PM7100_Operaters_Manual_en_PT00097962B00
PM7100_Operaters_Manual_en_PT00097962B00
PM7100_Operaters_Manual_en_PT00097962B00
TM
INVOS
Patient Monitor
©2019 Medtronic. All rights reserved. Medtronic and Medtronic logo are trademarks of Medtronic.
™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company.
Table of Contents
1 Introduction
2 Product Overview
3 Installation
iv Operator’s Manual
Table of Contents
5 Data Management
6 Performance Considerations
7 Product Maintenance
Operator’s Manual v
Table of Contents
9 Accessories
10 Theory of Operations
11 Product Specifications
A Clinical Studies
vi Operator’s Manual
List of Tables
Table 1-1. Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Table 2-1. Label Symbol Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Table 3-1. Monitoring System - Standard Shipped Items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Table 3-2. AC Power Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Table 3-3. Battery Status Indicators - Monitoring System Operating on Battery Power . . . . . . . . . . . 3-6
Table 3-4. Institutional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Table 4-1. Quick Start Steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Table 4-2. Operational Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Table 4-3. Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Table 4-4. System Memory - Common Situations and Recommended Actions . . . . . . . . . . . . . . . . . 4-10
Table 4-5. Alarm Audio Button States . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-36
Table 4-6. Audible and Visual Alarm Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-37
Table 4-7. Alarm Limits Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-38
Table 5-1. Serial Port Pin-Out Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Table 5-2. Hardware Requirements for Philips VueLink™* and IntelliBridge™* Communication . . . 5-5
Table 5-3. Case History Downloads - Data Fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Table 5-4. Case Data Downloads (PC LINK 1 and PC LINK 2) - Data Fields . . . . . . . . . . . . . . . . . . . . . . . 5-12
Table 5-5. Data Downloads - Event Codes 1 to 40 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
Table 5-6. Data Downloads - Event Codes 41 to 80. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Table 5-7. Data Downloads - Event Codes 81 to 120 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Table 5-8. Data Downloads - Event Codes 121 to 254 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
Table 5-9. Data Downloads - Status Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
Table 7-1. Maintenance Schedule. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Table 8-1. Alarm Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Table 8-2. Error Conditions and Resolutions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
Table 9-1. Monitoring System Accessory Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Table 11-1. Transport, Storage, and Operating Condition Ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
Table 11-2. Alarm Audio Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
Table 11-3. Average Alarm Sound Pressure Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
Table 11-4. Alarm Visual Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
Table 11-5. POST Pass Tone Characteristics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6
Table 11-6. INVOS™ Adult rSO2 Sensor Operating Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6
Table 11-7. Electromagnetic Emissions Guidelines and Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-9
Table 11-8. Electromagnetic Immunity Guidelines and Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-10
Table 11-9. Recommended Separation Distance Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-11
Table 11-10. Test Specifications for Enclosure Port Immunity to RF Wireless Communications
Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-13
Table 11-11. Recommended Separation Distances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-14
Table A-1. Performance Characteristics: Accuracy Bias, Error (SD), RMSD, and Correlation (R2)
for rSO2 and fSO2, 20 Subjects. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Table A-2. Trend Bias, Error (SD), and Correlation (R2) for ΔrSO2 and ΔfSO2, 20 subjects . . . . . . . . . . A-3
viii
List of Figures
Figure 1-1. Start-up Screen Showing Code Version. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Figure 2-1. INVOS™ Patient Monitor Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Figure 2-2. Monitor - Front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Figure 2-3. Monitor - Left Side . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Figure 2-4. Monitor - Right Side. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Figure 2-5. Monitor - Back . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Figure 2-6. Docking Station - Bottom View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Figure 2-7. Docking Station - Connector for Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Figure 2-8. Sample Monitoring Screen Elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Figure 3-1. Power Input Connection at Bottom of Docking Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Figure 3-2. Docking Station Power Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Figure 3-3. Battery Location at Back of Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Figure 3-4. Battery Slot and Latch at Back of Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Figure 3-5. Battery Installed in Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Figure 3-6. Monitor Mounted in Docking Station. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Figure 3-7. Connecting the Preamplifier Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Figure 3-8. Connecting the RSC to the Preamplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Figure 3-9. Power Input Connection at Bottom of Docking Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Figure 3-10. Power Button on Top of Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Figure 3-11. Set-Up Screen - Preamplifier(s) Connected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Figure 3-12. Set-Up Screen - No Preamplifier(s) Connected. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Figure 3-13. Monitoring System on Stand with Preamplifier Hung on Rail . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Figure 4-1. Set-Up Screen - Prompt for Sensor Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Figure 4-2. Monitoring Screen - New Case - No Baselines Set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Figure 4-3. Default Sensor Placement Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
Figure 4-4. Optional Sensor Placement Sequences. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Figure 4-5. Highlighting an On-Screen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Figure 4-6. Sensor Icons Next to CURRENT Values (Press to Locate Sensor on Patient) . . . . . . . . . . . 4-18
Figure 4-7. Patient Icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Figure 4-8. Entering a Patient ID Before Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
Figure 4-9. Patient Icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
Figure 4-10. Entering a Patient ID During Monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
Figure 4-11. Sensor Label Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
Figure 4-12. SETTINGS - GENERAL - SCREEN BRIGHTNESS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23
Figure 4-13. Monitoring Screen - New Data Set - No Baselines Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
Figure 4-14. SET BASELINES - No Baselines Previously Set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
Figure 4-15. SET BASELINES - Baselines Previously Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
Figure 4-16. Monitoring Screen with Baselines Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
Figure 4-17. Trend View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
Figure 4-18. Trend View - One Graph (Left, Default) and Two Graphs (Right) . . . . . . . . . . . . . . . . . . . . . . 4-29
Figure 4-19. SETTINGS - TRENDS - TWO AXIS VIEW vs. ONE AXIS VIEW. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
Figure 4-20. Trend Line Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
Figure 4-21. SETTINGS - TRENDS - TREND LINE AVERAGING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
Operator’s Manual ix
List of Figures
Figure 4-22. Trend View Review Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
Figure 4-23. Trend Graph Pop-Up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
Figure 4-24. Monitoring Screen - Trend Graph Pop-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
Figure 4-25. Visual Alarm Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-36
Figure 4-26. SET ALARM LIMITS - Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-39
Figure 4-27. SET ALARM LIMITS - Upper Limits On. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-39
Figure 4-28. SETTINGS - ALARMS - ALARM VOLUME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-41
Figure 4-29. Dismissible Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-43
Figure 4-30. Events Marked in Trend View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44
Figure 4-31. ADD EVENT Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-45
Figure 4-32. Event Mark Label. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-46
Figure 4-33. EDIT EVENT Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-47
Figure 4-34. SETTINGS - EVENTS (Screen 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-48
Figure 4-35. AUC SUMMARY Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-49
Figure 4-36. CASE HISTORY List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-53
Figure 4-37. Case History Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-54
Figure 5-1. VGA Port on Docking Station. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Figure 5-2. Serial Port on Docking Station. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Figure 5-3. Serial Port Pin-Outs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Figure 5-4. Null Modem Cable Diagram for Connection to Serial Port (Two Options) . . . . . . . . . . . . . 5-7
Figure A-1. Accuracy rSO2 and fSO2, 20 Subjects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Figure A-2. Trend rSO2 and fSO2, 20 Subjects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
x Operator’s Manual
1 Introduction
1.1 Overview
This manual contains information for operating the INVOS™ patient monitor (the
“monitoring system”).
This manual applies to the following product:
PM7100
1-1
Introduction
Symbol Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse events) to
the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the product.
Note
Notes provide additional guidelines or information.
WARNING:
Explosion hazard — Do not use the monitoring system in the presence of flammable
anesthetics.
WARNING:
Explosion hazard — When replacing the battery, do not use the incorrect type. Use
only the battery available from Medtronic. See Accessories/Parts List, page 9-2.
WARNING:
Shock hazard — Ensure the monitoring system is properly grounded when operating
on AC power.
WARNING:
Shock hazard — When connecting the monitoring system to any instrument, verify
proper operation before clinical use. Any equipment connected to the data interface
must be certified according to the latest IEC/EN 60950-1 standard for data-
WARNING:
Toxicity hazard — The LCD panel (screen) contains toxic chemicals. Do not touch
broken LCD panels. Physical contact with a broken LCD panel can result in
transmission or ingestion of toxic substances.
WARNING:
The monitoring system should not be used as the sole basis for diagnosis or therapy.
It is intended only as an adjunct in patient assessment. It must be used in
conjunction with clinical signs and symptoms.
WARNING:
Always disconnect and remove the monitoring system and sensors during magnetic
resonance imaging (MRI) scanning. Attempting to use the monitoring system during
an MRI procedure could cause burns or adversely affect the MRI image or the
monitoring system's performance.
WARNING:
As with all medical equipment, carefully route patient cabling to reduce the
possibility of patient entanglement or strangulation.
WARNING:
Do not set alarm limits to extreme values that render the monitoring system
ineffective. Ensure alarm limits are appropriate for each patient.
WARNING:
Do not silence, pause, or decrease the volume of audible alarms if patient safety
could be compromised.
WARNING:
An external multi-parameter system will not generate an alarm or error message if
remote communication between the multi-parameter system and the monitoring
system has been broken. During this period of no remote communication, the
monitoring system will continue to monitor, generate alarms, and display status
messages. The multi-parameter system operator should not rely on the multi-
parameter system for generating alarms.
WARNING:
Choking hazard — The reusable sensor cables (RSCs) include a strain-relief clip that,
if detached, may pose a choking hazard.
Caution:
If two sensors are placed in close proximity to each other on a patient, the same
preamplifier should be connected to both sensors to avoid poor performance.
1.3.4 Operation
WARNING:
Inspect the monitoring system and all accessories before use to ensure there are no
signs of physical damage or improper function. Do not use if damaged.
WARNING:
If you do not hear a tone at system start-up, discontinue use of the monitoring
system and contact Medtronic or a local Medtronic representative.
WARNING:
Monitoring system readings can be affected by certain patient conditions. Refer to
Patient Conditions, page 6-2.
WARNING:
Explosion hazard — When replacing the battery, do not use the incorrect type. Use
only the battery available from Medtronic. See Accessories/Parts List, page 9-2.
WARNING:
To ensure proper performance, avoid shock, and prevent device damage or failure,
do not expose the monitoring system to extreme moisture, such as direct exposure
Caution:
Dispose of the battery in accordance with local guidelines and regulations.
WARNING:
The use of accessories, sensors, and cables other than those specified may result in
poor performance of the monitoring system and increased electromagnetic
emissions or decreased electromagnetic immunity of the monitoring system.
WARNING:
Failure to cover the sensor site with opaque material when operating under high
ambient light conditions may result in poor performance.
Caution:
Make sure all connectors are fully engaged and free from moisture. Moisture
intrusion may cause poor performance or no readings at all.
WARNING:
Electromagnetic emissions from the monitoring system may interfere with other
critical devices.
WARNING:
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the monitoring system, including cables. Otherwise, degradation of
monitoring system performance may result.
WARNING:
The use of accessories, sensors, and cables other than those specified may result in
poor performance of the monitoring system and increased electromagnetic
emissions or decreased electromagnetic immunity of the monitoring system.
WARNING:
The monitoring system is intended for use by healthcare professionals only. It may
cause radio interference or may disrupt the operation of nearby equipment.
Mitigation for such disruption may require re-orienting or relocating the monitoring
system or shielding the location.
WARNING:
Any radio frequency transmitting equipment or other nearby sources of electrical
noise may result in disruption of the monitoring system.
