggers can cover more time slots than software triggers. The
software trigger is triggered in the middle of the buffer by default, and the hardware trigger
mechanism can choose where to trigger in the buffer.
If you purchase the OSC482 series products, as shown in the figure below, the time gear
marked with a black dot on the time gear wheel can be triggered, while other gears cannot be
triggered:
If you purchase other series of products, such as OSC802, OSCA02, OSC2002, OSCH02,
- 72 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
OSC980, almost all time slots smaller than 100MS can be triggered by hardware, as shown in
the following figure:
The triggering function is enabled only when the triggering module is checked, and
terminated only when the triggering module is checked.
Some large time gear trigger functions are disabled, and the check box is gray and
unavailable. The display time of a screen waveform in a large time gear is much longer than the
observation speed. For example, it takes 30 seconds to display a screen waveform. Therefore,
it is unnecessary to use the trigger function. You can completely pause the oscilloscope to
collect and observe and analyze the waveform when necessary.
The trigger function can be used only on channel A, so when you need to apply the trigger
to a signal, the signal should be connected to channel A. When the trigger is enabled, a triangle
with ‘T’ inside will appear on the right of the waveform display area to allow the trigger level
setting by dragging this blue triangle up and down with the mouse.
- 73 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The trigger level is combined with the selected rising or falling edge to define the trigger
condition. Considering as example the Normal Trigger with rising edge, the trigger condition is
met when the input voltage of channel A goes from low to high respect to the voltage value set
by the trigger level. When the trigger condition is met, the screen will display the entire
waveform before and after the trigger condition, keeping it until the trigger function is checked or
the next trigger occurs.
The Single trigger can easily capture accidental events, such as glitches with sudden
changes in waveform amplitude. You can set the trigger level just above the maximum value of
the normal signal, then click the Single button and wait for the trigger event to occur. When the
signal fluctuates beyond the trigger level, the oscilloscope automatically records the waveform
for a period before and after the trigger, and show on the software display area as shown in the
following figure:
When the signal fluctuation is frequent, if you want to continuously capture and show the
signal when matching the trigger criteria, then you need to use the Normal trigger with the
Normal button. During trigger scanning, the Stop button with appear for stopping the operation.
Some model device may not have this normal trigger feature.
- 74 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Take the rising edge trigger as an example, when the input signal voltage of channel A goes
from low to high and exceeds the voltage value set by the trigger level, the trigger condition is
met. At this time, the screen will display the entire waveform state before and after the trigger
condition occurs, and stay in this state without being updated until the trigger function is
checked, or the next allowed trigger occurs.
The trigger function of LOTO oscilloscope is divided into two ways: regular trigger and
single trigger.
The mechanism of the conventional trigger mode is that each time the waveform that meets
the trigger condition is displayed on the screen, the next time a new waveform that meets the
trigger condition arrives, the new waveform will be displayed, and the next tie the trigger
condition is met, the latest waveform will be displayed. Always replace the last triggered
waveform display with the latest triggered waveform display.
The mechanism of single trigger is that after clicking the "single trigger" button, only the
waveform that meets the trigger condition for the first time will be displayed on the screen, and
then the screen will not be updated until you click "single trigger" again to capture for the next
signal, or you click the "General Trigger" button to switch to a new trigger mechanism, or cancel
the trigger function, new waveforms will be collected and updated on the screen in real time.
- 75 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Some products support buffer trigger position selection and pulse width trigger. Later, in
order to add other resources, we removed the pulse width trigger function from the standard
products. Customers who need this function need to customize.
The OSC482 series does not support the function of setting the trigger position of the buffer.
Other hardware departure models support this function. As shown in the figure below, in the
software of models that support trigger position setting, you can drag the slider in the trigger
setting area to set the trigger horizontal position of the waveform in the entire buffer. By default,
we are at the middle 50% of the buffer. After triggering, we can see the waveform data of the
front and back half of the buffer. If you want to see more waveforms after triggering, you can set
the trigger position to a horizontal position near the front of the buffer.
Trigger Position
In the new version of hardware, the trigger sensitivity function is added to the hardware
trigger mechanism. Some hardware versions do not support this function:
Select L (low sensitivity) in low gear and H (high sensitivity) in high gear The software
defaults to low sensitivity trigger. This function is for the convenience of balancing the precise
trigger position and anti noise interference.
If the sensitivity is too high, for some signals with dense small burrs, the up and down jitter
of the burrs at the trigger level will cause false triggering. I choose the rising edge trigger. Maybe
when the falling signal reaches the trigger level, a small burr just rises. This is because the
trigger is too sensitive, and the falling edge signal may be captured as the rising edge.
Therefore, when we observe that the edge trigger will often capture the trigger waveform of the
- 76 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
reverse edge, it should be because the sensitivity is too high. At this time, we should set the
sensitivity to low L.
On the other hand, when the sampling rate is very high to collect high-speed signals, the
sampling points generally appear sparse. At this time, if the trigger sensitivity is not high, the
trigger may shift several sampling points, or even dozens of sampling points. Then the
waveform we see is not at the set trigger position, or even far away from the feeling. At this time,
we should choose high sensitivity, as shown in the following figure:
Triggered variable edge selection, generally there are two kinds of rising edge and falling
edge. The version that supports hardware triggering adds double edge triggering, that is, as
long as the jump edge crosses the trigger level, both the rising edge and the falling edge are
considered to meet the trigger conditions.
For a more detailed description of the relevant trigger sensitivity, please refer to Appendix V:
LOTO Oscilloscope Trigger Sensitivity.
When the OSCA02, OSC2002 and OSCH02 series need to trigger the third signal and
monitor the waveform of the other two analog signals at the trigger time, the external trigger
module E01 can be purchased. The external trigger module needs to be plugged into the side
expansion interface DE2 of the LOTO oscilloscope.
- 77 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Parameter ET01
Channel Number 1
Input Resistance 1MΩ
Output interface DE-15
Output signal 3.3V ttl
0.2V~1.7V Continuously adjustable (0V~5V range, other ranges
Trigger level
are expanded by multiples)
Trigger edge Rising edge/falling edge selectable
bandwidth 10K Hz
No
X1 Input range: 0V~5V
attenuation
2 times
X2 Input range: 0V~10V
attenuation
Input range(4 gears)
5 times
X5 Input range: 0V~25V
attenuation
X1 10 times
Input range: 0V~50V
0 attenuation
The external trigger only provides a trigger source different from chA in hardware, and the
use method is the same in software. We only need to turn on the external trigger function in the
following positions. As shown in the figure below:
- 78 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The waveforms persistence effect superimposes on the screen the data acquired in
successive times and allow the observation of the spatial concentration of the waveform energy.
When the Glow checkbox is set, the oscilloscope continues to acquire and display new data, but
do not erase the previous collected data. The waveform parts with higher occurrence will have
higher brightness; parts with fewer occurrences will appear less bright. The waveforms
persistence superimposition has a significance as a statistical measurement, since it can show
in intuitive way the distribution of noise on the time and space. This makes it suitable for the
preliminary analysis of random noise.
- 79 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Normal mode:
This is the most common sampling mode. The oscilloscope store and display the samples
according to the sampling rate.
The oscilloscope samples always at the maximum sampling rate to find the maximum and
minimum values within the time interval for each sample, and then it displays the maximum and
minimum values on the waveform curve for each sample position.
Peak Detect mode is best used for detecting glitches, viewing very narrow pulses or
measures the amplitude range of the noise. For example, if the actual sampling interval of the
oscilloscope is 2ms for one sample, the oscilloscope internally will use the maximum sampling
rate and will collect / show 2 values every 2ms. These 2 values are the maximum and minimum
values among n data points sampled within the interval of 2ms.
High-Resolution mode:
- 80 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Note
For displays with low resolution since the display area is too small, the software will move the
mode settings window to the Advanced tab.
- 81 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The High-Resolution sampling mode uses oversampling to improve vertical resolution from
8 Bit to about 13 Bit.
The input is sampled at a rate higher than the minimum required Nyquist sampling rate, fs.
For example, when using an N-bit ADC without oversampling, an input signal of 100 Hz is
sampled at 200 Hz (2 × 100 Hz) to get the digital output with the native ENOB of the ADC.
