B40i Technical Reference Manual Rev C
B40i Technical Reference Manual Rev C
B40i Technical Reference Manual Rev C
0537
Conformity according to the Council Directive 93/42/EEC concerning Medical Devices amended by 2007/47/EC.
Revision C
10 October, 2013
Manufacturer address:
GE Healthcare Finland Oy Kuortaneenkatu 2
FI-00510 Helsinki
Finland
www.gehealthcare.com
Copyright © 2013 General Electric Company. All rights reserved.
Classifications
In accordance with IEC 60601-1
− Class I and internally powered equipment - the type of protection against electric shock.
− Type BF or CF equipment. The degree of protection against electric shock is indicated by a
symbol on each parameter module.
− Equipment is not suitable for use in the presence of a flammable anesthetic mixture with
air or with oxygen or nitrous oxide.
− Continuous operation according to the mode of operation.
− Portable Monitor
In accordance with IEC 60529
− IP21 - degree of protection against harmful ingress of water.
In accordance with EU Medical Device Directive
− IIb.
In accordance with CISPR 11:
− Group 1 Class A;
• Group 1 contains all ISM (Industrial, scientific and medical) equipment in which there
is intentionally generated and/or used conductively coupled radio-frequency energy
which is necessary for the internal functioning of the equipment itself.
• Class A equipment is equipment suitable for use in all establishments other than
domestic and those directly connected to a low-voltage power supply network
which supplies buildings used for domestic purposes.
Trademarks
Listed below are GE Medical Systems Information Technologies and GE Healthcare Finland Oy
trademarks used in this document. All other product and company names contained herein
are the property of their respective owners.
Datex, Ohmeda, DINAMAP, Trim Knob, Unity Network, CARESCAPE, EK-Pro, TruSignal, Entropy,
GE Healthcare, GE Medical system, General Electric Company.
1 Introduction
About this manual 1
1 Overview 3
1.1 Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.1 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.2 Safety message signal words . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.3 Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.2.4 ESD precautionary procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.2.5 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3 Service information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3.1 Service requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3.2 Equipment identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2 System description 11
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2 Bus structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.3 Distributed processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4 Module communication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4.1 Serial communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.5 Parameter modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.6 Software loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3 Frame functional description 14
3.1 Main components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1.1 Keyboards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1.3 CPU board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.1.4 Power board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1.5 AC/DC unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.1.6 Batteries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2 Interfacing computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.3 Connectors and signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.3.1 External connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4 Hemo-dynamic module introduction 23
4.1 Monitor software compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2 Main components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.1 Hemo-dynamic module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.2 Signals and isolation barrier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.2.3 Power supply section. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.2.4 NIBP board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.2.5 ECG board in 5-lead measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.2.6 STP board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.3 Connectors and signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.3.1 Front panel connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.4 Measurement principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.4.1 NIBP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.4.2 ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.4.3 Pulse oximetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.4.4 Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.4.5 Invasive blood pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
i
Document no. 2069392-002
B40i Patient Monitor
4.4.6 Respiration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
2 Hardware installation
1 Installation 1
1.1 Unpacking instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2 Choosing location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Mounting the monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.4 Connection to power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.5 Check configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.6 Connection to Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.6.1 Pre-installation requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.6.2 To connect the network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.6.3 Network configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.7 Inserting and removing the E modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.8 Monitor connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.9 Visual indicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.10 Installation checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3 Maintenance
1 Instructions 1
1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Recommended tools. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2 Electrical Safety Tests 3
2.1 Test setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Power Outlet Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.3 Power cord and plug. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.4 Ground (earth) integrity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.4.1 Ground Continuity Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.4.2 Impedance of Protective Earth Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.5 Earth leakage current test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.6 Enclosure (Touch) leakage current test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.7 Patient leakage current tests - overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.7.1 Patient (source) leakage current test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.7.2 Patient (sink) leakage current tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.8 Test completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3 Installation checkout 13
3.1 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2.1 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.3 Time and date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.4 Parameters measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.5 Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.6 MC or S/5 Network connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.7 Conclusion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4 Maintenance and checkout 16
ii
Document no. 2069392-002
4.1 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2 Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.1 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.3 Keyboard(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.4 Time and date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.5 Hemo Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.6 Loudspeaker. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.2.7 Monitor software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.2.8 Watchdog circuitry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.2.9 Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.2.10 Final cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.3 Monitor battery maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.3.1 Use recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.3.2 Storage recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.3.3 Test the battery charge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.3.4 Charge the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.3.5 Condition the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.3.6 Replace the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5 Adjustments and calibrations 24
5.1 NIBP calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.2 Temperature calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.3 Invasive pressure calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4 Troubleshooting
1 Introduction 1
1.1 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 General troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Software troubleshooting chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Frame troubleshooting 4
2.1 Battery indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2 NET section troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3 Hemo Troubleshooting 9
3.1 NIBP troubleshooting flowchart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2 ECG troubleshooting flowchart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3 STP troubleshooting flowchart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.4 NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4.1 NIBP toubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4.2 NIBP error code explanation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.5 ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.6 Impedance respiration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.7 Pulse oximetry (SpO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.8 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.9 Invasive blood pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5 Service Menu
iii
Document no. 2069392-002
B40i Patient Monitor
1 Introduction 1
1.1 Service Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Service Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2 SW Management 3
2.1 SW Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Active Inactive SW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.3 NIBP SW Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.4 Country Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.5 License. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.6 Enter/Exit Demo Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 Frame 6
3.1 Country Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1.1 Languages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.2 Network. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2.1 Network Config (MC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.2.2 TCP/IP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.2.3 HL7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2.4 Dri Config (S/5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2.5 Dri Comm (in S/5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.3 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.3.1 Power page. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.3.2 WPM Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4 Keyboard 20
4.1 Keyboard Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5 Parameters 22
5.1 Gas Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.1.1 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.1.2 Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.2 ECG Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.2.1 ECG Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.3 STP Module (for GE SpO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.3.1 Calibrations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.4 NIBP Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.4.1 NIBP Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.4.2 NIBP Safety Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.4.3 NIBP Pneumatics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.5 SpO2 (for Masimo/Nellcor SpO2). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.6 Entropy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
6 Set/Test 36
7 Service Log 37
iv
Document no. 2069392-002
1.5 Hemo box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.6 FRU parts list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.7 Other parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2 Disassembly 6
2.1 ESD precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2 Reassembly precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Required tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.4 Before disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.5 To separate the frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.6 To disassemble the frame. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.7 To disassemble the extended rack and the recorder. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.8 Handling and storage of display component . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.9 To replace the fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.10 To download the software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
7 Technical specification
1 General Specifications 1
1.1 Genenral specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Defibrillator synchronization connector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Parameters specifications 4
2.1 ECG specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 Impedance respiration specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 GE TruSignal SpO2 specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.4 Nellcor SpO2 specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.5 Masimo SpO2 specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.6 NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.7 Invasive blood pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.8 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.9 Airway gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.10 Entropy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
8 E-miniC Module
1 Product overview 1
1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Measurement principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2.1 CO2 measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.3 Main components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3.1 Gas sampling system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.2 MiniC sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3.3 CPU board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.4 Connectors and signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2 Maintenance and checkout 7
2.1 Replacement of planned maintenance parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.1 Required parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.2 Replacement procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2 Visual inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
v
Document no. 2069392-002
B40i Patient Monitor
vi
Document no. 2069392-002
2.1.2 Planned Maintenance Kits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.1.3 Replacement procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.2 Visual inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.3 Functional check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.3.1 Test setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.3.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.3.3 Test completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3 Calibration and adjustments 21
3.1 Sample Flow Rate Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1.1 Calibration setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1.2 Sample Flow Rate Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2 Gas Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.2.1 Calibration setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.2.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4 Troubleshooting 24
4.1 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.2 Troubleshooting checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.2.1 Gas sampling system troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.2.2 MiniOM Measuring unit troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.2.3 MiniTPX Measuring unit troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.2.4 CPU board troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.3 Service Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.4 Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.4.1 Gas measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.5 Troubleshooting charts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
4.5.1 Gas measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5 Disassembly and reassembly 31
5.1 Disassembly guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.1.1 Serviceable parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.1.2 Service limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.1.3 ESD precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.1.4 Protection from dust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.1.5 Before disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.1.6 Required tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.2 Disassembly and reassembly procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.2.1 Disassembly workflow. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.2.2 Detaching the Front Cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.2.3 Detaching the Module Casing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.2.4 Replacement of Planned Maintenance Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.2.5 Replacement of CO2 Absorber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.2.6 Detaching the Latch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.2.7 Detaching the Front Chassis Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.2.8 Detaching the Main Flow Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.2.9 Detaching the Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.2.10 Detaching the OM holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6 Service parts 39
6.1 Ordering parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.1.1 Planned Maintenance Kits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.2 Spare parts for E-sCAiO, E-sCO, N-CAiO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.2.1 Front covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
vii
Document no. 2069392-002
B40i Patient Monitor
10 E-Entropy Module
1 Product overview 1
1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Measurement principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2.1 EntrEEG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.2 FEMG. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.3 RE and SE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.4 Impedance measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Main components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.1 Controls and connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.2 Entropy board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 Maintenance and checkout 5
2.1 Visual inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Electrical safety tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Functional check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3.1 Test setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3.2 Entropy tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3.3 Test completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3 Troubleshooting 8
3.1 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2 Troubleshooting chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4 Disassembly and reassembly 10
4.1 Disassembly guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.1.1 ESD precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.1.2 Before disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.1.3 Required tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.2 Disassembling and reassembling procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.2.1 To replace the front cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.2.2 To replace the Entropy board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.2.3 Reassembling the module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5 Service parts 12
5.1 Ordering parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.2 Spare parts for E-ENTROPY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
viii
Document no. 2069392-002
1 Introduction
About this manual
Intended audience
This Technical reference manual is meant for service representatives and technical personnel
who install, configure, maintain, administer, troubleshoot or repair B40i monitor running the
software license VSP-B.
1-1
− The electrical installation of the relevant room complies with the requirements of the
appropriate regulations.
− The equipment is used in accordance with the “User's Guide.”
− The equipment is installed, maintained and serviced in accordance with this manual.
Product availability
Some of the product parts and accessories mentioned in this manual may not be available in
all countries.
Please, consult your local representative for the availability.
Related documentation
− Clinical aspects, basic methods of measurement and technical background: Patient
Monitor B40i User’s Reference Manual
− Options and selections of the software: Patient Monitor B40i Default Configuration
Worksheet
− Compatible supplies and accessories: Patient Monitor B40i Supplies and Accessories
− Other devices closely related to the monitor:
• iCentral and iCentral Client User's Reference Manual
• CIC Pro Clinical Information Center Operator's Manual
Conventions used
To help you find and interpret information easily, the manual uses consistent text formats:
Within this manual, special styles and formats are used to distinguish between terms viewed
on screen, a button you must press, or a list of menu commands you must select:
− Names of hardware keys on the keypad are written in bold typeface: NIBP
Start/Cancel.
− Menu items are written in bold italic typeface: Monitor Setup.
− Emphasized text is in italic typeface.
− When referring to different sections in this manual, section names are enclosed in double
quotes: “Cleaning and care”.
− The word “select” means choosing and confirming.
− Messages (alarm messages, informative messages) displayed on the screen are written
inside single quotes: 'Learning.'
− Note statements provide application tips or other useful information.
1-2
1 Overview
The B40i is a modular multiparameter patient monitor. The monitor is especially designed for
monitoring in intensive care units. It can also be used during transportation within the hospital.
The modular design makes the system flexible and easy to upgrade.
NOTE: Your system may not include all these components. Consult your local representative for
the available components.
2
1, 4
1-3
1.1 Symbols
- On the rear panel:
- Electric shock hazard. Do not open the cover or the back.
Refer servicing to qualified personnel.
- For continued protection against fire hazard, replace the fuse
only with one of the same type and rating.
- Disconnect from the power supply before servicing.
- Do not use the monitor without manufacturer approved
mounting attached.
- On the hemo module: Protection against cardiac defibrillator
discharge is due in part to the accessories for pulse oximetry
(SpO2), temperature (T) and invasive pressure (P) measurement
Follow operating instructions
Alternating current
1-4
Audio pause.
ON/OFF.
Fuse. Replace the fuse only with one of the same type and rating
Gas inlet.
Gas outlet.
Storage temperature
Humidity limitations.
1-5
A blinking heart next to the heart rate or pulse rate value indicates
the beats detected.
1-6
Alarm off indicator - Displays in the digit field and in the Alarms
Setup menu when physiological alarms for this parameter are
turned off.
Audio alarms off indicator - Displays in the upper left corner of the
alarm area when physiological audible alarms are turned off.
1-7
Warnings
• Use only GE recommended power cords.
• When disconnecting the system from the power line, remove the plug from the wall outlet
first.
• Due to high voltage, use insulated screw driver.
• High voltage on test body; do not touch it during the test.
• To avoid the risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
• Always check that power cord and plug are intact and undamaged.
• All system devices must be connected to the same power supply circuit.
• Only interconnect devices when determined safe by qualified biomedical personnel.
• Only devices that are specified compliant with IEC 60950-1 or IEC 60601-1 may be
connected to the Ethernet MC ports.
• Biomed must determine interconnected parts are safe.
• Do not use without manufacturer approved mounting.
• If the software package is changed, all clinical settings will reset to factory defaults.
• Do not use with iCentral software V5.0.3 and earlier.
• Do not use Mobile Care Server software V5.2 and earlier.
• Verify compatibility of all system components prior to installation.
• Use only approved accessories, including mounts, and defibrillator-proof cables and
invasive pressure transducers. For a list of approved accessories, see the supplies and
accessories list delivered with the monitor. Other cables, transducers and accessories
may cause a safety hazard, damage the equipment or system, result in increased
emissions or decreased immunity of the equipment or system or interfere with the
measurement.
• Regular preventive maintenance should be carried out annually.
• Do not use multiple modules with identical measurements in the same monitor.
• The user may only perform maintenance procedures specifically described in this
manual.
• Incorrect power line frequency setting could adversely affect ECG processing.
• Make sure patient is not being monitored while servicing the equipment.
• Don’t press power key when changing language or doing factory reset.
• Never store the monitor with the batteries inside. Storing the monitor with the batteries
inside may result in damage to the monitor.
1-8
Cautions
• Set the time of a newly added network device as close as possible to the time of devices
already on the network.
1.2.5 Disposal
Dispose of the whole device, parts of it and its packing material and manuals in accordance
with local environmental and waste disposal regulations.
1-9
1-10
2 System description
2.1 Introduction
The B40i monitor build up a freely configurable modular system. The architecture is designed
to enable different module combinations so that the user is able to get the desirable parameter
and feature set. This modular approach makes it possible to add new features when they are
needed.
NAND
Ethernet
Flash
DATA BUS
AT91SA
USB HOSE RS485
M7s256
SDRAM
AT91 ARM IIC BUS Sound
Address bus
1-11
Marker Out
Frame
Hemo
Recorder E-module
module
1-12
+5V +13...16V
Isolation VMOD
Analog transformer
electronics
MODULE BUS
Patient isolation
A/D
PATIENT
CPU
convert
Data
RAM
EEPROM
Peripheral RS485
drivers Opto isolation
drivers
1-13
3.1.2 Display
The B40i use 12.1” LED backlight display with SVGA 800 x 600 resolution has bright long life
light bar and a wide viewing angle.
Backlights
The backlight of LCD is made up of 1 light bar, LED to be 3020, 20 pieces, 10 serials and 2
parallels. The backlight unit is driven by a separate driver board.
1-14
I/O connector
AC INLET Cable or wires
100-240 Vac
50/60 Hz Pin-to-pin connection
Alarm light
Speaker
board
AC/DC
Unit
LCD
display
SVGA
Nurse call LED
Serial data CPU driver
Power Board XY/CW Board board
Module bus
CW
Trim Knob
Battery board User interface
board XY Keyboard/Memb Power
External interface rane switch indicator
board
SMBUS
SMBUS
BAT1
BAT2
1-15
The board is based on AT91 ARM microprocessor. Other functions include LVDS display driver,
10/100Mbps on board Ethernet, WLAN communication, Alarm Light function, KEY board and
rotor encoder control, audio driver function, nurse call function, defibrillation function, module
bus function.
The CPU section takes care of the central processing.
The main features are:
• AT91 ARM
• 266 MHz Main CPU clock
• 64MBytes SDRAM
• 128 MBytes minimum NAND flash memory
• 4 MBytes Data flash memory
• Main CPU Provides one standard UART communication
Connectors
Ethernet communication connector
WLAN card operation connector
Color LED baclight display operation connector
Audio operation connector
Alarm Light indicator operation connector
Power board connection connector
Voltage supervision
There are two voltage supervision chips that control the system reset signals.
