MT812E - Cell Master - Anritsu
MT812E - Cell Master - Anritsu
MT812E - Cell Master - Anritsu
Cell Master™
Compact Handheld
Base Station Analyzer
MT8212E MT8213E
2 MHz to 4 GHz 2 MHz to 6 GHz Cable & Antenna Analyzer
100 kHz to 4 GHz 100 kHz to 6 GHz Spectrum Analyzer
10 MHz to 4 GHz 10 MHz to 6 GHz Power Meter
Introduction
Anritsu introduces its latest generation compact handheld Base Station Analyzer for installation and maintenance of wireless networks.
Designed as a lightweight base station analyzer meeting virtually all the testing needs of an RF technician. The Cell Master features
Signal Analyzer options for 2G, 3G, and 4G cellular networks including LTE, and WiMAX, and for digital broadcast.
Cable and Antenna Analyzer Highlights Spectrum and Interference Analyzer Highlights
• Measurements: RL, VSWR, Cable Loss, DTF, Phase • Measurements: Occupied Bandwidth, Channel Power, ACPR, C/I
• 2-port Transmission Measurement: High/Low Power • Interference Analyzer: Spectrogram, Signal Strength, RSSI,
• Sweep Speed: 1 msec/data point, typical Interference Mapping
• Display: Single or Dual Measurement Touchscreen • Dynamic Range: > 95 dB in 10 Hz RBW
• Calibration: OSL, InstaCal™, and FlexCal™ • DANL: –152 dBm in 10 Hz RBW
• Bias Tee: 32 V internal • Phase Noise: –100 dBc/Hz max @ 10 kHz offset at 1 GHz
• Frequency Accuracy: ± 50 ppb with GPS On
Capabilities and Functional Highlights
• LTE (10 MHz B/W) • Fixed, Mobile WiMAX • Built-in Bias Tee • E1, T1, T3 Backhaul Analyzer
• GSM/EDGE • ISDB-T, ISDB-T SFN • Internal Power Meter • 3 hour battery operation time
• W-CDMA/HSDPA • PIM Analyzer • High Accuracy Power Meter • USB Data Transfer
• TD-SCDMA/HSDPA • Interference Analyzer • USB Power Sensors, 4 to 26 GHz • Master Software Tools
• CDMA, EV-DO • GPS tagging of stored traces • Coverage Mapping • Line Sweep Tools
Cell Master™ MT8212E/MT8213E Base Station Analyzer featuring 8.4 in. Daylight Viewable Touchscreen
Compact Size: 273 mm x 199 mm x 91 mm, (10.7 in x 7.8 in x 3.6 in), Lightweight: 3.71 kg, (8.2 lbs)
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Cable and Antenna Analyzer
Measurements
Measurements VSWR
Return Loss
Cable Loss
Distance-to-Fault (DTF) Return Loss
Distance-to-Fault (DTF) VSWR
1-Port Phase
Smith Chart
Setup Parameters
Measurement Display Single/Dual Measurement Display with independent markers
Frequency Start/Stop, Signal Standard, Start Cal
DTF Start/Stop, DTF Aid, Units (m/ft), Cable Loss, Propagation Velocity, Cable, Windowing
Windowing Rectangular, Normal Side Lobe, Low Side Lobe, Minimum Side Lobe
Amplitude Top, Bottom Auto Scale, Full Scale
Sweep Run/Hold, Single/Continuous, RF Immunity (High/Low), Data Points, Averaging/
Smoothing,Output Power (High/Low)
Data Points 137, 275, 551, 1102, 2204
Markers Markers 1-6 (On/Off), Delta Makers 1-6 (On/Off), Marker to Peak/Valley, Marker Table
Traces Recall, Copy to Display Memory, No Trace Math, Trace ± Memory, Trace Overlay
Limit Line On/Off, Single Limit, Multi-segment (41), Limit Alarm, Clear
Calibration Start Cal, Cal Type (Standard/FlexCal™)
Save/Recall Setups, Measurements, Screen Shots Jpeg (save only)
Frequency
Frequency Range 2 MHz to 4 GHz (MT8212E), 2 MHz to 6 GHz (MT8213E)
Frequency Accuracy ≤ ± 2.5 ppm @ 25 °C
Frequency Resolution 1 kHz, (RF immunity low)
100 kHz, (RF immunity high)
Output Power
High 0 dBm, typical
Low –30 dBm, typical
Interference Immunity
On-Channel +17 dBm @ > 1.0 MHz from carrier frequency
On-Frequency 0 dBm within ± 10 kHz of the carrier frequency
Measurement Speed
Return Loss ≤ 1.00 msec/data point, RF immunity low, typical
Distance-to-Fault ≤ 1.25 msec/data point, RF immunity low, typical
Return Loss
Measurement Range 0 dB to 60 dB
Resolution 0.01 dB
VSWR
Measurement Range 1:1 to 65:1
Resolution 0.01
Cable Loss
Measurement Range 0 dB to 30 dB
Resolution 0.01 dB
Distance-to-Fault
Vertical Range Return Loss 0 dB to 60 dB
Vertical Range VSWR 1:1 to 65:1
Fault Resolution (meters) (1.5 x 108 x vp) / ∆F (vp = velocity propagation constant, ∆F is F2-F1 in Hz)
Horizontal Range (meters) 0 to (Data Points-1) x Fault Resolution, to a maximum of 1500 meters (4921 ft)
1-Port Phase
Measurement Range –180° to +180°
Resolution 0.01°
Smith Chart
Resolution 0.01
Page 2 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Cable and Antenna Analyzer (continued)
Measurement Accuracy
Corrected Directivity > 42 dB, OSL Calibration
> 38 dB, InstaCal™ Calibration
Measurement Uncertainty
Event Resolution 10.2 cm (0.335 ft) maximum, or 150 / (n*ΔF), ΔF in MHz, n is IOR
Horizontal Range 1020 meter (3345 ft) maximum, or (#dp-1) *Event Resolution
RF Connector N(m)
Page 3 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Spectrum Analyzer
Measurements
Smart Measurements Field Strength (uses antenna calibration tables to measure dBm/m2 or dBmV/m)
Occupied Bandwidth (measures 99% to 1% power channel of a signal)
Channel Power (measures the total power in a specified bandwidth)
ACPR (adjacent channel power ratio)
AM/FM/SSB Demodulation (wide/narrow FM, USB and LSB), (audio out only)
C/I (carrier-to-interference ratio)
Coverage Mapping (requires Option 0431)
Setup Parameters
Frequency Center/Start/Stop, Span, Frequency Step, Signal Standard, Channel #
Amplitude Reference Level (RL), Scale, Attenuation Auto/Level, RL Offset, Pre-Amp On/Off, Detection
Span Span, Span Up/Down (1-2-5), Full Span, Zero Span, Last Span
Bandwidth RBW, Auto RBW, VBW, Auto VBW, RBW/VBW, Span/RBW
File Save, Recall, Delete, Directory Management
Save/Recall Setups, Measurements, Limit Lines, Screen Shots Jpeg (save only), Save-on-Event
Save-on-Event Crossing Limit Line, Sweep Complete, Save-then-Stop, Clear All
Delete Selected File, All Measurements, All Mode Files, All Content
Directory Management Sort Method (Name/Type/Date), Ascend/Descend, Internal/USB, Copy, Format USB
Application Options Bias-Tee (On/Off), Impedance (50 Ω, 75 Ω, Other)
Sweep Functions
Sweep Single/Continuous, Sweep Mode (Fast, Performance, No FFT), Reset, Detection,
Minimum Sweep Time, Trigger Type, Gated Sweep (see Option 0090)
Detection Peak, RMS, Negative, Sample, Quasi-peak
