Abstract
Quartz-enhanced photoacoustic spectroscopy (QEPAS) is a sensitive gas detection technique which requires frequent calibration and has a long response time. Here we report beat frequency (BF) QEPAS that can be used for ultra-sensitive calibration-free trace-gas detection and fast spectral scan applications. The resonance frequency and Q-factor of the quartz tuning fork (QTF) as well as the trace-gas concentration can be obtained simultaneously by detecting the beat frequency signal generated when the transient response signal of the QTF is demodulated at its non-resonance frequency. Hence, BF-QEPAS avoids a calibration process and permits continuous monitoring of a targeted trace gas. Three semiconductor lasers were selected as the excitation source to verify the performance of the BF-QEPAS technique. The BF-QEPAS method is capable of measuring lower trace-gas concentration levels with shorter averaging times as compared to conventional PAS and QEPAS techniques and determines the electrical QTF parameters precisely.
Similar content being viewed by others
Introduction
Chemical gas phase analysis is significant in physics, chemistry, atmospheric science, space science, bioengineering, the life sciences as well as in medical applications1,2,3,4. Fundamental requirements for chemical sensing are sensitivity, selectivity and affordability. Quartz-enhanced photoacoustic spectroscopy (QEPAS), a variant of PAS, is useful as it meets these criterias5. Compared to PAS, the key innovation of QEPAS is to convert the acoustic wave induced by a modulated laser beam absorbed by the analyte into an electrical signal by means of a low-cost quartz tuning fork (QTF) based on its piezoelectric properties6. Such QTFs are mass-produced as a frequency reference and widely employed in cell phones as well as watches. QEPAS-based sensors combine the two main characteristics of PAS, that is, excitation wavelength independence and a detection sensitivity, which is proportional to the incident laser power7. QEPAS benefits from a set of unique piezoelectric properties of a QTF, such as a small size and a narrow acoustic resonance, resulting in an extremely high quality factor (10,000â15,000 at atmospheric pressure) as well as immunity to environmental acoustic noise6.
QEPAS-based sensors have been demonstrated for the detection of numerous inorganic and organic trace gases using a variety of laser sources in the wavelength range that covers the ultraviolet8, visible9, near-infrared10, mid-infrared11 and the terahertz spectral regions12,13. In recent years, several variants of QEPAS have been developed, including SO-QEPAS14, double acoustic micro-resonator (AmR) QEPAS15, intra-cavity QEPAS16,17 and evanescent-wave QEPAS18. The performance of different QEPAS-based sensor systems can be evaluated and compared based on the normalized noise equivalent absorption coefficient (NNEA) with respect to the optical laser power, the absorption line intensity and the detection bandwidth19. To date the lowest NNEA coefficient of QEPAS sensors, â¼10â10âcmâ1âWâHzâ1/2 was obtained for SF6 detection, which is one order of magnitude lower than the best NNEA values reported for conventional PAS results20,21.
Requirements for on-line monitoring of trace gases are uninterrupted operation, fast response, high selectivity and high detection sensitivity as well as compact size with a small gas cell. In the majority of conventional QEPAS sensor systems reported to date, a commercially available 33âkHz QTF resonance frequency is employed. Two important parameters of the QTF must be determined repetitively via an electric excitation method8, which defines the QEPAS sensor system calibration process. The first parameter is the resonance frequency f0 of the QTF. The QTF has a frequency resonance width of 3â5âHz at atmospheric pressure. Only frequency components in this resonance width can excite the vibrations of the QTF prongs efficiently6. The modulation of the laser radiation frequency (f) must accurately match with the QTF resonance frequency to obtain the highest signal amplitude. The second QTF parameter that must be measured is the Q-factor, as the magnitude of the Q-factor is proportional to the QTF signal amplitude22.