The +3.3V supervision chip outputs reset signals for +3.3V devices. 3.3V Reset Threshold will be
Falling: min 3.00 V; max 3.15 V.
1-16
The +1.2V supervision chip outputs reset signals for +1.2V devices. 1.2V Reset Threshold will be
Falling: min 1.08V; max 1.14 V.
1-17
MUX
+5V_PMC
Battery 2
Over Voltage
VSYS Enable
Protection Power Fail Alarm
Switch (Crowbar)
Over Current
15V Boost
Protection & Current VMOD
Regulator
Sense
LCD Backlight
Backlight_VCC
Enable Switch
1-18
AC/DC unit
The AC/DC unit is a compact medical power supply based on high-efficiency technology. It is
designed for 65 watt continuous output power, universal AC input and 15V output voltage.
3.1.6 Batteries
The B40i has two lithium-ion batteries, located in the battery compartment. The power board
connects one of the batteries to be the power source, if no power is received from the AC/DC
unit.The battery charging is controlled by the power board.
The batteries can be charged separately, and screen symbols and monitor frame LED
indicators indicate their charging level and possible failure.
NOTE: When the monitor is battery powered, the green battery LED is on. When the monitor is
mains powered, the green mains LED is on.
1-19
Network connector
RJ45 connector Pin Signal
1 Tx +
2 Tx -
1 2 3 4 5 6 7 8
3 Rx +
4 N/C
5 N/C
6 Rx -
7 N/C
8 N/C
1-20
Pin 1
Pin 2,3
1-21
Serial port
9 pin female connector Pin Signal
1 GND
2 SERIAL_TXD
5 1 3 SERIAL_RXD
4 SERIAL+3.3V
6
9
5 GND
6 N/C
7 SERIAL_CTS
8 SERIAL_RTS
9 N/C
Main power
Mains connector Pin Signal
L Live
PE Protected earth
N Neutral
1-22
1 1. InvBP connector
2
2. Temperature connector
3. SpO2 connector
4. ECG connector
3
5. NIBP connector
4
1-23
The monitor displays waveforms and measurement readings, and handles the trending and
alarm management. The ECG (e.g. heart beat and arrhythmia detection) and the Impedance
Respiration algorithms are in the monitor software. The modules measure signals and send
them to the monitor. The NIBP, SpO2, Temperature and Invasive Pressure algorithms are in the
module.
There are four parameter circuit boards inside the hemo-dynamic module for processing the
measurement signals. Each processing board has a microcontroller with software.
The NIBP parameter measurement requires one signal processing board, pneumatic system,
valve and pump unit connected to NIBP parameter board.
The second parameter board is the optional board, for Nellcor or Masimo SpO2 measurement,
it’s Masimo MS-2011 board or Covidien NELL1GE-S board at different configuration.
The third parameter board is for GE SpO2, IBP and Temperature measurement including input
board. All these three parameter is optional, according to different configuration, using
different board: it’s STP board, TP board for Nellcor, TP board for Masimo, GE SpO2 board.
The fourth parameter board is for 3/5-lead ECG with the Impedance Respiration measurement
including ECG input unit connected to the ECG parameter board.
All parameter boards are connected together via module bus flex board connecting voltage
and module communication, the module communicates with frame through RS-485 bus.
1-24
Valves
connector
EEPROM
1024Bytes
connector
Pump
Main CPU Pneumatic
AT91SAM7S256 control
RS485
Module bus connector
interface
256KBytes Flash
64KBytes SRAM
10bits ADC
Safety CPU
MSP430F2013
2KB+256B Flash
128B RAM
16bits Sigma-
Delta ADC
Signal processing
Two signals from the pressure transducers are amplified and sent to the A/D converter. After
the converter, digitized signals are sent to the microprocessor for data processing.
The NIBP board is controlled with an ARM7 microprocessor at 16 MHz oscillator frequency.
Memory
The NIBP program memory (processor flash memory) size is 256k x 8. The processor has 64
kBytes RAM. The EEPROM size is 8K x 8 and it is used to store the calibration values for the
pressure transducers, the pulse valve constants gained during measurements, the PC board
identification, and the module serial number.
Software control
The software controls valves and a pump. In addition to the individual on/off signals for each
component there is a common power switch for the valves and the pump that can be used at
pump/valve failures.
Safety circuit
The NIBP board is equipped with an independent safety circuit to disconnect supply voltages
from the pump and the valves if the cuff has been pressurized longer than the preset
maximum measurement time, or if the pressure of the cuff is inflated over the specified
pressure limit. The maximum measurement time values and pressure limits for different
measurement modes have been specified in the technical specification section of this manual.
1-25
Pneumatics
1
5, 6
3
4 1
1-26
ECG CABLE
- ECG LEAD SET
- ECG TRUNK CABLE
BASELINE
RESTORATION
ECG CPU
12_lead_ECG_meas_blck_dgrm.vsd
RS 485 POWER SUPPLY
COMMUNICATION NV
MEMORY
ISOLATION ISOLATION
ECG preamplifiers
The buffer amplifiers are used for each lead. The “Leads off” detection is implemented by
measuring the output level of the input buffer amplifiers with the A/D converter of the CPU. The
ECG signals are measured using differential amplifiers.
1-27
Pacemaker detection
Pacer detection has been made by using four slew rate detector circuits. The pacer detection
amplifiers have been realized at the front of the slew rate detectors independently of the ECG
measuring channels.
ECG CPU
The CPU is a 16 bit H8/3052 single-chip microcomputer. It contains 128 kbytes of flash memory
and 4 kbytes of RAM. The clock frequency is 16 MHz.
RS485 communication
The communication to the CPU board of the monitor uses RS485 protocol. The RS485 driver
circuits are optically isolated from the processor of the module.
Power supply
The ECG board has a driver-controlled half-bridge switching power supply with 5 kV isolation.
The supply voltages have been regulated with linear regulators.
ECG filtering
B40i monitors have three ECG filtering modes:
MONITORING 0.5 to 40 Hz
DIAGNOSTIC 0.05 to 150 Hz
ST FILTER 0.05 to 40 Hz
The purpose of filtering is to reduce high frequency noise and low frequency (e.g. respiratory)
movement artifacts.
The monitor filter is used in normal monitoring. The diagnostic filter is used if more accurate
diagnostic information is needed. The ST filter gives more accurate information of ST segment,
but reduces high frequency noise.
The high-pass filters 0.5 Hz and 0.05 Hz are done with software. The monitor sends a command
to the hemodynamic module determining which of the corner frequencies 0.5 Hz or 0.05 Hz is
to be used.
The 50 Hz and 60 Hz reject filters are both low-pass filters with zero at 50 Hz or 60 Hz
correspondingly. They are software based filters used for the mains supply filtering.
In diagnostic mode the upper frequency is 150 Hz and it is limited by software.
1-28
A/D CONVERSION
STP CPU
COMMUNICATION NV
MEMORY
ISOLATION ISOLATION
1-29
NV MEMORY
RS 485
POWER SUPPLY COMMUNICATION
ISOLATION ISOLATION
Microprocessor unit
The CPU is a 16 bit H8/3052 single-chip microcomputer. It contains 128 kbytes of flash memory
and 4 kbytes of RAM. The clock frequency is 16 MHz.
High speed I/O is used to obtain a pulse control sequence necessary for pulse oximetry
measurement. Timing for the clock is from the oscillator.
1-30
constant
Defibrillation/ESD protection current
resistors and diodes source
Temperature
sensors T1, T2
R Ref1 Ref2
T1 T2
Differential
R amplifier
0C:7k36 d/dt 0
0 To A/D
15C:3k54
converter
R
25C:2k53 0
38C:1k30
45C: 984 0
PSM_temp_meas_principle.vsd
R
1-31
Vin
Current
measurement G
to AD converter
Pressure
transducer
Instrum entation
amplifier
PSM_pressure_meas_principle.vsd
Vout
Input filter G
to A D converter
1-32
LP Oximeter channel 1
Gain=7.5
DC-
suppression
LP Oximeter channel 2
Gain=7.5
LP Oximeter channel 3
Amplifier: Gain=7.5
Preamplifier: Gain = 2
Current-to-voltage type
Spo2_measurement_blck_diagr.vsd
Bipolar/single-ended modes DC-
Adjustable gain suppression
Oximeter channel 4
LP
DE-MUX Amplifiers
Analog Digital
1-33
1-34
1-35
4.4.2 ECG
Electrocardiography analyzes the electrical activity of the heart by measuring the electrical
potential produced with electrodes placed on the surface of the body.
ECG reflects:
• electrical activity of the heart
• normal/abnormal function of the heart
• effects of anesthesia on heart function
• effects of surgery on heart function
See the “User's Guide” or the “User’s Reference Manual” for electrodes’ positions and other
information.
1-36
HbO2
SaO2 frac =
HbO2 + Hb + Dyshemoglo bin Formula 1
or against functional saturation SaO2func;
HbO 2
SaO 2 func =
HbO 2 + Hb Formula 2
Functional saturation is more insensitive to changes of carboxyhemoglobin and
methemoglobin concentrations in blood.
The oxygen saturation percentage SpO2 measured by the Datex-Ohmeda module is calibrated
against functional saturation SaO2func. The advantage of this method is that the accuracy of
SpO2 measurement relative to SaO2func can be maintained even at rather high concentrations
of carboxyhemoglobin in blood. Independent of the calibration method, pulse oximeters are
not able to correctly measure oxygen content of the arterial blood at elevated
carboxyhemoglobin or methemoglobin levels.
Pulse rate
The pulse rate calculation is done by peak detection of the plethysmographic pulse wave. The
signals are filtered to reduce noise and checked to separate artifacts.
Intensity of
Imax (DC-component) Imax
transmitted
light AC-component
Imin
Arterial blood
Venous blood
absorption_of_light.vsd
Tissue
Time
No pulsation Pulsatile blood
Incident light
1-37
IRED
Emitter
RED
PSM_absorption_of_infrared.vsd
Detector
1-38
Test methods used to establish the accuracy claims during low perfusion
Nellcor Oximax Technology with Oximax sensors
Nellcor Oximax Technology with Oximax sensors have been validated for SpO2 low perfusion
accuracy in bench top testing using Nellcor's PS II simulator with signal strength setting of
0.03% modulation and oxygen saturation levels of 70 to 100% at a pulse rate of 90 BPM.
Nellcor Oximax Technology with Oximax sensors have been validated for low perfusion pulse
rate accuracy in bench top testing using Nellcor's PS II simulator at a simulated low perfusion
level of 0.10% modulation in the pulse rate range of 40 to 250 beats per minutes (BPM) at a
SpO2 of 95%.
Masimo SET Technology with LNOP and LNCS sensors
Masimo SET Technology with LNOP and LNCS sensors have been validated for low perfusion
accuracy in bench top testing against Biotek Index 2 Simulator and Masimo's simulator with
signal strength setting of greater than 0.02% and a % transmission of greater than 5% for
saturation ranging from 70%-100%. One percent was added to accuracies for neonatal/infant
sensors to account for accuracy variation due to properties of fetal hemoglobin.
1-39
position. Pulse oximeter Sp02 readings were compared to SaO2 values of drawn blood samples
measured by CO-oximetry. Subjects comprised both adult men and women and spanned a
range of skin pigmentations.
4.4.4 Temperature
The temperature is measured by a probe whose resistance varies when the temperature
changes, called NTC (Negative Temperature Coefficient) resistor.
The resistance can be measured by two complementary methods:
• Applying a constant voltage across the resistor and measuring the current that flows
through it.
• Applying a constant current through the resistor and measuring the voltage that is
generated across it.
Hemo module uses the constant current method. The NTC-resistor is connected in series with a
normal resistor and a constant current is applied through them. The temperature dependent
voltage can be detected at the junction of the resistors, thus producing the temperature signal
from the patient. The signal is amplified by analog amplifiers and further processed by digital
electronics.
4.4.6 Respiration
Impedance respiration is measured across the thorax between ECG electrodes. The respiration
signal is made by supplying current between the electrodes and by measuring the differential
current from the electrodes. The signal measured is the impedance change caused by
breathing. The respiration rate is calculated from these impedance changes, and the
respiration waveform is displayed on the screen.
1-40
1 Installation
Safety precautions
Warnings
• The monitor or its components should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the monitor and its components
should be observed to verify normal operation in the configuration in which it will be used.
• After transferring or reinstalling the monitor, always check that it is properly connected
and all parts are securely attached. Pay special attention to this in case of stacked
mounting.
• Use only manufacturer approved mounts.
• Don’t install the monitor above the patient.
• Do not use the monitor in high electromagnetic fields (for example, during magnetic
resonance imaging).
• EXPLOSION OR FIRE -Using non-recommended batteries could result in injury/burns to
the patients or users. Only use batteries recommended or manufactured by GE.
• Do not incinerate a battery or store at high temperatures. Serious injury or death could
result.
• Allow temperature of the monitor to stabilize to operating temperature range before
connecting to power line.
• NETWORK INTEGRITY — The monitor resides on the CARESCAPE, S/5 network. It is possible
that inadvertent or malicious network activity could adversely affect patient monitoring.
The integrity of the network is the responsibility of the hospital.
• INTERFACING OTHER EQUIPMENT - Devices may only be interconnected with each other
or to parts of the system when it has been determined by qualified biomedical personnel
that there is no danger to the patient, the operator, or the environment as a result. In
those instances where there is any element of doubt concerning the safety of the
connected devices, the user must contact the manufacturers concerned (or other
informed experts) for proper use. In all cases, safe and proper operation should be verified
with the applicable manufacturer’s instructions for use, and system standards IEC
60601-1-1/EN 60601-1-1 must be complied with.
• The medical electrical equipment or medical electrical system may be interfered with by
other equipment, even if that other equipment complies with CISPR EMISSION
requirements.
Cautions
• The monitor display is fragile. Ensure that it is not placed near a heat source or exposed to
mechanical shocks, pressure, moisture or direct sunlight.
2-1
2-2
Connecting to mains
Connect the power cord to the mains power inlet at the back of the monitor and to the wall
socket.
NOTE: Before taking the monitor into use for the first time, the batteries should be fully charged.
Keep the monitor connected to the mains until the Battery charging symbol disappears (may
take up to 5 hours if the batteries are fully discharged).
WARNING The power cord may only be connected to a three-wire, grounded, hospital
grade receptacle
2-3
2-4
CARESCAPE Network
1
Hospital Intranet
1
Internet
RX
Network
1 1
HL7 PCD-01 HL7 PCD-01
Outbound Outbound
ApexPro® Telemetry
Dash® B40i B40i
Patient Monitor Patient Monitor Patient Monitor
1. HL7 outbound from monitor though CNI V2 switch, need ‘Unity and HL7’ license
Hospital Network
CIS/HIS Systems
2
Hospital Intranet
HL7 PCD-01
Outbound
B40i
Patient Monitor
2. HL7 outbound from monitor though hospital switch, need ‘HL7’ license
Notes
• This switch should support intervlan routing function.
• The hospital or the third party can choose GE offered switch or other types. If you need to
use GE offered switch, please contact to GE ND&I Team to setup CARESCAPE Network
Infrastucture version 2.
• If hospital or the third party design networks by themselves, the hospital or the third party
has full responsibility for the function and the risk of their network.
2-5
MC Network
• The MC Network infrastructure shall be installed according to the "CARESCAPE Network
Configuration Guide".
• The installation site of the patient monitor shall have a wall jack and a network patch
cable for the MC Network.
S/5 Network
• The S/5 Network shall be installed according to the "S/5 Network Installation Guide". Refer
to the "iCentral and iCentral Client Service Manual" for iCentral installation instructions.
• The installation site of the patient monitor shall have a wall jack and a network patch
cable for the S/5 Network.
HL7 network
The HL7 Network shall be installed according to the "B40i patient monitor HL7 reference
manual".
2-6
• Bed Name:
It is used for setting the bed name in the monitor. The default bed name is the last
five characters of the MAC address, excluding the colon delimiter characters.
• MCS IP Address
It is used for setting Mobile Care Server’s IP address which the monitor will talk to.
It is gray when network license is not “Unity” or “Unity+HL7”.
• Save Changes:
It is gray if the changes have not been done to the menu information.
NOTE: When save changes the monitor need restart.
4. Set up the TCP/IP configuration: TCP/IP Config > TCP/IP Config
• Configure the monitor’s IP address, subnet mask, default gateway.
NOTE: If the IP address or both of unit name and bed name duplicate with other
CARESCAPE network devices, the alarm will be given to the new added monitors.
