Abstract
Spectroscopy of whispering-gallery mode microresonators has become a powerful scientific tool, enabling the detection of single viruses, nanoparticles and even single molecules. Yet the demonstrated timescale of these schemes has been limited so far to milliseconds or more. Here we introduce a scheme that is orders of magnitude faster, capable of capturing complete spectral snapshots at nanosecond timescalesâcavity ring-up spectroscopy. Based on sharply rising detuned probe pulses, cavity ring-up spectroscopy combines the sensitivity of heterodyne measurements with the highest-possible, transform-limited acquisition rate. As a demonstration, we capture spectra of microtoroid resonators at time intervals as short as 16âns, directly monitoring submicrosecond dynamics of their optomechanical vibrations, thermorefractive response and Kerr nonlinearity. Cavity ring-up spectroscopy holds promise for the study of fast biological processes such as enzyme kinetics, protein folding and light harvesting, with applications in other fields such as cavity quantum electrodynamics and pulsed optomechanics.
Similar content being viewed by others
Introduction
The remarkable sensitivity of whispering-gallery mode (WGM) microresonators1,2 stems from their small mode volume (Vâ¼100âμm3), which enhances the electric field associated with every photon, and their ultra-high quality factor (Qâ¼108), which enables very long cavity lifetimes. This leads to the enhancement of the interaction with any object within the mode3,4,5,6,7 by a factor proportional to Q/V. As a result, the presence of even nanoscale objects within the evanescent field of the WGM can lead to detectable variations in its spectrum8. Specifically, these variations include broadening (due to shortening of the cavity lifetime)9,10, splitting (due to induced crosstalk between its two counter-propagating modes)4,8 or a shift (due to modification of the optical path length)5,11. The typical scale of all these spectral variations is of the order of a few MHz or more, in principle allowing their detection on a timescale of less than a microsecond.
However, although there are fast methods to separately measure splitting6,12, broadening10 or a shift13, the only method for obtaining the full spectrum had so far been laser scanning. Laser scanning is slow not only because of technical reasons. It is inherently slow since to accurately capture the cavity spectrum, the laser must scan over the cavity resonance (of width 2κ) during a time long enough for the intracavity field to continuously maintain a steady state. The timescale for this duration is determined by the cavity lifetime Ï (=1/4Ïκ), meaning that the scan rate needs to be significantly slower than κ/Ï. Accordingly, to acquire a spectral scan over a total bandwidth B, the total scan time needs to exceed NÏ, with N=B/κ representing the effective number of frequency points acquired during the scan. Yet this scan time is not optimalâthe fundamental limit for the duration of such a spectral measurement is defined by the timeâbandwidth uncertainty relation (or the Fourier limit) and is set by the desired spectral resolution (κ) to be roughly Ï. In other words, a scan measurement is at least N times slower than a Fourier-limited one.
Here we present cavity ring-up spectroscopy (CRUS) as an ultrafast method for acquiring complete spectral snapshots of WGM resonators at the minimal timescale set by the Fourier limit. CRUS relies on the abrupt turn-on of far-detuned probe laser pulses (δâ«2κ), which are coupled to the WGM microresonator (a microtoroid14 in our case) through a tapered optical fibre15. At such a large detuning, most of the light is transmitted through the fibre without entering the cavity. However, the short rise time tr of the pulse corresponds to a broad spectral distribution of bandwidth Bâ¼1/tr. Accordingly, if the cavity resonance lies within this bandwidth (δ<B), this turn-on leads to the build-up of a small, transient field in the cavity. This field then radiates back into the fibre, where it interferes with the transmitted field, creating a ring-up signal. The spectrum can now be retrieved with a resolution of â¼Îº (or better) by acquiring this ring-up signal during a time of Ï (or longer).
CRUS therefore enables monitoring kinetics and events in the nanosecond regime, drastically expanding the capabilities demonstrated in previous schemes3,4,5,6, which have focused on the slow detection of long-term events, such as binding of viruses or nanoparticles to the surface of the microresonator. As a proof of principle, we demonstrate how CRUS captures variations of WGM spectra due to different nonlinear mechanisms, induced by a separate pump laser. In particular, we directly observe the spectrum of a resonator oscillating mechanically at 12.5âMHz.
