Manual Versana Active
Manual Versana Active
Manual Versana Active
Technical Publication
Versana ActiveTM
Basic Service Manual
Direction Number: 5808595-100 English
Rev.8
This Manual covers the software version of R1.x.x for Versana Active ultrasound
system.
GE
DATE
REV (YYYY/MM/DD) REASON FOR CHANGE
Rev.4 2020/04/03 Update the Peripheral list and description for cart
grounding;
Update manual for R1.0.4 and higher;
Update the power cord list
REVISION REVISION
CHAPTER NUMBER NUMBER CHAPTER NUMBER NUMBER
Chapter 5 Rev.8
Translation policy
Proprietary to GE
Permission to use this Advanced Service Software and related
documentation (herein called the material) by persons other
than GE employees is provided only under an Advanced
Service Package License relating specifically to this Proprietary
Material. This is a different agreement from the one under which
operating and basic service software is licensed. A license to
use operating or basic service software does not extend to or
cover this software or related documentation.
If you are a GE employee or a customer who has entered into
such a license agreement with GE to use this proprietary
software, you are authorized to use this Material according to
the conditions stated in your license agreement.
However, you do not have the permission of GE to alter,
decompose or reverse-assemble the software, and unless you
are a GE employee, you may not copy the Material. The
Material is protected by Copyright and Trade Secret laws; the
violation of which can result in civil damages and criminal
prosecution.
If you are not party to such a license agreement or a GE
Employee, you must exit this Material now.
Trademarks
All products and their name brands are trademarks of their
respective holders.
Copyrights
All Material Copyright © 2019-2022 by
General Electric Company Inc. All Rights Reserved.
Introduction
1-1 Overview
1-3-1-2 Icons
ELECTRICAL
MECHANICAL
RADIATION
LASER
HEAT
PINCH
Table 1-4: Standard Icons that indicates that a special procedure is to be used
Avoid Static Electricity Tag and Lock Out Wear Eye Protection
NOTE: If the new label is needed during the service activities, please
click “Ask an Expert” to submit the case in the support central:
http://supportcentral.ge.com/products/sup_products.asp?prod_
id=44177. And please use Internet Explorer to open the link.
1-6-2 Introduction
The following safety precautions must be observed during all
phases of operation, service and repair of this equipment.
Failure to comply with these precautions or with specific
warnings elsewhere in this manual, violates safety standards of
design, manufacture and intended use of the equipment.
WARNING Have two people available to deliver and unpack the Versana
Active.
Attempts to move the Ultrasound system considerable
distances or on an incline by one person could result in injury
or damage or both.
WARNING Ensure that the Ultrasound system is turned off and unplugged.
Wait for at least 20 seconds for capacitors to discharge as
there are no test points to verify isolation. The amber light on
the OP panel ON/OFF button will turn off.
Ultrasound system components may be energized. Always
refer to the Ultrasound system's Proprietary Service Manual for
LOTO warnings and cautions.
WARNING Never use a probe that has fallen to the floor. Even if it looks
OK, it may be damaged.
CAUTION To avoid inquiry when you move the monitor and the monitor
arm, do not put your finger, hand, or object on the joint of the
monitor or the monitor arm.
CAUTION Use safety shoes when doing work where there is any chance
of foot injury.
CAUTION Use protective gloves when working with sharp edges or when
directed to wear PPE during a removal/replacement procedure.
1-6-5-2 Probes
1-6-5-3 Peripherals
NOTE: The battery shall be shipped in about 30% charged state. Those
packs have to be fully charged and discharged up to 3 times to
utilize Li-lon smart packs before use.
NOTE: The label on service part will provided the information of expiry
date. This date is only used for shelf life management. If it is
expired, the battery needs to do a complete discharge and
charge operation to extend shelf life. After the battery is installed
on ultrasound system for using, the expiry date is no longer
valid.
1-10-2 CE Compliance
Versana Active conforms to all applicable conducted and
radiated emission limits and to immunity from electrostatic
discharge, radiated and conducted RF fields, magnetic fields
and power line transient requirements.
For applicable standards, refer to the Safety Chapter of the
Ultrasound system User’s Manual.
NOTE: For CE Compliance, it is critical that all covers, screws,
shielding, gaskets, mesh, clamps, are in good condition,
installed tightly without skew or stress. Proper installation
following all comments noted in this service manual is required
in order to achieve full EMC performance.
GE Medical Systems (China) Co., Ltd. +86 510 85225888 +86 510 85226688
No.19 Changjiang Road
WuXi National Hi-Tech Dev.Zone
214028 Jiangsu
China
TELEPHONE / FAX
AUTHORIZED REPRESENTATIVE NUMBER
The location of the CE marking is shown in the Safety chapter of the Basic User
Manual.
Authorized EU Representative:
GE Medical Systems SCS
283 rue de la Minière 78530 BUC, France
TELEPHONE / FAX
Factory Site NUMBER
Site Preparations
2-1 Overview
CAUTION If the Ultrasound system is very cold or hot, do not turn on its
power until it has had a chance to acclimate to its operating
environment.
Degree F 23.9 28.4 32.9 37.4 104 108.5 113 117.5 122 131 140
hours 3 2 1 0 0 1 2 3 4 6 8
CAUTION Ensure that the probe face temperature does not exceed the
normal operation temperature range.
2-2-2-3 Cooling
2-2-2-4 Lighting
IEC 61000-4-11 < 5%T (> 95% dip) for < 5%T (> 95% dip) for Mains power
0.5 cycle; 0.5 cycle; quality should be
Voltage dips, short
interruptions and 40%T (60% dip) for 5 40%T (60% dip) for 5 that of a typical
voltage variations cycles; cycles; commercial or
on mains supply hospital
70%T (30 dip) for 25 70%T (30 dip) for 25
cycles; cycles;
environment.
< 5%T (>95% dip) for 5 < 5%T (>95% dip) for 5
sec sec
Be aware of Radio • Keep the Ultrasound system at least 5 meters (15 feet) away from other EMI
Frequency sources sources.
• Special shielding may be required to eliminate interference problems caused by
high frequency, high powered radio or video broadcast signals.
Ground the Poor grounding is the most likely reason an Ultrasound system will have noisy
Ultrasound system images. Check grounding of the power cord and power outlet.
Replace all screws, • After you finish repairing or updating the Ultrasound system, replace all covers
Radio Frequency and tighten all screws.
gaskets, covers, • Any cable with an external connection requires a magnet wrap at each end.
cores • Install all covers. Loose or missing covers or Radio Frequency gaskets allow
radio frequencies to interfere with the ultrasound signals.
Replace broken If more than 20% or a pair of the fingers on an Radio Frequency gasket are
Radio Frequency broken, replace the gasket. Do not turn on the Ultrasound system until any loose
gaskets metallic part is removed.
Do not place labels Where applicable, never place a label where Radio Frequency gaskets meet the
where Radio Ultrasound system. Otherwise, the gap created will permit Radio Frequency
Frequency gaskets leakage. Or, if a label has been found in such a position, move the label.
touch metal
Use GE specified The interconnect cables are grounded and require ferrite beads and other
harnesses and shielding. Also, cable length, material, and routing are all important; do not change
peripherals from what is specified.
Take care with Cellular phones may transmit a 5 V/m signal; that could cause image artifacts.
cellular phones
Properly route Where applicable, do not allow cables to lie across the top of the Card Rack or
peripheral cables hang out of the peripheral bays. Loop the excess length for peripheral cables
inside the peripheral bays. Attach the monitor cables to the frame.
CAUTION Ensure that the probe face temperature does not exceed the
normal operation temperature range.
CAUTION Check the room temperature before you use the probe.
CAUTION Ensure that the probe face temperature does not exceed the
normal operation temperature range.
11
10
9
8 12
13
7
6 14
1
1. Door – at least 762 mm (30 6. Examination Table – 1930 x 12. Network Interface
inches) 610 mm (76 x 24 inches) 13. 457 mm (18 inches) distance
2. Film Viewer 7. Footswitch of Ultrasound system from
3. Counter Top, Sink with hot 8. Stool wall or objects
and cold water and Supplies 9. Ultrasound system 14. GE Cabinet for Software and
Storage 10. External Peripherals Manuals
4. Linen Supply 11. Dedicated Power Outlet -
5. Probes/Supplies Circuit Breaker protected and
easily accessible
10 11 15
12
13 14 16
2 3 4 5
7
1
17
18
Figure 2-3. Suggested Room with EchoPAC PC workstation and Ultrasound Scanner
None.