Triggers Free Run, External, Video, Change Position, Manual
Trace Functions
Traces Up to three Traces (A, B, C), View/Blank, Write/Hold, Trace A/B/C Operations
Trace A Operations Normal, Max Hold, Min Hold, Average, # of Averages, (always the live trace)
Trace B Operations A B, B C, Max Hold, Min Hold
Trace C Operations A C, B C, Max Hold, Min Hold, A – B C, B – A C, Relative Reference (dB), Scale
Marker Functions
Markers Markers 1-6 each with a Delta Marker, or Marker 1 Reference with Six Delta Markers,
Marker Table (On/Off), All Markers Off
Marker Types Style (Fixed/Tracking), Noise Marker, Frequency Counter Marker
Marker Auto-Position Peak Search, Next Peak (Right/Left), Peak Threshold %, Set Marker to Channel,
Marker Frequency to Center, Delta Marker to Span, Marker to Reference Level
Marker Table 1-6 markers frequency and amplitude plus delta markers frequency amplitude and offset
Limit Line Functions
Limit Lines Upper/Lower, On/Off, Edit, Move, Envelope, Advanced, Limit Alarm, Default Limit
Limit Line Edit Frequency, Amplitude, Add Point, Add Vertical, Delete Point, Next Point Left/Right
Limit Line Move To Current Center Frequency, By dB or Hz, To Marker 1, Offset from Marker 1
Limit Line Envelope Create Envelope, Update Amplitude, Points (41 max), Offset, Shape Square/Slope
Limit Line Advanced Type (Absolute/Relative), Mirror, Save/Recall
Frequency
Frequency Range 100 kHz to 4 GHz (MT8212E), 100 kHz to 6 GHz (MT8213E)
Tuning Resolution 1 Hz
Frequency Reference Aging: ± 1.0 ppm/year
Accuracy: ± 1.5 ppm (25 °C ± 25 °C) + aging, ± 50 ppb with GPS On
Frequency Span 10 Hz to 4 GHz including zero span (MT8212E),
10 Hz to 6 GHz including zero span (MT8213E)
Sweep Time Minimum 100 ms, 10 µs to 600 seconds in zero span
Sweep Time Accuracy ± 2% in zero span
Bandwidth
Resolution Bandwidth (RBW) 10 Hz to 3 MHz in 1–3 sequence ± 10% (1 MHz max in zero-span) (–3 dB bandwidth)
Video Bandwidth (VBW) 1 Hz to 3 MHz in 1–3 sequence (–3 dB bandwidth)
RBW with Quasi-Peak Detection 200 Hz, 9 KHz, 120 kHz (–6 dB bandwidth)
VBW with Quasi-Peak Detection Auto VBW is On, RBW/VBW = 1
Page 4 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Spectrum Analyzer (continued)
Spectral Purity
SSB Phase Noise @ 1 GHz –100 dBc/Hz, –110 dBc/Hz typical @ 10 kHz offset
–105 dBc/Hz, –112 dBc/Hz typical @ 100 kHz offset
–115 dBc/Hz, –121 dBc/Hz typical @ 1 MHz offset
Amplitude Ranges
Dynamic Range > 95 dB (2.4 GHz), 2/3 (TOI-DANL) in 10 Hz RBW
Measurement Range DANL to +26 dBm
Display Range 1 dB to 15 dB/div in 1 dB steps, ten divisions displayed
Reference Level Range –120 dBm to +30 dBm
Maximum Continuous Input +35 dBm
Attenuator Resolution 0 dB to 55 dB in 5 dB steps
Amplitude Units Log Scale Modes: dBm, dBV, dBmv, dBμV
Linear Scale Modes: nV, μV, mV, V, kV, nW, μW, mW, W, kW
Amplitude Accuracy
100 kHz to 4.0 GHz ± 1.25 dB, ± 0.5 dB typical
> 4.0 GHz to 6 GHz ± 1.50 dB, ± 0.5 dB typical
Displayed Average Noise Level (DANL)
Preamp Off Preamp On
(Reference level -20 dBm) (Reference level -50 dBm)
(RBW Normalized to 1 Hz, 0 dB attenuation) Maximum Typical Maximum Typical
10 MHz to 2.4 GHz –141 dBm –146 dBm –157 dBm –162 dBm
> 2.4 GHz to 4 GHz –136 dBm –141 dBm –154 dBm -159 dBm
> 4 GHz to 5 GHz –133 dBm –138 dBm –154 dBm –155 dBm
> 5 GHz to 6 GHz –125 dBm –131 dBm –146 dBm –150 dBm
(RBW = 10 Hz, 0 dB attenuation)
10 MHz to 2.4 GHz –131 dBm –136 dBm –147 dBm –152 dBm
> 2.4 GHz to 4 GHz –126 dBm –131 dBm –144 dBm –149 dBm
> 4 GHz to 5 GHz –123 dBm –128 dBm –144 dBm –145 dBm
> 5 GHz to 6 GHz –115 dBm –121 dBm –136 dBm –140 dBm
Spurs
Residual Spurious < –90 dBm (RF input terminated, 0 dB input attenuation, > 10 MHz)
Input-Related Spurious < –75 dBc (0 dB attenuation, –30 dBm input, span < 1.7 GHz, carrier offset > 4.5 MHz)
Exceptions, typical < –70 dBc @ < 2.5 GHz, with 2072.5 MHz Input
< –68 dBc @ F1-280 MHz with F1 Input
< –70 dBc @ F1 + 190 MHz with F1 Input
< –52 dBc @ 7349-2F2 MHz, with F2 Input, where F2 < 2424.5 MHz
< –55 dBc @ 190.5 ± F1/2 MHz, with F1 Input, where F1 < 1 GHz
Third-Order Intercept (TOI)
Preamp Off (–20 dBm tones 100 kHz apart, 10 dB attenuation)
800 MHz +16 dBm
2400 MHz +20 dBm
200 MHz to 2200 MHz +25 dBm, typical
> 2.2 GHz to 5.0 GHz +28 dBm, typical
> 5.0 GHz to 6.0 GHz +33 dBm, typical
Second Harmonic Distortion
Preamp Off, 0 dB input attenuation, –30 dBm input
50 MHz –56 dBc
> 50 MHz to 200 MHz –60 dBc, typical
> 200 MHz to 3000 MHz –70 dBc, typical
VSWR
2:1, typical
Page 5 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Coverage Mapping (Options 0431)
Measurements
Indoor Mapping Outdoor Mapping
RSSI RSSI
ACPR ACPR
Setup Parameters
Frequency Center/Start/Stop, Span, Freq Step, Signal Standard, Channel #, Channel Increment
Amplitude Reference Level (RL), Scale, Attenuation Auto/Level, RL Offset, Pre-Amp On/Off, Detection
Span Span, Span Up/Down (1-2-5), Full Span, Zero Span, Last Span
BW RBW, Auto RBW, VBW, Auto VBW, RBW/VBW, Span/VBW
Measurement Setup ACPR, RSSI
Point Distance / Time Setup Repeat Type Time Distance
Save Points Map Save KML, JPEG
Recall Points Map Recall Map, Recall KML Points only, Recall KML Points with Map, Recall Default Grid
GPS Receiver Option (Option 0031) (Antenna sold separately, P/N 2000-1528-R)
Setup On/Off, Antenna Voltage 3.3/5.0 V, GPS Info
GPS Time/Location Indicator Time, Latitude, Longitude and Altitude on display
Time, Latitude, Longitude and Altitude with trace storage
High Frequency Accuracy Spectrum Analyzer, Interference Analyzer, CW Signal Analyzers
when GPS Antenna is connected < ± 50 ppb with GPS On, 3 minutes after satellite lock in selected mode
Connector SMA, Female
Page 6 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Channel Scanner (Option 0027)
Number of Channels 1 to 20 Channels
Measurements Graph/Table, Max Hold (On/5 sec/Off), Freq/Channel, Current/Max, Single/Dual Color
Scanner Scan Channels, Scan Frequencies, Scan Customer List, Scan Script Master™
Amplitude Reference Level, Scale
Custom Scan Signal Standard, Channel, # of Channels, Channel Step Size, Custom Scan
Frequency Range 100 kHz to 4 GHz (MT8212E),
100 kHz to 6 GHz (MT8213E)