• Save the configuration after configuring the TCP/IP.
NOTE: After saving changes the monitor need restart.
WARNING If only B40i monitors and Dash monitors in the Unity network, do not set up
B40i monitors with the highest IP address. For this may cause the Unity time
sync function can’t work with Dash monitors.
5. If the monitor does not connect to network correctly, you may also need to set up the
speed and duplex configuration:
TCP/IP Config > TCP/IP Config > Speed and Duplex
Choose the suitable settings according to your network environment. The default setting
is AUTO.
2-7
• Save Changes:
Select “Save Changes“ to take in use changes made in this page.
NOTE: In S/5 network protocol, no need to set up the TCP/IP settings.
NOTE: For other the network menu’s instruction, please refer to 3.2. Network in “Service
Menu” chapter for more details.
To use the E module, your monitor need pre-configure the extension rack from manufacture.
The E module options: airway gas module (E-miniC, E-sCO, E-sCAiO and N-CAiO), E-Entropy
module.
2-8
To insert module:
1. Align the module with the insertion guides
2. Push the module into the monitor frame until it clicks and stops.
3. Pull the module outwards to insure the module is firmly seated.
To remove module:
1. Pressing the release latch, on the bottom of the module.
2. Grasp the module firmly and pull out of the Frame. Make sure not to drop it when it comes
out.
WARNING When detaching modules, be careful not to drop them. Always support with
one hand while pulling out with other.
11
10
5 3
4
8
7
6
2-9
2-10
1 Instructions
1.1 Introduction
These instructions include procedures of system maintenance for the B40i. It’s include four
sections:
• Electrical safety tests.
• Installation checkout, which should be performed after installation and service
configuration.
• Maintenance and checkout, which should be performed every 12 months.
• Adjustments and calibrations
NOTE: Please complete the check form when performing the corresponding procedures.
NOTE: For the Gas modules maintenance and calibration, please refer to “8. E-miniC Module”
and “9. E-sCO, E-sCAiO and N-CAiO Modules“.
NOTE: For the Entropy module maintenance and calibration, please refer to “10. E-Entropy
Module“.
The symbol in the instructions means that the procedure performed should be signed in
the check form.
The procedures should be performed in ascending order, bypassing those that are not
applicable for a particular monitor.
To enter the service menus, you need following passwords:
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8)
In case you evaluate the measurement accuracy with a patient simulator, add the simulator’s
accuracy specification to the one for the monitor.
CAUTION Failure on the part of all responsible individuals, hospitals or institutions,
employing the use of this device, to implement the recommended
maintenance schedule may cause equipment failure. The manufacturer does
not, in any manner, assume the responsibility for performing the
recommended maintenance schedule, unless an equipment maintenance
agreement exists. The sole responsibility rests with the individuals, hospitals, or
institutions utilizing the device.
CAUTION Only trained personnel with appropriate equipment should perform the tests
and repairs outlined in this section. Unauthorized service may void warranty of
the unit.
CAUTION Wear a static control wrist strap when handling PCB boards. Electrostatic
discharge may damage components on the board.
3-1
Accessories
A rigid cylinder or pipe
NIBP cuff
Adult NIBP cuff hose with cuff ID
Infant NIBP cuff hose with cuff ID
Tubing parts to connect a manometer and a pump to
the NIPB cuff and hose.
Dual invasive pressure adapter cable
ECG accessories, IEC or AHA
- Multi-link 3-lead integrated cable and leadwire
- Multi-link 5-leadwire set
- Multi-link 3/5-lead ECG trunk cable
SpO2 finger probe
SpO2 Interconnect Cable
Temperature dual cable
CO2 Sampling line 3m/10 ft
Tool
A multiparameter patient simulator with IBP, Temp
adpter cables
Screwdrivers PH1, PH2
NOTE: For details on recommended accessories see “Supplies and Accessories“ catalog.
3-2
Perform electrical safety tests using an electrical safety analyzer per IEC 60601-1, UL 60601-1,
EN 60601-1 or CSA C22.2 No. 601.1. The schematics in the section provide a general
understanding of the test equipment. Actual configuration of test equipment may vary.
The patient monitor being tested should be placed on an insulating surface.
NOTE: Before proceeding, make sure that all test equipment is properly calibrated, maintained
and functioning.
NOTE: Refer to the instructions delivered with the safety analyzer to perform each test.
NOTE: GE recommends that the qualified personnel performing the tests should record the test
results of each electrical safety test, for example by using the installation / maintenance check
forms included in this manual.
System setup
These instructions are intended for every component in the system. Ensure that all system
components are properly connected to the patient monitor as described in chapter “2.
Hardware installation”.
3-3
WARNING Use only AC power cords recommended or manufactured by GE.
NOTE: The measuring device (MD) represents the network and voltage measuring instrument
and its frequency characteristics per IEC 60601-1.
3-4
Acceptance criteria:
• For equipment without a power supply cord, the impedance between the protective earth
terminal and any accessible metal part which is protectively earthed shall not exceed 0.1
ohms.
• For equipment with a power supply cord, the impedance between the protective earth
pin in the mains plug and any accessible metal part which is protectively earthed shall
not exceed 0.2 ohms.
Acceptance criteria:
• For equipment without a power supply cord, the impedance between the protective earth
terminal any accessible metal part which is protectively earthed shall not exceed 0.1
Ohms.
• For equipment with a power supply cord, the impedance between the protective earth
pin in the mains plug and any accessible metal part which is protectively earthed shall
not exceed 0.2 ohms.
3-5
Leakage Tester
Power Cord
HIGH NORM Power Cord
LOW
GND
GND
RVS Device
Under
Test
0.15µF
DMM 1K
10
NOTE: The measuring device (MD) represents the network and voltage measuring instrument
and its frequency characteristics per IEC 60601-1.
(1) Configure the safety analyzer as follows (NC):
• Polarity - NORMAL
• Neutral - CLOSED
(2) Power on the device under test.
(3) Read and record the current leakage indicated on the tester.
(4) Configure the safety analyzer as follows (SFC):
• Polarity - NORMAL
• Neutral - OPEN
(5) Read and record the current leakage indicated on the tester.
(6) Configure the safety analyzer as follows (SFC):
• Polarity - REVERSED
• Neutral - OPEN
(7) Read and record the current leakage indicated on the tester.
(8) Configure the safety analyzer as follows (NC):
• Polarity - REVERSED
• Neutral - CLOSED
(9) Read and record the current leakage indicated on the tester.
(10) Power off the device under test.
Acceptance criteria in NC (Normal condition):
• All readings shall be less than or equal to 300 μA for installations that require compliance
to UL 60601-1 requirements.
• All readings shall be less than or equal to 500 μA for installations that require compliance
to EN 60601-1 / IEC 60601-1 requirements.
3-6
Acceptance criteria in SFC (Single fault condition) - one of the supply conductors open at a time:
• All readings shall be less than or equal to 1 mA.
Perform the test in Normal Condition (NC) and in two different Single Fault Conditions (SFC): 1)
earth open and 2) one of the supply conductors open at a time. Perform the test with normal
and reverse polarity.
NOTE: Refer to the instructions delivered with the safety analyzer to perform this test.
Leakage Tester
Power Cord
RVS
Device
Open
Under
Closed GND Test
0.15µF
DMM 1K
10
Probe to exposed conductive chassis
NOTE: The MD represents the network and voltage measuring instrument and its frequency
characteristics per IEC 60601-1.
(1) Configure the safety analyzer as follows (NC):
• Polarity - NORMAL
• Neutral - CLOSED
• GND (Earth) - CLOSED
(2) Power on device under test.
(3) Read and record the current leakage indicated on tester.
(4) Configure the safety analyzer as follows (SFC):
• Polarity - NORMAL
• Neutral - OPEN
• GND (Earth) - CLOSED
(5) Read and record the current leakage indicated on the tester.
(6) Configure the safety analyzer as follows (SFC):
• Polarity - NORMAL
• Neutral - CLOSED
• GND (Earth) - OPEN
3-7
(7) Read and record the current leakage indicated on the tester.
(8) Configure the safety analyzer as follows (SFC):
• Polarity - REVERSED
• Neutral - CLOSED
• GND (Earth) - OPEN
(9) Read and record the current leakage indicated on the tester.
(10) Configure the safety analyzer as follows (SFC):
• Polarity - REVERSED
• Neutral - OPEN
• GND (Earth) - CLOSED
(11) Read and record the current leakage indicated on the tester.
(12) Configure the safety analyzer as follows (NC):
• Polarity - REVERSED
• Neutral - CLOSED
• GND (Earth) - CLOSED
(13) Read and record the current leakage indicated on the tester.
Acceptance criteria in NC:
• All readings shall be less than or equal to 100 µA
Acceptance criteria in SFC - earth open or one of the supply conductors open at a time:
• All readings shall be less than or equal to 300 μA for installations that require compliance
to UL 60601-1 requirements.
• All readings shall be less than or equal to 500 μA for installations that require compliance
to EN 60601-1 / IEC 60601-1 requirements.
3-8
Leakage Tester
1(875$/
($57+
RVS
Device
120K
Under
Closed GND
Test
Patient connector
0.15μF
Test Body
DMM 1K
10
NOTE: The MD represents the network and voltage measuring instrument and its frequency
characteristics per IEC 60601-1.
NOTE: Perform this test for all the connected parameter modules and patient connectors
specified in Table 2.
(1) Configure the safety analyzer as follows (NC):
• Polarity - NORMAL
• Neutral - CLOSED
• GND (Earth) - CLOSED
(2) Power on the device under test.
(3) Read and record the current leakage indicated on the tester.
(4) Configure the safety analyzer as follows (SFC):
• Polarity - NORMAL
• Neutral - OPEN
• GND (Earth) - CLOSED
3-9
(5) Read and record the current leakage indicated on the tester.
(6) Configure the safety analyzer as follows (SFC):
• Polarity - NORMAL
• Neutral - CLOSED
• GND (Earth) - OPEN
(7) Read and record the current leakage indicated on the tester.
(8) Configure the safety analyzer as follows (SFC):
• Polarity - REVERSED
• Neutral - CLOSED
• GND (Earth) - OPEN
(9) Read and record the current leakage indicated on the tester.
(10) Configure the safety analyzer as follows (SFC):
• Polarity - REVERSED
• Neutral - OPEN
• GND (Earth) - CLOSED
(11) Read and record the current leakage indicated on the tester.
(12) Configure the safety analyzer as follows (NC):
• Polarity - REVERSED
• Neutral - CLOSED
• GND (Earth) - CLOSED
(13) Read and record the current leakage indicated on the tester.
(14) Power off the device under test.
(15) Repeat this test for all the connected parameter modules and patient connectors
specified in Table 2.
Acceptance criteria in NC:
• All readings shall be less than or equal to 10 µA
Acceptance criteria in SFC - earth open or one of the supply conductors open at a time
• All readings shall be less than or equal to 50 µA
3-10
Leakage Tester
Power Cord
LINE NORM Power Cord
NEUTRAL
EARTH
RVS
Device
120K
Under
Closed GND
Test
NOTE: The MD represents the network and voltage measuring instrument and its frequency
characteristics per IEC 60601-1.
NOTE: Per IEC 60601-1, the impedance to protect the circuitry and the person performing the
test, but low enough to accept currents higher than the allowable values of the LEAKAGE
CURRENT to be measured.
WARNING Shock hazard. The following step causes high voltage at the test body. Do
not touch the test body.
NOTE: Perform this test for all the connected parameter modules and patient connectors
specified in Table 2.
(1) Configure the safety analyzer as follows:
• Polarity - NORMAL
• Neutral - CLOSED
• GND (Earth) - CLOSED
(2) Power on the device under test.
(3) Read and record the current leakage indicated on the tester.
(4) Configure the safety analyzer as follows:
• Polarity - REVERSED
• Neutral - CLOSED
• GND (Earth) - CLOSED
(5) Read and record the current leakage indicated on the tester.
(6) Power off the device under test.
(7) Repeat this test for all the connected parameter modules and patient connectors
specified in Table 2.
Acceptance criteria:
• All readings shall be less than or equal to 50 µA
3-11
3-12
3 Installation checkout
The purpose of the installation checkout procedure is to ensure that the system is properly
installed and configured for use.
Service personnel shall perform the following checkout procedure for the monitoring system
after the hardware installation and service configuration is completed:
1. 3.1. Visual inspection
2. 2. Electrical Safety Tests
3. 3.2. Functional inspection
Skip the tests that are not applicable for the installed monitor.
These instructions include a "Installation and checkout form, B40i" on page C-1 to be filled in
when performing the procedures.
3-13
3.2.2 Display
1. Verify that all text is readable and all images are clear.
2. Verify that the brightness is good. Adjust if necessary.
3.2.3 Time and date
1. Check that the clock on the screen shows correct time. Adjust the time and date, if
necessary.
Monitor Setup < Time and Date
NOTE: The B40i can’t be set as the TIME MASTER in network. You should adjust the time
and date from the central station.
3.2.4 Parameters measurements
Connect the accessories (no need to connect simulator/patient), check the monitor displays the
following messages or activities.
• ECG: After connecting ECG cable, ‘leads off’ will display in the Waveform Field
• NIBP: After connecting NIBP hose to module, ‘Adult/Pediatric’ or ‘Neonatal’ will display in
NIBP Digital Field for several seconds
• SpO2: After connecting SpO2 cable and sensor, SpO2 sensor will be lit.
• Temperature: After connecting Temp cable and sensor, ‘Performing temp test:’ will display
in Temp Digital Field for several seconds.
• IBP: After connecting IBP cable and transducer, ‘InvBP’s not Zeroed’ will display in
Message Field.
• Gas: After installing the gas module, ‘Calibrating gas sensor’ will display in CO2 waveform
field for about 1 minutes.
• Entropy: After installing the E-Entropy module and cable, 'No sensor' message will display
in Entropy digital field.
3.2.5 Recorder
1. Press the Recorder Start/Stop key and check that the module starts recording the
selected waveforms. Press the Recorder Start/Stop key again to stop recording.
2. Check that the quality of the recordings is acceptable.
3.2.6 MC or S/5 Network connection
NOTE: Pre-configure the network when install the monitor.
3-14
1. Check that the CAT-5 cable connector is clean and intact, then connect it to the Network
connector on the backside of the monitor.
Check that the monitor connects to the network, i.e. the network symbol appears on the
upper right-hand corner of the screen.
3.2.7 Conclusion
• Power off the monitor
• Perform final cleaning
• Fill in all necessary documents, refer to "Installation and checkout form, B40i" on page C-1
3-15
Before starting
• Save the patient data and monitor settings if necessary.
• Make sure that the monitor is turned off.
• Disconnect the mains power cord.
• If the monitor is connected to the network, disconnect the CAT-5 cable from the monitor.
4.2 Functional inspection
4.2.1 Start-up
1. Turn on the patient monitor.
Verify that the monitor starts up normally:
• The red, yellow and cyan alarm lights are lit in sequence.
• The speaker gives an audible beep.
3-16
• Check that the GE logo screen is displayed, followed by the notes screen and the
normal monitoring screen appears.
• Check and there are no error messages on the screen.
NOTE: Refer to section "Condition the battery" on page 3-23 to see the procedure for battery
conditioning if you receive a a Condition Battery X message.
NOTE: Batteries are recommended to be conditioned every six months.
4.2.2 Display
1. Verify that all text is readable and all images are clear.
2. Verify that the brightness is good. Adjust if necessary.
4.2.3 Keyboard(s)
1. Tests with the keyboard/membrane switch:
− Enter the service menu:
Monitor Setup < Install/Service (16-4-34) < Service (password 26-23-8) -
Keyboard
− Check the keys one by one, turn trim knob one full turn clockwise and one full turn
counter clockwise.
4.2.4 Time and date
1. Check that the clock on the screen shows correct time. Adjust the time and date, if
necessary.
Monitor Setup < Time and Date
NOTE: The B40i can’t be set as the TIME MASTER in network. You should adjust the time
and date from the central station.
4.2.5 Hemo Module
ECG measurement
• Configure ECG settings in monitor:
Connect 5 lead ECG cable to monitor and simulator.
Monitor Setup < Screen Setup < Waveform Fields < Field 1: ECG1; Field 2: ECG2;
Field 3: ECG3
ECG < ECG1 Lead: II; ECG2 Lead: V1; ECG3 Lead: aVL; ECG Size: 1.0
< ECG Setup < Beat Sound Volume: 1 or greater; Pacemaker: Show; HR Source: AUTO
• Configure ECG settings in simulator:
ECG rhythm: a normal sinus rhythm
3-17
2. Pacemaker Detection
− Configure the simulator to output "Asynchronous Pacemaker Pulse"
− Check that pacemaker spikes are shown on the ECG waveform.