WARNING:
The monitoring system is designed for use in environments in which the signal can
be obscured by electromagnetic interference. During such interference,
measurements may seem inappropriate or the monitoring system may not seem to
operate correctly.
WARNING:
EMI disruption can cause cessation of operation or other incorrect functioning.
WARNING:
The monitoring system should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, observe the monitoring system to
verify normal operation in the desired configuration. Technical alarms may indicate
that the configuration is not appropriate for the monitoring system.
Caution:
This device has been tested and found to comply with the limits for medical devices
related to IEC 60601-1-2: 2007 and IEC 60601-1-2:2014. These limits are designed to
provide reasonable protection against harmful interference in a typical medical
installation.
Caution:
When operating medical electrical equipment, special precautions related to
electromagnetic compatibility (EMC) are required. Install the monitoring system
according to the EMC information included in this manual.
Caution:
The monitoring system generates, uses, and can radiate radio frequency energy and,
if not installed and used in accordance with the instructions, may cause harmful
interference to other devices in the vicinity. If interference is suspected, move
monitoring system cables away from the susceptible device.
Caution:
The use of an electrosurgical or electrocautery instrument in the vicinity of the
monitoring system may interfere with the signal and cause poor performance or no
readings at all.
1.3.7 Cleaning
Caution:
Do not autoclave or gas sterilize any components of the monitoring system.
Caution:
To prevent device damage or failure, do not expose the monitor to isopropyl alcohol.
1.800.635.5267 or 1.925.463.4635
or contact a local Medtronic representative
www.medtronic.com
on the back of the monitor and the preamplifiers. The software version for the mon-
itoring system is displayed on the start-up screen at power-on.
Note:
An authorized technician can view serial numbers and software versions through the
monitoring system’s service mode. Refer to the monitoring system’s service manual.
• INVOS™ Adult rSO2 Sensor Instructions for Use — Provides important information
about sensor selection and use.
2.1 Overview
This chapter contains basic information about the INVOS™ patient monitor (the
“monitoring system”).
• Product Description, page 2-1
• A VESA™* docking station for the monitor, with USB, Serial (RS-232), and VGA ports
2-1
Product Overview
• User configurable data display (rSO2, change from baseline, sensor labels, and trend
data) (page 2-10)
1 Sensors 4 Monitor
3 Docking station
Monitor - Front
2 I/O port cover with latch (slide to access ports inside) 5 Docking station
Monitor - Back
Docking Station
3 VGA port
Monitoring Screen
1 Message area Provides information about alarm conditions. The background indicates the
severity of the condition. If alarm audio is silenced, an alarm audio silenced
symbol appears next to the alarm message. See Alarm Management, page 4-
35. If a “?” appears next to the alarm message, press the “?” for additional
information.
N/A Dismiss button Available in the message area when an alarm is dismissible. See Dismissing
(not shown) an Alarm, page 4-43.
2 Sensor icons Each sensor icon has a label indicating its location on the patient. See Chang-
ing the On-Screen Sensor Labels, page 4-21.
3 Date and time Current date in MM/DD/YYYY format (by default) and current time in 12-
hour HH:MM:SS format (by default). The date and time format can be
changed by an authorized technician.
5 Battery charging Indicates that the monitoring system is connected to AC power and the
indicator battery is charging.
AC power When the battery is fully charged and the monitoring system remains con-
indicator nected to AC power, the AC power indicator replaces the battery charging
indicator.
7 CHANGE The percent change between the current rSO2 reading and the established
baseline for each sensor, color-coded to the corresponding sensor icon.
8 Sensor/readings in If a sensor alarm or loss of readings occurs, the corresponding sensor is high-
alarm state lighted on the screen. If readings are lost, the CURRENT and CHANGE values
are replaced by dashes. Check the message area for information about the
condition.
9 Alarm limits The current alarm limits for each sensor, color-coded to the corresponding
sensor icon. A horizontal red line indicates each alarm limit in the trend view
graph. See Setting Alarm Limits, page 4-38.
10 Event mark Event marks indicate significant occurrences during monitoring. Event marks
can be added to the trend graph at any time during monitoring and are dis-
played as vertical lines with flags. See Event Marks, page 4-44.
11 Baseline values The current rSO2 baseline established for each tissue region being moni-
tored, color-coded to the corresponding sensor icon. Dotted lines in the
trend view graph also indicate the baseline values, color-coded to the corre-
sponding sensor icon. See Baselines, page 4-24.
12 Alarm Audio button Indicates whether alarm audio is on, silenced, or paused. Press to silence or
pause alarm audio or to turn alarm audio back on. When an alarm occurs and
alarm audio is silenced or paused, the button’s color corresponds to the
highest-level alarm. See Alarm Indicators, page 4-35, and Silencing or Pausing
Alarms, page 4-41.
13 MARK EVENT Press to add an event mark to the trend graph to indicate a significant occur-
button rence during monitoring. Choose from a customizable list of events. See
Event Marks, page 4-44.
14 Patient button While monitoring, press to return to the Set-up screen, reposition on-screen
sensors, or assign or modify a patient ID. See Set Up for Patient Monitoring,
page 4-11, Repositioning On-Screen Sensors, page 4-17, and Assigning or Mod-
ifying the Patient ID, page 4-19.
15 Trend view The trend view shows the progression of rSO2 values over the course of a
case. Data is color-coded to the corresponding sensor icon. rSO2 values are
presented on the vertical (y) axis. Time is presented on the horizontal (x) axis.
See Trend View Management, page 4-27.
16 MENU button Press to access a variety of settings and functions depending on whether or
not sensors are connected and monitoring has begun. See Menu Structure,
page 4-9.
Humidity limits (see Environmental Proper waste disposal for electrical and
Conditions, page 11-4) electronic equipment
3.1 Overview
This chapter contains information for the installation and set up of the INVOS™
patient monitor (the “monitoring system”) prior to first-time use.
• Safety Reminders, page 3-1
WARNING:
Inspect the monitoring system and all accessories before use to ensure there are no
signs of physical damage or improper function. Do not use if damaged.
WARNING:
Shock hazard — Ensure the monitoring system is properly grounded when operating
on AC power.
WARNING:
If you do not hear a tone at system start-up, discontinue use of the monitoring
system and contact Medtronic or a local Medtronic representative.
Caution:
When operating medical electrical equipment, special precautions related to
electromagnetic compatibility (EMC) are required. Install the monitoring system
according to the EMC information included in this manual.
3-1
Installation
Caution:
Make sure all connectors are fully engaged and free from moisture. Moisture
intrusion may cause poor performance or no readings at all.
Note:
Save the cartons and packing material for any future transport.
Item Quantity
Patient monitor 1
Lithium-ion battery 1
Preamplifiers 1 or 2
AC power supply 1
Power cord 1
1. The manual may be printed from the included USB drive. Order a printed Operator’s Manual or Service Manual
from Medtronic Technical Services or your local Medtronic representative. See Technical Services, page 1-7.
3.4.1 AC Power
WARNING:
Shock hazard — Ensure the monitoring system is properly grounded when operating
on AC power.
The power input connection for the monitoring system is at the bottom of the
docking station. Always use the provided AC power cord and power supply.
When the power supply is connected to the AC outlet and the docking station, the
power indicator on the docking station illuminates blue (Figure 3-2).
AC connected (battery
charging)
AC connected (battery
fully charged)
Note:
Whenever the monitoring system is connected to AC power, the battery will charge as
needed.
The monitoring system includes a rechargeable lithium-ion battery located in the
back of the monitor.
If using the monitoring system on battery power, be sure to frequently check the
battery status (Table 3-3).
Table 3-3. Battery Status Indicators - Monitoring System Operating on Battery Power
Battery Life
Caution:
Dispose of the battery in accordance with local guidelines and regulations.
For maximum battery reliability, replace the battery on a regular basis. Medtronic
recommends replacing the battery once every 2 years. If the battery will not sustain
monitoring for more than approximately 1 hour when fully charged, replace the
battery immediately.
Contact Medtronic or a local Medtronic representative for information about obtain-
ing replacement batteries. See Technical Services, page 1-7.
3.5 Setup
Use the following instructions to prepare the monitoring system for first use. See
Operation, page 4-1 for additional steps to be performed by the clinician when mon-
itoring a patient.
• Install the Battery, page 3-7
The monitoring system’s lithium-ion battery is shipped with the monitor, but not
installed. Install and fully charge the battery prior to clinical use (see Apply Power,
page 3-13 for charging information).
To install the battery:
1. Remove the battery from its packaging.
2. At the back of the monitor, make sure that the battery latch, located above the battery
slot, is to the right (Figure 3-4).
2 Battery slot
3. Insert the battery into the slot, bottom edge (with three tabs) first.
4. With the battery completely inserted, slide the battery latch to the left to lock the
battery in place (Figure 3-5).
The docking station is intended to hold the monitor during typical use of the mon-
itoring system. You can mount the docking station on the provided stand or on
other VESA™*-compatible equipment capable of supporting at least 5.1 lb (2.3 kg)
(VESA™* FDMI MIS-D, 75).
Monitor stand use: If you intend to use the monitor stand, attach the docking
station to the stand using the provided hardware. Refer to the instructions provided
with the monitor stand.
To insert the monitor into the docking station:
1. Align the connector at the bottom of the monitor with the connector in the docking
station (Figure 3-6).
2. Press the monitor down into the docking station until the connectors engage. Pegs and
guide holes in the docking station and monitor ensure proper alignment.
3. Press the monitor back into the clip at the top of the docking station until it fully engag-
es.
Caution:
Make sure all connectors are fully engaged and free from moisture. Moisture
intrusion may cause poor performance or no readings at all.
To connect the preamplifier(s) to the monitor:
1. Align the red dot on the preamplifier cable connector with the red marking on the
monitor input connection (Figure 3-7).
2. Insert the cable connector directly upward into the connection until the locking sleeve
rotates and clicks into place.
Note:
To disconnect a preamplifier cable from the monitor, grasp the locking sleeve and rotate it
in the direction indicated by the arrow while pulling downward.
Each preamplifier allows connection of up to two sensors via reusable sensor cables
(RSCs). Using two preamplifiers, the monitoring system can accommodate up to
four RSCs and four sensors.
To connect the RSCs:
1. Align the RSC’s male connector with the connection slot on the preamplifier. The con-
nector and slot are keyed to guide insertion.
2. Press firmly until the connector snaps into place. Ensure that the clip on the connector
engages completely with the connection slot.
WARNING:
If you do not hear a tone at system start-up, discontinue use of the monitoring
system and contact Medtronic or a local Medtronic representative.
Before applying power, refer to Power Options, page 3-3 to understand the require-
ments for using AC power and battery power.
To power on the monitoring system:
1. Ensure that the AC outlet is properly grounded and supplies the specified voltage and
frequency. See Electrical, page 11-3 for voltage and frequency specifications.
2. Connect the power supply to the power input connection on the bottom of the
docking station.
3. Connect the AC power cord to the power supply and the AC outlet. Verify that the blue
power indicator on the docking station is illuminated (Figure 3-2 on page 3-4).
The battery indicator at the top left of the monitor indicates the charge status. The
battery will charge as necessary when the monitoring system is connected to AC power.
Charging
Fully charged
While the monitoring system performs its power-on self-test (POST), a progress bar
appears at the bottom of the screen.
5. Ensure that the POST pass tone sounds when POST completes. See POST Pass Tone,
page 11-6 for tone specifications.
The POST pass tone is an audible confirmation of proper speaker performance. If the
speaker does not function, alarm warning sounds will not be audible.
Note:
If no preamplifiers are connected to the monitoring system, the screen prompts you to
connect them (Figure 3-12). See Connect the Preamplifier(s), page 3-11.
