GHB K G8 Met 508
GHB K G8 Met 508
GHB K G8 Met 508
This document details the Lab Protocol for testing the items listed in the following table:
2 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
The glucose aldehyde group and the free amino group on the valine in the N-
terminus of the hemoglobin beta chain react to form the Schiff base, aldimine
(also known as labile HbA1c or LA1c).
3 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
The analyzer dilutes the whole blood specimen with Hemolysis & Wash Solution,
and then injects a small volume of this specimen onto the TSKgel G8 Variant HSi
Column. Specimens may also be diluted offline using the dilution procedure
below. Separation is achieved by utilizing differences in ionic interactions
between the cation exchange group on the column resin surface and the
hemoglobin components. The hemoglobin fractions (designated as A1a, A1b, F,
LA1c+, SA1c, A0, and H-V0, H-V1, H-V2) are subsequently removed from the
column by performing a step-wise elution using the varied salt concentrations in
the Variant Elution Buffers HSi 1, 2, and 3.
The time from injection of the sample to the time the specific peak elutes off the
column is called Retention Time. The Tosoh Automated Glycohemoglobin
Analyzer HLC-723G8 software has been written so that each of the expected
fractions has a window of acceptable retention times. If the designated peak falls
within the expected window, the chromatogram peaks will be properly identified.
When a peak elutes at a retention time not within a specified window, an
unknown peak (P00) results. If more than one peak elutes at times not specified
by the software windows, each is given a sequential P0x title. In order to keep
the peaks within their appropriate windows, it may be necessary to change how
fast or slow the buffers are moving through the system by changing the pump
flow rate.
The separated hemoglobin components pass through the LED photometer flow
cell where the analyzer measures changes in absorbance at 415 nm. The
analyzer integrates and reduces the raw data, and then calculates the relative
percentages of each hemoglobin fraction. The Total Area of the SA1c is divided
by the sum of the total areas of all peaks up to and including the A0 to obtain a
raw SA1c percentage. This uncorrected result is substituted as the “x” value in
the linear regression formula determined during calibration. The analyzer prints
the final numerical results and plots a chromatogram showing changes in
4 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
absorbance versus retention time for each peak fraction. Specimens that show a
deterioration peak, an abnormal looking chromatogram, or a presumptive
hemoglobin variant are reflexed to the boronate affinity method. In addition, G8
results ≤ 10 % HbA1C that have a LA1C ≥ 1/2 SA1C are reflexed to the boronate
affinity HPLC method for verification.
2. SAFETY PRECAUTIONS
1. Laboratory services are requested through the Westat system operations via
an email notification containing a unique manifest list of the samples and
sample analysis type (e.g. GHB), which confirms that specimens have been
shipped to DDL.
2. Each Manifest Form should include and be verified against each sample
received:
a. Patient Sample ID #
5 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
b. Test Name
c. Date Collected
d. Shipment ID #
e. Shipment Date
f. Lab Name
g. Lab ID
h. Survey Year
3. Once specimens are received and verified the corresponding file is imported
electronically into the SQL server database via secure transfer.
4. After analysis the results, date analyzed and tech initials are imported from
the instrument into the SQL server database via secure transfer.
5. Data check sheets are printed out and checked against the instrument
printouts by the supervisor.
6. After results are cleared by the supervisor a results file in the specified
format is exported and uploaded to Westat via secure transfer.
6 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
1. Equipment:
a. Tosoh G8 Glycohemoglobin Analyzer (Tosoh Bioscience, Inc., South San
Francisco,CA).
Parameter Setting
Wavelength Sample: 415nm
Reference: 500nm
Column Temperature 25 °C
Injection Interval 1.6 min
Calibration Two-point
Working Temperature 15°C - 27°C
7 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
8 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
i. During pumping, if the pressure does not rise, air may be present on
the outlet side of the pump.
ii. Use the following procedure to remove the air.
iii. Verify analyzer is in STAND-BY mode.
iv. If the analyzer is not in STAND-BY mode, press the STOP key and
wait until ‘STAND-BY’ appears on the Status screen.
v. Press the REAGENT CHANGE key on the MAINTE screen.
vi. Press the DRAIN FLUSH key.
vii. The following message will be displayed requesting that the drain
valve be opened: “Open the door on the left side of the analyzer and
turn the drain valve 90 degrees in the counterclockwise direction to
open the valve.”
viii. Turn the valve ONLY 90 degrees counterclockwise.
ix. Press the OK key.
x. A confirmation message will appear. Ensure the drain valve is open.
