MARS-MD: rejection based image domain material decomposition
Authors:
C. J. Bateman,
D. Knight,
B. Brandwacht,
J. Mc Mahon,
J. Healy,
R. Panta,
R. Aamir,
K. Rajendran,
M. Moghiseh,
M. Ramyar,
D. Rundle,
J. Bennett,
N. de Ruiter,
D. Smithies,
S. T. Bell,
R. Doesburg,
A. Chernoglazov,
V. B. H. Mandalika,
M. Walsh,
M. Shamshad,
M. Anjomrouz,
A. Atharifard,
L. Vanden Broeke,
S. Bheesette,
T. Kirkbride
, et al. (6 additional authors not shown)
Abstract:
This paper outlines image domain material decomposition algorithms that have been routinely used in MARS spectral CT systems. These algorithms (known collectively as MARS-MD) are based on a pragmatic heuristic for solving the under-determined problem where there are more materials than energy bins. This heuristic contains three parts: (1) splitting the problem into a number of possible sub-problem…
▽ More
This paper outlines image domain material decomposition algorithms that have been routinely used in MARS spectral CT systems. These algorithms (known collectively as MARS-MD) are based on a pragmatic heuristic for solving the under-determined problem where there are more materials than energy bins. This heuristic contains three parts: (1) splitting the problem into a number of possible sub-problems, each containing fewer materials; (2) solving each sub-problem; and (3) applying rejection criteria to eliminate all but one sub-problem's solution. An advantage of this process is that different constraints can be applied to each sub-problem if necessary. In addition, the result of this process is that solutions will be sparse in the material domain, which reduces crossover of signal between material images. Two algorithms based on this process are presented: the Segmentation variant, which uses segmented material classes to define each sub-problem; and the Angular Rejection variant, which defines the rejection criteria using the angle between reconstructed attenuation vectors.
△ Less
Submitted 14 February, 2018;
originally announced February 2018.
-
Alignment and resolution studies of a MARS scanner
Authors:
A. P. Butler,
P. H. Butler,
S. T. Bell,
G. Chelkov,
M. Demichev,
A. Gongadze,
S. Kotov,
D. Kozhevnikov,
U. Kruchonak,
I. Potrap,
P. Smolyanskiy,
A. Zhemchugov
Abstract:
The MARS scanner is designed for the x-ray spectroscopic study of samples with the aid of computer tomography methods. Computer tomography allows the reconstruction of slices of an investigated sample using a set of shadow projections obtained for different angles. Projections in the MARS scanner are produced using a cone x-ray beam geometry. Correct reconstruction in this scheme requires precise…
▽ More
The MARS scanner is designed for the x-ray spectroscopic study of samples with the aid of computer tomography methods. Computer tomography allows the reconstruction of slices of an investigated sample using a set of shadow projections obtained for different angles. Projections in the MARS scanner are produced using a cone x-ray beam geometry. Correct reconstruction in this scheme requires precise knowledge of several geometrical parameters of a tomograph, such as displacement of a rotation axis, x-ray source position with respect to a camera, and camera inclinations. Use of inaccurate parameters leads to a poor sample reconstruction. Non-ideal positioning of camera, x-ray source and cylindrical rotating frame (gantry) itself on which these parts are located, leads to the need for tomograph alignment. In this note we describe the alignment procedure that was used to get different geometrical corrections for the reconstruction. Also, several different estimations of the final spatial resolution for reconstructed images are presented.
△ Less
Submitted 23 March, 2015; v1 submitted 29 January, 2015;
originally announced January 2015.
-
Measurement of the energy resolution and calibration of hybrid pixel detectors with GaAs:Cr sensor and Timepix readout chip
Authors:
A. P. Butler,
P. H. Butler,
S. T. Bell,
G. A. Chelkov,
D. V. Dedovich,
M. A. Demichev,
V. G. Elkin,
M. I. Gostkin,
S. A. Kotov,
D. A. Kozhevnikov,
U. G. Kruchonak,
A. A. Nozdrin,
S. Yu. Porokhovoy,
I. N. Potrap,
P. I. Smolyanskiy,
M. M. Zakhvatkin,
A. S. Zhemchugov
Abstract:
This paper describes an iterative method of per-pixel energy calibration of hybrid pixel detectors with GaAs:Cr sensor and Timepix readout chip. A convolution of precisely measured spectra of characteristic X-rays of different metals with the resolution and the efficiency of the pixel detector is used for the calibration. The energy resolution of the detector is also measured during the calibratio…
▽ More
This paper describes an iterative method of per-pixel energy calibration of hybrid pixel detectors with GaAs:Cr sensor and Timepix readout chip. A convolution of precisely measured spectra of characteristic X-rays of different metals with the resolution and the efficiency of the pixel detector is used for the calibration. The energy resolution of the detector is also measured during the calibration. The use of per-pixel calibration allows to achieve a good energy resolution of the Timepix detector with GaAs:Cr sensor: 8% and 13% at 60 keV and 20 keV, respectively.
