Channel Characterization of Implantable Intrabody Communication through Experimental Measurements

Kayhan Ateş University of Akdeniz
Department of Electrical and Electronics Engineering
AntalyaTürkiye
kayhanates@akdeniz.edu.tr
Anna Marcucci Fondazione I.R.C.C.S. Policlinico
San Matteo
PaviaItaly
anna.marcucci01@universitadipavia.it
Pietro Savazzi University of Pavia
&\&& CNIT Consorzio Nazionale Interuniversitario per le Telecomunicazioni - Unità di Pavia
PaviaItaly
pietro.savazzi@unipv.it
Şükrü Özen University of Akdeniz
Department of Electrical and Electronics Engineering
AntalyaTürkiye
sukruozen@akdeniz.edu.tr
Fabio Dell’Acqua University of Pavia
&\&& CNIT Consorzio Nazionale Interuniversitario per le Telecomunicazioni - Unità di Pavia
PaviaItaly
fabio.dellacqua@unipv.it
 and  Anna Vizziello University of Pavia
&\&& CNIT Consorzio Nazionale Interuniversitario per le Telecomunicazioni - Unità di Pavia
PaviaItaly
anna.vizziello@unipv.it
(2024)
Abstract.

Intrabody communication (IBC), is a promising technology that can be utilized for data transmission across the human body. In this study, a galvanic coupled (GC)-based IBC channel has been investigated for implantable configuration both theoretically and experimentally in the frequency range of 0 to 2.5 MHz. Theoretical studies were performed by using finite element method (FEM) based simulation software, called Comsol Multiphysics. A cylindrical human arm was modeled with realistic values. Experimental studies were carried out with chicken breast tissue as a substitute for human tissue. The pseudorandom noise (PN) sequences were transmitted to investigate the correlative channel sounder of tissue model. Results showed that the frequency affects signal propagation through the tissue model. Additionally, it is crucial to cancel common-mode noise in the IBC channel to enhance communication quality.

Intrabody communication, galvanic coupling, implantable sensor, channel impulse response, finite element method, simulation and measurement
journalyear: 2024copyright: acmcopyrightconference: The Eleventh Annual ACM International Conference on Nanoscale Computing and Communication (ACM NanoCom 2024); October 28–30, 2024; Milan, Italybooktitle: The Eleventh Annual ACM International Conference on Nanoscale Computing and Communication (ACM NanoCom 2024), October 28–30, 2024, Milan, Italyccs: Hardware Wireless integrated network sensorsccs: Mathematics of computing Partial differential equationsccs: Hardware Bio-embedded electronicsccs: Hardware Sensor devices and platformsccs: Hardware Wireless devicesccs: Hardware Noise reduction

1. Introduction

An increasingly aging population, the rise in chronic diseases, and the consequent cost soaring are among the major issues that health care systems are facing globally. This is why research is investigating innovative technologies to improve patient care, promote early diagnosis, and improve disease management (Zheng2014, ). Indeed, next-generation health care is moving toward new communication techniques that can interconnect devices: wireless body networks that can connect sensors inside, outside, or near the human body. This paradigm, which has been gaining increasing interest in recent years, is known as intra-body communication (IBC).

Radio frequency (RF) waves are most commonly used in IBC techniques. However, several studies have shown that applications within living tissue incur high losses. As a consequence, coverage is limited to short distances and care must be placed in avoiding heating that may cause damage to the tissues traversed by the waves (Vizziello2024, ). These are the main reasons that have driven research toward alternative technologies for subcutaneous communication between implantable devices. These include galvanic coupling (GC) technique. GC uses one pair of electrodes as the transmitter (TX) and another pair as the receiver (RX). The electrodes can be fixed or implanted in the body, and transmit weak currents (¡1 mA) modulated with data at low-to-medium frequencies (1 kHz-100 MHz) (Vizziello2022b, ).

In designing an intrabody communication system, characterization of the body channel is crucial, and methods based on impulse response show some potential (Vizziello2022b, ) in modeling IBC channels. Experiments require the use of different types of equipment: a signal generator on the TX side and the spectrum analyzer or oscilloscope on the RX side, or a vector network analyzer (VNA) on both TX and RX sides. To avoid distortion of experimental results, ground electrodes (GND electrodes) in TX and RX sides must not be connected through the equipment system, on the contrary their mutual electrical isolation must be ensured (Maamir2018, ).