When oversampling with a factor of k = 16, the same 100-Hz input signal is sampled at 3200 Hz
(k × 2 × 100 Hz). The samples obtained by oversampling are low-pass filtered and decimated
using a digital filter to achieve a reduction of the quantization noise. The signal at the frequency
band of interest is not affected by the filter, and the result is an improved SNR. The improved
SNR results in a higher ENOB performance.
The group averaging in the high resolution sampling mode mentioned above is also
essentially a digital filtering. The higher the sample extraction factor, the more the resolution is
increased. When the extraction multiple is 16, it can be improved from 8 bit to 10 bit. Working in
HiRes mode, the user cannot directly select how many bit resolutions to enhance, which
dynamically changes with the time scale of the oscilloscope. The user does not know how much
resolution is currently enhanced and how much bandwidth is dropped.
The steeper edges of the signal are easier to be under sampled in HiRes mode. For
example, the signal edge originally has 16 sampling points. In order to improve a 2-bit
resolution, it is necessary to average every 16 points. As a result, only one sample is left on the
edge. Group averaging also significantly reduces the oscilloscope waveform update rate.
Group averaging can conditionally increase resolution without improving the accuracy of
the oscilloscope. For example, a DC gain accuracy with a hardware 12-bit ADC oscilloscope
can reach 0.5% of the ground. Using an oversampling and linear noise reduction technique to
boost the oscilloscope to 12 bit resolution, the DC gain accuracy is still the level of an 8-bit ADC
oscilloscope: ±2%.
- 82 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The samples interpolation consists in filling the gaps between the ADC samples by
- 83 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
On the advanced page or the button in the lower right corner of the display page , you
can choose among Linear or Sine interpolation algorithm. The selected algorithm will use the
time slots (0.2us and 50ns) in the black sector on the time knob.
- 84 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The green dots in the above figure are the actual ADC samples, the red curve is the
waveform curve generated by the linear interpolation algorithm; the blue curve is the curve
generated by the sine interpolation 4x.
In the waveform display area, click the right mouse button and the following menu will
appear.
Restore the A channel baseline to the default position. (The default position is the middle
position of the Y axis of the waveform display area.)
- 85 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Only several model devices support this feature. If you cannot find this feature in your
software, it is not available for your device model. Usually, we may find this decoding feature in
a “Decode” tab panel on the right.
After using the trigger capture communication signal, the oscilloscope software enters the
decode mode by selecting the corresponding communication protocol, such as “UART” or “IIC”.
Clicking on the protocol menu item, “On” appears in front of the menu item, indicating that
the communication protocol is used for decoding. Clicking “OFF” will terminate the decoding.
After the communication protocol is selected, click the right mouse button in the waveform
display area. The “Setting the decode parameters” option in the right-click menu will be
activated and changed to the usable state. Click “Setting the decode parameters” to pop up the
protocol parameter setting window.
Usually, the software will display this setting window actively when the selected protocol is
selected for the first time to remind the user to complete the parameter setting. Then, during the
use, you can also use the options in the right mouse button menu to call out this window for
parameter adjustment.
- 86 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
For different protocols, the setting items appearing in the protocol parameter setting window
are different;
"Decode Options" panel item, select the basic parameters of the protocol.
The "Result" panel item displays all the decoding results of the entire buffer.
The order in which each frame of data is sent is from the high bit or from the low bit.
For example: 0x01, 8-binary digits: 0000 0001(The leftmost 0 is the high position, and the
rightmost 1 is the low position)
data bits invert LSB, the result is: 1000 0000(send from low position)
data bits invert MSB, the result is: 0000 0001(send from high position)
- 87 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
When there is a communication data signal, whether the data signal is high level or low level.
(if IO port defaults to low level when no data is sent, the data is high level when sending data;
then you should choose "enable voltage high"). For example: RS232.
(if IO port defaults to high level when no data is sent, the data is low level when sending data;
then you should choose "enable voltage low"). For example: TTL.
8.2.20.5 Examples
The following uses several practical examples to illustrate the decoding steps and
the results displayed.
- 88 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Procedure Description
BaudRate is 9600, which is set according to the frequency when the hardware
sends data in this example.
Calculated by: 1.0 / 9600 ≈ 0.00010416666666666667. More accurate
1
calculations are given to the oscilloscope software. We estimate that the final
value is approximately: 104.17us. (How much time each bit takes to measure
the length in seconds)
DataBits is 8bit, and the value is obtained according to the hardware circuit
2
parameters in this example.
Parity is Odd. (If not, it can be set to "None" and the oscilloscope software will
3
ignore it).
StopBits is One, and the value is obtained according to the hardware circuit
4
parameters in this example.
5 Start bit + DataBits + Parity + StopBits = 1 + 8 + 1 + 1 = 11 bit.
6 11bit * 104.17us ≈ 1145.87us. (Time of each frame of data )
In the oscilloscope software Time gear, the X axis of the 1ms gear position takes
1ms per grid.
The X-axis grid is about 1145us, which is a relatively appropriate time gear. The
7
periodic waveform density interval in the screen is appropriate, which can read
the decoded data well, and the success rate of oscilloscope software in
analyzing data will be greatly increased.
8 For transient discontinuous signals, it is recommended to use the trigger function
- 89 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
to capture;
Use the software's pause and left and right move shortcuts (Ctrl + <-) / (Ctrl + —
>) to help analyze the signal.
In this example, when RS232 sends a signal, each bit will be “bitwise inverted”,
9
so select: “bit invert Yes”
In this example, each frame of data is sent from the high bit of 8 bits, so choose
10
"databits invert No".
In this example, when no data is sent, the IO port defaults to low level and the
11
data sent is high level, so choose "enable voltage high".
The data sent is: 0xFF 0xEF 0x01 0x02 0x03. Finally, according to the display
12
results, the sending is correct.
What needs to be emphasized is Parity. If the check bit is found to be
inconsistent with the check bit after analyzing the received waveform, it will
13
automatically assume that this is a frame of error data and it will be discarded
without the block diagram showing the parsed result (but with the waveform).
Procedure Description
1 Probe of channel chA is connected to the SCL circuit.
2 Probe of channel chB is connected to the SDA circuit.
3 Select the appropriate oscilloscope software Time gear.
For transient discontinuous signals, it is recommended to use the trigger function
to capture;
4
Use the software's pause and left and right move shortcuts (Ctrl + <-) / (Ctrl + —
>) to help analyze the signal.
5 In this example of IIC communication, each bit has not been "bit-reversed", so
- 90 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Automatically identifying the communication rate of the measured CAN signal during
decoding; Supports any one or more of the five frame types of the CAN protocol for
simultaneous decoding;
- 91 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Automatic recognition of CAN_H line and CAN_L line is supported during decoding;
- 92 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
https://www.youtube.com/watch?v=1vRoUId8orU&t=783s
This function allows users to select picture files from PC and import them into waveform
plotting area as the background of acquisition waveform drawing. Used for real-time acquisition
of waveforms and imported pictures for reference or comparison. The reference comparison
function is shown in the tab page on the right side of the software as follows :
Select the "browse" button, import the picture file, and select the "on" button, the picture will
be displayed in the drawing area as the background. The imported image can be moved left and
right or zoomed in and edited, and changed to grayscale or transparency.
- 93 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
In the lower right corner of the display interface, click the red new update button to
see the latest version of all series software of Loto oscilloscope, and the user can
download the latest version of corresponding software.
At the bottom of the lower right corner of the software interface, click the blue button
of the waveform refresh rate to adjust the refresh rate of the oscilloscope measured
waveform by moving the cursor left or right. A higher refresh rate can be selected on the
display with better screen display effect to better display the waveform.
The actual display speed is related to the machine performance.
- 94 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
In the lower right part of the software display interface, click the button can
realize the statistical function of any variable of the signal, for example: measure the
change of the maximum value and peak value variable of a signal as shown in the
figure below.
