Good Laboratory Practice Regulations Management Briefings
Good Laboratory Practice Regulations Management Briefings
Good Laboratory Practice Regulations Management Briefings
Good Laboratory Practice Regulations Management Briefings, Post Conference Report, Aug 1979 | FDA
INTRODUCTION
The questions received pertained to general and specific issues concerning the
provisions of the GLPs, inspectional procedures, and FDA's enforcement policies. Many
of the questions and their answers have been consolidated to eliminate redundancy and
to focus more sharply on the issues.
No.
AAALAC accreditation does not substitute for Agency inspection nor does it
guarantee automatic compliance with the applicable GLP sections. It is of value,
however, in that it demonstrates that the facility has favorably passed a peer
group review.
3. Results of the quality assurance unit inspections are not routinely available to an
Agency investigator. However, the conforming amendments require that GLP
deviations are to be reported in detail with each submission to the FDA. Are we
required to send the contents of the quality assurance unit inspection report to
the FDA?
No. The GLP compliance statement in the conforming amendments to the GLPs
was included for several reasons:
(a) to provide an orderly transition across the effective date of the regulations. It
was understood that applications for research and marketing permits submitted
to the Agency for some period of time after the GLP effective date of June 20,
1979, would contain final reports of nonclinical studies begun and completed
prior to the effective date, begun prior to the effective date and completed
thereafter, and begun and completed after the effective date. Studies begun and
completed prior to the effective date are not required to comply with the GLPs
and accordingly, the conforming amendments require that differences be noted.
Similar considerations apply to studies begun prior to and completed after the
effective date, although in these studies, those portions underway as of the
effective date are required to comply.
(b) to provide for the submission of final reports of studies, which were not
required to comply with the GLPs but which otherwise, contribute to safety
evaluation. The GLPs do not apply to safety studies conducted by independent
investigators studying regulated products. Such studies are not sponsored by the
product manufacturer, nor is there any intention to submit the results to the
Agency. The study results are published in the open literature. The sponsor is
required red to submit the study to the Agency but could in no way control the
research. If the sponsor wishes to use the data in support of the application, the
conforming amendments provide a mechanism by which the sponsor can prove
that the study was not compromised. A similar situation exists for preliminary
exploratory safety studies done by the sponsor.
This statement is provided by the applicant for the research or marketing permit.
5. What is the degree of compliance with GLPs, which the FDA will require for INDs
submitted after June 20, 1979, but which include toxicology studies initiated
before June 20, 1979, and completed after June 20, 1979?
Those portions of the studies underway as of the effective date will have to be
done in accord with the applicable provisions of the GLPs.
These studies would not have to have been conducted under the GLPs but the
conforming amendment statement of compliance is required.
The Agency has not compiled such a list. A membership list is available from
NALSI, 1747 Pennsylvania Avenue, NW, Suite 300, Washington, D.C. 20O06. All
members who conduct nonclinical laboratory studies are subject to the GLPs.
8. Should a contract laboratory ask a sponsor if the article they are testing is subject
to FDA regulations? Should these studies then be listed as a separate master list
of studies to comply with the GLP regulations?
Contract laboratories should ask sponsors to identify studies which are associated
with FDA regulated products, although the GLPs place this responsibility on the
sponsor. A separate listing of such studies, apart from the firm's master list of all
studies undertaken by the firm will satisfy the requirements of the GLPs.
9. What impact have the GLP regulations had on the cost of performing toxicology
studies?
The president of a large contracting laboratory has stated that three years ago a
chronic rat study could be done for about $80,000; and that the current cost is
closer $250,000. He estimated that half of the increased cost is due to GLPs, 30%
to larger numbers of test animals per study on present day protocols and 20% to
inflation. The Agency has not developed cost estimates.
SUBPART A
GENERAL PROVISIONS
58.1 SCOPE
2. Does the Agency intend to audit analytical data collected on a test article?
3. Does the Agency intend to audit draft final protocols and draft final reports?
The regulations do not require that such materials be retained, however, if draft
reports are available, they may be audited in order to help the Agency follow the
process from raw data to final report.
4. Explain why the GLPs apply to "microorganisms or subparts thereof." How are
microorganisms currently used by FDA in assessment of safety?
For certain products, FDA does request that microbial tests be done for the
purpose of obtaining information on potential neoplastic and mutagenic activity.
Likewise, microsomal preparations (subparts thereof) are used as activating
systems for certain in vitro tests. When this happens, the tests should be done in
accord with the GLPs.
No.
The GLPs apply to safety studies submitted to the Agency in order to obtain the
license. They do not apply to such studies conducted for the purpose of obtaining
batch release of licensed biologicals.
If the medical device application for a research or marketing permit does not
require the submission of safety data for approval, then the GLPs do not apply.
No.
9. Are studies performed for label purposes as required by the Federal Hazardous
Substances Act considered to be nonclinical laboratory studies under the GLPs?
No.
10. When an application for Premarket Approval for a Class III Device is scrutinized,
would a GLP audit by FDA become a criterion for premarket approval?
Safety data are required for Class III Devices and such data are to be collected
under the GLPs, but an FDA audit will not automatically become part of the
premarket approval mechanisms.
11. Are Class I, II and III Devices regulated products within the meaning of the
GLPs?
Yes.
12. Are data contained in a 510(k) notification subject to the GLPs?
No.
13. How do the GLPs apply to the testing of electromechanical medical devices (non-
animal work)?
It is presumed that the question refers to engineering tests and in vitro tests of
such devices conducted to assess functionality. In these cases, the GLPs do not
apply.
14. Please elaborate on the preamble statement (43 FR 59989) that studies involving
"diagnostic products" and "medical devices, which do not come in contact with or
are implanted in man" are not within the scope of the GLPs.
No, the enzyme is part of the lens manufacturing process and its analysis would
be covered by the GMPs and not the GLPs. If, however, the proteolytic enzyme is
sold as a means of cleaning lenses after purchase by a person, the enzyme is an
accessory to a medical device and the safety studies supporting the use of the
enzyme would be subject to the GLPs.
No.
17. Are safety tests conducted on biological products exempt from the GLPs?
Two kinds of safety tests are performed on human biological products. Those
which are performed by the manufacturer prior to licensing, and those performed
post licensing. The tests performed prior to licensing establish the basic safety
profile of the product and they are covered by the GLPs. The safety tests
performed post licensing are part of the required quality control assays, which
permit the release of each batch of product. These tests are not covered by the
GLPs. Safety testing of interstate biological products for use in animals is not
covered by the GLPs since these products are not regulated by FDA.
