Here are potential responses to the practice questions:
1. Limitations to Beer's law include:
- Only applicable at low concentrations where absorbance is directly proportional to concentration
- Interference from other absorbing species
- Non-linear response at high concentrations
- Variations in pathlength between samples
- Temperature effects on absorbance
- Solvent effects on molar absorptivity
2. (a) Single beam spectrophotometer diagram and description
(b) Double beam spectrophotometer diagram showing two beams, one through sample one through reference. Description of how it measures both sample and reference simultaneously correcting for fluctuations.
3. Molar absorptivity - ability of a substance to absorb light at
Here are potential responses to the practice questions:
1. Limitations to Beer's law include:
- Only applicable at low concentrations where absorbance is directly proportional to concentration
- Interference from other absorbing species
- Non-linear response at high concentrations
- Variations in pathlength between samples
- Temperature effects on absorbance
- Solvent effects on molar absorptivity
2. (a) Single beam spectrophotometer diagram and description
(b) Double beam spectrophotometer diagram showing two beams, one through sample one through reference. Description of how it measures both sample and reference simultaneously correcting for fluctuations.
3. Molar absorptivity - ability of a substance to absorb light at
Here are potential responses to the practice questions:
1. Limitations to Beer's law include:
- Only applicable at low concentrations where absorbance is directly proportional to concentration
- Interference from other absorbing species
- Non-linear response at high concentrations
- Variations in pathlength between samples
- Temperature effects on absorbance
- Solvent effects on molar absorptivity
2. (a) Single beam spectrophotometer diagram and description
(b) Double beam spectrophotometer diagram showing two beams, one through sample one through reference. Description of how it measures both sample and reference simultaneously correcting for fluctuations.
3. Molar absorptivity - ability of a substance to absorb light at
Here are potential responses to the practice questions:
1. Limitations to Beer's law include:
- Only applicable at low concentrations where absorbance is directly proportional to concentration
- Interference from other absorbing species
- Non-linear response at high concentrations
- Variations in pathlength between samples
- Temperature effects on absorbance
- Solvent effects on molar absorptivity
2. (a) Single beam spectrophotometer diagram and description
(b) Double beam spectrophotometer diagram showing two beams, one through sample one through reference. Description of how it measures both sample and reference simultaneously correcting for fluctuations.
3. Molar absorptivity - ability of a substance to absorb light at
Download as PPTX, PDF, TXT or read online from Scribd
Download as pptx, pdf, or txt
You are on page 1of 44
CHE 312_LECTURE 3
Chemical analysis using a
spectrophotometers
Single-beam spectrophotometer- uses only one
beam in which the sample and reference cuvettes are alternately placed in the beam path Double-beam spectrophotometer- uses two beams where one beam is passed through the sample and the other beam is passed through the reference Block diagrams for both a single-beam and double-beam spectrophotometer Single-beam spectrophotometers Uses only one beam, the sample and reference are alternately placed in the beam (i.e in the sample slot) at each wavelength to be measured. If the radiant power of the radiation emerging from the sample is P and that emerging from the reference is Po then A=log Po -log P The purpose of comparing the sample to the reference solution compensate for reflection, scattering and absorption by both the cuvette and the solvent Disadvantages of single-beam spectrophotometers The inconvenience of alternately removing the reference to place the sample cuvette for each wavelength being measured Not suitable for measuring absorbance as a function of time especially in kinetic experiments This is because both the beam intensity and the detector response may drift slightly with time Double-beam spectrophotometers It uses two beams in which one beam is passed through the sample and the second one through the reference solution This is achieved by the use of a mirror rotated by a motor to direct the beam alternately to sample and reference cuvettes The radiant powers P and Po are compared: A=log Po /P Advantages of double-beam spectrophotometer Automatic correction for any changes in beam intensity and detector response Automatic wavelength