Frequency Accuracy ± 10 Hz + Time base error
Measurement Range –110 dBm to +26 dBm
Application Options Bias-Tee (On/Off), Impedance (50 Ω, 75 Ω, Other)
CW Signal Generator Option (Option 0028) (Requires CW Signal Generator Kit, P/N 69793)
Setup Parameters
Frequency Frequency, Signal Standard, Channel Number, Display Setup Help
Amplitude Power Level (Low/High), Offset (dB)
Frequency Range 2 MHz to 2 GHz
Frequency Reference Accuracy: ± 1.5 ppm (25 °C ± 25 °C) + aging, < ± 50 ppb with GPS On
Output Power High 0 dBm typical, Low –30 dBm typical
Attenuator (included in kit 69793): 0 to 90 dB in 1 dB steps
Page 7 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Power Meter
Frequency Center/Start/Stop, Span, Frequency Step, Signal Standard, Channel #, Full Band
Amplitude Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale
Average Acquisition Fast/Med/Slow, # of Running Averages
Limits Limit On/Off, Limit Upper/Lower
Frequency Range 10 MHz to 4 GHz (MT8212E), 10 MHz to 6 GHz (MT8213E)
Span 1 kHz to 100 MHz
Display Range –140 dBm to +30 dBm, ≤ 40 dB span
Measurement Range –120 dBm to +26 dBm
Offset Range 0 dB to +100 dB
VSWR 2:1 typical
Maximum Continuous Input +35 dBm without attenuator
Accuracy Same as Spectrum Analyzer
Application Options Impedance (50 Ω, 75 Ω, Other)
High Accuracy Power Meter (Option 0019) (Requires external USB Power Sensor(s))
Amplitude Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale
Average # of Running Averages, Max Hold
Zero/Cal Zero On/Off, Cal Factor (Center Frequency, Signal Standard)
Limits Limit On/Off, Limit Upper/Lower
Power Sensor Model PSN50 MA24104A MA24106A MA24108/18/26A
Description High Accuracy Inline High High Accuracy Microwave
RF Power Sensor Power Sensor RF Power Sensor USB Power Sensor
Frequency Range 50 MHz to 6 GHz 600 MHz to 4 GHz 50 MHz to 6 GHz 10 MHz to 8 GHz
(MA24108A)
10 MHz to 18 GHz
(MA24118A)
10 MHz to 26 GHz
(MA24126A)
Dynamic Range –30 dBm to +20 dBm +3 dBm to +51.76 dBm –40 dBm to +23 dBm –40 dBm to +20 dBm
(0.001 mW to 100 mW) (2 mW to 150 W) (0.1 µW to 200 mW) (0.1 µW to 100 mW)
VBW 100 Hz 100 Hz 100 Hz 50 kHz
Measurand True-RMS True-RMS True-RMS True-RMS, Slot Power,
Burst Average Power
Measurement Uncertainty ± 0.16 dB 1
± 0.17 dB 2
± 0.16 dB 1
± 0.18 dB3
Datasheet 11410-00414 11410-00483 11410-00424 11410-00504
(for complete specifications)
Notes: 1) Total RSS measurement uncertainty (0 ºC to 50 ºC) for power measurements of a CW signal greater than –20 dBm with
zero mismatch errors.
2) Expanded uncertainty with K=2 for power measurements of a CW signal greater than +20 dBm with a matched load.
Measurement results referenced to the input side of the sensor.
3) Expanded uncertainty with K=2 for power measurements of a CW signal greater than –20 dBm with zero mismatch errors.
Page 8 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
LTE Signal Analyzers (Options 0541, 0542, 0546)
Measurements
RF Modulation Over-the-Air (OTA) Pass/Fail
(Option 0541) (Option 0542) (Option 0546) (User Editable)
Channel Spectrum Constellation Scanner Pass Fail All
Channel Power Reference Signal Power (Up to six Sync Signals) Pass/Fail RF
Occupied Bandwidth Sync Signal Power Sync Signal Power Pass Fail Demod
ACPR EVM Cell ID Measurements
Spectral Emission Mask Frequency Error Sector ID Channel Power
RF Summary Carrier Frequency Group ID Occupied Bandwidth
Cell ID Dominance ACLR
Sector ID Modulation Results Frequency Error
Group ID (strongest sync signal) Carrier Frequency
Control Channel Power Dominance
RS EVM (peak)
P-SS EVM (rms)
S-SS RS Power
PBCH SS Power
PCFICH P-SS Power
Modulation Summary S-SS Power
PBCH Power
PCFICH Power
Cell ID
Group ID
Sector ID
Setup Parameters
Frequency E-UTRA bands 1 - 21 (tunable 10 MHz to 4.0 GHz)
Center, Signal Standard, Channel #, Closest Channel, Decrement/Increment Channel
Bandwidth 1.4, 3, 5, 10 MHz
Span Auto, 1.4, 3, 5, 10, 15, 20 MHz
Amplitude Scale/Division, Power Offset, Auto Range, Adjust Range
Sweep Single/Continuous, Trigger Sweep
EVM Mode Auto, PBCH only
Save/Recall Setup, Measurement, Screen Shot (save only), to Internal/External Memory
Measurement Summary Screens Overall Measurements, RF Measurements, Signal Quality Measurements
RF Measurements (Option 0541)
RF Channel Power Accuracy ± 1.5 dB, ± 1.0 dB typical, (RF input –50 dBm to +10 dBm)
Page 9 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
GSM/GPRS/EDGE Signal Analyzers (Options 0040, 0041)
Measurements
RF Demodulation Over-the-Air (OTA) Pass/Fail
(Option 0040) (Option 0041) (User Editable)
Channel Spectrum Phase Error RF Measurements and Channel Power
Channel Power EVM Demodulation can be made OTA. Occupied Bandwidth
Occupied Bandwidth Origin Offset Burst Power
Burst Power C/I There are no additional OTA Average Burst power
Average Burst Power Modulation Type Measurements. Frequency Error
Frequency Error Magnitude Error Phase Error
Modulation Type BSIC (NCC, BCC) EVM
BSIC (NCC, BCC) Origin Offset
Multi-channel Spectrum C/I
Power vs. Time (Frame/Slot) Magnitude Error
Channel Power
Occupied Bandwidth
Burst Power
Average Burst Power
Frequency Error
Modulation Type
BSIC (NCC, BCC)
Setup Parameters
GSM/EDGE Select Auto, GSM, EDGE
Frequency Center, Signal Standard, Channel #, Closest Channel, Decrement/Increment Channel
Amplitude Power Offset, Auto Range, Adjust Range
Sweep Single/Continuous, Trigger Sweep
Save/Recall Setup, Measurement, Screen Shot (save only), to Internal/External Memory
Measurement Summary Screens Overall Measurements, RF Measurements, Signal Quality Measurements
RF Measurements (Option 0040) (temperature range 15 °C to 35 °C)
Frequency Error ± 10 Hz + time base error, 99% confidence level
Occupied Bandwidth Bandwidth within which 99% of the power transmitted on a single channel lies
Burst Power Error ± 1.5 dB, ± 1 dB typical, (–50 dBm to +20 dBm)
Page 10 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
W-CDMA/HSDPA Signal Analyzers (Options 0044, 0045 or 0065, 0035)
Measurements
RF Demodulation Over-the-Air (OTA) Pass/Fail
(Option 0044) (Option 0045 or 0065) (Option 0035) (User Editable)
