Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Measurement of CENTRAL VENOUS PRESSURE Via A Transducer

Download as pptx, pdf, or txt
Download as pptx, pdf, or txt
You are on page 1of 22

Measurement of CENTRAL

VENOUS PRESSURE via a


Transducer

PRESENTED BY:
C. SAM ILA KYNDIAH
M S C ( N ) 1 ST Y E A R
NINE, PGIMER
Key terms

1. Central venous pressure


2. Central venous pressure monitoring
3. Phlebostatic axis
Introduction

 CVP is considered equivalent to right atrial pressure

 It is used to estimate a patient’s cardiac function,


venous return to the heart, and gauge how well the right
ventricle of the heart function, if the heart cannot
accept blood, then it will back up into the venous
system affecting the intravascular fluid volume status.
How does a CVP works?

Because no major valves lie at


the cavo-atrial junction,
pressures during systole and
diastole reflect back to the
catheter allowing for CVP
measurement. Normal CVP
values range 2-6mmHg
or 4-12cmH20
Wheatstone principle of Transducer

Superman take care of this!


What does CVP values mean?

increase decrease

 Fluid overload
 Right heart failure  Shock
 Cardiac tamponade  Hypovolemia
 Pleural effusion  Forced inhalation
 Tension pneumothorax
 Forced exhalation
 Mechanical ventilation
Equipments needed

 Transducer
 Transducer cable
 Transducer holder
 Pressure bag
 Disposable pressure tubing
 Yard stick
 Hepsaline
Ensuring accuracy

1. Priming of the pressure tubing

2. Levelling and zeroing

3. Dynamic response testing


Priming the pressure tubing

 Use 500 mL Hep Saline


 Use aseptic technique to spike bag , prime entire tubing
(stopcocks, luer-locks, transducer)
 Eliminate all air bubbles as they can be a main factor in
waveform blunting or overdamping
 Insert IV fluid bag into pressure bag and inflate the
pressure bag to 300 mmHg
 Label IV bag with date and time solution is hung +
initials
Priming the tubing cont.....

• Check all connectors on tubing as they may be loose.


Make sure that the connectors are secure but don’t
over tighten them b/c they can become stripped

 Insert IV fluid bag into pressure bag and inflate the


pressure bag to 300 mmHg. Why?

Prevents air from going into the solution and catheter


from clotting, allows 3ml/hr flush solution to be
delivered through the catheter
Priming the tubing cont.....

 Insert transducer into the transducer holder that


mounts onto the IV pole
 Avoid over tightening and stripping of connectors
 Prime entire tubing system including stopcock, luer-
locks, and transducer
Priming the tubing cont....

 Clamp CVC lumen to be used


 Scrub CVC port with alcohol swab (15 sec)
 Connect transducer directly to CVC port
When is ZEROING needed?

Whenever the air-fluid interface and whenever the


reference point changes. Position change, when
accuracy of waveform reading is questionable
Zeroing the transducer

 Place HOB from zero to 45 degrees. Supine is


recommended
 Position the patient and the transducer at the same
level - Make sure the transducer is located at the
phlebostatic axis
 With a carpenter’s level or yard stick locate the
phlebostatic axis (right atrium of the heart – 4th
intercostal space, midaxillary line). This ensures the
accuracy of the readings by eliminating hydrostatic
forces on the transducer.
 If transducer is too high will have falsely low BP readings.
 If the transducer is too low will have falsely elevated BP
readings.
Location of the phlebostatic axis
Phlebostatic axis
Zeroing the transducer

 Turn the stopcock just above the transducer off to the


patient’s arterial catheter … “off to the patient”
 Zero Balance & Calibrate the Transducer by:
 Open stopcock on transducer to port or “air”

 Remove dead-end cap

 Activate flush device

Press zero button on bedside monitor (will read 0)


 Return stopcock back to port/monitoring position

 Replace dead-end cap


Maintaining a CVP line

 Assess flush system every 4 hours to ensure pressure


bag is inflated to 300mmHg and that fluid is present in
flush solution
 Evaluate pressure monitoring system regularly for air
bubble formation and remove if present
 Evaluate the patient regularly for signs or symptoms of
catheter-related infection
Precautions & key points

 Monitor alarms set at appropriate limits


 Obtain baseline data including vital signs, level of
consciousness, and hemodynamic stability to help
identify acute changes in the patient.
 Ensure that pt is still while CVP reading is being
taken – measure at end expiration
 If CVP fluctuates by more than 2mmHg suspect
change in clinical status and report
Precautions and key points cont ...

 Ensure that the patient is still while the CVP reading


is being taken to prevent artifacts.
 Transparent dressings should be changed every 7
days and SOS
 Assess catheter necessity daily
 Complications of CVP monitoring include sepsis,
thrombus, vessel puncture, and air embolism
Documentation

 Position for zeroing the transducer


 CVP readings, interventions, outcomes, and if MD
was notified
 Dressing, tubing, flush solution changes, and
discontinuation of line
References

 Lippincott Williams & Wilkins (2011). Lippincott’s Nursing


Procedure and Skills. Central venous pressure monitoring,
transducer. Retrieved July 24, 2011 from,
http://procedures.lww.com/lnp/view.do?searchQuery=Arterial%20
pressure%20monitoring&pId=912702
 Lippincott Williams & Wilkins (2011). Lippincott’s Nursing
Procedure and Skills. Transducer system setup. Retrieved June 30,
2011 from,
http://procedures.lww.com/lnp/view.do?searchQuery=Transducer
%20system%20setup&pId=164403
 Pittman, J. A.L., Ping, J.S., Mark, J.B (2006). Arterial and central
venous pressure monitoring. Anesthesiology Clin, 24(4), 717-35.
 Rauen, C.A., Makic,m.B., & Bridges, E. (2009). Evidence-based
practice habits: Transforming research into bedside practice.
Critical Care Nurse 29(2), 46-59

You might also like