Click the fault button in the lower right corner of the signal variable statistics page,
and a fault linkage dialog box will pop up. The user can freely set the fault range and
select the alarm mode, such as IO warning, software alarm and RS232 alarm. It is very
convenient for us to measure whether the signal is normal or not. The dialog page is
shown in the figure below:
- 95 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The user clicks the reset button in the lower right corner of the software interface,
which will restore the oscilloscope to the factory settings. When the oscilloscope
crashes or other unexpected circumstances occur, the reset button can be used to reset
the oscilloscope, and the reset button is shown in the figure below:
- 96 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
When this button is clicked, a program for measuring the frequency response will be
started. If there is a customer whose screen resolution is insufficient to display the
button at this location, the customer can find the independent program in the
oscilloscope software directory and start it manually. The program running interface is
shown below:
- 97 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The frequency response curve function is generally used together with the frequency
sweep. The frequency sweep function of the signal generator gradually changes the
signal frequency from the initial value to the set value. Coordinate, at the same time,
you can select the amplitude of the detection channel or the amplitude ratio of the two
channels as the ordinate, so as to draw the frequency response curve. It should be
noted that this function is used in conjunction with the oscilloscope function. You can
pause at any time during the mapping process to allow time for the oscilloscope to
switch gears or do other preparations.
In the signal generator sweep function, we have introduced the "Freq response auto"
function. During the scanning process after checking, the oscilloscope will cooperate
with the signal generator to automatically switch the appropriate time base gear.
Note: At the beginning of the scan, we need to click the "Clear" button several times
to clean up the unnecessary parts that are not expected.
After the sweep is completed, click the "Fit" button to fit the scanned data points for
better observation.
When using the frequency response curve function, you need to turn on the
[Measurement Value Sharing] function:
- 98 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
After the oscilloscope is successfully powered on and connected to the software. The 6-
hole DE1-15 of the oscilloscope defaults to output a 1Khz square wave of about 1.5V, which is
convenient for users to perform probe compensation and calibration.
- 99 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Click the PWM/Pulse button, and the PWM setting window will pop up. In the window, you
can adjust the standard square wave to a PWM wave, or you can modify the PWM wave to
send a single pulse.
- 100 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Click the FIR button to open the FIR digital filter window. This digital filter is based on FFT.
In the pop-up window, you can choose to set Band-pass or Band-stop, or a combination of
both. After setting the filter frequency range, click the set button to take effect, and the waveform
display area of the oscilloscope will display the filtered waveform.
Click the Reset button to restore the default state. Close the window to close the FIR digital
filter function.
- 101 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Among LOTO's oscilloscope products, some products support eye diagram analysis. It is
used to analyze the influence of inter-symbol interference and noise of signal waveform on
system performance, and observe the transmission performance of the system.
Click the button to turn on the eye diagram analysis function, and click the
button again to turn off the eye diagram function.
According to the performance difference of different customers' computers, the final visual
effect of the eye diagram will be different. As the opening time continues to increase, the effect
will get better and better. The color gamut level of the eye diagram:
increasing from left to right.
- 102 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
This function can export historical video files or historical PC cache files into voltage format
Txt or Csv file sets. The user can use this function to convert the data collected by the
oscilloscope to other data formats, which is convenient for reading and analyzing with other
data analysis tools .
The function open button is located at the bottom right of the multi-frame display window of
the PC cache function, as shown in the figure:
When the button is clicked, the transcoding export function window will start
and display, as shown in the figure:
Click the button to browse the files saved by the historical oscilloscope. After
completing the corresponding settings, click the button to complete the transcoding
export.
- 103 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
During the running process of the oscilloscope software, when the user is interested in a
certain measurement value at a certain moment, the measurement value value at the current
moment can be recorded, and the software will draw and display multiple measurement value
data points; when necessary At that time, it is possible to perform (exponential, logarithmic, 3rd
degree polynomial, 2nd degree polynomial, linear) algorithm fitting on the curve set composed
of data points, and calculate a curve with a certain meaning. It is also possible to manually add
and delete data point sets.
During the running of the oscilloscope software, click the "Single Capture" button to add
data value points to the single acquisition drawing interface.
- 104 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Click the button , the data collected by the oscilloscope will be saved in the user's
computer in text format (.txt/.CSV), and the saving path can be customized by the user.
As shown in the figure above, click the save button to save the waveform data of part A
displayed in the drawing area, that is, part of the data in the cache, and the waveform data of
part B in the lower right corner of the figure, to the computer in the form of a text file superior.
Click the save button, and the following dialog box will pop up by default for the user to
choose to save as a TXT file or a CSV file. The former can be opened with the Notepad or
notebook that comes with the computer, or it can be opened with a third-party text editor. The
latter can be directly opened into an excel table style.
In the figure above, the slider is used to select data points to discard. The default is 1,
which means that every data point is saved. If it is 100, it means that when saving, one data
point is randomly selected for every 100 data points to be saved. This reduces the file size, but
at the same time equates to a several times lower sample rate.
- 105 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
We can also make further settings for file saving, as shown in the figure below, the
advanced settings for file saving can be found in the "Advanced" tab:
As shown in the figure above, A is our default save path, which is in the same directory of
the oscilloscope PC software, and B is a customer’s own save path that can be specified.
Choose one of these two paths, and at C When "Save without popup every time" is
selected, the oscilloscope software will store the waveform data displayed on the screen after
the customer clicks the save button in this alternative path.
- 106 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
When C is selected as "Every time you save, a dialog box for selecting a path will pop up",
the settings in A and B are invalid, and the software will save according to the path selected by
the customer.
After the data is saved, whether a prompt dialog box will pop up is set at D. After all the
settings are completed, click the setting at E to save the setting.
The upper computer software of the LOTO oscilloscope computer can set some colors
related to the waveform display. These colors can be saved after setting, and can also be reset
to factory default. As shown below:
The opening position of the color setting is at 1 in the figure, and clicking it will open the
setting dialog box shown in 2. We can set the corresponding color, and then click the 5 next to it
to make the set color take effect immediately. Finally, the color effect has reached the
expectation.
Click the "Save" button at 3 to save your color settings to memory and use them by default
in future software.
We set a "reset" button at 4, and clicking it will restore the color configuration we set.
- 107 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Clicking the button pointed by the arrow in the above figure will open the data sharing page,
as shown in the figure below:
After opening the Data sharing interface, we check the measured value option, as shown in
the figure below:
- 108 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
At this time, the waveform data collected by the oscilloscope in real time locks the various
measured values calculated (automatic measurement introduced in 8.2.10).
If we find that the data does not change, it may be because we have not enabled the data
in the oscilloscope software by default. Output switch, we are in the "Advanced" tab, make sure
to enable the following settings:
These measured values are read from the host computer software of the LOTO
oscilloscope in real time. Therefore, it is necessary to ensure that the LOTO oscilloscope
software is running normally so that there will be real-time data presentation.
We can set the refresh rate of the data, which can also be understood as how often we go
to the oscilloscope software to get the latest measurement data and update it.
We can set the serial port to send these measured values. Generally, we can connect a
serial port on the PC through USB to serial port, and then set the serial port to communicate
with external third-party devices, as shown in the figure below:
- 109 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Which measurement values we want to send can be selected in the tree directory on
the left. We preset different command codes for each measurement value, so that after
receiving the data frame, the third-party device can parse out the command code to distinguish
different Measurement category:
Address Command byte count Data high byte (1 Data low byte (1 CRC CRC
code (1 code (1 (1 byte) byte) byte) check check low
byte) byte high byte byte
The high byte and low byte of the data form a 16-bit data, which is organized as follows: the
lowest bit (bit0) represents the symbol, 0 means that the number is positive, and 1 means that
the number is negative number. The penultimate 3rd bit (bit2) and penultimate 2nd bit (bit1)
indicate how many digits of the entire data are decimals. for example:
- 110 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
1 0 1 1 0 0 1 0 1 1 1 0 0 1 0 1
The binary data 1011001011100 converted to decimal is: 5724, the lowest bit is 1 to
represent a negative number, so it is -5724, the third and second last digits are 10, representing
2 decimal places, so it is -57.24.
If the data sent by the customer requires a special command code or a special data format,
you can set a custom frame format for sending, as follows:
Since the oscilloscope automatically selects the sampling rate and storage length at a
selected time slot, these parameters are fixed, and it is very convenient for the user to view the
waveform in some scenarios. Users don't even have to care about or understand concepts and
theories like sampling rate and memory depth.
