Cobas c303
Cobas c303
Cobas c303
MDN2
ONLINE DAT Methadone II
Order information
Analyzer(s) on which cobas c pack(s) can be used
08058008190 ONLINE DAT Methadone II (850 tests) System‑ID 2088 001 cobas c 303, cobas c 503
03304671190 Preciset DAT Plus I calibrator CAL 5 Code 20435
07978766190 Serum DAT Control Low (ACQ Partner Channel*)
07978740190 Serum DAT Control High (ACQ Partner Channel*)
08063494190 NaCl Diluent 9 % (123 mL) System‑ID 2906 001
*Roche does not hold the product registration for Partner Channels. The legal manufacturer indicated on the kit is solely responsible for all of the design,
legal, and regulatory aspects of the product.
MDN2
ONLINE DAT Methadone II
Materials required (but not provided) Values obtained should fall within the defined limits. Each laboratory should
See “Order information” section establish corrective measures to be taken if values fall outside the defined
limits.
General laboratory equipment
Follow the applicable government regulations and local guidelines for
Assay quality control.
For optimum performance of the assay follow the directions given in this
document for the analyzer concerned. Refer to the appropriate operator’s Results
manual for analyzer‑specific assay instructions. The cutoff calibrator is used as a reference in distinguishing between
preliminary positive and negative samples. Samples producing a positive or
The performance of applications not validated by Roche is not warranted “0” absorbance value are considered preliminary positive. Preliminary
and must be defined by the user. positive samples are flagged with >Test. Samples producing a negative
Application for serum and plasma absorbance value are considered negative. Negative samples are preceded
by a minus sign.
Test definition Preliminary positive results should be confirmed by another method.
Qualitative Limitations - interference
Reporting time 10 min Criterion: No cross‑over at initial values of samples of 150 ng/mL and
450 ng/mL (control levels).
Wavelength (sub/main) – /546 nm
See the "Specific performance data" section of this document for
Reagent pipetting information on substances tested with this assay. There is the possibility
R1 59 µL that other substances and/or factors may interfere with the test and cause
erroneous results (e.g., technical or procedural errors).
R2 26 µL A preliminary positive result with this assay indicates the presence of
Sample volumes Sample methadone and/or its metabolites in serum. It does not measure the level of
intoxication.
300 ng/mL cutoff
Icterus:12 No significant interference up to an I index of 60 for conjugated
Normal 2.3 µL and unconjugated bilirubin (approximate conjugated and unconjugated
bilirubin concentration: 1026 µmol/L or 60 mg/dL).
Decreased 2.3 µL
Hemolysis:12 No significant interference up to an H index of 1000
Increased 2.3 µL (approximate hemoglobin concentration: 622 µmol/L or 1000 mg/dL).
For further information about the assay test definitions refer to the Lipemia (native lipaemic samples):12 No significant interference up to an
application parameters setting screen of the corresponding analyzer and L index of 100. There is poor correlation between the L index (corresponds
assay. to turbidity) and triglycerides concentration.
Calibration Rheumatoid factors: No significant interference from rheumatoid factors up
to a concentration of 1200 IU/mL.
Calibrators Qualitative application Immunoglobulins: No significant interference from immunoglobulins up to a
300 ng/mL cutoff assay concentration of 16 g/L (simulated by human immunoglobulin A), up to a
concentration of 70 g/L (simulated by human immunoglobulin G) and up to
S1: Preciset DAT Plus I calibrator ‑ CAL 5, a concentration of 10 g/L (simulated by human immunoglobulin M).
2000 ng/mL with automatic pre‑dilution Albumin: No significant interference from human serum albumin up to a
The drug concentration of the calibrator has been concentration of 70 g/L.
verified by GC‑MS. As with any assay employing mouse antibodies, the possibility exists for
interference by human anti‑mouse antibodies (HAMA) in the sample, which
Calibration K factor For the qualitative application a K factor of -1000 is could cause falsely lowered results.
predefined in the application settings.
For diagnostic purposes, the results should always be assessed in
Calibration mode Qualitative application conjunction with the patient’s medical history, clinical examination and other
findings.
Linear
ACTION REQUIRED
Calibration frequency Full calibration Special Wash Programming: The use of special wash steps is mandatory
- after reagent lot change when certain test combinations are run together on cobas c systems. All
- every 84 days special wash programming necessary for avoiding carry‑over is available
- as required following quality control procedures via the cobas link. The latest version of the carry‑over evasion list can be
found with the NaOHD/SMS/SCCS Method Sheet for information. For
For the cutoff calibrator a value of "0" is encoded in the e‑barcode in order further instructions refer to the operator’s manual.
to ensure flagging of positive samples with >Test and negative absorbance
values for negative samples. Expected values
Calibration interval may be extended based on acceptable verification of Qualitative assay
calibration by the laboratory. Results of this assay distinguish preliminary positive (≥ 300 ng/mL) from
Traceability: This method has been standardized against a primary negative samples only. The amount of drug detected in a preliminary
reference method (GC‑MS). positive sample cannot be estimated.
MDN2
ONLINE DAT Methadone II
(2 aliquots per run, 2 runs per day, 21 days). Results for repeatability and Samples were tested and the following results were obtained on a cobas c
intermediate precision were obtained on the cobas c 503 analyzer. 501 analyzer.
