Clinical Electrophysiology
Electroretinogram (ERG): Flash, pattern, multifocal, bright flash Electro-oculogram (EOG)
Visually Evoked Cortical Potentials (VECP): pattern, flash, multifocal, multichannel
History
1877-Year of birth of Human ERG (First record by Dewar)
1945 -Karpe reported 64 normal and 87 abnormal ERG results and introduced routine ERG in his clinic
Electroretinogram (ERG)
Recording of the bioelectric potential and the summed activity of the retinal cells, when stimulated by light
Physiology of ERG
Light induced changes in the trans membrane movements of ions in retinal cells, mainly sodium and potassium, making the cells hyperpolarized i.e. more negative to the extracellular space than in the dark
b-wave -Corneal positive - Muller cells due to lightinduced changes of extra cellular potassium
Implicit time (i.t.)= time from stimulus until peak of activity Amplitude (amp)= voltage magnitude at peak of activity
b amp a amp
a i.t.
a-wave
Scotopic
20-26 m sec
A-wave amplitude
B-wave amplitude
Ganzfeld stimulator
*Brigell M et al. Guidelines for calibration - -Doc. Ophthalmol 1998; 95: 1-4
blue (470 nm) green (525 nm) red (640 nm) 7 log unit luminance range Optoelectronic stimulation can be more precisely controlled and calibrated than xenon discharge tubes
ELECTRODES
Active Electrodes are placed in contact with the cornea or nearby bulbar conjunctiva. Corneal electrodes give higher amplitude than noncorneal electrodes Corneal Burian -Allen Electrode, Jet Electrodes
Non corneal- DTL, Gold foil, Conjunctival loop and LVP Electrode
Corneal Electrodes
Jet
Burian-Allen
ERG-Jet Electrodes
Disposable mono-polar contact lens electrode. Reduced possibility of nosocomial eye infections. Gold foil on polymethylmethacrylate with lead wire. Ease of insertion Disadvantage -movement of lens during recording and absence of lid speculum contribute to variability of recordings
Burian-Allen Electrodes
Bipolar, reusable contact lens electrode with built-in lid speculum and Reference electrodes
LVP-Zari electrodes
DTL (Dawson-Trick-Litzkow) LVP-Zari electrodes
Additional Electrode
Ear-clip electrode
Reference electrode is unipolar electrodes-in outer canthus Ground electrode-- Ear clip electrode
Instrumental: amplification, gain, stimulus, electrodes. Artifacts: Blinking, tearing, eye movements, air bubbles
under electrode.
Types of ERG
Single flash ERG
response)
Oscillatory potentials
ERG to a strong flash (arising from the cones) in the light-adapted eye (Single flash cone response) ERGs to a rapidly repeated stimulus (30 Hz flicker)
1989 a Basic protocol 1999 updated
A sharp a wave and a much larger, rapidly rising peaked b-wave which comes to baseline very slowly
Oscillatory potentials
Originate in middle/inner retina Higher frequency oscillations (100-160Hz) Scotopic condition , Standard Flash (0 dB-white) Higher frequency filter (75 to 300 Hz) 15 seconds interval Flashes
30 Hz flicker responses
Standard Flash 30 stimuli per second suppress the rod system The normal responses -the peak falls just before the stimulus onset for the next response
ROD RESPONSE
RODS
RODS + CONES
OSCILLATORY POTENTIALS
CONES
30 Hz FLICKER
CONES
ROD RESPONSE
Dark adaptation
Dim stimulus Blue stimulus
b wave
amplitude microvoltage
176.47281.87 310.04545.28
COMBINED REPSONSE
12.3315.87
29.36-62.36
27.2332.09
28.8330.00
121.40224.08
93.16-170.12
Pattern ERG
N35 photoreceptors P50 muller + bipolar cells N95 ganglion cells Loss of N95 in optic nerve disorders Ratio of P50 /N95 Normal < 2 Abnormal > 2
PERG - Application
1. Glaucoma
2. Ocular hypertension
3. Optic nerve disease
FERG - Applications
Unexplained visual loss Idiopathic macular hole - fellow eye Idiopathic central serous chorioretinopathy
Allows topographic mapping of retinal function in the central of the retina. Sensitive test for quantifying the visual function of maculopathies.