Supported networks:
100/1000 Mbit Ethernet/DICOM network (option)
Commercial devices such as laser cameras, printers, VCRs and external monitors,
usually exceed allowable leakage current limits and, when plugged into separate AC
outlets, are in violation of patient safety standards. Suitable electrical isolation of such
external AC outlets, or providing the device with extra protective earth, will be required
in order to meet UL60601-1 and IEC60601-1 / IEC60601-1-1 standards for electrical
leakage.
2.12.20DV (UL60601-1:2003)
In area in which patients are normally cared for, the patient vicinity is the space with
surfaces likely to be contacted by the patient or attendant who can touch the patient.
This encloses a space within the room 1.83 m (6 ft.) beyond the perimeter of the bed
(examination table, dental chair, treatment booth, and the like) in its intended location,
and extending vertically 2.29 m (7.5 ft.) above the floor.
1. Patient environment
Sub Clause 3.79 and figure A.9 (IEC60601-1:2005 and ANSI AAMI
ES60601-1:2005)
Such an area is an environment in which medical diagnosis, monitoring or treatment is
carried out. It is very difficult to attach unique dimensions to the PATIENT
ENVIROMENT.
In practice a distance of 2,5 m (8.2 ft.) above the floor on which the medical personnel
stand and a horizontal distance of 1,5 m (4.9 ft.) have justified themselves as
indicative of the dimensions of the Patient Environment.
The patient environment/vicinity will be depicted as a dashed line in this procedure.
See example below.
1. Patient environment
System Setup
3-1 Overview
CAUTION DO NOT wear the ESD wrist strap when you work on live
circuits and more than 30 V peak is present.
CAUTION DO NOT operate this unit unless all board covers and frame
panels are securely in place. System performance and cooling
require this.
CAUTION If the Ultrasound system is very cold or hot, do not turn on its power
until it has had a chance to acclimate to its operating environment.
Degree F 23.9 28.4 32.9 37.4 104 108.5 113 117.5 122 131 140
hours 3 2 1 0 0 1 2 3 4 6 8
3-3-4-1 Overview
CAUTION Please carefully unpack the system, and do not dispose the
package of Versana Active, so that it can be reused for service.
NOTE: Please check the Versana Active console is well assembly after
unpacking the system.
1. Cut the adhesive tape and open top covers of paper carton.
CAUTION Do not lift the unit by the rubber band. Equipment damage
may result.
CAUTION Do not lift the unit by the rubber band. Use handle to move
system.
58 362 390 mm
2.28 14.25 l5.35 Inches
Verify that the mains voltage specified for the Versana Active is
available on-site.
Refer to the latest revision of the Basic User Manual for a full list
of product labels prior to serving the system.
Power
Model Name Voltage Tolerances consumption Frequency
Versana Active VA
100-240V ±10% Max.200VA 50/60 Hz
Versana Active VS
NOTE: USB Flash Drive approved for Versana Active are verified for
EMC performance according to EN55011 class B. The use of
any other USB Flash Drive will compromise this verification, and
may cause interference on Versana Active itself, or on other
electronic devices.
For approved models, please refer to Chapter 9.
Insert the USB Flash Drive in one of the USB ports on the
Versana Active.
3-7-3 LDAP
Versana Active Ultrasound System support integration with
LDAP compliant directory servers. If the system is configured to
use LDAP (Utility/LDAP), the access control and user
authentication is performed with use of services from the
connected directory server. The authentication will be done by
the directory server, while the access control to the system is
done with data received from the directory server.
3-8-1 Overview
This section describes how to install and configure the
peripherals validated for the Versana Active.
About the operation check-out of peripherals, See
4-4-18 ‘Peripheral checks’ on page 4-72 for more information.
2 Paper Roll 1
3 USB cable 1
2 Paper Roll 1
3 USB cable 1
4 USB cable 1
• USB Stick
Table 3-9: Materials furnished with USB Stick
2 USB Cable 1
• Wireless Adaptor
1 1 Wireless Adaptor 1
• Bluetooth Adaptor
1 1 Bluetooth Adaptor 1
• Footswitch
1 1 Pedal Footswitch 1
2 3 Pedal Footswitch 1
• DVD RW
1 DVD RW 1
2 USB cable 1
3-8-3-1-1 Tools
No special tools needed.
3-8-3-1-2 Manpower
One person 5 minutes.
3-8-3-1-3 Preparations
1. Unpack B/W Printer.
2. Ensure no physical damage.
5. Press Utility-->Connectivity-->Service.
6. Select Standard Print under Service box by default or
Select Standard Print from drop-down menu and click Add.
NOTE: The print key configured in this step also can be used as
Print key in Worksheet.
7. Select the printer from the Printer pull-down Properties
menu. Type the printer name in the Name Field. This name
is used on the Button screen. Press Save.
3-8-3-2-1 Tools
No special tools needed.
3-8-3-2-2 Manpower
One person 5 minutes.
3-8-3-2-3 Preparations
1. Unpack the Printer.
2. Ensure no physical damage.
3-8-3-3-1 Tools
No special tools needed.
3-8-3-3-2 Manpower
One person 5 minutes.
3-8-3-3-3 Preparations
1. Unpack the Sony UP-D25MD Printer.
2. Ensure no physical damage.
3-8-3-4-1 Tools
No special tools needed.
3-8-3-4-2 Manpower
One person 5 minutes.
3-8-3-4-3 Preparations
1. Prepare the network connection.
3-8-3-5-1 Tools
No special tools needed.
3-8-3-5-2 Manpower
One person 2 minutes.
3-8-3-5-3 Preparations
1. Unpack the Footswitch.
2. Ensure no physical damage.
3-8-3-6-1 Tools
No special tools needed.
3-8-3-6-2 Manpower
One person 1 minute.
3-8-3-6-3 Preparations
1. Unpack the USB HDD.
2. Ensure no physical damage.
3-8-3-7-1 Tools
No special tools needed.
3-8-3-7-2 Manpower
One person 1 minute.
3-8-3-7-3 Preparations
1. Unpack the ECG Assy.
2. Ensure no physical damage.
3-8-3-8-1 Tools
No special tools needed.
3-8-3-8-2 Manpower
One person 1 minute.
3-8-3-8-3 Preparations
1. Unpack the DVDRW.
2. Ensure no physical damage.
3-8-3-9-1 Tools
No special tools needed.
3-8-3-9-2 Manpower
One person 1 minute.
3-8-3-9-3 Preparations
1. Unpack the wireless adapter
2. Ensure no physical damage.
3-8-3-10-1 Tools
No special tools needed.
3-8-3-10-2 Manpower
One person 1 minute.
3-8-3-10-3 Preparations
1. Unpack the bluetooth adaptor
2. Ensure no physical damage.
NOTE: Versana Active system does not support to pair with Apple
device.
5. The Pincode is shown in the dialogue, click Accept to
continue.
3-8-3-11-1 Tools
No special tools needed.
3-8-3-11-2 Manpower
One person 5 min.
3-8-3-11-3 Preparations
1. Unpack the AC isolation USB.
2. Ensure no physical damage.
Figure 3-46. Insert the USB cable into the Ultrasound system
2. For R1.0.3, skip this step. For R1.0.4 and higher, set ADM
password in Admin Registration. And then set up 3
password security questions and answers.
ADM is the administrator of the ultrasound system and has
the highest authority for the software. ADM can create other
accounts, import and export data and make some sensitive
settings.
NOTE: The passwords are critical for your system. Please store the
passwords safely.
NOTE: Record your answers for the 3 questions. The answers are
required when you need to reset ADM password.
NOTE: ADM password and security questions can be empty if
password policy level is selected as LOWEST.
Element DESCRIPTION
CRM No Customer Relationship Management (CRM) number. System identifier assigned to the
customer unit by the service region. CRM is pre-populated by adding Versana Active to the
CRM number. The CRM number of the Versana Active is editable.
Serial No Serial number of the agent (read-only). If the agent is not registered with a serial number,
this field is populated with the serial number of the Versana Active. The serial number of
the agent is tied to the serial number of the Versana Active.
Proxy Server When Enable Proxy is selected, name of the proxy server IP.
Proxy Port When Enable Proxy is selected, number of the proxy server port.
11. After having set the InSite - Remote Assistant, press Next to
continue.