This study focuses on intra-body communication based on GC for body channel modeling. Both the in-silico studies, through the use of Comsol Multiphysics FEM software, and the experimental measurements were conducted to evaluate compatibility of the communication channel through the tissue with implantable technologies. In in-silico tests, a cylindrical human arm composed of layers of different tissues was modeled with implanted pairs of TX and RX electrodes. Realistic values were assigned to all components and properties of the system. For experimental measurements, optocoupler-based isolation circuits were used to ensure electrical isolation between TX and RX ports. Then, pseudorandom noise (PN) sequences were generated and transmitted. The channel impulse response (CIR) was evaluated by analyzing the sequence transmitted from the output port of the TX isolation circuit through a chicken breast and the sequence recorded at the output port of the RX isolation circuit. Results indicated that the tissue communication channel shows a high-pass behavior.

2. Channel Modeling Overview

In this study, the channel characterization of intrabody communication through both in-silico and experimental investigations has been carried out on GC-based technology (Vizziello2022a, ). The selection of GC technology was driven by several factors, including interference and security concerns, electrical safety for biological tissue, and the simplicity of the transceiver architecture compared to other IBC technologies (Vizziello2023, ).

2.1. Finite Element Method Based Simulation Space

In-silico simulations were carried out with FEM-based commercial software, Comsol Multiphysics 6.1. In the simulation software, the quasi-static solution and the coupling of the electronic circuits were carried out with the Electric𝐸𝑙𝑒𝑐𝑡𝑟𝑖𝑐Electricitalic_E italic_l italic_e italic_c italic_t italic_r italic_i italic_c Currents𝐶𝑢𝑟𝑟𝑒𝑛𝑡𝑠Currentsitalic_C italic_u italic_r italic_r italic_e italic_n italic_t italic_s and Electric𝐸𝑙𝑒𝑐𝑡𝑟𝑖𝑐Electricitalic_E italic_l italic_e italic_c italic_t italic_r italic_i italic_c Circuits𝐶𝑖𝑟𝑐𝑢𝑖𝑡𝑠Circuitsitalic_C italic_i italic_r italic_c italic_u italic_i italic_t italic_s interfaces of the AC/DC𝐴𝐶𝐷𝐶AC/DCitalic_A italic_C / italic_D italic_C Module𝑀𝑜𝑑𝑢𝑙𝑒Moduleitalic_M italic_o italic_d italic_u italic_l italic_e, respectively (Pola2023, ). The simulation space was established on the basis of the following conditions (Il2023, ; Ates2022, ):

  • The problem space was modeled as three-dimensional.

  • Biological tissues were directly affected by electrical signals.

  • The outside of the problem space was considered as an infinite space.

  • The dielectric properties of tissues were modeled as homogeneous at the frequencies of interest.

The properties of the tissue layers of the arm model are presented in Table 1. The length of the arm model is 60 cm. The implanted TX and RX electrode pairs were modeled as copper with dimensions of 1 mm radius and 2 cm height. The distance between the TX and RX electrodes is 10 cm, and the inter-electrode distances within each pair are 4 cm for both TX and RX.

Table 1. Different tissues and properties to simulate the human arm model (Ateş2019, ).
Tissue Thickness (mm)
Skin 1.5
Fat 8.5
Muscle 27.5
Cortical Bone 6
Cancellous Bone 6.5

The general equation of the problem space is stated in Eq. (1) (Li2019, ):

(1) .(jωϵ0ϵV)+.(σV)+.Jsource=0formulae-sequence𝑗𝜔subscriptitalic-ϵ0superscriptitalic-ϵ𝑉𝜎𝑉subscript𝐽𝑠𝑜𝑢𝑟𝑐𝑒0\nabla.(j\omega\epsilon_{0}\epsilon^{\prime}\nabla V)+\nabla.(\sigma\nabla V)+% \nabla.J_{source}=0∇ . ( italic_j italic_ω italic_ϵ start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT italic_ϵ start_POSTSUPERSCRIPT ′ end_POSTSUPERSCRIPT ∇ italic_V ) + ∇ . ( italic_σ ∇ italic_V ) + ∇ . italic_J start_POSTSUBSCRIPT italic_s italic_o italic_u italic_r italic_c italic_e end_POSTSUBSCRIPT = 0

Through this latter equation, the behavior of electrical signals in tissue has been examined considering initial and boundary conditions through FEM-based simulations.

