Method 3 - Gas Analysis For The Determination of Dry Molecular Weight
Method 3 - Gas Analysis For The Determination of Dry Molecular Weight
Method 3 - Gas Analysis For The Determination of Dry Molecular Weight
version. Please refer to the official version in the FR publication, which appears on the Government Printing
Office's FDSys eCFR website:
(http://www.ecfr.gov/cgi-bin/text-idx?c=ecfr&tpl=/ecfrbrowse/Title40/40cfr60_main_02.tpl).
Note: This method does not include all of the specifications (e.g., equipment and supplies) and
procedures (e.g., sampling) essential to its performance. Some material is incorporated by
reference from other methods in this part. Therefore, to obtain reliable results, persons using this
method should also have a thorough knowledge of Method 1.
1.1 Analytes.
1.2 Applicability. This method is applicable for the determination of CO 2 and O2 concentrations
and dry molecular weight of a sample from an effluent gas stream of a fossil-fuel combustion
process or other process.
1.3 Other methods, as well as modifications to the procedure described herein, are also
applicable for all of the above determinations. Examples of specific methods and modifications
include: (1) A multi-point grab sampling method using an Orsat analyzer to analyze the
individual grab sample obtained at each point; (2) a method for measuring either CO 2 or O2 and
using stoichiometric calculations to determine dry molecular weight; and (3) assigning a value of
30.0 for dry molecular weight, in lieu of actual measurements, for processes burning natural gas,
coal, or oil. These methods and modifications may be used, but are subject to the approval of the
Administrator. The method may also be applicable to other processes where it has been
determined that compounds other than CO2, O2, carbon monoxide (CO), and nitrogen (N2) are
not present in concentrations sufficient to affect the results.
1.4 Data Quality Objectives. Adherence to the requirements of this method will enhance the
quality of the data obtained from air pollutant sampling methods.
2.1 A gas sample is extracted from a stack by one of the following methods: (1) single-point,
grab sampling; (2) single-point, integrated sampling; or (3) multi-point, integrated sampling. The
gas sample is analyzed for percent CO2 and percent O2. For dry molecular weight determination,
either an Orsat or a Fyrite analyzer may be used for the analysis.
3.0 Definitions[Reserved]
4.0 Interferences
4.1 Several compounds can interfere, to varying degrees, with the results of Orsat or Fyrite
analyses. Compounds that interfere with CO2 concentration measurement include acid gases
(e.g., sulfur dioxide, hydrogen chloride); compounds that interfere with O 2 concentration
measurement include unsaturated hydrocarbons (e.g., acetone, acetylene), nitrous oxide, and
ammonia. Ammonia reacts chemically with the O2 absorbing solution, and when present in the
effluent gas stream must be removed before analysis.
5.0 Safety
5.1 Disclaimer. This method may involve hazardous materials, operations, and equipment. This
test method may not address all of the safety problems associated with its use. It is the
responsibility of the user of this test method to establish appropriate safety and health practices
and determine the applicability of regulatory limitations prior to performing this test method.
5.2.1 A typical Orsat analyzer requires four reagents: a gas-confining solution, CO2 absorbent,
O2 absorbent, and CO absorbent. These reagents may contain potassium hydroxide, sodium
hydroxide, cuprous chloride, cuprous sulfate, alkaline pyrogallic acid, and/or chromous chloride.
Follow manufacturer's operating instructions and observe all warning labels for reagent use.
5.2.2 A typical Fyrite analyzer contains zinc chloride, hydrochloric acid, and either potassium
hydroxide or chromous chloride. Follow manufacturer's operating instructions and observe all
warning labels for reagent use.
Note: As an alternative to the sampling apparatus and systems described herein, other sampling
systems (e.g., liquid displacement) may be used, provided such systems are capable of obtaining
a representative sample and maintaining a constant sampling rate, and are, otherwise, capable of
yielding acceptable results. Use of such systems is subject to the approval of the Administrator.
6.1.1 Probe. Stainless steel or borosilicate glass tubing equipped with an in-stack or out-of-stack
filter to remove particulate matter (a plug of glass wool is satisfactory for this purpose). Any
other materials, resistant to temperature at sampling conditions and inert to all components of the
3
gas stream, may be used for the probe. Examples of such materials may include aluminum,
copper, quartz glass, and Teflon.
6.1.2 Pump. A one-way squeeze bulb, or equivalent, to transport the gas sample to the analyzer.
6.2.4 Pump. A leak-free, diaphragm-type pump, or equivalent, to transport sample gas to the
flexible bag. Install a small surge tank between the pump and rate meter to eliminate the
pulsation effect of the diaphragm pump on the rate meter.