Note:
Make sure that the battery is fully charged prior to clinical use. Table 3-2, page 3-5,
describes the charging status indicators.
Table 3-4 describes the institutional settings available for the monitoring system.
Institutional defaults must be set by an authorized technician. Refer to the monitor-
ing system's service manual. Refer to Table 4-2 on page 4-7 for information about
additional options that can be set or changed by the clinician.
Silence alarms at startup YES, NO NO - Alarms will sound unless the user
presses the Alarm Audio button
2. Observe that the screen goes completely dark. The battery indicator on the monitor and
the power indicator on the docking station remain illuminated as long as power is con-
nected.
WARNING:
As with all medical equipment, carefully route patient cabling to reduce the
possibility of patient entanglement or strangulation.
Users may choose from a variety of mounting configurations, including the provid-
ed monitor stand or other VESA™*-compatible mounting options capable of sup-
porting at least 5.1 lb (2.3 kg) (VESA™* FDMI MIS-D, 75). In addition, the preamplifiers
have a fold-away swivel hook for attachment to rails or poles (Figure 3-13).
• Monitor stand use — When using the monitor stand, place the stand on a sturdy flat
surface. Use the lever on the back of the stand to adjust the angle of the monitor for best
visibility.
• Power outlet access and power cord position — Ensure that the power outlet used
for the monitoring system is easily accessible.
• Cable routing — Position the monitoring system for ease of access to all cabling. You
can use the fold-away hooks on the preamplifiers to hang them from rails or poles.
Ensure that cables are routed so that there is no risk of patient entanglement or stran-
gulation.
• Monitor access and visibility — Ensure that the operator can easily access and view
the monitor while monitoring. Choose a location that allows visual communication of
rSO2 values and alarms. See Physical Characteristics, page 11-2 for visibility specifications.
4.1 Overview
This chapter explains how to use the INVOS™ patient monitor (the “monitoring sys-
tem”) to view and collect patient regional oxygen saturation (rSO2) data.
This chapter assumes that the monitoring system has been installed at the location
of use and tested by the institution. See Installation, page 3-1 for full installation
instructions.
• Safety Reminders, page 4-2
4-1
Operation
WARNING:
The monitoring system should not be used as the sole basis for diagnosis or therapy.
It is intended only as an adjunct in patient assessment. It must be used in
conjunction with clinical signs and symptoms.
WARNING:
Inspect the monitoring system and all accessories before use to ensure there are no
signs of physical damage or improper function. Do not use if damaged.
WARNING:
Explosion hazard — Do not use the monitoring system in the presence of flammable
anesthetics.
WARNING:
Always disconnect and remove the monitoring system and sensors during magnetic
resonance imaging (MRI) scanning. Attempting to use the monitoring system during
an MRI procedure could cause burns or adversely affect the MRI image or the
monitoring system's performance.
WARNING:
As with all medical equipment, carefully route patient cabling to reduce the
possibility of patient entanglement or strangulation.
WARNING:
Do not set alarm limits to extreme values that render the monitoring system
ineffective. Ensure alarm limits are appropriate for each patient.
WARNING:
Do not silence, pause, or decrease the volume of audible alarms if patient safety
could be compromised.
WARNING:
If you do not hear a tone at system start-up, discontinue use of the monitoring
system and contact Medtronic or a local Medtronic representative.
WARNING:
Monitoring system readings can be affected by certain patient conditions. Refer to
Patient Conditions, page 6-2.
WARNING:
Failure to cover the sensor site with opaque material when operating under high
ambient light conditions may result in poor performance.
WARNING:
Electromagnetic emissions from the monitoring system may interfere with other
critical devices.
WARNING:
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the monitoring system, including cables. Otherwise, degradation of
monitoring system performance may result.
WARNING:
The use of accessories, sensors, and cables other than those specified may result in
poor performance of the monitoring system and increased electromagnetic
emissions or decreased electromagnetic immunity of the monitoring system.
WARNING:
The monitoring system is intended for use by healthcare professionals only. It may
cause radio interference or may disrupt the operation of nearby equipment.
Mitigation for such disruption may require re-orienting or relocating the monitoring
system or shielding the location.
WARNING:
Any radio frequency transmitting equipment or other nearby sources of electrical
noise may result in disruption of the monitoring system.
WARNING:
The monitoring system is designed for use in environments in which the signal can
be obscured by electromagnetic interference. During such interference,
measurements may seem inappropriate or the monitoring system may not seem to
operate correctly.
WARNING:
EMI disruption can cause cessation of operation or other incorrect functioning.
WARNING:
The monitoring system should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, observe the monitoring system to
verify normal operation in the desired configuration. Technical alarms may indicate
that the configuration is not appropriate for the monitoring system.
Caution:
If two sensors are placed in close proximity to each other on a patient, the same
preamplifier should be connected to both sensors to avoid poor performance.
Caution:
Make sure all connectors are fully engaged and free from moisture. Moisture
intrusion may cause poor performance or no readings at all.
Caution:
The monitoring system generates, uses, and can radiate radio frequency energy and,
if not installed and used in accordance with the instructions, may cause harmful
interference to other devices in the vicinity. If interference is suspected, move
monitoring system cables away from the susceptible device.
Caution:
The use of an electrosurgical or electrocautery instrument in the vicinity of the
monitoring system may interfere with the signal and cause poor performance or no
readings at all.
Refer to Safety Information, page 1-2 for additional warnings and cautions regarding
use of the monitoring system.
1 Make sure all components of the monitoring system See Set Up for Patient Monitoring, page 4-
are present: 11
Monitor, docking station, power supply, power cord,
preamplifier(s), reusable sensor cables (RSCs), INVOS™
rSO2 sensors
2 Attach up to four INVOS™ rSO2 sensors to the See the Instructions for Use provided with
patient the sensors
3 Power on the monitoring system See Set Up for Patient Monitoring, page 4-
11
4 Optionally, change the sensor placement sequence See Changing the Sensor Placement
Sequence, page 4-16
5 Connect the sensors to the RSCs, following the on- See Set Up for Patient Monitoring, page 4-
screen prompts, and verify the sensor locations 11
6 Secure all cables to avoid patient skin injury or See Set Up for Patient Monitoring, page 4-
entanglement 11
10 Set or change alarm limits, as desired See Setting Alarm Limits, page 4-38
Press MENU, ALARM LIMITS
11 Optionally, change the trend view: See Trend View Management, page 4-27
- Number of graphs
- Trend-line averaging
- Time scale
12 Optionally, change the on-screen sensor labels See Changing the On-Screen Sensor Labels,
page 4-21
13 Optionally, check and modify the list of event See Changing the Event List, page 4-47
markers
14 Optionally, change Area Under the Curve (AUC) See Changing the AUC Threshold, page 4-
thresholds 50
15 Optionally, adjust the screen brightness See Changing the Screen Brightness, page
4-23
16 Optionally, adjust the alarm volume See Changing the Alarm Volume, page 4-40
Alarm audio off/ INDEFINITE, 2 MINS INDEFINITE - When silenced, alarms remain silent
pause duration until the user unsilences them. See Alarm Silence
Duration, page 4-42.
This setting may be changed by an authorized
technician.
Alarm reminder signal ON, OFF OFF - No reminder when alarms are silenced
This setting may be changed by an authorized
technician.
Alarm volume 1 - 10 5
See Changing the Alarm Volume, page 4-40
Screen brightness 1 - 10 7
See Changing the Screen Brightness, page 4-23
Trend axis view TWO AXIS VIEW, ONE AXIS ONE AXIS VIEW
VIEW See Changing the Trend View - Two Graphs or One
Graph, page 4-29
LOWER LIMITS
EVENTS
CASE HISTORIES1
SET BASELINES2
THRESHOLD
1. Available only when no sensors are connected to the monitoring system.
2. Available only when monitoring has begun.
– The monitoring system remembers the last sensors that were connected, the loca-
tions they were assigned (for example, left cerebral), and their last baseline values.
– The monitoring system also remembers the last reusable sensor cable (RSC) and
preamplifier port assignments (for example, which RSC and preamplifier the left
cerebral sensor was connected to).
– The previously used sensor location overrides the last RSC/preamplifier port assign-
ment. For example, if you plug what was the right cerebral sensor into what was the
left cerebral RSC/preamplifier port, it will show up on screen as the right cerebral
sensor.
Because the monitoring system retains this information for an ongoing case, you
can disconnect sensors from the system and reconnect the same sensors without
having to reassign sensor locations or perform the sensor set-up routine.
Table 4-4 describes some common situations, the actions you should perform, and
how the system reacts.
You have started monitoring a patient Navigate to the sensor Set-up screen (see Repositioning On-Screen
but notice that you have the sensor Sensors, page 4-17), drag and drop the sensors to the correct loca-
locations swapped. tions on the screen, confirm the new locations, then press APPEND
OLD DATA to resume monitoring.
Sensors need to be disconnected Disconnect and reconnect the affected sensors. The monitoring
from the monitoring system during a system automatically assigns the sensors back to the proper loca-
case (for example, to reroute the tions.
cables).
The monitoring system is powered off No action is necessary. The monitoring system automatically
and back on during a case. assigns sensors to the proper location and resumes monitoring.
A case has ended. Unplug and discard all sensors that were used during the case.
Dismiss the “Sensor Not Connected” alarm. The monitoring system
returns to the Set-up screen.
• Monitor
• Power supply and power cord - AC power is recommended for extended moni-
toring. You can use battery power briefly, if necessary, such as when transporting a
patient between locations within a hospital.
• Reusable sensor cables (RSCs) - Use one RSC per sensor, up to two per preampli-
fier. If no RSCs are connected, see Connect the Reusable Sensor Cables (RSCs), page 3-
12, for instructions.
• INVOS™ rSO2 sensors - Before use, carefully read the sensor Instructions for Use,
including all warnings, cautions, and instructions.
2. Determine the sites you will monitor on the patient. Select the correct type of INVOS™
rSO2 sensor for your patient; do not mix sensor types. Attach up to four sensors to the
patient. Refer to the Instructions for Use provided with the sensors for application instruc-
tions.
a. Make sure that the monitoring system’s power supply is plugged into the docking
station and a properly grounded hospital mains outlet (see AC Power, page 3-3). You
may operate the monitoring system briefly on battery power if necessary, but AC
power is recommended for extended monitoring. See Power Options, page 3-3 for
additional information about powering the monitoring system.
b. Press the Power button on top of the monitor (see Apply Power, page 3-13). If oper-
ating on battery power, check the battery status indicator on the screen to deter-
mine whether the battery needs charging (see Table 3-3, page 3-6).
c. Ensure that the POST pass tone sounds when POST completes. See POST Pass Tone,
page 11-6 for tone specifications.
The POST pass tone is an audible confirmation of proper speaker performance. If the
speaker does not function, alarm warning sounds will not be audible.
Once POST is complete, the Set-up screen appears (Figure 4-1).
4. Optionally, you can set the sensor sequence for the number of sensors you are using. If
you are using fewer than four sensors, you can set the number to three or two for on-
screen representations. See Changing the Sensor Placement Sequence, page 4-16.