Here, V𝑉Vitalic_V is the electrical potential, ϵsuperscriptitalic-ϵ\epsilon^{\prime}italic_ϵ start_POSTSUPERSCRIPT ′ end_POSTSUPERSCRIPT and σ𝜎\sigmaitalic_σ represent the dielectric constant and electrical conductivity of the tissue, respectively.

2.2. Experimental Measurement Testbed

The experimental system planned within the scope of this study is shown in Figure 1. Here, a computer-generated PN signal was imported to the signal generator (T3DSO1000-FG, Teledyne, USA). Then, the signal was transmitted through the chicken breast and recorded using the oscilloscope (T3DSO3204, Teledyne, USA) to analyze the communication channel. The investigations in this study were conducted within the frequency range of 0 to 2.5 MHz.

Refer to caption
Figure 1. Measurement testbed for intrabody communication channel.

Common-mode noise poses a significant challenge to intrabody communication (Shinagawa2014, ), and various methods have been proposed in the literature to mitigate it (Vasić2016, ; Maamir2018, ; Sasaki2013, ; Modak2022, ; Vizziello2020, ). This type of noise causes distorted measurement results; however, decoupling the measurement testbed from power networks eliminates the common-mode return path, thereby reducing this effect. Therefore, the HCPL-4562 optocoupler-based isolation circuits were utilized in this study. The bandwidth of the optocoupler is 17 MHz, as stated in the technical data sheets (HCPL4562, ). The designed circuits are shown in Figure 2.

Refer to caption
Figure 2. Optocoupler circuit diagram (HCPL4562, ).

2.3. Correlative Sounder of the Intrabody Communication Channel

In this study, intra-body communication channel modeling was carried out through analysis of channel impulse response (CIR), which employs the channel sounder (Molisch2011, ). One of the main advantages of this method is that the impulse response is recorded based on real values. Additionally, system simulations can be repeated based on the recorded data because the data remains accessible and can be used for different scenarios. In communication systems, the received signal is described as y(t)=x(t)h(t)+n(t)𝑦𝑡𝑥𝑡𝑡𝑛𝑡y(t)=x(t)\ast h(t)+n(t)italic_y ( italic_t ) = italic_x ( italic_t ) ∗ italic_h ( italic_t ) + italic_n ( italic_t ), where y(t)𝑦𝑡y(t)italic_y ( italic_t ) is the output signal in time domain, x(t)𝑥𝑡x(t)italic_x ( italic_t ) is the input signal in time domain, h(t)𝑡h(t)italic_h ( italic_t ) is the channel impulse response, \ast is the convolution operator, and n(t)𝑛𝑡n(t)italic_n ( italic_t ) is the additive noise (Vizziello2024, ). If correlation with x(t)𝑥𝑡x(t)italic_x ( italic_t ) is computed on each side of the mentioned equation, it yields the correlation function, as represented in Eq. 2:

(2) Rxy(τ)=h(τ)Rxx(τ)subscript𝑅𝑥𝑦𝜏𝜏subscript𝑅𝑥𝑥𝜏R_{xy}(\tau)=h(\tau)R_{xx}(\tau)italic_R start_POSTSUBSCRIPT italic_x italic_y end_POSTSUBSCRIPT ( italic_τ ) = italic_h ( italic_τ ) italic_R start_POSTSUBSCRIPT italic_x italic_x end_POSTSUBSCRIPT ( italic_τ )

where Rxysubscript𝑅𝑥𝑦R_{xy}italic_R start_POSTSUBSCRIPT italic_x italic_y end_POSTSUBSCRIPT is the cross-correlation between the input and the output signal, h(t)𝑡h(t)italic_h ( italic_t ) is the channel impulse response over time delay, Rxxsubscript𝑅𝑥𝑥R_{xx}italic_R start_POSTSUBSCRIPT italic_x italic_x end_POSTSUBSCRIPT is the auto-correlation function of x(t)𝑥𝑡x(t)italic_x ( italic_t ), and τ𝜏\tauitalic_τ is the delay time. It is a well-established fact that the impulse response varies over time due to multiple channel effects (Hwang2016, ). Since each channel exhibits its own random signal losses and delays, the impulse response is specified as a function of time and delay. The slowly time-varying channel is expressed as follows (Molisch2011, ):