Band Spectrum Code Domain Power Graph Scrambling Code Scanner (Six) Max Output Power
Channel Spectrum P-CPICH Power Scrambling Codes Frequency Error
Channel Power Channel Power CPICH EVM
Occupied Bandwidth Noise Floor Ec/Io CPICH
Peak-to-Average Power EVM Ec Occupied Bandwidth
Spectral Emission Mask Carrier Feed Through Pilot Dominance Spectral Mask
Single carrier ACLR Peak Code Domain Error OTA Total Power ACLR
Multi-carrier ACLR Carrier Frequency Multipath Scanner (Six) PCDE
Frequency Error Six Multipaths P-CCPCH
Control Channel Power Tau S-CCPCH
Abs/Rel/Delta Power Distance Code Spread 3
CPICH, P-CCPCH RSCP PICH
S-CCPCH, PICH Relative Power Code 128
P-SCH, S-SCH Multipath Power
HSDPA Test Models
Power vs. Time 1 (16), (32), (64)
Constellation 2
Code Domain Power Table 3 (16), (32)
Code, Status 4 (+CPICH), (-CIPCH)
EVM, Modulation Type 5 (2 HS), (4 HS), (8 HS)
Power, Code Utilization
Power Amplifier Capacity
Codogram
Setup Parameters
Scrambling Code, Threshold Auto, Manual
User Selectable Scrambling Code, S-CCPCH Spread, S-CCPCH Code, PICH Code, Threshold,
Max Amp Power, CPICH Power, Frequency Error Average
Maximum Spreading Factor 256, 512
Frequency Center, Signal Standard, Channel #, Closest Channel, Decrement/Increment Channel
Amplitude Scale/Division, Power Offset, Auto Range, Adjust Range, Units (dBm/Watts)
Marker Six Markers, Table On/Off
Sweep Single/Continuous, Trigger Sweep
Save/Recall Setup, Measurement, Screen Shot (save only), to Internal/External Memory
Measurement Summary Screens Overall Measurements, RF Measurements, Signal Quality Measurements
RF Measurements (Option 0044) (temperature range 15 °C to 35 °C)
Frequency Range Bands I – XIV, XVII
RF Channel Power Accuracy ± 1.25 dB, ± 0.7 dB typical, (temperature range 15 °C to 35 °C)
Occupied Bandwidth Accuracy ± 100 kHz
Adjacent Channel Leakage Ratio (ACLR) –54 dB/–59 dB ± 0.8 dB @ 5 MHz/10 MHz offset, typical, Bands I – VI, VIII – XIV, XVII
–54 dB/–57 dB ± 1.0 dB @ 5 MHz/10 MHz offset, typical, Band VII
Demodulation (Option 0045 for W-CDMA only or 0065 for W-CDMA and HSDPA) (temperature range 15 °C to 35 °C)
Frequency Error ± 10 Hz + time base error, 99% confidence level
EVM Accuracy ± 2.5%, 6% ≤ EVM ≤ 25%
Residual EVM (RMS) 3.25% typical
Code Domain Power ± 0.5 dB for code channel power > –25 dB,
16, 32, 64 DCPH (test model 1),
16, 32 DCPH (test model 2, 3)
Page 11 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
cdmaOne/CDMA2000 1X Signal Analyzers (Option 0042, 0043, 0033)
Measurements
RF Demodulation Over-the-Air (OTA) Pass/Fail
(Option 0042) (Option 0043) (0ption 0033) (User Editable)
Channel Spectrum Code Domain Power Graph Pilot Scanner (Nine) Channel Power
Channel Power Pilot Power PN Occupied Bandwidth
Occupied Bandwidth Channel Power Ec/Io Peak-to-Average Power
Peak-to-Average Power Noise Floor Tau Spectral Mask Test
Spectral Emission Mask Rho Pilot Power Frequency Error
Multi-carrier ACPR Carrier Feed Through Channel Power Channel Frequency
Tau Pilot Dominance Frequency error
RMS Phase Error Multipath Scanner (Six) Pilot Power
Frequency Error Ec/Io Noise Floor
Abs/Rel/ Power Tau Rho
Pilot Channel Power Carrier Feed Through
Page Multipath Power Tau
Sync Limit Test – 10 Tests Averaged RMS Phase Error
Q Page Rho Code Utilization
Code Domain Power Table Adjusted Rho Measured PN
Code Multipath Pilot Dominance
Status Pilot Dominance Multipath Power
Power Pilot Power
Multiple Codes Pass/Fail Status
Code Utilization
Setup Parameters
PN Setup PN Trigger (No Trigger, GPS, External), PN Search Type (Auto, Manual), PN Offset
Walsh Codes 64, 128
Measurement Speed Fast, Normal, Slow
External Trigger Polarity Rising, Falling
Number of Carriers 1 to 5
Carrier Bandwidth 1.23, 1.24, 1.25 MHz
Frequency Center, Signal Standard, Channel #, Closest Channel, Decrement/Increment Channel
Amplitude Scale/Division, Power Offset, Auto Range, Adjust Range, Units (dBm/Watts)
Sweep Single/Continuous, Trigger Sweep
Save/Recall Setup, Measurement, Screen Shot (save only), to Internal/External Memory
Measurement Summary Screens Overall Measurements, RF Measurements, Signal Quality Measurements
RF Measurements (Option 0042) (temperature range 15 °C to 35 °C)
RF Channel Power Accuracy ± 1.5 dB, ± 1.0 dB typical, (RF input -50 dBm to +20 dBm)
Demodulation (Option 0043) (temperature range 15 °C to 35 °C)
Frequency Error ± 10 Hz + time base error, 99% confidence level (in slow mode)
Rho Accuracy ± 0.005, for Rho > 0.9
Residual Rho > 0.995, typical, > 0.99 maximum, (RF input –50 dBm to +20 dBm)
PN Offset 1 x 64 chips
Pilot Power Accuracy ± 1.0 dB typical, relative to channel power
Tau ± 0.5 µs typical, ± 1.0 µs maximum
Over-the-Air (OTA) Measurements (Option 0033)
Page 12 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
CDMA2000 1xEV-DO Signal Analyzers (Option 0062, 0063, 0034)
Measurements
RF Demodulation Over-the-Air (OTA) Pass/Fail
(Option 0062) (Option 0063) (Option 0034) (User Editable)
Channel Spectrum MAC Code Domain Power Graph Pilot Scanner (Nine) Channel Power
Channel Power Pilot & MAC Power PN Occupied Bandwidth
Occupied Bandwidth Channel Power Ec/Io Peak-to-Average Power
Peak-to-Average Power Frequency Error Tau Carrier Frequency
Power vs. Time Rho Pilot Pilot Power Frequency Error
Pilot & MAC Power Rho Overall Channel Power Spectral Mask
Channel Power Data Modulation Pilot Dominance Noise Floor
Frequency Error Noise Floor Multipath Scanner (Six) Pilot Power
Idle Activity MAC Code Domain Power Table Ec/Io RMS Phase Error
On/Off Ratio Code Tau Tau
Spectral Emission Mask Status Channel Power Code Utilization
Multi-carrier ACPR Power Multipath Power Measured PN
Code Utilization Pilot Dominance
Data Code Domain Power Multipath Power
Active Data Power
Data Modulation
Rho Pilot
Rho Overall
Maximum Data CDP
Minimum Data CDP
Setup Parameters
PN Setup PN Trigger (No Trigger, GPS, External), PN Search Type (Auto, Manual), PN Offset
Walsh Codes 64, 128
Measurement Speed Fast, Normal, Slow
External Trigger Polarity Rising, Falling
Slot Type Auto, Active, Idle
Number of Carriers 1 to 5
Carrier Bandwidth 1.23, 1.24, 1.25 MHz