5. Connect the RSCs to the sensors that have been applied to the patient:
Caution:
If two sensors are placed in close proximity to each other on a patient, the same
preamplifier should be connected to both sensors to avoid poor performance.
a. Note the highlighted sensor location on screen (Figure 4-1). Locate the correspond-
ing sensor applied to the patient.
b. Look for the flashing blue light on the preamplifier and at the end of the corre-
sponding RSC.
c. Align the sensor’s male connector with the connection slot on the RSC. The connec-
tor and slot are keyed to guide insertion.
d. Press firmly until the connector snaps into place. The monitoring system indicates
proper connection by displaying an rSO2 reading at the corresponding sensor loca-
tion. If the sensor was previously used on the monitoring system, the last baseline
obtained on the sensor is also displayed.
e. Look for the next highlighted sensor location on screen and the next flashing blue
light on the preamplifier and RSC. Connect the next sensor to the corresponding
RSC.
f. Repeat these steps for each sensor applied to the patient.
g. Verify the placement of all sensors by briefly pressing each on-screen sensor loca-
tion and noting the blue flashing light on the RSC cable. Make sure that the on-
screen location matches the sensor location on the patient. If the location does not
match, you can reposition the on-screen sensors rather than disconnecting the
RSCs. See Repositioning On-Screen Sensors, page 4-17.
Note:
As you connect the RSCs to sensors applied to the patient, readings are displayed on the
Set-up screen. However, trends are not tracked and physiological alarms are disabled.
Do not attempt to monitor the patient from the Set-up screen.
6. Check the location of all cables connected to the monitoring system. Make sure that the
patient is not lying on any cables or connectors. To prevent entanglement and pro-
longed contact with patient skin, you can secure the RSCs with the strain-relief clips
attached to the cables. Do not place the RSCs, preamplifiers, or cables connecting the
preamplifiers to the monitor in contact with the patient.
7. Optionally, enter a patient ID. See Assigning or Modifying the Patient ID, page 4-19.
Note:
If necessary, you can return to the Set-up screen while monitoring by pressing the
Patient button:
10. Set or change alarm limits, as desired. See Setting Alarm Limits, page 4-38.
11. Optionally, change the trend view. You can view all sensor readings on one graph or
two, use trend line averaging, and change the time scale. See Trend View Management,
page 4-27.
12. Optionally, change the on-screen sensor labels. See Changing the On-Screen Sensor
Labels, page 4-21.
13. Optionally, check the list of available event markers and change it if necessary. See
Changing the Event List, page 4-47
14. Optionally, change the Area Under the Curve (AUC) threshold. See Changing the AUC
Threshold, page 4-50.
15. Optionally, adjust the screen brightness. See Changing the Screen Brightness, page 4-23.
16. Optionally, adjust the alarm volume. See Changing the Alarm Volume, page 4-40.
17. As you monitor the patient, you may need to perform the following tasks:
• Respond to alarms - See Alarm Indicators, page 4-35, Alarm Messages, page 8-2,
Silencing or Pausing Alarms, page 4-41, and Dismissing an Alarm, page 4-43.
Note:
In the event of defibrillation, the monitoring system will recover operation without
intervention within 30 seconds after the defibrillation voltage is removed.
18. When the monitoring session is complete, follow the steps in Finish Monitoring, page 4-
51.
By default, the monitoring system assumes that you will use two cerebral sensors
and two somatic sensors per case. You are prompted to connect the sensors in the
following order:
If you monitor a different combination of sites, you can change the prompts to one
of the following sequences:
2. Press SETTINGS.
3. Press GENERAL.
5. Press CLOSE.
Figure 4-6. Sensor Icons Next to CURRENT Values (Press to Locate Sensor on Patient)
4. At the Set-up screen, drag the on-screen sensor to a new position and release.
6. Press APPEND OLD DATA to resume monitoring. Note that there will be a brief gap in
the data on the trend graph.
By default, cases are identified by date and start/end time, but do not have patient
IDs assigned. As an option, you can assign an alphanumeric patient ID to a case
before or during monitoring. You can also modify a patient ID at any time while
monitoring.
Note:
Follow institutional protocols for assigning patient IDs. Avoid using patient-sensitive
information in IDs.
To assign a patient ID before monitoring:
1. Set up the monitoring system as described in Set Up for Patient Monitoring, but do not
begin monitoring.
You can change the sensor labels to any alphabetic value (A - Z) or S1 - S4.
Note:
You can verify the location of a sensor by briefly pressing the sensor label next to the
CURRENT value on the monitoring screen. The blue LEDs on the RSC light to indicate the
corresponding sensor.
2. Scroll through the choices by pressing anywhere within the list and dragging up or
down. When you see the choice you want, press it to highlight it. Note that duplicate
labels are not permitted.
3. Press CLOSE.
2. Press SETTINGS.
3. Press GENERAL.
5. Press CLOSE.
4.9 Baselines
The monitoring system requires an rSO2 baseline for each tissue region being mon-
itored so that changes from the baseline can be reported. Changes in rSO2 values of
>20% from baseline are considered clinically significant and cause for concern and
possible interventions.
When measured rSO2 values are below or above specific limits, the monitoring
system issues an alarm. You can use the monitoring system’s default alarm limits or
set custom limits, as described in Setting Alarm Limits, page 4-38.
It is recommended that you obtain baselines while the patient is stable and awake
(for example, prior to surgical induction). You can set baselines for all sensors at once
or for individual sensors. If necessary, you can retake baselines at any time during
monitoring.
Automatic Baselines: If you do not actively set baselines, the monitoring system
automatically sets them approximately five minutes after monitoring begins. You
can use these automatic baselines, or retake the baselines. Be aware that sudden
large changes in a patient’s saturation values during the initiation of monitoring can
result in unrepresentative automatically calculated baseline values. If you use auto-
matic baselines, be sure to check the values to make sure they are appropriate. Man-
ually retake the baselines, if necessary.
Note:
If you reposition or replace a sensor during monitoring, be sure to retake the baseline for
that sensor to ensure a valid representation of rSO2 at that location.
To set or retake baselines:
1. Set up the monitoring system and begin monitoring as described in Set Up for Patient
Monitoring, page 4-11. While baselines have not yet been set, the monitoring system
displays rotating arrows for the CHANGE values next to the CURRENT rSO2 readings.
2. Press MENU.
3. Press SET BASELINES. The SET BASELINES screen either indicates that no baselines have
been set (no value next to “BL” as in Figure 4-14), or, if automatic baselines have taken
effect, indicates the values and times they were taken (Figure 4-15).
• Individual sensor - Press the on-screen circle representing the sensor. The baseline
is set to the current rSO2 reading at that sensor site. The new baseline value is dis-
played with the date and time it was taken.
• All sensors - Press RETAKE ALL BASELINES. The baselines are set to the current rSO2
readings at all sensor sites. The new baseline values are displayed with the date and
time they were taken.
5. Press CLOSE.
The baseline values are indicated to the left of the trend graph, color-coded to the
corresponding sensor label. An event marker in the trend graph indicates the point
at which the baselines were taken.
The CHANGE values reflect the difference between the baselines and the CURRENT
readings.
Note:
If you disconnect and re-connect the same sensor while monitoring, the baseline for that
sensor is maintained. If you replace a sensor with a new sensor while monitoring, the
baseline calculated with the previous sensor is applied to the new sensor. If the monitoring
system is powered off and back on while monitoring, baselines are maintained for all
sensors connected to the system.
By default, the monitoring system displays all trends on one graph (Figure 4-18, left).
If desired, you can display trends on two graphs - one for cerebral sensors, one for
somatic sensors (Figure 4-18, right).
Figure 4-18. Trend View - One Graph (Left, Default) and Two Graphs (Right)
2. Press SETTINGS.
3. Press TRENDS.
Figure 4-19. SETTINGS - TRENDS - TWO AXIS VIEW vs. ONE AXIS VIEW
4. Press TWO AXIS VIEW or ONE AXIS VIEW, as desired. Your choice is highlighted by a
white square.
5. Press CLOSE.
Note:
The trend view setting remains in effect across power cycles.
Trend line averaging provides a 60-minute rolling average of rSO2 values. Viewing a
rolling average can be useful in situations where there is frequent and wide variabil-
ity in rSO2 values. The averaged data is displayed as a bold line superimposed over
the real-time rSO2 values in the graph. The trend line is the same color as the real-
time values for each sensor. The numerical rSO2 values and percent change from
baseline continue to be displayed in real-time.
2. Press SETTINGS.
3. Press TRENDS.
5. Press CLOSE.
Note:
The trend line averaging setting remains in effect across power cycles.
By default, the trend view shows 1 hour of data at a time. You can zoom the trend
view to see different time intervals. Available intervals are 1, 2, 4, 8, 12, and 24 hours.
Note that only the horizontal axis (time) zooms, not the vertical axis (rSO2 values).
To zoom the trend view:
1. To zoom out and see a longer time range, place two fingers within the trend graph, 1-
2 inches apart horizontally, and slide together.
2. To zoom in and see a shorter time range, place two fingers within the trend graph, side-
by-side horizontally, and slide apart.
Note:
The zoom setting remains in effect across power cycles.
4.10.5 Reviewing Data that has Scrolled Off the Trend View
During extended monitoring, trend data scrolls off the trend view to the left. You can
review this trend data while monitoring by swiping the trend view to the right.
Current rSO2 values are still displayed while you review older trends, but current
trends are not displayed until you swipe the trend view back to the left. As an indi-
cation that current trends are not shown, “REVIEW MODE” appears on the trend
view.
2. Repeat swipes until the data you want to view appears. Note that the amount of shift
corresponds to the length of the swipe.
3. To return to current trend data, swipe to the left until the message “REVIEW MODE” no
longer appears.
While monitoring a patient, you can view rSO2 values that were recorded earlier in
the case. Values appear in a pop-up corresponding to the point of interest (Figure 4-
23).
2. Slide your finger in either direction on the graph to see values at different times.
• Dismissing an Alarm
Note:
Refer to Table 8-1 on page 8-2 for a complete list of alarm messages, priorities, and
resolutions.
The monitoring system uses audio and visual indicators to identify alarms.
The message area at the top of the monitoring screen indicates active alarms
(Figure 4-25). If multiple alarms occur, the message area shows the color of the
highest priority alarm and indicates the total number of alarms currently active. By
pressing the arrow in the message area, you can expand the list and view all active
alarms.
When an rSO2 value crosses an alarm threshold, the background of the affected rSO2
reading flashes yellow.
The Alarm Audio button indicates alarm audio status. When alarms are silenced or
paused, the Alarm Audio button also indicates alarm status (Table 4-5 on page 4-36).
2 Message area
Silenced No alarms
Medium 3 pulse Message area: Flashing yellow bar with text BATTERY LOW
burst message LOW RSO2
Battery icon: For battery alarm, flashing yellow; oth-
erwise, no change
rSO2 readings area: Yellow flashing background on
affected reading for sensor alarms (dashes appear if
readings are lost)
Alarm Audio button: Yellow if alarms silenced or
paused
Low 2 pulse Message area: Yellow bar with text message (not TREND DATA LOST
burst flashing)
Battery icon: No change
rSO2 readings area: No change
Alarm Audio button: Yellow if alarms silenced or
paused
1. By default, alarm audio is on, but may be silenced or paused by the user. See Silencing or Pausing Alarms, page 4-41. See Table 11-
2 on page 11-5 for alarm tone specifications.
2. See Table 11-4 on page 11-5 for alarm visual specifications.
3. See Table 8-1 on page 8-2 for a complete list of alarm messages.
WARNING:
Do not set alarm limits to extreme values that render the monitoring system
ineffective. Ensure alarm limits are appropriate for each patient.
The monitoring system provides default alarm limit settings, as listed in Table 4-7.
You can change the alarm limits before or after setting baselines, or you can use the
existing alarm limits. Alarm limits are retained across power cycles, so check the
current settings to determine if they are appropriate for the patient.
