Technical Features FT1800

Why need Blood and Infusion Warmer?

Hypothermia

 Anesthesia 1st hour : decrease core temp. 1 – 1.6ºC

 Intra- and post- surgery: if patient core temp. < 35ºc,


need 5h to back to normal

 Massive fluid infusion

 Elderly, weak, infant and children patients


Why need Blood and Infusion Warmer

Complications
 Shivering
 Decreased tissue oxygen

delivery,
 Cardiac arrhythmias
 Impair coagulation, so

increase blood loss


 Wound infection
 Prolong postoperative

recovery
 Impairs immune function

disorder
Why need Blood and Infusion Warmer?

 Enzymes moderate &


metabolize ⇨ temp.
sensitive
 Warmed Infusion

effectively
increase /maintain
core temp.
 Less thermal stress ⇨

infusion comfort
Advanced features

o Advanced control: microprocessor + temperature sensors

o User friendly interface: smart design, easy to operate

o High safety system:


 24h continuous operating & self-testing

 over-heating automatic cut off

 Visual & audio alarms

o Open system:
 Accept the standard IV tube hospitals using now

 No extra disposable costs

 The most economical solution, especially welcomed by developing countries


The most advanced tube heating
system in this sector worldwide
FT1800
FT1800
 1.4m warming profile
active & gentle warming
up to the patient

 Using standard IV tubing

 2min reach to 36ºC


*room temp. 20ºC
 warming capacity:
 1.5 L/H – one channel

The most effective heating


Single channels
system using standard tubing
Open system, heating up to the patient

IV tube is completely wrapped in,


Soft warming profile fits
no heat loss wide range of tubing size
FT1800

 Set temperature freely


33 - 41°C

 Alarming & automatic


cut-off protections

 Easy to operate

Big LED screen


User friendly interface
FT1800
General illustration

Infusion
Stand
Infusion
Tubing
Main Control
Body

Warming
Profile
LED
Control Screen
Panel

Power
Connection
Cable
Cable
Illustration of the holder and the warming profile
Illustration of the control panel

Key to set Key to set


temp. up temp. down

Key to switch between


Celsius and Fahrenheit
Key
Standby/Oper
ating
Illustration of LED screen
- the symbol of the alarm will appear on the screen only when the alarm is triggered
Installation & Operation

Step 1:
- Use the clamp to fix
the device firmly to the
infusion stand at the
position of 20-30cm
below the
infusion/blood bag
- Screw the fastening
screw
- The device is almost
equal to the drip bucket
Fastening
Screw Infusion Clamp
Stand
Installation & Operation

Step 2:

roll clamp
- Glide the roll clamp
next to the drip bucket
- Empty the air in
tubing to make
infusion channel ready
Installation & Operation

Step 3:

- Take the primed


infusion line at 2-3cm to
the end of patient side
to insert in the tail
(small end) of the
warming profile

2-3CM
Installation & Operation

Step 4:

- Move your thumb over


the infusion line and
press the line gently into
the groove towards the
head(big end) of the
warming profile, until the
Big end
infusion line is
completely wrapped in
the warming profile
Installation & Operation

Step 5:

- Place the ends of the


warming profile into the
holder to prevent the
infusion tube
contacting with the Big
ground Small End
End
Installation & Operation

Note :
If the Infusion tubing is too
short to fill the warming
profile from the beginning to
the end, it is necessary to
ensure that the patient end
(where the fluid flows out) of
the warming profile is filled.
The head of the warming
profile can be left unfilled. Of
course, a long enough tubing
ensures the heating effect.
Installation & Operation

Step 6:
- Press the key or to
select the desired
temperature
- Adjust the appropriate
infusion rate and press the
key to start

Note: Resetting of
temperature can only be
performed in Standly mode.
Installation & Operation

Note :
If the infusion needs to be stopped for more
than 3 minutes, press the key to stop
warming (Standby). By press the key again
to resume it.
Installation & Operation
Shutting down the device
 press the key to enter Standby mode.
 Release all pressure in the infusion system by

switching off the pressure cuffs or infusion


pumps.
 Remove the consumable material from the

warming profile and dispose it according to the


relevantregulations.
 Disconnect the device from the power supply.

Clean and disinfect the equipment according to


instructions.
Maintenance
 After each use, if possible, please clean device
surface, fastening clamp and warming profile. Use
disinfectant to clean them periodically.
 Wring soft cloth with water or soap lye and use it to
clean device.
 Don’t allow cleaning fluids and disinfectant enter
ventilator of device.
 Disinfectants should be based on alcohol liquid
without oxygen releasing (with sodium hypochlorite and
1.4% bleach solution or < 0.2% acetaldehyde).
 It can’t be immersed in liquid and can’t be disinfected
in autoclave or by heat chemistry method.
Used in clinical settings
( major dept. )

Anesthesia, OT :
Blood transfusion & infusion during operation

ICU,SICU,CCU,NICU:
Blood , infusion , enteral/parenteral nutrition ,
blood return feeds

Emergency care, ambulance :


Blood transfusion & infusion
Used in clinical settings
( major dept. )

Blood bank :
Whole blood from cold storage

Dialysis:
Blood return feeds
Peritoneal dialysis
Used in clinical settings
( major dept. )

Urinary surgery:
Rinse solution for prostate or bladder

Oncology
Intraperitoneal perfusion chemotherapy
Used in clinical settings
( major dept. )

Obstetrics and Gynecology :


Maternal

Neonates, paediatrics:
Blood transfusion & infusion
Tenderness
&
Excellence
Keep every life
well…….