Frequency Center, Signal Standard, Channel #, Closest Channel, Decrement/Increment Channel
Amplitude Scale/Division, Power Offset, Auto Range, Adjust Range, Units (dBm/Watts)
Sweep Single/Continuous, Trigger Sweep
Save/Recall Setup, Measurement, Screen Shot (save only), to Internal/External Memory
Measurement Summary Screens Overall Measurements, RF Measurements, Signal Quality Measurements
RF Measurements (Option 0042) (temperature range 15 °C to 35 °C)
RF Channel Power Accuracy ± 1.5 dB, ±1.0 dB typical, (RF input -50 dBm to +20 dBm)
Demodulation (Option 0063) (temperature range 15 °C to 35 °C)
EV-DO Compatibility Rev 0 and Rev A
Frequency Error ± 10 Hz + time base error, 99% confidence level
Rho Accuracy ± 0.01, for Rho > 0.9
Residual Rho > 0.995 typical, > 0.99, maximum (RF input –50 dBm to +20 dBm)
PN Offset Within 1 x 64 chips
Page 13 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
IEEE 802.16 Fixed WiMAX Signal Analyzers (Options 0046, 0047)
Measurements
RF Demodulation Over-the-Air (OTA) Pass/Fail
(Option 0046) (Option 0047) (User Editable)
Channel Spectrum Constellation RF Measurements and Channel Power
Channel Power RCE (RMS/Peak) Demodulation can be made OTA. Occupied Bandwidth
Occupied Bandwidth EVM (RMS/Peak) Burst Power
Power vs. Time Frequency Error There are no additional OTA Preamble Power
Channel Power Carrier Frequency Measurements. Crest Factor
Preamble Power Base Station ID Frequency Error
Data Burst Power Spectral Flatness Carrier Frequency
Crest Factor Adjacent Subcarrier Flatness EVM
ACPR EVM vs. Subcarrier/Symbol RCE
RCE Base Station ID
EVM
Frequency Error
Carrier Frequency
Base Station ID
Setup Parameters
Bandwidth 1.25, 1.50, 2.50, 3.50, 5.00, 5.50, 6.00, 7.00, 10.00 MHz
Cyclic Prefix Ratio (CP) 1/4, 1/8, 1/16, 1/32
Span 5, 10, 15, 20 MHz
Frame Length 2.5, 5.0, 10.0 msec
Frequency Center, Signal Standard, Channel #, Closest Channel, Decrement/Increment Channel
Amplitude Scale/Division, Power Offset, Auto Range, Adjust Range
Sweep Single/Continuous, Trigger Sweep
Save/Recall Setup, Measurement, Screen Shot (save only), to Internal/External Memory
Measurement Summary Screens Overall Measurements, RF Measurements, Signal Quality Measurements
RF Measurements (Option 0046) (temperature range 15 °C to 35 °C)
RF Channel Power Accuracy ± 1.5 dB, ± 1.0 dB typical, (RF input –50 dBm to +20 dBm)
Demodulation (Option 0047) (temperature range 15 °C to 35 °C)
Frequency Error 0.07 ppm + time base error, 99% confidence level
Residual EVM (rms) 3% typical, 3.5% maximum (RF Input –50 dBm to +20 dBm)
Page 14 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
IEEE 802.16 Mobile WiMAX Signal Analyzers (Options 0066, 0067, 0037)
Measurements
RF Demodulation Over-the-Air (OTA) Pass/Fail
(Option 0066) (Option 0067) (Option 0037) (User Editable)
Channel Spectrum Constellation Channel Power Monitor Channel Power
Channel Power RCE (RMS/Peak) Preamble Scanner (Six) Occupied Bandwidth
Occupied Bandwidth EVM (RMS/Peak) Preamble Downlink Bust Power
Power vs. Time Frequency Error Relative Power Uplink Burst Power
Channel Power CINR Cell ID Preamble Power
Preamble Power Base Station ID Sector ID Crest Factor
Downlink Burst Power Sector ID PCINR Frequency Error
Uplink Burst Power Spectral Flatness Dominant Preamble Carrier Frequency
ACPR Adjacent Subcarrier Flatness Base Station ID EVM
EVM vs. Subcarrier/Symbol RCE
RCE (RMS/Peak) Sector ID
EVM (RMS/Peak)
Frequency Error
CINR
Base Station ID
Sector ID
DL-MAP (Tree View)
Setup Parameters
Zone Type PUSC
DL-MAP Auto Decoding Convolutional Coding (CC), Convolutional Turbo Coding (CTC)
Bandwidths 3.50, 5.00, 7.00, 8.75, 10.00 MHz
Cyclic Prefix Ratio (CP) 1/8
Span 5, 10, 20, 30 MHz
Frame Lengths 5, 10 msec
Demodulation Auto, Manual, FCH
Frequency Center, Signal Standard, Channel #, Closest Channel, Decrement/Increment Channel
Amplitude Scale/Division, Power Offset, Auto Range, Adjust Range
Sweep Single/Continuous, Trigger Sweep
Save/Recall Setup, Measurement, Screen Shot (save only), to Internal/External Memory
Measurement Summary Screens Overall Measurements, RF Measurements, Signal Quality Measurements
Channel Power Monitor Over time (one week), measurement time interval 1 to 60 sec
Preamble Scanner Six Strongest Preambles
Auto Save Yes
GPS Logging Yes
Page 15 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
TD-SCDMA/HSDPA Signal Analyzers (Options 0060, 0061, 0038)
Measurements
RF Demodulation Over-the-Air (OTA) Pass/Fail
(Option 0060) (Option 0061) (Option 0038) (User Editable)
Channel Spectrum Code Domain Power/Error Code Scan (32) Occupied Bandwidth
Channel Power (QPSK/8 PSK/16 QAM) Scrambling Code Group Channel Power
Occupied Bandwidth Slot Power Tau Channel Power RCC
Left Channel Power DwPTS Power Ec/Io On/Off Ratio
Left Channel Occ B/W Noise Floor Pilot Dominance Peak-to-Average Ratio
Right Channel Power Frequency Error Tau Scan (Six) Frequency Error
Right Channel Occ B/W Tau Sync-DL# EVM
Power vs. Time Scrambling Code Tau Peak EVM
Six Slot Powers EVM Ec/Io Peak Code Domain Error
Channel Power (RRC) Peak EVM DwPTS Power Tau
DL-UL Delta Power Peak Code Domain Error Pilot Dominance Noise Floor
UpPTS Power
DwPTS Power
On/Off Ratio
Slot Peak-to-Average Power
Spectral Emission
Setup Parameters
Slot Selection Auto, 0-6
Trigger Trigger Type (No Trigger/GPS/External), External Trigger (Rising/Falling), Tau Offset
SYNC-DL Code Auto, 0-31
Scrambling/Midamble Code Auto, 0-127
Maximum Users Auto, 2, 4, 6, 8, 10, 12, 14, 16
Measurement Speed Fast, Normal, Slow
User Selectable Uplink Switch Point, Number of Carriers (1, 3), Tau Offset
Demodulation Type Auto, QPSK, 8 PSK, 16 QAM
Frequency Center, Signal Standard, Channel #, Closest Channel, Decrement/Increment Channel
Amplitude Scale/Division, Power Offset, Auto Range, Adjust Range, Units (dBm/Watts)
Sweep Hold/Run, Trigger Sweep
Save/Recall Setup, Measurement, Screen Shot (save only), to Internal/External Memory
Measurement Summary Screens Overall Measurements, RF Measurements, Signal Quality Measurements
RF Measurements (Option 0060) (temperature range 15 °C to 35 °C)