But this also brings another inconvenient application problem. In some applications, the
user has already clearly known the situation of the signal to be collected. According to the plan,
the user needs to set the sampling rate, and then an exact long-term data. At this time, the
original oscilloscope mode is not very convenient. LOTO's acquisition card mode is to
supplement the deficiency of the oscilloscope mode and provide a new data acquisition method.
Oscilloscope mode and acquisition card mode have different application scenarios and
different operations. They will not be used at the same time. They are either oscilloscope mode
or acquisition card mode. If the user selects the "OSC" tab, it is using the oscilloscope mode. If
you choose If you open the capture card tab, you are using the capture card mode.
- 111 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The capture card mode is shown in the figure below, we can’t see the time gear knob
(roulette), but we can see the capture card icon:
In the setting interface of capture card mode, we can choose the required sampling rate
and buffer size. The sampling rate can have several options, and it is not completely possible to
enter the settings by yourself. The important thing is that there are many buffer size options, and
the maximum buffer size can be 248M bytes.
- 112 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
After selecting the sampling rate and buffer size, we need to click the "Settings" button to
make the sampling rate and buffer size take effect, as shown in the figure below:
- 113 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Next, we can use the channel start/stop button to control the acquisition of data waveforms
as in the oscilloscope mode, as shown in the figure below:
Other operations are similar to the oscilloscope mode, the following points need to be noted.
In the acquisition card mode, there is no trigger function. After clicking the start button
above, the oscilloscope collects the data of the specified buffer size according to the set
sampling rate, and the acquisition is completed. Whether the oscilloscope automatically stops
after one acquisition or automatically repeats multiple acquisitions depends on our settings, as
shown in the figure below:
In the capture card mode, we usually set a relatively large buffer, so the time required to
complete the entire screen capture is generally relatively long, and the entire time can be
calculated, and the capture time = half of the buffer size/sampling rate. Under the setting of long
time, the acquisition card mode needs to wait until the data buffer set by the user is completely
collected before displaying the waveform. The user needs to wait patiently, which is also
different from the oscilloscope mode.
In capture card mode, we will open up an area in the computer memory for buffer use. If the
buffer setting is relatively large, the following reminder will pop up. No need to contact us. This
is just a reminder for users with low computer performance. It may take up relatively large
memory resources. If you can use a small amount of buffer, you don’t need a large one, so as
not to make the computer freeze.
- 114 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
If the collected data is to be saved as a text file, then the acquisition card mode can use the
same text save button as the oscilloscope mode to save the collected data as a TXT or CSV
format file. As shown below:
The maximum file size of about 480M can be saved, as shown in the figure below after
opening:
- 115 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
We can also save the collected data as the PC cache file of the oscilloscope. For the use of
the PC cache file, please refer to chapter 8.2.5.1. After collecting the required data, we can click
the "Export" button in the figure below. Export the data as a PC cache file of the oscilloscope.
We can also save the collected data as the PC cache file of the oscilloscope. For the use of
the PC cache file, please refer to chapter 8.2.5.1. After collecting the required data, we can click
the "Export" button in the figure below. Export the data as a PC cache file of the oscilloscope.
We can select the "Browse" button in the PC cache setting area in the oscilloscope mode,
select the PC cache file stored above, and then import it into the PC cache. As shown below:
- 116 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
This chapter introduces the basic concepts and usage of the paperless recorder, a software
functional extension of the virtual oscilloscope hardware.
- 117 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The Recorder software has an interface consistent with the oscilloscope software, so to be
familiar to the user.
Equipment monitoring
After the software interface is opened, the OSCxxx device status will be monitored in real
time. When the hardware device is connected to the USB port of the PC, the software interface
will display a blue background and a message.
If the hardware get disconnected or the software have no access to the device, then the
display background will become yellow, showing a text alert.
The sampling rate of the paperless recorder is not high. It is suitable for scenes with long
and slow electrical signal change cycles, such as the detection of a slow discharge curve during
the battery discharge process for 24 hours a day. For scenes where the signal changes quickly,
it is recommended to use oscilloscope software for inspection and observation.
- 118 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Note
The paperless recorder software has exclusive access to the OSCxxx device. Therefore, the
device can be used just with one software at a time.
This panel allow to set the recording duration; it can be set in the range 1 minute ~ 3 days.
This panel allow to set the sampling interval from 1 second to 1 hour.
Note
The sampling interval cannot be greater than the total recording duration, otherwise it is
meaningless.
After the driver has been successfully installed, if the oscilloscope is connected when the
software is opened, the software automatically turns on the device. If the device is connected
before opening the software, it is required to start the device manually by clicking on the Start
Device button in the lower right corner of the software.
- 119 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The acquisition can be stopped by clicking the Stop button. The Start and Stop button are
actually the same button that change according the status of the oscilloscope. When the USB
cable is suddenly unplugged during the operations, the software will automatically stop.
When the device is not used it is recommended to close the software and then to unplug
the device's USB cable.
Signal generator S02 is a single-channel signal source module, which belongs to the
expandable module of LOTO oscilloscope. It is a signal source module made around the DDS
chip to facilitate customers to quickly generate periodic regular waveforms.
Some types of oscilloscope products support the extended functions of the signal
generator module, and you can check the hardware data sheet of the product. If the purchased
product does not contain or support the signal generator module, this part of the software
function will not produce the expected effect.
At present, there are many types of oscilloscope products, and the plugging positions of
the extended signal generator modules of different types of oscilloscopes are different.
The GPIO function of the OSC482 series is extended at the DE1 port, and the GPIO
function of other series of LOTO oscilloscopes is extended at the DE2 port. The specific
connection location is as shown in the figure below:
- 120 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Some devices have the ability to be integrated with signal generator moudle. So if you
receive a Signal generator module, you may find the corresponding feature in the host devices’
software:
- 121 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The two-channel standard product of LOTO oscilloscope can work with the extended S02
signal source at the same time, so that one analog signal can be output and two input analog
signals can be monitored at the same time. On the software, it can also be displayed and set at
the same time, as shown in the figure below:
- 122 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
As shown in the figure above, the software setting area corresponding to the signal source
module is in the tab page "Extended". Click the "Extend" button of the software, and the
extended function interface will cover the "Settings" function interface of the oscilloscope. If the
signal generator is not selected in the "Extended" interface, the signal generator module will not
start and will not work.
- 123 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The waveform type (sine, triangle, square wave) and waveform frequency of the signal
generator module S02 are set by software, and the amplitude and offset of the waveform are
controlled by the hardware knob on the module. Since the LOTO oscilloscope can be used
simultaneously with the signal source module, the amplitude and offset of the signal can be
observed through self-sending and self-receiving.
Click the "Signal Generator Module" button (marked by the red arrow) to select the signal
generator, and the signal generator module will start. Then you can click to select the desired
waveform button, the icon on the button will change from green to orange, and the signal
generator will generate a waveform to provide a stable signal source.
- 124 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
After getting the signal generator module, in order to quickly check whether the signal
generator module is normal. You can use the probe of the oscilloscope to measure the output
interface of the signal generator module. If the waveform state, frequency, and amplitude of the
signal in the display area of the oscilloscope are consistent with the settings of the signal
generator module, it means that the module is functioning normally. (The figure below is the
waveform measured by the oscilloscope and set to 1KHZ)
The signal generator module can set the frequency of the waveform according to the needs,
the frequency setting range:
- 125 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The frequency can be directly input in the input box, and the unit is HZ. You can also adjust
the triangle button attached to the input box. Due to the different ranges of the frequency value
in the display box, the step value is also different each time the button is clicked. For details,
refer to the following table:
Frequency Click the triangle button (up or down) to change the frequency
step by step
1HZ-99HZ 1HZ
100HZ-990HZ 10HZ
1000HZ-9900HZ 100HZ
≥10000HZ 1000HZ
The signal generator module adds a software-controlled shutdown function, which may not
be supported by older versions of the oscilloscope hardware. Click the OFF button to cut off the
output of the signal waveform, and click the ON button to enable the output of the signal:
- 126 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
It is also possible to set the step frequency of the sweep, that is to say, how much the
frequency should be increased based on the current output signal frequency each time the
sweep changes the frequency, and the unit is also HZ.