3. Asystole Detection
− Configure the simulator to output "Asystole".
− Check that the 'Asystole' alarm appears to the monitor screen.
− Configure the simulator to show "80 beats per minute, Normal Sinus Rhythm".
4. Leads Off Detection
− Detach the RA/R leadwire from the simulator.
− Check that the Lead II waveform disappears from the ECG1 waveform field and a
message 'RA/R lead off'' is shown momentarily.
− Check that Lead II is replaced by Lead III in the ECG1 waveform field after a while
and a message 'Lead changed' is followed by a message 'Learning'.
− Reconnect the RA/R leadwire to the simulator.
− Check that Lead III is replaced back to Lead II in the ECG waveform field.
Respiration measurement
• Configure RESP settings in monitor:
Set up the Resp waveform field to the monitor screen:
Monitor Setup < Screen Setup < Waveform Fields < Field 4: Resp
And
Others < RespSetup < Resp Rate Source: Imped.
< Measurement: ON
• Configure RESP settings in simulator:
Baseline impedance: 1000
Amplitude: 1
Respiration rate: 20 breaths per minute
Lead selection: II (or LL)
5. Respiration Rate
− Check that the RESP waveform is shown and the RR value is 20 (±5).
− Configure the simulator's Apnea Simulation to "32 sec".
3-18
6. Apnea Detection
− Check that the monitor activates the Apnea alarm.
− Configure the simulator's Apnea Simulation to "OFF"
Temperature measurement
• Configure the “T1+T2” digit field to the monitor screen.
• Configure the simulator’s temperature channels as follows:
Temperature : 37 °C/98.6 °F
7. Temperature detection
− Check the corresponding temperature value appears and that no error messages
are shown on the monitor screen.
NOTE: If the deviation on a temperature reading on the screen is more than 0.1°C,
calibrate the temperature channels according to the instructions in chapter
"Temperature calibration" on page 3-25.
Invasive blood pressure measurement
• Configure the simulator’s IBP channels as follows:
Sensitivity: 5 µV/V/mmHg
InvBP outputs: "0 mmHg static pressure" or "atmosphere"
8. Zeroing
− Press IBP Zero All key.
− Check that a message "Zeroing" followed by a message "Zeroed" is shown in the IBP
parameter window.
9. Static Pressure
− Configure the simulator's InvBP output to "200 mmHg static pressure".
− Check that the flat pressure line appears on the related waveform field. The reading
in the parameter window shall be 200 ±10 mmHg.
NOTE: Recalibration is required, if the measured value is not within the specification.
Calibrate the invasive pressure channels according to the instructions in "Invasive
pressure calibration" on page 3-26.
10. Pressure Waveforms
− Configure the simulator's InvBP output to "Arterial 120/80".
− Check that the pressure waveform for tested invasive pressure channel appears in
the IBP waveform field and the Sys/Dia (Mean) pressure values are shown in the
related parameter window.
3-19
SpO2 measurement
11. Test measurement
Connect the SpO2 probe onto your finger. Check that the reading of 90-100 and SpO2
waveform appears. Check that the HR value is calculated from SpO2 when ECG and InvBP
(ABP or Art) are not measured.
NOTE: a functional tester cannot be used to assess the accuracy of a pulse oximeter probe or a
pulse oximeter monitor
Non Invasive Blood Pressure measurement
• Connect and set up the NIBP integrated tool as following picture, ensure all the
connections made are leak-proof.
13. NIBP calibration
− Calibrate the Non-invasive blood pressure (NIBP) channel according to the
instructions in "NIBP calibrations" on page 3-24.
14. NIBP hose detection
− Disconnect the calibration test equipment.
3-20
4.2.6 Loudspeaker
Check the loudspeaker by setting the alarm volume:
Alarms Setup < Alarm Volume
Test the whole volume scale from 1 to 10 by turning the Trim Knob and check that the
alarm volume changes correspondingly. The alarm sound should be clear and audible
with all the settings.
4.2.7 Monitor software
Enter the service menu:
Monitor Setup < Install Service (16-4-34) < Service (26-23-8)
Take down the information regarding Monitor software.
4.2.8 Watchdog circuitry
Select:
Monitor Setup < Install Service (16-4-34) < Service (26-23-8) < Set/Test
Perform the tests Watchdog and WD by Overload. Check that the monitor restarts in
each case.
NOTE: When selecting Watchdog and WD by Overload, auto restarting should take place
approximately after 10 seconds.
4.2.9 Network
NOTE: Pre-configure the network when install the monitor.
1. Check that the CAT-5 cable connector is clean and intact, then connect it to the Network
connector on the backside of the monitor.
Check that the monitor connects to the network, i.e. the network symbol appears on the
upper right-hand corner of the screen.
3-21
4.3 Monitor battery maintenance
The lithium-ion (Li-Ion) battery is a rechargeable battery containing lithium-ion cells. Each
battery contains an integrated electronic fuel gauge and a safety protection circuit.
The following are facts about lithium-ion battery technology:
• The battery discharges on its own, even when it is not installed in the equipment. This
discharge is the result of the lithium-ion cells and the bias current required for the
integrated electronics.
• The capacity loss of the battery degrades significantly at higher temperatures.
• As the battery ages, the full-charge capacity of the battery degrades and is permanently
lost. As a result, the amount of charge that is stored and available for use is reduced.
3-22
3-23
Calibration check
• Refer to the "Non Invasive Blood Pressure measurement" on page 3-20 in maintenance
checkout section for NIBP integrated tool set up.
1. Enter Calibration menu:
Monitor Setup - Install/Service (password 16-4-34) -Service (26-23-8) - Parameters -
NIBP - Calibrations
Calibration
Active Leak Test OFF
Calibration Check OFF
Protection OFF
Calibrate
Previous Menu
3-24
There is a reading of B1 and B2 in help filed under NIBP calibration menu. If the error of
pressure channel B1 and B2 is larger than specified above, the module should be recalibrated.
The error of B2 has no effect on blood pressure measurement accuracy.
NOTE: If the monitor pass the calibration check, no need to do NIBP calibration.
Calibration
1. Enter Calibration menu.
2. Disconnect the hoses from the NIBP connector to enable proper zeroing.
3. Select Calibrate. If it is not available, perform the steps a and b.
a. Select Protection OFF in the Calibration menu and push the Trim Knob.
b. Menu selection Calibrate is now enabled.
• Start calibration by pushing the Trim Knob. Messages ‘Zeroing’ and ‘Zeroed’ will be
displayed in the NIBP message field.
• Connect an external manometer with a pump to the module through the both tubes of
the hose - both transducers B1 and B2 will be calibrated simultaneously. Pump up to a
pressure of about 200 mmHg according to the manometer. Calibration is possible in the
range of 150 to 250 mmHg.
• Verify that both pressure values in the prompt field match the manometer reading. If not,
adjust by turning the Trim Knob. When the values of the pressure bar and the manometer
are equal, push the Trim Knob to confirm the calibration. The message ‘Calibrating’ and
‘Calibrated’ will be displayed on the NIBP digit field after a few seconds.
• To set the protection on:
Select Protection ON and push the Trim Knob.
NOTE: If a long time not choose ON manually, this menu will automatically return to
Protection ON.
3-25
Using a transducer
1. Connect the IBP accessories to the monitor, use a pressure manometer with a pressure
pump
2. Enter STP service menu.
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) -
Parameters - STP.
3. Enter Calibrations menu.
4. Choose Protection OFF in protect mode.
5. Connect a pressure transducer with a pressure manometer to the P1/P2 connector.
Choose Calibrate P1 or Calibrate P2. Leave the transducer to room air pressure.
6. Push the Trim Knob to start zeroing.
7. Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended
pressure is 200 mmHg.
8. Set the pressure on the display to match the pressure reading on the manometer and
push the Trim Knob. A tolerance of ±1 mmHg is allowed.
9. The message ‘Calibrated’ will be displayed on the display.
• Check Protection ON in protect mode.
NOTE: If a long time not choose ON manually, this menu will automatically return to
Protection ON.
3-26
Using a simulator
1. Connect the IBP accessories to the monitor and simulator.
2. Enter STP service menu.
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) -
Parameters - STP.
3. Enter Calibrations menu.
4. Choose Protection OFF in protect mode.
5. Choose Calibrate P1 or Calibrate P2. Set the P1 or P2 channel to 0 mmHg on the
simulator.
6. Push the Trim Knob to start zeroing.
7. Set a pressure of 100 mmHg to 300 mmHg on the simulator. The recommended pressure
is 200 mmHg.
8. Set the pressure on the display to match the pressure reading on the simulator and push
the Trim Knob. A tolerance of ±1 mmHg is allowed.
9. The message ‘Calibrated’ will be displayed on the display.
10. Check Protection ON in protect mode.
3-27
3-28
1 Introduction
The problems and solutions in this section represent only a few of the faults that you may
encounter and are not intended to cover every possible problem that may occur.
This chapter focuses on troubleshooting technical problems. For clinical and operation issues
refer to the "User’s Guide" for troubleshooting.
NOTE: For the Gas modules troubleshooting, please refer to “9. E-sCO, E-sCAiO and N-CAiO
Modules“.
NOTE: For the Entropy module troubleshooting, please refer to “10. E-Entropy Module“.
If the problem remains, call technical support for service. To ensure accurate problem solving,
please be prepared to provide the following information:
• Problem description and the troubleshooting done so far.
• Device information.
• Error messages displayed, if any.
• Other information, as requested.
4-1
Mains power
LED lit?
No
No
Yes
Disconnect the mains
Possible AC/DC unit
power cord and batteries.
failure. Replace the Power board failure
Reconnect and press the
AC/DC unit.
ON/OFF key.
Yes
Mains power
No
LED lit?
4-2
Start-up with GE
logo image?
Yes
No
Try with another
Wait for 120 software CD to
seconds. No download again. Please
read the note below.
Start-up display
No
appears?
OK? Yes
Yes
No
NOTE:
The software CD may be
defective.
4-3
2 Frame troubleshooting
Problem Cause What to do
Monitor is not starting. 1. The batteries are empty. 1. Connect the power cord.
2. Fuses may be blown. 2. Replace fuses.
3. If power cord connected, AC/DC 3. Replace the AC/DC power unit.
unit may be faulty.
4. If AC/DC unit is working, the 4. Replace the power board unit.
power board may be faulty.
5. On/Stby key may be faulty. 5. Replace the keyboard/membrane
switch panel or the
interconnection cable.
Monitor is not starting. 1. The connection between power 1. Check connection between power
board and CPU board may be board and CPU board.
faulty.
2. Faulty CPU board. 2. Replace the CPU board.
The monitor starts (alarm beep 1. The display cables are loose. 1. Check the display connection
is heard), but the display board connectors.
remains black. 2. The backlights are not lit. 2. Check inverter cable.
Backlight Inverter may be faulty.
Replace the Inverter board.
Display and monitor operating 1. Loudspeaker connector or wires 1. Check loudspeaker connector and
but no audible beep in start-up. loose or faulty. wires.
Display is too dim. 1. Incorrect brightness adjustment. 1. Adjust display brightness higher.
2. Backlight faulty. 2. Check inverter cable.
3. Backlight inverter faulty. 3. Replace backlight.
Stripes or white areas on screen. 1. Loose faulty display connection 1. Check display connection cable in
cable in CPU and display. CPU and display.
Module data disappears from Parameter module current (in Detach and change parameter
the screen. ‘Module power module bus) too high. module.
supply overload’ message.
Module data disappears. 1. Module bus voltage or signals 1. Replace module interface board.
path broken. 2. Replace the cable between
module interface board and
Power board.
3. Change the parameter module.
‘Replace Battery’ message on 1. Problem in communication 1. Replace battery.
the screen. between battery and power
board.
2. Battery too old or defected. 2. Replace battery.
3. power board may be faulty. 3. Replace the power board.
4. Problem in communication
between power board and CPU
'Frame temperature high' The temperature inside the frame is Check monitor ventilation holes.
message. too high.
‘Battery temperature high’ Battery SMBus temperature is too Check monitor ventilation holes.
message on the screen. high. Replace battery.
4-4
4-5
4-6
4-7
The vitial sign receiver system Check the Service Log, if have “AR” or “AE”
have error or reject to receive. message from vitial sign receiver system, this is
the receiver system’s issue.
4-8
3 Hemo Troubleshooting
3.1 NIBP troubleshooting flowchart
NIBP module not
working
Insert Hemo
module to a good
monitor
NIBP
Replace NIBP
No parameter display
board
on the screen?
Yes
Check if NIBP
NIBP Cuff loose connector spring
ENd Yes Start NIBP
function well? message appears No contact well with
without hose
on screen? the button of cuff
ID board
Yes
Pump check
Check pump
in Service Menu No
connector.
OK?
NOTE: Please refer to 5.4. NIBP Module in Chapter 5 for Service Menu instruction.
4-9
Does NIBP
Refer to NIBP
module work No
flowcharts
well?
Yes
Replace ECG
board
ECG function
No End
well?
No
4-10
Does NIBP
Refer to NIBP
module work No
flowcharts
well?
Yes
Go to service
menu
Is STP parameter
Select the right
configuration No
configuration
correct?
Yes
Replace STP/TP
board
End
4-11
3.4 NIBP
3.4.1 NIBP toubleshooting
4-12
4-13
4-14
3.5 ECG
Problem Cause What to do
HR numerical display No heart rate available. If no ECG waveform, check LEADS OFF
shows ‘---’ message and connect the leads.
If ECG waveform exists, check heart rate
source e.g. in the ECG Setup menu behind
ECG key.
Unacceptable ECG Poor electrode or poor electrode skin Electrodes from different manufacturers
waveform contact. are used. /Too much/little gel is used.
Poor electrode condition. Electrodes are dried out.
Improper site of electrodes. Check that electrodes are not placed over
bones, active muscles, or layers of fat.
Improper skin preparation. Remove body hair. Clean attachment site
carefully with alcohol.
Improper bandwidth filter. Check filter.
Dirty ECG cable. Clean the cable.
Faulty ECG cable Change the cable.
No ECG trace Waveform not selected on screen. Press the Monitor Setup key and make
adjustments.
Module not plugged in correctly. Check the hemo module’s installation.
Noise-message High frequency or 50/60 Hz noise. Isolate noise source.
4-15
4-16
4-17
3.8 Temperature
Problem Cause What to do
No temperature displayed Wrong type of probe. Use correct probe.
Temperature out of measurable range. The range is between 10 and 45 °C.
Temperature calibration not protected. Set the protection ON in the Service
Menu.
The monitor not configure TEMP. Check the monitor’s configuration.
4-18
1 Introduction
The monitor has a Service Menu, which is a useful tool to examine monitor functions and
troubleshoot in case a fault occurs.
Active Inactive SW
License
STP Calibrations
NIBP Calibrations
Safety Valve
SpO2
Pneumatics
Entropy
Set/Test
Service Log
5-1
Scroll Vers
Press Scroll Vers to scroll and view the detail version for monitor and modules on the extension
window.
Record Vers
Press Record Vers to record the detail version for monitor and modules on the extension
window.
Clear Password
The monitor will remember the service menu’s password you first enter. If you not restart the
monitor, the next time you enter to service menu, any random password can get the access.
Press Clear Password to clear the memory of password for service menu. When you next time
enter to service menu, can use other level password for different service menu’s access.
5-2
2 SW Management
The SW Management menu includes Software-specific service menus.
----------------------------
Previous Menu
----------------------------
number:-----------------
200
----------------------------
2.1 SW Download
The SW Download submenu show IP Address and enable software download.
Enable SW Download
Previous Menu
5-3
Previous Menu
5-4
2.5 License
The license submenu is used to set up monitor configuration for features.
MAC Address: Show MAC address for this monitor, this item is
fixed, can’t be selected. License
License key: Show License key, this item can’t be selected.
MAC Address 0425FEED0107
Produce type, IBP, OCRG, TEMP, ST, SpO2, Network are the
monitor’s configuration which to be setup. These menus need a Licence key
license key to open access.
Product type B40i
Enter key: To enter the license key to open access for setup
IBP
configuration.
OCRG
TEMP
If you want to upgrade the monitor, please contact GE
ST
representative and provide the MAC address to order a license
key. NeoResp
NOTE: After setting the monitor’s configuration via license key, SpO2 GE
monitor will perform a cold start. Network Unity
NOTE: If disable the IBP or TEMP function, the IBP and TEMP Enter key
related settings in each mode will turn OFF, include the following
settings: Previous Menu
- Waveform Fields
- Digit Fields
- Graphical Trends pages
- Snapshot fields
- Record Waveforms
The IBP and TEMP settings for Record Trends will change to Pleth.