Both f0 and the Q-factor of a QTF are subject to the fabrication process23 and also depend on the operating environment, which includes the gas pressure, the temperature and the gas composition24. Thus in any trace-gas monitoring application, these two QTF parameters have to be determined prior to a QEPAS measurement and verified after a sensor operated for more than 24âh or if there is an environmental temperature change of >3â°C (ref. 25). Such a calibration process, which usually requires about 90âs to complete, interrupts a continuous QEPAS measurement. Furthermore, an inherently high Q-factor of the QTF provides the high sensitivity of conventional QEPAS, but also leads to a long accumulation time of the acoustic energy26. According to classical oscillator theory, the response time Ï is given by
The QTF response time is typically about 100âms and the measurement period can be up to 300â400âms for uncorrelated trace-gas concentration measurements. This time interval is too long to perform rapid trace-gas measurements which are important in a wide range of industrial applications26, especially for real-time monitoring of trace gases.
In this article, we report an innovative spectroscopic measurement technique that relies on the beat frequency (BF) signal between the QTF resonance frequency f0 and the laser modulation frequency f. The basic concept of BF-QEPAS requires that the laser modulation frequency has a frequency difference of Îf with respect to the QTF resonance frequency f0. For this condition, a BF-QEPAS signal with a period of Îf is generated and observed when the wavelength of the excitation laser is rapidly scanned across a targeted absorption line via a pulsed ramp current and the QTF signal is demodulated at the laser modulation frequency f. This technique can acquire the resonance frequency f0 and Q-factor of the QTF as well as the trace-gas concentration information in a measurement time of about 30âms.
Results
Theory of BF-QEPAS
The QTF can be modelled as a series resistanceâinductorâcapacitor circuit (based on the RLC ButterworthâVan Dyke model) using an electrical analogy of a mechanical damped oscillator27,28,29. This can be expressed by Newtonâs law of motion:
in which the resistance (R) represents the acoustic losses in the material and its environment, the inductor (L) represents the mass of the oscillator, the inverse capacity (1/C) and the electrical charge (q) represent the stiffness and displacement of the mechanical oscillator. The voltage (U) represents the force applied to the mechanical oscillator. The solution of equation (2) has the following form:
The term qs(t) accounts for the steady state response and qt(t) provides the transient response of the system. Figure 1 depicts a side-by-side principle comparison for conventional QEPAS and BF-QEPAS techniques using the steady and transient response of the QTF, respectively.
In conventional QEPAS, a slowly varying continuous acoustic wave causes forced vibrations of the QTF. The transient response of the QTF is neglected and only the steady state behaviour is taken into account due to the long averaging time (>300âms) used during which the transient response is averaged to be zero. In BF-QEPAS, an acoustic pulse induced by the target gas absorption is generated as a result of rapid wavelength scanning (>30âcmâ1âsâ1), which causes the prongs of the QTF to vibrate in a short period of time. Subsequently the QTF prong changes to a free vibration mode after the acoustic pulse terminates rather than the continuous forced vibrations caused by a continuous acoustic wave as in conventional QEPAS. The vibration energy will be dissipated via extrinsic and intrinsic QTF loss mechanisms30. At this point, the QTF is vibrating at its resonance frequency and not at the laser modulation frequency. The QTF signal is demodulated at the laser modulation frequency f. When the averaging time is short enough (<100âms) to provide a sufficient system detection bandwidth, a BF signal with an exponential decay envelope is generated from the QTF-transient response. The concentration of the trace gas, the resonance frequency and Q-factor of the QTF can be obtained by measuring the peak amplitude, the BF Îf and the decay time Ï of the BF-QEPAS signal, respectively.