UPPER LIMITS OFF - No alarm sounds for high rSO2 values OFF
ON - Specify a high rSO2 value that will generate an alarm
LOWER LIMITS MANUAL - Specify a low rSO2 value that will generate an alarm MANUAL
AUTO - Specify a percentage below baseline that will generate
an alarm
Note:
Upper and lower alarm limits for each sensor are not allowed to cross.
To set alarm limits:
1. Press MENU.
2. Press ALARM LIMITS. The SET ALARM LIMITS screen shows the most recent limits set (or
the defaults if limits have not previously been set).
Note:
Four sensors are shown even if fewer are in use.
a. Press ON for UPPER LIMITS. A second scale appears for each sensor.
b. Next to a sensor, press and hold the slider for the UPPER limit, drag it to the desired
value, and release. You can set different values for each sensor.
Note:
Alarm limit settings remain in effect across power cycles.
WARNING:
Do not silence, pause, or decrease the volume of audible alarms if patient safety
could be compromised.
When alarm audio is on, the default alarm volume is set to 5 with a range of 1 to 10.
See Table 11-2 on page 11-5 for alarm tone specifications.
To change the alarm volume:
1. Press MENU.
2. Press SETTINGS.
3. Press ALARMS.
5. Press CLOSE.
Note:
The alarm volume setting remains in effect across power cycles.
WARNING:
Do not silence, pause, or decrease the volume of audible alarms if patient safety
could be compromised.
Depending on your institutional default, alarm audio may be on or off when you
start a case. When alarm audio is on, you can silence or pause the audio.
A white Alarm Audio button indicates alarm audio is on. You can silence or pause
alarms by pressing the button. The button changes to indicate whether alarms are
silenced or paused (see Table 4-5 on page 4-36). Whether an alarm is silenced or
paused depends on institutional default settings (see Alarm Silence Duration, page 4-
42).
If an alarm condition occurs when alarm audio is silenced or paused, the Alarm
Audio button changes color to match the current highest-priority alarm condition
Note:
When alarms are silenced or paused, the following alarms are not silenced:
- BATTERY CRITICALLY LOW (while monitoring system is operating on battery power)
- SYSTEM FAILURE (under certain circumstances)
Note:
After a power cycle, the last setting used (alarm audio on or silenced) is retained.
Pressing the Alarm Audio button either silences alarms or pauses them temporarily,
depending on institutional defaults. The factory default is permanent silence of
alarms.
The pause option, which must be set by an authorized technician, is 2 minutes. If the
pause option is implemented, a countdown timer appears below the Alarm Audio
button when the button is pressed to pause an alarm. After 2 minutes, alarm audio
resumes.
By default, there is no reminder signal to indicate that alarms are silenced or paused.
Alarm silence reminders may be turned on by an authorized technician. See Alarm
Audio and Visual Characteristics, page 11-4 for alarm reminder signal specifications.
Some alarms, such as BATTERY LOW and SENSOR NOT CONNECTED, can be dis-
missed without resolving the situation. Others, such as LOW RSO2 and CHECK
SENSOR, clear only when the condition is resolved. Table 8-1 on page 8-2 indicates
which alarms are dismissible and which are not.
To dismiss an alarm:
1. Note whether the DISMISS button is available in the alarm message area and whether
there are multiple alarms occurring. When there are multiple alarms, the DISMISS
button applies only to dismissible alarms.
2. If there are multiple alarms, press the down arrow to view all current alarms and deter-
mine which are appropriate to be dismissed.
3. Press DISMISS next to the appropriate alarm. Dismissible alarms can only be dismissed
individually.
• Marking Events
• Renaming an Event
To mark an event:
1. While monitoring, press MARK EVENT. The ADD EVENT menu lists the available events.
2. Scroll through the list by pressing anywhere within the list and dragging up or down.
When you see the event you want, press it to highlight it. (If you don’t see the event you
want, see Changing the Event List, page 4-47.)
Event mark labels indicate the event name and the time the event was recorded.
You can view event mark labels while monitoring or reviewing a case history.
To view an event mark label:
1. At the trend graph, press the flag below an event mark line. The event name and time
appear for approximately 5 seconds.
If an event is mis-named, you can change it at any time while monitoring or review-
ing a case history.
To rename an event:
1. At the trend graph, press the flag below an event mark line.
2. When the event mark label appears, release the flag and press the label or the flag again.
The EDIT EVENT menu appears.
3. Scroll through the list by pressing anywhere within the list and dragging up or down.
When you see the event you want, press it to highlight it. (If you don’t see the event you
want, see Changing the Event List, page 4-47.)
4. Press OK.
5. Press the flag below the event mark line to verify the change.
A default list of common events is provided. You can change the list before or during
a case.
Note:
Be aware that changes to the list are retained across power cycles, so the list you see may
reflect a previous case. Make sure that the list of events is appropriate for your patient.
To change the event list:
1. Press MENU.
2. Press SETTINGS.
3. Press EVENTS. One of several screens of available events appears. An X next to an event
indicates that it currently appears in the ADD EVENT and EDIT EVENT menus.
4. Press NEXT or BACK to view all of the available events in alphabetical order.
5. To select or de-select an individual event, press the checkbox next to the event or the
event itself. If necessary, you can restore the list to its default values by pressing
RESTORE.
6. Press CLOSE.
AUC (Area Under the Curve), also referred to as cumulative saturation below thresh-
old, quantifies the depth and duration of desaturation below a specific value.
AUC was originally a metric in The STS (Society of Thoracic Surgeons) Adult Cardiac
Surgery Database and Congenital Heart Surgery Database. High AUC calculated
from a threshold of 25% below baseline rSO2 has been found to correlate with
increased morbidity.
By default, the monitoring system uses a threshold of 50 for AUC calculations. If
desired, you can change the AUC threshold to any value between 30 and 60, or you
can specify a percentage below baseline from 0% to 30%.
The monitoring system automatically calculates AUC by multiplying the difference
between the threshold and the current rSO2 values times the duration that rSO2 is
below the threshold. Units are minute-%. Values are accumulated throughout the
case. The AUC threshold applies to all sensors being monitored.
Note:
If you retake baselines at any point while monitoring, the monitoring system does not reset
AUC data collection to zero. If desired, you can reset AUC data collection as described in
Resetting AUC Collection, page 4-51.
By default, rSO2 values must fall below a fixed threshold of 50 to be included in AUC
totals. You can change this threshold any time after monitoring begins.
Note:
Be aware that changes to the AUC threshold are retained across power cycles, so the current
setting may reflect a previous case. Check the setting to make sure that it is appropriate for
your patient.
To change the AUC threshold:
1. After monitoring begins, press MENU.
3. At the AUC SUMMARY screen, the type of threshold currently in use is indicated (see
Figure 4-35 on page 4-49):
• FIXED, or
• % BELOW BASELINE
4. Adjust the value for the threshold by pressing the arrow buttons. The ranges are:
• FIXED: 30 - 60
5. Press CLOSE.
You can reset AUC data collection to zero at any point while monitoring. AUC data
collection restarts at the time of reset.
To reset AUC Collection:
1. After monitoring begins, press MENU.
3. At the AUC SUMMARY screen, press RESET AUC (see Figure 4-35 on page 4-49).
4. Press CLOSE.
2. Disconnect the RSCs from the sensors, and unclip the RSC strain-relief clips, if used. You
can leave the RSCs connected to the preamplifier(s).
3. Carefully remove the sensors from the patient. Dispose of the sensors according to insti-
tution procedures for single-use devices.
4. Clean the monitoring system as required by your institution. See Cleaning the Monitor-
ing System, page 7-2.
While not actively monitoring a patient, you can view case histories stored on the
monitoring system. You can also export case histories to a USB flash drive and review
or store them off-line.
The monitoring system automatically records trend data every 5 seconds and stores
up to 30 days (720 hours) of data in any combination of cases. When storage capac-
ity is reached, the monitoring system overwrites the oldest data to make room for
new data.
2. Press MENU.
• Sort the list by pressing DATE or PATIENT ID. Press your choice again to reverse the
sort order.
• Scroll through the cases by pressing anywhere within the list and dragging up or
down.
5. Press FULL CASE next to the case you want to view. The trend data for the case is dis-
played.
• Zoom the trend graph. See Zooming the Trend View, page 4-32.
• Press the trend graph at any location to see the readings at that point in the case.
The vertical line with the magnifier moves to that point. Slide your finger along the
graph to move the vertical line left or right to see additional values. If an alarm
occurred at any point, the alarm message is displayed and the sensor reading is
highlighted, if applicable.
• Scroll backward and forward in time by briefly touching within the trend graph and
swiping to the right or left. The shift is indicated by the time scale. Repeat swipes
until the data you want to view appears. Note that the amount of shift corresponds
to the length of the swipe.
• View an event mark label by pressing the flag below the vertical event line. See
Viewing an Event Mark Label, page 4-45.
• Add a new event mark by pressing the trend graph at the location of the event, then
pressing the Add Event button at the top of the screen. Select an event as described
in Marking Events, page 4-45.
• Export the case history, if desired. See Exporting Case Histories, page 4-55.
Using a USB flash drive, you can export individual or multiple cases from the case
history list, or you can export the case that you are viewing. After exporting cases,
you can upload them to a computer. See Downloading Case Histories to a USB Drive,
page 5-2 for information about data format, file names, and working with the data.
Note:
Medtronic recommends incorporating appropriate security measures for any external
devices receiving patient data from the monitoring system.
To export one or more case histories from the case history list:
1. Make sure that the monitoring system is powered on but that no sensors are connect-
ed.
2. Press MENU.
5. Insert a USB flash drive into a USB port on the monitor or docking station. See Figure 2-
3 on page 2-5 and Figure 2-6 on page 2-8.
6. Press the Export button. A progress bar appears. Do not remove the USB drive while the
export is taking place.
7. When the export is complete, press FINISH or remove the USB drive.
9. If you have not already done so, remove the USB drive.
2. Insert a USB flash drive into a USB port on the monitor or docking station. See Figure 2-
3 on page 2-5 and Figure 2-6 on page 2-8.
3. Press the Export button. A progress bar appears. Do not remove the USB drive while the
export is taking place.
4. When the export is complete, press FINISH or remove the USB drive to return to the case
history list.
6. If you have not already done so, remove the USB drive.
5.1 Overview
This chapter explains how to display INVOS™ patient monitor screens on an external
monitor, download case data for storage and analysis on a computer, and transmit
real-time data to external devices such as a Philips multi-parameter system.
• Safety Reminders, page 5-1
• Transmitting Monitoring System Data to External Devices via the Serial Port, page 5-3
WARNING:
Shock hazard — When connecting the monitoring system to any instrument, verify
proper operation before clinical use. Any equipment connected to the data interface
must be certified according to the latest IEC/EN 60950-1 standard for data-
processing equipment, the latest IEC/EN 60601-1 standard for electromedical
equipment, or the latest IEC/EN safety standards relevant to that equipment. All
combinations of equipment must be in compliance with Requirements for Medical
Electrical Systems within IEC/EN Standard 60601-1. Anyone who connects
equipment to the data interface is configuring a medical system and, therefore, is
responsible for ensuring the system complies with the Requirements for Medical
Electrical Systems IEC/EN Standard 60601-1 and the electromagnetic compatibility
IEC/EN Standard 60601-1-2. Performance may degrade if it is connected to
secondary I/O devices when the equipment is not connected to earth reference.
5-1
Data Management
WARNING:
An external multi-parameter system will not generate an alarm or error message if
remote communication between the multi-parameter system and the monitoring
system has been broken. During this period of no remote communication, the
monitoring system will continue to monitor, generate alarms, and display status
messages. The multi-parameter system operator should not rely on the multi-
parameter system for generating alarms.