1 Enter the Versana Active computer name. This may be the same as the station name.
2 Enter the Versana Active IP address, subnet mask, default gateway and network speed. For
automatic assignment of IP address, subnet mask and default gateway, select DHCP.
Note: If possible, set the Versana Active network Speed to match that of the Network
switch. if in doubt, set it to AutoDetect. Otherwise, transfer times can be two or five times
longer, during which the Versana Active will appear to be locked up. (If the Hard Drive
activity light on the front of the console is lit steady or blinks quickly, the Versana Active is
most likely not hung.)
1 Select MyComputer.
2 Assign an AE title to the Versana Active. (AE stands for Application Entity. DICOM
services use this to identify the Versana Active.) AE title is case-sensitive. This title
may contain the Computer Name from the TCP/IP page, if desired.
Note: It is NOT recommended to use the factory default. This is not prohibited, but
more than one system with the same AE title can cause confusion.
3 Edit Port Number if needed. 104 is typical. Save your changes and reboot the system.
Save your changes and then press Ping. A “Smiley Face” indicates successful
communication between your Versana Active and the device. A “Frown” indicates failed
communication. Check the following:
• Is the device running?
• Is it connected to the network?
• Did you enter the right IP address?
1 In Utility -> Connectivity -> Service, from the Destination Device drop-down menu, select the
device on which the service resides. (This assumes you have already setup the device in the
Device tab in ‘How to Setup and Use a DICOM Image Storage Service’ on page 3-81 .
2 From the Select Service Type to add drop-down menu, select DICOM Image Storage and press
Add.
4 Enter the AE Title and port Number of the service. AE Title is case-sensitive.
5 Save your changes and then press Verify. A “Smiley Face” indicates successful communication
with the service.
Note: If you get a successful Ping (Smiley Face) at the device level but not at the
service level, it is possible that the AE Title or Port Number of the service settings are
incorrect. Ensure that these are correct then re-verify. Be sure that the service type
(Store, Print, etc) is correct and supported by the device.
3-9-7 Dataflow
3 Select so that this dataflow does not appear as a Dataflow on the Patient menu.
4 Select to use this dataflow as the default dataflow when you start the system.
3-9-8 Button
You can assign print buttons via the Utility --> Connectivity -->
Button page.
Assigning print buttons. First select the print button to configure
on the upper, left corner of the page. Then select the device you
want to add in the middle part of the page, under Available Input/
Outputs. Then click on the right arrow in the top right corner of
the page.
NOTE: You can configure each print key to multiple output devices/
dataflows.
NOTE: Only attach one DICOM service per print key (e.g., PACS and
DICOM printer). Multiple DICOM devices should be configured
via a dataflow.
NOTE: When using a print key to send an image directly to a DICOM
device, this causes a single DICOM association per image. Most
devices (all known printers) work fine with this. However, some
storage devices, such as ALI, Kodak Access, and Cemax,
assume that the end of each association is the end of the exam
and can result in a new folder for each image. In the Utility
menu, select a single association or open PR for the desired
DICOM storage device.
3-9-8 Button(continued)
Table 3-23: Physical Print Buttons
M&A only (no images) Configures the system to send a DICOM structured report only; no image
isgenerated or sent.
Quick Format To format the media quickly, check this box. If you uncheck this box, the
media is formatted with a full format. New media should always be formatted
with a full format.
3-9-10 Miscellaneous
The Miscellaneous tab allows you to configure tools related to
patient management and print and store options. You can
specify default system functionality, such as whether patient ID
is required when you archive data, or if you want the system to
automatically search the archive for a patient when you enter
patient data.
Use birthdate In the Patient information window, enter either the patient age or the
birthdate:When selected, enter birth date, then the age is
calculated.Whencleared, enter age (birth date field not available).
Auto search for patient In the Search/Create Patient window: When selected, the system
automatically searches through the selected patient archive, while the user
enters patient information.When cleared, the automatic search tool is turned
off. If you are trying to keep the past patient data confidential, DO NOT use
this feature.
Automatic generation of In the Search/Create Patient window: When selected, the Patient ID is
patient ID notrequired when entering a new patient in the archive. The
systemautomatically generates an ID number. When cleared, the Patient ID
isrequired when entering a new patient in the archive.
After [End Current Patient], Select go to Worklist screen or Patient screen when ending the current
go to patient.
Show BBT Show BBT field on the OB patient screen to input the basal body
temperature.
Double click on patient list Select Review or New Exam to display each time you double click on the
to start patient name in the patient list on the Patient menu.
Detail Mode Select to display Detail Mode, rather than Exam View, when you select the
patient name in the patient list on the Patient menu. You can also type
comments while in Detail Mode.
Export to USB HDD: Create DICOMDIR is a DICOM file format which contains how the directory
Create DICOMDIR and DICOM files structured for diagnostic portable media behave. It is
important for portability between the Versana Active to PACS. If you want to
Export to Network storage: save exams to the USB Hard drive and look at it on the PACS, the
Create DICOMDIR DICOMDIR is a must.
Remember Cursor Position To set a default cursor location on the Data Transfer screen:
on the Transfer Screen 1. Select the “Remember cursor position in the Transfer screen” preset and
press Save.
2. On the Data Transfer screen, move the cursor to the desired field.
3. Exit out of the Data Transfer screen. When returning to the Data Transfer
screen, the cursor location is in the position your selected.
Quick New Patient Entry Select to store new patient automatically by pressing the Patient key.
Request acknowledge of When selected, the user is asked to confirm action when ending an
End Exam action examination.
Warn Image Store without Select to receive a warning when you press the Print key without an active
Patient patient.
Warn Register to No Select to receive a warning when you register a patient to the “No Archive”
Archive data flow. Select a different data flow for permanent storage of patient data.
Warn image store to Read The system posts a warning message if you attempt to store images to a
Only dataflow read-only Dataflow.
Warn video titles exist in The system posts a warning if the video titles exist on the internal DVR flash
the internal storage memory.
P[1-2] Key Sound Select None, Click, Chimes, Ding, Ding-Dong, or Whoosh.
Store Dual as Dicom Only Select to always store dual images as a DICOM (secondary capture) store,
rather than Raw DICOM.
Dual When Color Support Dataflow Mixed is not available. While transferring dual images to the PACS,
is Mixed send black and white images as gray; send color images as color. Set up 2
services (one gray and one color), set up 2 dataflows, and set up 2 buttons.
Each button needs to be tied to a different service.
Select if you want to keep the user preset for Color Photometric Interpretation
while in Dual mode.
Store Multiframe for Sec Select if you want the CINE loop stored as secondary capture.
Capture Loops
Image Order Scheme Select to Direct Store images in Acquisition Order, Scan Coach/Assistant
Order, or Off.
• Off. The clipboard on the Ultrasound system shows the image in the order it
was acquired. Therefore, re-stored images appear where you’d expect.
However, on the PACS system, images appear in arrival order or in image
number order.
• Acquisition Order. From the Ultrasound system perspective, the same as
“Off.” But on the PACS system (if based on image number order), images
are displayed consistently with the way they are stored on the Ultrasound
system.
• Scan Coach/Assistant Order. You can define the storage order (reading
order) via Scan Coach/Assistant Creator. Therefore, based on the order
defined in Scan Coach/Assistant, images are re-ordered and displayed in
this manner both on the Clipboard and on the PACS system.
Validation Format If the Enable Other ID preset is selected, the system validates the format of
“Other ID” when an ID is entered. Choose: NHS Number *** ** *****, Letters
and Numbers, Numbers, or Any (no restriction)
3-9-11 Bluetooth
To add a new bluetooth device,
1. Press Add New Device.
2. Type the device name in the Name field.
Add New Device/Remove Press Add new device to add a new device; press Remove to delete a
device.
Properties: MAC Address Unique network card address. NOTE: Only available for MyComputer.
3-9-12 Tricefy
New GE Versana ultrasound systems have Tricefy pre-installed.
Additional software is not necessary for using Tricefy; simply
activate your account using the following steps.
1. Press Utility -> Connectivity -> Tricefy.
NOTE: The Tricefy tab is only available on machines that have
Tricefy option.
3-9-12 Tricefy(continued)
3. Add Tricefy to Print workflow.
• Click Button to configure the P- Bottons for sending
and sharing images.
3-9-12 Tricefy(continued)
4. If you want to share the images to patient, please add Trice
Patient to Printflow View. And also remember input the
patient’s Phone number or email address.
NOTE: If you have any problem with Tricefy settings, please contact
your GE service/sales representative.