Experimental apparatus
A schematic of the experimental set-up for demonstrating the performance of the BF-QEPAS spectrophone is shown in Fig. 2. A 1,368.7ânm distributed feedback diode laser (DFB-DL; NTT Electronics, Inc. Model NLK1E5E1AA) was employed as the excitation source. The laser wavelength can be tuned by varying its current and temperature. The laser temperature was stabilized by a temperature controller (THORLABS, Inc. Model TED 200C). The laser drive current consisted of three components: a direct current (d.c.), an alternating current (a.c.) and a ramp signal. The d.c. component from a current source (ILX Lightwave Corp. LDX 3220) determined the centre wavelength of the DFB-DL. The continuous sinusoidal a.c. component, generated from the function generator 1 (Stanford Model DS345) realized the wavelength modulation while a ramp signal, provided by a second function generator 2 (Tektronix, Inc. Model AFG 3022), scanned across the targeted trace-gas absorption line and then remained at a constant value to complete the induced BF-QEPAS signal detection. The scanning cycle is the sum of the scanning time and the waiting time. The repetition rate of the ramp signal is 12âHz. The waiting time of a single scanning cycle is 0.05âs. A fibre-coupled collimator (OZ optics Ltd. Model LPC-01) produced a diode laser beam with a 200âμm diameter, which was directed to an acoustic detection module (ADM) and avoided touching the QTF31. For comparison, the ADM with an on beam configuration (Methods section), which is widely used in conventional QEPAS sensors, is placed in the BF-QEPAS experimental set-up. Although there is no standing acoustic wave built up in the acoustic resonator with the excitation of an acoustic pulse, the behaviour of signal enhancement is expected as the non-resonant micro-tube can effectively confine the acoustic pulse21. The ADM is enclosed in a gas enclosure with two CaF2 windows. A vacuum pump and a pressure controller (MKS Instruments Inc. Model 649B13TS1M22M) controlled and maintained the pressure inside the ADM. The gas flow was set to 120âs.c.c.m. by a pin valve to minimize flow noise. The piezoelectric signal generated by the QTF was processed by a transimpedance amplifier with a 10âMΩ feedback resistor Rg and directed to a lock-in amplifier (LIA; Stanford Research Systems, Model SR830). The harmonic signal demodulated by the LIA was then transmitted to computer via a DAQ card (National Instruments, Inc., Model BNC-2110).
BF-QEPAS signal simulation and analysis
The performance of the BF-QEPAS technique was simulated using MATLAB software (Supplementary Note 1). In Fig. 3a, a QTF resonance frequency, f0=32,760âHz, was employed as the laser modulation frequency and the interaction time ta between the acoustic wave and the QTF was long enough (ta=10âs) to provide a continuous acoustic wave with a slow amplitude variation. The integration time was set to 300âms corresponding to a detection bandwidth of 0.417âHz. An excellent 1f-based conventional QEPAS signal can be obtained as shown in Fig. 3a. In Fig. 3b, the value of ta was decreased to 500âms and the laser modulation frequency was set to the QTF resonance frequency, f0. Two decaying tails, related to the peak and the valley of the first harmonic signal were observed when the LIA detection bandwidth was 0.417âHz as previously, which implies that the QTF-transient response gradually dominates in the excitation of a quasi-pulsed acoustic signal of the QTF. But the first decaying tail was influenced by the valley of the first harmonic signal and only the second free-decaying tail caused by the valley reflects a correct response time. When the interaction time was decreased to 10âms, an acoustic pulse was produced due to an interaction time that is much shorter than the QTF response time. When the modulation frequency (f=32,960âHz) was detuned from the QTF resonance frequency, the BF-QEPAS signal as shown in Fig. 3c was observed. In this case, the averaging time was reduced to 100âμs (a detection bandwidth of 1,250âHz) to provide a sufficient response bandwidth and to maintain efficient background noise suppression. In addition, the result showed that the amplitude of the BF-QEPAS signal is proportional to the strength of the acoustic force, which is determined by the target gas concentration. This gas concentration can be obtained by detecting the peak value of the BF-QEPAS signal. The value of f0 can be obtained since the BF is equal to the difference between f0 and f values:
where Ît is the time interval between two adjacent peaks or valleys of the BF signal. As the Q-factor can be defined as the ratio of the energy accumulated in a resonator to the energy loss per cycle, the Q can be calculated using equation (1), from which Ï is obtained by an exponential decay fitting of the BF-QEPAS envelope as in ring-down spectroscopy32. The simulated results, fs=32,762âHz and Qs=1,882, are consistent with the experimental results, f0=32,760âHz and Q=1,846, obtained by the electric excitation method8.