18. Do the GLPs apply to veterinary drug and biological manufacturers even when
the end products are strictly for veterinary use?
The GLPs apply to animal drugs used on a prescription basis but they do not
apply to interstate veterinary biologicals since these products are regulated by
USDA. Intrastate veterinary biologicals, which are considered to be new animal
drugs, are also covered by the GLPs.
19. If an organization has separate divisions for basic research and for toxicological
safety testing, will the basic research division be subject to inspection under the
GLPs?
No, as long as the basic research division is not providing any service function for
the safety-testing unit.
20.Do the GLP requirements apply to an equal degree to acute, medium-term, and
long-term studies?
21. Are preliminary protocol development or design studies that employ laboratory
animals covered by the GLPs?
22. If an acute oral toxicity study, a 90-day oral toxicity study, and a two-year chronic
study are done, is only the two-year study required to be done under the GLPs?
24. Does the Agency agree that the GLPs are applicable to safety studies intended for
submission to the Agency in support of the approval of a regulated product and
that they are not applicable to preliminary exploratory studies, screening studies,
and range-finding studies whose purpose is to develop or improve the
experimental design of a planned nonclinical laboratory study?
Yes.
25. Many toxicological studies are conducted on products or formulations, which are
comprised entirely of materials which are known to be safe. Such studies are
intended to be a quality control measure to determine lack of product integrity or
to detect adulteration. Do the GLPs apply to such studies?
No. The Agency considers such studies to be quality control studies, which are
not subject to the GLPs.
Generally no. The GLPs apply to safety studies intended for submission to the
Agency in support of product approval. Food microbiology studies are quality
control studies not subject to the regulations.
27. Do the GLPs apply to laboratories, which perform routine sterility analyses on
marketable medical devices, which have been treated with gas for the purpose of
sterilization?
No.
No.
30.It is not clear whether a laboratory involved solely in chemical analysis support of
a nonclinical laboratory study would be required to comply with the GLPs. Can
this be clarified?
Yes. Analytical laboratories must comply with the GLPs to the extent that they
provide data, which support the nonclinical laboratory study. Only those portions
of the laboratory, those procedures and those personnel involved are required to
be in compliance with the GLPs.
31. What is FDA's position regarding the testing of "medical foods" according to GLP
requirements?
By "medical foods," it is assumed that you mean either diets, which complement
human therapy, or dietary products used for nutritional purposes. Such products
usually do not require an application for a research or marketing permit and
therefore they do not fall under the scope of the GLPs. If an application is
required, the safety tests would be within the scope.
32. How do previous GLP inspections prior to these new regulations affect our being
accredited by AAALAC?
Not at all. AAALAC accreditation deals with animal care practices and is a
process, which is independent from FDA's GLP inspections.
33. What about the special problems university laboratories have with complying to
the GLPs? Are these laboratories expected to comply to the same degree as
industry laboratories?
In crafting the final order, the Agency was cognizant of the problems of university
laboratories and certain changes were made which would simplify compliance for
all laboratories without frustrating the intent of the GLPs. All laboratories are
expected to comply to the same degree since product safety decisions are of equal
importance regardless of the size or of the organizational structure of the
laboratory doing the study.
Yes, the laboratories are subject to inspection at the request of the headquarters
bureau, which is evaluating the nonclinical laboratory studies on that substance.
The kind of inspection will be a data audit which will include only those records,
personnel and portions of the laboratory which collected the data on that
substance.
35. Does the definition of nonclinical laboratory study include electrical safety of
medical devices or evaluation of "safe" operation of equipment, i.e., fail-safe
studies for a critical device?
No, functionality studies do not fall within the scope of the GLPs.
For drugs and feed additives used in food producing animals, metabolism studies
come under the GLPs. In these cases, the studies are intended to define the tissue
residues of toxicological concern as well as to estimate tissue depletion. Such
studies on other regulated products are usually conducted as part of the
pharmacological evaluation and would not be covered. However, metabolism
studies on food additives are covered.
37. Does the FDA have a list of laboratories, which do and do not comply with the
GLPs?
No, but the Agency maintains a list of the laboratories which have been
inspected. Copies of individual inspection reports may be obtained as a Freedom
of Information request.
38.Does the term "nonclinical laboratory study" include animal laboratory studies,
which are designed for the explicit purpose of determining whether a test article
has reasonable promise of clinical effectiveness, and in which observations
bearing on clinical safety are only incidental or fragmentary, or at most, clearly
secondary?
No.
39. With regard to the Submission of foreign toxicity data to the Agency, must a
sponsor monitor and inspect the foreign laboratories and audit the final study
report?
40.If a sponsor company utilizes a contract laboratory, who is responsible for the
GLP compliance of the contract laboratory? Should a sponsor have its own
quality assurance unit to monitor contracted studies? If a contract laboratory has
its own quality assurance unit, is it necessary for the sponsor to audit these
studies also? How does a sponsor validate a report of a study performed at a
contract lab?
The ultimate responsibility for assuring the quality and integrity of a nonclinical
laboratory study rests with the person (sponsor) who submits the application for
a research or marketing permit to the Agency. This responsibility can be
discharged as follows:
Case 1. The contract laboratory has a fully functional quality assurance unit and is
operating in conformance with the GLPs. In this case, the sponsor should assure
itself that the contract facility has adequate personnel, facilities, equipment and
standard operating procedures to perform the study properly. Likewise, the
sponsor should examine the procedures used by the contract facility's quality
assurance unit and make a determination that such procedures are adequate to
obtain GLP compliance. Finally, the sponsor should review the final report (not
audit since this has already been done by the contract facility) for consistency and
accuracy.
Case II. The contract laboratory does not have a quality assurance unit and may
or may not be operating in conformance with the other provisions of the GLPs. In
this case, the sponsor must perform all quality assurance functions and take
whatever steps are required to promote the GLP compliance of the contract
facility. The final report will have to be audited since this has not been done by
the contractor.
SUBPART B
ORGANIZATION AND PERSONNEL
58.29 PERSONNEL
1. Must an employee with a cold or the flu be removed from the study?
3. Under what circumstances may QAU audit reports be inspected by FDA? Is there
any requirement to maintain these reports or can they be discarded?