NOTE: During and after setup, the documentation (i.e. CDs with
documentation, User Manuals, Installation Manuals, etc.) for the
Versana Active and the peripherals must be kept as part of the
original Ultrasound system documentation. This ensures that all
relevant safety and user information is available during the
operation and service of the complete Ultrasound system.
CAUTION Care must be used when moving the Cart or replacing its parts.
Failure to follow the precautions listed below could result in
injury, uncontrolled motion and costly damage.
ALWAYS:
• be sure the pathway is clear
• use slow, careful motions
• Limit movement to a slow careful walk.
Need two people to work together when moving on inclines or
lifting more than 16 kg (35 lbs).
WARNING When the cart is raised for a repair or moved along any incline,
use external caution since it may become unstable and tip
over.
Do
CAUTION The Cart is not water proof. Do not expose the Cart to water or
any kind of liquid.
Never set liquids on the Cart to ensure that liquid does not drip
into the unit.
CAUTION Two people should unpack the unit because of its weight. Two
people are required whenever a part weighing 19kg (42 lb.) or
more must be lifted.
When a new Cart arrives, check that any components are not
damaged and are not in short supply. If shipping damage or
shortage occurs, contact the address shown in Chapter 1.
CAUTION Do not tilt the unit more than 5 degrees to avoid tipping it over.
CAUTION Be sure that the system is fixed correctly on the Cart to avoid
falling down.
WARNING Serious injuries may result. To prevent the Cart from becoming
unstable, please use the peripherals with the recommended
specifications. DO NOT install peripheral devices that are too
large or too heavy for the location on the cart.
3-12-6-2-1 Tools
• Common Hex driver
• common Phillips screwdriver
3-12-6-2-3 Preparations
• Turn off all the power supply.
3-12-6-3-1 Tools
• NA
3-12-6-3-3 Preparations
• Turn off all the power supply.
3-12-7 Paperwork
NOTE: During and after setup, the documentation (i.e. User Manuals...)
for the peripheral units must be kept as part of the original
system documentation. This will ensure that all relevant safety
and user information is available during the operation and
service of the complete system.
4-1 Overview
4-2-1 Overview
Some procedures are used more often than other. The intention
with this section is to keep the most used procedures in one
place.
4-2-2-1 Warnings
CAUTION Use only power supply cords, cables and plugs provided by or
designated by GE.
Press once on the On/Off key on the Operator Panel to boot the
unit.
During a normal boot, you may observe that:
a. The unit’s ventilation fan starts on full speed, but slows
down after a few seconds (listen to the fan sound).
b. Power is distributed to the peripherals, Operator Panel
(Console), Monitor, Front End Processor and Back End
Processor.
c. Back End Processor and rest of scanner starts with the
sequence listed in the next steps:
d. Back End Processor is turned ON and starts to load the
software.
e. The Start Screen is displayed on the monitor.
f. A start-up bar indicating the time used for software
loading, is displayed on the monitor.
g. The software initiates and sets up the Front End
electronics and the rest of the instrument.
h. The backlight in the keyboard is lit.
i. As soon as the software has been loaded, either a 2D
screen is displayed on the screen, indicating that a
probe has been connected, or a No Mode screen is
displayed, indicating that no probe has been connected.
NOTE: Total time used for start-up is typical one and a half minutes
or less. If starting after a power loss or a lock-up, the
start-up time may be up to four minutes.
NOTE: Diagnostic LEDs on the side connector panel indicate the
boot up status, refer to “Normal Status” in Table 7-12 on
page 7-56.
NOTE: Set up Drive D password and System password. See
3-9-1 ‘EZ configuration Wizard’ on page 3-58 for more
information.
• Exit
(Only available when logged in as GE Service with Service
Dongle)
Select this button when you want to exit to Windows.
• Standby
Use this button to select Standby mode.
CAUTION DO NOT unplug and/or transport the unit until after the power
off sequence has been completed. Failure to do so may result
in corrupted patient files.
Move the cursor to OK and press Set key on the control panel to
select OK. The system enters.
Check the system date and time. If it is incorrect, follow below
steps to reset the system date and time.
• Enter Utility -> System -> General -> Date/Time(restart
needed).
• Reset the system date and time.
• Select Apply and then select OK.
• Select Save.
For R1.0.4 and higher, if user types the wrong user name or
has not set up password security questions yet, a warning
message will appear on the screen. Press Cancel to exit to
Operator Login window and the following steps will be
skipped.
7. Press EZBackup/EZMove.
9. It indicates the size of the data and the storage. Select Next
to continue.
NOTE: The calculation for the number of backup CD is only an
estimate. Allow for one additional CD when performing an
EZBack/EZMove.
4-2-14 Backup
For more information, refer to the latest revision of the Versana
Active Basic User Manual.
4-2-17 Where are the User Manuals and the Service Manual?
Online versions of the User Manuals are available via the help
function.
Both the User Manuals and the Service Manual are delivered as
PDF files on a CD-ROM. Paper copies may be ordered from
GE.
4-2-18 How to display or print the PDF files from the Manual CD?
1. Insert the CD-R disk (CD-ROM) into the CD-drive on a PC
or Laptop with Adobe Acrobat Reader.
CAUTION Do not try to use the Versana Active to read these files, it will
not work!
4-2-19-1 Manpower
4-2-19-2 Tools
4-2-19-3 Preparations
To get access to the trackball for cleaning, you must perform the
following steps:
1. Power down the system.
2. Disconnect the mains power cable from the wall outlet.
Follow these links if you need more information: See
4-2-3 ‘Power off’ on page 4-8 for more information.
3. Separate the trackball and the gasket. Wipe off any oil or
dust from the trackball, gasket and the trackball housing
using a cleaner or cotton swab.
4. Assemble the trackball and gasket, then put it into the
housing and rotate it clockwise until its notches are set in
the position.
CAUTION When cleaning, make sure not to spill or spray any liquid into
the trackball housing (keyboard or system).
Power up the system and test that the trackball now works as
intended.
CAUTION DO NOT operate the unit without the air filter in place.
Allow the air filter to dry thoroughly before re-installing it on the
unit.
4-2-20-1 Cleaning
CAUTION If the gel leaked in the gel warmer always cannot be cleaned
on time, it may cause a short circuit in the internal board.
4-2-22 Monitor
To clean the monitor face:
Use a soft, folded cloth. Gently wipe the monitor face.
DO NOT use a glass cleaner that has a hydrocarbon base (such
as Benzene, Methyl Alcohol or Methyl Ethyl Kentone) on
monitors with the filter (anti-glare shield). Hard rubbing will also
damage the filer.
NOTE: When cleaning the screen, make sure not to scratch the
monitor.
CAUTION Before cleaning the control panel, make sure the key cap is
firmly in place.
4-3-1 Overview
Disk Encryption is a function to protect the patient information on
the device and prevent unauthorized access to PI/PHI,
especially when the device is stolen. The encryption AES is 256
bit.
NOTE: Disk Encryption can also protect the user data stored on
removable device.
NOTE: The steps for Removable Media Encryption and Local Patient
Data Drive Encryption are the same.
CAUTION The user must make backups and take care of the encryption
password/passphrase and recovery key. It is the customer’s
responsibility for storing the data. GE will have no back door or
any responsibility or possibility of recovering the data.
• Press OK.
3. Press OK.
4-4-1 Overview
In this section, the functional checks for Versana Active are
described. Functional checks are used to verify that the product
works as intended. Functional checks may also be used during
troubleshooting.
4-4-3 Preparation
Turn on power to Versana Active. For detailed description, See
4-2-2 ‘Power ON/Boot Up’ on page 4-5 for more information.
2. Press Utility -> Preset Manager. Select IQC from the left
column, and select to move it to the Imaging
Preset Selections. Then Select Exit.
4-4-7-1 Introduction
4-4-7-2 Preparations
4-4-8-1 Preparations
4-4-9-1 Introduction
4-4-9-2 Preparations
• Adjust the Active mode gain to set the gain in the color flow
area.
Adjust Scale to the highest setting that provides adequate
flow detection.
NOTE: The scale value may affect FPS, Low Velocity Reject, and
Sample Volume.
Adjust Wall Filter to remove low velocity blood flow and
tissue movement that reduces image quality.
Adjust Sample Volume to a low setting for better flow
resolution, or a higher setting to more easily locate disturbed
flows.