The time interval of the frequency sweep refers to how long the interval is to increase the
frequency of the current signal by one step during the frequency sweep process. The unit of this
time interval is milliseconds, as shown in the figure below:
After the setting is complete, you can click the frequency sweep button to sweep the
frequency:
During the sweeping process, you can pause/resume or stop at any time:
- 127 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The default frequency sweep is one-way after it starts, that is to say, after sweeping from
the initial frequency to the final frequency, the frequency sweep stops, and the frequency is
fixed at the final frequency. If you need to perform the entire frequency sweep process cyclically,
you can check the "Circular frequency sweep option".
As shown in the figure below, the frequency response curve automatic scanning function is
used together with the "frequency response curve drawing function", please read the description
in chapter 8.2.27 Frequency Response Curve Function.
The starting value of the fully automatic frequency range supported by different models of
this function is different, the OSC482 series is 25Hz, and other models are 2Hz.
When using the automatic frequency sweep function of the frequency response curve, the
[measurement value sharing] function needs to be enabled. As shown below:
- 128 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
For the introduction of using the signal generator to sweep the frequency response curve,
please refer to "Appendix 4: LOTO Oscilloscope Detection Frequency Response Characteristic
Curve" and chapter 8.2.6 Frequency Response Curve Drawing Function of this document.
Some devices have this GPIO control function, allow us to implement several IO pins TTL
voltage output control or input state detect.These GPIO follow 3.3V TTL level logic.
- 129 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
12.2 DO output
Set the output of the IO to high level respectively, and observe the signal status of the IO
port with an oscilloscope:
- 130 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
- 131 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
12.3 DI input
Click the "Input" button to disable the "Output" function, and click the "Refresh" button.
When there is no signal connected, the display box defaults to the input high level, and when
the signal is connected, click Refresh again, and the current input status is displayed in the
display box.
Note
When "IO DIY" is selected, the signal generator module is prohibited to use. Similarly, select
the signal generating module, IO DIY is also prohibited.
- 132 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
13 Custom probes
13.1 The role of the custom probe function
Usually we use corresponding sensors to convert various measured physical quantities into
voltages for measurement. Therefore, each conversion requires a corresponding relationship
between the measured physical quantity and the converted voltage value.
Generally, this relationship will be made into a linear relationship, which is convenient for
conversion and calculation. Therefore, it is very useful to have a custom probe function to set
the linear correspondence between voltage and physical quantity.
Most of the standard LOTO oscilloscopes are two-channel, and each channel has a custom
probe function.
When customers use other probes or sensors other than the standard voltage probe to
measure physical quantity waveform data, they need to use this probe or enable the custom
probe function on the corresponding oscilloscope channel of the sensor.
This custom probe function is in the setting area of each channel. For example, the
custom probe settings of channel A of the oscilloscope are as follows:
- 133 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
As mentioned above, after we enable the custom probe function of the corresponding
channel, there will be two sets of values that we need to fill in.
Two values at value 1 and two values at value 2. They are actually two sets of
points with a linear relationship between the physical quantity and the switching voltage. Two
sets of points define a straight line.
To give a simple example, if we are using a current probe now, when the current is 0A, the
voltage measured by our current probe is 0.11V, and when the current is 1A, the voltage we
measure is 0.68V, such Two sets of data are enough to determine the correspondence between
the voltage output by the current probe and the current it actually measures, that is the following
settings:
Note that the unit is written in the unit of the actual measured physical quantity. After setting
in this way, click OK, the data and waveform measured by channel A of the entire oscilloscope
will be the actual current data waveform of the current probe, not the data and waveform of the
voltage converted by the current probe.
- 134 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Of course, the same principle and setting method are used for probes and sensors of other
physical quantities. As long as we get a probe or a sensor and know what physical quantity and
what ratio it converts into voltage, or know how many physical quantities are converted into two
sets of voltages, we can directly test the data and waveform of the physical quantity on the
LOTO oscilloscope up.
As mentioned in the previous section, we can already set parameters for any sensor or
probe. But the above approach can only be effective at that time.
After the software is closed, the parameter setting of this probe/sensor will disappear, and if
it is used again, it needs to be set again. It will be a little troublesome. So we need a function to
save the parameters of a fixed probe/sensor.
The PC-side software of the oscilloscope supports users to save customization. You can
create an XML file in the specified node format under the Probes folder in the application root
directory of the virtual oscilloscope software to complete the addition of custom probes.
After adding, as shown in the figure below, directly select the probe you saved and added
in the drop-down list, without having to manually set the parameters again.
Let's see how to make a custom probe for your own oscilloscope.
The files of custom probes are all placed in the <Probes> folder under the directory of the
oscilloscope host computer software on the computer, so we find this folder and enter:
- 135 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The first is the file name, and the file name will appear in the drop-down selection list of the
channel custom probe, so this name is easy to remember and it is enough to identify, preferably
the model of the probe, try to avoid using Chinese, so as not to cause garbled file paths in some
systems. But do not change the suffix of the file, it must be .xml.
After modifying the file name, which is the name of our probe, you can use the system’s
notepad or wordpad to open the file and modify the content of the file.
Only a few places need to be modified, and other places do not need to be changed.
- 136 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
- 137 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
</Probe>
After the XML file is added and saved, close the oscilloscope software and run the software
again, and the newly added custom probe will appear in the custom probe drop-down menu.
After understanding the saving of custom probes introduced in the previous section, we can
save our own probes. In the same way, LOTO officially makes corresponding XML files for its
various current probes, high-voltage differential probes, etc., based on the function of custom
probes to save XML files.
In this way, when customers use various probes and sensors of LOTO, they can directly
find the corresponding model from the drop-down list of custom probes and use them directly.
Don't need to care about conversion parameters anymore.
Customers can also send us the parameters of their own probes, and we can make a file as
long as it is stored in the specified directory, and it can directly become an officially supported
custom probe.
If the customer feels that the parameters of the previous custom probe are not accurate
enough and needs to be adjusted, the method in the previous section can also be used to
directly modify the XML file to correct the probe/sensor parameters.
- 138 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
At present, all models of LOTO oscilloscope products support the expansion function of the
isolated differential expansion module.
For specific models and parameters, please refer to the hardware data sheet of the product.
If the purchased product does not include a high-voltage isolation differential module, the
software functions in this part will not be effective.
Let's take IDM02 (a high-voltage isolated differential module with a bandwidth of 100K) as
an example. For example, we connected the oscilloscope and IDM02 to test the motor
waveform as shown in the figure below:
We will calibrate the zero position of the module when making the module. Customers
generally do not need to deal with it, but after long-term use, or with seasonal temperature
changes or component aging, zero position offset may occur. Therefore, before signal
measurement, it is sometimes necessary to calibrate the zero position of the differential module.
In order to avoid measurement value errors in the measurement results.
The step of zero position calibration is, after connecting the isolated differential module,
observe that the position of the zero noise line of channel chB does not coincide with the
position of the reference line of channel chB, and then click the zeroing button to clear. (Note:
When calibrating the zero position, the input BNC connector of the module is required, without
any external electrical signal)
- 139 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
After calibrating the zero voltage position, it can be used normally. If the user changes any
of the following settings for channel chB, the zero position may needs to be recalibrated:
2. The coupling mode of the channel chB has changed (AC / DC);
Note: If you are using the channel chB coupling method for AC coupling after isolating the
differential module from the connection, you don’t need to calibrate the zero position.
The range of the isolated differential module is parasitic on the channel chB and has its
own range. Take one of the isolated differential modules IDM02 as an example:
- 140 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
- 141 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Note:
The differential module gear position switch and the software differential module gear position are
mechanical operation and software operation, and the two cooperate with each other. It is inevitable to forget to
keep their settings consistent at some time. This causes waveforms or measured values, resulting in ploidy errors.
This requires more seriousness and care.
Some models of products support the extended functions of the logic analyzer module, and
you can check the hardware data sheet of the product. If the purchased product does not
contain or does not support the logic analyzer module, this part of the software function will not
be able to realize the logic analyzer function.