RF Channel Power Accuracy (RRC) ± 1.5 dB, ± 1.0 dB typical, (slot power –40 dBm to +10 dBm)
Frequency Error ± 10 Hz + time base error, in the presence of a downlink slot
Demodulation (Option 0061) (temperature range 15 °C to 35 °C)
Supported Modulation QPSK, 8 PSK, 16 QAM
Residual EVM (rms) 3% typical, P-CCPH slot power > –50 dBm
PN Offset Within 1 x 64 chips
Pilot Power Accuracy ± 1.0 dB typical
Timing Error (Tau) for Dominant SYNC-DL ± 0.2 µs (external trigger)
Spreading Factor 1, 16
Over-the-Air (OTA) Measurements (Option 0038)
Code Scanner 32 Sync Codes and associated Scrambling Code Groups
Tau Scanner Six strongest Sync Codes
Auto Save Yes
GPS Logging Yes
Page 16 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
ISDB-T (Options 0030, 0032) For full specifications, see Technical Data Sheet 11410-00436
Measurements
ISDB-T RF ISDB-T Signal Analysis ISDB-T Measurement Modes ISDB-T SFN Analysis
(Option 0030) (Option 0030) (Option 0030) (Option 0032)
Signal Power Constellation (w/zoom) Custom Delay Profile (w/zoom)
Channel Power Layer A, B, C, TMCC User specified measurement Inband Spectrum
Termination Voltage Sub-carrier MER and setup parameters Measured Data
Open Terminal Voltage Delay Profile (w/zoom) Easy Channel Power
Field Strength Frequency Response User specified measurements. Delay
Spectrum Monitor Measured Data Some setup parameters are DU Ratio
Channel Power Frequency automatically set or detected Power
Zone Center Channel Frequency Offset Batch Field Strength
Zone Center Frequency MER (Total, Layer A/B/C, User specified measurements
Spectrum Mask TMCC, AC1) and channels for automatic
Mask (Standard A) Japan Modulation (Layer A/B/C) measurement, display of
Mask (Standard B) Japan Mode, GI results and storage
Mask (Critical) Brazil Sub-carrier MER w/marker
Mask (Sub-critical) Brazil Delay w/marker
Mask (Non-critical) Brazil Frequency Response w/
Phase Noise marker
Spurious Emissions
Setup Parameters
Channel Map UHF (Japan), UHF (Brazil), None
Channel 13 to 62 (Japan), 14 to 69 (Brazil)
Frequency 35 MHz to 806 MHz
Pre-amp On, Off
Reference Level Setting –25 dBm to +20 dBm / 5 dB steps (Pre Amp: Off), –50 dBm to –10 dBm/10 dB steps (Pre Amp: On)
Page 17 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Backhaul Analyzers (Options 0051, 0052, 0053)
T1 Bit-Error-Rate Tester (BERT) (Option 0051)
Measurements
Error Detection Frame Bits, Bit Errors, BER, BPV, CRC, PATLS
Error Analysis (ITU G-821) Errored Seconds (ES), Error Free Seconds (EFS), Severely Errored Seconds (SES),
Unavailable Seconds (UAS), Available Seconds (AS), Degraded Minutes (DGRM)
Rx Signal Frequency (± 5 ppm, Max/Min), Vpp (± 5%) (Max/Min), dBdsx, Clock Slips, Frame Slips
VF Frequency (100 Hz to 3000 Hz, ± 3 Hz), Power (-40.0 dBm to +3.0 dBm, ± 0.2 dBm)
Status (Historical and Current) Rx (Signal, Frame Sync, Pattern Sync), DS1 (Alarms, Errors, B8ZS)
Status (Current) Tx (Alarm On, Error On, Loop On)
Setup
BERT Display Table, Histogram, Event List, Clear History
VF Tx (Off/On), Channel (1-24), Tx Freq, Tx Level (–30 dBm to 0 dBm), Volume, Audio, Clear
Line Code AMI, B8ZS
Tx Clock Internal (1.544 MHz ± 5 ppm), Recovered, External
Tx LBO 0.0 dB, –7.5 dB, –15.0 dB
Rx Input Terminate (Bantam connector 100 Ω balanced),
Monitor (Connect via 20 dB pad in DSX, 20 dB flat gain)
Bridge (≥ 1000 Ω, –36 dB to +6 dB)
Framing ESF, SF-D4
Payload T1 (1.544 Mbps), Fractional T1 (Nx64, 64, 56, 16, 8 kbps)
Pulse Shapes Conform to ANSI T1.403 and ITU G.703
Patterns QRSS, PRBS (2-9, 2-11, 2-15, 2-20, 2-23), All Ones, All Zeros,
1-in-8 (1-in-7), 2-in-8, 3-in-24 T1 Daly, Six User defined (≤ 32 bits),
Inverse Patterns (On/Off), Remote Loop Up/Down
Loop Codes CSU, NIU, Link Type (In-Band, Data-Link), Self Loop Up/Down, Loop Code User Defined
Error Insertion Bit Error, Bit Error Rate (BER), BPV, Frame Bit Error, Error (On/Off)
Alarm Insertion AIS On/Off (Blue Alarm), RAI On/Off (Yellow Alarm)
Data Log 1 minute to 3 days
Page 18 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Backhaul Analyzers (Options 0051, 0052, 0053)
T3 Bit-Error-Rate Tester (BERT) (Option 0053)
Measurements
Error Detection Frame Bits, Bit Errors, BER, BPV, Lof Count, P-bit Errors, C-bit Errors, FEBE Errors
Error Analysis (ITU G-821) Excess Zeros, Errored Seconds (ES), Error Free Seconds (EFS),
Severely Errored Seconds (SES), Unavailable Seconds (UAS),
Available Seconds (AS), Degraded Minutes (DGRM), Pattern Loss Seconds (PATLS)
Rx Signal Frequency (± 5 ppm, Max/Min), Vpp (± 5%) (Max/Min), dBdsx
VF Frequency (100 Hz to 3000 Hz, ± 3 Hz), Power (–30.0 dBm to +0.0 dBm, ± 0.2 dBm)
Status (Historical and Current) Rx (Signal, Frame Sync, Pattern Sync), DS3 (Alarms, Errors, DS3ZS)
Status (Current) Insert (Alarm On, Error On, Loop On)
Setup
BERT Display Table, Histogram, Event List, Clear History
VF Tx (Off/On), Channel #, Tx Freq, Tx, Level, Volume, Audio (On/Off)
Line Code AMI, B3ZS
Tx Clock Internal (44.736 MHz ± 5 ppm), Recovered
Tx LBO Low, DSX
Rx Input DSX3 (Bantam connector 100 Ω balanced)
Monitor (Connect via 20 dB pad in DSX)
Framing M13, C-Bit, Unframed
Test Mode Auto, DS3, DS1
Pulse Shapes Carrier present, Frame ID and Sync, Pattern ID and Sync
Patterns QRSS, PRBS (2-9, 2-11, 2-15, 2-20, 2-23), All Ones, All Zeros, 1010,
1-in-8 (1-in-7), 2-in-8, 3-in-24 T1 Daly, Six User defined (≤ 32 bits),
Inverse Patterns (On/Off), Loop Up/Down
Loop Codes Stuff Bit, DS3 C-Bit FEAC, DS3 Self Loop
Error Insertion Bit Error, BPV, DS3 Frame Bit Error, C-bit, P-bit, FEBE, Error Insert (On/Off)
Alarm Insertion AIS (Blue Alarm), RAI (Yellow Alarm), Idle Alarm, Alarm (On/Off)
Data Log 1 minute to 3 days
Page 19 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
General Specifications
All specifications and characteristics apply under the following conditions, unless otherwise stated: 1) After 5 minutes of warm-up time, where the instrument is
left in the ON state; 2) All specifications apply when using internal reference; 3) All specifications subject to change without notice; 4) Typical performance is the
measured performance of an average unit; 5) Recommended calibration cycle is 12 months; 6) Performance Sweep Mode.
