Model1700 Rev6
Model1700 Rev6
Model1700 Rev6
1 Cover Page
MODEL 1700
LIQUID LEVEL INSTRUMENT
Safety Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .x
Equipment Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xi
Introduction ...............................................................................1
Model 1700 Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Cryogenic (Capacitance-Based) Liquid Level Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Superconducting Level Sensors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Digitally-Controlled . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
System Flexibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Intuitive Human-Interface Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Measurement Flexibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Real Time Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Valve Control Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Analog Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Signal Relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Connectivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
REV 6 I
AMERICAN MAGNETICS, INC. TABLE OF CONTENTS
Installation............................................................................... 11
Unpacking and Inspecting the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Power Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Operation................................................................................ 23
Energizing the Model 1700 Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Screen Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Home Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Home Screen Footer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Editing a Field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
ii REV 6
AMERICAN MAGNETICS, INC. TABLE OF CONTENTS
Level-Based Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Time-Based Fill Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Multiple alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Configuring Alarm Setpoints. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Acknowledging an Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Muting an Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Configuring Relay Setpoints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Damping Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Serial Connectivity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Configuring the
RS-232 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Calibration ...............................................................................53
Setting the System Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
REV 6 III
AMERICAN MAGNETICS, INC. TABLE OF CONTENTS
Approximate Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
RS-232 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Serial port connector and cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Command/return termination characters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Ethernet Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Ethernet Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Termination Characters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Port Assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Command Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
System Related Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Display Configuration Commands and Queries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Relay Configuration Commands and Queries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Alarm Configuration Commands and Queries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Measurement Commands and Queries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Fill Control and Queries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
HE Channel Sampling Commands and Queries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Assignment Commands and Queries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
N2 Channel Calibration Commands and Queries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
HE Channel Calibration Commands and Queries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Remote Units Commands and Queries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
iv REV 6
AMERICAN MAGNETICS, INC. TABLE OF CONTENTS
Appendix................................................................................ 109
Serial (RS-232) Connector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .109
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Instrument Displays “LOSS OF SENSOR” Condition for LN2 Level Measurement . . . . . . . . . . . . . 114
Instrument Displays “LOSS OF SENSOR” Condition for LHe Level Measurement . . . . . . . . . . . . . 114
Instrument Displays “SENSOR SHORTED” Condition for LN2 Level Measurement . . . . . . . . . . . . 114
Instrument Displays “SENSOR PROTECTED” Condition for LHe Measurement . . . . . . . . . . . . . . 115
System Test Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
System Logs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Glossary .................................................................................123
Abbreviations and Acronyms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Index ......................................................................................127
REV 6 V
AMERICAN MAGNETICS, INC. TABLE OF CONTENTS
vi REV 6
AMERICAN MAGNETICS, INC. FOREWORD
FOREWORD
This manual contains the operation and maintenance instructions for the
American Magnetics, Inc. Model 1700 Liquid Level Control Instrument
and outlines applications for various system configurations. Since it is
not possible to cover all equipment combinations for all magnet systems,
only the most common configurations are discussed. The user is
encouraged to contact an authorized AMI Technical Support
Representative for information regarding specific configurations not
explicitly covered in this manual. This manual refers to the instrument as
configured for both liquid nitrogen and liquid helium operation.
REV 6 VII
AMERICAN MAGNETICS, INC. FOREWORD : GENERAL PRECAUTIONS
GENERAL PRECAUTIONS
Cryogen Safety
Cryogenic liquids, due to their extremely low temperatures, will also burn
the skin in a similar manner as would hot liquids. Never permit cryogenic
liquids to come into contact with the skin or allow liquid nitrogen to soak
clothing. Serious burns may result from careless handling. Never touch
uninsulated pipes or vessels containing cryogenic liquids. Flesh will stick
to extremely cold materials. Even nonmetallic materials are dangerous
to touch at low temperatures. The vapors expelled during the venting
VIII REV 6
AMERICAN MAGNETICS, INC. FOREWORD : GENERAL PRECAUTIONS
1. If any cryogenic liquid contacts the skin or eyes, immediately flush the
affected area gently with tepid water (102°F 105°F, 38.9°C 40.5°C) and
then apply cold compresses.
2. Do not apply heat. Loosen any clothing that may restrict circulation. Apply a
sterile protective dressing to the affected area.
3. If the skin is blistered or there is any chance that the eyes have been
affected, get the patient immediately to a physician for treatment.
Containers of cryogenic liquids are self pressurizing (as the liquid boils
off, vapor pressure increases). Hoses or lines used to transfer these
liquids should never be sealed at both ends (i.e. by closing valves at
both ends).
REV 6 IX
AMERICAN MAGNETICS, INC. FOREWORD : CRYOGEN SAFETY SUMMARY
SAFETY LEGEND
X REV 6
AMERICAN MAGNETICS, INC. FOREWORD : EQUIPMENT WARNINGS
EQUIPMENT WARNINGS
In the event that the ground path becomes less than sufficient to carry
the rated current of the power circuit, the equipment should be
disconnected from power, labeled as unsafe, and removed from place of
operation.
REV 6 XI
AMERICAN MAGNETICS, INC. FOREWORD : INSTRUMENT CONFIGURATION
cryogenic liquid, with the exception of liquid helium (LHe) which requires
a superconductor-based sensor.
INSTRUMENT CONFIGURATION
1700-A-B-C-D-E where
Tbl = tabletop
DRL = dual rack mount, 19” wide rack standard, instrument on left
side
XII REV 6
AMERICAN MAGNETICS, INC. FOREWORD : INSTRUMENT CONFIGURATION
N = North American
E = European, Schuko
A = Australian/New Zealand
C = Chinese
U = United Kingdom
P = India/pigtailed
N = not configured
The sensor active length and length units are appended to the
configuration code, ie ...CAP-10.4CM-...
N = not configured
The sensor active length and length units are appended to the
configuration code, ie ...He2K-40IN-...
C = instrument modified.
REV 6 XIII
AMERICAN MAGNETICS, INC. FOREWORD : INSTRUMENT CONFIGURATION
XIV REV 6
AMERICAN MAGNETICS, INC. INTRODUCTION
INTRODUCTION
The Model 1700 instrument will measure liquid helium level using a
superconducting level sensor. The instrument will be factory configured
for either 4.2K or 2K liquid helium level sensors at time of purchase.
REV 6 1
AMERICAN MAGNETICS, INC. INTRODUCTION : MODEL 1700 INSTRUMENT
2 REV 6
AMERICAN MAGNETICS, INC. INTRODUCTION : MODEL 1700 INSTRUMENT
OSCILLATOR/TRANSMITTER
DIGITALLY- The Model 1700 contains a microcomputer which controls analog data
conversion, display/keypad functions, communications I/O, dry contact
CONTROLLED closures, generation of analog output signals and relay control of a
mains power outlet for solenoid valve autofill applications.
SYSTEM FLEXIBILITY The Model 1700 instrument incorporates data converters to translate
signals between the analog and digital domains. Precision
instrumentation techniques and potentiometer-free designs are
employed throughout the Model 1700 Instrument to ensure long term
stability and accurate signal translation for a wide range of conditions.
REV 6 3
AMERICAN MAGNETICS, INC. INTRODUCTION : MODEL 1700 INSTRUMENT
DISPLAY The Model 1700 Instrument has a 4.3” diagonal measure TFT (Thin Film
Transistor) color liquid crystal display of 480 x 272 pixels. The display
has a 4-wire resistive touch overlay for easy operator input.
INTUITIVE HUMAN- The Model 1700 instrument is designed to simplify the touch-screen
based user interface. All functions were analyzed and subsequently
INTERFACE DESIGN programmed so that the most commonly used functions are addressed
with the least number of keystrokes. The menus are presented in a
logical fashion so that the operation of the Model 1700 is intuitive to the
user. Context-sensitive Help screens are also provided.
MEASUREMENT Depending on the instrument version purchased, the Model 1700 can be
configured to monitor and display one capacitance-based liquid level
FLEXIBILITY sensor (LN2 version) or one liquid helium level sensor (LHe version) or
both simultaneously. Set points can be assigned to control two dry-
contact relay outputs. The make or break function of each relay can be
set independently. A solid-state relay allows mains power to be switched
according to set points to operate a solenoid-operated valve or other
load.
REAL TIME CLOCK The Model 1700 Instrument incorporates a real time clock, which can be
manually set or synchronized via NTP, with support for worldwide
timezones including automatic DST adjustment.
VALVE CONTROL The Model 1700 Instrument has a switched 2 ampere at mains voltage
output for energizing a solenoid operated flow valve or other loads. This
OUTPUT output is controlled by a zero-crossing solid state relay.
The fill start is triggered by a level reading. The fill is stopped by reaching
a user-set level, or exceeding a maximum fill time.
ANALOG OUTPUTS The Model 1700 Instrument has two analog outputs, a 0-10 VDC voltage
output and a 4-20 mADC current loop output. The 4-20 mADC loop output
has 1500 VPK circuit isolation. The outputs can be used simultaneously.
The sources for either of the analog outputs can be either one of the two
configured sensors. Both outputs can be driven from the same source, if
desired.
SIGNAL RELAYS The Model 1700 Instrument has two signal relays that change state
based on a measured input. The set points of these relays are user-
selectable as is their function, alarm on level above or below set point.
4 REV 6
AMERICAN MAGNETICS, INC. INTRODUCTION : MODEL 1700 INSTRUMENT
For the dual channel instrument, both relays can be assigned to either
channel. For the single channel instrument, both relays are assigned to
the only channel.
CONNECTIVITY The Model 1700 Instrument has a 10Base-T Ethernet connection as well
as an RS-232 port for connecting to other equipment. The instrument
communicates with a SCPI-based command set. The command set is
100% backward compatible with the AMI Model 135, 136, 185, and 186
instruments when configured as a single channel instrument. When
configured as a dual channel instrument, additional commands are
incorporated into the command set.