The experiment based on the BF-QEPAS technique was carried out using the corresponding detection bandwidths used in simulations and the apparatus depicted in Fig. 2. The experimental results, shown in Fig. 3dâf, are in excellent agreement with the simulated results in Fig. 3aâc. As the DFB-DL wavelength scan rate varied from 0.12âcmâ1âsâ1 to 3âcmâ1âsâ1, the conventional QEPAS signal changed from a standard first harmonic signal to an irregular signal with an exponentially decaying tail. A 200âHz beat signal with a response time of 0.01797, s was observed, when the wavelength was scanned at the rate of 72âcmâ1âsâ1 (equivalent to an interaction time of about 10âms) and modulated at a frequency of 32,960âHz,. The experimental results of the QTF parameter, ft=32,760âHz and Qt=1,848 are in excellent agreement with the simulated results.
Experimental optimization
The performance of the BF-QEPAS sensor was optimized using 2.5% water vapour in N2 (Methods section) at room temperature and atmosphere pressure. The selected H2O absorption line is located at 7,306.75âcmâ1 with a line intensity of 1.8 à 10â20âcmâmolâ1. The laser temperature was set to 9.65â°C and the d.c. current was set to 120âmA to reach the H2O target wavelength. The diode laser output power was 13.0âmW for these operating conditions. The filter slope and time constant of the LIA were set to 12âdB (Supplementary Fig. 2, Supplementary Note 2) and 100âμs (Supplementary Fig. 3 and Supplementary Note 2), corresponding to a detection bandwidth of 1,250âHz. The experimental results verified that these settings do not distort the line shape of the BF-QEPAS signal and maintain the narrow bandwidth required for noise reduction.
In a conventional QEPAS-based sensor, the largest amplitude of the 2f harmonic component is generally lower than the 1f signal amplitude, as shown in Fig. 4a,b. However, the 2f harmonic component is often used as the detection signal. This is due to the fact that the peak position of the 2f signal corresponds to that of the target absorption line33. Furthermore, the long response time Ï of the QTF requires that the conventional QEPAS-based sensors must be operated in a line-locking mode21. In the novel BF-QEPAS-based sensor, the BF-QEPAS signal from the first, second, third harmonic have the same wave shape. The optimum harmonic must be determined experimentally. The BF-QEPAS signals as shown in Fig. 4dâf from the first, second, third harmonics at a detection bandwidth of 1,250âHz were measured. Further evaluation tests of the BF-QEPAS were performed at the first harmonic, since the 1f BF-QEPAS signal has the largest amplitude and obtains an improved minimum detection limit.
The modulation frequency and depth have a significant impact on BF-QEPAS signals. Optimization of the modulation frequency was carried out with the same modulation depth (MD=29âmA) and an optimal wavelength scan rate, 36âcmâ1âsâ1 (Supplementary Fig. 4 and Supplementary Note 2). The optimization results are plotted in Fig. 5a. The measurement was implemented from 32,440âHz to 33,075âHz as the QTF does not respond to frequencies outside of this range. The BF-QEPAS signal amplitudes between 32,440 and 32,705âHz were analyzed, since the response curve of the BF-QEPAS signal amplitude is symmetric and centred at the QTF resonance frequency. Two factors determine the amplitude of the BF-QEPAS signals. The first factor is the response of the QTF to the acoustic wave frequency which is equal to the laser modulation frequency. The QTF is easier excited to vibrate, when the laser modulation frequency f approaches the QTF resonance frequency f0. The second factor is the value of the BF, Îf. A small BF causes a large interval between two constructive interferences. The longer the interval is, the more vibrating energy of the QTF is dissipated during the interval. These two factors work in unison to generate a maximum amplitude of the BF signal. In Fig. 5a, the amplitude of the BF-QEPAS signal increases in the frequency range from 32,440 to 32,640âHz as the response of the QTF is enhanced, while the amplitude decreases between 32,640 and 32,705âHz, because the second factor gradually dominates. In addition, the MD was optimized at the optimal modulation frequency, f=32,640âHz. A maximum BF-QEPAS signal was observed with a MD=29âmA as shown in Fig. 5b. Hence, the following evaluation tests of this sensor were performed with a MD=29âmA and f=32,640âHz.