(3) h(t,τ)=α(t)δ(τ)𝑡𝜏𝛼𝑡𝛿𝜏h(t,\tau)=\alpha(t)\delta(\tau)italic_h ( italic_t , italic_τ ) = italic_α ( italic_t ) italic_δ ( italic_τ )

In Eq. 3, α(t)𝛼𝑡\alpha(t)italic_α ( italic_t ) is the attenuation over time, and δ(τ)𝛿𝜏\delta(\tau)italic_δ ( italic_τ ) is the Dirac delta function in delay time. Since the h(t,τ)𝑡𝜏h(t,\tau)italic_h ( italic_t , italic_τ ) of the GC intra-body communication channel changes slowly; the impulse response expression is considered as a time-invariant (Vizziello2024, ). Therefore, the GC channel impulse response becomes time-independent. The corresponding power delay profile may be expressed as in Eq. 4.

(4) P(τ)=|h(τ)|2=|Rxy(τ)|2𝑃𝜏superscript𝜏2superscriptsubscript𝑅𝑥𝑦𝜏2P(\tau)=|h(\tau)|^{2}=|R_{xy}(\tau)|^{2}italic_P ( italic_τ ) = | italic_h ( italic_τ ) | start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT = | italic_R start_POSTSUBSCRIPT italic_x italic_y end_POSTSUBSCRIPT ( italic_τ ) | start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT

A channel sounding signal is comprised of pulses transmitted at regular intervals. While this pulse signals are detected, these should be processed (Tomlinson2015, ). In this study, maximal-length pseudorandom noise (PN) sequences were utilized due to their favourable auto-correlation behavior, characterized by a high main peak and low side lobes. This features expedites detection of any multi-path component in the channel (Vizziello2022b, ).

3. Results and Channel Characterization

3.1. Simulation Results

The simulation model of this study was designed according to the implantable GC intrabody communication method in the frequency domain, allowing for the proper modeling of the dispersive behavior of tissues. Dielectric properties of each material were imported from (Hasgall2022, ) in the frequency region of interest. Simulations were performed on an Intel i7-11800H 2.3 GHz processor with 16 GB of RAM. A finer-grained mesh was selected, resulting into 140667 individual elements.

The simulation results are presented in Figure 3 and Figure 4. In Figure 3, three dimensional voltage and electric field distribution are represented at specific frequencies, such as 100 kHz, 1 MHz, and 2.5 MHz. Figure 3(a) shows the voltage distribution along the arm model. Due to the modeling of the galvanic coupled configuration, the voltage amplitude of the signal electrode and ground electrode of TX are 0.5 V and -0.5 V, respectively. In here, the maximum value of the induced voltage decreased when the frequency of signal is increasing. Figure 3(b) states the electric field distribution. It can be noticed that the electric field was confined at the muscle tissue, mostly due to the location of the electrodes. Furthermore, the maximum value of the induced electric field decreased with frequency.

Refer to caption
Figure 3. FEM-based simulation results for three dimensional multilayered human arm model: voltage distribution (a), and electric field distribution (b)

As shown in Figure 4, the voltage gain was calculated in the frequency range of 0 to 2.5 MHz, using the formula 20log10(VR/VT)20subscript10subscript𝑉𝑅subscript𝑉𝑇20\log_{10}(V_{R}/V_{T})20 roman_log start_POSTSUBSCRIPT 10 end_POSTSUBSCRIPT ( italic_V start_POSTSUBSCRIPT italic_R end_POSTSUBSCRIPT / italic_V start_POSTSUBSCRIPT italic_T end_POSTSUBSCRIPT ), where VRsubscript𝑉𝑅V_{R}italic_V start_POSTSUBSCRIPT italic_R end_POSTSUBSCRIPT and VTsubscript𝑉𝑇V_{T}italic_V start_POSTSUBSCRIPT italic_T end_POSTSUBSCRIPT show the received and the transmitted voltage, respectively. The gain varies between -54 dB to -44 dB, with simulated values of -52.2 dB, -43.75 dB, and -43.2 dB at 100 kHz, 1 MHz, and 2.5 MHz, respectively. When the frequency increases, the gain also increases, in a non-linear fashion. The received voltage increased up to around 345 kHz, while the transmitted voltage decreased, thus increasing the channel gain in the considered frequency range. From 345 kHz to 2.5 MHz, both the received and the transmitted voltages decreased, although with different slops. This circumstance demonstrated a slowly increasing channel gain at the higher frequency range. Naturally, other factors such as excitation signal, electrode features, tissue parameters, can also affect the gain (Callejon2018, ). Results showed that the galvanic coupled implantable system features a high-pass behavior with a cut-off frequency of 370 kHz.