The Model 1700 Instrument allows for remote operation with an external
browser via TCP/IP connection. All commands that are available by
touching the local screen are available via the web browser
REV 6 5
AMERICAN MAGNETICS, INC. INTRODUCTION : MODEL 1700 FRONT PANEL LAYOUT
6 REV 6
AMERICAN MAGNETICS, INC. INTRODUCTION : MODEL 1700 REAR PANEL LAYOUT
REV 6 7
AMERICAN MAGNETICS, INC. INTRODUCTION : MODEL 1700 SPECIFICATIONS @ 25°C
System Architecture
Display: 4.3” 24-bit color TFT display, 480x272 pixel
with resistive touch screen
Sensor Types: Capacitance-based liquid level
Superconductivity-based liquid level
Maximum Length Readout: Capacitance-based liquid level up to 999 in
Superconductivity-based liquid level (LHe) up to 80 in
Superconducting (LHe) Sensor Excitation: Continuous reading or Sample and Hold mode
System Operating Firmware Storage: microSD card
System Clock: Real time clock with timezone support, automatic DST
adjustment, and NTP synchronization
Display Measurement Units: Liquid level in cm, in or percent
Level Measurement
Resolution: 0.1%, 0.1 cm, 0.1 in
Accuracy: ±0.5% of active sensor length
Linearity: ±0.1% or 1 mm (whichever is greater)
Capacitance Sensor Excitation Voltage: 5 VDC
Capacitance Transmitter Measurement Resolution: 0.7 pF
Extension Cable Limits with External Oscillator: 6-500 ft
Superconducting LHe Sensor Current: 4.2K LHe Temperature: 75 mADC nominal
2K LHe Temperature: 57 mADC nominal
Dirty Sensor Mode: Approximately twice normal current for 1 second prior
to normal measurement excitation
Superconducting LHe Sensor Voltage: 4.2K Sensor: approximately 0.87 VDC per inch of
sensor active length @ 10K
2K Sensor: approximately 0.66 VDC per inch of sensor
active length @ 10K
Maximum Open Circuit Voltage: 96 VDC, galvanically isolated
Operating Parameters
Alarm Set Points: 0% to 100%, adjustable; Alarm condition settable to
above or below set point; Assignable to any channel as
high or low alarm for dual instrument
Controller Output: Line voltage @ 2 AAC (maximum)
Sample and Hold Period: 1 second to 86,400 seconds (24 hrs)
Audible alarm: 3500 ± 500 Hz, 73 to 86 dB(A)
8 REV 6
AMERICAN MAGNETICS, INC. INTRODUCTION : MODEL 1700 SPECIFICATIONS @ 25°C
Analog Outputs
Output Types: 0-10 VDC and simultaneous 4 - 20 mADC;
Each assignable to either level channel for dual
instrument
4-20 mA Current Loop Power Supply Voltage: 12-32 VDC
0-10 VDC Recorder Output Load: 50k ohms or greater
0-10 VDC Recorder Output 0% Max Offset: +20 mV
0-10 VDC Recorder Output 100% Max Error: ± 80 mV
Voltage and Current Output Converter Resolution: 12 bits
Integral Non-linearity: ±1LSB
Differential Non-linearitya: ±1LSB
Relays
№1 and/or №2 (W171DIP-7, or equivalent): Contact Form: 1 Form A (SPST-NO)
Maximum Switched Current: 3 A
Switching Voltage: 60 VAC / 100 VDC
Level Control (Solid State): Rated Load Voltage: 2 A at 100 to 240 VAC
Load Voltage Range: 75 to 264 VAC (Line Dependent)
Load Current: 0.1 to 2 A
Surge Current: 30 A (60 Hz, 1 cycle)
Auto Fill
Controller Output Socket: IEC 60320-13 socket on rear panel
Controller Output Power: 2 AAC at line voltage
Autofill Start/Stop Triggering: Level-based; 0% to 100%, adjustable;
Assignable to either nitrogen or helium channel
Fill Timeout Period: 1 minute to 99 hours, 59 minutes
Fill Error Alarm: Fill time out
Communication Protocol
Host Computer Network Protocol: Ehternet 10Base-T TCP/IP and RS-232
IP Addressing: DHCP or static, IPv4
Network Connectivity and Traffic Indication: Link and Activity LEDs on instrument rear panel
RS-232 Connector Specifications: 9-pin D-sub female connector to connect standard
DTE 9-pin D-sub male connector using a standard
straight cable
Communication Command Set: SCPI-based.
REV 6 9
AMERICAN MAGNETICS, INC. INTRODUCTION : MODEL 1700 SPECIFICATIONS @ 25°C
Power Requirements
Primary: 100-240 ±10% VAC, 50-60 Hz, 2.2 A maximum
(200 VA plus sum of controller output)
Backup Battery for RTC: CR2032
Physical
Dimensionsb: Table top configuration:
3.8” H x 8.4” W x 11.4” D
[97 mm H x 213 mm W x 290 mm D]
Single rack mount configuration: 3.5” H x 19.0” W x
11.4” D
[89 mm H x 483 mm W x 290 mm D]
Weight: Table-top configuration: 3.3 lbm [1.5 kG];
Single rack-mount configuration: 4.0 lbm [1.8 kG]
Environmental Limits
Ambient Temperature: Operating: 0°C to 40°C
[32°F to 104°F];
Non-operating: -20 °C to 60 °C
[-4 °F to 140 °F]
Relative Humidity: 0 to 95%; non-condensing
Maximum Instrument Background Field: Gauss (TBD)
Standards
Test Standards Testing of Equipment for Measurement, Control, and
Laboratory Use (IEC 61326-1:2012, EN 61326-1)
Electrostatic Discharge (ESD) (EN 61000-4-2)
Radiated Immunity (EN 61000-4-3)
Fast Transient Burst (EN 61000-4-4)
Surges (EN 61000-4-5)
Conducted Immunity (EN 61000-4-6)
Power Frequency Magnetic Field
(EN 61000-4-8)
Voltage Dips and Interrupts
(EN 61000-4-11)
Harmonics (EN 61000-3-2)
Flicker (EN 6100-3-3)
Conducted Emissions (EN 55011/IEC/CISPR 11)
Radiated Emissions (EN 55011/IEC/CISPR 11)
Safety Requirements for Electrical Equipment for Mea-
surement, Control, and Laboratory Use (IEC 61010-1)
10 REV 6
AMERICAN MAGNETICS, INC. INSTALLATION
INSTALLATION
WARNING Equipment warnings apply to all system installation
configurations. Refer to “Equipment Warnings” on page xi,
in the Foreword to be familiar with the safety requirements
for a system installation.
NOTE If there is any shipping damage, save all packing material and
contact the shipping company representative to file a damage
claim. Do not return to AMI unless prior authorization has been
received.
REV 6 11
AMERICAN MAGNETICS, INC. INSTALLATION : CAPACITANCE-BASED LIQUID LEVEL SENSOR INSTALLATION
1. Carefully remove the sensor from the shipping container and remove all
packaging material.
NOTE If there is any shipping damage, save all packing material and
contact the shipping representative to file a damage claim. Do not
return the instrument to AMI unless prior authorization has been
received.
2. Install the sensor in the vessel using the specified fitting of the sensor.
CAUTION Ensure the sensor is mounted with the top vent hole
located inside of the cryostat.
12 REV 6
AMERICAN MAGNETICS, INC. INSTALLATION : CAPACITANCE-BASED LIQUID LEVEL SENSOR INSTALLATION
CAUTION Exercise care when installing the sensor since dents, crimps,
bends or other physical distortions in the thin wall capacitor
will change electrical characteristics possibly causing
calibration errors and/or disruption of proper instrument
operation.
CONNECTING THE The capacitance sensor may be connected directly to the BNC
connector on the instrument rear panel if the length of the coaxial cable
CAPACITANCE SENSOR is 15 feet or less. If the sensor is greater than 15 feet from the
instrument, an external oscillator/transmitter unit must be used. Refer to
figures “Model 1700 Instrument using internal oscillator/transmitter” on
page 2 and “Model 1700 Instrument using external oscillator/transmitter”
on page 3 as appropriate.
REV 6 13
AMERICAN MAGNETICS, INC. INSTALLATION : LIQUID HELIUM LEVEL SENSOR INSTALLATION
1. Apply a small amount of ECL to any of the BNC connectors that may be
exposed to moisture.
2. Mate the ECL-coated connectors then remove any excess ECL from the
outside of the connector.
NOTE MSDS sheets for the ECL are available upon request.
2. For minimum losses, mount the liquid helium sensor so that warm helium
gas rising from the sensor can pass directly out of the dewar without con-
tacting surfaces at 4.2K.
3. Do not mount the sensor in restricted areas (tubes, etc.) where the liquid
level around the sensor might be depressed by pressure differences in the
gas. Do not cover the holes in the sensor.
5. Avoid installing in a location where icing (frozen water or gas) may occur
since ice formations may cause erratic operation. Ice formation on the NbTi
filament may stop the propagation of the normal (resistive) zone before it
actually reaches the liquid/gas interface. This will give an indication of a
higher helium level than actually exists.
14 REV 6
AMERICAN MAGNETICS, INC. INSTALLATION : LIQUID HELIUM LEVEL SENSOR INSTALLATION
6. Connect the sensor to the Model 1700 LHe Level Sensor connector on the
instrument rear panel (refer to “Liquid Helium Connector J1 Wiring” on
page 111). The liquid helium level sensor leads are color coded:
I+ Red Red 1
V+ Blue Green 8
V- Yellow Natural 6
I- Black Blue 7
CONNECTING THE The instrument is connected to the liquid helium level sensor with a 4-
conductor cable which has a 9-pin D-sub male connector on one end
LIQUID HELIUM LEVEL that mates with the connector used at the instrumentation feed through
SENSOR connector on the cryostat. This connector is typically a multi-pin circular
type connector.
The lead wire for the sensor may be sized by the following equation:
R = 420 – 5.21 L A
where R is the maximum allowable resistance (in ohms) for each lead
wire from the instrument to the sensor, and LA is the active length of the
REV 6 15
AMERICAN MAGNETICS, INC. INSTALLATION : LIQUID HELIUM LEVEL SENSOR INSTALLATION
connected helium level sensor in inches. Values for active sensor length
vs. lead wire distance are provided in the table below.
10 ft. 34 AWG
20 ft.
30 AWG
30 ft.
16 REV 6
AMERICAN MAGNETICS, INC. INSTALLATION : CONFIGURING AN AUTOFILL SYSTEM
LN2 AUTOFILL SYSTEM For LN2 autofill, the system consists of a Model 1700 Instrument with a
liquid level sensor, and a solenoid-operated flow valve. The instrument
DESCRIPTION makes continuous level measurements and based on level, energizes
the valve to begin liquid transfer. The transfer is stopped when the
measured level reaches a user-determined point.
11
13
10
12
LEVEL-CONTROLLED DEWAR
4 7
5
3
1 2
14 6
MODEL 1700 LIQUID
LEVEL CONTROLLER
REV 6 17
AMERICAN MAGNETICS, INC. INSTALLATION : CONFIGURING AN AUTOFILL SYSTEM
Number Item
1 Model 1700 Liquid Level Instrument (Level Controller)
4 Solenoid-operated flow control valve line cord with IEC60320 C14 plug
9 Coaxial cable connecting the Oscillator / Transmitter and the liquid level sensor
(optional; refer to the figure on page 13)
11 Transfer line attached to the target dewar and the fill solenoid valve on the
source dewar
14 Supply dewar
LN2 AUTOFILL SYSTEM CAUTION A relief valve must be used in autofill systems to ensure no
cryogenic liquid can be trapped in a transfer line volume
SETUP where expansion can cause damaging pressure. This can
occur if the solenoid operated fill valve and the supply
dewar isolation valve are closed, trapping a cryogenic liquid
in a confined volume. All AMI LN2 transfer line systems
include a relief valve to preclude this sort of event.
CAUTION The switched valve outlet socket switches the line voltage
to which the instrument mains power is connected. Ensure
any connected valve is rated to support the local VAC line
voltage.
2. Connect the transfer line (11) and fill solenoid valve (12) or supply manifold
to the source dewar.
3. Connect the other end of the transfer line to the fill port (10) on the valve/
manifold of the target dewar.
18 REV 6
AMERICAN MAGNETICS, INC. INSTALLATION : CONFIGURING AN AUTOFILL SYSTEM
a. For distances of 15 feet and less, connect the coaxial cable (6)
between the BNC connector on the liquid level sensor and the BNC
connector on the back of the instrument labeled O/T (3).
b. For distances greater than 15 feet, connect the coaxial cable (9)
between the BNC connector on the liquid level sensor and the BNC
connector on the oscillator / transmitter (7). Use a second length of
coaxial cable (6) to connect between the oscillator / transmitter (7)
and the BNC connector on the back of the instrument labeled O/T
(3).
5. Connect the solenoid valve (12) to the IEC60320 C13 valve socket1 on the
Model 1700 Instrument rear panel labeled VALVE CONTROL (2).
Connect the IEC320 C14 inlet connector and instrument power cord (5)
to an appropriate power source (see below).