6.2.5 Rate Meter. A rotameter, or equivalent, capable of measuring flow rate to ±2 percent of the
selected flow rate. A flow rate range of 500 to 1000 ml/min is suggested.
6.2.6 Flexible Bag. Any leak-free plastic (e.g., Tedlar, Mylar, Teflon) or plastic-coated
aluminum (e.g., aluminized Mylar) bag, or equivalent, having a capacity consistent with the
selected flow rate and duration of the test run. A capacity in the range of 55 to 90 liters (1.9 to
3.2 ft3) is suggested. To leak-check the bag, connect it to a water manometer, and pressurize the
bag to 5 to 10 cm H2O (2 to 4 in. H2O). Allow to stand for 10 minutes. Any displacement in the
water manometer indicates a leak. An alternative leak-check method is to pressurize the bag to 5
to 10 cm (2 to 4 in.) H2O and allow to stand overnight. A deflated bag indicates a leak.
6.2.7 Pressure Gauge. A water-filled U-tube manometer, or equivalent, of about 30 cm (12 in.),
for the flexible bag leak-check.
6.2.8 Vacuum Gauge. A mercury manometer, or equivalent, of at least 760 mm (30 in.) Hg, for
the sampling train leak-check.
7.2 Standards. Two standard gas mixtures, traceable to National Institute of Standards and
Technology (NIST) standards, to be used in auditing the accuracy of the analyzer and the
analyzer operator technique:
4
7.2.2. Gas cylinder containing 2 to 4 percent CO2 and about 15 percent O2.
8.1.1 The sampling point in the duct shall either be at the centroid of the cross section or at a
point no closer to the walls than 1.0 m (3.3 ft), unless otherwise specified by the Administrator.
8.1.2 Set up the equipment as shown in Figure 3–1, making sure all connections ahead of the
analyzer are tight. If an Orsat analyzer is used, it is recommended that the analyzer be leak-
checked by following the procedure in Section 11.5; however, the leak-check is optional.
8.1.3 Place the probe in the stack, with the tip of the probe positioned at the sampling point.
Purge the sampling line long enough to allow at least five exchanges. Draw a sample into the
analyzer, and immediately analyze it for percent CO2 and percent O2 according to Section 11.2.
8.2.1 The sampling point in the duct shall be located as specified in Section 8.1.1.
8.2.2 Leak-check (optional) the flexible bag as in Section 6.2.6. Set up the equipment as shown
in Figure 3–2. Just before sampling, leak-check (optional) the train by placing a vacuum gauge at
the condenser inlet, pulling a vacuum of at least 250 mm Hg (10 in. Hg), plugging the outlet at
the quick disconnect, and then turning off the pump. The vacuum should remain stable for at
least 0.5 minute. Evacuate the flexible bag. Connect the probe, and place it in the stack, with the
tip of the probe positioned at the sampling point. Purge the sampling line. Next, connect the bag,
and make sure that all connections are tight.
8.2.3 Sample Collection. Sample at a constant rate (±10 percent). The sampling run should be
simultaneous with, and for the same total length of time as, the pollutant emission rate
determination. Collection of at least 28 liters (1.0 ft 3) of sample gas is recommended; however,
smaller volumes may be collected, if desired.
8.2.4 Obtain one integrated flue gas sample during each pollutant emission rate determination.
Within 8 hours after the sample is taken, analyze it for percent CO 2 and percent O2 using either
an Orsat analyzer or a Fyrite type combustion gas analyzer according to Section 11.3.
Note: When using an Orsat analyzer, periodic Fyrite readings may be taken to verify/confirm the
results obtained from the Orsat.
8.3.2 Follow the procedures outlined in Sections 8.2.2 through 8.2.4, except for the following:
Traverse all sampling points, and sample at each point for an equal length of time. Record
sampling data as shown in Figure 3–3.
10.1 Analyzer. The analyzer and analyzer operator's technique should be audited periodically as
follows: take a sample from a manifold containing a known mixture of CO2 and O2, and analyze
according to the procedure in Section 11.3. Repeat this procedure until the measured
concentration of three consecutive samples agrees with the stated value ±0.5 percent. If
necessary, take corrective action, as specified in the analyzer users manual.
10.2 Rotameter. The rotameter need not be calibrated, but should be cleaned and maintained
according to the manufacturer's instruction.
11.1 Maintenance. The Orsat or Fyrite-type analyzer should be maintained and operated
according to the manufacturers specifications.