Adjust Frequency to optimize the color flow display. Higher
settings improve resolution. Lower settings improve depth
penetration and sensitivity. This does not affect the
frequency used for 2D and M-Mode.
NOTE: Frequency setting may affect FPS, SV and Low Velocity
Reject.
Adjust PowerOutput to obtain an acceptable image using
the lowest setting possible.
NOTE: The Power setting affects all other operating modes.
Adjust the following settings to further optimize display of the
image:
• Use Invert to reverse the color assignments in the color flow
area of the display.
Use Threshold to emphasize either the color flow overlay,
or the underlying grey scale tissue detail.
Use Baseline to emphasize flow either toward or away from
the probe.
Use Spatial Filter to reduce noise in the color flow area.
Spatial Filter smooths the image by averaging collected data
along the same horizontal line. An increase of the spatial
filter will reduce noise, but this will also reduce the lateral
resolution.
1. Select M Mode.
2. Use the trackball to position the ROI frame over the area to
be examined.
3. Press Set. The instruction Size should be highlighted in the
trackball status bar. Use the trackball to adjust the
dimension of the ROI.
• Adjust the Active mode gain to set the gain in the color flow
area.
Adjust Scale to the highest setting that provides adequate
flow detection.
NOTE: The scale value may affect FPS, Low Velocity Reject, and
Sample Volume.
Adjust Wall Filter to remove low velocity blood flow and
tissue movement that reduces image quality.
Adjust Sample volume (SV) to a low setting for better flow
resolution, or a higher setting to more easily locate disturbed
flows
Adjust Frequency to optimize the color flow display. Higher
settings improve resolution. Lower settings improve depth
penetration and sensitivity. This does not affect the
frequency used for 2D and M-Mode.
NOTE: NOTE: Frequency setting may affect FPS, SV and Low
Velocity Reject.
Adjust PowerOutput to obtain an acceptable image using
the lowest setting possible.
NOTE: The Power setting affects all other operating modes.
Adjust the following settings to further optimize display of the
image:
• Use Invert to reverse the color assignments in the color flow
area of the display.
Use Threshold to emphasize either the color flow overlay,
or the underlying grey scale tissue detail.
Use Baseline to emphasize flow either toward or away from
the probe.
4-4-10-1 Introduction
4-4-10-2 Preparations
Adjust the Active mode gain to set the gain in the spectral
Doppler area.
• Adjust Wall Filter to reduce unwanted low velocity blood
flow and tissue movement.
In PW mode, adjust Volume to low setting for better
resolution, or higher setting to more easily locate the
disturbed flows.
Adjust the Compression setting to balance the effect of
stronger and weaker echoes and obtain the desired intensity
display.
Adjust Frequency to optimize flow display. Higher setting
will improve resolution and the lower setting will increase the
depth penetration.
Adjust Frame rate to a higher setting to improve motion
detection, or to a lower setting to improve resolution.
NOTE: Frequency and Frame rate settings may affect the Low
Velocity Reject.
Adjust Power Output to obtain an acceptable image using
the lowest setting possible. This is particularly important in
CW mode, as the energy duty cycle is 100% (constant).
NOTE: The Doppler Power setting affects only Doppler operating
modes.
Adjust the following settings to further optimize the display
of the image.
Use the Sweep Speed to optimize the sweep speed.
To view signal detail, adjust Scale to enlarge the vertical
spectral Doppler trace.
Use Invert to reverse the vertical component of the spectral
Doppler area of the display.
Use Angle Correct to steer the ultrasound beam to the
blood flow to be measured.
4-4-11-1 Introduction
4-4-11-2 Preparations
CAUTION DO NOT allow the probe head to hang free. Impact to the
probe head could result in irreparable damage.
1 Select the appropriate connected probe from the The probe activates in the currently-selected
probe indicators. operating mode. The probe’s default settings
for the mode and selected exam are used
automatically.
4 Test the probe in each active connector slot. It will display pictorial data each time.
WARNING Take the following precautions with the probe cables: Do not
bend, be sure to keep probe cables free from the wheels.
4-4-14-1 Introduction
• ECG Module
4-4-14-3 Preparations
None
1 Connect the ECG harness to the connector The unit displays a straight curve along the
under the control panel. bottom edge of the image sector on the screen.
2 Connect the three leads to an ECG simulator, When connecting, the signal on the screen will
or be noisy.
Fasten the three ECG Pads to your body and When the connection is completed, a typical
connect the three leads to respective ECG clean ECG signal is displayed.
Pad.
4-4-15-1 Introduction
4-4-15-2 Preparation
• Press Freeze.
The left and right markers are displayed on either side of the
last detected heart cycle on the ECG trace.
• Press Freeze.
The selected heart beat is played back.
• Press Freeze to freeze the cineloop.
Use the trackball to scroll through the acquisition and find
the sequence of interest.
• Adjust Cycle select to move from heart beat to heart beat
and select the heart cycle of interest.
Adjust Num cycles to increase or decrease the number of
heart beats to be played back.
Adjust Left marker and Right marker to trim or expand the
cineloop boundaries.
Wheel Characteristics
5 Verify all of the customer’s options are set up Demo Option strings turn on
correctly.
4-8 My Trainer
4-8-1 Overview
My Trainer provides a quick guide to operate the system.
To access My Trainer,
1. Press Alt + H to enter My Trainer.
Or,
1. Press Utility -> System -> User Configurable Key. Check
Enable Numeric Hot Key, then set My Trainer in the
Keyboard Key or User Defined Key.
4-8-1 Overview(continued)
There are four sections in My Trainer. The four sections are
displayed on the left side of the My Trainer interface.
4-8-1 Overview(continued)
4-8-1 Overview(continued)
Table 4-7: Page turning
9. Step Description
5-1 Overview
5-2-3-2 Introduction
5-3-1 Oveview
The AC Power assyís main tasks are to isolate and output to the
DC/DC unit which is inside the system console. The input of AC
power pack will be the AC outlet and itís universal, the range is
AC 100V-240V, 50-60Hz. And no main power switch located on
this power pack.
5-3-2 AC Power
Versana Active Power supply is an ACDC adapter which use
100V-240V,(50-60Hz) as input and DC 19V (7.9A) as output.
5-4-1 Introduction
The Service Desktop is an interface that provides access to
system information, status and diagnostics.
The Service Desktop has different content or views depending
on the access level. The access level is determined by the user
profile as well as the service options enabled on the Versana
Active.
• Basic view is the standard view, restricted only by the user
through the user profile settings. Administrator default user
has access to the Service Desktop. Any user with "local
Service access" in their user profile can have access to this
view.
• Class C view is the view enabled by the service options
purchased.
• Service Advanced
• Service Expert (requires Service Advanced)
• Service Pro (requires Service Advanced)
Service Adjustments
6-1 Overview
Europe EU Preset
Diagnostics/Troubleshooting
7-1 Overview
7-2-1 Overview
There may be a time when it would be advantageous to capture
trouble images and system data (logs) for acquisition to be sent
back to the manufacturer for analysis. There are different
options to acquire this data that would give different results.
7-5-4 Licenses
With Service Basic Access (Class A), these are the available
options:
• HOME
• Utilities
• Change Password
• Data Transfer
• Delete Files
• Gather Logs
• Network Capture
• SSA License
• Thirty Party Licenses
• Options
• Agent Configuration
With Service Advanced (Class C), these are the available
options:
NOTE: With a Class C license, options display according to these
purchased level of access.
• HOME
• Diags
• Run Diags
• Diag History
• Utilities
• Change Password
• Checkpoints
• Delete Files
• Disk Defragment
• Data Transfer
• Gather Logs
• Network Capture
• SSA License
7-5-4 Licenses(continued)
• System Shutdown
• Thirty Party Licenses
• Options
• Agent Configuration
With Service Advanced plus Service Expert (Class C), the Clean
Userdefs, Reset Patient Database,and Software Reload utilities
are added to the Service Advanced options listed.
With Service Advanced and Service Expert plus Service PRO
(Class C), the probe assessment tool (ePAT) diagnostic is
added to the Service Advanced and Service Expert options
listed.
With GE Service access (Class M) and an SSA key, these are
the available options:
• HOME
• Diags
• Run Diags
• Diag History
• Utilities
• Change Password (not available through a remote
connection)
• Checkpoints
• Clean Userdefs
• Data Transfer
• Delete Files
• Disk Defragment (not available through a remote
connection)
• Disruptive Mode Utility
• Gather Logs
• Network Capture
• Reset Patient Database
• Software Reload
• SSA License
• SSH
• Thirty Party Licenses
• Virtual Console Observation
• Options
• Agent Configuration
7-5-5 Home
Home configurations vary depending upon the purchased
service level.