LHE AUTOFILL SYSTEM An autofill system for LHe application is a more complex issue than
simply selecting a solenoid-controlled valve that can be used for other
SETUP cryogens such as LN2. The valving control must be highly-insulated to
avoid vaporizing the transferred LHe. Contact an AMI Technical Support
Representative to inquire about available LHe autofill solutions as this is
a continually changing marketplace.
REV 6 19
AMERICAN MAGNETICS, INC. INSTALLATION : POWER REQUIREMENTS
POWER REQUIREMENTS
Ensure the power switch is in the OFF (O) position. Plug the Model 1700
Instrument line cord into the power entry module on the instrument rear
panel and into the appropriate power receptacle.
The Model 1700 instrument provides 0-10 VDC and 4-20 mADC outputs
as standard. These outputs may be optionally connected to external
equipment per the instructions below.
0-10 VDC RECORDER The Model 1700 instrument provides a 0-10 VDC output on the AUX I/O
connector (refer to “Aux I/O Pin Definitions” on page 112) corresponding
OUTPUT to 0-100% of liquid level. This output can be assigned to either level
measurement channel (refer to “Analog output signals” on page 47).
Connect the voltage output on the AUX I/O connector to a suitable
receiver being careful not to exceed the impedance restrictions (refer to
“Model 1700 Specifications @ 25°C” on page 8).
4-20 MA CURRENT The Model 1700 instrument provides a 4-20 mADC output on the AUX I/
O connector (refer to “Aux I/O Pin Definitions” on page 112)
LOOP OUTPUT corresponding to 0-100% of liquid level. This output can be assigned to
either level measurement channel (refer to “Analog output signals” on
page 47).
20 REV 6
AMERICAN MAGNETICS, INC. INSTALLATION : CONNECTING THE ANALOG OUTPUTS
4-20 mA
LOOP
RECEIVER
REV 6 21
AMERICAN MAGNETICS, INC. INSTALLATION : CONNECTING THE ANALOG OUTPUTS
22 REV 6
AMERICAN MAGNETICS, INC. OPERATION
OPERATION
This section describes the operation of the Model 1700 Instrument.
ENERGIZING THE 1. Turn the power switch on the front panel of the instrument to the On (I) posi-
tion. The display will briefly show a boot image and then display the home
MODEL 1700 screen showing level(s).
INSTRUMENT 2. The boot process takes approximately 30 seconds. Boot time can be longer
(approximately 2 minutes) if the instrument has been configured for a net-
work connection and then is booted without the network present.
3. When the boot process is complete, the instrument will display the home
(level) screen.
SCREEN NAVIGATION
HOME SCREEN NOTE If the instrument was purchased with a level sensor, the
instrument will be pre-configured and calibrated at the factory.
REV 6 23
AMERICAN MAGNETICS, INC. OPERATION : SCREEN NAVIGATION
HOME SCREEN FOOTER Every screen has a footer. The level home screen displays both status
information and buttons in the footer to navigate to other screens.
1. If the instrument was purchased with level sensor(s), the instrument will be shipped set up
and calibrated.
24 REV 6
AMERICAN MAGNETICS, INC. OPERATION : NAVIGATING THE INSTRUMENT MENUS
The functions of the ALARM, AUTOFILL, and Help footer items are as
described in the previous section.
REV 6 25
AMERICAN MAGNETICS, INC. OPERATION : NAVIGATING THE INSTRUMENT MENUS
MENU STRUCTURE
MENU
NITROGEN OSCILLATOR USED: ALARMS ADDRESSING SCHEME Instrument Serial LN2 ON/OFF
Number <value> SHUTDOWN
<INTERNAL | EXTERNAL> {STATIC | DYNAMIC |
1
ALARM No.1 SOURCE: {DISABLED | DISABLED} LHe ON/OFF
NITROGEN LEVEL ON HOME SCREEN? NITROGEN | HELIUM} HW: <value>
{YES | NO} ADDRESS < value>
<
ALARM No.1 WHEN LEVEL {>= | <=} FPGA: <value>
NITROGEN LEVEL DAMPING <level>% NETMASK < value> IN
FW: <value>
Undamped O/T Period: <period> μs ALARM No.2 SOURCE: {DISABLED | GATEWAY < value> OUT
1
NITROGEN | HELIUM} Date of
Undamped LN2 Level: <level> % HOSTNAME <value> Manufacture: >
1 ALARM No.2 WHEN LEVEL {>= | <=} <value>
<level>% MAC ADDRESS <value>
Damping Time Period: <period> >>
Helium Power
seconds
RELAYS Supply: <value>
Damped O/T Period: <period> μs
1 RELAY No.1 SOURCE: {DISABLED | SET TIME
NITROGEN | HELIUM}
Damping LN2 Level: <level> %
1 CURRENT SYSTEM DATE AND TIME: < value>
RELAY No.1 CLOSES WHEN LEVEL
HELIUM SENSOR TYPE: <2 K | 4.2 K> {>= | <=} < level>%
1 SYSTEM TIMEZONE: < value>
HELIUM LEVEL ON HOME SCREEN? RELAY No.2 SOURCE: {DISABLED | BATTERY GOOD <NOTIFY | NO NOTIFY}>
{YES | NO} NITROGEN | HELIUM}
SET MANUALLY
CAL N2 RELAY No.2 CLOSES WHEN LEVEL
{>= | <=} < level>%
TIMEZONE: < value>
OSCILLATOR PERIOD:
<value> μs AUTOFILL
1 YEAR: <value>
SENSOR ACTIVE LENGTH AutoFill CONTROL: {DISABLED |
MONTH: <value>
<nnn.n> cm NITROGEN | HELIUM}
DAY: <value>
PERFORM MAX CAL START: <level > %
HOUR: <value>
PERFORM MIN CAL STOP: <level> %
MIN: <value>
LOSS OF SENSOR CAL FILL TIMEOUT: < m> MINUTES
SET FROM NTP SERVER
APPROX. (MAX) CAL MULTIPLIER ANALOG OUTPUTS
<value> UPDATE
OUTPUT 0-10 SOURCE: {DISABLED |
APPROX. (MAX) CAL MULTIPLIER NITROGEN | HELIUM}
<Choose Firmware Version >
{NOTIFY | NO NOTIFY}
OUTPUT 4-20 mA SOURCE: RS-232 SETUP
SENSOR NAME(S) {DISABLED | NITROGEN | HELIUM}
RS-232 PORT CONFIGURATION {LINUX LOGIN | SCPI}
NITROGEN SENSOR NAME:
<value> ECHO {ON | OFF}
{ENABLED | DISABLED}
VALVE {OPEN | CLOSED}
Analog Outputs
DRAG
2
DRAW
2
QUIT
2
FACTORY RESTORE
1
INDICATION ONLY
RESTORE
2
LOCAL OPERATON ONLY 2
26 REV 6
AMERICAN MAGNETICS, INC. OPERATION : NAVIGATING THE INSTRUMENT MENUS
SCREEN DESCRIPTIONS The Model 1700 Instrument displays various screens on the graphic
display when a MENU item is chosen. The items for each screen are
described in the following table.
LIQUID NITROGEN LEVEL DAMPING Undamped O/T Period: Information: <value>b μs (present
value)
REV 6 27
AMERICAN MAGNETICS, INC. OPERATION : NAVIGATING THE INSTRUMENT MENUS
LIQUID HELIUM SENSOR SENSOR SAMPLE INTERVAL: Data entry: <value> min
CALIBRATION
MEASURE SENSOR V/A Performs a function: <values> V mAb
SENSOR ACTIVE LENGTH: Data entry: <value> cm or in (tap
units to change)
28 REV 6
AMERICAN MAGNETICS, INC. OPERATION : NAVIGATING THE INSTRUMENT MENUS
REV 6 29
AMERICAN MAGNETICS, INC. OPERATION : NAVIGATING THE INSTRUMENT MENUS
30 REV 6
AMERICAN MAGNETICS, INC. OPERATION : NAVIGATING THE INSTRUMENT MENUS
FIRMWARE UPDATE INSTALL Select Firmware Version: Selection list: <values> and then
moves to INSTALL verification on
selection
REV 6 31
AMERICAN MAGNETICS, INC. OPERATION : NAVIGATING THE INSTRUMENT MENUS
SYSTEM TEST TURN ON ALL LCD PIXELS Move to multi-color test screen
32 REV 6
AMERICAN MAGNETICS, INC. OPERATION : CAPACITANCE (LIQUID NITROGEN) LEVEL
CONFIGURE THE NOTE If the instrument was purchased with a capacitance-based level
sensor, Steps 1 through 5, below have already been performed.
INSTRUMENT TO
DISPLAY NITROGEN 1. From the main screen, choose the following: MENU > SENSORS.
LEVEL
2. The instrument has a built-in
sensor oscillator which will be
used if the level sensor is con-
nected to the BNC connector on
the rear panel of the instrument
with a maximum of 15 feet of
RG59/U coaxial cable (refer to
the figure on page 2). Ensure the
NITROGEN OSCILLATOR
USED field displays INTERNAL. INTERNAL Oscillator Indicated
a. If the sensor is greater
than 15 feet from the instrument, an external oscillator/transmitter is
required and the NITROGEN OSCILLATOR USED will display
EXTERNAL (refer to the figure on page 3).
REV 6 33
AMERICAN MAGNETICS, INC. OPERATION : SUPERCONDUCTING (LIQUID HELIUM) LEVEL
CONFIGURE THE NOTE If the instrument was purchased with an AMI LHe level sensor,
Steps 1 through 6, below have already been performed.
INSTRUMENT TO
DISPLAY HELIUM LEVEL 1. From the main screen, choose the following: MENU > SENSORS.
Choose CAL He
34 REV 6
AMERICAN MAGNETICS, INC. OPERATION : SUPERCONDUCTING (LIQUID HELIUM) LEVEL
NOTE The entered sensor active length must match the actual active
length of the connected LHe sensor for accurate results.
Numeric Keypad
REV 6 35
AMERICAN MAGNETICS, INC. OPERATION : SUPERCONDUCTING (LIQUID HELIUM) LEVEL
36 REV 6
AMERICAN MAGNETICS, INC. OPERATION : SUPERCONDUCTING (LIQUID HELIUM) LEVEL
REV 6 37
AMERICAN MAGNETICS, INC. OPERATION : SUPERCONDUCTING (LIQUID HELIUM) LEVEL
AMI expects the helium level sensor to be reasonably clean and free
from oil, water, ice, etc. for proper operation. However, it is recognized
that some experiments might result in some material being deposited on
the sensor wire. Ice formation at some point on the sensor is a typical
occurrence. Therefore, the Model 1700 has the capability of increasing
the current for a short period of time at the beginning of the
measurement cycle (in the HOLDING mode only) to try and drive the
resistive zone of the sensor wire past the dirty region. This is termed
DIRTY SENSOR MODE. This operation may or may not be successful
depending on the degree of sensor contamination. This mode should be
viewed as a stopgap measure only. If correct readings cannot be
reestablished, the only choice is to warm the sensor or remove for
cleaning or replacement.
NOTE Operation in the dirty sensor mode increases liquid helium losses.
Consequently, operation in this mode should not be used unless
the sensor is known or anticipated to become dirty or the helium
level measurement is in question due to unclean operation.
38 REV 6
AMERICAN MAGNETICS, INC. OPERATION : ALARMS AND RELAYS
The Model 1700 Instrument has two types of alarms, level-based and
time-based alarms.
ALARM STATUS Prior revisions of the Model 1700 have attempted to indicate the alarm
conditions in the main screen footer region, including cycling through
SCREEN multiple alarm texts in the footer. This has been replaced by an Alarm
Status screen that summarizes all alarm states and provides a button for
muting.