Results
Principles of CRUS
The configuration of CRUS, outlined in Fig. 1a, is reminiscent of the well-known method of cavity ring-down spectroscopy (CRDS), in which the field emitted from a cavity is directly measured by abruptly turning off a resonant probe16. CRDS is widely used as a sensitive method for absorption spectroscopy of gaseous samples in FabryâPérot resonators. Since CRDS captures only the intensity of the ring-down signal, it provides no knowledge of the centre frequency of the resonance, and therefore cannot be used to detect spectral shifts (nor does it enable reconstruction of asymmetric spectra). Accordingly, its application to WGM microresonators has so far been limited to Q-factor measurements14.
In contrast, in CRUS, the far-detuned probe, which continues in the fibre, serves as a local oscillator as it interferes with the emitted cavity field. The result is therefore a robust heterodyne measurement of both the amplitude and the phase of the cavity field, which allows the unambiguous extraction of the complete spectrum, including its centre frequency. The difference between CRUS and CRDS is therefore analogous to the difference between a hologram, which captures both the amplitude and the directionality of the object wavefront, to a picture, which captures only its intensity. In this view, the large detuning between the probe and the resonance is analogous to the large angular separation between the reference beam and the object in an off-axis hologram, which prevents overlap between positive and negative frequencies (the object and its conjugate) during the reconstruction.
Spectrum retrieval
Figure 1b presents the measured ring-up signal, defined as s(t)=(|y|2â|x|2)/(x), where x and y are the input and output fields, respectively. The main features of the WGM resonance already appear in this signal. Accordingly, s(t) enables complete retrieval of the transmission spectrum T(Ï)=|Y(Ï)/X(Ï)|2, which by the final value theorem is identical to that obtained by frequency scanning (where X,Y are the Fourier transforms of x,y). This is done by noting that the output field results from interference between the input field and the emission from the cavity, that is, Y(Ï)=X(Ï)+X(Ï)H(Ï), where H(Ï) is the cavityâs transfer function. By Fourier transforming the ring-up signal s(t) and using the fact that due to the large detuning H(Ï) contains only positive frequencies, we get H(Ï)=S(Ï)/X(Ï), finally yielding (see Methods)
Figure 1c presents the spectrum obtained from a single-shot, 500-ns pulse performed on a resonance with Q=1 Ã 108 at 780ânm, showing an excellent agreement with the much-slower (5âms) single-shot scan measurement. The fundamental origin of the drastic difference between the speeds of CRUS and a regular scan is clearly illustrated in Fig. 2, which simulates scans of the spectrum of Fig. 1c at various speeds. The fact that CRUS is a transform-limited measurement, whereas a scan measurement is not, is clearly exhibited by the ringing effects that appear at short scan times17.
The ability to rapidly capture the entire spectrum also makes CRUS more resilient to various noise mechanisms (such as amplitude noise) that typically occur at lower frequencies and can distort a scan measurement, which samples different parts of the spectrum at different times. Indeed, as evident from Fig. 1c, the signal-to-noise ratio of CRUS is comparable to that of the much-slower scan. CRUS also has the inherent advantage of heterodyne measurements, which enable shot-noise limited sensitivities (as shown in the extraction of decay rates from beat-note signals generated either by sweeping18 or shifting the probe frequency19,20, or by non-linearity-induced shifts of the resonance itself21). Sensitive detection of weak fields is extremely important with WGM microresonators, in which the intracavity power must be kept very low, to prevent the onset of undesired non-linearities.
Demonstration of rapid spectral acquisition using CRUS
As a first demonstration of the potential of CRUS for capturing ultrafast phenomena, we applied it to fully characterize the non-linearities of the microtoroid. For this purpose, we used a modulated pump laser (see Methods) to induce spectral variations in the probed WGM resonance through thermal22, Kerr23,24 and optomechanical effects25.