The logic analyzer and the channel B of the oscilloscope are mutually exclusive in
hardware, choose one of the two. However, the logic analyzer can be used and displayed
simultaneously with channel A of the oscilloscope, as shown in the figure below:
- 142 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The oscilloscope and logic analyzer are synchronized in time, they have the same sampling
rate. Use the same time slot roulette. For example, the highest sampling rate of OSCA02 is
100M, then all channels of the logic analyzer and channel A of the oscilloscope can reach a
maximum sampling rate of 100M. Of course, the sampling rate may change with the change of
the time wheel, but the change is for the oscilloscope and logic Analyzer is consistent.
The logic analyzer module and the oscilloscope share the same ground. The yellow wire of
the logic analyzer cable is GND. Pay attention to good grounding during testing. The logic
analyzer module is compatible with 3.3V TTL and 5V CMOS level standards, and cannot be
used to measure voltage logic higher than 5V. The extension port used by the logic analyzer
module is the DE1 port of the oscilloscope, as shown in the figure below:
- 143 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The logic analyzer channel is multiplexed from the ADC data line of the B channel. So we
will see that the switch of the logic analyzer function is in the setting area of channel B of
the oscilloscope, as shown in the figure below:
- 144 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
After checking the function switch of the logic analyzer, the display area of the logic
analyzer will appear in the lower part of the waveform display area of the oscilloscope PC
software:
In the figure, L shows the display area of the logic analyzer, and A shows the
waveform of channel A of the oscilloscope. So if the logic analyzer module you purchased has n
channels, it is actually equivalent to n+1 channels when used, and channel A is an analog
channel, which can see more detailed details than the logic analyzer channel.
Channel A is the oscilloscope channel, which is also used as the trigger channel of the
logic analyzer. For the setting of the trigger function, refer to Section 8.2.14.
The number of channels of the logic analyzer may be different for different types of
oscilloscopes, and may be 4 channels, 6 channels or 8 channels. The logic channel to be
displayed can be selected on the host computer software. Of course, the actual function needs
to match the number of hardware channels of the module. For example, if you buy a 4-channel
logic analyzer, it supports up to 4-channel waveform display (L0-L3).
- 145 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
After selecting the displayed logic analyzer channel, it will take effect in the logic analyzer
display area. It will be displayed according to the enabled logic channel. The leftmost is the
channel number of the logic analyzer, and its color block is the corresponding logic analyzer
cable. color. As shown in the figure below, for example, the blue channel number is 2, that is to
say, the logical channel of L2, its waveform, on the hardware wiring, corresponds to the blue
cable in the cable.
- 146 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
As shown in the figure above, the logic analyzer and the oscilloscope are used at the same
time. The green cursor is the hash cursor of the oscilloscope. As mentioned in the part of the
oscilloscope, these oscilloscope cursors can also be used for the logic analyzer at the same
time. The logic analyzer will also have orange 3-bar logic cursors. Like the green cursor of the
oscilloscope, you can drag the mouse to the desired position, and then the time axis information
of the cursor and the time difference information between the cursors will be displayed directly
next to the cursor.
- 147 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Because the LOTO oscilloscope and logic analyzer are used at the same time, we
recommend using the oscilloscope first to decode the communication signals of one line and
two lines. We have described oscilloscope decoding in chapter 8.2.20.
The decoding function of the logic analyzer is the same, and you can also refer to the
protocol decoding part of 8.2.20. The difference is that due to the logic level input limitation of
the logic analyzer, some decoding cannot be performed on the logic analyzer, such as 232,
RS485, RS422, and CAN bus that are not at the TTL level, all of which need to be on the
oscilloscope channel decoding.
The decoding on the logic analyzer currently supports RS232, iic, and SPI of TTL.
With the above foundation, I took a video of the process of using a USB virtual oscilloscope
+ logic analyzer to do SPI testing and decoding, which is more intuitive than graphics, as shown
below:
16 Four-channel oscilloscope
The LOTO oscilloscope has launched the cascading function, which can cascade
multiple 2-channel standard oscilloscopes into a multi-channel oscilloscope, which will
be introduced in detail in Chapter 17. This chapter introduces an integrated 4-channel
product dedicated to automotive circuit diagnosis. Customers in other application fields
can skip this chapter and read the introduction of the next chapter directly.
LOTO oscilloscope has specially launched a 4-channel product for the test
application of automotive sensor waveforms: OSC984, which has 4 analog input
channels and can measure 4 signals at the same time. As shown below:
- 148 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The supporting software of the 4-channel oscilloscope is the same as that of the 2-
channel OSC980. When using two channels, most of the operations are similar to those
of other models. You can refer to other chapters of this manual.
- 149 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
If the oscilloscope software is opened first, then plug in the oscilloscope device.
The software will run automatically. The software selects 2 channels by default, a two-
channel oscilloscope. At this time, you need to click the Start Device button to stop the
oscilloscope software, and in the software menu bar, select 4 channels 4 channels, and
then click Start Device to start the oscilloscope software.
If the 4 channels cannot be started normally, please check whether the cables of
USB1 and USB2 have been plugged into the USB ports of the computer.
- 150 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The 4 channels are acquired synchronously in time, so they share the same time
wheel and use the same sampling rate.
17 Multi-channel oscilloscope
LOTO's standard oscilloscopes are generally 2-channel, except for the 4-channel
that is specially introduced for automotive diagnostics mentioned in the above chapter.
Cascading is just adding cascading lines to the hardware, and some internal
synchronous designs, appearance and external wiring are the same as the original
standard products.
- 152 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
1. When multiple oscilloscopes are cascaded, the software enforces the hardware
ID of each device. Although we can program the hardware ID of the oscilloscope online,
we need to program the specified number, otherwise the cascading will fail.
ID. 1, the main equipment number, (main control) oscilloscope. The
channels are chA and chB on the software.
ID. 2, auxiliary equipment number, (auxiliary cascade) oscilloscope. The
channels are chC and chD on the software.
ID. 3, auxiliary equipment number, (auxiliary cascade) oscilloscope. In
software, the channels are ch5 and ch6.
ID. 4, auxiliary equipment number, (auxiliary cascade) oscilloscope.
Channels are ch7 and ch8 in software.
If you want to cascade 2 oscilloscopes to form a 4-channel oscilloscope, the
programming ID should be 1 and 2. Similarly, if you want to cascade 3
oscilloscopes to form a 6-channel oscilloscope, the device ID to be programmed
should be 1/2/3.
2. When multiple oscilloscopes are cascaded, the software requires the hardware
type, and each device needs to program the hardware device type.
ID. 1, (main control) oscilloscope. The device type must be written as "Host",
the master device.
- 153 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Other ID, (auxiliary cascade) oscilloscope. The device type must be written
as "Guests", auxiliary device.
HW ID Host/guest channels
1 Host chA, chB
2 guest chC, chD
3 guest Ch5, ch6
4 guest Ch7, ch8
1. Connect the oscilloscope to be programmed with the ID to the USB port of the
PC. Note that only one oscilloscope can be connected to each programming, to prevent
connecting multiple oscilloscopes to open the device and programming confusion.
2. After the USB of the PC is plugged into the oscilloscope device, the same as the
usual oscilloscope startup steps, click the button to start and connect
the oscilloscope.
3. Enter the "Advanced" option page of the oscilloscope and find the "Device
concatenation" group operation area.
For the main control oscilloscope, write "1" for the number and "Host" for the
type.
- 154 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Note, do not jump between IDs. There has to be a number 2 in order to have a
number 3.
When the number of the oscilloscope is programmed as 2/3/4.., not after the "1"
number. Plugging these oscilloscopes into a PC alone will not work as a single two-
channel oscilloscope.
If you need to use it alone, you need to write its number as "1" and the device type
as "Host".
Plug all the oscilloscopes with ID and device types into the USB of the PC and
open the oscilloscope software. First, check "Oscilloscope Cascading", select the
number of channels, and then click (Start Device) to start the oscilloscope software.
The auxiliary cascaded oscilloscope can be controlled in conjunction with the main
oscilloscope (voltage and channel), or it can be controlled independently. The location
of the controls is in the Oscilloscope Cascading tab:
- 155 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Note that when you want to switch the number of channels or adjust and change the
number of cascaded oscilloscopes, you need to suspend the oscilloscope software to
adjust the number of channels.