Press the OK key again.
xi. Air stuck in the pump will automatically be removed. This procedure
takes approximately 7 minutes to complete and is finished when the
“FLUSHING…” message disappears.
xii. A message will be displayed requesting that the drain valve be
closed. Turn the valve back 90 degrees in the clockwise direction to
securely close it.
xiii. Press the OK key.
xiv. Press the EXIT key to return to the main screen second page. Press
the PUMP key.
xv. If a pressure of within the acceptable range for the filter is displayed
in the HbA1c mode with no pressure fluctuation, air removal is
complete. Press the PUMP key again to stop the pump motor. If the
pressure does not rise 5Mpa or is unstable, stop the pump and
repeat the air removal procedure again.
e. Replacing the Filter Element
i. Replace the filter element after 400 injections or when the pressure
exceeds limits as indicated on the column inspection report +4MPa.
ii. Verify analyzer is in STAND-BY mode.
iii. If the analyzer is not in STAND-BY, press the STOP key and wait
until ‘STAND-BY’ appears on the Status screen.
iv. Open the door below the display.
v. Confirm that the SV1 key is open (O) on the second page of the main
screen.
vi. Remove the filter outlet (peek) tubing from the top of the filter
assembly.
vii. Loosen the top of the filter holder assembly by turning it
counterclockwise. Remove the filter holder by pulling it straight out.
viii. Lightly press the top of the holder to remove the old filter element. If
salt crystals are present in the holder, rinse with distilled or deionized
9 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
10 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
7) Check for leaks. Press the PUMP key to start the pump and
confirm there is no fluid leakage.
8) Check for fluid leaks at the connections. If leaks occur, tighten
fittings.
9) Verify that pressure stabilizes. The pressure should rise to the
pressure level that is indicated on the column inspection
report + 4 MPa. If leaks occur, tighten fittings.
10) Reference pressure and limits should be recorded onto a label
attached to the instrument. This label should include the data,
reference MPa, low and high MPa limits, and column serial
number.
11) After verifying connections are secure, stop the pump by
pressing the PUMP key.
12) Close column oven.
13) Close front doors of the analyzer.
14) After connecting a new column, reset (zero) to the column
counter in the REAGENT CHANGE screen.
15) Record change on daily diary sheet.
16) Run at least three whole blood samples to prime the new
column. Verify that the retention time for the SA1c peak is
between 0.57 – 0.61 minutes. The ideal the retention time for
SA1c is 0.59 minutes.5
17) If necessary, adjust the flow rate to match the retention time
for the SA1c peak on the reference chromatogram included
with the column.
18) Once the retention time matches within +/- 0.2 min, print off
the chromatogram and submit it along with the included
chromatogram from the manufacturer to the supervisor or
delegee for usage approval.
19) In the event of column lot change, all reagents need to be
replaced to lots corresponding to the new column lot. A
comparison needs to be done between the old and new lots
of columns/reagents (n=40). Comparison must meet these
criteria:
a) XY plot (current lot on x-axis) with linear regression
performed.
b) Slope = 1.0+/- 0.1
c) Intercept = 0.0+/- 0.1
d) R2 > 0.98
e) 95% CI of the differences between x and y within 0 +/-
0.5% HbA1c. Overall mean bias within +/- 0.2%
HbA1c. If any outliers (>1% HbA1c difference between
X-Y) occur, investigate further.
g. Replacing printer paper.
i. Lift the printer cover (upper lid) to the back to open.
11 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
ii. Push the paper holding lever down to the very front and wrap the
remaining paper onto the roll.
iii. Lift the roll up and remove the mandrel.
iv. Insert the mandrel into the new roll with attention to the direction.
v. Return the paper holding lever to the very back and insert the paper
into the printer. Press the feed switch to feed the paper.
vi. Check for twisted paper. If the paper is twisted, push the paper
holder lever to the front, adjust the paper, and return the lever to the
back.
h. Replacing the sampling needle.
i. Replace the needle if it is bent or broken. Although needle
replacement is normally done by field service personnel, the
procedure below may be performed by the operator.
ii. Put on protective clothing (goggles, gloves, etc.) and take care not
to touch the end of the sampling needle during handling.
iii. Press the POWER key to switch off the analyzer.
iv. Use a screwdriver to remove the sampling cover screws.
v. Remove the sampling needle cover.
vi. The sampling needle unit is located behind this cover. Grasp the
upper part of the sampling needle unit by hand and slowly pull the
unit forward as far as possible.
vii. A small volume of reagent may leak during needle replacement.