5-5
3 Frame
The Frame menu includes frame-specific service menus.
----------------------------
----------------------------
----------------------------
number:-----------------
200
----------------------------
5-6
Power Frequency: Set the power frequency (50/60 National Reqs None
Hz). This setting is used to filter out possible power Power Frequency 60Hz
frequency interference from parameter Time Format 24 h
measurements.
Previous Menu
Time Format: Set the time format of the real-time
clock (12 h or 24 h).
3.1.1 Languages
Language
Country Settings
Select a language to be used during monitoring.
NOTE: Service pages will always appear in English despite Languages
of this selection.
Language ENG.LNG
NOTE: For language codes, see the table below.
Delete Language Delete Language ENG.LNG
Delete a language file from the permanent memory of the Previous Menu POR.LNG
monitor. Cancel
5-7
Abbreviation Language
CHI Chinese
CZE Czech
DAN Danish
ENG English
FRE French
GER German
HUN Hungarian
ITA Italian
NLB Dutch
NOR Norwegian
POL Polish
POR Portuguese
RUS Russian
SPA Spanish
SWE Swedish
TUR Turkish
FIN Finnish
5-8
3.2 Network
Different network configuration (S/5 or MC) will have different menus.
MC Network Network Status
Unity: Licensed
Network Config S5: Non-Licensed
TCP/I P HL7: Non-Licensed
S/5
Network Network Status
Unity: Non-Licensed
Dri Config S5: Licensed
Dri Comm HL7: Non-Licensed
Interfaces:
Ethernet Not Connected
WLAN Not Connected
Current Interface: Eth_B40
Connection:
5-9
3.2.2 TCP/IP
TCP/IP Status
The TCP/IP Status view shows the general status of the TCP/IP set up.
Data link: “NIC” for ehternet,
TCP/IP TCP/IP Status (Ethernet)
DHCP Status: “Disabled“ or “Enabled“ depending
on the DHCP status. Data Link: NIC
TCP/IP Confing
IP Address 172.16.1.54
The DHCP status related information displayed Ping Subnet Mask 255.255.0.0
below. Gateway IP 172.16.254.254
Previous Menu DNS Server Not available
DHCP Server Not available
SLP Not available
Logins 0
5-10
TCP/IP config
IP address: For setting the static IP address of monitor.
Subnet Mask: For setting the static subnet mask of
monitor.
Default Gateway: For setting the static default gateway
of monitor.
Speed and Duplex: For setting speed and dupliex of
monitor.
Save Changes: Close menu and save the current menu
information to permanent memory.
Cancel Changes: Close menu without saving changes
Ping
IP Address: For configuring the destination IP address for the ping command.
Ping: Enable the ping command by sending Internet Control Message Protocol (ICMP) echo
request packets to the target host and wait for an ICMP response. The response should be
show in the Ping Status view.
3.2.3 HL7
The HL7 setup information is on HL7 Status view.
5-11
HL7 config
5-12
5-13
Ethernet
The Ethernet Status view shows the general status of the ethernet network communication.
5-14
Voltages
ACDC AC/DC converter’s output voltage, used as
monitor input voltage when the mains cord is
Power Supply Voltages Min Mean Max
ACDC 14.90 14.90 14.90
connected. Range when present: 15.25…16.55V Power page EXTDC 0.00 0.00 0.00
(Note: this includes the measurement inaccuracy). Bat1 12.00 12.00 12.08
WPM Battery Bat2 12.24 12.24 12.24
EXTDC Not used.
VSvs 14.92 14.92 14.92
Bat1 Battery A voltage measured at power Board. Previous Menu
0.00
VBoost 0.00 0.00
Range 9…12.6V for Li-ION battery VMod 14.98 14.98 14.98
Bat2 Battery B voltage measured at power 1/0 Mod 0.00 0.00 0.00
Board. Range 9…12.6V for Li-ION battery +5V 5.04 5.04 5.04
+3.3V 3.35 3.36 3.37
VSys System voltage at power board. This is the Currents
monitor input voltage measured at the power Mod Current 0.31 0.33 0.39
board after input voltage selection. Range Sys Current 0.87 1.23 1.34
9…16.5V. Bat Current 0.00 0.00 0.00
Powers
VBoost Not used. System Power 12.58 13.87 13.38
VMod Supply voltage for modules. VBOOST is fed Module Power 4.35 4.40 4.84
through a circuit breaker to VMOD. Range Temp C
Power 38.77 38.77 38.77
13.8…16V. Therm Not CHG 0.00 0.00 0.00
I/O Mod Not used. Dummy CHG 0.00 0.00 0.00
CPU Board A/D
+5V At power board. Range 4.8...5.3V.
CPU temp ( C ) 0.00 0.00 0.00
+3.3V At power board. Range 3.15…3.45V. VSvs Out V 0.00 0.00 0.00
Currents +3.3V 0.00 0.00 0.00
Mod Current Current from module bus voltage +5V 0.00 0.00 0.00
VMOD. Depends on the module configuration.
Sys Current Current from system voltage VSYS. Depends on the system configuration and battery charging.
Bat current Current from or to the battery selected (discharge or charge). Measured at power Board. Depends on
the system configuration and battery charging.
Powers
System Power Power from VSYS, calculated by power Board software. System power = VSYS * SYS Current
Module Power Power from VMOD, calculated by power Board software. Module power = VMOD * Mod Current
Temp
Power1, Power2 Power supply unit temperature, measured at power Board.
Therm Not CHG Not used.
Dummy CHG Not used.
CPU Board A/D
CPU Temp Not used.
VSYS_OUT Not used.
+3.3VMeasured at CPU Board. Range 3.15…3.45V.
+5V Measured at CPU Board. Range 4.75…5.25V.
5-15
5-16
5-17
Smart Batt1
Battx information from SMBus (this information Batt1 information from SMBus
is received from the battery via SMBus) WPM Battery Temperature 30.75 C
Temperature: Battery temperature Batts Voltage 11903 mV
Current 0 mA
Voltage: Battery voltage Smart Batt1 Avg. Current 0 mA
Rel. State of Charge 97 %
Current: Battery current (discharge or charge) Smart Batt2
Abs. State of Charge 72 %
Clear Temp Maxs Remaining Capacity 2532 mAh
Avg. Current: Rolling average of the battery Full Charge Capacity 2604 mAh
current Previous Menu Cvcle Count 33
Design Capacity 3520 mAh
Rel. State of Charge: Relative state of charge of Design Voltage 11100 mV
the battery. Expressed as a percentage of Rem.
Manufacture Date(DD:MM:YY) 20/12/24
Cap. (mAh) / Full Cap.(mAh).
Abs. State of Charge: Absolute state of charge. Manufacture Name NPC A07A90
Device Name SM201-6
Expressed as a percentage of Rem. Cap. (mAh)/ Device Chemistry LION
Design Capacity (mAh). Therm. Status from charger
5-18
Smart Batt2
If Smart Battx is selected for a battery NOT connected to the SMBus, the menu contains the
following:
Battx information from Memory and PMC (This information comes from the power Board
memory or is measured by the power Board. SMBus data in this menu is not real time, because
this battery is not connected to the SMBus).
NOTE: This page may not contain information if SMBus has been connected only to the other
battery. Page can be updated by battery disconnection and reconnection, if desired.
5-19
4 Keyboard
The service menu for testing the command board functions.
Service Data
Message Count counts the number of messages that are sent out to the main CPU board.
Leds upper and lower indicate the states of the alarm lights on the monitor.
Direct action keys texts are indications to the command board membrane keys. When a key
on the command board is pressed, the corresponding text in the menu changes its color.
Control wheel, Press counts the Trim Knob pressings.
Control wheel, Clockwise and Counter clockwise indicate the Trim Knob turnings.
5-20
Scroll Log: enables to scroll the keyboard log. Keyboard Keyboard Log
Scroll Stat: enables to scroll the keyboard Scroll Log
presses events. Scroll Stat Keyboard : Keyboard Log
2004- Jan- 29 05:40:54
Record Log: to record keyboard log.
Record Log Service Menu : Keyboard
Record Stat: to record keyboard stat. 2004- Jan- 29 05:40:40
Record Stat
Reset Log: to reset the keyboard log. Display : Previous Menu
Reset Log 2004- Jan- 29 05:40:39
Service menu : Display
Previous Menu
2004- Jan- 29 05:40:11
Frame : Previous Menu
2004- Jan- 29 05:40:09
Network : Previous Menu
2004- Jan- 29 05:40:05
Network Config : Previous Menu
2004- Jan- 29 05:40:00
TCP/IP : Previous Menu
2004- Jan- 29 05:39:57
TCP/IP Config : Previous Menu
2004- Jan- 29 05:39:40
5-21
5 Parameters
NOTE: Parameter values in Service Data fields are only for reference in this section.
5-22
Timeouts -12867
Bad checksums 0
Bad c-s by moc 0
5-23
5.1.2 Gases
Service Data
O2, CO2, N2O, AA % shows real-time concentrations.
noise-% is standard deviation of concentration.
O2, CO2, N2O, AA channels A-E mV: signal is scaled to mV.
Gain User gain. It is scaled as (User gain)/(Factory gain).
ID Identified agent.
ID unrel. The shown value tells how unreliable the identification is. With pure agent the value is normally
<50.
MAC MAC value
Sample Flow calculated from differential pressure and adjusted by the module. Zero value as measured
during initialization when the pump is off. Gain: sample flow measurement can be calibrated
by adjusting the gain.
Ambient Ambient pressure is measured every 30 min.
Amb-Work Ambient pressure - sampling system internal pressure.
Fall time CO2 and O2 in ms. For N2O and AA same as CO2.
CO2-O2 Delay In ms. No delay between CO2, N2O, and agents.
Pump Can be toggled ON/OFF. PWM output 0-100% is shown. Pump voltage is also shown.
Lamp The state, PWM control, and current of the lamp are shown.
Fan The state of the fan is shown.
Zero and Occl Valve Can be toggled between measurement state (MEAS) and zeroing/occlusion states
(ZERO/OCCL).
Temp Temperatures measured by the module from TPX, CPU, and OM.
Time after power on In minutes after power on.
5-24
Service Data
5-25
Bad c-s by mod is a cumulative number that indicates how many communication errors the
module has detected.
The monitor starts counting these items at power up and resets to zero at power off. The values
may also be reset when a module is attached to the monitor frame and be set to 32769 or
continuous counting may be started when the module is removed from the monitor frame.
The nonzero values do not indicate a failure, but the continuous counting (more than 5 per
second) or value 32769 indicates either a serial communication failure or a module not in place.
Also failures in other modules may cause these numbers to rise or be set to 32769.
RAM indicates the state of the RAM memory.
ROM indicates whether the checksum at the EPROM is in accordance with the one the software
has calculated.
EEPROM indicates if the values stored in the permanent memory are valid.
The state is either OK, Fail or ? (module not in place or a communication error).
LeadsOff indicates whether the monitor can measure ECG even if one or more leadwires are off:
1 = measurement is not possible, 0 = measurement can be done.
MissedPkgs indicates the number of packages missed.
5-26
5-27
Temp error shows the status of the temperature test. No errors found show the
status (OFF) and errors found (ON).
Protect key and Protect mode show normally ON but can be turned to OFF for
the temperature calibration in Calibration Menu.
Configuration shows the chosen module configuration: T, P, TP, ST, or STP.
Timeouts is a cumulative number that indicates how many times the module
has not responded to the monitor’s inquiry. Bad checksums is a cumulative
number that indicates how many times communication from the module to the
monitor has broken down.
Bad c-s by mod is a cumulative number that indicates how many
communication errors the module has detected.
The monitor starts counting these items at power up and resets to zero at
power off. The nonzero values do not indicate a failure, but the continuous
counting (more than 5 per second) indicates either a serial communication
failure, or a module not in place. Also other modules can cause communication
errors that cause these numbers rise.
RAM indicates the state of the RAM memory.
ROM indicates whether the checksum at the EPROM is in accordance with the
one the software has calculated.
EEPROM indicates if the values stored in the permanent memory are valid.
The state is either OK, Fail or ? (module not in place or a communication error).
5-28
5.3.1 Calibrations
5-29
5-30
How to calibrate
NOTE: Perform NIBP Calibration Check first to evaluate if calibration is needed or not.
NOTE: Both transducers B1 and B2 will be calibrated simultaneously.
NOTE: The module must be in the frame during the whole procedure.
NOTE: Calibration selection is available only when protection is OFF.
1. Change the protection setting from ON to OFF to enable the Calibrate selection - the
color of the Calibrate selection changes from grey to white.
2. Zeroing:
− Disconnect the NIBP hose from the module connector.
− Select Calibrate and push the Trim Knob.
NOTE: Messages ‘Zeroing’ and ‘Zeroed’ is shown in the NIBP message field and next to the
Calibrate selection momentarily. After this, a pressure bar will appear beside the menu.
3. Calibration:
− Connect the NIBP hose to the module connector.
− Connect an external manometer with a pump to both tubes of the hose.
− Pump about 200 mmHg pressure.
− Verify that both pressure values, B1 and B2, shown in the prompt field of the
calibration menu match the manometer reading. If not, adjust the by turning the
Trim KnobTrim Knob.
− Press Trim Knob to complete the calibration.
5-31
NOTE: Messages ‘Calibrating’ and ‘Calibrated’ are shown in the NIBP message field and next to
the Calibrate selection.
NOTE: When calibrating NIBP, always change the displayed pressure value slightly with the
Trim Knob, even in cases where the value would be correct. For example, change the value one
step higher and then back one step lower. The ‘Calibrated’ text should appear in the display.
This ensures that the calibration procedure is correctly registered and stored by the module.
4. Change the Protection setting from OFF to ON to disable Calibrate selection - the color of
the Calibrate selection changes from white to grey.
NOTE: If a long time not choose ON manually, this menu will automatically return to
Protection ON.
Start test is for starting and Stop test for stopping the
Safety Valve test. Safety Valve Safety Valve Data
B1 B2
Pressure 000000 000000
Zero 000000 000000
Safety Valve Data: ADULT
Start Test St1 0000 AD0 -16
For information on general items Pressure, Zero, St1 to St2 0000 AD1 6
Previous Menu
St4, AD0 to AD7 as well as Timeouts etc., see service St3 0400 AD2 -1
St4 0000 AD3 1502
data descriptions in section 4.4 NIBP Module. AD4 2
Max. press and 2 s after stop show the measured AD5 -1643
AD6 4
values at Safety Valve test. AD7 -1505
B1 B2
Max press 0 0
2 s after stop 0 0
Timeouts 2 RAM OK
Bad checksums 0 ROM OK
Bad c-s by mod 0 EEPROM OK
5-32
9. Check the pressure values ‘Max press’ for both transducers. All the values should be
within 150 - 165 mmHg.
10. Return to the normal monitoring mode by pressing Normal Screen.
5-33
5.6 Entropy
When the Entropy Service Menu is open, the function of Entropy module keys will be disabled.
Check Sensor: to replace module keys function to check the entropy sensor.
Module: to show the Module status in the right window.
Sensor: to show the Sensor status in the right window.
5-34
Module status
Measurement: show messurement status
Last Service: Service log
Entropy, Check sensor: to check the
module keyboard. Press the module keys,
these 2 items will highlight.
Main: If the pulse search is being done.
BSR status: show the status for Burst
supression classification. There are 4
status: Artefact; Supression; Burst; Invalid
(not used)
Artefact: artefect whether detected.
Diathermy: diathermy whether detected.
AD clipped: AD-converted value whether
clipped
Alg.ver.(Mon), Alg.ver.(Mod), Alg.ver.used:
These 3 items show version of entropy
algorithm, B40i always use algorithm from
monitor.
For information on Timeouts etc., see
section "NIBP Module."
Sensor status
Entropy, Check sensor: to check the
module keyboard. Press the module keys,
these 2 items will highlight.
Cable, Sensor: whether detect cable and
sensor.
Sensor S/N, Sensor Lot: show the sensor
information
Lead: whether detected each of 3 points
of sensor.
Imped.: imped. value for each of 3 points
of sensor.
Imp. Meas: whether Imped. measurement
on.
Imp.: whether is ok for Imped
measurement.
5-35
6 Set/Test
The system contains a watchdog circuitry, which needs refreshment at every 10 seconds. If the
refreshment did not occur, the watchdog will reset the main CPU.
The purpose of the watchdog is to restart the monitor, if there was a serious malfunction. This
feature is useful in two cases: when the main CPU is not able to control the monitor, or when
the CPU controls the monitor but detects a serious malfunction. Watchdog tests check proper
functionality of the watchdog in various conditions.
5-36
7 Service Log
Error, event and alarm data is stored in the Service Log.