The Alarm Status screen can be accessed at any time by tapping on the
ALARM text in the footer. All alarms are summarized in the Alarm Status
screen. If the Alarm Status screen shows the “ALARMING” button (as
illustrated above), then touching the button will mute the alarm(s). This
mute function will remain in effect until a new alarm is activated, or the
status of any active alarm changes. Shortcuts may also be provided on
the Alarm Status screen to enter the Level Alarms and Autofill
configuration screens.
LEVEL-BASED ALARMS The Model 1700 Instrument has two user-configurable level alarms.
Each alarm can be triggered by either level measurement (for dual level
configured instruments). Each alarm can be configured to be active
above or below a user-defined setpoint.
When an alarm condition occurs, an audible alert will sound and visual
indication will appear on the front panel.
REV 6 39
AMERICAN MAGNETICS, INC. OPERATION : ALARMS AND RELAYS
TIME-BASED FILL The Model 1700 Instrument has an alarm to indicate that there is a
problem with the autofill function. If enabled, the instrument will start a
ALARM timer when an autofill condition is initiated, and if the level has not
reached the fill stop level within the user-set period of time, an Autofill
Timeout alarm will occur.
1. The de-energizing of the fill valve socket on the instrument rear panel,
MULTIPLE ALARMS More than one alarm condition can occur at the same time. If that is the
case, the Alarm Status screen summarizes all the alarm states. Access
the Alarm Status screen by touching the ALARM text in the footer.
40 REV 6
AMERICAN MAGNETICS, INC. OPERATION : ALARMS AND RELAYS
ACKNOWLEDGING AN NOTE The alarms are not “latched” so if the alarm condition clears itself,
the instrument will remove the alarm condition.
ALARM
1. When an alarm is initiated, several things will occur:
REV 6 41
AMERICAN MAGNETICS, INC. OPERATION : ALARMS AND RELAYS
CONFIGURING RELAY The relay outputs are available via the AUX I/O connector on the rear
panel of the instrument per the pinout provided on page 112.
SETPOINTS
1. From the MENU screen, select
OUTPUTS and then choose
RELAYS.
42 REV 6
AMERICAN MAGNETICS, INC. OPERATION : ALARMS AND RELAYS
REV 6 43
AMERICAN MAGNETICS, INC. OPERATION : CONFIGURE THE AUTOFILL FUNCTION
SETTING THE AUTOFILL 1. From the MENU screen, choose OUTPUTS and then choose AUTOFILL.
44 REV 6
AMERICAN MAGNETICS, INC. OPERATION : CONFIGURE THE AUTOFILL FUNCTION
ENABLE THE AUTOFILL The autofill function must first be enabled. After the autofill control loop
has been configured (the parameters in the prior section) the instrument
FUNCTION will enter the AUTO-OFF state.
1. Touch the AUTO-OFF text in the display footer until it reads AUTO-ON.
AUTOFILL States
AUTO-ON Automatically maintains level Autofill will alarm and cease if fill
between fill START and STOP set- valve stays open for ≥ the
point (i.e. autonomous mode). FILL TIMEOUT setting .
M-OPEN Manual Open: Energizes the valve Any active ALARM with
control socket on the rear panel. >= setpoint defined for the fill liquid
will close valve.
CLEARING THE NOTE A Loss of Sensor error condition (see page 19) on the nitrogen
channel will also halt an active autofill in addition to the autofill
AUTOFILL TIMEOUT timeout feature.
ALARM
1. Touch the TIMEOUT text in the HOME screen or in the footer until AUTO-
ON is displayed.
2. Touch SAVE in the footer. This will restart the autonomous autofill function.
3. To clear the TIMEOUT without restarting the autofill function, choose the M-
CLOSED state and then choose SAVE in the footer.
REV 6 45
AMERICAN MAGNETICS, INC. OPERATION : SELECT THE APPROPRIATE UNITS ON THE DISPLAY
DAMPING CONFIGURATION
1. From the main screen, choose the following: MENU > SENSORS. Ensure
the nitrogen level measurement is enabled and displayed on the HOME
screen.
46 REV 6
AMERICAN MAGNETICS, INC. OPERATION : ANALOG OUTPUT SIGNALS
Refer to page 112 of the Appendix for the AUX connector pin-out.
REV 6 47
AMERICAN MAGNETICS, INC. OPERATION : ETHERNET CONNECTIVITY
ETHERNET CONNECTIVITY
6. If changes are made, select the SAVE button and then the instrument will
reconfigure itself with the chosen network settings.
48 REV 6
AMERICAN MAGNETICS, INC. OPERATION : SERIAL CONNECTIVITY
SERIAL CONNECTIVITY
REV 6 49
AMERICAN MAGNETICS, INC. OPERATION : ABNORMAL OPERATION
ABNORMAL OPERATION
CAPACITANCE SENSOR To ensure proper instrument calibration and operation, care must be
taken to ensure the sensor is kept free of contaminants and not
CONTAMINATION subjected to any force which would physically distort the sensor. Water
or other electrically conducting substances in the sensor will disturb the
measured capacitance and the instrument’s response. Physically
distorting the sensor in any way will also cause abnormal instrument
operation by introducing variations in the sensor capacitance not due to
liquid level. The absolute calibration of the instrument can be inaccurate
if care is not taken to ensure the sensor is in the proper environment.
50 REV 6
AMERICAN MAGNETICS, INC. OPERATION : ABNORMAL OPERATION
Cold sensors exposed to humidified air can show erroneous high level
readings because the air contains moisture that can condense between
the cold sensing tubes. A minute film of water can cause a shorted or
partially shorted condition, which results in false level readings. As the
sensor warms, the moisture may evaporate and the sensor will again
read correctly. This is a physical phenomenon and does not indicate any
problem with your AMI level equipment. Limit or eliminate exposure of
cold sensors to humidified air to avoid this condition.
If a sensor should require cleaning and the sensor is for use with liquids
other than liquid oxygen, flushing with pure alcohol is recommended.
The sensor cannot be used again until all the alcohol has been
evaporated. Under no circumstances should the sensor be
disassembled.
For sensors to be used with liquid oxygen (LOX), although measures are
taken to minimize oils and greases during manufacture, no special
cleaning required for LOX service is provided by AMI. Certified LOX
cleaning is the responsibility of the customer.
DIRTY HELIUM SENSOR Refer to “Other Liquid Helium Functions” on page 37.
OPERATIONAL MODE
2. Press RESTORE.
REV 6 51
AMERICAN MAGNETICS, INC. OPERATION : SHUTTING THE INSTRUMENT DOWN
52 REV 6
AMERICAN MAGNETICS, INC. CALIBRATION
CALIBRATION
Model 1700 instrument is calibrated at the factory for a specific length
sensor(s) for use in a specific liquid(s). The calibration length(s) and
calibration liquid(s) are listed on the calibration sticker on the bottom of
the instrument. For capacitance sensors, if the factory calibration
method utilized was approximate, the calibration length will be noted as
an approximate value.
REV 6 53
AMERICAN MAGNETICS, INC. CALIBRATION : SETTING THE SYSTEM DATE AND TIME
NOTE The clock is set to GMT at the factory and is battery backed. The
TIMEZONE field must be set to a value from the TZ database.1 The
list of TIMEZONE values can be found at:
https://en.wikipedia.org/wiki/List_of_tz_database_time_zones
The TIMEZONE entry must exactly match the entire field as listed
in the above URL in the TZ database name column and it is case
sensitive. An instrument reboot may be required for the
TIMEZONE to take effect. If the entered TIMEZONE supports
Daylight Savings Time (DST), then the local time will be
automatically adjusted to DST.
6. Choose SAVE in the footer after all the fields have been edited as neces-
sary.
1. The tz database is a collaborative compilation of information about the world's time zones,
primarily intended for use with computer programs and operating systems. The topic of
worldwide timezones is a complex issue so please refer to the on-line resource noted above.
54 REV 6
AMERICAN MAGNETICS, INC. CALIBRATION : CAPACITANCE-BASED LEVEL CALIBRATION
UNDERSTANDING THE American Magnetics, Inc. fabricates the liquid level sensor with two vent
holes; a lower vent hole in the side wall near the bottom which is typically
SENSOR ACTIVE the minimum liquid level calibration point and the upper vent hole in the
LENGTH sensor side typically near the top of the sensor. The liquid level location
approximately 2.5 cm (1 in) below the upper vent hole is typically the
100% calibration point.
NOTE Without entry of the active length, the instrument will not be
able to read out in units other than percent.
&211(&725
The Model 1700
Instrument requires
6(1625 the user to enter the
6(5,$/180%(5 calibrated, or active
length, (physical
distance between the
MIN and MAX
calibration locations
on the sensor) in
order for the absolute
units function
1371</21 (inches, cm) to be
)(('7+528*+
727$/
27+(502817,1* displayed if desired.
237,216$9$,/$%/(
6(1625
/(1*7+
The user must enter
the sensor length in
centimeters. Use the
Active Length value
noted on the level
sensor
documentation or
6(1625
measure the distance
$&7,9( between the lower
6(1625
/(1*7+ vent hole on the
63(&,)< sensor and 1.0 inch
(2.5 cm) below the
upper vent hole on
the sensor as
illustrated at left.
REV 6 55
AMERICAN MAGNETICS, INC. CALIBRATION : CAPACITANCE-BASED LEVEL CALIBRATION
VARIATIONS IN THE For cryogenic liquids, the dielectric of the liquid will change with a
change in density. The amount of change is dependent on the properties
DIELECTRIC WITH of the specific liquid. The figure below illustrates the variations in
CHANGING DENSITY dielectric for nitrogen vs. pressure under saturated conditions.1 Since
the instrument uses a capacitance-based method for determining liquid
level, such a change in the dielectric of the liquid will result in a shift in
the level reading of the instrument. The calibration procedures described
herein are most accurate when applied in situations where the operating
conditions of the cryo-vessel are relatively constant, i.e. the operating
pressure and temperature of the cryo-vessel are relatively constant.
56 REV 6
AMERICAN MAGNETICS, INC. CALIBRATION : CAPACITANCE-BASED LEVEL CALIBRATION
0.45
0.4
0.35
0.3
Diel - 1
0.25
0.2
0.15
0.1
0.05
0
0 50 100 150 200 250 300 350 400 450 500
Pressure (psi)
For the case of the target liquid being unavailable, AMI uses liquid
nitrogen as the reference liquid and an Approximate Calibration is
performed using mathematical manipulation of the ratio of the dielectric
constants between liquid nitrogen and the desired liquid. This procedure
is outlined in the Approximate Calibration section beginning on page 70.
The technique is intended to provide the instrument with an approximate
calibration so that it can be used immediately by the customer. However,
the customer is still expected to perform a more accurate calibration
where feasible, such as the open dewar or closed dewar calibration, with
the target liquid.
REV 6 57
AMERICAN MAGNETICS, INC. CALIBRATION : CAPACITANCE-BASED LEVEL CALIBRATION
For the case where a sensor is too long to be calibrated in AMI facilities,
AMI will perform a partial length open dewar calibration in liquid nitrogen,
and then calculate the MAX calibration point. A dielectric ratio may also
be subsequently utilized to adjust for a target liquid other than liquid
nitrogen. The customer is expected to perform a more accurate open
dewar or closed dewar calibration if feasible.
START
N
N
Is factory
calibration
Approximate?
Can Open Y**
Dewar Calibration be
Y
performed in target
Can Open liquid?