Place a tissue or plastic pad under the sampling needle tip to absorb
any leakage.
viii. By hand, loosen and remove the tubing connected to the 3-way
block.
ix. Remove the screws on the upper section of the sampling needle. Be
careful not to drop the screws or the holding plate inside the machine
during this operation.
x. Remove the screws that hold the guide through which the tubing
passes.
xi. Slowly lift up the sampling needle to remove it. Place immediately
into a sharps container.
xii. Insert the new sampling needle with the bevel facing forward. The
sampling needle must be positioned with the bevel facing forward or
the needle will not correctly dilute the sample.
xiii. Secure the holding plate with the screws.
xiv. Pass the tubing through the guide, secure with the screw, and
securely connect the tubing to the 3-way block.
xv. Move the sampling unit back and forth and confirm that the tubing
does not catch. If necessary, loosen the screws and change the
guide direction to prevent the tubing from being obstructed. Push
the sampling unit back; close the blue cover by following the above
procedure in reverse. Secure the screws.
12 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
xvi. Turn on the Main Power Switch. Press the POWER key on the
control panel and allow the analyzer to complete the WARMUP
process then to the STAND-BY state.
xvii. Assay 3 whole blood samples to confirm the sample is aspirated
correctly. The Total Area for these samples should be approximately
the same as it was before the sampling needle replacement.
xviii. Adjusting the Flow Rate: The flow factor is generally 1.00 mL/min,
but can be 1.02/1.03 mL/min dependent on instrument factory
setting. The flow factor should only be adjusted +/- 0.05 of the default
factory setting.1
3. Instrument Preventative Maintenance (PM)
a. A monthly PM is performed by a trained member of our DDL staff using the
monthly PM checklist (located on the H:\DDL\D150\Templates drive). The
completed checklist is reviewed by the lab supervisor or delegee for
verification.The Monthly PM is performed using the below steps;
i. With G8 powered on, press the menu button, then utilities, then
password.
ii. Enter “MAINTE” in password screen, then press enter.
iii. Exit back to main screen.
iv. Press Mainte button.
v. Press Sampler Mech button.
vi. Remove the fitting from the top of the filter holder assembly.
vii. Remove the top portion of the filter assembly.
viii. Eject the filter.
ix. Wet a cotton swab with DI water. Clean inside of filter holder
assembly, including the threads. Clean the top portion of the
assembly as well.
x. Reassemble the filter assembly.
xi. Remove the two screws from the top of the blue cover door, and pull
open cover.
xii. Press the “Move Y stat” button. The needle should move to the stat
position.
xiii. Remove the two screws holding down the needle. Remove the small
metal plate under the screws.
xiv. Remove the needle.
xv. Remove the screws from the bottom portion of the needle assembly.
Remove the metal plate beneath these screws.
xvi. Remove the blue o-ring.
xvii. Clean the screws, plates, and o-ring.
xviii. Using an alcohol wipe, clean the needle to remove any dried blood.
Use the corners of the alcohol wipe’s package to clean out the
grooves on two sides of the needle.
xix. Use the alcohol wipe to clean the positioning wheel behind the stat
well.
xx. Replace the o-ring. Place one drop of TriFlow lubricant in the middle
of the o-ring.
13 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
xxi. Replace the metal plate that holds the o-ring in place, and replace
the screws.
xxii. Insert the needle into the top hole, followed by the lower hole. Assure
that the bevel of the needle tip is facing you.
xxiii. While holding needle in place, replace the metal plate at the top, and
screw it into place.
xxiv. Assure needle is secure and press “Move Y Dil” button.
xxv. Use compressed air to blow out dust from interior, metal facing under
blue cover, under the printer cover, and over any vents and fans on
the exterior.
xxvi. Replace blue cover and screw into place.
xxvii. Using Oxivir, clean all exterior surfaces.
xxviii. Remove buffers from chrome holder, remove chrome holder.
xxix. Rinse chrome holder under tap water, dry with paper towel, and
replace holder and buffers.
xxx. Clean sysmex racks with Oxivir, rinse, and replace broken adapter
rings.
b. For every 20K injections, a PM is performed by a Tosoh Service
representative. 10
4. Pipette Preventative Maintenance
a. Hamilton Autodilutor 500/600
i. Verify calibration of the device according to Autodilutor SOP.
ii. Instrument should be cleaned with disinfectant and inspected for
proper functioning daily.
b. Adjustable volume pipettes
i. After each use, the pipette should be wiped with disinfect with soaked
gauze.
ii. Pipettes are calibrated annually by trained field service personnel.
