S Bio Cabinet
S Bio Cabinet
S Bio Cabinet
(Revised 1/24)
BIOSAFETY CABINETS
Scope
This SOP describes the classification and design of biological safety cabinets. Guidance on
operation and use procedures of biosafety cabinets is provided in the EHS SOP Working in a
Biosafety Cabinet.
References
Information in this SOP is derived from the following nationally recognized guidelines and
standards:
• Biosafety in Microbiological and Biomedical Laboratories, (BMBL, 6th Ed., Centers
for Disease Control and Prevention, National Institutes of Health
• NSF/ ANSI 49 – 2022 Ed., Biosafety Cabinetry: Design, Construction, Performance, and
Field Certification
NOTE: All work with RG-2 and RG-3 organisms and recombinant or synthetic
nucleic acids requires UNL IBC review and approval prior to initiation of
experiments. Consult the UNL Biosafety Guidelines for additional information.
The inward face velocities of Class I and II BSCs (75-100 linear feet per minute) provide
comparable levels of containment to protect laboratory workers and the immediate
environment from infectious aerosols generated within the cabinet. Class II BSCs also provide
product protection through high-efficiency particulate air filtration (HEPA filtration) of the intake
air. Class III cabinets offer the maximum protection to laboratory personnel, the community,
(Created 11/00; Revised 5/08, 2/10, 9/14, 4/15, 12/18, 5/20) 1
and the environment because all hazardous materials are contained in a completely enclosed,
ventilated cabinet.
All classes of BSCs are equipped with exhaust HEPA filters, which distinguishes them from
laminar flow “clean benches.” Additionally, Class II and III BSCs along with “clean benches”
protect the products/cultures within from environmental contamination by using a supply air
HEPA filter. In contrast, Class I BSCs are not equipped with supply HEPA filters and as such
provide no product protection. Design differences between various classes of biosafety
cabinets are summarized in Table 1 below.
NOTE: A “clean bench,” although providing either vertical or horizontal laminar
airflow, is not a BSC and users should be aware of the limitations of these devices.
Their design and use is discussed later in this document (Figures 9, 10).
Class I BSC
Figure 1 Class I Biological Safety Cabinet
(Note: Class I BSCs are manufactured on a
limited basis. Most have been replaced
with Class II BSCs.)
The Class I BSC provides personnel and
environmental protection, but no product
protection. Air movement is similar to a
chemical fume hood, but it has a HEPA
filter in the exhaust system to protect the
environment (Figure 1) from bioaerosols.
In the Class I BSC, unfiltered room air is
drawn across the work surface. Personnel
protection is provided by this inward airflow
as long as a minimum velocity of 75 linear
feet per minute (lfpm) is maintained
through the front opening and appropriate
work practices are observed.
The classical Class I BSC is hard-ducted
(i.e., direct connection) to the building
exhaust system, and the building A. front opening, B. sash, C. exhaust HEPA filter,
exhaust fan provides the negative D. blower.
pressure necessary to draw room air Note: The cabinet needs to be hard connected to
into the cabinet. the building exhaust system if toxic vapors are to
be used.
Photo courtesy NuAire, Inc. Laboratory Equipment Supply
Class II BSC
Class II BSCs are partial barrier systems that rely on laminar movement of air to provide
containment. If the air curtain is disrupted (e.g., movement of materials in and out of a cabinet,
rapid or sweeping movement of the arms) potential for contaminant release into the laboratory
work environment is increased as is the risk of product contamination.
Class II (Types A1, A2, B1, B2 and C1) 1 BSCs provide personnel, environment, and product
protection. Airflow is drawn into the front grille of the cabinet, providing personnel protection. In
addition, the downward laminar flow of HEPA-filtered air provides product protection by
minimizing the chance of cross-contamination across the work surface of the cabinet. Because
cabinet exhaust air is passed through a HEPA filter, it is particulate-free (environmental
protection), and may be recirculated to the laboratory (Type A1 and A2 BSCs) (Figures 2, 3) or
discharged from the building via a canopy connection (Figure 4). Exhaust air from Types B1
and B2 BSCs (Figures 5, 6) must be discharged to the outdoors via a hard connection.
An internal blower (Figure 2) draws sufficient Figure 2
room air through the front grille to maintain a Class II Type A1 Biological Safety Cabinet
minimum average inflow velocity of at least
75 lfpm at the face opening of the cabinet.
The supply air flows through a HEPA filter
and provides particulate-free air to the work
surface. Laminar (unidirectional, constant
velocity) airflow reduces turbulence in the
work zone and minimizes the potential for
cross-contamination.
The downward moving air "splits" as it
approaches the work surface; the blower
draws part of the air to the front grille and
the rest is drawn to the rear grille. Although
there are variations among different
cabinets, this split generally occurs about
halfway between the front and rear grilles
and 2-6 inches above the work surface.