Dewar Calibration be N
Y
performed in target
liquid?
N
Perform Open
Dewar Calibration
N
Can Closed
Dewar Calibration be
Y N
Can Closed performed in target
Dewar Calibration be liquid?
Y
performed in target
liquid?
Perform Closed
Dewar Calibration
N
Perform
Approximate
Calibration
Go to Operation
58 REV 6
AMERICAN MAGNETICS, INC. CALIBRATION : CAPACITANCE-BASED LEVEL CALIBRATION
If the instrument and sensor are purchased as a unit from AMI, then the
factory calibration, including sensor serial number and sensor physical
parameter information entered at the factory, will be adequate in most
cases. However, for the exceptions noted in the previous paragraphs
(which are approximate calibrations), the customer should perform a
more accurate open dewar or closed dewar calibration if at all possible.
A customer-performed calibration is also required for sensors that are
purchased as a separate item from the instrument, since the instrument
and sensor were not both available for calibration at AMI facilities.
MENU Selection
REV 6 59
AMERICAN MAGNETICS, INC. CALIBRATION : CAPACITANCE-BASED LEVEL CALIBRATION
CAL N2 Selection
60 REV 6
AMERICAN MAGNETICS, INC. CALIBRATION : CAPACITANCE-BASED LEVEL CALIBRATION
HOME Selection
For a system where the internal oscillator is used, connect the coaxial
cable to the BNC connector on the instrument rear panel but leave the
cable disconnected from the sensor BNC connector. For a system where
an external oscillator is used, connect the coaxial cable between the
instrument and the oscillator/transmitter unit. Connect the second (6 ft)
coaxial cable to the input of the oscillator/transmitter unit but leave the
cable disconnected from the sensor BNC connector.
REV 6 61
AMERICAN MAGNETICS, INC. CALIBRATION : CAPACITANCE-BASED LEVEL CALIBRATION
MENU Selection
SENSORS Selection
CAL N2 Selection
62 REV 6
AMERICAN MAGNETICS, INC. CALIBRATION : OPEN DEWAR CALIBRATION
MENU Selection
SENSORS Selection
CAL N2 Selection
REV 6 63
AMERICAN MAGNETICS, INC. CALIBRATION : OPEN DEWAR CALIBRATION
NOTE Note that the periods listed to the right of the PERFORM MAX
CAL and PERFORM MIN CAL buttons are updated as well as
the date and time stamps of the calibration points.
64 REV 6
AMERICAN MAGNETICS, INC. CALIBRATION : CLOSED DEWAR CALIBRATION
PRESETTING THE The following procedure should be performed before installation of the
sensor in the target cryo-vessel.
MAXIMUM AND
MINIMUM 1. Connect the sensor coaxial cable to the BNC connector on the rear panel of
the instrument (see page 13). Do not connect the sensor. Energize the
CALIBRATION POINTS instrument.
MENU Selection
SENSORS Selection
REV 6 65
AMERICAN MAGNETICS, INC. CALIBRATION : CLOSED DEWAR CALIBRATION
CAL N2 Selection
7. Connect the sensor to the oscillator coaxial cable that is connected to the
instrument.
2.1 L active – 1
C adj = 1.20 1 + --------------------------- -------------
5.2 L total 0.454
where Ltotal is the total sensor length in inches, Lactive is the active sensor
length in inches, and ε is the dielectric constant of the target liquid.
66 REV 6
AMERICAN MAGNETICS, INC. CALIBRATION : CLOSED DEWAR CALIBRATION
COMPLETING THE 1. Install the sensor in the dewar and energize the instrument with the sensor
connected to the instrument via the oscillator (if required) and extension
CLOSED DEWAR cable(s) (see the system diagram on page 2).
CALIBRATION
2. Connect a strip chart recorder or graphical data logging application to the
PROCEDURE recorder output terminals on the rear panel of the instrument. If the recorder
output is not available, the 4-20 mA current loop output may be used if
installed, or an installed communications option can be used to query the
instrument for the liquid level at regular time intervals during the calibration
procedure. If no remote monitoring or communication option is installed, the
level display must be manually plotted vs. time during the procedure.
3. Refer to “Analog output signals” on page 47. to configure the recorder out-
put or current loop output.
4. Commence filling the dewar. While the sensor is cooling down, there may be
a slow drift in the displayed liquid level. However, when the liquid actually
touches the bottom of the sensor, contact with the liquid surface may
become apparent by virtue of more random and frequent fluctuations in the
displayed liquid level. The liquid level trace will also start to show an increas-
ing profile with positive slope.
Once the indications of the contact between the sensor and liquid become
readily apparent, use the PERFORM MIN CAL procedure below to save the
new minimum calibration point. This point is the 0% level of the sensor when
the PERFORM MIN CAL procedure is finished and saved becomes the 0%
level.
NOTE If the sensor is installed in the dewar with some small amount of
liquid already in contact with the sensor, then the final minimum
calibration point can be set before filling begins but after any
thermally induced fluctuations in the observed output have
diminished. However, note that the measured span of the liquid
level is reduced by the initial level of liquid in contact with the
sensor.
REV 6 67
AMERICAN MAGNETICS, INC. CALIBRATION : CLOSED DEWAR CALIBRATION
7. Continue the transfer while observing the liquid level trace on the strip chart
recorder or computer display, whose slope is proportional to the transfer
rate. The slope of the liquid level trace should decrease significantly when
the liquid reaches the hole in the top of the sensor.
When the break in the slope of the level trace occurs (i.e. the slope of the
level trace becomes 0 or horizontal), perform a PERFORM MAX CAL pro-
cedure below. The level on the sensor when the PERFORM MAX CAL pro-
cedure is finished and saved becomes the 100% level.
68 REV 6
AMERICAN MAGNETICS, INC. CALIBRATION : CLOSED DEWAR CALIBRATION
10. To achieve a standard calibration of the sensor with the active region
located from the lower hole to one inch below the upper hole, use the level
data from the instrument to recalibrate the maximum point when the percent
level corresponds to one inch below the upper hole. Use the following equa-
tion to determine the percent level at which to reset the maximum calibration
point:
1
MAX percent = 100 – 100 ---------------
L active
where Lactive is the active length of the sensor in inches. This technique can
be used assuming the sensor was built as a standard sensor. If the sensor
was made in a custom configuration, refer to the sensor documentation and/
or drawing or contact AMI.
When the sensor is calibrated by the closed dewar procedure, the actual
length of calibration will be 21" (distance between the bottom and top holes
in the sensor). When the liquid is 1" below the upper hole, the display will
show 95.2% [e.g. 100% (1"/21" x 100%) ]. When the liquid level reaches
this point during usage, perform the PERFORM MAX CAL operation and
SAVE the result. The instrument and sensor are now calibrated with a stan-
dard active region of 20". The length setting of the sensor in the instrument
should also be configured for 50.8 cm (20").
REV 6 69
AMERICAN MAGNETICS, INC. CALIBRATION : APPROXIMATE CALIBRATION
APPROXIMATE CALIBRATION
MENU Selection
SENSORS Selection
CAL N2 Selection
70 REV 6
AMERICAN MAGNETICS, INC. CALIBRATION : APPROXIMATE CALIBRATION
NOTE Note that the periods listed to the right of the PERFORM MAX
CAL and PERFORM MIN CAL buttons are updated as well as
the date and time stamps of the calibration points.
8. Measure the distance between the bottom hole of the sensor and the loca-
tion of the liquid level dipped for max calibration. This measured length is
Ldipped.
9. The dielectric constant for the reference liquid, ε1, and the target liquid, ε2,
must be known to complete the approximate calibration. These values must
REV 6 71
AMERICAN MAGNETICS, INC. CALIBRATION : APPROXIMATE CALIBRATION
where Ldipped is the length of the sensor dipped in the reference liquid and
Lactive is the active sensor length.
NOTE If the target liquid is available for dipping (i.e. the reference liquid
and target liquid are the same), then the dielectric ratio, (ε2 1)/
(ε11), becomes 1. If the full active length of the sensor can be
dipped, then the length ratio, Lactive / Ldipped , becomes 1.
First, the sensor is dipped as far as possible into the liquid nitrogen and
cooled. The minimum point is then set as outlined in step 2. The
maximum point is set as outlined in step 3 while the sensor is
submerged 30" in liquid nitrogen. The dielectric constant for liquid
nitrogen is 1.454 and for liquid argon is 1.53. Substituting all values into
the approximate calibration factor equation yields:
1.53 – 1 100
APPROX MAX CAL = ---------------------- --------- = 3.891
1.454 – 1 30
72 REV 6
AMERICAN MAGNETICS, INC. CALIBRATION : APPROXIMATE CALIBRATION
11. The sensor can now be installed in the dewar containing the target liquid.
The approximate calibration can be used until an open dewar or closed
dewar calibration can be performed with the target liquid.
REV 6 73
AMERICAN MAGNETICS, INC. CALIBRATION : SUPERCONDUCTIVITY-BASED (LIQUID HELIUM) LEVEL CALIBRATION
The instrument has been internally calibrated for use with AMI liquid
helium level sensors at the AMI facility. However, it is still necessary to
verify, and if necessary enter, the correct active length for the connected
sensor.
NOTE If the instrument was purchased with a helium level sensor, the
active length will be set at AMI prior to shipping.
MENU Selection
SENSORS Selection
74 REV 6
AMERICAN MAGNETICS, INC. CALIBRATION : SUPERCONDUCTIVITY-BASED (LIQUID HELIUM) LEVEL CALIBRATION
MENU Selection
SENSORS Selection
CAL He Selection
REV 6 75
AMERICAN MAGNETICS, INC. CALIBRATION : SUPERCONDUCTIVITY-BASED (LIQUID HELIUM) LEVEL CALIBRATION
CONTINUOUS The Continuous measure time limit feature sets a maximum time that the
instrument will keep the liquid helium level sensor energized. If the
MEASURE TIME LIMIT instrument is left in the CONTINUOUS mode for the CONTIN.
MEASURE TIME LIMIT interval, the instrument will revert back to the
sample and HOLDING mode to prevent excessive liquid helium boil off.
76 REV 6
AMERICAN MAGNETICS, INC. CALIBRATION : SUPERCONDUCTIVITY-BASED (LIQUID HELIUM) LEVEL CALIBRATION
REV 6 77
AMERICAN MAGNETICS, INC. CALIBRATION : SUPERCONDUCTIVITY-BASED (LIQUID HELIUM) LEVEL CALIBRATION
78 REV 6
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE
REMOTE INTERFACE
REFERENCE
The Model 1700 Instrument provides both serial (RS-232) and Ethernet
interfaces as standard features. The serial and Ethernet interfaces can
be operated simultaneously. Separate output buffers are provided for the
serial and Ethernet return data. The serial and Ethernet interfaces
accept SCPI syntax for commands and queries.
The Model 1700 provides a web browser connection (port 80) via TCP/
IP and presents a mirror of the local display which has hot spots that can
be clicked or tapped, and which also accepts keyboard entry for textual
and numerical input. The instrument’s IP address can be ascertained by
referring to the section titled “IP Addressing Scheme” on page 48. By
using a web browser to connect to the instrument, all functionality of the
Model 1700 can be controlled via the remote connection.
REV 6 79
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : SCPI COMMAND SUMMARY
• Triangle brackets < > indicate that you must supply a value.
• Parentheses ( ) within < > indicate alternative units are available.
• Capitalized portions of the commands indicate acceptable abbreviations.
System-Related Commands
N2?
HE?
SERial_NUMber?