5. Materials:
a. Reagents—Supplied by Tosoh Bioscience (South San Francisco, CA). Part
numbers are subjected to change. Any questions or concerns about the
materials used for this assay please refer to your supervisor/delegee or call
Tosoh Scientific hotline at 1-800-248-6764, customer # 1208. Refer to
Operator’s Manual for additional details relating to the Tosoh G8 HPLC
instrument components and items used for testing.
b. Other Materials
i. Powder free nitrile exam gloves. (Fisher Scientific, Waltham, MA)
ii. Oxivir disinfectant. (AHP Technology, Sturtevant, WI)
iii. Gauze Sponges 4x4 not sterilized (Fisher Scientific, Waltham, MA)
iv. Kim Wipe lintless tissues. (Fisher Scientific, Waltham, MA).
6. Storage Requirements:
a. Unopened Elution Buffer 1, 2, and 3 are stable at room temperature until
the expiration date printed on the label. After opening, Elution Buffers are
stable for three months. Store at 4-30 ºC.
14 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
b. Unopened Hemolysis & Wash Solution is stable until the expiration date
printed on the label. After opening, Hemolysis & Wash Solution is stable
for three months. Store at 4-30ºC.
c. The unopened TSKgel G8 Variant HSi column should be stored at 4-15 °C
in a cool location away from direct sunlight. The column is stable until the
expiration date printed on the label. Replace column after 2500 injections.
d. Reagents must be brought to room temperature prior to use.
e. Reagent labeling: Reagents, calibrators, controls, and solutions should be
traceably identified to indicate the following:
i. Content and concentration
ii. Storage requirements (see above stability requirements for whole
blood).
iii. The below should be followed for working reagents; (open reagents
in use for testing):
1) Preparation date or opened date and the identity of the
preparer.
2) Tech’s initials.
3) Expiration date.
iv. Controls and calibrators obtained from freezer for daily
use should be initialed and dated with the pulled date.
v. Prior to processing controls and calibrators, sample cups
should be labeled (with the name of the control or
calibrator), dated, and initialed (tech responsible for
adding hemolysis wash to the sample cups).
vi. Processed controls and calibrators are immediately
returned to the refrigerator after samples are assayed
from the morning run, and are discarded after daily use.
vii. Buffers and wash should be labeled with the name of the
reagent and reagent source (name of vendor e.g. Tosoh),
preparation date or opened date, expiration date (from
open or preparation date), and tech’s initials.
15 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
16 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
17 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
18 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
ii. Insert tip of autodilutor into blood specimen and press the button on
top of the handle to draw 10ul of specimen into the tip.
iii. Insert tip into corresponding 1.5mL sample cup and press button
again to dispense sample and reagent into sample cup.
iv. Wipe outside of tip again with gauze wetted with distilled water.
v. Repeat procedure for all QCs and specimens not suitable for direct
sampling on the instrument from the primary tube.
b. Load calibrators.
Place the sample vials in the rack with PLC in position 1 (on the left) and
PHC in position 2.
c. Load controls: Controls are performed at the beginning of the run, every 50
samples, and at the end of a run.
d. Place controls in the rack with low in position 3 and high in position 4.
e. Place low and high controls in the end positions.
f. Load samples.
i. Mix each sample by gently inverting each capped sample tube.
ii. Place capped sample tubes in the rack in order from left to right. If
you are using barcoded tubes, verify that the labels face the
analyzer.
iii. Position each rack in the rack guide, starting with the right side.
iv. Place a blank rack after the last rack of samples. The blank rack
serves as an end marker.