It is possible to exhaust the air from a Type
A1 or A2 cabinet outside of the building.
However, it must be done so as not to alter
the balance of the cabinet exhaust system,
and thereby disturbing the internal
A. front opening, B. sash, C. exhaust HEPA filter,
cabinet airflow. The proper method of D. supply HEPA filter, E. common plenum, F. blower.
connecting a Type A1 or A2 cabinet to
the building exhaust system is through use of a canopy hood,1 which provides a small opening
(usually 1 inch) around the cabinet exhaust filter housing (Figure 4). The current NSF/ANSI 49
Standard requires Class II Type A1 or A2 cabinets with canopy connections be fitted with an
audible and visible exhaust alarm to indicate when air flow is being returned to the room rather
than being exhausted out. Annual certification of cabinets with these “thimble connections”
must include alarm function certification.
Figure 3 Figure 4 Canopy (thimble) unit for ducting a Class II,
Type A BSC2
Photo courtesy
NuAire, Inc. Laboratory Equipment Supply
In a Class II, Type B1 cabinet, the supply blowers in the cabinet draw room air (plus a
percentage of the cabinet’s recirculated air) through the front grille. This air is taken up through
plenums in the side of the cabinet, 30% is forced through the supply HEPA filter, and 70% is
drawn through the exhaust HEPA by the building exhaust system. Approximately 70% of the
down flow supply air is drawn into the rear grille and is directly exhausted through a HEPA to
the building exhaust. Since air near the rear of the cabinet is directly exhausted and not
recirculated, small quantities of volatile chemicals may be worked with in the cabinet, but
should be confined to the rear 50% of the cabinet.
Figure 5 Figure 6
Photo courtesy
Photo courtesy
NuAire, Inc. Laboratory Equipment Supply
NuAire, Inc. Laboratory Equipment Supply
Note: The cabinet exhaust needs to be hard Note: The cabinet exhaust needs to be hard
connected to the building exhaust system. connected to the building exhaust system.
Class II, Type B2, BSCs (Figure 6) are total exhaust cabinets; no air is recirculated. This
cabinet provides both biological and chemical containment. Consideration must be given to the
chemicals used in BSCs since some chemicals can destroy the filter medium, housings, and/or
gaskets causing loss of containment. The supply blower draws either room or outside air in at
the top of the cabinet, passes it through a HEPA filter and down into the work area of the
cabinet. The building exhaust system draws air through both the rear and front grills producing
an inflow face velocity of 100 lfpm. All air entering this cabinet is exhausted through a HEPA
filter.
A. glove ports with O-ring for attaching arm-length gloves to cabinet, B. sash, C. exhaust
HEPA filter, D. supply HEPA filter, E. double-ended autoclave or pass-through box.
Note: A chemical dunk tank may be installed which would be located beneath the work
surface of the BSC with access from above. The cabinet exhaust needs to be hard
connected to an independent dedicated exhaust system. The exhaust air must be double
HEPA filtered or HEPA filtered and incinerated.
Long, heavy-duty rubber gloves are attached in a gas-tight manner to ports in the cabinet and
allow direct manipulation of the materials isolated inside. Due to the design of the cabinet,
airflow can be turbulent within the cabinet and laminar airflow is not considered a characteristic
of a Class III BSC.
Figure 9 Horizontal laminar flow “clean Figure 10 Vertical laminar flow “clean
bench” bench”
Photo courtesy
Photo courtesy
NuAire, Inc. Laboratory Equipment Supply
NuAire, Inc. Laboratory Equipment Supply
A. front opening, B. supply grille, C. supply
HEPA filter, D. supply plenum, E. blower, F. A. front opening, B. sash, C. supply HEPA
grille. filter, D. blower.
All BSCs should be placed in a laboratory at a location that provides a minimum of:
6 inches from adjacent walls or columns
6 inches between two BSCs
6 inches space between both sides of the cabinet and 6 inches behind the BSC to allow
for service operations
40 inches of open space in front of the BSC
60 inches from opposing walls, bench tops and areas of occasional traffic
20 inches between BSC and bench tops along a perpendicular wall
100 inches between two BSCs facing each other
60 inches from behind a doorway
40 inches from an adjacent doorway swing side and
6 inches from an adjacent doorway hinge side
1NSF International (NSF); American National Standards Institute (ANSI). (2022). Biosafety Cabinetry: Design,
Construction, Performance, and Field Certification (NSF/ANSI Standard 49-2022).
2Centers for Disease Control and Prevention, N. I., Biosafety in Microbiological and Biomedical Laboratories (6th
edition). Washington, DC: U.S. Government Printing Office.