Note:
Medtronic recommends incorporating appropriate security measures for any external
devices receiving patient data from the monitoring system.
Note:
The monitoring system will also interface with multi-parameter systems that accept the
Philips VOI B module specified in Transmitting Real-Time Data to a Philips IntelliBridge™* and
VueLink™* Open Interface (IVOI) Module, page 5-5. Contact Medtronic Technical Services for
information about compatibility with other commercial devices. See Technical Services, page
1-7.
• No parity
• 8 data bits
• 1 stop bit
Pin-outs for the serial port are shown in Figure 5-3 and described in Table 5-1.
5 Ground
WARNING:
An external multi-parameter system will not generate an alarm or error message if
remote communication between the multi-parameter system and the monitoring
system has been broken. During this period of no remote communication, the
monitoring system will continue to monitor, generate alarms, and display status
messages. The multi-parameter system operator should not rely on the multi-
parameter system for generating alarms.
The following instructions explain how to transmit real-time data to a Philips IVOI
module for display on a Philips multi-parameter system. The data transmitted
includes rSO2 values, alarms, and status messages.
Hardware requirements for communicating with a Philips VueLink™* module or
Philips IntelliBridge™* EC10 module are described in Table 5-2.
Table 5-2. Hardware Requirements for Philips VueLink™* and IntelliBridge™* Communication
Philips VueLink™* System - Philips VOI B module (Philips VueLink™* P/N M1032A #A05)
- Philips VOI/RS-232 Interface Cable (Philips VueLink™* P/N M1032 #K6B)
- VueLink™* adapter cable, DB25F to DB9F (Medtronic P/N VLI)
3. Connect the VOI/RS-232 Interface Cable via the VueLink™* adapter cable to the serial
port on the monitoring system docking station (Figure 5-2 on page 5-3).
5. Ensure that the Philips monitor and VOI B module are properly configured per Philips
instructions.
6. Power on the monitoring system and set the serial port protocol to VUE LINK, as
described in the monitoring system's service manual. After the monitoring system
restarts, it begins communicating with the Philips monitor.
3. Connect the other end of Ethernet cable to the IntelliBridge™* EC5 ID module.
4. Using the IntelliBridge™* EC5 Open Interface RS232 cable, connect the IntelliBridge™*
EC5 ID module to the serial port on the monitoring system docking station (Figure 5-2
on page 5-3).
6. Ensure that the Philips monitor and IntelliBridge™* module are properly configured per
Philips instructions.
7. Power on the monitoring system and set the serial port protocol to VUE LINK, as
described in the monitoring system's service manual. After the monitoring system
restarts, it begins communicating with the Philips monitor.
The following instructions explain how to transmit case data to a computer using a
null modem cable and a terminal emulation program such as Tera Term™* or Hyper-
Terminal™* programs.
Data is transmitted approximately once every second as the case progresses. Data
can be transmitted in two formats: PC LINK 1and PC LINK 2. Instructions for selecting
a format are provided in the monitoring system's service manual. See Case Data
Downloaded via Serial Port, page 5-10 for descriptions of the formats.
A 9-pin to 9-pin null modem cable is required for transmitting case data to a com-
puter via the serial port.
The null modem cable must be no longer than 50 feet (15.24 m).
Refer to Figure 5-4 for acceptable 9-pin to 9-pin wiring configurations.
Figure 5-4. Null Modem Cable Diagram for Connection to Serial Port (Two Options)
The following procedure presents general steps for transmitting case data to a com-
puter using a terminal emulation program and a null modem cable. Steps will vary
among programs. Refer to the terminal emulation program’s instructions for specific
steps.
2. Power on the monitoring system and set the serial port protocol to PC LINK 1 or PC LINK
2, as described in the monitoring system's service manual.
• Data bits: 8
• Parity: None
• Stop bits: 1
5. When you are ready to transmit data from the monitoring system, open the terminal
emulation program.
6. In the terminal emulation program, begin the transfer and text capture. Specify a file
name and directory to save the file.
7. When monitoring is complete, end the transfer, disconnect, and close the terminal
emulation program.
Downloaded case history data can be accessed and graphed on a computer using
a commercial spreadsheet program.
Case histories downloaded to a USB flash drive are stored in individual files named
as follows (if not assigned a custom name):
CaseID__MonitorSerialNumber.H3
CaseID is the date and time the case started (YYYYMMDD_HHMM in 24-hour format).
For example, 20171023_0841.
An example file name is:
20171023_0841__GBA12P3023.H3
Downloaded case history data is stored in ASCII format as a single line as shown
below. Each field is separated by two space characters. Data is transmitted on a per
channel basis (Channel 1 through Channel 4). See Table 5-3, page 5-10 for field
descriptions.
Column: A B C D E F G
Column (continued): H I J K
Column (continued): L M N O
Column (continued): P Q R S
Field Description
During a case, data is transmitted in ASCII text-based streams in either of two user-
selectable formats: PC LINK 1 and PC LINK 2.
Data is transmitted as a single line of text, approximately once per second, as shown
below. Each field is separated by two space characters. Data is transmitted on a per
channel basis (Channel 1 through Channel 4). If a particular channel is not active, a
zero is sent for all the fields corresponding to that channel. See Table 5-4, page 5-12
for field descriptions.
Case data can be accessed and graphed on a computer using a commercial spread-
sheet program.
Description: Version Date Time Ch label rSO2 Event Status Baseline AUC UAL LAL A B Rsvd
Column (continued): O P Q R S T U V W X Y
Description (continued): Ch label rSO2 Event Status Baseline AUC UAL LAL A B Rsvd
Column (continued): Z AA AB AC AD AE AF AG AH AI AJ
Description (continued): Ch label rSO2 Event Status Baseline AUC UAL LAL A B Rsvd
Column (continued): AK AL AM AN AO AP AQ AR AS AT AU
Description (continued): Ch label rSO2 Event Status Baseline AUC UAL LAL A B Rsvd
Column (continued): AV AW AX AY
Column: A B C D E F G H I
Column (continued): J K L M N O P
Column (continued): Q R S T U V W
Column (continued): X Y Z AA AB AC AD
Column (continued): AE AF AG AH
Table 5-4. Case Data Downloads (PC LINK 1 and PC LINK 2) - Data Fields
Ch label Range: A to Z, S1 to S4
0 = Channel not active
Event An event that was marked between the last transmission and
this transmission. (Integer)
Range: 0 to 160, 252 to 254
See Event Codes for Data Downloads, page 5-13 for descrip-
tions.
0 = No event
Table 5-4. Case Data Downloads (PC LINK 1 and PC LINK 2) - Data Fields (Continued)
The following event codes apply to case histories downloaded from the monitoring
system via USB and case data transmitted via the monitoring system’s serial port.
3 Induction 23 Cardioversion
6 On CPB 26 ECLS On
10 Warming 30 Inotrope
44 Extubated 64 Diuretic
45 Intubated 65 ECLS On
50 Bradycardia 70 FFP/Platelets
58 Anti-Arrhythmic 78 Sedation
59 Anti-Epileptic 79 Vasopressor
The following status codes apply to case histories downloaded from the monitoring
system via USB and case data transmitted via the monitoring system’s serial port.
2 CHECK SENSOR
4 SYSTEM SIGNAL OK
5 HIGH rSO2
6 LOW rSO2
17 REPLACE SENSOR
19 INTERFERENCE DETECTED
Note:
The following alarms are not reported in downloaded data:
• BATTERY CRITICALLY LOW
• BATTERY FAILURE
• BATTERY LOW
• PREAMP FAILURE
• SYSTEM FAILURE
6.1 Overview
This chapter contains information about optimizing the performance of the INVOS™
patient monitor (the “monitoring system”).
• Safety Reminders, page 6-1
WARNING:
The use of accessories, sensors, and cables other than those specified may result in
poor performance of the monitoring system and increased electromagnetic
emissions or decreased electromagnetic immunity of the monitoring system.
WARNING:
Monitoring system readings can be affected by certain patient conditions. Refer to
Patient Conditions, page 6-2.
6-1
Performance Considerations
• Hemoglobinopathies
WARNING:
Electromagnetic emissions from the monitoring system may interfere with other
critical devices.
WARNING:
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the monitoring system, including cables. Otherwise, degradation of
monitoring system performance may result.
WARNING:
The use of accessories, sensors, and cables other than those specified may result in
poor performance of the monitoring system and increased electromagnetic
emissions or decreased electromagnetic immunity of the monitoring system.
WARNING:
The monitoring system is intended for use by healthcare professionals only. It may
cause radio interference or may disrupt the operation of nearby equipment.
Mitigation for such disruption may require re-orienting or relocating the monitoring
system or shielding the location.
WARNING:
Any radio frequency transmitting equipment or other nearby sources of electrical
noise may result in disruption of the monitoring system.
WARNING:
The monitoring system is designed for use in environments in which the signal can
be obscured by electromagnetic interference. During such interference,
measurements may seem inappropriate or the monitoring system may not seem to
operate correctly.
WARNING:
EMI disruption can cause cessation of operation or other incorrect functioning.
WARNING:
The monitoring system should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, observe the monitoring system to
verify normal operation in the desired configuration. Technical alarms may indicate
that the configuration is not appropriate for the monitoring system.
Caution:
This device has been tested and found to comply with the limits for medical devices
related to IEC 60601-1-2: 2007 and IEC 60601-1-2:2014. These limits are designed to
provide reasonable protection against harmful interference in a typical medical
installation.
Caution:
When operating medical electrical equipment, special precautions related to
electromagnetic compatibility (EMC) are required. Install the monitoring system
according to the EMC information included in this manual.
Caution:
The monitoring system generates, uses, and can radiate radio frequency energy and,
if not installed and used in accordance with the instructions, may cause harmful
interference to other devices in the vicinity. If interference is suspected, move
monitoring system cables away from the susceptible device.
Caution:
The use of an electrosurgical or electrocautery instrument in the vicinity of the
monitoring system may interfere with the signal and cause poor performance or no
readings at all.
Because of the proliferation of radio frequency transmitting equipment and other
sources of electrical noise in health care environments (for example, electrosurgical
units, cellular phones, mobile two-way radios, electrical appliances, and high-defini-
tion television), it is possible that high levels of such interference due to close prox-
imity or strength of a source may result in disruption of monitoring system
performance. Reference Manufacturer’s Declaration, p. 11-7.
Disruption may be evidenced by cessation of operation or other incorrect function-
ing. If this occurs, survey the site of use to determine the source of this disruption,
then take the appropriate actions to eliminate the source.
• Turn equipment in the vicinity off and on to isolate the interfering equipment.
• Increase the separation between the interfering equipment and the monitoring system.
• Connect the monitoring system to an outlet on a different circuit from the other
device(s).
• Have an authorized technician check the line frequency setting for the monitoring
system. The setting should match the AC power line input. Refer to the monitoring sys-
tem's service manual.
The monitoring system generates, uses, and can radiate radio frequency energy and,
if not installed and used in accordance with these instructions, may cause harmful
interference with other susceptible devices in the vicinity. Contact Technical Ser-
vices for assistance. Reference Obtaining Technical Assistance, p. 1-7.
7.1 Overview
This chapter describes the steps required to properly clean the INVOS™ patient
monitor (the “monitoring system”). It also provides information about periodic safety
checks, service, software and firmware upgrades, and component disposal.
• Safety Reminders, page 7-1
WARNING:
Inspect the monitoring system and all accessories before use to ensure there are no
signs of physical damage or improper function. Do not use if damaged.