Refer to caption
Figure 4. Simulated gain results

3.2. Experimental Measurements

Implantable GC intrabody communication were analyzed with experimental studies to measure the CIR. For this purpose, a linear generated polynominal PN sequences were transmitted with the degree of m=13𝑚13m=13italic_m = 13, the value being selected to cope with hardware limitations. These parameters correspond to a frequency bandwidth up to 2.5 MHz. The sequence was zero-padded to ensure a clearer CIR analysis. The generated signal, sampled at 5 MHz, was imported into the signal generator and applied to the input port of the isolation circuit on the TX side. The generated signal as measured on oscilloscope is visible in Figure 5. The amplitude of the signal was adjusted as a 1Vpp1subscript𝑉𝑝𝑝1V_{pp}1 italic_V start_POSTSUBSCRIPT italic_p italic_p end_POSTSUBSCRIPT.

Refer to caption
Figure 5. Measured transmitted signal

More than 40 measurements were performed under in indoor, room conditions. The generated signal was transmitted from the output port of the TX isolation circuit to the input port of the RX isolation circuit through the chicken breast, which served as a tissue model due to the human-like dielectric properties (Vizziello2024, ). The transmitted signal was recorded at the output port of the RX isolation circuit as a .csv file format. The measured signal is depicted in Figure 6.

Refer to caption
Figure 6. Measured received signal

The measured CIR is shown in Figure 7, as derived from experiments. Clearly, a high peak-to-off peak ratio was achieved.

Refer to caption
Figure 7. Measured channel impulse response

A comparison of measured and simulated gain results is offered in Figure 8, where, considering the relationship between gain and the channel frequency response (CFR) (Oliveira2016, ), the simulated frequency response is shown in the same graph with measurement samples. The CFR was derived from the measured CIR, whereas the red line represents simulation results reported for reference. It can be noted that, despite unavoidable differences due also to the different nature of data, a similar, high-pass trend can be observed in both cases. Furthermore, differences can be explained as the problem space was modeled with a heterogeneous multilayered cylinder while the experiments were performed with a single-layer chicken breast.

Refer to caption
Figure 8. Comparison between measurements and simulations

4. Conclusion

In this study, galvanic coupled based intrabody communication was investigated for implantable configurations in the 0 to 2.5 MHz frequency region. In this manner, the communication channel of the human tissue was modeled with both in-silico and experimental studies. In-silico investigations were performed with FEM-based simulation software, called Comsol Multiphysics. In simulations, the human arm was modeled as a multilayered cylindrical model with the realistic dielectric properties and thicknesses. Simulation results revealed that the gain increased with frequency. Also, both voltage and electric field distributions varied in the tissues. Experimental studies showed that the communication channel through the tissue is applicable for implantable technologies. A correlation relationship was observed between the TX and RX signals. Future work will include comprehensive analysis of both galvanic- and capacitive-coupled configurations of the TX and RX electrodes, and different parameters for simulation studies will be analyzed for possible intra-body communication channel. Other research directions include implementation of systems to increase the data rate in GC communications while keeping linear modulation schemes (Vizziello2016, ), as well as the development of compressive sensing methods to reduce the complexity of recovering data (Alesii2015, ). Furthermore, specific development of modulation and coding schemes considering multiple-input and multiple-output (MIMO) configurations (Kulsoom2018, ) are planned to be tailored for intra-body networks.

Acknowledgements

This study was partially supported by the Scientific and Technological Research Council of Türkiye (Project No: 1059B142201521), by the Ricerca Corrente funding from the Italian Ministry of Health to the Fondazione IRCCS Policlinico San Matteo (project 08073822), and by the European Union under the Italian National Recovery and Resilience Plan (NRRP) of NextGenerationEU, partnership on “Telecommunications of the Future” (PE00000001 - program “RESTART”).