QAU audit reports as a matter of administrative policy are exempt from routine
FDA inspection. FDA's access to QAU audit reports would be through the Courts
should the subject matter of those reports be litigated. Since there is no FDA
requirement that these reports be maintained, the disposition of these reports is
up to the firm's management. FDA advises that such records not be destroyed
without the firm seeking advice from its legal counsel.
4. What are the quality assurance unit inspection requirements for acute and short-
term studies?
For studies lasting less than 4 weeks, each final report should be reviewed by the
quality assurance unit for accuracy. With regard to the in process phases (dose
preparation, dose administration, in vivo observation and measurement,
necropsy, etc.), a random sampling approach could be used so that over a series
of studies each critical phase has been monitored. The random sampling
approach should be statistically designed so that it is adequate for revealing GLP
deviations. The approach and its justification should be made a part of the
standard operating procedures of the quality assurance unit.
(d) the purpose of the inspections is to identify significant problems, which may
affect study integrity, and to determine that no changes from approved protocols
or standard operating procedures were made without proper authorization.
The phases of a particular study will be determined by the nature of the study.
For example, the phases of a typical feeding study include the following:
- animal observations
- food consumption
- body weights
The type of inspection will depend on the nature of the phase. Each phase
must be inspected at least once during the study; the times selected for
inspection should be those most likely to reveal problems before the
quality of the data generated could be adversely affected.
6. Could you take a typical subacute 14-day study and define the phases?
Phases in a short term study (depending on the type) would include protocol
preparation, dose preparation, animal allocation, test system dosage, animal
observation, necropsy, data recording, data analysis and final report writing.
7. By what authority may the Agency examine master schedule sheets for studies,
which may never be used in support of an application for a research or marketing
permit?
Studies that are not intended to be used to support an application for a research
or marketing permit are not covered by the GLPs and need not appear on the
master schedule sheet. If however, the studies are intended to be submitted, then
they should be listed and can be inspected by the Agency under its authority to
evaluate the results of studies designed to demonstrate product safety.
9. In regard to the master schedule sheet, can the "current status of each study" be
satisfied by listing the starting date and completion date of the study? Can the
"status of the final report" be satisfied by listing the estimated or actual date of
issuance of the final report?
Although the GLPs do not specify entries for "current status of each study," dates
alone would not be adequate. Suggested entries that are possible include "study
proceeding according to protocol," "study proceeding according to protocol as
amended on such-and-such date," "study terminated due to such-and-such," etc.
Likewise, entries for the status of the final report might include "awaiting final
hematology report," "data in statistical analysis," "first draft prepared," "draft
under circulation for review and comment," etc.
10. In our laboratory, critical operations for all studies are carried out by the same
individuals using essentially similar procedures. Would it be adequate for the
quality assurance unit to inspect a set of representative operations for GLP and
standard operating procedure compliance that would incorporate a good cross-
section of studies?
The quality assurance unit review is to ensure that the final report accurately
reflects the raw data. Inasmuch as final reports of certain long-term studies can
encompass several hundred thousand observations, it would be a prodigious
exercise for the quality assurance unit to verify and trace all raw data. Further,
the Agency did not mean to require that the quality assurance unit review would
include a check of the accuracy of the calculations used to arrive at the final
report. This activity would be redundant since the contributing scientists would
have already done so in preparing their reports. Rather, the review was expected
to be of sufficient depth to reveal inaccuracies in the final report. Consequently,
the Agency envisioned the development of a statistically based system, whereby;
a random sample of the results in the final report is traced. The procedure should
be made a part of the standard operating procedures.
The Agency has not established an acceptable level of accuracy of the trace.
13. Does the master schedule sheet have to list studies on compounds for which no
data has yet been submitted to the Agency?
Yes. The GLPs cover all nonclinical laboratory studies of Agency regulated
products that support or are intended to support applications for research or
marketing permits.
14. The GLPs state that the quality assurance unit should assure that the final report
reflects the study results. Is it required that every final report be reviewed by the
quality assurance unit?
Yes. This procedure helps to ensure the accuracy of the final report.
15. Does the quality assurance unit review of each final study report have to be
reported to management?
Yes. The quality assurance unit must make periodic reports to management and
the study director on each study. These reports should include the results of the
final report review.
16. At our facility the quality assurance unit reports directly to the executive vice
president of the company and not to the vice president of research and
development. Is it necessary for us to formulate a separate quality assurance unit
within the research and development department?
The GLPs require that the quality assurance unit director and the study director
cannot be the same person. The quality assurance unit must report to a level of
management that has the authority to effect the corrective action as indicated by
the quality assurance unit inspection reports. How this is accomplished
organizationally is a management prerogative.
17. Is it acceptable for the quality assurance unit to report to the management person
who is also responsible for drug safety evaluation?
This is acceptable provided that the management person is not the study director
for the studies being inspected by the quality assurance unit.
The GLPs state that a person may not perform both quality assurance functions
and study direction and conduct functions for the same study. Thus, a
pharmacologist in a research division could serve as the director of the quality
assurance unit as long as he or she did not otherwise participate in the studies
under review by the quality assurance unit.
19. How is the requirement for a quality assurance unit to be interpreted when the
testing facility is itself a quality assurance unit?
By definition, a testing facility could not be a quality assurance unit. A quality
assurance unit, which conducts nonclinical laboratory studies, should make
separate provision for the performance of the GLP quality assurance functions.
This decision rests with facility management but such a choice is acceptable.
In the cited example, Company A would be the study sponsor while Company B
would be a contract laboratory performing a portion of a nonclinical laboratory
study. Both companies would be considered testing facilities, but, since the GLPs
require a single study director for each study, Company A would designate the
study director. Company B would, no doubt, designate a participating scientist in
charge of the animal work and would have the responsibility of submitting a
participating scientist's report to Company A for inclusion into the final report.
22. Is it acceptable to have two study directors for a single study at the same time?
No. The regulations require a single point of study control, which has been vested
in the study director.
23. Do the GLPs permit the designation of a "deputy" or "acting" study director to be
in charge of a nonclinical laboratory study when the study director is out of town,
on vacation, etc.?
Yes.
24. Must the study director personally verify all observations made during a
nonclinical laboratory study?
No. The study director must assure that study procedures are adequate to ensure
the collection of valid data.
25. A study is only as good as the people who perform it and most importantly as the
person who directs it. What does the Agency do to assess the training and
experience of toxicologists?
