Zetasizer DLS
Zetasizer DLS
Zetasizer DLS
CHAPTER 14
Introduction
The aim of this chapter is to describe the basic Size principles behind the Zetasizer Nano series. This will help in understanding the meaning of the results achieved. The chapter is divided into two major sections. What is Dynamic light scattering? and Operation of the Zetasizer Nano - Size measurements. The first section describes the theory, while the second describes the physical operation of how a size measurement is performed.
The speckle pattern will consist of areas of bright light and dark areas where no light is detected.
Z e t a s i z e r N a n o S e r i e s
Page 14.1
ILL 6090
CHAPTER 14
Z e t a s i z e r
N a n o
S e r i e s
What causes these bright and dark areas? The diagram below shows the propagated waves from the light scattered by the particles. The bright areas of light are where the light scattered by the particles arrive at the screen with the same phase and interferes constructively to form a bright patch. The dark areas are where the phase additions are mutually destructive and cancel each other out.
From Laser
Detector
Average intensity
ILL 4022
In the above example we said that the particles are not moving. In this situation the speckle pattern will also be stationary - in terms of both speckle position and speckle size. In practice, particles suspended in a liquid are never stationary. The particles are constantly moving due to Brownian motion. Brownian motion is the movement of particles due to the random collision with the molecules of the liquid that surrounds the particle. An important feature of Brownian motion for DLS is that small particles move quickly and large particles move more slowly. The relationship between the size of a particle and its speed due to Brownian motion is defined in the Stokes-Einstein equation. As the particles are constantly in motion the speckle pattern will also appear to move. As the particles move around, the constructive and destructive phase addition of the scattered light will cause the bright and dark areas to grow and diminish in intensity - or to put it another way, the intensity appears to fluctuate. The Zetasizer Nano system measures the rate of the intensity fluctuation and then uses this to calculate the size of the particles.
Page 14.2
CHAPTER 14 Within the instrument is a component called a digital correlator. A correlator basically measures the degree of similarity between two signals over a period of time. If we compared the intensity signal of a particular part of the speckle pattern at one point in time (say time = t) to the intensity signal a very short time later (t+dt) we would see that the two signals are very similar - or strongly correlated. If we then compared the original signal a little further ahead in time (t+2dt), there would still be a relatively good comparison between the two signals, but it will not be as good as at t+ dt. The correlation is therefore reducing with time. Now consider the intensity of the signal at t, with the intensity at a much later time - the two signals will have no relation to each other as the particles are moving in random directions (due to Brownian motion). In this situation it is said that there is no correlation between the two signals. With DLS we are dealing with very small time scales. In a typical speckle pattern the length of time it takes for the correlation to reduce to zero is in the order of 1 to 10's of milliseconds. The "short time later" (dt) will be in the order of nanoseconds or microseconds! If we compare the signal intensity at (t) with itself then we would have perfect correlation as the signals are identical. Perfect correlation is reported as 1 and no correlation is reported as 0. If we continue to measure the correlation at (t+3dt), (t+4dt), (t+5dt), (t+6dt), etc, the correlation will eventually reach zero. A typical correlation function of correlation against time is shown below.
1.00
Correlation
0 t=0
Time
t=
Page 14.3
ILL 4022
CHAPTER 14
Z e t a s i z e r
N a n o
S e r i e s
related to size of the particles (Stokes-Einstein equation). Large particles move slowly, while smaller particles move quickly. What effect will this have on the speckle pattern? If large particles are being measured, then, as they are moving slowly, the intensity of the speckle pattern will also fluctuate slowly. And simarlarly if small particles are being measured then, as they are moving quickly, the intensity of the speckle pattern will also fluctuate quickly. The graph below shows the correlation function for large and small particles. As can be seen, the rate of decay for the correlation function is related to particle size as the rate of decay is much faster for small particles than it is for large.