References

  • (1) Ya-Li Zheng, Xiao-Rong Ding, Carmen Chung Yan Poon, Benny Ping Lai Lo, Heye Zhang, Xiao-Lin Zhou, Guang-Zhong Yang, Ni Zhao, and Yuan-Ting Zhang. 2014. Unobtrusive sensing and wearable devices for health informatics. IEEE Trans. Biomed. Eng. 61, 5 (May 2014), 1538-1554. https://doi.org/10.1109/TBME.2014.2309951
  • (2) Anna Vizziello, Pietro Savazzi, Renata Rojas Guerra, and Fabio Dell’Acqua. 2024. Experimental channel characterization of human body communication based on measured impulse response. IEEE Trans. Commun. (2024). To appear. https://doi.org/10.1109/TCOMM.2024.3370468
  • (3) Anna Vizziello, Pietro Savazzi, and Fabio Dell’Acqua. 2022. Data driven channel characterization of human body communication. In Proceedings of the 9th ACM International Conference on Nanoscale Computing and Communication. ACM Press, New York, NY, 1-2. https://doi.org/10.1145/3558583.3558869
  • (4) Fouad Maamir, Mohamed Guiatni, Smail Tedjini, Said Gaoua, Rachida Touhami, and Farid Abed. 2018. Optocoupler effect on in-vivo measurements characterization for human body communication. 2018 18th Mediterranean Microwave Symposium (MMS). IEEE, 345-348. https://doi.org/10.1109/MMS.2018.8612129
  • (5) Anna Vizziello, Laura Galluccio, Maurizio Magarini, Pietro Savazzi, Federico Biglioli, Federico Bolognesi, Francesca Talpo, Gerardo Biella, and Giovanni Magenes. 2022. An implantable system for neural communication and stimulation: Design and implementation. IEEE Commun. Mag. 60, 8 (August 2022), 74-79. https://doi.org/10.1109/MCOM.005.2101090
  • (6) Anna Vizziello, Maurizio Magarini, Pietro Savazzi, and Laura Galluccio. 2023. Intra-body communications for nervous system applications: Current technologies and future directions. Comput. Networks 227 (May 2023): 109718. https://doi.org/10.1016/j.comnet.2023.109718
  • (7) Matteo Pola, Vasiliki Giagka, Wouter A. Serdijn, Danilo Demarchi, and Amin Rashidi. 2023.Galvanic brain-coupled communication among freely floating micro-scale implants. 2023 IEEE Biomedical Circuits and Systems Conference (BioCAS). IEEE, Piscataway, NJ, 1-5. https://doi.org/10.1109/BioCAS58349.2023.10388903
  • (8) Niyazi İl, Kayhan Ateş, and Şükrü Özen. 2023. Electromagnetic field exposure to human head model with various metal objects at sub-6 GHz frequencies. Electromagn. Biol. Med. 42, 3 (2023), 114-122. https://doi.org/10.1080/15368378.2023.2220736
  • (9) Kayhan Ates, Hamza F. Carlak, and Sukru Ozen. 2022. Dosimetry analysis of the magnetic field of underground power cables and magnetic field mitigation using an electromagnetic shielding technique. Int. J. Occup. Saf. Ergon. 28, 3 (2022), 1672-1682. https://doi.org/10.1080/10803548.2021.1918923
  • (10) Kayhan Ateş and Şükrü Özen. 2019. Modelling of galvanic coupled intrabody communication due to finite element method. 2019 Medical Technologies Congress (TIPTEKNO). IEEE, 1-4. https://doi.org/10.1109/TIPTEKNO.2019.8895216
  • (11) Maoyuan Li, Yong Song, Yongtao Hou, Ning Li, Yurong Jiang, Muhammad Sulaman, and Qun Hao. 2019. Comparable investigation of characteristics for implant intra-body communication based on galvanic and capacitive coupling. IEEE Trans. Biomed. Circuits Syst. 13, 6 (December 2019): 1747-1758. https://doi.org/10.1109/TBCAS.2019.2940827
  • (12) Mitsuru Shinagawa, Jun Katsuyama, Kazuki Matsumoto, Shinya Hasegawa, Ryo Sugiyama, and Yuichi Kado. 2014. Noise analysis for intra-body communication based on parasitic capacitance measurement. Measurement 51 (May 2014), 206-213. https://doi.org/10.1016/j.measurement.2014.02.020
  • (13) Željka Lučev Vasić, Igor Krois, and Mario Cifrek. 2016. Effect of transformer symmetry on intrabody communication channel measurements using grounded instruments. Automatika 57, 1 (2016), 15-26. https://doi.org/10.7305/automatika.2016.01.1583