The Agency recognizes the serious shortage of trained and certified pathologists
as well as toxicologists. The GLPs require that personnel possess the appropriate
combination of education, training and experience needed to do their jobs.
Therefore, it is permissible to have non-veterinarians conduct necropsies
provided their training and experience are adequate. The GLPs do not require
board certification for either pathologists or toxicologists.
27. What does the agency consider to be the minimal acceptable educational
requirements for someone appointed as "study director? "
Due to the wide range of nonclinical laboratory studies and the numerous
combinations of education, training and experience, which would be acceptable,
the Agency did not specify minimal educational requirements for nonclinical
laboratory study participants. The GLPs specify that the study director should
have the appropriate mixture of education, training and experience to permit the
performance of the assigned functions.
SUBPART C
FACILITIES
58.41 GENERAL
1. Would there be any criticism of a laboratory where animals of the same species,
used concurrently in 6-8 short-term eye or dermal irritation studies, were housed
in the same room, assuming there is sufficient spatial separation?
No. This procedure would be acceptable provided that precautions were taken to
prevent animal and experimental mix-ups and cross-contamination.
2. What is the relationship between the FDA and the USDA inspection of animal
facilities?
The USDA inspection is directed towards ensuring the humane care of animals
used in research whereas the FDA inspection is directed towards ensuring the
quality of data obtained from safety experiments that involve animals.
3. We feel that storage of test article - diet mixtures in animal rooms in well-labeled,
vermin proof containers will lead to fewer errors than storage in a central
common area. Is this permissible in light of section 58.47(b)?
Yes. Section 58.47(b) requires separate areas for test article diet mixtures, which
need not be a separate common area or a separate room. In the cited example,
each animal room could have a separate area devoted to feed storage.
4. Is it necessary to provide space for the isolation of diseased animals if they are
immediately removed from the study and sacrificed?
No. The intent of the regulations is to ensure that diseased animals are handled
in a manner that will not adversely impact on the nonclinical laboratory study.
Yes.
SUBPART D
EQUIPMENT
Specific standard operating procedures are required for each piece of equipment.
These procedures can incorporate verbatim the instructions contained in the
equipment manuals.
SUBPART E
TESTING FACILITIES OPERATION
2. With regard to section 58.90(c), does "separate" mean a separate air supply as
well as space?
3. There are many common reagents used in safety studies (e.g. glucose, sodium
chloride, etc.). Do the GLPs intend that these reagents be labeled with storage
conditions and expiration dates?
4. What are the environmental requirements for large animal (cattle/horses) safety
studies?
5. How long do animal care records (cage cards, vendor information, etc.) need to
be retained?
These records should be retained in the archives for the terms specified in section
58.195.
6. Does approximate age of the test system need to be listed on the cage cards?
No.
8. In the absence of the "Guide for the Care of Laboratory Animals," what reference
will FDA use in inspection of facilities for determining appropriate cage sizes,
animal environment, animal facilities, veterinary care, and animal care practices?
References to the guide and regulations promulgated by other agencies have been
deleted from the final order on the GLPs. Nonetheless, these materials do provide
guidance on the current state-of-the-art for animal care and they are helpful both
to the laboratory and to the Agency in determining the adequacy of animal care
practices.
9. Are expiration dates required on purchased chemicals and reagents present in the
laboratory?
Yes, expiration dates are required on such chemicals and reagents when they are
used in a nonclinical laboratory study.
10. Are expiration dates required on prepared solutions made from purchased
chemicals and reagents?
Yes.
11. Are stability data required to substantiate the expiration dates of reagents and
solutions?
12. With respect to evaluating the effectiveness of reagents and solutions throughout
their shelf life, what requirements are there on the certification of efficacy of the
test reagents used to evaluate the effectiveness of the GLP reagents and
solutions?
Standard operating procedures for the analyses should provide such efficacy tests
for reagents and solutions as the scientific literature, the manufacturer's
literature, and the laboratory experience indicate are necessary.
13. What does the Agency expect in the area of analysis of feed and drinking water
for known interfering contaminants?
The GLPs require analysis for and control of contaminants known to be capable
of interfering with the nonclinical laboratory study and which are reasonably
expected to be present in the feed and water. Certain contaminants may affect
study outcome by masking the effects of the test article, as was the case in recent
toxicological studies of pentachlorophenol and diethylstilbestrol. In these studies
the feeds used as carriers of the test article were found to contain varying
quantities of pentachlorophenol and estrogenic activity. These contaminants
invalidated the studies by producing erratic results. The use of positive and
negative controls in these studies was insufficient to compensate for the
variability in the concentration of the contaminants.
To implement this provision of the GLPs, the study director and associated
scientists should consider each study in the light of its length, the expected
toxicological endpoints and pharmacological activity of the test article, the test
system, the route of administration, and other relevant factors to determine what
contaminants could reasonably be expected to interfere. These considerations
coupled with scientific literature, experience and anticipated levels of
contamination should be used to determine which contaminants should be
controlled and analyzed.
For acute studies in which the test article dosage is sufficiently high, in most
instances, to overcome any effects from feed or water contaminants, the
analytical requirement would be minimized.
14. Study directors are frequently unfamiliar with certain aspects of their studies
(e.g. chemical analyses, histopathology, etc.). Is it appropriate for the study
director to authorize all deviations from standard operating procedures?
Yes. As the focal point for study direction and conduct, the study director must be
made aware of and react positively to any deviation from a standard operating
procedure. Where necessary, a study director should consult with other scientists
to determine the impact of a deviation on the study.
15. Is it required that the quality assurance unit test the reagents used in a
nonclinical laboratory study?
Whatever testing is required by section 58.83 of the GLPs for reagents and
solutions may be accomplished by those organizational units that normally
conduct such testing. It need not be done by the quality assurance unit.
16. May reagent grade chemicals be used in a study on the basis of label analysis
declaration?
17. If animals do not have some form of unique identification actually attached to the
animal, is identification using only cage cards appropriate? If the test system is
housed in individual cages, which are uniquely identified, must each and every
animal be identified?
Section 58.90(d) requires that animals which are to be removed from their home
cages or which are to be observed over a long period of time have appropriate
identification. Therefore, identification using only cage cards is not sufficient in
most cases and each animal should be identified.
SUBPART F
TEST AND CONTROL ARTICLES
No, heroic measures need not be taken. Storage conditions should be consistent
with the knowledge of the stability of the mixture under conditions of use and
reasonable so as not to permit accelerated decomposition.