Figure 3a shows the response of the microtoroid to an abrupt turn-on of a 300-kHz modulated pump, as exhibited by the variations in the centre frequency of a 54âMHz WGM resonance (full width at half maximum), captured by CRUS at intervals of 200âns. As demonstrated in previous works22,26, for modulation frequencies below â¼1âMHz, the non-linearity of the microtoroid is dominated by its thermal response, which is governed by two processes with different timescales22,26. The short timescale is set by the conduction rate of heat from the optical mode to the rest of the cavity, and the longer corresponds to dissipation of heat from the entire cavity to the silicon substrate through the pillar. The high rate of CRUS allows the accurate extraction of these timescales directly from the cavityâs temporal response to abrupt turn-off of the pump, as shown in Fig. 3b.
The CRUS measurements also reveal a phase lag of 50±1° between the cavity oscillations and the pump modulation (inset of Fig. 3a), which is a result of the interplay between the two capacitive thermal mechanisms and the much weaker, frequency-independent Kerr nonlinearity. By measuring this phase shift as a function of modulation frequency, we find that the heating per intracavity energy is 67âK/nJ for the slow mechanism and 7âK/nJ for the fast mechanism, in good agreement with theory27.
Figure 3c shows the cavityâs centre frequency response to abrupt turn-on of a pump laser modulated at 5.5âMHz, extracted from spectral snapshots taken every 37âns. At this high frequency, the non-linearity of the cavity is dominated by the Kerr effect23,26 (superimposed on the slower thermal drift), with the corresponding instantaneous response exhibited by a nearly zero phase shift (4±1°) with respect to the pump modulation (see inset).
As a final demonstration of the capabilities of CRUS, we directly monitor optomechanical oscillations25,28,29 of the microtoroid at a rate much faster than the oscillation frequency. Such pulsed measurements are of particular interest in the context of quantum non-demolition measurements of mechanical motion30. For this purpose, we tuned the pump to excite the second-order crown mode of the microtoroid (inset of Fig. 4a).
Figure 4a presents the measured time-varying spectrum of a 90-MHz WGM resonance, captured at intervals of 25âns. The graph clearly displays the variations of the resonance resulting from the oscillations of the microtoroidâs circumference at 8âMHz. Figure 4b displays the centre frequency shift extracted from CRUS spectra of another microtoroid, in which the frequency of the second-order crown mode was 12.5âMHz. Here the spectra were taken at intervals of 16âns. The extracted phase difference between the mechanical oscillations and the pump is 89±1°, in agreement with the expected value of Ï/2 for a resonantly driven oscillator. The resonant behaviour of these oscillations is also exhibited by the fact that they continue after the pump is abruptly turned off. By measuring the decay rate of these oscillations, we obtain a mechanical quality factor of 300. These results provide the first direct observation of optomechanical oscillations by continuous monitoring of the entire WGM spectrumâa task far beyond the ability of laser scanning.
Discussion
By acquiring maximal knowledge on the spectrum at the shortest possible timescale, CRUS introduces the promising field of optical sensing with WGM microresonators into the nanosecond regime. As a specific example, Fig. 4c presents a numerical simulation of a nanoparticle with a radius of 75ânm flying by the microtoroid. As evident in this example, CRUS enables the simultaneous observation of the shift, splitting and broadening induced by the interaction with the nanoparticle. Looking forward, these abilities open the path for the study of ultrafast phenomena in various fields, such as micromechanical vibrations30,31, cavity quantum electrodynamics32 and fast biological processes such as enzyme kinetics33, light harvesting34 and protein folding35.
Methods
Laser characteristics
The probe laser (at 780ânm) was intensity modulated with a LiNbO3 electro-optic modulator, yielding pulses with rise times typically below 1âns. The pulse length was set by the required spectral resolution, and the cavity lifetime was tuned to be significantly shorter by increasing the coupling to the tapered fibre. To allow separation from the probe, the pump laser was tuned to a WGM resonance at 770ânm (different from the probed resonance) and set to be counter-propagating to the probe.
Assumptions in the retrieval formula
Owing to the large detuning, the term quadratic in the cavity emission is neglected in obtaining equation (1), as it is much weaker than the probe. In addition, it was assumed that the input field x has a constant phase, and accordingly it was treated as real. This assumption typically holds except, in some cases, during the pulse rise. If the pulse rise time is significantly shorter than 뫉1, this effect can still be neglected. Otherwise, accurate retrieval of the spectrum requires knowledge of this transient phase, which depends on the specific modulation technique used (in our case a Z-cut modulator with a chirp factor of 0.7).