Standard ERG - uses fairly large stimuli to test large retinal areas at a time, focal damages cannot be detected this way
Instead of a single stimulus and a single response, retina divide into several areas each which are stimulated with a pseudo random probability seq ( M SEQUENCE) Cover up to +/- 25 degrees of the visual field Underlying mathematical system derives the true signal from each location and form retinal map of ERG
M sequence
Multifocal ERG
N1 - contributions from the same components as the awave of full field cone ERG
P1 - contribution from the components of the cone bwave and OPs
P1
N1
N2
20
arteries veins
APPLICATIONS OF MFERG
MACULOPATHIESCentral responses are lost or reduced ,surrounded by normal response.eg. ARMD, macular dystrophy ,CSR - crater or volcano like appearance in 3-D plot of response density VASCULAR DISORDERSoverall decreases in amplitude and delay in implicit time DISORDERS OF GANGLION CELLS AND VISUAL PATHWAY
Retinitis Pigmentosa
Diagnostic Prognostic
Detection of carriers
Abnormal even in early stages Reduced scotopic & photopic b-wave Delayed implicit time Av 16 18% loss of ERG amplitudes per year
Fundus albipunctatus
OGUCHIS DISEASE
Early H/o defective night vision Vision and peripheral fields are generaly normal in daylight and manifest defect is seen in dim illumination.
OGUCHIS DISEASE
FUNDUS Light adapted - a yellow gray sheen from optic disc to equator. Dark adapted after 2 to 3 hours, fundus appears normal (MIZUO PHENOMENON)
Oguchis disease
Scotopic b wave is diminished in amplitude, producing a negative ERG pattern. OPS normal or reduced. Photopic functions are normal
Fundus Albipunctatus
CONE DYSTROPHY
The rod specific ERG and the cone derived 30 Hz flicker and single flash photopic ERGs are undetectable.
Ischemic CRVO
TOXIC STATES
SIDEROSIS :supranormal in early stage and subnormal in advanced stage
CHALCOSIS CHLOROQUINE / QUININE TOXICITY LEAD PHENOTHIAZINES
GLAUCOMA
PERG is the test of choice Reflects ganglion cell dysfunction not detected by perimetry Can help in decision regarding treatment of ocular hypertension Outer retinal damage to rods and blue cones detected by dark adapted flash/light adapted flicker ERG
NEGATIVE ERG
a-wave is larger than the b-wave in scotopic maximal response
Diseases sparing the photoreceptors and involving the middle retinal layers
Congenital stationary blindness X linked retinoshisis Central retinal artery obstruction Quinine toxicity Melanoma associated retinopathy Batten disease Cone Rod dystrophy
Thank you!
Excellenc
Equity
Efficiency
Electroretinography (ERG) is an objective method of evaluating retinal function. Since its introduction to clinical practice in the 1940s, it has become a useful and routine diagnostic clinical tool in ophthalmology. This review summarizes the role of ERG as a clinical technique for evaluating the progression of diabetic retinopathy and as a research tool for increasing our understanding of the pathophysiology of diabetic retinopathy
. Most studies show unequivocally that the different types of ERG tests detect local abnormalities or widespread pathology, even in very early stages of the disease. It seems plausible that measurements from ERG recordings, particularly the oscillatory potentials, may be useful for predicting progression from nonproliferative to the more sight-threatening stagespreproliferative or proliferativeof diabetic retinopathy. Some recent work implies that the ERG can also be a useful diagnostic method for discriminating between eyes with diabetic retinopathy and those without the condition. (
MILD AND MODERATE NONPROLIFERATIVE DIABETIC RETINOPATHY Virtually all studies using OPs report distinctive
changes in cases with mild NPRD (Table 1). Recently, more detailed studies have revealed that several ERG parameters can be affected in cases of mild NPDR: scotopic b-wave implicit times, 30-Hz flicker amplitude and implicit times, OP amplitudes to both blue and white flashes (including selective reduction of OP3 and OP2), OP implicit times, and photopic ERG implicit times.
