SAFETY PRECAUTIONS
DANGER: High voltages exist and are accessible in the Console Drive. Use extreme caution when
servicing internal components.
WARNINGS: Tubing breakage may result in fluid being sprayed from pump. Use appropriate measures to
protect operator and equipment.
Turn drive off before removing or installing tubing. Fingers or loose clothing could get caught
in drive mechanism.
CAUTION: Power must be turned off before connecting the external remote control cable to prevent
damage to the drive.
WARNING
CAUTION: To avoid electrical shock, the power cord protective grounding conductor must be connected
to ground. Not for operation in wet locations as defined by EN 61010-1.
Trademarks bearing the ® symbol in this publication are registered in the U.S. and in other countries.
i
INTRODUCTION
The Console Drive controls the speed of MASTERFLEX® I/P® Pump Heads to provide flow rates from 0.12 to
17 L/min, or 26 L/min with two pump heads.
CONTROL/DISPLAY FUNCTIONS
Press keys to activate function. Use UP/DOWN (▲, ▼) arrow keys to correct/change a flashing display.
Press any key to enter new values.
I. DISPENSE/COPY—Select dispensed volume, number of copies of a dispensed volume, or the dispense interval.
J. CAL CONTROL—Refine built-in calibration using a measured volume.
K. TUBING SELECT—Select tubing size.
1
SETUP AND DRIVE OPERATION
1. Mount pump head and load tubing. (See pump head manual.)
2. Turn pump on and select tubing size by pressing “TUBING SELECT.”
NOTE: If CAL LED is lit, that tubing size has been previously field calibrated. If LED is not lit, the drive is operating
with the built-in factory calibration. To clear a field calibration, press and hold the CAL key until the CAL
light goes out. This will take about 3 seconds. To recalibrate for better accuracy, see Calibration section.
CALIBRATION
Use only MASTERFLEX precision tubing with MASTERFLEX pumps to ensure optimum performance.
Use of other tubing may void applicable warranties.
1. Select correct tubing size and flow rate.
2. Press CAL; calibration volume appears.
3. Press START/STOP; the pump will use its stored memory to dispense the specified calibration sample quantity.
The pump will stop automatically.
4. Weigh/measure the sample.
5. Use UP/DOWN (▲, ▼) arrow keys to correct the flashing display.
NOTE: If the adjusted calibration is too great, “Err” will appear in the display. If this occurs, press the CAL control
and repeat the calibration procedure. The microprocessor will retain one special calibration value per
tubing size, even when power is turned off. The next calibration will replace the existing value.
2
DISPENSE/COPY
A first press of the DISP key results in the last entered dispense volume being displayed. The VOLUME annunciator
will illuminate and flash. The INC/DEC (▲, ▼) keys are used to change the dispense volume, if desired. The
START/STOP key then initiates delivery of the set volume. The amount remaining to be dispensed will be displayed
during countdown. Resetting cumulative dispensed volume can be done by pressing START/STOP while holding
TOTAL VOLUME. Display flashes “999999” when cumulative volume is greater than 999999. The dispense function
is exited by pressing any key except Increment (▲), Decrement (▼), DISP, or START/STOP.
A second press of the DISP key causes the COPY annunciator to illuminate and flash. The START/STOP key is then
used to deliver the desired volume without the need to know the volume in specific units. A third press of the DISP
key enters the volume dispensed. The COPY annunciator stops flashing. The START/STOP key is then used to
initiate delivery of the copied volume. The number of copies dispensed is displayed after each dispense. The
maximum number of copies is 999999. The START/STOP key is used to pause the copy dispense during
dispensing; copy dispense can then be continued using the START/STOP key.
A fourth press of the DISP key results in the last entered dispense interval being displayed. The SECONDS
annunciator will illuminate and flash. The INC/DEC (▲, ▼) keys are used to change the dispense time, if desired,
from 1 to 999999 seconds. The START/STOP key then initiates delivery for the set time interval. The remaining time
will be displayed during countdown. Pressing the DISP key a fifth time exits this mode.
REMOTE CONTROL
Selectable input (0–20 mA, 4–20 mA, 0–10V DC )
±0.5% linearity control
±50V common mode range
START/STOP; CW/CCW; PRIME via contact closure
CAUTION: Power must be turned off before connecting the external remote control cable to prevent
damage to the drive.
2. Connect the cable from the external remote control to the mating receptacle on the rear panel.
3. Select type of remote control input and output required as follows:
a. Press and hold the MODE control while turning the power switch to the on position. After two seconds,
release the MODE control. The initial display will show: “inP”. After two seconds, the display shows either
0–20 or 4–20.
NOTE: Press the UP (▲, increment) or DOWN (▼, decrement) arrow keys to select between 4–20 and 0–20 for
current loop control.
b. Press the MODE control again. The initial display will show: “out”. After two seconds, the display shows either
0–10 or 0–20 and 4–20.
NOTE: Press the UP (▲, increment) or DOWN (▼, decrement) arrow keys to select between 4–20 or 0–20 and
0–10 for current loop or voltage output.
4. Press the MODE control to select mode of operation. The LEDs indicate the selected mode. Select either mA or V.
NOTE: If using only remote START/STOP, PRIME and/or CW/CCW, the MODE control can be set to any of the
three positions.