2. What are the details of the Agency's reserve sample retention policy?
Reserve samples are to be retained from each batch of test and control article
prepared in accord with section 58.105(a) for all nonclinical laboratory studies
lasting more than 4 weeks. For the purposes of these sections, the 4-week period
includes initial dosing to the final in vivo observations. Only sufficient sample
need be retained to permit meaningful reanalysis. The samples need be retained
either for the terms specified in section 58.195 or for the useful life of the sample
(dependent on the stability or the quality of the sample) whichever is shorter.
Storage conditions should be those commonly accepted as minimizing the
deterioration of sample quality and need not require exhaustive study to
determine those which maximize stability. All batches of test and control article
mixtures are to be retained even if they are prepared daily.
3. For medical devices, how can stability be demonstrated any more effectively than
by the continued functioning of a device within specifications during an in
vivo nonclinical study?
No. The test article must be sufficiently characterized to ensure that the same
article is used in any further studies.
5. Does FDA expect a firm to conduct long-term stability tests on test article-carrier
mixtures, which are used within a day of preparation?
The firm must determine the stability of the mixtures over the period of their use.
The GLPs require retention of samples of all batches of test article-carrier
mixtures for studies that last longer than 4 weeks. The regulations do not require
stability studies on such samples. Samples placed in storage may be analyzed
periodically to determine their useful storage life.
The GLPs do not cover the basic exploratory chemical tests done to derive the
specifications of the test article. They do cover those chemical tests done on
discrete batches of test article to determine identity, strength, purity and
composition.
7. Does the phrase "mixtures of articles and carriers" also refer to solutions and
suspensions, e.g., a solution of a test article in distilled water?
Yes.
8. For acute studies, is it necessary for the laboratory to analyze each batch of test
article-carrier mixture prior to dosing the test system?
No. Uniformity of the mixture must be known and periodic batch analyses need
to be done.
9. Will dialogues such as this and recent inspectional experience bring about
substantive changes in the final regulations through FDA initiated proposed
amendments? What changes are anticipated in the reserve sample retention
requirements?
The Agency does not believe the initiative to change the GLPs rests with FDA.
Petitions for change may be submitted to the Agency in accord with the 21 CFR
10.30. As was mentioned at the meeting, the Agency recognizes that the reserve
sample retention requirements are extensive and expensive and a petition for
change would be considered.
The Agency has not established guidelines with regard to the frequency of
periodic reanalysis of test article-carrier mixtures. Enough batches should be
analyzed to assure that the test systems are being exposed to the quantities of test
article in the specified protocol.
11. How long must one retain samples of feed used in nonclinical laboratory studies
and should they be frozen?
The sample retention period differs for the various regulated products and the
periods are listed in section 58.195. Feed samples need not be frozen for storage.
Carrier is the material with which the test article is mixed for administration to
the test system. It can be feed, water, solvents and excipients depending on
dosage form and route of administration.
All samples listed in a - d and f above should be retained for the term listed in
section 58.195.
No. Once stability has been determined in accord with good science, it is not
necessary to continually replicate the stability determination.
SUBPART G
PROTOCOL FOR AND CONDUCT OF A NONCLINICAL
LABORATORY STUDY
58.120 PROTOCOL
1. In as much as only wet tissues, blocks and slides are necessary to reconstruct the
histopathologic aspects of a study by a third party, are written notes, tapes, etc. of
the histopathologist's thought process in arriving at a final report legitimately
considered "raw data" in the presence of a signed and dated final report? Does
the Agency have the right to inspect the written notes from the pathologist?
Raw data in this case, refers only to the signed and dated final report of the
pathologist. Agency investigators may wish to examine the interim notes and
reports in an attempt to reconstruct the study but not to second-guess the
scientific process used to arrive at the final report. The GLPs do not require that
these interim reports and notes be retained.
2. What is considered to be raw data in computer systems when the data is
generated from dictated results?
Transcribed dictation, which has been proofread and corrected for typographical
-and transcription errors, is raw data.
3. Do the GLPs require that the protocol be amended to reflect the actual starting
date of the study?
4. It is said that raw data may be any verified exact copy of the original data. In a
computerized data system where data is put directly on disc thence to tape, what
documentation of the program performing this transfer is required to assure that
the tape copy is exact?
The Agency can not precisely answer this question without further details of the
new data format.
6. Are initials and dates on data printouts (e.g., scintillation counters, gas
chromatographs), when these printouts include standards, sufficient
documentation for standardization?
Yes.
7. Is there a time limit for submission of the final report of a nonclinical laboratory
study after its conclusion?
Generally no. On occasion, for marketed products, the Agency may establish time
frames for study conduct. Of course alarming findings on marketed products
should be reported as soon as possible.
Yes.
9. Does "studies in progress on June 20, 1979" refer to the phase of dosing of the
test system or the phase post-dosing but not yet reported?
The quotation pertains to all studies for which the final report has not yet been
completed. Included are all post-dosing phases.
10. The final report requires a list of participants. Should this include technicians as
well as people who perform support functions?
The final report should include the name of the study director, the names of other
scientists or professionals, and the names of all supervisory personnel involved in
the study.
11. When an analysis protocol is developed for the first time by using standard
scientific technique, who shall validate the protocol?
The Agency does not per se validate protocols. Persons developing new protocols
may submit them to the responsible bureau for review and comment prior to
initiating a nonclinical laboratory study.
12. One of the testing deficiencies found in the early Agency investigations of
nonclinical studies was protocol changes that were made without informing the
sponsor. The changes prejudiced the validity of the studies. Accordingly, the
GLPs require that each study have a specific protocol, which is attested to by the
sponsor.
13. The identity of the individual collecting data entered into a computer can be
recorded via the use of a code known only to the individual but directly
identifying the individual; similarly the identity of the individuals witnessing or
reviewing the data can be recorded. Is this acceptable?
Yes, this procedure is acceptable. The key to the code must be made available to
Agency investigators. Do note, however, the final GLPs do not require that data
entries need be witnessed by a second person.
14. Does the following proposal on data entry to computer files satisfy the GLP
intent?
No audit trail is kept for changes to "temporary" file. All changes to permanent
file are recorded in a change file with appropriate data, personnel code, and
comments regarding reason for change and original entry.
No. This method would permit unauthorized tampering with the temporary file
before the raw data are transferred to the permanent file.