scanning and continuous recording of absorbance Chemical analysis In spectrophotometric analysis, the procedure involves; First obtaining a scan (spectrum) of the pure solvent or reagent blank or reference in the expected wavelength range Then the sample is scanned and hence True absorbance=Sample absorbance-blank absorbance The absorbance readings are done at the wavelength of maximum absorbance (λmax) NB: Why are absorbance readings taken at wavelength of maximum absorbance? Absorbance is constant at λmax because there is minimum drift in detector response/source intensity For a given concentration, ε is maximum at λmax Because Amax=εmax bc Most spectrophotometers exhibit their minimum relative uncertainty (i.e less drift) in the range A=0.4-0.9 and this because; If P<< Po, i.e high absorbance, A>0.9 If P is almost equal to Po, i.e low absorbance, hence it is difficult to distinguish between the sample and the reference for very low absorbance values Precautions Prevent stray light from entering the detector-the sample compartment must be light-tight Stray light will interfere with absorbance measurements Prevent dust from entering the beam path to avoid scattering of light Fingerprints on the faces of silica cuvettes absorb radiation-handle the cuvettes with tissue paper Avoid mismatch of the sample and reference cuvettes since this will lead to systematic error in comparing P and Po Avoid misplacement of the cuvette in the sample compartment as this will lead to random error Principles of Uv/Vis Absorption of radiation in the visible(Vis) and ultraviolet(UV) regions of the electromagnetic spectrum results in electronic transitions between molecular orbitals. The UV range extends from 100-400 nm of which 100-190 nm is known as far UV and 190-400 nm is called near UV Visible range extends from 400-800 nm Commercial UV/Vis instruments cover 200-1000 nm region Nature of Electronic Transitions When two atoms react to form a compound, electrons from both atoms participate in the bond and occupy a new molecular orbital where bonding electrons are associated with the molecule as a whole This is called a bonding orbital and represents the lowest energy level Simultaneously, a corresponding antibonding obital is also formed, which are vacant in the unexcited or groundstate Thus a covalent bond may be formed by combination of S orbitals or P orbitals. This are designated as σ and π bonds respectively Corresponding antibonding orbitals are designated as σ* and π* Valence electrons NOT involved in bonding are referred to as non-bonding electrons and designated n In organic molecules, n electrons (lone pairs)are located in the atomic orbitals of nitrogen, oxygen, sulphur and halogen atoms regions include The possible electronic transitions involved in uv/vis include σ → σ*, n → σ*, n → π* and π → π* And their relative energies in decreasing order σ → σ* > n → σ* > π → π*>n → π* See Figures in the next two slides which depict the relative energies of the possible electronic transitions • σ → σ* and n → σ* transitions -These are high- energy transitions and involve very short wavelength ultraviolet light (< 150 nm) e.g observed for alkanes. • These transitions are NOT useful analytically as they fall outside the generally available measurable range of UV-visible spectrophotometers (200-1000 nm) • π →π* and n → π* transitions- these are low- energy transitions and fall within measurable range of UV-visible spectrophotometers (200- 1000 nm) The most intense band (large ε) for compounds is mostly due to π →π* transition. The n → π* transition is of lowest energy but is of low intensity (weak absorption or low ε ) as it is symmetry forbidden. Why Broad bands In UV/Vis Spectra The transition of an electron from one energy level to another is accompanied by simultaneous change in vibrational and rotational states and causes transitions between various vibrational and rotational levels of lower and higher energy electronic states Therefore many radiations of closely placed frequencies are absorbed and a broad absorption band is obtained (see caffeine Absorption spectrum in the next slide) The most probable transition would appear to involve the promotion of one electron from the highest occupied molecular orbital (HOMO) to the lowest unoccupied molecular orbital (LUMO), but in many cases several transitions can be observed, giving several absorption bands in the spectrum important terms and definitions Chromophore: The group of atoms within a molecule containing electrons responsible for the absorption of radiation that causes electronic excitation. Examples – C=C, C=O, C=S, N=N, N=O Auxochrome: The substituents that themselves do not absorb ultraviolet radiations but their presence shifts the absorption maximum to longer wavelength (red shift) with a corresponding increase in absorption intensity. Some examples are substituents like, hydroxyl, alkoxy, halogen, amino group etc. All auxochromes have one or more non-bonding pairs of electrons. If an auxochromes is attached to a chromophore, it helps is extending the conjugation by sharing of non-bonding pair of electrons Bathochromic Shift or Red shift: A shift of an absorption maximum towards longer wavelength or lower energy. Hypsochromic Shift or Blue Shift: A shift of an absorption maximum towards shorter wavelength or higher energy Factors affecting absorption by a chromophore Solvents Effects Absorption bands arising from n → π* transitions suffer hypsochromic shifts on increasing the solvent polarity, whilst those of π → π* transitions are shifted bathochromically Explanations lie in the fact that the energy of the non-bonding orbital n is lowered by hydrogen bonding in the more polar solvent thus increasing the energy of the n → π* transition but the energy of the π* orbital is decreased relative to the n orbital Non-polar vs polar solvents • There is little or no effect on the uv/vis absorption spectrum of an analyte dissolved in a non-polar solvent • However, If the SAME analyte dissolved in a polar solvent , the absorption bands are broadened, which leads to a loss in structural resolution and reduction in εmax. • The absorption spectrum of phenol (analyte) in isooctane (non-polar solvent) and ethanol (polar solvent) is illustrating this effect (see next slide) Spectra of phenol in isooctane and ethanol Conjugation Effects Absorption bands due to conjugated chromophores are shifted to longer wavelengths (bathochromic or red shift) and intensified (large ε )relative to an isolated chromophore The shift can be explained in terms of interaction or delocalization of the π and π* orbitals of each chromophore to produce new orbitals in which the highest π orbital and the lowest π* orbital are closer in energy e.g Effect of conjugation on absorption of ethylene The longer the conjugated carbon chain in the absorbing system, the greater the intensity of the absorption( large ε) . This is shown by the spectra of the polyenes CH3- (CH=CH)n-CH3, where n=3,4 and 5 (next slide). • β-carotene, a vitamin found in carrots, and used in food colouring, has eleven conjugated double bonds (Fig below) and its absorption maximum has shifted out of the ultraviolet and into the blue region of the visible spectrum, hence it appears bright orange Some Uv/Vis Analytes (find more examples of your own) Nicotine Practice questions 2 1. Discuss limitations to Beers law in quantitative analysis 2. With the aid/help of a labeled block/schematic diagrams, describe how a) a single-beam spectrophotometer b) double-beam spectrophotometer works. 3. Define the following terms as applied to spectroscopy; molar absorptivity, spectrometry, Absorption spectrum, transmittance 4. Pure hexane has negligible ultraviolet absorbance above a wavelength of 200 nm. A solution prepared by dissolving 25.6 mg of benzene (C6H6, MM=78.114 g/mol) in hexane and diluting to 250.0 ml had an absorption peak at 256 nm and an absorbance of 0.266 in a 1.00 cm cell. Find the molar absorptivity of benzene at this wavelength (256 nm) 5. Describe the changes in matter when it interacts with: X-rays, UV/Vis, IR, microwave and radiofrequency electromagnetic radiation 6. Calculate the frequency in hertz of; i. An emission line of copper at 324.7 nm ii. An X-ray beam with a wavelength of 2.97 Å
7. Calculate the wavelength in nm of
iii. An airport tower transmitting at 118.6 MHz iv. An infra-red absorption peak with a wavenumber of 1375 cm-1 8. Calculate the frequency in hertz and the energy in joules of an X-ray photon with a wavelength of 2.35 Å 9. Calculate the wavelength and the energy in joules associated with a signal at 220 MHz 10. Express the following absorbances in terms of percent transmittance i. O.936 ii. 0.494 iii. 0.0350 10. Convert the accompanying transmittance data to absorbances; i. 22.7 % ii. 31.5 % iii. 0.103 iv. 0.567 11. The complex formed between Cu(I) and 1,10- phenanthroline has a molar absorptivity of 7000 L cm-1 mol-1 at 435 nm, the wavelength of absorption. Calculate i. The absorbance of a 6.77 x 10-5 M solution of the complex when measured in a 1.00 cm cell(cuvet) at 435 nm. ii. The percent transmittance of the solution in (i) iii. The concentration of a solution in a 5.00 cm cell that has the same absorbance as the solution in (i) iv. The path length through a 3.40 x 10-5 M solution of the complex that is needed for an absorbance that is the same as the solution in (i)