For a comparison, the amplitude of the conventional 2f QEPAS signal as a function of frequency obtained with the same apparatus (Methods section) is also plotted in Fig. 5a. The peak of the 2f QEPAS signal at the resonance frequency is about 16% higher than the optimum amplitude of the BF-QEPAS signal. However, the averaging time of the 1f BF-QEPAS is three orders of magnitude shorter than that for 2f QEPAS.
Detection limit
The detection sensitivity of a trace-gas sensor is determined by the signal-to-noise-ratio. Previous experimental studies verified that the background noise of a conventional QEPAS-based sensor, , is dominated by thermal noise of the QTF34, which can be expressed as:
where kB is the Boltzmann constant, T is QTF temperature and Îfdet is the total detection bandwidth of the system for the two LIA detection channels. For a BF-QEPAS-based sensor, the thermal noise of the QTF cannot be calculated using equation (5). Instead, the noise power density should be integrated over the QTF resonant curve, since the LIA bandwidth is significantly wider than the resonant curve of the QTF. The equivalent noise detection bandwidth (ENBW) of the QTF is Ïf0/2Q. Hence the thermal noise of the QTF is expressed as
The total fundamental noise also includes the feedback resistor noise and should be integrated over the full lock-in detector bandwidth:
Hence the total noise can be expressed as:
The coefficient reflects the fact that the noise is calculated only for one detection channel. The theoretically estimated noise level is 13.9âμV. The experimentally measured noise level was in the 11â15âμV range, which is in excellent agreement with the theoretically estimated result.
The linear response of the BF-QEPAS sensor was investigated by measuring the second peak amplitude of the 1f-based BF-QEPAS (Supplementary Fig. 5 and Supplementary Note 3) for different H2O concentrations (Methods section) with the optimized parameters reported above. The results were plotted in Fig. 6. The linearity of the sensor response to the H2O concentration levels was confirmed by the high R2 value (>0.999) of the linear fitting.
The noise-equivalent concentration (NEC) of the BF-QEPAS-based sensor for 0.1âms integration time was estimated to be about 5.9âp.p.m. based on the background noise level and the data depicted in Fig. 6. Hence, the NNEA coefficient for H2O is 2.45 Ã 10â9âcmâ1âWâHzâ1/2 (Methods section), which is one order of magnitude higher than that obtained by the conventional QEPAS technique.
An AllanâWerle deviation analysis was performed in order to evaluate the long term stability of the BF-QEPAS sensor, as shown in Fig. 7. The ADM was filled with pure N2 at atmospheric pressure. The sensor system was operated at room temperature in the BF-QEPAS mode and the laser wavelength was scanned at the optimized rate. The AllanâWerle deviation indicates that the thermal noise of the feedback resistor (Rg) and the QTF are the dominant noise sources and the BF-QEPAS-based sensor allows data averaging without a base line or sensitivity drift for a time scale of >70âs.
Discussion
Three semiconductor laser sources were chosen as the excitation source with the same ADM to verify the performance of the BF-QEPAS technique. The laser beam from the three different excitation sources was focused by suitable lenses that ensure that the beam passes through the AmR tubes and QTF without touching. The experimental parameters, such as pressure, scan rate and modulation depth of the laser current were also optimized (Supplementary Figs 6 and 7 and Supplementary Note 4).
A side-by-side comparison of detection limits for the BF-QEPAS-based sensors and QEPAS-based sensors is shown in Table 1. The NEC was normalized to a 1âs integration time to facilitate inter-comparison. With three different excitation sources, the experimental results of the frequency and Q-factor of the QTF are consistent with the value obtained by conventional electric excitation method. The BF-QEPAS technique has an obviously lower NEC and NNEA value than the conventional QEPAS technique for the fast relaxing molecule (H2O), while for the slow relaxing molecules (CH4 and CO), the NEC and NNEA for BF-QEPAS technique are somewhat better than the conventional QEPAS technique. The BF-QEPAS technique similar to the conventional 2f wavelength modulation-based QEPAS technique provides high selectivity together with background-free signals. Its selectivity is mainly determined by an initial rapid wavelength scanning crossing the target absorption line. The target absorption line produces an acoustic pulse to push the QTF prongs to vibrate, ensuring high selectivity. Subsequently the high Q-factor of the QTF vibrating freely at its resonant frequency effectively suppresses the background noise from the non-resonant acoustic pulse. Only the intensity information of the acoustic pulse remains in the BF-QEPAS signal. As a result, the BF-QEPAS technique is dominated by the QTF thermal noise using either 1f or 2f detection.