3. Operation of the Tosoh Analyzer System.
a. Press the POWER button. (Already in standby when performing instrument
setup.)
b. Follow the calibration procedure in the CALIBRATION section of this
manual.
c. Press the START button.
d. Check chromatograms from normal and abnormal QC specimens in
positions 1, 2, 3 and 4.
e. When measurement ends, the analyzer washes the column by pumping
buffer for 15 minutes, and then enters STAND-BY mode.
f. After run is completed, verify and input, when necessary, accession
numbers for each individual specimen in the G8 Reporting Software under
the heading “barcode”.
g. Examine individual chromatograms.
i. Each chromatogram should include six peaks identified as A1A,
A1B, F, LA1C+, SA1C, and A0.
ii. All peaks should be clearly resolved.
iii. The acceptable retention time for SA1c is 0.57 – 0.61.1
iv. The acceptable retention time for A0 is 0.87 – 0.91.1
v. The acceptable range of TOTAL AREA is from 500 to 4000.
However, optimal results are obtained in the TOTAL AREA (TA)
range from 700 to 3000. Do not report results with a TA < 500 and >
4000. Low or high hematocrit samples may display Total Area above
or below the Total Area linear reportable range;
19 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
• Samples that exhibit TA < 500; reassay the specimen using less
Hemolysis & Wash Solution and centrifuge lightly to yield a TA
between 500 – 4000.1
• Samples with TA > 4000; reassay the specimen using more
Hemolysis & Wash Solution.1
vi. Repeat any specimens with %HbA1c values less than 4.0% or
greater than 14.0% for verification.
vii. Any samples which show a deterioration peak (extra peak between
the SA1C and AO peaks usually designated as P00 or P01) are to
be marked “X-PK”. Enter into the Primus data base the specimen ID,
and in the comment section “X-PK”. Also, record the specimen into
the Primus Sample Tracking from Tosoh G8 log and rerun by the
Trinity Premier boronate affinity method. Report the boronate affinity
HbA1c result.
viii. Samples with heterozygous HbD will exhibit an additional peak or
peaks after the A0 peak, and the instrument will designate the main
peak HV-0, and minor peaks Pxx. The chromatogram is labeled with
“HbAD”, and the %HbA1c cannot be reported for these specimens.
Enter into the Primus database the specimen ID, and in the comment
section “Hemoglobin D”. Also, record the specimen into the Primus
Sample Tracking from Tosoh G8 log and rerun by Trinity Premier
boronate affinity method. Report the boronate affinity HbA1c result.
ix. Samples with heterozygous HbS will exhibit an additional peak or
peaks after the A0 peak, and the instrument will designate the main
peak HV-1, and minor peaks Pxx. The chromatogram is labeled with
“HbAS”, and the %HbA1c cannot be reported for these specimens.
Enter into the Primus database the specimen ID, and in the comment
section “Hemoglobin S”. Also, record the specimen into the Primus
Sample Tracking from Tosoh G8 log and rerun by Trinity Premier
boronate affinity method. Report the boronate affinity HbA1c
result.
x. Samples with heterozygous HbC will exhibit an additional peak or
peaks after the A0 peak, and the instrument will designate the main
peak HV-2, and minor peaks Pxx. The chromatogram is labeled with
“HbAC”, and the %HbA1c cannot be reported for these specimens.
Enter into the Primus data base the specimen ID, and in the
comment section “Hemoglobin C”. Also, record the specimen into
the Primus Sample Tracking from Tosoh G8 log and rerun
by Trinity Premier boronate affinity method. Report the boronate
affinity HbA1c result.
xi. Samples with heterozygous HbE will exhibit an additional small
peak between the SA1C and A0 peaks. The instrument will
designate the HbE peak as PHV-E. The chromatogram is labeled
with “HbAE”, and the %HbA1c cannot be reported for these
specimens. Enter into the Primus data base the specimen ID, and
20 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
21 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
22 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
7. Supervisor Responsibility:
a. The supervisor or delegate ensures quality control passes within the
acceptable ranges prior to releasing patient results.
b. The supervisor or delegate checks every individual chromatogram to
ensure all peaks are resolved and reportable.
c. All chromatograms requiring further evaluation are noted and are not
uploaded into the database.
8. Procedure Notes
a. To avoid an error condition during calibration, be sure to place PLC and
PHC in the first sample rack in positions 1 and 2 respectively.
b. Each reagent lot number supplied by Tosoh is performance matched to
the supplied TSKgel G8 Hsi Columns. Following any announced change
in supplied Tosoh TSKgel G8 HSi Columns, contact Tosoh to determine
suitability of existing reagents.
c. The reagents must be at room temperature prior to use.
d. If the column is not to be used for more than one week, remove it from the
analyzer, seal the ends with the protective plugs and store in cool place at
4-15 ºC. Avoid direct sunlight.
e. The relationship between HbA1c results from NGSP network and the
IFCC network is expressed by using the following equation: NGSP (%) =
0.09148 x IFCC (mmol/mol) + 2.1522
f. Any changes to procedure must be documented. Major changes to the
SOP may include the way a procedure is performed or calculations and
requires the approval of the Medical Director. Minor changes include
typographical errors or other minor corrections that do not change the way
the procedure or calculation is performed and do not require approval of
the laboratory director. Major SOP changes must be reviewed by the Lab
Director prior to SOP update.
g. Any changes to the SOP will be communicated to technical staff via verbal
communication and email notification. Technical staff after reading the
changes made to the SOP will review, sign, and date the SOP.