DATE_MANUfacture?
HardWare_VERsion?
FirmWare_VERsion?
Scpi_VERsion?
SYStem:BEEPer:IMMediate <seconds>
SYStem:BEEPer:STATe {0|1}
SYStem:BEEPer:STATe?
SYStem:KLOCK {0|1}
SYStem:KLOCK?
SYStem:DATE <yyyy>,<mm>,<dd>
SYStem:DATE?
SYStem:TIME <hh>,<mm>,<ss>
SYStem:TIME?
SYStem:REBOOT
IPV4addr?
80 REV 6
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : SCPI COMMAND SUMMARY
MACADDR?
GATEWAY?
HOSTname?
SCREENCAP
SYStem:LOcal
SYStem:REMote
DISPlay:N2?
CONFigure:NAME:SENSor:N2 "<string>"
NAME:SENSor:N2?
DISPlay:HE?
CONFigure:NAME:SENSor:HE "<string>"
NAME:SENSor:HE?
CONFigure:RELay1:CHannel {0|1|2}
RELay1:CHannel?
CONFigure:RELay1:SETpoint <level>
RELay1:SETpoint?
CONFigure:RELay1:OPeration {0|1}
RELay1:OPeration?
CONFigure:RELay2:CHannel {0|1|2}
RELay2:CHannel?
CONFigure:RELay2:SETpoint <level>
RELay2:SETpoint?
CONFigure:RELay2:OPeration {0|1}
RELay2:OPeration?
{RELay1|RELay2}:STATus?
REV 6 81
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : SCPI COMMAND SUMMARY
CONFigure:ALArm1:CHannel {0|1|2}
ALArm1:CHannel?
CONFigure:ALArm1:SETpoint <level>
ALArm1:SETpoint?
CONFigure:ALArm1:OPeration {0|1}
ALArm1:OPeration?
CONFigure:ALArm2:CHannel {0|1|2}
ALArm2:CHannel?
CONFigure:ALArm2:SETpoint <level>
ALArm2:SETpoint?
CONFigure:ALArm2:OPeration {0|1}
ALArm2:OPeration?
{ALArm1|ALArm2}:STATus?
MEASure:N2:LEVel?
MEASure:N2:PERIod?
MEASure:HE:LEVel?
MEASure:HE:HOLD
MEASure:HE:CONTinuous
MEASure:HE:SAMPle
MEASure:HE:VOLTage?
MEASure:ADC0?
MEASure:ADC1?
MEASure:ADC2?
82 REV 6
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : SCPI COMMAND SUMMARY
CONFigure:FILL:CHannel {0|1|2}
FILL:CHannel?
CONFigure:FILL:A <level>
FILL:A?
CONFigure:FILL:B <level>
FILL:B?
CONFigure:INTerval:FILL <minutes>
INTerval:FILL?
CONFigure:INTerval:SAMPle <minutes>
INTerval:SAMPle?
CONFigure:HE:TIME_limit <minutes>
HE:TIME_limit?
LOG_LHE {0|1}
LOG_LHE?
HElium_PowerSupply?
DIRTy_sen_mode?
CONFigure:SOURCE:REC_out {0|1|2}
SOURCE:REC_out?
CONFigure:SOURCE:CURrent_LOOP {0|1|2}
SOURCE:CURrent_LOOP?
REV 6 83
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : RS-232 CONFIGURATION
CONFigure:N2:LENgth <value>
N2:LENgth?
MINCAL
MINCAL?
MAXCAL
MAXCAL?
NOSENSorCAL
NOSENSorCAL?
APPROXMAXCAL <value>
APPROXMAXCAL?
CONFigure:HE:LENgth <value>
HE:LENgth?
RS-232 CONFIGURATION
84 REV 6
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : ETHERNET CONFIGURATION
The Model 1700 uses the following fixed parameters related to the RS-
232 interface:
• Parity: No Parity
• Data Bits: 8 Data Bits
• Number of Start Bits: 1 bit
• Number of Stop Bits: 1 bit
• Flow Control: None
SERIAL PORT A PC-compatible computer’s serial port can be directly connected to the
Model 1700 via a standard USB-to-serial cable. Refer to the computer’s
CONNECTOR AND documentation to determine which ports are available on a computer
CABLING and the required connector type.
The Model 1700 uses only three wires of the rear-panel DB9 connector:
pin 2 (transmit), pin 3 (receive), and pin 7 (common). There are no
software or hardware handshaking options. The Model 1700 is classified
as a DCE (Data Communication Equipment) device since it transmits
data on pin 3 and receives data on pin 2. The instrument to which the
Model 1700 is attached must do the opposite, i.e., transmit on pin 2 and
receive on pin 3 (the requirements for a DTE, or Data Terminal
Equipment device).
COMMAND/RETURN All commands are transmitted and received as ASCII values and are
case insensitive. The Model 1700 can be configured to return <CR> (i.e.
TERMINATION a carriage return) or <CR><LF> (i.e. a carriage return followed by a
CHARACTERS linefeed) at the end of a serial transmission (see page 49). The Model
1700 can accept <CR>, <LF>, <CR><LF>, or <LF><CR> as termination
characters from an external computer.
ETHERNET CONFIGURATION
The Model 1700 allows its IP address, subnet mask and gateway IP
address to be assigned either statically or dynamically. To make these
values static and assign them manually, set IP Address Assignment to
Static (see page 48) and then set the values using the Edit parameter
REV 6 85
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : ETHERNET CONFIGURATION
The system name (also known as host name or computer name), can be
set using the Network screen (see page 48) or via remote
communications (either Ethernet or RS-232).
ETHERNET CONNECTOR The Model 1700 uses a standard RJ-45 jack for Ethernet
communications. The Ethernet jack pinout is fully documented on
page 110 in the Appendix.
TERMINATION All commands and queries are transmitted and received as ASCII values
and are case insensitive. The Model 1700 always transmits <CR><LF>
CHARACTERS (a carriage return followed by a linefeed) at the end of an Ethernet
transmission. The Model 430 can accept <CR>, <LF>, <CR><LF>, or
<LF><CR> as termination characters from an external computer.
PORT ASSIGNMENT The Model 1700 accepts remote connections to port 7180. Multiple
connections to port 7180 are allowed.
86 REV 6
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : COMMAND REFERENCE
COMMAND REFERENCE
All commands sent to the Model 1700 are processed and the Model
1700 responds with a return value (if applicable) and termination. If the
command is invalid, the Model 1700 will respond with an error code (see
the Error Codes section). All return values including error codes are
terminated with <CR><LF> (i.e. a carriage return followed by a linefeed).
For those commands that do not return a value, the Model 1700 will
return the <CR><LF> termination only.
The remote units settings are saved in non-volatile memory and are
restored at power-up.
COMMANDS Returns the identification string of the Model 1700. The identification
string contains the manufacturer name, model number, serial number,
and firmware revision code. Example output:
• *RST
• *TST?
• N2?
• HE?
Returns “0” if the instrument is not configured to read liquid helium level,
“1” if the instrument is configured to read 4.2K liquid helium level for
sensors of active length ≤ 40 inches, “2” if the instrument is configured to
REV 6 87
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : COMMAND REFERENCE
read 4.2K liquid helium level for sensors of active length ≤ 80 inches, “3”
if the instrument is configured to read 2K liquid helium level for sensors
of active length ≤ 40 inches, or “4” if the instrument is configured to read
2K liquid helium level for sensors of active length ≤ 80 inches.
• SERial_NUMber?
• DATE_MANUfacture?
• HardWare_VERsion?
• FirmWare_VERsion?
• Scpi_VERsion?
• SYStem:BEEPer:IMMediate <seconds>
• SYStem:BEEPer:STATe {0|1}
• SYStem:BEEPer:STATe?
• SYStem:KLOCK {0|1}
88 REV 6
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : COMMAND REFERENCE
interfaces. The state of the lock defaults to “0” (OFF) when the
instrument is restarted.
• SYStem:KLOCK?
Returns the state of the instrument local controls lock. A return value of
“0” indicates the lock is OFF. A return value of “1” indicates the local
interfaces, such as the touch display, of the instrument are locked.
• SYStem:DATE <yyyy>,<mm>,<dd>
Sets the date per <yyyy> as a four-digit number; <mm> month with
range of 1 to 12 inclusive; <dd> day of the month. Do not include the <>
characters in the command.
• SYStem:DATE?
• SYStem:TIME <hh>,<mm>,<ss>
• SYStem:TIME?
• SYStem:REBOOT
• SYStem:RESTORE
Reboots the instrument and sets all parameters back to factory defaults.
All prior settings are lost!
• IPV4ADDR?
• MACADDR?
REV 6 89
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : COMMAND REFERENCE
• GATEWAY?
• HOSTname?
• SCREENCAP
• SYSTem:LOCal
Enables the front panel touchscreen. All front panels controls are
enabled by default after a power-up or REBOOT command.
• SYSTem:REMote
DISPLAY • DISPLAY:N2?
CONFIGURATION Returns a “0” if the instrument is not configured to display liquid nitrogen
COMMANDS AND level on the home screen and a “1” if it is.
QUERIES
• CONFigure:NAME:SENSor:N2 "<string>"
• NAME:SENSor:N2?
• DISPLAY:HE?
• CONFigure:NAME:SENSor:HE "<string>"
• NAME:SENSor:HE?
90 REV 6
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : COMMAND REFERENCE
COMMANDS AND Assigns relay №1 as disabled (0), or to nitrogen (1) or helium (2).
QUERIES
• RELay1:CHannel?
• CONFigure:RELay1:SETpoint <level>
• RELay1:SETpoint?
• CONFigure:RELay1:OPeration {0|1}
Configures relay №1 such that it closes when the level is ≤ the setpoint
(0) or ≥ the setpoint (1).
• RELay1:OPeration?
Returns a “0” if relay №1 closes when the level is ≤ the setpoint and a “1”
if the relay closes when the relay is ≥ the setpoint.
• CONFigure:RELay2:CHannel {0|1|2}
• RELay2:CHannel?
• CONFigure:RELay2:SETpoint <level>
• RELay2:SETpoint?
• CONFigure:RELay2:OPeration {0|1}
Configures relay №2 such that it closes when the level is ≤ the setpoint
(0) or ≥ the setpoint (1)
REV 6 91
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : COMMAND REFERENCE
• RELay2:OPeration?
Returns a “0” if relay №2 closes when the level is ≤ the setpoint and a “1”
if the relay closes when the level is ≥ the setpoint.
• {RELay1|RELay2}:STATus?
Returns the present status of the specified relay. A return value of “0”
indicates the relay is not closed. A value of “1” indicates the relay is
closed.
• CONFigure:ALArm1:SETpoint <level>
• ALArm1:SETpoint?
• CONFigure:ALArm1:OPeration {0|1}
Configures alarm №1 such that it alarms when the level is ≤ the setpoint
(0) or ≥ the setpoint (1).
• ALArm1:OPeration?
Returns a “0” if alarm №1 alarms when the level is ≤ the setpoint and a
“1” if it alarms when the level is ≥ the setpoint. Alarm №1 is considered
as the “legacy” HI level alarm when the alarm condition is set to level ≥ a
setpoint.
• CONFigure:ALArm2:CHannel {0|1|2}
92 REV 6
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : COMMAND REFERENCE
• ALArm2:CHannel?
• CONFigure:ALArm2:SETpoint <level>
• ALArm2:SETpoint?
• CONFigure:ALArm2:OPeration {0|1}
Configures alarm №2 such that it alarms when the level is ≤ the setpoint
(0) or ≥ the setpoint (1)
• ALArm2:OPeration?
