Parameters of Method Development (Using HPLC)
Parameters of Method Development (Using HPLC)
phases first thoroughly investigate binary mobile phases before going on to ternary think of the factors that are likely to be significant in achieving the desired resolution
operational parameters involved makes HPLC method development a complex process, which follows the following steps:
step 1 - selection of the HPLC method and initial
system step 2 - selection of initial conditions step 3 - selectivity optimization step 4 - system optimization step 5 - method validation.
always to consult the literature to ascertain whether the separation has been previously performed and if so, under what conditions. Consideration must be given to the following:
1.1 Sample preparation. 1.2 Types of chromatography. 1.3 Gradient HPLC. 1.4 Column dimensions. 1.5 Detectors. 1.6 UV wavelength. 1.7 Fluorescence wavelength.
extraction, pre-concentration or clean up? Is chemical derivatization required to assist detection sensitivity or selectivity?
samples, but if acidic or basic analytes are present then reverse phase ion suppression (for weak acids or bases) or reverse phase ion pairing (for strong acids or bases) should be used.
with a large number of components (2030) because the maximum number of peaks that can be resolved with a given resolution is much higher than in isocratic HPLC. Gradient HPLC will also give greater sensitivity, particularly for analytes with longer retention times, because of the more constant peak width (for a given peak area, peak height is inversely proportional to peak width).
short columns (1015 cm) are recommended to reduce method development time. Such columns afford shorter retention and equilibration times. A flow rate of 1-1.5 mL/min should be used initially. Packing particle size should be 3 or 5 m.
1.5 Detectors
Consideration must be given to the following: Do the analytes have chromophores to enable UV detection? Is more selective/sensitive detection required (Table I)? What detection limits are necessary? Will the sample require chemical derivatization to enhance detectability and/or improve the chromatography?
1.6 UV wavelength
For the greatest sensitivity max should be used,
which detects all sample components that contain chromophores. UV wavelengths below 200 nm should be avoided because detector noise increases in this region. Higher wavelengths give greater selectivity.
maximum; that is, the wavelength that gives the maximum emission intensity. The excitation is set to the maximum value then the emission is scanned to locate the emission intensity.
adequately retain all analytes; that is, ensures no analyte has a capacity factor of less than 0.5 (poor retention could result in peak overlapping) and no analyte has a capacity factor greater than 1015 (excessive retention leads to long analysis time and broad peaks with poor detectability). Selection of the following is then required. 2.1 Mobile phase solvent strength 2.2 Gradient HPLC 2.3 Determination of initial conditions
pull analytes from the column. It is generally controlled by the concentration of the solvent with the highest strength
analytes (2030) or with a wide range of analyte retentivities, gradient elution will be necessary to avoid excessive retention.
two gradient runs differing only in the run time. A binary system based on either acetonitrile/water (or aqueous buffer) or methanol/water (or aqueous buffer) should be used.
(peak spacing). The mobile phase and stationary phase compositions need to be taken into account. To minimize the number of trial chromatograms involved, only the parameters that are likely to have a significant effect on selectivity in the optimization must be examined. Note that the optimization of mobile phase parameters is always considered first as this is much easier and convenient than stationary phase optimization. Selectivity optimization in gradient HPLC. Initially, gradient conditions should be optimized using a binary system based on either acetonitrile/water (or aqueous buffer) or methanol/water (or aqueous buffer). If there is a serious lack of selectivity, a different organic modifier should be considered.
resolution and analysis time after satisfactory selectivity has been achieved. The parameters involved include column dimensions, column-packing particle size and flow rate. These parameters may be changed without affecting capacity factors or selectivity.
important for pharmaceutical analysis when ensurance of the continuing efficacy and safety of each batch manufactured relies solely on the determination of quality. Method development and validation can be simultaneous, but they are two different processes, both downstream of method selection. Analytical methods should be used within GMP and GLP environments. It must be developed using the protocols set out in the ICH guidelines (Q2A and Q2B).
Accuracy Precision (repeatability and intermediate precision) Specificity Detection limit Quantitation limit Linearity Range Robustness Stability of analytical solutions.
Accuracy
Specificity Precision Limits of detection and quantitation Analytical solution stability Robustness
Conclusion
Method development involves a series of sample
steps;
A column and detector are chosen The sample is dissolved, extracted, purified and filtered
as required An eluent survey (isocratic or gradient) is run The type of final separation (isocratic or gradient) is determined from the survey Preliminary conditions are determined for the final separation Retention efficiency and selectivity are optimized as required for the purpose of the separation (quantitative, qualitative or preparation) The method is validated using ICH guidelines. The validated method and data can then be documented.