The service log contains information about the occurred monitor errors, events and alarms
since the last factory reset or service log reset.
Error Log is for selecting the error history view onto
the right side of the menu. Error Log shows also Service Log Error History
some monitoring events like warm and cold starts.
Error Log
Alarm Log is for selecting the alarm history view onto Last errors:
Alarm Log
the right side of the menu.
Scroll Log 2010-Nov-27 13:50:08
Scroll Log is for scrolling the error/alarm information UMBC handler: LOG Msg<Umbc reset> at
on the right side of the menu. Record Log UmbcChannel. cpp(189)
5-37
5-38
1 Spare part
The following B40i parts will be available as field replaceable spare parts.
6-1
6-2
1.3 Frame
6-3
6-4
* NOTE: 2061869-040 FRU B40i keyboard/Membrane includes multi-languages labels. When service people replace front
cover, please order both 2061869-002 and 2061869-040.
6-5
2 Disassembly
WARNING A short circuit may cause internal damage. Do not touch any exposed
wiring or conductive surface inside, this may cause an electric shock.
CAUTION Perform leakage current measurement whenever service or repair has been
done in the monitor.
Field repair of the patient monitor is limited to replacing Field Replaceable Units (FRUs). See “1.
Spare part” for a detailed list of available FRUs. Attempting a field repair on a printed circuit
board or a factory sealed component or assembly could jeopardize the safe and effective
operation of the monitor.
NOTE: Only a qualified service technician should perform field replacement procedures.
NOTE: Perform the checkout procedure described in chapter 3 “4. Maintenance and checkout”
after you have disassembled and reassembled the module.
6-6
3. From the back side of the monitor, pull the extension rack
out of frame, grasp it firmly.
6-7
7. Use the screwdriver to prize, pulling out the multi I/O at the
same time.
6-8
6-9
To remove the alarm light board After opening the front cover
To remove the user interface board After opening the front cover
6-10
6-11
- Remove screw
- Disconnect the cable to power board.
- Disconnect the cable to ethenet board
- Slightly lift the front of the ACDC unit, push the whole ACDC
unit to top direction.
6-12
6-13
6-14
6-15
6-16
6-17
6-18
1 General Specifications
1.1 Genenral specifications
Size
Without extension module 312±5 mm (H) * 312±5 mm (W) * 158±5 mm (D)
With extension module 312±5 mm (H) * 352±5 mm (W) * 178±5 mm (D)
Weight
With extension module ≤ 7 kg
Monitor Environment
Operating temperature Normal operation: 5 to 40°C (41 to 104°F)
Charging batteries: 5 to 35°C (41 to 95°F)
Storage and transport -20 to 60°C (-4 to 140°F)
temperature
Operating humidity 20 to 90% noncondensing
Storage and transport 10 to 90% noncondensing
humidity
Operating atmospheric 700 to 1060 hPa
pressure (525 to 795 mmHg)
Storage and transport 700 to 1060 hPa
atmospheric pressure (525 to 795 mmHg)
E-miniC module Environment
Operating temperature 10 to 40°C (50 to 104°F)
Non-operating -25 to 70°C (-13 to 158°F)
temperature
Operating humidity 10 to 95% noncondensing
Non-operating humidity 10 to 95% noncondensing
Operating altitude 666 to 1060 mbar
E-sCAiO, E-sCO, N-CAiO modules Environment
Operating temperature 10 to 40°C (50 to 104°F)
Non-operating -25 to 60°C (-13 to 140°F)
temperature
Operating humidity 10 to 98% noncondensing
Non-operating humidity 10 to 90% noncondensing
Operating altitude 660 to 1060 mbar
E-Entropy module Environment
Operating temperature 10 to 40°C (50 to 104°F)
7-1
7-2
7-3
2 Parameters specifications
2.1 ECG specifications
7-4
NOTE: Pacemaker detector may not operate correctly during the use of high-frequency (HF)
surgical equipment. The disturbances of HF surgical equipment typically cause false positive
pacer detection.
7-5
7-6
* Because SpO2 measurements are statistically distributed, only about 2/3 of the measurements can be expected to fall
within ±1 Arms of the value measured by a CO-oximeter
Test methods used to establish SpO2 accuracy: GE TruSignal SpO2 measurement have been validated for no motion and
motion accuracy in a controlled hypoxia studies with healthy non-smoking adult volunteers over the specified SpO2 rage.
SpO2 readings were compared to SaO2 values of drawn blood samples measured by CO-oximetry. Subjects comprised
both adult men and women and spanned a range of skin pigmentation.
** GE TruSignal technology have been validated for low perfusion SpO2 accuracy over the specified range in a bench top
testing against BioTek Index 2 patient simulator with 0.3% signal amplitude.
Table 2: The table below shows Arms values measured using GE SpO2 sensors with
GE CARESCAPETM V100 in a clinical study.*
GE SpO2 Sensor 70 - 80% 80 - 90% 90 - 100%
OXY-E 2.3 digits 1.4 digits 1.3 digits
OXY-SE 2.5 digits 2.0 digits 1.1 digits
OXY-F 1.3 digits 1.0 digits 1.1 digits
OXY-W 2.9 digits 1.8 digits 1.0 digits
OXY-AP 2.0 digits 1.9 digits 1.7 digits
OXY-AF 2.5 digits 1.4 digits 0.9 digits
* The sensors were clinically tested for accuracy with the following sensors: OXY-E (equivalent to OXY-E-UN, TS-E-D,
TS-E2-GE, TS-E4-GE), OXY-SE (equivalent to OXY-SE-3, TS-SE-3), OXY-F (equivalent to OXY-F-UN, TS-F-D, TS-F2-GE,
TS-F4-GE, TS-SA-D, TS-SA4-GE), OXY-W (equivalent to OXY-W-UN, TS-W-D), OXY-AP (equivalent to OXY-AP-10, OXY-AP-25,
TS-AP-10, TS-AP-25), OXY-AF (equivalent to OXY-AF-10, TS-AF-10, TS-AF-25).
The sensors were clinically tested for neonatal accuracy with the following sensors: OXY-SE (equivalent to OXY-SE-3,
TS-SE-3), OXY-AF (equivalent to OXY-AF-10, TS-AF-10, TS-AF-25).
7-7
7-8
2.6 NIBP
7-9
2.8 Temperature
7-10
7-11
7-12
Anesthetic agents rise time Hal, Enf, Iso, Des: < 420 ms with a 3 m sampling line
Hal: < 800 ms with a 3 m sampling line
Hal, Enf, Iso, Des: < 700 ms with a 6 m sampling line
Hal: < 1800 ms with a 6 m sampling line
Hal drift < 0.1 vol%
Enf drift < 0.1 vol%
Iso drift < 0.1 vol%
Sev drift < 0.1 vol%
Des drift < 0.3 vol%
Effects of interfering gases and vapors
E-miniC module
Non-disturbing gases of which - Ethanol C2H5OH (<0.3%)
effect on CO2 (5 vol%) readings - Acetone (<0.1% )
< 0.2 vol%:
- Methane CH4 (<0.2%)
- Nitrogen N2 (0-100%)
- water vapor (0-100%)
- Trichloromonofluoromethane (<1%)
- Dichlorotetrafluoroethane (<1%)
- Dichlorofluromethane (<1%)
Disturbing gas and its effect • Halotane (4%): increases CO2 (5 vol%) < 0.3 vol%
• Isoflurane (5%): increases CO2 (5 vol%) < 0.4 vol%
• Enflurane (5%): increases CO2 (5 vol%) < 0.4 vol%
• Desflurane (24%): increases CO2 (5 vol%) < 1.2
vol%
• Sevoflurane (6%): increases CO2 (5 vol%) < 0.4 vol%
• N2O (40%): increases CO2 (5 vol%) < 0.4 vol%
• Helium (50%): decreases CO2 (5 vol%) < 0.3 vol%
• If O2 compensation is not activated: O2 (40-95%)
decreases < 0.3 vol%
• If O2 compensation is activated: O2 (40-95%) error
< 0.15 vol%
• If N2O compensation is not activated: N2O
(40-80%) increases < 0.8 vol%
• If N2O compensation is activated: N2O (40-80%)
error < 0.3 vol%
E-sCO, E-sCAiO, N-CAiO module
7-13
Non-disturbing gases:
- Ethanol C2H5OH (< 0.036%)
- Acetone (< 0.2%)
- Methane CH4 (< 0.3%)
- Isopropanol (< 0.48%)
- Nitrogen N2
- Carton Monoxide CO (< 100 ppm)
- Nitrous Oxide NO (< 200 ppm)
- Freon R134A (< 1%) (for CO2, O2 and N2O)
- Water vapor
Effects of a non-disturbing gas • CO2 < 0.2 vol%
to the measured gas
concentrations • N2O < 2 vol%
• O2 < 2 vol%
• Anesthetic agents < 0.15 vol%
Gas cross effects • Helium (50 vol%):
Decreases CO2 (5 vol%) readings < 0.5 vol%
Decreases O2 (50 vol%) readings < 2 vol%
• Xenon (80 vol%):
Decreases CO2 (5 vol%) readings < 0.5 vol%
Decreases O2 (14 vol%) readings <1.5 vol%
NOTE: E-miniC measurement is intended for patients weighing over 5 kg (11 lb).
2.10 Entropy
Display range
Response Entropy (RE) 0 to 100
State Entropy (SE) 0 to 91
Burst Supression Ratio (BSR) 0 to 100%
Display accuracy ±1 or ±1%
Amplifier 1 MΩ @ 50 Hz
input impedance
Defibrillation protection 3000 V, 130 J
Amplifier frequency range 0.5 to 118 Hz
7-14
1 Product overview
1.1 Introduction
This chapter provides information for the maintenance and service of the E-miniC module. The
module is a single-width plug-in module designed for use with the GE modular monitors.
The E-miniC module provides airway measurements.
Letter C in the module name stands for CO2.
NOTE: E-miniC is intended for patients weighing over 5kg (11lb).
8-1
8-2
Mini D-fendTM
The sample is drawn through the sampling line. The gas then enters the module through the
water trap, where it is divided into two flows, a main flow and a side flow. The main flow goes
into the analyzer. This flow is separated from the patient side by a hydrophobic filter. The side
flow creates a slight subatmospheric pressure within the Mini D-fend water trap which causes
fluid removed by the hydrophobic filter to collect in the bottle.
Zero valve
The main flow passes through a magnetic valve before proceeding to the analyzer. This valve is
activated to establish the zero point for the MiniC measuring unit. When the valve is activated,
room air is drawn through a filter into the internal system and the gas sensor.
8-3
NafionTM tube 1)
A Nafion tube is used between the water trap and the zero valve to balance the sample gas
humidity with that of ambient air. The tube prevents errors caused by the effect of water vapor
on gas partial pressure when humid gases are measured after calibration with dry gases.
Gas analyzers
After the zero valve and Nafion tube, the gas passes through the miniC sensor.
Sampling pump
The gas sampling pump is a membrane pump run by a DC-motor. Sample flow is measured
with a differential pressure transducer across a known restriction. The motor is automatically
controlled to maintain a constant flow even when the D-fend water trap ages and starts to get
occluded. It also enables the use of sample tubes with varying lengths and diameters.
NOTE: In no occasion is the flow reversed towards the patient.
Zero
valve
MiniC
Diff. Abs.
Pressure Pressure
Sensor Sensor
Diff.
Pressure
Sensor
minic_fem_gastubing.vsd
Pump
Gas out
8-4
8-5
13 1
1 I RESET RS485
25 14
2 I -15 VDC (not used)
3 I +15 VDIRTY
4 I +15VDC (not used)
5 I/O -DATA RS485
6 I/O DATA RS485
7 Ground and Shield
8 I -RESET RS485
9 n/c
10 n/c
11 n/c
12 n/c
13 Ground and Shield
14 I +24/+32 VDIRTY depends on power supply (not
used)
15 I Ground DIRTY
16 n/c
17 n/c
18 n/c
19 n/c
20 I GASFR (not used)
21 I CTSD (not used)
22 I TXDD (not used)
23 O RXDD (not used)
24 I +5 VDC (not used)
25 I +5 VDC DIRTY (not used)
8-6
Planned maintenance
Service personnel shall perform the following checkout procedure completely every 12 months
after installation:
1. "Replacement of planned maintenance parts"
2. "Visual inspections"
3. "Functional checkout"
The instructions include a check form ("APPENDIX E") which should be filled in when performing
the procedures.
The symbol in the instructions means that the check form should be signed after
performing the procedure.
8-7
It is also recommended to replace the Mini D-fend water trap and the gas sampling line as part
of the planned maintenance procedure.
NOTE: See the supplies and accessories document delivered with the manual for compatible
accessories.
Nafion tube
Zero line air filter
2. Replace the special tube (Nafion) and check the condition of the internal tubing.
• Replace the 300 mm nafion tube in the sample gas line between the Mini D-fend
water trap and the zero valve unit.
• Check that the tubing inside the module is not contaminated. Any contamination
inside the tubing may indicate that the valve or sensor is contaminated too. This can
increase a risk of faulty operation in valve or sensor. The gas sensor is not possible
to clean in the field. Therefore, replace the whole miniC assembly with a new one.
NOTE: The nafion tube do not include the silicon fittings they connect to. Use the original silicon
fittings unless they are not damaged or leaking.
3. Replace the Mini D-fend O-rings:
• Detach the Mini D-fend.
• Detach the old rubber O-rings that are around the metal Mini D-fend connectors e.g.
using a small flat blade screwdriver. Pay special attention not to scratch the metal
Mini D-fend connectors and thus causing leaking.
• Set the new rubber O-rings into place and attach a new Mini D-fend.
8-8
2.3 Functional checkout
2.3.1 Test setup
Required tools
− A barometer
− A mass flowmeter for measuring air flow, minimum measurement range 100-300 ml/min,
accuracy 5% or better in the 100-300ml/min range.
− P/N: 755534-HEL Calibration Gas Regulator
− P/N M1006864 Calibration Gas Regulator, (US only)
− P/N: 755580 Calibration gas 5% CO2 and air, package of 4 cans
− P/N 755587 QUICK CAL calibration gas, (US only)
− 3 m / 10 ft gas sampling line
− A pressure manometer with either an integrated or a separate pressure pump
NOTE: See the supplies and accessories document delivered with the manual for compatible
accessories.
Connections
Disconnect the module from the monitor, if connected.
Monitor configuration
1. Configure the CO2 waveform field to the monitor screen.
2. check the CO2 unit: Airway Gas > CO2 Setup > Unit, make sure the unit is mmHg.
2.3.2 Procedure
1. Gas sampling system leak test
Check the gas sampling system for possible leakages.
• Disconnect the module from the monitor.
• Connect a new Mini D-fend water trap to the module.
8-9
• Connect a new gas sampling line to the sampling line connector in the water trap.
• Connect the other end of the gas sampling line to a pressure manometer and a
pressure pump.
• Block the "Sample Gas Out" connector.
• Pump 100 mmHg ± 20 mmHg pressure to the gas sampling system. Let the pressure
stabilize for approximately 10 seconds.
• Check that the pressure reading does not drop more than 6 mmHg during 1 minute.
NOTE: The gas module shall be disconnected from the monitor during the leak test.
2. Sample flow check
Check the sample flow rate.
NOTE: Gas measurement is not available during the first 1 minute after the module is
connected due to warm-up. Message 'Calibrating gas sensor' is shown in the waveform field.
Wait until warm-up is completed before proceeding with the next steps.
• Connect the module with the gas sampling line to the monitor.
• Connect the gas sampling line to the sampling line connector in the water trap.
• Connect the other end of the gas sampling line to a flowmeter.
• Check the sample flow rate reading from the flowmeter. The flow rate shall be within
the specification limit 150 ± 25 ml/min.
NOTE: Readjustment is needed, if the measured value is not within the specification limits.
Adjust the sample gas flow rate according to the instructions in section “3.1. Sample flow rate
adjustment”.
NOTE: If the sampling pump is noisy it indicates possible problems with motor bearing. In this
case it is recommended to replace the noisy sampling pump with a new one.
3. Zero valve operation
Test the zero valve functionality.
• Connect the gas regulator to the calibration gas container.
• Connect the end of the gas sampling line to the regulator on the gas container.
Leave the regulator overflow port open to room air.
• Enter to gas service menu: select Monitor Setup > Install/Service (16-4-34) >
Service (26-23-8) > Parameters > Gas Unit > Gases.
• Start feeding calibration gas. Wait until the CO2 value rises to approximately 5%.
• Open the zero valve to room air by selecting Zero valve ctrl.
• Check that the CO2 values drop back near 0%.
• Stop feeding the calibration gas.