Additional information
How to cite this article: Rosenblum, S. et al. Cavity ring-up spectroscopy for ultrafast sensing with optical microresonators. Nat. Commun. 6:6788 doi: 10.1038/ncomms7788 (2015).
References
Braginsky, V., Gorodetsky, M. & Ilchenko, V. Quality-factor and nonlinear properties of optical whispering-gallery modes. Phys. Lett. A 137, 393â397 (1989).
Vahala, K. J. Optical microcavities. Nature 424, 839â846 (2003).
Vollmer, F., Arnold, S. & Keng, D. Single virus detection from the reactive shift of a whispering-gallery mode. Proc. Natl Acad. Sci. USA 105, 20701â20704 (2008).
Zhu, J. et al. On-chip single nanoparticle detection and sizing by mode splitting in an ultrahigh-q microresonator. Nat. Photonics 4, 46â49 (2009).
Lu, T. et al. High sensitivity nanoparticle detection using optical microcavities. Proc. Natl Acad. Sci. USA 108, 5976â5979 (2011).
He, L., Ãzdemir, S. K., Zhu, J., Kim, W. & Yang, L. Detecting single viruses and nanoparticles using whispering gallery microlasers. Nat. Nanotechnol. 6, 428â432 (2011).
Dantham, V. R. et al. Label-free detection of single protein using a nanoplasmonic-photonic hybrid microcavity. Nano Lett. 13, 3347â3351 (2013).
Mazzei, A. et al. Controlled coupling of counterpropagating whispering-gallery modes by a single rayleigh scatterer: a classical problem in a quantum optical light. Phys. Rev. Lett. 99, 173603 (2007).
Shao, L. et al. Detection of single nanoparticles and lentiviruses using microcavity resonance broadening. Adv. Mater. 25, 5616â5620 (2013).
Truong, G.-W. et al. Frequency-agile, rapid scanning spectroscopy. Nat. Photonics 7, 532â534 (2013).
Arnold, S., Khoshsima, M., Teraoka, I., Holler, S. & Vollmer, F. Shift of whispering-gallery modes in microspheres by protein adsorption. Opt. Lett. 28, 272â274 (2003).
Knittel, J., Swaim, J. D., McAuslan, D. L., Brawley, G. A. & Bowen, W. P. Back-scatter based whispering gallery mode sensing. Sci. Rep. 3, 2974 (2013).
Keng, D., McAnanama, S., Teraoka, I. & Arnold, S. Resonance fluctuations of a whispering gallery mode biosensor by particles undergoing brownian motion. Appl. Phys. Lett. 91, 103902â103902 (2007).
Armani, D., Kippenberg, T., Spillane, S. & Vahala, K. Ultra-high-q toroid microcavity on a chip. Nature 421, 925â928 (2003).
Knight, J., Cheung, G., Jacques, F. & Birks, T. Phase-matched excitation of whispering-gallery-mode resonances by a fiber taper. Opt. Lett. 22, 1129â1131 (1997).
OâKeefe, A. & Deacon, D. A. Cavity ring-down optical spectrometer for absorption measurements using pulsed laser sources. Rev. Sci. Instrum. 59, 2544â2551 (1988).
Matone, L. et al. Finesse and mirror speed measurement for a suspended Fabry-Perot cavity using the ringing effect. Phys. Lett. A 271, 314â318 (2000).
Savchenkov, A. A., Matsko, A. B., Mohageg, M. & Maleki, L. Ringdown spectroscopy of stimulated Raman scattering in a whispering gallery mode resonator. Opt. Lett. 32, 497â499 (2007).
Levenson, M. D. et al. Optical heterodyne detection in cavity ring-down spectroscopy. Chem. Phys. Lett. 290, 335â340 (1998).
Levenson, M. D. et al. Frequency-switched heterodyne cavity ringdown spectroscopy. Opt. Lett. 25, 920â922 (2000).