In summary, an innovative modification of quartz enhanced photoacoustic spectroscopy PAS, BF-QEPAS, was demonstrated for the first time, which can simultaneously measure trace-gas concentrations, the QTF resonance frequency and Q-factor, thereby avoiding the calibration process and ensuring continuous real-time trace-gas monitoring. The BF-QEPAS technique acquires spectral data by detecting the beat signal generated by the transient response of the QEPAS system at a much faster rate than those reported in previous publications. The results show that this new technique is capable of providing reduced data acquisition times and improved detection sensitivity than those reported for conventional QEPAS, especially for fast relaxing molecules. The BF-QEPAS technique can be applied to all QEPAS-based sensor systems without any hardware changes of a conventional QEPAS sensor system. Further improvement of the detection sensitivity can be achieved either by optimizing the geometrical parameters of the non-resonant micro-tube or by combining with the overtone resonance mode of a custom fabricated QTF.
Methods
Preparation of an on-beam configuration ADM
The on-beam configuration ADM consists of a standard commercial QTF and a set of AmRs (Supplementary Fig. 1). Two identical metallic tubes of 4.0âmm in length and 0.8âmm inner diameter were placed close to, but not touching, the QTF, with 20âμm gaps on both sides of the QTF. The vertical distance between the centre of the tubes and the QTF opening was 0.7âmm. In this case, the Q-factor of the QTF decreased to 1,850 as a result of coupling between the QTF and the AmRs21,35. The QTF and the AmR were mounted on an aluminum plate.
Preparation of water vapour with different concentrations
The highest water vapour mixing ratio reported in this paper, 2.524%, was obtained by controlling pure N2 passing through a humidifier (PermSelect, PDMSXA-2500) at a stable flow rate, constant temperature and pressure. The other water vapour mixing ratios were generated by mixing H2O with pure N2 at different flow rate ratios. Two mass flow metres (Alicat Scientific, Inc. Model M-2SLPM-D/5M) were used to control the flow rate. The concentration of water vapour was determined by direct absorption spectroscopy using a multi-pass gas cell with an effective path-length of 57.6âm described in ref. 36.
Calculation of NNEA
The NNEA coefficient can be determined by the following equation:
αmin is the minimum detectable absorption coefficient, which can be calculated using the HITRAN database (http://www.hitran.com), if the absorption wavelength of the target trace gas and the NEC of the sensor are known. P is the laser power. The ENBW is the ENBW which should be 1,250âHz, when the time constant and filter slope of the LIA are set to 100âμs and 12âdB, respectively. Hence, the NNEA=2.45 à 10â9âcmâ1âWâHzâ1/2 can be calculated if αmin=9.3 à 10â6âcmâ1 and P=13.04âmW.
QTF response profile as a function of frequency
The apparatus mentioned in the experimental apparatus section was used to measure the QTF response profile as a function of frequency by measuring the peak value of the conventional 2f QEPAS signal at room temperature and atmosphere pressure with 2.5% water vapour. The time constant and filter slope of the LIA were set to 300âms and 12âdB. The wavelength scanning rate was 0.18âcmâ1âsâ1.
Data availability
The authors declare that all data supporting the findings of this study can be found within the paper and its Supplementary Information Files. Additional data supporting the findings of this study are available from the corresponding author (L.D.) upon reasonable request.
Additional information
How to cite this article: Wu, H. et al. Beat frequency quartz-enhanced photoacoustic spectroscopy for fast and calibration-free continuous trace-gas monitoring. Nat. Commun. 8, 15331 doi: 10.1038/ncomms15331 (2017).
Publisherâs note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
Anthony, K. W. et al. Methane emissions proportional to permafrost carbon thawed in Arctic lakes since the 1950s. Nat. Geosci. 9, 679â682 (2016).