1.00
Perfect Correlation
Correlation
0 t=0
Time
t=
After the correlation function has been measured this information can then be used to calculate the size distribution. The Zetasizer software uses algorithms to extract the decay rates for a number of size classes to produce a size distribution. A typical size distribution graph is shown below. The X axis shows a distribution of size classes, while the Y axis shows the relative intensity of the scattered light. This is therefore known as an intensity distribution.
Size distribution by Intensity
1.00
0.75
Amplitude
0.50
0.25
Diameter (nm)
Page 14.4
ILL 4022
CHAPTER 14 Although the fundamental size distribution generated by DLS is an intensity distribution, this can be converted, using Mie theory, to a volume distribution. This volume distribution can also be further converted to a number distribution. However, number distributions are of limited use as small errors in gathering data for the correlation function will lead to huge errors in distribution by number.
Volume
1000
Intensity
Relative % in class
1,000,000
Diameter (nm)
Diameter (nm)
Diameter (nm)
It is worth repeating that the basic distribution obtained from a DLS measurement is intensity - all other distributions are generated from this.
Page 14.5
ILL 4022
5 10
50 100
5 10
50 100
5 10
50 100
CHAPTER 14
Z e t a s i z e r
N a n o
S e r i e s
Attenuator
17
4 3
A
3
Detector
Detector
2
90 Cell
Correlator
The intensity of the scattered light must be within a specific range for the detector to successfully measure it. If too much light is detected then the detector will become overloaded. To overcome this an attenuator . is used to reduce the intensity of the laser and hence reduce the intensity of the scattering.
Page 14.6
ILL 6720
CHAPTER 14 . For samples that do not scatter much light, such as very small particles or samples of low concentration, the amount of scattered light must be increased. In this situation, the attenuator will allow more laser light through to the sample. . For samples that scatter more light, such as large particles or samples of higher concentration, the amount of scattered light must be decreased. This is achieved by using the attenuator to reduce the amount of laser light that passes through to the sample. The appropriate attenuator position is automatically determined by the Zetasizer during the measurement sequence. The scattering intensity signal for the detector is passed to a digital signal processing board called a correlator /. The correlator compares the scattering intensity at successive time intervals to derive the rate at which the intensity is varying. This correlator information is then passed to a computer 0, where the specialist Zetasizer software will analyse the data and derive size information. As mentioned earlier, depending upon the Zetasizer Nano model the detection optics will either be arranged at positions of either 173 or 90.
173
Laser Sample
ILL 6097
Why measure backscatter? There are several advantages to doing this: . Because the backscatter is being measured, the incident beam does not have to travel through the entire sample. This reduces an effect known as multiple scattering, where the scattered light from one particle is itself scattered by other particles. As the light passes through a shorter path length of the sample, then higher concentrations of sample can be measured. . Contaminants such as dust particles within the dispersant are typically large compared to the sample size. Large particles mainly scatter in the forward
Page 14.7
CHAPTER 14
Z e t a s i z e r
N a n o
S e r i e s
direction. Therefore, by measuring the backscatter, the effect of dust is greatly reduced. . The effect of multiple scattering is at a minimum at 180 - again, this allows higher concentrations to be measured.
Moveable lens
Within the Zetasizer Nano system, a movable lens allows the focus position within the cell to be changed. This allows a much larger range of sample concentrations to be measured.
For small particles, or samples of low concentration, it will be beneficial to maximise the amount of scattering from the sample. As the laser passes through the wall of the cell and into the dispersant, the cell wall will cause flare. This flare may swamp the scattering signal. Moving the measurement point away from the cell wall, towards the centre of the cell will remove this effect.
Sample
ILL 6098
Sample
Large particles or samples of high concentration, scatter much more light. In this situation, measuring closer to the cell wall will reduce the effect of multiple scattering. In this instance the flare from the cell wall will have less impact. Any flare will be proportionally reduced compared to the scattering signal.
Page 14.8
ILL 6099