Qualitative precision - 300 ng/mL Compound Comp. conc. Neg. Pos.
Cutoff (300) Number Correct Confidence level mg/L level level
tested results Acetaminophen 200 neg pos
Serum -75 % 84 84 > 95 % negative reading Acetylcysteine 1660 neg pos
ACQ‑L 84 84 > 95 % negative reading Acetylsalicylic acid 1000 neg pos
Cutoff serum 84 n.a.* n.a.* Amitriptyline 1.00 neg pos
ACQ‑H 84 84 > 95 % negative reading Ampicillin‑Na 1000 neg pos
Serum +75 % 84 84 > 95 % negative reading Ascorbic acid 300 neg pos
*n.a. = not applicable Caffeine 59.8 neg pos
Accuracy Cefoxitin 2500 neg pos
103 serum samples, screened negative for methadone on a cobas c 501 Cyclosporine 5.00 neg pos
analyzer were evaluated with the Methadone II assay on a cobas c 503
analyzer. 100 % of these normal serum samples were negative with the d‑Amphetamine 1.36 neg pos
Methadone II assay on a cobas c 503 analyzer. 51 serum samples
screened positive for methadone relative to the 300 ng/mL cutoff on a Diazepam 5.13 neg pos
cobas c 501 analyzer were evaluated with the Methadone II assay on a Doxycycline 50.0 neg pos
cobas c 503 analyzer. At the 300 ng/mL cutoff, 100 % of the samples were
positive on both the cobas c 501 analyzer and the cobas c 503 analyzer. d‑Pseudoephedrine 9.98 neg pos
Erythromycin 59.9 neg pos
Methadone II correlation (cutoff = 300 ng/mL)
Fenoprofen 195 neg pos
cobas c 501 analyzer
Furosemide 59.9 neg pos
+ -
Gentisic acid 18.0 neg pos
cobas c 503 analyzer + 51 0
Heparin 5000 U/L neg pos
- 0 103
Hydrochlorothiazide 6.02 neg pos
52 serum samples, screened negative for methadone on a cobas c 501
analyzer were evaluated with the Methadone II assay on a cobas c 303 l‑Amphetamine 1.00 neg pos
analyzer. 100 % of these normal serum samples were negative with the Ibuprofen 500 neg pos
Methadone II assay on a cobas c 303 analyzer. 52 serum samples
screened positive for methadone relative to the 300 ng/mL cutoff on a Imipramine 0.70 neg pos
cobas c 501 analyzer were evaluated with the Methadone II assay on a Ketamine 10.0 neg pos
cobas c 303 analyzer. At the 300 ng/mL cutoff, 100 % of the samples were
positive on both the cobas c 501 analyzer and the cobas c 303 analyzer. Levodopa 20.0 neg pos
Methadone II correlation (cutoff = 300 ng/mL) Lidocaine 12.0 neg pos
cobas c 501 analyzer Methyldopa + 1.5 H2O 20.0 neg pos
+ - Metronidazole 200 neg pos
cobas c 303 analyzer + 52 0 Morphine 0.50 neg pos
- 0 52 Naproxen 499 neg pos
Phenylbutazone 400 neg pos
Analytical specificity
The specificity of this assay for structurally similar compounds was Procaine 39.9 neg pos
determined by generating inhibition curves for each of the compounds listed Promethazine 1.20 neg pos
and determining the approximate quantity of each compound that is
equivalent in assay reactivity to a 300 ng/mL assay cutoff. Caution should Quinidine 12.0 neg pos
be taken when interpreting results of patient samples containing structurally Quinine 48.0 neg pos
related compounds having greater than 0.5 % cross‑reactivity. The
following results were obtained on a cobas c 501 analyzer. Rifampicin 60.0 neg pos
Compound ng/mL Approximate % Tetracycline 15.1 neg pos
Equivalent to cross‑reactivity Theophylline 100 neg pos
300 ng/mL Trifluoperazine 1.00 neg pos
methadone
In very rare cases, gammopathy, in particular type IgM (Waldenström’s
Chlorpromazine 70752 0.42 macroglobulinemia), may cause unreliable results.13
Metabolite Lu AA34443 6341 4.73 References
Methadone 308 97.3 1 Karch SB, ed. Drug Abuse Handbook. Boca Raton, FL: CRC Press
LLC 1998.
Vortioxetine 8344 3.60
2 Council Reports: Treatment of morphine-type dependence by
Drug interference withdrawal methods. JAMA 1972;219(12):1611-1615.
Interfering substances were added to serum containing methadone at 3 Smialek JE, Monforte JR, Aronow R, et al. Methadone deaths in
‑50 % and +50 % of the cutoff level at the concentration listed below. children: A continuing problem JAMA 1977;238(23):2516-2517.
MDN2
ONLINE DAT Methadone II
ABUSCREEN, COBAS, COBAS C, ONLINE DAT and PRECISET are trademarks of Roche.
All other product names and trademarks are the property of their respective owners.
Additions, deletions or changes are indicated by a change bar in the margin.
© 2021, Roche Diagnostics