WARNING:
Explosion hazard — When replacing the battery, do not use the incorrect type. Use
only the battery available from Medtronic. See Accessories/Parts List, page 9-2.
WARNING:
To ensure proper performance, avoid shock, and prevent device damage or failure,
do not expose the monitoring system to extreme moisture, such as direct exposure
to rain. Do not immerse in water, solvents, or cleaning solutions, since the
monitoring system and connectors are not waterproof.
7-1
Product Maintenance
Caution:
Do not autoclave or gas sterilize any components of the monitoring system.
Caution:
To prevent device damage or failure, do not expose the monitor to isopropyl alcohol.
Caution:
Dispose of the battery in accordance with local guidelines and regulations.
WARNING:
To ensure proper performance, avoid shock, and prevent device damage or failure,
do not expose the monitoring system to extreme moisture, such as direct exposure
to rain. Do not immerse in water, solvents, or cleaning solutions, since the
monitoring system and connectors are not waterproof.
Caution:
Do not autoclave or gas sterilize any components of the monitoring system.
Caution:
To prevent device damage or failure, do not expose the monitor to isopropyl alcohol.
7.3.1 Materials
• Lint-free cloths
• Water
• Sodium hypochlorite (8.25% household bleach diluted 1:500 with tap water)
• Phenolic germicidal detergent (Lysol™* concentrate diluted 1:100 with tap water)
7.3.2 Procedure
Note:
Follow recommended cleaning procedures for your institution.
To clean the monitoring system:
1. Power off the monitoring system and disconnect AC power.
2. Clean the outside surface of all monitoring system components with a cloth dampened
with a cleaning agent or a pre-moistened wipe.
Each power-up Confirm speaker operation. The POST pass tone is an audible confirmation of
proper speaker performance. If the speaker does not function, alarm warning
sounds will not be audible. See Apply Power, page 3-13.
Every 3 months If the monitoring system is not in use, apply power and charge the battery.
For long-term storage, Medtronic recommends removing the battery pack
from the monitor.
Annually Check the INVOS™ system. The monitoring system's service manual provides
annual inspection procedures to be followed by an authorized technician.
Annually Inspect all safety relevant labels for legibility. Contact Medtronic or a local
Medtronic representative, if labels are damaged or illegible.
Annually Replace the reusable sensor cables (RSCs). See Accessories/Parts List, page 9-2
for part number and ordering information.
Every 2 years Replace the monitor’s battery. See Accessories/Parts List, page 9-2 for part
number and ordering information.
After approximately Replace the monitor. See Service Life, page 7-4 for more information.
25,000 hours of operation
• The reusable sensor cables have a service life of approximately one year.
• The monitor’s LCD panel (screen) has a service life of approximately 25,000 hours of
operation before the brightness is reduced to approximately 50% of initial brightness.
To avoid difficulty reading the screen, Medtronic recommends replacing the monitor at
approximately 25,000 hours of operation.
Caution:
Dispose of the battery in accordance with local guidelines and regulations.
Follow local government ordinances and recycling instructions regarding disposal
or recycling of the monitoring system and its components, including its battery and
accessories.
8.1 Overview
This chapter describes alarms generated by the INVOS™ patient monitor (the “mon-
itoring system”) and explains how to troubleshoot other errors that might occur.
• Alarm Messages, page 8-2
8-1
Alarms and Troubleshooting
• HIGH RSO2
BATTERY Battery is malfunction- Low Visual No Make sure that the battery is
FAILURE ing. only present and that the latch is in the
locked position. Try removing
and reinstalling the battery. See
Install the Battery, page 3-7.
If the battery is in place and
locked, discontinue use of the
monitor.
BATTERY LOW System is using battery Medium Audio Yes Connect to AC power as soon as
power AND battery and visual possible.
level is Low (<33% If the battery does not charge,
charge remaining). contact Medtronic or a local
Medtronic representative.
CHECK SENSOR Sensor is not attached Medium Audio No Make sure the sensor is securely
to patient. and visual adhered to the patient. If neces-
OR sary, replace the sensor.
Excessive light is detect- Under high ambient light condi-
ed on a sensor channel. tions, loosely cover the sensor
with opaque material.
OR
Identify any possible sources of
Electrical interference.
interference, and take the actions
OR recommended in EMI (Electro-
Moisture in connector. magnetic Interference), page 6-2.
OR Try another hospital-grade elec-
Sensor applied over trical outlet.
nevi, sinus cavities, the Have an authorized technician
superior sagittal sinus, check the line frequency setting
subdural or epidural for the monitoring system. The
hematomas, arteriove- setting should match the AC
nous malformations, power line input. Refer to the
broken skin, thick fatty monitoring system's service
deposits, hair, bony pro- manual.
tuberances, or areas Ensure sensor and RSC connec-
contaminated with resi- tors are moisture-free.
due.
Check sensor site. Refer to the
sensor Instructions for Use for site
selection information.
If the problem persists, contact
Medtronic or a local Medtronic
representative.
HIGH RSO2 Upper rSO2 alarm limit is Medium Audio No Immediately check the patient.
exceeded. and visual
INTERFERENCE Interference on a sensor Medium Audio No Move or turn off the source of
DETECTED channel is corrupting and visual interference. Normal operation
the rSO2 data. will resume when the excessive
noise stops. See EMI (Electromag-
netic Interference), page 6-2.
Have an authorized technician
check the line frequency setting
for the monitoring system. The
setting should match the AC
power line input. Refer to the
monitoring system's service
manual.
LOW RSO2 Lower rSO2 alarm limit is Medium Audio No Immediately check the patient.
exceeded. and visual
POOR SIGNAL Poor signal quality is Medium Audio No Identify any possible sources of
QUALITY detected on a sensor and visual interference, and take the actions
channel. rSO2 values will recommended in EMI (Electro-
not be displayed due to: magnetic Interference), page 6-2.
Electrical interference. Try another hospital-grade elec-
trical outlet.
OR
Under high ambient light condi-
Incorrect line frequency tions, loosely cover the sensor
setting. with opaque material.
OR Have an authorized technician
Moisture in connector. check the line frequency setting
OR for the monitoring system. The
setting should match the AC
Sensor applied over
power line input. Refer to the
nevi, sinus cavities, the
monitoring system's service
superior sagittal sinus,
manual.
subdural or epidural
hematomas, arteriove- Ensure sensor and RSC connec-
nous malformations, tors are moisture-free.
broken skin, thick fatty Check sensor site. Refer to the
deposits, hair, bony pro- sensor Instructions for Use for site
tuberances, or areas selection information.
contaminated with resi- Replace sensor.
due.
Replace RSC.
OR
If the problem persists, contact
Defective sensor. Medtronic or a local Medtronic
OR representative.
Defective RSC.
PREAMP Preamplifier is not com- Medium Audio1 No If you are using two preamplifiers,
FAILURE municating with the and visual determine which preamplifier is
system as expected. generating the alarm by touching
OR each sensor label on the monitor-
ing screen and noting whether
Preamplifier is generat-
the corresponding LED on the
ing unrecoverable
preamplifier flashes. If an LED
errors.
does not flash on a preamplifier,
disconnect the preamplifier cable
from the monitor and reconnect
it. See Connect the Preamplifier(s),
page 3-11.
If the problem persists, replace
the preamplifier. Contact
Medtronic or a local Medtronic
representative.
PREAMP NOT Preamplifier becomes Medium Audio Yes Make sure the preamplifier con-
CONNECTED disconnected from the and visual nector is locked into the monitor.
monitor during use. See Connect the Preamplifier(s),
page 3-11. Normal operation will
resume when the preamplifier is
reconnected.
REPLACE Sensor incompatible Medium Audio No Make sure that all sensors match
SENSOR OR and visual and are compatible with the
monitoring system. See the
Invalid calibration data
Instructions for Use provided with
OR the sensors.
Bad sensor Replace the sensor(s) if necessary.
OR Replace the RSC(s) if necessary.
Open/short condition
OR
Unable to access sensor
memory
SENSOR NOT Sensor is disconnected Medium Audio Yes Check the connection between
CONNECTED from the RSC. and visual the sensor and RSC and between
OR the RSC and preamplifier. If the
problem persists, replace the RSC
RSC is disconnected
and/or the sensor.
from the preamplifier.
SYSTEM ERROR Sensor is connected to a Medium Audio1 No Power the monitoring system off
patient, but the rSO2 and visual and back on.
data display has not If the problem persists, contact
been updated for over Medtronic or a local Medtronic
30 seconds. representative.
The system may indi-
cate that settings are
corrupted.
SYSTEM Monitoring system was Medium Audio1 Yes Power the monitoring system off
FAILURE reset unexpectedly. and visual and back on.
Note: All settings may If the problem persists, contact
be lost. Medtronic or a local Medtronic
representative.
TREND DATA Software was upgraded. Low Audio Yes If the problem persists, contact
LOST OR and visual Medtronic or a local Medtronic
representative.
Time setting was
changed.
OR
Corrupt trend data was
detected at start up.
1. Sounds even when alarms are silenced or paused.
Problem Resolution
No AC power to monitoring system • Make sure that the monitoring system’s power supply is
when plugged in and powered on plugged into a hospital-grade power outlet.
(runs on battery power only)
• Make sure that the power cord is securely connected to the
power supply and that the power supply is securely connected
to the docking station.
Does not power up when power • Press the power button for more than 3 seconds.
button is pressed
• If operating on AC power, refer to the resolutions above.
System indicates a preamplifier has • Check the preamplifier connection to the monitor. See Connect
been disconnected when it is still the Preamplifier(s), page 3-11.
physically connected
• Try a different preamplifier.
Error message:
PREAMP NOT CONNECTED • Contact Medtronic or a local Medtronic representative.
System does not recognize and • Check the preamplifier connection to the monitor. See Connect
gives no indication a preamplifier the Preamplifier(s), page 3-11.
has been connected after it has
been physically connected • Try a different preamplifier.
Problem Resolution
Does not power off when power • If the monitoring system does not power off within approxi-
button is pressed mately 30 seconds after you press the Power button, press and
hold the Power button for at least 10 seconds to power off the
monitoring system.
Electrical noise on AC power line • Have an authorized technician check the line frequency setting
causing poor signal quality for the monitoring system. The setting should match the AC
power line input. Refer to the monitoring system's service
manual.
Multiple, consecutive sensor alarms If sensor alarms continue to occur after you have performed the rec-
ommended corrections for the alarms (see Table 8-1, page 8-2),
replace the following:
• RSCs
• Preamplifiers
Frequent battery alarms • Replace the battery. See Install the Battery, page 3-7.
Touch screen is unresponsive • Power the monitoring system off and back on.
Problem Resolution
A USB error occurs during any of the • Follow the resolution indicated in the error message.
following situations:
• If a USB flash drive is present but isn’t recognized by the system,
• Exporting case histories
make sure that the drive has been formatted. If necessary, you
• Exporting logs can format the drive on a computer running Microsoft Win-
dows™*.
• Exporting or importing system
settings
• INSUFFICIENT STORAGE
SPACE. INSERT NEW USB
DRIVE AND TRY AGAIN.
• NO FIRMWARE UPDATE
FOUND. CHECK USB DRIVE
AND TRY AGAIN.
• NO SOFTWARE UPDATE
FOUND. CHECK USB DRIVE
AND TRY AGAIN.
Problem Resolution
Unable to consistently display • Make sure that the monitor is completely seated in the docking
monitoring system screens to an station. See Insert the Monitor into the Docking Station, page 3-9.
external monitor
• Disconnect the VGA cable from the docking station and external
monitor. Reconnect the cable, making sure it is completely
seated at both connections.