Element DESCRIPTION
Agent CRM Verified CRM verified status of the agent. Valid values are:
• Yes - The agent is verified in the back office.
• No - The agent is not verified in the back office.
• Not Available - The agent is not running or has not
been configured.
Model Number GE part number for the Versana Active. The same
number as listed on the rating plate.
Element DESCRIPTION
System Date Current date in the format <day>, <month> <date> <year>.
System Time Local time based on the last time the system desktop was
refreshed in the format <hh:mm:ss>.
Application Built Date Date the application software was built. The application software
includes the Versana Active product-specific software.
Base Image Built Date Date the base image software was built. The base image software
includes the Windows operating system and other supporting
software.
Element DESCRIPTION
System On Time Displays the time the Versana Active was turned on.
Time Since Last Reboot Displays the time the Versana Active was last
rebooted.
Element DESCRIPTION
7-5-5-5 Temperature
This table shows all the elements available on FRU Status with
descriptions.
Element DESCRIPTION
7-5-5-7 Graphs
7-5-5-8 Voltage
This table shows all the elements available on FRU Status with
descriptions.
Element DESCRIPTION
Status Status of the voltage for the FRU. Valid values are:
• Normal - Indicates that the fan is within the allowable range.
• Warning - Indicates that the voltage is close to the limit of the
allowable range.
• Error - Indicates that the voltage is outside the allowable range.
7-5-5-10 AC Voltage
AC Voltage graphically displays the voltage input over a period
of time to help identify instabilities,or fault conditions, over time.
To access this page, under System Status, select Voltage and
then select AC Voltage.
Use Disk Drive Space to view the hard drive partitions including
total size and available free space in GB. Includes the overall
health of the drive in one of these colors:
• Red - Available free space is less than 10% of the total size.
• Orange - Available free space is more than 10% and less
than 20% of the total size.
• Green - Available free space is more than 20% of the total
size.
The information on Disk Drive Space is available to Class C
and Class M licenses.
To access Disk Drive Space, navigate to Utility > Service >
Home.
Element DESCRIPTION
Element DESCRIPTION
Host Name Name of the local host for the Versana Active.
Wired Connection
MAC Address Address for the MAC for the wired network.
Wireless Connection
MAC Address Address for the MAC for the wireless connection.
Element DESCRIPTION
Active Probe Temperature (Celsius) When available, temperature of the active probe.
Not all probes report temperature. The most common probe
to report temperature is the TEE probe.
4. After the system restarts, the network icon on the right top of
the screen shows that the network is available.
5. The network icon at the left bottom of screen shows that the
wireless network is available.
7-6-3-4 PC Test
7-6-3-4 PC Test(continued)
7-6-3-4 PC Test(continued)
• Trackball
• The Mouse Status Test verifies the cursor position and
mouse button state. When a mouse button is pressed,
the corresponding button on the picture will change
color. If the mouse is a wheel or scroll mouse, an arrow
will indcate the direction the wheel is being rotated.
Clicking the wheel will flash the picture of the mouse in
the Mouse Status Test area.
• The Drag and Drop Test verifies a mouse can
successfully perform drag and drop operations. Left
click the picture of the CD and drag is onto the picture of
the drive. If successful, the picture will change.
• The Double Click Test verifies a mouse can successfully
perform double-click operations. Double-click on the
picture of the monitor. If successful, the picture will
change.
• Sound Test generates sounds for testing the speakers.
• USB Ports Test lists USB Devices.
51-75% ON ON ON OFF
76%-100% ON ON ON ON
7-6-3-6-1e-PAT
e-PAT is a probe assessment tool that evaluates each probe
element response scanning on the air. Probes have to be clean,
any gel residue will provide incorrect data. This test is an
assessment that is intended to be used comparatively during the
life of the probe to evaluate possible probe deterioration over
time.
NOTE: Probe diagnostics is available to Class C (Service Pro) license.
1. Check that the probe to be tested is thoroughly clean and
dry. Connect it to the probe port on the scanner, then hold
the probe in the air ready for testing.
CAUTION DO NOT allow the probe head to hang free. Impact to the
probe head could result in irreparable damage.
7-6-3-6-1e-PAT(continued)
2. Insert the SSA key and press Insite ExC > Service
Desktop > Diags > Run Diags > Probe Assessment.
3. Select Start.
The Probe Diagnostics commence. While the test proceeds,
the name of the test currently in progress is displayed in the
Selected Diagnostics field. As the testing sequence
progresses, the progress bar in the Total Progress field will
advance to reflect the test progress.
4. At any stage, press Stop or Reset button, if required.
5. When the probe diagnostics is completed, there will be
Pass/Fail information listed in the Diognostic Results
Summary field to indicate whether the probe diagnostic is
completed successfully or not.
6. If you want to review all Reports to ensure the probe is fully
functional and has no damaged elements, go to In the Insite
ExC > Service Desktop > Diags > Diags History to
check the probe diagnostic results.
7-6-3-6-1e-PAT(continued)
7. All the probe diagnostics you have run will be displayed in
Runs field. Select one and press Get History button, the
corresponding reports will be shown in the Diagnostic
Results History field.
8. Press Details button if you want to see the detailed Reports
of the selected diagnostic.
NOTE: The above detailed information is just for reference and not
used to decide probe defect or not.
1. 4C-RS H4000SR
2. 8C-RS H40402LS
3. E8C-RS H40402LN
4. E8Cs-RS H48062AF
5. 9L-RS H40442LL
6. 3Sc-RS H45041DL
7. 6S-RS H45021RP
8. 12S-RS H44901AB
9. 12L-RS H40402LY
7-7 Troubleshooting
KBD On/
Off
button Checklist (only
LED2 LED1 HDD LED LED for reference) Note
After Blink Blink Slowly Not Green 1. Check KBD 1. KBD board LED
pressing Slowly board failure
power 2. Check KBD cable 2. KBD cable failure
On/Off 3. Check CWI PWA 3. CWI board boot
button, up failure
the
system Blink Blink 1. Check CWI PWA 1. Remove eDP
monitor Quickly Quickly 2. Check the cable from CWI
still monitor and PWA. Then power
keeps related eDP on system and
black. cable observe the LED1
(After at and LED2 status
least 2 again.
minutes)
Blink Blink Slowly Green 1. Check MST PWA 1. Remove MST
Slowly 2. Check CWI PWA board from CWI
board. Then
power on system
and observe the
LED1 and LED2
status again.
Blink OFF Blink when Green 1. Check CWI PWA 1. CPU PWM_EN
Slowly reading and signal failure.
writing data. 2. Connect a normal
external monitor to
HDMI port.
Replacement Procedures
8-1 Overview
8-2-1 Warnings
8-2-1 Warnings(continued)
NOTE: Use an ESD compatible work space or the ESD-kit during parts
replacement.
8-3 Disassembly/Re-assembly
CAUTION Do not wear the ESD wrist strap when you work on live circuits
and more than 30V peak is present.
Screwdriver
No Part Name Part No. QTY Screw Description Description
Montior Assy
Keyboard Assy
Bottom Assy
8-4-6-1 Overview
NOTE: For the upgrade option Load the complete disk or Load the
bootable C: partition only, the information for Computer name
and IP setting will be lost. Before upgrade, backup these
information.
Computer Name
IP-Address
Subnet Mask
Default Gateway
Network Speed
9. Select Select All, and then select Transfer to backup all the
patient information and images.
WARNING If selecting Load the complete disk, all existing software and
data will be erased. If backup has not been performed, all data
like Patient Data, System Configuration and User
Configurations (Customer Presets) will be lost. Please ensure
that any patient data on the disk has been backed up before
upgrading the system.
Reload Apps No No No No
Secure wipe is intended to erase all the patient data with the
software on the system before the system will be shipped for
service.
NOTE: This tool is not BAM approved.
NOTE: Before starting this procedure, remove all probes and
peripherals and remove them from the system.
NOTE: While it is believed to be unnecessary, it would not hurt to
disconnect the system from the network and remove all
transducer.
NOTE: Please ensure AC adapter is connected during system upgrade!
1. Insert the USB memory stick labeled “System & Application
Software” to the system.
2. Properly turn off the scanner by momentarily pressing the
Power On/Off Switch. In System-Exit window, select
Shutdown to shutdown the system.