Returns a “0” if alarm №2 alarms when the level is ≤ the setpoint and a
“1” if it alarms when the level is ≥ the setpoint. Alarm №2 is considered
as the “legacy” LO level alarm when the alarm condition is set to level ≤
a setpoint.
• {ALArm1|ALArm2}:STATus?
Returns the present status of the specified alarm. A return value of “0”
indicates the alarm is inactive. A value of “1” indicates the alarm is
active.
Mutes the audible alarm for any active alarm(s) when set to “1” or “YES”.
Unmutes the audible alarm if set to “0” or “NO”.
• ALARm:MUTE?
MEASUREMENT • MEASure:N2:LEVel?
COMMANDS AND Returns the liquid nitrogen level in the current units.
QUERIES
• MEASure:N2:PERIod?
REV 6 93
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : COMMAND REFERENCE
• MEASure:HE:LEVel?
• MEASure:HE:HOLD
• MEASure:HE:CONTinuous
• MEASure:HE:SAMPle
Energizes the liquid helium level sensor, makes a reading and returns to
Sample and Hold mode.
• MEASure:HE:VOLTage?
Returns the last measured liquid helium level sensor voltage in volts.
• MEASure:ADC0?
Returns the last measured liquid helium level sensor voltage in volts.
• MEASure:ADC1?
• MEASure:ADC2?
QUERIES Assigns the auto fill control relay to either no channel (disabled) (0),
nitrogen (1), or helium (2).
• FILL:CHannel?
Returns a “0” if the auto fill relay is disabled, a “1” if the relay is assigned
to the nitrogen channel, and a “2” if the relay is assigned to the helium
channel.
94 REV 6
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : COMMAND REFERENCE
• CONFigure:FILL:A <level>
Sets the A setpoint (control band upper limit) in the assigned channel’s
current units. The A setpoint (autofill stop) must be greater than the B
setpoint (fill start) and must also be between 0% and 100%.
• FILL:A?
• CONFigure:FILL:B <level>
Sets the B setpoint (control band lower limit) in the assigned channel’s
current units. The B setpoint (autofill start) must be less than the A
setpoint (fill stop) and must also be between 0% and 100%.
• FILL:B?
• CONFigure:INTerval:FILL <minutes>
Sets the fill timer in minutes. If the level does not reach or exceed the A
setpoint within the fill time, the fill is terminated. Setting the value of FILL
to “0” disables the fill timer function.
• INTerval:FILL?
• CONFigure:HE:TIME_limit <minutes>
REV 6 95
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : COMMAND REFERENCE
• HE:TIME_limit?
• LOG_LHE {0|1}
• LOG_LHE?
Returns “0” if the liquid helium level logging is disabled, or “1” if enabled.
• HElium_PowerSupply?
• DIRTy_sen_mode?
Returns a “0” if the instrument’s dirty sensor mode is disabled for the
helium channel and a “1” if it is enabled.
COMMANDS AND Configures the 0-10 VDC Recorder Output source to disabled (0),
QUERIES assigned to the nitrogen channel (1), or the helium channel (2).
• SOURCE:REC_out?
• CONFigure:SOURCE:CURrent_LOOP {0|1|2}
• SOURCE:CURrent_LOOP?
96 REV 6
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : COMMAND REFERENCE
CALIBRATION Configures the liquid nitrogen sensor active length in current units.
COMMANDS AND Returns an error code if the current units are percent.
QUERIES
• N2:LENgth?
• MINCAL
Performs and saves a MIN calibration point calibration for the liquid
nitrogen sensor.
• MINCAL?
Returns the last saved MIN calibration point in microseconds for the
liquid nitrogen sensor.
• MAXCAL
Performs and saves a MAX calibration point calibration for the liquid
nitrogen sensor.
• MAXCAL?
Returns the last saved MAX calibration point in microseconds for the
liquid nitrogen sensor.
• NOSENSorCAL
Calibrates and saves the LOSS OF SENSOR condition for the liquid
nitrogen sensor.
• NOSENSorCAL?
• APPROXMAXCAL <value>
Sets the Approx Cal (Max) Multiplier factor (see page 70) to a decimal
value. The value is nominally 1.000 for no modification to the MAX
calibration point.
• APPROXMAXCAL?
Returns the Approx Cal (Max) Multiplier factor (see page 70) as a
decimal value.
REV 6 97
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : COMMAND REFERENCE
CALIBRATION Configures the liquid helium sensor active length in current units.
COMMANDS AND Returns an error if the current units are percent.
QUERIES
• HE:LENgth?
COMMANDS AND Sets the liquid nitrogen level units of measurement to percent (0 or
QUERIES PERCENT), inches (1 or INCH), or centimeters (2 or CM). The default is
PERCENT units.
• N2:UNIT?
• HE:UNIT?
98 REV 6
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : ERROR CODES
ERROR CODES
The Model 1700 returns specific error codes for invalid commands and/
or arguments. If an error condition is returned, the command is not
processed and the configuration of the instrument is not modified. The
table below provides a list of error codes, their meaning, and any
associated limits.
Error
Code Meaning Valid Range
-1 LO (or relay №2a) setpoint out of range 0 LO (or relay №2a) LENGTH
-4 HI (or relay №1a) setpoint out of range 0 HI (or relay №1a) LENGTH
-8 Unrecognized command
-11 Command exceeds SCPI input buffer limit 256 characters, including spaces, etc.
REV 6 99
AMERICAN MAGNETICS, INC. REMOTE INTERFACE REFERENCE : ERROR CODES
100 REV 6
AMERICAN MAGNETICS, INC. SERVICE AND REPAIR
CLEANING
Replacement parts for the instrument are listed in the table below.
EF1700 Fuse, 3 A, 250 Vac, 5x20 mm, fast acting, UL/CSA recognized.
REV 6 101
AMERICAN MAGNETICS, INC. SERVICE AND REPAIR : BATTERY REPLACEMENT
BATTERY REPLACEMENT
This section describes the procedure for replacing the battery on the
instrument's main circuit board.
LOW BATTERY Battery status is displayed in the SYSTEM DATE & TIME dialog. To
access the dialog:
INDICATION
1. From the home screen, choose
MENU.
102 REV 6
AMERICAN MAGNETICS, INC. SERVICE AND REPAIR : BATTERY REPLACEMENT
TOOLS REQUIRED • Torx Plus (T.M.) size 10 driver (Wera 028034 or equivalent)
• Torx Plus (T.M.) size 15 driver (Wera 028035 or equivalent)
• Small, flat-blade screw driver (for prying)
2. Using the T-15 driver, remove the four 8-32 machine screws on the sides of
the instrument cover. Set these screws aside as they will be re-used.
3. Using the T-10 driver, remove the four 6-32 machine screws on the rear of
the instrument cover. Set these screws aside as they will be re-used.
4. Lift the instrument cover off of the instrument chassis and set aside.
REV 6 103
AMERICAN MAGNETICS, INC. SERVICE AND REPAIR : FUSE REPLACEMENT
5. Using the small, flat-blade screwdriver, carefully pry the battery from the
holder BH1.
7. Replace the top cover and secure using the eight machine screws which
were removed previously.
FUSE REPLACEMENT
This section describes the procedure for replacing the two fuses on the
instrument's main circuit board.
TOOLS REQUIRED • Torx Plus (T.M.) size 10 driver (Wera 028034 or equivalent)
• Torx Plus (T.M.) size 15 driver (Wera 028035 or equivalent)
• Small, flat-blade screw driver (for prying)
2. Using the T-15 driver, remove the four 8-32 machine screws on the sides of
the instrument cover. Set these screws aside as they will be re-used.
3. Using the T-10 driver, remove the four 6-32 machine screws on the rear of
the instrument cover. Set these screws aside as they will be re-used.
4. Lift the instrument cover off of the instrument chassis and set aside.
5. Using the small, flat-blade screwdriver, carefully pry the fuse(s) from the
fuse holders F1 and/or F2.
6. Install the new fuse(s) into the fuse holder(s) F1 and/or F2.
7. Replace the top cover and secure using the eight machine screws which
were removed previously.
104 REV 6
AMERICAN MAGNETICS, INC. SERVICE AND REPAIR : FIRMWARE UPGRADE VIA ETHERNET
1. Connect the instrument to a computer network via Ethernet cable and ener-
gize the instrument.
3. Determine the currently installed version of firmware. From the home page,
press MENU > SYSTEM > FW Version.
5. Start Windows Explorer and enter the instrument's IP address noted from
Step 2 above in the address field. Be sure to use the format such as
\\192.168.1.101. Ensure the Explorer screen looks similar to the
following.
REV 6 105
AMERICAN MAGNETICS, INC. SERVICE AND REPAIR : FIRMWARE UPGRADE VIA ETHERNET
7. Drag the firmware file to the instrument's Application Update folder in the
first Explorer screen.
8. Install the new firmware by pressing MENU > SYSTEM > UPDATE on the
instrument and choose the name of the firmware file copied to the
instrument. It is likely the top choice of the files listed.
10. The instrument will install the new firmware and reboot when completed.
UPGRADE VIA SCP If the instrument’s file shares are not visible on the local network, it is
possible to copy the firmware upgrade directly to the instrument using
the root login and password using SCP (Secure Copy) over Ethernet.
1. Connect the instrument to the local computer network via Ethernet cable
and energize the instrument.
3. Determine the currently installed version of firmware. From the home page,
press MENU > SYSTEM > FW Version.
106 REV 6
AMERICAN MAGNETICS, INC. SERVICE AND REPAIR : FIRMWARE UPGRADE VIA ETHERNET
5. Open a command line application and set the current directory to the folder
where the newly download firmware from the prior step resides.
Note that you will prompted to enter the root account password to complete
the file transfer. If you do not know the password, contact an Authorized AMI
Technical Support Representative for assistance.
7. Install the new firmware by pressing MENU > SYSTEM > UPDATE on the
instrument and choose the name of the firmware file copied to the
instrument via SCP. It is likely the top choice of the files listed.
The instrument will install the new firmware and reboot when completed.
REV 6 107
AMERICAN MAGNETICS, INC. SERVICE AND REPAIR : FIRMWARE UPGRADE VIA ETHERNET
108 REV 6
AMERICAN MAGNETICS, INC. APPENDIX
APPENDIX
1 N/C
6 N/C
7 N/C
8 N/C
9 N/C
REV 6 109
AMERICAN MAGNETICS, INC. APPENDIX : ETHERNET CONNECTOR
ETHERNET CONNECTOR
Pin 1
4
not used
5
110 REV 6
AMERICAN MAGNETICS, INC. APPENDIX : LIQUID HELIUM CONNECTOR J1 WIRING
5
1
9
6
1 Sensor I+ Red
2 +48 VDC P/S Test Point
6 Sensor V- Yellow
7 Sensor I- Black
8 Sensor V+ Blue
REV 6 111
AMERICAN MAGNETICS, INC. APPENDIX : AUX I/O CONNECTOR
1
4-20 mA Current Loop
2
3
0-10 VDC Output
4
5
Relay № 1 Dry Contact N/A
6
7
Relay № 2 Dry Contact N/A
8
9 N/A
112 REV 6
AMERICAN MAGNETICS, INC. APPENDIX : DIELECTRIC CONSTANTS FOR CRYOGENIC LIQUIDS
REV 6 113
AMERICAN MAGNETICS, INC. APPENDIX : TROUBLESHOOTING
TROUBLESHOOTING
support@americanmagnetics.com
1. The threshold is the midpoint between the MINimum CALibration and the NO SENSOR CALi-
bration.