11.2 Grab Sample Analysis. Use either an Orsat analyzer or a Fyrite-type combustion gas
analyzer to measure O2 and CO2 concentration for dry molecular weight determination, using
procedures as specified in the analyzer user's manual. If an Orsat analyzer is used, it is
recommended that the Orsat leak-check, described in Section 11.5, be performed before this
determination; however, the check is optional. Calculate the dry molecular weight as indicated in
Section 12.0. Repeat the sampling, analysis, and calculation procedures until the dry molecular
weights of any three grab samples differ from their mean by no more than 0.3 g/g-mole (0.3
6
lb/lb-mole). Average these three molecular weights, and report the results to the nearest 0.1 g/g-
mole (0.1 lb/lb-mole).
11.3 Integrated Sample Analysis. Use either an Orsat analyzer or a Fyrite-type combustion gas
analyzer to measure O2 and CO2 concentration for dry molecular weight determination, using
procedures as specified in the analyzer user's manual. If an Orsat analyzer is used, it is
recommended that the Orsat leak-check, described in Section 11.5, be performed before this
determination; however, the check is optional. Calculate the dry molecular weight as indicated in
Section 12.0. Repeat the analysis and calculation procedures until the individual dry molecular
weights for any three analyses differ from their mean by no more than 0.3 g/g-mole (0.3 lb/lb-
mole). Average these three molecular weights, and report the results to the nearest 0.1 g/g-mole
(0.1 lb/lb-mole).
11.4 Standardization. A periodic check of the reagents and of operator technique should be
conducted at least once every three series of test runs as outlined in Section 10.1.
11.5 Leak-Check Procedure for Orsat Analyzer. Moving an Orsat analyzer frequently causes it
to leak. Therefore, an Orsat analyzer should be thoroughly leak-checked on site before the flue
gas sample is introduced into it. The procedure for leak-checking an Orsat analyzer is as follows:
11.5.1 Bring the liquid level in each pipette up to the reference mark on the capillary tubing, and
then close the pipette stopcock.
11.5.2 Raise the leveling bulb sufficiently to bring the confining liquid meniscus onto the
graduated portion of the burette, and then close the manifold stopcock.
11.5.4 Observe the meniscus in the burette and the liquid level in the pipette for movement over
the next 4 minutes.
11.5.5 For the Orsat analyzer to pass the leak-check, two conditions must be met:
11.5.5.1 The liquid level in each pipette must not fall below the bottom of the capillary tubing
during this 4-minute interval.
11.5.5.2 The meniscus in the burette must not change by more than 0.2 ml during this 4-minute
interval.
11.5.6 If the analyzer fails the leak-check procedure, check all rubber connections and stopcocks
to determine whether they might be the cause of the leak. Disassemble, clean, and regrease any
leaking stopcocks. Replace leaking rubber connections. After the analyzer is reassembled, repeat
the leak-check procedure.
12.1 Nomenclature.
12.2 Nitrogen, Carbon Monoxide Concentration. Determine the percentage of the gas that is N 2
and CO by subtracting the sum of the percent CO2 and percent O2 from 100 percent.
12.3 Dry Molecular Weight. Use Equation 3–1 to calculate the dry molecular weight of the stack
gas.
Note: The above Equation 3–1 does not consider the effect on calculated dry molecular weight of
argon in the effluent gas. The concentration of argon, with a molecular weight of 39.9, in
ambient air is about 0.9 percent. A negative error of approximately 0.4 percent is introduced. The
tester may choose to include argon in the analysis using procedures subject to approval of the
Administrator.
16.0 References
1. Altshuller, A.P. Storage of Gases and Vapors in Plastic Bags. International Journal of Air and
Water Pollution. 6 :75–81. 1963.
2. Conner, William D. and J.S. Nader. Air Sampling with Plastic Bags. Journal of the American
Industrial Hygiene Association. 25 :291–297. 1964.
8
3. Burrell Manual for Gas Analysts, Seventh edition. Burrell Corporation, 2223 Fifth Avenue,
Pittsburgh, PA. 15219. 1951.
4. Mitchell, W.J. and M.R. Midgett. Field Reliability of the Orsat Analyzer. Journal of Air
Pollution Control Association. 26 :491–495. May 1976.
5. Shigehara, R.T., R.M. Neulicht, and W.S. Smith. Validating Orsat Analysis Data from Fossil
Fuel-Fired Units. Stack Sampling News. 4 (2):21–26. August 1976.
Average
a
% Dev.=[(Q−Qavg)/Qavg]×100 (Must be ≤±10%)