• Verify that the power cord is fully plugged into the monitor’s
docking station.
No data received or data scrambled • Make sure that the monitor is completely seated in the docking
during serial port transmissions station. See Insert the Monitor into the Docking Station, page 3-9.
• Disconnect the serial cable from the docking station and exter-
nal device. Reconnect the cable, making sure it is completely
seated at both connections.
• Verify that the power cord is fully plugged into the monitor’s
docking station.
• If transmitting data, verify that the serial cable meets the require-
ments described in Serial Port Specifications, page 5-3 and Null
Modem Cable Requirements, page 5-6.
9.1 Overview
This chapter contains information for selecting the appropriate rSO2 sensor and
other accessories for use with the INVOS™ patient monitor (the “monitoring sys-
tem”).
• Safety Reminder, page 9-1
WARNING:
The use of accessories, sensors, and cables other than those specified may result in
poor performance of the monitoring system and increased electromagnetic
emissions or decreased electromagnetic immunity of the monitoring system.
9-1
Accessories
VLI INVOS™ rSO2 interface cable (VueLink™* adapter cable, DB25F to DB9F)
10.1 Overview
This chapter explains the theory behind operations of the INVOS™ patient monitor
(the “monitoring system”).
10-1
Theory of Operations
11.1 Overview
This chapter contains physical and operational specifications for the
INVOS™ patient monitor (the “monitoring system”).
• Physical Characteristics, page 11-2
11-1
Product Specifications
Monitor
Docking Station
Type VESA™*
Indicator Power
Ports DC In, USB 2.0 (x2), serial (RS-232), and VGA ports
Stand
Preamplifier
Dimensions 12.8 x 8.7 x 2.8 cm (5.04 x 3.43 x 1.1 in.) with hook folded down
Power cable
11.3 Electrical
Input frequency 50 Hz to 60 Hz
11.4 Battery
Type Lithium-ion
Recharge 24 hours
Note:
The system may not meet its performance specifications if stored or used outside the
specified temperature and humidity range.
Alarms
Low 2 Pulse 1: 660 186.9 Rise: 19.2 Between pulse 1 and 2: >4
priority Pulse 2: 525 Fall: 12 188.1
alarm
Medium priority 21 dB 33 dB 38 dB
Low priority 19 dB 29 dB 35 dB
Repetitions No repeat
Operating Range
Degree of safety Not suitable for use in the presence of flammable anes-
thetics
WARNING:
Electromagnetic emissions from the monitoring system may interfere with other
critical devices.
WARNING:
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the monitoring system, including cables. Otherwise, degradation of
monitoring system performance may result.
WARNING:
The use of accessories, sensors, and cables other than those specified may result in
poor performance of the monitoring system and increased electromagnetic
emissions or decreased electromagnetic immunity of the monitoring system.
WARNING:
The monitoring system is intended for use by healthcare professionals only. It may
cause radio interference or may disrupt the operation of nearby equipment.
Mitigation for such disruption may require re-orienting or relocating the monitoring
system or shielding the location.
WARNING:
Any radio frequency transmitting equipment or other nearby sources of electrical
noise may result in disruption of the monitoring system.
WARNING:
The monitoring system is designed for use in environments in which the signal can
be obscured by electromagnetic interference. During such interference,
measurements may seem inappropriate or the monitoring system may not seem to
operate correctly.
WARNING:
EMI disruption can cause cessation of operation or other incorrect functioning.
WARNING:
The monitoring system should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, observe the monitoring system to
verify normal operation in the desired configuration. Technical alarms may indicate
that the configuration is not appropriate for the monitoring system.
Caution:
This device has been tested and found to comply with the limits for medical devices
related to IEC 60601-1-2: 2007 and IEC 60601-1-2:2014. These limits are designed to
provide reasonable protection against harmful interference in a typical medical
installation.
Caution:
When operating medical electrical equipment, special precautions related to
electromagnetic compatibility (EMC) are required. Install the monitoring system
according to the EMC information included in this manual.
Caution:
The monitoring system generates, uses, and can radiate radio frequency energy and,
if not installed and used in accordance with the instructions, may cause harmful
interference to other devices in the vicinity. If interference is suspected, move
monitoring system cables away from the susceptible device.
Caution:
The use of an electrosurgical or electrocautery instrument in the vicinity of the
monitoring system may interfere with the signal and cause poor performance or no
readings at all.
The monitoring system is suitable for prescription use only in the specified electro-
magnetic environments, in accordance with the IEC/EN 60601-1-2:2007 and IEC/EN
60601-1-2:2014 standards. The monitoring system requires special precautions
during installation and operation for electromagnetic compatibility. In particular, the
use of nearby mobile or portable communications equipment may influence mon-
itoring system performance.
Note:
The emissions characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11
class B is normally required) this equipment might not offer adequate protection to radio-
frequency communication services. The user might need to take mitigation measures, such
as relocating or re-orienting the equipment.
The monitoring system is intended for use in the electromagnetic environment specified below. The
customer or the user of the monitoring system should assure that it is used in such an environment.
Radiated emission CISPR 11/EN The monitoring system uses RF energy only for its internal
55011, Class A, function. Therefore, its RF emissions are very low and are not
Group 1 likely to cause any interference in nearby electronic equip-
ment.
Conducted emission CISPR 11/EN The monitoring system is suitable for use in all establish-
55011, Class A, ments other than domestic and those directly connected to
Group 1 the public low-voltage power network that supplies build-
ings used for domestic purposes.
Harmonic emission IEC/EN 61000-3-2,
Class A
The monitoring system is intended for use in the electromagnetic environment specified below. The
customer or the user of the monitoring system should assure that it is used in such an environment.
Electric fast ± 2kV (100 kHz repeti- ± 2kV (100 kHz repeti- Mains power quality should be that
transient/burst tion rate) AC mains tion rate) AC mains of a typical commercial or hospital
IEC/EN 61000-4-4 ± 1kV (100 kHz repeti- ± 1kV (100 kHz repeti- environment.
tion rate) input/ tion rate) input/
output >3 m output >3 m
Surge ± 1kV line-line, AC ± 1kV line-line, AC Mains power quality should be that
IEC/EN 61000-4-5 mains mains of a typical commercial or hospital
± 2kV line-ground, AC ± 2kV line-ground, AC environment.
mains mains
Voltage dips and 100% reduction for 0.5 100% reduction for 0.5 Mains power quality should be
interrupts cycles (at 0°, 45°, 90°, cycles (at 0°, 45°, 90°, that of a typical commercial or
IEC/EN 61000-4-11 135°, 180°, 225°, 270°, 135°, 180°, 225°, 270°, hospital environment.
and 315°) and 315°) If the user requires continued
100% reduction for 1.0 100% reduction for 1.0 operation during power mains
cycle (at 0°) cycle (at 0°) interruptions, it is recommended
30% reduction for 25/ 30% reduction for 25/ that the monitoring system be
30 cycles (at 0°) 30 cycles (at 0°) powered from an uninterruptible
power supply or battery.
100% reduction for 100% reduction for
250/300 cycles (at 0°) 250/300 cycles (at 0°)
Power frequency 50 and 60 Hz, 30 A/m, 50 and 60 Hz, 30 A/m, Power frequency magnetic fields
H-field immunity x-, y-, and z-axes x-, y-, and z-axes should be at levels characteristic of
IEC/EN 61000-4-8 a typical location in a typical com-
mercial or hospital environment.
d = 1.2 P
80 MHz to 800 MHz
d = 2.3 P
800 MHz to 2.7 GHz
where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site surveya, should be less than the compliance
level in each frequency rangeb.
Interference may occur in the vicinity of equipment marked with
the following symbol:
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the mea-
sured field strength in the location in which the monitoring system is used exceeds the applicable RF compliance level above,
the monitoring system should be observed to verify normal operation. If abnormal performance is observed, additional mea-
sures may be necessary, such as re-orienting or relocating the monitoring system.
bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Table 11-10. Test Specifications for Enclosure Port Immunity to RF Wireless Communications Equipment
710 704 to 787 LTE Band 13, 17 Pulse modulation 0.2 0.3 9
217 Hz
745
780
5785
The monitoring system is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the monitoring system can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the monitoring system as recommended below,
according to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance
(d) in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufac-
turer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects, and people.
A.1 Overview
This appendix provides information on clinical studies using the INVOS™ regional saturation
patient monitoring system. A reference bibliography containing additional clinical studies is
available on Medtronic’s website (www.medtronic.com) or by contacting Medtronic or a local
Medtronic representative. See Technical Services, page 1-7.
An adult volunteer study was performed comparing the regional oxygen saturation (rSO2) from
the system to blood sample analysis in 20 volunteers (hypoxia study). No adverse events
attributable to the system were reported during the study.
• Hypoxia Study, page A-1
A-1
Clinical Studies
Results
Table A-1. Performance Characteristics: Accuracy Bias, Error (SD), RMSD, and Correlation (R2) for rSO2 and fSO2, 20 Subjects
One subject had a bias value of 37.2%, which corresponded with a very low signal quality index
number (SQI=1). The remaining 41 subjects had SQIs that started at 10 and remained greater than
four throughout the study.
Subject 206’s absolute bias data was excluded from the analysis due to the low signal quality, but
the trend data was left in as trending is not dependent on SQI.
Table A-2. Trend Bias, Error (SD), and Correlation (R2) for ΔrSO2 and ΔfSO2, 20 subjects
2. Roberts KW, Crnkowic AP, Linneman LJ: Near infrared spectroscopy detects critical cerebral hypoxia
during carotid endarterectomy in awake patients. Anesthesiology 1998;89(3A):A934.
4. Murkin JM, Adams SJ, Novick RJ, Quantz M, Bainbridge D, Iglesias I, Cleland A, Schaefer B, Irwin B, Fox S.
Monitoring brain oxygen saturation during coronary bypass surgery: a randomized, prospective study.
Anesth Analg. 2007 Jan;104(1):51-8.
7. Edmonds HL Jr, Ganzel BL, Austin EH 3rd. Cerebral oximetry for cardiac and vascular surgery. Semin
Cardiothorac Vasc Anesth. 2004;8:147-66.
8. Goldman S, Sutter F, Ferdinand F, Trace C. Optimizing intraoperative cerebral oxygen delivery using
noninvasive cerebral oximetry decreases the incidence of stroke for cardiac surgical patients. Heart
Surgery Forum 2004;7(5):#2004-1062.
9. Murkin JM, Iglesias I, Bainbridge D, Adams S, Schaefer B, Irwin B, Fox S. Monitoring cerebral oxygen
saturation significantly decreases major organ morbidity in CABG patients: A randomized blinded
study. The Heart Surgery Forum 2004;7(6):515.
10. Yao FSF, Tseng CCA, Ho CYA, Levin SK, Illner P. Cerebral oxygen desaturation is associated with early
postoperative neuropsychological dysfunction in patients undergoing cardiac surgery. J Cardiothorac
Vasc Anesth 2004;18(5):552-558.
2. Kolb JC, Ainslie PN, Ide K, Poulin MJ. Effects of five consecutive nocturnal hypoxic exposures on the
cerebrovascular responses to acute hypoxia and hypercapnia in humans. J Appl Physiol. 2004
May;96(5):1745-54. Epub 2004 Jan 16.
Covidien llc
15 Hampshire Street, Mansfield, MA 02048 USA.
Covidien Ireland Limited, IDA Business & Technology Park, Tullamore, Ireland.
www.Medtronic.com