8-10
• Turn the zero valve back to the normal measurement position by selecting ZERO to
MEAS.
4. Gas calibration
Perform gas calibration according to the instructions in section “3.2. Gas calibration”.
5. Ambient pressure
Use a barometer to check the operation of the absolute pressure sensor.
• Check that the ambient pressure value shown in the Gas service menu does not
differ more than ± 13.33 mbar from the value shown by the barometer.
6. Occlusion detection
• Block the tip of the sampling line by your finger.
• Check that message 'Sample line blocked' appears to the parameter window within
30 seconds.
7. Air leak detection
• Detach the Mini D-fend water trap.
• Check that message 'Check D-Fend' appears to the parameter window within 30
seconds.
8. Airway gases
• Breathe a minimum of 5 times to the tip of the sampling line.
• Check that a normal CO2 waveform appears to the waveform field and the EtCO2
and FiCO2 values are updated to the parameter window.
9. Apnea detection
• Stop breathing to the gas sampling line.
• Check that an 'Apnea' alarm appears to the message field within 30 seconds.
8-11
8-12
Connections
• Ensure that the module is connected to the monitor.
• Ensure that you have a new Mini D-fend water trap in use.
• Connect a new gas sampling line to the sampling line connector in the water trap.
• Connect the other end of the gas sampling line to the flow meter.
NOTE: Before checking or adjusting the sample flow rate, make sure there is no leakage in the
sampling system.
8-13
8-14
3.2.2 Procedure
NOTE: Gas calibration is not available during the first 5 minutes after the module is connected.
For maximum accuracy, let the monitor to warm up for 30 minutes before starting calibration.
NOTE: Gas calibration is not available during a 'Sample line blocked', 'Check D-Fend' and
'Check sample gas out’ alarm condition. Resolve the alarm condition before starting
calibration.
1. Enter to gas calibration menu: Airway Gas > Gas Calibration
2. The monitor will start automatic zeroing of the gas sensors. Wait until the message
'Zeroing' is replaced by a message 'Zero OK' message.
3. Open the regulator after a message 'Feed gas' is shown. The measured gas
concentrations are shown in real-time in the gas calibration menu.
Continue feeding the calibration gas until the measured gas concentrations are stabilized
and a message 'Adjust' is shown. Close the regulator.
4. Use the Trim Knob to adjust the gas readings shown in the Calibration menu to match
with the gas readings in the labelling of the calibration gas container. Press Trim Knob to
accept the adjusted values when the gas readings match each other.
5. Wait until a message 'Ok' is shown. NOTE: A message 'Zero error' is shown in case the
zeroing fails.
NOTE: A message 'Calibr.error' is shown, if you do not start feeding gas within 1 minute after
the automatic zeroing is completed, or if the calibration fails due to too large gain adjustment.
NOTE: If zeroing or calibration failed, select the Recalibrate to restart the calibration procedure
from the beginning.
8-15
4 Troubleshooting
The problems and solutions in this chapter represent only a few of the faults that you may
encounter and are not intended to cover every possible problem that may occur.
This chapter focuses on troubleshooting technical problems. For clinical and operation issues
refer to the "User’s Guide" for troubleshooting.
NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout”
each time after you have opened the module casing.
If the problem remains, call technical support for service. To ensure accurate problem solving,
please be prepared to provide the following information:
• Problem description and the troubleshooting done so far.
• Device information.
• Error messages displayed, if any.
• Other information, as requested.
8-16
8-17
ETCO2 over scale >20% • abnormally high ETCO2 (permissive • let the module run without a
Shown until 32%, hypercapnia) sampling line until the CO2
specified range 0...20% • CO2 sensor contaminated sensor has dried out
• D-fend malfunction
• change D-fend
ETCO2>PaCO2 • unit is mmHg or kPa and ETCO2 is close to • change to “wet gas” by using
arterial PCO2 install/service menu
• “dry gas” as default
Abnormally high (or low) • Pressure transducer failure. • Replace miniCO2 assembly.
response to CO2, or
sudden occlusion
message.
Unable to perform gas • Gas calibration is not available during the • Wait until warm-up is
calibration. first 5 minute after the module is completed.
connected.
• Gas calibration is not available during • Resolve the alarm condition
‘Sample line blocked’, ‘Check D-fend’, and before starting calibration.
‘Check sample gas out’ alarm condition.
8-18
8-19
- pozidrive screwdrivers
- flat blade screwdriver
- pincers
- antistatic wristband
8-20
8-21
6 Service parts
6.1 Ordering parts
To order parts, Contact GE Healthcare. Contact information is available at
www.gehealthcare.com. Make sure you have all necessary information at hand.
NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout”
after you have disassembled and reassembled the module.
8-22
8-23
8-24
1 Product overview
1.1 Introduction
This document provides information for the maintenance and service of the E-sCO, E-sCAiO,
N-CAiO modules. These modules are single width plug-in modules.
The modules provide airway measurements. Letters in the module name stand for:
C = CO2 and N2O, O = patient O2, A = anesthetic agents, and i = agent identification
Table 1 Options for E-sCO, E-sCAiO and N-CAiO modules
9-1
9-2
1.2.2 O2 measurement
The differential oxygen measuring unit uses the paramagnetic principle in a pneumatic bridge
configuration. The signal picked up with a differential pressure transducer unit is generated in
a measuring cell with a strong magnetic field that is switched on and off at a main frequency of
164 Hz. The output signal is a DC voltage proportional to the O2 concentration difference
between the gas to be measured and the air reference.
9-3
1
2
9-4
9-5
Figure 8 Absorber
Nafion tubes 1
The Nafion tube between the water trap and the zero valve equalizes the humidity of the
sampled gas to ambient level. This will prevent calibration errors caused by the difference in
humilities in the sampled breathing gas and the totally dry calibration gas.
Another Nafion tube is used between the CO2 absorber and the zero valve to prevent
condensation of water generated in the CO2 absorber as by-product of CO2-absorption.
Gas sensors
After the zero valve, the gas flows trough the MiniTPX sensor that measures the concentrations
of all gases but oxygen.
The oxygen concentration is measured in the MiniOM sensor that has two inputs. One input
draws in a part of the main flow and the other draws in room air as reference gas for the O2
measurement.
Pneumatics unit
The pneumatics unit contains the zero valve, the occlusion valve and the pneumatics block
with tubing connections.
9-6
The zero valve is activated during the zero level calibrations of gas sensors. The occlusion valve
is activated when the sampling line or water trap is occluded. With the activated occlusion
valve, the gas pump generates maximal suction trough the “side flow” connector of the water
trap, thus maximizing the transfer of liquids from the wet side of the water trap to the
container.
The pneumatics block contains a network of constrictions to divide the sampled gas in correct
proportions to different parts in the module. The first branching takes place in the water trap
where incoming flow is divided to the “main flow” and “side flow”. The second branching takes
place before the MiniOM sensor.
The pneumatics block also contains a pneumatic low pass filter between gas sensors and gas
pump. The filter consists of constrictions (resistors) and volumes (capacitors) and it attenuates
the pressure pulsation generated in the gas pump so that they do not disturb the operation of
the gas sensors.
Pressure measurements
The four pressure sensors on the CPU board are used to measure ambient pressure, working
pressure of the MiniTPX and MiniOM sensors and pressure of the reference gas flow to the
MiniOM sensor.
9-7
The self-diagnostics of the gas sampling system sends the following status data to the patient
monitor when specific triggering conditions are met: ‘Check D-Fend’, ‘Sample line blocked’,
‘Replace D-Fend’, ‘Check sample gas out’ and ‘Continuous blockage’.
The gas pump is stopped when the 'Sample line blocked' has lasted for more than 1 minute.
The module automatically restarts the pump to check whether the abnormal situation has
been resolved so that normal gas sampling operation is possible.
The gas pump repeats 1 minute full pump, 30 seconds pump off when the ‘Continuous
blockage’ message is shown.
9-8
Pneumatic System
The pneumatic system, together with the gas sampling system of the module creates the gas
flows and pressures needed for the oxygen measurement and protection of the microphones
from excessive pressure. About 30 ml/min flow of sampled gas comes to the In connector on
9-9
the MiniOM sensor. Room air is drawn to the Ref input of MiniOM also at 30 ml/min rate. About
75% of these flows are conducted to a pressure equalization chamber so that only about a 8
ml/min flow of the two gas streams continue into the air gap of the magnet. All the internal gas
flows finally get to a volume enclosed by the sensor board and the sensor body, and then flow
out through the Out connection of the sensor. Some of the gas channels and flow restrictors
are integrated into the preamplifier electronics board utilizing the multi-layer structure of the
LTCC (Low Temperature Co-fired Ceramics) circuit board technology.
NOTE: It is very important to prevent dust or liquids from getting into the pneumatic circuit of
MiniOM and thus, the gas connections should always be closed with a protecting cap when the
sensor is not connected to the module pneumatics.
Amplifier Board
The amplifier board located in the sensor has two electric microphones for the differential
detection of pressure pulses generated in the magnet's air gap. The microphone signals are fed
to two identical signal conditioning channels with a band-pass filter and a digitally controlled
amplifier. The voltage gains of the amplifiers are set during factory calibration so that the
responses of the microphone channels match in spite of differences in microphone's
sensitivities. The amplifier board also has an amplifier for the thermistor measuring the
temperature of the magnet.
MiniOM Board
The MiniOM board has five functions
• Drive the magnet coil.
• Convert the microphone and temperature signals into digital format.
• Filter digitally the microphone signals and perform the RMS-conversion.
• Communicate digitally with the module CPU.
• Store factory calibration data in permanent memory and communicate them to the
module CPU.
The module CPU provides the coil drive and communication enabling signals and also clock
signal for MiniOM board. The FPGA takes care of the coil drive and has also back-up clock in
case of CPU clock does not work. The FPGA takes care of the A/Dconversions which are
performed with a serial controlled SAR A/D-converter.
The digital band pass filtering and RMS conversion of the microphone signals is made with
FPGA circuit controlled by the VHDL code stored in the circuit. In order to filter out the
disturbances caused by acoustic noise, mechanical vibration and amplifier noise, the band
pass filters are designed to have as narrow a pass band as possible without slowing down the
filter's response to changes in the amplitude of the 164 Hz signal.
The FPGA circuit takes care of the digital communication between the miniOM sensor and the
module CPU.
The factory calibration coefficients of the sensor are stored in an EEPROM memory on the
miniOM board. When the module starts up, the calibration coefficients are read to the module
CPU and then used for calculating the O2 concentration from the Oxygen raw data received
from the sensor.
9-10
The CPU board also contains sensors for pressures, the temperature and humidity as well as
drivers for valves, the fan and the pump. The module is connected to the module bus through
an RS-485 serial channel.
+3.3 V
mOM +1.8 V
Connector DCDC
VCC_OM 11.5…18 V Module
Power
+7 V Bus
Supply
+5 V
mTPX Connector
Connector D25
ModBus_Data RS-485 RS-485
Transceiver
Analog to
Humidity and
Digital
Temperature
sensor
coversion
Zero Valve
Connector
4 Pressure
sensors CPU Atmel ETC Valve
Valve
AT91RM9200 Connector
Control
Memories Occlusion
SDRAM Valve
32 Mbyte Connector
Fan Fan
Crystal Control Connector
16 MHz
User Button
Crystal Connector
32.678 KHz
9-11
9-12
Planned maintenance
Service personnel shall perform the following checkout procedure completely every 12 months
after installation:
1. 2.1. Replacement of planned maintenance parts
2. 2.2. Visual inspections
3. 2.3. Functional check
The instructions include a check form ("APPENDIX F") which should be filled in when performing
the procedures.
The symbol in the instructions means that the check form should be signed after
performing the procedure.
9-13
It is also recommended to replace the D-fend Pro water trap, the gas sampling line as part of
the planned maintenance procedure.
NOTE: See the supplies and accessories document delivered with the manual for compatible
accessories.
2. Replace the special tubes (Nafion) and check the condition of the internal tubing.
• Check that the tubing inside the module is not contaminated. Any contamination
inside the tubing may indicate that the valves or sensors are contaminated, too. This
can increase a risk of faulty operation in valves or sensors. The valves or gas sensors
9-14
are not possible to clean in the field. Therefore, if you noticed any contamination in
the module tubing, send the module to GE Healthcare for factory service.
NOTE: The nafion tubes do not include the silicon fittings they connect to. Use the original
silicon fittings unless they are damaged or leaking.
3. Replace the OM reference gas filter assembly.
4. Check that the fan and ventilation hole are not covered in dust.
9-15
2.3.2 Procedure
Mark each task as complete on the checkout form.
1. Gas Sampling System Leak Test
NOTE: The gas module shall be disconnected from the monitor during the leak test.
Check the gas sampling system for possible leakages.
• Disconnect the module from the monitor.
• Detach the module front cover and casing.
• Block the OM reference tube with the forceps. Correct positioning of the forceps is
indicated by the figure below.
NOTE: Be careful when attaching the pincers to the tube and avoid stretching the
tube. Short pieces of silicone tubing on the forcep jaws can be used to protect the
tube from breaks that may appear when the tube is compressed between the jaws.
9-16
2. Sample Flow Rate Check
Check the sample flow rate. Connect the module to the monitor.
NOTE: Gas measurement is not available during the first 1 to 5 minutes after the module
is connected due to warming up. A message 'Calibrating Gas Sensor' is shown in the
waveform field. Wait until warm-up is completed before proceeding with the next steps.
NOTE: The ambient temperature and air pressure influence the flow rate measured by the
flow meter. A flow meter, which has been calibrated at 21.11 °C (70 °F) and 760 mmHg
(1033 cmH2O), measures the flow rate correctly under the same conditions, i.e. in room
temperature at sea level. A flow rate correction as instructed by the manufacturer of the
flow meter needs to be performed when measuring flow rate under other conditions, for
example in high altitude.
• Connect the gas sampling line (3 m / 10 ft with E-sCO, E-sCAiO and N-CAiO) to the
sampling line connector.
• Connect the other end of the gas sampling line to a flowmeter.
• Check the sample flow rate reading from the flowmeter. The flow rate shall be within
the specification limit 120 ± 20 ml/min.
NOTE: Readjustment is needed, if the measured value is not within the specification limit.
Adjust the sample gas flow rate according to the instructions in section “3.1.2. Sample
Flow Rate Adjustment”.
9-17
3. Reference Gas Flow Rate Check
Check the flow rate in reference gas inlet:
• Connect the module to the monitor.
• Leave the other end of the gas sampling line open to room air.
• Connect the flowmeter to the OM reference gas inlet on the side of the module with
a piece of tubing.
• Check that the Reference Flow is within the following range:
10 - 50 ml/min with E-sCO, E-sCAiO and N-CAiO modules
• Detach the water trap.
• Attach the front cover.
4. Fan
• Check that the gas module's fan is running behind the D-fend Pro water trap.
• Attach the water trap
5. Zero Valve Operation
Test the zero valve functionality:
• Connect the gas regulator to the calibration gas container.
• Connect the end of the gas sampling line to the regulator on the gas container.
Leave the regulator overflow port open to room air.
• Enter to gas service menu: select Monitor Setup > Install/Service (16-4-34) >
Service (26-23-8) > Parameters > Gas Unit > Gases.
• Start feeding the specified calibration gas. Wait until the gas values shown in the
Gas calibration menu rise approximately to the level indicated in the labelling of the
calibration gas container.
NOTE: The gas values in the gas service menu is in percentages (%).
• Open the zero valve to room air by selecting Zero valve ctrl.
• Check that the CO2, N2O and anesthesia agent values drop back near 0% and the
O2 reading near 21% (room air).
• Stop feeding the calibration gas.
• Turn the zero valve back to the normal measurement position by selecting ZERO to
MEAS. (measurement position).
6. Gas Calibration
Perform gas calibration according to the instructions in section “3.2. Gas Calibration”.
9-18
7. Agent Identification
NOTE: Perform this test only for E-sCAiO and N-CAiO modules.
Check agent ID unreliability:
• Feed the specified calibration gas for at least 30 seconds.
• Check that the anesthesia agent is identified as Desflurane and the ID unrel. value
(=agent ID unreliability) shown in the gas service menu is lower than 75.
If the value is higher, repeat the gas calibration and check the value again.
8. Ambient Pressure
Use a barometer to check the operation of the absolute pressure sensor.
• Check that the ambient pressure value shown in the gas service menu does not
differ more than ± 13.33 mbar from the value shown by the barometer.
NOTE: The ambient pressure value in the gas service menu is in mmHg.
9. Occlusion detection
• Block the tip of the sampling line by your finger.
• Check that a 'Sample line blocked' and ‘Low gas sample flow’ message appears on
the screen within 30 seconds.
