3 20 2020 5 07 28 PM140@lec 3
3 20 2020 5 07 28 PM140@lec 3
3 20 2020 5 07 28 PM140@lec 3
2rd Stage
Lecturer: Ahmed Almhanna
2019-2020
1
Chapter Two
2.1 Introduction
The instrument is a device for determining the value or magnitude
of a quantity or variable. The performance of the measurement system is
strongly governed by the characteristics of the instruments and the
transducer used within them. Knowledge of these characteristics is
essential when designing and using of these instruments to ensure that the
measurement requirements are met.
The characteristics of any instruments or transducer are normally
given in the data sheet for the instrument supplied by the manufacturer.
Instrument characteristics can be divided into two categories: static and
dynamic characteristics.
Where
15
A = the accuracy
Xm = the measured value
XT = the true value
Example (2.1)
An ohmmeter us used to measure a resistance of 10 Determine
the accuracy of this instrument if its reading is 9.9 .
Solution
| |
A = [1 - ] * 100 %
| |
= [1 - ] * 100 %
| |
= [1 - ] * 100 %
| |
= [1 - ] * 100 %
| |
= [1 - ] * 100 %
= 99%
2. Inaccuracy: is the maximum error which may exist in measurement,
and it is sometimes known as uncertainty.
Example: (2.2)
For the instrument of Example (2.1), determine the inaccuracy.
Solution
| |
The inaccuracy = * 100 %
|
= [1 - ] * 100 %
= 10 %
16
3. Precision: Is the closeness of the measured value to each other. It is a
numerical value refers to the closeness of the measured values to the
mean value.
The precision is given as:
| ̅|
Pi = * 100 % ………….. (2.2)
̅
Where
P = the precision
Xi = the ith measured value.
̅ = the mean value of the measured values, i.e:
1 n
X Xi ....................(2.3)
n i1
Example (2.3)
A voltmeter is used to measure a voltage of 9V. the following
measurements are made: 8.9, 8.8 , 9.1 , 9.2 , 8.9 calculate the precision of
the third reading.
Solution
| ̅|
Pi = [1 - = * 100 %
̅
| ̅|
P3 = [1 - ] * 100%
̅
1 n
X Xi
n i1
5
= Xii 1
=
= = 8.98 V
17
| |
P3 = [1- ] X 100 %
= 99.73 %
The degree of repeatability or reproducibility in measurements
from an instrument is an alternative way of expressing its precision.
4. Repeatability: Is the closeness of the output readings of an instrument
when the same input is applied repetitively over a short period of time,
with the same instrument observer, method of measurement.
5. Reproducibility: Is the closeness of the output readings of an
instrument for the same input when there are changes in the measuring
instrument observer, method of measurement, location, condition of
use and time of measurement.
6. Sensitivity: Is the ratio of the change in the instrument output to the
change in the measured quantity.
The sensitivity of the indicating instrument is given as:
S= …………………. (2.4)
Output
reading
Slope = sensitivity of
measurement
Measured quantity
Fig (2.1). The sensitivity of measurement representation
18
Example: (2.4)
The following resistance values of a platinum-resistance
thermometer were measured at a range of temperature determine the
measurement sensitivity of instrument.
Resistance Temperature
( ) (C )
307 200
314 230
321 260
328 290
335 320
Solution:
Here, the instrument is a thermometer, therefore; the input to this
instrument is a temperature, while the output is the ohms, as shown
below:
ºC 𝜴
Thermometer
Input
Output
Therefore
S=
= = = 0.233 C
7. Zero drift: Is the description of the effect where the zero reading of
the instrument is modified by a change in ambient conditions as
illustrated in fig. (2.2)
19
Characteristic
Output with zero drift
reading
Normal
characteristic
Measured quantity
Fig (2.2). Description of Zero drift
Output
Characteristic with sensitivity drift
reading
Normal
characteristic
Measured quantity
Fig (2.3). The effect of sensitivity drift on the instrument
characteristic
If the instrument suffers both the zero drift and sensitivity drift at
the same time, then the modification of the output characteristic is shown
in fig. (2.4).
20
Output Characteristic with zero and sensitivity
reading drift
Normal characteristic
Measured quantity
Fig (2.4). The effect of both the Zero and sensitivity drift on the
output characteristic of instrument
Example: (2.5)
A spring balance is calibrated in an environment at a temperature
of 20 C and has the following deflection/toad characteristic:
Load (Kg) 0 1 2 3
Deflection (mm) 0 20 40 60
Therefore,
Zero drift = (5-0) mm = 5 mm
( (
Sensitivity drift = -
21
= 22 – 20 = 2 mm/kg
Thus:
Res = = = 0.2 A
Example (2.7)
If the ammeter of Example (2.6) can read up to 1/4 scale division,
determine the resolution of the instrument in mA.
Solution:
1 A = 1000 mA
10 A = 10000 mA
Res =
22
Example (2.8)
A resistor is chosen at random from a batch having a nominal value
of 1000 and a tolerance of 5 % determine the actual value of this
resistor.
Solution
5/100 * 1000 = 50
the tolerance of this resistor = 50
= 950 1050
11. Range and span: Are the maximum or minimum values of the
quantity that the instrument is designed to measure.
As an Example, for the instrument shown in Fig. (2.5)
The input range is from -10 C to 90 C
The output range Is from 2 V to 12 V
The input span = 90 – (-10) = 100 C
The output span = 12 – 2 = 10 V
-10 ºC 2V
Thermometer
90 ºC 12 V
23