2. The threshold is approximately 1 mA.
3. The threshold is approximately 200 kohms with a sensor capacitance of 2 nF.
114 REV 6
AMERICAN MAGNETICS, INC. APPENDIX : TROUBLESHOOTING
that the sensor resistance measured across the BNC connector should
be >107 ohms.1
SYSTEM TEST SCREENS The instrument provides various system test functions that can be
performed from the front panel touchscreen. The system test functions
are not available to the web browser interface nor the SCPI remote
interface over serial or Ethernet connections. To access the test
functions:
1. Any resistance between the sensor elements will cause the instrument to read higher than
actual level.
2. The threshold is 5% greater resistance than the SENSOR ACTIVE LENGTH resistance.
REV 6 115
AMERICAN MAGNETICS, INC. APPENDIX : TROUBLESHOOTING
116 REV 6
AMERICAN MAGNETICS, INC. APPENDIX : TROUBLESHOOTING
NOTE In the AUTO mode, the Test screen displays the real time
measurement or state of each labeled function.
NOTE When changing to MANUAL test mode, please be aware that the
automatic functions of the instrument will be overridden.
Functions such as autofill will be interrupted, or a fill initiated, by
the manual actions. Be sure the system is in a safe condition
before exercising manual control.
REV 6 117
AMERICAN MAGNETICS, INC. APPENDIX : TROUBLESHOOTING
NOTE The hardware control state will return to AUTO after a few
seconds if the operator does not explicitly return to automatic
control. However, the state of the hardware or autofill functions
are not guaranteed to be the same as when the MANUAL control
state was entered.
SYSTEM LOGS The Model 1700 provides a logging feature that saves detailed
information regarding the liquid level and system events. The information
can be very useful in troubleshooting suspected system errors.
The Model 17XX family of liquid level instruments keeps log files in a
local Ethernet network accessible folder via Samba file shares. The log
folder is labeled “Log files” and can found on a network by opening a
file explorer at the instrument IP address or hostname.
118 REV 6
AMERICAN MAGNETICS, INC. APPENDIX : TROUBLESHOOTING
The status bits contains information for both the channels, even if the
channels are disabled.
The status bits are a summation of the hardware flags, each flag has a
unique bit position, and so the status byte displayed is a hexadecimal
representation of those bits.
Refer to the Status Bits table on page 122 for the bit values assigned for
each hardware flag.
A log entry is made when either the level changes by 0.1% or if the
status bits change from last entry. Levels and bits are checked every
second.
REV 6 119
AMERICAN MAGNETICS, INC. APPENDIX : TROUBLESHOOTING
Entry #2: timestamped at Mon Jul 19 2021 10:58:22 with a last read
level of 45.1. The status bits (028) indicated that the external/internal N2
sensor oscillator selection was internal, and that the helium sensor was
active and dirty sensor mode was off. The level had changed between
entry #2 and #3, so when the unit stopped measuring (turning off the
helium sensor) it updated the level, which changed to 45.1
Refer to the Status Bits table on page 122 for the bit values assigned for
each hardware flag.
Entries #2-#6: shows the level changing over a 5 second span, settling
at 16.1%
Refer to the Log Status Bits table on page 122 for information on the
status bits.
120 REV 6
AMERICAN MAGNETICS, INC. APPENDIX : TROUBLESHOOTING
1626712173,LL,MENU,menu.aml,
Menu.aml indicated the screen page loaded was menu.aml, that is, the
MENU page.
REV 6 121
AMERICAN MAGNETICS, INC. APPENDIX : TROUBLESHOOTING
The class LL indicates that this was a link label button event, that is, the
label on the button was MENU, and that the linked page loaded was
menu.aml.
1626712191,LL,SHUTDOWN,../shutdown.aml,
The operator pressed the SHUTDOWN menu button which then loaded
the ../shutdown.aml page.
1626712200,LL,SHUTDOWN ,../shutdown.aml,
1626712200,SF,shmem_dump,checksum,6A7E48E4
The instrument saved the variables to the state file. Class SF indicates
this is a state file operation.
1626712200,CE,shutdown,,
The command executed (CE) was the shutdown command. The data
fields #2 and #3 in this line are empty as shown by the two commas after
the shutdown command.
122 REV 6
AMERICAN MAGNETICS, INC. GLOSSARY
GLOSSARY
Term Meaning
AC; ac Alternating Current; strictly, electrical current that periodically reverses direction.
Typically used also to describe an electrical power source in terms of the voltage. For
example, 240 VAC.
ASCII American Standard Code for Information Interchange; numerical representation of
characters such as 'a' or '@' or an action (such as line-feed); 'plain' raw text with no
formatting such as tabs, bold or underscoring
BNC connector A miniature quick connect/disconnect RF connector used for coaxial cable, featuring
two bayonet lugs on the female connector.
Cryogen A substance for obtaining low temperatures. In the case of use with the Model 1700
instrument, a cryogen is a liquefied gas such as liquid nitrogen or liquid helium.
D-Sub Term referring to the family of connectors containing an odd number of pins in two
parallel rows with a 1-pin difference in pins-per-row (DB9, DB15, and DB25 are most
common)
DB9 Type of electrical connector containing 9 pins arranged in two parallel rows of 4 pins
and 5 pins each
DCE Data Circuit-terminating Equipment - a device that sits between the Data Terminal
Equipment (DTE) and a data transmission circuit.
DHCP Dynamic Host Configuration Protocol; a computer networking protocol which dynam-
ically distributes the IP address to networked devices
dt Rate of change
DTE Data Terminal Equipment - an end instrument that converts user information into sig-
nals or reconverts received signals. A DTE device communicates with the Data Circuit-
terminating Equipment (DCE).
REV 6 123
AMERICAN MAGNETICS, INC. GLOSSARY : ABBREVIATIONS AND ACRONYMS
Term Meaning
ECL Electrical Connection Lubricant - also known as Dielectric Tune-up Grease, a protec-
tive lubricant that prevents corrosion.
I/O Input/Output; The hardware and associated protocol that implement communication
between information processing systems and/or devices. Inputs are the signals or
data received by the system or device, and outputs are the signals or data sent from
it.
IP Internet Protocol; when used with “address”, refers to a numerical Internet address
kG kilogauss: a magnetic field unit of measurement
LED Light-Emitting Diode; a semiconductor device that emits light when energized - used
for visual status indication
Max Maximum
Min Minimum
MSDS Material Safety Data Sheet - provides workers and emergency personnel with proce-
dures for handling or working with a specific substance in a safe manner and includes
information such as physical data, toxicity, health effects, first aid, reactivity, storage,
disposal, protective equipment, and spill-handling procedures.
RG-59/U A specific type of coaxial cable, often used for low-power video and RF signal connec-
tions, with a characteristic impedance of 75 ohms.
RS-232 RS-232 is a long-established standard and protocol for relatively low speed serial data
communication between computers and related devices; originally established for
teletypewriter communication.
V Volts
124 REV 6
AMERICAN MAGNETICS, INC. GLOSSARY : ABBREVIATIONS AND ACRONYMS
Term Meaning
Vm Magnet voltage
REV 6 125
AMERICAN MAGNETICS, INC. GLOSSARY : ABBREVIATIONS AND ACRONYMS
126 REV 6
AMERICAN MAGNETICS, INC.
Index INDEX
A C
abbreviations and acronyms 123 cables
capacitance sensor coax 13
abnormal operation 50 long-distance 13
Alarm Status screen 39 serial 85
alarms 39 calibration
alarm status screen 39 approximate 56, 70–73
configure source and level 41 approx cal equation 71
indication and acknowledgement 41 choosing a method 57
multiple 40 closed dewar 65–69
muting 39, 42 completing closed dewar cal 67
remote configuration 92 increasing the range 68
states 41 preset MIN/MAX 65
AMI contacts 114 diagram 58
dielectric constants 113
analog outputs LHe active length 76
connection 20 liquid helium sensors 74
connector 112 liquid level sensors 55
current loop diagram 21 loss of sensor detection 61
current loop output 20 open dewar 63–64
recorder output 20 presetting MAX/MIN 65
remote configuration 96 remote LHe calibration 98
simultaneous usage 20 remote LN2 calibration 97
source selection 47 stabilizing level during cal 67
specifications 9 substitute liquid or partial length 70
approximate calibration 70–73 touch screen 116
using analog outputs 67
autofill
AUTO-OFF 45 capacitance sensor
AUTO-ON 45 active length 55
home screen indicator 24 approximate calibration 70–73
loss of sensor response 19 avoiding icing 12
manual overrides 45 calibration diagram 58
M-CLOSED 45 calibration methods 55
M-OPEN 45 cleaning 51
output rating 9 closed dewar calibration 65–69
parameters 44 coax cable specs 13
relief valve caution 18 configuration 33
remote configuration 94 connecting 13
socket 9 dielectric variations 56
source selection 44 ECL coating 12
standard 18 exposure to humidity 51
start and stop levels 44 installation 12
states 45 loss of sensor 61, 114
system diagram 17 measurement method 56
system setup 17 open dewar calibration 63–64
timeout 9, 40, 44 sensor length 60
timeout reset 45 sensor name 61
shorted condition 114
AUTO-OFF 45 top vent hole 12
AUTO-ON 45 vent holes 55
choosing units 46
B closed dewar calibration 65, 65–69
battery backup 10 presetting MIN/MAX 65
REV 6 127
AMERICAN MAGNETICS, INC. INDEX
128 REV 6
AMERICAN MAGNETICS, INC. INDEX
I M
ice formation 12 M-CLOSED 45
installation measurement
autofill system 17 continuous mode time limit 35, 76
capacitance sensors 12 LHe burnout protection 36
liquid helium sensors 14 remote query 93
mounting 11 sampling interval for LHe 36
power 20 sampling mode for LHe 37, 75
unpacking 11
menu tree
installing overview 26
earth ground xi
menus 25
instrument descriptions 27
battery replacement 102 overall structure 26
cleaning 101
fuse replacement 104
method of measurement 56
part number definition xii moisture protection 12, 13, 51
replacement parts 101
troubleshooting 114 M-OPEN 45
instrument description 1 mounting
coaxial cable 12
ip address 85 external oscillator 12
helium sensor 14
L removing bottom feet 11
top vent hole 12
level damping 47
muting alarms 42
level units 46
liquid helium sensor
REV 6 129
AMERICAN MAGNETICS, INC. INDEX
130 REV 6
AMERICAN MAGNETICS, INC. INDEX
serial communications U
baud rate 9, 31, 50
commands 87 units 46
connector 9 remote configuration 98
connector pin out 109 user replaceable parts 101
factory reset 89
interactive communication 79
set date 89
V
set time 89 vacuum caution
system reboot 89 helium sensor 14
terminators 85
vent holes 55
serial port
baud rate 50
echo 50
W
function 50 warnings
setup 49 equipment xi
termination 50, 85
weight 10
serial port connector/cables 85
service
AMI contact 114
service - see instrument
shutdown 52
specifications
analog output 9
autofill 9
display 8
environmental limits 10
level measurements 8
physical 10
power 10
relays 9
standards conformance 10
system
date and time 53–54
logging 118
remote configuration 87
shutdown 52
system features 1
system specifications 8
system test
analog outputs 117
hardware controls 117
screen pixels 116
touch screen cal 116
T
test points for LHe sensor 111
TIMEZONE selection 54
troubleshooting common issues 114
system test screens 115
REV 6 131
AMERICAN MAGNETICS, INC. INDEX
132 REV 6