△ Less
Submitted 14 January, 2015;
originally announced January 2015.
-
Reducing beam hardening effects and metal artefacts using Medipix3RX: With applications from biomaterial science
Authors:
K. Rajendran,
M. F. Walsh,
N. J. A. de Ruiter,
A. I. Chernoglazov,
R. K. Panta,
A. P. H. Butler,
P. H. Butler,
S. T. Bell,
N. G. Anderson,
T. B. F. Woodfield,
S. J. Tredinnick,
J. L. Healy,
C. J. Bateman,
R. Aamir,
R. M. N. Doesburg,
P. F. Renaud,
S. P. Gieseg,
D. J. Smithies,
J. L. Mohr,
V. B. H. Mandalika,
A. M. T. Opie,
N. J. Cook,
J. P. Ronaldson,
S. J. Nik,
A. Atharifard
, et al. (6 additional authors not shown)
Abstract:
This paper discusses methods for reducing beam hardening effects using spectral data for biomaterial applications. A small-animal spectral scanner operating in the diagnostic energy range was used. We investigate the use of photon-processing features of the Medipix3RX ASIC in reducing beam hardening and associated artefacts. A fully operational charge summing mode was used during the imaging routi…
▽ More
This paper discusses methods for reducing beam hardening effects using spectral data for biomaterial applications. A small-animal spectral scanner operating in the diagnostic energy range was used. We investigate the use of photon-processing features of the Medipix3RX ASIC in reducing beam hardening and associated artefacts. A fully operational charge summing mode was used during the imaging routine. We present spectral data collected for metal alloy samples, its analysis using algebraic 3D reconstruction software and volume visualisation using a custom volume rendering software. Narrow high energy acquisition using the photon-processing detector revealed substantial reduction in beam hardening effects and metal artefacts.
△ Less
Submitted 20 November, 2013;
originally announced November 2013.
-
MARS spectral molecular imaging of lamb tissue: data collection and image analysis
Authors:
R Aamir,
A Chernoglazov,
C J Bateman,
A P H Butler,
P H Butler,
N G Anderson,
S T Bell,
R K Panta,
J L Healy,
J L Mohr,
K Rajendran,
M F Walsh,
N de Ruiter,
S P Gieseg,
T Woodfield,
P F Renaud,
L Brooke,
S Abdul-Majid,
M Clyne,
R Glendenning,
P J Bones,
M Billinghurst,
C Bartneck,
H Mandalika,
R Grasset
, et al. (13 additional authors not shown)
Abstract:
Spectral molecular imaging is a new imaging technique able to discriminate and quantify different components of tissue simultaneously at high spatial and high energy resolution. Our MARS scanner is an x-ray based small animal CT system designed to be used in the diagnostic energy range (20 to 140 keV). In this paper, we demonstrate the use of the MARS scanner, equipped with the Medipix3RX spectros…
▽ More
Spectral molecular imaging is a new imaging technique able to discriminate and quantify different components of tissue simultaneously at high spatial and high energy resolution. Our MARS scanner is an x-ray based small animal CT system designed to be used in the diagnostic energy range (20 to 140 keV). In this paper, we demonstrate the use of the MARS scanner, equipped with the Medipix3RX spectroscopic photon-processing detector, to discriminate fat, calcium, and water in tissue. We present data collected from a sample of lamb meat including bone as an illustrative example of human tissue imaging. The data is analyzed using our 3D Algebraic Reconstruction Algorithm (MARS-ART) and by material decomposition based on a constrained linear least squares algorithm. The results presented here clearly show the quantification of lipid-like, water-like and bone-like components of tissue. However, it is also clear to us that better algorithms could extract more information of clinical interest from our data. Because we are one of the first to present data from multi-energy photon-processing small animal CT systems, we make the raw, partial and fully processed data available with the intention that others can analyze it using their familiar routines. The raw, partially processed and fully processed data of lamb tissue along with the phantom calibration data can be found at [http://hdl.handle.net/10092/8531].
△ Less
Submitted 23 January, 2014; v1 submitted 18 November, 2013;
originally announced November 2013.