Lin, Q., Johnson, T. J., Michael, C. P. & Painter, O. Adiabatic self-tuning in a silicon microdisk optical resonator. Opt. Express 16, 14801â14811 (2008).
Ilchenko, V. & Gorodetskii, M. Thermal nonlinear effects in optical whispering gallery microresonators. Laser Phys. 2, 1004â1009 (1992).
Rokhsari, H. & Vahala, K. J. Observation of kerr nonlinearity in microcavities at room temperature. Opt. Lett. 30, 427â429 (2005).
Vukovic, N. et al. Ultrafast optical control using the Kerr nonlinearity in hydrogenated amorphous silicon microcylindrical resonators. Sci. Rep. 3, 02885 (2013).
Rokhsari, H., Kippenberg, T., Carmon, T. & Vahala, K. Radiation-pressure-driven micro-mechanical oscillator. Opt. Express 13, 5293â5301 (2005).
Schliesser, A., Del'Haye, P., Nooshi, N., Vahala, K. & Kippenberg, T. Radiation pressure cooling of a micromechanical oscillator using dynamical backaction. Phys. Rev. Lett. 97, 243905 (2006).
Schliesser, A. Cavity Optomechanics and Optical Frequency Comb Generation with silica Whispering-Gallery-Mode Microresonators PhD thesis Ludwig-Maximilians-Universität (2009).
Kippenberg, T. & Vahala, K. Cavity optomechanics: back-action at the mesoscale. Science 321, 1172â1176 (2008).
Aspelmeyer, M., Kippenberg, T. J. & Marquardt, F. Cavity optomechanics. Reviews of Modern Physics. 86, 1391 (2014).
Vanner, M. R. et al. Pulsed quantum optomechanics. Proc. Natl Acad. Sci. USA 108, 16182â16187 (2011).
Krause, A. G., Winger, M., Blasius, T. D., Lin, Q. & Painter, O. A high-resolution microchip optomechanical accelerometer. Nat. Photonics 6, 768â772 (2012).
Aoki, T. et al. Observation of strong coupling between one atom and a monolithic microresonator. Nature 443, 671â674 (2006).
Yasuda, R., Noji, H., Yoshida, M., Kinosita, K. & Itoh, H. Resolution of distinct rotational substeps by submillisecond kinetic analysis of f1-atpase. Nature 410, 898â904 (2001).
Perman, B. et al. Energy transduction on the nanosecond time scale: early structural events in a xanthopsin photocycle. Science 279, 1946â1950 (1998).
Mayor, U. et al. The complete folding pathway of a protein from nanoseconds to microseconds. Nature 421, 863â867 (2003).
Acknowledgements
This research was supported by the Israeli Science Foundation, the Joseph and Celia Reskin Career Development Chair in Physics, and the Crown Photonics Centre. This research was made possible, in part, by the historic generosity of the Harold Perlman family.
Author information
Authors and Affiliations
Contributions
Y.L., S.R. and B.D. conceived the idea. Y.L. and S.R. conducted the experiments and analysed the data. B.D., Y.L., S.R. and F.V. wrote the manuscript. All authors contributed to the design and construction of the experiment, discussed the results and commented on the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing financial interests.
Rights and permissions
This work is licensed under a Creative Commons Attribution 4.0 International License. The images or other third party material in this article are included in the articleâs Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/
About this article
Cite this article
Rosenblum, S., Lovsky, Y., Arazi, L. et al. Cavity ring-up spectroscopy for ultrafast sensing with optical microresonators. Nat Commun 6, 6788 (2015). https://doi.org/10.1038/ncomms7788
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/ncomms7788
This article is cited by
-
Non-planar graphene directly synthesized on intracavity optical microresonators for GHz repetition rate mode-locked lasers
npj 2D Materials and Applications (2024)
-
Microtubule WGM Sensor: Applications and Key Technologies
BioChip Journal (2023)
-
Dual-comb cavity ring-down spectroscopy
Scientific Reports (2022)
-
Cavity buildup dispersion spectroscopy
Communications Physics (2021)
-
Whispering-gallery-mode sensors for biological and physical sensing
Nature Reviews Methods Primers (2021)