Setup Parameters
System Status (Temperature, Battery Info, Serial Number, Firmware Version, Options Installed)
Self Test, Application Self Test
GPS (see Option 0031)
System Options Name, Date and Time, Brightness, Volume
Language (English, French, German, Spanish, Chinese, Japanese, Korean, Italian, User defined)
Reset (Factory Defaults, Master Reset, Update Firmware)
File Save, Recall, Delete, Directory Management
Save/Recall Setups, Measurements, Screen Shots Jpeg (save only)
Delete Selected File, All Measurements, All Mode Files, All Content
Directory Management Sort Method (Name/Type/Date), Ascend/Descend, Internal/USB, Copy, Format USB
Internal Trace/Setup Memory 2,000 traces, 2,000 setups
External Trace/Setup Memory Limited by size of USB Flash drive
Mode Switching Auto-Stores/Recalls most recently used Setup Parameters in the Mode
Connectors
RF Out Type N, female, 50 Ω (Reflection In)
RF Out Damage Level 23 dBm, ± 50 VDC
RF In Type N, female, 50 Ω
RF Input Damage Level +35 dBm peak, ± 50 VDC, Maximum Continuous Input (≥ 10 dB attenuation)
GPS SMA(f)
T1 Bantam Jacks
T3 BNC connectors
E1 RJ48C
External Power 5.5 mm barrel connector, 12.5 VDC to 15 VDC, < 4.0 Amps
USB Interface (2) Type A (Connect USB Flash Drive and Power Sensor)
USB Interface 5-pin mini-B, Connect to PC for data transfer and/or remote control
Headset Jack 2.5 mm mini-phone plug
External Reference In BNC, female, 50 Ω, Maximum Input +10 dBm
1 MHz, 5 MHz, 10 MHz, 13 MHz
External Trigger/Clock Recovery BNC, female, 50 Ω, Maximum Input ± 50 VDC
Display
Type Resistive Touchscreen
Size 8.4 in. daylight viewable color LCD
Resolution 800 x 600
Battery
Type Li-Ion
Battery Operation 3 hours, typical
Electromagnetic Compatibility
European Union CE Mark, EMC Directive 2004/108/EC
Low Voltage Directive 2006/95/EC
Australia and New Zealand C-tick N274
Interference EN 61326-1
Emissions EN 55011
Immunity EN 61000-4-2/-4-3/-4-4/-4-5/-4-6/-4-11
Safety
Safety Class EN 61010-1 Class 1
Product Safety IEC 60950-1 when used with Anritsu Company supplied Power Supply
Environmental
Operating Temperature –10 °C to 55 °C
Maximum Humidity 95% RH (none condensing) at 40 °C
Shock MIL-PRF-28800F Class 2
Storage –40 °C to 71 °C
Altitude 4600 meters, operating and non-operating
ESD
RF Port Center Pin Withstands up to ± 15 kV
Size and Weight
Size 273 mm x 199 mm x 91 mm, (10.7 in x 7.8 in x 3.6 in)
Weight 3.71 kg, (8.2 lbs)
Page 20 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Line Sweep Tools (for your PC)
Trace Capture
Browse to Instrument View and copy traces from the test equipment to your PC using Windows Explorer
Open legacy files Open DAT files captured with Hand Held Software Tools v6.61
Open Current files Open VNA or DAT files
Capture plots to: The Line Sweep Tools screen, DAT files, Database, or JPEG
Traces
Trace Types Return Loss, VSWR, DTF-RL, DTF-VSWR, Cable Loss, Smith Chart, and PIM
Trace formats DAT, VNA, CSV, PNG, BMP, JPG, HTML, Data Base, and PDF
Report Generation
Report Generator Includes GPS location along with measurements
Report Format Create reports in HTML or PDF format
Report setup Report Title, Company, Prepared for, Location, Date and Time, Filename, Company logo
Trace Setup 1 trace Portrait Mode, 2 Trace Portrait Mode, 1 Trace Landscape Mode
Trace Validation
Presets 7 presets allow “one click” setting of up to 6 markers and one limit line
Marker Controls 6 regular Markers, Marker Peak, Marker valley, Marker between, and frequency entry
Delta Markers 6 Delta markers
Limit Line Enable and drag or value entry. Also works with presets
Next Trace Button Next Trace and Previous trace arrow keys allow quick switching between traces
Tools
Cable Editor Allows creation of custom cable parameters
Distance to Fault Converts a Return Loss trace to a Distance to Fault trace
Measurement Calculator Converts Real, Imaginary, Magnitude, Phase, RL, VSWR, Rho, and Transmit power
Signal Standard Editor Creates new band and channel tables
Renaming Grid 36 user definable phrases for creation of file names, trace titles, and trace subtitles
Connectivity
Connections Ethernet, USB cable, USB Memory Stick, and RS-232 Serial Null Modem cable
Page 21 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Ordering Information
MT8212E MT8213E Description
2 MHz to 4 GHz 2 MHz to 6 GHz Cable and Antenna Analyzer
100 kHz to 4 GHz 100 kHz to 6 GHz Spectrum Analyzer
10 MHz to 4 GHz 10 MHz to 6 GHz Power Meter
Options Options
MT8212E-0419 MT8213E-0419 PIM Analyzer (requires PIM Master)
Page 22 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Power Sensors (For complete ordering information see the respective datasheets of each sensor)
Part Number Description
PSN50 High Accuracy RF Power Sensor, 50 MHz to 6 GHz, +20 dBm
MA24104A Inline High Power Sensor, 600 MHz to 4 GHz, +51.76 dBm
MA24106A High Accuracy RF Power Sensor, 50 MHz to 6 GHz, +23 dBm
MA24108A Microwave USB Power Sensor, 10 MHz to 8 GHz, +20 dBm
MA24118A Microwave USB Power Sensor, 10 MHz to 18 GHz, +20 dBm
MA24126A Microwave USB Power Sensor, 10 MHz to 26 GHz, +20 dBm
Manuals (soft copy included on Handheld Instruments Documentation Disc and at www.anritsu.com)
Part Number Description
10920-00060 Handheld Instruments Documentation Disc
Cell Master
MT8212E, MT8213E
10580-00244 Spectrum Analyzer Measurement Guide
- Interference Analyzer, Channel Scanner, Gated Sweep,
Compact Handheld Base Station Analyzer
CW Signal Generator, AM/FM/PM Analyzer,
Cable and Antenna Analyzer
Spectrum Analyzer Interference Mapping, Coverage Mapping
Power Meter
Page 23 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Standard Accessories (included with instrument)
Part Number Description
10920-00060 Handheld Instruments Documentation Disc
10580-00250 Cell Master User Guide (includes Bias-Tee, GPS Receiver)
3-68736 Soft Carrying Case
2300-498 Master Software Tools (MST) CD Disc
2300-530 Anritsu Tool Box with Line Sweep Tools (LST) DVD Disc
633-44 Rechargeable Li-Ion Battery
40-168-R AC-DC Adapter
806-141-R Automotive Cigarette Lighter 12 VDC Adapter
3-2000-1498 USB A/5-pin mini-B Cable, 10 feet/305 cm
11410-00485 Cell Master MT8212E/MT8213E Technical Data Sheet
One Year Warranty (Including battery, firmware, and software)
Certificate of Calibration and Conformance
Optional Accessories
Calibration Components, 50 Ω
Part Number Description
ICN50B InstaCal™ Calibration Module, 38 dB, 2 MHz to 6.0 GHz, N(m), 50 Ω
OSLN50-1 Precision Open/Short/Load, N(m), 42 dB, 6.0 GHz, 50 Ω
OSLNF50-1 Precision Open/Short/Load, N(f), 42 dB, 6.0 GHz, 50 Ω