Results are expressed on the report as % Hemoglobin A1c (HbA1c) and are
rounded to one decimal place. Results are reported throughout the entire
range of % HbA1c values verified by linearity studies. Results below 4.0% or
above 14.0% are reanalyzed for verification prior to results being reported.
Results outside of current linearity values are reported as < (low linearity
value) or > (high linearity value) as appropriate. Linearity is performed every
6 months. The established AMR is 3.1 – 19.5%.
23 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
1. Quality Control
a. Quality Control Preparation: Donors are recruited and compensated for
their donation of blood. Blood products are pooled together, mixed for at
least 30 min, and aliquoted under refrigerated conditions, see below.
i. Pooled Low Control
1) Single level low control was prepared from K2EDTA whole
blood whole blood were drawn from known non-diabetic
individuals (Normal level HbA1c).
2) The blood specimen were pooled, dispensed in 50 μL aliquots
into 400μL microtubes under refrigerated conditions. Batches
of low controls (aliquots) were assigned a lot number,
preparation date (date the controls were collected), and
stored at -70°C on the same day (working stock for daily
usage, no more than 50 tubes). The remaining aliquots were
assigned the same lot number/date and placed in a cryogenic
(liquid nitrogen) tank at -196°C in freezer boxes in the same
day (long term storage).
ii. Pooled High Control
1) The elevated (Abnormal) HbA1c level pooled whole blood
(K2EDTA) controls were purchased from Aalto Scientific.
Refer to Aalto Scientific product insert for additional details.
2) The blood was dispensed into 250 μL aliquots under
refrigerated conditions. Batches of high control, were
assigned a lot number, dated with the day the controls were
collected (prepared), and aliquots were stored at - 70°C on
the same day (working stock for daily usage, no more than 50
tubes). The remaining high control aliquots were assigned the
same lot number/date and were placed in a cryogenic (liquid
nitrogen) tank at -196°C in freezer boxes on the same day
(long term storage).
b. Preparation and stability.
i. Removed one box of calibrators from the liquid nitrogen phase to
the vapor nitrogen phase for at least 2 hours. After 2 hours, transfer
box into the -70°C freezer for at least 2 hours to allow materials to
acclimate. If box is already in the vapor nitrogen phase, move box
into the -70°C freezer and allow materials to acclimate.
ii. After the acclimation period, transfer 50 tubes (while working in the
freezer) into a box labeled the material’s name, date removed from
liquid nitrogen, and with an expiration date 365 days (one year) into
the future.
iii. Return the box with the remaining calibrator aliquots to the vapor
nitrogen phase of the liquid nitrogen.
iv. Low and high control aliquots are pulled from freezer daily and
placed in the refrigerator (2-8 °C) at start of the day. Prior to use,
thawed materials at room temperature, and mixed by inversion
gently.
24 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
25 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
26 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
27 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
1. Dilution studies demonstrate that the assay is linear from a Total Area of
500-4000.
2. For diagnostic purposes, the results obtained from this assay should be used
in conjunction with other data (for example, signs and symptoms, duration of
diabetes, results of other test, age of patient, clinical impressions, degree of
adherence to therapy, etc.).
3. The life span of red blood cells is shortened in patients with hemolytic
anemias, depending upon the severity of the anemia. As a consequence,
specimens from such patients may exhibit decreased HbA1c levels.1
4. The life span of red blood cells is lengthened in polycythemia or
postsplenectomy patients. Specimens from such patients may exhibit
increased HbA1c levels.1
5. Interference:
28 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
a. The presence of hemoglobin variants (e.g. HbC, HbF >38.4%, HbE, HbD,
HbS, etc.) may interfere with HbA1c results. 17
b. Refer to Operator’s manual Interference section for additional details.
