El Toxo G.
El Toxo G.
El Toxo G.
AccuDiag™ The organism is one of the most common latent infectious agents of man through out
the world. 6
Toxoplasma gondii IgG (Toxo IgG)
The Diagnostic Automation, Inc. Toxoplasma gondii IgG ELISA kit provides all the
ELISA Kit necessary reagents for the rapid quantitation of Toxoplasma gondiiIgG antibody in
human sera.
2. Samples that remain equivocal after repeat testing should be retested on an QUALITY CONTROL
alter nate method, e.g., immunofluorescence assay (IFA). If results remain
equivocal upon further testing, an additional sample should be taken. (See For the assay to be considered valid the following conditions must be met:
Limitation No. 4). 1. Calibrator and Controls must be run with each test run.
3. In the evaluation of paired sera, if the acute specimen is negative and the 2. Reagent blank (when read against air blank) must be < 0.150 Absorbance (A) at
convalescent specimen is positive, a seroconversion has taken place. This 450 nm.
indicates a significant change in antibody level and the patient is 3. Negative Control must be < 0.250 A at 450 nm (when read against reagent
undergoing a primary infection. blank).Each Calibrator must be > 0.250 A at 450 nm (when read against reagent
4. To evaluate paired sera for a significant change in antibody level or blank).
seroconversion, both samples must be tested in duplicate in the same assay. 4. Positive Control must be > 0.500 A at 450 nm (when read against reagent blank).
The mean ISR of both samples (acute and onvalescent) must be greater 5. The ISR(Immune Status Radio) Values for the Positive and Negative Control
than 1.00 to evaluate the paired sera for significant rise in antibody level. should be in their respective ranges printed on the vials. If the Control values are
5. Additional Quality Control for Paired Sera: (See NOTE under Assay not within their respective ranges, the test should be considered invalid and should
Procedure). As a check for acceptable reproducibility of both the acute be repeated.
sera (tested in duplicate) and the convalescent sera (tested in duplicate), 6. Additional Controls may be tested according to guidelines, or requirements of
the following criteria must be met for valid results: local, state, and/or federal regulations or accrediting organizations
7. Refer to NCCLS C24-A for guidance on appropriate QC practices. 19
8. If above criteria are not met upon repeat testing, contact Diagnostic Automation,
Inc. Technical Services.
ISO 13485
ISO 9001
Diagnostic Automation
/Cortez Diagnostics, Inc.
21250 Califa Street , Suite 102 and 116,
Woodland Hills, California 91367 USA