After clicking this button, the software will pop up a dialog box under the current
setting and display, as shown in the figure below:
- 156 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
This dialog will display the current date and time, and how much time has elapsed
since the screen was locked, and it will block any mouse and keyboard operations
unless the lock is unlocked by entering the unlock code, or someone forcefully quits the
current oscilloscope software.
The password to unlock the screen is 11235.
- 157 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
2. Enter in the Device Manager, and search for the device called OSCxxx inside the section Other
Devices (if not found, unplug and plug again the device and wait), then right click on the device
and select Update Driver Software:
- 158 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
3. In the next dialog box select “ Browse my computer for driver software ”:
4. In the next dialog box, select " Let me pick from a list of device drivers... ":
Note
In many cases, in post-patched systems or systems with strict permissions, this second option
is easier to succeed.
- 159 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
6. In the next dialog box, select Browse to select the driver path like "
Driver_XP_Win7\driver\UsbLjtMS.inf " and click OK.
- 160 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Note:
7. In the list of compatible hardware, select Oscillograph and then click on Next.
- 161 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Note:
In systems with more restrictions a warning window will appear. However, you can install the
driver anyway since it is safe.
- 162 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
(Installation finished)
Note
In some systems, our products will be shown as "WinUsb Device", but after the correct driver
installation it will be shown with the correct name.
The device is provided with two 40MHz probes with 1X/10X attenuation selector, as shown
below:
1. Attenuation Selector
When the attenuation selector at the probe position is toggled to the 1X position as shown
below, the input signal connected by the probe tip is sent to the oscilloscope without attenuation.
Since the input range of the device is ±5V, then just a signal in this range can be measured.
When the attenuation selector is moved to the 10X position, the input signal is sent to the
device attenuated by a factor 10. In this case you can measure signals within the range of ±50V.
- 163 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The 10X selector has better frequency characteristics and wider bandwidth, so you can
use it when dealing with bandwidth and frequency limitations, for better measurements.
2. Compensation Trimmer
When using the 10X attenuator, the frequency characteristic of the probe can be corrected
by adjusting the compensation capacitor on the probe. For this calibration set the probe to the
10X position and connect the probe to the reference signal generated by the device (square
wave). Adjust the compensation trimmer until you can see on the screen a proper square
waveform.
- 164 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The probe ground clip is directly connected to the ground of the virtual oscilloscope circuit
and connected to the PC ground via USB cable. When using a PC powered by the power outlet,
the PC ground is connected to the ground through the three-core power jack. Make sure that
the ground of the probe is connected to the ground of the circuit under test and to the ground
used by the PC; otherwise you can have current leakage, short circuits or measurement errors.
In severe circumstances, the device may also be damaged. See Chapter 4 for details.
Digital oscilloscopes all have dead time (Dead-time), as shown in the figure below, between two
acquisition time periods (Acquisition-time), there must be a dead time that has not been collected.
The signal waveform inside is not collected and displayed
Therefore, oscilloscope manufacturers are trying their best to shorten the dead time period
and lengthen the acquisition time period (gray part in the above figure). However, the dead time
period cannot be reduced indefinitely, and the acquisition time period cannot be increased
indefinitely, both of which need to be weighed in the actual use of the product.
Decreasing the sampling rate or increasing the storage depth are two common methods
to lengthen the acquisition time period. A large memory depth can ensure that the oscilloscope
can use the highest possible sampling rate to acquire waveforms for as long as possible at a
time, but it cannot eliminate dead time periods. Larger memory will reduce the processing and
waveform capture rate. This reduces the oscilloscope's response rate and increases the dead
time between each acquisition. The method of reducing the sampling rate reduces the number
- 165 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
of stored data points, which is beneficial to improve the capture rate and waveform refresh rate
of the waveform.
High-end oscilloscopes also have a third method: the segmented memory function. The
segmented storage feature divides storage into smaller segments. The user can specify that the
storage should be divided into segments, each segment of equal length.
When the oscilloscope observes the first trigger event, it begins storing samples in the first
segment of the acquisition memory until the first segment memory is full. The oscilloscope will
then re-trigger and start looking for the next trigger event to occur.
When a trigger event occurs, the oscilloscope stores the sample point in the next memory
segment. The oscilloscope repeats this process until all memory segments are full. The
segmented memory mode is especially useful when capturing burst signals with long dead
times. Many serial bus and communication signals fall into this category.
With segmented storage, the oscilloscope can maintain a high sample rate while capturing
minutes, hours or days.
The LOTO oscilloscope provides a variety of flexible storage methods for OSCA02
and higher models, and has made a trade-off and consideration for the performance of
the product. It can not only refresh quickly with a small storage depth, maintain a high
waveform update rate, but also set a long time. The acquisition mode is continuous
acquisition, and segmented storage is also possible.
In this mode, customers can adjust the frame rate of waveform refresh by
themselves. You can also set the trigger mode and position freely, and get fast
waveform response.
In the oscilloscope mode, the customer does not need to consider the sampling rate
and storage depth, but only needs to consider the time span of the current screen
display, that is, the time scale. The oscilloscope will automatically adjust the sampling
rate and storage depth according to the current time scale to achieve the most
appropriate effect, taking into account both efficiency and accuracy. Therefore, the
storage depth of the LOTO oscilloscope becomes longer.
In this mode, the LOTO oscilloscope also provides a PC buffering function of 500
frames, and also provides a persistence function. These two functions record data and
waveforms of multiple frames, and display them horizontally and vertically. The storage
is equivalent to 500 current storage depths. For example, if the current storage depth is
128K, turning on the PC cache function is equivalent to a depth of 60M. And these data
can be exported to a computer file, or imported into the software of the LOTO
oscilloscope to reproduce the waveform for analysis, or exported to a text file or
spreadsheet file frame by frame.
- 167 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
There is still a dead time between these 500 memory depths. When a high
waveform refresh rate is used, this function can greatly improve the discovery
probability of abnormal waveforms. This is a bit like the next mode segmented storage
mode, which we will introduce later.
2 Capture card mode (traditional long storage mode):
The LOTO oscilloscope can choose to switch the oscilloscope mode to the capture
card mode, as shown in the following figure:
In this mode, the customer manually selects the sampling rate and selects a
suitable buffer (equivalent to the storage depth) size for acquisition. After the selection
is made, click on the acquisition, and the acquisition will continue for a long time until
the entire selected buffer is filled. The buffer can be set to a maximum of 250
megabytes. If it is a dual-channel acquisition, the time required for the entire acquisition
process is the buffer size divided by 2 and then divided by the sampling rate. For
example, if the sampling rate is 2.4M, the buffer is 250M, and the sine wave of 1KHZ is
collected, the waveform data will be collected continuously for 60 seconds, about 1
- 168 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
minute, and drawn on the screen. During the collection process, it will appear very stuck.
Set a long time. If the waveform is not updated, there will be an in-progress indicator:
You need to wait patiently for the collection to end, as shown in the figure below.
Such 1-minute data is placed on the screen, and in most cases, it will appear very
dense. For example, there are more than 60,000 cycles of waveforms on the graph. We
can zoom in and observe and analyze one by one. Generally, with the naked eye and
the patience of engineers, people cannot see more than 1000 cycles will collapse, that
is to say, the maximum waveform cycle you can view is 1000, which is a probability of 1
in 1,000. For an anomaly of 1ppm, you will definitely not be able to see it. You can now
search by criteria, but what about criteria? Is it a pulse, an edge defect, or an overshoot?
You can't predict, and you can't search. Searching according to various conditions takes
a lot of time and mistakes. In some cases, use the oscilloscope mode of Mode 1 to
quickly refresh the waveform. Once you see that the waveform flashes abnormally, you
can immediately stop the acquisition and find it in the past 500 frame buffers.