NOTE: If the system will not shutdown normally, hold down the
Power On/Off Switch until the light turns off.
3. Power on the system. The system will detect the USB
memory stick automatically.
NOTE: All the patient data cannot be recovered after wipe process.
NOTE: After the system software loading completion, please check the
option strings to ensure that the options are activated and
working.
1. Reboot the system.
2. Press Utility -> Admin, the following dialog is displayed.
Select Acknowledge to continue.
4. Enter Utility -> Admin-> System Admin, ensure that all the
installed option keys are displayed and the status of Options
are valid.
• The status “Permanent” means the option keys are
activated and working.
NOTE: After the software upgrade process, the system language and
region preset will be set as the default. Change the settings if
needed.
NOTE: Refer to step 6 on 8-15 and step 7 on 8-16 for the system
Region Preset and Language settings.
1. When the system is powered on, go to Utility -> System ->
General. Select Regional Options.
NOTE: For R1.0.4 and higher, keep Input Language the same as
system language, otherwise corrupted characters will incur.
12. When the system boots up, the system appears in the
selected language.
NOTE: To have the settings take effect, you MUST restart the
system.
NOTE: Press Alt+Shift to change the input language.
12. Press Patient key on the control panel, and then select
Data Transfer. Select the task as Import and select the
source as where the patient information is exported to.
13. Select Select All, and then select Transfer to restore all the
patient information and images.
Renewal Parts
9-1 Overview
ABBREVIATION DESCRIPTION
3D THREE DIMENSIONAL
Assy ASSEMBLY
9-3-2 Cart
Probe 9-6
Peripheral 9-7
Manuals 9-11
9-3-4 Probe
Table 9-3: Probes for Versana Active
9-3-5 Peripheral
Table 9-4: Peripherals for Versana Active
Commercial
Item Part Number Part Number Description
Footswitch
USB Stick
Printer
ECG
917 5129487 H48492AY ECG module from Norav Isral, not applies to Muslim
country and China
920 5146739 H48502AB ECG detachable cable IEC type EURO and ASIA
Others
Commercial
Item Part Number Part Number Description
Part Commercial
Item Number Part Number Description
1005 6736102-2 H48512AD PWR SPLY CRD ANZ 10A 250V STRAIGHT 2.5M
1007 5177187-3 H48482AE AC Power Cord Australia and New Zealand Class-CRU
NOTE: Cable clip for Table 9-5 power cord should be installed in the
hole 1, refer to Figure 9-3.
Part Commercial
Item Number Part Number Description
Part Commercial
Item Number Part Number Description
1022 6736115-2 H48532AY PWR SPLY CRD DENMARK HOSPITAL GRADE 10A
250V STRAIGHT 2.5M
NOTE: Cable clip for Table 9-6 power cord should be installed in the
hole 1, refer to Figure 9-4.
9-3-7 Manuals
Table 9-7: Manuals for Versana Active
Commercial
Item Part Number Part Number Description
1103 5808593-100 H48412BW Versana Active Basic User Manual English version
1110 5807952-127 H48422BG Versana Active User Guide Brazilian Portuguese version
1114 5807952-141 H48412BZ Versana Active User Guide Simplified Chinese version
Commercial
Item Part Number Part Number Description
1133 5807952-177 H48432BE Versana Active User Guide European Portuguese version
1144 5808652-127 H48422BG Versana Active Release Notes Brazilian Portuguese version
1148 5808652-141 H48412BZ Versana Active Release Notes Simplified Chinese version
Commercial
Item Part Number Part Number Description
10-1 Overview
10-2 Warnings
CAUTION Do not pull out or insert circuit boards while power is on.
CAUTION Do not operate this Ultrasound system unless all board covers
and frame panels are securely in place. System performance
and cooling require this.
Per
Facilities
Service at Indicated QA
Time Daily Weekly Monthly Program Notes
Clean Console •
Clean Monitor •
Per
Facilities
Service at Indicated QA
Time Daily Weekly Monthly Program Notes
Clean Filters •
NOTE: For a list of required tools for servicing the Versana Active, refer
to chapter 8.
9-44579 Wrench Open End 1/2 - 9/16 9-XL9974MM Xcelite-hex Blade 2.5mm
9-44579 Wrench Open End 1/2 - 9/16 9-XL9975MM Xcelite-hex Blade 3mm
9-41581 Screwdriver, Blade 3/16 in. × 4 9-45341 Pliers Vice Grip 10 inch
in.
9-39451 20' Steel Tape, locking Spring 9-3001 Xacto Pen Knife
load
9-GH807 Ratchet, Offset, Slotted 9-HT62002 Solder Aid, Fork and Hook
9-41584 Screwdriver, slotted 1/4 in. × 6 9-GH300ME Steel Rule Metric 6 inch
in.
9-46677 Hex Keys, 20 pc., Metric 9-XL9924 Xcelite-hex Blade 1/8 inch
9-34701 1/4 in. Standard Socket set (19 9-XL9925 Xcelite-hex Blade 5/32 inch
pc)
9-43499 1/2 inch Socket 1/4 inch drive 9-XL9926 Xcelite-hex Blade 3/16 inch
9-65283 Case 8.5 in. × 4.5 in. × 2 in. 9-45072 Pliers 6 inch Diagonal
Deep
9-GH408 Tweezers
9-45381 Pliers, Arc Joint 9 1/2 inch 9-44067 Socket 1 1/16 in. for 1/2 in.
drive
9-45092 Pliers, Linesman 8 1/2 inch 9-42679 Socket 10MM Hex for 1/2 in.
drive (2273333)
9-42882 Punch, Pin 3/32 inch 9-44262 Extension 10 inch for 1/2 in.
drive (2273405)
9-42884 Punch, Pin 5/32 inch 9-4258 3/8 inch to 1/2 inch Adapter
9-42886 Punch, Pin 1/4 inch 9-34374 3/8 inch Metric Socket Set -
12 PT
9-31277 File, Half Round, Bastard 8 9-52068 Tap and Drill Set
inch
Anti Static Kit Kit includes anti–static mat, wrist strap and cables for 200 to
240V system
3M #2204 Large adjustable wrist strap
3M #2214 Small adjustable wrist strap
3M #3051 conductive ground cord
Safety Analyzer The safety Analyzer tool should be calibrated and compliant
with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
B/W Printer Cleaning Sheet See printer user manual for requirements
Color Printer Cleaning Sheet See printer user manual for requirements
Disposable Gloves
1. Ask and Listen Ask the customer if they have any problems or questions about the
equipment.
2. Paperwork Fill in the top of Ultrasound Inspection Certificate (see Figure 10-5 on
page 10-34). Record all probes and Ultrasound system options.
3. Power up • Turn the Ultrasound system power on and verify that all fans and
peripherals turn on.
• Watch the displays during power up to verify that no warning or error
messages are displayed.
• Where applicable, confirm that the battery is charged. If no AC Input
present, use the internal battery.
4. Probes Verify that the Ultrasound system properly recognizes all probes.
6. FFA (server) Where applicable, for Warranty and Contract Customers only:
• Verify that InSite is functioning properly.
• Ensure two-way remote communications.
7. Review Error Logs Where applicable, Error Logs can be reviewed via system diagnostics.
8. Diagnostics Optional.
10. Image Archive Back up the Image Archive onto appropriate media.
1. B-Mode Verify basic B-Mode (2D) operation. Check the basic Ultrasound system
controls that affect this mode of operation.
2. CF-Mode Verify basic CF-Mode (Color Flow Mode) operation. Check the basic
Ultrasound system controls that affect this mode of operation.
3. Doppler Modes Verify basic Doppler operation (PW and CW if available). Check the basic
Ultrasound system controls that affect this mode of operation.
4. M-Mode Verify basic M-Mode operation. Check the basic Ultrasound system
controls that affect this mode of operation.
5. Probe Elements Perform an Element Test on each probe to verify that all the probe
elements and system channels are functional.
6. Applicable Verify the basic operation of all optional modes such as Contrast. Check
Software Options the basic Ultrasound system controls that affect each options operation.
7. Xmit/Recv Use the Visual Channel Utility on the loop connect to verify that all system
Elements xmit/recv channels are functional.
10. Monitor Verify basic monitor display functions. Refer to Chapter 3 of the Basic
User Manual.
11. Software Menu Verify Software Menu display functions. Refer to Chapter 3 of the Basic
check User Manual.