16. Section 58.120 describes a sixteen-part protocol and section 58.185 describes a
fourteen part final report. Must all of these be included in protocols and reports
for LD 50's and other short-term tests?
Yes.
17. Is a protocol required for routine research and experimentation?
18. If all raw data are not required in a final report, does this mean, for example, that
weekly body weight or food intake averages can be in a report without the
individual animal data?
The data appearing in a final report depends on the type of study and the kind of
regulated product. Specific advice can be obtained by contacting the Agency
bureau, which has responsibility for the regulated product.
19. If a compound or formula is proprietary, must the final report describe its
detailed composition or chemical structure?
20.How does the requirement for "approval" of protocols apply to "in house" studies
which are' conducted in the laboratories of the actual "sponsor?" Who approves?
What is an "approved" protocol?
The word "approved" was retained in the final order to emphasize that a sponsor
should have a mechanism for evaluation and approval of initial protocols and all
amendments. The specifics of the mechanism can vary but a formal mechanism
should be in place.
21. Must the protocol contain both the name and the code number of the test article?
22. Section 58.120 states that the protocol shall contain the records to be maintained.
Is this intended as a detailed list of each data form to be generated?
No, in this case generalized statements would be satisfactory.
23. How much raw data must be entered into notebooks when performing well-
documented routine tests?
Basically, the GLPs define raw data as the immediate results of original
observations. All such immediate results must be entered.
24. What is meant by the statement in section 58.120(a)(12), which pertains to the
method by which the degree of absorption of the test and control articles by the
test system will be determined?
The GLPs do not mandate that absorption studies need be done, or which kind of
study is satisfactory. The GLPs do require, however, that the protocol describe
the method used if one is necessary to achieve the study objectives.
25. Please clarify the issue of having to provide reasons for all corrections to data
entries. It seems unreasonable to require reasons for "obvious" error corrections
such as misspell ings, transposed numbers, and wrong year early in a calendar
year.
26. How and to what extent is the selection of the test system to be justified in the
protocol?
27. Are we expected to label all specimens (e.g. serum, blood, urine, tissue slides)
with their exact nature?
Yes. Such information is useful in preventing mix-ups.
28.Why does "test system, study, nature and date of collection" have to be located on
a specimen container? Can such information be coded?
Specimen refers to any material derived from a test system for examination or
analysis. Consequently, blood, tissues, urine, feces, etc. are considered to be
specimens whose containers must carry the required label information. Such
information will help preclude mix-ups in the subsequent handling of the
specimens. Accession numbers or code numbers can be used for samples of
specimens, which are subjected to further analysis. For example, in
histopathology the excised fixed tissue is a specimen, which must carry all the
label information. However, the blocks and slides prepared from that tissue could
be identified by accession numbers. Similarly, in tissue residue analysis, the
excised tissue is a specimen; whereas, tissue samples, which are homogenized
and otherwise prepared for further analysis, are not specimens and need not
carry full labeling.
SUBPART J
RECORDS AND REPORTS
1. What types of storage conditions are required for the storage of retained
specimens?
The Agency has not developed guidelines for storage conditions. The Agency does
not expect heroic measures to be used, but conditions should be reasonable in
light of the nature of the specimen. Storage conditions, which foster accelerated
deterioration, should be avoided.
2. In section 58.185, it is stated that test and control article identification and
characterization must appear in the final report signed by the study director.
However, if the study director is affiliated with a contract laboratory, he/she has
no need to know such details of a proprietary test article. Do you agree that such
information can be appended to the final report by the sponsor rather than be
provided by the study director?
Yes.
3. Is the storage of archival material (tissues, slides, raw data) the responsibility of
the testing laboratory or can this responsibility be assigned to the sponsor of the
study?
The GLPs permit these materials to be stored in the archives of either the testing
laboratory or the sponsor. If they are stored in the sponsor's archives, the
archives of the testing laboratory must identify the storage location.
4. If a sponsor agrees to characterize and store test articles submitted for study to a
contractor, must the contractor also verify the characterization and provide
storage for the test articles?
The completion date is the date that the study director signs the final report.
Some discretion must be used however, since the protocol calls for a proposed
"completion date." In this case, it would be adequate for the protocol to list a
completion date for the in vivo phase and qualify it as such.
6. With respect to archival material, what is required to be listed as the date of the
study?
The study date would be the same as the completion date of the study.
8. What should be included in the signed and dated reports of the individual
scientists participating in the study?
The final report prepared by the study director should have appended to it all
reports written by other participating scientists. These reports should contain
sufficient detail to enable the study director to write a final report, which reflects
the results of the study.
SUBPART K
DISQUALIFICATION OF TESTING FACILITIES
58.200 PURPOSES
FACILITY
ENFORCEMENT STRATEGY
A laboratory will not be considered for disqualification unless all of the following
criteria are met:
The violations of the various provisions of the GLPs are evaluated to assess their
impact on the validity of the studies. It is impossible to assign weights to the
various provisions of the GLPs. Noncompliance with the various provisions must
be evaluated in the context of the entire laboratory operation and the kinds of
studies being performed. Thus, a violation of a specific provision may be critical
for one laboratory doing long-term studies and not for another laboratory
engaged in short term studies.
The disqualification will last until the laboratory submits in writing to the
Commissioner, reasons for reinstatement including a detailed description of the
corrective actions it has taken to assure that the violations which led to
disqualification will not recur. Reinstatement will depend upon one or more
inspections which show that the laboratory is in compliance with GLPs.
3. Paragraph 231 of the preamble to the GLPs states: "The order of disqualification
creates a rebuttable presumption that all studies previously conducted by the
facility are unacceptable. Paragraph 226 states: "Studies conducted at facilities
that are in substantial compliance will be presumed to be valid." Can we presume
that studies conducted during a period when a lab is found to be substantially in
compliance will be accepted by FDA as valid even if the laboratory is disqualified
at a later date?
FDA will not disqualify a laboratory on the basis of one invalid study.
Disqualification is viewed as a most serious regulatory sanction by FDA and will
only be imposed when the facts demonstrate that the laboratory is incapable of
producing valid scientific data and will not take adequate corrective measures. In
the event a laboratory is disqualified, all studies performed by the laboratory,
including those in progress are presumed to be unacceptable unless the sponsors
of those studies can establish, to the satisfaction of FDA, that the studies were not
affected by the circumstances that led to the disqualification.