  • (14) Ai-ichiro Sasaki, Takako Ishihara, Nobutaro Shibata, Ryusuke Kawano, Hiroki Morimura, and Mitsuru Shinagawa. 2013. Signal-to-noise ratio analysis of a noisy-channel model for a capacitively coupled personal area network. IEEE Trans. Antennas Propag. 61, 1 (January 2013), 390-402. https://doi.org/10.1109/TAP.2012.2220110
  • (15) Nirmoy Modak, Mayukh Nath, Baibhab Chatterjee, Shovan Maity, and Shreyas Sen. 2022 Bio-physical modeling of galvanic human body communication in electro-quasistatic regime. IEEE Trans. Biomed. Eng. 69, 12 (December 2022), 3717-3727. https://doi.org/10.1109/TBME.2022.3176541
  • (16) Anna Vizziello, Pietro Savazzi, Giovanni Magenes, and Paolo Gamba. 2020. PHY design and implementation of a galvanic coupling testbed for intra-body communication links. IEEE Access 8 (2020), 184585-184597. https://doi.org/10.1109/ACCESS.2020.3029862
  • (17) HCPL-4562 and HCNW4562, 2017. High Bandwidth, Analog/Video Optocouplers Data Sheet. Retrieved February 18, 2024 from https://docs.broadcom.com/doc/AV02-1361EN
  • (18) Andreas F. Molisch. 2011. Wireless Communications (2nd. ed.). Wiley, Chichester, U.K.
  • (19) Jung-Hwan Hwang, Tae-Wook Kang, Youn-Tae Kim, and Seong-Ook Park. 2016. Measurement of transmission properties of HBC channel and its impulse response model. IEEE Trans. Instrum. Meas. 65, 1, (January 2016), 177-188. https://doi.org/10.1109/TIM.2015.2476236
  • (20) William J. Tomlinson, Fabian Abarca, Kaushik R. Chowdhury, Milica Stojanovic, and Christopher Yu. 2015. Experimental assessment of human-body-like tissue as a communication channel for galvanic coupling. 2015 IEEE 12th International Conference on Wearable and Implantable Body Sensor Networks (BSN) IEEE, Piscataway, NJ, 1-6. https://doi.org/10.1109/BSN.2015.7299424
  • (21) Philippe A. Hasgall, Fabienne Di Gennaro, Christian Baumgartner, Esra Neufeld, Bryn A. Lloyd, Marie-Christine Gosselin, Davnah Payne, Anja Klingenböck, and Niels Kuster. Feb 22, 2022. IT’IS Database for Thermal and Electromagnetic Parameters of Biological Tissues, version 4.1. Retrieved March 06, 2024 from http://itis.swiss/database
  • (22) Maria A. Callejón, Pablo del Campo, Javier Reina-Tosina, and Laura M. Roa. 2018. A parametric computational analysis into galvanic coupling intrabody communication. IEEE J. Biomed. Health. Inf. 22, 4 (July 2018), 1087-1096. https://doi.org/10.1109/JBHI.2017.2734939
  • (23) Thiago R. Oliveira, Fernando J. A. Andrade, Antonio A. M. Picorone, Haniph A. Latchman, Sergio L. Netto, and Moises V. Ribeiro. 2016. Characterization of hybrid communication channel in indoor scenario. Journal of Communication and Information Systems 31, 1 (September 2016), 224-235. https://doi.org/10.14209/jcis.2016.20
  • (24) Anna Vizziello, Pietro Savazzi, and Roberta Borra. 2016. Joint phase recovery for XPIC system exploiting adaptive kalman filtering. IEEE Communications Letters 20, 5 (2016), 922–925. https://doi.org/10.1109/LCOMM.2016.2542798
  • (25) Roberto Alesii, Piergiuseppe Di Marco, Fortunato Santucci, Pietro Savazzi, Roberto Valentini and Anna Vizziello. 2015. Multi-reader multi-tag architecture for UWB/UHF radio frequency identification systems. 2015 International EURASIP Workshop on RFID Technology (EURFID), Rosenheim, Germany, 2015, 28-35, https://doi.org/10.1109/EURFID.2015.7332381.
  • (26) Farzana Kulsoom, Anna Vizziello, Hassan N. Chaudhry and Pietro Savazzi. 2018. Pilot reduction techniques for sparse channel estimation in massive MIMO systems. 2018 14th Annual Conference on Wireless On-demand Network Systems and Services (WONS), Isola 2000, France, 2018, 111-116, https://doi.org/10.23919/WONS.2018.8311671.