Jin, W., Cao, Y., Yang, F. & Ho, H. L. Ultra-sensitive all-fibre photothermal spectroscopy with large dynamic range. Nat. Commun. 6, 7767 (2015).
Royer, S. J. et al. A high-resolution time-depth view of dimethylsulphide cycling in the surface sea. Sci. Rep. 5, 17729 (2015).
Ren, W., Jiang, W. & Tittel, F. K. Single-QCL-based absorption sensor for simultaneous trace-gas detection of CH4 and N2O. Appl. Phys. B 117, 245â251 (2014).
Kosterev, A. A., Bakhirkin, Y. A., Curl, R. F. & Tittel, F. K. Quartz-enhanced photoacoustic spectroscopy. Opt. Lett. 27, 1902â1904 (2002).
Friedt, J.-M. & Carry, Ã. Introduction to the quartz tuning fork. Am. J. Phys. 75, 415â422 (2007).
Miklós, A., Hess, P. & Bozóki, Z. Application of acoustic resonators in photoacoustic trace gas analysis and metrology. Rev. Sci. Instrum. 72, 1937â1955 (2001).
Dong, L. et al. Ultra-sensitive carbon monoxide detection by using EC-QCL based quartz-enhanced photoacoustic spectroscopy. Appl. Phys. B 107, 275â283 (2012).
Köhring, M., Böttger, S., Willer, U. & Schade, W. LED-absorption-QEPAS sensor for biogas plants. Sensors 15, 12092â12102 (2015).
Wu, H. et al. Quartz enhanced photoacoustic H2S gas sensor based on a fiber-amplifier source and a custom tuning fork with large prong spacing. Appl. Phys. Lett. 107, 111104 (2015).
Cao, Y. et al. Simultaneous atmospheric nitrous oxide, methane and water vapor detection with a single continuous wave quantum cascade laser. Opt. Express 23, 2121â2132 (2015).
Borri, S. et al. Terahertz quartz enhanced photo-acoustic sensor. Appl. Phys. Lett. 103, 021105 (2013).
Spagnolo, V. et al. THz quartz-enhanced photoacoustic sensor for H2S trace gas detection. Opt. Express 23, 7574â7582 (2015).
Zheng, H. et al. Single-tube on-beam quartz-enhanced photoacoustic spectroscopy. Opt. Lett. 41, 978â981 (2016).
Dong, L. et al. Double acoustic micro-resonator quartz-enhanced photoacoustic spectroscopy. Opt. Lett. 39, 2479â2482 (2014).
Borri, S. et al. Intracavity quartz-enhanced photoacoustic sensor. Appl. Phys. Lett. 104, 091114 (2014).
Patimisco, P. et al. High finesse optical cavity coupled with a quartz-enhanced photoacoustic spectroscopic sensor. Analyst 140, 736â746 (2015).
Cao, Y., Jin, W., Ho, L. H. & Liu, Z. B. Evanescent-wave photoacoustic spectroscopy with optical micro/nano fibers. Opt. Lett. 37, 214â216 (2012).
Yi, H. et al. Application of a broadband blue laser diode to trace NO2 detection using off-beam quartz-enhanced photoacoustic spectroscopy. Opt. Lett. 36, 481â483 (2011).
Spagnolo, V. et al. Part-per-trillion level detection of SF6 using a single-mode fiber-coupled quantum cascade laser and a quartz enhanced photoacoustic sensor. Opt. Lett. 37, 4461â4463 (2013).
Dong, L., Kosterev, A. A., Thomazy, D. & Tittel, F. K. QEPAS spectrophones: design, optimization, and performance. Appl. Phys. B 100, 627â635 (2010).
Cremer, J. W., Thaler, K. M., Haisch, C. & Signorell, R. Photoacoustics of single laser-trapped nanodroplets for the direct observation of nanofocusing in aerosol photokinetics. Nat. Commun. 7, 10941 (2016).
Spagnolo, V. et al. THz quartz-enhanced photoacoustic sensor employing a quantum cascade laser source. Proc. SPIE 8993, 899320 (2014).