5. What steps must be taken by FDA prior to removal of a product from the market
because of a rejected study which was pivotal to the assessment of safety?
ENFORCEMENT STRATEGY
GENERAL POLICY
1. What is the regulatory basis for conducting GLP inspections? It would seem that
by making the GLPs regulations instead of guidelines, that the attorneys and
accountants are managing the studies. How does that produce good science?
The GLP regulations are process-oriented; they are designed to assure that the
data collected in a nonclinical laboratory study are valid and accurately reflect the
responses of the test system. The GLP inspections are necessary to assess the
degree of compliance with the GLPs. The science of a study depends on the
appropriateness of the design selected to answer the questions raised in the use
of the test article as well as the soundness of the conclusions drawn from the data
collected in the study. The assessment of the scientific merit of a study is made by
scientists.
2. Does FDA have the authority to audit an ongoing study of a product for which an
application for a research or marketing permit has not yet been submitted to
FDA?
Where the Agency has reason to believe that the laboratory is in fact conducting
nonclinical laboratory studies, a letter will issue to the laboratory stating that
FDA will not accept any future studies performed by that laboratory in support of
a research or marketing application. If the laboratory has not, or is not testing an
FDA regulated product, it is also advised to contact the local FDA district office to
arrange for an inspection should they anticipate engaging in such safety testing.
5. If GLP regulations are not retroactive, will FDA audit pre-June 1979 studies? If
so, will FDA investigators list non-conformance with GLPs on the FD-483 Notice
of Observations associated with those studies?
FDA will continue to audit pre-June 1979 studies for purposes of assessing not
only the quality of a particular study, but also the general performance of the
laboratory prior to the time when GLP regulations were first proposed in
November 1976. This is necessary because many of the marketing applications
pending before the Agency contain studies performed prior to 1976.
While deviations from the GLPs will be noted in the FD-483 associated with these
studies, the Agency will use this information only to make a judgment regarding
the scientific acceptability of those studies and will not use the deviations to
initiate regulatory action against the laboratory. After the June 1979 effective
date, however, deviations from the GLPs could result in regulatory action against
both the studies and the laboratories.
6. Will the GLPs apply to a study, which has been completed prior to the June 20,
1979, effective date for which a final report will not be prepared until after?
The GLP regulations became effective June 20, 1979, and those portions of
studies underway, as of that date, even if only the final report, became subject to
the regulations at that time.
Generally speaking, FDA inspects only those laboratories, which have conducted
studies submitted to the Agency. FDA strongly advises any laboratory which
intends to engage in the safety testing of a regulated product, and which has not
been previously inspected, to contact the local FDA district office and request a
GLP inspection.
8. Will FDA accept data from a study not conducted in accordance with GLPs for
regulatory purposes?
Even though a study has not been conducted totally in accordance with GLPs,
FDA may accept the data from such a study if it can be demonstrated that the
areas of non-compliance have not compromised the validity of that study. As a
special corollary to this policy, FDA will take note of positive findings of toxicity
in a study even though that study was not conducted in compliance with GLPs.
While a technically bad study can never establish absence of a safety risk, it may
establish the presence of an unsuspected hazard or untoward effect.
9. Where can the Inflationary Impact Assessment Report of the GLPs be obtained?
Rockville, MD 20857
10. How does FDA protect the confidentiality of valuable commercial or trade secret
information given to an investigator during a GLP inspection?
FDA employees are required by statute to protect the confidentiality of any trade
secret or confidential commercial information which they may acquire in the
performance of their duties. Thus any trade secret information which an FDA
investigator may receive from a laboratory being inspected is exempt from public
disclosure. Whenever the FDA receives a Freedom of Information Act request for
a copy of the laboratory inspection report, all information which falls under the
definition of trade secret or confidential commercial information will be purged
from the report before it's released.
11. Will FDA review non-GLP studies (range-finding, exploratory studies) in the
course of conducting GLP inspections of studies intended to be submitted to the
Agency? This is of particular concern in protecting proprietary research data. Will
there be an opportunity for the inspected firm to do an FOI review before the
final inspection report is written?
FDA may review on-going non-GLP studies as described in question 23 on
Subpart A and question 11 under "Inspections."
The inspected firm may not review a draft inspection report for purposes of
identifying what should not be released under FOI. Even if the Agency permitted
this, which it does not, the fact that the report was made available to someone
outside the Agency would immediately make that draft document available for
public disclosure under the provisions of the FOI regulations.
12. Will foreign laboratories be inspected to determine their compliance with GLPs?
13. What accords have been made with foreign countries regarding GLPs and
inspections?
14. Has FDA inspected its own animal research facilities for compliance with GLPs?
Other Federal laboratories?
Yes. To date, FDA has completed GLP inspections of all its animal research
facilities and is taking steps to bring all its laboratories into compliance. FDA has
also established contacts with the NIH, DOD and USDA for purposes of
scheduling inspections of laboratories performing safety studies intended to be
submitted to the Agency.
15. Has FDA established liaisons with other Federal agencies regarding the GLP
program?
Yes, liaisons have also been established with CPSC, EPA, and OSHA for purposes
of furthering the objectives of the GLP program, scheduling inspections of
Federal laboratories and sharing information resulting from the FDA program.
INSPECTIONS
Yes, it is possible. The testing facility may appeal any differences it has with the
investigator first to the FDA district office and, if this is not satisfactory to FDA
headquarters.
3. Will the inspectional training course at the National Center for Toxicological
Research be open to industry and academia?
No. The training of industry and academic personnel to enable them to properly
perform their duties is the responsibility of their employers. However, FDA is
prepared to participate in any training courses, which may be offered by industry
associations or the academic community to the extent that resources will allow.
4. If the GLPs are Phase I of Bioresearch Monitoring, what other phases are
anticipated by FDA?
The scheduling bureau makes the decision. During the past two years,
headquarters scientists have participated in about half of all GLP inspections and,
with rare exception, the Bureau of Biologics assigns a headquarters scientist to
each GLP inspection. Resources do not permit more extensive participation.
Laboratories are selected for inspection by bureaus within FDA. The criteria for
selection are actual or potential involvement in studies associated with products
regulated by FDA. Inspections will involve a specific study submitted to a bureau
or a study selected from the firm's master list which is of interest to FDA.
Routine surveillance inspections will occur at least once every two years or more
frequently depending upon findings of previous inspections. However, more
frequent inspections may occur when an audit of a specific study submitted to
FDA or EPA in support of a marketing application is required.
