INSTALLATION GUIDE

0604-021306 DSI Interface Unit QI-552V


DSI/AUXOUT Interface Unit QI-553V
First Edition: 30 Aug 2004 for TEC-5521, TEC-5531 Series Defibrillator
12345678901234567890123456789012123456789012345678901234567890121234567890123456789012345678901212345678901234567890123456789012123456789012345678901234567890121
12345678901234567890123456789012123456789012345678901234567890121234567890123456789012345678901212345678901234567890123456789012123456789012345678901234567890121
12345678901234567890123456789012123456789012345678901234567890121234567890123456789012345678901212345678901234567890123456789012123456789012345678901234567890121

The CE mark is a protected conformity mark Installation Procedure


of the European Community. The products
herewith comply with the requirements of 1. Make sure that the power of the defibrillator is turned off.
Medical Device Directive 93/42/EEC. 2. Disconnect the AC power cord from the AC outlet and
defibrillator.
General 3. Remove the battery pack if it is installed.
The QI-552V DSI Interface Unit and QI-553V DSI /AUXOUT
1) Turn the knob on the battery holder cover
Interface Unit are designed to measure CO2 and SpO2
waveforms for the TEC-5521, TEC-5531 Series Defibrillator counterclockwise and remove the battery pack holder
(The optional TG-901T3 or TG-921T3 CO2 Sensor Kit, JL- cover.
951T3 SpO2 Adaptor and sensor are necessary). The DSI/
AUXOUT interface unit outputs ECG waveforms. This
installation guide explains how to install the DSI interface unit
and DSI/AUXOUT interface unit into the defibrillator. For
installation, refer to the TEC-5500 Service manual. For safety
and use, read the TEC-5500 operator’s manual.

Model SpO2/CO2 AUX 2) Disconnect the battery cable by holding the battery
connectors connector cable connector and remove the battery from the
DSI Interface Unit 2* 0 defibrillator.
DSI/AUX OUT Interface Unit 1 1
4. Remove the DSI blank panel.
* The same parameter cannot be measured on two connectors.
1) Turn the defibrillator over.
2) Remove the two M3×2 binding head screws. These
WARNING screws are used when the interface unit is attached.
• Installation must be done by qualified service M3×2 binding head screw
personel.
• To avoid the possibility of injury to yourself or
damage to the defibrillator, do not install or remove
any component or change switch settings while the
power is on. When disassembling, make sure that the
defibrillator is off, the AC power cord is disconnected
from the defibrillator, and the battery pack is
removed from the defibrillator.

3) Remove the DSI blank panel by pulling it toward you.


Compositions
DSI blank panel
Description Q'ty
DSI interface unit
1
Or DSI/AUXOUT interface unit
Revision label 1

Installation Guide QI-552V/QI-553V 1


5. Attach the interface unit.
1) Connect the defibrillator and interface unit with the
cable.
Interface unit

3. Select “5. DSI Option” with the Item key.


4. Select “On” with the ← or → key.
5 Select “11. Flash Save” with the Item key.
6. Press the ← key. The setting changes from “Off” to “Exec”
2) Attach the interface unit to the defibrillator and secure and the function key changes as follows:
it with the two M3×2 binding head screws.

6. Install the battery pack. 7. Press the OK key.

Enabling the Interface Unit Operation Check


1. Open up the System Maintenance Screen For the defibrillator operation, refer to the TEC-5500 operator’s
1) Connect the AC power cord manual.
2) While pressing and holding the 1. Connect the AC power cord to the defibrillator and connect
Silence alarms key, turn the Energy/ the other end of the AC power cord to the AC outlet.
Mode Select control to the “SETUP” 2. Turn on the power of the defibrillator.
position. The System Setup - Menu 3. Check that no error message is displayed.
screen appears. 4. Connect the TG-901T3 or TG-921T CO2 Sensor Kit, or JL-
3) Release the Silence alarms key. 951T3 SpO2 Adaptor to the connector on the DSI Interface
4) While pressing the Multi-function Unit.
5. Check that “CO2” or “SpO2” is displayed on the screen
key, press the following keys on the (the waveforms and measurement data are displayed in
front panel one by one: dotted lines).
ECG lead key → Silence alarms key For the CO2 and SpO2 measurement, refer to the Operator’s
→ ECG sensitivity key. manual of the TG-901T3/TG-921T CO2 Sensor Kit and JL-
951T3 SpO2 Adaptor. For the ECG analog signal output,
refer to the TEC-5500 operator’s manual.
If “CO2” or “SpO2” is not displayed, check that the
The System Maintenance - Menu interface unit is correctly installed. Refer to “Installation
screen appears. Procedure”.
5) Release the Multi-function key. 6. Turn off the power of the defibrillator.

2. Select “Configuration” with the Item key and press the OK Attaching the Revision Label
key to open the System Maintenance - Configuration. After checking operation, attach the revision label to the rear
panel.

QI-55XV
00001 AA

Clean where the label IPX1


QI-55XV
00001 AA

will be attached with


cotton moistened with
alcohol.

2 Installation Guide QI-552V/QI-553V