Kosterev, A. A., Bakhirkin, Y. A. & Tittel, F. K. Ultrasensitive gas detection by quartz-enhanced photoacoustic spectroscopy in the fundamental molecular absorption bands region. Appl. Phys. B 80, 133â138 (2005).
Mordmüller, M., Köhring, M., Schade, W. & Willer, U. An electrically and optically cooperated QEPAS device for highly integrated gas sensors. Appl. Phys. B 119, 111â118 (2015).
Kosterev, A. A. et al. QEPAS detector for rapid spectral measurements. Appl. Phys. B 100, 173â180 (2010).
Patimisco, P. et al. A quartz enhanced photo-acoustic gas sensor based on a custom tuning fork and a terahertz quantum cascade laser. Analyst 139, 2079â2087 (2014).
Spagnolo, V. et al. Quartz tuning forks with novel geometries for optoacoustic gas sensing. Proc. SPIE 9755, 97752L (2016).
Waszczuk, K., Piasecki, T., Nitsch, K. & Gotszalk, T. Application of piezoelectric tuning forks in liquid viscosity and density measurements. Sens. Actuators B Chem. 160, 517â523 (2011).
Sampaolo, A. et al. Quartz-enhanced photoacoustic spectroscopy exploiting tuning fork overtone modes. Appl. Phys. Lett. 107, 231102 (2015).
Ma, Y. et al. HCl ppb-level detection based on QEPAS sensor using a low resonance frequency quartz tuning fork. Sens. Actuators B Chem. 233, 388â393 (2016).
Dupré, P. Birefringence-induced frequency beating in high-finesse cavities by continuous-wave cavity ring-down spectroscopy. Phys. Rev. A 92, 053817 (2015).
Schilt, S., Thevenaz, L. & Robert, P. Wavelength modulation spectroscopy: combined frequency and intensity laser modulation. Appl. Opt. 42, 6728â6738 (2003).
Liu, K. et al. Trace gas detection based on off-beam quartz enhanced photoacoustic spectroscopy: optimization and performance evaluation. Rev. Sci. Instrum. 81, 103103 (2010).
Wu, H. et al. Position effects of acoustic micro-resonator in quartz enhanced photoacoustic spectroscopy. Sens. Actuators B Chem. 206, 364â370 (2015).
Yu, Y., Sanchez, N. P., Griffin, R. J. & Tittel, F. K. CW EC-QCL-based sensor for simultaneous detection of H2O, HDO, N2O and CH4 using multi-pass absorption spectroscopy. Opt. Express 24, 10391â10401 (2016).
Acknowledgements
This material is based upon work supported by National Natural Science Foundation of China (Grants #61622503, 61575113, 61275213). Frank K. Tittel acknowledges Grant C-0586 by the Welch Foundation.
Author information
Authors and Affiliations
Contributions
H.W. and L.D. designed the experiments. F.K.T. supervised the project. H.W., Y.Y., W.M., L.Z., H.Z. performed the reported experiments. L.D., W.Y., L.X. and S.J. conducted the experimental analysis. H.W., L.D. and F.K.T. prepared the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing financial interests.
Supplementary information
Supplementary Information
Supplementary Figures, Supplementary Notes and Supplementary References. (PDF 948 kb)
Rights and permissions
This work is licensed under a Creative Commons Attribution 4.0 International License. The images or other third party material in this article are included in the articleâs Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/
About this article
Cite this article
Wu, H., Dong, L., Zheng, H. et al. Beat frequency quartz-enhanced photoacoustic spectroscopy for fast and calibration-free continuous trace-gas monitoring. Nat Commun 8, 15331 (2017). https://doi.org/10.1038/ncomms15331
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/ncomms15331
This article is cited by
-
Cavity-enhanced photoacoustic dual-comb spectroscopy
Light: Science & Applications (2024)
-
Quartz-enhanced multiheterodyne resonant photoacoustic spectroscopy
Light: Science & Applications (2024)
-
Dual-comb optomechanical spectroscopy
Nature Communications (2023)
-
Mid-infrared photoacoustic gas monitoring driven by a gas-filled hollow-core fiber laser
Scientific Reports (2021)
-
Low noise, open-source QEPAS system with instrumentation amplifier
Scientific Reports (2019)