Either type of inspection can result in more frequent visits if serious adverse
findings are reported. These latter visits are considered compliance or follow-up
inspections and are carried out to determine if correction of previous violative
conditions have been made.
Because of the comments received during the conferences and the experiences to
date with this program, laboratories will generally be notified prior to inspection.
However, compliance or special investigation inspections may not follow this
procedure.
A facility may request an inspection from either the local FDA district office or
from FDA headquarters. However, an inspection will be initiated only with
headquarters concurrence. Consideration will be given to the work schedules
under which district management is operating.
11. If a laboratory is not performing a study on an FDA regulated product at the time
the investigator arrives, will the inspection still be carried out?
Routinely, GLP inspections are not scheduled unless the Agency has received a
final report on a regulated product or has received submitted protocols, interim
study reports, or knows that a study on a regulated product is underway. In the
case of a laboratory that is not currently performing a study on a regulated
product the laboratory will be asked to consent to an inspection. The FDA
investigator will utilize an ongoing study, even though it is not associated with an
FDA regulated product, to document the laboratory's compliance with GLPs. In
such cases, the study will not be audited in terms of validating the raw data, and
specifics of the study will not be included in the inspection report.
12. Will inspections cover other areas such as chemistry, physical testing, metallurgy,
etc.?
To the extent that the protocol of a nonclinical laboratory study requires tests in
the field of metallurgy, clinical chemistry, etc., we will examine and evaluate
adherence to test specifications or protocol requirements.
13. Are firms notified of specific studies to be audited? Will sufficient time be allowed
to seek authorization from the sponsor of the study to disclose the data to the
FDA investigator? What happens if the sponsor of the study refuses to authorize
the laboratory to disclose the records?
As stated with respect to prior notification of inspection, where FDA has an
interest in auditing a study, ample time generally will be provided for the facility
to seek authorization from the sponsor to disclose the data. In some cases, FDA
investigators may begin inspecting the physical layout of the facilities while
authorization to release the study records is being obtained. If the sponsor
refuses to authorize disclosure of the records to the investigator, FDA will pursue
the matter directly with the sponsor.
14. Can FDA investigators ask for records to which they are not legally entitled; can
they engage in "fishing expeditions?"
15. Should the Form-FD-483, Notice of Observations issued by the FDA investigator
reflect current practices only; and should it include practices that were corrected
during the course of the inspection?
The FD-483 can include historical practices, which may have affected the
scientific validity of the nonclinical study in question even though
subsequent correction may have occurred. Any corrective action taken by the
facility will be noted by the investigator in the establishment inspection report.
16. At time of the observation, the management should discuss any differing
opinions and attempt to clarify the investigator's perceptions or observations.
The management may also, at the conclusion of the inspection, offer to explain
what the management considers to be erroneous 483 observations. Should the
matter in question remain unresolved, a written objection should be sent to the
local FDA district director or a meeting with district personnel should be
requested to attempt to resolve the issue.
17. What is the procedure for correcting errors in the FDA investigator's inspection
report? Such errors can be damaging to the laboratories since the reports are
ultimately available through FOI.
If in fact an error is made in an investigator's report, the matter should be
immediately brought to the attention of FDA district management. If district
management agrees with the complaint, the report will be amended and
amended reports will be sent to all outside persons who may have received the
erroneous report. It should be stressed, however, that the time to change what a
facility believes is an erroneous conclusion is when the FD-483 is discussed with
laboratory management because as soon as the FD-483 is presented to
management, it becomes available for public disclosure.
18. Does refusal to allow the FDA investigator access to certain information, which
the laboratory sincerely believes is not subject to FDA jurisdiction, constitute a
refusal of inspection? How can a disagreement of this kind be resolved?
Refusal to permit access to records which are associated with a study being
audited or which preclude a judgement being made regarding compliance with
GLPs, is considered a refusal of inspection with certain ensuing consequences.
However, a facility may legitimately question FDA authority to review certain
documents. Such objections and the reasons therefore, should be presented in
writing or by telephone to the FDA district office management where the
investigator is based. Each case will be individually reviewed both in the field
and, if necessary at headquarters and a decision will be communicated to the
inspected facility.
19. Will inspections and audits of foreign laboratories be carried out? Who pays for
these inspections?
20.In order for foreign laboratories to comply with the GLPs, do protocols, standard
operating procedures, records, etc. have to be in English? Do FDA investigators
bring interpreters with them to review records and data?
21. What kind of training does an FDA investigator have which qualifies him/her to
conduct a GLP inspection or data audit? Does the investigator draw conclusions
from his observations regarding the competence of the laboratory or quality of
the studies?
Along with education in one of the natural or physical sciences, the individuals
selected to conduct GLP inspections generally have had considerable experience
inspecting facilities involved in drug manufacturing, biologics production,
medical device assembly, food processing, and a range of other operations on
products regulated by the Agency. In addition, the investigators conducting
nonclinical laboratory inspections (GLPs) have undergone intensive training in
the normal operating procedures of nonclinical testing facilities. This training
which includes a full review of the Agency's policies and of the GLP regulations
National Center for Toxicological Research accomplished at FDA's National
located in Pine Bluff, Arkansas. Field investigators are encouraged to contact any
resource within the Agency, i.e., scientists and other personnel of the various
bureaus to resolve scientific questions that may arise during an inspection.
Bureau scientists and not the investigators, draw conclusions regarding the
competence of the laboratory of the quality of the study
22. Does a laboratory manager have the right to ask for the FDA investigator's
educational and experience qualifications prior to a GLP inspection?
Yes, questions regarding the formal training, educational experience, and on-the-
job training of an individual investigator may be addressed to the investigator
prior to a GLP inspection.
23. What can a laboratory manager do when he encounters an FDA investigator who
is overly antagonistic or uncertain as to what he is looking for?
24. What assurance does a firm have that confidential or trade secret information
given to the FDA investigator will be safeguarded by the Agency? What happens
when an FOI request for the inspection report is received by FDA?
Section 301(j) of the Food, Drug, and Cosmetic Act prohibits any employee from
revealing for his/her advantage any information obtained in the course of
carrying out his/her duties. Trade secrets and confidential commercial
information are deleted from documents before they are released under FOI.
Inspected firms may help by identifying information, which they consider to be
confidential when it is given to the investigator. FDA will however, exercise its
own judgment, in accordance with its FOI regulations as to whether such