This document provides information about central venous pressure (CVP) monitoring, including:
- CVP is measured in the right atrium and indicates right heart function and volume status. Normal CVP is 3-8 mmHg.
- CVP is monitored via a central venous catheter connected to a pressure transducer and cardiac monitor. The transducer must be zeroed at the phlebostatic axis.
- The CVP waveform has distinct A, C, and V waves that correlate with the ECG and events in the cardiac cycle. CVP readings are taken at the peak of the A wave or mid-QRS Z-point.
- Proper technique for CVP monitoring and
This document provides information about central venous pressure (CVP) monitoring, including:
- CVP is measured in the right atrium and indicates right heart function and volume status. Normal CVP is 3-8 mmHg.
- CVP is monitored via a central venous catheter connected to a pressure transducer and cardiac monitor. The transducer must be zeroed at the phlebostatic axis.
- The CVP waveform has distinct A, C, and V waves that correlate with the ECG and events in the cardiac cycle. CVP readings are taken at the peak of the A wave or mid-QRS Z-point.
- Proper technique for CVP monitoring and
This document provides information about central venous pressure (CVP) monitoring, including:
- CVP is measured in the right atrium and indicates right heart function and volume status. Normal CVP is 3-8 mmHg.
- CVP is monitored via a central venous catheter connected to a pressure transducer and cardiac monitor. The transducer must be zeroed at the phlebostatic axis.
- The CVP waveform has distinct A, C, and V waves that correlate with the ECG and events in the cardiac cycle. CVP readings are taken at the peak of the A wave or mid-QRS Z-point.
- Proper technique for CVP monitoring and
This document provides information about central venous pressure (CVP) monitoring, including:
- CVP is measured in the right atrium and indicates right heart function and volume status. Normal CVP is 3-8 mmHg.
- CVP is monitored via a central venous catheter connected to a pressure transducer and cardiac monitor. The transducer must be zeroed at the phlebostatic axis.
- The CVP waveform has distinct A, C, and V waves that correlate with the ECG and events in the cardiac cycle. CVP readings are taken at the peak of the A wave or mid-QRS Z-point.
- Proper technique for CVP monitoring and
Download as PPTX, PDF, TXT or read online from Scribd
Download as pptx, pdf, or txt
You are on page 1of 16
At a glance
Powered by AI
CVP measures blood pressure in the right atrium and vena cava, indicating right heart function and indirectly reflecting right ventricular pressure. It is an indicator of cardiac preload, afterload and contractility.
CVP is a direct measurement of the blood pressure in the right atrium and vena cava, indicating right heart function and indirectly reflecting right ventricular end-diastolic pressure.
CVP is monitored by inserting a catheter into the internal jugular or subclavian vein and advancing it to the superior vena cava near the right atrium. It is then connected to a pressure transducer and cardiac monitor.
By Nicole Bayuntara
Head Nurse/ Technical Advisor
August 2012, Updated July 2013 CVP = Central Venous Pressure CVP indicates Right Heart Function Is a direct measurement of the blood pressure in the right atrium and vena cava. CVP is an indicator of cardiac preload, afterload and contractility. (How well the heart is functioning) Indirectly reflects right ventricular end- diastolic pressure. Volume of blood returning to the right heart Vascular tone. Cardiac contractility. Patient position. Central venous access for CVP monitoring is obtained by inserting a catheter into a vein, typically the subclavian or jugular vein, and advancing it toward the heart until the catheter tip rests within the superior vena cava near its junction with the right atrium. Via a CVC. Connected with transducer, pressure bag, transducer cable and Cardiac Monitor. Pressure Bag with IV Normal Saline up to 300mmHg Connect to Brown/Distal (Wider) Lumen of the CVC Pause IV fluids running into this lumen of CVC while zeroing and taking CVP reading(No Inotropes should run through this lumen with CVP) Zero Transducer (Off to patient, open to air, press zero on monitor) Level with Phelbostatic Axis (Zero Point) 3 to 8 cm H2O or 2 to 6 mm Hg. CVP is elevated by : • overhydration which increases venous return • heart failure or PA stenosis which limit venous outflow and lead to venous congestion • positive pressure breathing (Ventilation), straining, CVP decreases with: • hypovolemic shock from hemorrhage, fluid shift, dehydration • negative pressure breathing which occurs when the patient demonstrates retractions or mechanical negative pressure which is sometimes used for high spinal cord injuries. A smaller-than-usual waveform can be caused by air bubbles in the system, thrombus formation, lodging of the catheter against the vessel wall, kinking of the catheter, incorrect calibration, or a loose connection in the tubing or transducer. An erratic waveform can result from movement of the catheter tip within the vessel lumen (the catheter may require repositioning). An absent waveform may indicate a large leak in the system (usually noted by reflux of blood in the tubing); a loose, cracked, or defective transducer; air in the transducer; stopcock turned to the wrong position; or thrombus occlusion of the catheter tip. The a wave is produced by RA systole (contraction) and occurs 80 to 100 ms after the P wave on the ECG. The c wave occurs with tricuspid valve closure; isovolemic ventricular contraction forces the tricuspid valve to bulge upward into the RA. The c wave follows the QRS on ECG. The v wave occurs as the RA continues to fill during against a closed tricuspid valve in late ventricular systole. The v wave correlates with the peak of the T wave on ECG. The high point of the A wave is the atrial pressure at maximum contraction and where to measure CVP. The Z-point coincides with the middle to end of the QRS wave. It occurs just before closure of the tricuspid valve. Therefore, it is a good indicator of right ventricular end diastolic pressure. The Z-point is useful when A waves are not visible, as in atrial fibrillation. Perform hand hygiene. Place the patient in a supine position and explain the procedure to patient. (If the patient can't tolerate being supine, make sure all CVP readings are taken with the patient in the same alternate position.) Locate the phlebostatic axis at the intersection of the mid-axillary line and fourth intercostal space (see illustration). If an I.V. solution is being infused through the CVP monitoring line, temporarily stop it and flush the line to prevent artifacts. Turn the three-way stopcock off to the patient and remove the cap from the three-way port to open the system to air. Press the zero button on the monitor and look for a display indicating the equipment has been zeroed. Replace the cap on the stopcock and turn the stopcock on to the patient. Observe the CVP waveform and document the CVP reading and patient position. Resume the I.V. infusion if indicated The CVP can also be measured manually using a manometer. A 3-way tap is used to connect the manometer to an intravenous drip set on one side, and, via extension tubing filled with intravenous fluid, to the patient on the other (Diagram 1). It is important to ensure that there are no air bubbles in the tubing, to avoid administering an air embolus to the patient. You should also check that the CVP catheter tubing is not kinked or blocked, that intravenous fluid can easily be flushed in and that blood can easily be aspirated from the line. The 3-way tap is then turned so that it is open to the fluid bag and the manometer but closed to the patient, allowing the manometer column to fill with fluid (Diagram 2). It is important not to overfill the manometer, so preventing the cotton wool bung at the manometer tip from getting wet. Once the manometer has filled adequately the 3-way tap is turned again – this time so it is open to the patient and the manometer, but closed to the fluid bag (Diagram 3). The fluid level within the manometer column will fall to the level of the CVP, the value of which can be read on the manometer scale which is marked in centimetres, therefore giving a value for the CVP in centimetres of water (cmH2O). The fluid level will continue to rise and fall slightly with respiration and the average reading should be recorded. http://www.nursingcenter.com/prodev/c e_article.asp?tid=1267859 http://www.rnceus.com/hemo/cvp.htm http://www.anaesthesia.hku.hk/LearNet/ measure.htm