2000-1618-R Precision Open/Short/Load, 7/16 DIN(m), DC to 4.0 GHz 50 Ω
2000-1619-R Precision Open/Short/Load, 7/16 DIN(f), DC to 4.0 GHz 50 Ω
22N50 Open/Short, N(m), DC to 18 GHz, 50 Ω
22NF50 Open/Short, N(f), DC to 18 GHz, 50 Ω
SM/PL-1 Precision Load, N(m), 42 dB, 6.0 GHz
SM/PLNF-1 Precision Load, N(f), 42 dB, 6.0 GHz
Calibration Components, 75 Ω
22N75 Open/Short, N(m), DC to 3 GHz, 75 Ω
22NF75 Open/Short, N(f), DC to 3 GHz, 75 Ω
26N75A Precision Termination, N(m), DC to 3 GHz, 75 Ω
26NF75A Precision Termination, N(f), DC to 3 GHz, 75 Ω
12N50-75B Matching Pad, DC to 3 GHz, 50 Ω to 75 Ω
Phase-Stable Test Port Cables, Armored w/ Reinforced Grip (recommended for cable & antenna line sweep applications)
InterChangeable Adaptor Phase Stable Test Port Cables, Armored w/Reinforced Grip (recommended for cable and antenna line sweep applications. It
uses the same ruggedized grip as the Reinforced grip series cables. Now you can also change the adaptor interface on the grip to four different connector types)
Phase-Stable Test Port Cables, Armored (ideal for use with tightly spaced connectors and other general use applications)
15NNF50-1.5C 1.5 m, DC to 6 GHz, N(m) to N(f), 50 Ω
15NN50-1.5C 1.5 m, DC to 6 GHz, N(m) to N(m), 50 Ω
15NDF50-1.5C 1.5 m, DC to 6 GHz, N(m) to 7/16 DIN(f), 50 Ω
15ND50-1.5C 1.5 m, DC to 6 GHz, N(m) to 7/16 DIN(m), 50 Ω
15NNF50-3.0C 3.0 m, DC to 6 GHz, N(m) to N(f), 50 Ω
15NN50-3.0C 3.0 m, DC to 6 GHz, N(m) to N(m), 50 Ω
Page 24 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Optional Accessories (continued)
Adapters
1091-26-R SMA(m) to N(m), DC to 18 GHz, 50 Ω
1091-27-R SMA(f) to N(m), DC to 18 GHz, 50 Ω
1091-80-R SMA(m) to N(f), DC to 18 GHz, 50 Ω
1091-81-R SMA(f) to N(f), DC to 18 GHz, 50 Ω
1091-172-R BNC(f) to N(m), DC to 1.3 GHz, 50 Ω
510-90 7/16 DIN(f) to N(m), DC to 7.5 GHz, 50 Ω
510-91 7/16 DIN(f) to N(f), DC to 7.5 GHz, 50 Ω
510-92 7/16 DIN(m) to N(m), DC to 7.5 GHz, 50 Ω
510-93 7/16 DIN(m) to N(f), DC to 7.5 GHz, 50 Ω
510-96 7/16 DIN(m) to 7/16 DIN (m), DC to 7.5 GHz, 50 Ω
510-97 7/16 DIN(f) to 7/16 DIN (f), DC to 7.5 GHz, 50 Ω
1091-379-R 7/16 DIN(f) to 7/16 DIN(f), DC to 6 GHz, 50 Ω, w/ Reinforced Grip
510-102-R N(m) to N(m), DC to 11 GHz, 50 Ω, 90 degrees right angle
Precision Adapters
34NN50A Precision Adapter, N(m) to N(m), DC to 18 GHz, 50 Ω
34NFNF50 Precision Adapter, N(f) to N(f), DC to 18 GHz, 50 Ω
Miscellaneous Accessories
2000-1528-R GPS Antenna, SMA(m)
69793 CW Signal Generator Kit
ODTF-1 Optical Distance-to-Fault Module, 1550 nm, Single Mode
2000-1520-R USB Flash Drive
2000-1374 External Charger for Li-lon Batteries
2300-532 Map Master CD
Directional Antennas
2000-1411-R 822 MHz to 900 MHz, N(f), 10 dBd, Yagi
2000-1412-R 885 MHz to 975 MHz, N(f), 10 dBd, Yagi
2000-1413-R 1710 MHz to 1880 MHz, N(f), 10 dBd. Yagi
2000-1414-R 1850 MHz to 1990 MHz, N(f), 9.3 dBd, Yagi
2000-1415-R 2400 MHz to 2500 MHz, N(f), 10 dBd, Yagi
2000-1416-R 1920 MHz to 2170 MHz, N(f), 10 dBd, Yagi
2000-1519-R 500 MHz to 3 GHz, log periodic
Portable Antennas
2000-1200-R 806 MHz to 866 MHz, SMA(m), 50 Ω
2000-1473-R 870 MHz to 960 MHz, SMA(m), 50 Ω
2000-1035-R 896 MHz to 941 MHz, SMA(m), 50 Ω (1/2 wave)
2000-1030-R 1710 MHz to 1880 MHz, SMA(m), 50 Ω (1/2 wave)
2000-1474-R 1710 MHz to 1880 MHz with knuckle elbow (1/2 wave)
2000-1031-R 1850 MHz to 1990 MHz, SMA(m), 50 Ω (1/2 wave)
2000-1475-R 1920 MHz to 1980 MHz and 2110 to 2170 MHz, SMA(m), 50 Ω
2000-1032-R 2400 MHz to 2500 MHz, SMA(m), 50 Ω (1/2 wave)
2000-1361-R 2400 MHz to 2500 MHz, 5000 MHz to 6000 MHz, SMA(m), 50 Ω
2000-1636-R Antenna Kit (Consists of: 2000-1030-R, 2000-1031-R, 2000-1032-R,
2000-1200-R, 2000-1035-R, 2000-1361-R, and carrying pouch)
Page 25 of 28
Cell Master™ MT8212E, MT8213E Base Station Analyzer Specifications
Optional Accessories (continued)
Mag Mount Broadband Antenna
2000-1647-R Cable 1: 698-1200 MHz 2 dBi peak gain, 1700-2700 MHz 5 dBi peak gain,
N(m), 50 Ω, 10 ft
Cable 2: 3000-6000 MHz 5 dBi peak gain, N(m), 50 Ω, 10 ft
Cable 3: GPS 26 db gain, SMA(m), 50 Ω, 10 ft
2000-1645-R 694-894 MHz 3 dBi peak gain, 1700-2700 MHz 3dBi peak gain,
N(m), 50 Ω, 10 ft
2000-1646-R 750-1250 MHz 3 dBi peak gain, 1650-2000 MHz 5 dBi peak gain,
2100-2700 MHz 3 dBi peak gain, N(m), 50 Ω, 10 ft
2000-1648-R 1700-6000 MHz 3 dBi peak gain,N(m), 50 Ω, 10 ft
Filters
1030-114-R 806 MHz to 869 MHz, N(m) to SMA(f), 50 Ω
1030-109-R 824 MHz to 849 MHz, N(m) to SMA(f), 50 Ω
1030-110-R 880 MHz to 915 MHz, N(m) to SMA(f), 50 Ω
1030-105-R 890 MHz to 915 MHz Band, 0.41 dB loss, N(m) to SMA(f), 50 Ω
1030-111-R 1850 MHz to 1910 MHz, N(m) to SMA(f), 50 Ω
1030-106-R 1710 MHz to 1790 MHz Band, 0.34 dB loss, N(m) to SMA(f), 50 Ω
1030-107-R 1910 MHz to 1990 MHz Band, 0.41 dB loss, N(m) to SMA(f), 50 Ω
1030-112-R 2400 MHz to 2484 MHz, N(m) to SMA(f), 50 Ω
1030-149-R High Pass, 150 MHz, N(m) to N(f), 50 Ω
1030-150-R High Pass, 400 MHz, N(m) to N(f), 50 Ω
1030-151-R High Pass, 700 MHz, N(m) to N(f), 50 Ω
1030-152-R Low Pass, 200 MHz, N(m) to N(f), 50 Ω
1030-153-R Low Pass, 550 MHz, N(m) to N(f), 50 Ω
1030-155-R 2500 MHz to 2700 MHz, N(m) to N(f), 50 Ω
Attenuators
3-1010-122 20 dB, 5 W, DC to 12.4 GHz, N(m) to N(f)
42N50-20 20 dB, 5 W, DC to 18 GHz, N(m) to N(f)
42N50A-30 30 dB, 50 W, DC to 18 GHz, N(m) to N(f)
3-1010-123 30 dB, 50 W, DC to 8.5 GHz, N(m) to N(f)
1010-127-R 30 dB, 150 W, DC to 3 GHz, N(m) to N(f)
3-1010-124 40 dB, 100 W, DC to 8.5 GHz, N(m) to N(f), Uni-directional
1010-121 40 dB, 100 W, DC to 18 GHz, N(m) to N(f), Uni-directional
1010-128-R 40 dB, 150 W, DC to 3 GHz, N(m) to N(f)
Page 26 of 28
Notes
Page 27 of 28
The Master Users Group is an organization dedicated to providing training, technical support, networking opportunities
and links to Master product development teams. As a member you will receive the Insite Quarterly Newsletter with user
stories, measurement tips, new product news and more.
Visit us to register today: www.anritsu.us/smiusignup
To receive a quote to purchase a product or order accessories visit our online ordering site: www.ShopAnritsu.com
Training at Anritsu
Anritsu has designed courses to help you stay up to date with technologies important to your job.
For available training courses visit: www.anritsu.com/training
®Anritsu All trademarks are registered trademarks of 11410-00485, Rev. F Printed in United States 2010-12
their respective companies. Data subject to change ©2010 Anritsu Company. All Rights Reserved.
without notice. For the most recent specifications visit:
www.anritsu.com