Oscillatory potential amplitudes correlate well with the severity of diabetic retinopathy. The few studies published to date that include patients with severe NPDR showed significantly reduced and sometimes absent Ops.
in all ERG parameters, including decrease of the OP amplitude (up to complete disappearance of the OPs) or delay of the implicit time of the OPs (Table 1), decrease of photopic and scotopic a- and b-waves, and 30-Hz flicker delay.
EARLY AND HIGH-RISK PROLIFERATIVE DIABETIC RETINOPATHY Significant abnormalities have been reported
Other changes
The amplitudes of N1 and P1 in 1 2 rings were decreased dramatically in patients with CSME.
The dramatic decrease of visual function was shown by the reduced visual acuity, subjectively, and the decreased average amplitudes of mfERG, objectively. It is suggested that the slight damage of outer retina may cause the decreased amplitude and that the more severe damage of the full-thickness retina could lead to further decrease in the amplitude.
The Multi focal-ERG reveals local retinal dysfunction in diabetic eyes even before retinopathy. The magnitude of delay of local ERG implicit time reflects the degree of local clinical abnormality in eyes with retinopathy. Local response delays found in some eyes without retinopathy suggest that the Multifocal-ERG detects subclinical local retinal dysfunction in diabetes. Analysis of Multifocal-ERG implicit time, independent of amplitude, improves the sensitivity of detection of local retinal dysfunction in diabetes.
Brad Fortune et al Investigative Ophthalmology and Visual Science. 1999
Palmowski et al. demonstrated that in some patients with diabetes, M-ERG responses (averaged across relatively large areas of retina) were smaller in amplitude and delayed in comparison with those in normal subjects. However, they did not determine the extent to which local abnormalities were detected (versus abnormalities present throughout the retina).
Review of literature
The implicit times of P1 and OPs were significantly delayed in the diabetics in the peripheral tested regions (2040) but not in the central area. After PRP, the amplitudes of P1 were markedly reduced in all areas tested; however, the changes of the P1 implicit time were not significant. The amplitudes of the OPs were significantly reduced in the peripheral regions after PRP; however, the changes in the implicit times were not significant. Retinal function in the posterior pole is markedly impaired in eyes with pre-PDR. PRP altered the mfERGs significantly, however, the reductions were limited to the amplitude.
Onozu H. et al Documenta Ohthalmologica , Volume 106, Number 3, May 2003
Full-field ERG,multifocal ERG and multifocal VEP in patients with retinitis pigmentosa and residual central visual fields Purpose: To evaluate (with three different electrophysiological methods) the residual retinal function in a selected group of patients with retinitis pigmentosa and remaining small central visual fields . Methods: Fourteen patients from several different genetic subgroups, who had been followed with visual acuity and visual field testing for periods up to 32 years,
were examined. Ophthalmological examination included full-field electroretinography (ERG), multifocal electroretinography (mfERG) and multifocal visual evoked potential (mfVEP). Results: The ERGs were severely reduced in all patients. The mfERGs demonstrated the residual central retinal function in five of the patients. The mfVEPs showed measurable amplitudes centrally in most of the patients. The follow-up
examinations demonstrated the slowly progressive course of the disease with preservation or only slight further loss of visual fields over a period of 732 years. Conclusion: Patients with retinitis pigmentosa may not always follow the typical natural course with progressive loss of visual fields, which may in some patients remain unaffected over several decades Multifocal ERG and mfVEP may be clinically useful for evaluating remaining visual function in these patients