Long storage will increase the computational processing requirements. With the
same processor, the processing time will be longer if there is more data, and the
waveform capture rate will become very slow. In order to facilitate the post-analysis of
such long-term and large data collection, we can export the collected data into a data
file, as shown in the following figure:
- 169 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
- 170 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
As shown in the figure above, the design of the LOTO oscilloscope takes into
account the advantages of long-term acquisition, and tries to make up for the
inconvenience of subsequent analysis and viewing. The LOTO oscilloscope can not
only use the acquisition card mode in the case of 2-channel oscilloscope analog signal
input, but also in the case of dual-channel combined acquisition, or in the case of logic
analyzer digital channel input, and can export files. And import it into the PC cache for
segmentation analysis. Once imported into the PC cache, the waveform can be
reproduced for analysis, or exported frame by frame as a text file or spreadsheet.
3-segment storage mode:
As mentioned earlier, in the oscilloscope mode, turning on the PC cache function is
basically equivalent to having a segmented storage capacity of 500 segments. This
segmented memory mode has a particularly typical application scenario: low duty cycle
pulse or burst signals. There is a long idle time between signals. In many cases, it is
difficult to achieve the desired acquisition time even with a large storage or by reducing
the sampling rate. Imagine that it may occur 100 times in a day. , every time the signal
that appears is very sharp and short, it needs a high sampling rate to capture, no matter
how large the oscilloscope has a large storage, it is impossible to store a day's data,
and no matter how low the sampling rate is, it is impossible to capture all 100 signals.
The segmented storage can be done very well.
- 171 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
As shown in the figure above, the segmented storage is triggered multiple times
during the high sampling rate acquisition process, and the data sampled for each trigger
is stored in a small storage that divides the storage space into segments. The
oscilloscope trigger fills one segment at a time, and the idle signals or parts of the signal
that are not of interest between segments are not acquired and stored.
Another common scenario that is particularly suitable for segmented storage is
serial bus analysis – the serial bus is transmitted in packets, and the idle time between
packets will occupy the precious storage resources of the oscilloscope. With segmented
storage, the oscilloscope can only collect data packets without sampling during idle time.
While maintaining a high sampling rate, more data packets can be collected to facilitate
decoding and analysis. Next, let's look at an example of a LOTO oscilloscope using
segmented memory to acquire and decode such a serial bus.
- 172 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
As shown in the figure above, the serial port RS232 will send a series of numbers
every 200ms (if you want the effect to be more obvious, you can set a longer interval,
such as 1 minute interval), 0000000 all the way to 999999999 and sent cyclically. We
started triggering and 500 frames of PC cache function as segment storage.
We see that the 6th frame of the segmented storage as shown in the figure above,
the data is 0000000, and it is decoded. In the 7th frame of the picture below, the data is
1111111, and it is decoded.
- 173 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The 8th frame of the segmented storage in the following figure, the data is 2222222,
and it is decoded...
- 174 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
In this way, until the end, each frame is captured by a data packet and decoded,
and 500 frames can be captured without omission, no matter how long the entire
sending process is.
- 175 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
In the figure above, we have prepared an op amp module as the circuit under test.
We connect the output of the signal source module to the input of the op amp, and the
software will control the sweep frequency to output a sine wave signal to excite the op
amp circuit. This input is connected in parallel with an oscilloscope probe, and the signal
is input to channel B of the oscilloscope, so that we can see the sweep output of the
signal source on the oscilloscope, that is, the input signal waveform of the circuit under
test. The operational amplifier circuit under test amplifies the input signal and outputs it,
and we directly connect the output signal to channel A of the oscilloscope.
This way the hardware and wiring are ready. If you don't have a suitable circuit
under test in hand to get familiar with the function, you can also manually adjust the
amplitude attenuation of the signal source during the frequency sweep to simulate. We
can measure the curve of the amplitude of the output signal of the op amp as a function
of frequency, and we can also measure the curve of the magnification of the op amp as
the frequency of the signal changes.
On the software, we first need to set the parameters of the frequency sweep, as
shown in the figure below, output sine wave, the start frequency of the frequency sweep,
the final frequency, the step size and the time interval of the step, etc. It is especially
important to note that, We need to select "Automatic sweep of frequency response
curve", this option will help us automatically set the time scale of the oscilloscope during
the frequency sweep process, so that we do not need to manually adjust the time scale
continuously, so that the waveform is not too dense or too sparse .
- 176 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Then, we click this button to open the frequency response curve function
interface:
We will open the control panel and graphical interface of the frequency response
characteristic curve:
- 177 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
In the curve fitting setting area, we need to select "None", that is, no curve fitting.
We use the real measurement points to represent the entire frequency response curve.
After completion, we can do curve fitting.
For example, what we want to measure is the response curve of the output
amplitude of the op amp with the frequency change, then we can choose the amplitude
of the channel corresponding to the oscilloscope to which the output signal of the op
amp is connected as the ordinate of the frequency response curve:
After setting, we switch back to the interface of oscilloscope and signal source, and
click the sweep button to start sweeping:
- 178 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The frequency sweep will be performed according to the set parameters, but there
will be a switching time of one or two seconds at the beginning. The signal output during
this time may not be stable, so we click the start of the frequency response curve
interface after the waveform is stable. button to start the drawing of the frequency
response curve, as shown in the following figure:
Next, we don't have to do anything, just observe the drawing of the frequency
response curve and wait for the end.
After we sweep to the end, or we observe that we have obtained enough curves,
we can click the pause button in the lower right corner to end the drawing. At this time,
even if the oscilloscope and the signal source software continue to sweep the frequency,
the interface of the frequency response curve will not continue to be updated.
At this time, we can click the "validation" button. The function of this button is to
normalize and detect these data points in the frequency sweep process, and remove
- 179 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
We can see that the frequency response curve has automatically identified the
position of -3DB and the corresponding cutoff frequency. The cutoff frequency is around
64K Hz. We can choose from a variety of curve fitting methods: linear, quadratic
polynomial, cubic polynomial, exponential fit, logarithmic fit.
Taking the logarithmic coordinate system as an example, the quadratic polynomial
fitting:
- 180 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Taking the logarithmic coordinate system as an example, all fitting options are
enabled:
- 181 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
Taking the linear coordinate system as an example, all fitting options are enabled:
- 182 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
We have made a video record and demonstrated the whole process of the above
process, you can refer to the following video link:
https://www.youtube.com/watch?v=frLPk0uoD_c&t=10s
- 183 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
The blue T line is the reference voltage of the trigger level. The interference of the
waveform in the figure is very small, but we zoom in and then zoom in on the edge of
the signal. As can be seen in the figure below, there is still a sawtooth shape. When the
noise is large, the jitter will be more severe. .
If a very sensitive trigger function is used, these glitches and noise disturbances
near the trigger level can cause triggering. The phenomenon shown is that the
waveform keeps shaking and shaking, and even the rising edge becomes the falling
edge, and the falling edge becomes the rising edge trigger. As shown in the figure
below, the rising edge trigger will occasionally trigger to the falling edge of the waveform,
because there are glitches in the above figure near the trigger level, and the setting with
high sensitivity is selected, then a certain value will be set at the trigger level. The rising
glitch is considered to meet the trigger condition, thus triggering a falling edge.
- 184 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
If you want to stabilize the rising edge of the trigger waveform, you need to use a
hysteresis comparison in the upper and lower range of the trigger level to filter
waveform jitter and glitches near the trigger level. This hysteresis range is the trigger
sensitivity.
- 185 -
LOTO PC Oscilloscope Software User Manual Version22.11.23
When the number of waveform points is dense or the noise is obvious, we can
choose low sensitivity (L). For the case where the noise is small or the number of
sampling points is small, if the low sensitivity is used, the hysteresis algorithm is used to
avoid the influence of noise, which will lead to The trigger position is imprecise, so
select High Sensitivity (H). For example the following situation:
In the above picture, due to the high frequency of the waveform, the sampling points
are relatively sparse. At this time, a high sensitivity trigger must be used to be stuck at
the trigger position at the appropriate sampling point as in the above picture. Otherwise,
due to the algorithm, a little A hysteretic comparison algorithm that removes the effects
of noise glitches will misplace many sampling points. When the sampling points are so
sparse, the displacement of more than 10 sampling points will obviously deviate from
the trigger position visually, and in this case, there is no burr and noise to be removed.
As shown in the figure below, it is a screenshot of the low sensitivity trigger: