Analyzer
Analyzer
Analyzer
Conductivity Level
PH Flow
Oxygen pressure
CO2 Temperature
CO Speed
Nitrogen
CnHm
…
:Common analytical parts
Sensor
Transmitter
Sampling system
Process line
Sample probe
To LER
Sensor
Transmitter
Sample box
Sample line (1/2"-1/4" tube) Cable Cable
tube"1/4
sample line
sample probe
sample box
Typical Sample Box
P.G
:Note
As an analyzer man (woman) you are
responsible to check and maintain not only
sensor and transmitter but also sample
!probe, sample line and sample box
Conductivity Measurement
Cell Constant (K) Values
k=L/a a=A*B
Units of Conductivity
Resistance(ohm)
Conductance(µs)
0.056 18,000,000
0.084 12,000,000
0.167 6,000,000
1 1,000,000
2.5 400,000
20 50,000
200 5,000
2000 500
20000 50
:Conductivity transmitters
:Calibration
sensor
:Caution
In some applications there is no chamber or
isolation valve to separate sensor from
process line and it is located directly in the
process line. In such situations if you
remove sensor, sample will come out with
high pressure and flow that can be very
dangerous. Please inform process people
.before any action
Temperature compensation
PH Measurement
PH means power of hydrogen ions
PH = -log10[H+]
sensor
Temperature
Calibration
:Temperature effects on PH sensor life time
?When is the time to replace a sensor
Unstable reading -2
To prevent explosion-1
To prevent polymerization-2
To prevent oxidation-3
Oxygen as impurity-4
Oxygen as final product-5
Measurement methods
Paramagnetic method (Is used for -1
percentage values)
Electrochemical method (Is used for-2
percentage, ppm, ppb values-percentage
is better not to be used)
Zirconium method (for places with high -3
temperature like furnaces or ovens)
Paramagnetic Oxygen Sensors
. O2 + 4e- = 2O •
- 2
The oxygen ions, O2-, are attracted •
though asolid electrolyte that is heated to
400°C and areconverted back to molecular
oxygen at the anodeelectrode. The
electrochemical oxidationreaction that
:takes place at the anode is
-
2O = O2+ 4 •
2- e
Life Time
Lifetime of the electrochemical sensor-1
can be increased by leaving it open circuit
when the instrument is switched off, but
this has the disadvantage that the sensor
takes about 20 minutes to settle down
.after reconnection
The commonest solution is to short the-2
.terminals of the sensor when not in use
Another possible solution is to disconnect-3
the sensor when the instrument will not be
used for a longer period of time, but this is
a somewhat unsatisfactory answer, and
raises questions about warranty
.conditions
Zirconium Oxide Oxygen Sensors
Hydrocarbon Measurement
Infrared sensors
Fundamentals
Of
Chromatography
Moisture (Dew point)
Measurement
Rosemount (GCX)
Siemens (Maxum)
Yokogawa (GC1000)
GCX - Modular GC Configuration
Mounting Plate with Lifting Hooks
Electronics Oven
GCX- Networking and Communications
GCX- 36
One Analogue
Output(Chromatogram)
4-20 mA DC
Isolated Outputs (6)
Alarm Relays
GCX- 3 GCX- 2 GCX-1 (TTL) (6)
5-24 VDC
Digital Inputs
Network
PC DCS
Controller
First Exposure To Chromatography
Was In Elementary School
Sample
Return
Slip Heater
Stream
T=10
T=20
seperation in column
GC Basics
Column Separation
TO
CARRIER GAS
COLUM DETECTOR
Carrier
Column Switching — Heart cut
Ethylene Tail
INJECT DETECTOR
Column # 1 Column # 2
Reference Measure
Vent Vent
Universal Detection
Sensitive Down To 100 ppm
Flame Ionization Detector
Vent
Igniter
Collector
Amplifier
Flame
Polarizing
voltage
Hydrogen
Carrier
Air
Flame Photometric Detector
Vent
Igniter
Flame Amplifier
Optical Filter
Photo-Multiplier Tube
Hydrogen
Carrier
Air
Sample Out
Peak Integration
Residence Time
Integration Integration
On Off
Baseline
Time
Typical chromatogram
Different resolutions
) Typical Chromatogram
Functions Of The Chromatograph Controller
Seldom agree
Differences in sample handling
Differences in calibration standards
What Makes A GC Complex
Multiple detectors