13. Measurements In measurement mode, make distance measurement, get result in result
window. Verify the distance by graduate rule. Distance Accuracy should
be within ±5%. (Name result from result window Result A, result from
graduate rule Result B; Distance Accuracy = (Result B-Result A)/Result A)
2. B/W Printer Verify hardcopy output of the B/W video page printer. Clean heads and
covers if necessary.
3. Color Printer Verify hardcopy output of the Color video page printer. Clean heads and
covers if necessary.
7. DVD Verify that the DVD is functioning properly. Clean heads and covers if
necessary.
1. Unplug Cord Disconnect the mains cable from the wall and Ultrasound system.
3. Verify Verify that the LINE, NEUTRAL and GROUND wires are properly attached
to the terminals, and that no strands may cause a short circuit.
10-6-2-4 Cleaning
1. Console Remove the battery. Use a fluid detergent in warm water on a soft, damp
cloth to carefully wipe the entire system. Be careful not to get the cloth too
wet so that moisture does not enter the console.
2. Probe Holder Clean probe holders. (they may need to be soaked to remove excess gel).
3. Monitor Use a soft, non-abrasive folder cloth. Gently wipe the monitor face. DO
NOT use a glass cleaner that has a hydrocarbon base (such as Benzene,
Methy Alcohol or Methy Ethyl Ketone) on monitor with the filter (anti-glare
shield).
Wet Wipe X X X X
Disinfectants [1]
Protex X X X X
Tristel Wipes X X X X
Clorox wipes X X X X
CaviWipes X X X X
PDI Sani-Cloth HB X X X X
[1]: Effective Disinfection is always a balance between safe inactivation of infectious agents and undesirable side
effects. Due to the generally uneven and irregular surface of Ultrasound consoles, a comprehensive surface
disinfection process cannot be recommended by the manufacturer.
1. Labeling Verify that all Ultrasound system labeling is present and in readable
condition.
2. Scratches & Dents Inspect the exterior for dents, scratches or cracks.
3. Input Power Refer to: 10-6-2-3 ‘Mains cable inspection’ on page 10-14.
4. Cables & Check all internal cable harnesses and connectors for wear and secure
Connectors connector seating. Pay special attention to footswitch assembly and probe
strain or bend reliefs.
5. Shielding & Check to ensure that all EMI shielding, internal covers, air flow panels and
Covers screws are in place. Missing covers and hardware could cause EMI/RFI
problems while scanning.
6. Control Panel Inspect keyboard and control panel. Note any damaged or missing items.
7. Control Panel Check for proper operation of all operator panel and Freeze Key light.
Lighting
8. Monitor Inspect the monitor Display for scratches and bad pixels.
Verify proper operation of Contrast and Brightness controls.
Where applicable, confirm that the monitor arm allows:
• swivelling the screen to the left and to the right
• folding the screen to the locked position
• release and adjustment backwards and forwards
• can be adjusted in the up/down positions.
Note: Monitor Arm movement may vary and is not applicable to all
Ultrasound systems.
10. Power and Check for proper operation of all Power and System Status Indicators.
System Status
Indicators
11. Battery Where applicable, check that the battery is not damaged, does not leak,
does not emit an odor, and is not deformed or discolored. Observe all
warnings and cautions for battery handling, recharging, storing, and/or
disposal,
1. Probe Holder Clean probe holders. (they may need to be soaked to remove excess gel).
2. Probes Thoroughly check the Ultrasound system probe connectors and remove
dust from inside the connector sockets if necessary. Visually check for
bent, damaged or missing pins.
3. Probes Verify that the Ultrasound system properly recognizes all probes.
CAUTION Do not submerge the probe above the probe as the strain relief
is not sealed. Do not allow fluid into or onto the connector of
the probe.
CAUTION Take extra care when handling the lens face of the Ultrasound
transducer. The lens face is especially sensitive and can easily
be damaged by rough handling. NEVER use excessive force
when cleaning the lens face.
CAUTION Disinfect a defective probe before you return it. Be sure to tag
the probe as being disinfected.
CAUTION Review the probe care card that is packed with each probe.
The following website contains the most current and up-to-date
recommendations:
http://www.gehealthcare.com/usen/ultrasound/products/
probe_care.html
Normal Reverse
Country Condition Open Ground Polarity Open Neutral
Table 10-14: Type BF Applied Part Leakage Current Limits - Probes Surface
NOTE: *Mains Applied refers to the sink leakage test where mains
(supply) voltage is applied to the part to determine the amount of
current that will pass (or sink) to ground if a patient contacted
mains voltage.
The following tests are performed at the factory and should be
performed at the site. These tests are: chassis leakage current,
and probe leakage current. All measurements are made with an
electrical safety analyzer which should be calibrated and
compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
NOTE: No outlet tester can detect the condition where the Neutral
(grounded supply) conductor and the Grounding (protective
earth) conductor are reversed. If later tests indicate high leakage
currents, this should be suspected as a possible cause and the
outlet wiring should be visually inspected.
Measure the resistance from the third pin of the attachment plug
to the exposed metal parts of the case. The ground wire
resistance should be less than 0.2 ohms. Reference the
procedure in the IEC60601-1.
NOTE: User needs to unscrew the GROUND PIN before releasing the
Versana Active.
1. GROUND PIN
2. OHMMETER
3. Versana Active
4. ACCESSIBLE METAL PART:
• MONITOR HOUSING
• PEAR PANEL CONNECTOR
• ANY CASTER/WHEEL SUPPORT
The test verifies the isolation of the power line from the chassis.
The testing meter is connected from accessible metal parts of
the case to ground. Measurements should be made with the unit
ON and OFF, with the power line polarity Normal and Reversed.
Record the highest reading of current.
Figure 10-3. Set Up for Chassis Source Leakage Current, IEC 601-1 Clause 19 -
Continuos Leakage Currents and Patient, Auxiliary Currents
The test passes when all readings measure less than the value
shown in Table 10-13 on page 10-23. Record all data on the PM
Inspection Certificate.
Table 10-17: Typical Data Sheet for enclosure Source Leakage Current
Tester
Neutral
Tester or Test 1 Test 2 Real
Unit Polarity Ground Speaker Panel Metal Optional Optional
Power Switch Switch Cover Parts Test 3 Test 4
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
10-7-6-1 Definition
This test measures the current that would flow to ground from
any of the probes through a patient who is being scanned and
becomes grounded by touching some other grounded surface.
NOTE: Some leakage current is expected on each probe, depending on
its design. Small variations in probe leakage currents are normal
from probe to probe. Other variations will result from differences
in line voltage and test lead placement. It is abnormal if no
leakage current is measured. If no leakage current is detected,
check the configuration of the test equipment.
10-7-6-2 Tools
CONSOLE
POWER N (WHITE)
OUTLET
G (GREEN)
MOMENTARY
SWITCH LEAKAGE TEST
METER
The test passes when all readings measure less than the values
shown in Table 10-14 on page 10-23. Record all data on the PM
Inspection Certificate.
Table 10-18: Typical Data Sheet For Transducer Source Leakage Current
Transducer Tested:
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
* Scan Format: Phased Array, Linear Array, Curved Array, Mechanical Array or Other
A pplic a ble H a rdwa re O ptions L a be ling (see User Manual for Labeling)
Control Panel
LCD
Measurement Accuracy
GE Approved Peripherals
COMMENTS:
_________________________ ____________________________ _________________________
_________________________ ____________________________ _________________________
_________________________ ____________________________ _________________________
_________________________ ____________________________ _________________________
ELECTRICAL SAFETY
Max Value Value
Electrical Test Performed Allowed Measured OK? Comments
PROBES
Probe 1:
Probe 2:
Probe 3:
Final Check. All system covers are in place. System scans with all probes as expected.
System ground
continuity
Chassis source
leakage current -
probe
Chassis source
leakage current -
wheel
Chassis source
leakage current -
monitor
Peripheral 1 leakage
current
Peripheral 1 ground
continuity
Peripheral 2 leakage
current
Peripheral 2 ground
continuity
Peripheral 3 leakage
current
Peripheral 3 ground
continuity
S V
safety verification
human, 1-14 system requirements, 3-15
safety precaution messages, 1-8 verify customer order, 3-13
safety test overview, 10-21 verifying
service safety considerations, i-16 removable media, 3-86
setup voltage settings, 3-16
completion, 3-14
preparations, 3-13 W
reminders, 3-3
warnings, 10-3