10. Air Leak detection
• Detach the D-fend Pro water trap.
• Check that the message 'Check D-Fend' appears on the screen within 30 seconds.
• Attach the water trap.
11. Gas exhaust blockage
• Block the gas exhaust connector with your finger.
• Check that the message 'Check sample gas out' appears on the screen within 30
seconds.
12. Airway Gases
• Breathe a minimum of 5 times to the tip of the sampling line.
• Check that a normal CO2 waveform appears to the waveform field and the EtCO2
and FiCO2 values are updated on the screen.
9-19
13. Apnea detection
• Stop breathing to the gas sampling line.
• Check that an 'Apnea' alarm appears to the message field within 30 seconds.
2.3.3 Test completion
• Select Discharge patient to discard any changes made to the monitor configuration
during checkout.
• Disconnect and reconnect the module before starting a new case.
• Fill in all necessary documents.
9-20
9-21
NOTE: Adjust the flow rate according to the reading in the flow meter. The flow rate reading in
the Gases menu is measured by the internal electronics and settles always back to the nominal
120 ml /min independent on the real flow rate.
Connections
1. Ensure that the module is connected to the monitor.
2. Ensure that you have a new D-fend Pro water trap in use.
3. Connect the gas regulator to the calibration gas container.
4. Connect a new gas sampling line to the sampling line connector in the water trap.
5. Connect the other end of the gas sampling line to the regulator on the gas container.
Leave the regulator overflow port open to room air.
9-22
3.2.2 Procedure
NOTE: Gas calibration is not available during the first 5 minutes after the module is connected.
For maximum accuracy, let the monitor to warm up for 30 minutes before starting calibration.
NOTE: Gas calibration is not available during a 'Sample line blocked', 'Check D-Fend' and
'Check sample gas out’ alarm condition. Resolve the alarm condition before starting
calibration.
1. Enter to gas calibration menu: Airway Gas > Gas Calibration
2. The monitor will start automatic zeroing of the gas sensors. Wait until the message
'Zeroing' is replaced by a message 'Zero Ok' for all measured gases.
3. Open the regulator after a message 'Feed gas' is shown for all measured gases. The
measured gas concentrations are shown in real-time in the gas calibration menu.
Continue feeding the calibration gas until the measured gas concentrations are stabilized
and a message 'Adjust' is shown for all measured gases. Close the regulator.
4. Use the Trim Knob to adjust the gas readings shown in the Calibration menu to match
with the gas readings in the labelling of the calibration gas container. Press Trim Knob to
accept the adjusted values when the gas readings match each other.
5. Wait until a message 'Ok' is shown for all measured gases.
NOTE: A message 'Zero Error' is shown in case the zeroing fails.
NOTE: A message 'Calibration Error' is shown, if you do not start feeding gas within 1 minute
after the automatic zeroing is completed, or if the calibration fails due to too large gain
adjustment.
NOTE: If zeroing or calibration failed, select the Recalibrate to restart the calibration procedure
from the beginning.
9-23
4 Troubleshooting
The problems and solutions in this chapter represent only a few of the faults that you may
encounter and are not intended to cover every possible problem that may occur.
This chapter focuses on troubleshooting technical problems. Refer also to the troubleshooting
hints on the “User’s Guide” for troubleshooting monitoring problems, performance issues and
clinical configuration issues.
NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout”
each time after you have opened the module casing.
If the problem remains, call technical support for service. To ensure accurate problem solving,
please be prepared to provide the following information:
• Problem description and the troubleshooting done so far.
• Device information.
• Error messages displayed, if any.
• Other information, as requested.
If in doubt of having any loose parts or cable connections inside the module, detach the
module box by removing the four screws from the back of the module and check that:
• all screws are tightened properly
• all cables are connected properly
• tubes are not pinched and there are no sharp bends on them
• all tubes are connected properly
• there are no loose objects inside the module
9-24
9-25
9-26
4.4 Messages
4.4.1 Gas measurement
Check D-Fend Water trap is not connected. Connect the water trap and sampling
Check D-Fend and line to the module.
sample gas out. Wait for Air leak inside the internal tubing. Check the internal tubing for leakages.
30 sec and press Normal Replace, if needed.
screen to continue.
Sample line blocked Gas sampling line is blocked. Check the external gas sampling line for
Continuous blockage. blockages. Replace, if needed.
Check sampling line Water trap container is full. Empty the water trap container.
and D-Fend. Water trap is occluded. Replace the water trap.
Internal tubing is blocked. Check the internal tubing for blockages.
Replace, if needed.
Check sampling gas out The sample gas outflow is blocked. Check the sample gas out connector in
Check D-Fend and the front panel and the exhaust line for
gas return or scavenging for blockages.
sample gas out. Wait for
30 sec and press Normal - If the sample gas is returned to the
patient circuit, check that there is no
screen to continue.
occlusion in the tubing.
- If the sample gas outlet is connected
to a scavenging system, make sure
an open system is used where gas is
removed in room pressure.
Check the internal tubing for blockages.
Internal tubing is blocked.
Check sample pump operation by
Pump failure.
measuring the sample gas flow rate.
Replace pump, if needed.
Replace D-Fend Defective or contaminated D-fend Pro. Replace the D-fend Pro water trap.
Occlusion in internal tubing. Check sample and reference flows.
Perform a visual check for the internal
tubing. Remove the cause for occlusion.
Calibrating Gas calibration is in progress. Wait until the calibration is completed
successfully.
Failure in Agent ID Agent ID has failed. Perform gas calibration. Check agent ID
unreliability (see functional check). If it
does not help, send the module to GE
Healthcare for factory repair.
Zeroing Zeroing is in progress. Wait until zeroing is completed
successfully.
9-27
9-28
9-29
9-30
9-31
improperly grounded test equipment. The following guidelines may not guarantee a 100%
static-free workstation, but can greatly reduce the potential for failure of any electronic
assemblies being serviced:
• Discharge any static charge you may have built up before handling semiconductors or
assemblies containing semiconductors.
• A grounded, antistatic wristband or heel strap should be worn at all times while handling
or repairing assemblies containing semiconductors.
• Use properly grounded test equipment.
• Use a static-free work surface while handling or working on assemblies containing
semiconductors.
• Do not remove semiconductors or assemblies containing semiconductors from antistatic
containers until absolutely necessary.
• Do not slide semiconductors or electrical/electronic assemblies across any surface.
• Do not touch semiconductor leads unless absolutely necessary.
• Semiconductors and electronic assemblies should be stored only in antistatic bags or
boxes.
• Handle all PCB assemblies by their edges.
• Do not flex or twist a circuit board.
9-32
Module casing
9-33
1. Carefully remove the main flow nafion tube and every 4th
year the shorter zero line nafion tube.
NOTE: Remember the route of the tubes and reassemble
correctly.
NOTE: Make sure that the Nafion tubes are routed in such a way
that they don’t come near the fan, and there is no risk of the fan
being obstructed by the tubes. An obstructed fan will result in
degraded ventilation inside the module, and sensor inop
message being displayed.
NOTE: The nafion tubes do not include the silicon fittings they
are connected to. Use the original silicon fittings unless they are
not damaged or leaking.
9-34
9-35
9-36
9-37
4. Lift the OM unit with the holder from the metal brackets.
5. Detach the holder.
9-38
6 Service parts
6.1 Ordering parts
To order parts, Contact GE Healthcare. Contact information is available at
www.gehealthcare.com. Make sure you have all necessary information at hand.
NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout”
after you have disassembled and reassembled the module.
9-39
9-40
9-41
9-42
1 Product overview
1.1 Introduction
This document provides information for the maintenance and service of the Entropy Module,
E-ENTROPY. The Entropy module uses an electroencephalography (EEG) signal, together with
spontaneous facial muscular activity with a frontal electromyography (FEMG) signal to
measure:
• Response Entropy (RE)
• State Entropy (SE)
• Burst Suppression Ratio (BSR)
The Entropy module is responsible for EEG and FEMG signal acquisition, amplification, filtering
and digitization, and electrode impedance measurement. All the calculated parameters can be
selected on the display and trended.
10-1
1.2.1 EntrEEG
EEG is a differential voltage signal measured from electrodes attached to the patient’s skin.
EEG measures the spontaneous electrical activity of the brain. This electrical activity reflects
the state of the brain. In referential measurement, the referential electrode delivers its potential
to every channel’s minus-input. The signal is the potential difference between this common
reference electrode (electrode #3) and the electrode connected to the plus input (electrode #1).
The purpose of the ground electrode (electrode #2) is to reduce common mode noise.
The EntrEEG signal is amplified, antialias filtered, digitized and software filtered. After that, the
EntrEEG signal is shown on the screen and the RE, SE and BSR characteristics are calculated
from it.
1.2.2 FEMG
FEMG is an electrical signal originating from facial muscles. The FEMG signal has much broader
spectrum than EEG and it overlaps with EEG at low frequencies. The FEMG signal gives its
contribution to the RE values (see “RE and SE”). Mains power frequency and its harmonics are
digitally filtered away to reject interference noise from power lines.
1.2.3 RE and SE
Entropy numbers range from 100 to zero (RE 0-100, SE 0 – 91), correlating to the patient's
anesthetic state. High values of Entropy indicate high irregularity of the signal, signifying that
the patient is awake. There are two Entropy parameters: the fast-reacting Response Entropy
and the more steady and robust State Entropy. State Entropy consists of the entropy of EEG
signal calculated up to 32 Hz. Response Entropy includes additional high frequencies up to 47
Hz and, consequently, the fast frontalis EMG (FEMG) signals enable a fast response time.
Table 1 Frequency ranges for Entropy calculation
10-2
Figure 3 Front panel of Entropy Module, E-ENTROPY, and the back of the
module
10-3
A/D Converter
Filtering
Module Bus
&
10-4
Corrective maintenance
Service personnel shall perform the following checkout procedure after any corrective
maintenance, before taking the module back into clinical use:
Planned maintenance
Service personnel shall perform the following checkout procedure completely every 12 months
after installation:
1. Visual inspections
2. Electrical safety tests
3. Functional check
10-5
2.3 Functional check
Turn the monitor on. Wait until the normal monitoring screen appears.
2. Sensor checking
• Connect the Entropy simulator to the Entropy sensor cable.
• Check that a 'Checking sensor' message appears after a while and wait until all 3
electrodes are checked. The sensor check is passed, if a green circle with a check
mark changes to orange with result text (PASS/FAIL) for each tested electrode is
shown.
10-6
3. Functional check with a simulator
• Check that the EntrEEG waveform and RE & SE values appear on the monitor screen
approximately 30 seconds after the sensor check is completed.
2.3.3 Test completion
• Fill in all necessary documents, refer to "Service check form, E-Entropy" on page G-1
10-7
3 Troubleshooting
The problems and solutions in this chapter represent only a few of the faults that you may
encounter and are not intended to cover every possible problem that may occur.
This chapter focuses on troubleshooting technical problems. For clinical and operation issues
refer to the "User’s Guide" for troubleshooting.
NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout”
each time after you have opened the module casing.
If the problem remains, call technical support for service. To ensure accurate problem solving,
please be prepared to provide the following information:
• Problem description and the troubleshooting done so far.
• Device information.
• Error messages displayed, if any.
• Other information, as requested.
10-8
10-9
10-10
- antistatic wristband
- pincers
- torx screwdriver, T10
- flat blade screwdriver
10-11
5 Service parts
5.1 Ordering parts
To order parts, Contact GE Healthcare. Contact information is available at
www.gehealthcare.com. Make sure you have all necessary information at hand.
NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout”
after you have disassembled and reassembled the module.
10-12
10-13
10-14
A-1
1.5 Workflow
The following sections describe the sequential workflow for the software installation:
Sections 1.6 through 1.8 describe how to prepare the patient monitor(s) and the service PC for
the software installation.
Sections 1.9 through 1.11 describe the software transfer process using the GE Healthcare
Software Transfer Utility.
Sections 1.12 and 1.13 describe the software activation process and the functional check
procedure after activating the new software.
A-2
− Monitor Setup > Install/Service (16,4,34) > Service (26, 23, 8) > SW Management >
Country Settings > Languages > Language
4. Set the patient type to NEO
− Admit/Discharge > Admit Patient > Patient Type
5. Record the IBP1 settings in “1.15. User settings record”
− IBP > IBP1 ‘XXX’ Setup
6. Discharge the patient
− Admit/Discharge > Discharge
NOTE: Language configuration and in Neonatal mode IBP1 settings will be lost after installation
of new monitor software.
7. Configure the patient monitor(s) for the software installation. Follow the appropriate
procedure below.
NOTE: When the patient monitor is in the download mode, the patient monitor will stop
communicating with central station(s) and all the parameter modules will be powered off.
NOTE: If replace the CPU board, which already has the run-in software baseline. The password
should be Install/Service (20), Service (20), skip step 3 to 6 of above process, other operations
are the same as normal upgrade.
A-3
A-4
A-5
c. Select File -> Save. Specify the destination and file name of this text file (.txt) so you can
easily navigate to it.
d. From the GE Healthcare Software Transfer Utility window, select File ->Import IP File.
Navigate to the .txt file containing the IP addresses. The IP addresses display in the Add
Target IP Addresses list.
3. Verify these IP address(es) are the destination(s) where you want to transfer the software.
NOTE: To remove an incorrect IP address from the displayed list, select the IP address
and click the up arrow button.
A-6
4. Verify the transfer status of all selected software packages indicates Success in PC.
NOTE: If the transfer status show the Failed information, please do not activate the
software. You should redo the software download.
5. Select Exit.
Software Transfer to destination monitor(s) is now completed. The next step is to activate the
transferred software.
NOTE: Please do not restart the monitor manually. Activate the software first. The monitor will
automatically restart by activation.
A-7
A-8
B-1
<5% UT <5% UT
(>95% dip in UT) (>95% dip in UT)
for 5 sec for 5 sec
Power frequency 3 A/m 3 A/m Power frequency magnetic field
(50/60 Hz) magnetic should be at levels characteristic of a
field typical location in a typical
IEC 61000-4-8 commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
B-2
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicated theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the equipment is used exceeds the applicable RF compliance level above, the
equipment should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the equipment.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
B-3
B-4
Monitor license
MAC Address License key
Modules
Gas module: Entropy module:
C-1
Notes
3.2.1. Start-up
3.2.2. Display
3.2.5. Recorder
3.2.7. Conclusion
3.2.7. Conclusion
Notes
Signature
C-2
Monitor Installation
IBP: SpO2 module: TEMP: Network:
Gas module: Entropy module:
Notes
D-1
< 500 µA
Normal Condition (NC)
< 300 µA
< 500 µA
Single Fault Condition (SFC)
< 300 µA
Notes
4.2.1. Start-up
4.2.2. Display
4.2.3. Keyboard(s)
Notes
. ECG measurement
2. Pacemaker Detection
3. Asystole Detection
D-2
Notes
. Respiration measurement
5. Respiration Rate
6. Apnea Detection
Notes
. Temperature measurement
7. Temperature detection
Notes
8. Zeroing
9. Static Pressure
Notes
. SpO2 measurement
Notes
Notes
4.2.6. Loudspeaker
4.2.9. Network
D-3
Notes
Notes
Notes
Signature
D-4
Prior to testing verify all equipment is calibrated via “Cal” labeling and record Cal Due Dates
Notes
4. Gas calibration
5. Ambient pressure
6. Occlusion detection
8. Airway gases
E-1
Document no. 2069392-002
B40i Patient Monitor
Notes
Signature
E-2
Prior to testing verify all equipment is calibrated via “Cal” labeling and record Cal Due Dates
Notes
4. Fan
6. Gas Calibration
7. Agent Identification
8. Ambient Pressure
F-1
Document no. 2069392-002
B40i Patient Monitor
9. Occlusion detection
Notes
Signature
F-2
Prior to testing verify all equipment is calibrated via “Cal” labeling and record Cal Due Dates
Notes
Notes
G-1
Document no. 2069392-002
B40i Patient Monitor
2. Sensor checking
Notes
Signature
G-2
H-1
Document no. 2069392-002
B40i Patient Monitor
H-2
Licensed/optional/required Required
Communication Partner Device/IP Printer
address/network
Middle layer protocols DRI
Application layer protocol and encoding PCL
Ports NA
Traffic characterization and bandwidth Max: 47 Kbps, Avg: 27 Kbps
requirements
H-3
Document no. 2069392-002
B40i Patient Monitor
In addition to the hazardous situations identified above, connection of the B40i monitor to a
network that includes other equipment could result in other unidentified risks to patients,
operators or third parties. The responsible organization should identify, analyze, evaluate and
control these risks on an ongoing basis including after changes to the network, which could
introduce new risks and require additional analysis.
H-4
0537