3
5. To adjust the voltage or current scaling for other than zero to full scale, Press and hold the MODE key while
pressing the FLOW key. This display will show “LO” for 2 seconds and then a flow rate. Use the UP/DOWN (▲,
▼) arrow keys to set the minimum control level. Press the FLOW key again. The display will show “HI” for 2
seconds and then a flow rate. Use the UP/DOWN (▲, ▼) arrow keys to set the maximum control level. Press any
other key to exit. The same scaling will be used for both input and output levels.
NOTE: Tubing size should be
selected before adjusting the
voltage or current scaling.
77420-00
RED/YELLOW
14
NOTE: Colors are those of Remote Cable,
Catalog number 77300-32.
BLUE
6
GREEN
4
GREY
10
PRIME CW/CCW START/STOP
TACH OUTPUT
YELLOW TAN
8 12
WHITE PINK
3 11
INPUT MOTOR RUNNING
0-20mA; 4-20mA N.O. CONTACT
ORANGE RED/GREEN
5 13
INPUT MOTOR RUNNING
0-10V N.C. CONTACT
BLACK RED/BLACK
1 15
BROWN
7 16 N.C.
OUTPUT
0-20mA; 4-20mA
VIOLET
9 17 N.C.
OUTPUT
0-10V
RED
2 18 N.C.
4
TROUBLESHOOTING AND MAINTENANCE
Fuse Replacement
1. Place the power switch in the off position.
2. Disconnect the AC power input line cord from the receptacle.
3. Remove and check the fuse and replace if defective.
4. Reconnect the AC power input line cord to the receptacle.
C. External Receptacle
D. IEC 320 Power Cord
Troubleshooting
SYMPTOM CAUSE REMEDY
A. Motor does not rotate. Display A. No power. 1. Check fuse and replace if
does not light. defective.
2. Check that unit is plugged into a
live line.
3. Check connection of power cord.
4. Check the line cord for continuity
and replace if defective.
5. Return for servicing.
B. Motor does not rotate. Display B1. Defective remote control. 1. Place power switch in off
lights. position.
2. Check that remote cable
connector is inserted fully into
the receptacle.
3. If motor still does not rotate,
select INTERNAL with the
MODE control and press the
START/STOP control.
4. If the motor rotates, replace the
remote control with similar unit. If
motor does not rotate, return
drive for servicing.
5
Troubleshooting (cont.)
If an error message is displayed, refer to the following list for possible corrective actions you can take. If these do
not correct the problem, contact your dealer.
6
Cleaning
Keep the drive enclosure clean with mild detergents. Do not immerse.
SPECIFICATIONS
Output:
Speed: 1 to 650 r/min
Torque output, Maximum:
Continuous 580 oz-in (41.8 kg•cm)
Start-up 1750 oz-in (126 kg•cm)
Speed regulation:
Line ±0.25% F.S.
Load ±0.25% F.S.
Drift ±0.25% F.S.
Display: Six-digit, seven-segment LED
Remote outputs: Voltage speed output (0–10V DC @ 1 kΩ min)
Current speed output (0–20 mA or 4–20 mA @ 0–600 Ω)
Tach output (TTL, 10 to 6500 Hz, 50% duty cycle 10 Hz/rpm)
Motor running output
(N.O. & N.C. contact closure, 1A @ 28V AC/DC)
Input:
Supply voltage limits: Dual voltage—Automatically selected
90 to 130 Vrms @ 50/60 Hz, or 200 to 260 Vrms @ 50/60 Hz
Current, max.: 4.5A @ 115 Vrms, or 2.5A @ 230 Vrms
Remote Inputs: START/STOP, CW/CCW, PRIME (Contact closure)
Voltage input (0–10V DC @ 10 kΩ)
Current input (0–20 mA or 4–20 mA @ 250 Ω)
Construction:
Dimensions (L 3 W 3 H): 18 3/8 in 3 11 in 3 17 in (467 3 279 3 432 mm)
Weight: 70 lb (31.7 kg)
Enclosure Rating: IP 66 per IEC 529/NEMA 4v (Indoor use)
Environment:
Temperature, Operating: 0 to 40 C (32° to 104°F)
Temperature, Storage: –25° to 65 C (–13° to 149°F)
Humidity (non-condensing): 10% to 100%
Altitude: Less than 2000 m
Pollution Degree: Pollution Degree 3
(Indoor use—sheltered locations)
Chemical Resistance: Enclosure is 316 stainless steel
Compliance: UL508C, CSA C22.2, No. 14
(For CE Mark):
EN61010-1 (EU Low Voltage Directive) and
EN61326 (EU EMC Directive)
7
WARRANTY
Use only MASTERFLEX precision tubing with MASTERFLEX pumps to ensure optimum performance. Use of
other tubing may void applicable warranties.
The Manufacturer warrants this product to be free from significant deviations from published specifications. If repair
or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to
misuse or abuse on your part, as determined by the Manufacturer. Repair costs outside the warranty period, or those
resulting from product misuse or abuse, may be invoiced to you.
The warranty period for this product is noted on the Warranty Card.
PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer for authorization and shipping instructions before
returning the product, either within or outside of the warranty period. When returning the product, please state the
reason for the return. For your protection, pack the product carefully and insure it against possible damage or loss.
Any damages resulting from improper packaging are your responsibility.
TECHNICAL ASSISTANCE
If you have any questions about the use of this product, contact the Manufacturer or authorized seller.
®
Cole
Parmer