Mean 5.0%
Range 4 - 6 % (equivalent of mean blood glucose of 60 - 120 mg/dL)
The normal range for the HbA1c test was established at the Diabetes Diagnostic
Laboratory in February 2000 based on 181 non-diabetic subjects collected from
the continental United States. Subjects were confirmed to have fasting blood
glucose less than 110 mg/dL (2000 standard for non-diabetic classification).
The mean HbA1c was 5.0%, with a 99% Confidence Interval of 4 to 6%.
Reference Range studies were repeated in 2004, 2009 and 2012 in the same
manner except that subjects were included only if their fasting glucose was less
than 100 mg/dL (Current ADA criteria for non-diabetic classification). In all
studies the original range of 4-6% was confirmed.
N/A
Any specimens not analyzed on the day of arrival in the laboratory are stored in
the refrigerator (2°C - 8°C). Upon completion of analysis, NHANES specimens
are frozen at -70°C and discarded after 1 year.
29 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
NHANES data files with results are exported from the NHANES database in the
specified format and uploaded to Westat via secure transfer weekly.
All shipments are recorded on the NHANES Shipping Log upon receipt. Actions
taken during the course of analysis, result reporting, and specimen retention are
also recorded on the log. Specimens are stored frozen at -70oC or colder after
analysis; specimen locations are recorded according to sequential DDL
accession number and box number. After one year specimens may be
discarded.
30 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
31 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
32 of 33
Glycohemoglobin (HbA1c)
NHANES 2019-2020
REFERENCES
(1) Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 Variant
Analysis Mode Operator’s Manual v. 3.0, Tosoh Bioscience, Inc. 2014 and
Training Manual.
(2) National Glycohemoglobin Standardization Program (NGSP) website:
http://www.ngsp.org
(3) Ultra 2 SOP/Premier SOP.
(4) Validation Study Binder for HbA1C.
(5) Rohlfing CL, Hanson S, Tennill AL, Little RR. Effects of whole blood
storage on hemoglobin a1c measurements with five current assay
methods. Diabetes Technol Ther. 2012 Mar;14(3):271-5. doi:
10.1089/dia.2011.0136. Epub 2011 Oct 27
(6) Little RR1, Rohlfing CL, Tennill AL, Connolly S, Hanson S. Effects of
Sample Storage Conditions on Glycated Hemoglobin Measurement:
Evaluation of Five Different High Performance Liquid Chromatography
Methods. Diabetes Technol Ther. 2007 Feb:9(1),36-42.
(7) Stability of whole blood at -70°C for measurement of hemoglobin A1c in
healthy individuals. Clin Chem 2004;50:2460-2461 Jones W, Scott J,
Leary S, Stratton F, Jones R; ALSPAC Study Team.
(8) HCCS Biorad Package Insert.
(9) Tosoh G8 Variant Analysis Mode Chromatogram Interpretative Guide.
(10) Verbal Communication - Tosoh Bioscience Field Service Engineer, Mark
Scheckel Tel: (650)-636-8350 Email: mark.scheckel@tosoh.com
(11) Little RR, Rohlfing CL, Hanson SE, Schmidt RL, Lin CN, Madsen RW,
Roberts WL. “The Effect of Increased Fetal Hemoglobin on 7 Common Hb
A1c Assay Methods.” Clin Chem. 2012 May 1: 58(5), 945.
(12) Shu I, Devaraj S, Hanson SE, Little RR, Wang P. Comparison of
hemoglobin A1c Measurements of samples with elevated fetal hemoglobin
by three commercial assays. Clin Chim Acta 2012: 413, 1712-13.
(13) Proficiency Testing Process (CAP) – Guideline version # 3. Paula Bullock,
approved 5/11/2016.
(14) Trinity Biotech HbA1c Carryover Testing Protocol-Jon Davis R&D
Technical Manager
(15) Tosoh Bioscience Carryover Testing Protocol- Eric Shaw, Clinical Support
Specialist.
(16) Autodilutor Carryover Procedure. Approved by Dr. Hammer on 2/15/2014,
Dr. Lay filed on 3/17/2015, and Dr. Little 6/21/2016.
(17) Rohlfing C, Hanson S, Weykamp C, Siebelder C, Higgins T, Molinaro R,
Yip PM, Little R. Effects of hemoglobin C,D,E and S traits on
measurements of hemoglobin A1c by twelve methods. Clin Chim Acta
2016: 455,80-83.
33 of 33